Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                

X TMF Gf100 Series MFC Eng

Download as pdf or txt
Download as pdf or txt
You are on page 1of 82

Installation and Operation Manual

X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

GF100 Series

Brooks GF100 Series


High Purity/Ultra-High Purity Digital Thermal
Mass Flow Devices

Model GF125 Analog I/O


Model GF125 Digital I/O

Model GF135 Analog I/O

GF100 Series

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

GF100 Series

Dear Customer,
We appreciate this opportunity to service your flow measurement and control requirements with a Brooks
Instrument device. Every day, flow customers all over the world turn to Brooks Instrument for solutions to their gas
and liquid low-flow applications. Brooks provides an array of flow measurement and control products for various
industries from biopharmaceuticals, oil and gas, fuel cell research and chemicals, to medical devices, analytical
instrumentation, semiconductor manufacturing, and more.
The Brooks product you have just received is of the highest quality available, offering superior performance,
reliability and value to the user. It is designed with the ever changing process conditions, accuracy requirements
and hostile process environments in mind to provide you with a lifetime of dependable service.
We recommend that you read this manual in its entirety. Should you require any additional information concerning
Brooks products and services, please contact your local Brooks Sales and Service Office listed on the back cover
of this manual or visit www.BrooksInstrument.com.
Yours sincerely,
Brooks Instrument

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

GF100 Series

THIS PAGE WAS


INTENTIONALLY
LEFT BLANK

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013
Paragraph
Number

Contents

GF100 Series
Page
Number

Section 1 General Information


1-1
Introduction ................................................................................................................................... 1-1
1-2
How to Use This Manual ............................................................................................................... 1-1
1-3
Product Support References ......................................................................................................... 1-2
1-4
Notice and Caution Statements ..................................................................................................... 1-2
1-5
Product Warranty .......................................................................................................................... 1-2
1-6
How to Order a GF100 Series Device ........................................................................................... 1-2
1-7
Industry Standard References ....................................................................................................... 1-2
1-8
GF100 Series Gas Table ............................................................................................................... 1-2
1-9
Glossary of Terms and Acornyms ................................................................................................. 1-2
1-10
Description .................................................................................................................................... 1-5
1-11
Specifications for GF100 Series Devices ...................................................................................... 1-8
Section 2 Installation
2-1
General ......................................................................................................................................... 2-1
2-2
Receipt of Equipment .................................................................................................................... 2-1
2-3
Recommended Storage Practice .................................................................................................. 2-1
2-4
Return Shipment ........................................................................................................................... 2-2
2-5
Transit Precautions ....................................................................................................................... 2-2
2-6
Removal from Storage .................................................................................................................. 2-2
2-7
Gas Connections ........................................................................................................................... 2-2
2-8
In-Line Filter .................................................................................................................................. 2-2
2-9
Mechanical Installation .................................................................................................................. 2-3
2-10
Flow Controller Installation Arrangement ...................................................................................... 2-4
2-11
Purge the Gas Supply Line Before GF100 Series Installation ....................................................... 2-5
2-12
Position and Mount the GF100 Series .......................................................................................... 2-6
2-13
Perform a Leak Test ...................................................................................................................... 2-8
2-14
Zeroing Setup Process .................................................................................................................. 2-8
2-15
Zeroing the GF100 Series ............................................................................................................. 2-9
2-15-1 Zeroing the GF100 Series Pressure Transducer from the LCD Display Panel ............................. 2-9
2-15-2 Zeroing GF100 Series Flow from LCD Display Panel .................................................................. 2-10
2-16
Performance Checks ................................................................................................................... 2-11
2-17
Introduction to the MultiFloTM Configurator ................................................................................... 2-11
2-18
Using the MultiFlo Configurator .................................................................................................... 2-11
2-19
Electrical Connections .................................................................................................................. 2-14
2-19-1 DeviceNet Connections ................................................................................................................ 2-14
2-19-2 Analog/RS485 Connections ......................................................................................................... 2-15
2-20
GF135 Inlet Valve Installation ....................................................................................................... 2-17
2-21
GF135 Commissioning ................................................................................................................. 2-18
Section 3 Operation
3-1
General ......................................................................................................................................... 3-1
3-2
Theory of Operation for Flow Measurement ................................................................................. 3-1
Section 4 Maintenance & Troubleshooting
4-1
Overview ....................................................................................................................................... 4-1
4-2
Maintenance .................................................................................................................................. 4-1
4-3
Troubleshooting............................................................................................................................. 4-3
4-4
Troubleshooting Checklist ............................................................................................................. 4-3
4-5
GF100 Series Troubleshooting Guide ........................................................................................... 4-4
Section 5 Product Description Code
5-1
Overview ....................................................................................................................................... 5-1
i

Contents

GF100 Series

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

Appendix A: GF100 Series Gas Table


A-1
Overview ....................................................................................................................................... A-1
Appendix B: GF100 Series Patents
B-1
Overview ....................................................................................................................................... B-1
Appendix C: Essential Instructions
C-1
Overview ....................................................................................................................................... C-1
Warranty, Local Sales/Service Contact Information ....................................................................... Back Cover
Figure
Figure
Number
Number
1-1
1-2
1-3
1-4
1-5
2-1
2-2
2-3
2-4
2-5
2-6
2-7
2-8
2-9
2-10
2-11
2-12
2-13
2-14
2-15
2-16
2-17
3-1

GF100 Series High Performance Gas Flow Controller Analog and Digital .................................... 1-1
LED Display .................................................................................................................................. 1-7
Dimensions - GF100 Series ......................................................................................................... 1-12
Dimensions - GF135, DeviceNet .................................................................................................. 1-13
Dimensions - GF135, RS485 ....................................................................................................... 1-14
Typical Gas Supply Arrangement with Non-PTI MFC.................................................................... 2-4
Typical Gas Supply Arrangement with PTI MFC ........................................................................... 2-4
GF100 Series Mounting Attitude Positions .................................................................................... 2-6
GF100 Series Mounted to K1 Series Substrate Blocks ................................................................. 2-7
Mounting Screws Torque Pattern .................................................................................................. 2-8
Display with PSI Reading .............................................................................................................. 2-9
Display Reading Zero PSI ............................................................................................................ 2-10
Display Set to %FS ...................................................................................................................... 2-10
%Flow Display Set to Zero ........................................................................................................... 2-10
MultiFlo Cable Adapter ................................................................................................................. 2-12
RS-232/RS-485 Converter (P/N A33231) ..................................................................................... 2-13
Diagnostic Port ............................................................................................................................. 2-13
GF100 Series DeviceNet Connection .......................................................................................... 2-14
GF100 Series with 9-Pin Analog Connector ................................................................................. 2-15
Analog 9-Pin Connector (M) ......................................................................................................... 2-15
GF100 Series Electrical Interface Options ................................................................................... 2-16
GF135 Series Showing Inlet/Outlet Pneumatic Isolation Valve Lines........................................... 2-17
GF100 Series Operating Principles ............................................................................................... 3-2

Table
Number

ii

1-1
1-2
1-3
1-4
1-5
1-6
2-1
2-2
2-3
4-1
4-2
5-1
5-2
A-1A
A-1B
A-1C
B-1

Table
Number

Industry Standard References ....................................................................................................... 1-3


Terms and Acronyms .................................................................................................................... 1-4
Specifications for Standard GF100 Series .................................................................................... 1-9
Specifications for High Accuracy (HA) GF100 Series .................................................................. 1-10
Specifications for Safe Delivery System (SDS) GF100 Series ..................................................... 1-10
Specifications for GF135 Series ................................................................................................... 1-11
K1 Series Fasteners ...................................................................................................................... 2-7
K1 Substrate Torque Data ............................................................................................................. 2-7
Gas and Flow Ranges - MultiFlo Configurable - N2 Equivalent .................................................... 2-9
Environmental Factors .................................................................................................................. 4-3
GF100 Series Troubleshooting Guide ........................................................................................... 4-4
GF100 Series Product Description Code Table ............................................................................. 5-1
Model GF135 Product Description Code Table ............................................................................. 5-3
GF100 Series Gas Table - Codes 1-110, Bins SH40 to SH50 ....................................................... A-1
GF100 Series Gas Table - Codes 118-953, Bins SH40 to SH50 ................................................... A-2
GF100 Series Gas Table - Codes 958-5022, Bins SH40 to SH50 ................................................ A-3
GF100 Series Patents ................................................................................................................... B-1

Section 1 Introduction

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

GF100 Series

1-1 Introduction
Designed for semiconductor, MOCVD , and other gas flow control
applications that require a high purity all-metal flow path, the Brooks
GF100 Series mass flow controllers and meters deliver outstanding
performance, reliability, and flexibility. Process throughput and yield are
maximized while process costs are reduced by the GF100 Series featuring:
Ultra fast settling time for quick start up and very rapid process steps
MultiFlo gas and range configurability enabling reconfiguration without
removing device from the gas line
An independent diagnostic/service port to troubleshoot or change flow
conditions without removing the mass flow controller from service
Long-term stability due to extremely low wetted surface area, and
corrosion resistant Hastelloy sensor and valve seat
Optional model variations includingpressure transient insensitivity (PTI),
High Accuracy (HA), Safe Delivery System (SDS)

Figure 1-1 GF100 Series High Performance Gas Flow Controller Analog and Digital

1-2 How to Use This Manual


This manual is intended to provide the user with all the information
necessary to install, operate, troubleshoot and maintain these thermal
mass flow devices. The manual is organized in the following sections:
Section 1 Introduction
Section 2 Installation
Section 3 Operation
Section 4 Maintenance and Troubleshooting
Section 5 Product Description Code
Appendix A GF100 Series Gas Table
Appendix B GF100 Series Patents
Appendix C Essential Instructions
Warranty, Local Sales/Service Contact Information
It is recommended that this manual be read in its entirety before attempting
to operate or repair these devices.

1-1

Section 1 Introduction

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

GF100 Series
1-3 Product Support References

Refer to www.BrooksInstrument.com for Brooks sales and service


locations and to obtain other documents that support the GF100 Series.
Those documents include:

- Brooks MultiFloTM Configurator Quick Start Manual:


X-SW-MultiFlo-Config-QS-eng. part Number 541B167AAG

- Brooks GF100 Series data sheets:


DS-TMF-GF100-Series-MFC-eng
DS-TMF-GF135-Series-MFC-eng

1-4 Notice and Caution Statements


Warning, caution and notice statements are located throughout this manual
in the ANSI format. A WARNING statement indicates a potentially
hazardous situation which, if not avoided, COULD result in death or serious
injury. A CAUTION statement indicates a potentially hazardous situation
which, if not avoided, MAY result in minor or moderate injury. It may also be
used to alert against unsafe practices. A NOTICE statement describes
specific information that requires special attention.
1-5 Product Warranty
Product warranty information can be found on the Back Cover of this
Manual and on the Brooks website at www.BrooksInstrument.com. This
information provides general warranty information, limitations, disclaimers,
and applicable warranty periods according to product group.
1-6 How to Order a GF100 Series Device
Refer to Section 5.
1-7 Industry Standard References
Refer to Table 1-1.
1-8 GF100 Series Gas Table
Refer to Appendix A.
1-9 Glossary of Terms and Acronyms
Refer to Table 1-2

1-2

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

Section 1 Introduction

GF100 Series

Table 1-1 Industry Standard References


Reference Number
MIL-STD-810
SEMI E12
SEMI E16
SEMI E17
SEMI E18
SEMI E27
SEMI E28
SEMI E52
SEMI E54
SEMI E56
SEMI E66
SEMI E67
SEMI E68
SEMI E69
SEMI E80
SEMI E16-90
SEMI F19
SEMI F20

SEMI F36
SEMI F37
SEMI F44
SEMI F45
SEMI F47
SEMI S2
SEMI S9
SEMI S10
SEMI S12

Reference Description
Method 514.4, Category 1, Transportation Requirement
Method 516.4, Procedure 1, Functional Shock Test Requirement
Standard temperature and pressure
Guideline for determining and describing MFC leak rates
Guideline for MFC transient characteristics tests
Guideline for temperature specifications of the MFC
Standard for MFC and MFM linearity
Guideline for pressure specifications for the MFC
Practice for referencing gases used in digital MFCs
Sensor actuator network connections for DeviceNet
Test method for detemining accuracy, linearity, repeatability, short-term
reproducibility, hystereses of thermal MFCs
Test method for determining particle contribution by MFCs
Test method for determining reliabilty of MFCs
Test method for determining warm-up time of MFCs
Test method for reproducibilty and zero drift for thermal MFCs
Test method for determining attitude sensitivity of MFCs
Guidelines for determining and describing mass flow controllers leak rates
Specification for the finish of the wetted surface of electro polished
216L stainless steel components
Specifications for 316L stainless steel bar, extruded shapes, plate, and
investment castings for components used in ultra-high purity semi
manufacturing applications
Guide for dimensions and connections of gas distribution components
Method for determination of surface roughness parameters for gas
distribution system components
Guideline for standardization of machined stainless steel weld fittings
Guideline for standardization of machined stainless steel reducing fittings
Specifications for semiconductor processing equipment
voltage sag immunity
Environmental, Health and Safety Guidelines
Dielectric testing
Risk assessment
Decontamination of fielded products

1-3

Section 1 Introduction

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

GF100 Series
Table 1-2 Terms and Acronyms

1-4

Term or Acronym

Definition

CSR

Customer Special Requirement

CVD

Chemical Vapor Deposition

DeviceNet

A 5-wire local network I/O communication device


that employs a command/response
communication protocol

DSP

Digital Signal Processor

EPI Epitaxy (EPI).

A process technology where a pure silicon


crystalline structure is deposited or grown on a
bare wafer, enabling a high-purity starting point
for building the semiconductor device.

HBD

Horizontal Base Down

GF100 Series

Integrated Flow Controller

F.S.

Full Scale

LED

Light Emitting Diode

MFC

Mass Flow Controller

MultiFlo Configurator

I/O communication software package that


configures gas and flow ranges

MultiFlo Technology

A physics-based calibration methodology that


enables gas and flow range configuration within
a defined standard configuration

PID

Proportional Integral Derivative Controller

PSIA

Pounds per Square Inch Absolute

PSID

Pounds per Square Inch Differential

PSIG

Pounds per Square Inch Gauge

PTI

Pressure Transient Insensitive. Reduces the


effect of pressure fluctuations in gas flow.
Applicable to GF125 only.

ROR

As pressure increases, flow increases at a


pressure rate of rise, or ROR.

HC

Standard Configuration w/ Hastelloy sensors


(to reduce reaction to corrosive gases)

S.P.

Setpoint

Step Technology

Enables fast set point control through a high


speed DSP and low volume drive circuit

VIU

Vertical mounting attitude with inlet side facing up

Section 1 Introduction

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

GF100 Series

1-10 Description
Ultra Fast Response
The Brooks GF Series patented flow sensor combined with a high speed
ARM processor and fast acting diaphragm-free valve assembly enables:
Faster response and settling time for improved wafer throughput
Ultra-fast 1-2 second etch step processing
Reduced diverted gas consumption and associated abatement costs
Time sensitive gas delivery steps in Atomic Layer Deposition
User programmable start-up function for processes requiring a slow
ramped gas turn-on or time critical transitions between flow rates
MultiFlo Gas and Range Configurability
The Brooks MultiFlo technology delivers exceptional improvement in
process gas accuracy for linear and non-linear gases. This is achieved
through advanced gas modeling and optimized through actual gas testing.
Brooks MultiFloTM allows the device to be quickly and easily configured for
another gas and/or flow range without sacrificing accuracy or rangability.
Selecting a new gas automatically creates a new calibration curve,
establishes optimized PID settings for dynamic control, automatically
compensates for gas density effects, and ensures smooth overshoot-free
transitions between flow rates with excellent steady-state stability. Brooks
MultiFlo technology offers unparalleled flexibility. An extensive gas
database is provided and a single device can be quickly programmed for
thousands of different gas and flow range combinations. Process benefits
achieved:
Mass flow controller full scale full range can be rescaled down typically by a
factor of 3:1 with no impact on accuracy, turndown or leak by specifications
Optimum process and inventory flexibility resulting in dramatically reduced
inventory costs
Fewer configurations/bin sizes required to support many different processes
Less down-time with rapid process recovery
MultiFloTM Support References: Brooks MultiFlow Configurator Quick
Start Guide (X-SW-MultiFlo-Config-QS-eng (Part Number: 541B167AAG))
MultiFloTM Configurator Accessory Kits:
778Z010ZZZ

Basic MultiFlo Configurator Kit


*Software, MultiFlo Configurator

A331710003

Cable Assembly 2.5mm

A332300001

Converter 232/485

778Z011ZZZ

Basic MultiFlo Configurator Kit w/Power Supply and


Adapter Cables *Software, MultiFlo Configurator

A331710003

Cable Assembly 2.5mm

A332300001

Converter 232/485

A332295001

Power Supply MFC

A332297002

Cable, Power, 9-Pin

A332297001

Cable, Power, DeviceNet

* MultiFlo Configurator Software is available on the Brooks Instrument


website at: www.BrooksInstrument.com/MultiFlo
www.BrooksInstrument.com/Documentation&Downloads

1-5

Section 1 Introduction

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

GF100 Series

Pressure Transient Insensitivity (PTI) Included with all GF125 &


GF135 models
Cost and space constraints are driving gas panel designers to remove
point-of-use pressure regulators and pressure monitoring components from
the process design and rely solely on the mass flow controller to accurately
control the process under dynamic pressure conditions. The Brooks GF125
& GF135 (PTI) devices utilize a patented control algorithm that inverts the
pressure signal, compares it to the pre-fluctuation signal and drives realtime valve position compensation to maintain stable flow. Enhanced
pressure transient is achieved through faster sensing, quicker processing,
and a reduction in internal dead-volume between the sensors and valve
orifice.
Optional High Accuracy (HA) GF125 model only
The GF125 High Accuracy (HA) is a gas and range specific model for
critical gas process applications requiring the widest working range with
tightest flow control accuracy. A typical application is for multi-step
processes requiring a high flow rate, up to 10 slpm, and a very accurate
low flow rate.
Optional Safe Delivery System (SDS) GF120 model only
The Safe Delivery System (SDS) enhanced GF120 model is a state-of-theart low pressure drop devices for the delivery of sub atmospheric safe
delivery system gases used in Implant and Etch processes. The Brooks
GF120 SDS models are available in full scale flow ranges 4 sccm -1 slpm.
Advanced Thermal Flow Measurement Sensor
Brooks proprietary highly corrosion resistant Hastelloy C-22 sensor with an
enhanced sensor manufacturing and burn in process incorporates a unique
orthogonal sensor mounting orientation to eliminate sensor drift caused by
valve heating effects and eliminates thermal siphoning effects. This unique
sensor configuration includes an optimized temperature profile for gases
prone to thermal decomposition. This design results in:
Enhanced signal to noise performance for improved accuracy at low
set points
Superior reproducibility at elevated temperature through new
isothermal packaging and onboard conditioning electronics with
ambient temperature sensing and compensation
Improved long-term stability
High Purity Flow Path
The Brooks GF Series has an all metal, corrosion resistant Semi F20
compliant wetted flow path with highly corrosion resistant Hastelloy C-22
valve seat and jet orifice. The GF120, GF125 & GF135 devices are
constructed with a 4 inch Ra max surface finish while the GF100 is
constructed with a 10 inch Ra.
Overall reduced surface area and un-swept volumes for faster dry-down
during purge steps
Long-term sensor and device stability for maximum yield and throughput
1-6

Section 1 Introduction

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

GF100 Series
Extensive Mechanical Configurations
GF Series supports all metal seal/UHP industry gas connection interface
standards for full OEM and process coverage.
Downport 79.8mm and 92mm C-seal and W-seal on 1-1/8" and 1-1/2"
bodies
124mm 4 VCR on 1-1/2" body
Enhanced Diagnostics and User Interface
The mass flow controller is one of the most complex and critical
components in gas delivery systems. When dealing with ultra-high-purity
gas distribution or highly toxic or corrosive gases, removing the mass flow
controller to assess functionality should be the last resort. Brooks GF
Series devices provide for in-line device evaluation and instantaneous
troubleshooting through:
Embedded self test routines and independent diagnostic/service port
High visibility LCD display with easy accessible push button for local
indication of Flow (%), Temperature (oC), Pressure (PSIA/kPa) and
Network Address
Zero button to easily re-zero the device during scheduled maintenance
This combination of features results in limiting service interruption and
reduced downtime.

Figure 1-2 LED Display

Communication Interfaces
The GF100 Series supports analog 0-5 Vdc, RS485, and DeviceNet
communication protocols. Analog connections can be accessed via the
DeviceNet or RS485 or Analog only connector options. DeviceNet and
RS485 are multi-drop connections that allow a maximum of 64 devices for
DeviceNet and 32 devices for RS485 to be connected on the same
network. Brooks Instruments DeviceNet profile has been certified by the
ODVA (Open DeviceNet Vendors Association). A range of low profile
adapter cables facilitate replacing previously installed devices eliminating
the need to carry multiple devices of the same gas/range but different
electrical connectors.

1-7

Section 1 Introduction

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

GF100 Series
1-11 Specifications for GF100 Series Devices

See Table 1-3 for specifications for standard GF100 Series.


See Table 1-4 for specifications for the high accuracy (HA) GF100 Series.
See Table 1-5 for specifications for the Safe Delivery System (SDS) GF100
Series.
See Table 1-6 for specifications for the GF135 Series.
See Figure 1-3 for dimensions for the GF100 Series.
See Figures 1-4 & 1-5 for dimensions for the GF135 Series.

1-8

Section 1 Introduction

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

GF100 Series

Table 1-3 Specifications for Standard GF100 Series


Performance

GF100

GF120

Full Scale Flow Range (N 2 Eq.)

3 sccm to 55 slm

Flow Accur
acy
Accuracy

+1% S.P. > 35-100%,

Repeatability & Repr


oducibility
Reproducibility

+ 0.5% F.S. (included in accuracy)


< 1 sec

<1.5 sec

Pr
essur
Pressur
essuree Insensitivity

MultiFlo

300ms (3-860 sccm N2 Eq.)


400ms (861-7200 sccm N2 Eq.)
500ms (7201-30000 sccm N2 Eq.)
<700ms (30001-55000 sccm N2 Eq.)

<1 sec

Normally Open Valve

Con
tr
ol Range
Contr
trol

+0.35% F.S. 2-35%

< + 0.15% S.P.

Linearity
Response Time (Settling Time)
Normally Closed Valve

GF125

Not Applicable
2-100% (Normally Closed Valve)

< 5% SP up to 5 psi/sec upstream press. spike


3-100% (Normally Open Valve)

optional

standard

#of Bins

11 bins

Valve Shut Down (N.C. Valve)


Valve Shut Down (N.O. Valve)

< 1% of F.S.
2% of F.S.

Zer
eroo Stability

< + 0.5% F.S. per year

Temper
atur
emperatur
aturee Coefficien
Coefficientt

Span: 0.005% full scale per oC, Zero: 0.001% full scale per oC

Ratings
Oper
ating Temper
atur
Operating
emperatur
aturee Range
Differ
en
tial Pr
essur
Differen
ential
Pressur
essuree Range*

10-50 oC
3-860 sccm = 7-45 psid, 861- 7200 sccm = 10-45 psid, 7201-55000 sccm = 15-45 psid
*Argon gas applications require an additional 10 psid differential pressure.
Low vapor pressure gases require an inlet pressure of > 100 Torr, with vacuum on outlet
(example SiCl4). Contact Brooks Technical Support for more information.

Maximum Oper
ating Pr
essur
Operating
Pressur
essuree

500 psia max

100 psia max

Pr
oof Pr
essur
Proof
Pressur
essuree

700 psia max

140 psia max

Burst Pr
essur
Pressur
essuree

3000 psia max

500 psia max

Leak In
tegrity (external)
Integrity

1x10-11 atm. cc/sec He

Mec
hanical
Mechanical
Valve Type

Wetted Materials
Surface Finish

Normally Closed
Normally Open (>93sccm flow rate N2 Equivalent)
Meter (no valve)
GF100: SEMI F20 GP Compliant, 316L VIM/VAR, Hastelloy C-22, 316L Stainless Steel, 304 Stainless Steel, KM-45
GF120/GF125: SEMI F20 UHP Compliant, 316L VIM/VAR, Hastelloy C-22, 316L Stainless Steel, 304 Stainless Steel, KM-45
4 inch Ra (0.1 m Ra)

10 inch Ra

Diagnostics & Display


Status Lights

MFC Health, Network Status

Alarms

Sensor Output, Control Valve Output, Over Temperature, Power Surge/Sag, Network Interruption

Display Type
Viewing Angle / Viewing Distance
Units Displayed / Resolution

Top Mount Integrated LCD


Fixed / 10 feet
Flow (%), Temp. (oC), Pressure (psia, kPa) / 0.1 (unit)

Electrical
Electrical Connection

RS485/Analog via 9-Pin D connector, DeviceNetTM via 5-Pin M12 connector

Digital Communication

RS485+ (model specific), DeviceNet (model specific), RS485 Diagnostic Port (all models)

Diagnostic /Service PPort


ort

RS485 via 2.5mm jack

PPower
ower Supply/Consumption

DeviceNet: 545mA max. @ +11-25 Vdc., 250mA max. @ 24Vdc RS485/Analog: 6 Watts max @ +15Vdc. (+10%)

1-9

Section 1 Introduction

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

GF100 Series
Table 1-4 Specifications for High Accuracy (HA) GF100 Series
Performance
GF125(HA)
Full Scale Flow Range*
Gases Supported
Progr
ogrammable*
MultiFlo Pr
ogr
ammable*
Flow Accur
acy*
Accuracy*

5 sccm - 10 slpm N2 equivalent


N 2, O2, Cl2, HBr, SiCl4, H 2
Not Configurable
10-100%F.S.=+1.0%S.P.
2-10%F.S.=+1% S.P. plus +0.04% F.S.

Repeatability & Repr


oducibility
Reproducibility

< +0.15% S.P.

Turn Down Ratio


Zer
eroo Stability

100:1
< 0.5% F.S./OC

Settling Time (to within +2% FF.S.)*


.S.)*

300ms (<860 sccm N2 Equivalent)


400ms (861-7200 sccm N2 Equivalent)
500ms (7201-55000 sccm N2 Equivalent)

Warm Up Time

Minimum of 30 minutes

Leak In
tegrity
Integrity

1X10-11 atm. cc/sec He

Valve Shut Down (Leak by)

<0.5% F.S.

Oper
ating Conditions
Operating
Max. Oper
ating Pr
essur
Operating
Pressur
essuree

GF125(HA)
100 psia max

Pr
oof Pr
essur
Proof
Pressur
essuree

140 psia max

Burst Pr
essur
Pressur
essuree

500 psia max

Pr
essur
Pressur
essuree Insensitivity

+5% S.P. for up to 5 psi/sec. upstream press. spike

Differ
en
tialPr
essur
e**
Differen
entialPr
tialPressur
essure**
High Pr
essur
Pressur
essuree Gases
Low Pr
essur
Pressur
essuree Gases

7-45 psid (N2, O 2, Cl2, HBr, H2)


>100 Torr (SiCl4)

Valve Con
figur
ation*
Configur
figuration*

Normally Closed

Ambien
atur
Ambientt Temper
emperatur
aturee Range
Zer
atur
Coefficientt
eroo Temper
emperatur
aturee Coefficien

10 oC-50 OC
Span: 0.005% full scale per OC, Zero: 0.001% full scale per OC

Table 1-5 Specifications for Safe Delivery System (SDS) GF100 Series
Performance
GF120XSL
GF120XSD
Full Scale Flow Range (N2 Eq.)
Gases Supported
MultiFlo Pr
ogr
ammable
Progr
ogrammable
Flow Accur
acy
Accuracy

4 - 25 sccm
>25 to 1 slpm
AsH3, PH3, BF3, SiF4, Ar, Xe, N2O, N2***
Not Configurable
+/-0.35% S.P. <35% F.S.
+/-1% S.P. >35% F.S.

Repeatability & Repr


oducibility
Reproducibility
Zer
eroo Stability

<+/- 0.15% S.P.


<=0.6% F.S. per year

Settling Time (to within +2% FF.S.)


.S.)

< 3 sec

Warm Up Time
Leak In
tegrity
Integrity

minimum of 30 minutes
1X10-11 atm. cc/sec He

Valve Shut Down (Leaky by)

Oper
ating Conditions
Operating
Minimum Oper
ating Inlet Pr
essur
e*
Operating
Pressur
essure*

Maximum Pr
essur
Pressur
essuree
Pr
oof Pr
essur
Proof
Pressur
essuree
Burst Pr
Pressur
essuree
essur

GF120XSD

GF120XSL
4 to 20 sccm < 10 Torr
>20 to 50 sccm < 20 Torr
>50 sccm to 1 slpm < 50 Torr
500 psia max
700 psia max
3000 psia max

Pr
essur
Pressur
essuree Insensitivity

Not Available
10 Torr-30 psid typical (1.33-207 kPa typical)

Valve Con
figur
ation
Configur
figuration

1-10

<1% F.S.

Differ
en
tialPr
essur
e**
Differen
entialPr
tialPressur
essure**
Ambien
atur
Ambientt Temper
emperatur
aturee Range
Zer
atur
eroo Temper
emperatur
aturee Coefficien
Coefficientt

*Consult Technical Support for details.


**Typical pressure drop. Actual pressure
drop will be gas and flow dependent.

Normally Closed
10oC-50oC
Span: 0.005% full scale per oC, Zero: 0.001% full scale per oC

*Performance at minimum inlet pressure will


be gas and flow range dependent. Consult
Technical Support for details.
**Typical pressure drop. Actual pressure drop
will be gas and flow dependent. Consult
Technical Support for details.
***Consult factory for other gases.

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

Section 1 Introduction

GF100 Series

Table 1-6 Specifications for GF135 Series

Performance

Full Scale Flow Range:

3 sccm to 5 slm (N2 Eq.)

Gasses Supported:

N2, O2, Ar, H2, SF6, NH3, CO2, Cl2, HBr, NF3, CF4, CH4, CH3F, CH2F2, SiCl4 (@ 100 Torr),
C4F6-q (@ 800 Torr), C4F8 (@ 1200 Torr)

Flow Accur
acy:
Accuracy:

+/-1.0%S.P. (10-100%F.S.), +/-1% S.P. plus +/-0.04% F.S. (2-10%F.S.)

Repeatability & Repr


oducibility:
Reproducibility:

< +/- 0.15% S.P.

