2002 Calibration and Repair of Ohaus Triple Beam Balance
2002 Calibration and Repair of Ohaus Triple Beam Balance
2002 Calibration and Repair of Ohaus Triple Beam Balance
4. Attach the 500g hanger weight and balance the scale using
laboratory test weights. If the scale doesn't read 500 ± 0.5g,
adjust the attachment weight pivot. Repeat step "E" using both
front and back pivots. When pivots are adjusted, tighten the
lock-nuts securely.
4. Remove the load, return poise to zero and check for zero
balance.
4. Remove the load, return poises to zero and check for zero
balance.
4. Remove the weights, return poises to zero and check for zero
balance.
A. Check rod assembly - The check rod beneath the scale must swing
freely. Any binding will cause inaccurate readings.
3. Carefully slide the "V" block out and replace it with a new one.
Replace all four blocks, one at a time.
C. Balance cup - A balance that will not "zero" may need adjustment of
the ballast weight.
SCALE NO.
0g
1g
2g
3g
4g
5g
6g
7g
8g
9g
10g
10g
20g
30g
40g
50g
60g
70g
80g
90g
100g
100g
200g
300g
400g
500g
1000g
1500g
2000g
2500g
2610g
PASS/FAIL
DATE
CHECKED
BY