Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                

HPLC Troubleshooting: Problem Possible Cause Solution

Download as docx, pdf, or txt
Download as docx, pdf, or txt
You are on page 1of 9

HPLC Troubleshooting

Abhishek Raj
Deputy Manager (R&D)
Quest Pharmaceuticals Pvt. Ltd, Chhata Pipara, Bara, Birgunj, Nepal.
abhishekraj4@gmail.com

Start up - Preliminary checks


Problem

Possible cause
Detector off

No peaks or
very small
peaks

Solution
Check detector

Broken connections
Check connections
to recorder
No sample/Wrong
sample

Check sample. Be sure it is not deteriorated.


Check for bubbles in the vials

Wrong settings on
Check attenuation. Check gain
recorder or detector
Pump off

Start Pump

Flow interrupted

Check reservoirs. Check position of the inlet


tubing. Check loop for obstruction or air. Check
degasing of mobile phase. Check compatibility of
the mobile phase components.

Leak

Check fittings. Check pump for leaks and


precipitates. Check pump seals.

Air trapped in the


system

Disconnect column and prime pump. Flush


system with 100% methanol or isopropanol.
Contact servicing if necessary.

No Flow

Column and Fittings Leaks


Problem
Column end

Possible cause
Loose fitting

abhishekraj4@gmail.com

Solution
Tighten or replace fitting

leaks

White powder at
loose fitting

Cut tubing and replace ferrule; disassemble


fitting, rinse and reassemble.

Leak at detector Detector-seal failure Replace detector seal or gaskets.


Leak at injection Worn or scratched
valve
valve rotor

Replace valve rotor

Leak at pump

Replace pump seal; check piston for scratches


and, if necessary, replace

Pump seal failure

Change in Retention time


Problem

Possible cause

Solution

Buffer retention times

Use buffer with concentration greater


than 20 mM.

Contamination buildup

Flush column occasionally with


strong solvent

Equilibration time
Pass at least 10 column volumes
insufficient for gradient run
through the column for gradient
or changes in isocratic mobile
regeneration or after solvent changes
phase
First few injections - active
Changing
sites
Retention Times

Condition column by injecting


concentrated sample

Inconsistent on-line mobilephase mixing

Ensure gradient system is delivering a


constant composition; compare with
manually prepared mobile phase;
partially premix mobile phase

Selective evaporation of
mobile-phase component

Cover solvent reservoirs; use lessvigorous helium purging; prepare


fresh mobile phase

Thermostat or insulate column;


Varying column temperature ensure laboratory temperature is
constant.
Decreasing
Retention Times Active sites on column
packing

Use mobil-phase modifier, competing


base (basic compounds), or increase
buffer strength; use higher coverage
column packing.

Column overloaded with


sample

Decrease sample amount or use


larger-diameter column.

Increasing flow rate

Check and reset pump flow rate.

Loss of bonded stationary

Use mobile-phase pH between pH 2

abhishekraj4@gmail.com

phase or base silica

and pH 8

Thermostat or insulate column;


Varying column temperature ensure laboratory temperature is
constant
Decreasing flow rate
Increasing
Retention Times Changing mobile-phase
composition
Loss of bonded stationary
phase
Slow column
equilibration
time

Check and reset pump flow rate;


check for pump cavitation; check for
leaking pump seals and other leaks in
system
Cover solvent reservoirs; ensure that
gradient system is delivering correct
composition.
Use mobile-phase pH between pH 2
and pH 8

Reversed phase ion pairing long chain ion pairing


Use ion-pairing reagent with shorter
reagents require longer
alkyl chain length
equilibration time
Baseline

Problem

Possible cause
Air bubbles in mobile phase

Void Time Positive-negative noise


difference in refractive
index of injection solvent
and mobile phase
Drifting
baseline

Solution
Degas or use back pressure restricor on
detector
Normal with many samples; use mobile
phase as sample solvent

Use non-UV absorbing mobile phase


Negative direction (gradient
solvents; use HPLC grade mobile phase
elution) - absorbance of
solvents; add UV absorbing compound to
mobile-phase A
mobile phase B.
Use higher UV absorbance detector
Positive direction (gradient wavelength; use non-UV absorbing mobile
elution) - absorbance of
phase solvents; use HPLC grade mobile
mobile phase B
phase solvents; add UV absorbing compound
to modile phase A.
Positive direction contamination buildup and
elution

Flush column with strong solvent; clean up


sample; use HPLC grade solvents

Wavy or undulating -

Monitor and control changes in room

abhishekraj4@gmail.com

Baseline
noise

temperature changes in
room

temperature; insulate column or use column


oven; cover refractive index detector and
keep it out of air currents.

