Varian GC450 User Manual English PDF
Varian GC450 User Manual English PDF
Varian GC450 User Manual English PDF
Herculesweg 8
4330 EA Middelburg
The Netherlands
450-GC
User Manual
North/South America
2700 Mitchell Drive
Walnut Creek
94598 California, USA
Tel: ++(1)9259392400
Fax: ++(1)9259452360 or
++(1)9259452344
2008 Varian, Inc. All Rights Reserved
Europe
P.O. Box 8033
4330 EA Middelburg
The Netherlands
Tel: ++(31)118671000
Fax: ++(31)118623193
Produced in the Netherlands
Australia/East Asia
679 Springvale Road
Mulgrave, Victoria 3171
Australia
Tel: ++(61)395607133
Fax: ++(61)395607950
CP501411 Rev:1.1
March 2008
Warranty
If your system has a TCD, gas must be flowing through both channels to protect
the filaments. Turn power to filaments off when not in use.
Page:II
Using cryogenic column oven cooling, on page 73 leave always the Injectror(s)
/ Detector(s) heating ON (enable) and at a temperature of at least 100 C.
Varian, Inc.
Warranty
HARDWARE PRODUCTS
All analytical instruments sold by Varian are warranted to be free from defects in material and workmanship for the periods specified and
in accordance with the terms on the face of Varian's quotation or as otherwise agreed upon in writing between Varian and the Customer.
The warranty period begins on the date of shipment from Varian to the original Customer. However, where installation is paid for by the
Customer or included in the purchase price, the warranty period begins upon completion of installation. If the Customer schedules
installation to start later than 30 days after delivery or if such delay is caused through the Customer's inability to provide adequate
facilities or utilities or through failure to comply with Varian's reasonable pre-installation instructions or through other omissions by
Customer, then the warranty period starts on the 31st day from date of shipment. Moreover Varian will charge the Customer for labor and
other expenses involved in making multiple or follow-up installation service calls.
SOFTWARE PRODUCTS
Where software is provided within the frame of a license agreement concluded between the Customer and Varian, any warranty shall be
strictly in accordance with the terms of such agreement.
In the absence of a license agreement and unless an alternate warranty period is agreed upon in writing between Varian and the
Customer, the warranty period is as specified on the face of Varian's quotation. Varian warrants such software products, if used with and
properly installed on Varian hardware or other hardware as specified by Varian to perform as described in the accompanying Operator's
Manual and to be substantially free of those defects which cause failure to execute respective programming instructions; however, Varian
does not warrant uninterrupted or error-free operation.
REMEDIES
The sole and exclusive remedy under hardware warranty shall be repair of instrument malfunctions which in Varian's opinion are due or
traceable to defects in original materials or workmanship or, at Varian's option, replacement of the respective defective parts, provided
that Varian may as an alternative elect to refund an equitable portion of the purchase price of the instrument or accessory.
Repair or replacement under warranty does not extend the original warranty period.
Repair or replacement under warranty claims shall be made in Varian's sole discretion either by sending a Customer Support
Representative to the site or by authorizing the Customer to return the defective accessory or instrument to Varian or to send it to a
designated service facility. The Customer shall be responsible for loss or damage in transit and shall prepay shipping cost. Varian will
return the accessory or instrument to the Customer prepaid and insured. Claims for loss or damage in transit shall be filed by the
Customer. To correct software operation anomalies, Varian will issue software revisions where such revisions exist and where, in
Varian's opinion, this is the most efficient remedy.
LIMITATION OF WARRANTY
This warranty does not cover software supplied by the Customer, equipment and software warranted by another manufacturer or
replacement of expendable items and those of limited life, such as but not limited to: Filters, glassware, instrument status lamps, source
lamps, septa, columns, fuses, chart paper and ink, nebulizers, flow cells, pistons, seals, fittings, valves, burners, sample tubes, probe
inserts, print heads, glass lined tubing, pipe and tube fittings, variable temperature dewars, transfer lines, flexible discs, magnetic tape
cassettes, electron multipliers, filaments, vacuum gaskets, seats and all parts exposed to samples and mobile phases.
This warranty shall be void in the event of accident, abuse, alteration, misuse, neglect, breakage, improper operation or maintenance,
unauthorized or improper modifications or tampering, use in an unsuitable physical environment, use with a marginal power supply or use
with other inadequate facilities or utilities. Reasonable care must be used to avoid hazards.
This warranty is expressly in lieu of and excludes all other express or implied warranties, including but not limited to warranties
of merchantability and of fitness for particular purpose, use or application and all other obligations or liabilities on the part of
Varian, unless such other warranties, obligations or liabilities are expressly agreed to in writing by Varian.
Varian, Inc.
Page: III
Safety Information
SAFETY INFORMATION
For Safety Information in German, French, Italian or Spanish please refer to the 430-GC,
431-GC and 450-GC User Manual CD-rom (CP501414) which is included with the GC.
INFORMATION
To prevent any injury to the user or any damage to the instrument it is essential that you
read the information in this chapter and the GC user manual.
If this manual is not in your native language and if you have problems understanding the
text, we advise you to contact your Varian office for assistance. Varian cannot accept
responsibility for any damage or injury caused by misunderstanding of the information in
this manual.
OPERATING INSTRUCTIONS
This user manual is provided to help you establish operating conditions, which will
permit safe and efficient use of your equipment.
Special considerations and precautions are described in the manual, which appear in
the form of NOTES, CAUTIONS and WARNINGS as described on the next page.
It is important that you operate your equipment in accordance with this user manual and
any additional information, which may be provided by Varian. Address any questions
regarding the safe and proper use of your equipment to your local Varian office.
Page: IV
Varian, Inc.
Safety information
Warning Symbol
Warning Description
Hazardous voltages are present inside instrument.
Disconnect from main power before removing screwattached panels.
Hazardous chemicals may be present. Avoid contact,
especially when replenishing reservoirs. Use proper eye
and skin protection.
Very hot or cryogenically cold surfaces may be exposed.
Use proper skin protection.
Eye damage could occur either from flying particles,
chemicals, or UV radiation. Use proper eye and face
protection.
The potential for fire may be present. Follow manual
instructions for safe operation.
Varian, Inc.
Page: V
Safety Information
Perform periodic leak checks on all supply lines and pneumatic plumbing.
Do not allow gas lines to become kinked or punctured. Place lines away
from foot traffic and extreme heat or cold.
Page: VI
Varian, Inc.
Safety information
ELECTRICAL HAZARDS
Disconnect the instrument from all power sources before removing protective
panels to avoid exposure to potentially dangerous voltages.
When it is necessary to use a non-original power cord plug, make sure the
replacement cord adheres to the color coding and polarity described in the preinstallation manual (P/N: CP01410) all local building safety codes.
Replace blown fuses with fuses of the size and rating stipulated on the fuse panel
or in the manual on page 44.
Make sure that voltage sources and line voltage match the value for which the
instrument is wired.
Replace faulty or frayed power cords immediately with the same type and rating.
Use only connector tubing that is chromatographically clean (Varian Part Number
03-918326-00) and has a pressure rating significantly greater than the highest
outlet pressure from the regulator.
Varian, Inc.
Page: VII
Safety Information
GC SAFETY PRACTICES
Exhaust System
No special exhaust ducting is necessary for GC detectors installed in a well-ventilated room
except when the detectors are used to test hazardous chemicals. If you do install ducting:
Secure air duct intakes such that their vibration or air movement does not effect
detector operation.
Burn Hazard
Heated or cryogenically cooled zones of the GC can remain hot or cold for a considerable
time after instrument power is turned off. To prevent painful burns, ensure that all heated or
cooled areas have returned to room temperature or wear adequate hand protection before
you touch potentially hot or cold surfaces.
Page: VIII
Varian, Inc.
Safety information
SERVICE AVAILABILITY
It is the policy of Varian to provide operational spare parts for any instrument and
major accessory for a period of five (5) years after shipment of the final production
run of that instrument. Spare parts will be available after this five (5) year period
but on an as available basis. Operational spare parts are defined as those
individual electrical or mechanical parts that are susceptible to failure during their
normal operation. Examples include relays, lamps, temperature probes, detector
elements, motors, etc. Sheet metal parts, structural members or assemblies and
castings, printed circuit boards and functional modules are normally capable of
being rebuilt to like-new condition throughout their useful life and therefore will be
supplied only on an as available basis after the final production run of the
instrument.
France
Les Ulis Cdex
Tel. +33.1.6986.3838
Russian Federation
Moscow
Tel. +7.095.937.4280
Australia
Mulgrave, Victoria
Tel. +61. 3.9560.7133
Germany
Darmstadt
Tel. +49.6151.7030
Spain
Madrid
Tel. +34.91.472.7612
Austria
Poettelsdorf
Tel. +43.2626.20090
India
Mumbai
Tel.
+91.22.2570.8595/97
Sweden
Solna
Tel. +46.8.445.1620
Benelux
Middelburg
Tel. +31.118.671500
Brazil and Latin America (S)
So Paulo
Tel. +55.11.32380400
Canada
Mississauga, Ontario
Tel. 800.387.2216
China
Beijing
Tel. +86.106310.8550
Europe
Middelburg, The Netherlands
Tel. +31.118.671.000
Varian, Inc.
Italy
Torino
Tel. +39.011.997.9111
Japan
Tokyo
Tel. +81.3.5232.1239
Korea
Seoul
Tel. +82.333.665.5171
Mexico and Latin
America (N)
Mexico City
Tel.
+52.5.55.5239465/026
United States
Walnut Creek, California,
USA
Tel. +1.800.926.3000
(GC and GC/MS)
Tel. +1.800.367.4752 (LC)
Switzerland
Steinhausen
Tel. +41.848.803.800
Taiwan
Shih-Chi
Tel. +886.22.698.9555
United Kingdom and
Ireland
Oxford
Tel. +44.1865.291500
http://www.varianinc.com/
Venezuela
Caracas
Tel.
+58.212.285.0320/2494
Page: IX
Safety Information
CE DECLARATION OF CONFORMITY
Page: X
Varian, Inc.
Table of Contents
TABLE OF CONTENTS
VARIAN ANALYTICAL INSTRUMENT WARRANTY................................................................. III
HARDWARE PRODUCTS .............................................................................................................. III
SOFTWARE PRODUCTS ............................................................................................................... III
REMEDIES .................................................................................................................................. III
LIMITATION OF WARRANTY .......................................................................................................... III
LIMITATION OF REMEDIES AND LIABILITY ...................................................................................... III
SAFETY INFORMATION.............................................................................................................IV
INFORMATION .............................................................................................................................IV
OPERATING INSTRUCTIONS .........................................................................................................IV
GENERAL SAFETY PRECAUTIONS ................................................................................................VI
ELECTRICAL HAZARDS ...............................................................................................................VII
COMPRESSED GAS CYLINDERS ..................................................................................................VII
GC SAFETY PRACTICES ...........................................................................................................VIII
Exhaust System.................................................................................................................VIII
Radioactive Source Detectors ...........................................................................................VIII
Burn Hazard.......................................................................................................................VIII
SPARE PARTS AVAILABILITY ........................................................................................................IX
SERVICE AVAILABILITY ................................................................................................................IX
CE DECLARATION OF CONFORMITY ..............................................................................................X
TABLE OF CONTENTS...............................................................................................................XI
FIGURES.................................................................................................................................. XXV
TABLES ................................................................................................................................. XXVII
INTRODUCTION ........................................................................................................................... 1
GUIDE TO THE 450-GC USER MANUAL ......................................................................................... 1
Abbreviations ........................................................................................................................ 1
BRIEF DESCRIPTION OF THE 450-GC ........................................................................................... 2
CONFIGURATION AND OPTIONS .................................................................................................... 3
INSTALLATION ............................................................................................................................ 5
NOTES ABOUT SETTING UP AND USING YOUR 450-GC................................................................. 5
PREPARING FOR INSTALLATION .................................................................................................... 5
Inspection ............................................................................................................................. 5
Unpacking............................................................................................................................. 5
Remove Protective Devices ................................................................................................. 6
Connect Data Handling Peripherals ..................................................................................... 6
Select Gases ........................................................................................................................ 7
Hydrogen as Carrier Gas...................................................................................................... 8
Varian, Inc.
Page: XI
Table of Contents
Varian, Inc.
Table of Contents
Fuses .................................................................................................................................. 44
DISPLAY AND KEYBOARD ...................................................................................................... 45
STATUS INDICATION LEDS ........................................................................................................ 46
Overview Analytical Phases ............................................................................................... 46
Error LED............................................................................................................................ 47
MOST USED FUNCTIONS ............................................................................................................ 48
Status page......................................................................................................................... 49
Version Information............................................................................................................. 50
ADVANCED FUNCTIONS .............................................................................................................. 51
DISPLAY.................................................................................................................................... 52
Title Bar............................................................................................................................... 53
Main Operation screen ....................................................................................................... 54
Sample Delivery Screen ..................................................................................................... 55
Injector Screen.................................................................................................................... 57
Pressure and Flow screen .................................................................................................. 58
Oven Screen....................................................................................................................... 60
Methods Screen.................................................................................................................. 61
Detector Screen.................................................................................................................. 62
Analog Output ................................................................................................................ 63
Setup/Calibration ................................................................................................................ 64
Calibration ...................................................................................................................... 64
Setup .............................................................................................................................. 65
Configuration Tab........................................................................................................... 66
External Device .............................................................................................................. 67
Sample Delivery Tab ...................................................................................................... 68
Oven Tab........................................................................................................................ 73
Column Tab .................................................................................................................... 74
Detector Tab................................................................................................................... 75
Injector Tab..................................................................................................................... 77
Preferences Tab............................................................................................................. 78
System Tab .................................................................................................................... 79
Touch Screen Calibration............................................................................................... 81
Error/Log Handling.............................................................................................................. 84
LOCAL AUTOMATION............................................................................................................... 87
COMMUNICATIONS................................................................................................................... 89
Introduction ......................................................................................................................... 89
Ethernet Communications .................................................................................................. 89
How to set IP address in Galaxie Workstation using BOOTP (DHCP) .......................... 90
How to set a Fixed IP address in the 450-GC................................................................ 93
ANALOG CABLES ....................................................................................................................... 94
SAMPLE INTRODUCTION......................................................................................................... 97
AUTOSAMPLERS........................................................................................................................ 97
LIQUID INJECTORS ..................................................................................................................... 97
SAMPLING VALVES .................................................................................................................... 99
Heated Zones for Valves.............................................................................................. 100
Varian, Inc.
Page: XIII
Table of Contents
Varian, Inc.
Table of Contents
Page: XV
Table of Contents
Varian, Inc.
Table of Contents
Page: XVII
Table of Contents
Varian, Inc.
Table of Contents
Page: XIX
Table of Contents
Varian, Inc.
Table of Contents
Page: XXI
Table of Contents
Varian, Inc.
Table of Contents
Page: XXIII
Table of Contents
Page: XXIV
Varian, Inc.
Table of Contents
FIGURES
FIGURE 1: COLUMN ENDS .............................................................................................................. 15
FIGURE 2: CONNECTORS FOR PERIPHERAL EQUIPMENT .................................................................. 34
FIGURE 3: CONNECTOR J504-J505................................................................................................ 35
FIGURE 4: CONNECTOR J401......................................................................................................... 36
FIGURE 5: CONNECTOR J302......................................................................................................... 37
FIGURE 6: CONNECTOR J45/J46 .................................................................................................... 38
FIGURE 7: CONNECTOR J44........................................................................................................... 39
FIGURE 8: CONNECTOR J42/J43 .................................................................................................... 40
FIGURE 9: CONNECTOR J41........................................................................................................... 41
FIGURE 10: OPTION BOARD ........................................................................................................... 43
FIGURE 11: 450-GC BACK-PANEL ETHERNET CONNECTION ............................................................ 89
FIGURE 12: PNEUMATIC SYSTEMS FOR LIQUID INJECTORS AND SAMPLING VALVES ......................... 101
FIGURE 13: TOP-COVER FOR SAMPLER ........................................................................................ 105
FIGURE 14: SPME SYRINGE AND HOLDER FIGURE 15: SPME/STANDARD SWITCH ....................... 142
FIGURE 16: SPME SYRINGE ........................................................................................................ 145
FIGURE 17: LIQUID SAMLING SYRINGE AND HOLDER FIGURE 18: SPME/STANDARD SWITCH ....... 149
FIGURE 19: EFC21 AND 1177 IN SPLIT MODE FLOW DIAGRAM....................................................... 157
FIGURE 20: EFC21 AND 1177 IN SPLITLESS MODE FLOW DIAGRAM ............................................... 158
FIGURE 21: EFC25 AND 1177 FLOW DIAGRAM ............................................................................. 159
FIGURE 22: 1177 INJECTOR ASSEMBLY ........................................................................................ 170
FIGURE 23: 1079 PTV INJECTOR ASSEMBLY ................................................................................ 177
FIGURE 24: EFC21 AND 1079 IN SPLIT MODE FLOW DIAGRAM....................................................... 181
FIGURE 25: EFC25 AND 1079 FLOW DIAGRAM ............................................................................. 182
FIGURE 26: CHROMATOPROBE ASSEMBLY .................................................................................... 199
FIGURE 27: 1041 ON-COLUMN INJECTOR WITH INSERT ................................................................. 207
FIGURE 28: EFC23 AND 1041 FLOW DIAGRAM ............................................................................. 209
FIGURE 29: EFC24 AND 1041 FLOW DIAGRAM ............................................................................. 210
FIGURE 30: 1061 FLASH VAPORIZATION INJECTOR WITH INSERT .................................................... 221
FIGURE 31: EFC23 AND 1061 FLOW DIAGRAM ............................................................................. 222
FIGURE 32: EFC24 AND 1061 FLOW DIAGRAM ............................................................................. 223
FIGURE 33: 1093 INJECTOR ASSEMBLY ........................................................................................ 234
FIGURE 34: EFC23 AND 1093 FLOW DIAGRAM ............................................................................. 235
FIGURE 35: EFC24 AND 1093 FLOW DIAGRAM ............................................................................. 236
FIGURE 36: CONNECTION CO2 TANK TO GC.................................................................................. 238
FIGURE 37: SPT ELECTRICAL DIAGRAM........................................................................................ 256
FIGURE 38: LCO2 COOLANT FLOW DIAGRAM ................................................................................ 257
FIGURE 39: LN2 COOLANT FLOW DIAGRAM ................................................................................... 260
FIGURE 40: SPT COMPONENTS (SHORT TRAP ASSEMBLY, LCO2 OPTION SHOWN.......................... 265
FIGURE 41: REMOVING SPRINGS FROM TRAP ASSEMBLY ............................................................... 267
FIGURE 42: CRYO NOZZLE ASSEMBLY .......................................................................................... 268
FIGURE 43: FID AUTO-IGNITE FLOW DIAGRAM ............................................................................... 287
FIGURE 44: FID EXPLODED VIEW ................................................................................................. 288
FIGURE 45: ORIENTATION OF IGNITOR PROBE ELEMENTS IN FID TOWER ........................................ 291
FIGURE 46: FID CROSS-SECTIONAL VIEW ..................................................................................... 291
Varian, Inc.
Page: XXV
Table of Contents
Page: XXVI
Varian, Inc.
Table of Contents
TABLES
TABLE 1: HYDROGEN AS CARRIER GAS DURING POWER OFF.............................................................. 8
TABLE 2: GAS INLET PRESSURE ...................................................................................................... 10
TABLE 3: FACTORY-INSTALLED INJECTOR INSERTS .......................................................................... 13
TABLE 4: 1/16 COLUMN FERRULES................................................................................................ 14
TABLE 5: MAXIMUM INSERT DEPTH FOR CAPILLARY COLUMN IN INJECTOR ........................................ 16
TABLE 6: GENERAL TIGHTENING AND RE-TIGHTENING FOR COMMON FERRULES ............................... 17
TABLE 7: VARIAN TEST COLUMNS ................................................................................................... 19
TABLE 8: INJECTOR AND EFC TYPE COMBINATION ........................................................................... 19
TABLE 9: COLUMN CONDITION HEAD PRESSURE ............................................................................... 22
TABLE 10: FLOW RATES FOR CARRIER AND DETECTOR GASES ........................................................ 26
TABLE 11: DETECTOR MANUAL FLOW RATES ................................................................................... 27
TABLE 12: VARIAN TEST SAMPLES ................................................................................................. 29
TABLE 13: VARIAN TEST SAMPLE 1041 AND 1061 CONDITIONS ....................................................... 29
TABLE 14: VARIAN TEST SAMPLE 1177 AND 1079 CONDITIONS ....................................................... 30
TABLE 15: RECOMMENDED SOLVENTS FOR GAS CHROMATOGRAPHY ............................................... 31
TABLE 16: USED FUSES ................................................................................................................. 44
TABLE 17: STATUS INDICATION LED'S ............................................................................................ 46
TABLE 18: ERROR LED.................................................................................................................. 47
TABLE 19: INJECTOR EFC MODULES .............................................................................................. 59
TABLE 20: DETECTOR EFC MODULES ............................................................................................ 63
TABLE 21: SAMPLER/INJECTOR POSITION ........................................................................................ 69
TABLE 22: VALVE TYPE NAMES ...................................................................................................... 72
TABLE 23: ANALOG CABLES ........................................................................................................... 95
TABLE 24: INJECTION MODE PARAMETERS FOR ON-COLUMN .......................................................... 125
TABLE 25: INJECTION MODE PARAMETERS FOR SPLIT-/SPLITLESS .................................................. 126
TABLE 26: INJECTION MODE PARAMETERS FOR NEAT ..................................................................... 128
TABLE 27: INJECTION MODE PARAMETERS FOR VISCOUS ............................................................... 129
TABLE 28: INJECTION MODE PARAMETERS FOR VOLATILE............................................................... 130
TABLE 29: SPLIT INJECTION TYPICAL CONDITIONS ......................................................................... 167
TABLE 30: SPLITLESS INJECTION METHOD PARAMETERS ............................................................... 168
TABLE 31: DETECTOR TEST SAMPLES .......................................................................................... 169
TABLE 32: 1177 COLUMN FERRULES 1/16"................................................................................... 170
TABLE 33: 1079 COLUMN FERRULES 1/16"................................................................................... 177
TABLE 34: SPLIT INJECTION TYPICAL CONDITIONS ......................................................................... 190
TABLE 35: ISOTHERMAL SPLITLESS INJECTION METHOD PARAMETERS ........................................... 191
TABLE 36: TEMPERATURE RAMPED SPLITLESS INJECTION ............................................................. 192
TABLE 37: ON-COLUMN INJECTION ............................................................................................... 193
TABLE 38: LARGE VOLUME INJECTION .......................................................................................... 194
TABLE 39: 1079 PTV TEST SAMPLES ........................................................................................... 194
TABLE 40: 1041 COLUMN FERRULES 1/16" X 0.8 MM .................................................................... 207
TABLE 41: CARRIER GAS FLOW RATES FOR 1041 ......................................................................... 212
TABLE 42: 1041 INJECTOR CHROMATOGRAPHIC TESTING CONDITIONS ........................................... 214
TABLE 43: 1061 COLUMN FERRULES 1/16" X 0.8 MM .................................................................... 221
TABLE 44: CARRIER GAS FLOW RATES FOR 1061 FLASH INJECTOR ............................................... 226
Varian, Inc.
Page: XXVII
Table of Contents
Page: XXVIII
Varian, Inc.
Introduction
INTRODUCTION
GUIDE TO THE 450-GC USER MANUAL
The purpose of this User Manual is to provide instructions for the safe
installation and operation of the Varian 450-GC (Gas Chromatograph).
Abbreviations
The following abbreviations can appear in this Manual:
Varian, Inc.
AC
DC
alternating current
direct current
ECD
EFC
FID
PFPD
TCD
TSD
GC
GC/MS
Gas Chromatograph
Gas Chromatograph/Mass Spectrometer
ID
OD
Inner Diameter
Outer Diameter
P/N
Part Number
LUI
S.I.D.
kPa
psi
Page: 1
Introduction
Page: 2
The 1079 PTV Injector provides 5 modes of injection - isothermal split and splitless,
temperature ramped splitless, on-column and large volume.
A full suite of universal and selective detectors - FID, TCD, ECD, TSD (N, P), PFPD
(S, P, N, C, Metals), PDHID and Ion Trap MS.
7 heated zones and 7 external events (24 Volt) for maximum flexibility and custom
configurations.
Option board with 8 external events (24 Volt) available, also 3 analog outputs with
output voltage 0-1 and 0-10 Volt.
Intuitive, function-based keyboard, LCD display with touch-screen for ease of use in
instrument handling.
Large column oven for easy access and installation of up to three analytical
columns.
Varian, Inc.
Introduction
1041 On-Column Injector for wide bore (0.53 mm ID) or packed columns.
Methanizer.
Any combination of three Injectors may be installed on the 450-GC. If a valve oven is
installed, it normally replaces 2 physical Injector positions.
Any combination of three detectors can be installed with the following exceptions: Only
2 TCD or PFPD detectors can be installed. If 2 standard TCDs are installed, no other
detectors can be accommodated (physical space). However, a dual TCD (two cells in
one housing) is available as a Custom Solution which allows installation of another
detector, such as an FID, behind it.
The methanizer typically occupies the same location as one of the detectors.
Installed options on the 450-GC are generally identified by their location. The options
mounted on the top of the instrument, such as those listed above, are designated as
Front, Middle and Rear to coincide with their relative location or the location of their
respective electronic control modules.
Varian, Inc.
Page: 3
Introduction
Page: 4
Varian, Inc.
Installation
INSTALLATION
NOTES ABOUT SETTING UP AND USING YOUR 450-GC
The purity of the gases used with your 450-GC are essential for optimum
operation.
Varian Gas Clean Filter Kit (CP736530) is highly recommended.
All GCs require regular maintenance in order to operate at their highest
efficiency.
You want to change the Injector septum, Injector liner, FID flame tip, gas filters
and AutoSampler vials, wash bottle septa, syringes and solvents regularly.
Varian has an extensive supply of quality maintenance parts for your 450-GC
and all of your chromatography needs including a comprehensive selection of
the highest quality capillary columns available to chromatographers. Please
visit the Varian internet site (www.varianinc.com), contact your Varian sales
representative, or call your local Varian distribution center.
Inspection
The GC will arrive packed in one large palletized box and one or more smaller
cartons. Inspect the cartons carefully for damage or signs of rough handling.
Report damage to the carrier and to your local Varian office.
Unpacking
Unpack the GC and accessories carefully and transfer to the work area, using
proper handling techniques. The GC should be lifted from the bottom. Inspect
the GC and accessories and immediately report any damage to the carrier and
your local Varian office.
Avoid back strain or injury by following all safety precautions when lifting
heavy objects.
Varian, Inc.
Page: 5
Installation
Page: 6
Varian, Inc.
Installation
Select Gases
The following gases are commonly used in Gas Chromatography for the
following functions. Note that the recommended purity for carrier make-up and
detector fuel gases is 99.999%. In addition, the use of gas filters to remove
hydrocarbons, oxygen and water from carrier gases is recommended.
Gas
He, N2, H2, Ar, Ar/CH4
H2, Air
He, N2, Ar/CH4
Air
LN2, LCO2
Function
Carrier gas
Detector fuel gases
Make-up gas for capillary column use
AutoSampler operation
Cryogenics
Do not exceed MAXIMUM inlet pressures listed on the rear panel of the
instrument.
Varian, Inc.
Page: 7
Installation
Hydrogen that is allowed to expand rapidly from high pressure can self
ignites.
Mechanism
Results
GC off
Power failure
If the power fails, the GC shuts down. The carrier gas is shut down (in case of EFC).
Table 1: Hydrogen as Carrier Gas during Power off
Page: 8
Varian, Inc.
Installation
Operating precautions
Varian, Inc.
Turn off the hydrogen at its source any time you shut down the GC.
Page: 9
Installation
Pressure
Carrier gas
500 kPa, 5 Bar, 80 psi (maximum 1000 kPa/ 10 Bar / 150 psi)
Make-up
500 kPa, 5 Bar, 80 psi (maximum 1000 kPa/ 10 Bar / 150 psi)
Air
400 kPa, 4 Bar, 60 psi (maximum 1000 kPa/ 10 Bar / 150 psi)
H2
300 kPa, 3 Bar, 40 psi (maximum 650 kPa/ 6.5 Bar / 100 psi)
Table 2: Gas inlet pressure
Calculate a GC pressure drop (restrictions, tubing and EFC) of about 200 kPa,
2 Bar, 30 psi for well working EFCs.
Example: Column pressure needed 100 kPa + 200 kPa (pressure drop) = Inlet
pressure needed at least 100+200= 300kPa.
Connect Gases
Follow the steps below to connect gas supplies to your GC:
1. Cut required lengths of chromatographically-cleaned 1/8 copper tubing
(P/N: 03-918326-00) for the carrier and fuel gas plumbing. Clean the ends
of the tubing with a metal file.
2. Slide a 1/8 Swagelok nut, back ferrule and front ferrule over one end of the
copper tubing and attach to the outlet fitting on the cylinder regulator.
3. Push the copper tubing into the outlet fitting as far as it will go, then pull
back very slightly and tighten 3/4-turn past finger-tight with a 7/16 wrench.
Page: 10
Varian, Inc.
Installation
4. Slide a 1/8 Swagelok nut, back ferrule and front ferrule over the other end
of the copper tubing and connect it to the corresponding gas inlet on the
GC rear panel.
1/8" Copper Tubing
Cylinder
Regulator
5. Push the copper tubing into the bulkhead fitting as far as it will go, then pull
back very slightly and tighten 3/4-turn past finger-tight with a 7/16" wrench.
6. Make sure the gas supply control valve, which controls gas pressure to the
GC, is completely closed. Open this valve slowly and monitor the pressure
gauges on the dual stage regulator. The first pressure gauge should now
read the pressure of the cylinder.
7. Cautiously turn the regulator valve to supply gas to the GC. Watch the
pressure gauge closely and adjust the pressure according to Table 2 on
page 10.
8. Leak checks all fittings.
Varian, Inc.
Page: 11
Installation
Connect GC to Power
Plug the GC power-cord into an appropriate source of power.
Varian supplies a number of power-cords qualified to be used
with the power load of the GC. Depending on the local
electrical network different power-cords are available.
Do not use power-cords with to low capacity. This may result in
overheating of the power-cord.
Turn on the GC power switch located at the top left of the GC.
The GC top cover (power switch) of the GC must be free of any obstacles.
This will give in case of emergency quick access to the power switch.
The 450-GC automatically enters the OPERATION screen after power up.
Press the SETUP
configuration.
Your GC is shipped from the factory with default settings. The following is
relevant information on the factory default states and settings:
Page: 12
When the GC is turned the first time on, all installed heated zones except
for the column oven default to power OFF.
When installed components that use a heated zone are turned ON in the
active method, they default set to a temperature of 50C.
Varian, Inc.
Installation
You may need to install and condition a new column after your GC has been in
use for some time. Reset all Injector, detector and column ovens to 50C to
avoid burn hazards when connecting column to Injector and detector parts
inside the oven. See also Column Conditioning on page 23.
Prior to installing your column, check that the factory-installed Injector is
equipped with the proper insert for your application. Factory-installed Injectors,
other than the 1041, have a glass insert suitable for a variety of applications. If
your application calls for a different insert, replace the factory-installed insert
prior to installing your column. Refer to the Injector section of the Manual for
instructions to replace the insert on page 97.
The factory-installed inserts for the Injectors are shown below, more Injector
insert you can find here, on page 443.
Injector
Insert Information
Part Number
1061
1079
PTV
1093
03-92611943
03-92611945
1041
03-92543101
1177
03-92611927
SG092034
Varian, Inc.
Page: 13
Installation
Reversed
Ferrule
Capillary
Column
Nut
Fused Silica
Column
Column ID
0.18 mm ID and smaller
0.25 mm ID
0.25 mm ID and smaller
0.32 mm ID
0.53 mm ID
Page: 14
Holes
1
1
2
1
1
2
1
Vespel
40% Graphite
/60% Vespel
Max 250C
Max 350C
Max 400C
CR214103
CR212103
CR213103
CR214104
CR212104
CR213104
CR212124
CR213124
CR214105
CR212105
CR213105
CR212125
CR213125
CR214108
CR212108
CR213108
Table 4: 1/16 Column Ferrules
Graphite
SilTite
Max 450C
CR211104
CR211124
CR211105
CR211125
CR211108
Metal, GC/MS
SG073300
SG073301
SG073302
Varian, Inc.
Installation
1
2
3
4
5
Varian, Inc.
Page: 15
Installation
Insert depth
Mark at
1041
1061
1177
1079 PTV
Maximum*
None
Maximum*
None
3.7 cm
2.7 cm
7.5 cm
5 cm
1093
On-Column
8.6 cm
5 cm
7. * Only for 1041 and 1061Injectors, for the other Injectors continue at 12.
8. Gently push the column fully up into the Injector (9 cm) until the column
comes to a firm stop. This correctly positions the column against the upper
end of the Injector insert. The column must go in the full distance. If it goes
not, the Injector will not function properly. Tighten the column nut just
enough to prevent column movement.
9. Manually insert the syringe that will be used with the column into the 1041
Injector, checking that the syringe goes into the Injector completely with no
binding or interference. If interference occurs, repeat the above column
insertion procedure until the syringe moves smoothly into the Injector.
10. Continue to hold the column firmly in place as you tighten the column nut
carefully and only enough to hold the column firmly in place and obtain a
good seal. Remove the syringe from the Injector.
11. After the column is installed at the Injector end, make several dry injections
to ensure that the needle enters the column easily.
12. Partially insert the column into the lower end of the Injector. Thread up the
capillary column nut and ferrule finger-tight.
Page: 16
Varian, Inc.
Installation
13. Gently push the column into the Injector until the mark on the column is
aligned with the bottom edge of the column nut.
14. Tighten the column nut by hand until it is finger-tight. Hold the column and
use a 5/16" wrench to tighten the column nut in place. Tighten the column
nut only enough to seal the column and hold it firmly in place.
An optional Capillary Column Quick Connect Kit is available that
simplifies the installation of a capillary column into Injectors and
detectors. It contains split capillary column nuts, re-usable graphite
ferrules (for 250 m, 320 m, 530 m ID columns) and a column
depth scale. The ferrule is set on the column using this depth scale.
No felt pen or typewriter correction fluid is needed (The part number
of the Quick Connect Kit is 03-925751-90).
Ferrule Type
Recommendation
Comments
- to -turn past finger-tight may be
required to achieve a seal.
If a different size of column is used,
more tightening may be required.
May require retightening after first or
second programmed run. Over
tightening will destroy ferrule and seal.
New graphite-Vespel
ferrule
Used graphite-Vespel
ferrule
Page: 17
Installation
Page: 18
Varian, Inc.
Installation
Part Number
-
CP8907
CP8907 (ECD)
1041/1061
CP7620
Carrier gas flow is adjusted at the user interface for EFC or at the
pneumatics panel for manual pneumatics. For manual pneumatics carrier
gas systems, identify the controls for carrier gas flow at the pneumatics
panel for your Injector system. These pneumatic controls are identified by
their location, i.e. front, middle, rear.
Turn on carrier gas flow by adjusting the manual flow controller or setting a
positive pressure or flow on the EFC system.
Varian, Inc.
Page: 19
Installation
3. When EFC is installed, in addition to programming the split state, the split
ratio can also be programmed.
4. Set the initial split state to ON and set the split ratio to 100.
key. Turn the column oven off by pushing
5. Press the COLUMN OVEN
on the button Turn Oven Power Off.
6. Allow the oven to cool down. Open the column
oven door and adjust the septum purge needle
valve located above the column oven until the
flow rate is in the range 3 - 5 mL/min. For EFC
equipped systems this value has to be entered
in the septum purge calibration routine.
Septum purge calibration
Page: 20
Varian, Inc.
Installation
Select the Injector you are setting up by specifying the location (front,
middle or rear) or type.
