Microscope Instructon Manual
Microscope Instructon Manual
Microscope Instructon Manual
CX21
EDUCATION MICROSCOPE
This instruction
manual is for the
Olympus Education
Microscopes Model
CX21. To ensure the
safety, obtain optimum
performance and to
familiarize yourself
fully with the use of
this microscope, we
recommend that you
study this manual
thoroughly before
operating the
microscope. Retain
this instruction manual
in an easily accessible
place near the work
desk for future
reference.
IVD
AX7251
CX21
CONTENTS
IMPORTANT Be sure to read this section for safe use of the equipment.
1-3
1
STANDARD COMPONENT UNITS
4
2
NOMENCLATURE
5
7-11
1
Turning the Lamp ON ...........................................
7
2 Placing Specimen on the Stage
.................... 7
3
Adjusting the Focus ................................................
8
4 Adjusting the Interpupillary Distance.....................
9
5
Adjusting the Diopter ............................................
6
Adjusting the Condenser Position and Aperture Iris Diaphragm .............................................................
10
7
Switching the Objectives ............................
10
8 Using the 100X Immersion Objective.....................
11
ONE-POINT ADVICE
TROUBLESHOOTING GUIDE
SPECI
FICAT
IONS
OPTICAL CHARACTERISTICS
15
ASSEMBLY
16
10 OPTIONAL ACCESSORIES
12
17-18
13-14
19-21
10-1
System Diagram of Optional Accessories ..............................................................................
19
10-2
Installation and Operation of Optional Accessories .......................................
19-21
1
Cord Hanger CH3-CH ............................................................................................................................................................................................
19
2
Filter Holder CH2-FH ...............................................................................................................................................................................................
20
3
Darkfield Ring CH2-DS ........................................................................................................................................................................................
20
4
Reflection Mirror CH20-MM ..........................................................................................................................................................................
20
5
Eyepieces WHC15X/WHC15X-H.......................................................................................................................................................................................................
21
6
Using the eyepiece micrometer ........................................................................................................................................................................................................
21
.......................................................
IMPORTANT
SAFETY PRECAUTIONS
1.
After the equipment has been used in an observation of a specimen
uct.
2.
To avoid potential shock hazards and burns when replacing the lamp
@
bulb, set the main switch 1 to (OFF) then disconnect the power cord
from the wall outlet in advance. Whenever you replace the bulb during
Fig. 1
use or right after use, allow the lamp socket 2 and bulb to cool before
touching. (Fig. 1)
3.
Install the microscope on a sturdy, level table or bench so as not to block
4.
Always use the power cord provided by Olympus. If no power cord is
provided, please select the proper power cord by referring to the section
this instruction manual. If the proper power cord is not used, product
5.
When installing the microscope, route the power cord away from the
microscope frame. Should the power cord come in contact with a hot
part, the power cord could melt and cause electric shock.
6.
Always ensure that the grounding terminal of the microscope and that
of the wall outlet are properly connected. If the equipment is not grounded,
equipment.
7.
Never allow metallic objects penetrate into the air vents of the micro-
scope frame as this could result in electrical shock, personal injury and
equipment damage.
8.
After operation or in case of abnormality, be sure to disconnect the power
cord from the connector on the microscope or from the wall power outlet.
Safety Symbols
The following symbols are found on the microscope. Study the meaning of the symbols and always use the
equipment in the safest possible manner.
Symbol
Explanation
Indicates that the surface becomes hot, and should not be touched with bare hands.
Before use, carefully read the instruction manual. Improper use could result in personal injury to the user and/or
damage to the equipment.
CX21
Warning Label
A warning indication label is attached to every part where special precaution is required when handling and using the
microscope. Always heed the warnings.
Warning label
Bottom of microscope
[Warning against high temperature in
position
frame
lamp bulb replacement]
Getting Ready
1.
A microscope is a precision instrument. Handle it with care and avoid
2.
Do not use the microscope where it is subjected to direct sunlight,
@
3.
Always use the tension adjustment ring to adjust the rotation tension of
4.
The microscope is ventilated by natural convection. Be sure to leave
5.
When carrying the microscope, hold both sides around the hole of the
#To prevent damage, do not hold the microscope by the stage 1 or observation tube 2.
