In Vivo: Nanometer Size Wear Debris Generated From Ultra High Molecular Weight Polyethylene
In Vivo: Nanometer Size Wear Debris Generated From Ultra High Molecular Weight Polyethylene
In Vivo: Nanometer Size Wear Debris Generated From Ultra High Molecular Weight Polyethylene
C.
samples.
2.1.4. Acid hydrolysis and washing 2.2.2. Energy-dispersive analysis of X-rays (EDS)
The delipidated samples of tissue (obtained from 0.3 g of freeze- The same samples as those used for SEM were observed in the
dried sample) were hydrolyzed with 5 ml of 65% HNO3 at room
microscope Quanta 200 FEG as described in the previous section,
temperature for 24 h. Only the upper 2 ml layer of the suspen- using an EDS detector (EDAX). EDS spectra were taken at 30 kV. The
sion was used for further isolation. The remaining lower part was signal was collected both from the whole areas of SEM micrographs
removed by aspiration and discarded. The upper layer was washed and from selected places containing large agglomerates of wear
twice with 5 ml of 65% HNO3 and twice with 5 ml of distilled
particles.
water: 5 ml of solvent (HNO3 or H2O) were added, the sample was
centrifuged (500 g, 1 min) and the lower 5 ml were removed by 2.2.3. Fourier-transform infrared spectroscopy (FTIR)
aspiration. The resulting washed 2 ml layer was neutralized with
Infrared-spectra were recorded with an IFS-55 spec-
12 mol/l KOH and washed twice with water in the same way as
trometer (Bruker, Germany) equipped with MCT detector
described above. The nal 2 ml of the sample was used for further
(256 scan/spectrum, resolution 4 cm
1
). IR spectra of selected
purication.
PC membranes with PE particles were measured before sputtering
with platinum and investigation by SEM and EDS. The measure-
2.1.5. Separation of wear debris by ltration ment was performed in two ways: (i) several locations on each
The wear debris suspension obtained by acid hydrolysis was membrane were measured by the attenuated total reection (ATR)
mixed with 2 ml of iPrOH and ltered through a 10- m Teon mem- technique using a Golden Gate TM Heated Diamond ATR Top-Plate
brane Omnipore (Millipore, USA). The ltrate was ll ed up to 6 ml (Specac Ltd.) and (ii) the whole membrane was also measured in
with iPrOH. Then these 6 ml of the suspension were divided into the transmission mode.
four parts: (i) 4.5 ml of the suspension was ltered through a 0.1-
m polycarbonate membrane Isopore (Millipore, USA), (ii) 0.2 ml of
2.2.4. Image analysis of FEGSEM micrographs
the suspension was ltered through another 0.1- m polycarbonate
Image analysis of FEGSEM micrographs was performed using
membrane Isopore (Millipore, USA), (iii) another 0.2 ml of the sus-
the Lucia program (Laboratory Imaging, Czech Republic). FEGSEM
pension was sonicated (5 min, 30 kHz, amplitude 112 m, energy
density 125 W/cm
2
) just before the ltrati on through the 0.1 m
micrographs with magnications of 50,000 were used to
distinguish and measure individual wear nanoparticles. Lower
membrane Isopore and (iv) the remaining 1.5 ml of the suspension
magnication micrographs were not used because the two ana-
was left as a reserve.
lyzed samples contained only nanoparticles, as demonstrated
below. Both image analyses were based on more than 500 particles.
2.1.6. Usage of lter membranes containing wear particles The particles were described by means of equivalent circle diam-
All membranes were investigated by all the methods described eter (D), circularity (C) and elongation (E). Equivalent diameter of
below, however, the image analysis results were based exclusively an object is equal to the diameter of a circle with the same area as
on membranes with lower concentrations of particles (0.2 ml of the corresponding object: D=(4 Area/ )
1/2
. Circularity is calcu-
the suspension), whereas infrared spectroscopy and microanalysis lated from the object area and perimeter: C =4 Area/Perimeter
2
;
were performed on all the membranes (both 0.2 and 4.5 ml). The C =1 for circles and C <1 for all other shapes. Elongation is dened
low-concentration membranes showed better particle separation as the ratio of maximum and minimum diameter of the parti-
on electron micrographs, while the high-concentration membranes cle: E =MaxDiameter/MinDiameter; E =1 for circles and E >1 for all
gave higher signals in infrared spectroscopy and microanalysis. other shapes.
