Scanner 2000 Quick Start Guide

Download as pdf or txt
Download as pdf or txt
You are on page 1of 8

NUFLO

Part No. 30165024, Rev. A

Scanner 2000 microEFM

QuickStart

Installing the Scanner 2000


H L

H L

Flow

Direct Mount

Remote Mount

To install the Scanner 2000 microEFM using a direct mount to


an orifice or cone meter run, perform the following steps:

To install the Scanner 2000 microEFM remotely from the


flow run, perform the following steps:

1. Determine the best location for the Scanner 2000.


a. For horizontal meter runs, locate process connections at
the top of the line, and mount the Scanner 2000 above
the pressure connections at the pipe.
b. For vertical meter runs, make sure pipeline flow is upward.
2. Locate the H and L markings on the Scanner 2000 transmitter designating high-pressure and low-pressure ports, and
connect a valve manifold (if desired) to the transmitter.
3. Connect the orifice flanges to the Scanner 2000/manifold
assembly, matching the upstream tap to the high-pressure
manifold port and the downstream tap to the low-pressure
manifold port.

1. Mount the Scanner to a flat, vertical surface, or bolt


to a 2-in. pipe using optional pole-mount hardware.
2. Install process tubing between the orifice run and
the manifold, making sure that the upstream pressure tap is connected to the high-pressure manifold
port and the downstream pressure tap is connected
to the low-pressure manifold port. Slope all piping
at least 1-inch per linear foot to avoid trapping
liquid.
3. Use a suitable compound or tape on all threaded
process connections.
4. Position all shutoff and bypass valves so they are
accessible from the front of the instrument.
5. Check the manifold for leaks.

ADJUSTABLE HEAD DESIGN.


When the sensor is oriented ap- Set screw
propriately for pressure connections, the position of the Scanner
display may need to be adjusted.
To change the orientation of the
head, loosen the set screw in the
MVT adapter, rotate the head
no more than 180 degrees and
tighten the set screw.

POLE-MOUNT OPTION.
An optional pole mount kit
(Part No. 9A-30028004)
allows the Scanner 2000 to
be mounted to a 2-in.
vertical or horizontal pipe.
The kit includes a mounting
bracket, two U-bolts and
screws for connecting the
bracket to the Scanner 2000
enclosure.

EXTERNAL POWER INPUT


6 TO 30 VDC
POWER SUPPLY

POWER

+ + +

Wiring the Scanner 2000

TURBINE INPUT

4
5

BATTERY

T1

1
2
3
4

Scanner 2000 microEFMs connect to a laptop three ways:


using an optional USB adapter and user-supplied USB cable
(external port for easy access)
using an RS-485 to RS-232 converter cable (wires to circuit board)
using an optional COM adapter and RS-485 to RS-232 converter
cable (external port for easy access)

RS-485 TO RS-232 CONVERTER

TX-

PORT 1

232/485
CONVERTER

PORT 2

RS-232 9-PIN
CONNECTOR

POWER

TD(A)
GND

TB2

Part No. 9A-101283116

232/485
CONVERTER

RS-232 9-PIN
CONNECTOR

PORT 1

Part No. 101283116

TX-

PORT 2

GND

POWER

TD(A)

TX+

+ + +

TD(B)

TB2

- IRTD

R+ R

WHITE

RTD -

WHITE

RTD+

RED

I+

RED

TB1
DIGITAL OUTPUT
PULSE READOUT
DEVICE

Leave this
end of shield
disconnected.

7
8
9
10

13

14

TB3
Resistor may be included in pulse readout
device. Size the resistor to limit the current
to 60 mA.

11
12

USB ADAPTER (OPTIONAL)


PORT 1 OR PORT 2 CAN BE
CONNECTED TO THE 232/485
CONVERTER

POWER SUPPLY
5 to 30 VDC

+ + +

TD(B)

TX+

I-

Scanner/Laptop Connection Options

TURBINE
MAGNETIC
PICKUP

RTD INPUT
- I-

Field wiring enters the Scanner 2000 microEFM through either of two
-in. NPT conduit openings in the enclosure and connects to three
terminal blocks on the main circuit assembly (TB1, TB2 and TB3). For
instructions on wiring an optional expansion circuit board, see the
Scanner 2000 Hardware Manual.

