Agilent 700 Series - Icp
Agilent 700 Series - Icp
Agilent 700 Series - Icp
Optical Emission
Spectrometers
Users Guide
Notices Warranty contract clause. Use, duplication or
disclosure of Software is subject to
The material contained in this Agilent Technologies standard
Agilent Technologies, Inc. document is provided as is, and is commercial license terms, and non-
2006, 2009 and 2010 subject to being changed, without DOD Departments and Agencies of the
No part of this manual may be notice, in future editions. Further, to U.S. Government will receive no
reproduced in any form or by any the maximum extent permitted by greater than Restricted Rights as
means (including electronic storage applicable law, Agilent disclaims all defined in FAR 52.227-19(c)(1-2) (June
and retrieval or translation into a warranties, either express or implied, 1987). U.S. Government users will
foreign language) without prior with regard to this manual and any receive no greater than Limited Rights
agreement and written consent from information contained herein, as defined in FAR 52.227-14 (June
Agilent Technologies, Inc. as governed including but not limited to the 1987) or DFAR 252.227-7015 (b)(2)
by United States and international implied warranties of merchantability (November 1995), as applicable in any
copyright laws. and fitness for a particular purpose. technical data.
Agilent shall not be liable for errors
Manual Part Number or for incidental or consequential Safety Notices
damages in connection with the
8510230100
furnishing, use, or performance of
this document or of any information CAUTION
Edition contained herein. Should Agilent and
A CAUTION notice denotes a hazard.
Third edition, August 2010 the user have a separate written
It calls attention to an operating
agreement with warranty terms
Printed in Australia procedure, practice, or the like that, if
covering the material in this
Agilent Technologies, Inc. not correctly performed or adhered to,
document that conflict with these
could result in damage to the product
terms, the warranty terms in the
or loss of important data. Do not
Errata Statement separate agreement shall control.
proceed beyond a CAUTION notice
until the indicated conditions are fully
NOTICE: This document contains Technology Licenses understood and met.
references to Varian. Please note that
Varian, Inc. is now part of Agilent The hardware and/or software
Technologies. For more information, described in this document are
go to www.agilent.com/chem. furnished under a license and may be WARNING
used or copied only in accordance
with the terms of such license.
A WARNING notice denotes a
Restricted Rights Legend hazard. It calls attention to an
If software is for use in the operating procedure, practice, or the
performance of a U.S. Government like that, if not correctly performed or
prime contract or subcontract, adhered to, could result in personal
Software is delivered and licensed as injury or death. Do not proceed
Commercial computer software as beyond a WARNING notice until the
defined in DFAR 252.227-7014 (June indicated conditions are fully
1995), or as a commercial item as understood and met.
defined in FAR 2.101(a) or as
Restricted computer software as
defined in FAR 52.227-19 (June 1987)
or any equivalent agency regulation or
Contents
1 Safety Practices and Hazards 7
General 7
Plasma 8
Heat, Vapors and Fumes 9
Compressed Gas Hazards 9
Electrical Hazards 10
Other Precautions 10
Warning Symbols 12
Color Coding 14
US FCC Advisory Statement 14
Spurious and Harmonic Radiation Compliance 14
CE Compliance 15
2 Introduction 17
Pre-installation Requirements 17
User Documentation 18
Conventions 18
Notes and Tips 18
Specifications 19
Environmental Conditions 19
Electrical Power Supplies 19
Gas Supplies 22
Water Cooling System 24
Drain Vessel 27
Laboratory Conditions 27
Temperature Control 27
Cleanliness 28
Instrument Cooling Air Supply 28
Weights and Dimensions 29
3 Installation 31
Personal Computer Requirements 31
Installing the ICP Expert II Software 32
PCI-GPIB Driver Installation 35
PCI-GPIB Communications Card Installation 36
Hardware Installation 40
4 Operation 41
Analysis Checklist 42
Turning on the Instrument and Software 42
Turning On the Instrument for the First Time (or From Shutdown
Mode) 42
Turning On the Instrument From Standby Mode 43
Creating/Opening a Worksheet 44
Developing a Method 45
Preparing for Analysis 46
Calibrating the Detector 47
Aligning the Torch 49
Setting Operating Conditions 50
Setting up Standards 52
Running Samples 53
Printing a Report 54
Turning the Instrument Off 55
Using Standby Mode 56
Using Shutdown Mode 58
5 Maintenance 61
Routine Maintenance 62
Maintenance of Consumables 63
Cleaning 64
Spare Parts 64
General
Operation of an Agilent 710/715/720/725/730/735 ICP Optical
Emission Spectrometer involves the use of compressed gases, high
voltage radio frequency energy and hazardous materials including
corrosive fluids and flammable liquids. Careless, improper or
unskilled use of this spectrometer can cause death or serious injury
to personnel, and/or severe damage to equipment and property.
