HPLC Troubleshooting: Problem Possible Cause Solution
HPLC Troubleshooting: Problem Possible Cause Solution
HPLC Troubleshooting: Problem Possible Cause Solution
Abhishek Raj
Deputy Manager (R&D)
Quest Pharmaceuticals Pvt. Ltd, Chhata Pipara, Bara, Birgunj, Nepal.
abhishekraj4@gmail.com
Possible cause
Detector off
No peaks or
very small
peaks
Solution
Check detector
Broken connections
Check connections
to recorder
No sample/Wrong
sample
Wrong settings on
Check attenuation. Check gain
recorder or detector
Pump off
Start Pump
Flow interrupted
Leak
No Flow
Possible cause
Loose fitting
abhishekraj4@gmail.com
Solution
Tighten or replace fitting
leaks
White powder at
loose fitting
Leak at pump
Possible cause
Solution
Contamination buildup
Equilibration time
Pass at least 10 column volumes
insufficient for gradient run
through the column for gradient
or changes in isocratic mobile
regeneration or after solvent changes
phase
First few injections - active
Changing
sites
Retention Times
Selective evaporation of
mobile-phase component
abhishekraj4@gmail.com
and pH 8
Problem
Possible cause
Air bubbles in mobile phase
Solution
Degas or use back pressure restricor on
detector
Normal with many samples; use mobile
phase as sample solvent
Wavy or undulating -
abhishekraj4@gmail.com
Baseline
noise
temperature changes in
room
Gradient or isocratic
proportioning - lack of
solvent mixing
Gradient or isocratic
proportioning malfunctioning
proportioning valvesl
Random - contamination
buildup
Spikes - column
temperature higher than
boiling point of solvent
Pressure
Problem
Decreasing
Pressure
Possible cause
Solution
Pump cavitation
abhishekraj4@gmail.com
Fluctuating
pressurre
Bubble in pump
Column particle size too small (for Use larger particle size (for
example 3 micrometers)
example 5 micrometer)
High Back
Pressure
Systematically disconnect
components from detector end to
column end to find blockage;
replace or clean blocked
component
Increasing
Pressure
abhishekraj4@gmail.com
column
Peaks
Problem
Possible cause
Solution
Broad
peaks
Ghost
Mobile-phase solvent
viscosity too high
Peak dispersion in
injector valve
Contamination
abhishekraj4@gmail.com
HPLC-grade solven
Elution of analytes
retained from previous
injection
peaks
Negative
peaks
Oxidation of
trifluoroacetic acid in
peptide mapping
Reversed-phase
chromatography contaminated water
Unknown interferences
in sample
Refractive index
detection - refractive
index of solute less than
that of mobile phase
UV-absorbance detection
Use mobile phase with lower UV absorbance;
- absorbance of solute
if recycling solvent, stop recycling when
less than that of mobile
recycled solvent affects detection
phase
Peak
Doubling
Blocked Frit
Coelution of interfering
compound
Peak
Fronting
Tailing
Peaks
Channeling in column
Column overloaded
Beginning of peak
doubling
abhishekraj4@gmail.com
abhishekraj4@gmail.com
Store column tightly capped; flush reversedphase columns with degassed methanol