88488008EN Manual 848 Titrino Plus PDF
88488008EN Manual 848 Titrino Plus PDF
88488008EN Manual 848 Titrino Plus PDF
Manual
8.848.8008EN / 2015-01-19
Metrohm AG
CH-9100 Herisau
Switzerland
Phone +41 71 353 85 85
Fax +41 71 353 89 01
info@metrohm.com
www.metrohm.com
Manual
Table of contents
1 Introduction 1
1.1 Instrument description ......................................................... 1
1.1.1 Titration and measuring modes ................................................ 1
1.1.2 Connectors .............................................................................. 2
1.1.3 Intended use ........................................................................... 2
1.2 About the documentation ................................................... 2
1.2.1 Symbols and conventions ........................................................ 3
1.3 Safety instructions ................................................................ 4
1.3.1 General notes on safety ........................................................... 4
1.3.2 Electrical safety ........................................................................ 4
1.3.3 Tubing and capillary connections ............................................. 5
1.3.4 Flammable solvents and chemicals ........................................... 5
1.3.5 Recycling and disposal ............................................................. 5
4 Titrations 17
4.1 Dynamic equivalence point titration (DET) ....................... 17
4.2 Monotonic equivalence point titration (MET) .................. 17
4.3 Endpoint titration (SET) ..................................................... 18
5 Operation 19
5.1 Switching the instrument on and off ............................... 19
5.2 Fundamentals of operation ............................................... 20
5.2.1 The keypad ............................................................................ 20
5.2.2 Structure of the dialog windows ............................................ 21
6 System settings 52
6.1 Basic settings ...................................................................... 52
6.2 Managing sensors .............................................................. 56
6.2.1 General ................................................................................. 56
6.2.2 Editing the sensor data .......................................................... 56
6.3 Managing solutions ............................................................ 58
6.3.1 General ................................................................................. 58
6.3.2 Editing the solution data ........................................................ 58
6.4 Managing common variables ............................................ 60
6.4.1 General ................................................................................. 60
6.4.2 Editing common variables ...................................................... 61
6.5 File management ................................................................ 62
7 Parameters 68
7.1 Dynamic equivalence point titrations (DET) ..................... 68
7.1.1 Start conditions ..................................................................... 68
7.1.2 Titration parameters .............................................................. 69
7.1.3 Stop conditions ..................................................................... 75
7.1.4 Evaluation ............................................................................. 76
7.1.5 Calculation ............................................................................ 79
7.1.6 Statistics ................................................................................ 81
7.1.7 Reports .................................................................................. 81
7.2 Monotonic equivalence point titrations (MET) ................ 83
7.2.1 Start conditions ..................................................................... 83
7.2.2 Titration parameters .............................................................. 84
7.2.3 Stop conditions ..................................................................... 89
7.2.4 Evaluation ............................................................................. 90
7.2.5 Calculation ............................................................................ 93
7.2.6 Statistics ................................................................................ 95
7.2.7 Reports .................................................................................. 95
7.3 Endpoint titrations (SET) .................................................... 97
7.3.1 Conditioning ......................................................................... 97
7.3.2 Start conditions ..................................................................... 98
7.3.3 Titration parameters ............................................................ 100
7.3.4 Control parameters EP1 ....................................................... 102
7.3.5 Control parameters EP2 ....................................................... 105
7.3.6 Stop conditions ................................................................... 105
7.3.7 Calculation .......................................................................... 106
7.3.8 Statistics .............................................................................. 108
7.3.9 Reports ................................................................................ 108
7.4 pH calibration (CAL) ......................................................... 110
7.4.1 Calibration parameters ........................................................ 110
7.4.2 Buffers ................................................................................. 112
7.4.3 Reports ................................................................................ 113
9 Troubleshooting 115
9.1 SET titration ...................................................................... 115
9.2 Miscellaneous ................................................................... 116
10 Appendix 117
10.1 Exchange unit ................................................................... 117
10.1.1 Maximum dosing and filling rate .......................................... 117
10.1.2 Parameters for the preparing (PREP) ..................................... 117
10.2 Stirring rate ....................................................................... 118
10.3 Balance .............................................................................. 118
10.4 USB devices ....................................................................... 119
10.4.1 6.2147.000 numerical USB keypad ...................................... 119
10.4.2 Key assignment of a USB keyboard ...................................... 120
10.4.3 PC mouse ............................................................................ 120
10.4.4 Printer ................................................................................. 121
10.5 System initialization ......................................................... 121
10.6 Remote interface .............................................................. 122
10.6.1 Pin assignment of the remote interface ................................ 122
10.6.2 Status diagram of the remote interface ................................ 124
10.7 Remote control via an RS-232 connection ..................... 125
10.7.1 Commands and variables ..................................................... 126
10.8 Arithmetic algorithms in the 848 Titrino plus ............... 127
12 Accessories 135
Index 137
Table of figures
Figure 1 Front 848 Titrino plus ........................................................................ 7
Figure 2 Rear 848 Titrino plus ......................................................................... 8
Figure 3 Connecting pH or redox electrodes ................................................... 9
Figure 4 Connecting a reference electrode .................................................... 10
Figure 5 Connecting a temperature sensor .................................................... 10
Figure 6 Connecting a polarizable electrode .................................................. 11
Figure 7 Connecting a stirrer ......................................................................... 11
Figure 8 Connecting a balance ...................................................................... 12
Figure 9 Connecting USB devices .................................................................. 13
Figure 10 Connecting the USB stick ................................................................ 14
Figure 11 Connecting the 6.2147.000 USB keyboard with USB stick and
printer ............................................................................................. 14
Figure 12 Connecting the USB hub with USB stick, printer and the 6.2148.030
RS-232/USB Box (for connecting balances). ...................................... 15
Figure 13 Connecting a remote cable ............................................................. 15
Figure 14 Attaching the exchange unit ........................................................... 16
Figure 15 Reagent dosing for DET ................................................................... 17
Figure 16 Reagent dosing for MET .................................................................. 17
Figure 17 Reagent dosing for SET ................................................................... 18
Figure 18 Keypad 848 Titrino plus ................................................................... 20
Figure 19 Directory structure on the USB flash drive ........................................ 62
Figure 20 Tubbs method for determining the equivalence point ...................... 79
Figure 21 Rotational speed depending on stirring rate .................................. 118
Figure 22 Pin assignment of remote socket and remote plug ........................ 122
Figure 23 Remote status diagram DET/MET ................................................... 124
Figure 24 Remote status diagram SET ........................................................... 124
Figure 25 Remote status diagram CAL .......................................................... 125
Figure 26 Connecting the RS-232/USB Box to the PC .................................... 125
1 Introduction
CAL
Electrode calibration.
Measuring mode:
pH (calibration of pH electrodes)
1.1.2 Connectors
The instrument is equipped with the following connectors:
MSB connector (Metrohm Serial Bus)
For connecting a stirrer.
USB (OTG) connector
The 6.2151.100 adapter can be used to connect, for example, a
printer, a USB flash drive or a USB keyboard.
Sensor connectors
Four connectors for the following sensor types:
pH or redox electrodes
Reference electrodes
Polarizable electrodes
Temperature sensors (Pt1000 or NTC)
Remote connector
For connecting a Dosimat or sample changer. The Dosimat enables
automatic addition of an auxiliary solution.
CAUTION
WARNING
This instrument has left the factory in a flawless state in terms of technical
safety. To maintain this state and ensure non-hazardous operation of the
instrument, the following instructions must be observed carefully.
WARNING
WARNING
Mains voltage
WARNING
WARNING
CAUTION
Leaks in tubing and capillary connections are a safety risk. Tighten all
connections well by hand. Avoid applying excessive force to tubing
connections. Damaged tubing ends lead to leakage. Appropriate tools
can be used to loosen connections.
Check the connections regularly for leakage. If the instrument is used
mainly in unattended operation, then weekly inspections are manda-
tory.
WARNING
More details about the disposal of your old equipment can be obtained
from your local authorities, from waste disposal companies or from your
local dealer.
1
2
3
3 Coupling 4 Display
For switching the flat stopcock.
5 Keypad
5
2
6
3
7
4 8
3 Installation
3.1.2 Checks
Immediately after receipt, check whether the shipment has arrived com-
plete and without damage by comparing it with the delivery note.
3.1.3 Location
The instrument has been developed for operation indoors and may not be
used in explosive environments.
Place the instrument in a location of the laboratory which is suitable for
operation, free of vibrations, protected from corrosive atmosphere, and
contamination by chemicals.
The instrument should be protected against excessive temperature fluctu-
ations and direct sunlight.
NOTE
NOTE
The red plug must always be plugged into the red socket at the temper-
ature sensor for the purpose of shielding against disruptions.
If you use an electrode with an integrated NTC probe, then you must
plug the red plug into the red socket.
NOTE
CAUTION
Make sure that the flat side of the plug matches the marking on the
socket.
6.2151.100
CAUTION
NOTE
Most of the USB devices need a so-called hub in order to work cor-
rectly.
A USB hub is a distributor to which several USB devices can be con-
nected. USB hubs are available in specialty stores in a number of differ-
ent models.
The USB (OTG) connector of the 848 Titrino plus has no such hub. The
6.2147.000 numerical USB keypad has a USB hub and two USB con-
nectors.
Examples:
Figure 11 Connecting the 6.2147.000 USB keyboard with USB stick and
printer
USB stick
Printer
Figure 12 Connecting the USB hub with USB stick, printer and the
6.2148.030 RS-232/USB Box (for connecting balances).
1 Slide the exchange unit from the front onto the 848 Titrino plus and
push all the way to the rear.
It must snap in audibly.
4 Titrations
V/mL
Figure 15 Reagent dosing for DET
V/mL
Figure 16 Reagent dosing for MET
Control range
V/mL
t/s
5 Operation
Proceed as follows:
All tubings and the cylinder are rinsed with the PREP (Preparing)
function. The preparing of the buret unit is described in chapter
"Preparing the buret unit (PREP)", page 45.
Confirm the message with [OK].
The display of this message can be deactivated in the system set-
tings (see "PREP warning", page 54).
