HTL Single- and

Multi-channel
Pipettors
Quick Start Guide

Discovery Pro
Discovery Comfort
Labmate Pro
Optipette
Clinipet+, fixed volume

This quick guide is intended to provide


a brief overview of the main features
and operation of all HTL Single- and
Multi-channel Pipettors.
For detailed information, please refer to
the User Manuals that can be found at
www.htl.com.pl in several languages.
Getting started
Know your pipettor

Intended Use
HTL Pipettors are volumetric instruments designed to
measure and transfer liquids precisely and safely with the They can accommodate volumes from 0.1 µL to 10,000 µL
use of disposable pipet tips. The pipettors operate on the depending on the model. They are available in single-,
air-cushion principle (i.e., the aspirated liquid does not 8- and 12-channel variable volume as well as fixed volume
come in contact with the shaft or plunger of the pipettor). single-channel versions.

Variable volume pipettors Clinipet+, fixed volume

Pushbutton
Pushbutton
Ejector button
Ejector button
Volume lock
(Discovery Comfort
and Discovery Pro only)
Volume display
Handle

Multi-channel
manifold

Tip ejector
Tip ejector

Shaft Shaft
Step 1 Step 2
Set the volume Insert the tip

Discovery Comfort Labmate Pro Optipette

✓ ✕
2 1
1 2
6 5
2.16 µL 1.25 mL

Discovery Pro

2 1
0 0
4 0
2 0 ✓ ✕
20.42 µL 1,000 µL

Lock the volume


Discovery Comfort Discovery Pro

NOTE: Labmate Pro and Optipette Single- and Multi-channel


Pipettors do not feature a volume lock function.
Ejector adjustment

Discovery Comfort Labmate Pro Pipettors 5,000 and 10,000 µL

Optipette NOTE: Ejector adjustment is not necessary in Discovery Pro


and Clinipet+ pipettors as they feature a long ejector stroke.

+1 mm

+2 mm
Step 3
Aspirate the liquid Recommended depth

0.1-1 µL 1-100 µL 101-1,000 µL > 1,000 µL

≤1 mm 2-3 mm 2-4 mm 5-7 mm

Step 4
Dispense the liquid

~1 sec.

10-40°
Step 5 Step 6
Eject the tip Calibration

Discovery Comfort Labmate Pro Optipette Clinipet+

Calibration key

NOTE: Before calibration, in the Discovery Comfort and


Discovery Pro lines, always remember to lock the volume
using the lock function and for the Labmate Pro and
Optipette hold the calibration knob.
Safety Notes
• The pipettor is designed for the transfer of liquids only
using the tip. Do not aspirate liquids without the tip
attached. The aspirated liquid should not enter the
Discovery Pro pipettor, as it may cause damage.
• Single-use tips reduce the risk of contamination of
samples.
• Keep the pipettor clean, avoiding the use of abrasive or
Calibration key – flat end corrosive cleaning agents (e.g., acetone).
• Keep the pipettor upright when there is liquid in the tip.
• Only using the pipettor in accordance with the
“Calibration” manufacturer’s instructions ensures the correct pipettor
position parameters are maintained.
• After replacing the plunger or the shaft, the pipettor
should be calibrated.
“Operating”
• In the case of incorrect operation, the device should
position
be cleaned in accordance with the Instructions for Use
or transferred to a service point.
• Ambient operating temperature is +5°C to 45°C.
• Ambient storage conditions (in the original packaging
during transport and short storage) is -25°C to 55°C.

• Follow general work safety regulations regarding hazards


Calibration key – cross end
related to work in the laboratory.
• Take special care when pipetting aggressive substances.
• Use appropriate protective attire (e.g. clothing, goggles
and gloves).
• Avoid pointing the pipettor at yourself or others during use.
• Only use parts and accessories recommended by the
manufacturer.

Cleaning
• External parts may be cleaned with a swab moistened
with isopropyl alcohol.
• Pipettors can be autoclaved at 121°C for 20 minutes.
Warranty Prior to autoclaving, untighten the pipettor shaft,
HTL Single- and Multi-channel Pipettors are covered by remove the filter from 5 and 10 mL pipettor shafts.
a three (3) year limited warranty. For more information on • The outer body of the pipettor is UV-resistant. The
the warranty limitations refer to the full version of the User recommended distance from the radiation source to
Manuals available at www.htl.com.pl. exposed element should not be less than 50 cm.
• Prolonged and very intense UV exposure can cause
de-coloration of pipettor parts but does not affect its
performance.
For additional product or technical information, visit
www.htl.com.pl or contact your local sales office.

ASIA/PACIFIC Singapore
t 65 6572-9740
Australia/New Zealand
t 61 427286832 f 65 6735-2913

Chinese Mainland Taiwan


t 86 21 3338 4338 t 886 2-2716-0338
f 86 21 3338 4300 f 886 2-2516-7500

India EUROPE
t 91 124 4604000 htlcs@corning.com
f 91 124 4604099
LATIN AMERICA
Japan
grupoLA@corning.com
t 81 3-3586 1996
f 81 3-3586 1291 Brazil
t 55 (11) 3089-7400
Korea
t 82 2-796-9500 Mexico
f 82 2-796-9300 t (52-81) 8158-8400

© 2021 Corning HTL SA. All rights reserved.


2/21 CLSHTL-AN-1006DOC REV2

Corning HTL SA, Daniszewska 4, 03-230 Warsaw


www.htl.com.pl

Made in Poland

For a listing of trademarks, visit www.corning.com/clstrademarks.