PAL User Manual Installation and Operation Combi Pal / Pal LHX Pal Combi-Xt / Pal LHX-XT
PAL User Manual Installation and Operation Combi Pal / Pal LHX Pal Combi-Xt / Pal LHX-XT
PAL User Manual Installation and Operation Combi Pal / Pal LHX Pal Combi-Xt / Pal LHX-XT
Printing History
Original Instructions
© 1999 – 2013 CTC Analytics AG. All rights reserved. Neither this
publication nor any part hereof may be copied, photocopied, reproduced,
translated, distributed or reduced to electronic medium or machine readable
form without the prior written permission from CTC Analytics AG, except as
permitted under copyright laws.
CTC Analytics AG acknowledges all trade names and trademarks used as the
property of their respective owners.
1•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
General Considerations
The PAL System User Manual and related documents must be consulted
by the user under all circumstances before a unit is put to use.
The user should be aware that if the equipment is used in a manner not
specified by the manufacturer, the protective and safety features of the
equipment may be impaired.
Environmental Conditions
The environmental conditions in which the PAL System can be operated are
listed in the corresponding section of the PAL User Manual; see the section
‘Specifications’.
2•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
Electrical Hazards
The installation category (over-voltage category) for this instrument is Level II.
The Level II category pertains to equipment receiving its electrical power from
a local level, such as an electrical wall outlet.
Connect the PAL System only to instruments complying with IEC 61010 safety
regulations.
The power line and the connections between the PAL System and other
instruments used in the configuration setup of the total analytical system must
maintain good electrical grounding. Poor grounding represents a danger for
the operator and may seriously affect instrument performance.
Do not connect the PAL System to power lines that supply devices of a heavy
duty nature, such as motors, refrigerators and other devices that can generate
electrical disturbances.
Use only fuses of the type and current rating specified. Do not use repaired
fuses and do not short-circuit the fuse holder.
The supplied power cord must be inserted into a power outlet with a
protective earth contact (ground). When using an extension cord, make sure
that the cord also has an earth contact.
If the supplied power cord does not fit the local electrical socket and a
replacement or adapter has to be purchased locally, make sure that only a
certified power cord is used. Any power cord used should be certified by the
appropriate local authorities.
Pay attention not to leave any cable connecting the PAL System and the
chromatographic system or the power cord close to heated zones, such as the
injector or detector heating blocks or the GC hot air vents.
Always replace any cable showing signs of damage with another one
provided by the manufacturer. Safety regulations must be respected.
3•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
The combination of a PAL System with an LC/MS System also requires that
the safety measures described by the LC/MS System manufacturer be
followed. Detailed instructions for the safety grounding on the LC/MS System
are outlined in the corresponding operating/installation manual.
Do not turn the instrument on if you suspect that it has incurred any kind of
electrical damage. Instead, disconnect the power cord and contact a CTC
Analytics representative for a product evaluation. Do not attempt to use the
instrument until it has been evaluated. Electrical damage may have occurred if
the PAL System shows visible signs of damage, exposure to any liquids or has
been transported under severe stress.
Damage can also result if the instrument is stored for prolonged periods
under unfavorable conditions (e.g. subjected to heat, moisture, etc.).
Ensure that the power supply/controller unit is always placed in a clean and
dry position. Avoid any liquid spills in the vicinity.
4•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
Never try to repair or replace any components of the instrument that are not
described in the PAL User Manual without the assistance of a CTC Analytics
representative.
The power supplies for the PAL System and the Peltier Stack Module have the
symbols I/0 on the power switch label to indicate ON/OFF.
5•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
Warning The installed Laser device is a CDHR/IEC Class II/2 Laser Product.
Class II/2 Laser devices are not considered to be hazardous when used for
their intended purpose. The following statement is required to comply with
US and international regulations:
For more details see label on the product or manufacturer specifications and
safety warnings.
6•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
Other Hazards
Do not operate the PAL System without the safety guard. The safety guard
must be installed for safe operation.
Do not place any objects inside the area of the safety guard. Keep away from
the area around the safety guard during operation of the PAL System.
Danger of crushing to fingers and hands. To avoid injury keep your hands
away from moving parts during operation. Turn off the power to the PAL
System if you need to reach inside a mechanically powered system with
moving parts.
When using the PAL System, follow the generally accepted procedures for
quality control and method development.
7•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
Use caution when working with any polymer tubing under pressure:
Do not reuse headspace vials. During the process of washing the vial, micro-
cracks can form which will weaken the glass wall and increase the chances of
the vial breaking.
In case of a single fault situation where the temperature control of the Agitator
fails, there is the potential danger that the device will heat up in an un-
controlled manner until it reaches the cut-off temperature of the over-
temperature fuse, in this case, 240 °C.
8•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
Based on this single fault scenario, when working with flammable solvents,
the user must ensure that the solvent used has a flash point which is 25 °C
higher than the maximum potential temperature (240 °C) of the Agitator.
Before using dangerous substances (toxic, harmful, etc.) please read the
hazard indications and information reported in the Material Safety Data Sheet
(MSDS) supplied by the manufacturer referring to the relevant CAS (Chemical
Abstract Service) number. The PAL System requires the use of several
chemical products with different hazard characteristics, which are present in
vials and syringes. Before using these substances or replacing the syringe,
please read the hazard indications and information reported in the MSDS
supplied by the manufacturer referring to the relevant CAS number.
During sample preparation, please refer to local regulations for the ventilation
conditions of the work room.
All waste materials must be collected and eliminated in compliance with the
local regulations and directives in the country where the instrument is used.
9•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
Biological Hazards
10•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
Maintenance
11•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
Disposal
Warning The customer has to ensure that the PAL System has not been contaminated
by any hazardous chemical or biological compounds including (but not
limited to) bacteria or viruses.
Any part which has had direct contact with the analytical sample must be
identified and must undergo an appropriate decontamination procedure
prior to shipping for disposal. Potentially dangerous components are:
Syringes, Valves, Loops, Tubing, Vials and Well Plates.
Any critical parts sent for disposal must be handled according to national
laws for hazardous compounds.
The customer and the service engineer/technician are fully responsible for
enforcing these requirements. CTC Analytics AG will hold the representative
and/or customer responsible if these regulations are not observed.
12•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
Regulatory Compliance
CTC Analytics declares, under sole responsibility, that the product as originally
delivered complies with the requirements of the following applicable
European Directives and carries the CE marking accordingly:
Safety:
13•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
CTC Analytics AG hereby declares that our product has been tested and
complies with the requirements of FCC Part 15 Subpart B (Class B).
This device complies with part 15 of the FCC Rules. Operation is subject to
the following two conditions:
14•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
Referenced Documents
Directive 2002/95/EC on
Restriction of the use of certain Hazardous Substances, RoHS
Directive 2002/96/EC on
Waste Electrical and Electronic Equipment, WEEE
Confirmation
We herewith confirm that the aforementioned product(s) does/do not
contain any of the reportable substances (substances banned or to be
declared) as listed in the referenced documents in concentrations above the
limits as specified therein, or – if there are no concentration limits specified –
in concentrations being intentionally introduced.
The design and production of the electrical and electronic equipment takes
into account the dismantling and recovery of WEEE and its components and
materials.
WEEE Marking
The products that are subject to the WEEE Directive shipped after July 1st
2006 (date of enforcement of WEEE Directive by the European Union) are
compliant with the WEEE marking requirements.
See WEEE symbol above, section ‘Disposal’. This symbol indicates that the
product must not be disposed of with household waste.
Declaration of Compliance
All PAL Systems including the optional modules comply with RoHS and WEEE
requirements. There is no substance to be declared.
Disclaimer
All information in this declaration is given to the best of our present
knowledge and beliefs. The information given does not constitute a warranty
within the meaning of warranty law.
15•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
Make sure that you follow the precautionary notices presented in this guide
and in any additional documentation such as the PAL User Manual. Safety
and other special notices include the following:
Warning For your safety, and in compliance with international regulations, the
physical handling of this CTC Analytics instrument requires a team effort to
lift and/or move the equipment.
This instrument is too heavy and bulky for one person alone to handle
safely.
Before lifting or moving the PAL System, secure all moving parts first to avoid
crushing hands or tipping over the unit.
Warning This is the general safety warning symbol and safety alert word to prevent
actions that could cause personal injury. It highlights a general hazard or
signifies the presence of another specific safety symbol. This is usually
followed by an appropriate safety precautionary message. When you see a
safety alert on your instrument or in the documentation, please carefully
read the safety instructions before proceeding.
16•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
Symbol Description
Biological Hazard
Direct Current
Alternating Current
Fuse
Electrical Power ON
Used with Main PAL Power Supply
17•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
B. Table of Contents
A. Safety Information and Warnings for
Users of the PAL System ................................................................... 2
B. Table of Contents ............................................................................. 18
C. List of Figures .................................................................................... 21
D. How to Use this Manual .................................................................. 23
E. COMBI PAL Operating Instructions ................................................ 24
1. Using the Control Terminal ..................................................................... 24
1.1. Menu Screens ................................................................................................24
1.2. Function Keys.................................................................................................25
1.3. ESCape and STOP Keys ...............................................................................25
1.4. Scroll Knob and ENTER Button ..................................................................25
2. Methods...................................................................................................... 26
2.1. Creating Methods..........................................................................................26
2.2. Edit / View Methods .....................................................................................27
2.3. Delete Methods .............................................................................................27
3. Job and Job Queue ................................................................................... 28
3.1. Building and Starting a Job Queue ...........................................................28
3.2. Aborting a Job Queue ..................................................................................29
3.3. Restarting an aborted Job Queue .............................................................30
4. Utility Functions......................................................................................... 30
4.1. Syringe .............................................................................................................31
4.2. Tray ...................................................................................................................32
4.3. Injector .............................................................................................................33
4.4. Wash Station ..................................................................................................33
4.5. Vial ....................................................................................................................34
4.6. Dilutors ............................................................................................................34
4.7. Tools .................................................................................................................35
5. Logfile ......................................................................................................... 35
6. Info Functions ............................................................................................ 36
6.1. Hardware.........................................................................................................36
6.2. Software ..........................................................................................................36
6.3. Maintenance...................................................................................................37
6.4. Free Objects / Free Items ............................................................................37
7. Setup Functions ......................................................................................... 37
7.1. Sounds .............................................................................................................38
7.2. Time .................................................................................................................38
7.3. Objects.............................................................................................................38
18•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
19•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
20•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
C. List of Figures
Figure 1 PAL Control Terminal and Conventions
Figure 2 Accessing a Method Screen
Figure 3 Verifying Tray with corresponding Tray Type
Figure 4 Example of Job Queue Screen
Figure 5 Selecting Utilities Functions
Figure 6 Selecting Info Functions
Figure 7 Selecting Setup Functions
Figure 8 COMBI PAL Major System Components
Figure 9 Attachment of Mounting Claws
Figure 10 Attaching the PAL Injection Unit
Figure 11 Connecting the Injection Unit Ribbon Cable
Figure 12 Inserting the Injection Unit Mounting Torx Screws
Figure 13 COMBI PAL with Injection Unit and Standalone Supports
Figure 14 Installing the Keypad Terminal
Figure 15 Installing the Flush Gas Regulator
Figure 16 Installing the Wash Station
Figure 17 Installing the Agitator
Figure 18 Installing a Tray Holder
Figure 19 Electrical Connections for Combi PAL System
Figure 20 Electrical Connections for PAL COMBI-xt System
Figure 21 Object Reference Position
Figure 22 Menu Screen Object Tray Holder
Figure 23 Inclined Tray, Corrections for X-,Y-,Z-Axes
Figure 24 Tray Holder Reference Position
Figure 25 Demonstrating a possible Inclination of Tray in X-,Y-,Z-Axes.
