Manual Dinamometro Imada
Manual Dinamometro Imada
Manual Dinamometro Imada
Imada, Inc. warrants its products to the original purchaser to be free from defects
in workmanship and material under normal use and proper maintenance for two
years (one year for adapters, attachments and cables) from original purchase. This
warranty shall not be effective if the product has been subject to overload, shock
load, misuse, negligence, accident or repairs attempted by others than Imada, Inc.
During the warranty period, we will, at our option, either repair or replace
defective products. Please call our customer service department for a return
authorization number and return the defective product to us with freight prepaid.
The foregoing warranty constitutes the SOLE AND EXCLUSIVE WARRANTY, and we
hereby disclaim all other warranties, express, statutory or implied, applicable to the
products and/or software, including but not limited to all implied warranties of
merchantability, fitness, non-infringement, results, accuracy, security and freedom
from computer virus. In no event shall Imada, Inc. and/or its affiliated companies be
liable for any incidental, consequential or punitive damages in connection with the
use of its products and/or software.
Complete color
catalog available.
Call 800-373-9989
or E-mail us.
Model: DS2
INSTRUCTION MANUAL
INTRODUCTION
DS2 Series are state-of-the-art basic performance, easy-to-use
digital force gauges which feature giant, easy-to-read LCD displays
and provide RS-232, Digimatic and analog output. The Real time
measuring mode displays force transients and the Peak measuring
mode captures the peak force achieved during a test. Selectable
lbf(ozf), kgf(gf), and N measuring units.
IMPORTANT
Make sure to read this manual before operating.
1. WARNING!! Test samples and fixtures can break or shatter,
wear eye and body protection to avoid injury.
2. WARNING!! REGARDLESS of whether the unit is ON or OFF,
DO NOT exceed the capacity of the gauge.At 110% of the rated
capacity, the display flashes to warn. NEVER exceed 200% of the
rated capacity, or the load cell will be damaged.Avoid shock
load.
3. The gauge can be used between 30 100F (0 40C)
For the most accurate results, use the gauge at temperatures
close to ones on the certificate of calibration.
4. When mounting DS2 Series, use M4 mounting screws with a
maximum insertion depth of 8 mm into the gauge.
5. Measure in line tension and compression
forces only. DO NOT attempt to measure forces at an angle to the measuring
shaft damage to load cell and/or shaft
may result.
6. Hand tighten attachments. DO NOT use
tools.
7. Make sure this gauge and all peripherals are powered down
before attaching any cables.
8. DO NOT disassemble the gauge. Disassembly voids warranty.
9. Use only Imada AD120 (or AD230) charger/adapter.
page 2
AP-001
IMADA GAUGE
GAUGE MOUNTING
SCREWS (4)
PEM NUTS
DS2 DIMENSIONS
page 11
Measuring Shaft
Attach any of the included
standard attachments or
optional special attachments to measure tension
or compression forces.
up to 110 lbf #10-32
220 lbf
M6
On/Off
Power ON, capacity is displayed.After 10 minutes, unit
powers down if no key is
pressed.
Zero
Tares weight of attachment
and resets peak value
Send
Transmit display value
Peak
Press to activate Peak mode
ACCESSORIES
Optional Handle
The OH-1 optional handle fits DS2 gauges.
Constructed of high quality steel for rugged use,
the handle facilitates measurements of heavy
loads. Complete with mounting screws.
Charger Port
Recharge the internal
NiMH batttery or use the
Imada AD120 or AD230
adapters only.
RS-232 Port
See page 8 for pin assignments
LCD DISPLAY
Optional SW-1 Data Acquisition Software
Use SW-1software to capture and analyze peak data from DS-2 force
gauges.A running log of all the data is displayed along with a chart.
Calculate max/min, average, and standard deviation.
Optional Cables
CB-101 Analog cable (10')
2
3
page 10
Compression icon
Indicates compression measurement
Tension icon
Indicates tension measurement
Units indicator
Displays selected measuring unit. (ozf,
lbf, kgf or N).
Battery icon
Flashes when gauge needs to be
recharged
GO/NG Indicator
under low setpoint
O between low and high setpoints
+ over high setpoint
Negative sign
Displayed when measuring tension
Peak icon
Displays continuously when peak
mode is active
page 3
2. Mitutoyo Digimatic
Connect the CB-301 cable to the communications port and the
device receiving the data. Set up parameters as instructed from the
Mitutoyo processor manual.
GENERAL OPERATION
Selecting
Selecting Units
Units
Press
and press
briefly shows the gauge capacity, then zero with the newly
selected measuring unit, which is retained as a default.
Model
DS2-0.4
DS2-1
DS2-4
DS2-11
DS2-44
DS2-110
DS2-220
Ounces (ozf)/
Pounds (lbf)
7.00 (0.01 ozf)
18.00 (0.01 ozf)
4.400 (0.001 lbf)
11.00 (0.01 lbf)
44.00 (0.01 lbf)
110.0 (0.1 lbf)
220.0 (0.1 lbf)
Capacity (Resolution)
Grams (gf)/
Kilograms (kgf)
200.0 (0.1 gf)
500.0 (0.1 gf)
2.000 (0.001 kgf)
5.000 (0.001 kgf)
20.00 (0.01 kgf)
50.00 (0.01 kgf)
100.0 (0.1 kgf)
Newtons
2.000 (0.001 N)
5.000 (0.001 N)
20.00 (0.01 N)
50.00 (0.01 N)
200.0 (0.1 N)
500.0 (0.1 N)
1000 (1 N)
Reversing
Reversing the
the Display
Display
Press
DS2 Specifications
and press
once. Press
REVERSE
DISPLAY
to cycle between
to select. The
LCD display briefly shows gauge capacity and then zero with
Accuracy
Selectable Units
Overload Capacity
Data processing speed
Display Update
Power
Battery Indicator
CPU
Setpoints
Outputs
Operating Temp.
Accessories included
32 to 100F (0 to 40C)
AC adapter/charger, hook, flat tip, conical tip, chisel tip,
notched tip, extension shaft
page 4
page 9
COMMUNICATIONS PORT
Hand
Hand Tighten
TightenAttachments
Attachments
Definition
RS-232C and Digimatic Ground
RS-232C Transmit Data
Analog Output 1VDC
Digimatic Data Request
RS-232C Receive Data
Analog Ground
Digimatic Clock
Digimatic Transmit Data
Selecting
Selecting Peak
Peak or
or Real
Real time
TimeMeasuring
MeasuringMode
Mode
Press
to turn on the gauge and enter real time mode. For peak
measurement press
Function
Select kgf/gf units
Select N units
Select lbf/ozf units
Select peak mode
Select real time mode
Tare Display
Turn off power
Set high/low setpoints(4 digit)
HHHH=High, LLLL=Low
Read high/low setpoints
D[CR]
page 8
Response*
Apply
Apply Force
Force and
and Measure
Measure
Make sure to apply force in line with the
R[CR] executed
E[CR] error*
EHHHHLLLL[CR]**
setpoint values (4 digit)
HHHH=High, LLLL=Low
[value][units][mode][CR]
Connect the gauge and PC with a CB203 cable. Measure and press
to transmit data to the PC. Use optional SW-1 (see page 10) or
other software to collect and display data.
page 5
OPTIONAL SETTINGS
The following steps are not necessary for most test applications and
may be skipped entirely.
and press
or
briefly show HI, then the High setpoint value with flashing
then press
HI. Press
to increase and
to increase and
to select.
page 6
page 7