DC-Voltage Gradient (DCVG) Surveys Using MCM's Integrated Pipeline Survey Test Equipment and Database Management Package
DC-Voltage Gradient (DCVG) Surveys Using MCM's Integrated Pipeline Survey Test Equipment and Database Management Package
DC-Voltage Gradient (DCVG) Surveys Using MCM's Integrated Pipeline Survey Test Equipment and Database Management Package
MANUAL CONTENTS
Page
SECTION I:
INTRODUCTION
SECTION II:
PHYSICAL PRINCIPLES
SECTION III:
5
9
12
12
13
14
14
16
16
17
20
SECTION IV:
41
42
44
44
46
48
50
VI. 1 Introduction
VI. 2 The Manual Approach ...
VI. 3 Using the Driver in the ProActive Program ...
54
55
56
59
SECTION I: INTRODUCTION
This manual is the sister training manual to MCMs CIS Training Manual
- G1 Version, herein referred to as the CIS Training Manual.
In the CIS Training Manual, our concern was recording the potential
difference between a cathodically-protected buried pipeline and the soil
above it at ground level, as a function of distance down the length of the
pipe.
Since the pipe potential can be high at the site of a defect (due to large
impressed-current flow to the defect), the potential at ground level above the
pipe (above the defect site) can be significantly higher than zero (the remote
earth potential). Consequently, the measurement of a significant above the
defect soil potential (voltage) with respect to remote earth, is a good
indication that a coating-defect has actually been located, especially if the
soil potential above the pipe away from the defect location is essentially zero
with respect to remote earth.
Since the defect site is surrounded by soil, a voltage gradient will exist
between the defect site and the soil. A voltage gradient will actually exist in
3 dimensions. Consider the defect site as being the center of a series of
concentric spheres with the surface of each sphere being a surface of
constant potential. Near the defect, the concentric spheres will be closely-
spaced which means that the potential will drop-off quickly with distance
from the epicenter (the defect site). The spheres (constant potential
surfaces) will be spaced further and further apart, implying a slower
potential drop-off rate, with distance away from the defect site.
directly above the defect is 100% of the maximum potential, ie, the largest
potential measurable for this defect.
Since such surface potential (voltage) gradients exist above pipeline defects,
we can perform soil-to-soil potential difference surveys to detect the defects.
Such surveys are know as DCVG surveys (Direct-Current Voltage Gradient
surveys).
10
The +80mV reading would occur since the voltmeter would subtract
-100mV from -20mV to give +80mV [-20mV - (-100mV) = +80mV].
Remember that the right-hand electrode is connected to the positive side of
the voltmeter and the left-hand electrode is connected to the negative side of
the voltmeter.
If we continue walking past the location of the defect, the difference
between the soil potentials at the reference electrode positions will begin to
decrease in magnitude and eventually will be zero again as we exit the
voltage gradient field. Consequently, a profile such as that shown below
in Figure 5 will be observed in the case of Perpendicular DCVG surveys as a
coating defect is encountered.
DCVG surveys are performed in the rectifier-current ON/OFF mode, ie, the
rectifier current is switched ON and OFF in a cyclic fashion. This allows
soil-to-soil potential differences to be recorded during the ON portion of the
current cycle and also during the OFF portion of the current cycle.
For DCVG surveys, the MCM data loggers software calculates the
difference between the soil-to-soil potential difference recorded during the
ON part of the cycle [delta V(ON)] and the soil-to-soil potential difference
recorded during the OFF part of the cycle [delta V(OFF)]. The polarities of
both delta V(ON) and delta V(OFF) are also recorded at defect sites (more
on these polarities later).
For the remainder of this manual, the term DCVG voltage will be taken to
mean the difference between the CP current ON soil-to-soil voltage
reading and the CP current OFF soil-to-soil voltage reading.
12
13
the maximum DCVG voltage that will be recorded for this defect. This Max
mV value becomes the first component of the Total mV determination.