Figure 26 Tray Type VT 32
Figure 27 Tray Type DW96
Figure 28 Utilities Function ‘Trays’
Figure 29 Adjusting ‘Offset X,Y, Z’ for Tray Position 001
Figure 30 Adjusting Row Inclination at Second Tray Corner Position
Figure 31 Adjusting Column Inclination at Third Tray Corner Position
Figure 32 Wash Station Wash1/Wash2 Reference Point
Figure 33 Wash Station Waste/Waste2 Reference Position
Figure 34 Wash Station Standard Reference Position
Figure 35 GC Injectors (e.g. GC Inj1) Reference Position
Figure 36 Agitator Reference Position
Figure 37 Syringe and Syringe Adapter for Liquid Injections
Figure 38 Installing and Removing a Syringe for Liquid Injections
Figure 39 Headspace Syringe and heated Syringe Adapter
Figure 40 Installing and Removing a Headspace Syringe
Figure 41 Removing a Headspace Syringe from the
Heated Syringe Adapter
Figure 42 COM Port Settings
Figure 43 Reading MAC Address from Ethernet Communication Port
Figure 44 PAL-xt Communication Settings
Figure 45 Setting Communication Mode
21•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
22•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
Follow the description given in this manual to Install and operate the PAL
model ‘LHX PAL’, which differs only by its X-axis length of 120 cm as
compared to 80 cm for the COMBI PAL model and thus is not handled
separately. Thus, model PAL LHX-xt is also not handled separately.
Unless stated otherwise, this manual refers to all COMBI PAL models.
Safety Information
COMBI PAL Operating Instructions
Specifications
Product Warranty
COMBI PAL Description and Installation
Troubleshooting
Replacing Parts
Maintaining PAL System
Appendices
note The COMBI PAL must be installed and set up properly before the Operating
Instructions in Section E can be used.
Users who are installing a COMBI PAL system or COMBI PAL accessories, or
who need to adjust an installed system, should consult ‘COMBI PAL
Description and Installation’ in Section F.
The Appendices provide useful information such as the Software Flow Chart,
Definition of Terms, Injection Cycle parameter descriptions, or the COMBI PAL
accessories guide.
23•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
Figure 1 illustrates the PAL control terminal and the conventions used to
enter, edit, and view information.
24•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
Pressing the function key labeled Home will always return to the Job Queue
menu.
25•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
2. Methods
2.1. Creating Methods
Methods can be defined by the user and assigned names up to eight
characters in length. Methods can be created, copied, edited, and viewed
from the Methods menu. Methods can be viewed (but not edited) from the
Job Queue menus.
Use the scroll knob and the left-right arrow function keys (F2 and F3) to select
among alphanumeric characters and spaces. Press the ENTER function key
(F4) to accept the name.
1. Assign and enter a new Method name as above. After a copy of the
Method has been created, the Method parameters will display and can
be edited. The Cycle and Syringe entries cannot be changed;
2. If the Method is new (i.e. added), select and enter a Cycle that is
appropriate for the application;
note Once a Method has been created and saved, the Cycle and Syringe cannot
be changed. To use a different Cycle or Syringe, a new Method must be
created.
26•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
1. Scroll to and select the parameter to be changed. Assign the new value
and press the ENTER key;
2. Exit from Parameter List by pressing either the Home function key (F4)
to return to the top-level Job Queue menu or the ESCape key to return
to the previous menu;
27•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
Before a Job can be activated the operator must verify that the Tray Type
matches the specified Tray and vial size (type). This step is done in Utilities
class.
Select the corresponding Tray Type. By activating function key “F3” the
injection unit can be moved to the first position and to another two corner
positions to verify the correct selection of the Tray Type and teaching position.
28•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
2. Load a sample Tray onto an available location in a Tray Holder. Note the
corresponding Tray name;
note If a sample Tray contains 10 mL sample vials and the "HS-Inj" cycle is used,
insert the supplied adapters into the Agitator sample positions.
3. Add a new Job for the Tray. Press the Add Job key to bring up the default
Job;
4. For TRAY, select the Tray name (e.g. Tray1) that corresponds to the
location of the Tray that was just loaded;
5. Enter the First and Last sample number for this Job;
6. Select and enter the sample processing Method for this Job;
7. Press the Home function key (F4) to return to the JOB QUEUE screen;
10. If only one Job is to be processed, select the Job with the scroll knob.
Press Start key. In the dialog box "Select Job(s) to Process" select one
of the following options:
29•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
5. Select Job to abort processing all samples in the current Job. Processing
will resume with the next Job. The aborted Job is marked with an X;
6. Select Job Queue to abort processing of all Jobs. The JOB QUEUE screen
will be displayed. The aborted Job is marked with an X.
2. Select the Resume option. The job after the last one marked as aborted
will be started.
4. Utility Functions
Utility functions, selectable from the Menu screen, provide quick access to
checking operations and parameters that may need to be changed. These
functions are available for the actual Syringe, Trays, Injectors, and the Wash
Station. They allow access to key functions without having to set up and
execute a Method and Job.
note If an item is used in the sample processing cycle, the appropriate Utility
value will be overwritten by the Method value.
30•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
4.1. Syringe
The following functions are available by pressing a Function Key:
The following Syringe items may be changed by selecting the particular item:
Item Description
Actual ID Indicates the identification number (ID) of the currently inserted syringe. If the syringe detection
system is set to manual, the message “Syringe: No syringe" is displayed.
Fill Volume This parameter serves to control the filling of the syringe. It can occur that air bubbles remain below
the plunger after the first pull up. If the plunger is moved up and down several times (see Fill
Strokes), these air bubbles are worked out. With this operation the syringe can be completely filled
even when using very small sample volumes.
Fill Strokes Number of fill strokes. All fill strokes, except the last one, use the selected fill volume. If the
selected sample volume is higher than the fill volume, the sample volume is used for all fill
strokes. If zero is selected the plunger is pulled up only once using the sample volume value.
Pullup Del By using this item a delay time can be selected between sample pullup and ejection while filling the
syringe. When the plunger reaches the zero position during the fill strokes, the system waits half the
Pullup Del time. This allows for an air bubble to float away from the needle tip. This feature is especially
useful for removing any air bubble in the syringe and handling viscous fluids.
Fill Speed Speed of plunger movement used in all syringe filling operations.
Eject Speed Speed of plunger movement used in all syringe eject operations except sample injection.
Inject Speed Speed of plunger movement for sample injection. Typically used for “Fill Strokes”.
Standby Temp* Sets the temperature for a heated syringe during standby mode,
(e.g. before or after a Job/Job Queue is processed).
Actual Temp * Indicates the actual (current) syringe temperature
Plunger Chnge Pos Plunger position during Chang Syr operation. The syringe plunger is moved to a position where
the syringe can be removed and replaced. The value may be changed for different types of
syringes.
* Available for Agitator only
31•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
4.2. Tray
After selecting the particular Tray to be accessed, the following functions are
available:
The following Tray items may be changed by selecting the particular item:
Item Description
Needle Penetr Needle penetration depth into the sample vial. The needle penetration depth for the selected Tray
may be changed by entering the desired value.
Tray Type The Tray Type which is selected for the Tray is shown. If the Tray enables the use of different
Tray Types it can be changed at this position.
Tray Offset X If necessary, a correction to the ideal X-position of “Position 1” can be made by using
“Tray Offset X”.
Tray Offset Y If necessary, a correction to the ideal Y-position of “Position 1” can be made by using
“Tray Offset Y”.
Tray Offset Z If necessary, a correction to the ideal Z-position of “Position 1” can be made by using
“Tray Offset Z”.
dxRow Correction of any inclination of a Tray (plate) in the X-axis of a row (see Fig. 25 for details).
dyRow Correction of any inclination of a Tray (plate) in the Y-axis of a row (see Fig. 25 for details).
dzRow Correction of any inclination of a Tray (plate) in the Z-axis of a row (see Fig. 25 for details).
dxCol Correction of any inclination of a Tray (plate) in the X-axis of a column (see Fig. 25 for details).
dyCol Correction of any inclination of a Tray (plate) in the Y-axis of a column (see Fig. 25 for details).
dzCol Correction of any inclination of a Tray (plate) in the Z-axis of a column (see Fig. 25 for details).
Standby Temp* Sets the temperature for the Agitator during standby mode,
(e.g. before or after a Job/Job Queue is processed).
Actual Temp * Indicates the actual agitator temperature.
Speed * Sets the Agitator spinning speed. Usually it has to be set as high as possible, but not so high that
the liquid does not move up to the septum. Check the correct speed by inserting a spacer below
the vial inside the agitator so the liquid level is visible. Try different speeds and monitor the liquid
level.
Agi On Time * Sets how long the Agitator spins in one direction.
Agi Off Time * Sets how long the Agitator stops before it reverses the direction of spin. If this value is set to 0,
only the clockwise direction with no interrupts will be used.
*Available for Agitator only
Items displayed in “italic” are available beginning with PAL Firmware Version 4.1.X.
note A “staggered pattern” is selectable in firmware class “Tray Type”. For details see section F. “Description
and Installation”, point 9.3.1 “Tray Type”.
32•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
4.3. Injector
The following Injector item may be changed by selecting the particular item:
Item Description
Needle Penetr By selecting the parameter "Needle Penetr" the Injector Needle Penetration value can be checked
and/or changed. To ensure reproducible sample injections and minimize carryover it is critical
that the needle penetration depth be accurately set.
Item Description
Needle Penetr By selecting the Parameter "Needle Penetr" the Wash Station Needle Penetration value can be
checked and/or changed.
Rinse Time If rinse time is activated (value > 0), the solenoid of a Fast or Active Washstation opens for the
specified time after the syringe needle has been removed from the wash port (after completion
of the syringe wash cycle).
The solvent flows into the wash port without the restriction of the needle; be aware of higher
solvent consumption.
The following Wash Station items may be changed by selecting the particular
item:
Item Description
Needle Penetr By selecting the parameter "Needle Penetr" the Wash Station needle penetration value can be
checked and/or changed.
Rinse Time If rinse time is activated (value > 0), the solenoid of a Fast or Active Washstation opens for the
specified time after the syringe needle has been removed from the wash port (after completion
of the syringe wash cycle).
The solvent flows into the wash port without the restriction of the needle; be aware of higher
solvent consumption.
33•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
4.5. Vial
After selecting the specific Vial type*, the following functions are available:
Item Description
Needle Penetr By selecting the parameter "Needle Penetr" the wash station needle penetration value can be
checked and/or changed.
4.6. Dilutors
After selecting the specific Dilutor, the following functions are available:
The following Dilutor items can be changed by selecting the particular item:
Item Description
Syringe Indicates the Dilutor side-port syringe inserted in the Z-axis. This function allows
selecting another syringe size.
Syr Dilut Pos Activating this function moves the plunger of the side port syringe up by the specified
distance. This allows adjustment of the plunger tip of the side port syringe exactly
above the lower side port (solvent inlet). This fine tuning is necessary after changing a
side port syringe to allow unrestricted solvent flow.
Dilutor Syr Indicates Dilutor Syringe is installed. If the syringe size is changed, this item must be
adapted accordingly. The syringe-specific dimensions are coordinated by this function.
Prime Volume A volume to prime the Dilutor Syringe can be specified. The allowed range can be
fourfold higher than the actual syringe volume. This allows filling and emptying a
dilutor syringe more often than just once.
Pullup Delay Using this item permits selecting a delay time between solvent filling and ejection
while filling (or priming) the syringe.
Fill Speed Speed of plunger movement of Dilutor Syringe used in all syringe filling operations.
Eject Speed Speed of Dilutor Syringe plunger movement used in all dilutor syringe eject
operations.
Eject Delay Using this item permits selecting a delay time between solvent ejection and filling
while filling (or priming) the syringe.
34•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
4.7. Tools
note The Object class “Tools” is available with PAL Firmware version
3.0.X , 4.1.X. or higher.
After selecting the specific tool, the following functions become available:
The following tool items may be changed by selecting the particular item:
Item Description
Teach Point The reference point at which to teach the Object (Tool; MHETool) is selectable. In the case of
the MHETool, no extra position is necessary. The Parking Station is the fix point. Select “None.”