Clean all glass components by wiping gently with gauze. To remove fingerprints or oil smudges, wipe with gauze slightly
moistened with a mixture of ether (70%) and alcohol (30%).
Since solvents such as ether and alcohol are highly flammable, they must be handled carefully. Be sure to keep these
chemicals away from open flames or potential sources of electrical sparks for example, electrical equip-ment that is
being switched on or off. Also remember to always use these chemicals only in a well-ventilated room.
Do not attempt to use organic solvents to clean the microscope components other than the glass components. To clean
them, use a lint-free, soft cloth slightly moistened with a diluted neutral detergent.
Do not disassemble any part of the microscope as this could result in malfunction or reduced performance.
When not using the microscope, ensure that the frame is cooled down and store it in a dry locker or cover it with a dust
cover.
To clean the condenser, fully loosen the securing knob 1, then remove the condenser by lowering it with the condenser
height adjustment knob 2, and wipe the front lens of the condenser.
The condenser can be attached by reversing the above removal
procedure.
Fig. 3
Caution
If the
microscope is
used in a
manner not
specified by
this manual, the
safety of the
user may be
imperiled. In
addition, the
equipment may
also be
damaged.
Always use the
equipment as
outlined in this
instruction
manual.
The following
symbols are
used to set off
text in this
instruction
manual.
: Indicates
that failure to
follow the
instructions in
the warning
could result in
bodily harm to
the user and/or
damage to
equipment
(including
objects in the
vicinity of the
equipment).
# : Indicates
that failure to
follow the
instructions
could result in
damage to
equipment. } :
Indicates
commentary
(for ease of
operation and
maintenance).
This device
complies with
the
requirements of
directive
98/79/EC
concerning in
vitro diagnostic
medical
devices. CE
marking means
the conformity
to the directive.
NOTE: This
equipment has
been tested
and found to
comply with the
limits for a
Class A digital
device,
pursuant to
Part 15 of the
FCC Rules.
These limits are
designed to
provide
reasonable
protection
against harmful
interference
when the
equipment is
operated in a
commercial
environment.
This
equipment
generates, uses,
and can radiate
radio frequency
energy and, if
not installed and
used in
accordance with
the instruction
manual, may
cause harmful
interference to
radio
communications
.
Operation of this
equipment in a
residential area
is likely to cause
harmful
interference in
which case
FCC WARNING:
Changes or
modifications
not expressly
approved by the
party
responsible for
compliance
CX21
STANDARDCOMPONENTUNITS
}After opening the package, make sure that the correct component units for the selected set are present.
The differences between the CX21FS1 and CX21FS2 lie in the number of objectives and presence of immersion oil.
#The objectives have been screwed in tightly to prevent them from being loosened during transportation. To remove an objective, turn
it counterclockwise while holding it with a rubber sheet, etc., so that your fingers dont slip.
Binocular observation
Eyepieces
tube
(Fixed at 10X)
Revolving nosepiece
Objectives
Microscope frame
Stage
Condenser
Immersion oil
Optional Accessories
Reflection Mirror
Cord Hanger
Filter Holder
Darkfield Ring
CH3-CH
CH2-FH
CH2-DS
CH20-MM
Eyepieces
WHC15X
WHC15X-H
NOMENCLATURE
}The following items have been attached at the factory to prevent deterioration during transport. Remove these
items and retain them for future use. Then loosen the observation tube clamping knob and correct the orientation
of the eyepieces as shown in the illustration below.
}Attach the lamp bulb and power cord as described in chapter 9, ASSEMBLY on pages 17-18.
: Power ON
: Power OFF
}For detailed description of each control, refer to the page indicated inside parentheses.
CX21
SUMMARYOFBRIGHTFIELDOBSERVATIONPROCEDURE
(Controls Used)
(Page)
the brightness.
2Light intensity adjustment knob
(P. 7)
3Specimen holder
(P. 7)
5Revolving nosepiece
(P. 10)
5Revolving nosepiece
(P. 10)
(P. 20)
Start observation.
