Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                

DC-Voltage Gradient (DCVG) Surveys Using MCM's Integrated Pipeline Survey Test Equipment and Database Management Package

Download as pdf or txt
Download as pdf or txt
You are on page 1of 60

DC-Voltage Gradient (DCVG) Surveys Using MCMs

Integrated Pipeline Survey Test Equipment


and Database Management Package

DCVG Training Manual G1 Version

M. C. Miller Co., Inc.


11640 U.S. Highway 1, Sebastian, FL 32958
Telephone: 772 794 9448; Website: www.mcmiller.com

MANUAL CONTENTS

Page
SECTION I:

INTRODUCTION

SECTION II:

PHYSICAL PRINCIPLES

II. 1 Potential (Voltage) Gradients


II. 2 Perpendicular DCVG Survey Technique ..
II. 3 Definition of DCVG Voltage ...
II. 4 Definition of Total mV .
II. 5 Procedure to Determine Total mV
II. 6 Defect Size Considerations ....
II. 7 Definition of % IR .
II. 8 Summary of DCVG Parameters at Defect Locations
II. 9 Polarity Considerations ..
II. 10 In-Line (Parallel) DCVG Survey Technique .

SECTION III:

HOW TO SETUP THE G1 DATA-LOGGER


FOR DCVG SURVEYS

5
9
12
12
13
14
14
16
16
17

20

SECTION IV:

TEST EQUIPMENT HOOK-UPS FOR


DCVG SURVEYS

IV. 1 How to Make Cable Hook-Ups for DCVG Surveys ..


IV. 2 How to Attach Cables and Accessories to the G1 ......

41
42

SECTION V: HOW TO PERFORM DCVG SURVEYS

V.1 How to Carry Test Equipment During a DCVG Survey .


V.2 Readings to Take at Beginning of a DCVG Survey
V.3 How to Record the Pipe-To-Soil Voltage Waveform ..
V.4 How to Record the Pipe-To-Soil Voltages at a Test Station .
V.5 How to Locate and Mark Defects .

44
44
46
48
50

SECTION VI: HOW TO COPY SURVEY FILES FROM THE


G1 TO YOUR PC

VI. 1 Introduction
VI. 2 The Manual Approach ...
VI. 3 Using the Driver in the ProActive Program ...

54
55
56

APPENDIX 1: How to Delete Survey Files from the G1 ..........

59

SECTION I: INTRODUCTION

This manual is the sister training manual to MCMs CIS Training Manual
- G1 Version, herein referred to as the CIS Training Manual.

In the CIS Training Manual, our concern was recording the potential
difference between a cathodically-protected buried pipeline and the soil
above it at ground level, as a function of distance down the length of the
pipe.

In the case of DC-Voltage Gradient (DCVG) surveys, however, our concern


is recording soil-to-soil potential differences at ground level above a buried
pipeline, as a function of distance down the length of the pipe. So, in this
case, we do not make electrical contact with the pipe itself (other than to
determine IR drop values see later).

DCVG surveys are typically performed on coated pipelines with a view to


determining the location of holidays (coating defects) or other defects. As
discussed in subsequent sections, such surveys can be used to not only pin-
point the location of defects, but they can also be used to provide a measure
of the size (severity) of a particular defect.

Since DCVG surveys are close-interval in nature, they can be considered


complementary to pipe-to-soil close interval surveys (CIS) and both types of
surveys are typically performed on the same pipeline sections as part of the
ECDA (External Corrosion Direct Assessment) protocol regarding Indirect
Inspections. With regard to the ECDA protocol, at least 2 types of close-
interval Indirect Inspection surveys are required to be performed on all
sections of a buried pipeline and typically DCVG and CIS surveys are
employed to satisfy this requirement on coated-pipelines.

The physical principles associated with DCVG surveys are discussed in


Section II and the setup procedures of MCMs data loggers for the
performance of DCVG surveys are described in Section III. Section IV
details the equipment hookup requirements for DCVG surveys and how to
actually conduct a DCVG survey is presented in Section V. Finally, Section
VI describes the procedures involved in copying DCVG survey files to a PC.

SECTION II: PHYSICAL PRINCIPLES


II. 1 Potential (Voltage) Gradients
As discussed in our CIS Training Manual, when a buried pipeline is under
cathodic protection, the DC rectifier current that flows to (and in) the
pipeline (impressed current) causes the pipeline to be negatively-charged
with respect to the surrounding soil. In fact, as illustrated in Figure 1 below,
a potential gradient exists between the pipeline and the surrounding soil, the
potential being largest at the pipe itself (largest negative value) and dropping
off rapidly with distance away from the pipe. The potential at remote
earth is zero and the potential gradient is the potential difference between
the pipe potential (negative potential at the pipe) and remote earth.

Figure 1: Potential Gradient Associated with a Cathodically-Protected


Pipeline (Equipotential Lines are closer together near the
pipe, indicating that the potential drops quickly (initially)
with distance away from the pipe).

In the case of a well-coated section of pipeline, the potential at ground level


in the vicinity of the pipe will be close to zero (close to remote earth
potential).
However, when coating defects (holidays) are present, the local potential at
the defect locations increases significantly, since the impressed current flows
through the soil primarily to the defects. This situation is illustrated
schematically in Figure 2 below.

Figure 2: The Influence of Coating Defects on Impressed-Current Flow


Distribution along a Pipeline (locally high potentials occur
at defect locations due to locally large currents flowing to the
defects).
A small amount of current will also flow to the pipeline through the coating
(the coating will not be a perfect electrical insulator), however, this current
will be negligible compared to that flowing to the coating defect areas.
6

Since the pipe potential can be high at the site of a defect (due to large
impressed-current flow to the defect), the potential at ground level above the
pipe (above the defect site) can be significantly higher than zero (the remote
earth potential). Consequently, the measurement of a significant above the
defect soil potential (voltage) with respect to remote earth, is a good
indication that a coating-defect has actually been located, especially if the
soil potential above the pipe away from the defect location is essentially zero
with respect to remote earth.

Since the defect site is surrounded by soil, a voltage gradient will exist
between the defect site and the soil. A voltage gradient will actually exist in
3 dimensions. Consider the defect site as being the center of a series of
concentric spheres with the surface of each sphere being a surface of
constant potential. Near the defect, the concentric spheres will be closely-
spaced which means that the potential will drop-off quickly with distance
from the epicenter (the defect site). The spheres (constant potential
surfaces) will be spaced further and further apart, implying a slower
potential drop-off rate, with distance away from the defect site.

If we consider the surface of the ground above the pipeline to be a plane,


the ground will have an in-plane potential gradient in the region above the
site of the defect. Imagine taking a slice through the ball of concentric
spheres (spherical surfaces of constant potential). The slice would represent
the surface of the ground, with the distance from the epicenter of the
concentric spheres to the slice being the distance from the defect to the
ground, ie, the depth of the pipeline.

Figure 3 illustrates the in-plane potential gradient (voltage gradient)


associated with the surface of the ground in the vicinity of a coating defect
on a buried pipeline. The defect location is indicated by the X symbol on
the buried pipeline and the voltage gradient on the surface of the ground,
arising as a consequence of the defect, is represented by the concentric
circles of constant potential. The separation between constant potential
circles is small in the vicinity of the defect, indicating a rapid drop-off in
potential near the defect while the separation between constant potential
circles increases with distance from the defect, indicating a less rapid drop-
off rate with distance as the remote earth potential is approached. In order to
illustrate this point, the constant potential circles have been labeled as
percentages. It is assumed that the potential on the surface of the ground

directly above the defect is 100% of the maximum potential, ie, the largest
potential measurable for this defect.