ToolOffset X If necessary, a correction to the ideal X-position of the MHETool can be made by using
“ToolOffsetXX”.
ToolOffset Y If necessary, a correction to the ideal Y-position of the MHETool can be made by using
“ToolOffsetY”
ToolOffset Z If necessary, a correction to the ideal Z-position of the MHETool can be made by using
“ToolOffsetZ”
A “ToolOffsetZ” of -48.0 mm is necessary to position the MHETool on the sample vial.
5. Logfile
The PAL logfile cannot be read-out directly on the terminal display.
However, the function key “F3” does allow a print out. A serial printer has to
be connected to port “SER2” for the PAL System and to the port “SER1” for
PAL-xt System. If a serial printer is not available use a serial/parallel converter.
35•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
6. Info Functions
Info functions, selectable from the Menu screen, provide quick access to
information to be viewed. These info functions are available for the Hardware,
Software Maintenance, Free Objects, and Free Items.
All items with the exception of “Maintenance” are read only.
6.1. Hardware
Item Description
CPU SNo The serial number (SNo.) of the PCB “APR CPU” is displayed-
CPU ID Version number of the PCB “APR CPU”.
MOTIO ID Version number of the PCB “APR CPU”.
6.2. Software
Item Description
PAL Firmware Firmware version of PAL System.
Head Firmware Firmware version PAL Injection Unit.
Terminal FW Firmware version PAL Terminal.
Altera Firmware Firmware version of the Altera component.
36•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
6.3. Maintenance
Item Description
PlgStrokeCnt This is a counter for syringe plunger movements. The actual number of strokes is displayed. If the
counter reaches the set limit (PlgStrkeLim), a warning is displayed at the next start of a job (run).
The system continues but signals the user to verify syringe conditions. The counter can be set
back to zero to restart.
There is only one counter for a syringe. If syringe types are changed the system continues to
count as if it were the same type.
PlgStrokeLim An upper limit for the syringe plunger strokes can be set.
Inject Count This counter monitors the number of injections. The number of penetrations into the GC injector
is a helpful tool for the user to decide on a replacement of the injector system. The actual number
of injections is displayed. If the counter reaches the set limit (PlgStrokeLim), a warning is
displayed at the next start of a job (run) (Inject Limit). The same counter is used for valve
switches (HPLC technique).
Inject Limit An upper limit for the number of injections can be set.
note Counters for the plunger movement and injector penetrations are
available with PAL Firmware version 2.5.X or higher.
7. Setup Functions
The Setup functions, selectable from the Menu screen, allow accessing
various functions for the PAL System. The “Sound”, “Time” and “Objects” are
basic functions used at installation or if changes have been made over time.
37•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
7.1. Sounds
Item Description
Message Box A specific dual beep tone signals that a pop-up window (Message Box) for user intervention
appears on the screen.
This beep signal can be turned on or off.
Warn Move A beep sound is heard at the start of the PAL movement. It is advisable to keep this function turned
on for safety reasons.
End Cycle A beep sound is heard at the end of a cycle. Select as desired.
End Job A beep sound is heard at the end of a cycle. Select as desired.
7.2. Time
Item Description
Year The “year” can be set for the PAL internal clock
Month The “month” can be set for the PAL internal clock.
Day of Month The “day” can be set for the PAL internal clock.
Hours The “hours” can be set for the PAL internal clock.
Minutes The “minutes” can be set for the PAL internal clock.
Seconds The “seconds” can be set for the PAL internal clock.
After setting or resetting the date and time, use the function key “F1”
“Set Time” to store.
7.3. Objects
The various PAL Firmware Object classes can be selected and the functions
most used by the user are directly accessible. For detailed listing see PAL
Firmware overview in the appendices.
38•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
39•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
1.1. Specifications
1.1.1. Sample Capacity
40•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
a: Protection Class I:
Protection class describes the insulating scheme used in the instrument to protect the
user from electrical shock. Class I identifies a single level of insulation between live
parts (wires) and exposed conductive parts (metal panels), in which the exposed
conductive parts are connected to a grounding system. In turn this grounding system
is connected to the third pin (ground pin) on the electrical power plug.
c: Pollution Degree 2:
This is a measure of pollution on electrical circuits that may produce a reduction of
the dielectric strength or surface resistivity.
Degree 2 refers to normally only non-conductive pollution.
Occasionally, however, a temporary conductivity caused by condensation must be
expected.
d: Moisture Protection:
Normal (IPXO) – IPXO means that there is NO Ingress Protection against any type of
dripping or sprayed water. The X is a place holder to identify protection against dust
if applicable.
41•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
42•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
A PAL System can be upgraded to a PAL-xt System; for details contact your
CTC Analytics representative.
The various PAL Software programs, such as PAL Loader or PAL Object
Manager, are operated with Microsoft Windows Operating systems, such as
Windows XP, Vista or Windows 7 Enterprise 32 or 64 bit. For details, see the
corresponding ‘Addendum to the PAL User Manual’.
The PAL System can be controlled using PAL control software, the Cycle
Composer, or any data system software that controls the PAL using the Cycle
Editor for PAL ICC interpretation (e.g. Analyst, ChemStation, Empower,
EZChrom, MassLynx, Xcalibur).
For details on software requirements for integration of the PAL System with
other data handling systems (CDS), contact your CTC Analytics representative
or the manufacturer’s representative of the other CDS software program.
The PAL System allows a serial communication protocol with a PC. As a new
feature, the PAL-xt System provides LAN communication, also referred to as
‘TCP/IP’ or ‘Ethernet’.
The PAL-xt System requires a dedicated ‘APR Control-xt’ board and PAL
Firmware version 4.1.X or higher.
Setup within Corporate Network requires the support of the system
administrator who has local administrator privileges for Windows and network
setup. For the network port 60000 the access has to be provided.
43•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
A quick way to check whether serial communication with the PAL System can
be established is to use the PAL Loader software. For details see ‘Addendum
to PAL User Manual PAL Loader Software’.
note CTC Analytics AG reserves the right to make improvements and/or changes
to the product specifications without notice.
44•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
2. Product Warranty
2.1. Statement of Limited Product Warranty
This warranty does not cover defects or failures resulting from damage
caused by accidents, neglects, misuse or abuse, including (but not limited to):
Force Majeure: CTC Analytics AG is not liable for any events beyond
reasonable control, including, but not limited to, fire, storm, flood,
earthquake, explosions, riots, strikes, labor disputes, transportation
embargoes or other contingencies beyond the control of
CTC Analytics AG (i.e. ‘acts of God’ or ‘Force Majeure’).
45•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
This warranty remains in effect for a period of one year from the date of
installation. Additionally, the warranty request must be filed within the
aforementioned warranty period.
The replacement of a module does not extend the warranty period as set out
under point ‘What is the Period of Coverage’.
46•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
To keep this warranty in effect, the user must take care to avoid accidents,
misuse and abuse, as described under point above ‘What is NOT covered by
this Warranty?’
The maintenance guidelines outlined in the PAL User Manual must be
followed. Preventative Maintenance (PM) or any service or repair task on the
instrument must only be carried out by authorized CTC Analytics personnel.
The user must report any problem with the equipment immediately after
discovery to the representative of CTC Analytics AG and assist technical
support personnel in diagnosing and attempting to resolve the instrument
malfunction. It is expected that the user will cooperate by installing any
replacement parts (modules) that have been sent and will return any
exchanges or unused replacement parts to the representative of CTC
Analytics AG, or according to prior agreement.
This warranty statement from CTC Analytics AG gives the user specific rights.
The user may also have other rights which vary from country to country or
from state to state.
This warranty statement from CTC Analytics AG gives the user specific rights. If
the warranty statement from your retailer, distributor, or OEM is broader, the
added value of that warranty may be valid for the user. This, however, is not
the responsibility of CTC Analytics AG, but solely that of the specific distributor
or OEM partner.
47•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
3. Installation
3.1. Unpacking the Components
A COMBI PAL system is shipped in one box. The box contains the X-,Y-axes
assembly, the Injection Unit, the heated Agitator, the Keypad Terminal, GC
mounting kit (optional), connecting cables, power supply, (heated) Syringe
kits, Tray Holders, Trays, Wash Station assembly, Safety Guard and
miscellaneous parts.
1. Open the box and first remove the accessory boxes and the Injection Unit
before attempting to remove the X-, Y-axes assembly;
2. Carefully lift the X-,Y-axes assembly and remove it from the box. Hold
the Y-axis in place while the assembly is being removed from the box.
Set the X-,Y-axes assembly on a bench;
4. When placing the COMBI PAL onto a GC, make sure that no objects
interfere with either the Y-axis or the Injection Unit throughout the entire
range of potential movement.
note During headspace operation, the syringe is cleaned after each injection with
clean, inert gas (e.g. N2 or He). Therefore, a 1/8 inch gas supply line must
be available close to the COMBI PAL.
2. Loosen the two mounting clamp Torx screws of the vertical legs;
3. Place the X-, Y-axes assembly on top and fit the mounting kit legs into the
groove in the X- axis. Be sure that the clamps fit completely into the
grooves. Alternately tighten the two Torx screws until the legs are firmly in
place;
4. Double check whether the leg claws are correctly attached to the X-axis
(see Fig. 9).
48•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
note Installation of the Injection Unit should be done carefully. When installing it
for the first time, have someone hold it in place while the mounting screws
are inserted.
1. Remove the three Torx mounting screws A, B, and C, used to fix the
Injection Unit to the Y-axis;
2. Connect the ribbon cable A protruding from the front end of the Y-
axis to the corresponding connector on the Injection Unit (see Fig. 11);
49•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
3. Hold the Injection Unit in place against the Y-axis. Make sure the two
locating pins on the Y-axis fit into the two guide pin holes on the Injection
Unit;
4. Place one of the screws onto the end of the supplied Torx driver. Slide the
clear plastic cover on the Injection Unit all the way to the top. Locate the
three large holes in the black anodized frame attached to the Z-axis inside
the Injection Unit. Slide the frame upwards until the top hole is centered
on the top threaded hole at the end of the Y-axis. Insert and
securely tighten the Torx screw A. (See Fig. 12.);
5. Install the two remaining Torx screws B, C in the left and right mounting
holes, respectively. It may be necessary to move the elastic cord slightly to
the left to insert the Torx screw C into the right-hand hole.
Figure 13. COMBI PAL with Injection Unit and Standalone Supports
50•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
1. Install the safety shield on the left and right sides to the outside of the X-
axis. Use the provided, longer thumbscrew on the side where you plan to
install the keypad;
2. Install the Keypad mounting bracket on either the right or left side of the
X-axis;
3. Connect one end of the white coiled cable (Part No. SS8J-700) to the
Keypad and the other end to the TERMINAL (SER3) interface jack on the
rear side of the X-axis. For details see Figures 19 or 20;
1. Locate the power supply, the DC power cable (Part No. RS3M-2000),
and the AC power cable;
3. Connect one end of the DC power cable to the power supply and the
other end to the POWER connector at the rear side of the X-axis;
4. Connect the female end of the AC power cable to the power supply.
Then connect the male end to an AC power outlet.
note Before proceeding with the remaining steps, make sure the power to the
PAL System is switched off.
51•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
note Safety Warning: Use an inert gas, such as helium or nitrogen, for the flush
gas. Hydrogen is highly flammable and explosive at certain concentrations.
1. Attach the OUT-side of the supplied pressure regulator at rear right hand
side of the COMBI PAL. (See Fig. 15.);
2. Connect a 1/8 inch clean gas supply line to the IN-side of the pressure
regulator. Use a separate gas line to avoid any pressure drop on other
equipment (e.g. GC);
3. Adjust the pressure of the gas supply source to 1.0 • 1.5 bar.
If a gas line is teed-up from an installed gas chromatograph, the preset
secondary pressure will be approx. 5 bar;
4. Adjust the COMBI PAL gas pressure regulator to approx. 0.5 bar;
note Pressure higher than 0.5 bar will not yield higher flow rates through the
heated syringe. The syringe needle is the point of flow restriction in the
gas flow system.