Continous - detector lamp


problem or dirty cell

Replace UV lamp( each should last 2000 h;


clean and flush flow cell.

Gradient or isocratic
proportioning - lack of
solvent mixing

Use proper mixing device; check


proportioning precision by spiking one
solvent with UV absorbing compound and
mointor UV absorbance detector outputl.

Gradient or isocratic
proportioning malfunctioning
proportioning valvesl

Clean or replace proportioning precision


valves; partially remix solventsl.

Occasional sharp spikes external electrical


interference

Use voltage stabilizer for LC system; use


independent electrical circuit.

Periodic - pump pulses

Service or replace pulse damper; purge air


from pump; clean or replace check valves.

Random - contamination
buildup

Flush column with strong solvent; clean up


sample; use HPLC grade solvent

Spikes - bubble in detector

Degas mobile phase; use back pressure


restrictor at detector outlet.

Spikes - column
temperature higher than
boiling point of solvent

Use lower column temperature.

Pressure
Problem

Decreasing
Pressure

Possible cause

Solution

Insufficient flow from pump

Loosen cap on mobile phase


reservior

Leak in hydralic lines from pump


to column

Tighten or replace fittings; tighten


rotor in injection valve

Leaking pump check valve or seals

Replace or clean check valves;


replace pump seals.

Pump cavitation

Degas solvent; check for


obstruction in line from solvent
reservoir to pump; replace inlet-line
frit

abhishekraj4@gmail.com

Fluctuating
pressurre

Bubble in pump

Degas solvent; purge solvent with


helium

Leaking pump check valve or seals

Replace or clean check valves;


replace pump seals

Column blocked wth irreversibly


adorbed sample

Improve sample cleanup; use guard


column; reverse-flush column with
strong solvent to dissolve blockage

Column particle size too small (for Use larger particle size (for
example 3 micrometers)
example 5 micrometer)

High Back
Pressure

Microbial growth on column

Use at least 10% organic modifier


in mobile phase; use fresh buffer
daily; add 0.02% sodium azide to
aqueous mobile phase; store
column in at least 25% organic
solvent without buffer

Mobile phase viscosity too high

Use lower viscosity solvents or


higher temperature

Plugged frit in in-line filter or


guard column

Replace frit or guard column

Plugged inlet frit

Replace endfitting or frit assembly

Use correct solvent with column;


change to proper solvent
compositionl consult
Polymetric columns - solvent
manufacturer's solventchange causes swelling of packing
compatibility chartl use a column
with a higher percentage of crosslinking
Salt precipitation (especially in
reversed-phase chromatography
with high concentration of organic
solvent in mobile phase)
concentration of organic solvent in
mobile phase)

Ensure mobile phase compatibility


with buffer concentration; decrease
ionic strength and water-organic
solvent ratio; premix mobile phase

When injector disconnected from


column - blockage in injector

Clean injector or replace rotor

Blocked flow lines

Systematically disconnect
components from detector end to
column end to find blockage;
replace or clean blocked
component

Particulate buildup at head of

Filter sample; use .5 micrometer in-

Increasing
Pressure

abhishekraj4@gmail.com

column

line filter; disconnect and backflush


column; replace inlet frit

Water-organic solvent systems buffer precipitation

Ensure mobile phase compatibility


with buffer concentration; decrease
ionic strength or water organic
solvent ratio

Peaks
Problem

Possible cause

Solution

Analytes eluted early due


Dilute sample 1:10 and reinject
to sample overload
Use smallest possible cell volume consistent
Detector-cell volume too
with sensitivity needs; use detector with no heat
large
exchanger in system

Broad
peaks

Ghost

Injection volume too


large

Decrease solvent strength of injection solvent to


focus solute; inject smaller volume

Large extra column


volume

Use low- or zero-dead-volume endfittings and


connectors; use smallest possible diameter of
connecting tubing (<0.10 in. i.d.); connect
tubing with matched fittings

Mobile-phase solvent
viscosity too high

Increase column temperature; change to lower


viscosity solvent

Peak dispersion in
injector valve

Decrease injector sample loop size; introduce


air bubble in front and back of sample in loop

Poor column efficiency

Use smaller-particle-diameter packing, lowerviscosity mobile phase, higher column


temperature, or lower flow rate

Retention time too long

Use gradient elution or stronger isocratic


mobile phase

Sampling rate of data


system too low

Increase sampling frequency.