Varian, Inc.
You will be prompted to enter the Column Head Pressure and Adjusted
Septum Purge Flow Rate. Fill in the column head pressure and adjust and
measure the Septum Purge Flow rate.
Fill in the actual Septum Purge Flow and press the OK button.
Page: 21
Installation
Column ID
Head Pressure
250 m
320 m
530 m
35 kPa (5 psi)
Injector Conditioning
It may be necessary to bake out the Injector if it has become contaminated
during the re-assembly process, i.e. if the internal parts were placed on dirty
surfaces or touched with bare hands. Then, both the Injector and column
should be conditioned. See section Maintenance, Instructions for HighTemperature bakeout of Injectors on page 433.
To prevent contamination and overheating of the fused silica
column, the column must NOT be connected to the Injector during
the Injector bake-out procedure.
Page: 22
Varian, Inc.
Installation
Column Conditioning
In general the column should be conditioned at 20C above the maximum
temperature used in the analytical method, or 20C below the maximum
operating temperature of the column, whichever is lower. Check with the
column supplier for the correct conditioned details.
New Columns
To minimize chemical contamination of the detector, all new columns should be
baked out before they are connected to the detector. Refer to the supplier's
recommendations for temperatures.
Used Columns
Periodically (especially during extended isothermal operation), the column
should be baked out to remove accumulated contaminants. Column
conditioning is required when column efficiency begins to fall off, i.e. loss of
resolution, peak tailing, severely drifting baseline, retention of extra peaks, etc.
For fused silica columns, it is recommended that 1 to 2 turns at the beginning of
the column (approximately 1/2 meter) be cut off prior to reconditioning, as this is
the end part of the column where the stationary phase has most likely been
disturbed. Attach the column to the Injector and, with carrier gas flowing, bake
out the column for a minimum of 3 hours.
It is important to establish the maximum recommended conditioning
temperature for the installed column. This information is provided with the
column or can be obtained from the column manufacturer. The maximum
allowable temperature for the column does not have to be used for conditioning
purposes.
A GC column may be irreversibly damaged if it is heated with air or
moisture present in the pneumatic system. It is important to purge the
column with inert carrier gas before heating the column. After the
column flow is established, allow the carrier gas to purge through the
column for 15 - 20 minutes before heating it.
Varian, Inc.
Page: 23
Installation
Check that the capillary column nut and ferrule are installed on the column.
Note that if the TCD is used with capillary columns, a special capillary
adapter (PN 03-925605-01) is used on the TCD inlet.
Page: 24
Varian, Inc.
Installation
If using a detector other than TCD, locate the packed column detector
adapter in the Packed Column Kit and install on the inlet to the detector.
The TCD inlet is designed for packed column installation.
Insert the open end of the packed column into the detector inlet or adapter
and push until it reaches a stop.
Hold the column in place and finger-tighten the column nut. Then, using 2
7/16 wrenches, tighten 3/4-turn past finger-tight. On subsequent
installations, tighten the nut 1/4 to 1/2-turn past finger-tight.
Varian, Inc.
Page: 25
Installation
BASIC OPERATION
At points throughout this section, specific values and conditions are given
for analysis of the Varian test samples on Varian test columns. Your
analytical column and application may require different values and
conditions. Please ascertain the range of operating temperatures for your
analytical column and make sure your settings are within this range.
Detector
Carrier
Gas
FID
N2, H2 or
He
20 1
PDHID
1 - 15
Make-Up
Gas
N2 or He
30 1
300 10
30 1
TSD
N2 or He
20 1
1 - 15
Same as
Carrier
Gas
30 1
175 10
4 0.2
TCD a
N2, He or
H2
20 1
1 - 15
Same as
Carrier
Gas
30 1
ECD
N2 or
Ar/CH4
20 1
1 - 15
N2 or
Ar/CH4
30 1
PFPD
N2, He or
H2
20 1 b
1 - 10
None
__
17 1
10 1
13 1
a TCD Reference Flow The flow rate for carrier gas through the reference cell is 30 1 mL/min.
b PFPD with packed columns An effluent splitter is used to reduce the carrier flow to the detector to < 5
mL/min.
Page: 26
Varian, Inc.
Installation
Follow the steps below to set all gas flow rates on the 450-GC.
Flows are typically set with carrier gas first followed by detector make-up and
fuel gases. The most convenient way to measure or verify gas flows is at the
detector exit.
Adapters are provided in the individual detector accessory kits for this purpose.
Note, however, that the most accurate method to determine FID and TSD flows
is at the flame tip after removing the tower.
Detector
Carrier +
make-up
FID
30
60
360
TSD
30
34
209
PFPD
15
32
Carrier + make-up +
hydrogen +air1 + air2
42
Varian, Inc.
Page: 27
Installation
Page: 28
Varian, Inc.
Installation
Part Number
TCD
82-005048-01
ECD
82-005048-02
PFPD
82-005048-03
TSD
82-005048-04
82-005048-07
82-005048-00
Note: If the FID test sample is not available, the TCD test sample can be used if first diluted 100:1.
Table 12: Varian Test Samples
Note:
These conditions may not conform to the requirements of your analytical column. Do not exceed your
columns Tmax. See Column Manufacturers specifications for your column.
FID
TCD a
TSD
PFPD
80
80
80
80
80
20
20
20
20
20
200
200
200
200
200
190
140
140
175
175
15
15
15
15
15
220
220
220
220
220
300
300
220
300
300 b
12
0.05
12
10 c
On
On
On
On
On
Range
Initial Autozero
3.0 0.3
a TCD Reference Flow Rate for carrier gas flow through the reference cell is 20 1 mL/min.
b Depends on application: S-mode 200C, P-mode 300C and N-mode 250C.
c Normal range: 10 if Varian testmix is being used.
User Manual 450-GC
Varian, Inc.
Page: 29
Installation
FID
TCD a
TSD
PFPD
80
80
80
80
80
200
200
200
200
200
20
20
20
20
20
220
220
220
220
220
300
300
220
300
300 b
12
0.05
12
10
On
On
On
On
On
3.0 0.3
Range
Initial Autozero
TSD bead current (amps)
Injection mode
Splitless
ON
0.01
OFF
1.00
ON
20
Table 14: Varian Test Sample 1177 and 1079 Conditions
a TCD Reference Flow Rate for carrier gas flow through the reference cell is 20 1 mL/min.
b Depends on application: S-mode 200C, P-mode 300C and N-mode 250C.
Page: 30
Varian, Inc.
Installation
Solvents for GC
Following are some recommended solvents for non-polar, intermediate and
polar phase columns.
Boiling Point ( C)
Column Phase
Recommended Solvent
Non-Polar
Pentane
36.1
100% Methyl
n-Hexane
69.0
Cyclohexane
80.7
Iso-octane
99.3
Benzene
80.1
Toluene
110.6
Ethyl Ether
34.6
55.2
Methylene Chloride
39.8
Chloroform
61.2
Carbon Disulfide
46.5
Intermediate
Ethyl Acetate
77.0
Acetone
56.5
127.0
Acetonitrile
81.6
Polar
Methanol
64.7
Polyethyene Glycol
Ethanol
78.5
2-Propanol
82.5
n-Butanol
117.7
Varian, Inc.
Page: 31
Installation
On-Column, Flash
Vaporizing (1041/1061)
Wide bore Capillary Columns
(0.53 mm ID)
Packed Columns
0.12 Wide bore Capillary
1-5 Packed Columns
0.5-5 Wide bore Capillary
Column
Split
(1079 PTV, 1177)
Splitless
(1079 PTV, 1177)
All Capillary
Columns
0.2-2
0.2-1
25-50 for
CP-8400/8410 A/S
1 for
CP-8400/8410 A/S
To inject your sample, insert the syringe needle into the Injector inlet, press
the syringe barrel against the Injector switch and depress it completely,
simultaneously pressing down the syringe plunger. Remove the syringe from
the Injector immediately.
The blue RUN light starts to blink until the analysis ends. The run may be
stopped by pressing the STOP
If you are analyzing the Varian Test Sample on a Varian Test Column, you
may compare your resulting chromatogram with the appropriate test
chromatogram in the Detector Test Chromatograms, on page 419. The
retention times will be close to those shown in the example chromatogram.
The peak heights may be considerably different.
Page: 32
key.
Varian, Inc.
Peripheral Connectors
REFERENCE
PERIPHERAL CONNECTORS
The 450-GC Gas Chromatograph is equipped with various connectors on the rear panel and top
of the instrument intended for use with peripheral equipment such as AutoSamplers and data
handling devices. The following are the standard connectors on the 450-GC and their intended
uses:
Varian, Inc.
Page: 33
Peripheral Connectors
Top View
PERIPHERAL CONNECTORS
Peripheral connectors are located underneath the detector cover.
Lift the detector cover up at the on/off power-switch and slide
the cover to the rear of the GC. Lift the cover up and remove it.
J301 Analog IN
Not in use
Page: 34
Varian, Inc.
Peripheral Connectors
Page: 35
Peripheral Connectors
Top View
Page: 36
Varian, Inc.
Peripheral Connectors
Top View
Page: 37
Peripheral Connectors
Page: 38
Varian, Inc.
Peripheral Connectors
Varian, Inc.
Page: 39
Peripheral Connectors
Top View
Page: 40
Varian, Inc.
Peripheral Connectors
Top View
Page: 41
Peripheral Connectors
Rear View
REAR PANEL
Not in use.
Page: 42
Varian, Inc.
Peripheral Connectors
OPTION BOARD
The Option Board is situated below the top-cover in the Option Slot.
The Option Board have 8 Time-Events contacts (24 Volt DC) available, also 3
Analog Outputs are present and available via connector J401 Analog Out.
Top View
Page: 43
Peripheral Connectors
FUSES COMMON
Fuses have been installed within the instrument for the protection of major
power circuits.
They are designed to open as quickly as possible to prevent damage to other
components within the circuit.
Occasionally, an opened fuse may have been caused by a short one-time
surge; however, it is far more common that a component within the protected
circuit has failed.
Before replacing fuses, turn off the oven and any heated zones and
let them cool down. Turn off any detector gases at their supply.
Then turn off the main power switch and unplug the power cord.
Fuses
The major fuses are located on the power module.
There are 4 fuses on the power module on page 34
Fuse F5 is an electronic fuse and situated on the electronic main board (not
accessible or replaceable).
ID
Description
F1
Glass/Type T
F2
System
Power Rating
Partnumber
6.3A/250V 5 x 20 mm
5550033400
Glass/Type T
6.3A/250V 5 x 20 mm
5550033400
F3
Glass/Type F
Heaters/Ovenfan (except
column oven)
10A/250V 5 x 20 mm
5550034800
F4
Glass/Type T
6.3A/250V 5 x 20 mm
5550033400
F5
Electronic
0.5 A
Page: 44
Varian, Inc.
on page 54
on page 51
on page 48
on page 52
The 450-GC is carried out with a so-called touch screen. Touch simplifies
the human / machine interface by enabling the operator reactions to be
natural and intuitive. Varian touch solution also helps reduce the complexity
of operations by eliminating the often cumbersome traditional operator
interfaces. They simplify the task and reduce training time for operators while
adding more capabilities to the machine itself.
Touching the display (with Stylus-Pen or Finger) will simplified the operation
of the 450-GC.
Varian, Inc.
Page: 45
Status on Main
Screen
LED
Not Ready
LED
Ready
LED
Run
Creation
Creating
ON
OFF
OFF
Initialization
Initialization
ON
OFF
OFF
Equilibration
Not Ready
ON
OFF
OFF
Stabilization
Stabilizing
BLINKING
OFF
OFF
Monitor
Monitor
BLINKING
OFF
OFF
Pre-Run
PreRun
BLINKING
OFF
OFF
Sample Preparation
Sample Prep
OFF
BLINKING
OFF
Ready
Ready
OFF
ON
OFF
Run
OFF
OFF
BLINKING
ON
OFF
BLINKING
Standby
Standby
ON
OFF
OFF
Fault
Error
ON
OFF
OFF
Page: 46
Varian, Inc.
Error LED
Led
Error
Event
OFF
Alert
BLINKING
Error
ON
Table 18: Error LED
Varian, Inc.
Page: 47
Start/Stop GC Run
Page: 48
Varian, Inc.
Status page
Allows the user to view the current status of the various components of the
450-GC.
It gives an overview about the status of the GC components, split up in the
positions front, middle and rear.
Varian, Inc.
The
The
The
The
Page: 49
Version
Information
This screen will give information about the Software Versions used in the
450-GC. The About 450GC tab shows the Hostname and IP- and MAC
address.
The Detailed tab give detailed firmware information about Hardware (mainand displayboard) Injector and detector EFCs, AutoSampler (if present)
firmware.
Page: 50
Varian, Inc.
ADVANCED FUNCTIONS
Throughout the keyboard direct access is possible to advanced GC functions.
Enter
Varian, Inc.
Page: 51
DISPLAY
The display is a LCD screen in combination with a touch-screen.
By touching the display (Stylus-Pen or finger) it is possible to control the GC.
The use of the Stylus-Pen is advised, it is more accurate and will
keep the display free of fingerprints. The Stylus-Pen is located at
the bottom right side of the Display/Keyboard, on page 45.
Page: 52
Varian, Inc.
Title Bar
This bar is always visible with identical information, only the page name will
change.
Automation Indicator
Varian, Inc.
Page: 53
Local Automation,
run 2 of 20 is busy
Sampler information,
indicating run sample
Goto GC Log
Needs customer attention
on page 84
Goto Methods page,
method open, save page,
on page 61
Page: 54
Varian, Inc.
Varian, Inc.
Page: 55
The Valve Table screen gives timed control over the valves (defined in the
Setup/Sample Delivery).
All valves are connected (hardware) to connector J42/J43 External
events, on page 40 or to the option board connector J302/J30,on page 43.
The position of the valve is indicated by its energized (+) or de-energized (-)
state.
Only the enabled (on page 71) valves are visible.
Page: 56
Varian, Inc.
Injector Screen
screen allows you to modify the Injector parameters and
The INJECTOR
view the status of individual Injectors. If more than one Injector is installed,
pressing the corresponding tab will guide you to other Injector page.
The above example is for an 1177 Injector. Further details for programming
Injectors can be found in the Injector screen on page 97 of this manual.
Varian, Inc.
Page: 57
EFC modules are identified by type (EFC21 in this example) and location
(front , middle , rear ) to coincide with the sample delivery or injection
device with which they are associated. Type EFC21 is used with the 1177 or
1079 PTV. The above screen is a typical EFC program for type EFC21. In this
case, a pressure program is applied to keep the column flow rate constant while
the column temperature is ramped.
Four types of Injector EFC modules exist and each type is associated with a
specific GC Injector system.
The Setup menu identifies each EFC type and its associated Injector system by
location In Table 19, on page 59 is a description of each EFC type with used
Injector.
EFC21 setup calibration (1177 and 1079 PTV Injectors) also contains a Septum
Purge Calibration button and Splitless Vent Flow.
Page: 58
Varian, Inc.
Septum Purge Calibration involves measuring a flow rate at a set pressure and
entering those values into the system to correctly calculate the settings for split
and column flow. The Splitless Vent Flow provides a vent purge for runs in
which the split valve remains closed for the duration of the run. It should be set
to at least 20 mL/min.
Sample Introduction
System
EFC
Module
1177
EFC21
1079 PTV
EFC21
1041 Injector
EFC23
1041 Injector
EFC24
1061 Injector
EFC23
1061 Injector
EFC24
EFC24
EFC25
EFC25
EFC23
Varian, Inc.
Page: 59
Oven Screen
allows you to establish the column oven temperature,
The OVEN screen
build a temperature program, or view the column oven status.
The status field for the OVEN indicates the programmed set temperature, the
actual measured temperature and whether the column oven is ready, not ready
or in error (colored dot).
Similar to other programming tables, the initial or isothermal line will always
appear on the first line of the table. You can then build a temperature program
by adding new lines. Below is a column oven program with an initial
temperature, initial hold time, programmed ramp and final hold. The ranges of
the Column Oven temperature are dynamically depending on injectors,
detectors, columns and cryo options.
Stabilization time is the period the column oven will equilibrate for before
allowing the column oven to go READY. It will count down (visible in the Main
Operation screen) from the set value to 0.00 min when the stabilization period
is over. The Oven Power checkbox is used to turn the column oven heater and
fan ON or OFF.
The End Stabilization button is used to bypass the remainder of the stabilization
time.
Page: 60
Varian, Inc.
Methods Screen
By the method screen it is possible to load, save, save as, create (new) and
delete methods.
During first (after factory) startup a Default Method will be loaded, all heated
zones will be set at 50C.
The top line of the screen identifies the method that is currently active, the
current run time and the end time of the active method and whether the GC is
under local or remote control.
The * behind the method name is an indication that the present method
parameters are different then one saved (example: initial oven temperature has
been changed manually).
Saving the method under the same or other name will let the * disappear.
The * will also disappear when the loaded method parameters are in conflict
with the setup parameters. Example: the maximum column temperature in the
setup screen is changed (other column used), if a (older) method is loaded the
system will detect the difference in maximum column temperature and let the *
appear. Saving the method again will let the * disappear. The method will be
automatically adjusted to the present setup parameters.
Touching the Method name in the Title Bar will switch to the Method
screen.
Varian, Inc.
Page: 61
Detector Screen
The above example is for an FID detector. Further details for programming
detectors can be found in the detector screen on page 279 of this manual.
Six types of detector EFC modules exist and each type is associated with a
specific GC detector system.
In Table 20, on page 63 is a description of each EFC type with the used
detector.
Page: 62
Varian, Inc.
Detector Type
FID
(Flame Ionization Detector)
EFC
Module
Type EFC11
TSD
(Thermionic Specific
Detector)
ECD
(Electron Capture Detector)
Type EFC12
Type EFC13
TCD
(Thermal Conductivity
Detector)
Type EFC14/16
PFPD
(Pulsed Flame Photometric
Detector)
Type EFC15
TCD Make-up
EFC14: 0 60 mL/min (2x)
Reference: TCD Reference
EFC16: makeup: 0 60 mL/min
Ref: 0 55 mL/min
Hydrogen:
Flow controller 0-33 mL/min
Air 1:
Flow controller 0-33 mL/min
Air 2:
Flow controller 0-55 mL/min
Analog Output
For detailed description of the analog outputs click here, on page 280.
Varian, Inc.
Page: 63
Setup/Calibration
key allows access to the instrument configuration
The Instrument SETUP
screens for the purpose of viewing or modifying the 450-GC configuration, it
also allows calibrating parts of the GC hardware.
The 450-GC has already been configured at the factory and requires only
minor input from the user.
Access to SETUP screen is only possible if:
GC in not in run state
No error(s) are present (in case of an error, goto error log on page
84 and reset error before continue.
Calibration
The Calibration page will appear:
Page: 64
Varian, Inc.
Setup
More details concerning the Setup Tabs just click on the links below:
Configuration Tab on page 66, Sample Delivery Tab on page 68,
Oven Tab on page 73,
Column Tab on page 74,
Detector Tab on page 75,
Injector Tab on page 77,
Preferences Tab on page 78, System Tab on page 79
Varian, Inc.
Page: 65
Configuration Tab
In the Configuration Tab it is possible to set the type of injector, Injector options
and Oven in relation with each of the heated zones (their physical location) of
the 450-GC.
The Column oven always occupies 1 heated zone (zone 7). The other available
heated zones are numbered 1 through 6 and can be occupied by 1 of the
following devices: 1177 Injector, 1079 PTV Injector, 1041 Injector, 1061
Injector, 1093 Injector, SPT, FID, TCD, ECD, TSD, PFPD, large valve oven,
small valve oven, or methanizer. SPT is allowed only in zone 3. Dual TCD is
using 1 heated zone for 2 detectors.
The location typically refers to whether the device is installed in the front,
middle, or rear position on the top of the 450-GC. In the case of detectors the
location also refers to the relative positions of the detector electrometers
installed on the main electronics board. By default, zones 1, 2 and 3 are
identified as front, middle and rear and zones 4, 5 and 6 are identified as front,
middle and rear.
Zones 1, 2 and 3 are normally used for Injectors and zones 4, 5 and 6 are used
for detectors.
Page: 66
Varian, Inc.
External Device
If an external device is connected that needs to be ready prior GC start, the GC
state (ready, not ready) can include the state of the external device.
When the external device state is included a selection must be made if the
ready-in (pin-1 on connector J401 Analog Out on page 36 ) of the GC is
normally closed or normally open.
Enable
check box
Ready-In
check box
Grayed out
Varian, Inc.
Page: 67
Varian, Inc.
Injector position
Front
Front
Middle
Rear
Middle
Rear
Table 21: Sampler/Injector position
Rear Injector
Middle Injector
Middle Injector
Front Injector
In case the sampler is mounted in the front position it can inject in the front and
middle Injector. In case the sampler is mounted in the rear position it can inject
in the middle and rear Injector. In the sample delivery page of Setup, the front
sampler Injection position is the first Injector (clock wise rotation of the sampler)
the sampler is able to reach. The rear sampler Injection position is the second
Injector the sampler is able to reach.
If 1 Injector is installed, make certain to specify the correct Injector. An incorrect
setting may cause the CP-8400/8410 to inject into the top cover resulting in
Varian, Inc.
Page: 69
damage to the syringe and the top cover. Select both setting if you have 2
Injectors that you wish to use with the sampler. Note that selecting both allows
the sampler to use 2 Injector volumes when dual/duplicate mode is used. If you
configure a single Injector, you will see a single injection volume in the
automation tables.
Syringe Size (1 L, 2L, 5 L, 10 L, 50 L, 100 L and 250 L). (Note that
this field is not visible when the AutoSampler is in the SPME mode). Enter the
syringe volume you will install and use with your CP-8400/8410. The 450-GC
will use this value to set the correct entry selections and any needed internal
settings to correctly use the indicated syringe size.
If you are simply changing a syringe with a different volume syringe you can
press OK and Exit at this point to accept the new syringe size.
Front/Rear Injection Depth (0-100 %). The Injector Needle Depth parameters
define how far into the designated Injector the needle will penetrate. 100%
causes the needle to be completely inserted to its 2-inch length. The valid
range is settable in 1% steps. The default is 90%.
Front/Rear Speed (0-100 cm/s) The Injector Injection Speed parameters define
the rate at which the needle will penetrate the Injector septum and descend to
the set Injector Needle Depth. The valid range is settable in 1 steps. The default
is 30 cm/s.
Page: 70
Varian, Inc.
Valving
This screen configures the 7 standard external events available on the 450-GC
that are typically used to control the 1177 and 1079 split valve with manual
pneumatics (non-EFC) or actuators for sampling or switching valves. Each
valve is identified in SETUP by a name that indicates its application. When a
name is chosen this also defines the default states (Abbreviation, - DeEnergized and + Energized) of the valve.
The Option board has 8 optional external events available and will be
numbered from 8 till 15.
Only valves that are enabled are shown in the method pages.
Varian, Inc.
Page: 71
In Table 22 the complete set of valve names and associated states that are
available for programming.
Valve Type
Abbreviation
FSV
MSV
RSV
GSV
LSV
S
IS
SS
SB
BD
BV
CS
IBD
IBV
AI
SI
MB
SPT
EA
EB
EC
ED
EE
EF
EG
EH
EI
EJ
PFOS
QSV
De-Energized
State (-)
ON (split)
ON (split)
ON (split)
Fill
Fill
OFF
OFF
OFF
Series
Forward
Forward
Column 1
Fill + Backflush
Fill + Backflush
Column 1
Fill
Series
SPT Desorb
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
OFF
Page: 72
Varian, Inc.
Oven Tab
The default temperature limit of the column oven is set at 450C (If the column
limit is set lower then 450 C the oven limit is set accordantly, the column with
the lowest column temperature limit is used as maximum temperature limit), but
can be set to a lower value.
If the 450-GC is idle for more then the Standby Timeout time, the column oven
temperature is set to the Standby Temperature and 450-GC will go into
Standby mode, see operation screen. This standby temperature will be invoked
after the timer counts to zero, provided another run is not started before the
timer has timed out.
The default Standby Temperature is 60C.
The default Standby Timeout time is 120 minutes.
A value of 0 minutes in Standby Timeout will disable the Standby Temperature.
The Coolant checkbox turns on or off the cryogenic supply for the oven, if
installed. Coolant type can be LN2 or LCO2..
Coolant time-out refers to the period of time for which the coolant will remain on
after it is turned on. For example, if the set temperature has not been reached
or the run has not started within the specified time, the coolant supply will be
turned off.
Leave always the Injector(s) / Detector(s) heating ON (enable) and at a
temperature of at least 100 C when using cryogenic column oven cooling!
Varian, Inc.
Page: 73
Column Tab
Column parameters such as packed column, column length, internal diameter
and choice of carrier gas are used for the flow calculations of EFC.
The column limit temperature(s) have a direct relation with the maximum oven
temperature. If set at 350 C the oven temperature limit will be also 350 C.
The column with the lowest column temperature limit is used as maximum
temperature limit.
Carrier gas is configured by carrier gas selected for the corresponding Injector.
In case of a packed column the parameters column length and internal
diameter are not used.
Page: 74
Varian, Inc.
Detector Tab
The detector tab identifies each Detector system and its associated EFC type
detector system by location (front , middle , rear ). In case more information
is available this will be shown on next pages. The last page (in example
Page 2) Analog output is shown.
In Table 20, on page 63 is a description of each EFC type with the related
detector.
Detector EFC involves selecting the relevant carrier/make-up/reference gas
type.
The default temperature limit of all detectors is set at 450C, but can be set to a
lower value.
The 450-GC monitors the FID background current continuously and will attempt
to light the flame (only if the checkbox Check for Flameout is checked) if the
current drops below a specified threshold and the detector temperature is still
above 100 C (for all Ignite conditions click here on page 287. Up to three
attempts to ignite the flame will be made before the 450-GC reports a flame-out
fault.
Varian, Inc.
Page: 75
On the second page (Page 2 or Page 3) the Detector Analog Output screen will
be visible (only if the Analog option board is present).
The Port name can be changed by the button Edit.
The Port name will be used in the Detector screen, on page 280.
If the Range setting must be changed, make sure that the Range setting
matches the mainboard jumper settings!
Page: 76
Varian, Inc.
Injector Tab
The Setup menu identifies each Injector system and its associated EFC type
detector system by location (front , middle , rear ).
The default temperature limit of all Injectors is set at 450C, but can be set to a
lower value.
Injector EFC involves selecting the relevant carrier gas/outlet Pressure and
more parameters.
In Table 19, on page 59 is a description of each EFC type with used Injector.
Outlet Pressure is used to distinguish between a column venting to atmosphere
or a column venting to a vacuum source such as a mass spectrometer.
A specific Gas Saver Flow and a timeout period may be set. A time entry of 0
turns the Gas Saver feature off. If a run is not started or setup activated
before the timer expires, the set Gas Saver flow rate will be started. A message
will be logged in the instrument log.
The Splitless Vent Flow is the flow used for calculating the total flow when the
method split program contains only splitless events.
Varian, Inc.
Page: 77
Preferences Tab
In this tab it is possible to set General parameters.
General:
Page: 78
Varian, Inc.
System Tab
In System tab it is possible to set General and Network parameters.
The Network Change button is un-usable when the network-switch at the Rear
panel, on page 42 is set at DHCP a (BOOTP).
General:
Date and Time, pressing the change button, it is possible to change both values.
Page: 79
Network:
Network, pressing the edit button it is possible to change network settings.
Its possible to set a fixed IP address in the 450-GC.
Check before changing the network settings if on the Rear panel, on page 42
the dip-switch is set to the Fixed IP position. If not set switch in Fixed IP
position and restart the GC.
Fill in the IP address and Hostname, make sure the Subnet Mask and default
Gateway are compatible with the PC settings.
Page: 80
Varian, Inc.
Varian, Inc.
Page: 81
Final, as verification, touch the center of the cross in the center of the screen.
If the calibration results are accepted by the software, press the Start/stop
button
Page: 82
Varian, Inc.
If the touch screen is not accessible for screen calibration, it is possible to access via
the hard-keys.
Varian, Inc.
Page: 83
Error/Log Handling
will be visible.
a A list of all errors with error conditions, can be found in Appendix A: Error Codes on page 479
Page: 84
Varian, Inc.
Varian, Inc.
Page: 85
Page: 86
Varian, Inc.
Loacl Automation
LOCAL AUTOMATION
Local Automation is a tool to operate multiple analysis and single method
directly via the GC. The input is done via the Display and Keyboard and is only
to a Workstation.
button.
To start the local automation press the start
The run will start as soon as the 450-GC state is ready.
A local automation running button
will pause the local automation run
Varian, Inc.
Page: 87
Local Automation
Page: 88
Varian, Inc.
Communications
COMMUNICATIONS
Introduction
The 450-GC uses Ethernet as the method of communications for instrument
control and for sending digitized data to an external data system. In addition,
the 450-GC is equipped with (optional) analog output ports which allow analog
data to be sent to external integrators or other analog data systems.
The 450-GC is equipped to handle digital and analog data simultaneously. The
Ethernet communication option allows Workstation control of the 450-GC and
transmits up to three channels of detector data to the Workstation.
The (optional) analog connection can be configured with up to three detector
analog outputs and 8 external timed events.
Various cable options are available for sending analog data to integrators and
data.
Ethernet Communications
Ethernet refers to the type of hardware connection used to transfer digital data
between the 450-GC and the data system to which it is attached.
The 450-GC has a RJ45 connector located on the rear of the GC. The Ethernet
cable connects the 450-GC to the Workstation or switch on a company network.
Varian, Inc.
Page: 89
Communications
5. Make sure that the network interface is compatible with the GC settings.
6. Select Discover. .
7. Check if the BOOTP (DHCP) switch at the rear of the GC on page 42 is in
the upper position, switch the GC Off/On or On.
Page: 90
Varian, Inc.
Communications
8. When the MAC address (from the GC) appears select the address and
press Validate.
9. Fill in IP-address and the host name or description and press APPLY.
10. Select System in the organization bar to create a new system.
11. Create a new Instrument and select the Acquisition and Sequence server
and press Next.
12. Select Groups/Project and press Next.
13. Add the devices (example 450-GC and NO sampler).
Varian, Inc.
Page: 91
Communications
14. Only the 450-GC driver must be installed when a sampler is installed.
15. Press OK.
16. Press Yes to access the system.
17. Select communication and press the magnifying glass to see the instrument
or type in the IP-address.
Varian, Inc.
Communications
Varian, Inc.
Page: 93
Communications
ANALOG CABLES
The 450-GC has a number of analog cables options, depending on the device to which
the cable is connected. All cables have a 15-pin Dsub connector on one end to attach
to J401 (analog out) on page 36 on the 450-GC and have appropriate connectors on
the other end of the cable for the devices to which they are being attached.
Partnumber
Description
Picture
392567501
392567201
392567301
390793805
CBLV3800
Page: 94
Varian, Inc.
Communications
QU040.0115
CP742629
Varian, Inc.
Page: 95
Communications
Page: 96
Varian, Inc.
Sample Introduction
SAMPLE INTRODUCTION
Sample introduction devices refer to the hardware used to introduce the sample
into the Gas Chromatograph. These include AutoSamplers, liquid Injectors, gas
sampling valves and liquid sampling valves. This screen also covers the
pneumatic systems associated with the various sample introduction systems.
The various pneumatic systems are described with the individual sample
introduction systems.
AUTOSAMPLERS
Two Varian samplers are available for the 450-GC, the CP-8400 AutoSampler
and the CP-8410 AutoInjector, on page 103. Also a Combi PAL sampler
(Liquid, Headspace, SPME Injection System) is available for the 450-GC, more
details can be found at www.varianinc.com.
LIQUID INJECTORS
There are 5 liquid Injectors available on the 450-GC Gas Chromatograph:
1. The 1177 Split/Splitless Capillary Injector on page 153 is optimized
especially for split and splitless injections. The Injector is designed to accept a
variety of inserts allowing tailored configurations for either the Split or Splitless
injection method.
2. The 1079 PTV Injector on page 175 can be used in a variety of sampling
modes including isothermal split and splitless, temperature ramped splitless,
on-column and large volume injection. The 1079 PTV can be temperature
programmed and cooled by cryogenics.
3. The 1041 On-Column Injector on page 205 is designed for use with wide bore
capillary columns or packed columns. The standard Injector is supplied with an
adapter for wide bore capillary columns. Optional adapter kits are available for
1/8" and 1/4" packed columns.
4. The 1061 Flash Vaporization Injector on page 219 is also designed for use
with wide bore capillary or packed columns. This Injector uses a glass insert
inside the Injector body. The standard Injector is supplied with an insert for use
with wide bore capillary columns. Optional adapters are available for 1/8" and
1/4" packed columns.
5. The 1093 Septum-Equipped Programmable Injector (SPI), on page 231
gives optimum performance for non-vaporizing (cold) injection into conventional
Varian, Inc.
Page: 97
Sample Introduction
Page: 98
Varian, Inc.
Sample Introduction
SAMPLING VALVES
The 450-GC can be configured with a wide variety of gas and liquid sampling
valves. These sampling valves can be unheated or installed in a (valve) oven.
Typically a sampling valve is installed in series with a liquid Injector and thus
shares the pneumatic system of the Injector, but valves can also be installed
with their own pneumatic systems. Optional actuators are available to automate
the function on the sampling valves. In addition to sampling valves, Varian also
supplies a large number of switching valve configurations.
Load
Fill
SAMPLE OUT
SAMPLE OUT
SAMPLE IN
SAMPLE IN
LIQUID
SAMPLING
VALVE
CARRIER GAS
LIQUID
SAMPLING
VALVE
CARRIER GAS
INJECTOR
INJECTOR
SPLIT VENT
SPLIT VENT
COLUMN
COLUMN
DETECTOR
DETECTOR
Load
Fill
SAMPLE LOOP
SAMPLE OUT
SAMPLE OUT
SAMPLE IN
SAMPLE IN
SAMPLE LOOP
GAS
SAMPLING
VALVE
CARRIER GAS
GAS
SAMPLING
VALVE
CARRIER GAS
INJECTOR
SPLIT VENT
Varian, Inc.
INJECTOR
SPLIT VENT
COLUMN
COLUMN
DETECTOR
DETECTOR
Page: 99
Sample Introduction
The following are the typical locations for valves on the 450-GC:
Heated
Zones for
Valves
Unheated Valves located on the top panel of the instrument, beneath the
Injector cover. The area behind the Injectors can accommodate multiple (3possible) unheated valves.
Up to three valves can be installed on the right side of the column oven.
This is the least preferred location due to temperature cycling of the column
oven and will effect the valve live negatively.
If valves are located in a valve oven then they are configured in 450-GC as
front , middle , rear , valve ovens. The valve oven temperature is
. Note that when multiple valves are
programmed in the Oven tab
installed in one valve oven the temperature of the oven should always be set
lower than the lowest temperature rated valve. In general valves have upper
temperature ratings of 220 C or 350 C.
Page: 100
Varian, Inc.
Sample Introduction
Pneumatic Systems
Most sample introduction systems may be equipped with manual pneumatics or
Electronic Flow Control (EFC). The following table summarizes the available
pneumatics systems for liquid Injectors, sampling valves, purge and trap and
heated headspace systems.
Sample Introduction
System
EFC
Module
EFC21
EFC21
EFC23
EFC24
EFC23
EFC24
EFC24
EFC25
EFC25
EFC23
Optional Manual
Pneumatics System
Manual pneumatics - inlet flow
controller/back pressure regulator
with pressure gauge.
Manual pneumatics - inlet flow
controller/back pressure regulator
with pressure gauge.
Manual digital flow controller with
pressure gauge.
Manual pressure regulator with
gauge.
Manual digital flow controller with
pressure gauge.
Manual pressure regulator with
gauge.
Manual pressure regulator with
gauge.
Manual pneumatics - inlet flow
controller/backpressure regulator
with pressure gauge.