87
5
6
a
4
DETAILEDOBSERVATIONPROCEDURE
(Fig. 4)
Fig. 4
(Fig. 5)
@
#Place the specimen gently. If the bow-shaped lever is returned with
1.
Rotate the coarse adjustment knob 2 in the direction of the arrow to fully
2.
Open the bow-shaped lever 3 outward, place the specimen by sliding
the specimen glass plates on the stage from the front toward the rear.
|
3.
After sliding the specimen glass plates all the way, return the bow-shaped
lever 3 gently.
4. Rotating the upper knob which is the Y-axis feed knob 4 moves the
Fig. 5
specimen in the vertical direction. Rotating the lower knob which is the X-
#Do not move the specimen holder directly by hand, for this will
Cover glass
Cover glass
Fig. 6
Slide glass
Placing Specimen
Slide glass
CX21
1.
The horizontal coordinate can be read at position 1 on the specimen
holder.
2.
The vertical coordinate can be read at the position of index line 2.
Fig. 7
Focusing Procedure
Fig. 8
the stage.
Objective
4X
10X
40X
100X
Magnification
Fig. 9
(Fig. 8)
@
The tension of the coarse focus adjustment knob has been designed
WD (mm)
18.5
10.6
0.6
0.13
If the stage descends on its own or if the specimen gets out of focus
quickly even when it is brought into focus using the fine adjustment
knob, it means that the tension of the coarse adjustment knob is too
low. Turn the ring 1 in the direction of the arrow to increase the tension.
}The pre-focusing knob controls the mechanism for preventing collision between the specimen and objective.
1. After bringing the specimen into focus, turn the pre-focusing knob 1 inside the hole of the arm so that the prefocusing mechanism hits the stage guide.
To provide a certain margin for focusing, rotate the knob by about half turn backward from the stopped position.
#If the function of this mechanism is not required, set the pre-focusing
(Fig. 11)
}The interpupillary distance adjustment consists of regulating the two eye-pieces according to that between your eyes
so that you can observe a single microscopic image through two eyepieces. This greatly helps to reduce fatigue during
observation.
While looking through the eyepieces, move both eyepieces until the left and right fields of view coincide completely.
The position of index dot indicates the interpupillary distance value. }Note your interpupillary distance so that it can be quickly
duplicated.
Fig. 11
(Fig. 12)
}The diopter adjustment consists of compensating for the difference in eyesight between your eyes.
1. While looking through the right eyepiece with your right eye, rotate the coarse and fine focus adjustment knobs to
bring the specimen into focus.
While looking through the left eyepiece with your left eye, rotate only the diopter adjustment ring 1 to focus on the
specimen.
Fig. 12
Use with the eye shades in the normal, folded-down position. This will prevent the eyeglasses from being scratched.
Extend the folded eye shades in the direction of the arrow to prevent extraneous light from entering between the eyepieces
and eyes.
Fig. 13
CX21
(Fig. 14)
}The condenser is usually used in the highest position. If the entire observed field of view is not bright enough, brightness
may be improved by lowering the condenser slightly.
1. Rotate the condenser height adjustment knob 1 to move the condenser to the highest position.
2. The aperture iris diaphragm ring 2 has an objective magnification scale (4X, 10X, 40X, 100X). Rotate the ring so that the
magnification of the objective in use faces frontward.
Fig. 14
Fig. 15
10
8
Using the 100X Immersion Objective
(Fig. 16)
}The designated immersion oil should be attached to the top lens of the
to be focused on.
1.
Focus on the specimen using all objectives, starting from the lowest-
2.
Before engaging the immersion objective in the light path, place a drop
3. Rotate the revolving nosepiece to engage the immersion objective and rotate the fine adjustment knob to bring the
specimen into focus.
#Since air bubbles in the oil will affect the image quality, make sure
Fig. 16
To remove bubbles, rotate the revolving nosepiece slightly to move the oil-immersed objective by one reciprocation or
two.
}The condenser of this microscope manifests the full performance when oil is placed between the slide glass and the front lens of
condenser. If oil is not attached there, the observation image may become slightly dark.
After use, remove oil from the objective front lens by wiping with gauze slightly moistened with an ether (70%)/alcohol (30%)
mixture.
If immersion oil enters your eyes or contacts your skin, immediately take the following treatment.