Figure 3: Potential (Voltage) Gradient at the Surface of the Ground


associated with a Coating Defect on a Buried Pipeline
(concentric circles represent constant potential lines).
The X symbol represents location of coating defect.
If we imagine pushing a pole vertically from the surface of the ground
directly above the defect down to the defect itself, and then walking on the
surface of the ground away from the pole, well encounter lower potentials
(with respect to our maximum potential at the pole) as we move further and
further away from the pole. The percentages indicated in Figure 3, would
represent percentages of the maximum potential measured at the pole
location.

Since such surface potential (voltage) gradients exist above pipeline defects,
we can perform soil-to-soil potential difference surveys to detect the defects.
Such surveys are know as DCVG surveys (Direct-Current Voltage Gradient
surveys).

DCVG surveys can be performed in two different ways: Perpendicular Mode


and In-Line (or Parallel) Mode. In the case of Perpendicular DCVG
surveys, an imaginary line drawn between the reference electrodes on the
surface of the ground would be perpendicular to the direction of the
pipeline and, in the case of In-Line DCVG surveys, an imaginary line drawn
between the reference electrodes on the surface of the ground would
(ideally) be in-line with the center of the pipeline (directly above the
pipeline).

II. 2 Perpendicular DCVG Survey Technique:

As illustrated in Figure 4 below, this technique involves placing a matched-


pair of reference electrodes (canes) on the surface of the ground
perpendicular to the pipeline direction, with the left-hand cane (red cane)
positioned (ideally) over the centerline of the pipe and the right-hand cane
(green cane) spaced typically 4 or 5 feet from the left-hand cane. With this
cane spacing maintained, the operator walks down the length of the pipeline
section triggering voltage recordings every 2.5 feet (or some other length
interval, typically 5 to 10 feet). Since the left-hand cane connects to the
negative side of the voltmeter (data loggers voltmeter) and the right-hand
cane connects to the positive side of the voltmeter, the voltage recorded at
each triggering location is the difference between the soil potentials at the
two reference electrode positions in each case. For a detailed discussion on
the design and operation of MCMs reference electrodes used for surveys on
buried pipelines, please refer to our CIS Training Manual.

Figure 4: Reference Electrode Positioning for a Perpendicular DCVG


Survey. Potential difference measured at the green line will
represent the maximum voltage recorded.
As can be seen from Figure 4, prior to encountering the coating defect, the
potential difference measured between the reference electrodes will be
essentially zero, since we are currently outside the voltage gradient field.
As we enter the voltage gradient field, walking towards the defect, the
difference between the soil potentials at the reference electrode positions
will begin to increase in magnitude and reach a maximum value when we
are level with the defect (the green line location in Figure 4). As an
example, if the maximum potential at the defect is -100mV (with respect to
remote earth), the potential difference measured by the data loggers
voltmeter at the green line location (defect location) would be +80mV, since
the right-hand reference electrode would be positioned on the 20% constant
potential line.

10

The +80mV reading would occur since the voltmeter would subtract
-100mV from -20mV to give +80mV [-20mV - (-100mV) = +80mV].
Remember that the right-hand electrode is connected to the positive side of
the voltmeter and the left-hand electrode is connected to the negative side of
the voltmeter.
If we continue walking past the location of the defect, the difference
between the soil potentials at the reference electrode positions will begin to
decrease in magnitude and eventually will be zero again as we exit the
voltage gradient field. Consequently, a profile such as that shown below
in Figure 5 will be observed in the case of Perpendicular DCVG surveys as a
coating defect is encountered.

Figure 5: DCVG Voltage as a Function of Electrode Position along


Pipeline in the vicinity of a Coating Defect as measured with
electrodes in perpendicular configuration (see Figure 4)
11

II. 3 Definition of DCVG Voltage:

DCVG surveys are performed in the rectifier-current ON/OFF mode, ie, the
rectifier current is switched ON and OFF in a cyclic fashion. This allows
soil-to-soil potential differences to be recorded during the ON portion of the
current cycle and also during the OFF portion of the current cycle.

For DCVG surveys, the MCM data loggers software calculates the
difference between the soil-to-soil potential difference recorded during the
ON part of the cycle [delta V(ON)] and the soil-to-soil potential difference
recorded during the OFF part of the cycle [delta V(OFF)]. The polarities of
both delta V(ON) and delta V(OFF) are also recorded at defect sites (more
on these polarities later).

Since delta V(ON) represents the soil-to-soil potential difference with


current contributions from the pipelines CP system as well as from all other
sources (stray interference, foreign pipelines etc.), and, since delta V(OFF)
represents the soil-to-soil potential difference with only the other current
sources contributing, a DCVG recording will represent the soil-to-soil
potential difference with only the pipelines CP system contributing to the
current flow to the defect.

For the remainder of this manual, the term DCVG voltage will be taken to
mean the difference between the CP current ON soil-to-soil voltage
reading and the CP current OFF soil-to-soil voltage reading.

II. 4 Definition of Total mV:

The, so-called, Total mV is the difference between the maximum potential


at a defect location and the potential at remote earth, arising due to the CP
systems contribution to the current flow to the defect. Since, as shown in
Figure 4, a typical electrode spacing is 4 to 5 feet and the voltage gradient
field associated with a defect typically extends a significant distance
beyond the right-hand reference electrode location, several measurements
are actually required in order to determine the total voltage gradient (Total
mV) associated with a defect.

12

II. 5 Procedure to determine Total mV:


When the defect location has been identified as described above (ie, the
location where a maximum DCVG voltage reading has been observed), a set
of DCVG voltage recordings is made by moving in a straight line
perpendicular to the pipeline direction.

Figure 6: Reference Electrode Positions (black dot locations along


green line) for Total mV Determination
With reference to Figure 6 above, a recording is first made with the left-hand
reference electrode positioned over the defect and the right-hand reference
electrode positioned at the first black dot location along the green line. The
DCVG voltage recorded here is known as the Max mV recording, as this is

13

the maximum DCVG voltage that will be recorded for this defect. This Max
mV value becomes the first component of the Total mV determination.

Once the Max mV has been recorded, a DCVG voltage recording is made by
placing the left-hand reference electrode in the previous position of the right-
hand electrode and moving the right-hand electrode to the location of the
second black dot along the green line (all black dots shown in Figure 6 are
spaced at 4 to 5 feet intervals). Once a DCVG voltage recording has been
made with the electrodes in this position, the electrodes are moved in the
same fashion to the next measurement positions and a third DCVG voltage
recording is made. This process is continued (typically 3 or 4 measurements
are required for a significant defect) until the soil-to-soil potential difference
is basically zero, ie, you have reached remote earth.

The MCM data loggers software will calculate the sum of these DCVG
voltage recordings to generate the Total mV. This will represent the
voltage gradient on the surface of the ground associated with this defect.

II. 6 Defect Size Considerations:

Since Total mV represents the voltage gradient on the surface of the


ground associated with CP current flowing to a defect, if we can relate the
Total mV to a measure of the magnitude of the CP current, we can obtain a
measure of the size of the defect.

Since the CP current is being switched ON and OFF, we can measure the IR
drop associated with the CP current flowing in the soil (primarily to the
defect), which is just the difference in the pipe-to-soil voltage measured
during the ON portion of the cycle and the pipe-to-soil voltage measured
during the OFF portion of the cycle. Consequently, we can have a measure
of the current flowing to the defect (through the IR drop determination),
which we can relate to the voltage gradient (Total mV).

II. 7 Definition of % IR:

The magnitude of the voltage gradient on the surface of the ground (Total
mV), arising as a consequence of ionic current flowing to the pipe in the
vicinity of the defect (CP current magnitude = I), expressed in relationship

14

to the IR voltage drop in the soil (due to the CP current flowing), is a


quantity known as the % IR.