Linearity:

Included in accuracy

Settling Time:

<300ms (<860 sccm N2 Equivalent), <400ms (861-5000 sccm N2 Equivalent)

Pr
essur
Pressur
essuree Insensitivity:

< 1% SP up to 5 psi/sec upstream press. spike

Con
tr
ol Range:
Contr
trol

1-100%

Valve Shut Down:

< 0.5% of F.S. N2

Zer
eroo Stability:

< +/- 0.5% F.S. per year

Temper
atur
t:
emperatur
aturee Coefficien
Coefficient:

Span: +/-0.05% full scale per C, Zero: +/-0.005% full scale per C

Rate-of-Decay PPerformance
erformance
Flow Rate:

Maximum flow for which an ROD measurement can be obtained 800 sccm

Accur
acy:
Accuracy:

+/- 2% of setpoint down to 5% SP, then 0.1% FS from 5% SP to 1% SP

Temper
atur
t:
emperatur
aturee Coefficien
Coefficient:

+/- 0.04% SP/Deg C

Pr
essur
t:
Pressur
essuree Coefficien
Coefficient:

+/- 0.04% FS/psi

Valve Leakby Accur


acy:
Accuracy:

+/- 0.2% FS.

Ratings
Oper
ating Temper
atur
Operating
emperatur
aturee Range:

10-50C

Differ
en
tial Pr
Differen
ential
Pressur
essuree Range**:
essur

3-860 sccm = 7-45 psid, 861- 5000 sccm = 10-45 psid


**Typical pressure drop. Actual pressure drop will be gas and flow dependent.
Argon gas applications require higher differential pressure.
Low vapor pressure gases require an inlet pressure of > 100 Torr, with vacuum on outlet
(example SiCl4). Contact Brooks Technical Support for more information.

Maximum Oper
ating Pr
essur
e:
Operating
Pressur
essure:

100 psia max

Pneumatic Valve Oper


ating Pr
essur
e:
Operating
Pressur
essure:

43.5 psia - 72.5 psia

Pr
oof Pr
essur
e:
Proof
Pressur
essure:

140 psia max

Burst Pr
essur
e:
Pressur
essure:

500 psia max

Leak In
tegrity (external):
Integrity

1x10-10 atm. cc/sec He

Mec
hanical
Mechanical
Valve Type:

Normally Closed

Wetted Materials:

SEMI F20 UHP Compliant 316L VIM/VAR, Hastelloy C-22,316L Stainless Steel, 304 Stainless Steel, KM-45

Surface Finish:

4 inch Ra (0.1 m Ra)

Diagnostics & Display


Status Lights:

MFC Health, Network Status

Alarms:

Sensor Output, Control Valve Output, Over Temperature, Power Surge/Sag, Network Interruption,
Sensor Drift, Flow Error, Valve Leak

Display Type:

Top Mount Electronically Rotatable Integrated LCD

Viewing Distance:

Fixed / 10 feet

Units Displayed / Resolution:

Flow (%), Temp. (C), Pressure (psia, kPa) / 0.1 (unit)

Electrical
Electrical Connection:

Analog/RS-485 via 9-Pin D connector, DeviceNet via 5-Pin M12 connector

Digital Communication:

RS485+ (model specific), DeviceNet (model specific), RS485 Diagnostic Port (all models)

Diagnostic / Service PPort:


ort:

Micro USB

Power Supply/Consumption:

DeviceNet: +11-25 Vdc., 545 mA max. @ 11 Vdc., 250 mA (max.) @ 24 Vdc.,


Analog /RS485: +/-15 Vdc. (+10%), 6 Watts (max) or +24 Vdc +/-10%

Compliance
En
vir
onmen
tal Compliance:
Envir
vironmen
onmental

CE: EN61326: 2006 (FCC Part 15 & Canada IC-subset of CE testing)

Safety:

EN61010-1

DeviceNet

ODVA

1-11

Section 1 Introduction

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

GF100 Series
Downport Configurations
Electrical Connector
G1/GX/TX Specific Dimensions

Electrical Connector
SX Specific Dimensions

Electrical Connector
DX/BB Specific Dimensions

3.29in
84mm

3.67in
93mm

.98in
25mm

5.09in
129mm

4.45in
113mm

3.75in
95mm

Fitting Option Code

Dim A

Dim B

Dim D

Dim E

Dim F

Dim G

CX

C-SEAL

92mm [3.62in]

105mm [4.13in]

22mm [0.86in]

28mm [1.12in]

83mm [3.28in]

25mm [1.00in]

127mm [5.00in]

EX

Seal Type

W-SEAL

79.8mm[3.14in]

93mm [3.66in]

30mm [1.18in]

Dim C

38.7mm [1.525in]

72mm [2.82in]

25mm [1.00in]

127mm [5.00in]

WX

W-SEAL

92mm [3.62in]

105mm [4.13in]

22mm [0.86in]

28mm [1.12in]

83mm [3.28in]

25mm [1.00in]

127mm [5.00in]

DX

C-SEAL

79.8mm[3.14in]

93mm [3.66in]

22mm [0.86in]

28mm [1.12in]

72mm [2.82in]

25mm [1.00in]

127mm [5.00in]

YX

W-SEAL

79.8mm[3.14in]

93mm [3.66in]

22mm [0.86in]

28mm [1.12in]

72mm [2.82in]

25mm [1.00in]

127mm [5.00in]

AX

C-SEAL

92mm [3.62in]

105mm [4.13in]

30mm [1.18in]

39mm [1.53in]

83mm [3.28in]

25mm [1.00in]

127mm [5.00in]

BX

W-SEAL

92mm [3.62in]

105mm [4.13in]

30mm [1.18in]

39mm [1.53in]

83mm [3.28in]

25mm [1.00in]

127mm [5.00in]

LX

C-SEAL

92mm [3.62in]

105mm [4.13in]

22mm [0.86in]

28mm [1.12in]

83mm [3.28in]

25mm [1.00in]

127mm [5.00in]

VCR Configurations
Electrical Connector
G1/GX/TX Specific Dimensions

Electrical Connector
DX/BB Specific Dimensions

Electrical Connector
SX Specific Dimensions

3.66in
93mm

3.29in
84mm

.98in
25mm

5.09in
129mm

4.45in
113mm

3.75in
95mm

F
H
J
A

4X 8-32 UNC
0.12 (3.0mm)

Fitting Option Code

VCR

Dim A

Dim B

Dim C

Dim D

Dim E

Dim F

Dim G

Dim H

Dim J

VX

1/4"

69mm [2.72in]

82mm [3.24in]

18mm [0.72in]

38mm [1.48in]

83mm [3.28in]

31mm [1.21in]

127mm [5.0in]

13mm [0.50in]

124mm [4.88in]

1-12 Figure 1-3 Dimensions - GF100 Series

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

Section 1 Introduction

GF100 Series

Model GF135, DeviceNet, C-Seal Connections

Model GF135, DeviceNet, VCR Connections

Figure 1-4 Dimensions - GF135, DeviceNet

1-13

Section 1 Introduction

GF100 Series

Model GF135, RS485, C-Seal Connections

Model GF135, RS485, VCR Connections

1-14 Figure 1-5 Dimensions - GF135 Series, RS485

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

Section 2 Installation

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

GF100 Series

2-1 General
This section provides installation instructions for the Brooks GF100 Series
Thermal Mass Flow Devices. The installation process consists of purging
the gas supply line prior to installation, unpacking and inspecting the
device, connecting the device to the gas supply line and testing for leaks.
2-2 Receipt of Equipment
When the instrument is received, the outside packing case should be checked
for damage incurred during shipment. If the packing case is damaged, the
local carrier should be notified at once regarding their liability. A report should
be submitted to your nearest Product Service Department.
Brooks Instrument
407 W. Vine Street
P.O. Box 903
Hatfield, PA 19440 USA
Toll Free (888) 554 FLOW (3569)
Tel (215) 362 3700
Fax (215) 362 3745
E-mail: BrooksAm@BrooksInstrument.com
www.BrooksInstrument.com
Europe
Brooks Instrument GmbH
Zur Wetterwarte 50 Haus 377/B,
01109 Dresden
Germany
Tel +49 (0) 351 215 20 442

Asia
Brooks Instrument Korea, Ltd.
D-406 Bundang Techno Park 151
Sungnam, Kyungki-do, 463-070
Korea
Tel +82 31 708 2522

In case you need technical assistance:


USA
888 275 8946
Taiwan
Europe
+49 351 215 2040
China
Japan
+81 3 5633 7100
Singapore
Korea
+82 31 708 2521

+886 3 5590 988


+86 21 5079 8828
+6297 9741

Remove the envelope containing the packing list. Outside of your clean area,
carefully remove the equipment from the packing case. Make sure spare parts
are not discarded with the packing material. Inspect the contents for damaged
or missing parts.
This device has been assembled, calibrated, and double-vacuum bagged in
a Class 100 clean room. In your semi-clean area, remove the outer bag
only. Pass the device into your clean area. Remove the second clean room
compatible bag only when the device is ready to be tested and/or installed
in your clean system.
2-3 Recommended Storage Practice
If intermediate or long-term storage of the device is required, it is
recommended that it be stored in accordance with the following conditions:

Within the original shipping container.

Ambient temperature 21C (70F) nominal, 32C (90F) maximum,


7C (45F) minimum.

Relative humidity 45% nominal, 60% maximum, 25% minimum.

2-1

Section 2 Installation

GF100 Series

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

2-4 Return Shipment


Prior to returning any device to the factory, visit the Brooks web site
(www.BrooksInstrument.com) for a Return Materials Authorization Number
(RMA#), or contact one of the locations provided on p. 2-1.
Prior to returning the device, it must be purged in accordance with the
following:

All flow devices returned to Brooks require completion of Form RPR003-1,


Brooks Instrument Decontamination Statement, along with a Material
Safety Data Sheet (MSDS) for the fluid(s) used in the instrument. Failure
to provide this information will delay processing by Brooks personnel.
Copies of these forms can be downloaded from the Brooks website
(www.BrooksInstrument.com) or are available from any of the Brooks
Instrument locations provided on p. 2-1.

2-5 Transit Precautions


To safeguard against damage during transit, transport the device to the
installation site in the same container used for transportation from the
factory, if circumstances permit.
2-6 Removal from Storage
Upon removal of the device from storage, a visual inspection should be
conducted to verify its "as-received" condition. If the device has been
subject to storage conditions in excess of those recommended (refer to
"2-3 Recommended Storage Practice" on p. 2-1), it should be subjected
to a pneumatic pressure test in accordance with applicable vessel codes.
To maintain a devices ultraclean integrity, this service should be
performed by the factory or one of the certified service centers.
2-7 Gas Connections
Prior to installation, ensure that all piping is clean and free from
obstructions. Install piping in such a manner that permits easy access to
the device if removal becomes necessary.
2-8 In-Line Filter
It is recommended that an in-line filter be installed upstream from the
device to prevent the possibility of any foreign material entering the flow
sensor or control valve. The filtering element should be replaced
periodically or ultrasonically cleaned.

2-2

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

Section 2 Installation

GF100 Series

2-9 Mechanical Installation

The recommended installation procedure guidelines are as follows:


The device should be located in a clean, dry atmosphere relatively free
from shock and vibration.
Leave sufficient room for access to the user interface, display and MAC
ID and baud rate switches (if equipped) at the top of the device.
Install the device in such a manner that permits easy purge and removal if
the device requires servicing.

The GF100 Series is equipped with PTI technology, which reduces the
effect of pressure fluctuations on gas flow. In PTI technology, a signal from
an integrated pressure transducer is combined with the standard thermal
sensor output. The combined signals allow precise and stable flow, even
when the line pressure is fluctuating.
The GF100 Series also utilizes MultiFlo technology that allows the user to
configure standard configurations ("SHs") or "blanks" for a variety of pure
gases and mixtures. As a result, MultiFlo technology enables the user to
reduce unique inventory requirements.

2-3

Section 2 Installation

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

GF100 Series
2-10 Flow Controller Installation Arrangement

Typical gas supply arrangements are shown in Figures 2-1 and 2-2.
GF100's are often arranged inside a gas panel. Configure standard
configurations ("SHs") or "blanks" for a variety of pure gases and mixtures.
As a result, MultiFlo technology enables the user to reduce unique
inventory requirements.

Figure 2-1 Typical Gas Supply Arrangement with non-PTI MFC

Figure 2-2 Typical Gas Supply Arrangement with PTI MFC

2-4

Section 2 Installation

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

GF100 Series

2-11 Purge the Gas Supply Line Before GF100 Series Installation

Before operating the GF100 Series, the gas supply line must be completely
purged with nitrogen or argon to ensure the line is free from toxic or
flammable gases, contaminants, moisture, and oxygen. The purge gas must
be free of moisture and oxygen to less than 100 ppb. Purge the gas lines as
follows or in accordance to prescribed company and safety procedures.
1. Shut off the process gas supply valve(s) upstream of the GF100 Series.
If such a valve is not available, shut the valve on the gas panel. Tag the
valve at this point to prevent accidental re-exposure of the process gas
to the gas line.
2. Cycle purge the gas line with dry nitrogen or argon to fully flush out the
process gas. Cycle purging consists of evacuating to a low pressure
adequate to induce out-gassing and then purging to remove adhered
moisture and oxygen. If a toxic or reactive gas is present and a clogged
GF100 Series is suspected, then proceed with caution. Pump down and
purge the GF100 Series from both downstream and upstream lines. If
check valves are present in the gas line, both pumping down and
purging are required. Pumping down without purging is inadequate. If a
good vacuum source is not available, the GF100 Series can be decontaminated by purge only.
3. Repeat the purge cycle several times within 2-4 hours to complete the
cleaning. For toxic and corrosive gasses, it is recommended to use 100120 cycles.

2-5

Section 2 Installation

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

GF100 Series
2-12 Position and Mount the GF100 Series

Position the GF100 Series so that the gas flow is pointed in the direction of
the grey arrow on the GF100 Series label. The various mounting positions
are described in Figure 2-3

Figure 2-3 GF100 Series Mounting Attitude Positions

The standard orientation for the GF100 Series is Horizontal Base Down (HBD).
The GF125 & GF135 employ a proprietary algorithm that utilizes the internal
pressure sensor to compensate for potential orientation effects when the
MFC is used with certain higher density gases. Non HBD mounting
orientations can be selected by using the MultiFlo software.
In the case of the GF100/120 Series, which does not have an internal
pressure sensor, it is recommended that the MFC is re-zeroed with process
gas following the recommended Brooks procedure (see zeroing bulletin
FSB-001-0015 for futher information).
If your GF100 Series is configured with downported fittings, follow Steps 1
though 4 below. If your GF100 Series has VCR fittings, proceed to Step 5.
1. Refer to Figure 2-4. If downported fittings (1) are used, the GF100
Series is mounted to K1 Series substrate blocks (2) with four screws
(3). Metal C-seals or W-seals (4) (as provided by integrator) are
inserted between the GF100 Series and substrate blocks before the
screws are installed. These metal seals must be replaced after each
installation.
2. Select the mounting screws noted in Table 2-1 below for downported
devices. M4 screws are used on 1.125" devices, K1S. M5 screws are
used on 1.5" devices, K1R2 and K1H.
3. Refer to Figure 2-4. Insert the two mounting seals (4) over the gas flow
path of the K1 block. Carefully align the GF100 Series mounting holes
onto the K1 substrate blocks. Using your fingers, install the screws
through the GF100 Series fitting and hand tighten.
2-6

Section 2 Installation

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

GF100 Series

3
1
4
2

Figure 2-4 GF 100 Series Mounted to K1 Series Substrate Blocks

4. Using a torque wrench and a metric hex key, tighten the screws to the
torque value as described in Table 2-2 and Torque Pattern Figure 2-5.

Table 2-1 K1 Series Fasteners


Fastener Size

Connection
GF 100 Series
to
Subtstrate

K1S

K1R2

K1H

M4 x 34mm
or
M4 x 35mm

M5 x 30mm

M5 x 37mm

5. If your GF100 Series is configured with " VCR fittings, secure the
GF100 Series block to the gas panel with two, 8-32-UNC-2B" screws.
Then connect the inlet/ outlet fittings to the gas supply line using two
wrenches. Tighten the fittings to manufacturer recommendations.

Table 2-2 K1 Substrate Torque Data


Connection

Torque Pattern

GF125 to
Subtstrate

Use a square pattern as


shown in Figure 2-5.
Start at 25 inch-pounds and
increase in increments of
10 inch-pounds until proper
value is obtained.

Torque (Inch-Pounds)
K1S

K1R2

K1H

45

45

45

2-7

Section 2 Installation

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

GF100 Series

Figure 2-5 Mounting Screws Torque Pattern

2-13 Perform a Leak Test

! WARNING

It is critical to leak test the gas supply lines and GF100 Series connections
before turning on the process gas supply after any new installation. Check
for leaks using a helium leak detector or any other appropriate leak test
method. Follow leak test specifications as defined by integrator.
2-14 Zeroing Setup Process
The following steps are required before the GF100 Series is zeroed.
1. Make sure that the GF100 Series has been installed inside the
equipment (panel) for at least four hours and powered up at least one
hour prior to zeroing. This insures that the GF100 Series is in its "use
attitude" and is operating at normal temperature. If the GF100 Series is
subjected to a vacuum purge for more than one minute, turn off the
GF100 Series (ie., provide a zero setpoint) for a time period of twice the
vacuum purge time.
2. Refer to Figure 2-1. Open the upstream shut-off valve (5) and close the
downstream shut-off valve (7). This eliminates a pressure drop across
the GF100 Series and subsequent leakage from the PID control valve
inside the GF100 Series.
3. Provide a 100% setpoint to the GF100 Series for no longer than 60
seconds. This equalizes the pressure across the PID control valve.
4. Refer to Figure 2-1. Close the upstream shut-off valve (5) to prevent any
pressure effects from the regulator (3).
5. Close the GF100 Series and wait two minutes.
6. Read the output signal of the GF100 Series. This output signal is the
initial flow in percent of full scale. The output signal should be 0.0 (
0.1%). If the output signal is too high, re-zero the GF100 Series as
described in Section 2-15.

2-8

Section 2 Installation

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

GF100 Series

2-15 Zeroing the GF100 Series


Many high density gases exhibit slight changes in zero output as a function
of inlet pressure. Gases such as tungsten hexafluoride and many
fluorocarbons are especially sensitive to this problem. Since inlet pressure
is a potential source for zero errors, the pressure transducer on each
GF100 Series should be correctly set to zero after installation. The zeroing
process is performed from the backlight LCD display on top of the GF100
Series.
OEM tools using a microprocessor or computer for operating the GF100
Series should sequence the GF100 Series off between processes. To
accomplish this, simply provide a zero set point. The GF100 Series will shut
off automatically.

Shut-off valves, whether upstream or downstream from the GF100 Series,


should be programmed to turn on before the GF100 Series is turned on and
turned off after the GF100 Series is turned off.
2-15-1 Zeroing the GF100 Series Pressure Transducer from the LCD Display Panel
1. Place the GF100 Series under a strong vacuum with the GF100
Series set to 100% set point. Make sure that upstream valve is
closed and the downstream valve is open. Allow time for the
upstream pressure to bleed off.
2. Looking at the top of the GF100 Series, press the "Display" button,
starting at the MACID, four times to "PSI" or five times to "kPa" or until
the LCD displays the labels "PSI" or "kPa". The GF100 Series will
display pressure in units of PSIA or kPa. Press and hold down the Zero
button a minimum of 5 seconds or until the display reads 0.000, with the
last two digits flickering at different values. The pressure transducer
zeroing procedure can be done while the display is either in PSIA or
kPA output. Refer to Figure 2-6.

Figure 2-6 Display with PSI Reading


2-9

Section 2 Installation

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

GF100 Series

After completion of the pressure transducer zeroing, the LCD display will
read 0.0 with the last two digits flickering as shown in Figure 2-7.

Figure 2-7 Display Reading Zero PSI

2-15-2 Zeroing GF100 Series Flow from LCD Display Panel


1. Place the GF100 Series under normal inlet operating pressure. Close
the down stream valve to prevent any flow.
2. Looking at the top of the GF100 Series, press the "Display" button until
the LCD display label is "%FS" as shown in Figure 2-8. Three button
depressions from the MACID label display.

Figure 2-8 Display Set to %FS

3. Press and hold down the Zero button for a minimum of 5 seconds or until
the "%FS" display reads 0.0 as shown in Figure 2-9. The %FS label will
flash during this procedure.

Figure 2-9 % Flow Display Set to Zero


2-10

Section 2 Installation

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

GF100 Series

2-16 Performance Checks

This section describes how to zero and sequence the GF100 Series
devices for proper operation.

The GF100 Series must be warmed up for at least 30 minutes.


The active gas page must be correct.
The GF100 Series pressure transducer must be correctly zeroed.
The GF100 Series flow must be correctly zeroed.
This chapter describes how to zero and sequence the GF100 Series for
proper operation.
2-17 Introduction to the MultiFloTM Configurator
The MutliFlo Configurator application is used to configure the gas and
range of the GF Series devices. The following section describes the
MultiFlo Configurator and its uses.

2-18 Using the MultiFlo Configurator


The MultiFlo Configurator application allows communication to GF Series
devices through personal computer with serial COM Port and a Windows
XP or Windows 7 operating system. Its primary function is to configure
gas and flow ranges within six defined standard configurations. Flow
ranges are configured to the Nitrogen equivalent.
Using the MultiFlo Configurator software, configure the gas and flow rate
according to Table 2-3.
Table 2-3 Gas and Flow Ranges - MultiFlo Configurable - N2 Equivalent

Standard MG-MR Bin


Configurations
SH40
SH41
SH42
SH43
SH44
SH45
SH46
SH47
SH48
SH49
SH50

Flow range
Code
010C
030C
092C
280C
860C
2.6L
7.2L
015L
030L
040L
055L

Gas Flow Range


(N2 Equivalent)
3-10 sccm
11-30 sccm
31-92 sccm
93-280 sccm
281-860 sccm
861-2600 sccm
2601-7200 sccm
7201-15000 sccm
15001-30000 sccm
30001-40000 sccm
40001-55000 sccm
2-11

Section 2 Installation

GF100 Series

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

Figure 2-10 MultiFlo Cable Adaptor

The MultiFlo Configurator interfaces to the GF Series device through RS485


or DeviceNet communications. There are various ways to connect the
device regardless of device configuration. Devices may be connected
through the diagnostic port using cables in one of the two Basic MultiFlo
Configurator Kits or DeviceNet devices can alternatively be connected
using a National Instruments or SST DeviceNet scanner card.
778Z010ZZZ Basic MultiFlo Configurator Kit
A331710003 Cable Assembly 2.5mm
A332300001 Converter 232/485
778Z011ZZZ Basic MultiFlo Configurator Kit
w/Power Supply and Adapter Cables
A331710003 Cable Assembly 2.5mm
A332300001 Converter 232/485
A332295001 Power Supply MFC
A332297002 Cable, Power, 9-Pin
A332297001 Cable, Power, DeviceNet
Connect the MultiFlo Cable Adapter 2.5mm jack to the Diagnostic Port on
the top of the device. See Figure 2-12.
Connect the RS-485 end of the converter to the 9-Pin RS-485 end of the
MultiFlo Cable Adapter.
Connect the RS-232 end of the converter to the Serial Port of a laptop or PC.
The latest MutliFlo Configurator Software and Databases and MultiFlo
Configurator Quick Start Guide are available on the Brooks Instrument
website at: www.BrooksInstrument.com/MultiFlo. Please reference the

2-12

Section 2 Installation

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

GF100 Series

Figure 2-11 RS232/RS485 Converter (P/N A33231)

Diagnostic Port

Figure 2-12 Diagnostic Port

MultiFlo Configurator Quick Start Guide for installation and operation details.
Download the MultiFlo Configurator software into your computer from the
Brooks Instrument website www.BrooksInstrument.com/MultiFlo. Install the
MultiFlo Configurator as described in the MultiFlo Configurator Quick Start
Guide and use the guide as a reference for operation details.

2-13

Section 2 Installation

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

GF100 Series
2-19 Electrical Connections

2-19-1 DeviceNet Connections


DeviceNet is a 5 wire local network connection that employs a command
response communication protocol for communicating between a master and
slave. Obtain a DeviceNet communication cable (Micro M-12) and fasten it
to the 5-pin connector as shown in Figure 2-13.
DeviceNet (M-12)
Connector Cable

DeviceNet 5-Pin (M-12)


Connector (M)

1
5

Pin Connection
1
Drain
2 V+ (11-25 Vdc)
3
V-4
CAN-H
5
CAN-L

Figure 2-13 GF100 Series DeviceNet Connection

2-14

Section 2 Installation

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

GF100 Series

2-19-2 Analog/RS485 Connections


The GF100 Series devices are available with Analog 9-Pin D-Connectors
shown in Figure 2-14.

Analog 9-Pin
Connector

(Refer to Figure 2-15


for Pin-Out Details)

Figure 2-14 GF100 Series with 9-Pin Analog Connector

PIN
1
2
3
4
5
6
7
8
9

Connection
Valve Control
Output (0-5 Vdc)
+15 Vdc
Power Common
-15 Vdc
Setpoint (0-5 Vdc)
Signal Common
RS485 DX+
RS485 DX-

Figure 2-15 Analog 9-Pin Connector (M)


2-15

Section 2 Installation

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

GF100 Series
Base I/O Options
PDC Ordering Code G1
(GF135 use Code G2)
Description: Industry standard
Analog / RS485 interface
123

PDC Ordering Code GX


Description: OEM specific
Analog / RS485 interface.
Display and top plate
re-oriented 180o
123

PDC Ordering Code DX


(GF135 use Codes D0-D9
and DA-DX)
Description: Industry standard
ODVA compliant DeviceNet
interface

Pin No.
1
2
3
4
5
6
7
8
9

Pin No.
1
2
3
4
5
6
7
8
9

M12
Pin No.
1
2
3
4
5

Signals
Valve Control
Output (0-5 Vdc)
+15 Vdc
Power Common
-15 Vdc
Setpo int (0-5 Vdc)
Signal Commo n
RS-485 (DX+)
RS-485 (DX-)

Signals
Valve Control
Output (0-5 Vdc)
+15 Vd c
Power Common
-15 Vdc
Setpo int (0-5 Vdc)
Signal Commo n
RS-485 (DX+)
RS-485 (DX-)

Signals
Drain
V+ (11-25 Vdc)
VCAN-H
CAN-L

PDC Ordering Code TX


Description: Industry standard
Analog only interface

Pin No.
1
2
3
4
5
6
7
8
9

PDC Ordering Code SX


Description: Industry standard
Analog 9-Pin Sub D connector
and dual RJ11 RS485 ports

D-Sub
Pin No.
1
2
3
4
5
6
7
8
9
RJ11 J2
Pin No.
3
2

PDC Ordering Code BB


Description: Industry standard
ODVA compliant DeviceNet
interface, Plus a separate
Analog 0-5 Vdc Connector

M12
Pin No.
1
2
3
4
5
HIROSE
Pin No.
1
2
3
4

Signals
Valve Control
Output (0-5 Vdc)
+15 Vd c
Power Common
-15 Vdc
Setpo int (0-5 Vdc)
Signal Commo n
No Connection
No Connection

Signals
Valve Control
Output (0-5 Vdc)
+15 Vdc
Power Common
-15 Vdc
Setpo int (0-5 Vdc)
Signal Commo n
Signal Commo n
Va lve Test Point
Si gnals
RS-485 (DX+)
RS-485 (DX-)

Signals
Drain
V+ (11-25 Vdc)
VCAN-H
CAN-L
Signals
Flow Out
AGND
GPIO_CAP0
GHD_Earth

All Base I/O options include: Diagnostic port communication RS485 via 2.5mm jack

I/O Options Using Base Model and Adapter Cable


A range of low profile adapter cables have been deveoped to support replacing older
generation MFC's with different pinout configurations. The base MFC will be either a, G1,
TX or SX configuration, depending on the product being replaced.
PDC Ordering Code UX
Description: SX base I/O with 7003550
adapter for
compatability with
Unit UDU15
Pin No
9
6
4
7
11
15
1,13,14
2
12
8
3,5,10

Signals

Pin No

Signals

VALVE OFF
OUTPUT ( 0-5 VDC)
+ 15 VDC
POW ER COMMON
- 15 VDC
SETPOINT ( 0-5 VDC )
SIGNAL COMMON
ZERO ALARM
VALVE TEST POINT
CASE GROUND
NO CONNECTION

15
2
5
1
6
8
9
10
14
3,4,7
11,12,13

VALVE OFF
OUTPUT ( 0-5 VDC)
+ 15 VDC
COMMON
- 15 VDC
SETPOINT ( 0-5 VDC )
COMMON
COMMON
CASE GROUND
NO CONNECTION
NO CONNECTION

PDC Ordering Code: EX


Description: GX base I/O with
7003083 adapter for compatability
with Unit E, IN L, R
Signals
VALVE OFF
OUTPUT ( 0-5 VDC)
+ 15 VDC
POWER COMMON
- 15 VDC
SETPOINT ( 0-5 VDC )
SIGNAL COMMON
CASE GROUND
NOT CONNECTED
NOT CONNECTED
KEY WAY

Pin No
J
3
4
2
F
A
B,C,10
1
5, 6, 8, 9
I, D, E, H
7,G
RJ11 J2
Pin No

2
3

2-16

RJ11 J3
Pin No

3
4

PDC Ordering Code: T1


Description: TX base I/O with 7003551
adapter for
compatability with
IFlow DB15 & TN 15 pin

PDC Ordering Code: KX


Description: G1 base I/O with
7003298 adapter for compatability
with Unit UDK15
Pin No
3
2
7
5
6
8
11,12
15

Signals
VALVE CONTROL
OUTPUT ( 0-5 VDC)
+ 15 VDC
POWER COMMON
- 15 VDC
SETPOINT ( 0-5 VDC )
SIGNAL COMMON
CASE GROUND

1, 4, 9, 10,
13, 14

NO
CONNECTION

RS-485 (DX-)
RS-485 (DX+)

Figure 2-16 GF100 Series Electrical Interface Options

PDC Ordering Code: FX / JX


Description: SX base I/O with 7003069
(FX)/7001814 (JX)
adapter for compatability with
Unit UDF9/UDJ9
Pin No
1
2
3
4
5
6
7
8
9

Signals
VALVE CONTROL*
OUTPUT ( 0-5 VDC )
+ 15 VDC
POWER COMMON
- 15 VDC
SETPOINT ( 0-5 VDC )
SIGNAL COMMON
SIGNAL COMMON
VALVE TEST POINT

PDC Ordering Code: BX


Description: G1 base I/O with
7003590 adapter for
compatability with Brooks 15-Pin D
Pin No
12
2
5
9
6
8
1,10
3,4,7,11
13,14,15

Signals
VALVE OVERRIDE
OUTPUT ( 0-5 VDC)
+ 15 VDC
POWER COMMON
- 15 VDC
SETPOINT ( 0-5 VDC )
SIGNAL COMMON
NO CONNECTION
NO CONNECTION

Other adapter options are available for the GF 100 Series.


Please contact Brooks Customer Service for more information.