Once the Max mV has been recorded, a DCVG voltage recording is made by
placing the left-hand reference electrode in the previous position of the right-
hand electrode and moving the right-hand electrode to the location of the
second black dot along the green line (all black dots shown in Figure 6 are
spaced at 4 to 5 feet intervals). Once a DCVG voltage recording has been
made with the electrodes in this position, the electrodes are moved in the
same fashion to the next measurement positions and a third DCVG voltage
recording is made. This process is continued (typically 3 or 4 measurements
are required for a significant defect) until the soil-to-soil potential difference
is basically zero, ie, you have reached remote earth.
The MCM data loggers software will calculate the sum of these DCVG
voltage recordings to generate the Total mV. This will represent the
voltage gradient on the surface of the ground associated with this defect.
Since the CP current is being switched ON and OFF, we can measure the IR
drop associated with the CP current flowing in the soil (primarily to the
defect), which is just the difference in the pipe-to-soil voltage measured
during the ON portion of the cycle and the pipe-to-soil voltage measured
during the OFF portion of the cycle. Consequently, we can have a measure
of the current flowing to the defect (through the IR drop determination),
which we can relate to the voltage gradient (Total mV).
The magnitude of the voltage gradient on the surface of the ground (Total
mV), arising as a consequence of ionic current flowing to the pipe in the
vicinity of the defect (CP current magnitude = I), expressed in relationship
14
In order to determine the IR drop value to use in the above equation, pipe-to-
soil voltages are recorded at the test stations bracketing a defect location.
The data loggers software then calculates the difference between the ON
and the OFF pipe-to-soil voltages (IR drop values) at each of the test
stations. If the IR values recorded at both test stations are essentially the
same, the data logger uses this IR drop value to calculate the % IR value for
the defect using the above equation. If, on the other hand, the IR drop
values recorded at the two test stations bracketing the defect are significantly
different, the data loggers software will generate an average IR drop value
to use in the equation, based on the defects location relative to the locations
of the two test stations.
For example, if the two test stations bracketing a defect are 1 mile apart,
and the defect is located as shown below, and, if the IR drop values are
recorded as shown at the two test stations (300mV and 500mV), the data
logger will determine the IR drop value for the % IR calculation to be:
Consequently, if the Total mV has been determined (by the software) and
the IR drop has been determined, a value for the size of the defect, ie, the
% IR, can be calculated by the software.
For example, if the Total mV for the above defect was determined to be
100 mV, the % IR for the defect would be calculated by the software as
follows:
15
If net current is flowing to the defect, the potential at the defect will be
negative, relative to remote earth; however, the maximum delta V recording
will have a positive value, due to the positioning of the positive and
negative reference electrodes. This is a cathodic condition.
If net current is flowing away from the defect into the soil, the potential at
the defect will be positive with respect to remote earth; however, the
maximum delta V recording will have a negative value. This is an anodic
condition.
Since we are able to record the polarity of Max mV(ON) and Max
mV(OFF), we can determine the corrosion condition of the defect both when
the CP current is impressed and when the CP current is not flowing to the
defect. The best case scenario is cathodic/cathodic, however,
cathodic/anodic is also acceptable, since in this case, the CP systems
impressed current is doing its job, ie, it is preventing corrosion taking place
16
at the defect site. The worst case scenario would be anodic/anodic, since in
this case the defect is not protected even with CP current flowing.
As can be seen from Figure 7, as the operator enters the voltage gradient
field associated with the defect and places the reference electrodes as
indicated with the right-hand electrode connected to the positive side of the
data loggers voltmeter and the left-hand electrode connected to the negative
side of the voltmeter, the DCVG voltage measured will begin to increase and
have a positive polarity.
As the defect location is approached, the DCVG voltage will increase more
sharply, since the circles of constant potential are spaced closer together. A
maximum DCVG voltage will be measured when the left-hand reference
electrode is directly over the defect.
Continuing down the line, with the right-hand electrode directly over the
defect, the DCVG voltage will again have a maximum value, however, this
time the polarity will be negative.
Finally, as the operator continues down the line, the magnitude of the DCVG
voltage measured will decrease quickly near the defect and then decrease
more gradually moving further away from the defect.
17
18
19
Step 1:
Switch on the G1 by pressing the power button (red key on keyboard).
Assuming that the battery pack is charged (see Section II), the screen will
light up and will display the home screen (Today screen), in the case of
the Q100 units, and the desktop screen, in the case of the Q200 units.
