+GF+ SIGNET 8512 Flow Transmitter Instructions ENGLISH

3-8512.090-1
C-6/98

CAUTION!
Remove power to unit before wiring input and output connections.

1. Loop/System Power Connections


1.1 2-Wire operation (for +GF+SIGNET 515, 525, 2517, 3-8510-XX, 2536/3-8512-XX flow sensors).

A. Ground referenced PLC with internal transmitter power supply B. Power connection for display use only

PLC Power
8512 Transmitter supply 8512 Transmitter
4-20 mA Input Fuse** Fuse**
Terminals Terminals
+ Power out 1/8A
+ 4-20 mA - 24* + 1/8A
+ 4-20 mA
VDC
24 VDC
- CH +
-
+
Loop

17-30V *
- Loop

+ 17-30V *
250*
- Gnd
- Gnd

CH - (gnd)

C. 1 to 5 VDC recorder (C1), 4 to 20 mA indicator (C2), or ground referenced PLC (C3) connections without internal transmitter power supply

Recorder Indicator PLC


1-5 VDC Power 4-20 mA Power 4-20 mA Input Power
CH - supply Fuse** CH - supply CH - supply Fuse** 8512 Transmitter
Fuse**
(gnd) (gnd) Terminals
C1 - 24 + 1/8A C2 - 24 + 1/8A C3 (gnd)
- 24* + 1/8A
+ 4-20 mA
VDC* VDC* VDC
250*
CH +
250*
CH +
- Loop

+ 17-30V *
CH +
- Gnd

*Refer to maximum loop impedance specification for minimum operating voltage requirements (section 10).
**1/8A fuse recommended (customer supplied)

1.2 3-Wire operation (for +GF+ SIGNET 2000, 2507, 2530, 2535, 2540 flow sensors). This wiring is required for powered flow
sensors that consume more than 1.5 mA DC current.

A. Ground referenced PLC with internal transmitter power supply B. Differential input PLC with internal transmitter power supply
8512 Transmitter
PLC PLC Terminals
4-20 mA Input 8512 Transmitter 4-20 mA Input Fuse**
Fuse** + 4-20 mA
+ +
Terminals 1/4A
+ Power out 1/4A
+ 4-20 mA - 24 VDC - Loop

24 VDC - Loop
- + 17-30V *
- CH + + 17-30V * - Gnd

250*
- Gnd
250*
CH +
CH - (gnd) CH -

C. 1 to 5 VDC recorder (C1) and 4 to 20 mA indicator (C2) connections without internal transmitter D. Power connection for display use only
power supply
Recorder Power Indicator Power
1-5 VDC supply 4 -20 mA supply 8512 Transmitter Power 8512 Transmitter
CH - Fuse** CH - Fuse** supply Fuse**
Terminals Terminals
C1 (gnd)
- 24 + 1/4A C2 (gnd)
- 24
+ 1/4A
+ 4-20 mA - 24* + 1/4A
VDC* VDC*
VDC + 4-20 mA
250* CH + -
+
Loop
- Loop
17-30V * + 17-30V *
CH +
- Gnd
- Gnd

*Refer to maximum loop impedance specification for minimum operating voltage requirements (section 10).
**1/4 A fuse recommended (customer supplied).

2. Compatible Sensor Connections


2.1 2-Wire Operation (sec. 1.1) 2.2 3-Wire Operation (sec. 1.2)
+GF+ SIGNET Sensors:
8512 Transmitter 8512 Transmitter
Terminals Red Terminals
2536,
Black +GF+ SIGNET Sensors:
3-8512-XX
Frequency In Shld Frequency In Red 2000, 2507,
OPEN OR OPEN
COLLECTOR Sensor Power Shld
COLLECTOR Sensor Power Black 2530, 2535,
SENSOR
515, 525, SENSOR

Gnd Gnd Shld 2540


STANDARD
Black 2517, STANDARD
SENSOR Frequency In Red 3-8510-XX SENSOR Frequency In
3. Pulse Output Wiring
8512 Transmitter 8512 Transmitter + 5 to 30
Terminals Terminals 10 k - VDC
Pulse Gnd Shld +GF+ SIGNET Pulse Gnd Gnd. Other
Pulse Out Red instrument Pulse Out Input instrument

Use 3-2507.278 input module for +GF+ SIGNET instruments


+GF+ SIGNET Intelek-Pro, use 2535/2536 input card setting

4. Installation Options

4.1 Standard Panel Mount 4.3 Optional 3-8011 Integral Mounting Kit
Panel cutout template/instructions (included). NPT and DIN conduit port kits available.
Compatible with 3-8510-XX/3-8512-XX flow sensors (only).
Flow sensor and fitting purchased separately (see section 9).
8512 Flow Transmitter

Instrument
+GF+ SIGNET panel

Conduit ports

4.2 Optional 3-8010 Universal Mounting Kit


NPT and DIN conduit port kits available (see section 4.3).
See section 9 for ordering options.