The main dialog is displayed:
The instrument is switched off with the [STOP] key. The fact that the key
needs to be pressed down for an extended time prevents accidental
switch off.
Proceed as follows:
The current dialog title is displayed on the left-hand side of the title bar.
The current status of the system is displayed in the upper right-hand cor-
ner:
ready The instrument is in normal status.
cond.busy The working medium is being conditioned.
cond.ok The working medium is conditioned.
busy A method has been started.
hold A method has been paused.
Some dialogs have a so-called function bar on the bottom line. The func-
tions contained therein can be selected with the arrow keys [] or []
and executed with [OK].
In the editing dialog for text input or numerical input, select the individual
characters with the arrow keys. Use [OK] to apply the character in the
input field. The following functions are available:
In a selection list, select the individual entries with the arrow keys [] and
[]. Accept the selection with [OK] or [BACK].
Variable Description
C00 Sample size
EP# Volume of endpoint EP# (# = 19)
CI# Sample identification (# = 12)
R# Result (# = 15)
FP# Volume of fixed endpoint FP# (# = 19)
CV0# Common variable (# = 15)
SMN# Mean value of result R# (# = 15)
TITER Titer of selected solution
CONC Concentration of selected solution
Var List of additional variables (see "Variables", page 24)
Templates List of predefined calculation formulas (see "Calculation
templates", page 24)
"#" stands for a sequential number that you must enter manually. Exam-
ple: if you apply the variable EP# in the formula, only EP is entered. You
will still need to enter the number yourself.
The meanings of the editing functions are explained in chapter 5.2.4,
page 22.
Variables
Pressing Var displays a list with additional variables. You can enter these
variables either directly into the formula or also by selecting them from the
list and applying them with [OK].
Variable Description
MIM Initial measured value, i.e. measured value prior to the
processing of the start conditions
MSM Start measured value, i.e. measured value after the pro-
cessing of the start conditions
MCV End volume, i.e. total dosed volume at the end of the
titration
ET# Temperature at endpoint EP# (# = 19)
EM# Measured value of endpoint EP# (# = 19)
ED# Time at endpoint EP# (# = 19)
MSV Start volume
MEN Electrode zero point pH(0)
MSL Electrode slope
DD Duration of the entire determination
MST Start temperature
MCT End temperature
FT# Temperature at fixed endpoint FP# (# = 19)
FM# Measured value of fixed endpoint FP# (# = 19)
FD# Time at fixed endpoint FP# (# = 19)
Calculation templates
Pressing Templates displays a list with calculation templates. You can
apply these templates directly with [OK].
NOTE
Some templates contain the wildcard Molw, which stands for the
molar mass of the sample. You must replace this wildcard with the cor-
rect value in the calculation formula.
Template Description
Content % Content in %
Unit of the sample size = g
Content mmol/L Content in mmol/L
Unit of the sample size = mL
Content mol/L Content in mol/L
Unit of the sample size = mL
Content g/L Content in g/L
Unit of the sample size = mL
Content ppm Content in ppm
Unit of the sample size = g
Titer Titer calculation
Unit of the sample size = g
Blank mean value Blank value as mean value of single results
Blank single value Blank value as single value
5.4 Methods
5.4.1 Creating a new method
Proceed as follows to create a new method:
If a new method has been created, then the individual parameters can be
modified under Menu Parameters.
2 Select a method
Select the desired method.
NOTE
5.5 Control
Menu Control
In the dialog Control, the settings for the execution of a single determi-
nation or of one sample series are defined.
Sample table
If this parameter is activated, the sample data for a sample series can be
entered in a table (see Chapter 5.7, page 32).
Selection on | off
Default value off
Autostart
If this parameter is activated, a new determination is started automatically
at the end of a determination. This continues until the number specified
has been reached (see Number of autostarts).
Selection on | off
Default value off
Number of autostarts
This parameter is visible only when Autostart = on.
Number of automatic starts.
Input range 1 - 50
Selection table
Default value table
table
The number of automatic starts corresponds to the number of samples
in the sample table.
NOTE
ID1
Sample identification. The sample identification can be used in calculations
as the variable CI1.
Entry max. 10 characters
Default value empty
ID2
Sample identification. The sample identification can be used in calculations
as the variable CI2.
Entry max. 10 characters
Default value empty
Sample size
Sample size. The value of the sample size can be used in calculations as
the variable C00.
Input range 999999999 - 9999999999
Default value 1.0
Unit
Unit of the sample size.
Selection g | mg | g | mL | L | pieces | User-defined
Default value g
User-defined
A user-defined unit can be created. This will be added to the selection
list. The previous entry will be overwritten as soon as the new unit has
been defined.
The sample table contains numbered lines. The identification (ID1) and
the sample size of each sample are displayed.
Edit
Edit the data of the selected line, see following chapter.
Delete
Delete the selected line from the sample table.
Insert
Insert a new line above the line selected.
New
Delete the sample table completely. This function is visible only if the
instrument is in ready status.
You will see at the very bottom the line number of the selected line and
the line number of the last line containing data. In this example, the first
line is opened and the sample table contains four lines.
One can scroll between the individual data sets with the keys [] and [].
Method
Method used for processing the sample.
Selection Selection of stored methods | empty
Default value empty
empty
The currently loaded method is used.
ID1
Sample identification. The sample identification can be used in calculations
as the variable CI1.
Entry max. 10 characters
Default value empty
ID2
Sample identification. The sample identification can be used in calculations
as the variable CI2.
Entry max. 10 characters
Default value empty
Sample size
Sample size. The value of the sample size can be used in calculations as
the variable C00.
Input range 999999999 - 9999999999
Default value 1.0
Unit
Unit of the sample size.
Selection g | mg | g | mL | L | pieces | User-defined
Default value g
User-defined
A user-defined unit can be created. This will be added to the selection
list. The previous entry will be overwritten as soon as the new unit has
been defined.
If the sample size is sent directly from the balance, then it will always be
entered in a new line at the end of the sample table. It does not matter
which line is selected or whether the sample table is even opened. In the
above example, the sample size is entered in line 5.
NOTE
If you would like to enter the sample size in a particular line, then you
must open the corresponding editing dialog (i.e. the Sample data dia-
log is displayed).
If the editing dialog for the sample size is opened, then the sent value
will be ignored.
NOTE
The axes are scaled automatically. Using Hold pauses the run. Con-
tinue then takes its place on the display. It can be used to resume
the run.
The stirring rate can be modified during titration with the Stirrer
function. The following dialog is opened by pressing [OK]:
Stir- can be used to decrease the stirring rate while Stir+ increases
it. Off switches off the stirrer. On then takes its place on the display.
This can be used to switch the stirrer back on. The [BACK] key closes
this dialog.
After the completion of the titration, the results dialog is displayed:
NOTE
NOTE
NOTE
5.10 Results
Menu Results
After the completion of the titration, the results dialog is displayed:
The calculated result and details concerning the endpoint are shown in
the overview.
Curve
Display the curve of the current determination.
Recalculate
Recalculate the current determination. The procedure will be executed
immediately.
Statistics
Display the statistical overview of a determination series (see Chapter
5.11, page 41).
The arrow keys [] and [] can be used to move to the individual measur-
ing points. A cross hair is used to show the current position on the curve.
The data (volume, measured value, time, etc.) for the respective measur-
ing point is indicated on the right-hand side.
Recalculating
NOTE
All of the results are recalculated with the Recalculate function. This is
necessary if, for example, the calculation, the titer or the sample size has
been modified.
5.11 Statistics
Menu Results Statistics
The statistical overview of a determination series can be displayed in the
Results dialog with the Statistics function.
NOTE
The mean value (Mean), the absolute and the relative standard deviation
(s abs and s rel) are displayed in the overview. For the mean value, the
number of individual results from which it has been calculated is displayed
in parentheses. In this example, it is 3. The Statistics line shows how
many determinations have already been carried out and how many deter-
minations are to be carried out in total. All three determinations were car-
ried out in this example.
Details
Display additional data.
Reset
Delete all statistics data.
Increase
Add a further determination to the determination series.
The result and the sample size of each determination are shown.
On/Off
Remove the selected determination from the statistics. The line will then
be marked with an asterisk (*), the statistics will be recalculated automati-
cally. If several calculations are defined in the method, then all the results
will be removed from the statistics.
The available subfunctions are listed for each function in the function bar.
NOTE
5.13.1 Dosing
The following dosing functions are available in the manual control:
Prepare the buret Rinse the cylinder and tubings of the buret unit
unit (PREP) (see Chapter 10.1.2, page 117).
Continuous dosing Dose while the [START] key is pressed.
(DOS)
Dose a fixed vol- Dose a specified volume.
ume (ADD)
CAUTION
Make sure that the buret tip is directed into a vessel that can
accommodate the cylinder volume of your buret unit several times
over.
NOTE
The dosing and filling rates should be decreased for viscous liq-
uids.
The maximum dosing and filling rates depend on the cylinder
volume (see Chapter 10.1.1, page 117).
4 Start dosing
Press [START].
The status changes to busy, the dosed volume is displayed. When
the volume of one cylinder has been dosed, the dosing cylinder will
be refilled automatically.
NOTE
The dosing and filling rates should be decreased for viscous liq-
uids.
The maximum dosing and filling rates depend on the cylinder
volume (see Chapter 10.1.1, page 117).
4 Start dosing
Press [START].
The status changes to busy, the dosed volume is displayed. When
the volume of one cylinder has been dosed, the dosing cylinder will
be refilled automatically.
5.13.2 Measuring
Open the dialog for manual measurement as follows:
In the function bar, select the measuring mode and press [OK].
The status changes to busy. The current measured value and the
measuring temperature are displayed.
5.13.3 Stirring
You can control a connected stirrer manually.
Proceed as follows:
6 System settings
User name
A user name can be entered here for the report. This parameter will only
be printed if a user has been defined.
Entry max. 12 characters
Default value empty
Instrument name
A instrument name can be entered here for the report. This parameter will
only be printed if a designation has been defined.
Entry max. 10 characters
Default value empty
Serial number
Serial number of the instrument. This is printed as a part of the instrument
identification in the report header.