Manual pneumatics - inlet flow
controller/back pressure regulator
with pressure gauge.
Manual digital flow controller with
pressure gauge.
Figure 13: Pneumatic Systems for Liquid Injectors and Sampling Valves
Varian, Inc.
Page: 101
Sample Introduction
Page: 102
Varian, Inc.
AutoSamplers
CP-8400/8410 AUTOSAMPLER
INTRODUCTION
We strongly recommend that you read this section before you start using your
CP-8400 AutoSampler or CP-8410 AutoInjector.
The CP-8400 is a 100 vial capacity AutoSampler.
The CP-8410 is an AutoInjector based on the CP-8400; it has a manually
positioned tray and can accommodate up to 10, 6 and 5 each of 2 mL, 5 mL
and 10 mL vials respectively.
Both the AutoSampler and AutoInjector can inject into 2 Injectors; the Sampler
can inject into either one of the 2 Injectors or into both (either from the same
sample vial or from adjacent vials) during the same analysis.
Varian, Inc.
Page: 103
AutoSamplers
Both the CP-8400 and CP-8410 are preprogrammed with several injection
modes (Standard On-Column, Standard Split/Splitless, Viscous, Volatile and
Neat) that are optimized for most common types of samples. The User Defined
mode allows access to all parameters and should be used with discretion since
injection accuracy and precision may be adversely affected by incorrect choice
of certain parameters. The SPME mode requires different hardware (optional).
If the SPME mode is selected the other injection modes will not be available.
This section will cover AutoSampler installation, software setup and calibration
to the correct sampling and injection positions, a description of the operating
parameters and some examples of how you may wish to use the CP-8400/8410
to accomplish your analyses. If, after reading this section you still have
questions or encounter a problem that is not covered, please contact your local
Varian Technical Support for further assistance.
Page: 104
Varian, Inc.
AutoSamplers
Varian, Inc.
Page: 105
AutoSamplers
The front position allows the CP-8400/8410 to use the front and middle
Injectors.
The rear position allows the CP-8400/8410 to use the middle and rear
Injectors.
2. Remove any old equipment or old autosamplers from the top of the GC.
3. Place the CP-8400/8410 near the GC, manually rotate the Injector tower so
that it does not obstruct the Sample-carrousel/tray.
4. Remove the Sample-carrousel/tray.
5. Remove the old 450-GC Top-Cover by lifting up and removing the Detector
cover. Remove the Detector cover by lifting the rear of the cover and sliding
it towards the rear of the GC and up. Remove the 4 screws holding the
Injector cover down on the top of the GC. Retain the screws and set the old
Top-Cover aside.
6. Install the sampler Top-Cover so that the guide-pins on the cover fit into the
corresponding slots on the chassis. Screw the Sampler Top-Cover in place.
Page: 106
Varian, Inc.
AutoSamplers
Varian, Inc.
Page: 107
AutoSamplers
11. Place the CP-8400/8410 onto the Top-cover, aligning it with either the front
or rear mounting position (See Figure 14) as appropriate.
12. The base is secured to the Top-Cover via 2 (short) screws at the rear and
3 (long) screws underneath the AutoSampler/AutoInjector samplecarrousel/tray.
Page: 108
Varian, Inc.
AutoSamplers
14. Remove protective shipping materials from the Injector carriage and install
the syringe into the AutoSampler Injector tower by opening the clear plastic
door, carefully inserting the syringe into its holder and the needle through
its guide, then fasten the plunger screw on the top and snap the retaining
clip over the syringe.
15. Close the clear plastic
sampler door.
Plunger Retaining Nut
(unscrew to remove
syringe and tighten
when installing)
Varian, Inc.
Page: 109
AutoSamplers
Page: 110
Varian, Inc.
AutoSamplers
Varian, Inc.
Page: 111
AutoSamplers
CALIBRATION
Sampler calibration will align the Sampler components (Tower, Vials and
Needle) and the GC Injector ports.
Calibration procedure is important for a trouble free working GC/Sampler
combination.
Calibration must be carried out in the following cases:
Sampler has been removed from the GC and after maintenance/repair back
on the GC.
Do not recalibrate the Injector positions after normal Injector maintenance such
as changing septa or inserts.
Two calibration methods are present:
- Vial calibration, on page 113.
- GC Injector calibration, on page 116.
Page: 112
Varian, Inc.
AutoSamplers
Vial calibration
The Vial Calibration procedure will allow the AutoSampler/AutoInjector to learn
the tower and carrousel/tray positions so that the syringe is positioned directly
over the middle of the first vial in the sample-carrousel (Vial 0) or the middle of
the first 2 mL vial of the AutoInjector sample tray (Vial 1).
The CP-8400 may use a number of different sample vial form factors; however,
you should use all of the same vials for each session.
Mixing different vials is strongly discouraged. If you calibrate using one vial type
and then begin using another vial type with a different height, it is strongly
recommended that you recalibrate the Vial 0 Position to avoid bent needles and
other minor problems.
In order to accurately calibrate Vial positions use the calibration
tool (P/N: CP740889). The calibration tool simulates a standard
vial. The tool is used for both the CP-8400 and CP-8410
AutoSamplers.
The calibration tool can be placed (depending of the
carrousel/tray-material) in 2 different ways in the tray.
CP-8410 (polymer tray)
calibration tool in Vial position 1
Varian, Inc.
Page: 113
AutoSamplers
Varian, Inc.
AutoSamplers
CP-8400/8410 has found the approximate Vial 0 (or Vial 1) position, visually
check that the syringe sled is not resting on the Vial top.
If the syringe sled is resting on the vial top, press the Syringe Up button
until the syringe sled is clear of the Vial top by approximately 1 mm.
5. Start the calibration process by adjusting the tower position. Move the
Tower CCW or CW until the needle-guide is exactly cover the top of the
calibration tool.
6. For CP-8400, set the sample-carrousel position. You may have to switch
back and forth between the tower and sample-carrousel one or more times
to determine the optimum rotational positions.
7. Adjust the syringe sled height so that the needle is just barely above the
calibration tool top. You should be able to see about a papers thickness of
space between the syringe sled needle guide and the top of the calibration
tool. The penetration depth into the vial is measured from the bottom of the
needle guide, so this height will give you optimum penetration depth.
8. Press the Accept button if the calibration of Vial 0 or 1 is complete, the
calibration data will be saved.
9. Press Exit button to exit the calibration screen, without saving the
calibration data.
If the used Vial is shorter/longer then the calibration tool its
advisable to repeat only the Syringe calibration.
Varian, Inc.
Page: 115
AutoSamplers
GC Injector calibration
key and select CALIBRATION, select on the next
1. Press the SETUP
page Sampler CALIBRATE.
Sampler mounted in
the Front position
Injector 2
Injector 1
Page: 116
Varian, Inc.
AutoSamplers
Sampler mounted in
the Rear position
Injector 2
Injector 1
Page: 117
AutoSamplers
5. If the needle guide descends so that it is either resting on the inject switch
or below the needle cone on the inject switch, use the Up button move the
sled up (10 steps) until the needle guide is just above the Injector cone.
6. Using the CCW (counterclockwise) and CW (clockwise) button. Position
the tower directly over the Injector Inject Switch Locator cone.
7. Adjust the height (using the Up and Down button) of the needle guide
such that the sled is just resting on or very slightly depressing the Injector
Switch and is covering the locator cone.
8. Press the Accept button if the calibration of Injector 1 is complete, the
calibration data will be saved.
9. Calibrate Injector 2 (if present) using the above mentioned procedure.
10. Press the Exit button if calibration is finished or leaving the calibration
page without saving is needed.
The Reset Calibration button will restore the factory settings to all
calibration values.
Page: 118
Varian, Inc.
AutoSamplers
SYRINGE CHECK/REPLACE
For changing the syringe size continue on the next page.
If the syringe has to be replaced by the same size follow the procedure below:
key and select CALIBRATION, select on the next
1. Press the SETUP
page Sampler Change Syringe.
2. The next massage will pop-up and the Sampler Tower will go to his Syringe
Replace/Check Syringe position.
3. Check or Replace the Syringe when the above pop-up message appears.
4. After syringe replacement the syringe plunger stroke counter can be set to
zero, press OK.
5. After Syringe check (no reset of the counter necessary), press Cancel.
Varian, Inc.
Page: 119
AutoSamplers
Page: 120
Varian, Inc.
AutoSamplers
OPERATION
PARAMETERS AND QUICK REFERENCE
The layout of the screens and the placement of the parameters followed the philosophy
of keeping those parameters that an operator may wish to change for a series of
samples (such as the injection volume, vial range, GC method) within the Automation
pages for quick access and change.
Those parameters that drive the AutoSampler actions such as clean, viscosity settings
and needle residence times are located within the AutoSampler Method.
The CP-8400/8410 AutoSampler is preprogrammed with settings and actions that are
appropriate for most common analyses. The following diagrams represent the hierarchy
of screens that follow from selecting the indicated key on the 450-GC front panel.
Varian, Inc.
Page: 121
AutoSamplers
SAMPLER CONFIGURATION
The AutoSampler will be delivered in one of two (Liquid or SPME)
configurations.
Default is the Liquid Sampling configuration.
The SPME operation can be thought of as fundamentally reversed from liquid
sampling in terms of the actions the plunger performs during the sampling
process. In liquid sampling, the plunger is at its home state more or less fully
descended and sampling is done by withdrawing the plunger from the syringe
barrel in order to take an aliquot for injection. In SPME however, the plunger is
at the top of its travel when it is in the home position and the SPME fiber must
be exposed by inserting the plunger into the fiber holder barrel to expose the
fiber.
The SPME Sampling mode can not be selected by the 450-GC
display/keyboard. In order to configure the system to use SPME you
must set a switch on the syringe sled to the correct position and install
the special SPME fiber holder block. This means of setting up the
system to use SPME was adopted to avoid unintentional damage to
either SPME hardware or syringes. Ones the installation is completed,
the 450-GC will recognize the sampler mode.
Page: 122
Varian, Inc.
AutoSamplers
Varian, Inc.
Page: 123
AutoSamplers
Sampler
sequence
SyringeSize
(L)
10
50
- Volume (L)
0.75
1.5
3.75
7.5
0.5
2.5
187.5
20
50
50
62.5
10
25
50
25
0.75
1.5
3.75
7.5
37.5
75
187.5
0.5
2.5
10
20
50
10
25
50
25
50
62.5
Clean Strokes
(sample sample)
0.2
0.4
10
20
50
10
25
50
25
50
62.5
0.75
1.5
3.75
7.5
37.5
75
187.5
Page: 124
75
10
37.5
250
100
0.0
0.0
0.0
0.0
0.0
0.0
0.0
Draw Sample
- Plunger Speed During Fill (L/s)
0.2
0.4
10
20
50
0.0
0.0
0.0
0.0
0.0
0.0
0.0
Varian, Inc.
AutoSamplers
Sampler
sequence
5
6
SyringeSize
(L)
10
50
100
250
0.1
0.2
0.5
0.5
0.5
Inject Sample
- Plunger Speed During Injection (L/s)
0.5
2.5
10
12.5
0.0
0.0
0.0
0.0
0.0
0.0
0.0
6.0
6.0
6.0
6.0
6.0
6.0
6.0
0.75
1.5
3.75
7.5
37.5
75
187.5
0.5
2.5
10
20
50
10
25
50
25
50
62.5
10
50
Varian, Inc.
100
250
- Volume (L)
0.75
1.5
3.75
7.5
9.9
9.9
9.9
9.9
37.5
75
187.5
9.9
9.9
0.5
2.5
9.9
10
20
50
50
62.5
SyringeSize
(L)
10
25
50
25
0.75
1.5
3.75
7.5
37.5
75
187.5
9.9
9.9
9.9
9.9
9.9
9.9
9.9
0.5
2.5
10
20
50
Page: 125
AutoSamplers
Sampler
sequence
10
50
100
250
10
25
50
25
50
62.5
0.4
10
20
50
10
25
50
25
50
62.5
0.75
1.5
3.75
7.5
37.5
75
187.5
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.2
0.4
10
20
50
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.1
0.2
0.5
0.5
0.5
10
25
50
25
50
62.5
Draw Sample
0.2
Clean Strokes
(sample sample)
- Plunger Speed Up (L/s)
SyringeSize
(L)
0.0
0.0
0.0
0.0
0.0
0.0
0.0
6.0
6.0
6.0
6.0
6.0
6.0
6.0
0.75
1.5
3.75
7.5
37.5
75
187.5
0.5
2.5
10
20
50
10
25
50
25
50
62.5
Page: 126
Varian, Inc.
AutoSamplers
Sampler
sequence
SyringeSize
(L)
Neat - mode
10
50
100
250
0.75
1.5
3.75
7.5
37.5
75
187.5
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.5
2.5
10
20
50
10
25
50
25
50
62.5
0.75
1.5
3.75
7.5
37.5
75
187.5
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.5
2.5
10
20
50
10
25
50
25
50
62.5
Coating Strokes
(same movent as Pre-Injection Sample
Clean Flushes)
- Volume (L)
0.75
1.5
3.75
7.5
37.5
75
187.5
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.5
2.5
10
20
50
10
25
50
25
50
62.5
Clean Strokes
(sample sample)
- Plunger Speed Up (L/s)
0.2
0.4
10
20
50
0.0
0.0
0.0
0.0
0.0
0.0
0.0
10
25
50
25
50
62.5
10
50
100
250
Draw Sample
- Plunger Speed During Fill (L/s)
0.2
0.4
10
20
50
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.1
0.2
0.5
0.5
0.5
10
25
50
25
50
62.5
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.0
Inject Sample
- Plunger Speed During Injection (L/s)
Varian, Inc.
Page: 127
AutoSamplers
Sampler
sequence
SyringeSize
(L)
Neat - mode
10
50
100
250
- Volume (L)
0.75
1.5
3.75
7.5
0.5
2.5
37.5
75
187.5
10
20
50
10
25
50
25
50
62.5
Sampler
sequence
SyringeSize
(L)
Viscous - mode
10
50
100
250
- Volume (L)
0.75
1.5
3.75
7.5
37.5
75
187.5
0.0
0.0
0.0
0.0
0.0
0.0
0.0
0.5
2.5
10
20
50
10
25
50
25
50
62.5
0.75
1.5
3.75
7.5
37.5
75
187.5
9.9
9.9
9.9
9.9
9.9
9.9
9.9
0.1
0.2
0.5
10
25
10
25
50
25
50
62.5
Page: 128
Varian, Inc.
AutoSamplers
Sampler
sequence
SyringeSize
(L)
Viscous - mode
1
3
0.5
2.5
10
20
50
10
25
50
25
50
62.5
0.1
0.2
0.5
0.5
0.5
2.0
4.0
0.1
0.2
0.5
10
25
9.9
9.9
9.9
9.9
9.9
9.9
9.9
0.1
0.2
0.5
0.5
0.5
100
250
Draw Sample
50
10
Inject Sample
- Plunger Speed During Injection (L/s)
0.5
2.5
10
20
50
0.0
0.0
0.0
0.0
0.0
0.0
0.0
12.0
12.0
12.0
12.0
12.0
12.0
12.0
- Volume (L)
0.75
1.5
3.75
7.5
37.5
75
187.5
0.5
2.5
10
20
50
10
25
50
25
50
62.5
Varian, Inc.
Page: 129
AutoSamplers
Sampler
sequence
SyringeSize
(L)
Volatile - mode
1
1
1.5
3.75
7.5
37.5
75
187.5
0.0
0.0
0.0
0.0
0.0
0.5
2.5
10
20
50
50
62.5
10
25
50
25
0.75
1.5
3.75
7.5
37.5
75
187.5
9.9
9.9
9.9
9.9
9.9
9.9
9.9
0.1
0.2
0.5
10
25
10
25
50
25
50
62.5
0.5
2.5
10
20
50
10
25
50
25
50
62.5
0.1
0.2
0.5
0.5
0.5
2.0
4.0
0.1
0.2
0.5
10
25
9.9
9.9
9.9
9.9
9.9
9.9
9.9
0.1
0.2
0.5
0.5
0.5
Draw Sample
250
0.0
100
0.0
50
0.75
10
- Volume (L)
Inject Sample
- Plunger Speed During Injection (L/s)
0.5
2.5
10
20
50
0.0
0.0
0.0
0.0
0.0
0.0
0.0
12.0
12.0
12.0
12.0
12.0
12.0
12.0
- Volume (L)
0.75
1.5
3.75
7.5
37.5
75
187.5
0.5
2.5
10
20
50
10
25
50
25
50
62.5
Page: 130
Varian, Inc.
AutoSamplers
Page 1
Prep.Ahead
This parameter is used to select whether the AutoSampler will sample for the
next injection immediately after the current chromatographic run is completed or
to cause the AutoSampler to sample while the current run is in progress thus
improving throughput. Prep-Ahead works for either liquid or SPME sampling
modes. If you set the Prep-Ahead, the system will use the Prep-Ahead Delay
time entered into the method to determine when to begin sampling for the next
run. The Prep-Ahead timer will begin after the AutoSampler has completed all
Varian, Inc.
Page: 131
AutoSamplers
of its actions for the current run. For instance, if you have dual or duplicate
injections selected and the second injection is set to take place 0.5 minutes into
the run, the Prep-Ahead timer for the next sample will start once the second
injection is finished. So if you have set the Prep-Ahead timer to 10 minutes, the
system will actually start sampling at around 10.5 minutes into the run. If your
sample is either heat or light sensitive you may wish to leave the Prep-Ahead
blank. Note that the Prep-Ahead timer is independent of the run time; this
allows you to set a next sampling time while the instrument is either
equilibrating or stabilizing if the sample may be heat or light sensitive. You will
have to experiment with sample and solvent flushes along with any other preinjection settings for your AutoSampler to determine the correct Prep-Ahead
time for optimum cycle time and minimum wait before injection.
Page: 132
Varian, Inc.
AutoSamplers
Varian, Inc.
Page: 133
AutoSamplers
Page: 134
Varian, Inc.
AutoSamplers
Page 2
Clean Mode
Number of Pre-Injection Solvent Clean Flushes (0-99)
This parameter defines the number of times the syringe will draw up solvent,
rotate to the waste cup and expel the syringe contents before the sampling
process begins. The default value is 3.
Varian, Inc.
Page: 135
AutoSamplers
Internal Standard
Internal Standard
This parameter determines whether the system will use an Internal Standard as
part of the sampling procedure. Selecting will cause the CP-8400/8410 to draw
up the specified amount of Internal Standard from the specified vial.
The Internal Standard will always be withdrawn before the sample is withdrawn
to avoid as much as possible cross contamination of the Internal Standard with
analyses. The default value is blank.
Page: 136
Varian, Inc.
AutoSamplers
Varian, Inc.
Page: 137
AutoSamplers
Page 3
Solvent Plug
This parameter determines whether the system will us a solvent plug as part of
the sampling procedure. Selecting will cause the CP-8400/8410 to draw up the
specified amount of solvent from the specified vial. The default value is blank.
Vial for Solvent Plug (I, II, III)
This parameter defines which of the vials will be used as the source for the
solvent plug. The default selection is wash vial I.
Page: 138
Varian, Inc.
AutoSamplers
Abort Clean
Use in case Automation is stopped or a serious fault is detected.
Clean Vial (I, II, III)
This parameter specifies the wash solvent vial that will be used by the Default
Cleaning procedure. The default value is 1.
Clean Volume
This parameter defines the amount of cleaning solvent that will be drawn up
and expelled to waste. The volume is selectable in 0.1 L (depending of the
syringe size). The default value is 5.0 L.
Number of Clean Strokes (0-10)
This defines the number of times the cleaning solvent will be drawn up and
expelled into the waste cup. The default value is 1.
Clean Drawup Speed
Varian, Inc.
Page: 139
AutoSamplers
This parameter specifies how quickly the plunger will be withdrawn from the
syringe drawing solvent into the barrel. The default value is 5 L/sec
(depending of the syringe size). This same value should be used as the plunger
speed when expelling the syringe contents into the waste cup.
Varian, Inc.
AutoSamplers
For liquid sampling you will get best results if the level of the liquid in the
vial does not contact the fiber sheath thus potentially causing carry-over.
5. The AutoSampler will next perform the SPME sample introduction into the
Injector. The SPME fiber holder will descend and penetrate the Injector
septum. The fiber will then be extended the specified Desorb time on the
main SPME parameters page.
6. Upon withdrawal from the Injector, any post injection clean actions will be
performed.
The SPME mode screens and parameters are summarized below:
Varian, Inc.
Page: 141
AutoSamplers
AutoSamplers
Syringe bracket
Set to SPME
(UP-position)
Varian, Inc.
Page: 143
AutoSamplers
Installation of the fiber into the SPME syringe, i.e. the fiber holder, is explained
in the fiber-holder manual.
11. Close the cover on the tower and turn on the power to the GC. The GC will
automatically detect the new configuration of the switch and will respond
accordingly.
12. Recalibrate the sampler, on page 112.
13. Adjust the Injector insert, on page 145 to SPME injection.
14. The instrument is now in the SPME mode.
Page: 144
Varian, Inc.
AutoSamplers
Injector Insert
To assure optimum results when desorbing a SPME fiber, it is important to use
the correct Injector insert. Use a straight, unpacked insert with an internal
diameter of 0.8 mm (P/N: 392611948 for the 1079 Injector). Smaller inserts do
not allow the fiber sheath to penetrate the Injector, while large inserts will
broaden early-eluting analytes.
The SPME fiber is mounted in the fiber sheath. During septum
penetration, the fiber is completely retracted into the sheath for
protection. A blunt sheath with a relatively large diameter may result in
septum leakage. If you encounter frequent septum failure, use the
optional Merlin Microseal TM.
Plunger
Barrel
Fiber Sheath
Color-coded
Screw Hub
Sealing Septum
Retaining Nut
Fiber Sheath
Fiber
Attachment Rod
(pierces septum
of sample vial
and GC injector)
SPME Fiber
Fiber
Attachment Rod
SPME Fiber
When performing liquid SPME, make sure that the fiber sheath does NOT
contact the liquid. This may result in carry-over.
Varian, Inc.
Page: 145
AutoSamplers
Injector Temperature
To accelerate the desorption process of the analytes from the SPME fiber,
elevated injection temperatures are required. The ideal temperature is
approximately 10-20C below the maximum temperature of the fiber and GC
column (usually between 200 and 280C).
Agitation Strokes
SPME can be used in the agitation mode, which is useful when performing
liquid SPME. Agitation disrupts the depletion zone that may surround the fiber
during the extraction process. With CP-8400 and CP-8410 SPME, agitation is
achieved by moving the fiber in and out of the sample liquid, approximately
once every 2 seconds.
To use the agitation mode, fill in the number of Agitation Strokes (1 or more).
Zero will disable the Agitation mode.
Page: 146
Varian, Inc.
AutoSamplers
Typical value for the 2 mL vial (P/N: 392611549 and P/N: 392611979)
Sample volume 1 mL
Penetration depth 60%
sampling time.
Varian, Inc.
Page: 147
AutoSamplers
Page: 148
Varian, Inc.
AutoSamplers
Varian, Inc.
Syringe switch
Page: 149
AutoSamplers
Syringe bracket
Once the syringe-guide foot touches the top of the gas chromatograph,
gently compress it further. Observe the white switch, which will be just
exposed once the sled has reached its maximum downward travel.
Push sledge down
until switch appears
Page: 150
Varian, Inc.
AutoSamplers
6. Set the switch to the STD (Standard, liquid sample configuration) DOWNPosition.
7. Raise the sled until you hear a click, this fixes the sled.
8. Install the Liquid Sample syringe holder.
9. Install the Liquid Syringe, in the block. Lock the plunger head and fix the
syringe.
10. Close the cover on the tower and turn on the power to the gas
chromatograph. The gas chromatograph will automatically detect the new
configuration of the switch and will respond accordingly.
11. Recalibrate the sampler, on page 112.
12. Adjust the Injector insert to Liquid Injection.
13. The instrument is now in the Liquid Injection mode.
Varian, Inc.
Page: 151
AutoSamplers
Page: 152
Varian, Inc.
Split.
Splitless.
The 1177 Injector can be used with a wide range of narrow bore to wide bore
(100 to 530 m ID) capillary columns.
The basic pneumatics design is a wide range inlet flow controller and back
pressure regulated column head pressure. This is the case whether manual
pneumatics or Electronic Flow Control (EFC) is used.
Varian, Inc.
Page: 153
The internal Injector temperature profile keeps the septum cool while
maintaining the point of injection at the setpoint temperature.
The 1177 uses positive septum purge to minimize the adsorption of sample
onto the Injector septum and to prevent contaminants from the septum
entering the column.
The 1177 uses a unique dual split vent that allows effective sweeping of the
entire Injector body.
Page: 154
Varian, Inc.
Split Mode
The split injection mode is preferred for the analysis of relatively concentrated
samples. The sample is split in the Injector with a representative portion
entering the column. The split injection mode provides the shortest sampling
time which leads to sharp chromatographic peaks. Use the 4.0 mm open insert,
or the packed 4.0 mm ID insert when operating in the split mode.
In the split injection mode, the sample volume is typically 2 L or less. Early
eluting compounds usually appear as very sharp peaks. In some cases, the
peak width is less than one second. It is important that you inject the sample as
quickly as possible. If you are using the CP-8400 AutoSampler use the
Standard Split/Splitless mode of injection. If you are using a non-Varian
AutoSampler or you are using the User Defined Mode of operation on the
CP-8400 AutoSamplers, if the sample injection time (the time between the
insertion and removal of the syringe needle from the Injector) exceeds the peak
width, peaks can broaden, tail or chromatographic performance will be
degraded. With broader, later eluting peaks, it is less important that you inject
the sample quickly. The split ratio (fraction of sample that enters the column) is
the ratio of Splitvent flow to the column flow.
Splitless Mode
The 4 mm Goose-neck insert packed with glass wool is typically used for
isothermal splitless injection. The small capillary section of the glass insert fits
around the syringe needle and restricts backflush of the sample vapor during
injection. In the splitless injection mode, the sample enters the column during a
variable sampling time at the beginning of the analysis. This period is typically
30-90 seconds during which there is no flow from the Injector to the split vent.
After the sampling time, the Injector is vented to remove any residual solvent
and sample out of the Injector.
Varian, Inc.
Page: 155
Page: 156
Varian, Inc.
Page: 157
Page: 158
Varian, Inc.
The EFC25 is almost identical with the EFC21 the only exception is that the
pressure is monitored at the module itself rather than at the Injector. This allows
the EFC25 to be used with Purge and Trap devices between EFC and Injector.
Varian, Inc.
Page: 159
Page: 160
Varian, Inc.
Do not heat the column oven above 50 C without carrier gas flowing
through the column. The column phase can be irreversibly damaged.
EFC Pneumatics
Type EFC21 or EFC25 used with the 1177 Injector allows the user to set a
constant column head pressure, build a pressure program, or set a constant
column flow. In addition, a split ratio can be set or time programmed. A
pressure program is typically used to maintain the column flow at a constant
value while temperature programming the column oven. When Constant Flow
Programming is enabled, the pressure program needed to maintain constant
flow is derived whenever the method is loaded.
The 1177 is a pressure-controlled Injector; thus the column flow decreases with
increasing column temperature if the pressure remains constant. EFC method
parameters and status are accessible via the FLOW/PRESSURE
the 450-GC keyboard.
Varian, Inc.
key on
Page: 161
The Type EFC21 status field displays the actual column pressure (in the units
chosen in Setup), calculated column flow rate, calculated column linear velocity
and the split ratio. The split ratio status is either off if the split state is set to
OFF, or a whole number.
The lower part allows you to view/edit constant flow mode.
The following screen is an example of a type EFC21 method.
Varian, Inc.
Pressure Pulse
Once you have selected Constant Flow programming, a checkbox Pressure
Pulse becomes accessible.
Selecting Pressure Pulse gives you the option of enabling a pressure pulse,
setting the desired pressure and the duration for which the pressure should last.
Typically, the pressure pulse pressure is held for between 30 and 90 seconds.
The purpose of the pressure pulse is twofold. The higher pressure will prevent
the solvent vapor cloud from becoming excessively large allowing larger
injection volumes and providing more efficient passage of the sample into the
column. The higher pressure also causes higher flow rates into the column
making the transfer quicker and thus preventing excessive residence times in
the Injector with consequent decomposition of labile compounds. If you use
Pressure Pulse you need to perform a septum purge calibration using the
higher pressure pulse pressure.
Varian, Inc.
When a run is started the column pressure as defined in pressure pulse will
remain for the Pulse Duration time. After the pulse duration time the
column pressure is calculated from the constant flow conditions.
Page: 163
Gas Saver
For type EFC21 and EFC25 a specific Gas Saver Flow and a timeout period
key and click on the Injector tab now choose
may be set. Press the setup
the correct Injector (Page 1 or Page 2 or Page 3).
A time entry of 0 turns the Gas Saver feature off. If a non-zero time is
entered, the Gas Saver timer will start incrementing. If a run is not started or a
new method is not loaded or download by Workstation, the set Gas Saver flow
rate will be activated (the instrument will go into stand-by). A message will be
logged in the instrument log. Stand-by is canceled by loading a method or
pressing the end stand-by key.
Page: 164
Varian, Inc.
The Calibration page will appear, choose Calibrate for the correct Injector (front,
middle or rear).
Enter the desired Column Head pressure. After a few seconds measure the
septum purge flow rate and adjust the septum purge valve (located behind the
column oven door) to the desired flow, enter this value (typical 3-5 mL/min) in
the Adjusted Septum Purge Flow Rate field.
Press OK.
Varian, Inc.
Page: 165
Manual Pneumatics
Tools and equipment needed
3. Connect the flowmeter to the split vent on the left side of the GC and
measure the split vent flow rate. Turn the Split Flow Controller valve to
adjust the split vent flow rate to 50 mL/min.
4. Adjust the Septum Purge Needle valve to adjust the septum purge flow rate
to 3 - 5 mL/min. Readjust the split vent flow rate to 50 mL/min.
5. Before heating the column, purge the system with carrier gas for 10 - 15
minutes.
With manual pneumatics the split valve should be configured in the sample
delivery/valve table in set-up of the 450-GC. Split/splitless mode can be
programmed through the valve table in the sample delivery page of the GC
method.
Page: 166
Varian, Inc.
The following is a brief description of the method parameters used for each
injection mode. In all cases Electronic Flow Control (EFC) is used for carrier gas
control.
Injector Insert
Column Installation
3.7 cm from the bottom of the column nut at the base of the Injector.
Injector Temperature
Isothermal 250 C.
Column Temperature
Set the split mode to ON for the duration of the run in the 1177 method
section and set the split ratio to 100. The sample is split upon injection
and a representative portion representing 1/100th of the amount injected
enters the column. Note that the split ratio is a method specific parameter
and should be set appropriately for individual analyses. In addition the
Injector pressure or pressure ramp should be set to achieve the desired
column flow rate.
Table 29: Split Injection Typical Conditions
Varian, Inc.
Page: 167
Splitless Injection
The classical splitless injection technique involves vaporizing the sample in a
hot Injector and slow transfer to the column. The split state is OFF during the
sampling period; therefore, all of the injected sample should enter the column.
The initial column temperature is maintained for at least the sampling time to
trap all sample components at the head of the column. At the end of the
sampling period (typically 0.5 to 1.5 minutes) the split state is turned ON to vent
any residual sample or solvent from the Injector.
The following table describes typical method parameters for an isothermal
splitless injection.
Injector Insert
Column Position
3.7 cm from the bottom of the column nut at the base of the Injector.
Injector Temperature
Column Temperature
Set the initial split mode to OFF and time program it to ON after 0.75
minutes. In this case the splitless sampling time is 0.75 minutes. The split
ratio during the split ON period should be set to 50. Set the appropriate
pressure or pressure ramp to achieve the desired column flow rate.
Table 30: Splitless Injection Method Parameters
Note that in all the above cases the parameters given are generic and need
optimization (including column position) for specific applications. Particular
care should be taken with the large volume mode of injection where the initial
Injector temperature and timing of the split states have to be carefully selected.
Page: 168
Varian, Inc.
Test Sample
Part Number
TCD
82-005048-01
ECD
82-005048-02
PFPD
82-005048-03
TSD
82-005048-04
FID
82-005048-07
Note: If the FID test sample is not available, the TCD test sample can be used if first
diluted 100:1.
Table 31: Detector Test Samples
To run one of these test samples, use the chromatographic conditions listed
previously for the injection technique you are currently using. The detector
should be operated at the most sensitive range, e.g., 12 for FID and TSD, 0.05
for TCD, 10 for PFPD and 1 for ECD. The resultant chromatogram on page 419
should approximate that shown in the detector section of this manual.
Varian, Inc.
Page: 169
1.
2.
3.
4.
6
7
8
9
10
11
12
13
14
5.
Pneumatic Type
Stainless Steel
UltiMetal
EFC21
392597301
392597303
EFC25 or Manual
392597302
392597304
Pneumatics
6. Purge head screw (2x 391866308)
7. O-ring, Liner
Graphite, 6.5mm for Splitless (392611930)
Viton, 6.3mm (8850103100)
8. Spacer (39258101)
1
9. Glass Insert (default 392611927 More liners)
10. Injector Body
Stainless Steel (392599401)
UltiMetal (392599411)
Manual (392599501)
11. Column Ferrule (see table below)
12. Capillary Column Nut (394955100)
13. Insulation
14. Cover
Teflon
Column ID
0.18 mm ID and smaller
0.25 mm ID
0.25 mm ID and smaller
0.32 mm ID
0.53 mm ID
Holes
1
1
2
1
1
2
1
Vespel
40% Graphite
60% Vespel
Max 250C
Max 350C
Max 400C
CR214103
CR212103
CR213103
CR214104
CR212104
CR213104
CR212124
CR213124
CR214105
CR212105
CR213105
CR212125
CR213125
CR214108
CR212108
CR213108
Table 32: 1177 Column Ferrules 1/16"
Graphite
SilTite
Max 450C
CR211104
CR211124
CR211105
CR211125
CR211108
Metal, GC/MS
SG073300
SG073301
SG073302
1 In combination with the PFPD detector the Siltek liner (RT210462145) must be used.
Page: 170
Varian, Inc.
Maintenance
Before maintaining the 1177 Injector please read the common
Injector/septum maintenance information in the section Maintenance,
on page 432.
The 1177 is a Split/Splitless capillary Injector. Typically, to change from Split
operation to Splitless involves changing the Injector insert. Also replace the
insert on routine base. This is especially important when dirty samples are
analyzed.
Tools Required
Varian, Inc.
Page: 171
4. Condition the insert by setting the 1177 Injector to the split mode and
purging with carrier gas for 30 min at 300C.
Page: 172
Varian, Inc.
2. Use the 5/16 open-end wrench to loosen the capillary column nut.
3. By hand, carefully withdraw the fused silica column and nut from the
Injector assembly. Set the column nut and column end on the floor of the
GC oven.
Use proper eye and skin protection. Methanol and acetone are toxic and
flammable chemicals. Exercise appropriate precautions when these chemicals
are used.
4. Remove the Injector nut, septum and top of Injector.
5. Remove the Insert.
6. Moisten a cotton swab and a stick with methanol and gently swab center of
the Injector body.
7. Moisten a cotton swab on a stick with iso-octane and gently swab the
center of the Injector body.
8. Replace the insert, seal, Injector top, septum and Injector nut.
9. Examine the end of the fused silica column to ensure that it has not been
damaged. Slide the end of the fused silica column into the opening of the
Injector.
Varian, Inc.
Page: 173
10. Gently twist the column nut in the opposite direction used to tighten the nut.
Place the column nut on the end of the Injector and tighten the nut. This
procedure minimizes twisting the column.
11. Use the 5/16 open-end wrench and tighten with 1/6-turn to secure the
column in place.
12. To check whether the Injector is leak-tight, connect a flow meter to the split
vent. Set the Relay to the split mode. If the flow meter indicates a flow less
than previously measured, the Injector has a leak.