5. Check the installed Flush Gas pressure regulator connections for leaks.
Avoid any contamination of the pneumatic system (gas flow) by using
liquid soap or similar leak detection aids;
6. Verify the Flush Gas pressure again during the flush period in a
Headspace Cycle.
52•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
note If other types of Wash Stations are installed, follow the instructions in the
HTS PAL User Manual for the Fast Wash Station or for other types, such as
Active Washstation, in the specific Addendum of the PAL User Manual.
Observe safety measures if a waste line has to be installed in combination
with a gas chromatograph. Do not position the waste line above a heated
zone.
1. Loosen the two Torx screws on the two Agitator mounting brackets;
53•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
2. Attach the Agitator to the left or right end of the X-axis. Install it with the
mounting clamp teeth fit into the grooves on the bottom of the X-axis;
note Do not position the Agitator in the middle of the X-axis. The Agitator must be
installed on the right or left side. Crossing-over the Agitator with a 20 ml vial
would cause a collision.
3. Be sure that the clamps fit completely into the grooves. Alternately tighten
the two Torx screws until the two mounting clamps are firmly in place;
4. Double check whether the Agitator clamps are correctly attached to the X-
axis. (See Fig. 9.);
5. Connect the supplied Agitator control cable from AUX1 to the connector
on the rear side of the Agitator.
1. Loosen the two Torx screws on the two mounting clamps located on top
of the Tray Holder legs;
2. Install the Tray Holder with the mounting clamp teeth fit into the grooves
on the bottom of the X-axis;
3. Be sure the clamps fit completely into the grooves. Alternately tighten the
two torx screws until the two mounting clamps are firmly in place;
4. Double check if the two Tray Holder clamps are correctly attached to the
X-axis.
54•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
note Always switch OFF the PAL power supply before connecting or disconnecting
any COMBI PAL accessories cable! Do not interchange the Terminal with
LAN cable connectors.
Before defining the COMBI PAL object positions, make sure the Agitator and
Keypad Terminal are correctly connected to the COMBI PAL X-axis rear side.
(See Figures 19 or 20).
55•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
note Remove the Syringe Adapter from the Injection Unit before performing the
following steps.
The objective is to define the Reference Positions for all COMBI PAL Objects.
Make sure the Tray Holder(s), Agitator and Wash Station are properly
mounted to the PAL X-axis. The following description is an example of how to
teach the reference position for a Tray Holder. The described procedure is
common to all COMBI PAL Objects.
For a Tray Holder, the reference position is a hole (slightly larger than the
lower needle guide) in the base plate of the holder. The lower needle guide
should be centered in the hole with the bottom of the needle guide flush with
the bottom of the base plate (see Fig. 21).
2. Observe the keypad display. The model name COMBI PAL will display
along with the software version number. The Job Queue menu screen
will then display;
56•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
3. When the "Job Queue" menu displays, complete the following sequence
(common to all objects):
4. After selecting "Named Tray Holder", the X-, Y-, Z- positions associated
with the object will be displayed (see Fig. 22);
5. Highlight item Position X with the cursor bar and press ENTER. The
Injection Unit will move to the previously defined X-axis position;
6. Rotate the outer knob to adjust the X-axis position to the Tray Holder
reference position (see Fig. 21);
9. If the F3 button "Moveto Zero" is activated, the Injection Unit will move
to the HOME position;
10. Verify the defined X-, Y-, Z-positions by pressing F1 "Check Pos".
note PAL Firmware verson 3.0.X permits correcting for an inclined surface, for
example a Tray Holder or Tray where the Z-axis position in front differs from
the rear position. For applications such as the PAL MALDI Option, perfect
alignment of the MALDI tip is vital. This compensation for the Z-axis position
is provided in the firmware class “Tool”.
57•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
For the device “PAL MHE Tool” the PAL Firmware Object Class “Tool” is also
required. See specific Addendum to PAL User Manual.
The release of PAL Firmware version 4.1.X makes possible the correction of
all three axes, X-, Y-, Z, in the firmware class “Tray.” The correction of the
inclination can be made for a tilted Tray in the direction of the row and/or
the column. Teaching is possible in the ”Utilities” section by using the path:
Menu/Utilities/ Trays.
Note that these functions are not available for the GOMBI PAL if operated
solely on PAL Firmware version 2.4.X or higher, but neither with firmware
version 3.X.X. nor with version 4.X.X.
58•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
For a Tray Holder, the reference position is a hole in the base plate of the
Holder. The lower needle guide should be centered in the hole with the
bottom of the needle guide flush with the bottom of the base plate.
Figure 24 shows the example of a Tray Holder for a single Tray, e.g. “THldr1”
for “Tray 1” with Tray Type “VT98”.
4.2.2. Trays
In general it is not necessary to teach a Tray position. The Tray Holder has a
teaching position, as described above. A Tray is assigned to the Tray Holder.
The numbers (values) for the relevant items are stored in the firmware object
“Tray”. To complete the combination of a Tray Holder and a Tray, it is
necessary to assign the corresponding “Tray Type” to the “Tray.”
The “Tray Type” contains the geometric data for the Tray and information
such as how many samples are in a row and a column in the Tray. A Tray
Type defines the pattern and sampling sequence of sample location within a
Tray. For details see below.
As described above, the Tray is physically placed in the Tray Holder, the
firmware assigns the Tray to the Tray Holder and the Tray Type to the Tray.
This is routine procedure as long as no special circumstances need be
considered, such as customized Trays or Tray Types.
PAL Firmware version 4.1.X now makes a correction possible if the Tray
surface is not exactly horizontal and planar but is inclined in one or more
axes. At version 4.1.X this is standard and a dialog window pops up the
moment the Tray has been checked-out at the corners to verify vial positions
and heights. How to correct a possible inclination is explained below.
59•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
The PAL System defines Rows and Columns by the order in which samples
are treated. A “Row” is not associated with an X- or Y-axis. The definition is
visualized below with Tray Type VT32 and DW96:
Figure 26. Tray Type VT32 Figure 27. Tray Type DW96
The following description of the correction is done by using the path to the
“Utilities” section.
After selecting the Utilities function “Trays”, verify whether the Tray Type
assigned corresponds to the Tray placed into the Tray Holder. If this is
verified, the vial positions can be checked at the three corners of the Tray.
This is activated by the “F3” function key.
60•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
The Injection Unit moves to Tray position no. 001. With the help of the
inserted syringe is it possible to check the optimized position of the needle tip
matching with the vial or well position.
In case the position is off, make corrections in X-, Y- or Z-positions.
Continue with function key “F3” to move to the next corner to verify this
position. If the position matches, continue with function key “F3” to move the
unit to the third corner position.
If a deviation of the ideal position is observed in second or third corner, the
corrections have to be made on the position no. 001. The offset is bound to
position no. 001.
note Activating one of the parameters Offset X-, Y-, or Z-axis at the positions of
second or third corner can bend the syringe needle.
The following description of the correction is done by using the path to the
“Utilities” section.
After selecting the Utilities function “Trays”, verify whether the Tray Type
assigned corresponds to the Tray placed into the Tray Holder. If this is
verified, the vial positions can be checked at the three corners of the Tray.
This is activated by the “F3” function key.
61•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
A dialog window opens, displaying a note to “Adjust offset X-, Y-, Z for
position 001”. Continuing in this dialog will provide the user the opportunity
to adjust all three axes perfectly by the top of the vial. Please note that this
step is only valid for position number 001.
To check the next corner of the Tray, activate function key “F3” again, example
Tray Type “VT32”, position 008.
The last step will be to move to the next corner of the Tray by again activating
function key “F3”. This will be position 32 in the example of Tray Type “VT32”.
62•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
The three corner points are now adjusted for a possible inclination of the Tray
in any axis. A possible deviation from an ideal axis position for the other vials,
caused by variance from the horizontal, is interpolated by the system without
further teaching.
note For routine work using standard vials it is not necessary to compensate for a
possible inclination as long as the vial detection is within the “Z-Tolerance”
range.
note If positions of other types of Wash Stations have to be taught, follow the
instructions in the HTS PAL User manual for the Fast Wash Station or for
other types, such as Active Washstation, in the specific Addendum of the PAL
User Manual.
Observe safety measures if a waste line has to be installed in combination
with a gas chromatograph. Do not position the waste line above a heated
zone.
For the Wash Station, the reference positions are the two holes in the wash
vial caps (see Fig. 32, Wash1 and Wash2). The lower needle guide should be
centered in these holes with the bottom of the lower needle guide lightly
touching the surface of the caps.
63•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
note The Waste positions represent an "Injector" within the PAL software. It is
defined in the Object class "Injectors".
For Waste / Waste2 the reference positions are the two holes in the wash
vial caps (see Fig. 33). The lower needle guide should be centered in these
holes with the bottom of the lower needle guide lightly touching the surface
of the caps.
64•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
The reference position is the hole in the vial cap. Teaching this position is
identical for Wash or Waste vials.
note For certain GC models, centering rings are provided with the GC mounting
kit. If such a centering ring is available, it should not only be used for
teaching the injector position but also for routine work. The ring stabilizes the
needle guide of the PAL Injection Unit at the moment of needle penetration.
It acts as a guide tool and prevents needle bending.
For the GC Inj1 (GC Inj2) the reference position is the appropriate GC
Injection port. The lower needle guide should be centered on the port with
the bottom of the lower needle guide lightly touching its surface.
65•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
note Steps 2+3 are obsolete beginning with Firmware version 2.2.0 or higher.
4. Open the white Agitator cover manually and block it open with a
screwdriver, fitting it into the side of the cover;
6. Attach the red vial transport ring to the lower needle guide;
7. Define the X-, Y-, Z- positions common to all objects as described in topic
4.1. The reference position is located at the hole in the steel vial cap. The
lower needle guide should be centered on the vial cap and should touch
the cap;
note After all positions have been set, perform a firmware backup of the existing
configuration by using the PAL Loader software. For instructions, see below.
66•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
5. Syringes
5.1. Selecting Syringes
The standard COMBI PAL is shipped with two syringe kits:
The supplied syringe kits cover the standard liquid and headspace GC
injection techniques. However, certain applications may require different
syringe sizes. Consult "COMBI PAL Accessories" for detail.
note Before a syringe is inserted, the position “Change Syr” should first be
verified. The position “Change Syringe” should not be located above an
object where the needle could collide with a vial, as for example.
Path:
Menu/Utilities/Syringe/function key “F3” “Change Pos”.
Move the PAL Injection Unit to a location where a collision cannot occur by
selecting the X- and Y- axes. The value for the Z-axis is given as a default and
a change of this position is not necessary in standard operation.
1. Select Menu and press F1/Chang Syr. The Injection Unit will move to a
location that will facilitate installation of the syringe;
67•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
3. Place the syringe in the appropriate syringe adapter. Pull the plunger out
to approximately 20% of its length. (See Fig. 37.);
5. Move the syringe, installed in the syringe adapter, partially into the
Injection Unit. First, guide the needle into the upper needle guide and
then into the lower needle guide. (. Fig. 38);
6. Insert the plunger button into the plunger holder. Allow the syringe
adapter to "click" into place by magnetic force against the syringe carrier;
68•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
2. Loosen the plunger retaining screw. Move the plunger slightly out of the
plunger holder;
3. Pull the syringe adapter out and then carefully upward to remove the
syringe adapter, with the syringe, from the Injection Unit.
3. Place the syringe in the appropriate heated syringe adapter. Pull the black
plunger out to approximately 20% of its length. (See Fig. 39.);
69•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
5. Lift the lower needle guide, sliding it into the upper needle guide. Insert
the syringe needle tip and match the magnetic pins of the syringe holder
with the counter positions on the syringe slider. Let the lower needle
guide return slowly and make sure the needle tip does not catch on the
guide. (See Fig. 40.);
6. Press the heated cartridge firmly against the Z-axis slider to ensure that the
Syringe heater unit engages. Move the plunger up (plunger holder) until
the thread of the screw catches the thread of the plunger bushing. Tighten
the screw to fix the plunger holder. (See Fig. 40.);
note The clamp provided with the Headspace syringe kit must only be installed
with a 5 mL HS-syringe. See note enclosed in the syringe kit.