Program version
Version number of the instrument software. This is printed as a part of the
instrument identification in the report header.
Time
Current time. Only valid numbers can be entered.
Format: hh:mm:ss
Date
Current date. Only valid numbers can be entered.
Format: YYYY:MM:DD
Language
Setting the dialog language. An additional language can be selected
besides English.
NOTE
Dialog type
The user dialog can be limited for routine operations. One can operate
normally with methods in the limited dialog. However, no settings can be
made or methods deleted.
The resetting of the dialog will take effect as soon as you exit the main
menu.
The limitation of the dialog results in the following:
The menu items System, Parameters and Control are not shown in
the main menu.
Methods can only be loaded, but not deleted, exported or created.
NOTE
If the limited dialog for routine operation is activated, then the expert
dialog cannot be activated during ongoing operation. To change the
dialog type, the 848 Titrino plus must be switched off and then back on
again. The expert dialog can be forced as soon as the instrument is
started up again. Then it is possible to enter whatever settings one
wishes, e.g. the changing of the dialog type. If the instrument is
switched off again without changing the dialog type, then the routine
dialog will remain activated.
Routine
Limited dialog for routine operations.
Contrast
The contrast of the display can be adjusted with the arrow keys [] and
[].
[]: the contrast will be decreased by one step each time the key is
pressed.
[]: the contrast will be increased by one step each time the key is
pressed.
Input range 150 - 240
Default value 212
NOTE
Beep
If this parameter is activated, then a short beep will sound in the following
cases:
When a key is pressed.
At the end of the determination.
When the system remains conditioned without interruption for 10 sec-
onds.
Selection on | off
Default value on
PREP warning
If this parameter is activated, then the recommendation will be made to
carry out the function PREP (Preparing):
After the instrument is switched on.
Each time a buret unit is attached.
All tubing and the cylinder are rinsed with this function (see Chapter
10.1.2, page 117).
Selection on | off
Default value on
Temperature sensor
The instrument supports the use of two different temperature measure-
ment techniques:
NTC (Negative Temperature Coefficient)
Pt1000 (platinum resistance)
Select the type here that has been connected to the instrument. If you use
an NTC sensor, it is required that two characteristics for the sensor be
entered in addition. These characteristics are listed in the specifications of
the sensor.
Selection Pt1000 | NTC
Default value Pt1000
R (25 C)
This parameter is visible only when Temperature sensor = NTC.
Nominal resistance of the NTC sensor at 25 C.
Input range 1000 - 99999 ohm
Default value 30000 ohm
B value
This parameter is visible only when Temperature sensor = NTC.
Material constant of the NTC sensor. B values of NTC sensors are fre-
quently based on different reference temperatures (usually 25 C and
50100 C).
Input range 1000 - 9999 K
Default value 4100 K
Type
The sensor type is displayed.
Slope
This parameter is only visible with pH electrodes.
Slope of the pH electrode. With a 1-point calibration, only pH(0) can be
calculated, 100.0% is used as the slope.
Input range 999.9 - 999.9 %
Default value 100.0 %
pH(0)
This parameter is only visible with pH electrodes.
pH value of the pH electrode at 0 mV. Apart from the slope, pH(0) is the
second characteristic of the calibration curve.
Input range 20.000 - 20.000
Default value 7.000
Calibration temp.
This parameter is only visible with pH electrodes.
Temperature at which the last calibration was carried out.
Input range 20.0 - 150.0 C
Default value 25.0 C
Calibration date
This parameter is only visible with pH electrodes.
Date of the last calibration.
Monitoring
This parameter is only visible with pH electrodes.
Activating and deactivating the calibration monitoring.
Selection on | off
Default value off
Time interval
This parameter is visible only when Monitoring = on.
You will be notified that this time interval (in days) has elapsed when start-
ing a method. You can then select whether or not you would still like to
start the method.
Input range 1 - 999 d
Default value 999 d
The name and the type are specified for each solution in the solution list.
The asterisk (*) on the right-hand side indicates that this buret unit is
attached (only for intelligent buret units). An unlimited number of solu-
tions in buret units with data chip can be added to the solution list. The
number of solutions in buret units without data chip is limited to 10 items.
Meaning of the type:
EU: exchange unit without data chip
IEU: exchange unit with integrated data chip
Edit
Edit the data of the selected solution, see following chapter.
New
Add a new solution to the list, see following chapter.
Delete
Delete the selected solution from the list.
Type
The model of the buret unit is displayed.
Cylinder volume
Cylinder volume of the buret unit in mL. The cylinder volume is automati-
cally read out with intelligent buret units.
Selection 1 | 5 | 10 | 20 | 50
Default value 20
Concentration
Concentration of the solution.
Input range 999999999 - 9999999999
Default value 1.000
Concentration unit
Unit of the concentration.
Selection mol/mL | mmol/L | mol/L | g/L | mg/L | mg/mL |
g/L | ppm | % | mEq/L | User-defined
Default value mol/L
User-defined
A user-defined unit can be created. This will be added to the selection
list. The previous entry will be overwritten as soon as the new unit has
been defined. A blank entry can be generated this way as well.
Titer
Titer of the solution.
Input range 999999999 - 9999999999
Default value 1.000
Titer unit
Unit of the titer.
Selection mol/mL | mmol/L | mol/L | g/L | mg/L | mg/mL |
g/L | ppm | % | mEq/L | empty | User-defined
Default value empty
User-defined
A user-defined unit can be created. This will be added to the selection
list. The previous entry will be overwritten as soon as the new unit has
been defined. A blank entry can be generated this way as well.
Monitoring
Activating and deactivating the titer monitoring.
Selection on | off
Default value off
Time interval
This parameter is visible only when Monitoring = on.
You will be notified that this time interval (in days) has elapsed when start-
ing a method. You can then select whether or not you would still like to
start the method.
Input range 1 - 999 d
Default value 999 d
Proceed as follows:
NOTE
If you have set the parameter Statistics to on, then the mean value of
the results will be assigned to the respective common variable.
NOTE
This menu item is visible only if a USB flash drive has been connected as
an external storage medium.
Methods can be imported and deleted from a USB flash drive in this dia-
log. Only methods located in the Files directory are displayed in the list
(see "Directory structure on the USB flash drive", page 62).
A backup of the system can be created (all data and settings). Similarly, an
existing backup can be reloaded.
Import
Import the selected method.
Delete
Delete the selected method.
Backup
Create a backup of all data and settings on the USB flash drive.
NOTE
Only one backup can be created on the same USB flash drive.
If a backup is already stored on the flash drive, then this will be over-
written as soon as the function is performed once again.
Restore
Load the backup from a connected USB flash drive.
PC/LIMS report
Specification of the storage location for the PC/LIMS report. The PC/LIMS
report is a machine-readable report with all of the important data for a
determination. It can be saved as follows:
as a TXT file on a USB flash drive.
to a LIMS via an RS-232 interface. The 6.2148.030 RS-232/USB Box is
required for this purpose.
Selection COM2 | USB Stick
Default value USB Stick
COM2
The report is sent via the serial COM2 interface. The interface parame-
ters set in the dialog COM2 settings are used (see "Editing the COM2
settings", page 65).
USB Stick
The report will be saved as a TXT file on the USB flash drive in the
folder pc_lims_report.
Printer
If a printer is connected, then the printer type needs to be defined here in
order for the reports to be printed out correctly.
The printers that have the designation ESC-POS are so-called POS printers
(point-of-sale printers), i.e. they print on continuous paper.
Selection Citizen (ESC-POS) | Custom (ESC-POS) | Epson |
Epson (ESC-POS) | HP DeskJet | HP LaserJet |
Seiko (ESC-POS)
Default value HP DeskJet
Graphics width
Adjust the width of the curve to be printed out to the paper width of the
printer to be used. The default value depends on the selected printer. The
height of the curve is 2/3 of the width.
Input range 100 - 3000 Pixels
Keyboard layout
A commercially available USB keyboard can be connected to make it eas-
ier to enter text and numbers. Specify the country-specific keyboard lay-
out.
Selection English US | French FR | German CH | German
DE | Spanish ES
Default value English US
Balance
If you have connected a balance, then you must define the balance type
here.
Selection AND | Mettler | Mettler AT | Mettler AX |
Ohaus | Precisa | Sartorius | Shimadzu
Default value Sartorius
The following table indicates the balance type that needs to be selected
for the balance model:
The interface parameters for the connected balance are set under COM1
settings.
Baud rate
Transfer rate in characters per second.
Selection 1200 | 2400 | 4800 | 9600 | 19200 | 38400 |
57600 | 115200
Default value 9600
Data bits
Number of data bits.
Selection 7|8
Default value 8
Stop bits
Number of stop bits.
Selection 1|2
Default value 1
Parity
Type of parity testing.
Selection even | none | odd
Default value none
Handshake
Type of the data transfer protocol.
Selection hardware | software | none
Default value hardware
NOTE
Program files
They are instrument-specific. The file name has the following structure:
5XXXyyyy.bin where
XXX = Instrument type (e.g. 848 for the 848 Titrino plus)
yyyy = Program version
Language files
They can be recognized by means of the two-digit language code in the
file name. A language file contains the dialog texts for various instrument
types. It is not instrument-specific. The file name has the following struc-
ture:
5848xxxxYY.bin where
Loading a file
Proceed as follows:
7 Parameters
Activation pulse
Output of an activation pulse on a remote line. This activation pulse starts
a connected Dosimat.
Selection on | off
Default value off
Start volume
Volume that is dosed prior to the start of the titration.
Input range 0.00000 - 9,999.99 mL
Default value 0.00000 mL
Dosing rate
Rate at which the start volume is dosed. The maximum dosing rate
depends on the cylinder volume (see Chapter 10.1.1, page 117).
Input range 0.01 - 166.00 mL/min
Selection max.
Default value max.
Pause
Waiting time, e.g. for the electrode to settle down after the start or a
reaction time after the dosing of a start volume.
Input range 0 - 999,999 s
Default value 0s
Request sample ID
Selection of the sample identification that is queried in the method run.