Turn the column oven back on by checking the checkbox Power in the oven
screen
Page: 174
Varian, Inc.
Varian, Inc.
Page: 175
The Injector temperature profile keeps the septum cool while maintaining the
point of injection closer to the setpoint temperature.
The 1079 PTV Injector design facilitates large volume injection (5 - 250 L).
The 1079 PTV uses positive septum purge to minimize the adsorption of
sample onto the Injector septum and to prevent contaminants from the septum
entering the column.
The 1079 PTV combined with the ChromatoProbe on page 198 is an effective
tool for chromatographic analysis of dirty samples (liquids, solids and slurries,
blood, urine or milk), without cleanup or extraction. The ChromatoProbe is
based on intra-Injector thermal extraction of the semi volatile compounds in a
sample microvial, while non-volatile residue is retained in the microvial, which is
disposed of after the analysis.
Page: 176
Varian, Inc.
1.
4
2.
3.
5
6
4.
5.
6.
7.
Pneumatic Type
Stainless Steel
UltiMetal
EFC
392544001
392544011
Manual Pneumatics
392559601
392559611
8.
9.
8
9
Figure 24: 1079 PTV Injector Assembly
Teflon
Column ID
0.18 mm ID and smaller
0.25 mm ID
0.25 mm ID and smaller
0.32 mm ID
0.53 mm ID
Varian, Inc.
Holes
1
1
2
1
1
2
1
Vespel
40% Graphite
60% Vespel
Max 250C
Max 350C
Max 400C
CR214103
CR212103
CR213103
CR214104
CR212104
CR213104
CR212124
CR213124
CR214105
CR212105
CR213105
CR212125
CR213125
CR214108
CR212108
CR213108
Table 33: 1079 Column Ferrules 1/16"
Graphite
SilTite
Max 450C
CR211104
CR211124
CR211105
CR211125
CR211108
Metal, GC/MS
SG073300
SG073301
SG073302
Page: 177
Page: 178
Varian, Inc.
Split Mode
The split injection mode is preferred for the analysis of relatively concentrated
samples. The sample is split in the Injector with a representative portion
entering the column. The split injection mode provides the shortest sampling
time which leads to sharp chromatographic peaks. Use the 3.4 mm open insert,
the 3.4 mm ID fritted insert, or the packed 3.4 mm ID insert when operating in
the split mode.
In the split injection mode, the sample volume is typically 1 L or less. Early
eluting compounds usually appear as very sharp peaks. In some cases, the
peak width is less than one second. Thus, it is important that you inject the
sample as quickly as possible. If the sample injection time, (time between the
insertion and removal of the syringe needle from the Injector) exceeds the peak
width, peaks can broaden, tail or chromatographic performance will be
degraded. With broader, later eluting peaks, it is less important that you inject
the sample quickly. The split ratio (fraction of sample that enters the column) is
the ratio of the flow of carrier gas out the split vent to the flow through the
column. The 1177 Split/Splitless Injector on page 153 is optimized for Split
application.
On-Column Modes
In the on-column modes the column is sealed to the glass insert (0.18 - 0.53
mm columns). This ensures that there is maximum transfer of sample to the
column. In the on-column modes the sample is normally injected at or below
the solvent boiling point and then the Injector is ramped to transfer the sample
onto the column. The 1093 OC Injector, on page 231 is optimized for OnColumn applications. Especially for applications operating at an oven
temperature range that require short wire-bore columns.
Varian, Inc.
Page: 179
Page: 180
Varian, Inc.
Page: 181
The EFC25 is almost identical as the EFC21 the only exception is that the
pressure is monitored at the module itself rather than at the Injector. This allows
the EFC25 to be used with Purge and Trap devices upstream from the Injector.
Page: 182
Varian, Inc.
Varian, Inc.
Page: 183
EFC Pneumatics
Type EFC21 used with the 1079 PTV Injector allows the user to set a constant
column head pressure, build a pressure program, or set a constant column
flow. In addition, a split ratio can be set or time programmed. A pressure
program is typically used to maintain the column flow at a constant value while
temperature programming the column oven. When Constant Flow Programming
is enabled, the pressure program needed to maintain constant flow is derived
whenever the method is loaded.
The 1079 PTV is a pressure-controlled Injector; thus the column flow decreases
with increasing column temperature if the pressure remains constant. EFC
method parameters and status are accessed via the FLOW/PRESSURE
key on the 450-GC keyboard.
Page: 184
Varian, Inc.
The Type EFC21 status field displays the actual column head pressure (in the
units chosen in Setup), calculated column flow rate, calculated column linear
velocity and the split ratio. The split ratio status is either off if the split state is
set to OFF, or a whole number.
The lower part allows you to view/edit constant flow mode.
The following screen is an example of a type EFC21 method.
Page: 185
Pressure Pulse
Once you have selected Constant Flow programming, a checkbox Pressure
Pulse becomes accessible.
Selecting Pressure Pulse gives you the option of enabling a pressure pulse,
setting the desired pressure and the duration for which the pressure should last.
Typically, the pressure pulse pressure is held for between 30 and 90 seconds.
The purpose of the pressure pulse is two fold. The higher pressure will prevent
the solvent vapor cloud from becoming excessively large allowing larger
injection volumes and providing more efficient passage of the sample into the
column. The higher pressure also causes higher flow rates into the column
making the transfer quicker and thus preventing excessive residence times in
the Injector with consequent decomposition of labile compounds. If you use
Pressure Pulse you will need to perform septum purge calibration using
the higher pressure pulse pressure.
Page: 186
When a run is started the column pressure as defined in pressure pulse will
remain for the Pulse Duration time. After the pulse duration time the
column pressure is calculated from the constant flow conditions.
Varian, Inc.
Gas Saver
For type EFC21 a specific Gas Saver Flow and a timeout period may be set.
A time entry of 0 turns the Gas Saver feature off. If a non-zero time is
entered, the Gas Saver timer will start incrementing. If a run is not started or a
new method is not loaded or download by Workstation, the set Gas Saver flow
rate will be activated (the instrument will go into stand-by). A message will be
logged in the instrument log. Stand-by is canceled by loading a method or
pressing the end stand-by key.
Varian, Inc.
Page: 187
key and
The Calibration page will appear, choose Calibrate for the correct Injector (front,
middle or rear).
Enter the desired Column Head pressure. After a few seconds measure the
septum purge flow rate and adjust the septum purge valve (located behind the
column oven door) to the desired flow, enter this value (typical 3-5 mL/min) in
the Adjusted Septum Purge Flow Rate field.
Press OK.
Page: 188
Varian, Inc.
Manual Pneumatics
Tools and equipment needed
3. Connect the flowmeter to the split vent on the left side of the GC and
measure the split vent flow rate. Turn the Split Flow Controller valve to
adjust the split vent flow rate to 50 mL/min.
4. Adjust the Septum Purge Needle valve to adjust the septum purge flow rate
to 3-5 mL/min. Readjust the split vent flow rate to 50 mL/min.
5. Before heating the column, purge the system with carrier gas for 10 - 15
minutes.
Varian, Inc.
Page: 189
The Injector insert used and the position of the column within the insert.
The following is a brief description of the method parameters used for each
injection mode. In all cases Electronic Flow Control is used for carrier gas
control.
In many cases switching from one 1079 PTV mode to another involves
changing the glass insert. A detailed stepwise procedure to carry out
this task is given in the Maintenance section of this manual, on page
195. In most injection modes the insert is installed with a graphite
ferrule. Use the special fixture supplied with the 1079 PTV accessory kit
to install the ferrule correctly on the insert.
Isothermal Split Injection
The split mode is used when samples are relatively concentrated and for neat
samples. This mode of injection involves rapid vaporization of the sample
followed by sample splitting. Splitting involves directing a portion of the sample
into the column while the remainder is vented. The split ratio is defined as the
proportion of sample vented to the sample entering the column. With EFC
carrier gas control this parameter can be set automatically in the 1079 PTV
method.
Injector Insert
Install the fritted split insert (03-926119-46) or unpacked split insert (03926119-45). See instructions in the Maintenance section on page 195 for
changing the 1079 PTV insert.
Column Installation
7.5 cm from the bottom of the column nut at the base of the Injector
Injector Temperature
Isothermal 250 C.
Column Temperature
Set the split mode to ON for the duration of the run in the 1079 PTV method
section and set the split ratio to 100. The sample will be split upon injection
and a representative portion representing 1/100th of the amount injected will
enter the column. Note that the split ratio is a method specific parameter and
should be set appropriately for individual analyses. In addition the Injector
pressure or pressure ramp should be set to achieve the desired column flow
rate.
Table 34: Split Injection Typical Conditions
Page: 190
Varian, Inc.
Column Position
7.5 cm from the bottom of the column nut at the base of the Injector.
Injector Temperature
Column Temperature
Set the initial split mode to OFF and time program it to ON after 0.75
minutes. In this case the splitless sampling time is 0.75 minutes. The split
ratio during the split ON period should be set to 50. Set the appropriate
pressure or pressure ramp to achieve the desired column flow rate.
Table 35: Isothermal Splitless Injection Method Parameters
Varian, Inc.
Page: 191
Injector Insert
Column Position
Injector Temperature
Column Temperature
Carrier gas control
On-column Injection
The wide-bore on-column mode uses a wider bore insert with a taper to allow
insertion of a 0.53 mm ID capillary column inside the insert. A standard gauge
GC syringe (26s) is used to deposit sample inside the column.
In case narrow bore columns are required, a retention gap is used as noncoated 0.53 mm ID pre-column.
This enables true on-column injection in a narrow bore column while using
standard syringes.
The Injector is temperature ramped to minimize sample decomposition due to
thermal effects or active sites.
The on-column technique involves some special setup of the 1079 PTV
Injector.
In the on-column mode the column has to be sealed within the insert. The
Injector nut, septum, septum support and insert are removed from the Injector.
The column is pushed up through the Injector until it protrudes past the top of
the Injector. The tapered on-column insert is then pushed onto the column to
make a seal between the polyimide coating on the column and the glass
surface. The insert is then lowered into the Injector and the septum support
installed without a graphite ferrule. The septum support must not be screwed
all the way down or the septum will not seal.
Page: 192
Varian, Inc.
The following are typical conditions for carrying out an on-column injection. The
Injector should be maintained at 10 - 20 C below the solvent boiling point at
injection. In this example the solvent is hexane which has a boiling point of
68 C.
Injector Insert
Install the on-column insert (P/N 01-900109-07) for wide bore (0.53 mm
ID) columns.
Seal/Position Column
Injector Temperature
Column Temperature
Split does not apply in on-column mode. The Splitless vent flow must be
set to 20 mL/min or greater in Setup. Delete any splitter program and
set initial state to off in the Injector screen.
Table 37: On-Column Injection
Varian, Inc.
Page: 193
Injector Insert
Column Position
7.5 cm from the bottom of the column nut at the base of the Injector.
Injector Temperature
Column Temperature
Set the initial split mode to ON, time program it to OFF after 1.00 minutes
and then back to ON after 3.00 minutes. The split ratio should be set to 50
during the split ON periods.
Table 38: Large Volume Injection
Note that in all the above cases the parameters given are generic and will have
to be optimized for specific applications. Particular care should be taken with
the large volume mode of injection where the initial Injector temperature and
timing of the split states have to be carefully selected.
Part Number
TCD
82-005048-01
ECD
82-005048-02
PFPD
82-005048-03
TSD
82-005048-04
FID
82-005048-07
Note: If the FID test sample is not available, the TCD test sample can be used if first
diluted 100:1.
Table 39: 1079 PTV Test Samples
To run one of these test samples, use the chromatographic conditions listed previously
for the injection technique you are currently using. The detector should be operated at
the most sensitive range, e.g., 12 for FID and TSD, 0.05 for TCD, 10 for PFPD and 1 for
ECD. The resultant chromatogram should approximate that shown in the detector
section of this manual. Some chromatographic interpretation information is given in the
troubleshooting section of this manual.
Page: 194
Varian, Inc.
Maintenance
The 1079 PTV Injector that can be operated in several modes. These modes
include split, splitless, on-column and large volume injection. Typically, to
change from one mode of operation to another involves changing the Injector
insert. The insert should be replaced on a routine basis. This is especially
important when dirty samples are analyzed.
After prolonged use, the 1079 PTV Injector glass insert may need to be
replaced with a new insert.
Tools Required
Varian, Inc.
Page: 195
Replace the Injector septum each time the glass insert is replaced.
Page: 196
Varian, Inc.
Septum Support
Insert Ferrule
Glass Insert
Insert/ferrule
positioning tool (392538500)
2. Position the tool as shown on a flat, clean surface. Use clean laboratory
tissue on the surface. Tighten the septum support finger-tight. Holding the
tool with a 5/8" wrench, give the septum support an extra 1/3 to 1/2 turn
past finger-tight. Now unscrew the septum support which now has the
ferrule and insert seated in it. If there is any graphite extruded past bottom
of the septum support, cut it off with a blade or sharp knife. Carefully wipe
off any graphite flakes which may adhere to the insert or septum support.
Gripping the septum support unit with a piece of laboratory tissue, carefully
put this unit in the 1079 PTV Injector and tighten the septum support 1/6turn past finger-tight.
3. Use tweezers to place a new septum over the septum support.
If the septum has a Teflon face, place the Teflon face toward (down) the
column.
4. Place the Injector nut on the Injector and tighten by hand until you feel
some resistance, then tighten an extra 1/4 turn using the Injector nut
wrench.
5. Condition the insert by setting the 1079 PTV Injector to the split mode and
purging with carrier gas for 30 min at 300 C.
Varian, Inc.
Page: 197
ChromatoProbe
How to Assemble the ChromatoProbe
Refer to the exploded view on the next page.
The ChromatoProbe assembly consists of the adapter, guide, ChromatoProbe,
ChromatoProbe cap and the small and large O-rings. One small O-ring is
placed on the ChromatoProbe shaft; the other is placed on the ChromatoProbe
cap shaft. The large O-ring is placed in the groove in the bottom of the adapter
which connects to the 1079 Injector body.
The storage stand may be used to hold and protect the ChromatoProbe or
ChromatoProbe cap. The user may find the small holes in the top of the storage
stand useful for holding microvials during filling or just prior to running an
analysis.
Page: 198
Varian, Inc.
Once the ferrule is positioned and tightened, the guide and insert can be placed
into the 1079 Injector body. Screw the guide into the Injector body with the key
and tighten snugly.
At this point the ChromatoProbe is ready to be attached onto the Injector. The
adapter is attached to the Injector such that the portion containing the large Oring is screwed onto the Injector. Tighten this fingertight to prevent any leaks.
DO NOT use a tool to tighten this connection. Insert the ChromatoProbe cap
into the top of the adapter and hand tighten to cap off the Injector when not
using the ChromatoProbe.
1
2
3
4
ChromatoProbe Kit (03-925677-91)
5
6
8
10
11
9
1.
2.
3.
4.
5.
6.
7.
8.
9.
10.
11.
Varian, Inc.
Page: 199
Varian, Inc.
COLUMN OVEN:
Varian, Inc.
Page: 201
Page: 202
Varian, Inc.
6. Solvent vaporization takes about one minute for each 3-4 microliter sample.
The split vent can be opened during that time, with a split flow rate above
20 mL/min to speed up the solvent vaporization.
7. After approximately one minute solvent vaporization time, the Injector
temperature should be fast programmed to 250C (for pesticides) or the
appropriate temperature required for the thermal extraction of the
semivolatile compounds. The final temperature chosen is a compromise
between higher temperature for more effective and faster thermal extraction
and a lower temperature value that might be required to prevent thermal
degradation of delicate compounds.
8. During the Injector heatup and thermal desorption time, the split valve must
be closed and the carrier gas flow rate must be at least 4-5 mL/min for
effective thermal extraction. At lower carrier gas flow rates a higher Injector
temperature will be required.
9. The thermal desorption stage takes between 0.5-2 minutes. After that time,
the Injector is cooled down, the carrier gas flow rate is programmed to the
chromatographic optimal value and the GC oven program can begin in the
usual way. Note that the final GC oven upper temperature and time can be
significantly reduced since the less volatile compounds are retained in the
microvial.
10. At the end of the analysis, dispose of the sample microvial. Do not re-use
microvials. The microvial holder should only be removed from the adapter
after both the GC oven and Injector are cool and a helium purge flow from
the ChromatoProbe protects the column.
11. In addition to passivation of the microvial holder, analysis of thermally labile
sample compounds is facilitated with a shorter microvial (up to 6 mm long).
The standard 15 mm long microvials can be cut (like a column) to the
desirable length. Handle the microvial carefully while cutting to avoid
contamination.
12. The microvial volume is 30 microliters. Up to 20 microliters can be loaded
and the solvent evaporated outside the GC prior to sample introduction.
13. Test your method for thermal vaporization efficiency, reproducibility and
long term stability before beginning routine analysis.
Varian, Inc.
Page: 203
Cleaning
Should the ChromatoProbe become contaminated with time, it can be cleaned
by placing it in a solvent such as acetone and placed in a sonic bath. If it is
severely contaminated a mild abrasive can be used to clean it.
Recommended Reading
1. Aviv Amirav and Shai Dagan, A Direct Sample Introduction Device for Mass
Spectrometry Studies and GC/MS Analysis. European Mass Spectrometry,
3, 105-111 (1997).
2. Hongwu Jing and Aviv Amirav, Pesticide Analysis with the Pulsed Flame
Photometric Detector and a direct sample Introduction Device. Analytical
Chemistry, 69, 1426-1435 (1997).
3. Samuel B. Wainhaus, Shai Dagan, Mark L. Miller and Aviv Amirav, Fast
Drug Analysis in a Single Hair. Submitted for publication.
Page: 204
Varian, Inc.
The 1041 On-Column Injector is designed for on-column sample injection onto 530
micron I.D. fused silica columns, 1/4 inch and 1/8 inch O.D. packed columns (glass or
metal).
On-column injection provides complete sample transfer from Injector to column results
in good quantitation.
The fused silica column or glass column extends all the way from the Injector septum to
the detector, providing a clean inert system without cross-overs.
Quick and easy to switch from megabore to packed column or vise versa by changing
an adaptor. The 1041 is supplied with capillary mounting hardware as standard. If you
wish to operate the Injector in the packed column mode then a packed column adapter
kit must be ordered (03-925588-91). This kit contains Injector and detector hardware to
facilitate use of the Injector with 1/8" metal packed columns. In addition, adapter kits are
available for 1/4" columns. Full instructions on converting the 1041 to packed column
use are included with the packed column kit.
The 1041 On-Column Injector is available in manual pneumatics and electronic flow
control (EFC) for different laboratory requirement.
Choose from EFC23 (constant flow control) and EFC24 (constant pressure control and
constant total flow control) for different application requirement.
Varian, Inc.
Page: 205
Page: 206
Varian, Inc.
3
1.
2.
3.
4
4.
5
6
5.
6.
7.
8.
9.
8
9
Figure 28: 1041 On-Column Injector with Insert
Teflon
Vespel
Column ID
Max 250C
0.53 mm ID
CR214108
Max 350C
CR212108
40% Graphite
60% Vespel
Max 400C
CR213108
Graphite
Max
450C
CR211108
SilTite
Metal, GC/MS
SG073302
Varian, Inc.
Page: 207
Page: 208
Varian, Inc.
Varian, Inc.
Page: 209
The type EFC24 flow diagram shown below is an indication of the control
mechanism of this type of EFC module.
Page: 210
Varian, Inc.
Column Installation
The Connect Capillary Column to Injector on page 15 procedure describes the
installation (or reinstallation) of a capillary column in a Varian 450-GC equipped
with a 1041 Injector. Each step of the procedure is described in detail.
Condition the Column
For conditioning the capillary column see Column Installation and Conditioning
on page 23.
Column Installation in Detector
For connecting a capillary column into a detector, see Connect Capillary
Column to Detector on page 24.
EFC Pneumatics
Type EFC24 used with the 1041 or 1061 Injectors allows the user to set a
constant column head pressure, build a pressure program, or set a constant
column flow. With EFC23 equipped Injectors a flow range of 0 100 mL/min
may be set from the 450-GC keyboard. A pressure program is typically used to
maintain the column flow at a constant value while temperature programming
the column oven. When Constant Flow Programming is enabled, the pressure
program needed to maintain constant flow is derived whenever the method is
loaded. The only user settable parameter with the 1041 type of EFC is the
desired flow rate.
The 1041 is a pressure-controlled Injector; thus the column flow decreases with
increasing column temperature if the pressure remains constant. EFC method
parameters and status are accessed via the FLOW/PRESSURE
the 450-GC keyboard.
key on
Page: 211
Carrier Gas
Nitrogen
1.5 mL/min
3 - 15 mL/min
Helium
2.4-4.0 mL/min
4 - 15 mL/min
4.0-8.0 mL/min
8 - 20 mL/min
Hydrogen
Packed Mode
When operating in the packed column mode, adjust the carrier gas flow to
applicable packed column flow rates (20 to 50 mL/min), depending on carrier
gas type.
Page: 212
Varian, Inc.
1041 Operation
This screen contains the specific operating information and procedures required
for optimum performance of the 1041 on-column Injector. All installations for
both the instrument and the Injector must be completed before continuing
further.
Varian, Inc.
Page: 213
250 C
Column temperature
Injection Volume
1 L
Observe the eluting peaks for symmetry, separation and elution time.
Abnormally wide or skewed peaks, excessive elution times, abnormally small
peaks and noisy or drifting baselines indicate faulty performance.
Page: 214
Varian, Inc.
Maintenance
Hardware Replacement and Cleaning Procedures
Routine cleaning and maintenance of the 1041 On-Column Injector system
includes septum replacement and column conditioning. Pressure testing and
leak checking procedures are included to assure proper functioning of the
system.
Septum Replacement
Septum replacement represents the major part of routine chromatographic
maintenance. Septum damage from the needle penetrations can be avoided by
injecting into the same hole and not using syringes with needles having burrs or
bends at the tip which cut the septum.
1. Be sure the Injector nut is cool. Unscrew the Injector nut (use the Injector
Nut wrench) and place on a clean, uncontaminated surface.
2. Using tweezers, remove the septum, taking care not to scratch the internal
surfaces of the Injector.
3. Using tweezers, place a new high temperature septum in the Injector. If the
septum is TFE or FEP coated put that side down.
4. Replace the Injector nut and tighten until resistance is felt, then tighten an
extra 1/2 to full turn.
Varian, Inc.
Page: 215
Leak Checking
Leak checking methods used with the 1041 On-Column Injector are presented
in order of overall sensitivity and performance. When leak checking the Injector
only, be sure to use the appropriate procedure.
1. Remove the test column and plug the Injector outlet with a 1/16" no-hole
ferrule (P/N 28-694590-01). Remove the Injector nut, install a new septum,
then replace Injector nut and turn nut clockwise until it comes to a stop.
2. Pressurize the Injector to 400 kPa (60 psig) with carrier gas. Turn off the
carrier gas flow at the supply.
3. A pressure drop less than or equal to 3 kPa (0.5 psig) in 15 minutes is
acceptable. Locate leaks with an appropriate leak detector.
Any changes in temperature while performing this test may result in
false readings due to expansion/contraction of the gas with
temperature.
Commercial soap type leak detection fluids should not be used at any
point in a capillary system, since, if a leak is present, the fluid will
penetrate and contaminate the system. Column performance will be
degraded and a substantial period of time may be required to achieve a
clean system.
4. An alternate way to leak check the Injector only is to have the column
installed in the Injector end and the opposite column end sealed with a
flame or appropriate fitting.
Page: 216
Varian, Inc.
Commercial soap type leak detection fluids should not be used at any
point in a capillary system, since, if a leak is present, the fluid will
penetrate and contaminate the system. Column performance will be
degraded and a substantial period of time may be required to achieve a
clean system.
Varian, Inc.
Page: 217
Page: 218
Varian, Inc.
The 1061 flash vaporization Injector is designed for use with wide-bore columns
of 0.53 mm ID.
Used with packed glass liner with vaporization volume for dirty samples.
Glass liner is removable from the top. No need to disconnect the column.
The 1061 Injector can operate in 2 modes: the capillary column mode or the packed
column mode.
Switching from one mode to another involves changing the Injector and detector column
mounting hardware.
The 1061 is supplied with capillary mounting hardware as standard.
If you wish to operate the Injector in the packed column mode then a packed column
adapter kit must be ordered (03-925588-92). This kit contains Injector and detector
hardware to facilitate use of the Injector with 1/8 metal packed columns.
Complete instructions are provided in this kit for installing and operating the Injector in
the packed column mode.
Varian, Inc.
Page: 219
Page: 220
Varian, Inc.
1
2
3
4
5
6
7
8
9
10
6
7
8
9
10
Figure 31: 1061 Flash Vaporization Injector with Insert
Teflon
Vespel
Max 250C
Max 350C
Column ID
0.53 mm ID
CR214108
CR212108
40% Graphite
60% Vespel
Max 400C
CR213108
Graphite
Max
450C
CR211108
SilTite
Metal, GC/MS
SG073302
Varian, Inc.
Page: 221
Varian, Inc.
The type EFC24 flow diagram shown below is an indication of the control
mechanism of this type of EFC module.
Varian, Inc.
Page: 223
Column Installation
The following instructions apply to installing a 0.53 mm ID capillary column in
the 1061 Injector. Note that the thin polymeric coating on fused silica columns
will give some protection against breakage; however, fused silica columns are
somewhat fragile and must be handled with care. Alternative is to use UltiMetal
0.53 mm wide-bore columns.
Column Installation
The Connect Capillary Column to Injector on page 15 procedure describes the
installation of a capillary column in a Varian 450-GC equipped with a 1061
Flash Injector. Each step of the procedure is described in detail.
Condition the Column
For conditioning the capillary column see Column Installation and Conditioning
on page 23.
Column Installation in Detector
For connecting a capillary column into a detector, see Connect Capillary
Column to Detector on page 24.
Page: 224
Varian, Inc.
EFC Pneumatics
Type EFC24 used with the 1061 Injector allows the user to set a constant
column head pressure, build a pressure program, or emulate a constant column
flow when the column is temperature programmed. With EFC23 equipped
Injectors a flow range of 0 100 mL/min may be set from the 450-GC
keyboard. A pressure program is typically used to maintain the column flow at a
constant value while temperature programming the column oven. When
Constant Flow Programming is enabled, the pressure program needed to
maintain constant flow is derived whenever the method is loaded.
The 1061 is a pressure-controlled Injector; thus the column flow decreases with
increasing column temperature if the pressure remains constant. EFC method
parameters and status are accessed via the FLOW/PRESSURE
the 450-GC keyboard.
key on
Varian, Inc.
Page: 225
Nitrogen
1.5 mL/min
3 - 15 mL/min
Helium
2.4-4.0 mL/min
4 - 15 mL/min
4.0-8.0 mL/min
8 - 20 mL/min
Hydrogen
Table 44: Carrier Gas Flow Rates for 1061 Flash Injector
Page: 226
Varian, Inc.
Varian, Inc.
Page: 227
Injector temperature
250 C
Column temperature
Injection Volume
1 L
Observe the eluting peaks for symmetry, separation and elution time.
Abnormally wide or skewed peaks, excessive elution times, abnormally small
peaks and noisy or drifting baselines indicate faulty performance.
Page: 228
Varian, Inc.
MAINTENANCE
The 1061 is mechanically similar to the 1041 except that it uses a glass insert
for flash vaporization injection. Follow the same maintenance procedures for
the 1061 that are outlined above for the 1041, except in the case of changing or
cleaning glass inserts.
Varian, Inc.
Page: 229
3. Place the Injector nut on the Injector and tighten by hand until you feel
some resistance, then tighten an extra 1/2 to full turn using the Injector nut
wrench.
After the Injector nut has been replaced check the head pressure (EFC
status on display or manual pressure gauge on pneumatics panel)
increases to its normal value. If this does not happen there is an
indication of a leak.
4. Condition the insert by setting the 1061 Injector temperature to 300C and
allowing the system to condition for 30 minutes.
Page: 230
Varian, Inc.
The Varian 1093 Injector gives optimum performance for non-vaporizing (cold)
injection into conventional fused silica capillary columns (0.32-0.53 mm ID
columns).
It incorporates several unique design features to produce improved peak
resolution, quantitation and versatility over other cold on-column Injectors.
The result is not only a capillary on-column injection system ideally suited for
analyses that cannot be done with conventional split/splitless injection, but one
that gives highly accurate and precise results with a wide variety of trace to
semi-trace level samples.
The 1093 Injector can replace manually operated vaporizing capillary Injectors
in many cases and may provide improved quantitation of labile and high
molecular weight samples.
Temperature-programmed, cold injection: The Injector is held at a temperature
20 C to 30 C below the solvent boiling point during injection and then is
rapidly temperature programmed up to the final column temperature.
The oven is started at the same temperature as the Injector. Conventional
temperature programmed analysis is carried out in the GC column.
Varian, Inc.
Page: 231
Frequently, good results may be obtained and the run time may be reduced by
injecting into a cold Injector but with the column at 5C above the solvent boiling
point. In this case, the column is held at the initial temperature until the Injector
is at its final temperature.
Cold injection reduces or eliminates many of the problems of conventional
vaporizing split/splitless injection, such as molecular weight discrimination (loss
of high and low boiling substances), thermal decomposition of the sample and
sample adsorption onto hot, active Injector surfaces. These problems are
largely eliminated in cold injection because the sample is directly introduced
into the column as a liquid. In the split/splitless Injector, the sample is first
vaporized and then allowed to pass into the column, where it may be
condensed back into its liquid state.
The 1093 Injector may also be used as an isothermal, vaporizing Injector.
However, best performance is achieved using the temperature-programmed
mode. The system consists of three major components: a glass insert, the
heated Injector body and a standard syringe. Maximum syringe needle
diameter is 0.019" (0.49 mm) or 26 gauge.
All heated parts and all parts in direct contact with the carrier gas stream are
constructed of stainless steel, borosilicate glass, fused silica, silicone rubber
(septum) and Inconel X750.
The Injector body is mounted on an insulated enclosure. The Injector heating
rate is adjustable to allow optimization for thermally sensitive samples. A purge
exit is provided for bake-out and septum purge. The internal dead volume of the
Injector body has been minimized.
In operation, the fused silica column is inserted up into the Injector body from
the column oven and seated in the glass insert. The column is sealed to the
Injector with conventional polyimide/graphite or graphite ferrules. The syringe
needle enters the glass insert from the top. After the syringe is withdrawn, the
entire injection zone is rapidly heated, driving the sample into the column where
the peaks are sharpened by the solvent effect and cold trapping. Significant
peak broadening occurs (up to 2X) in Injectors without this rapid heating step,
especially when larger samples and high-boiling samples are used.
Of the 2 internal parts in the Injector body, the septum support and glass insert
require periodic removal for cleaning. A biweekly check is recommended during
regular use. The need for cleaning will vary with different samples and the glass
insert is easily replaced in case of non-removable deposits or breakage.
The 1093 Septum-Equipped Programmable Injector (SPI) is factory-installed or
available as a field upgrade (must be installed by trained personnel).
Page: 232
Varian, Inc.
Varian, Inc.
Page: 233
3
4
5
6
7
8
10
11
12
Page: 234
Varian, Inc.
Page: 235
The type EFC24 flow diagram shown below is an indication of the control
mechanism of this type of EFC module.
Page: 236
Varian, Inc.
Glass Insert
Different glass inserts are available for use in the 1093 Injector.
Click here to see all inserts of the 1093 Injector parts on page 454
The standard insert is the high performance (P/N 190010906).
Syringe
The required syringe needle length is 2.00" (excluding the hub) for both manual
injections and used in the CP-8400 and 8410 AutoSamplers.
Varian, Inc.
Page: 237
Column Installation
The Connect Capillary Column to Injector on page 15 procedure describes the
installation (or reinstallation) of a capillary column in a Varian 450-GC equipped
with a Varian 1093 Injector. Each step of the procedure is described in detail.
Teflon
washer
CGA320
Fitting
In-Line Filter
Hex Head
Reducer
Swagelok
Fitting
To CO2 Solenoid
Valve
Page: 238
Varian, Inc.
Approximately 100 chromatographic runs can be made per tank of CO2 when
the Injector is cooled to -50C for each run. Cooling to higher temperatures will
consume less coolant per run. The rate of use of CO2 is dependent on the
operational conditions, thus your actual use may vary from this estimate.
1. Use 1/8" copper or stainless steel tubing for the connection between the
CO2 solenoid valve and the CO2 tank. NOTE: If stainless steel tubing is
used, change all fittings to stainless steel.
2. Connect one end of the 1/8" Swagelok fitting to the GC and the other end of
the tubing to the in-line filter and CGA 320 adapter. All hardware is supplied
in the Start-Up Kit.
3. The CGA 320 fitting will be connected to the CO2 tank. A Teflon washer is
fitted in the nut that will connect to the tank and it must be inspected. Check
that the washer is in good condition. If it is not, replace that washer and
then connect the CGA 320 fitting to the tank. NOTE: Do not over tighten
the CGA 320 fitting. A high torque is not required, as the Teflon
washer is the seal.
Page: 239
Do not heat the column oven above 50 C without carrier gas flowing
through the column. The column phase can be irreversibly damaged.
Page: 240
Varian, Inc.
EFC Pneumatics
The Electronic Flow Control module used on a 1093 Injector can be type
EFC23 or EFC24.
The EFC23 is used to control the 1093 Injector under flow control to program
the column flow. EFC23 sets a required carrier flow into the Injector/Column
system. A leak in the system would be indicated to the user by a drop in inlet
pressure.
The EFC24 is used to control the 1093 Injector in combination with Headspace
sampler and other devices that have a long carrier gas line.
Both EFC types are designed specifically for the 1093 Injector to support its
various modes of operation. In simple terms it duplicates the behavior of the
1093 manual pneumatics system in that there is an inlet mass flow controller
supplying carrier gas to the Injector and a pressure control valve downstream
from the Injector which sets the Injector pressure. As Injector pressure
determines the rate of carrier gas flow through the column, this pressure is
monitored close to the point of injection.
Type EFC23 or EFC24 used with the 1093 Injector allows the user to set a
constant column head pressure, build a pressure program, or set a constant
column flow. A pressure program is typically used to maintain the column flow
at a constant value while temperature programming the column oven. When
Constant Flow Programming is enabled, the pressure program needed to
maintain constant flow is derived whenever the method is loaded.
The 1093 is a pressure-controlled Injector; thus the column flow decreases with
increasing column temperature if the pressure remains constant. EFC method
parameters and status are accessible via the FLOW/PRESSURE
the 450-GC keyboard.
key on
Varian, Inc.
Page: 241
Page: 242
Varian, Inc.
Test Sample
Part Number
TCD
82-005048-01
ECD
82-005048-02
PFPD
82-005048-03
TSD
82-005048-04
FID
82-005048-07
Note: If the FID test sample is not available, the TCD test sample can be used if first
diluted 100:1.
Table 46: Detector Test Samples
To run one of these test samples, use the chromatographic conditions listed
previously for the injection technique you are currently using. The detector
should be operated at the most sensitive range, e.g., 12 for FID and TSD, 0.05
for TCD, 10 for PFPD and 1 for ECD. The resultant chromatogram on page 419
should approximate that shown in the detector section of this manual.
Observe the eluting peaks for symmetry, separation and retention time.
Abnormally wide or skewed peaks, excessive retention times, abnormally small
peaks and noisy or drifting baselines indicate faulty performance.
When the instrument is installed a definitive indication of instrument
performance can be obtained by running a test chromatogram for the
detector(s) that will be used with the Injector. This initial test chromatogram
should be retained as a standard of comparison for later checking the
instrument if change in detector sensitivity is suspected.
Varian, Inc.
Page: 243
key and
The Calibration page will appear, choose Calibrate for the correct Injector (front,
middle or rear).