PORT 1 OR PORT 2 CAN BE


CONNECTED TO THE 232/485
CONVERTER

B
A

TB1

J1

RTD

TB3

RED

R+ R

OU

BLACK

I+

14

12

I+

TFM1

12

11

TB2

10
11

9
10

DIG

13

TC

I
SW

TFM1

RTD

R+ R
I+

TFM 1

TB2

TB1

PORT 1

J2

PORT 1

- I-

PORT 2

SCANNER 2000
Main Circuit Board
PN: 30160010

PORT 2

POWER

+ + +

7
8
9

CAUTIONAll field wiring must conform to the National


Electrical Code, NFPA 70, Article 501-4(b) for installations within
the United States or the Canadian Electric Code for installations within Canada. Local wiring ordinances may also apply.
All field wiring must have a wire range of 22 to 14 AWG and
terminal block screws must be tightened to a minimum torque
of 5 to 7 in-lbs. to secure the wiring within the terminal block.
Only personnel who are experienced with field wiring should
perform these procedures.

10
11
12

WARNINGTo prevent ignition of hazardous atmospheres, do


not remove the cover while circuits are alive. Under normal
conditions, the Scanner 2000 microEFM poses no hazard when
opened in a safe area.

Keypad Configuration & Controls


Remove the instrument cover to access the keypad. Each of the three keys are used for
configuration and operation.

CONTROL THE PARAMETER DISPLAY...


Press the LEFT ARROW (DISPLAY) button to
manually advance the display of parameters in the LCD window.

The following illustration defines the functions of each key.

CONFIGURATION:
Move between menus
and menu selections

TEST

CONFIGURATION:
Save configuration
settings

ACCESS

OPERATION:
View next parameter

PRESS

OPERATION:
Save totals

simultaneously to view
time/date, temperature,
and battery voltage

CONFIGURATION:
Change digits and
other menu selections
OPERATION:
View daily logs

VIEW DAILY FLOW LOGS....


Press the UP ARROW (LOG) button to
view daily logs. The two-digit log index
number to the left indicates the number
of days that have passed since the log was
recorded. A 1 for example will display
yesterdays log value; a 2 will display the
log value from 2 days ago. Up to 99 daily
logs can be viewed using the keypad.

Volume
PRESS

simultaneously to access
Configuration menu

Log index
(Days since log
was created)
Date stamp
(MMDDYY)

TO CONFIGURE THE INSTRUMENT....


Press the UP ARROW and ENTER buttons simultaneously to configure the following
slave address
contract hour
baud rate
plate size (if security
date and time
settings allow)
The following example demonstrates the steps
required to enter a slave address.

EXTERNAL SWITCH OPTION....


The external control switch option lets you
control the display of parameters on the
LCD and view daily logs without removing
the cover of the Scanner 2000 microEFM.

To Enter a Port 1 Slave Address:


Enter the Configuration
menu.

Press UP ARROW (LOG) and


ENTER (SAVE) simultaneously.

Locate the Slave Address


setting.

PORT 1 SLAVE ADDRESS will


appear in the lower display, and the
right-most digit in the top display will
begin blinking.

Enter the Slave Address.


(range: 1 to 65535,
excluding 252 through
255 and 64764)

Press UP ARROW (LOG) until the


correct digit is displayed.

00000000

Press LEFT ARROW (DISPLAY) to


select the next digit to the left.
Repeat using UP ARROW (LOG)
and LEFT ARROW (DISPLAY) to
enter all remaining digits.
Press ENTER (SAVE).

Press and release the push-button to


view the next parameter in the LCD
window.
Press and hold the push-button for
4 seconds to view the daily logs screen
(shown above). Then press and release
the push-button to view the next daily
log.

...

Up to 99 daily logs can be viewed from the


LCD using the keypad or external pushbutton switch.

Navigating the ModWorX Pro Interface


Main Display Screen
Dedicated button for quick and easy downloads

Four task-based menus


simplify navigation

The Scanner 2000 installation CD provided


with the instrument loads the ModWorX Pro
interface software for configuring and operating the Scanner 2000.
The Main Display screen (at left) is the control
center for the Scanner 2000 and provides quick
and easy access to the most commonly
performed tasks. From this screen, the user can:
view real-time totals and flow rates, input
data, and system information
download data
access configuration, calibration, and flow
run/turbine maintenance menus
get quick answers to task-specific questions
with links to user documents

Documentation for quick reference

Change Turbine
Flowmeter
Change K-Factors

Controls for all commonly performed tasks are conveniently grouped under four
menus near the top of the Main screen. Menus are named for the functions they
support: Calibrate Inputs, Maintain Flow Run, Maintain Turbine, or Configure.