Selection off | ID1 | ID2 | ID1&ID2
Default value off
Hold at request
If this parameter is activated, then the run will be paused during the
request. If the parameter is switched off, the titration will be started in the
background.
Selection on | off
Default value on
Titration rate
Three predefined sets of parameters can be selected for the titration rate.
Selection slow | optimal | fast | user
Default value optimal
slow
For titrations in which the finest details are also to be visible. This
could, however, also lead to an increase in noise, which could result in
unwanted equivalence points.
optimal
For all standard titrations. The parameters have been optimized for the
most frequent applications.
fast
For fast and less critical titrations.
user
The individual titration parameters can be modified.
NOTE
Select optimal as titration rate when you are developing a new titra-
tion method. This parameter is suitable for virtually all titrations and
only needs adaptation in special cases.
The settings of the individual titration rates are listed in table 1, page
72.
Min. increment
This parameter is visible only when Titration rate = user.
This smallest permitted volume increment is added at the start of the titra-
tion and with steep curves in the region of the equivalence point. Very
small values should only be used if a low titrant consumption is expected;
otherwise unwanted equivalence points could be evaluated.
Input range 0.05 - 999.90 L
Default value 10.00 L
Max. increment
This parameter is visible only when Titration rate = user.
A maximum volume increment should be selected in the following cases:
when titration consumption is very low up until the equivalence point
is reached.
when a start volume is dosed up until shortly before the equivalence
point is reached.
NOTE
It is not advisable to select similar volumes for the minimum and the
maximum increment. Monotonic equivalence point titration (MET) is
appropriate for these applications.
Dosing rate
This parameter is visible only when Titration rate = user.
Rate at which the volume increments are dosed. The maximum dosing
rate depends on the cylinder volume (see Chapter 10.1.1, page 117).
Input range 0.01 - 166.00 mL/min
Selection max.
Default value max.
Signal drift
This parameter is visible only when Titration rate = user.
Maximum permissible drift for the measured value acceptance, i.e. maxi-
mum change of the measured value per minute. This type of titration is
often referred to as equilibrium titration.
NOTE
off
Measured value acceptance will take place after the maximum waiting
time has elapsed. This can be useful when the titration reaction pro-
ceeds slowly or the electrode is slow to respond.
Titration rate
slow optimal fast
Signal drift
pH, U and Ipol 20.0 mV/min 50.0 mV/min 80.0 mV/min
Upol 20.0 A/min 50.0 A/min 80.0 A/min
Min. waiting time 0s 0s 0s
Max. waiting time 38 s 26 s 21 s
Temperature
Manually entered titration temperature. If a temperature sensor is con-
nected then the temperature will be measured continuously. This value is
used for temperature correction in pH measurements.
Input range 20.0 - 150.0 C
Default value 25.0 C
Sensor
Selection of the sensor from the sensor list. The selection depends on the
measuring mode. Sensors are defined under System Sensors.
Selection Selection of configured sensors
Solution
Selection of the solution from the solution list. We recommend always
selecting the solution. This ensures that the correct data (titer, concentra-
tion, etc.) is always used for the calculation. Solutions are defined under
System Solutions.
For buret units with integrated data chip, a check is made in the method
run to verify whether the correct solution has been attached and whether
the type of dosing drive matches. For buret units without integrated data
chip, the cylinder volume and the type of dosing drive are checked. The
validity of the titer is checked for the selected solution at the start of the
determination.
Selection Selection of configured solutions | not defined
Default value not defined
not defined
No check takes place.
I(pol)
The polarization current is the current that is applied to a polarizable elec-
trode during voltametric measurement. This parameter is available only
with I(pol) determinations.
U(pol)
The polarization voltage is the voltage applied to the polarizable electrode
during an amperometric measurement. This parameter is available only
with U(pol) determinations.
Input range 1,250 - 1,250 mV (Increment: 10)
Default value 400 mV
Electrode test
In the case of polarizable electrodes, an electrode test can be carried out.
A check is made that the electrode is properly connected and that no
short-circuit is present. The electrode test is carried out when the determi-
nation is started. This parameter is available only with I(pol) and U(pol)
determinations.
Selection on | off
Default value off
Stirrer
The stirrer is switched on at the start of the determination when this para-
meter is activated.
Selection on | off
Default value on
Stirring rate
Setting the stirring rate. It can be set in steps of 15 to +15. The default
setting 8 corresponds to 1,000 rpm. The formula for calculating the rota-
tional speed is specified in chapter 10.2, page 118. The optimum stirring
rate can be tested in the manual control.
The algebraic sign of the stirring rate changes the stirring direction. When
the stirrer is viewed from above, this means:
"+": counterclockwise rotation
"": clockwise rotation
Input range 15 - 15
Default value 8
Stop volume
The titration is canceled when the specified volume has been dosed since
the start of the titration. This volume should be adjusted to the size of the
titration vessel in order to prevent the contents from running over.
Input range 0.00000 - 9,999.99 mL
Default value 100.000 mL
Selection off
Stop EP
The titration is canceled when the specified number of equivalence points
has been found.
Input range 1-9
Default value 9
Selection off
Volume after EP
This volume will be added when the number of equivalence points
defined under Stop EP has been found. The curve shape after the equiva-
lence point can also be seen this way.
Stop time
The titration is canceled when the specified time has elapsed since the
start of the titration.
Input range 0 - 999,999 s
Selection off
Default value off
Filling rate
Rate at which the dosing cylinder is filled after the titration. The maximum
filling rate depends on the cylinder volume (see Chapter 10.1.1, page
117).
Input range 0.01 - 166.00 mL/min
Selection max.
Default value max.
7.1.4 Evaluation
Menu Parameters Evaluation
The parameters for the evaluation of the titration curve are defined under
Evaluation.
Window
Activate this parameter if equivalence points are to be recognized only in
a specific measured value range (window). Only one window can be
defined.
Selection on | off
Default value off
Lower limit
This parameter is visible only when Window = on.
Measured value for the lower limit.
Upper limit
This parameter is visible only when Window = on.
Measured value for the upper limit.
EP criterion
The equivalence point criterion found (ERC = Equivalence point Recogni-
tion Criterion) is compared with this value. Equivalence points whose ERC
is less than the value defined here will be ignored.
Input range 0 - 200
Default value 5
EP recognition
This parameter allows you to filter out only the equivalence points that are
being sought.
for Window = on
Selection first | greatest | last
Default value first
first
Only the first equivalence point will be recognized.
greatest
Only the equivalence point with the greatest ERC value, i.e. the steep-
est jump, will be recognized.
last
Only the last equivalence point will be recognized.
Fixed EP1 at
The associated volume will be interpolated from the measuring point list
for the measured value entered. The fixed endpoint must lie between the
first and the final entry in the measuring point list.
Fixed EP2 at
See Fixed EP1 at.
pH
VE V/mL
Figure 20 Tubbs method for determining the equivalence point
The figure shows that the evaluation still requires measured values from
the measuring point list also after the equivalence point.
For the recognition of the EPs found, the set EP criterion is compared to
the ERC (Equivalence point Recognition Criterion) found. The ERC is the
first derivative of the titration curve combined with a mathematical func-
tion which is more sensitive for flat jumps than for steeper ones. EPs
whose ERC is smaller than the defined EP criterion will not be recognized.
The ERC is displayed in the results dialog for each discovered and recog-
nized EP. If you adjust the EP criterion after the fact in order to recognize
more or fewer EPs, then you can initiate the reevaluation in the results dia-
log with the [Recalculate] key.
7.1.5 Calculation
7.1.5.1 General
Menu Parameters Calculation
A maximum of five calculations can be defined in one method. A series of
variables (raw data from the determination, previously calculated results) is
available for the calculations. A definition can be made for each calcula-
tion as to whether the result is to be saved as a titer or as a common vari-
able.
Edit
See the following chapter for editing the data of the selected calculation.
Delete
Delete the selected calculation.
Result name
The result name is the text which will be shown in the results display and
in the report.
Entry 12 characters
Default value empty
R1=R5=
Shows the calculation formula. A special editor is opened for the defini-
tion (see Chapter 5.3, page 23).
Entry 44 characters
Default value empty
Decimal places
Number of decimal places used to display the result.
Input range 0-5
Default value 2
Result unit
The result unit is displayed and saved along with the result.
Selection % | mol/L | mmol/L | g/L | mg/L | mg/mL | ppm |
g | mg | mL | mg/piece | C | L | mL/min | User-
defined
Default value %
User-defined
A user-defined unit can be created. This will be added to the selection
list. The previous entry will be overwritten as soon as the new unit has
been defined. A blank entry can be generated this way as well.
Save as titer
The result can be saved as titer for the selected solution. If statistics has
been switched on, then the current mean value of the determination
series will be saved.
Selection on | off
Default value off
Save as CV
The calculated result can be saved as a method-independent variable,
called a common variable. The result is then also available in other meth-
ods for calculations. If statistics has been switched on, then the current
mean value of the determination series will be saved.
Selection on | off
Default value off
7.1.6 Statistics
Menu Parameters Statistics
The statistics calculation of a multiple determination is activated under
Statistics and definition is made as to how many determinations the
series contains.
Statistics
If this function is activated, then statistics calculations will be carried out
for all of the defined results.
Selection on | off
Default value off
Number of determinations
The number of determinations that are carried out for the statistics calcu-
lations.
If an additional determination has to be added to the determination
series, because one determination has been incorrect, for example, then
this can be accomplished in the statistical overview (see Chapter 5.11,
page 41).
Input range 2 - 20
Default value 3
7.1.7 Reports
Menu Parameters Reports
The reports that will be printed out automatically in connection with a
determination are defined under Reports.
Results
The result report contains the calculated results, equivalence points and
endpoints, sample data, etc.
Selection on | off
Default value off
Curve
Curve report. The width of the curve is defined in the system settings (see
"Graphics width", page 64).
Selection on | off
Default value off
Calculations/Statistics
Output of the calculation formulas for the individual results. Results are
specified with full accuracy. This makes recalculation with an external pro-
gram possible (see Chapter 10.8, page 127). If Statistics has been acti-
vated, then the following data will be printed out as well:
Result and sample size of the individual determinations
Mean value as well as absolute and relative standard deviation
Selection on | off
Default value off
Parameters
All of the parameters of the current method are printed out in the para-
meter report.