Enter the desired Column Head pressure. After a few seconds measure the
septum purge flow rate and adjust the septum purge valve (located behind the
column oven door) to the desired flow, enter this value (typical 3-5 mL/min) in
the Adjusted Septum Purge Flow Rate field.
Page: 244
Varian, Inc.
Maintenance
Before maintaining the 1093 Injector please read the common
Injector/septum maintenance information in the section Maintenance,
on page 432.
Septum Replacement
Septum replacement represents the major part of routine chromatographic
maintenance. Septum damage from the needle penetrations can be avoided by
injecting into the same hole and not using syringes with needles having burrs or
bends at the tip which cut the septum.
Handling a septum with bare fingers may result in column
contamination. Use tweezers when installing a new septum.
Follow steps 1 through 7 below for replacing the septum in the Injector. Be sure
to cool all heated zones to less than 80C before opening the septum nut and
decrease the column head pressure to less than 2 psig. If pressure is too high,
the septum support and glass insert may pop up, rather than remain in their
correct positions.
Removal of Glass Insert
Use care when removing or replacing the glass inserts for the 1093 Injector.
The Injector nut may be hot. Lower the Injector temperature
to 50 C and permit the Injector nut to cool before
proceeding.
Removal
To avoid contaminating the capillary system, ALWAYS use
tweezers or the extraction tool to handle the Injector
internal parts.
1. Use the Injector nut wrench to remove the Injector
nut and place on a clean, uncontaminated surface,
such as a fresh cleaning tissue.
Varian, Inc.
Page: 245
Page: 246
Varian, Inc.
Leak Checking
Leak checking methods used to leak check the 1093 Injector are
presented in order of overall sensitivity and preference.
Leak Checking the Injector Only
1. Remove the (test) column and plug the Injector outlet with a 1/16" no-hole
ferrule (P/N: 28-649503-00). Turn Injector nut clockwise until it comes to a
stop. Turn off the septum purge.
2. Pressurize the Injector to 30 psig with carrier gas. Turn off the carrier gas
flow at the ON-OFF valve.
3. The pressure should hold for 30 minutes, as shown on the status screen. A
pressure drop less than or equal to 1 psig per minute at 30 psig is
acceptable. Locate leaks with an appropriate leak detector.
Commercial soap type leak detection fluids should not be used at any
point in a capillary system, since, if a leak is present, the fluid will
penetrate and contaminate the system. Column performance will be
degraded and a substantial period of time may be required to achieve a
clean system.
4.
An alternate way to leak check the Injector only is to have the column
installed in the Injector end and the opposite column end sealed.
Butane
Attenuation,
Range
16 x10-12
Maximum
Peak Size
30%
Dichloromethane
2x1
50%
Detector
Gas
FID
ECD
PFPD, Sulfur Mode
TSD
2 x 10
-10
50%
2 x 10
-12
2%
Varian, Inc.
Page: 247
Page: 248
Varian, Inc.
4. Set flow between 100 and 200 mL/min. Measure flow at the purge line exit,
then remove the no-hole ferrule from the column nut.
5. Set the Injector temperature to 450C.
6. Bake out the Injector for not less than 4 hours, to a maximum of 16 hours at
a time.
7. Cool the Injector to 50C and then install the capillary column, on page 15.
8. Replace the septum.
9. Readjust the septum purge to the desired flow rate.
Varian, Inc.
Page: 249
Page: 250
Varian, Inc.
Description
The SPT is a concentrating system for the determination of trace level volatile
organic components in air and gaseous matrices. Integrated into the top of the
GC, its design provides the following features:
Trap heating rates of up to 40 C/sec provide fast injection from the trap to the
capillary column. This helps to ensure good chromatographic performance
without the need of an isolation valve, even for low boiling analytes.
Fast injection also allows higher initial column oven temperatures, thereby
reducing or eliminating the need for column oven cryogenics. Cryogenic usage
for trap cooling is minimized because of the compact design of the SPT and the
specially designed algorithm that controls the cooldown and temperature of the
SPT.
Varian, Inc.
Page: 251
Each SPT option includes a SPT valve oven, installed either for operation with
2 gas switching valves or with one valve and one Injector. The SPT valve oven
is controlled by the 450-GC heated zone and provides a constant temperature
environment for the trap ends and tubing connecting them to an SPT valve.
The SPT is assigned to heated zone 3 only.
The multi-valve oven may be installed on the 450-GC for SPT systems which
require more than 2 valves.
See building an SPT method on page 263 for more details.
Page: 252
Varian, Inc.
The short trap is generally used for cryogenic trapping on glass beads
using LN2 or for highly adsorbing trap materials.
The long trap length is suitable for trapping with adsorbents such as
Tenax or charcoal-based materials at ambient or near ambient trapping
temperatures where LCO2 cryogen may be used.
Total Length
17.02 cm
39.47 cm
Varian, Inc.
Page: 253
Trap Dimensions
Once the trapping material is chosen, the trap size (long or short) should be
selected based on the volume of the sample to be passed through it and the
trapping temperature (all SPT traps are made from 1/8 o.d., type 316 stainless
steel tubing). Most traps should be able to accommodate samples of
approximately 400 milliliters at flow rates of about 40 mL/min in order to obtain
the detection levels required.
Trap Temperature
The temperatures required for cryogenic trapping on glass beads are easy to
determine. As noted above, a trap temperature of -160C has been used for air
toxics. When preconcentrating even lower boiling analytes, such as the
C2-hydrocarbons, trap temperatures as low as -180C are necessary. When
using adsorbents, the trapping temperature is determined by the breakthrough
volume of the sample. Lower activity adsorbents require lower temperatures.
Trapping and Desorption
The flow rate during trapping is generally from 10 to 40 mL/min. Trapping is
always more efficient at lower flow rates. Flow rate during desorption depends
on the analytical system to which the analytes are passed. Desorption flow
rates of 5 mL/min or greater are compatible with 0.53 mm i.d. columns and
usually produce narrow peaks for early eluting components. When the capillary
column i.d. does not allow this high flow, fast desorption flow rates in
combination with inlet splitters may be used.
To improve peak shape and quantitation in cases where narrow bore column
systems are in use but detectivity requirements prohibit the use of split inlet
systems, column oven cryogenics will be required. This allows refocussing of
early eluting components.
Page: 254
Varian, Inc.
Electrical System
The SPT option provides independent temperature control of 2 parts of the trap
assembly:
1. Heating and cooling of the trap.
2. Heating of the trap ends.
Trap temperature programming is accomplished in 450-GC heated zone 3,
controlling heating and cryogen cooling of the trap. The SPT trap is mounted on
the SPT valve oven, which maintains the SPT trap ends at a constant
temperature. The oven is powered by one of the other remaining heated zones.
Extremely fast heating of the trap is made possible by utilizing the wall of the
trap tubing as an electrical heating element. This eliminates the extraneous
thermal mass associated with conventional heater cartridges. To match the
electrical resistance characteristics of the trap tubing, a specific electrical
voltage and current are provided by the SPT transformer (mounted in the rear
of the GC). To provide the required voltages, the SPT trap harness connectors
for the short and long trap are different. Although low voltages (less than 8 volts
AC) are used to heat the trap, the SPT transformer is designed to deliver very
high currents.
The trap ends carry dangerous currents. Do not put any conducting
material, such as wrenches, screwdrivers, or metal tubing, across the trap
ends. Do not wear metallic jewelry while working near the trap.
Variations in line voltage supplied to the GC will have an effect on the
heating rate while the trap is desorbing.
Varian, Inc.
Page: 255
Circuit breaker
The circuit breaker reset button is located above the SPT transformer on the
rear panel of the GC. This pop-out button protects the SPT from abnormal
power usage by interrupting the electrical connection to the trap. Wait at least 5
minutes after the circuit breaker is tripped before resetting it.
Page: 256
Varian, Inc.
LCO2 Option
System Description
The SPT-LCO2 option allows trapping temperatures as low as -60C. This
system is very efficient and consumes a minimum amount of LCO2.
Because of low coolant consumption, conditions that affect thermal efficiency of
the coolant transfer system are very important to consider.
Varian, Inc.
Page: 257
Theory of Operation
As LCO2 expands into the gas phase in the SPT cryo Enclosure, the latent heat
of vaporization provides cooling power. The coolant flow rate is regulated by
varying the pulse rate of the cryo solenoid valve.
The latent heat of vaporization drops rapidly as the temperature increases to
the CO2 critical point of 31C. In order for the SPT to have adequate cooling
power, LCO2 should be delivered to the cryo valve at 27C or cooler.
Characteristics of LCO2 as a Coolant
The coolant for this option is liquid phase carbon dioxide (LCO2) supplied from
a pressurized cylinder a. At pressure, this cylinder contains both liquid and gas
phases coexisting at equilibrium. The cylinder pressure is then equal to the
saturation pressure and is dependent on the supply temperature. As the
cylinder temperature rises, the cylinder pressure rises. Thus if a constant
temperature is provided a constant pressure is maintained. This is true until the
liquid phase CO2 is depleted. For this reason it is advisable to keep the cylinder
at a fairly full condition. Do not let it drop more than 60% of its weight.
Inadequate Cooling
To insure reliable cooling power and trapping temperature stability, the CO2
must be delivered as liquid phase CO2. When cooling power is lost, the CO2 is
often being delivered as gas phase. This may be caused by 2 different
conditions.
1. CO2 supply is too low. There is insufficient CO2 remaining in the cylinder to
build supply pressure to the saturation pressure.
2. The CO2 has vaporized in the transfer line because the temperature of the
transfer line is higher then that of the supply. The supply temperature must
be equal to or warmer than the transfer line.
Locating the LCO2 supply cylinder near the GC in a temperature controlled
room is ideal.
Varian, Inc.
Transfer Line
The SPT option comes with a 3 meter long, small diameter, low volume transfer
line. The supply cylinder must be located then, less than 3 meter from the GC.
The low volume feature of the transfer line helps this option tolerate situations
where the GC and transfer line are slightly warmer than the supply cylinder.
Keep the transfer line cool and stable by insulating it with foam insulation
provided in the Accessory Kit.
Recommended Supply Conditions
Supply
Supply Conditions
Outlet Filter
Transfer Line (from LCO2
cylinder to the GC)
Varian, Inc.
Page: 259
LN2 Option
Description
The SPT - LN2 option allows trapping temperatures as low as -190C. For best
utilization of the high performance features of the SPT, it is recommended that
the transfer line from the LN2 source to the GC be very efficient. The LN2
transfer line should have the following characteristics:
Minimal length.
Should not have excessive restriction to flow.
Low thermal mass.
Well insulated.
In addition, the transfer line should not be routed near heat sources that could
reduce the cooling efficiency, such as the column oven exhaust fan.
If the SPT system is at ambient temperature (e.g., 23C) the time required
for initial cooling of the SPT from 50 to -190C might be longer
(approximately 5-10 minutes) than subsequent cooling cycles.
Page: 260
Varian, Inc.
Supply
Supply Pressure
19 to 25 psig
Parameter
450-GC Range
Recommended
Trapping Temperature
-60 to 450 C
-185 to 450
LCO2 - 60 to 50 C
LN2
- 185 to 40 C
Desorption Temperature
Program 1
Desorption Temperature
Program 2
SPT Temperature Limit
SPT Valve Oven
50 to 420C
50 to 420 C
-60 to 450 C
-185 to 450 C
--
-150 to 250 C
Page: 261
Page: 262
Varian, Inc.
Varian, Inc.
Page: 263
Page: 264
Varian, Inc.
Figure 41: SPT Components (Short Trap Assembly, LCO2 Option Shown
Varian, Inc.
Page: 265
Page: 266
Varian, Inc.
Page: 267
1. Unscrew the cryo nozzle insert using a large flat bladed screwdriver.
Do not remove the cryo nozzle insert from within the cryo nozzle
assembly.
2. Remove entire cryo nozzle assembly: hold both cryo nozzle and
cryo nozzle insert together while removing them from within the trap coil.
This prevents loss of the three belleville washers that are stacked inside
the cryo nozzle assembly.
If the used trap assembly is still functional, its ends should be capped off to
isolate it from atmospheric contamination during storage.
Page: 268
Varian, Inc.
4. Preassemble the LN2 cryo nozzle components (Figure 43, on page 268):
a. Using a narrow pointed tool (such as a small screwdriver), align the
stack of three belleville washers in alternating orientations,
placing the first washer with its convex side facing the pointed tool.
b. Holding the pointed tool pointing up so the washer stack doesnt fall
off, lower the cryo nozzle over the washer stack so that the
washers are inserted fully into the large cavity of the cryo nozzle.
c.
Tip the cryo nozzle so that its large cavity is facing upwards, then
remove the pointed tool so as to not disturb the orientation of the
washer stack.
d. Insert the cryo nozzle insert (slotted end facing out) into the cryo
nozzle cavity.
e. Keep the cryo nozzle insert closely engaged within the cryo
nozzle.
5. Align the LN2 end plate assembly with the right open end of the cryo
enclosure and engage the cryo nozzle assembly (Figure 43, on page 268):
a. Orient the LN2 end plate assembly so that its delivery tube is
pointing away from you. Vent barb and vent tube should be at the
bottom with vent tube exiting to the rear.
b. The LN2 end pilot assembly, which is attached to the inside of the
LN2 end plate assembly, should fit into the right open end of the
cryo enclosure. Take care not to catch the edge of the gasket
Varian, Inc.
Page: 269
against the rim of the cryo enclosure; the rim should bear evenly
against the gasket facing in the area surrounding the end pilot.
c.
From the left side of the trap coil, engage the cryo nozzle
assembly through the coil. Screw the cryo nozzle insert onto the
threaded stud of the LN2 end pilot assembly using a large flat
blade screwdriver. Take care not to cross-thread the cryo nozzle
insert. Tighten gently, then back off approximately 1/12th of a turn.
6. To align the LCO2 end plate assembly, with the right open end of the cryo
enclosure:
a. Orient the LCO2 end plate assembly so that its delivery tube is
near the bottom of the right face and pointing away from you. The
Vent barb and vent hose should exiting to the rear.
b. The LCO2 end pilot assembly, which is attached to the inside of
the CO2 end plate assembly, should fit into the right open end of
the cryo enclosure. Take care not to catch the edge of the gasket
against the rim of the cryo enclosure; the rim should bear evenly
against the gasket facing.
7. Align the blank end plate assembly (see Figure 42, on page 267) with the
left open end of the cryo enclosure. The four SPT extension springs
attached to this assembly will have to be held out of the way of the
enclosure. The blank end pilot should engage into the left rim of the cryo
enclosure. Take care not to catch the edge of the gasket against the rim of
the enclosure; the rim should bear evenly against the gasket facing.
8. Use needle nose pliers to gently grip the open hook end of the extension
springs, and extend the springs to engage their respective slots that
should be aligned on the right end of the cryo enclosure in the LCO2 or
LN2 end plate assembly. Do not overextend the springs. Overextending
them might result in their permanent distortion.
9. If both end plate assemblies are properly engaged into the cryo
enclosure, they should be parallel to each other and perpendicular to the
sides of the cryo enclosure. The SPT extension springs should firmly hold
the end plate assemblies against the rims of the cryo enclosure. The
SPT extension springs should not touch the trap ends.
10. The oven/trap lines that connect to the trap end fittings should be
preformed for proper alignment with the trap end fittings. The nut and the
reducing ferrules should be preplaced onto the line ends so that when the
trap is installed, there is minimal bending of the lines.
To maintain the ultra-clean condition of the reducing ferrules during
installation, the use of clean cotton gloves or clean forceps is
recommended.
Page: 270
Varian, Inc.
11. Confirm that the trap assembly with the trap ends pointing towards the
front of the GC, with the word TOP on the cryo enclosure facing
upwards.
12. Engage the oven/trap lines into trap ends.
13. Align both trap ends with the slots in the trap mounting block. Lower trap
ends to engage into the slots.
14. Place the trap clamping bar over the trap ends.
15. Make sure the belleville washer on the trap clamping bar screw has its
concave side facing the trap clamping bar. Tighten trap clamping bar
screw to anchor the trap assembly into the SPT valve oven. Do not
overtighten.
16. Reconnect the valve oven/trap lines to trap end fittings. The packing type
number is stamped on the inlet trap end fitting (see Figure 41 on page
265). Carefully engage each nut onto its respective fitting to prevent crossthreading. Using the faces of the hexagonal nut as a guide, tighten each nut
fingertight, then tighten an additional:
a. Using new ferrules, 4/6 - 5/6 of a turn.
b. Reusing ferrules, 2/6 of a turn.
17. Do not over tighten. Use 3/16 backup wrench to prevent rotation of the
trap end fittings (the flats on fittings should remain vertical).
To avoid the possibility shorting the SPT electrical circuit, do not insert
the valve oven/trap lines more than 8 mm into the trap ends. Inserting the
lines too far into the trap ends will cause them to come in contact with
the trap.
Do not plumb the valve in such a way that metal tubing shorts across
trap tube ends.
Varian, Inc.
Page: 271
19. Connect the LN2 delivery line (part of LN2 end plate assembly) to the
LN2 solenoid valve by engaging the Swagelok nut to the connector on the
valve (see on page 266). Do not over tighten Swagelok fitting. The Valve
mounting bracket might need to be anchored to the GC. Do not induce
strain onto the delivery line: carefully align the LN2 solenoid valve and its
bracket during tightening of the valve mounting bracket.
20. Check that the drip pan is located under the LN2 solenoid valve. The drip
pan should not excessively block vent holes in the column oven top.
21. Tighten down the SPT valve oven cover.
22. Install the SPT cover.
23. Connect the SPT trap harness connector to the SPT transformer harness.
24. Replace GC top cover.
25. Turn on coolant supply.
26. Power up GC.
Page: 272
Varian, Inc.
Trap Conditioning
With carrier or auxiliary gas passing through the trap, condition the trap
according to the following table:
Packing Type
Mesh Size
Conditioning Procedure
60-80
Tenax/GR Charcoal
60-80
4 to 16 hours @ 250 C
HayeSep D
60-80
4 to 16 hours @ 250 C
60-80
4 to 16 hours @ 270 C
60-80
4 to 16 hours @ 270 C
60-80
4 hours @ 225 C
Tenax TA,
60-80
4 hours @ 225 C
60-80
4 hours @ 200 C
60-80
4 hours @ 200 C
Carbopack C, Carbopack B
60-80
4 hours @ 270 C
60-80
4 hours @ 200 C
Tenax GR
60-80
4 hours @ 200 C
Varian, Inc.
Page: 273
Partnumber
Short Traps
Short Trap Assembly Kit (8 cm), silanized glass beads
03-925713-92
03-925713-91
Blank Trap
03-925713-01
03-925713-02
Long Traps
Long Trap Assembly Kit (30 cm), blank
03-925714-91
Blank Trap
03-925714-01
Tenax GR/Charcoal
03-925714-02
Tenax GR
03-925714-03
03-925714-04
03-925714-05
03-925714-06
03-925714-07
HayeSep D
03-925714-08
03-925714-09
Tenax TA
03-925714-10
03-925714-11
03-925714-91
03-925714-92
Kit, Tenax GR
03-925714-93
03-925714-94
03-925714-95
03-925714-96
03-925714-97
Kit, HayeSep D
03-925714-98
03-925714-99
Kit, Tenax TA
03-925714-80
03-925714-81
Varian, Inc.
Part Description
Cryo Valve:
LCO2
LN2
03-925555-01
03-925557-01
28-694502-00
28-694501-00
24-000467-00
27-590826-00
03-918850-00
03-918853-01
SPT PCB
03-918758-00
SPT Transformer
03-918788-00
37-000146-01
03-918765-00
03-918764-00
Union (LCO2)
28-211457-00
03-918852-00
28-693996-00
28-693997-00
28-693998-00
28-694027-00
28-694028-00
28-694029-00
03-918850-00
03-918801-01
03-918801-02
Table 51: SPT Replacement Parts
Varian, Inc.
Page: 275
Troubleshooting
This section provides troubleshooting information to aid the operator in locating
and correcting apparent problems when working with an SPT system. The
following table is geared specifically to SPT troubleshooting. If you have
questions about any of the corrective actions described below, contact your
Varian Customer Support Representative.
Symptom
Possible Cause
Remedy
SPT overheating.
Page: 276
Varian, Inc.
Symptom
Possible Cause
Remedy
No coolant flow.
Perform trapping
temperature program.
SPT overcooling.
SPT undercooling or
cooling too slowly.
Enable coolant.
Varian, Inc.
Symptom
Page: 278
Possible Cause
Remedy
Varian, Inc.
Detectors
DETECTORS
INTRODUCTION
The 450-GC can accommodate up to three installed detectors and have all
three running simultaneously.
The standard detectors available on the 450-GC are:
Except for the TCD and PFPD any combination of three detectors can be
installed.
In case of the PFPD, only two may be installed but any one additional detector
may also be installed including a TCD.
In the case of the standard TCD, only two can be installed and then no
additional detectors can be installed.
However, a dual TCD option is available as a custom solution which allows
installation of an additional ionization detector.
The dual TCD consists of 2 detector cells in one housing but using only one
heating zone.
Detectors are mounted on the top of the 450-GC, above the left side of the
column oven.
The position of the detector, however, is determined by the location of the
detectors electrometer in the electronics cabinet. Detector electrometers are
installed on the left side of the GC accessed by removing the left side panel.
All cables connecting the electrometers to the detectors are accessible by
removing the 2 covers on the top left of the 450-GC, i.e. the detector top cover
and the cover over the keyboard/display.
For connecting a capillary column to a detector, see Connect Capillary Column
to Detector on page 24.
Varian, Inc.
Page: 279
Detectors
ANALOG OUTPUT
The 450-GC can be equipped with an option board.
The analog output signals are located at connector J401, on page 36 on the top
of the GC, accessed by removing the top left cover.
This connector contains all 3 analog channels plus status signals for Ready-In
to the GC and Start Out from the GC.
page by the detector tab on the second page (Page 2 or
In the SETUP
page 3) the Detector Analog Output screen will be visible, only if the Analog
option board is present.
The Analog Output Port name can be changed by the button Edit.
If the Range setting must be changed, make sure that the Range setting
matches the mainboard jumper settings!
Page: 280
Varian, Inc.
Detectors
The analog output ports are programmed in the 450-GC method by pressing
key on the keyboard, select Analog Output.
the DETECTOR
This brings up a menu screen from which you choose one of the three ports to
program; Page 1, Page 2 or Page 3. The following screen appears when
Page 1 is chosen.
The Analog Output Port name is in this example is A, but can be changed in the
Setup screen. The individual analog ports can be assigned to any installed
detector on the 450-GC. The ports are identified as A, B and C with the default
assignment being front, middle and rear detector respectively. Default Analog
output voltage is 0-1 Volt (0-10 Volt is possible by selectable jumpers on the
mainboard (contact your local Varian office).
The right-side of the screen is used to time program the signal source of the
analog output. By default analog output A is assigned to the front detector. In
the example above, the analog port is programmed to transmit the Front
detector signal for the first 10 minutes of the run and then switch to the Middle
detector after 20 minutes the switch will be made to the rear detector for the
remainder of the run. This mode of operation is known as detector switching.
Note that the attenuation should normally be set to 0. However for optimum
resolution (less electronic noise) the best results are achieved by increasing the
attenuation if small peaks are detected. For overloaded peaks decrease
attenuation.
Varian, Inc.
Page: 281
Detectors
The 450-GC has a number of analog cable options on page 94, depending on
the device to which the cable is connected. All cables have a 15 pin D-shell
connector on one end to attach to J401 on the 450-GC and have the
appropriate connectors on the other end of the cable for the devices to which
they are being attached.
Page: 282
Varian, Inc.
The flame ionization detector (FID) is the most sensitive gas chromatographic
detector for hydrocarbons such as butane or hexane. With a linear range for 6
or 7 orders of magnitude (106 to 107) and limits of detection in the low picogram
or femtogram range, the FID is the gas chromatographic detector for volatile
hydrocarbons and many carbon containing compounds.
The FID detector employs hydrogen as the combustion gas which is mixed with
Air and the column eluent (helium, nitrogen or other appropriate gas) and burn
at a small jet situated inside a cylindrical electrode. A potential of a few hundred
volts is applied between the jet and the electrode and when a carbon containing
solute is burn in the jet, the electron/ion pairs that are formed are collected at
the jet and cylindrical electrode.
The following instructions refer to the operation of a 450-GC Flame Ionization
Detector (FID). The FID is installed on a detector base, directly above the
column oven. The FID electrometer is installed in the electronics cabinet on the
left of the instrument. The electronics of the FID are controlled from the 450-GC
Display/Keyboard, the gas flows are set and controlled either from the 450-GC
Display/Keyboard (if detector EFC is installed) or at the pneumatics panel.
Varian, Inc.
Page: 283
Initial Set-Up
To set-up the FID, proceed as follows. Note that when the 450-GC is first
powered up all heated zones with the exception of the column oven are
powered off and all detector electronics are turned off.
1. Refer to the Pre-Installation Instructions for additional information on gas
supply requirements.
key on the 450-GC Display/Keyboard to display
2. Press the DETECTOR
the DETECTOR screen. The FID should be listed in position Front ( ),
Middle ( ) or Rear ( ). If your GC is equipped with detector EFC verify the
makeup gas you will be using with your FID.
3. Install a chromatographic column in the GC, on page 13. If the analytical
column is not pre-conditioned, use a no-hole ferrule in the detector column
nut while conditioning the column and leave the detector end of the column
loose in the oven. If the analytical column is pre-conditioned, follow the
usual procedure for column installation.
Page: 284
Varian, Inc.
4. If detector EFC is present, set the following flow rates in the adjustments
section of the FID method:
Operation
1. Check the detector temperature status on the DETECTOR/FID screen.
Check the FID range setting and set to range 12, if necessary.
Generally, to prevent water condensation in the detector assembly, the
detector should be operated at a temperature above the column
temperature and not below 150 C. If the detector is operated at a lower
temperature, condensation can lead to excessive noise.
2. After the FID oven has reached its setpoint temperature, turn on the FID
electronics. Note that the FID will ignite automatically when auto-ignite
conditions are fulfilled (for detailed information about auto-ignite conditions
see here, on page 287). The 450-GC monitors the FID background current
continuously and will attempt to ignite the flame (only if in Setup/Detector
the checkbox on page 75 check for flame out is checked) if the current
drops below a specified threshold. Up to three attempts to ignite the flame
will be made before the 450-GC reports a flame-out fault. If a flame-out fault
is reported, correct the cause of the flame-out and then clear the fault in
Instruments Log.
The FID will automatically ignite after the fault is cleared.
Varian, Inc.
Page: 285
4. The Ignite button applies power to the FID igniter filament for 5
seconds.
5. The Autozero button applies an immediate zeroing of the detector
signal.
Page: 286
Varian, Inc.
1
Figure 44: FID auto-Ignite flow diagram
a In an automated sequence the 60 seconds delay helps to recover from FID flame faults and avoids
having the fault again in the next run. It also prevents a workstation starting a run while FID ignition is still
going on.
Varian, Inc.
Page: 287
Installation/Disassembly
At times, it is necessary to remove the detector components to inspect, to
clean, to replace parts, or to install another detector. Follow the
disassemble/reassemble instructions on the next pages for both the FID and
the detector oven.
Exploded View
6
1
8
9
4
10
1
2
3
4
5
6
7
8
9
10
11
11
12
13
14
15
16
-----
11
Signal Cable (200187700)
Signal Probe (200187300)
12
Ignitor Cable (392548701)
Ignitor Probe (200187200)
13
Screws, 8-32 x (1214200804)
Collector Tube (394958700)
Tower Screws, 8-32 x 3 (391866301)
14
Ceramic Insulator (2100003200)
Tower Top Plate (3400073100)
Detector Tower Body (100099300)
Flame Tip, 0.01 in. Capillary Tip (200187500)
15
Flame Tip, 0.02 in. Standard Tip (200193800)
Ferrule: Vespel (100%) max. 350C (CR212200)
16
Graphite/Vespel 40/60 max. 400C (CR213200)
Graphite (100%) max. 450C (CR211200)
Aluminum Seal Washer (1500334701)
Figure 45: FID Exploded View
Detector Base (392547501)
Column Ferrule
Column Nut (394955100)
Flow Tube (200187600)
Large Stopper (394958600)
Page: 288
Varian, Inc.
Varian, Inc.
Page: 289
Page: 290
Varian, Inc.
9. Insert the signal probe into the upper arm of the detector tower. The probe
clip should fit around the tapered section on the collector tube tightly
enough to exert a downward force (see Figure 47). Secure the signal probe
by tightening the knurled nut. Make sure the notch in the detector tower arm
does not cut the O-ring seal.
10. Connect the igniter cable to the igniter probe and the signal cable to the
signal probe.
Detector Tower
TOP VIEW
Ignitor Coil
Spring Clip
Flame Tip
Figure 46: Orientation of Igniter Probe Elements in FID Tower
Collector Tube
(03-949587-00)
Signal Probe
Insulator
Ignitor Coil
Notch
Spring Clip
Flame Tip
Assembly
Ignitor Probe
Varian, Inc.
Page: 291
MAINTENANCE
When the performance of the FID has degraded you may want to replace the
flame tip ferrule, or clean the deposits from the internal parts, including the
flame tip.
1. Remove the detector and internal parts (see disassemble the FID, on page
289).
2. Scrape the deposits from the bore of the collector tube, the insulator and
the metal part of the flame tip with emery cloth (e.g., SiO2).
3. If the flame tip is plugged, insert a wire through the flame tip orifice to clear
it.
4. If you have access to an ultrasonic cleaner, clean the collector, insulator
and flame tip with distilled water.
5. Rinse the cleaned components with methanol or acetone and air dry.
6. Wipe the detector tower with acetone.
7. Clean the probe arms with methanol and air dry or dry in an air oven
(maximum drying temperature of 150 C).
8. Reassemble the FID, see Reassemble the FID, on page 290.
Page: 292
Varian, Inc.
Varian, Inc.
Page: 293
Page: 294
Varian, Inc.
Varian, Inc.
Page: 295
Page: 296
Varian, Inc.
Cable overview
Varian, Inc.
Page: 297
Exploded View
1.
2.
3.
4.
5.
3
5
Page: 298
Varian, Inc.
Gas Connections
Installation
The installation of the PDHID detector is normally done at the factory, field
installations are not advised.
Varian, Inc.
Page: 299
General Precautions
Do not turn the unit on until helium discharge gas is flowing through the
detector.
Do not shut off or disconnect the discharge gas when the detector is hot, even if
the unit is turned off. Turn off the discharge power switch and allow the detector
to cool down naturally before disconnecting or shutting off the discharge gas.
Do not cover the unit with anything which could restrict air circulation.
Position the controller unit where the mains switch on the rear panel can be
reached easily.
Page: 300
Varian, Inc.
5. Observe the needle of the output pressure gauge for 15 minutes. There will
be a slight initial drop, but if it doesnt move after that, consider that all the
connections are tight.
6. If necessary, use an electronic leak detector to locate any leaks. If a leak
detector is not available, tighten all the fittings (including the output
pressure gauge) and repressurize the system for another test.
7. Uncap the outlet tube of the purifier and purge the system for 15 to 30
minutes at 60 - 80 mL/min to eliminate air from the purifier getter material.
Page: 301
Page: 302
Varian, Inc.
Varian, Inc.
Page: 303
Page: 304
Varian, Inc.
Varian, Inc.
Page: 305
Initial Set-Up
To set up the TSD for operation, proceed as follows. Note that when the 450GC is first powered up all heated zones with the exception of the column oven
are powered off and all detector electronics are turned off.
1. Refer to the Pre-Installation Instructions for additional information on gas
supply requirements.
key on the 450-GC Display/Keyboard to display
2. Press the DETECTOR
the DETECTOR screen. The TSD should be listed in location Front ( ),
Middle ( ) or Rear ( ). If your 450-GC is equipped with detector EFC,
verify the makeup gas you will be using with your TSD.
Page: 306
Varian, Inc.
4. Condition the TSD bead. The following bead conditioning and optimization
procedures assume use of a short, non-polar column such as the Varian
test column, 03-912300-30 (packed, COLUMN, 1/8" X 50 CM 5%, OV101)
or CP887098 (capillary CP-Sil 5 CB, 4M X .32). If you use a different type
of column, you should adjust conditions accordingly.
Page: 307
175 C (isothermal).
80 C (1 min), 20 C/min to 200 C (2 min).
230 C.
300 C.
5. Set the initial Column oven temperature. Set the desired Injector
temperature to which your TSD column is connected. Check that the
Injector oven power is ON. Detector temperature is set by pressing the
key, selecting the TSD, setting the temperature to 300 C
DETECTOR
and turning the Oven Power ON. Leave the Electronics and beadcurrent
OFF at this point. Set the detector to range 12 and AUTOZ. unchecked.
6. Allow the TSD detector to reach temperature and allow the bead to
thermally equilibrate for at least 15 minutes prior to turning on the
electronics.
7. Set the initial Autozero OFF. Set the bead current to 3.100 A. Turn on the
Bead Power. Allow the bead to condition for at least 15 minutes.
8. Increase the bead current in steps of 0.100 Ampere, waiting at least 15
minutes for the bead to equilibrate between increases. When increasing the
bead current causes the TSD Signal to go above 64 mV, the bead has
reached its ignition point. Allow the bead current to remain at this setting for
a minimum of 12 hours (preferably overnight). This completes the initial
conditioning of the TSD bead probe. Now optimization of the bead probe for
N/C selectivity is required.
The Clear Autozero button removes the Autozero offset so that you can
observe the detector background signal.
Page: 308
Varian, Inc.
Varian, Inc.
Page: 309
1. Azobenzene:
2 ng Azobenzene contains 310 pg nitrogen
2. Heptadecane:
4 g heptadecane contains 3.4 g carbon
3. Methylparathion:
Capillary Column
Packed Column
4. Malathion:
4 ng Malathion contains 380 pg phosphorus
Figure 52: Structure of Components in the TSD Test Sample & Chromatogram examples
Page: 310
Varian, Inc.
Column Bleed
To minimize contamination of the detector by column bleed, all new columns
should be conditioned before they are connected to the detector. Condition new
columns for a minimum of 24 hours at 40 C (or 2 hours at maximum analysis
top) below the maximum temperature of the column packing liquid phase.
Since the TSD detects compounds containing nitrogen or phosphorus, column
packing materials containing these elements should be avoided. Common liquid
phases containing nitrogen are OV-225, OV-275, FFAP, XE-60 and TCEP. If
you must use such columns, it is important that they be well conditioned and
operated at as low a temperature as possible in order to minimize column
bleed. High column bleed produces high background signals, making it difficult
or impossible to set the bead current and can result in the occurrence of
negative peaks in the chromatogram.
Solvent Considerations
When the TSD is used for trace level analysis, only solvents free of nitrogen,
phosphorus, sulfur, or halogenated compounds should be used. Depending on
the volatility of the impurity in a solvent, it can produce excessive solvent tailing
or extraneous chromatographic peaks.
Chlorinated Solvents
Chlorinated solvents can be used with the TSD. However, these solvents
generally cause an abrupt increase in both detector background signal and
sample response. These effects appear to be associated with a temporary
adsorption of chlorinated species onto the surface of the hot alkali-ceramic
bead. Repetitive injections of samples in chlorinated solvents yield constant
responses. However, once the injection of chlorinated solvents is stopped, the
background and sample response will decay back to the response levels that
existed before the initial injection of the chlorinated solvent. Therefore, when
chlorinated solvents are used, calibration standards should be run frequently.
You can minimize the effect of chlorinated solvents on the bead by turning off
bead current during the time that the solvent elutes, using the time
programming table of the detector method. The total time that the bead is
turned off should be less than 2 minutes in order to minimize thermal reequilibration times of the detector. (Any peaks eluting during this off time will not
be measured.)
Varian, Inc.
Page: 311
Page: 312
Varian, Inc.
TSD Assembly
It may be necessary to remove the TSD detector components for inspection,
cleaning, parts replacement, or installation of another detector. Disassembly/reassembly instructions are detailed below for both the TSD and the detector
oven.