Eyes: Rinse with fresh water (for 15 minutes or more). Skin : Rise with water and soap.
If the appearance of the eyes or skin is altered or pain persists, immediately see your doctor.
11
CX21
ONEPOINTADVICE
To track an image
moving leftwards
glass to the
specimen image
left.
rightwards):
To move the
Move the slide
specimen image
glass down.
upward:
}The image observed through the microscope moves in directions opposite to the actual up-down and left-right
movements of the specimen.
Total Magnification
The size of the specimen image for observation is obtained by multiplying the eyepiece magnification by the
objective magnification. This value is referred to as total magnification.
Resolution
Resolution determines how finely a lens is able to distinguish the details of a specimen.
The resolution of a microscope is mainly determined by the ability of the objective and is scarcely related to that of the
eyepieces. The only function of the eyepieces is to magnify an image already resolved by the objective.
Although both configurations above provide the same total magnification, the higher magnification of the objective
yields a better specimen resolution.
The field number is the diameter of the image observed through an eyepiece, represented in millimeters.
The actual field of view is the size on the specimen actually observed in the
microscope. This yields approximate size of the actual specimen.
Actual field of view
= 18/10 = 1.8 mm
Actual field of view = Field number/Objective magnification
12
TROUBLESHOOTINGGUIDE
Under certain conditions, performance of the unit may be adversely affected by factors other than defects. If problems occur,
please review the following list and take remedial action as needed. If you cannot solve the problem after checking the entire list,
please contact Olympus for assistance.
Problem
Cause
Remedy
Page
field.
path.
it clicks.
vation field.
and/or specimen glasses are dirty.
blurred or unclear.
path.
it clicks.
sion objective.
Olympus.
stage.
and secure using the specimen holder.
6. High-magnification objective
The specimen is upside down.
Set the specimen correctly with the
of the knob.
match.
properly.
not set.
13
CX21
Problem
Cause
Remedy
Page
magnification objective.
0.17 mm.
14
SPECIFICATIONS
Item
Specifications
1. Optical system
UIS (Universal Infinity System) optical system
2. Illumination
Built-in illumination system.
100-120/220-240 V
0.42/0.25A, 50/60 Hz
3. Focusing mechanism
Stage height adjustment mechanism.
4. Revolving nosepiece
Quadruple positions fixed (Front oriented)
18
30
Interpupillary distance
48 to 75 mm
adjustment range
6. Stage
Size
Movement range
76 (X-axis) x 30 (Y-axis) mm
Specimen holder
7. Condenser
Type
N. A .
Built in
9. Operating environment
Indoor use.
15
Focal depth : (Object side) : Field number (FN) : Actual field of view : Total power :
(Total magnification)
Resolution :
CX21
OPTICALCHARACTERISTICS
Magnification
Mechanical
tube length
The following table shows the optical characteristics of combinations of eyepieces and objectives. The figure on the
right shows the performance data engraved on the
objectives.
Color band
Number of
aperture
Cover glass
thickness
Plan Objective
Optical
Characteristics
W.D.
Cover
Resolution
Power
N.A.
Total
Focal
Actual
Remark
(mm)
Glass
(m)
Depth
Field of
Thickness
Power
Objectives
(m)
View
3.36
40X
175.0
4.5
10X
0.25
10.6
1.34
100X
28.0
1.8
40X
0.65
0.6
0.17
0.52
400X
3.04
0.45
100XO
1.25
0.13
0.27
1000X
0.69
0.18
Oil immersed
Legend
Working distance (WD) : Distance between the top surface of cover glass and the objective extremity.
Number of aperture (NA) : The figure corresponding to the F-number of the camera. This is associated with
the resolution and larger NA means higher resolution.
Ability of an objective for identifying adjacent two lines in the image, which is expressed in terms of the
minimum distance between two points on the specimen surface.
The depth range of a specimen, in which focusing is obtained at a time. Stopping down the aperture iris diaphragm
increases the focal depth and increasing the objective NA decreases it. The diameter of the image observed
through an eyepiece, represented in millimeters.
Diameter of the field of view, expressed as the size on the specimen surface. Objective magnification x
Eyepiece magnification.