The startup screens for the Q100 and the Q200 units are illustrated below:
Q100 Units:
20
Q200 Units:
Step 2:
In the case of the Q100 units, tap on the Start button on the Home
screen. This will pull up the main menu.
Step 3:
In the case of the Q100 units, tap on the G1 PLS button on the main menu.
In the case of the Q200 units, double-tap on the G1PLS icon on the desktop
or tap on the start icon (bottom left-hand corner of desktop), tap on the
Programs button and tap on the G1PLS button.
In either case, (Q100 or Q200 units), the survey screen shown below will be
displayed.
Note: In the case of the Q200 units, the menu items are arrayed along the
top of the screen.
21
Step 4:
Tap on Survey in the menu bar along the bottom of the above screen. The
screen shown below will appear.
Under Survey there are several options. If this is a new survey (not a
continuation of a previous survey) tap on New Survey. The screen shown
below will appear.
22
Step 6:
Enter a filename for the Survey.
Note: This is an important step as the filename is used to identify the survey
and, also, recorded data (voltages) will be stored in this named file for future
retrieval. It is highly recommended that a protocol be established for
selecting Survey Filenames. Critical information should be included in the
filename, such as pipeline companys name, city or state in which the
pipeline is located, pipeline number and section of pipeline number under
survey. The protocol developed should be applied consistently for each
survey.
For example, lets assume that pipeline company XYZ has a pipeline located
in Texas and that the pipeline is identified as pipeline 12 and a survey is
being performed on section 085 of this pipeline. A good filename for this
survey would be:
XYZ TX 12 085 DCVG
When this data file is later accessed, with this filename we know the name of
the pipeline owner, we know the state in which the pipeline is located, we
know the pipeline number, we know the section number of the pipeline that
was surveyed and we know that it was a DCVG survey.
23
Step 7:
Tap once on the OK button. The screen shown below will appear.
Step 8:
Select Survey Type.
The 5 available survey types can be seen by tapping on the pull-down list
arrow in the Survey Type box as shown on the screen below.
24
Tap on DCVG to select a DCVG survey type. The screen shown below will
appear.
Step 9:
Select Rectifier Mode.
25
Step 10:
Make Cane Button Functionality Choices
You can also use either cane button when recording voltages at D.C.P.s
(data collection points (devices)) to save a device reading, as opposed to
tapping on the save button on the device screen. For example, with the
selection as indicated on the above screen, device readings would be
recorded automatically by triggering either cane button, which is the save
option. If you prefer to tap buttons on the screen at Devices, select the
none option in the D.C.P. box.
Step11a:
Select Walking Direction
Finally on the above screen, you should indicate whether station numbers
will be increasing or decreasing as you proceed in the survey direction by
tapping on either Increasing or Decreasing in the Survey Walking
Direction box.
Step 11b:
Make selection of Metric units if required.
26
By checking off the box labeled Metric, the reading interval (distance
between voltage recordings) and the flag internal (flag spacing) will be
displayed in meters, as opposed to feet.
Tap on the Next button on the above screen.
Step 12:
Select GPS Type
By tapping on the pull-down-list arrow button in the GPS: field, you can
select the type of GPS unit you will be using (if any) from the list shown
below.
As can be seen from the above screen, there are 5 choices for GPS Type:
None:
MCM External:
NMEA:
Manual:
MCM Internal:
This means that the G1s internal GPS unit will be used
(a Garmin (Model 15L) WAAS-enabled receiver)
Step 13:
Select GPS Options:
After selecting the GPS Receiver Type and COM Port, choices need to be
made regarding GPS Options (see screen below).
28
If a GPS accessory has been selected for use with the data-logger for a
particular DCVG survey, all, or some of the functions available can be
enabled (box ticked). A box can be ticked or unticked by tapping inside
the box. The GPS options available are as follows:
Note: Logging Standard GPS position data is usually better than having
no data logged at all, which would be the case if you checked this box and
for some reason your GPS receiver was not outputting differentially-
corrected position data. Consequently, it may be preferable to leave this box
unchecked, unless it is imperative that you exclusively log differentially-
corrected GPS position data.