7. Press "Home". The plunger moves down until it hits the mechanical stop.
This position is stored as the syringes zero volume position. Then the
Injection Unit returns to the HOME position.
3. Pull the syringe adapter out and then carefully upward to remove the
syringe adapter, with the syringe, from the Injection Unit;
70•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
Figure 41. Removing a Headspace Syringe from the Heated Syringe Adapter
5. Place the supplied gray plastic tube over the needle. This plastic tube
protects the syringe needle at shipment, but it can be used as a syringe
removal tool;
6. Using the gray plastic tube, push the Headspace Syringe through the
heated Syringe Adapter.
note Gastight Headspace syringes have very tight plungers when new. After a
short operating time at elevated temperature, the plunger tip adjusts its
diameter. Should the syringe be used later at lower temperatures, there will
be a risk of leakage. The polymer tip never expands at lower temperatures.
Therefore, a particular Headspace syringe should always be used at the
same or higher temperatures.
Certain types of integration of the PAL system into data handling or control
software, such as ChemStation, Xcalibur, Galaxie or Chromeleon, accomplish
synchronization of the Ready/Start signal done directly via software control
71•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
PAL Systems are shipped with all Cycle events pre-defined, as shown in the
table below. If a different physical signal needs to be used, then it will be
necessary to assign a new Event to the Object signal.
Inject2 (Sync Signal) Inject the second sample for “GC- Immediat
Dual” cycle (Ignore)
72•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
The setting is provided with the schematic description of the GC Sync Cable
supplied with the specific GC Mounting Kit.
The PAL Loader software has a second function of loading the firmware or a
complete backup file to a PAL system.
The PAL Loader software has always been a part of the PAL System.
The following PAL Loader Software versions are in use:
- PAL Loader Version 1.0.1
- PAL Loader Version 1.1.1
A new PAL Loader software, “Loader” version 2.1.1 (or higher), has been
developed for the PAL-xt system.
The PAL-xt system requires PAL Firmware 4.1.X or higher, which calls for the
new “Loader” software.
The name of the *.exe file has been changed intentionally to avoid conflicts
when old and new versions are installed on the same computer:
Details are described in: “Addendum to PAL User Manual, PAL Loader
Software Version 2.1.X, Installation and Operation”.
73•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
The PAL Object Manager software is the software used to handle the so-called
PAL Object Lists.
The software version 2.2.1 or higher has been developed for the PAL-xt
System that, in addition to serial communication, also permits LAN
communication (TCP/IP). The software is backwards compatible and can be
used for all PAL Firmware versions.
The PAL Object Lists have to match the corresponding PAL Firmware version,
as shown in the list below:
The two Object Lists for the PAL and PAL-xt Systems can be loaded with the
same PAL Object Manager software. This simplifies the handling in cases
where the PAL and PAL-xt Systems are in use within the same working
environment.
74•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
A quick way to check whether serial communication with the PAL System can
be established is to use the PAL Loader software. For details see ‘Addendum
to PAL User Manual PAL Loader Software’.
75•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
Menu/ F3-Setup/Communication
The system automatically searches for the MAC address when the ‘Setup
Communication’ window is accessed.
Figure 43. Reading the MAC Address from the Ethernet Communication Port
This search for and reading of the MAC address takes approximately 10 to 15
seconds. If successful, the MAC Address is stored in the system firmware and
can be viewed on the terminal as follows.
76•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
Pressing function key “F1”, ‘Set Comm’ sets the selected communication
mode. The same is achieved by pressing ‘ESC’ or ‘Home’.
77•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
IPAddr1 to 4 Enter the IP address divided into The IP address is a fixed PAL-xt
four blocks, example IP default System address. The IP address is
Address: 192.168.0.1. often provided by the network
administrator.
SubNet1 to 4 SubNet mask for network The SubNet mask must be
configuration. provided by the network
administrator.
TCPPort This parameter is read only, no The network administrator must
changes possible. be informed about this setting.
The port must be cleared of
firewall protection to enable
communication.
StdGateway If Gateway is used in network, Gateway is used if independent
1 to 4 the Gateway IP address (switch, networks need to communicate
hub) has to be entered. interactively. Communication
within the network is provided by
a hub or switch, defined as
“Gateway”.
MACAddr1 to6 The unique MAC Address for the The MAC address is provided by
PAL-xt System is read and the Ethernet communication
entered into this PAL-xt hardware. The address is also
communication protocol. visible beside the read-out from
This is a read only parameter the terminal on the label affixed
field. to the Ethernet module on the
PCB “APR Control-xt”.
78•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
DHCP:
If “DHCP” mode is active, an automatic search for the IP address is activated.
If a free IP address is found, the communication between the PAL-xt System
and the network becomes established. The new IP address (found through
“DHCP” search) is immediately entered into the PAL-xt Communication
protocol as seen from the local terminal. The parameters “IPAddr1 to 4” are
not updated until one opens the communication menu again.
The Setup for the PAL Loader software requires entering the IP address or
hostname. The dynamically provided IP address is displayed on the local
terminal. As an alternative, the following setup procedure can be used:
Open the “Setup” window and enter the following name in the “Hostname
or IP Address” field: CXXXXXX. The “Xs” being the last six characters of the
MAC address, which is known from the parameter “MacAddr4, MacAddr5,
and MacAddr6.
If a MAC address only consists of one character, insert a 0 in front of it. If, for
example, MacAddr4=BA, MacAddr5=35 and macAddr=3, enter CBA3503 and
not CBA353.
79•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
9. Special Functions
This section describes special functions which can be activated by the “PAL
Extended User Mode”.
The purpose of having two software access levels is to display discrete items
and sections of the firmware at each level. Settings which need to be done
rarely, perhaps at the time of installation of the PAL System, are hidden at
level 1(User Level) and revealed at level 2 (Extended User Level). This also
protects the PAL System in group use. Nevertheless, the important points
should be known by the user and are described below.
In this PAL User Manual or in technical notes, only the following will be
referred to:
Menu/F3-Setup/.
80•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
Object Description
Ident The “Identification” of the system provides the opportunity to assign Users, site,
and system names. The serial number of the PAL System can be entered at this
level. Remark: After loading a fresh PAL Firmware Backup file, the SNo. Is
displayed as “XXXXXX”.
Access Access can be selectively opened or limited to different users for the “Job”,
“Method” or “Setup”.
Sounds The “Beep” sound can be adjusted to specific needs. For safety reasons it is
advisable to keep the default settings, especially the “Warn Move,” turned ON to
signal the start of the PAL System to the user.
Time The PAL internal clock can be set.
Remark: If the clock is fast, reset all items, from “Year” to “Seconds” and use the
mandatory function key “F1” “Set Time”.
Communication The communication mode serial or LAN can be selected and conditions are
defined.
For details see above point 8 or Addendum PAL Loader Software.
System Basic parameters can be predefined for specific needs. Details see below.
Service Basic service tests can be activated. Details see below.
Objects Accessing the PAL Firmware class “Objects” in the Extended User mode provides
more detailed possibilities for object items. At this level an Object can be copied
or deleted by using the specific function keys. This provides quick access to add
e.g. a second or third injector.
Items displayed in “italic” are available beginning with PAL Firmware Version 4.1.X.
Items displayed in “bold” are explained in detail below.
81•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
9.2.1. System
The following “System” Items can be changed by selecting the particular item.
Beginning with PAL Firmware version 4.1.X, an intermediate step was added.
After selecting “System”, the new “Settings” level follows. Activating “Settings”
opens up the described Items below.
Item Description
Syr Detect The syringe detection can be turned off in cases of sensor failure or if a special
syringe with unknown ID is installed.
Turning OFF the syringe identification sensor automatically displays a dialog box
enabling the selection of syringe identification manually.
Start Ref This item provides the choice of referencing all axes at the start of the system or
suppressing the referencing. It is advisable to use this function actively to detect a
possible loss of steps.
Stop XY Error This function allows turning off the automatic recovery of the PAL system after
detection of a collision (loss of steps of a stepper motor).
Stop XY Error = ON: Automated recovery turned OFF
Stop XY Error = OFF: Automated recovery turned ON
In PAL Firmware version 2.3.X the new feature that recovers automatically after
detecting a collision has been implemented. Firmware version 2.6.2 provides the
possibility of disabling automated recovery.
If recovery is turned OFF, the PAL system always checks its position before moving to
the injection port for injection. This can lead to an extra move, crossing over the X-,Y-
axes sensors – if the injection port is at the opposite end of the X-axis relative to the
sample location, (zero X-axis position in-between sample location and injection
port).
For critical application fields, e.g. clinical analyses, it is advisable to turn off the
automated recovery; Stop XY Error= ON.
PlgPathCheck The plunger travel path is monitored to ensure that the plunger tip reaches the
syringe zero point in all cases – compared to original plunger referencing and
defining the syringe zero point.
If an error is detected, the PAL System stops operation.
This feature is useful if sample solutions with particles need to be handled, or if the
syringe type is not adequate for the solution composition, or if the syringe plunger is
blocked or even bent, such an error will be recognized and the system will come to
an emergency stop.
PlgChnge Pos Changing a syringe using the Menu function “Chnge Syr” moves the plunger up to
the position specified by this item. For normal use, accept the default values.
Init Syr at The syringe plunger can be initialized (referenced) at “Home” or “Waste” position.
If samples with a high danger of toxicity are treated, it is advisable to use the “Waste”
position for referencing.
Inj Signal An Autosampler is usually the master in a total chromatographic system. The other
components send a “Ready” signal to enable a start of the Autosampler. At the
moment of injection a Start signal is sent from the PAL System to the other
component(-s) of the entire system. With this “Inj Signal” item the status at which
the signal is sent out can be defined. The syringe is filled with the defined sample
volume.
PlgUp = Start signal at the moment the plunger starts moving down for injection.
PlgDown = Start signal at the moment the plunger reached zero point from syringe.
(recommended for large volume injection, but not in combination with Agilent
Chemstation.)
ValveSw = Start signal sent at the moment the injection valve is switched, HPLC
technique.
PreInj = Start signal sent at a time before injection, as with a sample preparation
device started before a GC or LC system. Minus time relative to chromatographic
start time.
82•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
Vial PrePress This item can be activated to avoid a vacuum if a relatively large amount of sample is
taken out of a small volume vial. If e.g. a volume of 50 µL is specified for injection,
the PAL system first takes 50 µL ambient air and “injects” this volume into the sealed
vial, causing overpressure in the vial. After this step, the sample volume can be
aspirated without forming a vacuum.
This functionality is only valid for liquid sampling, not for the Headspace technique.
Starting with PAL Firmware version 4.01X this item is also active in combination with
Cycle Composer.
LC-Inj Standard cycle for HPLC technique. Details on method parameters see Appendices,
point 3, “Injection Cycles” and point 4, Flow Chart “Firmware Overview”
LC-Cut Cycle for column switching with two valves, HPLC technique. Both valve drives have
to be controlled through “AUX” interfaces.
Details on method parameters, see Appendices, point 4, Flow Chart “Firmware
Overview”
GC-Inj Standard cycle for GC technique. Details on method parameters, see Appendices,
point 3, “Injection Cycles” and point 4, Flow Chart “Firmware Overview”.
GC-InS Standard cycle covering GC Sandwich technique. Details on method parameters, see
Appendices, point 3, “Injection Cycles” and point 4, Flow Chart “Firmware
Overview”.
GC-Dual Cycle for GC technique which allows injecting two samples in two different injectors.
The start signal is sent after injection of both sample solutions.
Details on method parameters, see Appendices, point 3, “Injection Cycles” and point
4, Flow Chart “Firmware Overview”.