16
ASSEMBLY
}Each standard set can be assembled by simply attaching the lamp bulb, daylight (blue) filter and power cord.
(Fig. 17)
#Before attaching the lamp bulb, remove the parts that may drop
such as the filter and specimen from the microscope frame, and
@
place it on the back so that the bottom can be seen from the front.
1.
Pull the lock knob 1 on the bottom to open the lamp bulb replacement
cover.
2.
Hold the halogen lamp bulb 2 without taking it out of the polyethylene
bag so as not to stain the bulb with fingerprints or stains, and push the
bulb into the pin holes on the socket 3. After attaching, remove the
Fig. 17
polyethylene bag from it.
Always use the designated bulb. Using a bulb other than a specified one may lead to a fire hazard.
Fingerprints or stains on the lamp bulb reduce its service life. When it is contaminated, wipe with a cloth slightly moistened with
alcohol.
With the lock knob left in the pulled-out position, close the lamp bulb replacement cover. Then push in the lock knob to lock the cover.
#The cover cannot be closed if the lock knob is in the pushed-in position. Make sure that it is in the pulled-out position before
closing the cover.
The bulb, lamp socket and areas near these will be extremely hot during and right after use. Set the main switch to
(OFF),
discon-nect the power cord from the wall outlet, then allow the old bulb and lamp socket to cool before replacing the bulb with a
new bulb of the designated type.
}This filter modifies the color of observation light into a natural color (day-
@
light color).
Slide the daylight filter 1 into the bottom of the condenser 2 until it
Fig. 18
17
CX21
| 5
Fig.
20
F
i
g
.
1
9
3
Co
nne
ctin
g
the
Po
wer
Cor
d
(Figs
. 19
& 20)
T
h
e
p
o
w
e
r
c
o
r
d
i
s
v
u
l
n
e
r
a
b
l
e
w
h
e
n
b
e
n
t
o
r
t
w
i
s
t
e
d
.
N
e
v
e
r
s
u
b
j
ect it
to
exce
ssive
force
.
M
ake
sure
that
the
main
switc
h1
is
set
to
(OF
F)
befor
e
connecti
ng
the
pow
er
cord.
Al
ways
use
the
pow
er
cord
provi
ded
by
Oly
mpu
s. If
no
pow
er
cord
is
provi
ded,
plea
se
sele
ct
the
prop
er
pow
er
c
o
r
d
b
y
r
e
f
e
r
r
i
n
g
t
o
t
h
e
s
e
c
t
i
o
n
P
R
O
P
E
R
S
E
L
E
C
T
I
O
N
O
F
T
H
E
P
O
W
E
R
SUP
PLY
COR
D at
the
end
of
this
instr
uctio
n
man
ual.
1.
Con
nect
the
pow
er
cord
s
conn
ector
2 to
conn
ector
3
firml
y.
B
e
sure
to
supp
ly
pow
er
from
a
grou
nded
, 3cond
uctor
pow
er
outle
t
usin
g the
prop
er
pow
er
cord.
If the
pow
e
r
o
u
t
l
e
t
i
s
n
o
t
g
r
o
u
n
d
e
d
p
r
o
p
e
r
l
y
,
O
lymp
us
can
no
long
er
warr
ant
the
elect
rical
safet
y
perfo
rman
ce of
the
equi
pme
nt.
2.
Con
nect
the
pow
er
cord
s
plug
| to a
wall
pow
er
outle
t 5.
18
OPTIONALACCESSORIES
101SystemDiagramofOptionalAccessories
Eyepieces
WHC15X WHC15X-H
Eyepiece micrometer
CH2-FH
Microscope
Frame
CX21
Cord Hanger
Filter Holder
Reflection Mirror
45 mm filter
CH3-CH
CH2-DS
102InstallationandOperationofOptional
Accessories
Fig. 21
Cord Hanger
CH3-CH
@
|
Removal
Fig. 22
19
Filter Holder
CH2-FH
CX21
Fig. 23
(Fig. 23)
Darkfield Ring
CH2-DS
Filter
Application
32.5C
Changes the light of the built-in lamp bulb (which
is yellow-
32.5G533
Reduces fatigue or provides specimen with
contrast.