Selection on | off
Default value off
PC/LIMS
The PC/LIMS report is a machine-readable report with all of the data
important for a determination. The PC/LIMS report can be saved as a TXT
file on a USB storage medium or sent via an RS-232 interface to a LIMS.
The output location is defined in the system settings (see "PC/LIMS
report", page 63).
The file name of the TXT file is constructed as follows: PC_LIMS_Report-
ID1-YYYYMMDD-hhmmss.txt.
Selection on | off
Default value off
Activation pulse
Output of an activation pulse on a remote line. This activation pulse starts
a connected Dosimat.
Selection on | off
Default value off
Start volume
Volume that is dosed prior to the start of the titration.
Input range 0.00000 - 9,999.99 mL
Default value 0.00000 mL
Dosing rate
Rate at which the start volume is dosed. The maximum dosing rate
depends on the cylinder volume (see Chapter 10.1.1, page 117).
Input range 0.01 - 166.00 mL/min
Selection max.
Default value max.
Pause
Waiting time, e.g. for the electrode to settle down after the start or a
reaction time after the dosing of a start volume.
Input range 0 - 999,999 s
Default value 0s
Request sample ID
Selection of the sample identification that is queried in the method run.
Hold at request
If this parameter is activated, then the run will be paused during the
request. If the parameter is switched off, the titration will be started in the
background.
Selection on | off
Default value on
Titration rate
Three predefined sets of parameters can be selected for the titration rate.
Selection slow | optimal | fast | user
Default value optimal
slow
For titrations in which the finest details are also to be visible. This
could, however, also lead to an increase in noise, which could result in
unwanted equivalence points.
optimal
For all standard titrations. The parameters have been optimized for the
most frequent applications.
fast
For fast and less critical titrations.
user
The individual titration parameters can be modified.
NOTE
Select optimal as titration rate when you are developing a new titra-
tion method. This parameter is suitable for virtually all titrations and
only needs adaptation in special cases.
The settings of the individual titration rates are listed in table 2, page
87.
Volume increment
This parameter is visible only when Titration rate = user.
Volume dosed at each dosing increment. High accuracy requires using the
correct volume increment. A good guideline is 1/20 of the expected end-
point volume. The volume increment should be closer to 1/100 of the
endpoint volume for steep jumps and closer to 1/10 for flat jumps.
Small volume increments are used for determining blank values or with
very asymmetrical curves. The accuracy of the evaluation cannot be
increased by using smaller increments as the measured value changes
between two measuring points are then of the same order of magnitude
as the noise.
Input range 0.00005 - 999.900 mL
Default value 0.10000 mL
Dosing rate
This parameter is visible only when Titration rate = user.
Rate at which the volume increments are dosed. The maximum dosing
rate depends on the cylinder volume (see Chapter 10.1.1, page 117).
Input range 0.01 - 166.00 mL/min
Selection max.
Default value max.
Signal drift
This parameter is visible only when Titration rate = user.
Maximum permissible drift for the measured value acceptance, i.e. maxi-
mum change of the measured value per minute. This type of titration is
often referred to as equilibrium titration.
NOTE
Temperature
Manually entered titration temperature. If a temperature sensor is con-
nected then the temperature will be measured continuously. This value is
used for temperature correction in pH measurements.
Input range 20.0 - 150.0 C
Default value 25.0 C
Sensor
Selection of the sensor from the sensor list. The selection depends on the
measuring mode. Sensors are defined under System Sensors.
Selection Selection of configured sensors
Solution
Selection of the solution from the solution list. We recommend always
selecting the solution. This ensures that the correct data (titer, concentra-
tion, etc.) is always used for the calculation. Solutions are defined under
System Solutions.
For buret units with integrated data chip, a check is made in the method
run to verify whether the correct solution has been attached and whether
the type of dosing drive matches. For buret units without integrated data
chip, the cylinder volume and the type of dosing drive are checked. The
validity of the titer is checked for the selected solution at the start of the
determination.
Selection Selection of configured solutions | not defined
Default value not defined
not defined
No check takes place.
I(pol)
The polarization current is the current that is applied to a polarizable elec-
trode during voltametric measurement. This parameter is available only
with I(pol) determinations.
Input range 125 - 125 A (Increment: 1)
Default value 5 A
U(pol)
The polarization voltage is the voltage applied to the polarizable electrode
during an amperometric measurement. This parameter is available only
with U(pol) determinations.
Input range 1,250 - 1,250 mV (Increment: 10)
Default value 400 mV
Electrode test
In the case of polarizable electrodes, an electrode test can be carried out.
A check is made that the electrode is properly connected and that no
short-circuit is present. The electrode test is carried out when the determi-
nation is started. This parameter is available only with I(pol) and U(pol)
determinations.
Selection on | off
Default value off
Stirrer
The stirrer is switched on at the start of the determination when this para-
meter is activated.
Selection on | off
Default value on
Stirring rate
Setting the stirring rate. It can be set in steps of 15 to +15. The default
setting 8 corresponds to 1,000 rpm. The formula for calculating the rota-
tional speed is specified in chapter 10.2, page 118. The optimum stirring
rate can be tested in the manual control.
The algebraic sign of the stirring rate changes the stirring direction. When
the stirrer is viewed from above, this means:
"+": counterclockwise rotation
"": clockwise rotation
Input range 15 - 15
Default value 8
Stop volume
The titration is canceled when the specified volume has been dosed since
the start of the titration. This volume should be adjusted to the size of the
titration vessel in order to prevent the contents from running over.
Input range 0.00000 - 9,999.99 mL
Default value 100.000 mL
Selection off
Stop EP
The titration is canceled when the specified number of equivalence points
has been found.
Input range 1-9
Default value 9
Selection off
Volume after EP
This volume will be added when the number of equivalence points
defined under Stop EP has been found. The curve shape after the equiva-
lence point can also be seen this way.
Stop time
The titration is canceled when the specified time has elapsed since the
start of the titration.
Input range 0 - 999,999 s
Selection off
Default value off
Filling rate
Rate at which the dosing cylinder is filled after the titration. The maximum
filling rate depends on the cylinder volume (see Chapter 10.1.1, page
117).
Input range 0.01 - 166.00 mL/min
Selection max.
Default value max.
7.2.4 Evaluation
Menu Parameters Evaluation
The parameters for the evaluation of the titration curve are defined under
Evaluation.
Window
Activate this parameter if equivalence points are to be recognized only in
a specific measured value range (window). Only one window can be
defined.
Selection on | off
Default value off
Lower limit
This parameter is visible only when Window = on.
Measured value for the lower limit.
Upper limit
This parameter is visible only when Window = on.
Measured value for the upper limit.
EP criterion
The equivalence point criterion found (ERC = Equivalence point Recogni-
tion Criterion) is compared with this value. Equivalence points whose ERC
is less than the value defined here will be ignored.
EP recognition
This parameter allows you to filter out only the equivalence points that are
being sought.
all
All equivalence points will be recognized.
greatest
Only the equivalence point with the greatest ERC value, i.e. the steep-
est jump, will be recognized.
last
Only the last equivalence point will be recognized.
off
No evaluation takes place.
for Window = on
Selection first | greatest | last
Default value first
first
Only the first equivalence point will be recognized.
greatest
Only the equivalence point with the greatest ERC value, i.e. the steep-
est jump, will be recognized.
last
Only the last equivalence point will be recognized.
Fixed EP1 at
The associated volume will be interpolated from the measuring point list
for the measured value entered. The fixed endpoint must lie between the
first and the final entry in the measuring point list.
Fixed EP2 at
See Fixed EP1 at.
: EP volume
For the recognition of the EPs found, the set EP criterion is compared to
the ERC (Equivalence point Recognition Criterion) found. The ERC is the
sum of the measured value changes before and after the jump:
In certain cases only three or only a single summand are taken into
account.
EPs whose ERC is smaller than the defined EP criterion will not be recog-
nized. The ERC is displayed in the results dialog for each discovered and
recognized EP. If you adjust the EP criterion after the fact in order to rec-
ognize more or fewer EPs, then you can initiate the reevaluation in the
results dialog with the [Recalculate] key.
7.2.5 Calculation
7.2.5.1 General
Menu Parameters Calculation
A maximum of five calculations can be defined in one method. A series of
variables (raw data from the determination, previously calculated results) is
available for the calculations. A definition can be made for each calcula-
tion as to whether the result is to be saved as a titer or as a common vari-
able.
Result name
The result name is the text which will be shown in the results display and
in the report.
Entry 12 characters
Default value empty
R1=R5=
Shows the calculation formula. A special editor is opened for the defini-
tion (see Chapter 5.3, page 23).
Entry 44 characters
Default value empty
Decimal places
Number of decimal places used to display the result.
Input range 0-5
Default value 2
Result unit
The result unit is displayed and saved along with the result.
Selection % | mol/L | mmol/L | g/L | mg/L | mg/mL | ppm |
g | mg | mL | mg/piece | C | L | mL/min | User-
defined
Default value %
User-defined
A user-defined unit can be created. This will be added to the selection
list. The previous entry will be overwritten as soon as the new unit has
been defined. A blank entry can be generated this way as well.
Save as titer
The result can be saved as titer for the selected solution. If statistics has
been switched on, then the current mean value of the determination
series will be saved.
Selection on | off
Default value off
Save as CV
The calculated result can be saved as a method-independent variable,
called a common variable. The result is then also available in other meth-
ods for calculations. If statistics has been switched on, then the current
mean value of the determination series will be saved.
Selection on | off
Default value off
7.2.6 Statistics
Menu Parameters Statistics
The statistics calculation of a multiple determination is activated under
Statistics and definition is made as to how many determinations the
series contains.
Statistics
If this function is activated, then statistics calculations will be carried out
for all of the defined results.
Selection on | off
Default value off
Number of determinations
The number of determinations that are carried out for the statistics calcu-
lations.