Calibrate Inputs Menu


From the Calibrate Inputs menu, a user can calibrate differential pressure, static pressure, and process temperature. The user can load a
factory calibration, load a previous calibration, or enter new calibration points.

1 Select the task you


want to perform.
Then select the parameter
you want to calibrate.

2 Click Acquire Point to enter a new


calibration point and enter the applied
calibration value at the prompt.
Or click Load Calibration From... to load
the data from a factory calibration or
previous calibration.

Click on Change Task to


select a new parameter
or a new calibration task.

3 The measured value


can be viewed with or
without a trend graph.
Repeat Step 2 to enter
additional calibration
points.

4 Click Save Changes to


save the New Calibration
values to memory.

5 Click Exit
Maintenance Mode
to resume live data
readings.

Maintenance Mode
When a user enters a menu in which parameter changes
affect volume calculations, the Scanner 2000 is placed in
Maintenance Mode. This locks the input sensor readings at
their current values. When the user exits Maintenance Mode,
the inputs are unlocked.

Maintain Flow Run Menu


From the Maintain
Flow Run menu, a
user can change an
orifice plate size,
cone meter or
averaging pitot
tube parameters,
cone meter factors,
steam properties, or
gas composition.

1 Select the task you


want to perform.

2 Enter new parameters in the fields


provided.
Or click on Copy From Current to load current
parameters.
3 Click Save Changes to save the new
parameters to memory.

4 Click Exit
Maintenance Mode
to resume live data
readings.

Tasks will vary with


the flow calculation
method used.
Click Change Task
to select a new
maintenance task.

Maintain Turbine Menu


From the Maintain
Turbine menu, a
user can change the
K-factor or select
a different turbine
meter.
The user can enter a
linear K-factor or a
multi-point calibration factor (1 to 12
points allowed).

Click Change Task to


select a new turbine
meter or a new
maintenance task.

1 Select the task you


want to perform.

2 Select a new meter or enter a new


K-factor using the fields provided.

3 Click Save Changes to save


the new selections to memory.

4 Click Exit
Maintenance Mode
to resume live data
readings.

Configure Menu
From the Configuration
menu (right), a user can access
every setting required to
perform the initial setup for
the Scanner 2000.
1 Set the instrument date and
time.
2 Establish archive settings
parameters to be logged
contract hour
interval period for saving logs

The Configure Flow Run 1 screen below


shows configuration settings for an
AGA-3 gas run. Required parameters will
vary, depending on the flow calculation
method selected (step 6).
For information on configuring a steam
flow run, see the ModWorX Pro User
Manual.
If a liquid turbine input will be used in
addition to a compensated gas run, click
on Turbine Input 1 for configuration
options.

3 Configure flow run


parameters (as shown below).

4 Set the calculation period.

5 Set the units of measurement used for volume and rate.

6 Click Change to select the flow


calculation method and fluid property
calculation method to be used.
7 Click Change Plate to
specify plate parameters.
8 Specify pipe parameters.

9 Enter base conditions


specific to the location.
10 Click Change Composition
to specify the gas composition.
(Percentages must total 100%.)
11 Click OK to save the new
selections to memory.

Download Screens
To download data, simply press
the Download button and
select the range and type of
files to be downloaded. Daily
and interval logs can be viewed
or printed as a table or a trend
graph.

1 Click on the red


Download button to
download log data.

Each log type and setting type is


displayed under a separate tab.

2 Select the range of


data, the type of logs,
and any additional
settings to be included
in the report. Then click
Download Now.

3 Downloaded data will appear in


a table as shown. Click one of these
buttons to:
export data
print the tabular data
view and/or print a trend graph
Export formats:
.cfx (Flow-Cal)
.ANA, .VOL, and .EVT (PGAS)
.pdf, .xls, .rtf, .csv, and .html
4 The trend view offers two methods
of navigation:
(1) Use the scroll bar to move the trend
line left or right. The data point that intersects this vertical record index (white
dotted) line marks the record being
read; the values of all parameters in the
record appear in the table at left. Values
in the table update automatically as the
intersecting data point changes.

View all downloaded data


(gray box shows range of data
displayed in bottom graph).
Select the parameter
you want to view.

(2) Move the mouse to guide the


cursor along the trend line. The cursor
will snap to the nearest data point. The
corresponding parameter, date and
time stamp, and parameter value will
appear in a pop-up box on the screen.

5569.0501
ms-kl@c-a-m.com
INDIA

Part No. 9A-30165024, Rev. 03