HS-Inj Standard cyle for GC headspace technique. Details on method parameters, see
Appendices, point 3, “Injection Cycles” and point 4, Flow Chart “Firmware Overview”
SPME Standard cycle for GC SPME technique, Solid Phase Micro Extraction, Fiber
technique. Details on method parameters, see Appendices, point 4, Flow Chart
“Firmware Overview”
TTS The two cycles are dedicated cycles for “dissolution applications”. TTS cycle is used
Sotax for transdermal patches and the Sotax cycle is used for tablet dissolution
Items displayed in “italic” are available beginning with PAL Firmware Version 4.1.X.
note Standard injection cycles are selectable and activated directly from the PAL
System. Any deviation from a standard cycle, a Cycle Composer Macro or
Cycle Editor (ICC-CE) Cycle must be programmed for customized
requirements.
note In PAL Firmware versions < 4.0 the standard cycles are grouped in
Item “Cycles”.
83•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
9.2.2. Service
Item Description
Check Motors Every stepper motor can be moved separately. This function is useful to check
out a specific motor or to perform an endurance test after a repair.
The Start and Stop positions of the test path can be selected. The motor
Current, Speed, and Acceleration are selectable (accept default value if not
secure). A Pause time can be specified. A Cycle Limit and Actual Cycle Counter
are useful tools for endurance tests.
1
Test Head This tests the functionality of the PAL Injection Unit (“Head”).
Two tests can be activated by the function keys:
“F1” Check Plg: The stepper motor of the plunger drive is tested. Sensor
check.
“F2” Check Z: The vial detection sensor and “Needle Guide Blocking” function
is tested.
These tests are only required if any erratic behavior of the PAL System is
observed. In such cases contact a representative of CTC Analytics AG.
Test Switches X-Limit = Actual status of X-axis sensor. Move the X-Y-Carriage manually
across the limit switch to verify activation of the sensor.
Y-Limit = Actual status of X-axis sensor. Move the Y-axis manually to the Y-
zero point (at X-axis), crossing over the limit switch to verify activation of the
sensor.
Z-Limit = Actual status of Z-axis sensor. Move the syringe slider assembly
manually up to the zero Z-axis point, crossing over the limit switch to verify
activation of the sensor.
Aux1 Limit = Actual status of Aux1 limit sensor
Aux2 Limit = Actual status of Aux2 limit sensor
Plg –Pos = Actual status of Plg –Pos
Needle Guide = Actual status of needle guide sensor
Items displayed in “italic” are available beginning with PAL Firmware version 4.1.X.
1
Please note that the service test ‘Test Head’ is executed in the same way, independent of
the PAL Firmware version. However the test evaluation is different between firmware versions
2 - 3 and version 4.
The updated PAL Service Manual, Edition 1.1, will include the PAL-xt System and will explain
these differences.
84•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
Object Description
Syringes At this level all loaded syringes are selectable. At the “Utilities” user level only the
active syringe is visible.
Vials The Firmware Objects accessed on the PAL Extended User Mode have more specific
items selectable than at the User level. Tolerances or ranges such as, e.g. “Maximum
Penetration” can be defined and limited for the user such that this value cannot be
exceeded.
In the Object “Tray Types”, details for the various Items are described more closely.
Tray Types The object class “Tray Type” is only accessible via the PAL Extended User Level. All
loaded Tray Types are visible and selectable. Details, see below.
Tray Holders The Firmware Objects accessed on the PAL Extended User Level have more specific
items selectable than at the User level. Depending on the type of Tray Holder, e.g.
an Access and Restore Path or a Heater and Agitator (Motor) is assigned. Typical
example, the Tray Holder “Agitator”.
Trays All loaded Trays are visible and accessible. A Tray must be assigned to a Tray Holder.
A “TrayTypeGroup” identification checks if the specified Tray Type is allowed in this
combination. The offsets X-,Y-, Z and the inclination corrections for a row or column
are explained in the “Utilities Functions” section.
The “Path Offset X-, Y-, Z-” is used only if a “Path” is assigned to the specified Tray
Holder, typically a Stack or Agitator uses a Path.
Wash Stations The Wash station and its positions “Wash1” and “Wash2” are visible and selectable.
Details , see below.
Injectors The Firmware Objects accessed on the PAL Extended User Mode have more specific
items selectable than at the User level. Tolerances or ranges such as “Maximum
Penetration” can be defined and limited for the user so that this value cannot be
exceeded.
In the “Tray Types” Object details for the various Items are described. Furthermore,
in this Firmware Object class a “Valve” and/or a “Heater” can be assigned.
Note that the Wash Station Positions “Waste” and “Waste2” are listed in this
“Injectors” class.
85•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
Valves The Valves and Valve Drives connected and controlled by an “AUX” interface are
listed in this object class. For more details, see the Valves and Valve Drives
Addendum.
Note that the valves controlled by serial communication (Serial Valves or
Multiposition Valves) are in the PAL Firmware Object class “Serial Valves”.
Agitators At this level items such as “Motor Drive”, “Minimum Speed” and “Maximum
Speed” are selectable for the “Agitator” Object. The range can be set for the speed.
The other items are available directly as Method parameters.
Dilutors The Firmware Objects accessed in the PAL Extended User Mode have more specific
items selectable than at the User level. Tolerances or ranges such as “Maximum Fill
Speed” can be defined and limited for the user so that this value cannot be
exceeded.
Items such as “Prime Volume”, “Waste to”, “Motor Drives”, or “Scale Length” are
specific items that remain hidden at the User levels.
Tools The Object class “Tools” is used to handle another device, such as the MHE-Tool
(Multiple Headspace Extraction Tool) or MALDI Tool. For details, see the specific
Addendum for the PAL User Manual.
Positions The various positions used with the PAL System are selectable at this level. Positions
such as “Home”, “Change Syringe”, or a Path Point, reference the Paths for a
specified Tray Holder directly, such as “Agitator” or “Stack”.
Selecting an item, e.g. “Home” will provide the function key “F1”, Check Pos” on
the screen. Activating this function provides the possibility of verifying X-, Y-, Z-axes
values for this particular position.
Paths A Path defines specific movements for the Injection Unit, which represent a
repetitive task assigned to a Tray Holder. Typical examples are opening and closing
a drawer from a Stack or opening the lid of an Agitator.
The Path is composed of several steps (points) which are added up in a sequence.
Each individual step (point) can be positioned by teaching X-, Y-, Z-axes. In most
cases the return path is in the reverse order of the starting path. Example “AgiOpnL”
and “AgiClsL”. If one setting of a point is changed, this change will be valid for the
return path as well.
Sync Signals The various synchronization signals are described above in Section F. “Description
and Installation”, point 6.1. At this level no other functionality, as shown at the user
level, is provided.
Out Signals The various Out-Signals are described above in Section F. “Description and
Installation”, point 6.1. At this level no other functionality, as shown at the User level,
is provided.
Events The various synchronization signals are described above in Section F. “Description
and Installation”, point 6.1. At this level no other functionality, as shown at the user
level, is provided. Three events are newly available on this extended user level:
Pwr-Out1, Pwr-Out2 and FlushVlv. The main purpose of these three power-out
signals is to activate a solenoid, such as for the Wash station or the Gas Flush Valve.
At this level the functionality of the signal can be tested.
Serial Valves Serial Valves, or “Multiposition Valve Drives”, are powered and controlled by the
serial (RS232) control and not through the “AUX” interface. With firmware version
4.1.X a “Mode” is now selectable. The mode refers to the setting of the Valve Drive,
whether the current for the motor is set to “High Speed” or “High Torque”.
For more details, see Addendum to PAL User Manual “Valve and Valve Drive”.
Out Exp Box The “Out Expansion Box” provides 8 TTL contacts and 8 relay contacts
(24 V contact closure). The optional module (box) is connected and controlled by
“Interface 2”.
For more details, see Out Expansion Box Addendum to PAL User Manual.
RS232 This Object allows selecting the serial ports for the specific needs. Choices are:
Remote, Terminal, Printer, Barcode, ExtDev, VICI-Vlv (for serial valve drives), or
None.
86•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
Motors In the class ‘Motors’, the Objects ‘Motor-Y’ and ‘Motor-Z’ are accessible, to enable
the user to select specific motor parameters in case a specific application requires an
adjustment.
Motor-Y
Strategy
Three different settings are selectable for the Y-Motor movement strategy:
Retract = Y-axis moves from any point first to zero Y (back to the x-axis) before the
X-movement is started. This mode allows bypassing an object (such as a large GC
detector) which is in the travel path if the mode ‘Rectang’ is used.
Path Speed = Y-axis motor speed in conjunction with a “Path”, example opening
and closing a drawer of a Stack. In certain situations it is necessary to slow down the
speed to open and close a drawer. Typical application example is fraction collection,
collecting directly into a well plate. If the liquid level is high, the liquid can wash over
into the next row if drawer movement is too fast.
Remark: At PAL Firmware version 4.1.X these items of the Y motor have been made
available at the Extended User Level.
Motor-Z
For the ‘Motor-Z’, following speeds are selectable:
Penetr Speed = This is the speed used to penetrate the syringe needle into any
object, with the exception of the injector. Example: Vial penetration.
Inj Penet Spd = Injector Penetration Speed. With PAL Firmware version 2.X and 3.0
the injector penetration speed is identical to the injector retraction speed.
Inj Retr Spd = Injector Retract Speed. With PAL Firmware version 4.1.X, the injector
speed was separated into a penetration and a retraction speed. This allows safe
penetration of a vial by any type and diameter of syringe needle without bending
through the septum of the GC injector. The needle retraction speed can be
increased to speed up the entire injection process. This new feature enables the
user to avoid uncontrolled evaporation of any remaining sample solution in the
needle.
The visibility of the Motor-Z parameters at the Extended User Level’ was introduced
with PAL Firmware version 4.1.3.
Items displayed in “italic” are available beginning at PAL Firmware Version 4.1.X.
Items displayed in “bold” are explained in detail below.
87•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
The following object “Tray Type” Items may be changed by selecting the
particular item. First select the desired Tray Type, such as “VT32-20”.
note The definition of Row and Column is given above in Section F. “Description
and Installation, point 4.2.2.1.
By reversing the first six entries, the order of injection can be changed.
Example: Tray Type “VT32-20” has a “Row Length Y” and a “Col Length X”
defined. Reversing to “Row Length X” and “Col Length Y” would start with
sampling from left to right, as used e.g. with Tray Type “DW96”.
The values of the items “Spl Per Row” and “Spl Per Col” must also be
reversed.
Item Description
Row Length X Defines the Row length in the X-direction from the center of the first to the
center of the last vial position, measured in mm.
Row Length Y Defines the Row length in the Y-direction from the center of the first to the
center of the last vial position, measured in mm.
Col Length X Defines the Column length in the X-direction from the center of the first to the
center of the last vial position, measured in mm.
Col Length Y Defines the Column length in the Y-direction from the center of the first to the
center of the last vial position, measured in mm.
Spl Per Row Number of samples per Row.
Spl Per Col Number of samples per Column
Pattern The pattern of the Row or Column arrangement has to be defined.
Regular = square arrangement (standard)
Staggrd+ = Rows are arranged in staggered pattern.
Offset of second Row is shifted by +50% of hole pattern.
Staggrd-= Rows are arranged in staggered pattern.
Offset of second Row is shifted by -50% of hole pattern.
See Graphics below.
TrayTypeGroup Software protection so that not every Tray Type can be placed on any Tray
Holder. Example: VT32-20 cannot be fit into a Stack.
Plate Thickn Thickness of Tray plate bottom (not the total height of the Tray).
Measured in mm.
Vial Height Total height of the vial including cap to seal if applicable. Measured in mm.
Remark: The sum of the plate thickness and vial height adds up to the total
height This is the point where the Injection Unit expects an object.
Vial Trnsprt The mode to transport a vial is defined by:
None = No transport required. Example liquid sampling.
Magnet = Magnetic Transport. Requires magnetic vial caps. The Injection Unit
moves to the side to wipe off the vial when moved away.
Needle = Device is transported with the syringe needle inserted. Example:
special vials where a magnetic cap cannot be crimped, e.g. “Vacutainer”.