Exploded view
1
2
3
4
5
16
17
18
8
19
9
1
2
3
4
5
6
7
8
9
10
11
12
13
14
15
16
17
18
19
---
Varian, Inc.
10
11
12
13
14
15
Page: 313
TSD disassembly
1. The TSD is mounted on the ion detector base in the detector oven.
2. Turn the TSD Oven Power and Bead Power OFF. Allow the detector to cool
to ambient temperature.
3. Turn off the main GC gas supplies to the detector at the pneumatics panel.
4. Disconnect the signal and the igniter cables from the TSD tower.
5. Remove the 2 tower mounting screws and lift the detector tower vertically
from the detector base until clear of the flame tip. Remove the 2 screws
from the top clamp.
6.
7. Remove the signal probe. There may be some resistance to removal, due
to the O-ring seals. Do not rotate probe as it is withdrawn from electrical
contacts in tower.
8. Remove the collector and the lower insulator with tweezers.
9. Remove the bead probe. There may be some resistance to removal, due to
the O-ring seals. Do not rotate probe as it is withdrawn from electrical
contacts in tower.
10. Carefully remove the flame tip assembly from the detector base.
11. Remove and discard the aluminum seal washer from detector base. In
some cases this washer may remain in the detector tower. If so, be sure to
remove and discard this washer so that the detector is not accidentally
reassembled with both the old washer and a new washer.
Page: 314
Varian, Inc.
TSD Re-assembly
Refer Figure 53 for re-assembly orientation.
1. Re-assembly is the reverse of removal.
2. Install the flame tip in the detector base. This is a 0.020-inch ID ceramic
flame tip. The TSD and FID flame tips are identical. DO NOT use a
0.010-inch flame tip for the TSD.
a. If you are installing the flame tip assembly with a used Vespel
ferrule, tighten the assembly finger-tight plus an extra 1/6-turn.
b. If you are installing the flame tip assembly with a new Vespel
ferrule, tighten about 1/3-turn past finger-tight.
3. The Vespel ferrule supplied with the GC has a maximum temperature limit
of 350C. If it is ever necessary to operate the detector above 350 C,
replace the Vespel ferrule with a graphite ferrule.
4. Install a new aluminum seal washer onto the shoulder of the detector base.
For a reliable tower seal, use a new aluminum washer each time the
detector is installed.
5. Measuring the nominal 4.0 mL/min H2 flow for the TSD may be simplified at
this time by attaching a tube to the flame tip and then connecting the tube
to a flowmeter. Make sure the makeup gas is turned of and adjust the
hydrogen flow.
6. Replace the detector tower on the detector base and secure with the 2
screws. Tighten screws alternately as tower is tightened into place.
Varian, Inc.
Page: 315
10. Insert signal probe into upper arm of detector tower. Carefully align the
spring clip with the collector. Be sure the key on the probe and the tower
arm are aligned. Secure signal probe by tightening knurled nut.
11. Assemble the upper insulator and top clamp over the collector. The O-ring
will provide slight resistance to complete insertion. Take care not to tear the
O-ring or the tower will not seal properly.
12. Look down the tower assembly to verify that the bead does not touch the
collector. Carefully adjust the collector as necessary.
13. Assemble the top clamp to the tower and secure with the 2 screws.
14. Connect the signal cable to the signal probe (top) and the igniter cable to
the bead probe (bottom). If cables were totally disconnected from the GC,
reconnect signal cable and the bead current supply cable at the top of the
TSD Electrometer Board.
1
2
3
4
5
Page: 316
Varian, Inc.
MAINTENANCE
Cleaning / Maintenance
When the TSD exhibits low sensitivity or high background noise, cleaning is
indicated. Visual inspection may also indicate a need for cleaning. The ceramic
bead, collector, or flame tip are all parts that are deposit prone, depending on
the type of samples being used. Clean these parts as indicated in the following
paragraphs.
Always handle detector components with tweezers to avoid
contamination.
Bead Cleaning
Non-volatile deposits can be removed from the bead by gently scraping the
bead surface with a sharp tool or abrasive paper. The bead should be
supported while being cleaned to prevent bending the lead and cracking the
ceramic coating on the leads.
1. Remove the detector and internal parts. Handle internal parts with tweezers
and place on a clean, uncontaminated surface.
2. Clean deposits from the surface of the collector with water or abrasive
paper.
3. Remove deposits which form on top of the flame tip by scraping. If flame tip
is plugged, clear by inserting wire (approximately 0.012 inch diameter) into
flame tip orifice.
Varian, Inc.
Page: 317
Test Chromatogram
For an FID test chromatogram see section Detector Test Chromatograms,
on page 421.
Page: 318
Varian, Inc.
Varian, Inc.
Page: 319
Initial Set-Up
To set up the TCD, proceed as follows:
Note that the TCD is a 2 channel detector and requires both a sample flow of
carrier gas and a reference flow of the same gas. Either channel can be used
as the sample stream but there must always be a reference stream.
There are TCD parameters which need to be set before using the TCD for
analysis. These are the choice of carrier, make-up and reference gas and the
filament temperature limit.
1. Refer to the Pre-installation Instructions for additional information on gas
supply requirements.
2. Check the gas supply pressures. Flow controllers with turn dials are
calibrated for Helium at 80 psig. Recalibrate the flow controllers if you use a
different supply pressure or carrier gas.
key on the 450-GC Display/Keyboard. Select the
3. Press the SETUP
TCD by location. Select the carrier, Make-up gas and TCD oven
temperature limit. Press OK.
Page: 320
Varian, Inc.
Varian, Inc.
Page: 321
TCD Adjustments
With prolonged use, the filaments slowly oxidize and their resistances increase.
At some point this will cause the largest peaks in your chromatogram to
become flat topped, indicating that the filament protection software has been
activated. (Note that if the TCD signal is sufficiently large the output may go to
zero and become unresponsive. If this occurs, adjust your injection site and
conditions to avoid over range conditions.) To adjust for this increase in
resistance, set the Filament Temperature Limit to 490 C, even for filament
temperature setpoints below 390 C.
The TCD filament is automatically protected in the GC. The GC turns off the
filament current after 4 minutes if it detects that the carrier gas flow has stopped
or that the filament current is low. Also, the GC turns off filament current when it
detects that carrier gas is flowing through one side of the detector cell but not
through the other (e.g., a septum or column is not installed, or a leak exists in
one cell and not the other).
If you select Helium or H2 as the carrier gas, the GC operates in the standard
way to protect the TCD filament. However, because the filament protection
feature in the GC operates through the detection of air (or N2) in the detector
cells, the filament protection feature is disabled when you use nitrogen or argon
as the carrier gas.
Set the Filament Temperature Limit to 390 or 490 C, according to the highest
filament temperature setpoint required. With the Limit set to 390 C, the
maximum filament temperature is limited to approximately 450 C. The
temperature limit is approximately 550 C at the 490 C setting. The 450 C
limit protects filaments from oxidation when exposed to air over an extended
period of time. However, at the 550 C temperature limit the filament can
degrade after only a few minutes.
Page: 322
Varian, Inc.
Operation
To operate, the TCD requires only carrier gas. Helium, Argon or Hydrogen (only
with special EFC16) are recommended as carrier gases. You can use nitrogen
as a carrier gas, but you may lose some sensitivity and see an increase in the
detector noise.
Page: 323
5. Un-check AutoZ. (Autozero OFF). Press the Autozero button to set the
initial signal to 0. Monitor the detector signal. A continuous drift, either
positive or negative, may indicate an air leak in either the carrier or
reference gas stream.
6. Note that it is normal for the TCD baseline to drift until the detector
thermally equilibrates. The time it takes to reach equilibrium always
exceeds the time it takes for the detector oven to reach its temperature.
When the baseline is stable, you can begin your analysis. Check the AutoZ.
(Autozero ON). Autozero is automatically disabled when the GC method is
in RUN.
Page: 324
Varian, Inc.
If you inject a standard or sample into the reference side of the TCD, peaks
appear in the negative direction. If you want to integrate or process these peaks
with a data system, you must convert the polarity of the peaks from negative to
positive. Use the checkbox Polarity.
Use the time program when you would like the polarity to be automatically
reversed during a chromatographic run.
Range Setting
Use the 0.05 mV range in the following cases:
Varian, Inc.
Page: 325
A.
Filament current: 290 mA
Filament temperature: 300 C
At 1.5 minutes, range and attenuation
changes from 2 x 0.5 mV to 128 x 5.0
mV.
B.
Filament current: 151 mA
Filament temperature: 160 C
At 1.5 minutes, range and attenuation
changes from 2 x 0.05 mV to 128 x 0.5
mV.
C.
Filament current: 299 mA
Filament temperature: 300 C
At 1.5 minutes, range and attenuation
changes from 4 x 0.05 mV to 128 x 5.0
mV.
Page: 326
Varian, Inc.
S=
P Fc
W
Where:
S = Sensitivity in mV x mL/mg.
P = Integrated peak area in mV x min.
W = Weight of sample in carrier gas in mg.
FC = Carrier gas flow in mL/min corrected to detector temp
Td Pw
Fc = Fo
1
Pa
Ta
Where:
FO = Flow rate measured at detector outlet (at ambient temperature) in
mL/min.
TD = Detector temperature in K (Kelvin).
TA = Ambient temperature in K (Kelvin).
PW = Partial pressure of H2O at ambient temperature in torr.
PA = Ambient pressure in torr.
Note:
The
flowmeter.
Pw
1
Pa
S=
E
Cd
Where:
S = Sensitivity in mV x mL/mg
E = Detector signal (peak height) in mV
CD = Concentration of test substance in the measured
peak volume in mg/mL
Varian, Inc.
Page: 327
Peak volume =
Peak width at half height (in min.) x gas flow rate at detector
(in mL/min) = W h/2 x FC
Calculate Detector Detectivity
To calculate the detectivity of the TCD, proceed as follows:
Detectivity =
Adjust TCD Carrier Gas
2 Noise
Sensitivity
Flow Rates
All 450-GCs equipped with TCDs are factory tested at the flow rate listed below.
Carrier Gas:
Helium
Inlet Pressure:
80 psig
Purity:
99.999%
Column Flow Rate:
30 mL/min
Reference Column Flow Rate: 30 mL/min
MAINTENANCE
The Thermal Conductivity Detector is not user serviceable and has no
replacement parts. To condition the TCD, set the detector temperature to 350C
and allow it condition for 4 hours or overnight. The Electronics (filaments) box
must be Unchecked (switched OFF) during this procedure.
Page: 328
Varian, Inc.
The electron capture detector (ECD) is a highly sensitive detector capable of detecting
picogram amounts of specific types of compounds. The high selectivity of this detector
can be a great advantage in certain applications. Compared with the FID, it has much
more limited linear response. The response can vary significantly with temperature,
pressure and flowrate.
The ECD is installed on the detector oven, directly above the column oven. The ECD
electronic Board is installed in the electronics cabinet. The electronics of the ECD are
controlled through the 450-GC keyboard, whereas gas flows are set and controlled at
the pneumatics panel and in the Adjustments section of the Method if Detector EFC is
installed.
An ECD equipped with a detector cell is included in the ECD Kit. Varian is authorized to
distribute this kit to General Licensees, as defined in the Radiation Safety Manual.
The Radiation Safety Manual(s) can be found on the CD-rom CP501425 which is
included in the ECD package.
Due to the NRC regulations on the testing, packaging and labeling of radioactive
materials, additional or replacement ECD detector cells are available only as kits.
Varian, Inc.
Page: 329
Initial Set-up
To set up the ECD, proceed as follows. Note that when the 450-GC is first
powered up all heated zones with the exception of the column oven are
powered off and all detector electronics are turned off.
1. Refer to the Pre-Installation Instructions for additional information on gas
supply requirements.
key on the 450-GC Display/Keyboard to display
2. Press the DETECTOR
the DETECTOR screen. The ECD should be listed in Position Front ( ),
Middle ( ) or Rear ( ). If your GC is equipped with detector EFC verify the
makeup gas you will be using with your ECD.
Page: 330
Varian, Inc.
ECD Adjustments
There are ECD adjustments which should be set before operating the detector.
Cell-Current
In order to obtain the optimum linear and dynamic ranges of response, you
must select the proper ECD cell current. Your choice of cell current depends on
the specific carrier gas that is used and the baseline signal (frequency, f0) with
only carrier gas present. One of 4 values may be selected for the cell current:
In addition, a Zero cell current setting is provided for convenience in setting the
cell contact potential (see below).
For normal operation with nitrogen carrier gas, use the N2STD setting. When a
greater linear range is desired, select the Ar/CH4 setting and use 90%
Argon/10% Methane as the carrier gas.
Set the switch to CAP when using capillary columns. For capillary columns, the
column bleed is low and the baseline is usually below 1 mV when current is set
to N2STD on Range 1.
You may encounter applications in which high column bleed is unavoidable
when using nitrogen as the carrier gas. In such instances, the baseline signal
may be too high. Select the N2HIGH setting in these situations.
In general, the baseline signal should never exceed 25 mV when measured on
Range 1. A value above 25 mV indicates that you have selected the wrong
reference current for the chromatographic conditions (column, phase, sample
matrix). A baseline signal above 25 mV decreases the useful linear range of the
ECD detector.
Varian, Inc.
Page: 331
will be attenuated with a factor 10, the signal value will be 10 times lower then range setting 10.
Page: 332
Varian, Inc.
ECD Operation
Observe the following procedures when operating with an ECD.
Varian, Inc.
The higher the temperature of the foil, the greater the rate of migration of
the 63Ni into the backing material. Rapid migration of the 63Ni reduces the
electron flux and the lifetime of the foil.
Page: 333
To prevent column bleed from fouling the detector, wait for the detector to
reach its operating temperature before you increase the column
temperature.
ECD Pneumatics
The 450-GC is tested at the factory at the flow rates listed in Table 10 on page
26. Use these flow rates as a starting point. Measure the flow rate at the exit
tube of the ECD cell.
You may need to clean the detector cell periodically with carrier gas. Flow rates
for the thermal cleaning procedure are also listed in Table 10 on page 26.
Refer to the installation section for a list of gas filters recommended for packed
column or capillary applications.
Each ECD kit is equipped with a Teflon flow tube assembly. When
setting/checking the flow rate, attach the end of the flow tube with the
ferrule to the 1/16-inch OD exit tube of the ECD cell. Attach the other end
of the flow tube assembly to a flowmeter. To prevent crosscontamination between detectors, use the flow tube only with the ECD
and store the assembly with the ECD kit.
Inlet Pressure
at
Gas Cylinder
Minimum
Purity
Carrier:
N2, He, or Ar/CH4 (90:10)
150 psig
99.999%
Detector Make-Up
(capillary):
N2 or Ar/CH4
150 psig
99.999%
Recommended
Flow Rate
(mL/min)
30 (packed column)
5 (capillary
columns)
20 to 30
Varian, Inc.
Installation/Disassembly
In the ECD Kit (P/N 02-001972-00) the detector cell,
which contains the radioactive ionization source, is
fixed within the detector tower and cannot be
removed. Do not attempt to remove the cell from the
tower. Regulations for the handling of radioactive
materials strictly prohibit unlicensed disassembly.
Evidence of removal voids the warranty. Violations of
these regulations are reported to NRC.
Exploded View
1.
2.
3.
12
2
8
9
6
7
10
11
Figure 56: ECD Exploded View
Varian, Inc.
Page: 335
Page: 336
Varian, Inc.
ECD Installation
Because of the sensitivity of the ECD to oxygen, check the pneumatics
for leaks prior to its installation.
1. Turn the detector oven power OFF.
2. Remove the Detector cover and the GC Top-Cover.
3. Remove both probes (pulser and signal) from the tower arms (see Figure
56, on page 335) of the detector assembly.
Be sure both probes have been removed from the detector tower arms
before attempting to install the detector tower. If you insert the
installation tool before you have removed the probes you will damage
the probe electrode contacts.
4. Check to see whether the detector tower, base and detector cell fittings are
clean. Install the ECD shim on the detector base.
Poor metal-to-metal contact between the tower contact area and the
detector oven can produce a temperature gradient.
Page: 337
9. Carefully install the signal and pulser probes in the tower arms. The signal
probe (bottom) has a twist-lock cable connector and a narrow clip to
engage the collector cylinder. The pulser probe (top) has a threaded cable
connector and a wide clip to engage the foil cylinder. Insert the probes into
the tower arms with the keys on the probes lined up with the notches in the
tower arms. If you encounter any resistance during the insertion of the
probes, check that the probes are being installed in the correct positions.
10. Install the insulated detector tower cap. The label shown is factory installed
on a detector tower registered as a General License Device or Specific
License.
11. Connect the pulser and signal cables to the appropriate probes of the ECD.
Refer to the designation on each cable connector.
12. Cable connections for the operation of the ECD require installation of the
ECD Electrometer Board.
13. Connect the gas supplies at the rear of the GC.
14. Replace the GC Top-Cover and the Detector cover.
Page: 338
Varian, Inc.
15. Install the CAUTION RADIOACTIVE MATERIAL label (P/N 31-00034700) in a location on the front of the GC that is clearly visible. This label
describes the ECD radioactive isotope and the amount of radioactivity.
16. Return the installation tool to the ECD case. You will use this tool each time
you remove the ECD. Retain the ECD case should you need to store the
ECD or to return the detector to Varian. See the Radiation Safety Manual.
Varian, Inc.
Page: 339
MAINTENANCE
Chemical contaminants may deposit on the components of the detector
assembly as the ECD is operated. In many cases, you can remove
contaminants deposited on the radioactive foil or cell ceramic insulators by
heating.
Refer to the Radiation Safety Manual (P/N 03-913999-00) for detector repair,
radioactive decontamination, foil replacement and leak-test information.
Although all heated zones on the 450-GC have a potential upper
temperature of 450 C, Varian recommends that you do not operate the
ECD above 420 C. (The radioactive source temperature is less than
400 C even when the detector set temperature is 450 C).
Page: 340
Varian, Inc.
3. Monitor the output signal while the detector heats to its maximum
temperature. The signal will increase initially, then gradually decrease as
chemical contaminants vaporize from the detector surfaces.
4. Allow the DETECTOR TEMP to remain at 400C until the signal reaches a
stable level. A stable signal indicates that the foil has been cleaned as
much as it can be by this method.
5. Set the detector at normal operating temperatures and flow carrier gas
through the detector overnight. Check and readjust the Contact Potential
setting if necessary.
6. If the performance of the detector is not restored by these cleaning
methods, refer to the Radiation Safety Manual (P/N 03-913999-00) for
information on replacement of the foil.
Varian, Inc.
Page: 341
Page: 342
Varian, Inc.
Description
The Pulsed Flame Photometric Detector (PFPD) is the newest member of the
family of flame-based gas chromatographic detectors. The PFPD operates in a
pulsed-flame rather than continuous-flame mode and offers a number of
significant improvements over the conventional flame photometric detector:
Increased detectivity.
Varian, Inc.
PFPD assembly.
PFPD pneumatics.
Page: 343
Page: 344
Varian, Inc.
Item
Description
Function
Igniter cap
Detector tower
Quartz combustor
Detector base
Light pipe
Combustor support
10
11
12
Air2 inlet
13
14
15
16
17
Varian, Inc.
Page: 345
Description
Function
J1: Electrometer input signal cable BNC connector Photomultiplier signal output is directed to the
electrometer board through this BNC connector.
J2: Photomultiplier tube high voltage cable TNC
High voltage source on the electrometer board is
connector
directed to the photomultiplier tube through this
TNC connector.
J3: Igniter cable Molex connector
Current to the igniter coil is directed through this
Molex connector.
SIGNAL (TP1): Test point
With an oscilloscope, used to set the Gate Width
and Gate Delay.
TRIGGER (TP2): Test point
Used to monitor the gate trigger (threshold).
GATE (TP3): Test point
Used to monitor when the integration period
(Gate Width) begins.
J4: Top
These are the same signals provided at TP1, TP2
and TP3, organized to provide easy connection to
Signal
a cable. Used to connect to the PFPD Analytical
Trigger
Software.
Gate
Ground
Page: 346
Varian, Inc.
Varian, Inc.
Fill: Air and hydrogen mix and enter the combustion chamber at two points.
Part of the combustible gas stream mixes with the column effluent and
moves upward through the interior of the quartz combustor tube. A second
part of the gas stream sweeps along the outside of the quartz combustor
tube and into the igniter chamber.
Emit: During and after flame propagation, the sample atoms of interest
undergo further reaction to form electronically excited species, from which
light is emitted. The flame background emission is complete within less
than 3-4 milliseconds after propagation, whereas phosphorus and sulfur
molecular species emit over a much longer time. This difference in
emission times yields the enhanced detectivity and selectivity of the PFPD.
Page: 347
How Does the PFPD Achieve its Greater Detectivity and Selectivity?
The enhanced detectivity and selectivity of the PFPD are a consequence of two
main factors:
Separation (in time) of the specific element emission from the background
emission.
Page: 348
Varian, Inc.
Varian, Inc.
Page: 349
Page: 350
Varian, Inc.
The air:hydrogen mixture enters the PFPD at two points: around the outer wall
and through the center of the quartz combustor tube. The wall flow is used to
refill the upper igniter chamber of the detector after each pulse, while the
combustor flow mixes with the column effluent and refills the combustor. For
proper operation, the combustor should refill just slightly faster than the igniter.
The refill rates are balanced by use of a needle valve on the air:hydrogen
mixture and by the flow of Air2, which bypasses the combustor and goes
directly to the igniter.
Varian, Inc.
Page: 351
References
Refer to the following articles for additional information on the PFPD.
H. Jing and A. Amirav, Pesticide Analysis with the PFPD and a Novel
Direct Sample Introduction Device, Anal. Chem., 69, 1426-1435 (1997).
Specifications
Specification
Detectivity
Sulfur Performance
1 pg S/sec
Phosphorus Performance
0.1 pg P/sec
Selectivity
106 S/Carbon
NOTE: The selectivity of Sulfur/Carbon is a
function of the sulfur concentration at any
time (from 105 near its detection limit to 106
at higher sulfur concentrations).
106 P/Carbon
Linear Dynamic
Range
Drift (in 20 minutes)
103
105
Page: 352
Varian, Inc.
20 pg N/sec
Selectivity:
>105 N/C
>102
Drift:
Varian, Inc.
Page: 353
Installation
Inspecting the PFPD
The PFPD is installed and tested at the factory before the GC is shipped.
Carefully inspect the PFPD. If there is evidence of damage or if parts are
missing from the PFPD or its accessory kit, please contact your nearest Varian,
Inc. subsidiary or Varian, Inc. representative.
Item
Part Number
Qty
Description
Teflon combustor extraction device
(wide bore, 3 mm ID)
Teflon combustor extraction device
(narrow bore, 2 mm ID)
03-925180-00
03-925181-00
03-925192-00
29-898500-00
Teflon tweezers
03-925194-04
15-003347-00
15
03-925138-00
10
88-189610-00
16-000327-00
Page: 354
Varian, Inc.
Please note that Item 16, Sulfur Linearization Software, is not used with the
450-GC PFPD. Several important parts for Sulfur mode include Items 10, 11
and 12 while parts for Phosphorus mode include Items 13, 14 and 15.
Item
Part Number
Qty
Description
10
03-925176-00
11
03-925178-00
12
03-925151-01
13
03-925177-00
14
03-925179-00
15
03-925151-02
16
1
5
4
3
ltem
Part Number
Qty
Description
03-925132-00
Adapter seal
03-925195-00
03-925196-00
03-949786-00
Spring
03-925200-00
88-189610-00
Page: 355
Description
Capillary Column Nut
Capillary Column
Ferrule
Correct Column Cut
Incorrect Column Cut
Magnifying lens.
Follow these steps to prepare the end of the chromatographic column for
insertion into the PFPD.
1. Before cutting the end of the column, slide the capillary column nut (Item 1)
over the fused silica column end (Item 2).
2. Install the appropriate Vespel/Graphite or graphite ferrule (Item 3) over
the column end.
Make certain that the tapered edge of the ferrule is toward the inside of the
column nut (or away from the column end).
Page: 356
Varian, Inc.
Varian, Inc.
Phillips-head screwdriver.
Metal forceps.
Teflon tweezers.
Lab tissues.
Page: 357
8. Gently insert the appropriate Teflon combustor extraction device into the
opening in the top of the combustor. Wait 2-3 seconds for the Teflon
combustor extraction device to seat, then lift the quartz combustor from the
combustor support (Item 8). Place the quartz combustor on a clean surface
(e.g., a clean lab tissue).
9. Use the metal forceps to remove the top aluminum crunch washer (Item 13)
from the detector tower. Discard this used crunch washer.
Page: 358
Varian, Inc.
Varian, Inc.
Page: 359
Phillips-head screwdriver.
5/16" wrench.
Magnifying lens.
Varian, Inc.
contacts the detector tower. See the following figure. Tighten the column
nut to secure the column in this position.
3. Continue with step 4 in Reassemble the PFPD.
Detector Tower
Column
Combustor
Support
Varian, Inc.
Phillips-head screwdriver.
Metal forceps.
Page: 361
1. Use the clean metal forceps to place a new bottom aluminum crunch
washer (Item 14) on the detector base.
2. Hold the PMT assembly (Item 2) with one hand and the detector tower with
the other. Sight the top of the combustor support (Item 8) through the
detector tower. Keep the combustor support centered as you lower the
detector tower (Item 4) over the support.
Be very careful not to crush the end of the column.
3. Reinstall the 2 Allen screws (Item 16) that hold the detector tower to the
detector base (Item 6). Tighten the screws alternately. Note: If installing the
column with the aid of the column depth gauge, follow instructions on page
360.
4. Use the Teflon combustor extraction device (of the correct taper) to insert
the quartz combustor (Item 5).
5. Use the clean metal forceps to place a new top aluminum crunch washer
(Item 13) on the detector tower.
6. Replace the igniter cap (Item 3) on the detector tower. Use the 9/64" Allen
wrench to reinstall in the top of the igniter cap the 2 Allen screws (Item 15)
that secure the igniter cap to the detector tower.
7. Reinstall the Detector cover on the top of the GC.
8. Turn on the PFPD electronics and PFPD Oven Power. Be sure that the
detector gases are turned on. The detector may not begin pulsing normally
until the detector tower has begun to heat up somewhat. Allow the detector
to come to temperature and equilibrate for at least 15 minutes. During this
equilibration period the polyimide coating on the portion of the column
exposed to the flame is burned off, leaving a clean fused silica surface.
9. Did you install a quartz combustor with a different diameter?
Page: 362
Varian, Inc.
Operation
Adjusting the Gas Flow Rates
The following procedure describes the adjustment of the flow rates of the gases
to the PFPD. If you frequently adjust the column flow rate or change the carrier
gas, you may want to calibrate the air and hydrogen flow controllers. Once the
flow controllers are calibrated, you can quickly select new flow rates and use
the optimization tables that are provided.
If the PFPD Detector EFC option is installed on your 450-GC, detector gas
flows are set by pressing the Detector key, selecting the PFPD tab.
The Split Valve adjustment is described below.
Split Valve
A split valve is used to finely adjust the flow rate through the combustor section
of the PFPD combustor. Follow these directions when adjusting this split valve:
1. Disconnect the outlet line from the needle valve and replace with a
flowmeter.
2. With Air1 and Hydrogen turned on, close the needle valve by turning the
knob fully clockwise. The flowmeter should indicate a flow of less than
1 mL/minute. If this is the case, disconnect the flowmeter and reattach the
outlet line.
3. If the flow from the needle valve exceeds 1 mL/minute, remove the knob by
loosening the set screw. Rotate the valve stem clockwise gently until the
flow from the outlet drops below 1 mL/minute. Do not overtighten the valve
stem, as the needle valve does not need to shut off completely to function
properly and overtightening can damage the delicate needle in the valve.
After making the adjustment, reinstall the knob on the valve stem and
secure with the set screw.
Varian, Inc.
Page: 363
Page: 364
Item
Description
Split valve
Igniter cap
Varian, Inc.
Stopwatch
For best stability of flows, flow controllers should have pressure applied
to them for at least 24 hours before adjustments are made. A moderate
flow of gas through the flow controller during this conditioning period is
also helpful in stabilizing the flow controller.
1. Turn the igniter OFF.
2. Locate the Air1 mass flow controller valve (Item 1). Note the current dial
setting. Turn Air1 OFF.
3. Locate the Air2 mass flow controller valve (Item 2). Note the current dial
setting. Turn Air2 OFF.
4. Locate the Hydrogen mass flow controller valve (Item 4). Note the current
dial setting. Turn Hydrogen OFF.
5. Install the flow measuring adapter (Swagelok 1/8" to 1/16" reducing union)
to the threaded hole in the igniter cap. Connect a 2-3" piece of tube to 1/8"
end of the union. Use Teflon thread tape to make the connection to the
igniter cap gas tight. Measure and record the column flow rate, then turn
the column flow OFF.
A Varian Intelligent Digital Flowmeter is highly recommended for this
procedure.
6. Adjust the Air1 mass flow controller valve until flow rate is near 17 mL/min.
Write down and save the dial setting for the Air1 mass flow controller valve.
Turn the Air1 flow controller Off.
Varian, Inc.
Page: 365
7. Adjust the Air2 mass flow controller valve until flow rate is near 10 mL/min.
Write down and save the dial setting for the Air2 mass flow controller valve.
Turn the Air2 flow controller Off.
8. Adjust the Hydrogen mass flow controller valve until flow rate is near
12 mL/min.
If you will be using hydrogen as your carrier gas, the total hydrogen flow
to the detector should be approximately 12 mL/minute. The flow rate set
in this step should thus be 12 mL/min minus the anticipated column flow
rate. Write down and save the dial setting for the Hydrogen mass flow
controller valve.
9. Restore the Air1 and Air2 flow controllers to the settings recorded above.
Restore the carrier gas flow to the column. Remove the flow measurement
adapter, remove any Teflon thread tape residue from the igniter cap.
Leave the flow controllers ON at all times unless you must turn them off
to re-plumb or reconfigure the gas chromatograph. If you turn the flow
controllers off for an extended period then on again, they must
equilibrate overnight before the PFPD can achieve optimum
performance.
10. Turn the PFPD and thereby the igniter ON.
Stopwatch.
1. Turn off the column flow. Or, if you want flow to continue to the column,
disconnect the column from the detector and use a column nut and 1/16"
no-hole ferrule to seal the detector.
2. Write down the current dial settings for Air1, Air2 and Hydrogen, then turn
off each flow controller.
3. Install the flow measuring adapter (1/8" to 1/16" reducing union). Use
Teflon thread tape to make a gas-tight seal.
Page: 366
Varian, Inc.
Varian, Inc.
Page: 367
Detectivity
Detectivity is the minimum detectable quantity (MDQ) of Sulfur or Phosphorus
(usually expressed as picograms of Sulfur (or Phosphorus) per second.
Detectivity is the preferred method of expressing the sensitivity of a detection
method because it includes a consideration of the Signal-to-Noise ratio.
Sensitivity alone is insufficient in that the sensitivity of the PFPD can be
increased simply by increasing the voltage of the photomultiplier tube, but the
noise increases as the signal increases, yielding little enhancement in the
minimum amount detectable.
Selectivity
An important question when optimizing the selectivity of the PFPD is, "Selective
relative to what?" In some situations, the Sulfur or Phosphorus response is
best optimized relative to Carbon (the principal background emission in the airhydrogen flame). At other times, it may be more important to optimize the
Phosphorus response relative to the Sulfur response, as when Phosphorus is
analyzed in certain plant tissue extracts that are high in Sulfur content.
Throughput
The additional selectivity of the PFPD can result in shorter analysis times and
thereby greater sample throughput. As an example, by optimizing the sulfur
response relative to hydrocarbons, the PFPD may detect fewer peaks thereby
simplifying the chromatogram and increasing the chromatographic resolution.
With greater resolution, the chromatography can be adjusted such that the
peaks of interest elute faster thereby decreasing the analysis time. In addition,
when hydrogen is used as the carrier gas, the PFPD can be operated at higher
flow rates without greatly affecting the detectivity of the method. Higher carrier
gas flow rates translates into shorter analysis times and greater sample
throughput.
Page: 368
Varian, Inc.
H2
(mL/min)
Air1
(mL/min)
Air2
(mL/min)
13
17
10
13
20
10
10
20
34
10
11
17
10
18
10
10
15
31
10
14
17
10
16
21
10
10
24
34
10
Hydrogen
Nitrogen
Varian, Inc.
Page: 369
Element
Sulfur
Gate Delay
6
Gate Width
20
Trigger
200
Phosphorus
10
200
3.5
200
Nitrogen
Optimization Criteria
Refer to the following table when optimizing the PFPD.
Page: 370
Varian, Inc.
To Maximize ...
Consider ...
Phosphorus
response
Set hydrogen carrier gas flow rate to a value that is near, but
just above, the minimum of the van Deemter curve for the
column and maximum column temperature used.
Sulfur response
Adjust split valve such that detector is just out of tick-tock, then
turn counterclockwise an additional 1/2 turn.
Varian, Inc.
Page: 371
To Maximize ...
Consider ...
Sulfur-to-Carbon
selectivity
Adjust split valve such that detector is just out of tick-tock, then
adjust counterclockwise an additional 1/2 turn.
The trigger level should be set to trigger on the main flame peak.
Setting the trigger level correctly prevents the hydrocarbon and
sulfur responses from overlapping as the flame propagation slows
in hydrocarbon-rich flames.
The Sulfur response is proportional to the square of the Sulfur
concentration in the flame whereas the Carbon response is directly
proportional to the Carbon concentration. Therefore, when the
amount of sample that is injected is doubled, the Sulfur-Carbon
selectivity doubles (all other parameters being equal).
Phosphorus-toSulfur selectivity
Page: 372
Varian, Inc.
To Maximize ...
Analytical
throughput
(decrease analysis
time)
Consider ...
Varian, Inc.
Page: 373
To Maximize ...
Consider ...
200 C
300 C
Page: 374
Varian, Inc.
Establishing Tick-Tock
Tick-tock occurs when the combustible mixture ignites in the igniter chamber
but does not propagate into the combustor chamber because the combustor
has not yet filled with a combustible gas mixture. Optimum performance is
generally achieved when the detector is operating near, but not in, the tick-tock
mode.
Establishing Tick-Tock and subsequently the proper setting of the Air-Hydrogen
needle valve is accomplished using either the scope mode of the PFPD
Analytical Software or the procedure outlined below. The PFPD Analytical
Software is available for desktop (03-925943-92) or laptop PCs (03-92594391).
Follow these steps to adjust the gas flow rates for optimum performance of the
PFPD. This adjustment should be done with the column oven at the highest
temperature which will be used in the analysis. This will assure that the detector
will remain in the normal operating mode throughout the column temperature
program.
1. To monitor the detector baseline, connect a strip chart recorder or PC to the
detector analog output. Turn off the AutoZero function (in the Detector
section of the GC method). Set the Gate Delay to 4 msec and the Gate
Width to 10 msec. Leave the trigger level at the default value, 200 mV.
(This makes Tick-Tock easier to observe. Remember to switch back to the
desired gate delay and width settings after setting the needle valve.)
2. Adjust the gas flow rates so that the PFPD is pulsing at about 2-4 Hz
(pulses/second).
3. Adjust the Air-Hydrogen needle valve fully clockwise. This shuts off the flow
of air-hydrogen to the combustor chamber while permitting the air-hydrogen
mixture to pass around the outside of the combustor and enter the igniter
chamber. The pulse rate should increase. As the flame stops propagating
into the combustor, the detector baseline drops and becomes smoother
(less noisy).
At low column flow rates (particularly with the 3 mm combustor), there
may be enough diffusion of fuel mixture into the combustor chamber
that the baseline is not completely smooth (noiseless).