M. Lapcikova et al. / Wear 266 (2009) 349355 351
particles (Al, Zr), were found; a typical EDS spectrum is shown in
Fig. 2.
3.2. Separation of nanoparticles by sonication
Fig. 3 shows PC membranes with PE particles isolated from
periprosthetic tissue of the rst patient H1 (Table 1). On the mem-
branes, nanometer-sized wear particles tended to form aggregates
(Fig. 3a, c and e), which could be disintegrated by sonication
of the ltered suspension (Fig. 3b, d and f). The aggregates of
non-sonicated nanoparticles at medium magnications (Fig. 3a;
magnication 7500) and high magnications (Fig. 3c; magni-
cation 12,000) looked as if they were microparticles. Single
nanoparticles within the aggregates were only revealed at very
high magnications (Fig. 3e; magnication 50, 000). Sonication
resulted in nanoparticle separation, which could be observed even
Fig. 1. Comparison of IR spectra of pure PE (from a catalog), blank PC mem-
brane and PC membrane with isolated PE particles. The spectra were measured in
ATR mode (PE = polyethylene, PC = polycarbonate, ATR = attenuated total reection
infrared spectroscopy).
3. Results
3.1. Purity of isolated particles
Three facts indicated that observed nanoparticles were not
artifacts, but pure UHMWPE. Firstly, we used a well-established
isolation technique, which had already been employed and per-
fected in our previous studies [1215]. Secondly, the purity of the
isolated PE particles on PC membranes was investigated by means
of FTIR: comparison of IR spectra of pure PE, unused PC mem-
brane and PC membrane with PE nanoparticles showed that all
samples exhibited only peaks of PE and/or PC (Fig. 1). Thirdly, we
performed EDS analysis of all membranes, which were investi-
gated by FEGSEM. In all cases, we found only peaks of C (mostly
from PC membrane and partially from PE particles), O (from PC
membrane), Pt (10 nm thick platinum layer from vacuum sput-
ter coater), Ag and Cl (from conductive silver paste that xed the
membranes to the support). No traces of other elements, which
would indicate bone fragments (Ca, P), dust (Si), organic impu-
rities (N, S), metallic wear particles (Cr, Mo, Ti) or ceramic wear
Fig. 2. EDS spectrum of UHMWPE wear particles on a PC membrane. The peaks of
C and O come from the PC membrane and/or UHMWPE particles. The peaks of Pt,
Ag and Cl result from the fact that the membrane was sputter coated with platinum
and xed with a conductive silver paste containing chlorine before investigation
in an electron microscope (PC = polycarbonate, EDS = energy-dispersive analysis of
X-rays).
at medium magnications (Fig. 3b). At high magnications (Fig. 3d),
the single nanoparticles could be clearly resolved. The micrographs
of sonicated nanoparticles at very high magnications (Fig. 3f) were
suitable for image analysis.
3.3. Image analysis of separated nanoparticles
Fig. 4 shows nal FEGSEM micrographs of PC membranes with
in vivo PE wear particles, which were isolated from periprosthetic
tissues of both patients (Table 1). Both suspensions of the isolated
particles were sonicated immediately before ltrati on through the
membrane to separate particles. It is worth noting that FEGSEM
micrographs displayed neither bigger particles (>0.1 m) nor impu-
rities (such as bone fragments with their characteristic texture or
dust particles with sharp edges).
Image analysis of the wear particles of each patient was based
on nine FEGSEM micrographs from different parts of the mem-
brane. All nine micrographs in each set have the same magnication
(50, 000; the same as in Fig. 4e and f) and each set of micrographs
contained more than 700 well-separated particles for analysis. The
wear particles from patients H1 and H2 exhibited average sizes
18.5 and 21.2 nm, respectively. Complete histograms (Fig. 5) proved
that in both cases the wear particles were quite small, with rel-
atively narrow size distributions. Complete numerical results of
image analysis (Table 2) re-conrmed the small particle size, con-
rmed narrow size distributions and indicated that all particles
were spherical in shape.