% IR = [Total mV/IR drop] x 100

In order to determine the IR drop value to use in the above equation, pipe-to-
soil voltages are recorded at the test stations bracketing a defect location.
The data loggers software then calculates the difference between the ON
and the OFF pipe-to-soil voltages (IR drop values) at each of the test
stations. If the IR values recorded at both test stations are essentially the
same, the data logger uses this IR drop value to calculate the % IR value for
the defect using the above equation. If, on the other hand, the IR drop
values recorded at the two test stations bracketing the defect are significantly
different, the data loggers software will generate an average IR drop value
to use in the equation, based on the defects location relative to the locations
of the two test stations.

For example, if the two test stations bracketing a defect are 1 mile apart,
and the defect is located as shown below, and, if the IR drop values are

<---- 1/3 mile ---------> <--------------- 2/3 mile --------------------- >


TS1-----------------------X-------------------------------------------------TS2
300mV
Defect
500mV
(IR drop)
(IR drop)

recorded as shown at the two test stations (300mV and 500mV), the data
logger will determine the IR drop value for the % IR calculation to be:

300mV + [1/3 x (500mV 300mV)] = 366.67mV

Consequently, if the Total mV has been determined (by the software) and
the IR drop has been determined, a value for the size of the defect, ie, the
% IR, can be calculated by the software.

For example, if the Total mV for the above defect was determined to be
100 mV, the % IR for the defect would be calculated by the software as
follows:

% IR = (100mV/366.67mV) x 100 = 27.27%

15

II. 8 Summary of DCVG Parameters at Defect Locations

Assuming that IR drop recordings were properly made at bracketing test


stations and that the appropriate DCVG voltages were recorded at a defect
location (with reference to Figure 6 above), the software calculates and
stores the following important quantities (in addition to storing all regular
DCVG survey data).

For each marked defect site:


> Maximum delta V(ON) or Max mV(ON): Magnitude and Polarity
> Maximum delta V(OFF) or Max mV(OFF): Magnitude and Polarity
> Max mV (Max mV(ON) Max mV(OFF)): Magnitude (always positive)
> Total mV: Magnitude
> % IR: Magnitude

II. 9 Polarity Considerations

By noting the polarities of the maximum soil-to-soil potential difference


recorded with rectifier-current ON and rectifier-current OFF at a defect site,
we can determine the corrosion condition of the defect with the CP system
current flowing (ON state) and without the CP systems impressed current
flowing (OFF state).

If net current is flowing to the defect, the potential at the defect will be
negative, relative to remote earth; however, the maximum delta V recording
will have a positive value, due to the positioning of the positive and
negative reference electrodes. This is a cathodic condition.

If net current is flowing away from the defect into the soil, the potential at
the defect will be positive with respect to remote earth; however, the
maximum delta V recording will have a negative value. This is an anodic
condition.

Since we are able to record the polarity of Max mV(ON) and Max
mV(OFF), we can determine the corrosion condition of the defect both when
the CP current is impressed and when the CP current is not flowing to the
defect. The best case scenario is cathodic/cathodic, however,
cathodic/anodic is also acceptable, since in this case, the CP systems
impressed current is doing its job, ie, it is preventing corrosion taking place

16

at the defect site. The worst case scenario would be anodic/anodic, since in
this case the defect is not protected even with CP current flowing.

II. 10 In-Line (Parallel) DCVG Survey Technique:

There is a second method of performing DCVG pipeline surveys, known as


In-Line (or Parallel) mode DCVG surveys.

In this case, DCVG voltages are recorded in a close-interval fashion by


placing the reference electrodes in-line with the pipeline (directly over the
pipeline), as opposed to placing the reference electrodes perpendicular to the
pipeline direction.

This technique is illustrated in Figure 7 below.

As can be seen from Figure 7, as the operator enters the voltage gradient
field associated with the defect and places the reference electrodes as
indicated with the right-hand electrode connected to the positive side of the
data loggers voltmeter and the left-hand electrode connected to the negative
side of the voltmeter, the DCVG voltage measured will begin to increase and
have a positive polarity.

As the defect location is approached, the DCVG voltage will increase more
sharply, since the circles of constant potential are spaced closer together. A
maximum DCVG voltage will be measured when the left-hand reference
electrode is directly over the defect.

As indicated in Figure 7, however, the DCVG voltage will drop to zero


when the operator is straddling the defect, since in this position, the
reference electrodes will be measuring the same potential and, hence, the
potential difference between them will be zero.

Continuing down the line, with the right-hand electrode directly over the
defect, the DCVG voltage will again have a maximum value, however, this
time the polarity will be negative.

Finally, as the operator continues down the line, the magnitude of the DCVG
voltage measured will decrease quickly near the defect and then decrease
more gradually moving further away from the defect.

17

Figure 7: In-Line DCVG Survey Technique

Consequently, the DCVG voltage profile recorded in the vicinity of a defect


will be considerable different in appearance from that observed in the case
of a Perpendicular DCVG survey (see Figure 5).

18

Figure 8 represents an illustration of the type of DCVG voltage profile that


is recorded during an In-Line DCVG survey when a defect is encountered.

Figure 8: DCVG voltage in the vicinity of a coating defect as a function


of the position along the pipeline of the center point between
the electrodes with the electrodes configured as shown in
Figure 7 (In-Line or Parallel)
It should be noted that once a defect has been located during an In-Line
DCVG survey, the critical parameters noted above in Section II. 8 for
defect-size determination, need to be determined using the Perpendicular
DCVG technique illustrated in Figure 6.

19

SECTION III: HOW TO SETUP THE G1 DATA-LOGGER


FOR DCVG SURVEYS
The following section outlines the steps required to setup the G1 data-logger
to participate in DCVG pipeline survey applications. The setup process
establishes the conditions of the particular survey about to be performed and
identifies the section of pipeline that is about to be examined by the DCVG
application. The setup process (and this is very important) also establishes a
file in which the voltage recordings (survey data) will be stored. At the
completion of the survey, DCVG survey data can then be retrieved by a PC
that is in communication with the data-logger by accessing the file in which
the survey data are stored.

Step 1:
Switch on the G1 by pressing the power button (red key on keyboard).
Assuming that the battery pack is charged (see Section II), the screen will
light up and will display the home screen (Today screen), in the case of
the Q100 units, and the desktop screen, in the case of the Q200 units.
The startup screens for the Q100 and the Q200 units are illustrated below:
Q100 Units:

20

Q200 Units:

Step 2:
In the case of the Q100 units, tap on the Start button on the Home
screen. This will pull up the main menu.

Step 3:
In the case of the Q100 units, tap on the G1 PLS button on the main menu.
In the case of the Q200 units, double-tap on the G1PLS icon on the desktop
or tap on the start icon (bottom left-hand corner of desktop), tap on the
Programs button and tap on the G1PLS button.
In either case, (Q100 or Q200 units), the survey screen shown below will be
displayed.
Note: In the case of the Q200 units, the menu items are arrayed along the
top of the screen.

21

Step 4:
Tap on Survey in the menu bar along the bottom of the above screen. The
screen shown below will appear.