Section 2 Installation

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

GF100 Series

2-20 GF135 Inlet Valve Installation


When installing the GF135
1. The inlet pneumatic isolation valve supply line needs to be disconnected
from the inlet pneumatic isolation valve.
2. Next the inlet valve supply line needs to be connected to the inlet tube on
the MFC. The tube is labeled IN and a connector is provided with the
GF135 to connect the customer inlet valve supply tube to the MFC IN tube.
3. Next the MFC tube labeled OUT should be connected to the customer
inlet pneumatic isolation valve.
Note The air passes through a failsafe normally open valve inside the
GF135 and is connected back to the inlet isolation valve. For safe operation
and in line with S2 compliance, the GF135 cannot override and open the
inlet isolation valve.

Outlet - Pneumatic
Isolation Valve Line

Inlet - Pneumatic
Isolation Valve Line

Figure 2-17 GF135 Series Showing Inlet/Outlet Pneumatic Isolation Valve Lines

2-17

Section 2 Installation

GF100 Series

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

2-21 GF135 Commissioning

Download the GF135 commissioning software, part number 805-C-026,


from the Brooks Instrument website. The software will need to be
downloaded onto a computer with a Windows based operating system.
Necessary equipment:
Computer that runs a Windows operating system
Power (either DeviceNet or RS485) to the device
Need gas supply for pneumatic valve
BB 485 to 232 converter
9-pin to diagnostic cable
In order to ensure best results, commission the GF135 at operating
temperature and pressure on tool.

1. Start up the GF135 Commission Application and select a port. This will
bring up all of the devices on this port.

2-18

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

Section 2 Installation

GF100 Series

2. Select the single device that you want to perform the commissioning and
click on "Connect to Device". The Commissioning button becomes enabled.

3. Click on the "Commission" button to begin the process. Some


intermediate messages will appear on the bottom mesage bar. You will then
be presented the option to override the current inlet gas pressure. Selecting
"Yes" will keep the current inlet pressure. Selecting "No" will allow the user
to overrided the pressure. Selecting "Cancel" will end the commissioning
process.

2-19

Section 2 Installation

GF100 Series

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

4. The user will be presented the option to override a few of the parameters
being tested prior to actual commissioning. Selecting "Yes" will continue the
process. By selecting "No" the commissioning process will end.

5. After numerous checks are performed, the auto-volume will start. This will
take approximately 5 minutes to complete. The K1 setpoints will then begin.
This process will run for approximately 40-90 minutes.

2-20

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

Section 2 Installation

GF100 Series

6. Once the auto-volume and the K1 set points are complete, the program
will go back to the initial screen. In addition, the overall run time will be
displayed. From here, another device (or the same one again) may be
selected for commissioning.

2-21

Section 2 Installation

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

GF100 Series

THIS PAGE WAS


INTENTIONALLY
LEFT BLANK

2-22

Section 3 Operation

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

GF100 Series

3-1 General
After the device has been properly installed in the process, it is ready for
operation. When initiating flow, slowly open any upstream shutoff valve to
avoid a flow surge. A bypass is helpful in bringing the flow on smoothly.
Avoid starting a pump to supply the device without the use of a valve
upstream of the device.

3-2 Theory of Operation for Flow Measurement


The thermal mass flow measurement system consists of two components:
the restrictor or bypass and the flow sensor. Figure 3-1 is diagram of the
flow stream through the device, with an enlarged view of the flow sensor.
Gas flow entering the device is separated into two paths; one straight
through the restrictor and the other through the flow sensor. This is
represented in Figure 3-1, where the total flow A+B enters the device and
is separated into streams A and B. The streams are joined again at the far
side of the restrictor.
The separation of the flow streams is caused by the restrictor. During flow
conditions, there will be a pressure differential across the restrictor that
forces gas to flow in the sensor.
The pressure difference caused by the restrictor varies linearly with total
flow rate. The sensor has the same linear pressure difference versus flow
relationship. The ratio of sensor flow to the flow through the restrictor
remains constant over the range of the device (A/B = constant). The full
scale flow rate of the device is established by selecting a restrictor with the
correct pressure differential for the desired flow.
The flow sensor is a very narrow, thin-walled Hastelloy tube. Onto this
tube are built upstream and downstream temperature sensing and heating
elements. During no-flow conditions, the amount of heat reaching each
temperature sensor is equal, so temperatures T1 and T2 (Figure 3-1) are
equal. Gas flowing through the tube carries heat away from the upstream
temperature sensor and toward the downstream sensor. The temperature
difference, T2 T1, is directly proportional to the gas mass flow.
GF100 Series operating principles are described in Figure 3-1.

3-1

Section 3 Operation

GF100 Series

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

Operating principles are listed below by following #1-5.


1. The GF125 uses a pressure transducer and support circuitry that sums the transducer signal in the sensor path
(transducer not included in the GF100 and the GF120). Incoming pressure fluctuation results in a signal that is
proportional to fluctuation.
2. The signal from the pressure transducer is inverted and then summed in with the original sensor signal.
3. The two summed signals cancel each other out.
4. The sensor signal that is applied to the microprocessor is undisturbed.
5. The flow of process gas is undisturbed by incoming pressure fluctuations.
Figure 3-1 GF100 Series Operating Principles

3-2

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

Section 4 Maintenance & Troubleshooting

GF100 Series

4-1 Overview
No routine maintenance is required on the Brooks GF100 Series devices.
If an in-line filter is used, the filtering elements should be periodically
replaced or cleaned. Any precision unit such as a flow controller requires
occasional servicing, especially if it has been operating for an extended
period of time. If reactive gases are being used, it is recommended that
you send the device to a Brooks Service Center for cleaning and
recalibration. Please follow the instructions for removal, product packaging
and product return instructions found in Section 2- InstallationReturn
Shipment. All active process instrumentation and equipment is subject to
aging and wear from their environment. This includes temperature,
mechanical stress, component tolerance shift, contaminant buildup,
oxidation, and other influences. The effects are gradual, but over time the
changes can affect the accuracy of even the best equipment. Therefore, it
is recommended to re-zero the device at 6 month intervals. Refer to
Section 2-18 Performance Checks for re-zeroing instructions.

4-2 Maintenance

4-1

Section 4 Maintenance & Troubleshooting

GF100 Series

4-2

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

Section 4 Maintenance & Troubleshooting

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

GF100 Series

4-3 Troubleshooting

This section includes a Troubleshooting Checklist and a GF100 Series


Troubleshooting Guide that identifies symptoms, possible causes, and
corrective actions.

4-4 Troubleshooting Checklist


1. Check environmental factors that could affect changes to GF100 Series
performance. The most common environmental factors are listed in
Table 4-1.
Table 4-1 Environmental Factors
GF100 Series Performance

Possible Causes

Inaccurate flow.

Temperature shift (steady state or transient).


Inlet pressure shift (steady state or transient).
Power supply problem.
Electrical interference
Dirty gas chamber
Changes in gas.

Control problems. Can


not reach setpoint.
Oscillation.

Differential pressure not within operating range


Inlet pressure not stable

Zeroing problems, Indicated


zero is not stable.

Temperature shift (steady state or transient).


Inlet pressure shift (steady state or transient).
Power supply problem.
Electrical interference

2. Check supply voltage and check for a consistent ground.


3. Insure OEM tool setpoint matches the setpoint at the GF100 Series.
Observe for consistency.
4. Verify isolations valves are open and the gas supply is turned on. Then
verify operating pressures are within operating ranges.
5. Check GF100 Series voltage response by moving the setpoint back
and forth.
Observe for voltage changes.

4-3

Section 4 Maintenance & Troubleshooting

GF100 Series

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

4-5 GF100 Series Troubleshooting Guide

Table 4-2 GF100 Series Troubleshooting Guide

Symptoms & Possible Causes


1. No gas flow.
Is the gas supply turned on?
Is the regulator turned on at the
correct operating pressure?
Are any upstream or downstream
shut-off valves closed, either by
the system or because of failure?
Is the MOD LED light on the GF100 Series
lit solid green?
Is the commanded setpoint from
tool/system at 0.00 Vdc?
Has the been commanded
off by an active valve
closed input?
2. Flow out of range.
Is the gas inlet/outlet pressure differential
either too high or too low?
NOTE: If the differential pressure is too high,
voltage to the will be zero, which is
abnormally low for the setpoint. If the
differential is too low, voltage to will
be at its maximum value, which is abnormally
high for the set-point.
Is the MOD LED light on the GF100 Series
lit solid green?
Is the setpoint correct for the
required gas flow?
Is the calibrated for the
particular gas?
Is the zero correct?

3. No gas control; flow is at or


above maximum.
Is the gas pressure across the
too high?
Are system valves open, or is the
purge input activated?
Is the setpoint correct for the
required flow?

4-4

Corrective Action
Check shut-off valve and pressure readout. Open
the gas supply.
Turn off the regulator and reset it to the
recommended pressure as described in the
Data Sheet.
Verify that the valves are open and operating
properly.
Observe the LED display panel on top of to
verify. If the LED light is not lit, cycle power the
to reboot.
Use the tool software to verify.
Use the tool software to verify.

Verify that the pressure is correct for the gas


and range. If required, adjust inlet/outlet pressure
to achieve proper pressure reading.

Observe the LED display panel at top of .


If the LED light is not lit, cycle power the to
reboot.
Use the tool software to verify.
Check the side label. Run a
flow check to verify.
Zero the according to zeroing
procedure in Section 2-14. Verify leak check rates
are OK.

Verify that the pressure is correct for the gas and


range. If required, adjust inlet/outlet pressure to
achieve proper pressure reading.
Use tool software to verify.
Use tool software to verify.

Section 4 Maintenance & Troubleshooting

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

GF100 Series

Table 4-2 GF100 Series Troubleshooting Guide (Continued)

Symptoms & Possible Causes


4. No gas flow above some set-point.
NOTE: When the setpoint is
increased beyond this point,
the GF100 Series signal remains at
some value lower than the set-point.

Corrective Action

Is the gas inlet/outlet differential


pressure sufficient?

Verify that the pressure is correct for


the gas and range. If required, adjust
regulator to achieve proper pressure

NOTE: If the pressure reading is too low,


the valve voltage to the GF100 Series
will be at its maximum output. This condition
will cause internal GF valve heating and
inability to properly reach desired flow setpoints.
Is the GF100 Series calibrated for the
gas flow?

Check GF100 Series side label. Run a flow check to verify.


If flow is incorrect, replace the GF100 Series with a unit that
is calibrated properly.

5. No gas flow below some set-point.


NOTE: When the setpoint is
decreased below this point, the
GF100 Series signal remains at
some value higher than the
setpoint.
Is the gas inlet/outlet differential
pressure too high, or above published
setpoints?
NOTE: If the differential pressure reading.
is too high, voltage to the
GF100 Series will be at its maximum
value when the setpoint is
decreased below the point
where flow decreases.
Is the GF100 Series leaking?

6. Gas flow, or GF100 Series pressure


reading oscillates.
Is the GF100 Series calibrated for the
gas flowing?
Is there too much gas pressure
across the GF100 Series?

Verify that the pressure is correct for


this gas and range. If required, adjust
regulator to achieve proper pressure

Check for contamination. Test the GF100 Series


for leak integrity. Replace the Unit
GF100 Series if leakage is detected.

Check the GF100 Series side can label. Run


a flow check to verify. If flow is incorrect,
replace the GF100 Series.
Verify that the pressure is correct for
this gas and range. If required, adjust
regulator to achieve proper pressure
reading.

4-5

Section 4 Maintenance & Troubleshooting

GF100 Series

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

Table 4-2 GF100 Series Troubleshooting Guide (Continued)

Symptoms & Possible Causes


6. Gas flow, or GF100 Series pressure reading
oscillates.
Are inlet and outlet pressures stable?

NOTE: Most GF100 Series calibrated


with nitrogen will oscillate
with hydrogen or helium.

Corrective Action

If outlet pressure is unstable, check for


(no oscillation or hunting) a faulty vacuum pump,
or hunting at a downstream valve.
Check inlet pressure on tool. A faulty
pressure regulator can make the
GF100 Series appear to oscillate.
Adjust inlet pressure up or down by 2
psig increments until hunting disappears.
Verify common gas pressure is
within range.
NOTE: Hunting or oscillation can be
contributed to multiple GF100 Series sharing
a common gas manifold. Therefore,
inspect gas delivery sources to
the gas box. (for example; two tools
sharing a common gas bottle and
calling for gas at the same time.) Valve leak.
Unregulated gas pressure from Facilities.

7. GF100 Series does not read zero pressure


when gas is shut off.
Is the differential pressure across
the GF100 Series really zero?

Is the GF100 Series configured properly


in the tool software?
Is the GF100 Series mounted to the
proper attitude?
8. OEM tool does not read correct GF100 Series
zero reading.
Is the differential pressure across the
GF100 Series really zero?
Is the supply voltage within specified range?
Is the GF100 Series mounted in the proper attitude?
Is the flow output signal of the GF100 Series
really zero?
9. Zero Drift.
Improper zero of the GF100 Series?
Excessive Valve leakage?

4-6

Verify that the pressure is correct for


the gas and range. If the GF100 Series has
been contaminated, it may not be able
to close, and therefore, will not zero.
Equalize the pressure across the GF100 Series by
opening it briefly. Set up the GF100 Series for zeroing.
Then perform the zeroing procedure in Section 2-14.
Use the tool software to verify.
Refer to the side can label on the GF100 Series.
The GF100 Series should be calibrated
in the attitude it will be operating at.

GF100 Series valve leakage.


Incorrect MFC zero.

GF100 Series aging or sensor stabilization.


Zero is not correct.

Section 4 Maintenance & Troubleshooting

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

GF100 Series

Table 4-2 GF100 Series Troubleshooting Guide (Continued)

Symptoms & Possible Causes


10. Calibration Drift.
Gas box temperature too high?
Is it linear offset?
11. GF100 Series indicates Overshoot.
If the tool is idle for an extended period of time,
high inlet pressure or contamination will cause
overshoot on first use.
12.OEM tool indicates the wrong full scale
value for GF100 Series.
Older version of Multiflo Configurator used
to program GF100 Series.
13. GF100 Series dumps large volume of
gas into chamber when setpoint is
commanded from the tool.
The tool is commanding a setpoint before the
pneumatic valves are opened.
GF100 Series and pneumatic timing may be offset.
GF100 Series overshoots.
14. Tool display output doesnt match
GF100 Series flow output.
Cable resistance causing offset in the tools
display.

Corrective Action
Zero is not correct.

Check GF100 Series zero.

4-7

Section 4 Maintenance & Troubleshooting

GF100 Series

THIS PAGE WAS


INTENTIONALLY
LEFT BLANK

4-8

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

Section 5 Product Description Code

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

GF100 Series

Table 5-1 GF100, GF120 & GF125 Series Product Description Code Table
Code Description
I.
Base Model Code
II.

Package / Finish Specifications

III.

Configurability

IV
IV..

Code Option
GF
100
120
125

Option Description
High Purity/Ultra High Purity Digital Mass Flow Controllers
Flow range 3 sccm -55 slpm N2 Eq.; + 1.0% SP Accuracy; 1 sec Response; 10 Ra
Flow range 3 sccm -55 slpm N2 Eq.; + 1.0% SP Accuracy; 500 msec Response; 4 Ra
Pressure Transient Insensitive (PTI) Flow range 3 sccm -55 slpm N2 Eq.; + 1.0% SP Accuracy;
300-500 msec Response; 4 Ra

C
X

MultiFlo capable. Standard bins or specific gas/range may be selected.


Not MultiFlo capable. Specific gas/range required. (must select w/ SD, SL or HA special application)

Special Application

XX
HA
SL
SD

Standard
High Accuracy Calibration; (GF125 only)
Safe Delivery System (GF120 Only) Full scale flow range; 4 to 25 sccm, Nitrogen Equivalent
Safe Delivery System (GF120 Only) Full scale flow range; >25 sccm to 1 slpm, Nitrogen Equivalent

V.

Valve Configuration

O
C
M

Normally Open valve (not available with SD, SL or HA options)


Normally Closed valve (must select with SD, SL or HA special application)
Meter (No Valve)

VI.

Gas or SH MultiFlo Bin

XXXX XXXX
SH40 010C
SH41 030C
SH42 092C
SH43 280C
SH44 860C
SH45 2.6L
SH46 7.2L
SH47 015L
SH48 030L
SH49 040L
SH50 055L

VII. Fitting

VX
CX
DX
EX
WX
YX
AX
BX
LX

Specific Gas Code & Range, i.e. 0004 = Argon and 010L = 10 slpm (must select w/ SD, SL or
HA special application).
*Standard Configuration #40, 3-10 sccm Nitrogen Equivalent
*Standard Configuration #41, 11-30 sccm Nitrogen Equivalent
*Standard Configuration #42, 31-92 sccm Nitrogen Equivalent
Standard Configuration #43,93-280 sccm Nitrogen Equivalent
Standard Configuration #44, 281-860 sccm Nitrogen Equivalent
Standard Configuration #45, 861-2600 sccm Nitrogen Equivalent
Standard Configuration #46, 2601-7200 sccm Nitrogen Equivalent
Standard Configuration #47, 7201-15000 sccm Nitrogen Equivalent
Standard Configuration #48, 15001-30000 sccm Nitrogen Equivalent
Standard Configuration #49, 30001-40000 sccm Nitrogen Equivalent
Standard Configuration #50, 40001-55000 sccm Nitrogen Equivalent
1-1/2 body width, 124mm 1/4 VCR male
1-1/8 body width, 92mm C Seal
1-1/8 body width, 79.8mm C Seal
1-1/2 body width, 79.8mm W Seal
1-1/8 body width, 92mm W Seal
1-1/8 body width, 79.8mm W Seal
1-1/2 body width, 92mm C Seal
1-1/2 body width, 92mm W Seal
1-1/8 body width, 92mm C Seal w/Poke Yoke

VIII. Downstream Condition

A
V

Atmosphere
Vacuum; Default for SD, SL and HA special application

IX.

Default Sensor Orientation

Sensor

Note: *Bins SH40, SH42, SH42 are not avaiable with normally open valve.

5-1

Section 5 Product Description Code

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

GF100 Series

Table 5-1 GF100, GF120 & GF125 Series Product Description Code Table (Continued)
Code Description
X. Connector

Code Option
BX
DX
EX

Option Description
Cable adapter to 15 pin D Brooks (Unit B,N)
5-Pin DeviceNetTM micro (Unit D, IN D)
Cable adapter to Cardedge (w/out VTP), RS485 through RJ11 jacks (UnitE; IN L, R);
display and overlay 180O orientation
Cable adapter with 9 pin STEC pin-out & jack screws (w/VTP) (UnitF,O)
9-Pin D with RS485 (UnitG); display and overlay 180o orientation
9-Pin D with RS485 (UnitG)
Cable adapter with 9 pin STEC pin-out & jack screws (w/VTP) (UnitJ,W)
Cable adapter to MKS 15-Pin D (Unit K)
9 pin D with STEC pin-out (w/VTP) (UnitS,Q)
9 pin D with UDT9 pin-out (UDT9)
Cable adapter to 15 pin D (IFlow DB15 & TN 15 pin)
Cable adapter to 15 pin D (w/VTP) (Unit & TN U)
DeviceNetTM Analog (Not Available on 80mm fitting DX, YX, EX)

FX
GX
G1
JX
KX
SX
TX
T1
UX
BB

DeviceNet Standard Configuration Parameters


Poll IO
Poll IO
Poll IO External
Power On Full Scale Full Scale Full Scale Instance Instance
State
Baud
Connector
State
Setting
Setting
Setting Producer Consumer Transition
Rate

I/O
D0
D1
D2
D3
D4
D5
D6
D7
D8
D9
DA
DB
DC
DD
DE
Code Description
XI. Customer Special Request

DeviceNET
DeviceNET
DeviceNET
DeviceNET
DeviceNET
DeviceNET
DeviceNET
DeviceNET
DeviceNET
DeviceNET
DeviceNET
DeviceNET
DeviceNET
DeviceNET
DeviceNET
Code Option
XXXX
0924
0925

Idle
Idle
Idle
Idle
Executing
Idle
Idle
Idle
Idle
Executing
Idle
Idle
Idle
Executing
Executing

Count
Count
SCCM
Count
Count
Count
Count
Count
Count
Count
Count
Count
Count
Count
Sccm

Integer
Integer
Float
Integer
Integer
Integer
Integer
Integer
Integer
Integer
Integer
Integer
Integer
Integer
Float

6000h
6000h
7FFFh
6000h
6000h
6000h
7FFFh
7FFFh
6000h
6000h
7FFFh
6000h
7FFFh
7FFFh
6000h

2
21
13
22
22
6
3
6
3
2
22
22
3
22
15

A
X

Auto Shut-Off (Included) Default for SD and SL special application


Auto Shut-Off (Not Included)

XIII. Auto Zero

A
X

Auto Zero (Included)


Auto Zero (Not Included)

Sample Standar
d Model Code
Standard
I
II
III
GF
120
C

000

IV
XX

7
7
19
7
8
8
7
8
7
7
7
8
7
8
19

Executing
Executing
Executing
Executing
Executing
Executing
Executing
Executing
Executing
Executing
Executing
Executing
Idle
Executing
Executing

0oC Reference Calibration (Standard) - Default Setting

V
M

VI
VII
- SH40010C - VX

VIII
A

IX
0

X
GX

XI
XXXX

XII
A

XIII
X

VI
- XXXXXXXX -

VII
CX

VIII
V

IX
0

X
FX

XI
XXXX

XII
A

XIII
X -

XIV
000

Sample Safe Delivery System (SDS) Model Code


I
III
IV
VI
II
V
GF
120
X
SD
C - XXXXXXXX -

VII
EX

VIII
V

IX
0

X
SX

XI
XXXX

XII
A

XIII
X -

XIV
000

Sample High Accur


acy (HA) Model Code
Accuracy
I
III
IV V
II
GF
120
X
HA

500KB
500KB
500KB
500KB
500KB
500KB
500KB
500KB
500KB
500KB
500KB
500KB
500KB
500KB
500KB

Option Description
Customer Special Request Number; required with DX, BB Conn. Option to define DNet settings
Generic DeviceNet 2/7 6OOOH. Firmware = 1.10
Generic DeviceNet 6/8 7FFFH. Firmware = 1.10

XII. Auto Shut-Off

XIV
XIV.. Reference Temperature

5-2

5 Pin Micro
5 Pin Micro
5 Pin Micro
5 Pin Micro
5 Pin Micro
5 Pin Micro
5 Pin Micro
5 Pin Micro
5 Pin Micro
5 Pin Micro
5 Pin Micro
5 Pin Micro
5 Pin Micro
5 Pin Micro
5 Pin Micro

XIV
- 000

Section 5 Product Description Code

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

GF100 Series

Table 5-2 Model GF135 Product Description Code Table


Code Description

Code Option

I. Base Model Code

GF

II. Package/FinishSpecifications

135

Option Description
Pressure Transient Insensitive (PTI) Ultra High Purity Advanced Diagnostic MFC

III. Configurability

Gas specific

IV
IV.. Special Application

XX

Standard Application

V. Valve Configuration

Normally Closed Valve

VI.
VI.Specific Gas Code & Range, XXXX XXXX

Specify Gas Code & Range, i.e. 0004 = Argon and 010L = 10 slpm

VII. Fitting

1 1/2" VCR 1/4"

VX
CX

1 1/8" C Seal 92mm

WX

1 1/8" W Seal 92mm (FUTURE RELEASE)

VIII. Downstream Condition


IX. Sensor

Atmosphere

Vacuum

Default Orientation

X. Connector

DeviceNet Standard Configuration Parameters

I/O

XI. Customer Special Request

Poll IO
Poll IO
Poll IO External
Power On Full Scale Full Scale Full Scale Instance Instance
State
Baud
Connector
State
Setting
Setting
Setting Producer Consumer Transition Rate

D0

DeviceNET

5 Pin Micro

Idle

Count

Integer

6000h

Executing

500KB

D1

DeviceNET

5 Pin Micro

Idle

Count

Integer

6000h

21

Executing

500KB

D2

DeviceNET

5 Pin Micro

Idle

SCCM

Float

7FFFh

13

19

Executing

500KB

D3

DeviceNET

5 Pin Micro

Idle

Count

Integer

6000h

22

Executing

500KB

D4

DeviceNET

5 Pin Micro Executing

Count

Integer

6000h

22

Executing

500KB

D5

DeviceNET

5 Pin Micro

Idle

Count

Integer

6000h

Executing

500KB

D6

DeviceNET

5 Pin Micro

Idle

Count

Integer

7FFFh

Executing

500KB

D7

DeviceNET

5 Pin Micro

Idle

Count

Integer

7FFFh

Executing

500KB

D8

DeviceNET

5 Pin Micro

Idle

Count

Integer

6000h

Executing

500KB

D9

DeviceNET

5 Pin Micro Executing

Count

Integer

6000h

Executing

500KB

DA

DeviceNET

5 Pin Micro

Idle

Count

Integer

7FFFh

22

Executing

500KB

DB

DeviceNET

5 Pin Micro

Idle

Count

Integer

6000h

22

Executing

500KB

Idle

500KB

DC

DeviceNET

5 Pin Micro

Count

Integer

7FFFh

Idle

DD

DeviceNET

5 Pin Micro Executing

Count

Integer

7FFFh

22

Executing

500KB

DE

DeviceNET

5 Pin Micro Executing

Sccm

Float

6000h

15

19

Executing

500KB

DX

DeviceNET

5 Pin Micro

G2

Analog/RS485

9 Pin D

NA

NA

NA

NA

NA

XXXX

TO BE DEFINED BY CSR
NA

Auto Shut Off (Included)

Auto Shut Off (Not Included)

XIII. Auto Zero

Auto Zero (Included)

Auto Zero (Not Included)

Example Model Code


I
II
III
GF 135
X

IV
XX

000

V
C

NA

Customer Special Request Number

XII. Auto Shut-Off

XIV
XIV.. Reference Temperature

NA

0C Reference Calibration (Standard) - Default Setting

VI
XXXX XXXX

VII
VX

VIII
A

IX
O

X
D1

XI
XXXX

XII
A

XIII
A

XIV
000

5-3

Section 5 Product Description Code

GF100 Series

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

GF100 Series Ordering Instructions


Refer to the Product Description Codes on the previous pages. Starting from the left, choose the product code
options as follows:
1. Required performance model.
a.Standard Performance, non-PTI: GF100
b.High Performance, non-PTI: GF120
c.High Performance, w/ PTI: GF125
d.High Performance, w/ PTI, w/ROD: GF135
2. Configurability
a.Disabled: X
b.Enabled: C, Standard (not an option for GF135)
3. Specialty Application
a.High Accuracy, for GF125 only: HA
b.Safe Delivery, for GF120 only: SD
c.No Specialty App: XX
4. Valve Configuration
a.Normally Closed: C
b.Normally Open: O
5. Gas or SH MultiFlo Bin
a.If Gas Specific, enter SEMI gas code: ex. 0013, for N2
b.If SH MuliFlow Bin: SHnn, nn being the required SHBin, 40 - 50
6. Maximum Flow
a.If Gas Specific, enter maximum range in sccm, "C" or slm "L": ex. 500C
b.If SH Bin, enter defined maximum flow: ex. 860, choosing SH44, 281 - 860 sccm
7. Fitting
a.Enter 2-character option code as defined: ex. CX, 1 1/8" C Seal 92mm
8. Downstream Condition
a.Outlet to Vacuum: V
b.Outlet to Atmosphere: A
9. Sensor
a.Orthogonal: O, this is default, non-selectable
10.Connector
a.Enter 2-character option code as defined: ex. DX, DeviceNet
11.CSR
a.Customer Special Requirement, contact Brooks Apps Engineering for review of requirement and
creation of CSR: nnnn
b.If DNET connector, CSR required to define DNET attributes: 0924, generic, ODVA Std. configuration
c.None Required: XXXX
12.Auto Shut-Off.
a.Enabled: A
b.Disabled: X
13.Auto Zero
a.Enabled: A
b.Disabled: X
14.Reference Temperature, Operating Temperature in Degrees C
a.0oC Reference Calibration (Standard): 000, default

Here is an example of a configured Product Description Code (PDC) for a GF125, Configurable, no Specialty
Application, Valve Normally Closed, MultiFlo for 281-860 sccm, 1 1/8" C Seal 92mm, outlet to Vacuum, default
sensor, DNET, no CSR, Auto Shut Off enabled, Auto Zero disabled and Default Reference Temperature:
GF125CXXC-SH44860C-CXVODX-0924AX-000

5-4

Appendix A: GF100 Series Gas Table

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

GF100 Series

A-1 GF100 Series Gas Table


Table A-1 GF100 Series Gas Table
(Reference the following pages, A-1 and A-2).
Table A-1A GF100 Series Gas Table - Codes 1-108, Bins SH40 to SH50
Gas
Code
1
2
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
22
23
25
27
28
29
31
32
33
34
35
37
38
39
42
43
46
48
49
53
54
58
60
62
63
67
69
70
72
77
85
88
89
94
97
99
104
108

Gas Symbol
He
Ne
Ar
Kr
Xe
H2
Air
CO
HBr
HCl
HF
N2
D2
O2
NO
HI
F2
Cl2
H2S
H2Se
CO2
N2O
CH4
NH3
PH3
SO2
CH3F
COS
AsH3
ClCN
C2H4
SiH4
C2H2
GeH4
COF2
BF3
CHF3
NF3
C2H6
B2H6
COCl2
PF3
CF4
SiH2Cl2
C3H6-b)
BCl3
ClO3F
ClF3
C2H7N
SiF4
C3H8
C2F4
Si2H6
GeF4
C4H8-i)
SiCl4

Gas Name
Helium
Neon
Argon
Krypton
Xenon
Hydrogen
Air
Carbon Monoxide
Hydrogen Bromide
Hydrogen Chloride
Hydrogen Fluoride
Nitrogen
Deuterium
Oxygen
Nitric Oxide
Hydrogen Iodide
Fluorine
Chlorine
Hydrogen Sulfide
Hydrogen Selenide
Carbon Dioxide
Nitrous Oxide
Methane
Ammonia
Phosphine
Sulfur Dioxide
Methyl Fluoride
Carbonyl Sulfide
Arsine
Cyanogen Chloride
Ethylene
Silane
Acetylene
Germane
Carbonyl Fluoride
Boron Trifluoride
Fluoroform (Freon-23)
Nitrogen Trifluoride
Ethane
Diborane
Phosgene
Phosphorus Trifluoride
Carbon Tetrafluoride (Freon-14)
Dichlorosilane
Propylene
Boron Trichloride
Perchloryl Fluoride
Chlorine Trifluoride
Dimethylamine
Silicon Tetrafluoride
Propane
Tetrafluoroethylene
DISILANE
Germanium Tetrafluoride
Butene
Silicon Tetrachloride

Min inlet pressure for vac. exhaust


(PSIA)
SH40-SH47 SH48 SH49 SH50
19.7
24.7
29.7
24.7
24.7
29.7
39.7
44.7
29.7
24.7
14.7
14.7
19.7
24.7
29.7
24.7
29.7
24.7
24.7
29.7
14.7
17.7
24.7
29.7
29.7
32.7
14.7
14.7
24.7
29.7
29.7
34.7
24.7
29.7
24.7
29.7
24.7
29.7
24.7
29.7
24.7
29.7
24.7
29.7
24.7
29.7
29.7
29.7
24.7
29.7
29.7
29.7
24.7
24.7
29.7
24.7
24.7
29.7
19.7
24.7
19.7
24.7
24.7
29.7
24.7
29.7
24.7
29.7
14.7
21.7
24.7
29.7
24.7
29.7
16.7
19.7
24.7
29.7
19.7
19.7
21.7
19.7
24.7
24.7
24.7
24.7
26.7
24.7
29.7
24.7
26.7
17.7
29.7
19.7
19.7
12.7
14.7
19.7
24.7
24.7
24.7
24.7
26.7
14.7
19.7
19.7
19.7
11.7
14.7
14.7
20.7
14.7
20.7
9.7
14.7
19.7
24.7
9.7
11.7
19.7
19.7
19.7
19.7
20.7
24.7
20.7
24.7
8.7

SH40
SH41
SH42
SH43
SH44
Low High Low High Low High Low High Low High
5
14
15
42
43 128 129 400 401 1194
5
14
15
42
43 129 130 400 401 1207
5
14
15
42
43 130 131 400 401 1214
4
11
12
32
33 100 101 300 301
930
3
6
7
19
20
58
59 178 179
546
3
10
11
30
31
92
93 280 281
860
3
10
11
30
31
92
93 280 281
860
3
10
11
30
31
92
93 280 281
860
3
8
9
25
26
77
78 235 236
723
3
10
11
30
31
92
93 280 281
860
3
10
11
30
31
94
95 281 282
880
3
10
11
30
31
92
93 280 281
860
3
10
11
30
31
94
95 280 281
880
3
10
11
30
31
92
93 280 281
860
3
10
11
30
31
92
93 280 281
860
3
5
6
15
16
46
47 141 142
432
3
9
10
27
28
83
84 254 255
780
3
6
7
19
20
57
58 173 174
531
3
8
9
25
26
76
77 232 233
713
3
7
8
22
23
66
67 202 203
620
3
7
8
22
23
69
70 209 210
642
3
7
8
21
22
65
66 200 201
611
3
8
9
23
24
71
72 215 216
660
3
8
9
24
25
73
74 223 224
685
3
7
8
22
23
67
68 205 206
629
3
6
7
17
18
52
53 157 158
483
3
7
8
22
23
67
68 204 205
625
3
7
8
20
21
60
61 183 184
562
3
6
7
18
19
55
56 170 171
510
3
5
6
15
16
46
47 142 143
435
3
6
7
17
18
54
55 163 164
501
3
6
7
18
19
56
57 170 171
523
3
6
7
18
19
57
58 170 171
530
3
6
7
17
18
53
54 161 162
495
3
5
6
17
18
53
54 160 161
500
3
5
6
16
17
50
51 150 151
457
3
5
6
16
17
48
49 145 146
445
3
5
6
15
16
46
47 140 141
430
3
5
6
16
17
50
51 153 154
469
3
4
5
12
13
38
39 116 117
358
3
3
4
11
12
35
36 106 107
325
3
4
5
14
15
42
43 129 130
400
3
4
5
13
14
40
41 121 122
372
3
3
4
10
11
29
30
89
90
273
3
4
5
12
13
36
37 110 111
338
3
3
4
10
11
31
32
94
95
289
3
4
5
12
13
38
39 114 115
350
3
4
5
11
12
35
36 107 108
327
3
3
4
11
12
34
35 101 102
310
3
4
5
11
12
34
35 103 104
316
3
3
4
10
11
32
33 100 101
300
3
3
4
10
11
31
32 100 101
300
3
3
4
10
11
30
31
92
93
282
3
3
4
10
11
30
31
92
93
282
3
5
6
17
18
52
53
159
3
6
7
18
19
56
57
172

For reference only, consult factory on the latest tables.