29
By selecting one of these options, you can elect to have the GPS position
data logged automatically at every survey reading, at every second reading,
at every fifth reading, at every tenth reading, or not at all (never) at survey
readings.
Step 14:
Select Max mV Reading to be First Reading in Total mV Determination.
The last box in the above screen is actually not related to GPS Options.
If this box is ticked, the Max mV voltage value recorded at an anomaly
location will automatically become the first voltage value used by the data-
loggers software to calculate the Total mV (total voltage gradient).
Otherwise, if this box is not ticked, you will have to repeat the Max mV
recording a second time as part of the Total mV determination process.
Step 15:
Finally on the above screen, select the electricity supply operating frequency
of the country in which you are performing the surveys (60Hz or 50Hz). For
the U.S., select 60Hz.
30
Tap on the Next button on the above screen. The screen shown below will
appear.
Step 16:
Enter Pipeline Name and Survey Starting Location
Provide the Name of the Pipeline:
By tapping in the field labeled Name of P/L, you can type in the pipeline
name. Note: This is not the same as the filename for the DCVG survey that
was selected back at Step 6. This is the actual name of the pipeline.
Provide the Valve Segment Identification Number:
By tapping in the field labeled Valve Segment, you can type in the valve
segment name or number (if known).
Provide the Starting Location:
Provide the Starting Location for the Survey
You can select to have location information displayed on the survey screen
as station number, feet or milepost (station number, meters or kilometers for
the metric case).
31
Whichever selection you make here will determine how you enter your
starting location information.
For example, if your pipeline locations are represented by station numbers,
you would select Station Number from the drop down list and you would
enter a starting location for the survey in the form of a station number. [If
you do not know the station number where youre beginning your survey,
enter 0+0.0].
As an example, if you are working on pipeline ABC within valve segment
45 and you are about to begin a DCVG survey at station number 12+00.0,
your screen would be as shown below.
Step 17:
Select Voltage Recording Interval (Distance Between Recordings)
By tapping in the field on the above screen labeled Distance Per Reading,
you can type in the voltage reading interval (distance in feet (or in meters for
the metric case) expected between recordings) for the DCVG survey. In
DCVG work this expected interval distance is typically 5.0 feet.
32
Step 18:
Select Survey Flag Interval (Distance Between Survey Flags)
Step 19:
Select the maximum permissible error between the actual number of
recordings made between 2 survey flags and the expected number of
recordings.
By tapping in the field labeled, CIS Flag Read Count %, you can type in
the maximum permissible error. For example, the maximum permissible
error is indicated as 20% on the above screen. If the recording interval is
expected to be 5.0 feet and the survey flag separation is 100 feet, that means
that 20 recordings are expected. If, however, only 15 recordings are
actually made between survey flags, an error window will appear on the
screen, since there is a 25% difference between the expected and actual
number of recordings made. No error window will appear if the difference
is less than 20% for this example, ie, you could have a minimum of 16
recordings and a maximum of 24 recordings between survey flags to stay
within the 20% (max.) error allowance.
Step 20:
Select whether or not you would like the recordings to be uniformly spaced
between survey flags, in cases where less than or greater than 20 recordings
are made.
By tapping in the box labeled, Auto Pacing Mode, and inserting a tick in
the box, you will enable the data-logger to automatically adjust the actual
recordings and space them evenly over the flag spacing distance, regardless
of the actual number of recordings made. Again, enabling this selection is
recommended.
Tap on the Next button on the above screen. The screen shown below will
appear.
33
Step 21:
Provide the Work Order Number for the DCVG survey.
By tapping in the field on the above screen labeled, Work Order #, you
can type in the work order number for the DCVG survey.
Step 22:
Provide Your Name.
By tapping in the field labeled, Technician Name, you can type in your
name or the name of your supervisor. If you use your supervisors name
here, you might add your own name in the Comments Section.
Step 23:
Provide Comments.
By tapping in the field labeled, Comments/Description, you can enter any
comments you might have regarding the survey (perhaps weather conditions,
soil conditions etc.).
Also shown on the above screen are the Survey (File) Name and the Survey
Start Date and Time.