4. Slowly turn the Air-Hydrogen needle valve counterclockwise. With each turn
of the needle valve, more of the combustible mixture enters the combustor
chamber. As the amount of combustible mixture entering the combustor
chamber increases, the flame begins to propagate into the combustor
chamber. First, the flame propagates into the combustor chamber on every
third or fourth pulse and then, as the flow rate of the combustible mixture
increases, on alternate pulses (tick-tock mode) and then finally on every
Varian, Inc.
Page: 375
pulse. As you increase the flow rate of combustible mixture that enters the
combustor, the baseline should rise sharply and become very noisy as the
detector enters tick-tock. As the detector leaves the tick-tock mode, the
baseline will rise and then fall and become less noisy. The pulse rate will
also decrease as you pass through tick-tock, since more of the combustible
mixture is burned on each pulse as you adjust the needle valve for normal
operation.
5. Once the baseline reaches a steady value, you may adjust the needle valve
an additional 1/2-turn counterclockwise to ensure stable operation of the
PFPD. If you changed the gate delay and width, reset them to the proper
values for your analysis.
Page: 376
Varian, Inc.
Varian, Inc.
Page: 377
Within 10-20 volts of the nominal value for the desired mode of
operation, you may not need to make any further adjustment.
Not within this range, enter a new PMT voltage while observing the
PFPD output on a strip chart recorder or on the Workstation System
Control display.
4. If you want to optimize the detector sensitivity with respect to PMT voltage:
Page: 378
Set up the PFPD to operate in the desired mode. Set the detector to
Range 10.
Slowly adjust the PMT setting while watching the baseline noise
(displayed on a strip chart recorder or other output device).
Varian, Inc.
10 L syringe.
Page: 379
Page: 380
Varian, Inc.
Varian, Inc.
Page: 381
6. If necessary, adjust the Air1 and Hydrogen flow rates to give a flame pulse
rate 2-4 pulses/second. Alternately increase or decrease Air1 and
Hydrogen so that the flow rate remains approximately the same as the gas
composition is changed. Allow the system to equilibrate for 2-3 minutes
between flow rate adjustments so that the air and hydrogen filters are
completely flushed with the new gas composition.
When adjusting the flow rate, turn the Air and Hydrogen flow controller
dials in increments of 10 or 20, or make 0.2-0.5 mL/min adjustments
using PFPD Detector EFC.
7. Increase the flow rate of Hydrogen by 10 dial units. Examine the emission
on the stripchart recorder. Monitor the emission after each flow rate
adjustment. If the sulfur emission increases, continue to increase the
hydrogen flow rate until the response no longer increases. If the pulse rate
increases beyond 4 Hz, reduce the Air1 flow by 10 units and continue the
procedure. Typically, noise is relatively independent of gas composition so
that the relative gas composition can be adjusted with little or no effect on
the noise.
Sulfur emission is maximized when the gas mixture is very rich in
hydrogen.
8. Once the sulfur emission has been maximized, adjust the fill time of the
igniter volume so that it is just slightly less than the fill time of the combustor
volume. See Establishing Tick-Tock on Page 375.
9. Remove the column from the diffusion device holder, remove the diffusion
device holder from the Injector outlet and reinstall the test column or the
appropriate analytical column in the Injector.
10 L syringe.
Page: 382
Varian, Inc.
Set the Gate Delay and Gate Width for the element of interest.
Disconnect the test column, cap off the detector and operate the GC for at least
4 hours to bake out the column. Set the oven temperature to room temperature
and wait until the oven and column are cool. Reconnect the test column to the
detector.
4. Set the GC method (or GC Workstation method) for the following test
conditions:
Set PMT voltage to 510V for phosphorus or 550V for sulfur. If you are using the
R5070 (N mode) PMT, set voltage to 600V or use values obtained in on page
378.
5. Operate the GC for at least 1 hour to ensure that the PFPD and its
electronics are thermally stable. Re-optimize the PFPD as recommended in
Optimizing the PFPD on Page 368.
Varian, Inc.
Page: 383
6. If you are using a split injection, inject 1 L of FPD test mix with a split ratio
of 20:1. If you are using a non-splitting Injector, dilute the FPD Test Sample
1:20 in iso-octane and inject 1.0 L of the diluted FPD Test Sample.
7. The resulting chromatographic peaks should have similar retention times,
heights and resolution to those shown below.
8. Are the peak retention times substantially different?
Page: 384
If Yes, adjust the column temperature and/or column flow rate and
repeat steps 5-6.
If No, replace the test column with the column to be used in your
particular application, adjust the GC conditions to appropriate values
and proceed with the analysis.
Varian, Inc.
Varian, Inc.
Page: 385
Detectivity
Refer to the instructions below to calculate the detectivity for sulfur and
phosphorus with the PFPD:
Page: 386
Varian, Inc.
Page: 387
Page: 388
Varian, Inc.
Trigger Level: signal required to start (trigger) the delay and width timer.
Gate Width: time in milliseconds after the Gate Delay during which the
detector signal is integrated.
The appropriate values for the Gate Width and the Gate Delay are determined
from the emission profile (time x emission intensity) of the element of interest.
Appropriate values for a number of elements are listed in the following table.
Preset
Range
Trigger Level
200 mV
10 - 2000 mV
Gate Delay
4 msec
0.5 - 20 msec
Gate Width
10 msec
0.5 - 20 msec
Varian, Inc.
Page: 389
10 L syringe.
The trigger level is preset to 200 mV. You may need to set the trigger level
lower if you are using a narrow band interference filter which does not transmit
much of the flame background. You may need to set the trigger level higher
than 200 mV for optimum operation with some elements which have no delayed
emission, or very short emission delay times.
To set the trigger level, follow the steps below:
1. Press the Detector key and select the PFPD you wish to adjust. Turn OFF
initial Autozero in the active method.
2. Press the Adjustments button to access the integrator parameters. Set
Gate delay to 3.5 ms. This allows the tail of the prompt flame background
emission to be integrated as a test signal. This figure shows a typical flame
background emission profile.
Trigger Level
OK
NOISY
OK
Page: 390
Varian, Inc.
3. Set the trigger level to the minimum and observe the noise on the detector
output using a strip chart recorder or the Star Workstation status display.
Gradually increase the trigger level. The signal level should decrease and
the noise should also decrease. Select a trigger level which gives minimum
noise. This will usually be triggering on the preliminary emission pulse.
4. If you will be detecting an element such as nitrogen, which emits very
promptly after the main flame propagation, you may get better results by
continuing to increase the trigger level until triggering is occurring on the
main background emission pulse itself. As you increase the trigger level,
the noise will increase substantially at one point. This is due to the trigger
occurring sometimes on the apex of the preliminary emission pulse and
sometimes on the main pulse. Increasing the trigger level 10 to 20% above
this point should provide very stable triggering for detection of prompt
emitting elements. This setting is also preferred when the effects of large
amounts of hydrocarbons on emission timing must be minimized. The
following figure illustrates the signal observed during such an adjustment.
Varian, Inc.
Page: 391
Element
Gate Delay
Gate Width
Sulfur
20
Phosphorus
10
3.5
Nitrogen
Varian, Inc.
8. If the desired selectivity has not been achieved, increase the Gate Delay by
0.5 msec and inject the standard to check the effect of this change.
Continue to increase the Gate Delay until the desired selectivity is
achieved.
In the case of elements which emit very near the flame background peak,
selectivity may be improved by opening the combustor flow needle valve
somewhat beyond the position determined in the procedure Establishing
Tick Tock. An increase of 1/4 to 1/2 turn counterclockwise is usually
sufficient.
10 L syringe.
Varian, Inc.
Page: 393
Building a GC Method
The procedure for building a GC method involves 4 tasks:
Page: 394
Varian, Inc.
1. Open the method to which you wish to add the PFPD section. Select the
450-GC section and select the Detector page. Select the PFPD tab.
2. Set the Electronics and Detector Oven to ON.
3. Set the Detector Temperature to an appropriate value for your analysis.
4. Set the Range and Autozero as required. Note that these parameters are
time programmable.
5. Set the Trigger level, Gate Delay and Gate Width as required.
6. If you wish to use the Analog Output in addition to the digital signal used by
the Star Workstation, select the Analog Output page, select the output
channel you wish to use and select the PFPD position and attenuation.
Varian, Inc.
Page: 395
Quantitation
Quantitation with the PFPD
When using the PFPD, it is important to understand how its pulsing flame can
affect the accuracy and precision of a quantitative analysis.
As described previously, the combustor and igniter chambers of the PFPD fill
with a combustible mixture of air and hydrogen. The combustor chamber fills
with the combustible mixture at about the same time as the igniter chamber.
When the gas mixture enters the igniter, the mixture ignites and the flame
propagates back through the combustor and then extinguishes. The combustor
then begins to fill with the gas mixture and the process repeats. The rate of this
repetitive process can vary, but for optimum performance of the PFPD, you
must adjust the gas flow rates in order to strike a balance between the fill time
of the combustor chamber and the fill time of the igniter chamber. It is important
that the combustor chamber fill slightly before the igniter chamber in order to
ensure that the flame propagates completely back to the combustor chamber.
However, the combustor chamber should not fill too fast (relative to the igniter
chamber) or some of the sample exits the detector before it is combusted
resulting in poorer detectivity.
Because the PFPD pulses, the emission derived from the PFPD appears as
discrete signals separated in time. This requires electronics that are quite
different from the conventional non-pulsed FPD in which the emission signal
increases then decreases without interruption. That is, the PFPD electronics
are not sampling discrete points on the eluting peak, as in the case of a
conventional FPD. Rather, the PFPD stores the emission output derived during
each pulse and generates a peak (or waveform) that may or may not be
identical to the actual eluting peak shape, but which contains all the signal
information from the chromatographic peak.
Given these unique properties of the PFPD, there are differences between
quantitation based on the height or area of the chromatographic peaks.
Page: 396
Varian, Inc.
Varian, Inc.
Page: 397
Page: 398
Varian, Inc.
Advanced Operation
This section describes the operation of the PFPD in the Nitrogen mode,
including:
Installation of the nitrogen mode parts - for users who are adding the
nitrogen mode capability to an existing PFPD. Includes an exploded view of
the parts.
Theory of Operation
Nitrogen mode is based upon combustion of nitrogen containing compounds to
form NO. NO then reacts with hydrogen atoms remaining in the combustor after
the flame is extinguished to form HNO*, an electronically excited molecule
which emits light in the 700 - 800 nm range of the near infrared. Because these
emission wavelengths are beyond the range of the photomultiplier tube used in
the standard PFPD, the Nitrogen mode requires a new red-sensitive
photomultiplier and PMT holder hardware, along with an appropriate dark red
colored glass filter.
Because N mode is based on complete combustion of the sample molecule to
form HNO*, the PFPD response is much less dependent on the chemical
nature of the sample than is the TSD response, which depends on the
presence of a C-N bond in the molecule. Thus even inorganic nitrogen
compounds such as ammonia, hydrazine and nitrogen oxides are detectable by
the N mode PFPD, whereas these compounds give no response on the TSD.
The only nitrogen compound known to give no useful response on the PFPD is
molecular nitrogen, N2. Compounds such as nitroglycerin and other nitrate
esters, nitroaromatics and nitroso compounds give a somewhat enhanced
response when compared to more reduced nitrogen compounds, but the
enhancement amounts to only ~20%, so for most purposes the N mode PFPD
can be considered to give a uniform response to all nitrogen compounds, as it
Varian, Inc.
Page: 399
Helium
Nitrogen
Varian, Inc.
6. Press the Detector key, select the PFPD and press the Adjustments soft key.
Set a Gate Delay of 3.5 msec, a Gate width of 8 msec and a PMT Voltage of
600 volts.
Item
1
2
3
4
5
6
7
8
9
10
Description
Filter Mount
O-ring
O-ring, PMT Mounting
PMT Housing
Set Screws
O-ring
PMT
Socket
Retainer
Connectors with Mounting Plate
Varian, Inc.
Page: 401
Varian, Inc.
Description
03-925119-01
03-925119-02
03-925128-00
Varian, Inc.
Page: 403
Multi-element Detection
In some applications, it may be desirable to detect more than one heteroatom
simultaneously. Several combinations which may be of interest are: Sulfur plus
phosphorus mode use a WG345 filter with the standard PMT or WG345 +
RG39 filters with the R5070 PMT if nitrogen interference is undesirable. Use a
gate delay of 4 msec and adjust gate width to balance sulfur and phosphorus
response.
Sulfur + phosphorus + nitrogen mode WG345 filter with
R5070 PMT. This combination will also detect Sn, Se and As.
Sulfur + Nitrogen mode R5070 PMT with BG3 filter for
suppression of phosphorus response.
Emission Wavelengths
Arsenic exhibits a very broad emission spectrum, extending from 360 to
800 nm. The choice of the wavelength at which to observe As emission is
determined largely by the emission characteristics of interfering elements.
Emission Timing
Arsenic emission is longer delayed than sulfur, with a correspondingly longer
emission duration. Use the Gate Delay and Gate Width recommended for
sulfur, or increase delay even further for additional selectivity against nitrogen
and phosphorus. A 3 mm combustor provides best sensitivity.
Flame Composition
Delayed arsenic emission is optimized in hydrogen rich conditions similar to
those used for sulfur. Arsenic may also be detected in an air-rich flame, as
described below.
Page: 404
Varian, Inc.
Filters
Arsenic may be detected at a variety of wavelengths, depending upon the
requirements of the analysis. Good detectivity (~6 pg/second As) and selectivity
are achieved using an RG9 (nitrogen) or RG695 filter with the Nitrogen mode
R5070 photomultilplier tube. Interference by P and N compounds can be
reduced by using a long gate delay time. Tailing of the delayed emission can be
reduced by operating the detector at the highest practical temperature, or by
operating below 200 C.
An alternative approach to arsenic detection uses an air-rich flame, a GG495
(phosphorus) filter and a very short gate delay. This air-rich mode gives very
good detectivity, ~3 pg/second As. and very good selectivity against sulfur
interference, but at the expense of reduced selectivity against hydrocarbon
interference. This air-rich mode also eliminates the tendency to peak tailing
observed with the hydrogen-rich flame emission.
Other Considerations
Arsenic detection provides the most opportunities for optimization of any of the
common PFPD-detectable elements. Selectivity can be optimized by the choice
of air-rich or hydrogen-rich flame conditions, by the choice of gate delay and by
the choice of optical filter. In some applications, it may even be desirable to
reduce the selectivity among heteroatoms and instead look to selectivity only
against hydrocarbons. For example, the ability to detect simultaneously arsenic,
phosphorus, sulfur and nitrogen, while totally rejecting hydrocarbon response,
is extremely important in the detection of chemical warfare agents for treaty
verification purposes.
Emission Wavelengths
Selenium exhibits a fairly broad emission spectrum in the PFPD, extending
from ~360 to ~550 nm, attributed to emission of the Se2 molecule.
Varian, Inc.
Page: 405
Emission Timing
Selenium emission is somewhat longer delayed than sulfur, with a
correspondingly longer emission duration. Use the Gate Delay and Gate Width
recommended for sulfur.
Flame Composition
Selenium emission is optimized in cool, hydrogen rich conditions similar to
those used for sulfur. Since emission is due to Se2, use the 2 mm combustor
and keep all gas flows as low as practical to minimize dilution of the selenium
concentration in the flame.
Filters
Selenium may be detected using either the BG12 (sulfur) or the GG495
(phosphorus) filter. Best detectivity, ~10 pg/second Se, is achieved using the
BG12 filter, but this also gives the most interference from sulfur compounds.
Detectivity with the phosphorus filter is ~20 pg/second Se, but sulfur
interference is considerably reduced. Either filter gives essentially complete
specificity against hydrocarbon interference.
Other Considerations
Since selenium emission depends upon recombination of Se to form
electronically excited Se2 molecules, use the lowest detector temperature which
gives satisfactory chromatographic peak shapes. Selenium also exhibits a
prompt emission which may be used to distinguish it from sulfur.
Emission Wavelengths
Tin exhibits a fairly broad emission spectrum in the PFPD. A broad band
between ~380 and 540 nm is attributed in the literature to interaction between a
tin species and the quartz surface. The sharper bands at 610 and 695 nm have
been attributed to SnH emission. The relative and absolute amplitudes of these
bands may be a function of the PFPD flame composition.
Page: 406
Varian, Inc.
Emission Timing
Tin emission is somewhat longer delayed than phosphorus and emission
duration is intermediate between phosphorus and sulfur. Use the phosphorus
gate delay and width for best sensitivity; use the sulfur gate delay and width for
best selectivity against carbon interference.
Flame Composition
Tin emission is optimized in hydrogen rich conditions similar to those used for
phosphorus or sulfur. Increasing Air1, Air2 and hydrogen by 50 - 100% each
will help to reduce peak tailing. Either the 2 or 3 mm combustor may be used,
with a slight preference for the 3 mm.
Filters
Tin may be detected using either the BG12 (sulfur) or the OG590 filter. Best
detectivity, ~100 fg/second Sn, is achieved using the BG12 filter, but this also
gives the most interference from high levels of sulfur compounds. Detectivity
with the OG590 filter is ~1 pg/second Sn, but sulfur response is considerably
reduced. Response with the OG590 filter is best with the red sensitive R5070
photomultiplier tube. A 610 nm interference filter has also been found useful in
conjunction with the R5070 PMT.
Other Considerations
Tin compounds exhibit tailing in the PFPD, probably due to limited volatility of
the tin combustion products. This tailing can be reduced by using higher gas
flow rates and by raising the detector temperature to 350 C to help volatilize
the tin compounds. You may also find it helpful to insulate the detector tower
using the PFPD tower cover, or by wrapping the tower with fiberglass cloth,
being careful not to cover the outlet.
An alternative approach to tin tailing is to reduce the detector temperature as
much as possible without impairing the chromatography. This reduces the tail
amplitude, but spreads it out in time. Several minutes may be required for the
detector to clean itself of tin deposits. Refer also to: J. A. Jacobsen, F. StuerLauridsen and G. Pritzl, Organotin Speciation in Environmental Samples by
Capillary Gas Chromatography and Pulsed Flame Photometric Detection
(PFPD), Appl. Organometallic. Chem. 11, 737-741 (1997).
Page: 407
Emission Wavelengths
Manganese exhibits a strong atomic emission line at 403 nm, along with other
weaker lines and bands in the 380 - 600 nm range. The 403 nm line is the only
emission of analytical interest with the PFPD.
Emission Timing
Manganese shows no delayed emission, as expected for atomic emission. Use
the minimum delay time and a gate width of 2 to 4 msec, depending on your
flame propagation rate.
Flame Composition
Manganese emission is favored by a hot flame, which is achieved by a
stoichiometric or slightly air rich air1/hydrogen mixture and minimum dilution of
the flame by carrier gas flow. The manganese emission is not very sensitive to
the exact gas composition, so you may vary the composition somewhat to
enhance selectivity against other elements. Emission is enhanced by the use of
the 3 mm ID combustor.
Filters
Because the manganese emission is not separated in time from the flame
background emission, a narrow band interference filter is required to provide
adequate selectivity against hydrocarbons and other interfering elements. A
404.7 nm interference filter with 10 nm bandwidth was used in Prof. Amiravs
work. This filter is commonly available as part of a set of filters for the isolation
of mercury emission lines and may be obtained from many sources of
interference filters. Using this filter, detectivity of ~100 fg Mn/second and a
selectivity against hydrocarbons of ~104 have been observed by Prof. Amirav.
Other Considerations
Manganese containing peaks show some tailing due to incomplete clearance of
the manganese from the combustor. This tailing is minimized using nearstoichiometric flame conditions. It may be further reduced by using a detector
temperature below 225 C, but this may result in long term buildup of
manganese deposits in the combustor, although this was not observed by Prof.
Amirav.
Page: 408
Varian, Inc.
Item
1
2
3
4
5
6
7
8
Description
Ignitor cap
Photomultiplier tube (PMT) high voltage cable
Photomultiplier tube (PMT) signal cable
Photomultiplier tube (PMT) housing
Light pipe assembly
Glass filter
Filter holder
O-ring
Varian, Inc.
Page: 409
Phillips-head screwdriver.
Lab tissues.
Capillary pick.
Handle the filter holder by its outside edge. Do not touch the optical
surface of the filter.
Replace the Optical Filter
1. If you are cleaning the optical filter, remove the C-clip holding the filter in
the filter holder. Moisten a clean lab tissue with methanol or iso-propanol.
Hold the filter by its outer edge and gently wipe both surfaces of the filter
with the lab tissue. Wipe away the excess solvent with a clean dry lab
tissue. Place the filter in the filter holder and reinstall the C-clip.
2. Hold the filter that is to be installed by its outside edge and slide it into the
recess in the end of the PMT housing. Check that the filter is correctly
seated so that its surface is flush with the rim of the PMT housing and the
O-ring (Item 8) is properly seated in its groove.
3. Carefully join the PMT housing to the light pipe assembly and install and
tighten the 2 thumbscrews.
Page: 410
Varian, Inc.
For column flow rates over 5 mL/min, use the wide bore quartz
combustor.
Proceed as on page 363.
Varian, Inc.
Page: 411
Description
1/16" No-Hole Ferrule
Page: 412
SG0726841
82-005048-03
03-912805-99
10 L Syringe
47-000007-00
03-925195-00
03-925196-00
03-925157-00
03-925190-90
Spring
03-949786-00
03-925200-00
03-925490-01
03-925461-01
15-003347-00
03-925132-00
27-402826-00
03-925176-00
03-925177-00
03-925178-00
03-925179-00
03-925138-00
03-925192-00
03-925194-04
88-189610-00
16-000327-00
03-925180-00
03-925181-00
Teflon Tweezers
29-898500-00
01-900158-00
User Manual 450-GC
Varian, Inc.
Part Number
Description
Varian, Inc.
01-900115-00
Soap Flowmeter
03-949846-00
Magnifying Lens
00-997369-00
03-925151-01
03-925151-02
03-925151-03
03-925151-04
03-925151-05
03-925158-00
03-925194-02
03-925194-01
Page: 413
Troubleshooting
Overview
The following section describes how to troubleshoot and correct problems with
the Pulsed Flame Photometric Detector (PFPD).
Solution
Check that the carrier, air and hydrogen regulators read correct
pressures. Check that air and hydrogen are flowing through the
PFPD at the expected flow rates.
Turn the PFPD electrometer ON.
Be sure that detector oven power is On and allow enough time
for the tower to approach the operating temperature. This is
particularly important with very hydrogen rich conditions.
Use a mirror to check that the igniter is glowing. If the coil is not
glowing, the igniter coil may be broken. Disconnect the igniter
cable from the electrometer board. Check for continuity. Replace
the igniter coil and assembly if defective.
Check whether gas is leaking from around the aluminum crunch
washers or the light pipe seal.
If the igniter is glowing dimly, the gas composition may be too
rich in hydrogen. To adjust the air:hydrogen composition,
proceed as follows:
Page: 414
Varian, Inc.
Possible Cause
Carrier gas flow rate is too high.
Solution
This is a particular problem when using helium as a carrier gas,
as helium tends to cool the flame excessively, preventing the
flame from propagating into the combustor. Decrease carrier
gas flow rate to an appropriate value, or increase Air1 flow
and/or increase the combustor flow to create a hotter flame in
the combustor.
Check the hydrogen gas supply as follows:
Solution
Combustor is dirty.
The end of the column may be dirty, the column may have
broken off in the combustor, or the column may be inserted too
far into the combustor. If this is the case, splinters of glass may
be fluorescing, thereby increasing the background signal.
Reinstall the column.
The Gate Delay has been set too close to the hydrocarbon
emission. Reset the Gate Delay.
Varian, Inc.
Page: 415
Possible Cause
Solution
Check that the Square Root Mode is not On, unless you wish to
operate in that mode, in which case a high noise level is
expected.
Check for leaks. Check that the mass flow controllers are
operating properly. Replace if necessary.
NOTE: To prevent steps in the baseline when the
AutoSampler cycles, DO NOT drive the AutoSampler with
the PFPD air line.
Increase or decrease trigger level ~20%. You may need to reoptimize Gate Delay after changing trigger level.
A rather characteristic pulse pattern (short pulse-long pulsepause) suggests that the flame front is propagating past the end
of the combustor support. This may happen at high detector
temperatures or when using a narrow-bore column with
hydrogen as the carrier gas. In such situations, the polyimide
coating on the column end will be burned off below the
combustor support.
Increase the Hydrogen flow rate by 10%. This cools the flame
and limits the propagation of the flame below the end of the
combustor support.
Alternatively, you may connect your analytical column to a 15
cm length of 0.53 mm ID deactivated fused silica tubing using a
Press-Fit connector, inserting the deactivated tubing into the
PFPD in place of your analytical column. This minimizes the
clearance between the column and the combustor, preventing
unwanted flame propagation.
Page: 416
Varian, Inc.
Solution
Varian, Inc.
Page: 417
Page: 418
Varian, Inc.
The following table lists the series of test samples available for Varian GC
detectors.
Test Sample
Part Number
Compound Concentrations
TCD
82-005048-01
ECD
82-005048-02
FPD
82-005048-03
TSD
82-005048-04
FID
82-005048-07
Note: If the FID sample is not available, the TCD test sample can be used if first diluted 100:1
Table 61: Detector Test Samples
Varian, Inc.
Page: 419
The following figures show typical chromatograms of the Varian test mixes for
the detectors.
Column: Non-polar, e.g., 0.25m film thickness, 15m x 0.25 mm, CP-Sil 5 CB
Carrier: Nitrogen, 5 mL/minute
Injection: 1079 PTV, 200 C, 1 L splitless
Temperature Program: 80 , hold 1 minute, 20/minute to 200, hold 1 minute
Detector: 300 C, Range 12
Page: 420
Varian, Inc.
Column: Non-polar, e.g., 0.25m film thickness, 15m x 0.25 mm, CP-Sil 5 CB
Carrier: Nitrogen, 5 mL/minute
Injection: 1079 PTV, 200 C, 1 L splitless
Temperature Program: 80, hold 1 minute, 20/minute to 200, hold 1 minute
Detector: 300 C, Range 12
Varian, Inc.
Page: 421
Column: Non-polar, e.g., 0.25m film thickness, 15m x 0.25 mm, CP-Sil 5 CB
Carrier: Helium, 5 mL/minute
Injection: 1079 PTV, 200 C, 1 L splitless
Temperature Program: 80, hold 1 minute, 20/minute to 200, hold 1 minute
Detector: 220 C, Range 0.05, Filament Temperature 270 20, Filament current 200 10 mA
Page: 422
Varian, Inc.
Column: Non-polar, e.g., 0.25m film thickness, 15m x 0.25 mm, CP-Sil 5 CB
Carrier: Nitrogen, 5 mL/minute
Injection: 1079 PTV, 200 C, 1 L splitless
Temperature Program: 80, hold 1 minute, 20/minute to 200, hold 1 minute
Detector: 300 C, Range 1, Capillary current setting
Varian, Inc.
Page: 423
PFPD P Mode
Column: Non-polar, e.g., 0.25m film thickness, 15m x 0.25 mm, CP-Sil 5 CB
Carrier: Nitrogen, 2 mL/minute
Injection: 1079 PTV, 220 C, 1 L splitless of a 1:30 dilution of FPD test mix in iso-octane
Temperature Program: 80, hold 1 minute, 20/minute to 200, hold 1 minute
Detector: 300 C, Range 10, Air1 17 mL/min., Hydrogen 14 mL/min., Air2 10 mL/min.
Gate Delay 4 msec., Gate Width 10 msec, Trigger level 200 mV
Page: 424
Varian, Inc.
PFPD S Mode
Column: Non-polar, e.g., 0.25m film thickness, 15m x 0.25 mm, CP-Sil 5 CB
Carrier: Nitrogen, 2 mL/minute
Injection: 1079 PTV, 220 C, 1 L splitless of a 1:30 dilution of FPD test mix in iso-octane
Temperature Program: 80, hold 1 minute, 20/minute to 200, hold 1 minute
Detector: 300 C, Range 10, Air1 17 mL/min., Hydrogen 14 mL/min., Air2 10 mL/min.
Varian, Inc.
Page: 425
Page: 426
Varian, Inc.
Maintenance
GENERAL GC MAINTENANCE
Overview
In this section maintenance information is presented in 4 headings:
1.
2.
3.
4.
Frequency
Change septa
Condition column
Daily or as required
Weekly, or as required
Weekly
Leak check
Monthly
Quarterly or as required
Condition system
Semiannually
Annually or as required
Pencil Filters
On a yearly base.
Change/Replace column
Page: 427
Maintenance
General GC Maintenance
The common general GC maintenance tasks are checking and changing gas
supply cylinders, leak checking and changing gas purifier filters. These should
be performed at the frequency suggested above. The following procedures are
critical to the successful long term operation of a Gas Chromatograph. It is very
important to leak check the system on a routine basis and to ensure the quality
of gas supplies, particularly the carrier gas and H2 (Safety Reasons).
Check and Renew Gas Supplies
The pressure of the various GC gas supplies should be checked on a weekly
basis and the following guidelines used for frequency of renewing supplies:
Carrier gas: The carrier gas supply cylinder should be changed when its
pressure drops below 200 psi. This ensures that high purity carrier gas is
always supplied to the instrument. With typical usage on one Gas
Chromatograph an A-size cylinder of carrier gas should last for 3 to 6 months.
Note that this usage includes also using the carrier gas as the supply of makeup to the detectors. When a new cylinder is installed the regulator and tubing
should be purged with carrier gas before connecting to the GC. This will avoid
introducing a large amount of air into the GC. Flush the GC for a few minutes
before heating the Column.
Detector fuel gases: The typical detector fuel gases are hydrogen and air.
These should be changed when the supply pressure drops below 200 psi.
Typically an A size cylinder of hydrogen used on one GC with one Flame
Ionization Detector lasts about 6 months. However, due to the high flow rate of
air used in a Flame Ionization Detector this supply will need renewing more
frequently. For this reason it is worthwhile to consider a clean supply of
compressed air for FID equipped instruments.
Leak Check
Leak checking is one of the most important maintenance tasks for any Gas
Chromatograph. The following information refers to general leak checking
procedures for the 450-GC. Specific information regarding individual 450-GC
components will be presented in the relevant part of this maintenance section.
Leaks in the GC system can lead to poor chromatographic performance or
damage components such as the analytical column. The presence of oxygen in
the GC carrier gas at elevated temperatures can lead to permanent column
phase degradation. While the use of an oxygen filter on the carrier supply to the
instrument can help, leaks downstream of the filter are generally more likely to
be the problem.
Page: 428
Varian, Inc.
Maintenance
The use of soap-based leak detection fluids is NOT recommended for a high
performance capillary Gas Chromatograph due to the danger of introducing
contaminants into the system. If leak detection fluids are being used they
should only be used outside the Gas Chromatograph, i.e. to verify there are no
leaks in the plumbing to the Gas Chromatograph. If a leak detection fluid is
being used inside the Gas Chromatograph, an alcohol such as isopropanol or a
50:50 mixture of isopropanol and water should be used. Use a dropper or
syringe to place a few drops of leak detection fluid on the desired fitting and
monitor the area for bubble formation.
The most important step in leak checking is to verify that the GC system can
hold pressure. This is done by removing the column from the Injector, sealing
all exits from the Injector and pressurizing the system. Use the following
procedure to leak check 450-GC Injectors:
When conducting leak check procedures it is important to completely
seal all carrier gas exits, including septum purge and split vent outlets. If
more than one Injector is present, then the outlets of all the Injectors
must be sealed. The following procedure refers to one Injector but
should be duplicated for every Injector on the 450-GC.
1. Set the column oven and the Injector zone temperatures to 50C and allow
them to reach this temperature. This allows safe handling of Injector
surfaces.
2. Remove Injector septum nut and install a new septum. An older septum can
often be the source of a substantial leak.
3. Remove the column from the Injector. Use the appropriate nut and ferrules
to seal the base of the Injector. For packed column systems use a 1/4"
Swagelok blank off plug (SWB400P). For capillary systems use a capillary
nut (03-949551-00) with a no hole ferrule (28-694590-01).
4. If the Injector is either an 1177, 1093 or a 1079 PTV, then the septum purge
and split vent outlets must be sealed. The septum purge outlets are located
on the top frame surrounding the column oven, behind the column oven
door. The split vent outlets are located on the left side panel of the 450-GC.
Seal the septum purge by removing the outlet fitting from the septum purge
valve and replacing it with a blank off plug (P/N SWB200P). Seal the split
vent outlet by installing a Swagelok union on the outlet tube and sealing it
with a blank off plug.
5. With all outlet ports plugged, pressurize the system to 20 - 30 psi. This can
be accomplished by adjustment of manual pneumatic controls or Electronic
Flow Control (make certain that the split ratio is turned to off, constant flow
is not selected and the septum purge valve, if appropriate, is turned off).
For information on setting carrier gas pressure see the Basic Operation
section of this manual.
Varian, Inc.
Page: 429
Maintenance
6. Shut off the carrier supply at the source and monitor the displayed pressure
at the GC for 15 minutes. The pressure should not drop more than 0.5 psi
in 15 minutes.
Leak checking ensures that there are no leaks in the GC system up to and
including the Injector assembly. If the indicated pressure on the GC drops by
more than 0.5 psi (3 kPa) during the 15 minute test period, this indicates that
there is a positive leak in the system. Finding such a leak, particularly if it is a
small leak, can be very difficult. In general the best approach is to
systematically go through the pneumatic system from the source and tighten
each fitting until the leak is eliminated. It is important to note that leaks are often
found in the carrier gas supply to the GC.
Locate leaks preferably with an electronic leak detector or leak detection fluid.
An electronic helium leak detector is available from Varian (CP87610 110 VAC,
CP80610 220 VAC, CP83610 240VAC). This device can detect helium
concentration in the air as low as 2 ppm and is very effective at identifying the
precise location of a helium leak from a GC system.
Page: 430
Varian, Inc.
Maintenance
The carrier gas should first pass through a hydrocarbon trap, then a moisture
filter and then an oxygen trap. As moisture reduces the effectiveness of an
oxygen trap, a moisture trap should always be placed in front of an oxygen trap.
The moisture trap should also be changed immediately when it changes color
to prevent breakthrough of moisture to the oxygen trap.
The minimum recommended carrier gas purifier configuration is a moisture and
hydrocarbon trap. This configuration can be used with routine GC analyses
using FID, TCD, TSD and PFPD detectors. If an ECD or mass spectrometer is
being used, an oxygen trap should also be added. For trace analysis the
combination of hydrocarbon trap, moisture trap and oxygen traps should be
used.
To install or replace the Gas Clean filter cartridges: remove new filter from
sealed bag. Unscrew the ring on the filter base, lift up the filter cartridge. Always
replace the 2 upper O-rings. The base contains check valves that shut off flow
when the cartridge is removed. Place ring over new cartridge, press and hold
cartridge to base and engage retaining ring.
The following are the gas purifiers available from Varian.
Check the Varian website for more information: www.varianinc.com.
Part Number
CP17970
CP17972
CP17971
CP17973
Varian, Inc.
Page: 431
Maintenance
Injector Maintenance
The Injector is the component of the Gas Chromatograph that requires the most
frequent maintenance. This is due to the fact that the sample is deposited in the
Injector, thus leading to potential contamination and build up of non volatile
deposits. The most frequent Injector maintenance is septum replacement. In
addition insert replacement and Injector cleaning are very common. As septum
replacement is common to all liquid Injectors on the 450-GC this procedure is
presented first, followed by instructions for High-Temperature bake out of the
Injector, followed by specific maintenance procedures for the individual
Injectors.
Septum Replacement
The Injector septum is an expendable part and must be replaced on a routine
basis. The frequency of replacement depends on the number of injections and
whether the injections are by hand or with an autosampler. In general the
septum should be replaced every 50 - 100 injections or when symptoms of a
septum leak are seen. These symptoms include changing retention times,
reduced detector response or a drop in column head pressure. The latter
symptom is not always valid as some Injectors, such as the 1079 PTV, are
pressure controlled. With a pressure controlled Injector, the column head
pressure will remain constant even if a leak is present.