SH45
Low High
1195 3609
1208 3650
1215 3671
931 2800
547 1651
861 2600
861 2600
861 2600
724 2187
861 2600
881 2610
861 2600
881 2600
861 2600
861 2600
433 1305
781 2358
532 1604
714 2155
621 1874
643 1942
612 1849
661 2000
686 2072
630 1901
484 1459
626 1890
563 1700
511 1550
436 1320
502 1516
524 1581
531 1600
496 1500
501 1500
458 1381
446 1344
431 1300
470 1418
359 1082
326 1000
401 1200
373 1123
274
824
339 1022
290
874
351 1060
328 1000
311
960
317 1000
301
910
301
900
283
853
283
860
160
480
173
520

SH46
SH47
SH48
SH49
Low
High
Low High
Low
High
Low
High
3610 11100 11101 23100 23101 47000 47001 100000
3651 10700 10701 22100
3672 10000 10001 20200 20201 41000 41001 42000
2801 7160 7161 14900
1652 4210 4211 8760
2601 8000 8001 16400 16401 33000 33001 73000
2601 7400 7401 15000 15001 30000
2601 7300 7301 15000 15001 30000
2188 5610 5611 11700
2601 6900 6901 14200 14201 29000
2611 7400 7401 15000 15001 30000
2601 7200 7201 15000 15001 30000 30001 40000
2601 8100 8101 16500 16501 33000
2601 7200 7201 15000 15001 30000 30001 37000
2601 7200 7201 15000 15001 30000
1306 3340 3341 6960 6961 13900
2359 6700 6701 14000 14001 28000
1605 4850 4851 10100 10101 20200
2156 5900 5901 12100 12101 24100
1875 4770 4771 10000 10001 20000
1943 5300 5301 11000 11001 22000 22001 28000
1850 5100 5101 10400 10401 21000 21001 25000
2001 5800 5801 12000 12001 24000 24001 46000
2073 6000 6001 12200 12201 25000 25001 46000
1902 5200 5201 10700 10701 21300
1460 3800 3801 7920 7921 15800
1891 5200 5201 10600 10601 21200
1701 4500 4501 9400 9401 18300
1551 4000 4001 8500 8501 17000
1321 3400 3401 7060 7061 14100
1517 4400 4401 9300 9301 18200
1582 4400 4401 9300 9301 18200
1601 4400 4401 9300 9301 18200
1501 4000 4001 8400 8401 16400
1501 4000 4001 8400 8401 16500 16501 21000
1382 3800 3801 7900 7901 15500
1345 3600 3601 7600 7601 15000 15001 17000
1301 3600 3601 7500 7501 15000 15001 17000
1419 4000 4001 8300 8301 16300
1083 3100 3101 6400 6401 12600
1001 2520 2521 5250 5251 10500
1201 3200 3201 6800 6801 13300
1124 3010 3011 6400 6401 12600 12601 17000
825 2140 2141 4450 4451 8900
1023 2800 2801 5900 5901 11700
875 2230 2231 4650 4651 9300
1061 2800 2801 5800 5801 11500
1001 2560 2561 5340 5341 10700
961 2530 2531 5400 5401 10600
1001 2600 2601 5400 5401 10600
911 2420 2421 5100 5101 10100
901 2300 2301 4900 4901 9800
854 2300 2301 4900 4901 9800
861 2200 2201 4700 4701 9400
481 1420 1421 2950 2951 5900
521 1320 1321 2750

SH50
Low High

42001 55000

40001 55000
37001 55000

28001 39000
25001 37000

17001 26000
17001 26000

17001 22000

(990-100-0082-R17)

A-1

Appendix A: GF100 Series Gas Table

GF100 Series

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

Table A-1B GF100 Series Gas Table - Codes 109-875, Bins SH40 to SH50
Gas
Code
109
110
111
117
118
121
128
129
138
155
160
185
190
266
270
297
354
368
371
509
516
527
528
536
537
542
557
563
595
597
603
604
606
607
615
626
632
649
653
654
662
674
676
693
695
698
700
701
762
766
780
805
820
824
837
875

Gas Symbol
C3H9N
SF6
C4H10-d)
C4H10-c)
C2F6
WF6
C3F8
C4F8
C3F6
C2HF5
CH2F2
CH6Si
(CH3)3SiH
C5F8
C4F6
C4F6-q)
C5F8O
Si3H9N
C4H11N
10%GeH4/H2
10%PH3/H2
10%PH3/N2
4.5%PH3/N2
20%O2/He
1%PH3/Ar
5%H2/N2
1%B2H6/H2
1%PH3/H2
3%B2H6/N2
3%H2/N2
30%He/O2
30%O2/He
4%H2/He
4%H2/N2
5%B2H6/Ar
5%PH3/Ar
50%PH3/SiH4
10%O2/He
2%SiH4/N2
5%B2H6/N2
.8%B2H6/N2
10%PH3/He
7.5%PH3/SiH4
5%PH3/He
2%B2H6/N2
10%GeH4/Ar
4%H2/Ar
10%B2H6/H2
5%H2/He
5%B2H6/He
5%SiH4/He
1%B2H6/Ar
15%B2H6/H2
2%SO2/N2
30%GeH4/Ar
20%O2/Ar

Min inlet pressure for vac. exhaust


(PSIA)
Gas Name
SH40-SH47 SH48 SH49 SH50
Trimethylamine
6.7
8.7
Sulfur Hexafluoride
19.7
19.7
19.7
20.7
Isobutane
19.7
19.7
19.7
20.7
Butane
19.7
19.7
19.7
20.7
Hexafluoroethane (Freon-116)
19.7
19.7
Tungsten Hexafluoride
9.7
11.7
Perfluoropropane
19.7
19.7
Octafluorocyclobutane
19.7
19.7
Hexafluoropropylene
19.7
19.7
PENTAFLUOROETHANE (FREO
19.7
19.7
Difluoromethane
24.7
29.7
Methylsilane (MONO)
9.7
14.7
Trimethylsilane (TMSi)
6.7
Octafluorocyclopentene
9.7
14.7
Hexafluoro-2-butyne
9.7
Hexafluoro Butadiene-1-3
14.7
14.7
Epoxyperfluorocyclopentene
19.7
19.7
Trisilylamine (TSA)
6.0
7.2
Dimethylethylamine
8.7
8.7
10%Germane/Hydrogen
20.7
20.7
10%Phosphine/Hydrogen
14.7
14.7
10%Phosphine/Nitrogen
19.7
24.7
4.5%Phosphine/Nitrogen
20.7
29.7
20%Oxygen/Helium
19.7
29.7
1%Phosphine/Argon
19.7
29.7
5%Hydrogen/Nitrogen
24.7
29.7
1%Diborane Hydrogen
24.7
29.7
1%Phosphine/Hydrogen
19.7
19.7
3%Diborane/Nitrogen
20.7
29.7
3%Hydrogen/Nitrogen
24.7
29.7
30%Helium/Oxygen
24.7
29.7
30%Oxygen/Helium
19.7
29.7
4%Hydrogen/Helium
19.7
19.7
4%Hydrogen/Nitrogen
24.7
29.7
5%Diborane/Argon
24.7
5%Phosphine/Argon
27.7
32.7
50%Phosphine/Silane
16.7
24.7
10%Oxygen/Helium
19.7
24.7
2%SILANE/NITROGEN
19.7
24.7
5%Diborane/Nitrogen
24.7
29.7
.8%Diborane/Nitrogen
24.7
29.7
10%Phosphine/Helium
14.7
19.7
7.5%Phosphine/Silane
14.7
14.7
5%Phosphine/Helium
12.7
14.7
2%Diborane/Nitrogen
24.7
29.7
10%Germane/Argon
24.7
29.7
4%Hydrogen/Argon
24.7
29.7
10%Diborane/Hydrogen
14.7
14.7
5%Hydrogen/Helium
19.7
19.7
5%Diborane/Helium
14.7
16.7
5%Silane/Helium
14.7
16.7
1%Diborane/Argon
14.7
16.7
15%Diborane/Hydrogen
14.7
14.7
2%Sulfur Dioxide/Nitrogen
14.7
14.7
30%Germane/Argon
14.7
14.7
20%Oxygen/Argon
14.7
14.7

SH40
SH41
SH42
SH43
SH44
Low High Low High Low High Low High Low High
3
8
9
25
26
78
79
240
3
8
9
25
26
77
78
237
3
7
8
23
24
70
71
216
3
8
9
25
26
76
77
234
3
8
9
23
24
71
72
218
3
5
6
16
17
50
51
150
3
5
6
16
17
50
51
154
3
5
6
16
17
50
51
154
3
6
7
20
21
60
61
184
3
8
9
25
26
77
78
235
3
6
7
19
20
57
58 174 175
533
3
4
5
12
13
37
38 111 112
340
3
7
8
20
21
62
63
189
3
5
6
14
15
44
45
134
3
6
7
19
20
57
58
176
3
6
7
17
18
52
53
160
3
4
5
13
14
40
41
122
3
6
7
20
21
60
61
181
3
6
7
20
21
63
64
190
3
9
10
28
29
86
87 260 261
800
3
9
10
28
29
90
91 275 276
813
3
9
10
30
31
91
92 271 272
850
3
10
11
30
31
94
95 280 281
880
4
13
14
38
39 120 121 360 361 1102
5
14
15
42
43 128 129 390 391 1200
3
10
11
30
31
92
93 280 281
860
3
10
11
30
31
91
92 270 271
850
3
10
11
30
31
90
91 273 274
850
3
9
10
29
30
90
91 270 271
850
3
10
11
30
31
92
93 280 281
860
4
11
12
33
34 100 101 301 302
950
4
12
13
37
38 113 114 345 346 1060
5
14
15
41
42 126 127 400 401 1200
3
10
11
30
31
92
93 280 281
860
4
12
13
38
39 116 117 353 354 1084
5
13
14
41
42 125 126 380 381 1170
3
6
7
20
21
62
63 190 191
580
5
13
14
41
42 123 124 380 381 1150
3
10
11
30
31
93
94 280 281
870
3
9
10
28
29
86
87 262 263
804
3
10
11
30
31
93
94 280 281
870
4
13
14
40
41 120 121 370 371 1100
3
6
7
19
20
58
59 175 176
540
4
13
14
41
42 125 126 380 381 1150
3
10
11
30
31
93
94 280 281
870
4
12
13
38
39 114 115 350 351 1070
5
14
15
42
43 130 131 390 391 1200
3
8
9
26
27
81
82 241 242
760
5
14
15
41
42 125 126 400 401 1200
4
12
13
38
39 115 116 350 351 1100
4
13
14
40
41 122 123 380 381 1140
5
14
15
42
43 130 131 390 391 1200
3
8
9
25
26
76
77 230 231
710
3
10
11
30
31
93
94 280 281
870
3
10
11
30
31
93
94 280 281
860
4
13
14
39
40 120 121 370 371 1120

SH45
Low High
241
720
238
715
217
653
235
706
219
658
151
460
155
465
155
465
185
556
236
711
534 1612
341 1050
190
572
135
406
177
533
161
500
123
369
182
560
191
580
801 2400
814 2500
851 2510
881 2600
1103 3331
1201 3600
861 2600
851 2510
851 2531
851 2500
861 2600
951 2800
1061 3203
1201 3600
861 2600
1085 3278
1171 3600
581 1730
1151 3500
871 2600
805 2500
871 2600
1101 3400
541 1610
1151 3500
871 2600
1071 3300
1201 3700
761 2300
1201 3600
1101 3300
1141 3500
1201 3600
711 2120
871 2600
861 2600
1121 3400

SH46
Low High
721 1910
716 1900
654 1800
707 1910
659 1750
461 1200
466 1200
466 1170
557 1470
712 1900
1613 4300
1051 2800
573 1530
407 1050
534 1400
501 1270
370 1000
561 1410
581 1530
2401 7200
2501 7600
2511 7100
2601 7200
3332 10000
3601 9800
2601 7400
2511 7900
2532 7800
2501 7100
2601 7400
2801 8100
3204 9700
3601 11000
2601 7400
3279 8900
3601 9500
1731 4800
3501 10500
2601 7300
2501 7000
2601 7300
3401 10000
1611 4500
3501 10500
2601 7200
3301 8700
3701 9800
2301 7000
3601 11000
3301 10000
3501 10500
3601 9700
2121 6500
2601 7300
2601 6900
3401 9200

SH47
Low
High
1911 4000
1901 4000
1801 3800
1911 4000
1751 3700
1201 2500
1201 2500
1171 2430
1471 3050
1901 4000
4301 9000
2801 5900
1531 3200
1051 2200
1401 2900
1271 2640
1001 2100
1411 3000
1531 3200
7201 15000
7601 15500
7101 14500
7201 15000
10001 21000
9801 20000
7401 15100
7901 16100
7801 16000
7101 14500
7401 15100
8101 17000
9701 20000
11001 23000
7401 15100
8901 18200
9501 19400
4801 10000
10501 22000
7301 15000
7001 14100
7301 15000
10001 21000
4501 9400
10501 22000
7201 15000
8701 18000
9801 20000
7001 14200
11001 23000
10001 21000
10501 22000
9701 20000
6501 13300
7301 15000
6901 14100
9201 19000

SH45
Low High
1151 3500
1001 3000
1159 3502
851 2530
861 2600
851 2530
501 1500
761 2300
851 2520
861 2600
966 2918
851 2551
1201 3500
401 1200
651 2000
661 2000
401 1200
751 2240
1201 3700
1101 3400
861 2600
861 2600
1201 3600
861 2600
461 1400
791 2400
701 2100
881 2600
881 2600
851 2510
881 2600
881 2610
1121 3400
563 1700

SH46
Low High
3501 10700
3001 8100
3503 10700
2531 8000
2601 8000
2531 8000
1501 4100
2301 6400
2521 7200
2601 7400
2919 9000
2552 7800
3501 10700
1201 3400
2001 5400
2001 5900
1201 3100
2241 6600
3701 11000
3401 9000
2601 7200
2601 7200
3601 9700
2601 7200
1401 3600
2401 6500
2101 5900
2601 7400
2601 7400
2511 7000
2601 7300
2611 7700
3401 9400
1701 4500

SH47
Low
High
10701 22100
8101 17000
10701 22200
8001 16200
8001 16300
8001 16200
4101 8600
6401 13100
7201 15000
7401 15100
9001 18400
7801 16000
10701 22200
3401 7100
5401 11000
5901 12000
3101 6600
6601 13500
11001 23000
9001 18400
7201 15000
7201 15000
9701 20000
7201 15000
3601 7500
6501 13300
5901 12000
7401 15000
7401 15000
7001 14400
7301 15000
7701 16000
9401 19200
4501 9400

SH48
Low
High
4001 8100
4001 8000
3801 7600
4001 8100
3701 7400
2501 5000
2501 5100
2431 4900
3051 6110
4001 8000
9001 18000
5901 11600
2201

4500

2641
2101
3001
3201
15001
15501
14501
15001
21001
20001
15101
16101
16001
14501
15101
17001
20001
23001
15101

5270
4200
6000
6500
30000
31000
29000
30000
42000
40000
31000
33000
32000
29000
30100
34000
40000
46000
30100

19401
10001
22001
15001
14101
15001
21001
9401
22001
15001
18001
20001
14201
23001
21001
22001
20001
13301
15001
14101
19001

39000
20000
44000
30000
28100
30000
43000
18400
44000
30000
36000
40000
29000
46000
42000
44000
40000
27000
30000
28100
38000

SH49
Low
High
8001
7601
8101

8200
8200
8200

SH50
Low High
8201 13400
8201 13200
8201 14100

Table A-1C GF100 Series Gas Table - Codes 878-5022, Bins SH40 to SH50
Gas
Code
878
881
897
898
910
916
917
925
929
930
939
946
950
953
958
962
965
978
979
980
982
983
984
985
986
987
988
989
990
991
992
993
994
5022

A-2

Min inlet pressure for vac. exhaust


(PSIA)
Gas Symbol
Gas Name
SH40-SH47 SH48 SH49 SH50
3%C2H4/He
3%Ethylene/Helium
19.7
19.7
10%B2H6/Ar
5%Diborane/Argon
24.7
29.7
2.7%C2H4/He
2.7%Ethylene/Helium
19.7
19.7
1%GeH4/H2
1%Germane/Hydrogen
20.7
20.7
.5%GeH4/H2
0.5%Germane/Hydrogen
20.7
20.7
2%PH3/H2
2%Phosphine/Hydrogen
21.7
21.7
50%C3H6-b)/N2 50%Propylene/Nitrogen
19.7
21.7
10%C3H6-b)/N2 10%Propylene/Nitrogen
19.7
21.7
10%CH4/N2
10%Methane/Nitrogen
19.7
21.7
3.9%H2/N2
3.9%Hydrogen/Nitrogen
24.7
29.7
10%B2H6/He
10%Diborane/Helium
19.7
19.7
30%C2H4/He
30%Ethylene/Helium
19.7
24.7
10%H2/He
10%Hydrogen/Helium
19.7
19.7
15%H2/B2H6
15%Hydrogen/Diborane
19.7
19.7
17%CH4/CO2
17%Methane/Carbon Dioxide
24.7
29.7
20%CH6Si/H2
20%Methylsilane (MONO)/Hydrog
14.7
19.7
50%CH3SiHCl2/H250%Dichloromethylsilane/Hydroge
11.7
13.7
20%GeH4/H2
20%Germane/Hydrogen
19.7
24.7
0.5%B2H6/He
0.5%Diborane/Helium
14.7
19.7
20%F2/Ar
20%Fluorine/Argon
24.7
29.7
5%HF/N2
5%Hydrogen Fluoride/Nitrogen
24.7
29.7
3%H2S/CO
3%Hydrogen Sulfide/Carbon Mon
24.7
29.7
1.4%C2H2/Ar
1.4%Acetylene/Argon
24.7
29.7
2%Ge2H6/H2
2%Digermane/Hydrogen
24.7
29.7
38%SiCl4/O2
38%Silicon Tetrachloride/Oxygen
24.7
29.7
29.7
29.7
20%H2S/20%CO2/ 20%Hydrogen Sulfide/20%Carbon
24.7
29.7
20%SiH2Cl2/H2
20%Dichlorosilane/Hydrogen
24.7
29.7
0.1%Cl2/N2
0.1%Chlorine/Nitrogen
24.7
29.7
1%HCl/N2
1%Hydrogen Chloride/Nitrogen
24.7
29.7
3%BCl3/N2
3%Boron Trichloride/Nitrogen
24.7
29.7
18%NO/N2
18%Nitric Oxide/Nitrogen
24.7
29.7
9%NO/41%N2/H2 9%Nitric Oxide/41%Nitrogen/H2
24.7
29.7
20%H2/Ar
20%Hydrogen/Argon
24.7
29.7
COS-Special
Carbonyl Sulfide (Special)
24.7
29.7

SH40
SH41
SH42
SH43
SH44
Low High Low High Low High Low High Low High
4
13
14
40
41 125 126 375 376 1150
4
11
12
35
36 105 106 320 321 1000
4
13
14
40
41 125 126 377 378 1158
3
10
11
30
31
91
92 275 276
850
3
10
11
30
31
92
93 280 281
860
3
10
11
30
31
91
92 275 276
850
3
5
6
17
18
53
54 160 161
500
3
8
9
26
27
82
83 243 244
760
3
10
11
29
30
91
92 272 273
850
3
10
11
30
31
92
93 280 281
860
4
11
12
34
35 103 104 314 315
965
3
10
11
30
31
90
91 275 276
850
4
13
14
40
41 125 126 380 381 1200
3
4
5
14
15
42
43 130 131
400
3
7
8
23
24
70
71 210 211
650
3
7
8
23
24
71
72 212 213
660
3
4
5
14
15
42
43 130 131
400
3
8
9
26
27
81
82 241 242
750
5
14
15
42
43 130 131 400 401 1200
4
13
14
39
40 120 121 360 361 1100
4
10
11
30
31
92
93 280 281
860
3
10
11
30
31
92
93 280 281
860
5
14
15
42
43 130 131 390 391 1200
3
10
11
30
31
92
93 280 281
860
3
5
6
16
17
50
51 150 151
460
3
9
10
27
28
85
86 252 253
790
3
8
9
24
25
74
75 222 223
700
4
10
11
30
31
94
95 280 281
880
4
10
11
30
31
94
95 280 281
880
3
9
10
29
30
91
92 270 271
850
3
10
11
30
31
94
95 280 281
880
4
10
11
30
31
94
95 281 282
880
4
13
14
40
41 120 121 370 371 1120
3
7
8
20
21
60
61 183 184
562

For reference only, consult factory on the latest tables.

SH48
SH49
SH50
Low
High
Low High
Low
High
22101 45000
17001 34000
22201 45000
16201 33000
16301 33000
16201 33000
8601 17000
13101 26100
15001 30000
15101 30100
18401 37000
16001 32000
22201 45000
7101 14000
11001 22000
12001 24000
6601 13000
13501 27000
23001 47000
18401 37000
15001 30000
15001 30000
20001 40000
15001 30000
7501 15000 15001 15001 15002 25000
13301 27000
12001 24000
15001 30000
15001 30000
14401 29000
15001 30000
16001 32000
19201 39000
9401 18300

(990-100-0082-R17)

Appendix B: GF100 Series Patents

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

GF100 Series

B-1 GF100 Series Patents


The GF100 Series may be protected by the following US patents and their
international filings.
Table B-1 GF100 Series Patents
Patent/Pub. No.
6343617
6389364
6425281
6445980
6539792
6640822
6681787
6714878
6752166
6826953
6845659
6910381
6941965
6962164
7043374
7073392
7082824
7113895
7114511
7133785
7143774
7150201
7216019
7231931
7243035
7272512
7273063
7287434
7360551
7363182
7380564
7409871
7412986
7424894
7434477

Title
System and method of operation of a digital mass flow controller
System and method for a digital mass flow controller
Pressure insensitive gas control system
System and method for a variable gain proportional-integral (PI) controller
Method and apparatus for balancing resistance
System and method of operation of a digital mass flow controller
System and method of operation of a digital mass flow controller
System and method for a digital mass flow controller
Method and apparatus for providing a determined ratio of process fluids
Flow sensor
Variable resistance sensor with common reference leg
System and method of operation of an embedded system
for a digital capacitance diaphragm gauge
Method and apparatus for providing a determined ratio of process fluids
System and method for a mass flow controller
Flow sensor signal conversion
Methods and apparatus for pressure compensation in a mass flow controller
Variable resistance sensor with common reference leg
System and method for filtering output in mass flow controllers and mass flow meters
System and method for a mass flow controller
Valve control system and method
Method and apparatus for providing a determined ratio of process fluids
System and method for measuring flow
Method and system for a mass flow controller with reduced pressure sensitivity
System and method for a mass flow controller
System and method for mass flow detection device calibration
Flow sensor signal conversion
Methods and apparatus for pressure compensation in a mass flow controller
System and method for measuring flow
Method and apparatus for providing a determined ratio of process fluids
System and method for mass flow detection device calibration
System and method for a mass flow controller
Mass flow meter or controller with inclination sensor
Method and system for flow measurement and validation of a mass flow controller
Method and apparatus for providing a determined ratio of process fluids
Methods and apparatus for pressure compensation in a mass flow controller

B-1

Appendix B: GF100 Series Patents

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

GF100 Series

THIS PAGE WAS


INTENTIONALLY
LEFT BLANK

B-2

Section C Essential Instructions

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

Models MT3809G & MT3810G

Bulgarian


!
Brooks Instrument , ,
. , ,
.
, Brooks Instrument.
x
, , , ,
.
x
, .
, .
- .
: , .
/ .
x
, Brooks Instrument .
x
, .
x

. .
x
: (1) . . (2)
. ,
.
x
.
, Brooks Instrument .
, .
, .
x
, ,
, .
: ,
, .
, .

(PED)
0,5 bar (g) - 25 mm 1" (),
(PED).
x
PED .
x
97/23/.
x
Brooks Instrument 1.
x
- 25 mm 1" () I, II III PED.

25 mm 1" (), -, (SEP).

(EMC)
(/) Brooks Instrument
( EMC 2004/108/EC).
, , .
, :
Brooks Insturment , CE .
, , 100%- .
D .
.
360.
.
. 100%- ,
CE .
.
: .

(ESD)
: ,
. ,
.
:
1. .
2. , ,
.
3. .
. ,
, .
:
, (ESD) .
, - (NMOS, SMOS ..). ,
.
, .

A-1

Section C Essential Instructions

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

GF100 Series
Czech

Zkladn instrukce
Ped instalac si pette nsledujc instrukce!
Spolenost Brooks Instrument konstruuje, vyrb a testuje tento produkt tak, aby splnil mnoho nrodnch a mezinrodnch standard. Pstroje mus bt
dn nainstalovny, pouvny a udrovny tak, aby byl zajitn jejich nepetrit provoz v rmci normlnch technickch specifikac. Muste dodrovat
nsledujc pokyny a integrovat jejich obsah do svho bezpenostnho programu pi instalovn, pouvn a udrovn produkt spolenosti Brooks.
x
Pro zajitn sprvn funkce zazen mohou jeho instalaci, obsluhu, programovn, drbu a aktualizace firmwaru provdt vhradn
kvalifikovan osoby.
x
Ped instalac, provozem a drbou produktu si prostudujte vechny pokyny. Pokud tato pruka nen tou sprvnou prukou pro dan
zazen, informujte se na zadn stran oblky o kontaktu na mstn prodejn kancel. Uchovejte si tuto pruku pro pozdj potebu.
UPOZORNN: Neprovozujte zazen v rozsahu mimo dan rozsah v provozn pruce. Poruen tohoto upozornn me mt za nsledek
vn jmy na zdrav a vst k pokozen zazen.
x
x
x
x
x
x

Pokud nkterm pokynm nerozumte, kontaktujte svho prodejnho zstupce spolenosti Brooks a vydejte si objasnn.
Dodrujte vechny vstrahy, upozornn a pokyny, uveden a vyznaen na produktu, nebo s nm dodan.
Namontujte zazen specifikovanm zpsobem podle sprvn montn pruky a podle platnch mstnch a nrodnch pedpis. Pipojte
vechny produkty ke sprvnm zdrojm elektrick energie a stlaench mdi.
Postup: (1) Pomalu do systmu pivdjte mdium. Pro zabrnn vzniku rz v systmu otvrejte procesn ventily postupn. (2) Pekontrolujte
tsnost vstupnho a vstupnho pipojen prtokomru. Pokud nezjistte dn netsnosti, postupn zvedejte tlak na provozn hodnotu.
Ped provdnm servisnch prac zkontrolujte, zda systm nen pod tlakem. V ppad poteby vmny dl zajistte, aby byly pouity nhradn
sousti specifikovan spolenost Brooks Instrument a vmnu provdla kvalifikovan osoba. Pouit neschvlench dl a postup me negativn
ovlivnit efektivitu a bezpenost procesu. Pouit nhrad za originln dly me zpsobit por, raz elektrickm proudem nebo nesprvnou funkci.
Pokud nen zrovna provdna drba kvalifikovanou osobou, ujistte se, e zazen je opateno vemi pedepsanmi kryty.

UPOZORNN: Pokud je u zazen s prtokem kapalin nutno z jakhokoli dvodu uzavt vstupn a vstupn ventily, je nutn zazen
kompletn vyprzdnit. Pokud tak neuinte, me z dvodu teplotn roztanosti zbytk mdia v zazen dojt k jeho pokozen nebo k ohroen
zdrav osob.