34
Note: Do not attempt to change the File Name indicated here as this
identification will be required by the ProActive software to transfer
your DCVG survey data to your PC.
Tap in the Next button on the above screen. The screen shown below will
appear.
Step 24:
Select Voltmeter Settings and Current Interruption Cycle.
Voltmeter Reading Mode
DCVG surveys are performed with the CP systems rectifier-current
switched ON and OFF in a cyclic fashion and, as can be seen from the above
screen, there are two choices available for the data-loggers voltmeter Read
Mode. The choices are On/Off Pairs (DSP mode) and On/Off Pairs
(GPS Sync).
Note: On/Off Pairs (Min/Max) is not a option here since DCVG surveys
require a measurable difference between On and Off voltages, ie, they
require a significant IR drop.
35
First, select the type of GPS unit you will be using (the data-loggers
Internal unit or the MCM External unit ONLY).
Next, enter your On Delay and your Off Delay selections by tapping in
the appropriate box and typing in the requested delay time in milliseconds.
These delay times are employed so that any spiking in the voltage waveform
that might occur as the rectifier-current is switched from ON to OFF and
from OFF to ON does not become a factor in the voltmeters determination
of the true ON and OFF voltage values for each DCVG measurement. For
example, if 150 ms was selected for the Off Delay, the data logger would
record the voltage value sampled 150 ms after the rectifier-current was
switched from the ON to the OFF state. Also, if 150 ms was selected for the
On Delay the data logger would record the voltage value sampled 150 ms
after the rectifier-current was switched from the OFF to the ON state.
Finally, if your interruption cycle starts with the current in the ON state (the
first transition is from ON to OFF), place a tick in the Start Cycle box
(remove the tick if the opposite is true).
After making your GPS Settings selections, tap on the OK button which
will return you to the Voltmeter Settings screen.
37
Consequently, the rate at which DCVG voltages can be captured and saved
depends critically on the interruption waveform period. This is a particularly
important issue in the case of large period waveforms. For example, if a 10
second On and a 2 second Off waveform is employed (12 second waveform
period), stabilization times up to 12 seconds could be possible before each
DCVG voltage recording can be made. In contrast, the stabilization period
would be only on the order of one second for a 700ms On, 300ms Off
interruption cycle.
The recommended settings for the voltmeter for DCVG surveys are 5.7
Volts (full-scale) and an input impedance value of 400MOhm, These
settings provide a relatively-fast response time (~80 ms), which is important
in rectifier-current-interrupted applications and also a relatively-sensitive
scale with regard to the magnitude of the DCVG voltages that youll be
measuring.
Note: Caution has to be exercised if any of the ranges beginning with the
number 4 (for example, 4V, 400mV or 40mV) are selected for DCVG
applications, since these represent relatively-slow response settings. Such
settings provide a higher level of AC filtering (less noise), but at the expense
of speed of response. Rather than the response time of 80ms for the 5.7V
range, these settings have a response time greater than 1 second.
Consequently, such settings are not appropriate for fast current-interruption
cycles, ie, interruption On and Off times shorter than 1 second, as true On
and Off potentials would not be recorded.
38
Step 25:
Pull-up Active Survey Screen
By tapping on the OK button on the Voltmeter Settings screen, the
active Survey screen would appear and you would be ready to
proceed with your DCVG survey.
An example starting active Survey screen is shown below.
In the above active DCVG survey screen, the direction of the arrow
(pointing to the left or pointing to the right) depends on the polarity of the
DCVG voltage being measured. In the above case, the voltage being
measured was positive which produced an arrow pointing to the left.
The opposite would be true if the DCVG voltage reading had been negative
(ie, the arrow would point to the right).
With the reference electrodes properly positioned, the polarity of the DCVG
voltage will always be positive in the case of perpendicular DCVG surveys.
In this case, the arrow on the survey screen will always point back to the
pipe.
39
Also, with respect to the above active survey screen, as DCVG voltages
are recorded by the data-logger, the Total Distance (total distance from the
start of the survey) parameter will increase in increments of 5.0 feet , or
whatever the Distance Per Reading value was that was entered back at
Step 17. (Distances would be in meters for the metric case).