If an additional determination has to be added to the determination
series, because one determination has been incorrect, for example, then
this can be accomplished in the statistical overview (see Chapter 5.11,
page 41).
Input range 2 - 20
Default value 3
7.2.7 Reports
Menu Parameters Reports
The reports that will be printed out automatically in connection with a
determination are defined under Reports.
Results
The result report contains the calculated results, equivalence points and
endpoints, sample data, etc.
Selection on | off
Default value off
Curve
Curve report. The width of the curve is defined in the system settings (see
"Graphics width", page 64).
Selection on | off
Default value off
Calculations/Statistics
Output of the calculation formulas for the individual results. Results are
specified with full accuracy. This makes recalculation with an external pro-
gram possible (see Chapter 10.8, page 127). If Statistics has been acti-
vated, then the following data will be printed out as well:
Result and sample size of the individual determinations
Mean value as well as absolute and relative standard deviation
Selection on | off
Default value off
Parameters
All of the parameters of the current method are printed out in the para-
meter report.
Selection on | off
Default value off
PC/LIMS
The PC/LIMS report is a machine-readable report with all of the data
important for a determination. The PC/LIMS report can be saved as a TXT
file on a USB storage medium or sent via an RS-232 interface to a LIMS.
The output location is defined in the system settings (see "PC/LIMS
report", page 63).
The file name of the TXT file is constructed as follows: PC_LIMS_Report-
ID1-YYYYMMDD-hhmmss.txt.
Selection on | off
Default value off
Conditioning
If this parameter is switched on, then the first time the titration is started
the working medium will be titrated to the endpoint with the specified
control parameters. The status is kept stable. The actual method run does
not begin until [START] has been pressed once more. Conditioning will
be carried out again automatically after the titration.
Selection on | off
Default value off
Start drift
Conditioning OK will be displayed as soon as this volume drift has been
reached and the titration can be started.
Input range 1 - 999 L/min
Default value 20 L/min
Drift correction
The endpoint volume can be corrected for drift. This involves multiplying
the volume drift by the drift correction time and then subtracting the
resulting value from the endpoint volume. The drift correction time is the
time interval between the end of the conditioning process and the end of
the determination.
Selection auto | manual | off
Default value off
auto
The value of the current volume drift is automatically applied at the
start of the titration.
manual
If the volume drift is known throughout a longer period of time, this
can be entered manually.
off
No drift correction takes place.
Drift value
This parameter is visible only when Drift correction = manual.
Volume drift for manual drift correction.
Activation pulse
Output of an activation pulse on a remote line. This activation pulse starts
a connected Dosimat.
Selection on | off
Default value off
Start volume
Volume that is dosed prior to the start of the titration.
Dosing rate
Rate at which the start volume is dosed. The maximum dosing rate
depends on the cylinder volume (see Chapter 10.1.1, page 117).
Input range 0.01 - 166.00 mL/min
Selection max.
Default value max.
Pause
Waiting time, e.g. for the electrode to settle down after the start or a
reaction time after the dosing of a start volume.
Input range 0 - 999,999 s
Default value 0s
Request sample ID
Selection of the sample identification that is queried in the method run.
Selection off | ID1 | ID2 | ID1&ID2
Default value off
Hold at request
If this parameter is activated, then the run will be paused during the
request. If the parameter is switched off, the titration will be started in the
background.
Selection on | off
Default value on
Solution
Selection of the solution from the solution list. We recommend always
selecting the solution. This ensures that the correct data (titer, concentra-
tion, etc.) is always used for the calculation. Solutions are defined under
System Solutions.
For buret units with integrated data chip, a check is made in the method
run to verify whether the correct solution has been attached and whether
the type of dosing drive matches. For buret units without integrated data
chip, the cylinder volume and the type of dosing drive are checked. The
validity of the titer is checked for the selected solution at the start of the
determination.
Selection Selection of configured solutions | not defined
Default value not defined
not defined
No check takes place.
Sensor
Selection of the sensor from the sensor list. The selection depends on the
measuring mode. Sensors are defined under System Sensors.
Selection Selection of configured sensors
I(pol)
The polarization current is the current that is applied to a polarizable elec-
trode during voltametric measurement. This parameter is available only
with I(pol) determinations.
Input range 125 - 125 A (Increment: 1)
Default value 5 A
U(pol)
The polarization voltage is the voltage applied to the polarizable electrode
during an amperometric measurement. This parameter is available only
with U(pol) determinations.
Input range 1,250 - 1,250 mV (Increment: 10)
Default value 400 mV
Electrode test
In the case of polarizable electrodes, an electrode test can be carried out.
A check is made that the electrode is properly connected and that no
short-circuit is present. The electrode test is carried out when the determi-
nation is started. This parameter is available only with I(pol) and U(pol)
determinations.
Selection on | off
Default value off
Stirrer
The stirrer is switched on at the start of the determination when this para-
meter is activated.
Selection on | off
Default value on
Stirring rate
Setting the stirring rate. It can be set in steps of 15 to +15. The default
setting 8 corresponds to 1,000 rpm. The formula for calculating the rota-
tional speed is specified in chapter 10.2, page 118. The optimum stirring
rate can be tested in the manual control.
The algebraic sign of the stirring rate changes the stirring direction. When
the stirrer is viewed from above, this means:
"+": counterclockwise rotation
"": clockwise rotation
Input range 15 - 15
Default value 8
Temperature
Manually entered titration temperature. If a temperature sensor is con-
nected then the temperature will be measured continuously. This value is
used for temperature correction in pH measurements.
Input range 20.0 - 150.0 C
Default value 25.0 C
Titration direction
The titration direction is normally determined automatically from the initial
measured value and the set endpoint. It is recommended that you specify
whenever possible whether the change of the measured value is positive
or negative. If two endpoints have been set then the titration direction will
be defined automatically. In this case the setting will be ignored.
Selection + | - | auto
Default value auto
+
Positive measured value change, i.e. in the direction of a higher pH
value, greater voltage or greater current.
-
Negative measured value change, i.e. in the direction of a lower pH
value, lesser voltage or lesser current.
auto
The titration direction is determined automatically from the initial mea-
sured value and the set endpoint.
Extraction time
Minimum duration of the titration. The titration will not be canceled dur-
ing the extraction time, even if the endpoint has already been reached.
The titration is, however, canceled if a stop condition is fulfilled during this
time (see Chapter 7.3.6, page 105). The entry of an extraction time may
be advisable, for instance, for the titration of sparingly soluble samples.
Input range 0 - 999,999 s
Default value 0s
Time interval MP
Time interval for entering a measuring point in the measuring point list.
The measuring point list is limited to 1000 measuring points.
Input range 0.1 - 999,999.0 s
Default value 2.0 s
Endpoint 1 at
Measured value for the first endpoint.
Titration rate
Three predefined sets of parameters can be selected for the titration rate.
Selection slow | optimal | fast | user
Default value optimal
slow
For steep titration curves for which dosing must be carried out in small
steps at the endpoint.
optimal
For all standard titrations. The parameters have been optimized for the
most frequent applications.
fast
For flat titration curves for which the endpoint is reached only slowly.
user
The individual titration parameters can be modified.
The settings of the individual titration rates are listed in table 3, page
104.
Dynamics
This parameter is visible only when Titration rate = user.
This parameter defines the control range before the specified endpoint.
Individual volume steps are dosed in the control range, the dosing is finely
controlled. The closer the endpoint, the slower the dosing until the dosing
rate defined under Min. rate has been reached. The larger the control
range, the slower the titration. Outside the control range, dosing is carried
out continuously, and the dosing rate is defined under Max. rate.
Max. rate
This parameter is visible only when Titration rate = user.
Rate at which dosing is carried out outside of the control range. The maxi-
mum dosing rate depends on the cylinder volume (see Chapter 10.1.1,
page 117).
Input range 0.01 - 166.00 mL/min
Default value 10.00 mL/min
Selection max.
Min. rate
This parameter is visible only when Titration rate = user.
Rate at which dosing is carried out at the very beginning of the titration
and in the control range at the end of the titration. This parameter has a
decisive influence on the titration rate and thus also on the accuracy. The
smaller the selected minimum rate, the slower the titration.
Input range 0.01 - 9999.00 L/min
Default value 25.00 L/min
Stop criterion
The titration is canceled when the endpoint has been reached and this
stop criterion has been fulfilled. If no stop criterion has been selected then
the titration will not be canceled. The stop conditions (see Chapter 7.3.6,
page 105) always lead to a stop, even if the stop criterion has not been
reached.
Stop drift
This parameter is visible only when Stop criterion = drift.
The titration is canceled when the endpoint and the stop drift have been
reached.
Input range 1 - 999 L/min
Default value 20 L/min
Delay time
This parameter is visible only when Stop criterion = time.
When the endpoint has been reached, the specified time is allowed to
elapse after the last dosing and the titration is then stopped.
Input range 0 - 999 s
Default value 10 s
Stop volume
The titration is canceled when the specified volume has been dosed since
the start of the titration. This volume should be adjusted to the size of the
titration vessel in order to prevent the contents from running over.
Stop time
The titration is canceled when the specified time has elapsed following the
termination of the start conditions.
Input range 0 - 999,999 s
Selection off
Default value off
Filling rate
Rate at which the dosing cylinder is filled after the titration. The maximum
filling rate depends on the cylinder volume (see Chapter 10.1.1, page
117).
Input range 0.01 - 166.00 mL/min
Selection max.
Default value max.
7.3.7 Calculation
7.3.7.1 General
Menu Parameters Calculation
A maximum of five calculations can be defined in one method. A series of
variables (raw data from the determination, previously calculated results) is
available for the calculations. A definition can be made for each calcula-
tion as to whether the result is to be saved as a titer or as a common vari-
able.
Result name
The result name is the text which will be shown in the results display and
in the report.
Entry 12 characters
Default value empty
R1=R5=
Shows the calculation formula. A special editor is opened for the defini-
tion (see Chapter 5.3, page 23).
Entry 44 characters
Default value empty
Decimal places
Number of decimal places used to display the result.
Input range 0-5
Default value 2
Result unit
The result unit is displayed and saved along with the result.