Evropsk smrnice pro tlakov zazen (PED)


Na veker tlakov zazen s vnitnm tlakem vym ne 0,5 baru (g) a velikosti vt ne 25 mm nebo 1" (palec) se vztahuje platnost smrnice o
tlakovm zazen (PED).
Kapitola Technick daje v tto pruce obsahuje dleit bezpenostn a provozn pokyny souvisejc se smrnic PED.
x
Mic pstroje popsan v tto pruce jsou v souladu se smrnic EN 97/23/ES.
x
Vechny prtokomry spolenosti Brooks Instrument spadaj do rmce Kapaliny, skupina I.
x
Mic pstroje vt ne 25 mm nebo 1" (palec) jsou v souladu se smrnic PED, kategorii I, II nebo III
x
Mic pstroje s velikost 25 mm nebo 1" (palec) a men spadaj do rmce Sprvnch technickch postup (SEP).

Evropsk smrnice pro elektromagnetickou kompatibilitu (EMC)


Elektrick/elektronick zazen Brooks Instrument nesouc znaku CE bylo spn testovno dle pedpis pro elektromagnetickou kompatibilitu
(smrnice EMC . 2004/108/EC).
Vbru signlnch kabel pro pouit se zazenm oznaenm CE je nutn vnovat zvltn pozornost.
Kvalita signlnch kabel, kabelovch prchodek a konektor:
Brooks Instrument dodv vysoce jakostn kabely splujc poadavky kvalitativnho zaazen CE. Pokud chcete pout vlastn signln kabely, zvolte
typy s kvalitnm stnnm vech il a v cel dlce trasy.
V ppad pouit konektor kruhovch nebo tvaru D, mus mt tyto kovov stnn. V ppad jejich pouit, mus kovov kabelov prchodky bt
propojeny se stnnm kabelu.
Stnn by mlo bt pipojen ke kovovmu tlesu pstroje nebo krytu, na obou koncch kabelu a po celm jeho obvodu.
Stnn by mlo bt uzemnno.
Ppojky vedouc ke kartm podle norem jsou nekovov. Pro splnn poadavk pedpis CE mus bt pouit kabely kompletn stnny.
Stnn by mlo bt uzemnno.
Konfigurace kontakt je uvedena v piloenm nvodu k obsluze.

Elektrostatick vboj (ESD)


UPOZORNN: Tento pstroj obsahuje komponenty citliv na pokozen statickou elektinou. Pi monti, demonti i jin manipulaci
s vnitnmi elektronickmi obvody je poteba dodrovat pslun postupy.
Postup oetovn:
1. Odpojte pstroj od napjen.
2. Osoba provdjc drbu m bt pi instalaci, vyjmn i prci na desce plonch spoj nebo jin vnitn elektronice uzemnna zemncm nramkem,
nebo jinm vyhovujcm zpsobem.
3. Desky plonch spoj je nutn pepravovat v elektricky vodivm obalu. Plon spoje vyjmejte z vodivho obalu a bezprostedn ped instalac do
pstroje! Plon spoje vyjmut z pstroje a uren pro dal vyuit, opt neprodlen umstte do ochrannho obalu.
Poznmky:
Existence prvk, citliv na elektrostatick vboje (ESD) v pstrojch je astm jevem. Prvky s technologi oxidu kov (NMOS,SMOS) jsou pouvny u
vtiny modernch elektronickch zazen. Zkuenosti dokazuj, e i jen mal elektrostatick vboje mohou pokodit nebo zniit tyto zazen. U
pokozench soustek, jakkoli zdnliv pracujcch bezchybn, dochz brzy k porue.

A-2

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

Section C Essential Instructions

Models MT3809G & MT3810G

Dansk

Grundlggende vejledninger
Ls disse fr anvendelse!
Brooks Instruments designer, fremstiller og afprver sine produkter sledes, at de tilpasser sig bde de indenrigs og internationale standarder. Disse
udstyr br installeres, bruges og repareres omhyggeligt, s at de kan virke tilsvarende deres normale anvendelsesperiode. De flgende regler skal
overholdes og implementeres under installeringen samt ved brug og reparation.
x For at garantere den passende kapacitet, er udstyrets installation, anvendelse, opdatering, programmering kun tilladt for kvalificeret personale.
Alle vejledninger skal lses fr produktets installation, anvendelse og reparation.
x Hvis denne manual ikke er den passende udgave, kontakt venligst jeres leverandr for yderligere information.
x Det anbefales at gemme denne manual for senere brug.
OBS.: Udstyret m ikke anvendes til andet end det er angivet i brugsanvisningen. Hvis denne regel brydes, kan der forekomme alvorlige
personskader eller brist p udstyret.
Hvis vejledningerne ikke er forstelig, kontakt venligst Deres Brooks reprsentant for at afklare problemet som er opstet.
x
Overhold alle regler, som er markeret eller leveret sammen med udstyret.
x
Installer udstyret efter den angivne installationsvejledning og gldende lovgivning for anvendelsesomrde. Udstyret m kun tilsluttes med kabler
og stik som overholder kravspecifikationerne i vejledningen.
x
Ibrugtagning: (1) bn langsomt for trykket i systemet.. bn langsomt for alle procesventiler for at forhindre ustabil gas flow. (2) Tjek systemet for
lkage ved tilsluttet mleinstrumenter, samlinger og andet tilsluttet udstyr. Derefter g trykket i systemet indtil arbejdstrykket er opnet.
x
Fr reparation tjek altid at procesledningen ikke str under tryk. Hvis der er brug for at udskifte defekte dele m kun kvalificeret personale udfre
arbejdet og af sikkerhedsmssige rsager m der kun anvendes originale Brooks reservedele. Det er ikke tilladt at anvende reservedele eller
udfre arbejde der p nogen mde kan ndre produktet fra dens oprindelige specifikationer. Manglende overholdelse af de foreskrevne
procedurer kan resultere i brnd, og fare for elektrisk std eller kortslutning.
x
Af sikkerhedsmssige rsager srg for at alle sikkerhedsforanstaltninger er overholdt. Eksempelvis at alle afskrmninger eller anden form for
installationsbeskyttelse er lukket eller installeret ved normal drift.
Advarsel.: Ved brug af udstyr som anvendes til vske skal det sikres at indgangsventilen og udgangsventilen ikke bliver lukket p
samme tid i forbindelse med aftapning. Sfremt dette ikke overholdes, er der risiko for at vsken p grund af varmeudvikling ekspanderer
og dette kan forrsage skade p udstyr og personer.

Det Europiske direktiv for trykudstyr (PED)


P alt udstyr hvis indgangstryk er strre end 0,5bar (g) og strre end 25 mm eller en tomme, glder det europiske direktiv for trykapparater.
Manualens afsnit tekniske data indeholder anvisninger om PED direktivet.
x
I manualen angives mleinstrumenter der er tilpasset direktivet 97/23/EK EU.
x
Alle Brooks gennemstrmningsmlere tilhrer vskegruppen nr. 1.
x
Alle mlinstrumenter som er strre end 25 mm eller en tomme beskrives i direktivet PEDs kategorier I, II eller III.
x
Alle mlinstrumenter som er p 25 mm eller en tomme, eller mindre, beskrives i Sound Engineering Practice (SEP).

Det europiske direktiv for elektromagnetisk kompatibilitet (EMC)


Alle Brooks instrumenter (elektrisk/elektronisk) som har CE markering er godkendt og testet iflge om elektromagnetiske kompatibilitets forskrifter
(EMC direktiv 2004/108/EC).
OBS: Man skal vre opmrksom p hvilken type kabler der bruges til CE mrket udstyr..
Om kvalitet af signalkabler, kabeltilslutninger og koblinger:
Brooks tilbyder kabler af hjest kvalitet, som er tilpasset CEE kvalificeringens forskrifter. Hvis man vlger at bruge egne kabler, skal man vlge et
kabel som har den ndvendige afskrmning for at sikre 100 % mod udefra kommende stj.
Tilslutningerne D eller rundformede tilslutninger skal vre afskrmet med skal af metal.
Stikket skal vre afskrmet p alle sider. Al afskrmning skal jordes..
Card Edge tilslutninger er iflge standarden ikke metalliske.. De anvendte kabler og stik skal vre 100 % afskrmet for at opfylde CE kravene.. De
skal ligeledes jordes.
For stik konfigurationen se vedlagte brugsvejledning.

Elektrostatisk afladning (ESD)


OBS.: Udstyret indeholder tilbehr som kan skades ved elektrostatisk elektricitet. Alle forskrifter skal overholdes ved kontakt med alle
elektriske komponenter bde under drift og vedligeholdelse..
Behandlingsproceduren:
1. Sluk for al strmtilfrsel til udstyret.
2. Personer som skal i kontakt med udstyret skal vre jordet eller bre anden form for elektrisk beskyttende udstyr. Manglende overholdelse af
dette kan medfre skader p alle elektriske komponenter.
3. Alle elektriske komponenter skal opbevares eller transporteres i deres originale indpakning for at sikre komponenter mod statiske elektriske
skader. Emballagen m ikke bnes fr komponenten skal installeres i udstyret. Ved afslutning af vedligeholdelse/reparation af udstyret, skal
udstyret installeres med det samme eller pakkes forsvarligt hvis det skal p lager eller transporteres.
Bemrkninger:
Dette udstyr er ikke unik i den hensigt, at det indeholder for elektrostatisk afladning (ESD) sensitive reservedel. I de fleste elektronisk udstyr findes
der metaloxyd teknologiske reservedel (NMOS, SMOS m.m.). Erfaringerne viser at selv den mindste statiske elektricitet kan skade, eller delgge
disse instrumenter. Selv en fungerende elektrisk del kan have levetiden markant reduceret p grund af statisk elektricitet..

A-3

Section C Essential Instructions

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

GF100 Series
Dutch

Essentile instructies
Leeszevoordatuverdergaat!
Brooks Instrument ontwerpt, produceert en test haar producten zodanig dat ze voldoen aan vele nationale en internationale normen. Deze producten moeten
correct worden genstalleerd, bediend en onderhouden zodat ze binnen hun normale specificaties blijven werken. De volgende instructies moeten worden
toegevoegd aan en gentegreerd in uw veiligheidsprogramma als u producten van Brooks Instrument installeert, bedient en onderhoudt.

Om de juiste prestaties te kunnen garanderen mag alleen gekwalificeerd personeel het product installeren, bedienen, updaten, programmeren en
onderhouden.

Lees alle instructies voordat u het product gaat installeren, bedienen en onderhouden. Als dit niet de juiste handleiding is, kijk dan op de achterzijde voor
contactinformatie van uw vertegenwoordiger. Bewaar deze handleiding voor later.

WAARSCHUWING: gebruik dit instrument niet als niet is voldaan aan de specificaties in de handleiding. Het niet naleven van deze waarschuwing kan
ernstig letsel en/of schade aan de apparatuur tot gevolg hebben.

Als u n of meer instructies niet begrijpt, vraag dan om uitleg aan uw vertegenwoordiger van Brooks Instrument.

Bediening: (1) Laat het systeem langzaam volstromen. Open de procesafsluiters langzaam om drukstoten te voorkomen. (2) Controleer op lekkages rondom
de inlaat- en uitlaataansluitingen van de stromingsmeter. Als er geen lekkages zijn, kan het systeem op de bedrijfsdruk worden gebracht.

Zorg ervoor dat de procesleiding drukvrij is gemaakt voordat u servicewerkzaamheden gaat uitvoeren. Als vervangingsonderdelen nodig zijn, zorg er dan
voor dat gekwalificeerd personeel de door Brooks Instrument gespecificeerde vervangingsonderdelen gebruikt. Niet goedgekeurde onderdelen en
procedures kunnen de prestaties van het product en de veilige werking van uw proces in gevaar brengen. Niet goedgekeurde vervangingsonderdelen
kunnen brand, elektrische schokken of een onjuiste werking tot gevolg hebben.

Zorg ervoor dat alle deksels van de apparatuur gesloten zijn en de afdekkingen gemonteerd zijn om elektrische schokken en lichamelijk letsel te voorkomen,
behalve als gekwalificeerd personeel de onderhoudswerkzaamheden uitvoert.

Neem alle waarschuwingen, voorschriften en instructies in acht die op het product zijn aangebracht of bij het product zijn geleverd.
Installeer uw apparatuur volgens de instructies in de bijgeleverde handleiding en in overeenstemming met de geldende lokale en nationale voorschriften.
Sluit alle producten aan op de juiste elektrische voedings- en drukbronnen.

WAARSCHUWING: bij vloeistofstroomapparaten waarvan de inlaat- en uitlaatkleppen om welke reden dan ook gesloten zijn, moet de vloeistof
volledig worden afgetapt. Als dat wordt nagelaten, kan dit leiden tot thermische expansie van de vloeistof waardoor het apparaat kan barsten en
lichamelijk letsel kan veroorzaken.

PED-richtlijn (Pressure Equipment Directive)


Alle drukapparatuur met een interne druk van meer dan 0,5 barg en een diameter van meer dan 25 mm valt onder de PED-richtlijn.
x In het hoofdstuk Specificaties van deze handleiding staan aanwijzingen die verband houden met de PED-richtlijn.
x De meters die in deze handleiding worden beschreven, voldoen aan de Europese richtlijn 97/23/EG.
x Alle stromingsmeters van Brooks Instrument vallen in groep 1.
x Meters met een diameter van meer dan 25 mm voldoen aan de categorien I, II of III van de PED-richtlijn.
x Meters met een diameter van 25 mm of kleiner voldoen aan de regels van goed vakmanschap.

Elektromagnetische compatibiliteit (EMC)


De elektronische apparatuur van Brooks Instrument met de CE-markering is succesvol getest in overeenstemming met de EMC-voorschriften (richtlijn EMC
2004/108/EC).
De keuze van de signaalkabel voor gebruik in combinatie met apparatuur met CE-markering verdient speciale aandacht.
Kwaliteit van de signaalkabel, kabelafdichtingen en stekkers:
Brooks Instrument levert hoogwaardige kabels die voldoen aan de specificaties voor de CE-markering.
Als u zelf voor signaalkabel zorgt, moet u altijd een volledig afgeschermde kabel gebruiken.
Stekkers van het type D of ronde stekkers moeten zijn voorzien van een metalen afscherming. Indien nodig moeten metalen kabelafdichtingen worden
gebruikt waarvan de afscherming voor het klemmen van de kabel kan worden gebruikt.
Het kabelscherm moet met het metalen omhulsel of de metalen afdichting worden verbonden en aan beide uiteinden rondom volledig worden afgeschermd.
De afscherming moet aan de aardpotentiaal worden aangesloten.
Card Edge Connectors zijn standaard niet van metaal. De gebruikte kabels moeten volledig zijn afgeschermd om te voldoen aan de CE-markering.
De afscherming moet aan de aardpotentiaal worden aangesloten.
Voor de pinconfiguratie: Raadpleeg de bijgevoegde handleiding.

Elektrostatische ontlading
VOORZICHTIG: Dit instrument bevat elektronische componenten die gevoelig zijn voor statische elektriciteit. Neem de juiste procedures in acht bij
het verwijderen en installeren of bij andere werkzaamheden aan de interne printplaten of apparaten.
Procedure:
1. Schakel de voeding van de eenheid uit.
2. Het personeel moet zich met een polsbandje of ander veilig en geschikt hulpmiddel aarden voordat een printplaat of ander intern apparaat mag worden
genstalleerd, verwijderd of aangepast.
3. Printplaten moeten in een geleidende verpakking worden vervoerd. De platen mogen pas vlak voor de eigenlijke installatie uit de beschermende verpakking
worden gehaald. Verwijderde printplaten moeten onmiddellijk in de beschermende verpakking worden geplaatst om te worden getransporteerd, opgeslagen
of teruggestuurd naar de fabriek.
Opmerkingen
Dit instrument is niet uniek als het gaat om componenten die gevoelig zijn voor elektrostatische ontlading. De meeste moderne elektronische apparaten
bevatten componenten die gebruik maken van de metaaloxidetechnologie (NMOS, SMOS, enz.). Uit ervaring blijkt dat zelfs kleine hoeveelheden statische
elektriciteit deze apparaten al dan niet onherstelbaar kunnen beschadigen. Beschadigde componenten, zelfs als ze goed lijken te functioneren, raken eerder
defect.

A-4

Section C Essential Instructions

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

Models MT3809G & MT3810G

Estonian

Olulised juhised
Enne kasutamist lugege hoolikalt lbi!
Brooks Instrument konstrueerib, valmistab ja katsetab oma tooteid selliselt, et need vastaksid paljude erinevate riiklike ja rahvusvaheliste standardite nuetele.
Ainult nuetekohane paigaldamine, kasutamine ja hooldamine tagab toodete katkematu talitluse tavaspetsifikatsiooni raames. Brooks Instrument'i toodete
paigaldamisel, kasutamisel ja hooldamisel tuleb tita alljrgnevaid juhiseid ja integreerida need asjakohasesse ohutusprogrammi.
x
Nuetekohase talitluse tagamiseks tohib toodet paigaldada, kasutada, tiustada, programmeerida ja hooldada ainult kvalifitseeritud personal.
x
Enne toote paigaldamist, kasutamist ja hooldamist lugege kik kasutusjuhised hoolikalt lbi. Kui kesolev kasutusjuhend ei vasta teie tootele, prduge
kohaliku edasimja poole, kelle kontaktandmed leiate kasutusjuhendi tagakaanelt. Hoidke see kasutusjuhend edaspidiseks alles.
HOIATUS: rge kasutage seda instrumenti vljaspool kasutusjuhendis spetsifitseeritud piirvrtusi. Selle hoiatuse eiramine vib phjustada
tsiseid kehavigastusi ja/vi kahjustada seadet.
x
Kui te saa mne juhise mttest aru, prduge selgituste saamiseks kohaliku Brooks Instrumenti edasimja poole.
x
Jrgige kiki hoiatusi, thelepanule manitsusi ja juhiseid, mis on tootele peale kantud vi tootega kaasa antud.
x
Seadme paigaldamisel jrgige vastavas kasutusjuhendis toodud paigaldusjuhiseid ning asjakohaseid kohalikke ja riiklikke eeskirju. hendage tooted
nuetekohaste toite- ja surveallikatega.
x
Talitlus: (1) Avage aeglaselt vool ssteemi. Vooluimpulsside vltimiseks avage tventiilid aeglaselt. (2) Kontrollige, et voolukulumturi sisend- vi
vljundhenduste mber ei ole lekkeid. Kui lekkeid ei ole, laske ssteemil saavutada tsurve.
x
Enne seadme hooldamist veenduge, et kogu ssteem on surve alt vabastatud. Varuosasid tohib vahetada ainult kvalifitseeritud personal, kasutades selleks
Brooks Instrumenti poolt heakskiidetud varuosi. Mitteoriginaalvaruosade kasutamine ja ebapdev toimingute tegemine vivad kahjustada toote tomadusi
ja phjustada riski tootmistegevuse ohutuse tagamisel. Originaalvaruosadele sarnaste osade kasutamine vib phjustada tule- vi elektrilgiohtu vi
seadme vrtalitlust.
x
Elektrilgi- ja vigastuseohu vltimiseks peavad seadme luugid olema alati suletud ja kaitsekatted oma kohal, va seadme hooldamisel kvalifitseeritud isikute
poolt.
HOIATUS: voolava vedelikuga seadmete kasutamisel kui seadmega klgnevad sisend- ja vljundklapid on vaja mingil phjusel sulgeda,
tuleb seadmed vedelikust tiesti thjaks lasta. Vastasel korral vib vedelik soojuse mjul paisuda niivrd, et seade puruneb. See vib phjustada
tsiseid kehavigastusi.

Euroopa surveseadmete direktiiv (PED)


Euroopa surveseadmete direktiiv kohaldub kikidele surveseadmetele, mille sisesurve on le 0,5 baari (g) ja lbimt le 25 mm vi 1 tolli.
x
Kesoleva kasutusjuhendi spetsifikatsiooniosa sisaldab surveseadmete direktiiviga seonduvaid juhiseid.
x
Kesolevas kasutusjuhendis kirjeldatud mturid vastavad EN direktiivi 97/23/E nuetele.
x
Brooks Instrumenti voolukulumturid kuuluvad vedelike 1. gruppi.
x
Mturid lbimduga le 25 mm vi 1 tolli vastavad surveseadmete direktiivi kategooriale I, II vi III.
x
Mturitele lbimduga alla 25 mm vi 1 tolli kohaldatakse hid inseneritavasid.

Euroopa elektromagnetilise hilduvuse direktiiv (EM)


Brooks Instrumenti (elektrilised/elektroonilised) seadmed, millele on omistatud CE-this, on edukalt lbinud asjakohased katsed ja vastavad elektromagnetilise
hilduvuse nuetele (EM direktiiv 2004/108/EC).
Kuid signaalkaabli valimisel on vaja prata suurt thelepanu CE-thisega seadmetele.
Signaalkaabli, lbiviigutihendite ja konnektorite kvaliteet
Brooks Instrument turustab krgekvaliteedilisi kaableid, mis vastavad CE-sertifikaadi nuetele.
Olemasoleva kaabli kasutamisel jlgige, et kaabel oleks tielikult mbritsetud varjestusega.
D vi Ring-tpi konnektorid peavad olema varustatud metallvarjestusega. Kus kohaldatav, tuleb kasutada metallist lbiviike, mis tagavad kaabli varjestuse
henduse.
Kaabli varjestus hendatakse metallkesta vi lbiviigutihendiga ja on mlemast otsast kaitstud 360 ulatuses.
Varjestus peab olema maandatud.
Mikroskeemide servahendused on ldjuhul mittemetallist. Vastavuse tagamiseks CE-sertifikaadi nuetele peavad kasutatud kaablid olema 100% varjestatud.
Varjestus peab olema maandatud.
Klemmide konfigureerimine: vt komplekti kuuluvat kasutusjuhendit.

Elektrostaatiline laeng
THELEPANU: seade sisaldab staatilise elektri suhtes tundlikke elektroonikakomponente. Seadmesse paigaldatud trkkplaatide
eemaldamisel ja paigaldamisel, samuti trkkplaadi vi seadmega muude toimingute teostamisel jrgige nuetekohase ksitsemise juhiseid.
Ksitsemisjuhised
1.
Lahutage seade toiteallikast.
2.
Enne trkkplaadi vi mne muu siseelemendi paigaldamist, eemaldamist vi konfigureerimist peab personal olema maandatud lbi randmepaela vi mne
muu sobiva vahendi.
3.
Trkkplaate transporditakse voolujuhtivas konteineris. Vtke trkkplaat kaitsvast konteinerist vlja vahetult enne selle paigaldamist. Seadmest eemaldatud
trkkplaadid tuleb viivitamatult asetada kaitsvasse konteinerisse kas siis edasiseks transportimiseks, hoiustamiseks vi tehasesse tagasisaatmiseks.
Kommentaarid
See seade ei ole ainus, mis sisaldab staatilise elektri suhtes tundlikke elemente. Enamik kaasaegsetest elektroonikaseadmetest sisaldavad komponente, mille
valmistamiseks on kasutatud metalloksiidtehnoloogiat (NMOS, SMOS jne). Kogemused nitavad, et isegi vike kogus staatilist elektrit vib neid seadmeid
kahjustada vi isegi hvitada. Kuigi vib nida, et kahjustatud komponendid ttavad nuetekohaselt, hakkavad talitlushired ilmnema juba varakult.

A-5

Section C Essential Instructions

GF100 Series

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

Finnish

Perusohjeet
Lue ensin ohjeet huolellisesti!
Brooks Instrument suunnittelee, valmistaa ja testaa laitteensa vastaamaan useimpien kotimaisten ja kansainvlisten standardien
vaatimuksia. Tuotteet tulee asentaa, kytt ja huoltaa kyttohjeiden mukaan jotta niiden toimivuus taataan. Brooks Instrumentin
laitteiden asennuksessa, kytss ja huollossa on noudatettava soveltuvia mryksi ja ohjeita, lisksi mainitut ohjeet on
huomioitava tysuojelun ohjeistuksessa.

x
x
x
x
x

Oikean toiminnan varmistamiseksi vain valtuutettu huoltohenkil saa asentaa, kynnist, pivitt, ohjelmoida ja huoltaa
laitteita.
Lue kaikki kyttohjeet koskien tuotteen asennusta, kytt ja huoltoa. Jos kyttohje on puutteellinen, listietoja saa paikalliselta
jlleenmyyjlt. Yhteystiedot lytyvt oppaan kansilehdest. Silyt ohjeet.
VAROITUS!: Kyttohjeessa ilmoitettujen standardien mukaisia ohjeita ja raja-arvoja ei saa ylitt. Rajoitusten
laiminlynti voi aiheuttaa tuotteen rikkoutumisen ja/tai vakavan henkilvahingon vaaran.
Jos ohjeissa on epselvyytt, ota yhteytt Brooks Instrumentin edustajaan ongelman selvittmiseksi.
Noudata kaikkia laitteessa olevia tai siihen liittyvi ohjeita, mryksia ja varoituksia.
Laittteen asennuksessa on noudatettava erityisi asennusohjeita sek voimassa olevia paikallisia ja kansainvlisi mryksi.
Laitteet saa yhdist vain soveltuvaan shk- ja paineverkkoon.
Asennusohjeita: (1) Pst virtaus hitaasti jrjestelmn. Avaa venttiilit hitaasti, jotta virtaus pysyy tasaisena. (2) Tarkista, ettei
virtausmittarin sisn- ja ulosmenon vieress ole vuotoa. Jos jrjestelmss ei ole vuotoa, aseta oikea kyttpaine.
Tarkista, ett laitteeseen menev paine on katkaistu ennen laitteen korjaamista vlttksesi killisen painepstn aiheuttaman
loukkaantumisriskin. Mahdollisten varaosien tulee olla Brooks Instrumentin hyvksymi. Vain valtuutettu huoltohenkil saa
asentaa varaosat. Ei-hyvksyttyjen varaosien kytt voi vahingoittaa tuotteen toimintaa ja aiheuttaa turvallisuusriskin. Samoin eihyvksyttyjen varaosien kytt voi aiheuttaa tulipalon, shkiskun tai virhetoiminnan riskin.
Varmista ett kaikki kaikki laitteen ovet/luukut ovat suljettuina ja tarkista ett suojakannet ovat paikoillaan estksesi mahdollisen
shkisku- ja loukkaantumisvaaran.
VAROITUS!: Jos jrjestelmss virtaa neste ja laitteen sisn- ja ulosmenoventtiilit pit sulkea, laite on ensin
tyhjennettv kokonaan. Tyhjentmisen laiminlynti aiheuttaa nesteen lmplaajenemista, joka saattaa johtaa laitteen
rikkoutumiseen ja henkilvahingon vaaraan.

Eurooppalainen painelaitedirektiivi (PED)


Painelaitteet, joidenpaine on suurempi kuin 0,5 bar ja joiden koko on suurempi kuin 25 mm tai 1 tuuma , kuuluvat eurooppalaiseen
painelaitedirektiiviin (PED).
x
PED direktiivi koskevat mrykset lytyvt kyttoppaan Tekniset tiedot -luvusta.
x
Kyttoppaassa kuvatut mittarit ovat 97/23/EC EU-direktiivin mukaisia.
x
Kaikki Brooks Instrumentin virtausmittarit kuuluvat virtausryhmn 1. Laitteet jotka ovat suurempia, kuin 25 mm tai 1 tuuma, ovat
PED I, II, III kategorien mukaisia.

Mittarit joiden koko on alle 25 mm tai 1 tuuma ovat hyvn konepajakytnnn (SEP) mukaisia.

Eurooppalainen direktiivi shkmagneettisesta yhteensopivuudesta (EMC)


Brooks Instrumentin CE-merkin saaneet (shk/shkiset) laitteet tyttvt EMC direktiivin vaatimukset ja testit shkmagneettisesta
yhteensopivuudesta (2004/108/EC EMC direktiivi).
Erityist huomioita on kiinnitettv CE-merkittyjen laitteiden kytss olevien kaapelien valintaan.
Kaapelien, kiinnikkeiden ja liittimien laatu:
Brooks Insturmentin kaapelit ovat korkealaatuisia ja tyttvt CE-merkint direktiivin vaatimukset.
Muun valmistajan kaapelia kytettess on kytettv 100% suojattua kaapelia.
Liittimien tulee olla hirisuojattua tyyppi. Tarvittaessa kytetn metallisia kiinnikkeit kaapelin suojuksen kiinnittmiseen. Kaapelin
o
suojakuoren pit olla yhdistettyn metallisuojukseen tai laippaan ja sen pit olla molemmista pist suojattuna 360 . Suojaus
pttyy maadoitukseen.
Standardin mukaan korttien liittimet eivt ole metallisia. Kytettyjen kaapelien suojaus on oltava 100%, jotta se tyttisi CEmerkinnn direktiivin vaatimukset.
Suojaus pttyy maadoitukseen.
Napojen jrjestys: Katso liitteen oleva kyttopas.

Elektrostaattinen purkaus (ESD)


VAROITUS!: Tuote sislt elektroniikkakomponentteja jotka voivat vahingoittua staattisesta shkst. Sisisten
piirilevyjen purkamisessa,asennuksessa ja ksittelyss tulee noudattaa kaikkia mryksi ja ohjeita.
Asennusohjeet:
1. Jrjestelmn shkt katkaistaan.
2. Laitteen kanssa tyskentelev henkil on suojattava shkiskuilta rannehihnalla tai muulla suojavarustuksella ennen piirilevyn tai
muun sisosan asennusta, poistamista tai korjaamista.
3. Piirilevyt kuljetetaan konduktiivisessa pakkauksessa. Piirilevyt puretaan paketista juuri ennen asennusta. Poistettu piirilevy on heti
pakattava soveltuvaan suojapakkaukseen kuljettamista, varastoimista tai palautusta varten.
Huomautukset:
Tuotteen herkkyys elektrostaattiselle purkaukselle (ESD) ei ole eptavallista. Suurin osa elektroniikkatuotteista sislt
komponentteja jotka hydyntvt metallioksiditekniikkaa (NMOS, SMOS jne.) Kokemusten mukaan pienikin elektrostaattinen purkaus
voi aiheuttaa laitteiden virhetoiminnan tai vahingoittumisen. Vahingoittuneet komponentit saattavat aiheuttaa laitteen ennenaikaisen
rikkoutumisen vaikka laite nyttisi toimivan normaalisti.