Also, the Distance From Flag parameter will increase in the same
increments as voltages are recorded. The difference in this case, however,
will be that when each survey flag is registered, this distance parameter will
begin again at zero. In other words, this will show the distance you are
assumed by the data-logger to have traveled from the last flag that you
encountered (and registered).
40
41
As can be seen from Figure 9, a pair of canes (reference electrodes see our
CIS Training Manual) is illustrated, the left-hand cane (RED-handled cane)
and the right-hand cane (GREEN-handled cane). These canes, which will be
placed on the soil above the pipeline in either the Perpendicular or the
In-Line configuration (see Section II), have push buttons on top of the
handles so that the operator can trigger voltage recordings on his
command at each of the pipeline survey measurement locations as well as at
Devices and Geo-Features. See Section III, Step 10 for a discussion on
cane button functionality (Step 10 of the G1s Set Up Process).
The canes (data probes) are connected as shown to the input terminals of
the dual-probe adapter and the output terminal of the adapter is connected
to a 5-pin Data Probe Connector socket on the top side of the G1.
The GPS Antenna illustrated on the top side of the G1 in Figure 9 represents
the antenna employed by the data-loggers internal GPS receiver. External
GPS receiver units, if employed, would be connected to the data-logger via
one of the two USB COM terminals located on the bottom side of the G1.
(The right-hand-side COM Port (COM 4) is recommended. With either the
internal unit or an external GPS unit, the location of items such as flags,
devices, geo-features and anomalies can be recorded during the performance
of a DCVG survey, either manually by tapping on the Log GPS button on
the survey screen at each critical location or automatically, by pre-
programming the data-logger as described above in Section III (Step 13).
The terminals on the top side of the G1 data-logger for the various
connections described above are illustrated in Figure 10 below.
42
With the MCM test equipment connected as shown in Figure 9 (Section IV),
and the G1 data-logger set up as described in Section III, you are ready to
perform a DCVG survey.
With the harness assembly, the G1 sits on a tray at waist level allowing the
operator to view the screen at all times and to make any selections required
by tapping on the screen. Also, the dual-probe adapter shown in Figure 9, is
attached to the underside of the tray, allowing convenient (5 pin cable)
connection of the adapters output to the G1. [Again, this will effectively
connect the green-handled reference electrode cane to the positive side of the
data loggers voltmeter and the red-handled reference electrode cane to the
negative side of the voltmeter].
In addition, an external GPS receiver unit (together with its battery pack and
antenna), if the internal GPS unit is not being employed, can also be attached
to the waist band of the harness, typically on the operators back at waist
level, with the antenna rising to above-head height.
44
45
46
By tapping on the Read button on the above screen, you can view the
actual pipe-to-soil voltage waveform. A typical waveform would be as
shown in the screen below.
Since we set up a 700ms ON and a 300ms OFF rectifier cycle, the pipe-to-
soil voltage is High for 0.7 sec. and Low for 0.3 sec., as evidenced by
the 1 second snap shot of the voltage waveform shown in the above
screen. Since the voltmeter in the G1 has a finite response time (~80ms in
the case of the 5.7V, 400M setting), the transitions from High to Low and
Low to High are not perfect (right-angled) steps.
However, these voltage waveforms are very useful with regard to confirming
multiple-rectifier synchronization and to selecting On Delays and Off
Delays to avoid voltage spikes (see Section III, Step 24), for example.
To record a pipe-to-soil voltage waveform, tap on the save button on the
above screen. The screen shown below will appear.
47
You are being asked here to establish a file in which the waveform data will
be stored (saved). The name of this file will be the name you type into the
Name box on the above screen. For ease of future retrieval when you
uploaded your survey data to the Database Management software package
(ProActive) [see Section VI], it is recommended that you use the same name
as your survey filename with the addition of something like Wave TS 1 to
the file name. This would represent the waveform data recorded at Test
Station 1 at the beginning of your survey.
Note: Make sure that the target folder is G1_Data, as indicated above.
You should then tap on the OK button. This process will save the
waveform data in the G1s CompactFlash memory for future retrieval by
ProActives driver.