Use the following procedure to change a septum on any Varian Injector:
1. Cool the column oven and Injector oven to 50C. This ensures safe
handling of Injector parts and protects the column from elevated
temperatures while it is exposed to air.
2. Turn off the carrier gas supply to the Injector to prevent the septum from
being inadvertently ejected due to carrier gas pressure.
3. Remove the Injector nut by turning it counterclockwise.
4. Remove the old septum using metal tweezers or a septum pick (P/N 72000084-00). Do not to handle any internal parts of the Injector with hands or
gloves.
5. Install the new septum, again using tweezers to avoid contaminating the
Injector.
Page: 432
Varian, Inc.
Maintenance
6. Replace the Injector nut and tighten it finger-tight until you feel resistance
from the septum, then tighten an extra 1/4 turn. Turn on the carrier gas
supply.
On occasion to save time you may want to change a septum while the Injector
is hot. Use an Injector wrench (P/N 03-908423-00) to remove and reinstall the
Injector nut. In all instances the column oven should be cool before removing
the Injector nut.
Item
Part
Number
Description
Bakeout NUT
03-92521701
BLACK
Bakeout DISK
03-92573102
0.34"OD x
3/32"
Bakeout NUT
03-92577201
RED
Bakeout DISK
03-92521602
0.55"OD x
3/32"
1177
1079 PTV
1. Verify the GC is capable of normal operation and that clean, inert carrier
gas is supplied. Handle all parts using clean tweezers and place all parts to
be re-used on clean lab tissues. Remember that it is much easier to
Varian, Inc.
Page: 433
Maintenance
Page: 434
6.
Set the Injector temperature to 400 C and bake out for about 4 hours.
During this time you may want to maintain a column oven temperature of
about 300 C to assist in cleaning the Injector crossover. However, do not
heat the column oven above the temperature limit of any components,
such as valves, etc. which will remain in the oven during bakeout.
7.
Cool the Injector and column oven to about 50 C. Remove the bakeout
nut, bakeout seal, insert and sealing ferrule and scrap column and replace
with components appropriate to the intended analysis. Reset the column
parameters, Gas Saver Timeout and temperature limits you changed in
Setup. Restore the desired GC method parameters.
Varian, Inc.
Maintenance
8.
Varian, Inc.
Page: 435
Maintenance
Page: 436
Varian, Inc.
Maintenance
Pencil Filter
The Pencil filters are used to protect the Injector and/or EFC (Injector
depending).
The Pencil Filter(s) are located underneath the left-Top cover.
Both pencil filters must be replaced on a yearly base.
It is advised to have the filters replaced by a Varian service engineer on a
regular maintenance visit.
Pencil Filters
Varian, Inc.
Page: 437
Maintenance
Page: 438
Varian, Inc.
CLEANING INSTRUCTIONS
To keep the 450-GC surface clean, refer to the remarks given below:
DISPOSAL INSTRUCTIONS
When the lifetime of the 450-GC or parts of it has reached the end of its useful
life, disposal must be carried out in accordance with all (environment)
regulations applicable in your country.
Varian, Inc.
Page: 439
Page: 440
Varian, Inc.
Replacement Parts
REPLACEMENT PARTS
Replacement Parts for the CP-8400 AutoSampler, CP-8410 AutoInjector, 1177, 1041, 1061,
1079 PTV and 1093 Injectors, FID, TSD, ECD, TCD and PFPD detectors.
CP-8400/8410 Vials
2 mL Vials
392611979 100/pk
392612016 100/pk
392611549 100/pk
392611550 100/pk
5 mL Vials
AL666011 1 pc
10 mL Vials
Starter Kit
Varian, Inc.
4/pk
50/pk
100/pk
50/pk
100/pk
50/pk
100/pk
1 ea
Page: 441
Replacement Parts
CP-8400/8410 Syringes
1 L
HM202950 1 ea
2 L
HM202960 1 ea
HM87900
392590204 1 ea
5 L
1 ea
HM80300
1 ea
HM80366
6/pk
Hamilton 10 L 701 syringe, conical tip needle, 26 ga. SK0163520 1 ea
Hamilton 10 L 701N syringe, 26 gauge
10 L
SG002980
1 ea
100 L
SG005333
1 ea
250 L
SGS006157 1 ea
CP738642
1 ea
CP739656
1 ea
392538401 2/pk
392538402 2/pk
392538403 2/pk
392575001 1 ea
Fiber
391896308 3/pk
Test Sample
391896700 5 ea
Fiber Holder
392590210 1 ea
SPME Parts
Page: 442
Varian, Inc.
Replacement Parts
Injector Liners/Inserts
Insert
Type a
Description
1041
1061
1079 PTV
1093
1177
Splitless liner
10
Splitless liner
11
Splitless liner
12
13
14
15
16
17
18
19
20
21
22
X
X
Page: 443
Replacement Parts
23
Insert
Type a
Description
1041
1061
1079 PTV
1093
1177
24
Split/Splitless Liner
25
26
27
28
29
Splitless liner
30
31
Splitless insert
32
33
34
35
36
37
Splitless liner
38
Split/Splitless Liner
39
40
41
42
43
Split/Splitless Focusliner
44
Split/Splitless Focusliner
45
46
Varian, Inc.
Replacement Parts
47
Insert
Type a
Description
1041
1061
1079 PTV
48
Split/Splitless Focusliner
49
Split/Splitless Liner
50
51
52
1093
1177
Liner/Insert Descriptions
Insert
Part Number Description
Type
Type
ID
OD
Gooseneck
4 mm 6.3 mm 72 mm
1000
L
Glass
Inverted cup
4 mm 6.3 mm 72 mm
1000
L
Glass
yes
Gooseneck
4 mm 6.3 mm 72 mm
1000
L
Glass
Yes
Baffle
4 mm 6.3 mm 72 mm
1000
L
Glass
Yes
Gooseneck
2 mm 6.3 mm 72 mm 250 L
Glass
Yes
190010906
SG092021
Quartz
Yes
Tapered
0.5 mm 5 mm
54 mm
10 L
Glass
Yes
Tapered
0.8 mm 5 mm
54 mm
30 L
Glass
Yes
Tapered
0.8 mm 4.6 mm 54 mm
30 L
Glass
Yes
On-column
10
Straight
11
Straight
4 mm 6.5 mm 78.5 mm
12
Gooseneck
13
Gooseneck
4 mm 6.5 mm 78.5 mm
14
Gooseneck
4 mm 6.5 mm 78.5 mm
Metal
1000
L
Yes
Glass
Yes
Glass
Yes
1000
L
Glass
Yes
1000
L
Glass
Glass
Yes
Page: 445
Replacement Parts
15
Gooseneck
Insert
Part Number Description
Type
Type
4 mm 6.5 mm 78.5 mm
ID
OD
1000
L
Glass
Yes
16
Inverted Cup
4 mm 6.3 mm 78.5 mm
1000
L
Glass
17
Inverted Cup
4 mm 6.3 mm 78.5 mm
1000
L
Glass
18
392611933
Inverted cup
4 mm 6.3 mm 78.5 mm
1000
L
Glass
19
Straight
4 mm 6.3 mm 78.5 mm
1000
L
Glass
Quartz
yes
20
Gooseneck
4 mm 6.5 mm 78.5 mm
1000
L
Glass
Glass
Yes
21
Straight
4 mm 6.4 mm 78.5 mm
1000
L
Glass
Glass
yes
22
Straight
Glass
Glass
Yes
23
392611941
Gooseneck
4 mm 6.3 mm 72 mm
24
Gooseneck
1 mm 6.3 mm 72 mm 225 L
25
1000
L
Glass
26
Straight
1000
L
Glass
27
Gooseneck
3.4 mm 5 mm
54 mm 500 L
Glass
Yes
28
Gooseneck
3.4 mm 5 mm
54 mm 500 L
Glass
Yes
29
Gooseneck
54 mm 250 L
Glass
Yes
30
Straight
0.8 mm 5 mm
54 mm
30 L
Glass
Yes
31
Straight
0.5 mm 5 mm
54 mm
15 L
Glass
Yes
32
Tapered
0.8 mm 4.6 mm 54 mm
30 L
Glass
Yes
33
34
35
36
392611954
Straight
2 mm
5 mm
1000
L
Yes
Glass
Glass
54 mm 250 L
Glass
5 mm
54 mm
Glass
Yes
Glass
Siltek
Glass
Siltek
Straight
38
RT207732145
Split/Splitless Liner
5 pack
Straight
4 mm 6.5 mm 78.5 mm
1000
L
Glass
Yes
5 mm
RT207132145
Splitless liner
5 pack
Yes
Glass
37
Page: 446
Yes
Yes
Glass
Straight
Fused Silica
Glass
6 mm 63.5 mm
2 mm
Yes
Yes
Varian, Inc.
Replacement Parts
39
Insert
Part Number
Type
40
RT21079
PTV2265
41
Gooseneck
Description
Type
4 mm 6.2 mm 78.5 mm
ID
OD
1000
L
Glass
Siltek
Straight
SG092113
FocusLiner
42
SG092021
Gooseneck
4 mm 6.3 mm 72 mm
1000
L
Glass
Quartz
Yes
43
SG092022
Split/Splitless Focusliner
Focusliner
4 mm 6.3 mm 72 mm
1000
L
Glass
Quartz
Yes
44
SG092025
Split/Splitless Focusliner
Focus liner
4 mm 6.3 mm 72 mm
1000
L
Glass
Quartz
Yes
45
SG092028
Focus liner
4 mm 6.3 mm 72 mm
1000
L
Glass
Quartz
Yes
46
SG092031
Straight
0.5 mm 5 mm
54 mm
10 L
Glass
47
SG092036
FocusLiner
3.4 mm 5 mm
54 mm 500 L
Glass
Quartz
Yes
48
SG092037
Split/Splitless Focusliner
FocusLiner
3.4 mm 5 mm
54 mm 500 L
Glass
Quartz
Yes
49
SG092038
Split/Splitless Liner
Gooseneck
3.4 mm 5 mm
54 mm 500 L
Glass
Yes
50
SG092039
Tapered
2 mm
54 mm 170 L
Glass
Yes
51
SG092113
Straight
2 mm 6.3 mm 72 mm 250 L
Glass
Yes
Gooseneck
52
CP14988
Varian, Inc.
4 mm 6.5 mm 78.5 mm
1000
L
5 mm
3.4 mm 5 mm
54 mm 500 L
Glass
Quartz
Glass
Glass
Siltek
Yes
Yes
Quartz
Yes
Page: 447
Replacement Parts
1177 Injector
Parts and Supplies
Ceramic scoring wafer (for cutting fused silica columns)
0190015800
0394955100
Ferrule, no hole
2869459001
Septum pick
720000CP-8400
10/pk
RT210462145 5/pk
RT207992145
5/pk
RT207732145
5/pk
RT207132145
5/pk
392611931
5/pk
5/pk
5/pk
392611934
5/pk
392611929
5/pk
Page: 448
Varian, Inc.
Replacement Parts
1177 Injector
Deactivated Varian Injector Liners
Gooseneck split/splitless Injector Liner, 2 MM ID,
modified 2 MM
392611926
5/pk
392611936
5/pk
392611928
5/pk
392611927
5/pk
392611937
5/pk
5/pk
392611938
5/pk
5/pk
5/pk
392611998
Deactivated SGE Injector Liners (HP model 5890, 6890, 6850, 4890 GCS)
Varian, Inc.
SG092016
5/pk
SG092001
5/pk
Page: 449
Replacement Parts
1177 Injector
Deactivated SGE Injector Liners (HP model 5890, 6890, 6850, 4890 GCS)
No Picture available
SGE split inlet Liner, 4 MM ID, Straight-Through
SG092007
5/pk
SG092005
5/pk
5/pk
SG092002
5/pk
SG092111
5/pk
SG092018
5/pk
SG092010
5/pk
SG092017
5/pk
5/pk
5/pk
No Picture available
SGE split/less 2.3 MM ID fast focusliner inlet liner
No Picture available
No Picture available
SGE splitless inlet liner, 2 MM ID, Straight-through,
Quartz
Page: 450
SG092004
5/pk
Varian, Inc.
Replacement Parts
1177 Injector
Septa
Insert
Seals
CR298713
50/pk
CR246713
50/pk
CR239778
25/pk
Viton O-ring for 6.3 and 6.5mm OD liners, 260 C max 8850103100
25/pk
392611930
10/pk
392611940
10/pk
392611935
10/pk
0190015800
0394955100
Ferrule, no hole
2869459001
Septum pick
720000CP-8400
10/pk
392611948
5/pk
190010907
5/pk
190010906
5/pk
392611946
5/pk
Split Inserts
Varian, Inc.
Page: 451
Replacement Parts
392611945
5/pk
392611954
5/pk
392611947
5/pk
5/pk
5/pk
Page: 452
SG092245
5/pk
5/pk
SG092037
5/pk
SG092038
5/pk
SG092039
5/pk
Varian, Inc.
Replacement Parts
Septa
SG092031
5/pk
CR298777
50/pk
CR246725
50/pk
CR239787
25/pk
392534201
10/pk
CR298705
50/pk
CR239188
25/pk
CR246124
50/pk
392543101
392611943
5/pk
392611944
5/pk
CP742552
1/pk
1041/1061 Injector
Parts and Supplies
Septa
Inserts
No Picture available
1041 Stainless Steel Insert for Headspace injection
Packed Column Adapter Kits
Packed Column Adapter Kit for 1041, 1/8 SS columns
03-925588-91
03-925588-92
Packed Column Adapter Kit for 1041& 1061, 1/4 glass columns
03-925586-91
Varian, Inc.
Page: 453
Replacement Parts
1093 Injector
Parts and Supplies
Septa
Column
Ferrule
00-997630-02 25/pk
CR298777
50/pk
CR239787
25/pk
CR246725
50/pk
391833400
391833401
Septum Support
391821100
390820601
3600004601
2824707101
394966601
Polymide/Graphite 0.5mm ID
2869458101
10/pk
Polymide/Graphite 0.8mm ID
2869458200
10/pk
Graphite 0.5mm ID
SG072626
10/pk
Graphite 0.8mm ID
392611909
10/pk
190010906
5/pk
190010907
5/pk
392611950
5/pk
SG092027
5/pk
10/pk
Inserts
No Picture Available
0.5mm ID, SPI deactivated Injector liner
No Picture Available
0.8mm ID, 0.25mm restriction SPI deactivated Injector
SG092030
Liner
No Picture Available
0.8mm ID, 0.5mm restriction SPI deactivated Injector
Liner
Page: 454
SG092034
5/pk
5/pk
Varian, Inc.
Replacement Parts
Detectors
Flame Ionization Detector (FID)
Flame Tip Assembly (0.02 ID)
02-001938-00
15-003347-01
03-906074-00
03-906074-01
02-001972-01
ECD Wrench
72-000081-00
03-925605-01
15-003347-00
03-925176-00
03-925177-00
03-925151-01
03-925151-02
Igniter Assembly
03-925160-00
Combustor support, 2 mm
03-925178-00
Combustor support, 3 mm
03-925179-01
03-925138-00
Varian, Inc.
Page: 455
Replacement Parts
Page: 456
Varian, Inc.
Technical Tips
TECHNICAL TIPS
This section will give technical tips regarding common questions like:
1. Chromatographic Interpretations, on page 458.
2. To Bleed not to Bleed .. that is the question 1, on page 461.
3. Choosing the right GC Injection Technique 2, on page 466.
4. GC line Labyrinth How to choose the right Injector Insert 3, on page 472.
Page: 457
Technical Tips
CHROMATOGRAPHIC INTERPRETATIONS
The troubleshooting procedures outlined in this section are structured to
accommodate deviations from known responses of a standard sample,
including peak shapes, peak areas, retention times, baseline characteristics,
etc.
In these cases, the general characteristics of the chromatogram are well known
and even slight deviations can be discerned.
A vast majority of gas chromatographic malfunctions will be indicated by one or
more of the following symptoms; inspect the chromatogram for all of the
symptoms:
Cause
Remedy
Tailing
Contaminated insert.
Excess dead volume.
Injector not heating correctly.
Technical Tips
Symptom
Cause
Remedy
Check carrier gas flow.
Contaminated Injector.
Leading peak
Overloaded column.
Contamination.
Varian, Inc.
Page: 459
Technical Tips
Symptom
Cause
Instrument problem.
Page: 460
Remedy
detector body connection
Condition the column or
bake out the column as
described in Section 4. In
badly contaminated
systems, it may be
necessary to lift the
detector crossover from
the ion oven, then heat
the lines and crossover
with a flame while purging
with an inert gas.
Install an uncoated test
column and repeat the
temperature program.
Change to ceramic flame
tip,
P/N 03-918351-00.
Reverse glass insert.
Varian, Inc.
Technical Tips
Both the column and the Injector/detector system can play a role in baseline
rises and it is important to distinguish between the 2 sources.
Starting the troubleshooting exercise by immediately disconnecting the column
from Injector and/or detector can mean an unstable system for at least a couple
hours. It can be more effective to have a closer look at the baseline itself first.
Varian, Inc.
Page: 461
Technical Tips
Injector contamination
Unstable and noisy baselines, which include ghost peaks during a blank
analysis run, are hardly ever caused by column bleed.
Clear distinctive peaks eluting during such a blank analysis run are an
indication again of Injector contamination (figure 1).
Higher molecular weight residues inside the Injector are continuously being
introduced on to the column, trapped there at lower oven temperatures and
eluted again at higher temperatures. Injector contamination can be confirmed
by lowering the Injector temperature to e.g. 100 C. Evaporation from the
Injector of the higher molecular residues is significantly reduced at these lower
temperatures. The overall bleed level should drop and the majority of the ghost
peaks should disappear. In case of a split/splitless injection Injector
contamination can also be confirmed by extending the splitless time period
considerably. A contaminated Injector will expose itself by an increased bleed
level and/or higher ghost peaks. If the Injector is contributing to the bleed level
maintenance should be performed. In particular replace the septum, liner and
glass wool. Confirm the cleanliness of the Injector by performing a blank
injection.
If lowering the Injector temperature has little effect on the total bleed level the
focus of attention should be directed towards the column or to the purity of the
carrier gas.
Page: 462
Varian, Inc.
Technical Tips
The Column
Mass spectra taken from the bleed can be of great help in determining its
source. The main bleed products from all siloxane-based columns are some
smaller cyclic siloxane fragments formed by degradation of the liquid phase.
These fragments have m/z ions of 207, 281 and 341 (figure 2). However, the
presence of these ions in the spectrum is no confirmation of column bleed. Also
the septum is made of almost 100 % pure polydimethylsiloxane just like the
liquid phase of the capillary column. The total siloxane amount of a septum can
be more than a factor 40 of that present in a capillary column. Torn away pieces
of septum inside the liner or septum bleed can therefore be responsible for this
typical mass spectrum. In cases where the mass spectra of the bleed is
dominated by other m/z ions than 207 and 281 the cause of the bleed is most
likely Injector contamination due to sample residues. Degradation of the liquid
phase and formation of bleed products is often the result of low ppm level
presence of oxygen and moisture in the carrier gas. It is therefore imperative to
install high quality gas filters close to the GC.
Bleed formation from the column is a continuous process and usually exhibits
itself as a slow steady climb of the baseline. Exceptions to this general rule are
the older type cyanopropyl containing liquid phases. These columns may form a
wavy, sinusoid like, rising baseline close to their maximum operating
temperature.
Varian, Inc.
Page: 463
Technical Tips
Page: 464
Varian, Inc.
Technical Tips
Gas Impurities
Impurities in either the carrier gas or detector gases can cause unstable
baselines.
Figure 4 shows the effect on the baseline stability due to the presence of some
hydrocarbon impurities in the hydrogen gas flow for an FID. A noisy baseline
caused by gas flow impurities is independent of oven temperatures and visible
throughout the entire temperature program.
The installation of a good quality gas purifying system like the Varian GasClean filters will maintain a supply of high quality gasses to both the column
and detector. Overall baseline stability will improve. If positioned close enough
to the GC such a system will guarantee oxygen, moisture and hydrocarbon free
gas supply regardless of initial gas bottle quality or the possible presence of
leakage anywhere in the gas supply system.
Varian, Inc.
Page: 465
Technical Tips
Page: 466
Varian, Inc.
Technical Tips
Direct Injection
Description
Sample is vaporized in the hot insert. The entire sample enters the
column.
Volume(s)
0.1 L 1 L
Concentration Range
1 ppm - %-levels.
Application
Direct injection is the most commonly used injection technique for packed
columns and 0.53 mm ID columns. Its major advantages are its ease of use
and the wide concentration range that it can cover. It is especially suitable for
the lower concentrations in the ppm range, which are below the scope of the
split injection. Higher gas flow rates are needed for an efficient and quick
sample transfer from the Injector liner to the column. The minimum gas flow for
direct injection is about 5 mL/min. Direct injection can therefore only be applied
to columns, which can cater for higher column flow rates such as, packed
columns or 0.53 mm ID columns. These columns can also handle the higher
concentration levels because of their increased sample capacity. Direct
injection will produce poor peak shapes if too large injection volumes (> 1 L)
are applied. Discrimination effects are low compared to split injections. Direct
injection is sometimes confused with the splitless type injection described
further below. The Varian Injector models 1040/1041 and 1060/1061 are
designed for various types of direct injection.
Split Injection
Description
Volume(s)
0.1L 1 L
Concentration Range
50 ppm - %-level
Remark
Varian, Inc.
Page: 467
Technical Tips
Split ratio
530
1:5 to 1:15
15
320
1:20 to 1:250
25
250
1:50 to 1:250
40
150
1:150 to 1:500
75
100
> 1:500
150
The suggested minimum split ratios and split flows can result in band
broadening inside the Injector liner and therefore loss of separation efficiency is
likely.
The 1177 Injector model is the most recent split/splitless Injector of Varian.
Splitless Injection
Description
Volume(s)
0.5 L 2 L
Concentration Range
Remark
Page: 468
Varian, Inc.
Technical Tips
Boiling Point
[C]
n-Pentane
36
10 to ambient
Dichloromethane
36
10 to ambient
Carbon disulfide
40
10 to ambient
Chloroform
46
10 to ambient
Methanol
61
25
n-Hexane
65
35
Ethyl acetate
69
40
Acetonitrile
77
45
n-Heptane
82
50
Isooctane
98
70
Column flow and the volume of the liner determine the speed and effectiveness
of the sample transfer from Injector to column. Very low column flows prevent
the use of splitless and it is mainly for this reason that 100 m are not suited of
splitless type of injections. 150 m columns can be used but only if the splitless
time period is long enough and the liner volume limited to 250 L to permit
adequate sample transfer.
Splitless is not very suitable for samples where the components of interest elute
closely to the solvent either in front or behind. Insufficient focusing of the
analytes will produce poor peak shapes. In those cases a direct injection on a
0.53 mm ID column is often preferable.
The 1177 Injector model is the most recent split/splitless Injector of Varian.
Varian, Inc.
Page: 469
Technical Tips
Cool on-column
Description
Sample passes from the syringe into the column. The column is kept
cool during injection and is subsequently heated. The Injector is kept
cool during sample introduction.
Volume(s)
0.1 L 2 L
Concentration Range
Remark
Sample passes as a liquid from the syringe into a cooled inlet. The
sample is subsequently heated to vaporize the sample. Split or
splitless injection can be applied.
Volume(s)
0.1 L 250 L
Concentration Range
1 ppb 50 ppm
Remark
Page: 470
Varian, Inc.
Technical Tips
components starts as the insert is being heated. The PTV can operate in
various modes:
Cold Split Injection
is comparable with a split injection but with a cold transfer from syringe to
insert. This provides less discrimination effects and therefore more accurate
results. Refocusing using retention gaps and low oven temperatures may be
necessary for lower boiling compounds.
Cold Splitless Injection
is similar to hot splitless. It has the same advantages as cold split with
respect to the cold transfer from the syringe. The classic hot splitless technique
can result in thermal degradation due to the long residence time of components
in the hot insert. The PTV in cold splitless mode also avoids this phenomenon.
Refocusing using retention gaps and low oven temperatures may be necessary
for lower boiling compounds.
Solvent Venting
permits the injection of large sample volumes (250 L). It enables detection
limits in the ppb range. The majority of the solvent is removed through the split
vent. The components are retained in the liner by cold trapping, aided in some
cases by packed supports. At the end of the vent period the split vent is closed
and the insert is rapidly heated. This technique requires a slow sample
introduction in order to avoid overloading the insert. Lower boiling compounds
will unavoidably be partly vented of together with the solvent depending on
insert temperatures, type of packed support and vent times. In general, light
boiling solvents are preferred for better results. Some column refocusing on the
column supported by retention gaps and low initial oven temperatures is
needed.
The Varian 1079 PTV Injector can perform all of these modes.
Varian, Inc.
Page: 471
Technical Tips
Liner Types
Three major types of injection based on flash vaporization of the sample can be
distinguished:
1. Direct injection
After flash vaporization the entire sample is transferred to the column. The
sample transfer speed in the Injector is determined by column flow and
volume of the Injector liner. This injection type is most commonly applied for
0.53 mm ID columns or packed columns. Injection volumes usually do not
exceed 1 L. This maximum in injection volume is related to the internal
volume of the liner.
2. Split injection
After flash vaporization the sample is divided into 2 streams. One directed
towards the column, the other to the split vent. Here, the sample transfer
speed in the Injector is mainly determined by the flow of the split vent and
the liner volume. The column flow plays an insignificant role. A high transfer
speed of the sample is critical for the quick introduction of the compounds
into the capillary column maintaining the columns separation efficiency Split
injections are used for all capillary columns with IDs ranging from 100 m to
530 m. The maximum injection volume for split injections is about 1 L.
Page: 472
Varian, Inc.
Technical Tips
3. Split/Splitless
During the initial splitless period the vaporized sample is transferred to the
column. The capillary column flow and the liner volume determine the speed
and efficiency of this process. Larger injection volumes are possible for
splitless type injections; volumes up to 5 L are not uncommon. These large
injection volumes can never be injected at once because of possible liner
overload but must be careful introduced at a rate of about 1 L/sec.
Liner volume
The sample gas volume, which at one time resides in the Injector, should in all
cases be smaller than the internal volume of the liner. If not, the liner is
overloaded and back flash of the sample into various (metal) parts of the
Injector may occur. This phenomenon will cause poor injection profiles, bad
peak shapes, adsorption and possible sample carry over. Many liners used for
direct and split injection have dimensions of about 80 mm x 4 mm resulting in
an internal volume of ca. 1mL. The effective volume available for the sample is
however significantly lower as part of the liner volume will always be filled with
carrier gas. The list below indicates the solvent expansion volume per L at
250 C and 50 kPa
Methanol
0.65 mL
n-hexane
0.2 mL
Isopropanol
0.34 mL
Water
1.46 mL
Dichloromethane
0.41 mL
It is clear from the above that water and methanol in particular quickly can
cause liner overload and some care must taken if these solvents are used. For
these solvents smaller injection volumes than 1 L are to be recommended.
Varian, Inc.
sample volume
type of solvent and sample components
Injector temperature
Injector and liner geometry and design
presence of quartz wool or other liner fillings
Page: 473
Technical Tips
The last three determine the heat capacity of the Injector and its ability for heat
transfer to the sample. The below list compares the amount of energy needed for
evaporation of methanol and water relative to cyclohexane.
Water
5.22
Methanol
3.29
Cyclohexane
1.00
Solvents like methanol and water require much more energy than other
common solvents for their evaporation therefore they need a larger heat
capacity and better heat transfer system for a quick evaporation.
This can be accomplished by using higher Injector temperatures (>275 C) and
larger liner surface areas. Glass or quartz wool is often used to increase the
Injectors heat transfer capacity. The evaporation of higher boiling compounds
also requires more energy and also in those cases quartz wool is often used.
Besides benefits glass or quartz wool also has a few disadvantages. The wool
can become adsorptive especially if some fibers are broken or when it has
become dirty. It should be exchanged at a regular basis to prevent
chromatographic problems
Discrimination effects
The main aim and challenge of injections is to maintain sample integrity and
sample composition during its transition from the liquid to the gaseous state and
also during the sample transfer from Injector to the column. The evaporation
process of low boiling compounds is always quicker than that of high molecular
weight ones. This difference accounts for the often-mentioned discrimination
effects caused by injections. In general, the quicker and more complete the
evaporation of the sample, the less discrimination effects will be observed.
Higher Injector temperatures, special liner design and quartz wool will help to
minimize these effects. Inconsistent discrimination profiles will result in poorer
repeatability.
Page: 474
Varian, Inc.
Technical Tips
These are low cost liners for samples with a narrow boiling point range
and little risk of thermal decomposition. Can be packed with quartz
wool to aid evaporation, limit discrimination and trap non-volatile
materials. Do not use for high boiling samples. Comes usually in 2 or 4
mm ID. Use the 4 mm version for sample volumes > 2 L to prevent
liner overload.
Gooseneck
The single gooseneck can be easily packed with quartz or glass wool to
minimize discrimination. Wool is also a risk as it can become active and
adsorptive due to the trapping of non-volatile residues. Replace it at
regular intervals.
Double Gooseneck
Varian, Inc.
Page: 475
Technical Tips
Recessed Gooseneck
This model has the same advantages as the other goosenecks but in
addition it can be packed with glass wool at the top. It can also be used
for dual column analysis.
The ceramic frit will prevent dirt and non-volatiles from reaching the
column. The frit can become active. The gooseneck prevents back
flash.
Inverted -or Laminar Cup Splitter
The cup will promote evaporation and mixing of the sample with the
carrier gas. It is the best splitter for higher molecular weight
compounds. Slightly larger injection volumes are possible because the
cup will hold the sample liquid until it vaporizes. It is difficult to clean
and is relatively expensive.
Baffle Splitter
Page: 476
Varian, Inc.
Technical Tips
This type of liner is more suited for packed columns than it is for 0.53
mm ID columns. The wide-bore column should be positioned at the
bottom of the liner just piercing through the ferrule. This will minimize
the effects of possible dead volume during injections although Injector
parts, which are swept poorly with carrier gas will result in poor injection
profiles.
Gooseneck
Varian, Inc.
Page: 477
Technical Tips
Page: 478
Varian, Inc.
Error Description
Coolant timed out
Error level
Event
Cause Condition(s)
450-GC action
Error Led on for all
errors
Notification Message
Action needed
Activate method, clear
timeout
15
Alert
16
Alert
17
Alert
Notification Message
Auto-re-ignite
Notification Message
Auto-re-ignite
Notification Message
Auto-re-ignite
Event
Air Leak
19
Event
Air Leak
20
Event
Air Leak
35
Alert/Error w
Activated Safety
Method
36
Alert/Errora
Activated Safety
Method
37
Alert/Errora
Activated Safety
Method
Alert/Errora
Notification Message
w If the used gas is Hydrogen (H ) the Error level is Error, in all other cases an Alert is given.
2
Varian, Inc.
Page: 479
Alert/Errora
40
Alert/Error x
Notification Message
Notification Message
Alert/Errora
Activated Safety
Method
42
Alert/Errora
Activated Safety
Method
43
Alert/Errora
Activated Safety
Method
Alert/Errora
Activated Safety
Method
45
Alert/Errora
Activated Safety
Method
46
Alert/Errora
Activated Safety
Method
50
Alert
51
Alert
52
Alert
53
Event
54
Event
Event
IOC Failure
55
x If the used gas is Hydrogen (H ) the Error level is Error, in all other cases an Alert is given.
2
Page: 480
Varian, Inc.
Event
57
Event
58
Event
59
Event
60
IOC no Detector
Event
61
Event
62
Event
63
IOC_EFC_COMMUNICATION_ERROR, Unit
Rebooted
Event
Error
68
Error
69
Error
70
Error
71
Error
72
Error
73
Error
Error
75
Error
Varian, Inc.
Page: 481
Error
77
Error
78
Error
79
Error
80
Error
Error
82
Error
83
Error
84
Error
85
Error
86
Error
87
Error
Error
Defect Sensor
89
Error
Defect Sensor
90
Error
Defect Sensor
91
Error
Defect Sensor
Page: 482
Varian, Inc.
Error
Defect Sensor
93
Error
Defect Sensor
94
Error
Defect Sensor
Error
96
Error
97
Error
98
Error
99
Error
100
Error
101
Error
Error
110
Error
111
Error
112
Error
113
Error
114
Error
Varian, Inc.
Page: 483
Error
Error
117
Error
118
Error
119
Error
120
Error
121
Error
122
Error
Fan or Heater
123
Error
Hardware error
124
Error
Hardware error
Event
Notification Message
231
Event
Notification Message
232
Event
Notification Message
Alert
234
Alert
Page: 484
Varian, Inc.
Alert
Error
248
Error
249
Error
250
Error
251
Error
252
Error
253
Error
Alert/Error y
Flow is above max +10 % or below 5 mL/min for more then 10 sec.
305
Alert/Errora
Flow is above max +10 % or below 5 mL/min for more then 10 sec.
306
Alert/Errora
Flow is above max +10 % or below 5 mL/min for more then 10 sec.
307
Alert/Errora
Flow is above max +10 % or below 5 mL/min for more then 10 sec.
308
Alert/Errora
Flow is above max +10 % or below 5 mL/min for more then 10 sec.
309
Alert/Errora
Flow is above max +10 % or below 5 mL/min for more then 10 sec.
310
Alert/Errora
Flow is above max +10 % or below 5 mL/min for more then 10 sec.
311
Alert/Errora
Flow is above max +10 % or below 5 mL/min for more then 10 sec.
312
Alert/Errora
Flow is above max +10 % or below 5 mL/min for more then 10 sec.
y If the used gas is Hydrogen (H ) the Error level is Error, in all other cases an Alert is given.
2
Varian, Inc.
Page: 485
313
Alert/Errora
314
Alert/Error z
315
Alert/Errora
316
Alert/Errora
317
Alert/Errora
318
Alert/Errora
319
Alert/Errora
320
Alert/Errora
321
Alert/Errora
Event
371
Sampler Faulted
Alert
Mismatch AutoSampler
Notification Message
372
Alert
No communication AS
Notification Message
373
Alert
374
Alert
375
Alert
376
Alert
377
Alert
No Syringe available
Place Syringe
378
Alert
No Vial available
Vial detect should be turned on
Notification Message
z If the used gas is Hydrogen (H ) the Error level is Error, in all other cases an Alert is given.
2
Page: 486
Varian, Inc.
379
Alert
380
Alert
Notification Message?
restart
381
Alert
Notification Message?
383
Alert
384
Alert
385
Alert
386
Alert
387
Alert
Notification Message?
Event
Activate standby
method
389
Event
Activate standby
method
390
Event
Activate standby
method
391
Alert
Stopped
Notification Message?
Miscellaneous
412
Alert
Notification Message?
413
IOP_COMMUNICATION_FAILURE_AT_STARTUP_e
Error
Notification Message?
Replace/fix IOP
425
IOP_SKIPS_ADC_READING_e
Error
Notification Message?
Alert
Varian, Inc.
Page: 487
Alert
Notification Message?
418
Alert
Notification Message
423
Alert
Defective sensor
424
Event
426
Alert
427
Alert
428
Alert
429
Alert
434
Alert
436
Error
437
Event
439
440
Notification Message?
Replace/fix IOP
Error
Recalibrate PGA
Error
Recalibrate temp
442
Error
443
Error
GC main board
444
Error
445
Error
446
Error
447
Error
Page: 488
Varian, Inc.
Error
449
Error
450
Error
451
Error
452
Error
453
Error
454
Error
455
Error
456
Error
457
Error
458
Error
Error
460
Error
461
Error
462
Error
463
Error
464
Error
465
Error
466
Error
Page: 489
Alert
Buffer underrun
483
Alert
Buffer overrun
Notification Message
484
Alert
Notification Message
485
Alert
Notification Message
Page: 490
Varian, Inc.
Varian, Inc.
Page: 491