Remark: This item became active beginning at PAL Firmware version 2.5.X.
ZSlideOffRetr If magnetic transport is selected, the distance can be defined to move the
syringe slider up (Z-direction), lifting up the vial by magnetic force. Measured
in mm.
YSlideOffDist If magnetic transport is selected, the distance can be defined to move the
Injection Unit to the side, Y-direction, to wipe off the vial. Measured in mm.
Barcode If a Barcode Reader is installed, the mode has to be set for “AutoFix”.
None = No Barcode Reader option requested for this Tray Type.
88•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
Z Tolerance A tolerance window to give a plus/minus range (expressed in mm) where the
Injection Unit must expect an object.
If the value of this item is > 0, this is a relative detection mode and the sensor
from the Injection Unit needle guide is active.
If the value of this item is set to “0” the sensor of the needle guide is turned off
and the syringe slider moves to an absolute value, Z-axis position as specified in
the object.
Z Retract The distance to lift up the syringe slider before the Y-, Z-axes assembly moves
across the unit can be specified, expressed in mm.
Max Penetr The maximum allowed Needle Penetration can be defined as a safety so that
the user cannot destroy the needle tip by going too far down.
The Tray Type is prepared for a specific vial size; this value is therefore unique
to this combination. The maximum needle penetration must not exceed the
length of the syringe needle, considering the loss due to needle guide, etc..
Typical needle length is 51 mm, maximum penetration must not exceed
48 mm.
Needle Penetr This item is identical to that used on the User level or if available as a method
parameter of a local cycle. The value defined on these levels is mirrored to the
Extended User Level.
Spl1 Offset X The “X” offset is defined as the distance from the outer edge of the Tray in the
X-direction to the center of the vial position 1. Measured in mm.
Spl1 Offset Y The “Y” offset is defined as the distance from the outer edge of the Tray in
Y-direction to the center of the vial position 1. Measured in mm.
Spl1 Offset Z The “Z” offset can be selected for special cases where e.g. the syringe needle
has to reach a deeper point before penetration into a vial. A “Z” offset value = 0
considers the Tray surface or vial top as the reference point. Measured in mm.
Items displayed in “italic” are available beginning with PAL Firmware Version 4.1.X.
note The dimensions used for Row/Column Length or X-, Y-, Z-Offsets for Sample
Number 1, are specific for a particular Tray Type. If a general adjustment for
a Tray from a specific vendor – e.g. for a Deepwell Plate – has to be made,
this should be done in the “Tray Type” class of Object. For fine tuning, the
Offset items from the class “Trays” can be used.
89•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
90•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
9.3.2. Trays
The following object “Tray” Items can be changed by selecting the particular
item. First select the desired Tray, such as “Tray1”.
Item Description
Tray Holder The Tray Holder to which the Tray is assigned is specified.
TrayTypeGroup All Tray Type Groups which are allowed for this particular combination of Tray
and Tray Holder are displayed. The “TrayTypeGroup” for the Tray Type itself is
specified for each Tray Type.
Tray Type The Tray Type to be used is linked to the Tray.
This item is usually set at User level in “Utilities/Trays”. See above.
Offset X The “X” offset is defined as the distance from the center of the teaching position
from the Tray Holder to the edge of the Tray in X-direction. Measured in mm.
This item is available at User level in section “Utilities/Tray” and can be used for
fine tuning a Tray installed in a Tray Holder. Important for well-plates 384
installed in a Stack.
Remark: Position#1 is the only correction point.
Offset Y The “Y” offset is defined as the distance from the center of the teaching position
from the Tray Holder to the edge of the Tray in Y-direction. Measured in mm.
This item is available at User level in section “Utilities/Tray” and can be used for
fine tuning a Tray installed in a Tray Holder. Important for well-plates 384
installed in a Stack.
Remark: Position#1 is the only correction point.
Offset Z The “Z” offset is defined as the distance from the plate surface of the Tray
Holder to the teaching point, black needle guide flush underneath plate of Tray
Holder (which corresponds to plate thickness of Tray Holder). Measured in
mm.
This item is available at User level in section “Utilities/Tray” and can be used for
fine tuning a Tray installed in a Tray Holder. Important for well-plates 384
installed in a Stack.
Remark: Position#1 is the only correction point.
Path Offset X The Path Offset “X” can be specific for a Tray, adjustment in X-direction.
Expressed in mm.
Remark: This item is only valid if a Path is assigned to the Tray Holder as
specified in this combination.
Path Offset Y The Path Offset “Y” can be specific for a Tray, adjustment in Y-direction.
Expressed in mm.
Remark: This item is only valid if a Path is assigned to the Tray Holder as
specified in this combination.
Path Offset Z The Path Offset “Z” can be specific for a Tray, adjustment in Z-direction.
Expressed in mm.
Remark: This item is only valid if a Path is assigned to the Tray Holder as
specified in this combination.
dxRow Correction for inclination of a Tray in X-Row direction. For details, see Section
4.1.2.2 above.
dYRow Correction for inclination of a Tray in Y-Row direction. For details, see Section
4.1.2.2 above.
dzRow Correction for inclination of a Tray in Z-Row direction. For details, see Section F.
“Description and Installation”, point 4.1.2.2 above.
dxCol Correction for inclination of a Tray in X-Column direction.
For details, see Section F. “Description and Installation”, point 4.1.2.2 above.
dyCol Correction for inclination of a Tray in Y-Column direction.
For details, see Section F. “Description and Installation”, point 4.1.2.2 above.
dzCol Correction for inclination of a Tray in Z-Column direction.
For details, see Section F. “Description and Installation”, point 4.1.2.2 above.
Items displayed in “italic” are available beginning with PAL Firmware Version 4.1.X.
91•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
The following object “Wash Station” items may be changed by selecting the
particular item. First select first the desired Wash Station, such as “Wash1”.
Item Description
Position X X-axis position for Wash Station.
Position Y Y-axis position for Wash Station.
Position Z Z-axis position for Wash Station.
Z Tolerance A tolerance window for a plus/minus range (expressed in mm) where the
Injection Unit has to expect an object. If the value of this item is > 0, this is a
relative detection mode and the sensor from the Injection Unit needle guide is
active. If the value of this item is set to “0” the sensor of the needle guide is
turned off and the syringe slider moves to an absolute value, Z-axis position as
specified in the object.
Z Retract The distance to lift up the syringe slider before the Y-, Z-axes assembly moves
across the unit can be specified, expressed in mm.
Max Penetrat The maximum allowed Needle Penetration can be defined as a safety so that
the user cannot destroy the needle tip by going too far down.
The maximum needle penetration must not exceed the length of the syringe
needle, considering the loss due to needle guide, etc.. Typical needle length is
51 mm, maximum penetration must not exceed 48 mm.
Needle Penetr This item is identical to that at the User level. The value defined at this level is
mirrored to the Extended User Level.
Type The specific mode for the Wash station in use has to be defined:
Standard = Standard Wash Station. Syringe is filled with wash solvent and
expelled into Waste.
Pulse = A pulsed signal is sent to activate the device. Used for the “Fast Wash
Station” and “Active Wash Station”. See Type ‘Active’ for details.
Flow = Continuous flow of liquid by gravity. No electrical contact required for a
solenoid.
DLW = Dynamic Load & Wash. Item used to activate specific solenoid for DLW
Wash Station.
Active = Used for Active Wash Station in conjunction with the Wash Station
Print with PNo. ‘APR WASHPUMP B.A.’ (Revision B). Simplified print, activation
of pumps is controlled by PAL Firmware. Requires Object Lists
‘PAL WashActive_Rev. B’.
Remark: Type ‘Active’ is not used for Fast Wash Station operation and cannot
be used for first generation Active Wash Station with print Revision A installed.
For these cases is the type ‘Pulse’ required.
Flow Control The power source to activate a Wash Station is specified. For the Fast, Active
Wash Station and DLW the Event “Power-Out1” and “Power-Out2” is used.
Ndl Cln Path A Path can be assigned to the Wash Station.
Waste to The Waste position is assigned to the Wash Station.
Clean Volume A percentage of the syringe maximum volume is defined for the cleaning step.
Clean Count The counter monitors the number of wash cycles. The setting in the method
has the same functionality and the value is mirrored to the Extended User
Level.
Max Fill Spd The fill speed for the syringe is defined with this item.
If the maximum value of the Fill Speed (10.0 mL/s) is selected, the system takes
the injection syringe Fill Speed as defined in the method.
Max Eject Spd The Eject Speed is the speed used to eject the wash solvent during the wash
cycle. This speed is usually higher than the Fill Speed.
If the maximum value of the Eject Speed (10.0 mL/s) is selected, the system
takes the injection syringe Eject Speed as defined in the object class “Syringe”.
Rinse Time Rinse is the time to rinse the wash port with wash solvent after the wash cycle
of the syringe has been finished.
In the case of Fast/Active Wash Station, the syringe needle is pulled out of the
wash port. The wash solvent flows without restriction. As a result, higher solvent
consumption has to be taken into consideration.
Items displayed in “italic” are available beginning with PAL Firmware Version 4.1.X.
92•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
Symptom or
Possible Cause Recommended Action
Error Message
No or very low Clogged syringe. Remove syringe and aspirate/dispense liquid manually.
detector signal is Clean syringe.
observed. Bent needle. Inspect and/or change syringe.
No sample liquid is injected. Check and/or adjust Needle Penetration into sample vial
(see section 4.2).
Sample volume too low. Increase sample volume.
Syringe does not fill Air bubbles below Syringe Increase method parameter "Fill Strokes".
properly. plunger. It can happen that air
bubbles remain below the
plunger after the first pull up. If
the plunger is moved up and
down several times, these air
bubbles are worked out.
Sample Dirty syringe. Increase Pst Cln Slv1/Slv2 values in method.
peaks/responses are Use PreClnSlv1/Slv2 and PreClnSpl.
not reproducible. Syringe pressure differences. Increase Pullup Delay value.
Vacuum created in sample vial. Reduce sample volume in sample vial. Use setting under
”F3-Setup/System/PrePressureVial”.
Highly volatile solvent. Use gastight syringe and sample tray cooling.
Not properly crimped vials. Check vial cap by attempting to rotate by hand. Loose caps
may cause selective loss of lighter components from
sample. Adjust crimping tool correctly.
Method Parameters Check recommended Method Parameters in Flow Chart
“Firmware Overview”, Appendix 4.
The following parameters are crucial and should be
verified first:
- Fill speed
- Pullup delay
- Injection Speed
- Post Inj Delay
Excessive carryover Dirty syringe. Increase Pst Inj Slv1/Slv2 values in method.
between samples. Use PreClnSlv1/Slv2 and PreClnSpl.
Check and, if necessary, replace vial septa from
Wash/Waste vials.
Inappropriate wash solvent(s). Use appropriate wash solvent(s).
Wash Solvent and Waste vial Check if the syringe fills completely when washing. If
septa too tight. several cleaning cycles are programmed and many
samples are processed, then vacuum in the Wash Solvent
vials and overpressure in the Waste vials may build up.
Cut away a segment of the Solvent and Waste septa to
assure equalization of pressure.
Unexplained collisions Object(s) not defined correctly. Define Object(s) correctly. (See section F. “Describing and
of the Injection Unit. Installing”, point 4.1 “Defining Object Positions”).
Error Message Incorrect Z value for the affected Redefine Z value for the affected Object. (See section F.
"Object Not Object (e.g. GC Inj1, THldr1 or “Describing and Installing”, point 4.1 “Defining Object
Detected". Agitator). Positions”).
Error Message Agitator not connected. Switch OFF PAL and check connection between PAL and
"Invalid Motor Status Agitator.
Motor Aux1". Agitator Connection cable. Change connection cable.
MOTIO Board defective. Replace MOTIO or Control-xt Board.