Pipe Pipe Panel Panel


Conduit Port Options
DIN Ports NPT Ports

8512 Flow Transmitter 8512 Flow Transmitter

Blind holes
+GF+ SIGNET +GF+ SIGNET
for DIN conduit
connectors (8) PG13.5 metric 1/2 in. NPT
threaded conduit threaded conduit
ports (2) ports (2)

5. FUNCTIONS 6. VIEW (example)

1 2 3
Choose: Change: Save:
E

A.
Flow rate

2s OPTIONS = OPTIONS =
Press &
hold for B.
access: 1. Press and 1. Press keys in
Resettable total hold 2s sequence

VIEW 2s
2. 2.
C.
2s 5s Press and
CALIBRATE Permanent total To exit without hold 2s
changes:

OPTIONS D. quick 3.
Loop output press

E.
Last cal.
A
7. CALIBRATE (example) 8. OPTIONS (example)

Press keys in sequence to Press keys in sequence to


continue: continue:

1 2 3 1 2 3
Choose: Change: Save: Choose: Change: Save:
H E

A. A.
Contrast Totalizer VIEW= VIEW=

B. Unit/timebase:
Flow h,m,s,d B. Display low= =700 ms
averging hi= =3 s
off= 100 ms
C.
Flow K-factor Press &
C. hold
Flow display Decimal
D. Units
postion
Totalizer - label only - 2s
Press &
hold D. 4 mA
E. Output adjust
Totalizer K-factor
2s
E. 20 mA
F. 4 mA
Output adjust
Output setpoint A

To return to Return to
VIEW: To restore original value: VIEW before
G. 20 mA removing
Output setpoint quick quick power
press press
H.
Last calibration
A

To return to To restore original value: Return to


VIEW: VIEW before
removing
quick quick power
press press

Menu Functions A - H: Menu Functions A - E:


A. Selects display contrast: 4 levels A. Selects totalizer reset options: Lock on (enables) or lock off (disables) the
B. Sets flow units label (gpm) and timebase (gpm). Flow units label: A - Z, VIEW menu totalizer reset security code feature (RST: -- -- -- --).
a - z, 0 - 9; Timebase options: s=seconds, m=minutes, h=hours, d=days B. Selects display averaging: off = 100ms, low= =700ms, hi= =3s (also
(timebase effects flow display and 4 to 20 mA output) affects 4 to 20 mA output).
C. Sets flow K-factor: 000.01 to 99999. (see technical notes) C. Selects display decimal: **** . to ** . **
D. Set totalizer units: For label purposes only D. Adjusts 4 mA output: 3.9 to 4.1 mA (overrides 4.00 mA factory calibration)
E. Sets totalizer K-factor: 000.01 to 99999. (see technical notes) E. Adjusts 20 mA output: 19.8 to 21.0 mA (overrides 20.00 mA factory
F. Sets 4 mA setpoint (4 mA and 20 mA setpoints are reversible) calibration)
G. Sets 20 mA setpoint
H. Sets user defined date

Technical notes:
Flow and totalizer K-factors are independent of each other. These K-factors
represent the number of pulses generated by the +GF+ SIGNET flow sensor for
each engineering unit measured (published in flow sensor manual).