32.5LB45,
150, 200
light film).
(Fig. 24)
Insert the darkfield ring in the CH2-FH filter holder in the same way as inserting
a filter.
The darkfield ring enables darkfield observation using an objective from 4X to
40X.
CH2-DS
Fig. 24
(Fig. 25)
The eyepieces should be oriented toward the rear so that the reflec-
tion mirror receives the bright light. Loosen the observation tube
1.
Fit the reflection mirror in the window lens of the microscope frame by
2.
Remove the daylight filter because it renders the image bluish under
@
natural light.
3.
The reflection mirror 1 should be pointed toward the bright area. While
Fig. 25
observing the image through eyepieces, adjust the orientation of the
reflection mirror.
20
5 Eyepieces WHC15X/WHC15X-H
(Fig. 26)
1. Using a small flat-blade screwdriver, loosen the clamping screw 1 of a 10X eyepiece and remove it.
2. Insert the WHC15X or WHC15X-H into the eyepiece sleeve and tighten the clamping screw 1.
Fig. 26
@
as well as the WHC15X and WHC15X-H.
performed, so those with poor eyesight will have trouble in bringing the
Remove the micrometer sleeve and fit the micrometer @ into the eyepiece with the indication side facing down as
shown in Fig. 27. Replace the micrometer sleeve to use the micrometer.
Fig. 27
21
CX21
If no power supply cord is provided, please select the proper power supply cord for the equipment by referring to
Specifications and Certified Cord below:
CAUTION: In case you use a non-approved power supply cord for Olympus products, Olympus can no longer warrant the
electrical safety of the equipment.
Specifications
Voltage Rating
125V AC (for 100-120V AC area) or, 250V AC (for 220-240V AC area)
Current Rating
6A minimum
Temperature Rating
60C minimum
Length
3.05 m maximum
Fittings Configuration
Grounding type attachment plug cap. Opposite terminates in molded-on IEC con-
A power supply cord should be certified by one of the agencies listed in Table 1 , or comprised of cordage marked with an
agency marking per Table 1 or marked per Table 2. The fittings are to be marked with at least one of agencies listed in Table 1.
In case you are unable to buy locally in your country the power supply cord which is approved by one of the agencies
mentioned in Table 1, please use replacements approved by any other equivalent and authorized agencies in your country.
Country
Agency
Certification
Country
Agency
Certification
Mark
Mark
Argentina
IRAM
Italy
IMQ
Australia
SAA
Japan
JET, JQA, TV,
UL-APEX / MITI
Austria
VE
Netherlands
KEMA
Belgium
CEBEC
Norway
NEMKO
Canada
CSA
Spain
AEE
Denmark
DEMKO
Sweden
SEMKO
Finland
FEI
Switzerland
SEV
France
UTE
United
ASTA
Kingdom
BSI
Germany
VDE
U.S.A.
UL
Ireland
NSAI
22
Approval Organization
tion Marking (May be located on
Black-Red-Yellow Thread (Length
ing)
Black
Red
Yellow
30
10
(CEBEC)
10
10
10
30
(UTE)
30
50
Qualita (IMQ)
10
30
Cables (BASEC)
N.V. KEMA
KEMA-KEUR
<HAR>
10
30
30
SEMKO
<HAR>
10
10
50
Materielkontrollanstalter
sterreichischer Verband fr
<VE>
<HAR>
30
10
50
Elektrotechnik (VE)
10
30
(DEMKO)
30
50
(NSAI)
10
70
(NEMKO)
Asociacion Electrotecnica Y
<UNED>
<HAR>
30
10
70
<HAR>
30
30
70
Standardization (ELOT)
10
90
(IPQ)
Schweizerischer Elektro
SEV
<HAR>
10
30
90
Elektriska Inspektoratet
SETI
<HAR>
10
30
90
23
491B River Valley Road, #12-01/04 Valley Point Office Tower, Singapore 248373
2-8 Honduras Street, London EC1Y OTX, United Kingdom.
31 Gilby Road, Mt. Waverley, VIC 3149, Melbourne, Australia. 6100 Blue Lagoon
Drive, Suite 390 Miami, FL 33126-2087, U.S.A.