93•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
Headspace specific
Possible Cause Recommended Action
Symptoms
Sample peaks or Dirty syringe. Increase method value "Syr Flushing"
responses are not Set method value "Syringe Temp" 10 °C above "Incubat
reproducible. Temp"
Syringe pressure differences. Increase Pullup delay value
Vacuum created in sample vial. Reduce sample volume
Improperly crimped vials. Check vial cap by attempting to rotate manually. Loose
caps may cause selective loss of more volatile components
from sample. Adjust crimping tool correctly.
None or too low Syringe flush Check pressure at flush gas regulator (0.5 bar) during
gas pressure syringe flushing period. Check gas flow at syringe needle
tip.
Method Parameters. Check recommended Method Parameters in Flow Chart
“Firmware Overview”, Appendix 4.
The following parameters are crucial and should be
verified first:
- Fill Speed
- Pullup Delay
- Injection Speed
- Post Inj Delay
Peak distortion or Method Parameters. See recommended actions above.
tailing.
Needle penetration speed in Check the parameters for injection technique. If all
GC injector. changes give no positive result, check the ‘Injector Retract
Speed’. See F3-setup/Objects/Motors/Motor-Z.
Solvent not suitable for Check the boiling point of the solvent and try to change
application. the solvent.
GC Injector liner not suitable. Check the liner geometry according to recommendations
of the GC manufacturer.
GC related parameters. Any parameter like injector, oven or detector temperature
can contribute to bad peak shape. Check the column type
and connections (square cut connections).
Check the troubleshooting guide of the GC Manufacturer.
Excessive carryover Dirty syringe. Increase method value "Syr Flushing"
between samples. Set method value "Syringe Temp" 10 °C above "Incubat
Temp"
None or too low Syringe Flush Check pressure at flush gas regulator (0.5 bar) during
gas pressure. Syringe Flushing period.
Check gas flow at syringe needle tip.
Unexplained Dirty flush gas. Check flush gas for impurities.
chromatographic
peaks.
Faulty temperature HS-syringe Cartridge (holder) Contact your CTC Analytics representative.
read-out. temperature heater probe or
sensor broken.
e.g. value: -99 °C.
Agitator temperature heater
probe or sensor broken.
94•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
H. Replacing Parts
note Safety Warning
Always disconnect the power cord(s) from the power supply or from the
various power supplies if optional devices are installed before attempting
any type of maintenance.
Capacitors inside the instrument may still remain charged even if the
instrument is turned off.
1. MOTIO Board
PNo.: APR Motio
2. Carefully pull the two ribbon cable connectors upwards to free them
from the board;
note After reconnecting the two ribbon cables to the board, ensure that the cables
are folded flat and can slide inside the X-axis without damage.
95•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
96•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
2. CPU Board
PNo.: APR CPU
1. Remove the right end cover of the X-axis (facing the rear of the
chassis).
Remark: This step is only necessary if the CPU board needs to be
removed and the MOTIO Board remains in the system;
2. Slide the CPU board out and towards the right end of the X-axis. This is
necessary because the CPU board mates on its left side with a connector
on the MOTIO board;
97•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
3. Control-xt Board
PNo.: APR Control-xt
note The Control-xt board combines the MOTIO and the CPU boards in a single
control board.
This board is required for the PAL COMBI-xt System and operates only at
PAL Firmware Version 4.1.X or higher.
2. Carefully pull the two ribbon cable connectors upwards to free it from the
board;
note After reconnecting the two ribbon cables to the board, ensure that the cables
are folded flat and can slide inside the X-axis without damage.
98•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
4. Agitator
Follow the sequence of steps shown in section F. “Description and
Installation”, point 3.2, “Assembling the COMBI PAL”.
5. Injection Unit
Follow the sequence of steps shown in section F. “Description and
Installation”, point 3.2, “Assembling the COMBI PAL”.
6. Wash Station
Follow the sequence of steps shown in section F. “Description and
Installation”, point 3.2, “Assembling the COMBI PAL”.
99•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
Use only a soft lint-free cloth dampened with mild Weekly or as needed
soap and water.
Replace the syringe plunger Syringe Plunger for a liquid syringe (Metal
Plunger sealing against glass) should be
cleaned on a regular basis with an aqueous
alcohol solution.
Be aware that this plunger type should not be
used for samples with a high amount of
matrices, for clinical or residue analysis. A
polymer plunger tip would be more suitable.
100•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
J. Appendices
1. Definition of Terms
Job Queue
A Job Queue is a list of sample processing Jobs. Jobs are executed in the
order displayed on the JOB QUEUE menu screens. New Jobs may be added
to the queue while samples are being processed.
Job
Cycle
Method
A Method defines how the samples are processed. The elements of a Method
are a Cycle, a Syringe and a Parameter List. Methods have names with up to
eight characters and can be edited, copied, and deleted.
Method Parameters
Tray Holder
A Tray Holder can hold one or more trays. Each Tray Holder has a reference
position (X-, Y-, Z-coordinates) that defines its location.
101•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
Tray
A Tray holds multiple samples. Trays are defined by designating the Tray Type
(see below) and the Tray Holder. Tray names are used to identify the sample
source within a PAL Job.
Tray Type
A Tray Type defines the pattern and sampling sequence of sample locations
within a Tray.
Stack
Object Class
Object Item
102•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
Objects
Module
103•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
2. Naming Convention
This section recommends the standard naming convention for COMBI PAL
Trays, Tray Types, and Tray holders. Following these conventions will allow
the COMBI PAL to be pre-configured for certain applications, will simplify
software backups and application development, and will improve technical
support and training.
104•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
105•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
Parameter Description
CYCLE GC-Inj
SYRINGE Selected syringe
Sample Volume Selected sample volume
Air Volume Volume aspirated after the syringe needle is moved out of the sample liquid
Pre Cln Slv1 Pre injection syringe wash strokes with Solvent 1
Pre Cln Slv2 Pre injection syringe wash strokes with Solvent 2
Pre Cln Spl Pre injection syringe wash strokes with Sample
Fill Volume Fill Volume used for all filling strokes except the last one
Fill Speed Speed of plunger movement used in all syringe filling operations
Fill Strokes Number of fill strokes
Pullup Del Delay time between sample pullup and ejection
Inject to Selected Injector
Inject Speed Speed of plunger movement used during sample injection
Pre Inj Del Delay time prior to sample injection
Pst Inj Del Delay time after sample injection
Pst Cln Slv1 Post injection syringe wash strokes with Solvent 1
Pst Cln Slv2 Post injection syringe wash strokes with Solvent 2
106•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
107•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
Parameter Description
CYCLE GC-InjS
SYRINGE Selected syringe
Sample Volume Selected sample volume
Solvent Plug Volume aspirated to form sandwich with Solvent Plug
Slv Source Source Vial for Solvent Plug
IntStd Volume aspirated to form sandwich with Internal Standard or Reagent
Std Source Source Vial for Internal Standard or Reagent
Air Gap(s) Air Gap(-s) to separate Solvent Plug/IntStd/Reagent/Sample
AirVolumeNdl Volume aspirated after syringe needle is moved out of the sample liquid
Pre Cln Slv1 Pre injection syringe wash strokes with Solvent 1
Pre Cln Slv2 Pre injection syringe wash strokes with Solvent 2
Pre Cln Spl Pre injection syringe wash strokes with Sample
Fill Speed Speed of plunger movement used in all syringe filling operations
Pullup Del Delay time between sample pullup and ejection
Inject to Selected Injector
Inject Speed Speed of plunger movement used during sample injection
Pre Inj Del Delay time prior to sample injection
Pst Inj Del Delay time after sample injection
Pst Cln Slv1 Post injection syringe wash strokes with Solvent 1
Pst Cln Slv2 Post injection syringe wash strokes with Solvent 2
108•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
109•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
Parameter Description
CYCLE HS-Inj
SYRINGE Selected syringe
Sample Volume Sample volume
Incubat Temp Temperature during incubation time
Incubat Time Incubation time
Agi Speed Agitator speed during incubation time
Agi On Time On interval time during incubation time
Agi Off Time Off interval time during incubation time
Syringe Temp Syringe temperature
Fill Speed Speed of plunger movement used in all syringe filling operations
Pullup Del Delay time between sample pullup and ejection
Inject to Selected Injector
Inject Speed Speed of plunger movement used during sample injection
Pre Inj Del Delay time prior to sample injection
Syr Flushing Syringe flush time
GC-Runtime GC-Runtime (calculated from GC ready to GC ready)
110•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
111•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
Parameter Description
CYCLE GC-Dual
SYRINGE Selected syringe
1. Sample Vol Sample volume for first injection
1. Air Vol Volume aspirated after the syringe needle is moved out of the sample liquid
1. Inject to Injector used for first injection
Inj Time Diff Time difference between first and second injection
2. Sample Offs Offset from first sample number to second sample number
2. Sample Vol Sample volume for first injection
2. Air Vol Volume aspirated after the syringe needle is moved out of the sample liquid
2. Inject to Injector used for second injection
Pre Cln Slv1 Wash cycles with Wash1 before first sample is aspirated
Pre Cln Slv2 Wash cycles with Wash2 before first sample is aspirated
Pre Cln Spl Rinse cycles with sample before first sample is aspirated
Int Cln Slv1 Wash cycles with Wash1 before second sample is aspirated
Int Cln Slv2 Wash cycles with Wash2 before second sample is aspirated
Pst Cln Slv1 Wash cycles with Wash1 after second sample is aspirated
Pst Cln Slv2 Wash cycles with Wash2 after second sample is aspirated
Fill Volume Fill Volume used for all Filling strokes except the last one
Fill Speed Speed of plunger movement used to aspirate sample
Fill Strokes Number of filling strokes
Pullup Del Delay time between sample pullup and ejection
Inject Speed Speed of plunger movement used during sample injection
Pre Inj Del Delay time prior to sample injection
Pst Inj Del Delay time after sample injection
112•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
113•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
114•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
115•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
116•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
note Firmware version 4.X.X., as shown in the flow chart, corresponds to the
PAL-xt System. The changes from Firmware version 2.5.X., as used for
the PAL System, are emphasized in this PAL User Manual.
117•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
118•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
119•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
120•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
121•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
122•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
123•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
124•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
Environmental analysis
Quality control in the chemical and pharmaceutical industries
Drug discovery
Metabolic studies on agrochemicals or pharmaceutical products
The classical use of the PAL or PAL-xt Systems has been chiefly as an
autosampler in chromatography, with gas chromatography (GC) and high-
pressure liquid chromatography (HPLC) being the two main applications.
Sample preparation steps such as filtration, purification or separation by solid
phase extraction (SPE), derivatization, etc. are possible in combination with
chromatographic or other techniques.
The PAL and the PAL-xt System can also be used for a variety of other
techniques or areas of application.
X-, Y-, Z-robotic systems are typically used where sample transport, liquid
handling or sample preparation need to be automated. Well known examples
include tablet dissolution testing, viscosimetry, etc.
The PAL and the PAL-xt System have been developed following the
ISO Guide 9001.
125•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
By definition, CTC Analytics AG also excludes the use of PAL or the PAL-xt
Systems directly on living animals or humans, for example in patients
(invasive devices).
In addition, the PAL or the PAL-xt System may not be used in an environment
in which explosion protection is required for safety reasons.
The ‘Intended Use’ described above is valid only for the PAL or the PAL-xt
System and does not apply to the later models of the ‘PAL System’ product
family, such as the product groups ‘PAL3 System’, which include for example
the model PAL RTC.
The PAL and the PAL-xt System, like all CTC Analytics products, must only be
used in the manner described in the corresponding User Manual and the
corresponding CTC Analytics User Manuals for the other CTC Analytics
products.
CTC Analytics assumes no responsibility for any other usage which may result
in damage to the product or personal injury.
CTC Analytics assumes no responsibility for any usage of the PAL or the
PAL-xt System which violates applicable rules, laws or regulations.
126•
COMBI PAL / LHX PAL
PAL COMBI-xt / PAL LHX -xt
User Manual
Installation and Operation
7. PAL Accessories
The accessories for the PAL System are described in the attached ‘PAL
accessories Catalog’ or on the web page: www.ctc.ch.