A-6

Section C Essential Instructions

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

Models MT3809G & MT3810G

French

Instructions essentielles
A lire avant de commencer !
Brooks Instrument conoit, fabrique et teste ses produits pour rpondre de nombreuses normes nationales et internationales. Ces produits doivent tre
correctement installs, utiliss et entretenus pour pouvoir fonctionner dans le cadre de leurs spcifications normales. Les instructions qui suivent doivent tre
respectes et intgres votre programme de scurit lors de linstallation, lutilisation et lentretien des produits Brooks Instrument.
Afin dassurer un fonctionnement correct, faites appel du personnel qualifi pour linstallation, lutilisation, la mise jour, la programmation et lentretien du
produit.
Lisez toutes les instructions avant linstallation, lutilisation et lentretien du produit. Si le prsent manuel dutilisation nest pas le bon, consultez la dernire
page de la couverture pour connatre le point de vente le plus proche. Conservez ce manuel dutilisation pour pouvoir vous y reporter par la suite.
AVERTISSEMENT : nutilisez pas cet instrument au-del des spcifications numres dans le manuel dutilisation.
Le non-respect de cet avertissement peut entraner de graves blessures et / ou endommager lquipement.
Si vous ne comprenez pas lune des instructions, prenez contact avec un reprsentant de Brooks Instrument pour obtenir des explications.
Tenez compte de tous les avertissements, prcautions et instructions marques sur le produit et fournies avec celui-ci.
Installez votre quipement de la faon indique dans les instructions dinstallation du manuel dutilisation et conformment la lgislation en vigueur au niveau
local et national. Branchez tous les produits aux sources dlectricit et de pression agres.
Utilisation : (1) Faites lentement entrer le dbit dans le systme. Ouvrez progressivement les vannes de procd pour viter des pics de dbits. (2) Vrifiez
quil ny a pas de fuite au niveau des branchements dentre et de sortie du dbitmtre. Sil ny a pas de fuite, amenez le systme sa pression dutilisation.
Avant de procder lentretien, assurez-vous que la conduite de procd nest plus sous pression. Lorsquil faut remplacer une pice, assurez-vous que les
pices de rechange sont celles indiques par Brooks Instrument et que des personnes qualifies effectuent le remplacement. Les pices et procdures non
autorises peuvent porter atteinte au fonctionnement du produit et mettre en pril la scurit de votre procd. Les remplacements par des pices dapparence
similaire peuvent entraner des incendies, des risques lectriques ou un mauvais fonctionnement.
Vrifiez que toutes les trappes de lquipement sont fermes et que les couvercles de protection sont en place pour viter les chocs lectriques et les
blessures, sauf lorsque lentretien est ralis par des personnes qualifies.
AVERTISSEMENT : dans le cas dappareils coulement liquide, si les vannes dentre et de sortie adjacentes aux appareils doivent tre fermes pour une
raison quelconque, les appareils doivent tre compltement vidangs. Si cela nest pas fait, une ventuelle dilatation thermique du fluide peut casser lappareil
et provoquer des blessures.

Directive europenne quipements sous pression (PED)


Tous les quipements sous pression dont la pression interne est suprieure 0,5 bar (pression relative) et dont la taille dpasse 25 mm ou un pouce entrent
dans le cadre de la directive PED.
La section Spcifications de ce manuel contient les instructions relatives la directive PED.
Les appareils de mesure de ce manuel sont conformes la directive EN 97/23/EC.
Tous les dbitmtres Brooks Instrument fonctionnent avec des fluides de groupe 1.
Les appareils de mesure dune taille suprieure 25 mm ou un pouce entrent dans la catgorie PED I, II ou III.
Les appareils de mesure dune taille infrieure ou gale 25 mm ou un pouce relvent des bonnes pratiques dingnierie (SEP).

Compatibilit lectromagntique europenne (CEM)


Lquipement Brooks Instrument (lectrique / lectronique) portant le marquage CE rpond la rglementation en matire de compatibilit lectromagntique
(directive CEM 2004/108/EC).
Il faut cependant prter une grande attention au choix du cble dinterconnexion utiliser avec lquipement marqu CE.
Qualit du cble dinterconnexion, des presse-toupes et connecteurs :
Brooks Instrument fournit un ou des cbles de qualit suprieure qui rpondent aux spcifications exiges pour la certification CE.
Si vous utilisez votre propre cble dinterconnexion, ce cble doit tre protg par un blindage intgral.
Les connecteurs rectangulaires ou circulaires utiliss doivent avoir un blindage mtallique. Sil y a lieu, des presse-toupes mtalliques doivent faire office de
serre-cran de cble.
Lcran du cble doit tre raccord lenveloppe mtallique ou au presse-toupe et blind aux deux extrmits sur 360 degrs.
Le blindage doit sachever sur une prise de terre.
Les connecteurs de carte standards sont non mtalliques. Les cbles utiliss doivent tre protgs par un blindage intgral pour se conformer la certification
CE.
Le blindage doit sachever sur une prise de terre.
En ce qui concerne la configuration des broches, veuillez vous reporter au manuel dutilisation joint.

ESD (dcharge lectrostatique)


ATTENTION : cet instrument contient des composants lectroniques sensibles llectricit statique. Des procdures de manipulation adquates
doivent tre respectes pendant le retrait, linstallation ou la manipulation des cartes de circuits imprims ou des dispositifs internes.
Procdure de manipulation :
1. Lalimentation lectrique de lappareil doit tre coupe.
2. Le personnel doit tre mis la terre, au moyen dune bande de poignet ou dun autre moyen sr et adquat, avant linstallation, le retrait ou le rglage de
toutes les cartes de circuits imprims ou autres dispositifs internes.
3. Les cartes de circuits imprims doivent tre transportes dans un rcipient conducteur. Les cartes ne doivent enleves de cette enveloppe protectrice quau
dernier moment, juste avant linstallation. Les cartes retires doivent tre immdiatement places dans un rcipient de protection pour le transport, le stockage
ou le retour lusine.
Observations
Brooks Instrument nest pas le seul proposer des produits comportant des composants sensibles aux dcharges lectrostatiques. La plupart des produits
lectroniques modernes contiennent des composants qui utilisent des technologies oxydes mtalliques (NMOS, SMOS, etc.). Lexprience dmontre que
dinfimes quantits dlectricit statique suffisent endommager ou dtruire ces appareils. Les composants endommags, mme sils semblent fonctionner
correctement, tombent rapidement en panne.

A-7

Section C Essential Instructions

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

GF100 Series
German

Wichtige Anweisungen
Bitte zuerst lesen!
Brooks Instrument entwickelt, produziert und testet seine Produkte derart, dass sie viele nationale und internationale Standards erfllen. Nur bei korrektem Einbau
sowie richtiger Bedienung und Wartung dieser Produkte ist ein Betrieb unter Einhaltung der Standardvorgaben sichergestellt. Die folgenden Anweisungen mssen
eingehalten werden und in Ihr Sicherheitsprogramm integriert werden, wenn Sie Brooks Produkte installieren, bedienen und warten.
Um die entsprechende Leistung zu gewhrleisten, setzen Sie qualifiziertes Personal fr die Installation, den Betrieb, die Aktualisierung, Programmierung und
Wartung des Produkts ein.
Lesen Sie alle Anweisungen, bevor Sie das Produkt installieren, in Betrieb nehmen und warten. Falls es sich bei diesem Handbuch nicht um das richtige
Handbuch handelt, schauen Sie bitte auf der Rckseite nach den Kontaktdaten Ihres Vertriebsbros vor Ort. Bewahren Sie dieses Handbuch auf, falls Sie
spter etwas nachschauen mchten.
WARNUNG: Dieses Gert nicht auerhalb der in Bedienungsanleitung und Handbuch angegebenen Grenzen betreiben. Wird diese Warnung nicht
beachtet, kann dies zu schweren Personenschden bzw. Schden des Gertes fhren.
Falls Sie Anweisungen nicht verstehen, wenden Sie sich zur Klrung an Ihren Brooks Instrument Vertreter.
Befolgen Sie alle Warnhinweise und Anweisungen, die auf dem Produkt markiert sind oder zusammen mit diesem geliefert werden.
Installieren Sie Ihr Gert, wie in den Installationsanweisungen des entsprechenden Handbuchs angegeben und gem der gltigen regionalen und nationalen
Gesetze. Schlieen Sie alle Produkte an eine geeignete Strom- und Druckluftversorgung an.
Bedienung: (1) Langsam den Zufluss zum System starten. Die Ventile langsam ffnen, um einen sprunghaften Anstieg der Durchflussmenge zu verhindern. (2)
Bereich der Anschlsse (Zufluss und Ausfluss) des Durchflussmessers auf Undichtigkeiten berprfen. Wenn das System dicht ist, auf Betriebsdruck
hochfahren.
Sicherstellen, dass der Leitungsdruck vor Wartungsarbeiten heruntergefahren wird. Wenn Ersatzteile bentigt werden, stellen Sie sicher, dass qualifizierte
Personen Ersatzteile verwenden, die von Brooks Instrument vorgegeben sind. Nicht genehmigte Teile und Verfahren knnen die Leistungsfhigkeit des
Produkts beeintrchtigen und den sicheren Betrieb Ihres Prozesses gefhrden. hnlich aussehende Austauschteile knnen zu Brnden, elektrischen Gefahren
oder nicht sachgerechtem Betrieb fhren.
Stellen Sie sicher, dass alle Tren der Anlage geschlossen sind und dass alle Schutzabdeckungen angebracht sind, um Stromschlge und Personenschden
zu vermeiden, es sei denn die Wartungsaufgaben werden von qualifizierten Personen durchgefhrt.
WARNUNG: Werden die Ein- und Auslassventile neben Durchflussmessgerten aus irgendwelchen Grnden geschlossen, so mssen die Gerte
komplett entleert werden.
Durchflussmessgeraete muessen vor dem Schliessen von Ein- und Auslassventilen komplett entleert werden,
anderenfalls kann es zu einer thermischen Ausdehnung der Flssigkeit und damit zum Bruch des Gertes kommen; Personenschden knnen die
Folge sein.

Europische Druckgerterichtlinie (PED)


Alle Druckgerte mit einem internen Druck von mehr als 0,5 bar (g) und einer Gre von mehr als 1in ( 1 in = 25,4 mm) unterliegen der Druckgerterichtlinie.
x Das Kapitel zu den technischen Daten in dieser Anleitung enthlt wichtige Sicherheits- und Betriebsanweisungen in Bezug auf die Druckgerterichtlinie.
x Messgerte, die in diesem Handbuch beschrieben sind, erfllen die europische Richtlinie 97/23/EG.
x Alle Durchflussmesser von Brooks Instrument fallen unter die Fluidgruppe 1.
x Messgerte, die grer als 25 mm oder 1 (inch) sind, erfllen die Kategorien I, II oder III der Druckgerterichtlinie (PED).
x Messgerte mit einer Gre von 25 mm oder 1 (inch) oder kleiner sind Sound Engineering Practice (SEP).

Europische Verordnung zur elektromagnetischen Vertrglichkeit (EMV)


Gerte von Brooks Instrument (elektrischer und elektronischer Art) mit CE-Zeichen haben den Test auf Einhaltung der Verordnung zur elektromagnetischen
Vertrglichkeit (EMV Richtlinie 2004/108/EC) erfolgreich bestanden.
Dennoch muss bei der Wahl des Signalkabels fr das Gert mit CE-Zeichen auf folgende Dinge geachtet werden.
Qualitt von Signalkabel, Kabeldurchfhrung und Anschlsse:
Brooks Instrument liefert qualitativ hochwertige Kabel, die den Anforderungen fr eine CE-Zertifizierung entsprechen.
Sollten Sie eigene Kabel einsetzen, so sollte das Kabel berall mit einer 100%-Abschirmung versehen sein.
D- oder Rundstecker sollten eine Metallabschirmung aufweisen. Wenn mglich, mssen Kabeldurchfhrungen aus Metall mit Kabelschirmgeflechts-Klemmen
verwendet werden.
Der Kabelschirm sollte an die Metallhlle oder -durchfhrung angeschlossen werden und an beiden Enden rundherum (360 ) abgeschirmt werden.
Die Abschirmung sollte geerdet werden.
Randstecker auf Platinen sind standardmig nicht aus Metall. Die verwendeten Kabel mssen mit einer 100 % Abschirmung versehen werden, um die CEVorgaben zu erfllen.
Die Abschirmung sollte geerdet werden.
Klemmenbelegung: Siehe beigefgtes Handbuch.

ESD (Elektrostatische Entladung)


ACHTUNG: Dieses Gert enthlt elektronische Komponenten, die durch elektrostatische Entladungen beschdigt werden knnen. Ordnungsgeme
Verfahrensanweisungen mssen whrend des Ausbaus, der Installation oder anderer Handhabung der eingebauten Platinen oder Gerte eingehalten
werden.
Verfahrensanweisung:
1. Trennen Sie das Gert von der Stromversorgung.
2. Das Personal ist vor dem Einbau, Ausbau oder der Einstellung von Platinen oder anderen internen Komponenten durch ein entsprechendes Armband mit dem
Erdpotential zu verbinden.
3. Platinen sind in speziellen Behltern mit Schutz gegen elektrostatische Spannungen zu transportieren oder zu lagern. Platinen drfen erst kurz vor dem Einbau
aus der Schutzhlle entfernt werden. Ausgebaute Platinen mssen umgehend in Schutzbehlter zum Transport, zur Lagerung oder Rcksendung an das Werk
gelegt werden.
Anmerkung
Dieses Gert ist wie viele andere elektronische Gerte auch mit Komponenten bestckt, die anfllig fr elektrostatische Entladung sind. Die meisten modernen,
elektronischen Gerte enthalten Komponenten, die die Metalloxidtechnologie (NMOS, SMOS etc.) verwenden. Die Erfahrung hat gezeigt, dass schon geringe
Mengen elektrostatischer Energie ausreichen, um diese Gerte zu beschdigen oder zu zerstren. Beschdigte Teile fallen frh aus, obwohl sie funktionsfhig zu
sein scheinen.

A-8

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

Section C Essential Instructions

Models MT3809G & MT3810G

Greek


!
Brooks Instrument , .
, .
,
Brooks Instrument.
x
, , , .
x
, . ,
. .
:
. /
.
x
Brooks Instrument.
x
, .
x

. .
x
: (1) .
. (2) . ,
.
x
.
Brooks Instrument.
.
, .
x

, .
: ,
, .
,
.

(PED)
0,5 bar (g) 25 mm 1
(PED).
x
PED.
x
97/23/.
x
Brooks Instrument 1.
x
25 mm 1 , PED.
x
25 mm 1 (SEP).

(EMC)
(/) Brooks Instrument CE
( 2004/108/EC EMC).
CE.
, :
Brooks Instrument CE.
, 100% .
D .
.
360 .
.
. 100%
CE. .
: .

(ESD)
:
. ,
.
:
1. .
2. ,
.
3. .
.
, .
:
ESD ( ) .
(NMOS, SMOS .).
.
, , .

A-9

Section C Essential Instructions

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

GF100 Series
Hungarian

Alapvet utastsok
Elszr olvassa el ezeket!
A Brooks Instrument olyan mdon tervezi, gyrtja s teszteli termkeit, hogy azok megfeleljenek szmos belfldi s nemzetkzi szabvnynak. Ezeket
a berendezseket megfelelen kell telepteni, zemeltetni s karbantartani ahhoz, hogy mindenkppen a norml mkdsi tartomnyuknak
megfelelen zemelhessenek. Az albbi utastsokat be kell tartani, s be kell pteni a munkavdelmi programba a Brooks Instrument termkeinek
teleptse, zemeltetse s karbantartsa sorn.
A megfelel teljestmny garantlsa rdekben kizrlag szakkpzett szemlyzet vgezze a termk teleptst, zemeltetst, frisstst,
programozst s karbantartst.
Valamennyi utastst el kell olvasni a termk teleptse, zemeltetse s szervizelse eltt. Amennyiben ez a kziknyv nem a megfelel kiadvny, a
hts bortn keresse meg a helyi forgalmazt, s tovbbi tjkoztatsrt lpjen kapcsolatba vele. rizze meg ezt a kziknyvet ksbbi
tjkoztatsknt.
FIGYELEM: Ne mkdtesse a berendezst az zemeltetsi utastsban megadott zemi tartomnyokon tl. Ennek megsrtse slyos
szemlyi srlshez vagy a berendezs meghibsodshoz vezethet.
x
Amennyiben a gpknyv utastsai nem egyrtelmek, lpjen kapcsolatba Brooks Instrument kpviseljvel, hogy tisztzzk a problmt.
x
Tartsa be a berendezsen feltntetett vagy azzal egytt szlltott sszes figyelmeztetst, felhvst s utasts.
x
A megfelel teleptsi utastsban megadott utastsok valamint a hatlyos helyi s nemzeti elrsok szerint teleptse a berendezst. A
termkeket kizrlag a megfelel elektromos s nyomsellt forrsra ksse.
x
Menete: (1) Lassan helyezze nyoms al a rendszert. Lassanknt nyissa ki az zemi szelepeket az ramlsingadozs elkerlse rdekben. (2)
Ellenrizze, nincs-e szivrgs az ramlsmr be-, s kimeneti bektseinl. Ha nincs szivrgs, tltse fel a rendszert az zemi nyomsra.
x
Szervizels eltt mindenkppen ellenrizze, hogy az zemi vezetk nincs-e nyoms alatt. Amennyiben cserealkatrszekre van szksg,
mindenkppen szakkpzett szemlynek kell kezelnie a Brooks Instrument ltal meghatrozott cserealkatrszeket. A nem engedlyezett
alkatrszek s tevkenysgek befolysolhatjk a termk teljestmnyt, illetve veszlyeztethetik a biztonsgos zemeltetst. A pusztn hasonl
alkatrszekkel trtn helyettests tzet, ramtsveszlyt vagy elgtelen mkdst eredmnyezhet.
x
A berendezs sszes ajtaja mindenkppen legyen zrva, a vdburkolatok pedig legyenek a helykn az ramts s a szemlyi srlsek
elkerlse rdekben, kivve, ha szakkpzett szakember vgez rajta karbantartsi munkkat.
FIGYELEM: Folyadkot ramoltat berendezsek esetben, ha brmilyen okbl el kell zrni a berendezs melletti ki-, s belp
szelepeket, a berendezst teljesen le kell rteni. Ennek elmulasztsa a folyadk htgulst okozhatja, ami krosthatja a berendezst, s
szemlyi srlshez vezethet.

Nyomstart berendezsekre vonatkoz eurpai irnyelv (PED)


Minden 0,5 bar-nl (g) magasabb bels nyoms s 25 mm-nl vagy 1 hvelyknl nagyobb nyomstart berendezs a nyomstart berendezsekre
vonatkoz eurpai irnyelv (PED) hatlya al tartozik.
x
A gpknyv Mszaki adatok fejezete tartalmaz a PED irnyelvre vonatkoz utastsokat.
x
A gpknyvben megadott mreszkzk megfelelnek a 97/23/EK EU irnyelvnek.
x
Minden Brooks tfolysmr az 1-es folyadkcsoportba tartozik.
x
A 25 mm-nl vagy 1 hvelyknl nagyobb mreszkzk megfelelnek a PED I, II, vagy III kategrinak.
x
A 25 mm-es illetve 1 hvelykes vagy kisebb mreszkzk az elfogadott mrnki gyakorlatot (SEP) kvetik.

Elektromgneses kompatibilitsra vonatkoz eurpai irnyelv (EMC)


A Brooks Instrument CE jellst kirdemelt (elektromos/elektronikus) berendezsei sikeresen teljestettk az elektromgneses kompatibilitsi
kvetelmnyeket (2004/108/EC sz. EMC irnyelv) vizsgl teszteket.
Ugyanakkor klns figyelmet kell fordtani a CE jells berendezsekhez felhasznlt jelkbelek kivlasztsra.
A jelkbelek, kbelsszektk, csatlakozk minsge:
A Brooks Insturment magas minsg kbeleket knl, melyek megfelelnek a CE minsts kvetelmnyeinek.
Amennyiben sajt jelkbelt alkalmaznak, olyat kell vlasztani, amely 100%-os rnykolssal, teljes mrtkben szrt.
A D vagy kr alak csatlakozknak fmrnykolval rnykoltnak kell lennie. Szksg esetn fm kbelsszektket kell alkalmazni a kbelszr
rgztsre.
A kbelszrt a fm hzhoz vagy hvelyhez kell csatlakoztatni s mindkt feln 360-ban le kell rnykolni. Az rnykolsnak fldelsben kell
vgzdnie.
A krtykhoz tartoz csatlakozk szabvnyosan nem fmesek. Az alkalmazott kbeleknek 100%-s rnykolssal szrteknek kell lennik, hogy
megfeleljenek a CE minstsnek.
Az rnykolsnak fldelsben kell vgzdnie.
rintkez konfigurci: Lsd a mellkelt kezelsi utastst.

Elektrosztatikus kisls (ESD)


VIGYZAT: A kszlk olyan alkatrszeket tartalmaz, melyek hajlamosak a sztatikus elektromossg okozta krosodsra. Be kell tartani
a megfelel eljrsokat a bels ramkri krtyk s eszkzk eltvoltsa, behelyezse vagy egyb kezelse sorn.
Kezelsi eljrs:
1. A berendezst ramtalantani kell.
2. A szemlyt fldelni kell csuklpnttal vagy egyb biztonsgos s a clra alkalmas eszkzzel, mieltt ramkri krtyt vagy egyb bels eszkzt
teleptene, venne ki, vagy lltana be.
3. A nyomtatott ramkri krtykat vezetkpes csomagolsban kell szlltani. A krtyk kizrlag kzvetlenl a behelyezs eltt vehetk ki a
vdburkolatbl. A kiszerelt krtyt haladktalanul el kell helyezni a mozgatsra, raktrozsra vagy a gyri visszaszlltsra szolgl
vdcsomagolsba.
Megjegyzsek:
Nem egyedi jelensg, hogy a kszlkben elektrosztatikus kislsre (ESD) rzkeny alkatrszek tallhatk. A legtbb korszer elektronikus

eszkzben fmoxid technolgis alkatrszek (NMOS, SMOS stb.) tallhatk. A tapasztalatok azt igazoljk, hogy mg
kis mrtk sztatikus elektromossg is krosthatja, vagy tnkreteheti ezeket az eszkzket. A krosodott alkatrszek,
mg ha ltszlag megfelelen mkdnek is, kezdd hibra utalnak.
A-10

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

Section C Essential Instructions

Models MT3809G & MT3810G

Italian

Istruzioni fondamentali
Leggerle subito!
La Brooks Instrument progetta, fabbrica e collauda i propri prodotti in maniera tale che siano conformi ai vari standard nazionali ed internazionali. Tali
apparecchiature devono essere installate, messe in esercizio e tenute in manutenzione in maniera adeguata affinch operino in conformit alle loro normali
specifiche di funzionamento. Le seguenti istruzioni devono essere rispettate ed inserite nel programma di tutela sul lavoro durante linstallazione, il
funzionamento e la manutenzione dei prodotti Brooks Instrument.
x
Per garantire un adeguato rendimento l'installazione, il funzionamento, laggiornamento, la programmazione e la manutenzione del prodotto devono
essere eseguiti esclusivamente da personale specializzato.
x
Leggere tutte le istruzioni prima dellinstallazione, utilizzo e manutenzione del prodotto. Se questo manuale non quello relativo al Vostro prodotto,
cercare sul retro della copertina il distributore locale e contattarlo per ulteriori informazioni. Conservare il presente manuale per future consultazioni.
ATTENZIONE: Non utilizzare questo strumento in condizioni che eccedono le specifiche riportate nel Manuale dUso. Linosservanza pu
causare gravi lesioni alle persone e/o danni allapparecchiatura.
x
Qualora le istruzioni del manuale non siano chiare, contattare un rappresentante della Brooks Instrument per chiarire il problema.
x
Rispettare tutti gli avvisi, le istruzioni e gli avvertimenti riportati sullapparecchiatura o forniti insieme ad essa.
x
Installare lapparecchiatura in base alle istruzioni riportate nel Manuale dUso e alle prescrizioni locali e nazionali in vigore. Collegare i prodotti
esclusivamente ad unadeguata sorgente di pressione ed alimentazione elettrica.
x
Procedimento: (1) mettere lentamente sotto pressione il sistema. Aprire lentamente le valvole di servizio per evitare loscillazione del flusso. (2) Controllare
che non ci siano perdite nei punti di connessione in entrata e in uscita del misuratore di flusso. Se non ci sono perdite, caricare il sistema alla pressione
desercizio.
x
Prima di effettuare manutenzione controllare che la linea di processo non sia sotto pressione. Se avete bisogno di pezzi di ricambio, il personale
specializzato deve usare i pezzi di ricambio definiti dalla Brooks Instrument. Attivit e pezzi di ricambio non autorizzati possono influire sul rendimento del
prodotto e comprometterne il funzionamento in sicurezza. La sostituzione con pezzi di ricambio non originali pu causare incendi, pericolo di scosse
elettriche o funzionamento improprio.
x
Tutti gli sportelli dellimpianto devono essere chiusi, le cappe di protezione devono essere al loro posto per evitare scosse elettriche e lesioni personali,
tranne quando il personale specializzato esegue lavori di manutenzione.
ATTENZIONE: In caso di apparecchiature in cui scorre un liquido, se per qualsiasi motivo bisogna chiudere le valvole dentrata e duscita
accanto allapparecchiatura, allora si deve svuotare completamente lapparecchiatura. Linosservanza pu causare la dilatazione termica del liquido
che pu danneggiare lapparecchiatura e provocare lesioni alle persone.

Direttiva europea relativa alle apparecchiature a pressione (PED)


Ogni apparecchiatura a pressione con pressione interna maggiore di 0,5 bar (g) e pi grande di 25 mm o di 1 pollice ricade nellambito della Direttiva Europea
relativa alle apparecchiature a pressione (PED).
x
Il capitolo Dati tecnici del manuale contiene le disposizioni relative alla direttiva PED.
x
Gli strumenti di misura descritti nel presente manuale sono conformi alla Direttiva UE 97/23/CE.
x
Ogni flussimetro Brooks appartiene al gruppo di fluidi 1.
x
Gli strumenti di misura maggiori di 25 mm o di 1 pollice sono conformi alla categoria I, II o III della PED.

Gli strumenti di misurazione minori di 25 mm o di 1 pollice rientrano nella categoria SEP (Sound Engineering Practice).

Direttiva europea relativa alla compatibilit elettromagnetica (EMC)


Le apparecchiature (elettriche/elettroniche) Brooks Instrument dispongono del marchio CE ed hanno superato positivamente i test per i requisiti di compatibilit
elettromagnetica (Direttiva EMC 2004/108/EC).
In ogni caso bisogna prestare particolare attenzione alla scelta dei cavi di segnale utilizzati per le apparecchiature con marchio CE.
Qualit dei cavi di segnale, dei pressacavi e dei connettori:
La Brooks Instrument offre cavi dalta qualit conformi ai requisiti della certificazione CE.
Qualora vengano utilizzati cavi di segnale propri, devono essere scelti con schermatura al 100% e interamente filtrati.
I connettori D o rotondi devono essere schermati con schermatura metallica. In caso di necessit bisogna utilizzare pressacavi metallici di collegamento per
fissare la schermatura del cavo.
La schermatura del cavo deve far contatto col guscio metallico o col pressacavo; il cavo deve essere schermato su entrambi i lati a 360. La schermatura deve
essere effettuata con messa a terra.
I connettori Card Edge normalmente non sono di metallo. I cavi utilizzati devono essere filtrati con schermatura al 100% per essere conformi alla marcatura
CE.
La schermatura deve essere effettuata con messa a terra.
Configurazione pin: Vedi Manuale duso allegato.

Scarica elettrostatica (ESD)


ATTENZIONE: Il dispositivo contiene componenti elettronici che possono essere danneggiati da elettricit statica. Bisogna rispettare le
adeguate procedure durante la rimozione, linstallazione o altra manovra delle schede del circuito elettrico interno.
Procedura di manovra:
1. Togliere alimentazione elettrica allapparecchiatura.
2. La persona deve essere collegata a terra con una cerniera o con altri strumenti di sicurezza e adeguati allo scopo prima di installare, togliere o impostare
la scheda del circuito elettrico o altri dispositivi interni.
3. Le schede del circuito stampato devono essere spedite in contenitori conduttivi. Le schede devono essere tolte dal rivestimento protettivo esclusivamente
prima dellinstallazione. Le schede confezionate devono essere collocate immediatamente nellimballaggio protettivo per la movimentazione,
limmagazzinamento o resa alla fabbrica.
Note:
un fenomeno comune che nei dispositivi di questo tipo si trovino componenti sensibili alla scarica elettrostatica (ESD). Nella maggior parte degli strumenti
elettronici moderni si trovano componenti tecnologici metallo-ossido (NMOS, SMOS, ecc.). Le esperienze dimostrano che lelettrostaticit anche in piccola
misura pu danneggiare o rovinare gli strumenti. I componenti danneggiati, anche se allapparenza funzionano correttamente, potrebbero manifestare il difetto
rapidamente.

A-11

Section C Essential Instructions

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

GF100 Series
Latvian

Svarga instrukcija
Pirms turpint izlasiet!
Brooks Instrument projekt, rao un prbauda savus raojumus atbilstoi daudziem nacionlajiem un starptautiskajiem standartiem. Lai nodrointu o
izstrdjumu turpmku darbbu atbilstoi noteiktajiem parametriem, tie ir pareizi juzstda, jlieto un japkopj. Uzstdot, lietojot Brooks Instrument
izstrdjumus un veicot to apkopi, ir jievro ie nordjumi un jiekauj tie jsu drobas programm.
x
Lai nodrointu piencgu izstrdjuma sniegumu, izstrdjuma uzstdanu, lietoanu, atjauninanu, programmanu un apkopi uzticiet veikt tikai
kvalifictam personlam.
x
Pirms izstrdjuma uzstdanas, lietoanas un apkalpoanas izlasiet visus nordjumus. Ja instrukciju rokasgrmata nav pareiz, izstrdjumam
atbilsto rokasgrmata, ldzu skat. aizmugurjo vku, kur ir sniegta vietj tirdzniecbas biroja kontaktinformcija.
BRDINJUMS! Nelietot instrumentu rpus Instrukciju un lietoanas rokasgrmat nordtajiem parametriem. brdinjuma
neievroanas rezultt var rasties traumas un / vai aprkojuma bojjumi.
x
Ja js nesaprotat kdu no instrukcijm, sazinieties ar Brooks Instrument prstvi un ldziet izskaidrot to.
x
Ievrojiet visus brdinjumus, piesardzbas mrus un instrukcijas, kas nordti uz izstrdjuma vai piegdti kop ar to.
x
Uzstdiet aprkojumu t, k tas nordts attiecgaj instrukciju rokasgrmat iekautaj uzstdanas instrukcij un atbilstoi piemrojamajm vietjm
un nacionlajm normm. Pievienojiet visus izstrdjumus pareiziem elektriskajiem un spiediena avotiem.
x
Lietoana: (1) Lnm uzsciet plsmu sistm. Lai izvairtos no straujiem plsmas kpumiem, lnm atveriet procesa vrstus. (2) Prbaudiet, vai nav
nopldes ap plsmas mrtja iepldes un izpldes savienojumiem. Ja nopldes nav, uzstdiet sistm darba spiedienu.
x
Prliecinieties par to, lai pirms instrumenta tehnisks apkopes btu likvidts procesa lnijas spiediens. Ja ir nepiecieams veikt kdu dau nomaiu,
nodroiniet, lai tiktu izmantotas Brooks Instrument nordts daas un dau nomaiu veiktu kvalificts personls. Neatautu dau un procedru
izmantoana var ietekms raojuma sniegumu un samazint procesa drobu. Ldzgu, bet ne identisku dau nomaias lietoana var izraist
ugunsgrka, elektrisko traucjumu riskus un nepareizu izstrdjuma darbbu.
x
Nodroiniet, lai btu aizvrtas visas durvis un btu pareizi uzstdti visi aizsargprsegumi, tdjdi novrot elektrooka un traumu risku. Izmums ir
gadjumi, kad kvalificts personls veic raojuma apkopi.
BRDINJUMS! Ja idrs plsmas ieru tuvum esoos iepldes un izpldes vrstus kda iemesla d ir jaizver, no iercm ir jizlai
viss idrums. Pretj gadjum idrums var termiski izplesties, prraut ierci un radt traumas.