V. 4 How to Record the Pipe-To-Soil Voltages at a Test Station:
In order to record the High and the Low voltages (also known as the ON and
the OFF voltages), tap on the Device button on the Survey screen. The
Device screen shown below will appear with Single Test Station
selected.
48
By tapping the Next button on the above screen, the screen shown below
will appear.
As shown above, the High and Low voltage readings (for each cycle)
measured at the test station will be displayed on this screen.
The voltages shown above represent a snap-shot in time.
49
The software will calculate the difference between the High and the Low
voltage recordings (which is the IR drop) and use this quantity in additional
calculations (%IR calculations). The "Use reading for DCVG signal
strengh" setting defaults to "Yes" (box checked).
The IR drop is also referred to as the signal strength.
You should then tap on the Save button to save this data.
Note 1: You should perform pipe-to-soil measurements at each test
station you encounter in order that the software can perform the linear
interpolation calculation described in Section II. 7, when necessary, to
determine an average signal strength (IR drop) to apply at each defect
location between test stations.
Note 2: You can "edit" which device readings (pipe-to-soil readings) will
be used in signal strength calculations using the following procedure:
Tap on the "Options" menu, tap on "Edit Survey Data", tap on "Retrieve",
locate and highlight the device in question (via the "Mode" column) and
tap on the "View Details" button. Next, scroll across to the "Use for %IR"
column and highlight the cell. Next, select Y or N.
Once you have recorded the pipe-to-soil voltages and observed the
magnitude of the signal strength (IR drop), you should disconnect the test
cable from the G1s red banana plug terminal and reconnect the positive
(green-handled) cane to the dual-probe adapter. You are now ready to
record DCVG voltages down the length of the pipeline.
For Perpendicular DCVG surveys, you would place the reference electrodes
(canes) about 4 to 5 feet apart as indicated in Figure 4 (Section II) with the
left-hand cane (negative cane) directly over the pipe and the right-hand cane
(positive cane) off to the right-hand side. If you set up the cane push-button
functionality as indicated in Section III (Step 10), ie, with the right-hand
cane set to trigger voltage recordings and the left-hand cane set to designate
the location of survey flags, you would trigger the right-hand cane every 5
feet or whatever recording interval you set up in Step 17 of the set-up
process (see Section III).
50
As discussed in Section II. 2, if you are outside the voltage gradient field
of a coating defect, the DCVG voltages will essentially be zero. However,
as you enter a defects voltage gradient field, you will observe an increase in
the DCVG voltage values displayed on the G1s Survey screen (see Figure 5
in Section II. 2). When you observe a peak (maximum value) in the DCVG
voltage readings, you should interrupt your walking and Mark the location
of the maximum DCVG reading. This will represent a defect location.
In order to mark the defect location, you should tap on the mark button on
the Survey screen. The screen shown below will appear.
As discussed in Section III (Step 24), the arrow on the Survey screen will
always point back to the pipe in the perpendicular survey case.
If you selected automatic logging of GPS position data at
Sidedrains/Anomalies back at Step 13 (Section III), the G1 data-logger will
log the GPS position data for this DCVG Anomaly. If you did not select this
option, you should tap on the Log GPS button at this time to log the defect
location, if you are using a GPS Unit.
51
If you are satisfied that you are reading the maximum DCVG voltage value
for this defect, you should record the Max mV reading by tapping on the
radio button labeled Read Max mV and tapping on the Accept button
(do not tap on the Save button at this time). The screen show below will
appear.
In the above example, the Max mV reading was 65.1mV [delta V (ON) was
+104.7mV and delta V (OFF) was +39.7mV and the difference between
these values is 65.1mV; this was a cathodic/cathodic situation, since both
polarities were positive].
You will also notice that the software has automatically applied the Max
mV value to the Total mV determination, since we selected this option
back at Step 13 (Section III). If this were not the case, you would Accept
the Max mV reading a second time and this would become the first voltage
used in the Total mV determination.
52
You should then proceed to move the electrodes to their second positions
(see Section II. 5) and you should Accept the second reading. You should
proceed in this fashion until you are outside of the defects voltage gradient
field, ie, the DCVG voltage reading is essentially zero (typically DCVG
voltages less than about 5mV would be considered essentially zero, on the
5.7V, 400M voltmeter setting).
