Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                

EL-120-Brochure AGD

Download as pdf or txt
Download as pdf or txt
You are on page 1of 2

Principle Ion selective Electrode ( ISE )

Sample Types Serum, Plasma, Whole Blood, Urine, CSF


®
GREAT CHEMISTRY

Parameters Na+, K+ , CI-, iCa++ & pH

Sample volume < 90µL

Analysis Time < 60 secs

Throughput < 60 samples per hour

Calibration Automatic Calibration :


• User-defined two-point calibration interval - 4 hours/ 8 hours/ 12 hours
• One Point Calibration with each sample analysis

Normal Ranges User defined Normal Ranges for all Sample types with flagging for Abnormal Samples

Data Management Calibration History, Sample History and Control History of last 30 days

Quality Control Tri-Level Quality Control, with L J Plots of last 30 days

Display

Printer
5 color LCD with Touchscreen Display, Resolution: 480 x 272 pixels

Built-in 2 ( 48 mm ) Thermal Printer, 42 characters wide


120
External Interface USB ( Keyboard, Barcode reader, Mouse, Pendrive )

Operating Temp. 15 to 35 C

Relative Humidity 5 to 85% non-condensing


ISE - Tritrol
Power Supply 110 to 240 VAC 50/60 Hz +/- 10%, 35W

Dimensions 365mm x 255mm x 375mm

Net Weight Sample Storage 5 Kg 10,000 Na+, K+ , CI-, iCa++, pH


Calculated parameters:
Precision CV % - TCa, nCa
Na+ 2%, Low sample volume
K+ 2%,
Fast sample processing
Cl- 2%,
iCa++ 3%, Low cost per test
pH 2%,

* Optional Barcode reader & LIS available.

AGD Biomedicals Pvt. Ltd. P : +91-22-28231061/66 ISO : 9001 : 2015


Mehta Trade Center, +91-22-28257999 ISO : 13485 : 2016
Sir M.V. Road, Andheri East, E : sales@agdbio.com
www.agdbio.com
Mumbai - 400 099 INDIA W : www.agdbio.com

MKT-BRO-04-0 *Specifications are subject to change without the prior notice of manufacturer.
Erma EL - 120 is ideal for Medium to High Volume Laboratories and Hospitals. It is user Friendly & Reliable. Measuring Range
It also allows measurement of specific ions; Accurately and Precisely. This supports Fast Diagnosis.
SERUM /PLASMA
PARAMETER BLOOD / AQUEOUS

Two Point Calibration after 4 hrs/ 8 hrs or 12 hrs. Na+ 40 - 205


Programmable as per user
K+ 1.5 - 15
One Point Calibration with each Sample Analysis
CI- 50 - 200
Display Status of Calibration with Date and Time
for each parameter iCa++ 0.2 - 5.0
Also Displays the Date and Time of next Calibration
pH 5.0 - 9.0
to be Performed / Due
• Large Graphical Display affords convenience of Operation
and Minimizes Errors while Entering Data

• User Friendly Software with Interactive Graphical User


Interface

• Shortcut Menu Keys for Easy Operation Levy Jennings plot of daily and monthly QC data to assess
the performance

Helpful for laboratories incase of Accreditation and Quality


Assessment

• Aspiration System: Internal Peristaltic Stepper Motor Pump


for Accurate and Precise aspiration of Analytes

• Automatic flushing after aspiration of every sample to ISEPak


eliminate carryover

• Precise & Accurate Result


• Convenient to place ISEPak which contains reagents
required for all parameters
• Integrated reagent Pack for testing of All Parameters
ELECTRODES • In-Built Reagent Inventory Management Software

URINE DILUENT AND


• Electrodes are based on Ion Selective Principle DEPROTEINIZER
• Electrodes can be refilled.
• Long life and Assured Stability • Urine Diluent : Used for the Dilution of Urine Samples to
Quantitatively determine Urine Electrolytes.
• Deproteinizer : Used for the Cleaning and Maintenance
of Electrolyte Analyzer