Mastering Temporary Invasive Cardiac Pacing: Clinical
Mastering Temporary Invasive Cardiac Pacing: Clinical
Mastering Temporary Invasive Cardiac Pacing: Clinical
Mastering Temporary
Invasive Cardiac Pacing
Heart blocks
Drug-refractory dysrhythmia
Cardiovascular surgery
Authors
Devorah Overbay is a clinical transplant coordinator at Oregon Health & Science University in Portland, Ore. She is a former critical care and cardiovascular clinical nurse specialist.
Laura Criddle is a doctoral student at Oregon Health and Science University.
To purchase reprints, contact The InnoVision Group, 101 Columbia, Aliso Viejo, CA 92656. Phone, (800) 8092273 or (949) 362-2050 (ext 532); fax, (949) 362-2049; e-mail, reprints@aacn.org.
Sinus
Node
Atrioventricular
Node
Cardiac Anatomy
and Physiology
Bundle of His
Purkinje
Fibers
Microscopically,
myocardial tissue is difFigure 1 Cardiac conduction system.
ferentiated by the functions of its various cell
types. Working myocarthroughout the entire circulatory
dial cells contract, providing pumpsystem.3
ing forces; clusters of specialized
pacemaker cells initiate electrical
Many intrinsic and extrinsic facimpulses; and the Purkinje fibers
tors influence the genesis and the
provide rapid conduction of these
propagation of cardiac impulses.
3
impulses. The sinoatrial node,
Dysrhythmias may arise from abnormal initiation or conduction of
located high in the right atrium, is
impulses or from both.3 Bradydysthe cluster of cells that initiates
normal cardiac stimulation and
rhythmias can result from electrolyte
serves as the primary pacemaker.
imbalances, the toxic effects of drugs,
These signals then travel across the
inherent abnormalities in the conducatrium to the atrioventricular node,
tion system, ischemia, or myocardial
located close to the septal leaflet of
damage.2 By sustaining a rate suffithe tricuspid valve. Conducting
cient to allow filling and emptying of
fibers from the atrioventricular
the hearts chambers, artificial
node converge in the bundle of His,
mechanical pacemakers can be a lifeallowing rapid transmission of
saving adjunct for maintaining an
impulses to the Purkinje fibers of
adequate cardiac output.
the ventricles (Figure 1). The result
of this highly coordinated impulse
Components of Pacing
propagation is a cardiac contraction
The mechanics of pacing
that efficiently pumps blood
involves several fundamental com-
Pacemaker Concepts
The four basic concepts of pacemaker functioning are connection,
output, capture, and sensitivity.
Connection
The connection between the pacemaker generator and the heart is
made through either unipolar or bipolar electrode wires.1 In a unipolar
system, only the negative electrode
is in direct contact with the heart. In
a bipolar system, both negative and
positive electrodes lie within the
heart. Pacemakers can be either
unipolar or bipolar. Distinguishing
between the negative and positive
electrodes is important so that the
wires are connected appropriately to
the pulse generator.6
The 2 types of invasive temporary
pacing are epicardial and transvenous.3 The transvenous category
also includes devices that combine a
specialized pulmonary artery catheter
with a pacemaker.1 Transvenous pacing involves a pulse generator, which
is externally connected to 2 electrode
wires, threaded through a large vein
(generally the subclavian or internal
jugular) into either the right atrium
or the right ventricle.1 These wires
directly contact the endocardium
within the heart (Figure 2).
Pulmonary artery catheters are
the newest form of invasive temporary pacing. In specialized pulmonary
artery catheters, dedicated atrial and
ventricular ports provide sites for the
introduction of electrode wires while
still allowing routine thermodilution
hemodynamic monitoring.1 Unfortunately, inflating the balloon to measure pulmonary artery wedge pressure
may cause the electrodes to migrate
out of their pacing position. Consequently, simultaneous pacing and
determinations of wedge pressures
are infrequently done.
The second method of temporary invasive pacing involves directly
stimulating the epicardium (Figure 3).
To pulse generator
To pulse generator
Output
Once the wires from the patient
are connected to the generator, the
amount of
electrical output must be
Pacing wire ports on top
selected. The
sole function
of the generator is to supply sufficient
energy to the
heart muscle
to stimulate a
contraction.
Setting the
output has 3
components:
rate, amount,
and chamber.
The rate
determines
the number of
stimulations
to be delivered per
minute. The
amount conBattery compartment on bottom
trols the level
of energy proFigure 4 Pulse generator for a dual-chamber pacemaker.
vided, and the
chamber
depends on both the patients condition and the reason for pacing.1
Rates for a surgical patient can start
as high as 90 to 110 beats/min. In
medical patients, therapy is generally started at 70 to 90 beats/min.
In patients who have had cardiac
arrest, the initial rate is 80
beats/min. Pacing rates for overdrive suppression of tachydysrhythmias may greatly exceed these
values.1 The heart rate on a patients
rhythm strip should never be lower
than the patients set pacemaker
rate.3
Amount The output amount is the
Vs
Vs
Vs
Vp
Capture
Electrical capture, the ability of
the electrical impulse to initiate a
cardiac response, is detected by examining an electrocardiogram. Capture
is both an electrical and a mechanical
event. Electrical capture is indicated
by a pacer spike followed by a corresponding P wave or QRS complex,
depending on which chamber is
being paced (Figures 5 and 6). If the
atrium is paced, the spike appears
before the P wave.7 If the ventricle is
paced, the spike occurs before the
QRS complex.1
Because the pacemaker causes the
heart to depolarize in an artificial
fashion, the path of depolarization is
abnormal, resulting in widened P
waves and QRS complexes.3 A pacer
spike without a corresponding P wave
or QRS complex indicates failure to
capture1 (Figure 7). For a list of potential causes of loss of capture, see
Table 3.
If loss of capture occurs, the
patient is assessed first and then connections and settings are checked to
Vs
Vp
Vp
Ap
Ap
As
As
Ap
Figure 6 Normal electrical capture, atrial pacer. Normal atrial electrical capture is
demonstrated by a pacer spike (Ap) followed by a corresponding P wave.
Abbreviations: Ap, atrial pacing; As, atrial stimulation (intrinsic).
Rhythm strip courtesy of Medtronic, Inc, Minneapolis, Minn.
Capture
X
X
X
X
X
X
X
X
X
Sensing
X
X
X
X
X
X
X
Summary
Competent management of
patients with an invasive temporary
pacemaker is an important skill for
nurses who provide care for critically
ill patients with cardiac disease.
Such management requires familiarity with normal cardiovascular
anatomy and physiology, conduction
system defects, and rhythm interpretation. With an understanding of
the basic concepts of rate, output,
chambers, sensitivity, and capture,
pacing can be done with ease. Care
7. Paschall FE, McErlean ES. Temporary transvenous and epicardial pacing. In: LynnMcHale DJ, Carlson KK, eds. AACN
Procedure Manual for Critical Care. 4th ed.
Philadelphia, Pa: WB Saunders Co;
2001:285-297.