Eiropas spiedieniekrtu direktva (PED)


Uz vism spiedieniekrtm, kuru iekjais spriegums prsniedz 0,5 bar (g) un ir lielkas par 25 mm jeb 1 (collu), attiecas Eiropas spiedieniekrtu direktva
(PED).
x
s rokasgrmatas tehnisko parametru noda ir sniegtas a PED Direktvu saistts instrukcijas.
x
aj rokasgrmat apraksttie mrtji atbilst EN Direktvas 97/23/EK prasbm.
x
Visi Brooks Instrument plsmas mrtji ietilpst 1. idrumu grup.
x
Uz 25 mm jeb 1 (collu) maziem un mazkiem mrtjiem attiecas labas inenierijas prakse (SEP).
x
25 mm jeb 1 (collu) mazi vai mazki mrtji atbilst PED kategorijai I, II vai III.

Eiropas elektromagntisks savietojambas direktva (EMS)


Brooks Instrument (elektrisks / elektronisks) iekrtas ar CE zmi ir izturjuas prbaudi un atztas par atbilstom Eiropas elektromagntisks
savietojambas direktvas (EMS) prasbm (EMS 2004/108/EC)
Tomr, izvloties signlkabeli, kas tiks lietots kop ar CE marjuma iekrtu, ir jievro paa uzmanbaSignlkabea, kabea blvslgu un savienotju
kvalitte:
Brooks Instrument piegd augstas kvalittes kabeus, kas atbilst CE sertifikcijas tehniskajiem parametriem.
Ja js lietojat pats savu signlkabeli, tam ir jbt pilnb, 100% ekraniztam.
D un apa tipa savienotjiem ir jbt aprkotiem ar metla ekranizjumu. Ja nepiecieams, ir jizmanto metla blvslgi ar kabea ekranizjuma skavojumu.
Kabea ekranizjumam ir jbt savienotam ar metla apvalku un abs puss aizsargtam 360 grdu diapazon.
Ekranizjumam ir jbeidzas pie iezemjuma.
Card Edge savienotji standarta izpildjum ir nemetla. Kabeiem ir jbt prkltiem ar 100% ekranizjumu, lai tie atbilstu CE sertifikcijas prasbm.
Ekranizjumam ir jbeidzas pie iezemjuma.
Attiecb uz tapu konfigurciju: skat. pievienoto instrukciju rokasgrmatu.

ESD (elektrostatisk izlde)


IEVROT PIESARDZBU! is instruments satur elektriskos komponentus, kas ir jutgi pret statisko elektrbu. Izemot un uzstdot iekjs
des plates un ierces vai k citdi darbojoties ar tm, ir jievro noteikta darba krtba.
Darba krtba:
1.
Iekrta jatsldz no baroanas.
2.
Pirms jebkdas druktas shmas kartes vai citas iekjs ierces uzstdanas, izemanas vai regulanas personlam, kas veiks os darbus, ir jbt
iezemtam, piem., izmantojot aproces vai citus drous, piemrotus ldzekus.
3.
Drukts shmas kartes ir jtransport vado iepakojum. Plksnes no aizsargkorpusa drkst izemt tikai tiei pirms uzstdanas. Transportjot,
uzglabjot vai atgrieot rpnc no izemts plksnes ir nekavjoties jievieto aizsargiepakojum.
Komentri
Instruments nav unikls taj aspekt, ka tas satur pret ESD (elektrostatisko izldi) jutgus komponentus. Vairums msdienu elektroiekrtu satur komponentus,
kuru raoan izmantota metla oksdu tehnoloijas (NMOS, SMOS u.c.). Pieredze rda, ka pat neliels daudzums statisks elektrbas var nodart bojjumus
dm iercm vai pilnb sabojt ts. Bojtie komponenti pat tad, ja tie ietami darbojas pareizi, ir pakauti trkai atteicei.

A-12

Section C Essential Instructions

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

Models MT3809G & MT3810G

Lithuanian

Pagrindins instrukcijos
Perskaitykite prie tsdami!
Brooks Instrument projektuoja, gamina ir ibando savo gaminius, kad jie atitikt vairius nacionalinius ir tarptautinius standartus. ie gaminiai turi bti tinkamai
montuojami, eksploatuojami ir priirimi, kad ir toliau veikt pagal jiems bdingus techninius parametrus. Toliau pateikt nurodym reikia laikytis ir traukti juos
saugos program montuojant, eksploatuojant ir priirint Brooks Instrument produktus.
x
Siekiant utikrinti tinkam veikim, montuoti, eksploatuoti, naujinti, programuoti ir priirti gamin turi tik kvalifikuoti darbuotojai.
x
Perskaitykite visus nurodymus prie montuodami, eksploatuodami ir priirdami gamin. Jei gavote netinkam instrukcij, galiniame jos virelyje irkite
vietins prekybos atstovybs kontaktin informacij. Isaugokite i instrukcij pasiirjimui ateityje.

x
x
x
x
x
x

SPJIMAS: nenaudokite io prietaiso virydami instrukcijoje ir eksploatacijos vadove nurodytus techninius duomenis. Nesilaikydami io
spjimo galite sunkiai susieisti ir (arba) sugadinti rang.
Jei nesuprantate koki nors nurodym, kreipkits Brooks Instrument atstov, kad paaikint.
Paisykite vis spjimo, perspjim ir nurodym, paymt ant gaminio arba pateikt su juo.
rang montuokite taip, kaip nurodyta atitinkamos instrukcijos montavimo nurodymuose arba taikomuose vietiniuose ar nacionaliniuose kodeksuose. Visus
gaminius junkite prie tinkam elektros ir slgio altini.
Eksploatacija: (1) ltai junkite sraut sistem. Ltai atidarykite proceso votuvus, kad ivengtumte srauto antpldi. (2) Patikrinkite, ar nra nuotki
aplink srauto matuoklio leidimo ir ileidimo jungtis. Jei nuotki nra, sukurkite sistemoje darbin slg.
Prie atlikdami prieiros darbus btinai paalinkite slg proceso linijoje. Jei reikia pakeisti dalis, utikrinkite, kad kvalifikuoti darbuotojai naudot Brooks
Instrument nurodytas pakaitines dalis. Netinkamos dalys ir procedros gali pakenkti gaminio veikimui ir kelti pavoj saugiai js proceso eksploatacijai. Tik
panaiai atrodantys pakaitalai gali slygoti gaisr, elektros pavojus ar netinkam veikim.
Utikrinkite, kad visos rangos durels bt udarytos, o apsauginiai dangiai udti, kad ivengtumte elektros smgio ir sueidim, iskyrus kai kvalifikuoti
darbuotojai atlieka prieiros darbus.
SPJIMAS: naudojant skysto srauto renginius, jei dl kokios nors prieasties prireikia udaryti alia renginio esanius leidimo ir ileidimo
votuvus, i renginio reikia ileisti vis skyst. To nepadarius galimas iluminis skysio pltimasis, galintis sugadinti rengin ir sueisti mons.

Europos slgins rangos direktyva (PED)


Visa slgin ranga, kurios vidinis slgis didesnis nei 0,5 bar (g), o dydis didesnis nei 25 mm arba 1 colis, yra reglamentuojama slgins rangos direktyvos
(PED).
x
ios instrukcijos dalyje Techniniai duomenys pateikiami nurodymai, susij su PED direktyva.
x
ioje instrukcijoje aprayti matuokliai atitinka Europos Sjungos direktyv 97/23/EB.
x
Visi Brooks Instrument srauto matuokliai priklauso 1 skysi grupei.
x
Didesni nei 25 mm arba 1 colis matuokliai atitinka PED I, II arba III kategorij.

25 mm arba 1 colio ar maesni matuokliai atitinka tinkam ininerijos praktik (SEP).

Europoje taikomi elektromagnetinio suderinamumo (EMC) reikalavimai


CE enklu paymta Brooks Instrument (elektrin / elektronin) ranga buvo skmingai ibandyta pagal elektromagnetinio suderinamumo reikalavimus (EMC
direktyv 2004/108/EC).
Bet reikia ypatingo dmesio renkantis signalizavimo kabel, kuris bus naudojamas su CE enklu paymta ranga.
Signalizavimo kabelio, kabeli riebokli ir jungi kokyb:
Brooks Instrument tiekia kokybikus kabelius, kurie atitinka CE sertifikavimo specifikacijas.
Jei naudojate savo signalizavimo kabel, jis turi bti visikai ir visas ekranuotas 100 % ekranu.
Naudojamos D arba apskrito tipo jungtys turi bti ekranuotos metaliniu ekranu. Jei taikoma, reikia naudoti metalinius kabeli rieboklius, utikrinanius kabelio
ekrano suspaudim.
Kabelio ekran reikia jungti prie metalinio apvalkalo ar rieboklio ir ekranuoti abiejuose galuose 360 laipsni.
Ekranas turi baigtis eminimu.
Standartins kratins jungtys yra ne metalins. Naudojami kabeliai turi bti ekranuoti 100 % ekranu, kad atitikt CE sertifikavim. Ekranas turi baigtis eminimu.
Keturi kontakt konfigracija: r. pridt instrukcij.

ESD (elektrostatinis ilydis)


PERSPJIMAS: iame prietaise yra elektronini komponent, kuriuos gali sugadinti statin elektra. Iimant ar dedant vidines spausdintines
ploktes ar renginius, arba atliekant su jomis kitus darbus, reikia laikytis tinkam darbo procedr.
Darbo procedra:
1. Atjunkite renginio maitinim.
2. Darbuotojai turi pasirpinti eminimu naudodami rieo juostel ar kitas saugias tinkamas priemones prie ddami, iimdami ar reguliuodami bet koki
spausdintins plokts kortel ar kit vidin komponent.
3. Spausdintins plokts korteles reikia transportuoti laidiame konteineryje. Neleidiama iimti plokts i apsauginio dklo, nebent prie pat djim. Iimtas
ploktes reikia nedelsiant dti apsaugin konteiner transportavimui ar saugojimui arba grinti gamykl.
Pastabos
is instrumentas nra unikalus dl jame esani ESD (elektrostatiniam ilydiui) jautri komponent. Daugelyje iuolaikini elektronini gamini yra komponent,
kuriuose naudojama metalo oksid technologija (NMOS, SMOS ir pan.). Patirtis rodo, kad net ir maas statins elektros kiekis gali pakenkti tokiems gaminiams
ar juos sugadinti. Sugadinti komponentai, net jei atrodo, kad jie veikia tinkamai, anksti nustoja veikti.

A-13

Section C Essential Instructions

GF100 Series

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

Polish

Zalecenia wstpne
Prosimy przeczyta przed rozpoczciem uytkowania!
Brooks Instrument projektuje, wytwarza i testuje swoje produkty tak, aby speniay wymagania licznych norm krajowych i midzynarodowych. Te
produkty musz by poprawnie instalowane, obsugiwane oraz konserwowane, aby zapewni ich prawidowe dziaanie zgodnie ze specyfikacj
techniczn. Podczas instalowania, obsugiwania i konserwowania produktw firmy Brooks Instrument naley przestrzega nastpujcych zalece:
x
Aby zapewni waciwe dziaanie sprztu, instalacja, obsuga, aktualizacje, programowanie i konserwacja powinny by wykonywane przez
przeszkolony personel.
x
Przed instalacj, obsug i czynnociami serwisowymi naley zapozna si ze wszystkimi zaleceniami producenta. Aby uzyska instrukcj
obsugi odpowiedni dla danego sprztu naley skontaktowa si z lokalnym przedstawicielem handlowym producenta. Instrukcj obsugi
naley zachowa do pniejszego uycia.
OSTRZEENIE: Nie wolno przekracza podanych w instrukcji zakresw dziaania urzdzenia. Nieprzestrzeganie tego zalecenia
moe doprowadzi do powanego zagroenia ycia lub zdrowia personelu i / lub uszkodzenia sprztu.
x
x
x
x
x

Jeeli jakie zalecenia w instrukcji obsugi urzdzenia s niezrozumiae, prosimy o skontaktowanie si z przedstawicielem firmy Brooks
Instrument, aby wyjani problem.
Naley postpowa biorc pod uwag wszystkie ostrzeenia, uwagi i zalecenia umieszczone na produkcie lub doczone do niego.
Instalacj urzdzenia naley przeprowadzi zgodnie z zaleceniami zawartymi w instrukcji instalacji oraz z obowizujcymi lokalnymi i narodowymi
oznaczeniami. Wszystkie urzdzenia mona podcza wycznie do odpowiednich rde energii elektrycznej oraz cinienia.
Pierwsze czynnoci obsugowe: (1) Naley powoli wczy przepyw w instalacji. Nastpnie powoli otworzy zawory robocze tak, aby unikn
waha przepywu. (2) Naley teraz sprawdzi, czy nie wystpuj nieszczelnoci przy podczeniach wejciowym i wyjciowym miernika
przepywu. Jeeli nie ma adnych nieszczelnoci, mona zwikszy cinienie w instalacji do wartoci cinienia roboczego.
Przed przystpieniem do czynnoci serwisowych naley upewni si, e cinienie robocze jest odczone. Jeeli konieczna jest wymiana czci
zamiennych, naley zawsze stosowa czci zamienne specyfikowane przez firm Brooks Instrument a czynnoci ich wymiany powinien w
kadym przypadku dokonywa przeszkolony personel. Stosowanie nieautoryzowanych czci i procedur serwisowych moe niekorzystnie
wpyn na dziaanie produktu oraz zagrozi bezpieczestwu instalacji. Korzystanie z podobnie wygldajcych zamiennikw moe doprowadzi
do poaru, poraenia prdem lub nieprawidowego dziaania urzdzenia.
Naley upewni si, e wszystkie otwory urzdzenia s zamknite a osony umocowane na swoich miejscach, aby zapobiec obraeniom ciaa lub
poraeniu prdem personelu. Zalecenie to nie dotyczy przeszkolonego pracownika wykonujcego prace serwisowe lub konserwacyjne.
OSTRZEENIE: W przypadku miernikw przepywu cieczy, jeeli znajdujce si na nich zawory wejciowe i wyjciowe maj by z
jakiego powodu zamknite, to urzdzenie musi zosta cakowicie oprnione z ciekego medium. Niedopenienie tego zalecenia
moe doprowadzi do termicznego zwikszenia objtoci cieczy, co z kolei moe spowodowa uszkodzenie urzdzenia i obraenia
personelu.

Europejska dyrektywa dotyczca urzdze cinieniowych (PED)


Wszystkie urzdzenia cinieniowe pracujce przy cinieniu wewntrznym wzgldnym wikszym ni 0.5 bara i wielkoci powyej 25 mm lub 1 cala
podlegaj dyrektywie europejskiej dotyczcej urzdze cinieniowych (PED).
x
Rozdzia Specyfikacja techniczna niniejszej instrukcji zawiera zalecenia dotyczce dyrektywy PED.
x
Mierniki opisane w tej instrukcji s zgodne z dyrektyw EN 97/23/EC.
x
Wszystkie mierniki przepywu firmy Brooks Instrument nale do 1-szej grupy cieczy.
x
Urzdzenia pomiarowe o wielkoci powyej 25 mm lub 1 cala nale do kategorii I, II lub III dyrektywy PED.
x
Urzdzenia pomiarowe o wielkoci 25 mm lub 1 cala lub mniejsze podlegaj zaleceniom Uznanej Praktyki Inynierskiej (SEP).

Europejska dyrektywa dotyczca kompatybilnoci elektromagnetycznej (EMC)


Urzdzenia elektryczne / elektroniczne firmy Brooks Instrument posiadajce oznaczenie CE, przeszy pozytywnie testy pod ktem speniania przez
nich wymogw kompatybilnoci elektromagnetycznej (Dyrektywa EMC 2004/108/EC).
Jednake szczegln uwag naley powici przy doborze przewodw sygnaowych, ktre maj by stosowane z urzdzeniami ze znakiem CE.
Jako przewodu sygnaowego, dawic oraz zczy przewodu:
Firma Brooks Instrument dostarcza wysokiej jakoci przewody, ktre speniaj wymagania zawarte w specyfikacji dla certyfikatu CE.
Jeeli stosuje si wasne przewody sygnaowe, to powinny one by w caoci w peni ekranowane.
Zcza typu D lub okrge powinny zawiera metalowy ekran. Jeli to moliwe, naley stosowa metalowe dawice przewodu zapewniajce
mocowanie jego ekranu.
Ekran przewodu powinien by poczony z metalow oson lub dawic zapewniajc cakowite, dooklne ekranowanie na obu kocach przewodu.
Ekran przewodu powinien by uziemiony.
Zcza krawdziowe s standardowo niemetaliczne. Stosowane przewody musz by w peni ekranowane zgodnie z certyfikatem CE.
Ekran przewodu powinien by uziemiony.
Konfiguracja stykw jest podana w niniejszej instrukcji obsugi.

Wyadowania elektrostatyczne (ESD)


UWAGA: Urzdzenie zawiera czci elektroniczne podatne na uszkodzenia spowodowane adunkami elektrostatycznymi. Przy
obchodzeniu si z wewntrznymi podzespoami i czciami elektronicznymi naley przestrzega nastpujcych zasad
postpowania:
1. Naley odczy zasilanie od urzdzenia.
2. Osoba wykonujca czynnoci musi zosta uziemiona za pomoc opaski na przegubie doni lub w inny, bezpieczny sposb, zanim przystpi do
instalacji, wyjcia lub regulacji obwodw drukowanych lub innych wewntrznych podzespow elektronicznych urzdzenia.
3. Obwody drukowane naley transportowa w przewodzcym pojemniku. Pytki drukowane naley wyjmowa z opakowania ochronnego
bezporednio przed ich montaem. Wymontowane pytki naley niezwocznie umieci w opakowaniu ochronnym sucym do transportowania,
skadowania lub odsyania do producenta.
Uwagi:
Fakt, e urzdzenie zawiera czci nieodporne na wyadowania elektrostatyczne (ESD) jest rzecz normaln. Wikszo nowoczesnych urzdze
elektronicznych zawiera komponenty wykonane w technologii tlenkw metali (NMOS, SMOS itp.). Jak pokazuje praktyka, nawet niewielkie
wyadowanie elektrostatyczne moe uszkodzi lub zniszczy takie urzdzenie. Uszkodzone czci, nawet jeeli na pozr dziaaj poprawnie, szybko
doprowadzaj do nieprawidowej pracy urzdzenia.

A-14

Section C Essential Instructions

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

Models MT3809G & MT3810G

Portuguese

Instrues Bsicas
Ler antes de proceder!
A Brooks Instrument projecta, fabrica e testa os seus produtos de forma a satisfazer numerosas normas nacionais e internacionais. Estes
equipamentos devem ser instalados, utilizados e mantidos de forma adequada, e devem funcionar dentro da sua gama de utilizao. As instrues
seguintes devem ser, durante a instalao, uso e/ou manuteno dos equipamentos da Brooks Instrument, apreendidas e integradas no plano de
proteco e segurana no trabalho.
x
Para assegurar o desempenho adequado, a instalao, explorao, actualizao e/ou manuteno do equipamento deve ser realizada,
exclusivamente, por pessoal qualificado.
x
Antes de instalar, utilizar e/ou executar operaes de manuteno devem ser lidas todas as instrues do equipamento. No caso do presente
manual no ser apropriado procure, na capa traseira, o distribuidor mais prximo e contacte-o para obter informaes adicionais. Guarde este
manual para futura referncia.
ATENO: no sujeite o equipamento a condies fora das gamas de servios indicadas. Ao no respeitar esta advertncia poder
provocar avarias no equipamento e/ou danos pessoais.
x
Se as instrues deste manual no estiverem suficientemente claras, contacte o representante Brooks Instrument para esclarecer as suas
dvidas.
x
Tenha sempre presente todas as advertncias, apelos e instrues indicadas no equipamento e/ou fornecidas junto com o mesmo.
x
A instalao do equipamento dever ser efectuada cumprindo todas as instrues indicadas no manual assim como as normas e regulamentos
locais e nacionais vigentes. Ligue o equipamento exclusivamente a fontes de energia elctrica e/ou pneumtica adequadas.
x
Procedimento: (1) Pressurize lentamente o sistema. Abra lentamente as vlvulas para evitar variaes bruscas de caudal. (2) Verifique se h
fugas nas ligaes de entrada e sada do medidor de caudal. Se no detectar fugas, poder colocar o sistema presso de trabalho.
x
Antes de efectuar qualquer operao de manuteno verifique sempre a possibilidade do equipamento estar sob presso. No caso de ser
preciso substituir peas, estas devem ser as recomendadas pela Brooks Instrument e o trabalho de substituio dever ser efectuado por
tcnicos qualificados. Procedimentos e peas no conformes podero alterar o desempenho do equipamento, danific-lo ou colocar em risco a
sua segurana ou de outros. Substituir peas por outras no originais, meramente semelhantes, poder originar choques elctricos, fogo ou em
funcionamento inadequado.
x
Dever manter o equipamento intacto e fechado, verificando se as coberturas de proteco esto nos seus lugares devidos, de forma a evitar
choques elctricos e/ou danos pessoais, excepto no caso de se tratar de um tcnico qualificado e se estiver a executar trabalhos de
manuteno.
ATENO: Se por qualquer razo for necessrio fechar as vlvulas a montante e jusante do equipamento, tenha em ateno que o
mesmo dever ser previamente esvaziado do fludo que o atravessa. Esta obrigao surge do facto de existir a possibilidade de ocorrer
dilatao trmica do lquido, no interior do aparelho, podendo provocar danos pessoais ou materiais graves.

Directiva Europeia para equipamentos sob presso (PED)


Todos os equipamentos sujeitos a presso interior superior a 0,5 bar (g) e com calibre superior a 25 mm (1 polegada) esto sob a vigncia da
directiva europeia de equipamentos sob presso (PED).
x
O captulo "Dados tcnicos" do manual contm instrues relativas Directiva PED.
x
O caudalmetro objecto deste manual satisfaz a directiva 97/23/CE da UE.
x
Os caudalmetros Brooks pertencem ao grupo 1 de fluidos.
x
Os caudalmetros com calibre superior a 25 mm (1 polegada) pertencem s categorias PED I, II, ou III.

Os caudalmetros de 25 mm (1 polegada) ou menores seguem as Boas regras de engenharia (SEP).

Directiva Europeia sobre Compatibilidade electromagntica (EMC)


Os equipamentos (elctricos/electrnicos) da Brooks Instrument que tm a marcao CE passaram os testes comprovativos dos requisitos de
compatibilidade electromagntica (Directiva EMC nmero 2004/108/EC).
Todavia, ao utilizar os aparelhos compete-lhe a escolha dos cabos de sinal adequados para os equipamentos com marcao CE.
Qualidade dos cabos de sinal, bucins e conectores:
A Brooks Instrument fornece cabos de alta qualidade que cumprem todos os requisitos da marcao CE.
No caso de utilizar os seus prprios cabos de sinal, assegure uma blindagem a 100%.
Os conectores do tipo D ou "circular" tm que ser blindados por uma malha metlica. Se precisar de usar bucins, estes tm que permitir a
crimpagem da malha/blindagem do cabo.
A blindagem do cabo deve ser ligada ao corpo metlico ou bocal e assegurando a blindagem em 360. A blindagem deve terminar numa ligao
terra.
Os conectores ligados a cartes sero, em geral, no-metlicos. Os cabos utilizados devem ter fita de blindagem a 100% para satisfazer a marcao
CE.
A blindagem, dever terminar numa ligao terra.
Atribuio de pinos: Veja as instrues de operao anexas.

Descarga Electrosttica (ESD)


ATENO: Alguns componentes deste equipamento so susceptveis aco da electricidade esttica, podendo ficar danificados. Ao
remover, colocar ou manipular placas de circuitos electrnicos dever ter em ateno os seguintes procedimentos:
1. Desligar o equipamento da rede.
2. O utilizador, antes de qualquer interveno que envolva os cartes de circuitos ou outros dispositivos internos, ter que se ligar terra por meio
duma bracelete de pulso ou outro dispositivo adequado.
3. Os circuitos impressos devero ser transportados numa embalagem condutiva. Os cartes s devero ser retirados da embalagem protectora
imediatamente antes da sua insero. O carto retirado dever ser recolocado imediatamente na embalagem protectora que servir para o seu
transporte, armazenagem ou retorno a fbrica.
Observaes:
Tenha presente que este equipamento poder no ser o nico objecto capaz de ser portador de peas sensveis a descargas electrostticas (ESD).
Na maioria dos dispositivos electrnicos Brooks encontram-se peas de tecnologia de xidos metlicos (NMOS, SMOS, etc.). A experincia mostra
que at pequenas quantidades de electricidade esttica so capazes de danificar ou destruir esses dispositivos. Os componentes danificados,
embora inicialmente funcionem aparentemente bem, acabam por ter falhas prematuramente.

A-15

Section C Essential Instructions

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

GF100 Series
Romanian

Indicaii de referin
Citii-le nti pe acestea!
Brooks Instrument i proiecteaz, produce i testeaz produsele ntr-un mod ce respect un mare numr de standarde autohtone i internaionale.
Aceste instalaii trebuie amplasate, exploatate i ntreinute corespunztor, pentru ca n toate situaiile, domeniul lor de lucru s corespund operrii
normale. n ceea ce privete instalarea, operarea i ntreinerea produselor Brooks Instrument, indicaiile de mai jos trebuie respectate i trebuie
introduse in programul de protecia muncii.
x
Pentru garantarea prestaiei corecte, instalarea, operarea, actualizarea, programarea i ntreinerea produsului poate fi realizat doar de ctre
personal calificat.
x
Instruciunile de instalare ale produsului trebuie citite integral, nainte de punerea n serviciu i exploatarea sa. n msura n care ediia acestui
manual nu este cea adecvat, identificai pe ultima copert coordonatele distribuitorului local i pentru lmuriri suplimentare adresai-v acestuia.
Pstrai acest manual pentru referine ulterioare.
ATENIE: Nu utilizai instalaia n afara intervalului de funcionare indicat n instruciunile de operare. Nerespectarea acestui lucru se
poate solda cu rniri grave de persoane sau defectarea instalaiei.
x
n msura n care indicaiile crii mainii nu sunt suficient de lmuritoare, luai legtura cu reprezentantul Brooks Instrument pentru clarificarea
problemei.
x
Pstrai toate avertismentele, avizele i instruciunile livrate odat cu instalaia sau inscripionate pe aceasta.
x
Efectuai instalarea echipamentului n conformitate cu indicaiile de instalare corespunztoare, respectiv cu respectarea prevederilor naionale.
Echipamentul se conecteaz exclusiv la surse de energie electric i de presiune corespunztoare.
x
Succesiune: (1) Presurizai lent instalaia. Deschidei ncetul cu ncetul supapa de funcionare pentru evitarea fluctuaiilor de flux. (2) Controlai
dac nu sunt prelingeri la intrarea sau ieirea debitmetrului de branare. Dac nu sunt scurgeri, presurizai instalaia la presiunea de lucru.
x

naintea exploatrii/ ntreinerii, verificai neaprat dac conducta uzinal nu este sub presiune. n msura n care este nevoie de piese de
schimb, este neaprat necesar ca manevrarea pieselor de schimb s fie fcut de personal cu calificare profesional agreat de Brooks
Instrument. Utilizarea altor piese de schimb dect cele originale i liceniate poate avea efecte asupra performanelor instalaiei i asupra
siguranei sale n exploatare. Utilizarea de piese asemntoare de substituire poate avea ca rezultat pericol de incendiu i electrocutare.
x
n toate cazurile toate uile instalaiei trebuie s fie nchise, cutile de protecie s fie puse la locurile lor, pentru evitarea electrocutrii i rnirii de
persoane, exceptnd situaiile cnd un specialist efectueaz lucrri de ntreinere.
ATENIE: n cazul instalaiilor cu flux de fluide, dac din orice motiv este necesar nchiderea valvelor de intrare i ieire, limitrofe
instalaiei, instalaia trebuie complet golit. Neglijarea acestui lucru poate avea ca efect dilatarea termic a fluidului, care poate defecta
instalaia i poate produce rniri de persoane.

Directiva european pentru instalaiile sub presiune (PED)


Toate instalaiile i sistemele presurizate ce se afl sub presiuni interne ce depesc 0,5 mbar (g) i au mai mult de 25 mm sau 1 ol, cad sub
incidena normei europene corespunztoare (PED).
x
La capitolul "Date tehnice" din cartea mainii se gsesc indicaiile corespunztoare directivei PED.
x
Mijloacele de msurare menionate n cartea mainii corespund directivei 97/23/EK EU.
x
Toate debitmetrele Brooks corespund clasei 1 de fluide.
x
Mijloacele de msurare mai mari de 25 mm sau 1 ol corespund categoriei PED I, II sau III.
x
Mijloacele de msurare mai mici de 25 mm sau 1 ol se conformeaz practicii inginereti acceptate (SEP).

Directiva european privitoare la compatibilitatea electromagnetic (EMC).


Instalaiile (electrice /electronice) ce poart marca Brooks Instrument CE ndeplinesc cu succes cerinele testelor de verificare ale compatibilitii
electromagnetice (Cf. directivelor europene EMC cu nr. 2004/108/EC).
n acelai timp trebuie acordat o atenie deosebit la alegerea cablurilor de semnalizare utilizate pentru instalaiile ce poart marcajul CE.
Calitatea cablurilor de semnalizare, a legturilor prin cablu i a conectoarelor:
Brooks Instrument ofer cabluri de calitate ridicat, care corespund cerinelor calitative ale CE.
n msura n care folosii cabluri proprii, trebuie alese acelea care sunt 100% ecranate i prevzute cu filtre.
Conectoarele D sau cele circulare trebuie sa dispun de ecrane metalice. n caz de nevoie trebuie folosite conectoare metalice pentru montarea
filtrelor de cablu.
Filtrul de cablu trebuie conectat la carcasa metalic sau manon i n ambele cazuri trebuie asigurat ecranarea la 360. Ecranarea trebuie terminat
cu o legare la pmnt.
Conform standardului, conectoarele aparinnd plcilor electronice nu sunt metalice. Cablurile folosite trebuie s fie 100% ecranate i prevzute cu
filtre pentru a corespunde clasificrii CE.
Ecranarea trebuie terminat cu o legare la pmnt.
Configuraie de contact: Vezi instruciunile de operare ataate.