Under Survey there are several options. If this is a new survey (not a
continuation of a previous survey) tap on New Survey. The screen shown
below will appear.
22

Step 6:
Enter a filename for the Survey.
Note: This is an important step as the filename is used to identify the survey
and, also, recorded data (voltages) will be stored in this named file for future
retrieval. It is highly recommended that a protocol be established for
selecting Survey Filenames. Critical information should be included in the
filename, such as pipeline companys name, city or state in which the
pipeline is located, pipeline number and section of pipeline number under
survey. The protocol developed should be applied consistently for each
survey.
For example, lets assume that pipeline company XYZ has a pipeline located
in Texas and that the pipeline is identified as pipeline 12 and a survey is
being performed on section 085 of this pipeline. A good filename for this
survey would be:
XYZ TX 12 085 DCVG
When this data file is later accessed, with this filename we know the name of
the pipeline owner, we know the state in which the pipeline is located, we
know the pipeline number, we know the section number of the pipeline that
was surveyed and we know that it was a DCVG survey.
23

Note: You will not be permitted to use invalid characters, such as


slashes( / or \ ), as part of a filename. You will be alerted if you try to
use any invalid characters.

Step 7:
Tap once on the OK button. The screen shown below will appear.

Step 8:
Select Survey Type.
The 5 available survey types can be seen by tapping on the pull-down list
arrow in the Survey Type box as shown on the screen below.

24

Tap on DCVG to select a DCVG survey type. The screen shown below will
appear.

Step 9:
Select Rectifier Mode.

25

Select Cycle ON/OFF in the Rectifier Mode field by tapping on the


pull-down list arrow button and highlighting Cycle ON/OFF. This is the
only appropriate rectifier mode for DCVG surveys.

Step 10:
Make Cane Button Functionality Choices

Typically, for DCVG surveys, you would trigger DCVG voltage


recordings using the right-hand (positive) reference electrode and the left-
hand reference electrodes push button switch would be used to designate
the location of survey flags. For this case, your Survey selection for canes
would be Flag Read as shown in the above screen. The other options can
be viewed by tapping on the pull-down list arrow button in the Survey field.

You can also use either cane button when recording voltages at D.C.P.s
(data collection points (devices)) to save a device reading, as opposed to
tapping on the save button on the device screen. For example, with the
selection as indicated on the above screen, device readings would be
recorded automatically by triggering either cane button, which is the save
option. If you prefer to tap buttons on the screen at Devices, select the
none option in the D.C.P. box.

Note: When Marking DCVG anomalies, the cane push-button


functionality becomes accept, regardless of whether save or none is
selected here.

Step11a:
Select Walking Direction

Finally on the above screen, you should indicate whether station numbers
will be increasing or decreasing as you proceed in the survey direction by
tapping on either Increasing or Decreasing in the Survey Walking
Direction box.

Step 11b:
Make selection of Metric units if required.

26

By checking off the box labeled Metric, the reading interval (distance
between voltage recordings) and the flag internal (flag spacing) will be
displayed in meters, as opposed to feet.
Tap on the Next button on the above screen.

Step 12:
Select GPS Type
By tapping on the pull-down-list arrow button in the GPS: field, you can
select the type of GPS unit you will be using (if any) from the list shown
below.

As can be seen from the above screen, there are 5 choices for GPS Type:
None:
MCM External:
NMEA:
Manual:

This means that a GPS receiver is not being used


This means that an external MCM GPS Accessory is
being used
This means that an NMEA 0183 compatible GPS
receiver is being used
This means that location data will be entered manually
when the GPS button is pressed on the survey screen
during a DCVG survey.
27

MCM Internal:

This means that the G1s internal GPS unit will be used
(a Garmin (Model 15L) WAAS-enabled receiver)

Select the appropriate choice by tapping on your selection.


When using an external GPS unit, you need to select the Com Port that
youll be using on the G1 (see above screen). The connector terminal on the
left-hand side of the data-logger (facing the bottom side) is the COM 1 Port
and the terminal on the right-hand side is the COM 4 Port. It is
recommended that you select the COM 4 Port for your external GPS unit in
order to avoid potential conflicts with ActiveSync which uses COM 1.

Step 13:
Select GPS Options:
After selecting the GPS Receiver Type and COM Port, choices need to be
made regarding GPS Options (see screen below).

28

If a GPS accessory has been selected for use with the data-logger for a
particular DCVG survey, all, or some of the functions available can be
enabled (box ticked). A box can be ticked or unticked by tapping inside
the box. The GPS options available are as follows:

Differential GPS Required:


This box should be ticked if you only want differentially-corrected GPS
position data to be logged by the data-logger. If this box is left unticked, it
means that you will allow the G1 data-logger to log either standard GPS
position data or differentially-corrected GPS position data.

Note: Logging Standard GPS position data is usually better than having
no data logged at all, which would be the case if you checked this box and
for some reason your GPS receiver was not outputting differentially-
corrected position data. Consequently, it may be preferable to leave this box
unchecked, unless it is imperative that you exclusively log differentially-
corrected GPS position data.

Use GPS Altitude:


If this box is ticked, altitude data will be included with the position data
whenever GPS data is logged. (Note: Altitude data on some GPS units is not
particularly accurate in survey applications).

Log GPS at Flags:


If this box is ticked, GPS position data will be logged automatically at flags
when either the flag button is tapped (directly on the Survey screen) or when
the push-button on the designated flag cane is pressed.

Log GPS at DCP/Feature:


If this box is ticked, GPS position data will be logged automatically at
Devices or Geo-Features when either the Device button is tapped on
the Survey screen and a Device reading is logged or when the Geo-Feat.
button is tapped on the Survey screen and a geo-feature is registered.

Log GPS at Sidedrain/Anomaly:


If this box is ticked, GPS location data will be logged automatically when
DCVG defects (anomalies) are marked.

29

Auto Log GPS:


By tapping on the drop down menu button in the Auto Log GPS field, the
selections available will be displayed as indicated below.

By selecting one of these options, you can elect to have the GPS position
data logged automatically at every survey reading, at every second reading,
at every fifth reading, at every tenth reading, or not at all (never) at survey
readings.

Step 14:
Select Max mV Reading to be First Reading in Total mV Determination.
The last box in the above screen is actually not related to GPS Options.
If this box is ticked, the Max mV voltage value recorded at an anomaly
location will automatically become the first voltage value used by the data-
loggers software to calculate the Total mV (total voltage gradient).
Otherwise, if this box is not ticked, you will have to repeat the Max mV
recording a second time as part of the Total mV determination process.

Step 15:
Finally on the above screen, select the electricity supply operating frequency
of the country in which you are performing the surveys (60Hz or 50Hz). For
the U.S., select 60Hz.

30

Tap on the Next button on the above screen. The screen shown below will
appear.

Step 16:
Enter Pipeline Name and Survey Starting Location
Provide the Name of the Pipeline:
By tapping in the field labeled Name of P/L, you can type in the pipeline
name. Note: This is not the same as the filename for the DCVG survey that
was selected back at Step 6. This is the actual name of the pipeline.
Provide the Valve Segment Identification Number:
By tapping in the field labeled Valve Segment, you can type in the valve
segment name or number (if known).
Provide the Starting Location:
Provide the Starting Location for the Survey
You can select to have location information displayed on the survey screen
as station number, feet or milepost (station number, meters or kilometers for
the metric case).

31

Whichever selection you make here will determine how you enter your
starting location information.
For example, if your pipeline locations are represented by station numbers,
you would select Station Number from the drop down list and you would
enter a starting location for the survey in the form of a station number. [If
you do not know the station number where youre beginning your survey,
enter 0+0.0].
As an example, if you are working on pipeline ABC within valve segment
45 and you are about to begin a DCVG survey at station number 12+00.0,
your screen would be as shown below.