3.4. Nanoparticles in other samples
In the previous sections we have demonstrated that 2 samples
out of 100 contained mostly nanoparticles. Detailed microscopic
analysis of the two samples suggested that they contained
almost no microparticles. At higher magnications, practically all
microparticles were identied as the agglomerates of nanoparticles
as documented by FEGSEM (Figs. 3 and 4) and conrmed by image
analysis (Fig. 5 and Table 2).
The other samples exhibited no exceptional features and con-
tained mostly microparticles, in agreement with numerous previous
studies (e.g. [1120]). Particles coming from one of our typical
samples of periprosthetic tissue are shown in Fig. 6. The differ-
ence between the two samples containing mostly nanoparticles
(Figs. 3 and 4) and the typical samples containing mostly micropar-
ticles (Fig. 6) was visible even at medium magnication FEGSEM
micrographs (cf. Figs. 3a and b, 4a and b and 6a). In the two
samples containing nanoparticles there were no particles >0.1 m,
whereas a typical sample contained a number of particles with
dimensions >1 m. This was the main reason why we focused
our attention on the two exceptional samples and revealed wear
352 M. Lapcikova et al. / Wear 266 (2009) 349355
Fig. 3. FEGSEM micrographs of UHMWPE wear particles on 0.1 m PC membrane. All particles come from patient H1 in Table 1. Left column (a, c and e): particles without
sonication before l trati on; right column (b, d and f): particles with sonication before ltration. The selected micrographs were taken at medium (a and b), high (c and d)
and very high (e and f) magnication.
nanoparticles, which become clearly visible at higher magnica- 4. Discussion
tions (Figs. 3cf and 4cf).
Nevertheless, further analysis of the other samples, which con- In the previous section it has been demonstrated for the rst
tained mostly microparticles, showed certain small amounts of time that UHMWPE wear nanoparticles are generated in vivo. We
wear nanoparticles as well (Fig. 6b and c). Although FEGSEM should re-emphasize that in this study the term wear nanopar-
micrographs of typical samples re-conrmed clear predominance ticles denotes UHMWPE particles with sizes below 0.05 m, i.e.
of wear microparticles, small amounts of wear nanoparticles the particles whose size is even smaller than the lowest pore
could be detected (Fig. 6c), nanoparticles marked with white size in commonly used lters [9]. Three independent methods,
arrows. This indicated that a certain number of wear nanopar- FEGSEM, EDS and IR suggested that the observed nanoparticles
ticles might be present in all periprosthetic tissues. Accurate are a polyethylene and not an artifact. Moreover, the nanoparticles
determination of volume fraction of wear nanoparticles in typi- with average sizes 18.5 and 21.2 nm were observed in two indepen-
cal samples of periprosthetic tissues is a subject of our subsequent dent samples. On the other hand, many previous studies reported
study. much larger average sizes of in vivo UHMWPE wear nanoparticles.
M. Lapcikova et al. / Wear 266 (2009) 349355 353
Fig. 4. Final FEGSEM micrographs of PC membranes with UHMWPE wear particles of both studied patients, H1 and H2 (Table 1). Left column (a, c and e): particles from H1;
right column (b, d and f): particles from H2. The selected micrographs were taken at medium (a and b), high (c and d) and very high (e and f) magnications.
Why were similar wear nanoparticles not found in previous
works?
The rst reason why wear nanoparticles were not reported
before might consist in the fact that we investigated more than
100 samples and detected signicant volumes of wear nanopar-
ticles just in 2 of them. In the other samples the microparticles
predominated and the amounts of nanoparticles were either
small or negligible. Elck et al. [16] summarized the average
sizes of in vivo wear particles from various studies; their sizes
varied around 0.5 m. Also in vitro wear particles from joint
simulators showed average sizes of several tenths of microm-
eters [17]. Thus it seems that standard wear particles are
microparticles, whereas wear nanoparticles might be an excep-
tion.