Selection % | mol/L | mmol/L | g/L | mg/L | mg/mL | ppm |
g | mg | mL | mg/piece | C | L | mL/min | User-
defined
Default value %
User-defined
A user-defined unit can be created. This will be added to the selection
list. The previous entry will be overwritten as soon as the new unit has
been defined. A blank entry can be generated this way as well.
Save as titer
The result can be saved as titer for the selected solution. If statistics has
been switched on, then the current mean value of the determination
series will be saved.
Selection on | off
Default value off
Save as CV
The calculated result can be saved as a method-independent variable,
called a common variable. The result is then also available in other meth-
ods for calculations. If statistics has been switched on, then the current
mean value of the determination series will be saved.
Selection on | off
Default value off
7.3.8 Statistics
Menu Parameters Statistics
The statistics calculation of a multiple determination is activated under
Statistics and definition is made as to how many determinations the
series contains.
Statistics
If this function is activated, then statistics calculations will be carried out
for all of the defined results.
Selection on | off
Default value off
Number of determinations
The number of determinations that are carried out for the statistics calcu-
lations.
If an additional determination has to be added to the determination
series, because one determination has been incorrect, for example, then
this can be accomplished in the statistical overview (see Chapter 5.11,
page 41).
Input range 2 - 20
Default value 3
7.3.9 Reports
Menu Parameters Reports
The reports that will be printed out automatically in connection with a
determination are defined under Reports.
Results
The result report contains the calculated results, equivalence points and
endpoints, sample data, etc.
Selection on | off
Default value off
Curve
Curve report. The width of the curve is defined in the system settings (see
"Graphics width", page 64).
Selection on | off
Default value off
Calculations/Statistics
Output of the calculation formulas for the individual results. Results are
specified with full accuracy. This makes recalculation with an external pro-
gram possible (see Chapter 10.8, page 127). If Statistics has been acti-
vated, then the following data will be printed out as well:
Result and sample size of the individual determinations
Mean value as well as absolute and relative standard deviation
Selection on | off
Default value off
Parameters
All of the parameters of the current method are printed out in the para-
meter report.
Selection on | off
Default value off
PC/LIMS
The PC/LIMS report is a machine-readable report with all of the data
important for a determination. The PC/LIMS report can be saved as a TXT
file on a USB storage medium or sent via an RS-232 interface to a LIMS.
The output location is defined in the system settings (see "PC/LIMS
report", page 63).
The file name of the TXT file is constructed as follows: PC_LIMS_Report-
ID1-YYYYMMDD-hhmmss.txt.
Selection on | off
Default value off
Signal drift
Maximum permissible drift for the measured value acceptance, i.e. maxi-
mum change of the measured value per minute.
Input range 0.1 - 999.0 mV/min
Default value 2.0 mV/min
Selection off
off
Measured value acceptance will take place after the maximum waiting
time has elapsed. This can be useful when the electrode is slow to
respond.
Temperature
Manually entered calibration temperature. If a temperature sensor is con-
nected then the temperature will be measured continuously.
The temperature can also be entered during the run.
Input range 20.0 - 150.0 C
Default value 25.0 C
Sensor
Selection of the sensor from the sensor list. The selection depends on the
measuring mode. Sensors are defined under System Sensors.
Selection Selection of configured sensors
Stirrer
The stirrer is switched on at the start of the determination when this para-
meter is activated.
Selection on | off
Default value on
Stirring rate
Setting the stirring rate. It can be set in steps of 15 to +15. The default
setting 8 corresponds to 1,000 rpm. The formula for calculating the rota-
tional speed is specified in chapter 10.2, page 118. The optimum stirring
rate can be tested in the manual control.
The algebraic sign of the stirring rate changes the stirring direction. When
the stirrer is viewed from above, this means:
"+": counterclockwise rotation
"": clockwise rotation
Input range 15 - 15
Default value 8
Sample changer
This parameter must be switched on for automatic calibrations with sam-
ple changers (e.g. 869 Compact Sample Changer). Because the buffers are
changed by the sample changer, the run is not paused in order to request
values. The 848 Titrino plus sends a signal (EOD) to the sample changer
via the remote connection as soon as the measurement of one buffer has
been completed (see Figure 25, page 125). The sample changer switches
to the next buffer at that time.
The parameterization of the sample changer is described in the corre-
sponding manual.
NOTE
Selection on | off
Default value off
7.4.2 Buffers
Menu Parameters Buffers
The buffer type and the number of buffers is defined under Buffers.
Buffer type
Selection of a predefined buffer series or definition of special buffers. In
the case of predefined buffer series, the instrument automatically recogni-
zes which buffer is involved.
Selection Baker | Beckman | DIN | Fisher | Fluka Basel |
Hamilton | Merck CertiPUR | Merck Titrisol |
Metrohm | Mettler | NIST | Precisa | Radiome-
ter | Special
Merck CertiPUR
Reference temperature = 25 C. When using Merck CertiPUR buffers
(20 C) the buffer type Merck Titrisol must be selected.
Special
Up to five calibration buffers can be defined in the method. The auto-
matic buffer recognition is not activated in this case. The buffers must
be measured precisely in the specified sequence.
Number of buffers
Number of buffers that are used for calibration. If calibration is accom-
plished with more than two buffers, then they can be used repeatedly in
order to give them more statistical weight. The first two buffers must,
however, always be different from one another.
Selection 1|2|3|4|5
Default value 2
Buffer 1 pH
This parameter is visible only when Buffer type = Special.
Input range 20.000 - 20.000
Default value 7.000
Buffer 2 pH
This parameter is visible only when Buffer type = Special.
Input range 20.000 - 20.000
Default value 4.000
Selection off
Buffer 3 pH
This parameter is visible only when Buffer type = Special.
Buffer 4 pH
See Buffer 3 pH.
Buffer 5 pH
See Buffer 3 pH.
7.4.3 Reports
Menu Parameters Reports
The reports that will be printed out automatically in connection with a cal-
ibration are defined under Reports.
Results
The result report contains the specifications for the calibration (slope,
pH(0), etc.).
Selection on | off
Default value off
Parameters
All of the parameters of the current method are printed out in the para-
meter report.
Selection on | off
Default value off
PC/LIMS
The PC/LIMS report is a machine-readable report with all of the data
important for a determination. The PC/LIMS report can be saved as a TXT
file on a USB storage medium or sent via an RS-232 interface to a LIMS.
The output location is defined in the system settings (see "PC/LIMS
report", page 63).
The file name of the TXT file is constructed as follows: PC_LIMS_Report-
ID1-YYYYMMDD-hhmmss.txt.
Selection on | off
Default value off
Qualification
Please contact your local Metrohm representative for support in qualifica-
tion of instruments and software. The Installation Qualification (IQ)
and Operational Qualification (OQ) are offered by Metrohm represen-
tatives as a service. They are carried out by trained employees using stan-
dardized qualification documents and in accordance with the currently
applicable requirements of the regulated industry.
Maintenance
The electronic and mechanical functional groups of Metrohm instruments
can and should be checked by specialist personnel from Metrohm as part
of a regular preventive maintenance schedule. Please ask your local
Metrohm representative regarding the precise terms and conditions
involved in concluding a corresponding maintenance agreement.
For detailed information on this topic, please visit www.metrohm.com.
9 Troubleshooting
9.1 SET titration
Problem Cause Remedy
The titration will not The minimum dosing rate Define Titration rate = user and increase the
be finished. is too low. minimum rate (Min. rate) (see Chapter 7.3.4,
page 102).
The stop criterion is unsuit- Adjust the control parameters (see Chapter
able. 7.3.4, page 102):
The sample is over- The control parameters are Adjust the control parameters (see Chapter
titrated. unsuitable. 7.3.4, page 102):
The titration time is The control parameters are Adjust the control parameters (see Chapter
too long. unsuitable. 7.3.4, page 102):
The results are The minimum dosing rate Define Titration rate = user and reduce the
spread widely. is too high. minimum rate (Min. rate) (see Chapter 7.3.4,
page 102).
9.2 Miscellaneous
Problem Cause Remedy
No report is printed. The printer is not recog- Switch the 848 Titrino plus off and then
nized by the instrument. back on again.
Use a USB hub and connect the printer to
the USB hub.
The printer model is not Use a printer which fulfills the required specifi-
compatible. cations (see Chapter 10.4.4, page 121).
The USB keyboard The keyboard or mouse is Switch the 848 Titrino plus off and then
or the PC mouse not recognized by the back on again.
does not function. instrument. Use a USB hub and connect the keyboard
and mouse to the USB hub.
The keyboard or the mouse Use a model which fulfills the required specifi-
is not compatible. cations (see Chapter 10.4, page 119).
The display is no The contrast is set incor- Adjust the contrast correctly (see Chapter 6.1,
longer readable. rectly. page 52).
Message 020-511 The USB flash drive is no 1. Connect the USB flash drive.
"Action not possi- longer connected. 2. Switch the instrument off and then back
ble" is displayed. on again.
The USB flash drive is full. Use a different USB flash drive.
Delete files with the aid of a PC.
10 Appendix
Example:
Stirring rate set: 8
Rotational speed in rpm = 125 8 = 1000
r/min
2000
1500
1000
500
-15 -10 -5 0 5 10 15
Figure 21 Rotational speed depending on stirring rate
10.3 Balance
The sample size and the associated unit can be sent from a connected bal-
ance. The sample size is transmitted as a number with up to ten charac-
ters (including algebraic sign and decimal point).
Sample size and unit are sent as a single character string. They are separa-
ted by a space character. The string is terminated with the ASCII charac-
ters CR and LF.
If the balance sends a negative sample size (e.g. when you are reweighing
a sample), then the algebraic sign is adopted. The algebraic sign is, how-
ever, ignored for the calculations.
NOTE
With some balances, the sample identification and the method can be
sent in addition to the sample size.
Make sure that the balance does not send the sample size until the end.