9. Accessories Order no. Description Code


3-8010 Universal mounting kit, NPT ports 198 864 502
3-8010-D Universal mounting kit, DIN ports 198 864 503
3-8011 Integral sensor mounting kit, NPT ports 198 864 500
3-8011-D Integral sensor mounting kit, DIN ports 198 864 501
3-8510-P0 Integral sensor, 0.5 to 4 inch pipe, Polypropylene body & Titanium rotor pin 198 864 504
3-8510-P1 Integral sensor, 5 to 8 inch pipe, Polypropylene body & Titanium rotor pin 198 864 505
3-8510-V0 Integral sensor, 0.5 to 4 inch pipe - PVDF body & Hastelloy C rotor pin 198 864 506
3-8512-P0 Integral sensor, 0.5 to 4 inch pipe, Polypropylene body & Titanium rotor pin 198 864 513
3-8512-P1 Integral sensor, 5 to 8 inch pipe - Polypropylene body & Titanium rotor pin 198 864 514
3-8512-V0 Integral sensor, 0.5 to 4 inch pipe - PVDF body & Hastelloy C rotor pin 198 864 516
3-8512-T0 Integral sensor, 0.5 to 4 inch pipe - PVDF body & PVDF rotor pin 198 864 518
10. Specifications
General Data Electrical Data
Compatible Sensors: All current +GF+ SIGNET flow sensors Loop:
Display Accuracy: Flow, 0.1% of reading Impedance: 1 max. @ 17 VDC,
Totalizers, 0.03% of reading 300 max. @ 24 VDC,
Enclosure: 600 max. @ 30 VDC
Rating: NEMA 4X/IP65 Accuracy: 0.050 mA
Material: Glass-filled polypropylene Resolution: 5 A
Gasket: Silicone rubber (captive) Update rate: 100 ms
Screws: 8-32, self-tapping (captive)
Outputs:
Display: Current: 4 to 20 mA (adjustable & reversible)
Type: 8-digit alphanumeric dot matrix Pulse output: Sensor frequency, optically isolated
Update rate: Flow=1s, Totalizers=100 mS open-collector transistor, max. current sink
Contrast: Variable 10 mA @ 30 VDC
Ranges: Flow, 0.01 to 9999.
Resettable/permanent totalizers, 0 to Dimensions: 30 mm (1.2 in.)
99999999
Loop current, 3.90 to 21.00 mA

Environmental
Operating temperature: -15 to 70 C (5 to 158 F) 8512 Flow Transmitter

Storage temperature: -15 to 80 C (5 to 176 F)


Relative humidity: 0 to 95%, non-condensing 107 mm
(4.2 in.)
Quality Standards +GF+ SIGNET
CE
Manufactured under ISO 9001

Electrical Data
107 mm (4.2 in.)
Frequency range: 0.5 Hz to 500 Hz 36 mm (1.4 in.)
Loop/system power: (2-wire mode) 17 to 30 VDC @ 20 mA max.
(3-wire mode) 17 to 30 VDC @ 68 mA max. (front view) (side view)
Sensor power: (2-wire mode) 5 VDC @ 1.5 mA max.
(3-wire mode) 5 VDC @ 20 mA max.

11. Troubleshooting

Display Message Cause Solution


1) Input frequency too high 1) Reduce input frequency.
2) Display overrange 2) Move display decimal to right in OPTIONS menu.
3) Display timebase too large 3) Change display timebase (H,M,S,D) to smaller value
(e.g. LPH to LPM).

K-Factor cannot be zero Change K-Factor to a non-zero value.

2s Memory corrupted Press to restore normal operation. Settings will revert to


2s factory default. Recalibration is required.

+GF+ SIGNET
Sales Offices:
USA George Fischer, Inc., 2882 Dow Avenue, Tustin, CA 92780/USA, Tel. (714) 731-8800, Fax (714) 731-6201
Switzerland Georg Fischer Rohrleitungssysteme AG, P.O. Box 671, CH-8201 Schaffhausen/Switzerland, Tel. 052/631 1111, Fax 052/631 2830
Singapore George Fischer Pte. Ltd., 15 Kaki Bukit Road 2, KB Warehouse Complex, Singapore 1441, Tel. 65/747 0611, Fax 65/747 0577
Japan Kubota George Fischer, 2-47 Shikitsuhigashi, 1-Chome, Naniwa-Ku, Osaka, 556-91 Japan, Tel. 816/648 2545, Fax 816/648 2565
China Georg Fischer Ltd., Rm 1503, Business Residence Bldg. of Asia Plaza, 2-3 Bldg. No. 5th Qu Anzhenxili, Chaoyang Qu, Beijing 100029,
P.R. China, Tel. 86/10 6443 0577, Fax 86/10 6443 0578
Australia George Fischer Pty. Ltd., Suite 3, 41 Stamford Road, Oakleigh, Victoria 3166, Australia, Tel. 61/3 9568 0966, Fax 61/3 9568 0988

Signet Scientific Company, 3401 Aerojet Avenue, El Monte, CA 91731-2882 U.S.A., Tel. (626) 571-2770, Fax (626) 573-2057

GEORGE FISCHER +GF+ Piping Systems PRINTED ON RECYCLED PAPER

3-8512.090-1/(C-6/98), English Signet Scientific Company 1996 Printed in U.S.A.