The second reason why wear nanoparticles escaped the atten-
tion of previous investigators could be sonication. Wear particles
are usually isolated from periprosthetic tissues by means of
acid/base digestion method. Then the suspension of the particles
is ltered through a membrane and the particles are analyzed by
means of scanning electron microscopy. Our results showed that
the nanoparticles on the membrane tend to form agglomerates
looking like single microparticles (Fig. 3). Intensive sonication of
the suspension just before its l trati on through a membrane is nec-
essary to achieve nanoparticle separation. As this sonication was
not always used in other studies, the agglomerates of nanoparticles
could have been misinterpreted as single microparticles in partic-
ular at low and medium magnication micrographs. In this study
the nanoparticles on the membranes occurred as loose agglomer-
mated
standard
deviation
(
=
(1/n
i
xi
)
and
the
width
of
the
distribution,
,
was
calculated
as
esti-
2
of the distribution, n is the number of all elements in the distribution and summation
Fig. 6. FEGSEM micrographs of PC membranes with UHMWPE wear particles of a
typical patient, whose periprosthetic tissue contained mostly microparticles. The
micrographs were taken at medium (a), high (b) and very high (c) magnications.
A few nanoparticles are visible at very high magnications (marked with white
runs through all elements.
arrows).
M. Lapcikova et al. / Wear 266 (2009) 349355 355
In some other studies, the authors used techniques, which were
not able to detect the smallest particles. For example, some parti-
cle counters detected only particles with sizes above 0.3 m [19],
in transmission electron microscopy the nanoparticles might be
hardly observable due to insufcient contrast as their size may be
lower than the thickness of investigated ultrathin section (typically
around 60 nm) [20] and in the methods based on light scatter-
ing the signal from bigger particles may entirely predominate
[11,12]. Our results suggest that a combination of isolation method
(veried in previous studies) and good detection method (sonica-
tion +FEGSEM) are necessary to catch wear nanoparticles. In future
studies the in vivo UHMWPE wear nanoparticles might be observed
more and more frequently as the sampling, isolation and detection
methods are further improved.
5. Conclusion
Methods for isolation and visualization of UHMWPE wear par-
ticles are still being improved. As a result, it has been possible
to isolate and detect smaller and smaller wear particles. A few
years ago Scott et al. [9] proved that a signicant fraction of in
vitro UHMWPE wear particles can be smaller than 0.2 m. Recently
Galvin et al. [10] found in vitro UHMWPE wear nanoparticles with
sizes below 0.1 m. In the present work we found, for the rst time,
in vivo UHMWPE wear nanoparticles with sizes below 0.05 m.
These nanoparticles were isolated from periprosthetic tissues of
two different patients and had average particle size of 18.5 and
21.2 nm in the rst and the second case, respectively.
Acknowledgement
Financial support through grant 2B06096 (Ministry of Edu-
cation, Youth and Sports of the Czech Republic) is gratefully
acknowledged.
References
[1] G. Lewis, Properties of crosslinked ultra-high-molecular-weight polyethylene,
Biomaterials 22 (2001) 371401.
[2] Alternatives to conventional UHMWPE for hip arthroplasty, in: S.M. Kurtz (Ed.),
The UHMWPE Handbook, Elsevier, Amsterdam, 2004, p. 93 (Chapter 6).
[3] S.M. Kurtz (Ed.), The Clinical Performance of UHMWPE in Knee Replacements.
The UHMWPE Handbook, Elsevier, Amsterdam, 2004, p. 182 (Chapter 8).
[4] G. Lewis, Polyethylene wear in total hip and knee arthroplasties, J. Biomed.
Mater. Res. 38 (1997) 5575.
[5] J.L. Tipper, E. Ingham, J.L. Hailey, A.A. Besong, J. Fisher, B.M. Wroblewski, M.H.
Stone, Quantitative analysis of polyethylene wear debris, wear rate and head
damage in retrieved Charnley hip prostheses, J. Mater. Sci.: Mater. Med. 11
(1999) 117124.