Step 17:
Select Voltage Recording Interval (Distance Between Recordings)
By tapping in the field on the above screen labeled Distance Per Reading,
you can type in the voltage reading interval (distance in feet (or in meters for
the metric case) expected between recordings) for the DCVG survey. In
DCVG work this expected interval distance is typically 5.0 feet.
32

Step 18:
Select Survey Flag Interval (Distance Between Survey Flags)

By tapping in the box on the above screen labeled Distance Between


Flags, you can type in the survey flag interval (distance between survey
flags in feet, or in meters for the metric case) for the section of pipeline
being measured. Typically, survey flags are located at 100 feet intervals.

Step 19:
Select the maximum permissible error between the actual number of
recordings made between 2 survey flags and the expected number of
recordings.

By tapping in the field labeled, CIS Flag Read Count %, you can type in
the maximum permissible error. For example, the maximum permissible
error is indicated as 20% on the above screen. If the recording interval is
expected to be 5.0 feet and the survey flag separation is 100 feet, that means
that 20 recordings are expected. If, however, only 15 recordings are
actually made between survey flags, an error window will appear on the
screen, since there is a 25% difference between the expected and actual
number of recordings made. No error window will appear if the difference
is less than 20% for this example, ie, you could have a minimum of 16
recordings and a maximum of 24 recordings between survey flags to stay
within the 20% (max.) error allowance.

Step 20:
Select whether or not you would like the recordings to be uniformly spaced
between survey flags, in cases where less than or greater than 20 recordings
are made.

By tapping in the box labeled, Auto Pacing Mode, and inserting a tick in
the box, you will enable the data-logger to automatically adjust the actual
recordings and space them evenly over the flag spacing distance, regardless
of the actual number of recordings made. Again, enabling this selection is
recommended.

Tap on the Next button on the above screen. The screen shown below will
appear.

33

Step 21:
Provide the Work Order Number for the DCVG survey.
By tapping in the field on the above screen labeled, Work Order #, you
can type in the work order number for the DCVG survey.

Step 22:
Provide Your Name.
By tapping in the field labeled, Technician Name, you can type in your
name or the name of your supervisor. If you use your supervisors name
here, you might add your own name in the Comments Section.

Step 23:
Provide Comments.
By tapping in the field labeled, Comments/Description, you can enter any
comments you might have regarding the survey (perhaps weather conditions,
soil conditions etc.).
Also shown on the above screen are the Survey (File) Name and the Survey
Start Date and Time.

34

Note: Do not attempt to change the File Name indicated here as this
identification will be required by the ProActive software to transfer
your DCVG survey data to your PC.
Tap in the Next button on the above screen. The screen shown below will
appear.

Step 24:
Select Voltmeter Settings and Current Interruption Cycle.
Voltmeter Reading Mode
DCVG surveys are performed with the CP systems rectifier-current
switched ON and OFF in a cyclic fashion and, as can be seen from the above
screen, there are two choices available for the data-loggers voltmeter Read
Mode. The choices are On/Off Pairs (DSP mode) and On/Off Pairs
(GPS Sync).
Note: On/Off Pairs (Min/Max) is not a option here since DCVG surveys
require a measurable difference between On and Off voltages, ie, they
require a significant IR drop.

35

On/Off Pairs (DSP mode):


In this mode, the voltmeter uses digital signal processing techniques to
determine the flat portions (for both the On and Off parts of the cycle) of
the rectifier-current cycle. V values are then determined for the On and the
Off portions of the cycle from which the DCVG voltage is calculated
[V(On) minus V(Off)].
For these measurements to be successful, the data probes (reference
electrode canes) must remain in good electrical contact with the soil for at
least one rectifier-current waveform period following the initial contact of
the probes with the soil, for a stable DCVG voltage value to be displayed.
On/Off Pairs (GPS Sync):
This voltmeter mode could be selected if you were using an MCM GPS
receiver on your DA Meter and current-interrupter switches equipped with
GPS units were being employed on the rectifiers. In this case, V(On) and
V(Off) values are based on potentials measured at pre-determined
sampling times on each waveform cycle, which are synchronized with the
current-interruption switching timing.
Also, in this case, you would tap on the GPS Settings button, which
would bring up the screen shown below.

You are being asked here to make several selections.


36

First, select the type of GPS unit you will be using (the data-loggers
Internal unit or the MCM External unit ONLY).

Next, enter your On Delay and your Off Delay selections by tapping in
the appropriate box and typing in the requested delay time in milliseconds.
These delay times are employed so that any spiking in the voltage waveform
that might occur as the rectifier-current is switched from ON to OFF and
from OFF to ON does not become a factor in the voltmeters determination
of the true ON and OFF voltage values for each DCVG measurement. For
example, if 150 ms was selected for the Off Delay, the data logger would
record the voltage value sampled 150 ms after the rectifier-current was
switched from the ON to the OFF state. Also, if 150 ms was selected for the
On Delay the data logger would record the voltage value sampled 150 ms
after the rectifier-current was switched from the OFF to the ON state.

Next, select your current interrupters Downbeat timing. As indicated by


tapping on the pull-down-list arrow button in the Downbeat box, there are
3 industry-standard selection choices; Each Minute, Each Hour and
Midnight. If Each Minute is an option for your current interrupter, we
suggest making this selection. The Each Minute designation indicates to
the data-logger that a new interruption cycle will begin at the top of each
minute. Consequently, the data-logger can reference its timing to the top of
each minute. In this instance, your On plus Off times (your waveform
period) should be an even multiple of 60.

Again, for these measurements to be successful, the data probes (reference


electrode canes) must remain in good electrical contact with the soil for at
least one rectifier-current waveform period following the initial contact of
the probes with the soil, for a stable DCVG voltage value to be displayed.

Finally, if your interruption cycle starts with the current in the ON state (the
first transition is from ON to OFF), place a tick in the Start Cycle box
(remove the tick if the opposite is true).

After making your GPS Settings selections, tap on the OK button which
will return you to the Voltmeter Settings screen.

37

Select the Cycle On/Off Times:


The specific On and Off times for the particular current-interruption
waveform cycle that will be employed during your DCVG survey should be
entered in the Cycle (ms) fields. The On and the Off times should be
entered in milli-seconds.

Note: After a time period corresponding to at least one waveform period


(On time plus Off time), the DCVG voltage value displayed on the DA
Meter survey screen should stabilize (assuming that the data probes are
making good electrical contact with the soil) and a DCVG voltage recording
can be triggered.

Consequently, the rate at which DCVG voltages can be captured and saved
depends critically on the interruption waveform period. This is a particularly
important issue in the case of large period waveforms. For example, if a 10
second On and a 2 second Off waveform is employed (12 second waveform
period), stabilization times up to 12 seconds could be possible before each
DCVG voltage recording can be made. In contrast, the stabilization period
would be only on the order of one second for a 700ms On, 300ms Off
interruption cycle.

Select Voltmeter Range and Input Impedance Settings:


By tapping on the pull-down list arrow button in the Range field, you can
select the voltmeter range and input impedance settings for your application.

The recommended settings for the voltmeter for DCVG surveys are 5.7
Volts (full-scale) and an input impedance value of 400MOhm, These
settings provide a relatively-fast response time (~80 ms), which is important
in rectifier-current-interrupted applications and also a relatively-sensitive
scale with regard to the magnitude of the DCVG voltages that youll be
measuring.
Note: Caution has to be exercised if any of the ranges beginning with the
number 4 (for example, 4V, 400mV or 40mV) are selected for DCVG
applications, since these represent relatively-slow response settings. Such
settings provide a higher level of AC filtering (less noise), but at the expense
of speed of response. Rather than the response time of 80ms for the 5.7V
range, these settings have a response time greater than 1 second.
Consequently, such settings are not appropriate for fast current-interruption
cycles, ie, interruption On and Off times shorter than 1 second, as true On
and Off potentials would not be recorded.