Descrcare electrostatic (ESD)


ATENIE: Instalaia include piese care sunt predispuse la defectare sub influena electricitii statice. Trebuie respectate metodele
corespunztoare de extragere, instalare sau alte manipulri ale circuitelor electronice.
Procedur de manipulare:
1. Instalaia trebuie scoas de sub tensiune.
2. naintea de inserarea, scoaterea sau reglarea vreunei cartele electronice, sau a altui dispozitiv intern, persoana trebuie s se lege la pmnt cu
banda pentru articulaia minii sau alte dispozitive de siguran disponibile pentru acest scop.
3. Cartelele cu cablaje electronice imprimate trebuie transportate n ambalaje anti-electrostatice (conductoare). Cartelele se pot scoate din ambalaj,
doar nemijlocit naintea amplasrii lor. Cartela demontat trebuie pus nentrziat n ambalajul de protecie n vederea transportrii, a depozitrii
sau returnrii la productor.
Observaii:
n echipamente se gsesc adesea componente sensibile la descrcare electrostatic (ESD). Majoritatea echipamentelor moderne includ componente
electronice realizate n tehnologie metal-oxid semiconductor (NMOS, SMOS, etc.) Experiena a dovedit c acestea pot fi afectate sau deteriorate chiar
de energii electrostatice de slab intensitate. Componentele defectate, cu toate c n aparen sunt funcionale, duc n timp la defeciuni incipiente.

A-16

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

Section C Essential Instructions

Models MT3809G & MT3810G

Slovak

Zkladn prkazy
Preta pred intalciou!
Brooks Instrument svoje vrobky projektuje, vyrba a testuje takm spsobom, aby tieto vyhoveli domcim aj medzinrodnm normm. Tieto
zariadenia je potrebn predpsanm spsobom intalova, prevdzkova a udriava, na zabezpeenie ich spoahlivej a normlnej prevdzky v celom
pracovnom rozsahu. Niie uveden prkazy je potrebn dodriava a zaleni do programu bezpenostnch predpisov v priebehu intalcie,
prevdzky a drby vrobkov Brooks Instruments.

V zujme zabezpeenia vyhovujceho vkonu intalciu, prevdzku, programovanie, aktualizciu a drbu zariaden m vykonva vlune
odborne kvalifikovan personl.

Pred intalciou, prevdzkou a servisu zariaden je potrebn preta vetky prkazy. Ak tto prruka nie je sprvna, tak na zadnej strane treba
njs miestneho distribtora, kontaktova ho pre alie informcie. Pre neskorie informcie uschovajte prruku.
UPOZORNENIE: Neprevdzkova zariadenie v rozsahu mimo rozsahu uvedenom v prevdzkovej prruke. Poruenie tohto oznmenia
me ma za nsledok ak ujmu na zdrav a vedie k pokodeniu zariadenia.
x
Ak prkazy v nvode nie s jednoznan, kontaktujte zstupcu Brooks Instrument na objasnenie problmov.
x
Dodrujte vetky upozornenia, prkazy a usmernenia uveden na zariaden, alebo s nm dodan.
x
Zariadenia intalujte poda nvodu uvedenho v prkaze na intalciu, v slade s miestnymi a nrodnmi predpismi. Zariadenie pripojte vlune
len na vyhovujci elektrick a tlakov zdroj
x
Postup: (1) Pomaly natlakujte systm. Prevdzkov ventil otvorte pomaly na zamedzenie kolsania prietoku. (2) Prekontrolujte tesnos vstupnho
a vstupnho zapojenia prietokomeru. Ke nie je presakovanie, spoje s tesn, naplni systm na prevdzkov tlak.
x
Pred vykonvanm servisnch prc kontrolova, i systm nie je pod tlakom. V prpade, e je potrebn vmena siastky, vmenu dielov,
urench Brooks Instrument mus vykona kvalifikovan osoba. Pouitie nepovolench dielov a vykonvanie nepovolench aktivt ohrozuj
bezpenos prevdzvky a maj negatvny vplyv na vkon zariadenia. Nahradenie siastok len podobnmi komponentmi me ma za nsledok
poiar, raz elektrickm prdom alebo nedostaton funkciu zariadenia
x
Vetky ochrann kryty, dvierka zariadenia maj by zatvoren na zabezpeenie ochrany proti razu elektrickm prdom a proti poraneniam
obsluhy. Vnimku tvor vykonvanie drby kvalifikovanm odbornkom.
UPOZORNENIE: Pri zariadeniach s prietokom kvapaln, ke z akhokovek dvodu je nutn uzavrie vstupn a vstupn ventily,
zariadenie je potrebn plne vyprzdni. Zanedbanie vypania m za nsledok pokodenie zariadenia s monosou zranenia obsluhy z
dvodu tepelnej rozanosti nplne.

Eurpska smernica vzahujca sa na tlakov zariadenia (PED)


x
x
x
x
x
x

Vetky zariadenia s vym vntornm pretlakom ako 0,5 bar (g), a vieho rozmeru ako 25 mm alebo 1 anglick palec, podliehaj pod
Eurpsku smernicu vzahujcu sa na tlakov ndoby (PED).
Kapitola "Technick daje" nvodu na obsluhu obsahuje prkazy vzahujce sa na smernicu PED.
Meracie prostriedky uveden v nvode na obsluhu vyhovuj smernici 97/23/ES E.
Vetky prietokomery Brooks patria do 1. skupiny kvapaln.
Meracie prstroje presahujce rozmery 25 mm alebo 1 " spaj I., II., alebo III. kategriu PED.
Meracie prstroje menie alebo rovn ako 25 mm alebo 1 " zodpovedaj zauvanej meracej praxi (SEP).

Eurpska smernica vzahujca sa na elektromagnetick kompatibilitu (EMC)


Elektrick / elektronick zariadenia Brooks Instrument, ktor si zaslili znaku CE, spene splnili skobn testy poiadaviek elektromagnetickej
kompatibility (smernica EMC . 2004/108/EC).
Pritom treba venova zvltnu starostlivos na vber signlnych kblov zariadenia, s oznaenm CE.
Kvalita signlnych kblov, kblovch spojov a prpojov:
Brooks Instrument ponka vysoko kvalitn kble, ktor spaj poiadavky kvalitatvneho zaradenia CE.
Ak pouijete vlastn signlne kble, maj ma 100%-n tienenie, s plnm filtrovanm.
Prpojky "kruhov" alebo tvaru "D" maj ma kovov tienenie. V prpade potreby treba poui kovov kblov spojky k upevneniu kblovho filtra.
Kblov filter treba pripoji ku kovovmu telesu alebo k puzdru, na oboch stranch zabezpei tienenie v kruhu 360 . Tienenie m by ukonen
uzemnenm.
Prpojky vedce ku kartm poda noriem s nekovov. Pouit kble, pre vyhovenie predpisom CE musia ma 100%-n filtrovanie tienenm.
Tienenie m by ukonen uzemnenm.
Konfigurcia kontaktov: Vi priloen nvod na obsluhu. .

Elektrostatick vboj (ESD)


UPOZORNENIE: Prstroj obsahuje siastky, ktor mu by pokoden od elektrostatickch nbojov. Pri monti, odstraovan alebo
inej drby vntornch obvodovch kariet je potrebn dodriava prslun postupy.
Postup oetrovania:
1. Zariadenie odpoji od napjania.
2. Osoba vykonvajca drbu m by uzemnen uzemujcim nramkom, alebo inm, na tto prcu vyhovujcim spsobom pred vykonvanm
intalcie, demonte a nastavenia obvodovch kariet alebo inho vntornho prostriedku.
3. Karty obvodov sa musia prepravova v elektricky vodivom balen. Karty sa mu vybera z ochrannho obalu vlune len tesne pred montou,
zasunutm! Vybran kartu okamite treba umiestni do ochrannho obalu, urenho pre dopravu, skladovanie, alebo pre sptn prepravu do
vrobnho zvodu.
Poznmky:
Existencia prvkov, ktor s citliv na elektrostatick vboje (ESD) v prstroji je astm javom. U viny modernch elektronickch prostriedkov s
pouit prvky s technolgiou oxidu kovov (NMOS, SMOS, at.). Sksenosti dokazuj, e aj nepatrn elektrostatick vboje pokodzuj, zniia tieto
prostriedky. Pokoden siastky, aj ke zdanlivo pracuj bez chyby, odkazuj na vznikajce poruchy.

A-17

Section C Essential Instructions

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

GF100 Series
Slovene

Osnovna navodila
Najprej preberite jih
Brooks Instrument tako konstruira, izdeluje in terstira svoje izdelke, da oni ustrezajo tevilnim domaim in mednarodnim standardom. Te naprave se
morajo ustrezno instalirati, koristiti in vzdrevati, da vsekakor delajo ustrezno normalnom podroju funkcioniranja. Naslednjih navodil se mora drati in
potrebno je vgraditi v program varstva pri delu pri instaliranju, korienju in vzdravanju izdelkov proizvajalca Brooks Instrument.
x
Za jemstvo ustreznega uinka naj izkljuno strokovno osobje opravlja instaliranje, korienje, osveenje, programiranje in vzdrevanje izdelka.
x
Potrebno je prebrati vsa navodila pred instaliranjem, korienjem in servisiranjem izdelka. V kolikor ta prironik ni ustrezna publikacija, na zadnji
strani poiite lokalnega distributerja in za nadaljnje informacije stopite z njim v kontakt. Prihranite ta prironik za pozneje informacije.
OPOMBA: Ne koristite napravo izven podroja iz navodila za uporabo. Prekrek tega lahko privede do osebnih pokodb ali okvare
naprave.
x
V kolikor navodila prironika niso enosmiselna, stopite v kontakt z zastopnikom drutva Brooks Instrument, da razistite problem.
x
Drite se vseh opozoril, povabil, navodil, ki so navedena na napravi ali skupaj z njim izporoena.
x
Napravo instalirajte shodno navodilom in relevantnim lokalnim in nacionalnim navodilom iz navodila za instalirenje.. Izdelek spojite izkljuno na
ustrezni izvir elektrike in obskrbo tlaka.
x
Proces: (1) Poasi poloite sistem pod tlak. Poasi odpirajte pogonske ventile zaradi izogibanja nihanju pretoka. (2) Preverjajte, ima li curljanja pri
povezu vstopa in izstopa merilca toka. e ni curljanja, naplonite sistem na pogonski tlak.
x
Pred servisom vsekakor preverjajte, ali je pogonski vod pod tlakom. e je potreben rezervni del, svekakor strokovna oseba mora rokovati z
rezervnimi deli odrejenimi od Brooks Instrument. Nedovoljeni rezervni deli in dejavnosti lahko vplivajo na uinek izdelka, oziroma ugroavajo
varnostni pogon. Sprememba samo z podobnimi rezervnimi deli lahko ima za posledico poar, nevarnost elektinega oka ali nedovoljno
funkcioniranje.
x
Vsa vrata naprave svekakor morajo biti zaprta, zaitni ovitki morajo biti na svojem mestu zaradi izogibanja oku in osebnim pokodbam, razen,
e na njej strokovnjak opravlja dela vzdrevanja.
OPOMBA: V primeru naprave za tok tekoine, e je iz katera koli razloga potrebno zapreti vstopne in izstopne ventile pri napravi,
naprava se mora celotno izprazniti. Zamuda tega lahko provzroi toplotno proirjenje tekoine, kar pokoduje napravo in lahko povrzoi
osebno pokodbo.

Europske smernice za naprave ohranjanja pritiska (PED)


Vsaka naprava ohranjanja pritiska z notranjim pritiskom ve od 0,5 bara (g) in vejim od 25 mm ali 1 palca sodi pod Europskim smernicam za
ohranjanje pritiska (PED).
x
Poglavje prironika Tehnini podatki vsebuje navodila, ki se nanaajo na smernice PED.
x
Merilni instrumenti navajani v prironiku ustrezajo smernicam EU tevilke 97/23/EK
x
Vsaki merilec pretoka Brooks sodi v skupino tekoin tevilke 1.
x
Merilni instrumenti veji od 25 mm ali 1 palca ustrezajo kategriji PED I, II, ali III.
x
Merilni instrumenti manji od 25 mm oziroma 1 palca ali manji sledijo sprejeti inenjeski praksi (SEP).

Evropska smernica za Elektromagnetno kompatibilnost (EMC)


Naprave Brooks Instrument, ki so zasluile CE oznabo (elektrini/elektronski) so uspeno izpolnile testove zahteve o elektromangnetni
kompatibilnosti (t. 2004/108/EC smernice EMC) .
Obenem mora se posebno pozornost posvetiti na izbiro signalnega kabla, ki se uporabljajo za naprave z oznabo CE.
Kakaovost signalnih kablov, povezav kablov, prikljukov:
Brooks Insturment nuja kabele visoke kakovosti, ki ustrezajo zahtevam kvalificiranja CE.
V kolikor se uporbalja lastni signalni kabel, treba je izbrati, ki za 100 % senco v polni meri filtriran.
Prikljuki D olika kroga morajo biti zaseneni kovonskom sencom. e bo potrebno, uporabite kovinske povezave za pritrditev filtra kabla
Filtar kabla je treba prikljuiti na kovinsko ohije ali plat in na obe polovici je treba seniti v 360. Senenje mora zavriti v ozemljitvi.
Prikljuki, ki sodijo karitcam normalno nisu kovinskik. Korieni kabli moraju biti filtrirani s 100 % senco, da odgovarjajo CE kvalifikaciji.
Sencanje mora imat konec v ozemljitvi.
Konfiguracija stika: Vidi priloeno navodilo za uporabo.

Elektrostatino izpranjenje (ESD)


OPOZORILO: Naprava vsebuje take sestavne dele, ki so naklonjeni k pokodbi od statine elektrike. Treba se je drati ustreznih
postopkov pri odstranjevanju, vlogu ali drugega rokovanja kartic in sredstev notranjih tokovnih krogov.
Posotpek rokovanja:
1. napravo je treba izklopiti iz elektrinega toka.
2. Osebo je treba ozemljiti zapestnim paekom ali z drugim varnostnim in za ta namen primernim sredstvom preden bi instalirala, jemala, ali
nastavila kartico za tokovni krog ali drugo notranjo opremo.
3. Tiskane kartice tokovnega kroga je treba izporoiti v konduktivni embalai. Karte so odstranjive izkljuno neposredno pred vlogom iz zaitne
embalae ven. Izjemljeno kartico nemudoma staviti v zaitno embalao za gibanje, skladienje ali poiljanje nazaj.
Opombe:
Nije poseben pojav, da se v napravi nahajajo obutljivi sestavni deli na elektrostatino pranjenje (ESD). V veem delu modernih elektronskih
sredstev nahajajo se sestavni deli na kovinsko oksidno tehnologijo (NMOS, SMOS itd..). Izkustva potrjujejo, da majhna statina elektrika more
pokodovati ali uniiti ta sredstva. Okodovani rezervni deli, e na videz dobro delajo, nakaejo zaetno napako.

A-18

Section C Essential Instructions

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

Models MT3809G & MT3810G

Spanish

Instrucciones bsicas
Lalos primero!
El Brooks Instrument proyecta, fabrica y prueba sus productos de manera que stos respondan a numerosas normas nacionales e internacionales. Dichas
instalaciones deben ser emplazadas, operadas y mantenidas adecuadamente, para que puedan marchar de todas formas en conformidad con el alcance
normal de funcionamiento. Las siguientes instrucciones deben cumplirse y incorporadas en su programa de seguridad cuando instalando, operando y
mantenimiento los productos Brooks Instrument.
x
Para asegurar el adecuado rendimiento, para instalar, operar, actualizar, programar y mantener tiene que realizarse exclusivamente por una persona
calificada.
x
Antes de la instalacin, operacin y servicio del producto leer todas las respectivas instrucciones. Si el presente manual no es la adecuada publicacin,
busque al distribuidor local que figura en la contraportada y pngase en contacto con l para obtener informaciones. Guarde el presente manual para
tener informaciones tambin en el futuro.
ATENCIN: No haga funcionar los equipos fuera del rango indicado en las instrucciones de funcionamiento. El incumplimiento de estas ltimas
puede conducir a graves daos personales o a la avera del equipo.
x
Si las instrucciones del manual no son evidentes, pngase en contacto con el representante de Brooks Instrument para aclarar el problema
x
Observar todas las alertas, advertencias e instrucciones indicadas en el equipo o suministradas con el mismo.
x
Instale su equipo en conformidad con las recomendaciones indicadas en las respectivas instrucciones de instalacin y con las pautas de las normas
vigentes locales e internacionales. Conectar el producto exclusivamente a la adecuada fuente elctrica y presin.
x
Proceso: (1) Colocar lentamente flujo en el sistema. Abrir lentamente las vlvulas de proceso para evitar oscilacin del flujo. (2) Verificar si hay fuga
alrededor de las conexiones de entrada y salida del flujmetro, Si no hay, llenar el sistema con la presin de operacin.
x
Antes de efectuar el servicio, verificar si hay presin o no en la tubera de la red. Si se requiere realizar un recambio de piezas, solamente el personal
calificado puede manipular las piezas de repuesto determinadas por Brooks Instrument Las piezas y operaciones no autorizadas pueden afectar el
rendimiento del producto o arriesgar el funcionamiento seguro. El recambio realizado con piezas slo similares pueden traer como consecuencias
incendios, choques elctricos o funcionamiento bajo.
x
Todas las puertas de la instalacin deben estar cerradas, las cubiertas de proteccin tienen que hallarse en el debido sitio con el fin de evitar los daos
personales y los choques elctricos, salvo cuando un especialista efecta el mantenimiento.
ADVERTENCIA: En caso de instalaciones que circulen lquido, si por cualquier razn se hubiera de cerrar las vlvulas de entrada y salida
situadas al lado del equipo, dichas instalaciones debern ser completamente vaciadas. La omisin de esto ltimo puede provocar la dilatacin
trmica del lquido, lo que puede daar al equipo y conducir a daos personales.

Directriz Europea de los Equipos de Presin (PED)


Todos los equipos de presin, con una presin interna que supere a 0,5 bar (g) con tamao mayor a 25 mm o 1 pulgada entran el mbito de la Directriz
Europea de los Equipos de Presin (PED).
x
El captulo Datos Tcnicos del manual incluye las instrucciones respecto a las directivas de PED
x
Los instrumentos de medicin indicados en el Manual responden a las EN directivas 97/23/EC.
x
Todos los flujmetros Brooks pertenecen a la categora 1 del grupo de fluidos.
x
Los instrumentos de medicin ms grandes que 25 mm o 1 pulgada estn en conformidad con las categoras I, II o III de PED
x
Los instrumentos de medicin ms pequeos que 25 mm o 1 pulgada siguen la Prctica Aceptada de Ingeniera (SEP).

Directriz Europea respecto a la Compatibilidad Electromagntica (EMC)


Las instalaciones de Brooks Instrument (elctricas/electrnicas) merecedores de la categora CE cumplieron con xito las pruebas que verifican las exigencias
de la compatibilidad electromagntica (directiva de EMC 2004/108/EC).
Al mismo tiempo se ha de prestar una especial atencin en la seleccin de los cables de seal, utilizados con los equipos marcados con CE.
Calidad de los cables de seal, piezas de unin de cable y conectores:
El Brooks Instrument ofrece cables de alta calidad, que responden a los requerimientos de calificacin CE.
Si se utiliza cable propio de la firma, se ha de elegir uno que sea completamente filtrado con blindaje de 100%.
Las piezas de unin de forma D o circular deben ser blindadas mediante blindaje metlica. Si es necesario, aplicar piezas de unin de metal para sujetar el
filtro de cable.
Conectar el filtro de cable a la caja o manguito de metal blindndolo en ambas caras en 360.
El blindaje debe terminar en tierra.
Los conectores que pertenecen a las tarjetas normalmente no son metalizados. Los cables utilizados deben ser filtrados con una blindaje de 100% para
responder a la calificacin CE.
El blindaje debe terminarse en tierra.
Configuracin de contacto: Vase Instrucciones de operacin adjuntas.

Descarga Electroesttica (ESD)


PRECAUCIN: El aparato incluye piezas electrnicas que son susceptibles a los daos provocados por la electricidad esttica. Observar los
adecuados procesos para remover, instalar o manipular las tareas y medios de circuitos elctricos internos
Proceso de operacin:
1. Desconectar la fuente elctrica de la unidad.
2. La persona debe ponerse a tierra mediante una palanca acodada o por otro medio seguro y apropiado para dicho fin antes de instalar, sacar o ajustar el
circuito impreso elctrico u otro medio interno.
3. El circuito impreso debe ser transportado en embalaje conductivo. Las tarjetas no pueden sacarse de la cubierta protectora exclusivamente directamente
antes de la instalacin. Las tarjetas desmontadas deben colocarse sin tardar en el embalaje protector utilizado para manipulacin, almacenamiento o
devolucin a la fbrica.
Notas:
Esto equipo no es el nico contenido de piezas susceptibles a la descarga electroesttica (ESD). En la mayora de los medios electrnicos modernos se
encuentran piezas fabricadas por tecnologa de xido metlico. (NMOS, SMOS etc.). Las experiencias confirman que incluso una mnima electricidad esttica
puede daar o destruir dichos medios. Las piezas averiadas, aunque funcionen aparentemente bien, indican una falla inicial.

A-19

Section C Essential Instructions

GF100 Series

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

Swedish

Vsentliga anvisningar.
Ls detta innan du fortstter !
Brooks Instrument konstruerar, tillverkar och testar sina produkter med syfte att uppfylla alla nationella och internationella standarder. Dessa produkter mste
installeras p rtt stt, handhas och underhllas fr att de skall fungera kontinuerligt enligt deras normala specifikation. De fljande anvisningarna br fljas
och integreras till Ert skerhetsprogram varje gng nr Brooks Instruments produkter installeras, handhas och underhlls.
x
Fr att garantera angiven funktion, anvnd kvalificerad personal till att installera, handha, uppgradera, programmera och serva produkten.
x
Ls alla instruktioner innan produkten installeras, startas upp och underhlls. Om du finner att denna instruktionshandbok inte r den rtta
instruktionsboken, titta p i slutet av prmen fr information om hur man kan kontakta lokala representanter. Spara denna instruktions manual fr senare
behov.
VARNING: Kr inte detta instrument utanfr dess specifikationer som r angiven i Instruktionsboken.
Undvikande att ta denna varning
kan leda till allvarliga personliga skador och / eller skada utrustningen.
x
Om du inte frstr ngon av dessa instruktioner, kontakta din representant fr Brooks Instrument fr klarlggande.
x
Flj alla varningar och instruktioner som fljer med leveransen av denna produkt.
x
Installera din utrustning p sttet som anges i den gllande handbokens installationsanvisningar och enligt tillmpliga lokala och nationella freskrifter.
Koppla varje produkt till freskriven strm- och tryckkllan.
x
Igngsttning: (1) Koppla lngsamt p flde i systemet. ppna processventiler sakta fr att undvika fr hga flden. (2) Kontrollera lckor vid mtarens
anslutningar fr in- och utlopp. Om inget lckage frekommer, kr systemet upp till drifttrycket.
x
Kontrollera att processledningens tryck r bortkopplat fre service. I fall det behves kompletteras med nya delar, se till att komponenter freskrivna av
Brooks Instrument anvnds. Samt att kvalificerad personal utfr arbetet. Ej rekommenderade komponenter och tgrder kan pverka produktens
prestanda och stta din driftskerhet p spel. Felaktiga ersttningar kan orsaka eld, elektriska skador samt felaktig funktion.
x
Se till att anordningens kpor och skyddslock ligger p sin plats med syfte att frebygga elektriska kontakt och personliga skador; det enda undantag
gller nr underhllsarbete utfrs av kvalificerad personal.
VARNING: I fall av - fldesmtare / regulatorer fr vtskor: Ifall ventiler fre och efter skall stngas av, mste alla ledningar tmmas p all
vtska. Att ej tmma ledningar alt koppla bort trycket kan gra s att vtskans vrmeutvidgning kan sprcka / skada utrustningen och orsaka
personliga skador.

European Pressure Equipment Directive (PED) - (Rdets Direktiv 99/36/EG av den 29 april 1999[1] om transportabla
tryckbrande anordningar)
Alla utrustning fr tryck med ett tryck ver 0.5 Bar(g) bar och strre demensioner n 25 mm eller 1" (inch) faller under Tryck direktiv 99/36/EG av den 29 april
1999[1] om transportabla tryckbrande anordningar - PED.
x
Den hr Instruktionsbokens Sektion Specifikation innehller anvisningar gllande PED Direktivet.
x
Mtare som beskrivs i denna Instruktionsbok r i verensstmmelse med EN Direktivet 97/23/EC.
x
Brooks Instruments alla fldesmtare faller under fldesgrupp nr. I.
x
Mtare som r strre n 25 mm eller 1" (inch) verensstmmer med PED kategorier I, II eller III.
x
Mtare p 25mm eller 1" (inch) eller mindre faller under Sound Engineering Practice (SEP) (God Teknisk Praxis).

European Electromagnetic Compatibility (EMC) - Elektromagnetisk kompatibilitet


Brooks Instrument (elektriska/elektroniska) CE-mrkta anordningar har redan genomgtt ett framgngsrikt prov enligt regleringar under Electromagnetic
Compatibility (EMC directive 2004/108/EC). Man mstre dock gna srskild uppmrksamhet till valet av signalkabeln som skall anvnds fr CE-mrkta
anordningar.
Signalkablars, packboxars och kontakdons kvalitet:
Brooks Instrument levererar hgkvalitativa kablar som verensstmmer med specifikation fr CE-intygade produkter.
Om man anvnder sin egen signalkabel, d br man anvnda en kabel som r fullstndigt skrmad med en 100% avskrmning.
D eller Cirkelformiga kontakdon skall vara skrmade med metalliska avskrmningar. Om det r anvndbart, br metallpackboxar som ger en bra
fastspnning fr kabelskrmar anvndas.
Kabelavsrmningen mste kopplas till den metalliska skrmande anordningen eller packboxen och skrmas vid bda ndar runt omkring. Avskrmningens
avspnning mste jordas.
Card Edge Kontaktdon r icke metalliska. Fr att verensstmma med krav p CE-intyg, skall de kablarna som anvnds vara skrmade med 100%
skrmning.
Skrmningen mste jordas.
Vad gller stiftkonfigurationen: Se den bifogade Instruktionshandboken.
ESD (Elektrostatiska urladdningar)
OBS: Denna utrustning innehller elektroniska komponenter som r lttpverkade av skada orsakad av statisk elektricitet. Lmplig
hanteringsprocedur mste fljas nr man tar bort, installerar eller p ngot annat stt hanterar inre kretskort eller andra anordningar.
1. Strm till enheten mste kopplas frn.
2. Personalen mste jordas med hjlp av ett armband eller ngot annat skert medel innan ngot kretskort eller andra inre anordningar installerras, tas bort
eller justeras.
3. Kretskort mste transporteras i en speciell frpackning fr elektronik. Kort skall ej tas bort frn deras skydsskpa innan man skall installera dem. De
borttagna korten br omedelbart lggas i speciell frpackning fr transport, lagring eller terlmnande till fabriken.
Anmrkningar:
Dessa instrument r ej unika vad gller dess ESD (Elektrostatiska urladdningar) - knsliga komponenter. De flesta samtida konstruktioner innehller
komponenter som utnyttjar metalloxid teknologi (NMOS, SMOS, o.s.v.). Erfarenhet har visat att ven sm mngder av statisk elektricitet kan skada eller
frstra dess komponenter. Skadade komponenter - ven om de annars verkar fungera ordentligt har ofta en kortare livslngd. .

A-20

Installation and Operation Manual


X-TMF-GF100-Series-MFC-eng
Part Number: 541B137AAG
March, 2013

GF100 Series

THIS PAGE WAS


INTENTIONALLY
LEFT BLANK

Installation and Operation Manual


X-TMF-GF100-Series-eng
Part Number: 541B137AAG
March, 2013

GF100 Series
LIMITED WARRANTY

Seller warrants that the Goods manufactured by Seller will be free from defects in materials or workmanship under normal use
and service and that the Software will execute the programming instructions provided by Seller until the expiration of the
earlier of twelve (12) months from the date of initial installation or eighteen (18) months from the date of shipment by Seller.
Products purchased by Seller from a third party for resale to Buyer (Resale Products) shall carry only the warranty extended
by the original manufacturer.
All replacements or repairs necessitated by inadequate preventive maintenance, or by normal wear and usage, or by fault of
Buyer, or by unsuitable power sources or by attack or deterioration under unsuitable environmental conditions, or by abuse,
accident, alteration, misuse, improper installation, modification, repair, storage or handling, or any other cause not the fault of
Seller are not covered by this limited warranty, and shall be at Buyers expense.
Goods repaired and parts replaced during the warranty period shall be in warranty for the remainder of the original warranty
period or ninety (90) days, whichever is longer. This limited warranty is the only warranty made by Seller and can be
amended only in a writing signed by an authorized representative of Seller.
BROOKS SERVICE AND SUPPORT
Brooks is committed to assuring all of our customers receive the ideal flow solution for their application, along with
outstanding service and support to back it up. We operate first class repair facilities located around the world to provide
rapid response and support. Each location utilizes primary standard calibration equipment to ensure accuracy and reliability
for repairs and recalibration and is certified by our local Weights and Measures Authorities and traceable to the relevant
International Standards.
Visit www.BrooksInstrument.com to locate the service location nearest to you.
START-UP SERVICE AND IN-SITU CALIBRATION
Brooks Instrument can provide start-up service prior to operation when required.
For some process applications, where ISO-9001 Quality Certification is important, it is mandatory to verify and/or (re)calibrate
the products periodically. In many cases this service can be provided under in-situ conditions, and the results will be traceable
to the relevant international quality standards.
CUSTOMER SEMINARS AND TRAINING
Brooks Instrument can provide customer seminars and dedicated training to engineers, end users and maintenance persons.
Please contact your nearest sales representative for more details.
HELP DESK
In case you need technical assistance:
888 275 8946
USA
Europe
+49 351 215 2040
Japan
+81 3 5633 7100
Korea
+82 31 708 2521

Taiwan
China
Singapore

+886 3 5590 988


+86 21 5079 8828
+6297 9741

Due to Brooks Instrument's commitment to continuous improvement of our products, all specifications are subject to change
without notice.
TRADEMARKS
Brooks ........................................................... Brooks Instrument, LLC
DeviceNet ...................... Open DeviceNet Vendors Association, Inc.
Hastelloy ............................................................ Haynes International
MultiFlo ......................................................... Brooks Instrument, LLC
ODVA ............................. Open DeviceNet Vendors Association, Inc.
VCR .................................................................................... Cajon Co.

You might also like