[6] J.D. Mabrey, A. Afsar-Keshmiri, G.A. Engh, C.J. Sychterz, M.A. Wirth, C.A.
Rockwood, C.M. Agrawal, Standardized analysis of UHMWPE wear parti-
cles from failed total joint arthroplasties, J. Biomed. Mater. Res. 63 (2002)
475548.
[7] E. Ingham, J. Fisher, Proceedings of the Institution of Mechanical Engineers. Part
HJ. Eng. Med. 214 (2000) 2137.
[8] J.B. Matthews, A.A. Besong, T.R. Green, M.H. Stone, B.M. Wroblewski, J. Fisher,
E. Ingham, Evaluation of the response of primary human peripheral blood
mononuclear phagocytes to challenge with in vitro generated clinically rel-
evant UHMWPE particles of known size and dose, J. Biomed. Mater. Res. 52
(2000) 296307.
[9] M. Scott, K. Widding, S. Jani, Do current wear particle isolation procedures
underestimate the number of particles generated by prosthetic bearing com-
ponents? Wear 251 (2001) 12131217.
[10] A.L. Galvin, J.L. Tipper, E. Ingham, J. Fischer, Nanometre size wear debris gen-
erated from crosslinked and non-crosslinked ultra high molecular weight
polyethylene in articial joints, Wear 259 (2005) 977983.
[11] M. Slouf, I. Sloufova, G. Entlicher, Z. Horak, M. Krejc, P. Stepanek, T. Radonsky, D.
Pokorny, A. Sosna, New fast method for determination of numbers of UHMWPE
wear particles, J. Mater. Sci.: Mater. Med. 15 (2004) 12671278.
[12] D. Pokorny, M. Slouf, Z. Horak, D. Jahoda, G. Entlicher, S. Eklova, A. Sosna, Method
for assesment of distribution of UHMWPE wear particles in periprosthetic tis-
sues in total hip arthroplasty, Acta Chirurgiae Othopaedicae et Traumatologiae
Cechosl 73 (2006) 243250.
[13] M. Slouf, S. Eklova, J. Kumstatova, S. Berger, H. Synkova, A. Sosna, D. Poko-
rny, M. Spundova, G. Entlicher, Isolation, characterization and quantication of
polyethylene wear debris from periprosthetic tissues around total joint replace-
ments, Wear 262 (2007) 11711181.
[14] M. Slouf, D. Pokorny, G. Entlicher, J. Dybal, H. Synkova, M. Lapcikova, Z. Fej-
farkova, M. Spundova, F. Vesely, A. Sosna, Quantication of UHMWPE wear
in periprosthetic tissues of hip arthroplasty: description of a new method
based on IR and comparison with radiographic appearance, Wear 265 (2008)
674684.
[15] E. Zolotarevova, Z. Fejfarkova, G. Entlicher, M. Lapcikova, M. Slouf, D. Pokorny, A.
Sosna, Can centrifugation affect the morphology of polyethylene wear debris?
Wear (in press) doi:10.1016/j.wear.2008.04.005.
[16] A.P.D. Elck, S.M. Green, S. Krikler, A. Unsworth, The nature and dissemination
of UHMWPE wear debris retrieved from periprosthetic tissue of THR, J. Biomed.
Mater. Res. 65A (2003) 95108.
[17] O. Calonius, V. Saikko, Analysis of polyethylene particles produced in different
wear conditions in vitro, Clin. Orthop. Rel. Res. 399 (2002) 219230.
[18] ISO 17853:2003 (E).
[19] K.J. Margevicius, T.W. Bauer, J.T. McMahon, S.A. Brown, K. Merritt, Isolation and
characterization of debris in membranes around total joint prostheses, J. Bone
Joint Surg. 76-A (1994) 16641675.
[20] M. Visentin, S. Stea, S. Squarzoni, B. Antonietti, M. Reggiani, A. Toni, A new
method for isolation of polyethylene wear debris from tissue and synovial uid,
Biomaterials 25 (2004) 55315537.