38

Step 25:
Pull-up Active Survey Screen
By tapping on the OK button on the Voltmeter Settings screen, the
active Survey screen would appear and you would be ready to
proceed with your DCVG survey.
An example starting active Survey screen is shown below.

In the above active DCVG survey screen, the direction of the arrow
(pointing to the left or pointing to the right) depends on the polarity of the
DCVG voltage being measured. In the above case, the voltage being
measured was positive which produced an arrow pointing to the left.
The opposite would be true if the DCVG voltage reading had been negative
(ie, the arrow would point to the right).
With the reference electrodes properly positioned, the polarity of the DCVG
voltage will always be positive in the case of perpendicular DCVG surveys.
In this case, the arrow on the survey screen will always point back to the
pipe.

39

However, a change in arrow direction will be evident when performing


parallel, or in-line, measurements in the vicinity of a DCVG anomaly
(coating defect). In this case, the arrow will always point in the direction of
the defect location on the pipe with respect to where you are currently
standing. As you approach the defect, the arrow will point towards the
upcoming defect location and when you are past the defect, the arrow will
point back to the defect location. Consequently, performing in-line
measurements in the vicinity of a defect can be a good way to pin-point its
precise location, by monitoring the arrows polarity switches. When you are
actually straddling a defect location (with the reference electrodes in-line
with the pipe), the DCVG voltage will be zero (ie, neither positive nor
negative).

Also, with respect to the above active survey screen, as DCVG voltages
are recorded by the data-logger, the Total Distance (total distance from the
start of the survey) parameter will increase in increments of 5.0 feet , or
whatever the Distance Per Reading value was that was entered back at
Step 17. (Distances would be in meters for the metric case).

Also, the Distance From Flag parameter will increase in the same
increments as voltages are recorded. The difference in this case, however,
will be that when each survey flag is registered, this distance parameter will
begin again at zero. In other words, this will show the distance you are
assumed by the data-logger to have traveled from the last flag that you
encountered (and registered).

You are now ready to perform a DCVG Survey.

40

SECTION IV: TEST EQUIPMENT HOOK-UPS FOR


DCVG SURVEYS
IV. 1 How to make Cable Hook-Ups for DCVG Surveys
The cable connections for DCVG surveys employing MCM test equipment
are illustrated in Figure 9.

Figure 9: Cable Connections for Hook-Up of MCMs DCVG Survey


Test Equipment

41

As can be seen from Figure 9, a pair of canes (reference electrodes see our
CIS Training Manual) is illustrated, the left-hand cane (RED-handled cane)
and the right-hand cane (GREEN-handled cane). These canes, which will be
placed on the soil above the pipeline in either the Perpendicular or the
In-Line configuration (see Section II), have push buttons on top of the
handles so that the operator can trigger voltage recordings on his
command at each of the pipeline survey measurement locations as well as at
Devices and Geo-Features. See Section III, Step 10 for a discussion on
cane button functionality (Step 10 of the G1s Set Up Process).

The canes (data probes) are connected as shown to the input terminals of
the dual-probe adapter and the output terminal of the adapter is connected
to a 5-pin Data Probe Connector socket on the top side of the G1.

NOTE: The left-hand reference electrode (red-handled cane) is connected


to the dual probe adapter via a black-band cable while the right-hand
reference electrode (green-handled cane) is connected to the dual-probe
adapter via a red-band cable. This is important since the right-hand
reference electrode connects to the positive side of the data loggers
voltmeter, while the left-hand reference electrode connects to the negative
side of the voltmeter. The red-band and the black-band cables can be
connected to either one of the dual-probe adapters input terminals.

The GPS Antenna illustrated on the top side of the G1 in Figure 9 represents
the antenna employed by the data-loggers internal GPS receiver. External
GPS receiver units, if employed, would be connected to the data-logger via
one of the two USB COM terminals located on the bottom side of the G1.
(The right-hand-side COM Port (COM 4) is recommended. With either the
internal unit or an external GPS unit, the location of items such as flags,
devices, geo-features and anomalies can be recorded during the performance
of a DCVG survey, either manually by tapping on the Log GPS button on
the survey screen at each critical location or automatically, by pre-
programming the data-logger as described above in Section III (Step 13).

IV. 2 How to Attach Cables and Accessories to the G1

The terminals on the top side of the G1 data-logger for the various
connections described above are illustrated in Figure 10 below.

42

Figure 10: Connection terminals on top side of G1 Data-Logger


The 5-pin Data Probe Connector Socket is shown above as Terminal 1. As
discussed above, the reference electrode canes are connected to this terminal
via the dual probe adapter. Again, the reference electrodes are effectively
connected to the positive (green-handled cane) or negative (red-handled
cane) sides of the voltmeter with this connection.
Consequently, the red and black banana plug terminals (Terminals 2 and 3)
are not used in this DCVG survey application.
However, the red banana plug terminal (Terminal 3) is used to connect
(temporarily) a test cable between the G1 and a test station (TS), in order to
record pipe-to-soil voltages and voltage waveforms prior to beginning a
DCVG survey (see Sections V. 2 and V. 3).
The GPS Antenna shown above is employed by the data-loggers internal
GPS unit. External GPS units are connected to either the left-hand-side
COM Port (COM 1) or the right-hand-side COM Port (COM 4) located on
the bottom side of the data-logger. As mentioned above, in Section III, the
COM 4 Port is recommended to avoid conflict with ActiveSync.
43

SECTION V: HOW TO PERFORM DCVG SURVEYS

V. 1 How to Carry the Test Equipment During a DCVG Survey

With the MCM test equipment connected as shown in Figure 9 (Section IV),
and the G1 data-logger set up as described in Section III, you are ready to
perform a DCVG survey.

To make a pipeline survey more manageable, MCM has developed a special


harness (belt pack) which allows the G1 to be carried around the waist area
in a hands-free fashion, allowing the individual to be able to position the
reference electrodes (canes) on the ground (every 5.0 feet, or so, down the
length of the pipeline) and to be able to trigger the push button canes when
appropriate to do so.

With the harness assembly, the G1 sits on a tray at waist level allowing the
operator to view the screen at all times and to make any selections required
by tapping on the screen. Also, the dual-probe adapter shown in Figure 9, is
attached to the underside of the tray, allowing convenient (5 pin cable)
connection of the adapters output to the G1. [Again, this will effectively
connect the green-handled reference electrode cane to the positive side of the
data loggers voltmeter and the red-handled reference electrode cane to the
negative side of the voltmeter].

In addition, an external GPS receiver unit (together with its battery pack and
antenna), if the internal GPS unit is not being employed, can also be attached
to the waist band of the harness, typically on the operators back at waist
level, with the antenna rising to above-head height.

V. 2 Readings to Take at Beginning of a DCVG Survey

It is important to begin your DCVG survey at a test station or at some other


device that allows you to make electrical connection temporarily to the
pipeline. Connection to the pipeline is required so that an IR drop value
can be determined for use in the softwares calculation of % IR.
(see Section II. 7).

44

Note: A significant IR drop value is necessary in order to be able to


detect DCVG anomalies. For example, for the 5.7V, 400M voltmeter
setting, a minimum IR drop value of ~200mV is suggested for the
following reason:

If it is assumed that you would like to mark defects having, as a


minimum, a 10% IR size, the minimum Total mV value, in the
example case of a 200mV IR drop, would be 20mV. This means, in
turn, that the minimum Max mV value that youd be measuring
would only be ~10mV or so, since, typically, about 50% of the total
voltage gradient is accounted for in the first side drain reading (the Max
mV reading). Since the noise level associated with the 5.7V, 400M
voltmeter setting is around 5mV, you can see why ~200mV would be a
suggested minimum IR drop value. An IR drop value around 500mV
would make small size anomaly detection significantly easier.

