Silica Check
Silica Check
Silica Check
Summary of method
Modifications are necessary to adapt the Low Range Silica method for analyzing trace levels in the Ultra
Low Range method. Liquid reagents produce more reproducible readings and lower blank values by
eliminating slight turbidity that may remain when using powdered reagents. Use of liquid reagents in
continuous monitors for silica provides a means of confirming the analyzer performance.
Silica and phosphate in the sample react with molybdate ions under acidic conditions to form
yellow silicomolybdic acid complexes and phosphomolybdic acid complexes. Addition of citric acid
destroys the phosphate complexes. Amino Acid F Reagent is then added to reduce the yellow
silicomolybdic acid to an intense blue color, which is proportional to the silica concentration. Test results
are measured at 815 nm.