Mettler AX
For the Mettler AX balance, the fields that contain the sample identifica-
tion or the method must be designated as follows:
Designation for the field with the method name: METHOD
Designation for the field with sample identification 1: ID1
Designation for the field with sample identification 2: ID2
NOTE
NOTE
The lettering of the USB keyboard may differ from above lettering,
depending on the country-specific keyboard used.
10.4.3 PC mouse
In order to make navigating in the dialog of the 848 Titrino plus easier, a
PC mouse can be connected.
10.4.4 Printer
The range of USB printers available is extremely varied and constantly
changing. The following points must be taken into account when select-
ing a printer:
USB interface necessary
Printer language: HP-PCL, Canon BJL Commands, Epson ESC P/2 or
ESC/POS
NOTE
Inexpensive printers are often designed solely for use with a PC and
may not be equipped with one of the printer languages listed above.
Such models are not suitable for this reason.
CAUTION
All user data (methods, solutions, etc.) are deleted if a system initializa-
tion is carried out. Afterwards, the instrument will have the factory set-
tings again.
We recommend creating a backup of the system at regular intervals in
order to avoid data losses.
NOTE
9 6
6 9
Figure 22 Pin assignment of remote socket and remote plug
The above figure of the pin assignment applies for all Metrohm instru-
ments with 9-pin D-Sub remote connector.
Inputs
+5 V approx. 5 k Pull-up
tp
tp > 100 ms
active = low, inactive = high
Outputs
Open Collector
tp
tp > 200 ms
active = low, inactive = high
IC = 20 mA, VCEO = 40 V
+5 V: maximum load = 20 mA
Activ./Dos. Output 1
Titr/Determ Output 2
condOK Output 3
Error Output 4
Start Input 0
Stop Input 1
[STOP]
Figure 23 Remote status diagram DET/MET
Activ./Dos. Output 1
Titr/Determ Output 2
condOK Output 3
Error Output 4
Start Input 0
Stop Input 1
[STOP]
Figure 24 Remote status diagram SET
Activ./Dos. Output 1
Titr/Determ Output 2
condOK Output 3
Error Output 4
Start Input 0
Stop Input 1
[STOP]
Figure 25 Remote status diagram CAL
EOD = End of Determination
6.2134.040
Interface parameters
We recommend the following parameters for the RS-232 interface
COM2:
Baud rate: 19200
Data bits: 8
Stop bits: 1
Parity: none
Handshake: software
On the 848 Titrino plus, adjust the interface parameters under
Menu System External devices COM2 settings.
Transfer protocol
The data communication is synchronous. On each command there is an
acknowledgement by the instrument.
A command must be sent to the instrument with the control characters
CR LF as terminator. The acknowledgements of the instrument will also
be transferred with CR LF as terminator.
The instrument does not send spontaneous messages.
The values of the variables are only available after the end of a determina-
tion (in the status 'ready').
Acknowledgement of Comment
the instrument
OK Command executed
E1 Method not found
E2 Invalid variable
E3 Invalid command
Rounding-off process
Measured values and results are rounded to the defined number of deci-
mal places (commercial rounding, in accordance with the US Pharmaco-
peia USP). If the digit at the first dropped decimal place is 1, 2, 3 or 4,
then it will be rounded off; if this digit is 5, 6, 7, 8 or 9, then it will be
rounded up. Negative digits will be rounded in accordance with their
amount, i.e. away from zero.
Examples:
2.33 yields 2.3
2.35 yields 2.4
2.47 yields 2.5
2.38 yields 2.4
2.45 yields 2.5
Statistics
The arithmetic mean value and the absolute and relative standard devia-
tions of results are calculated:
You can statistically evaluate a maximum of five results (1 k 5) calcu-
lated in a determination. A statistical series can contain a maximum of 20
determinations (1 n 20).
The following convention applies to the subsequent formulas:
1 n 20 and 1 k 5.
Mean value:
Explanations
The individual values are incorporated in the statistics with full accuracy.
NOTE
11 Technical specifications
11.1.1 Potentiometry
A high-ohm measuring input (Ind.) for pH electrodes and redox electrodes
and a measuring input for separate reference electrodes (Ref.).
Input resistance > 1 1012
Offset current < 1 1012 A
Measuring mode
pH
Measuring 13 - +20
range
Resolution 0.001
Measuring 0.003
accuracy (1 digit, without sensor error, under reference conditions)
Measuring mode
U
Measuring 1,200 - +1,200 mV
range
Resolution 0.1 mV
Measuring 0.2 mV
accuracy (1 digit, without sensor error, under reference conditions)
11.1.2 Polarizer
One measuring input (Pol.) for polarizable electrodes.
Measuring mode Determination with adjustable polarization current.
Ipol
Polarization 120 - +120 A (increment: 1 A)
current 125 - 121 A / +121 - +125 A: non-guaranteed values, dependent
on reference voltage +2.5 V
Measuring 1,200 - +1,200 mV
range
Resolution 0.1 mV
Measuring 0.2 mV
accuracy (1 digit, without sensor error, under reference conditions)
11.1.3 Temperature
A measuring input (Temp.) for temperature sensors of the Pt1000 or NTC
type with automatic temperature compensation.
R (25 C) and B value can be configured for NTC sensors.
Measuring range
Pt1000 150 - +250 C
NTC 5 - +250 C
(For an NTC sensor with R (25 C) = 30,000 and B (25/50) = 4,100 K)
Resolution
Pt1000 0.1 C
NTC 0.1 C
Measuring accu-
racy
Pt1000 0.2 C
(applies for measuring range 20 - +150 C)
NTC 0.6 C
(applies for measuring range +10 - +40 C)
11.3 Interfaces
USB (OTG) connec- For connecting USB devices.
tor
MSB connector For connecting a stirrer.
Remote connector For connecting instruments with a remote interface.
Design and testing According to EN/IEC/UL 61010-1, CSA-C22.2 No. 61010-1, protection
class I, EN/IEC 60529, degree of protection IP40.
Safety instructions This document contains safety instructions which have to be followed
by the user in order to ensure safe operation of the instrument.
Immunity
Standards ful- EN/IEC 61326-1
filled EN/IEC 61000-6-2
EN/IEC 61000-4-2
EN/IEC 61000-4-3
EN/IEC 61000-4-4
EN/IEC 61000-4-5
EN/IEC 61000-4-6
EN/IEC 61000-4-11
EN/IEC 61000-4-14
11.9 Dimensions
Material of hous- Poly(butylene terephthalate) (PBT)
ing
Material of display Glass
cover
Width 142 mm
Height 164 mm
Depth 310 mm
Weight 2,950 g
12 Accessories
Up-to-date information on the scope of delivery and optional accessories
for your instrument can be found on the Internet.
NOTE
2
Click on .
The Search webpage will be displayed.
3 Enter a search term relating to the instrument into the search field
and click on Find.
The search results will be displayed.
4 In the search results, select the Devices tab (if it is not already
selected) and then click on the Metrohm article number of the
required instrument (e.g. 2.852.0050).
The page with information pertaining to the searched article is dis-
played.
6
Click on .
3 Enter the article number and click on the Generate PDF command.
The PDF file with the accessories data will be created in the language
selected.
Index
A Dialog type ............................... 53 Language file
Adapter Directory structure .................... 62 Load ................................... 66
Connect ............................. 12 Dosing Live modifications ..................... 37
Continuous (DOS) ............... 46 Live parameters ........................ 39
B Fixed volume (ADD) ............ 47
Backup ..................................... 62 Dosing rate M
Balance ............................ 35, 118 Maximum ......................... 117 Mains voltage ............................. 4
Connect ............................. 12 Maintenance agreement ........ 114
Interface parameters ........... 64 E Manual control
Balance type ............................. 64 Electrode Dosing ............................... 45
Buffers Connect ............................... 9 Measuring .......................... 49
CAL .................................. 112 Electrode connector Stirring ............................... 50
Buret unit pH electrode ........................ 8 MET ..................................... 1, 17
Prepare (PREP) .................... 45 Polarizable electrode ............. 8 Parameters ......................... 83
Redox electrode ................... 8 Method .................................... 25
C Reference electrode .............. 8 Create ................................ 25
CAL Electrostatic charge .................... 5 Delete ................................ 62
Parameters ....................... 110 End of Determination EOD ..... 125 Export ................................ 28
Calculation ................. 79, 93, 106 ERC .................................... 79, 93 Import ................................ 62
Delete .................. 80, 94, 106 Evaluation Load ................................... 27
Edit ...................... 80, 94, 106 DET .............................. 76, 78 Save ................................... 26
Numerical format ............. 127 MET ............................. 90, 93 MSB
Rounding-off process ....... 128 Exchange unit Connector ............................ 8
Calibration parameters Attach ................................ 16
CAL .................................. 110 Expert dialog ............................ 53 N
Common variable ..................... 60 Navigate ................................... 21
Edit .................................... 60 F Numerical input ........................ 22
Common Variable Filling rate
Delete ................................ 60 Maximum ......................... 117 O
Set to invalid ...................... 60 Formula editor .......................... 23 Operation
Conditioning General .............................. 20
SET ..................................... 97 G Operational qualification ........ 114
Contrast ................................... 54 Graphics width ......................... 64
Control ..................................... 29 P
Autostart ............................ 29 I Parameters
Sample table ...................... 29 Initialization ............................ 121 CAL .................................. 110
Control parameters Installation qualification .......... 114 DET .................................... 68
SET ................................... 102 Instrument MET ................................... 83
Curve ....................................... 41 Switch off ........................... 20 Modify live ......................... 39
Switch on ........................... 19 SET ..................................... 97
D Instrument diagnosis ................ 66 PC/LIMS report ......................... 63
DET ...................................... 1, 17 Pin assignment ....................... 122
Parameters ......................... 68 K PREP
Determination Keyboard Parameters ....................... 117
Cancel ................................ 37 Connect ............................. 12 Preparing
Carry out ............................ 35 Key assignment ........ 119, 120 Parameters ....................... 117
Result ................................. 40 Keyboard layout ....................... 64 Print ......................................... 43
Diagnosis ................................. 67 Printer .............................. 63, 121
Dialog language ....................... 52 L Connect ............................. 12
Load ................................... 66 Language ................................. 52 Program crash ........................ 121