Alternatively, a lower noise level voltmeter setting could be used, for


example, the 400mV, 10M setting, which as a noise level around
1mV. In such a case, the minimum IR drop value required to
reasonably detect a 10% anomaly would be around 40mV, by the above
reasoning. However, as indicated in Step 24 of Section III (Select
Voltmeter Range and Input Impedance Settings), caution has to be
exercised when selecting a range beginning with the number 4, such as
400mV or 40mV, as these settings have slower response times than the
5.7V range, for example. Since, the response time on the 400mV setting
is around 1 second (as opposed to 80ms for the 5.7V setting), the current
interruption On and Off times need to both be greater than 1 second, so
that the voltmeter will read accurate On/Off potentials for each cycle.

With the CP systems current being switched ON and OFF in a prescribed


cyclic fashion, you can make pipe-to-soil voltage recordings with a slight
adjustment to the equipment hook-up arrangement shown in Figure 9.

By temporarily disconnecting the positive (green-handled) cane from the


dual-probe adapter and connecting a test cable from the red banana plug
terminal on the top side of the G1 to the test station, you can record pipe-to-
soil voltages during the ON and the OFF cycles, which will allow the
software to determine the IR drop at this test station location.

45

For the pipe-to-soil measurements, the negative cane (red-handled) cane


should be positioned over the pipe. As described below (Section V. 4), a
device reading can then be taken by tapping on the Device button on the
Survey screen and proceeding as indicated. It is also recommended that you
examine and record the pipe-to-soil Voltage Waveform at this location as
described below (Section V. 3)
V. 3 How to Record the Pipe-To-Soil Voltage Waveform
When you are performing a DCVG survey, it is recommended that you
examine the pipe-to-soil voltage waveform at your starting location (starting
test station) and it is suggested that you make a recording of this voltage
waveform using your G1 data-logger. The nature of the voltage waveform
that you are examining will reflect the nature of the rectifier-current
waveform that is currently in effect on your pipeline.
With the red-handled reference electrode cane making good electrical
contact with the soil above the pipe at the first test station and a test cable
connected from the test station to the red banana plug terminal on the G1,
you can examine the pipe-to-soil voltage waveform, either by tapping on the
WAVE button at the bottom of the Survey screen or by tapping on
Options followed by Wave. The screen shown below will appear.

46

By tapping on the Read button on the above screen, you can view the
actual pipe-to-soil voltage waveform. A typical waveform would be as
shown in the screen below.

Since we set up a 700ms ON and a 300ms OFF rectifier cycle, the pipe-to-
soil voltage is High for 0.7 sec. and Low for 0.3 sec., as evidenced by
the 1 second snap shot of the voltage waveform shown in the above
screen. Since the voltmeter in the G1 has a finite response time (~80ms in
the case of the 5.7V, 400M setting), the transitions from High to Low and
Low to High are not perfect (right-angled) steps.
However, these voltage waveforms are very useful with regard to confirming
multiple-rectifier synchronization and to selecting On Delays and Off
Delays to avoid voltage spikes (see Section III, Step 24), for example.
To record a pipe-to-soil voltage waveform, tap on the save button on the
above screen. The screen shown below will appear.

47

You are being asked here to establish a file in which the waveform data will
be stored (saved). The name of this file will be the name you type into the
Name box on the above screen. For ease of future retrieval when you
uploaded your survey data to the Database Management software package
(ProActive) [see Section VI], it is recommended that you use the same name
as your survey filename with the addition of something like Wave TS 1 to
the file name. This would represent the waveform data recorded at Test
Station 1 at the beginning of your survey.
Note: Make sure that the target folder is G1_Data, as indicated above.
You should then tap on the OK button. This process will save the
waveform data in the G1s CompactFlash memory for future retrieval by
ProActives driver.
V. 4 How to Record the Pipe-To-Soil Voltages at a Test Station:
In order to record the High and the Low voltages (also known as the ON and
the OFF voltages), tap on the Device button on the Survey screen. The
Device screen shown below will appear with Single Test Station
selected.

48

By tapping the Next button on the above screen, the screen shown below
will appear.

As shown above, the High and Low voltage readings (for each cycle)
measured at the test station will be displayed on this screen.
The voltages shown above represent a snap-shot in time.

49

The software will calculate the difference between the High and the Low
voltage recordings (which is the IR drop) and use this quantity in additional
calculations (%IR calculations). The "Use reading for DCVG signal
strengh" setting defaults to "Yes" (box checked).
The IR drop is also referred to as the signal strength.
You should then tap on the Save button to save this data.
Note 1: You should perform pipe-to-soil measurements at each test
station you encounter in order that the software can perform the linear
interpolation calculation described in Section II. 7, when necessary, to
determine an average signal strength (IR drop) to apply at each defect
location between test stations.
Note 2: You can "edit" which device readings (pipe-to-soil readings) will
be used in signal strength calculations using the following procedure:
Tap on the "Options" menu, tap on "Edit Survey Data", tap on "Retrieve",
locate and highlight the device in question (via the "Mode" column) and
tap on the "View Details" button. Next, scroll across to the "Use for %IR"
column and highlight the cell. Next, select Y or N.

V. 5 How to Locate and Mark Defects

As mentioned above, when possible, the CP system should be adjusted to


provide a signal strength (IR drop) of at least 200mV, in the case of the
5.7V, 400M voltmeter setting. This will make it easier to find the smaller
defects as well as the larger ones.

Once you have recorded the pipe-to-soil voltages and observed the
magnitude of the signal strength (IR drop), you should disconnect the test
cable from the G1s red banana plug terminal and reconnect the positive
(green-handled) cane to the dual-probe adapter. You are now ready to
record DCVG voltages down the length of the pipeline.

For Perpendicular DCVG surveys, you would place the reference electrodes
(canes) about 4 to 5 feet apart as indicated in Figure 4 (Section II) with the
left-hand cane (negative cane) directly over the pipe and the right-hand cane
(positive cane) off to the right-hand side. If you set up the cane push-button
functionality as indicated in Section III (Step 10), ie, with the right-hand
cane set to trigger voltage recordings and the left-hand cane set to designate
the location of survey flags, you would trigger the right-hand cane every 5
feet or whatever recording interval you set up in Step 17 of the set-up
process (see Section III).

50

As discussed in Section II. 2, if you are outside the voltage gradient field
of a coating defect, the DCVG voltages will essentially be zero. However,
as you enter a defects voltage gradient field, you will observe an increase in
the DCVG voltage values displayed on the G1s Survey screen (see Figure 5
in Section II. 2). When you observe a peak (maximum value) in the DCVG
voltage readings, you should interrupt your walking and Mark the location
of the maximum DCVG reading. This will represent a defect location.
In order to mark the defect location, you should tap on the mark button on
the Survey screen. The screen shown below will appear.

As discussed in Section III (Step 24), the arrow on the Survey screen will
always point back to the pipe in the perpendicular survey case.
If you selected automatic logging of GPS position data at
Sidedrains/Anomalies back at Step 13 (Section III), the G1 data-logger will
log the GPS position data for this DCVG Anomaly. If you did not select this
option, you should tap on the Log GPS button at this time to log the defect
location, if you are using a GPS Unit.

51

If you are satisfied that you are reading the maximum DCVG voltage value
for this defect, you should record the Max mV reading by tapping on the
radio button labeled Read Max mV and tapping on the Accept button
(do not tap on the Save button at this time). The screen show below will
appear.