The screen will be as shown below.
As can be seen from the above screen, the software has generated a sum of
the voltage recordings required for us to reach remote earth, ie, be outside of
the voltage gradient field associated with this defect. This represents the
Total mV value, which in our example is 199.9 mV.
At this point, you should tap on the Save button which will save all of the
data associated with this defect.
You would then pick up the survey where you left off and continue
triggering DCVG voltage recordings until you encounter the voltage
gradient field of another defect.
53
VI. 1 Introduction
Survey data are stored in independent files (one file for each survey) on the
CompactFlash memory card on the G1 data-logger and you can copy survey
files to your PC using one of two approaches; manually or via the driver in
the ProActive software program.
If you have the ProActive software program installed on your PC, or you can
bring your G1 to a PC that has ProActive installed on it, you can use the
ProActive program to automatically access survey files on the G1. If either
of these situations applies, you would proceed to Section VI. 3.
If you do not have the ProActive software program installed on your PC and
you cannot bring your G1 to a PC that has ProActive installed on it, you can
copy survey files manually from your G1 to your PC and you can
subsequently send the copied files to a recipient that has ProActive installed
on their PC. If this situation applies, you would proceed to Section VI. 2.
Note: In order to copy survey files from your G1 to your PC, either
manually or via ProActive, you will need to have the Microsoft
ActiveSync software program installed on your PC.
54
Step 1:
Connect a communication cable between the left-hand side COM Port on the
G1 data-logger and your PC and switch on your G1. With Microsofts
ActiveSync program installed on your PC, the ActiveSync window will
appear on your desktop and the program will confirm that two-way
communication has been established.
Step 2:
Follow the procedures detailed below:
You would now be in a position to send the survey file to a recipient who
has access to the ProActive software program.
Note: Do not rename the survey file prior to sending the file to the
recipient, as the survey file must have the same name as the survey
itself.
55
Step 1:
Create a folder on your PCs hard-drive that will be used to permanently
save files copied from your data-logger. You might choose to name this
folder something like, Surveys.
Step 2:
Connect a communication cable between the left-hand side COM Port on the
G1 data-logger and your PC and switch on your G1. With Microsofts
ActiveSync program installed on your PC, the ActiveSync window will
appear on your desktop and the program will confirm that two-way
communication has been established.
Step 3:
Double-click on the ProActive icon on your PCs desktop screen.
Step 4:
Click on the Surveys button on the main menu bar at the top of the screen.
This will open a window labeled Data Logger: Get Pipeline Survey.
Step 5:
Select Gathererone (G1) Option.
Note: It may take a few seconds for the Driver Window to appear.
56
Step 6:
Select the Survey File to be Copied.
Select Pipeline Survey in the field labeled Data Type.
The Surveys field in the Gathererone Driver window will list all of the
survey files currently stored on your G1s CompactFlash memory
card. Highlight the survey file that you would like to be copied to your PC.
Also, place a tick in the box labeled Copy to Local Folder and identify the
folders location on your hard-drive in the field underneath.
This is the folder that you set up previously in which to save all of your
survey files copied from your G1. If you named the folder Surveys, the
folders location would be: C:\Surveys
Also, if you used the metric system on the data-logger, check off the box
labeled, Use Metric.
57
Step 7:
Examine the Survey Data Prior to Bringing the Data into the Database
Management Section of ProActive.
You will have actually completed the process of copying a survey file to a
local folder on your PC by this point, using the Driver in ProActive.
Survey Settings
Readings
Device Readings
Graph
By clicking on these page tabs, you can view information on the Survey
conditions or you can view actual survey data (see the ProActive Training
Manual).
58
Once you have copied your survey files to your PC (see Section VI), you can
(if you wish) delete the files from your G1s CompactFlash memory card if
you need, for example, to create space in the data-loggers file storage
memory to make room for future survey files.
Since your survey files are stored on the CompactFlash card, you will have
to access this memory in order to delete selected survey files. The procedure
would be as follows:
59
Tap and hold on the Survey File you wish to delete until a menu
appears
Tap on Delete. This will delete the highlighted survey file
60