Slow detector time


constant

Adjust time constant to match peak width

Some peaks broad - late


elution of analytes
retained from previous
injection

Flush column with strong solvent at end of


run; end gradient at higher solvent
concentration

Contamination

Flush column to remove contaminatint; use

abhishekraj4@gmail.com

HPLC-grade solven
Elution of analytes
retained from previous
injection

Flush column with strong solvent at end of


run; end gradient at higher solvent
concentration

Ion-pair chromatography Prepare sample in mobile phase; reduce


- upset equilibrium
injection volume

peaks

Negative
peaks

Oxidation of
trifluoroacetic acid in
peptide mapping

Prepare trifluoroacetic acid solutions fresh


daily; use antioxidant

Reversed-phase
chromatography contaminated water

Check suitability of water by running different


amounts through column and measure peak
height of interferences as function of
enrichment time; clean water by running it
through old reversed-phase column; use
HPLC-grade water.

Unknown interferences
in sample

Use sample cleanup or prefractionation before


injection.

Refractive index
detection - refractive
index of solute less than
that of mobile phase

Reverse polarity to make peak positive

UV-absorbance detection
Use mobile phase with lower UV absorbance;
- absorbance of solute
if recycling solvent, stop recycling when
less than that of mobile
recycled solvent affects detection
phase

Peak
Doubling
Blocked Frit

Replace or clean frit; install 0.5-um porosity


in-line filter between pump and injector to
eliminate mobile-phase contaminants or
between injector and column to eliminate
sample contaminants

Coelution of interfering
compound

Use sample cleanup or prefractionation; adjust


selectivity by changing mobile or stationary
phase

Coelution of interfering Flush column with strong solvent at end of


compound from previous ran; end gradient at higher solvent
injection
concentration
Column overloaded
Column void or
channeling
abhishekraj4@gmail.com

Use higher-capacity stationary phase; increase


column diameter; decrease sample amount
Replace column, or, if possible, open top
endfitting and clean and fill void with glass
beads or same column packing; repack column

Injection solvent too


strong

Use weaker injection solvent or stronger


mobile phase

Use injection volume equal to one-sixth of


Sample volume too large column volume when sample prepared in
mobile phase for injection

Peak
Fronting

Tailing
Peaks

Unswept injector flow


path

Replace injector rotor

Channeling in column

Replace or repack column

Column overloaded

Use higher-capacity stationary phase; increase


column diameter; decrease sample amount

Basic solutes - silanol


interactions

Use competing base such as triethylamine; use


a stronger mobile phase; use base-deactivated
silica-based reversed-phase column; use
polymeric column

Beginning of peak
doubling

See peak doubling

Chelating solutes - trace


metals in base silica

Use high purity silica-based column with low


trace-metal content; add EDTA or chelating
compound to mobile phase; use polymeric
column

Silica-based column degradation at high pH

Use polymeric, sterically protected, or highcoverage reversed-phase column; install silica


gel saturatorcolumn between pump and
injector

Silica-based column degradation at high


temperature

Reduce temperature to less than 50 C

Silica-based column silanol interactions

Decrease mobile-phase pH to suppress silanol


ionization; increase buffer concentration;
derivatize solute to change polar interactions

Unswept dead volume

Minimize number of connections; ensure


injector rotor seal is tight; ensure all
compression fittings arecorrectly seated

Replace column, or, if possible, open top


endfitting and clean and fill in void with glass
Void formation at head of
beads or samecolumn packing; rotate injection
column
valve quickly; use injection valve with
pressure bypass; avoid pressure shock
Spikes

Bubbles in mobile phase Degas mobile phase; use back-pressure


restrictor at detector outlet; ensure that all

abhishekraj4@gmail.com

fittings are tight


Column stored without
caps

abhishekraj4@gmail.com

Store column tightly capped; flush reversedphase columns with degassed methanol

You might also like