In the above example, the Max mV reading was 65.1mV [delta V (ON) was
+104.7mV and delta V (OFF) was +39.7mV and the difference between
these values is 65.1mV; this was a cathodic/cathodic situation, since both
polarities were positive].
You will also notice that the software has automatically applied the Max
mV value to the Total mV determination, since we selected this option
back at Step 13 (Section III). If this were not the case, you would Accept
the Max mV reading a second time and this would become the first voltage
used in the Total mV determination.

52

You should then proceed to move the electrodes to their second positions
(see Section II. 5) and you should Accept the second reading. You should
proceed in this fashion until you are outside of the defects voltage gradient
field, ie, the DCVG voltage reading is essentially zero (typically DCVG
voltages less than about 5mV would be considered essentially zero, on the
5.7V, 400M voltmeter setting).
The screen will be as shown below.

As can be seen from the above screen, the software has generated a sum of
the voltage recordings required for us to reach remote earth, ie, be outside of
the voltage gradient field associated with this defect. This represents the
Total mV value, which in our example is 199.9 mV.
At this point, you should tap on the Save button which will save all of the
data associated with this defect.
You would then pick up the survey where you left off and continue
triggering DCVG voltage recordings until you encounter the voltage
gradient field of another defect.

53

SECTION VI: HOW TO COPY SURVEY FILES FROM


THE G1 DATA-LOGGER TO YOUR PC

VI. 1 Introduction

Survey data are stored in independent files (one file for each survey) on the
CompactFlash memory card on the G1 data-logger and you can copy survey
files to your PC using one of two approaches; manually or via the driver in
the ProActive software program.

The ProActive software program represents MCMs CP data management


system and this program allows integration of pipeline survey data in a
database system and offers extensive reporting (both textual and graphical)
capabilities on the survey data.

If you have the ProActive software program installed on your PC, or you can
bring your G1 to a PC that has ProActive installed on it, you can use the
ProActive program to automatically access survey files on the G1. If either
of these situations applies, you would proceed to Section VI. 3.

Note: The ProActive software program is required to actually view and


analyze the survey data.

If you do not have the ProActive software program installed on your PC and
you cannot bring your G1 to a PC that has ProActive installed on it, you can
copy survey files manually from your G1 to your PC and you can
subsequently send the copied files to a recipient that has ProActive installed
on their PC. If this situation applies, you would proceed to Section VI. 2.

Note: In order to copy survey files from your G1 to your PC, either
manually or via ProActive, you will need to have the Microsoft
ActiveSync software program installed on your PC.

54

VI. 2 The Manual Approach

Step 1:
Connect a communication cable between the left-hand side COM Port on the
G1 data-logger and your PC and switch on your G1. With Microsofts
ActiveSync program installed on your PC, the ActiveSync window will
appear on your desktop and the program will confirm that two-way
communication has been established.

Step 2:
Follow the procedures detailed below:

For the Q100 units:


* Double-click on My Computer on your PC
Double-click on Mobile Device
Double-click on My Pocket PC
Double-click on CompactFlash
Double-click on My Documents
Double-click on G1_Data
Right-click on the survey file you wish to copy
Copy the file to a local folder

For the Q200 units:


* Double-click on My Computer on your PC
Double-click on Mobile Device
Double-click on System
Double-click on My Documents
Double-click on G1_Data
Right-click on the survey file you wish to copy
Copy the file to a local folder

You would now be in a position to send the survey file to a recipient who
has access to the ProActive software program.

Note: Do not rename the survey file prior to sending the file to the
recipient, as the survey file must have the same name as the survey
itself.

55

VI. 3 Using the Driver in the ProActive Software Program

Step 1:
Create a folder on your PCs hard-drive that will be used to permanently
save files copied from your data-logger. You might choose to name this
folder something like, Surveys.

Step 2:
Connect a communication cable between the left-hand side COM Port on the
G1 data-logger and your PC and switch on your G1. With Microsofts
ActiveSync program installed on your PC, the ActiveSync window will
appear on your desktop and the program will confirm that two-way
communication has been established.

Step 3:
Double-click on the ProActive icon on your PCs desktop screen.

This will open up ProActives main menu window. A window labeled


Entire Database will also be seen here. The suggested organization of
your Entire Database is discussed in the ProActive Training Manual.

Step 4:
Click on the Surveys button on the main menu bar at the top of the screen.

This will open a window labeled Data Logger: Get Pipeline Survey.

By clicking on the pull-down-list arrow button in the Data Logger field,


you can select the data-logger from which you are copying the survey file.
As shown in the drop-down-list, the various data-loggers currently supported
by ProActive are offered as choices.

Step 5:
Select Gathererone (G1) Option.

Highlight Gathererone in the drop-down-list and click on the Go button.


This will open up a window labeled, Driver (Pipeline Survey).

Note: It may take a few seconds for the Driver Window to appear.

56

Step 6:
Select the Survey File to be Copied.
Select Pipeline Survey in the field labeled Data Type.
The Surveys field in the Gathererone Driver window will list all of the
survey files currently stored on your G1s CompactFlash memory
card. Highlight the survey file that you would like to be copied to your PC.
Also, place a tick in the box labeled Copy to Local Folder and identify the
folders location on your hard-drive in the field underneath.
This is the folder that you set up previously in which to save all of your
survey files copied from your G1. If you named the folder Surveys, the
folders location would be: C:\Surveys

Also, if you used the metric system on the data-logger, check off the box
labeled, Use Metric.

57

Click on the Go button.

Step 7:
Examine the Survey Data Prior to Bringing the Data into the Database
Management Section of ProActive.

You will have actually completed the process of copying a survey file to a
local folder on your PC by this point, using the Driver in ProActive.

However, before exiting the Driver, it is recommended that you examine


your survey data prior to bringing the data into the Database Management
Section of ProActive. The process of bringing survey data into the database
management section of ProActive is detailed in the ProActive Training
Manual.

To do so, click on the cancel button on the window that is currently


showing (Data Logger: Get Pipeline Survey Window). The G1 Driver
window will again be shown but, this time, there will be a selection of page
tabs on the window, labeled as follows:

Survey Settings
Readings
Device Readings
Graph

By clicking on these page tabs, you can view information on the Survey
conditions or you can view actual survey data (see the ProActive Training
Manual).

58

APPENDIX 1: How to Delete Survey Files from the G1

Once you have copied your survey files to your PC (see Section VI), you can
(if you wish) delete the files from your G1s CompactFlash memory card if
you need, for example, to create space in the data-loggers file storage
memory to make room for future survey files.

Your survey files are stored in the G1 on a non-volatile CompactFlash


memory card. Since this memory is non-volatile, your survey data will be
safe, even if all power is lost to your G1. Please note, however, that the
G1s survey programs (for example, G1 PLS) are stored in a volatile
memory, and so, if all power was lost (ie, the battery-pack and back-up
battery were fully discharged), you would need to re-install the G1s
software application package see the G1 Users Manual.

Since your survey files are stored on the CompactFlash card, you will have
to access this memory in order to delete selected survey files. The procedure
would be as follows:

For the Q100 units:


Tap on the Start button
Tap on Programs
Tap on the File Explorer icon
Tap on the CompactFlash card icon at the bottom of the screen
Tap on My Documents
Tap on the G1_Data folder icon
Tap and hold on the Survey File you wish to delete until a menu
appears
Tap on Delete. This will delete the highlighted survey file

For the Q200 units:


Tap on the Start icon
Tap on Programs
Tap on Windows Explorer
Double-tap on the Systems folder
Double-tap on the My Documents folder
Double-tap on the G1_Data folder

59

Tap and hold on the Survey File you wish to delete until a menu
appears
Tap on Delete. This will delete the highlighted survey file

60

You might also like