The spectrometer incorporates interlocks and covers that are
designed to prevent inadvertent contact with any potential hazards.
If the instrument is used in any manner not specified by Agilent, this
protection may be impaired. It is good practice to develop safe
working habits that do not depend upon the correct operation of the
interlocks for safe operation. It is essential that no interlock or cover
is bypassed, damaged or removed.
The safety practices described below are provided to help the user
operate the instrument safely. Read each safety topic thoroughly
before attempting to operate the instrument and ALWAYS operate
the spectrometer in accordance with these safety practices.
Plasma
The plasma is extremely hot (about 10000 K) and radiates dangerous
levels of radio frequency (RF) and ultraviolet (UV) energy. The work
coil operates at 1500 V RMS and about 40 MHz. Exposure to the RF
and UV energy can cause severe skin damage and cataracts of the
eyes, while close contact with the operating plasma can result in
severe heat burns to the skin, and an electrical discharge which can
jump a considerable distance and may cause death, severe electric
shock or sub-surface skin burns.
The plasma must NEVER be operated unless:
the torch compartment door is closed, with the locking lever fully
latched; and
the space above the chimney is clear of objects.
The shielding around the torch compartment is designed to reduce
UV, visible and RF radiation to safe levels while still permitting easy
access to, as well as installation and viewing of, the torch. The
spectrometer has an interlock system that is designed to extinguish
the plasma if either the mains supply fails or the handle on the torch
compartment door is opened. DO NOT attempt to bypass the
interlock system.
Before opening the torch compartment door, ALWAYS extinguish the
plasma by pressing F4 on the keyboard or by clicking the Plasma Off
icon on the ICP Expert II software Toolbar.
The torch and its surroundings remain hot for up to five minutes
after the plasma is extinguished. Touching this area before it has
cooled sufficiently may result in burns. Allow the torch and torch
compartment to cool before carrying out any work in this area, or
wear heat-resistant gloves.
The plasma system has been carefully designed to operate safely and
effectively when using torches and related components that conform
to Agilents design criteria. Use of non-approved components in the
plasma compartment may render the system inoperative and/or
hazardous. It may also invalidate the warranty on the instrument.
Use only torches and related components supplied or authorized by
Agilent.
Electrical Hazards
The spectrometer system and some accessories contain electrical
circuits, devices and components operating at dangerous voltages.
Contact with these circuits, devices and components can cause death,
serious injury or painful electric shock. Panels or covers which are
retained by screws on the spectrometer and accessories may be
opened ONLY by Agilenttrained, Agilent-qualified or
Agilentapproved field service engineers (unless specified otherwise).
Consult the manuals or product labels supplied with your personal
computer (PC), monitor, printer/plotter and water-cooling system to
determine which parts are operator-accessible.
Replace blown fuses with ones of the size and rating shown in the
text near to the fuse holder.
Other Precautions
Use of the spectrometer system and accessories may involve
materials, solvents and solutions which are flammable, corrosive,
toxic or otherwise hazardous. Careless, improper or unskilled use of
such materials, solvents and solutions can create explosion hazards,
fire hazards, toxicity and other hazards that can result in death,
serious personal injury or damage to equipment.
Always ensure that laboratory safety practices governing the use,
handling and disposal of hazardous materials are strictly observed.
These safety practices should include wearing appropriate safety
clothing and safety glasses.
Air flow to the cooling fans of the spectrometer and accessories must
be unobstructed. Do not block the ventilation grills on the
spectrometer and accessories. Consult the manuals supplied with
your PC, monitor, printer and water-cooling system for their specific
ventilation requirements.
Great care should be taken when working with glass or quartz parts
to prevent breakage and cuts. This is especially important when
attaching plastic tubing to glass barbs, or removing and replacing
pieces of broken torch or bonnet.
The spectrometer weighs approximately 203 kg (448 lb). To avoid
injury to personnel or damage to the instrument or property, always
use a forklift or other suitable mechanical lifting device to move the
instrument.
Use only Agilent-supplied spares with your instrument.
Warning Symbols
The following is a list of symbols that appear in conjunction with
warnings in this manual and on the spectrometer. The hazard they
describe is also shown. The beginning of the warning text is noted by
a warning icon:
WARNING
A triangular symbol indicates a warning. The meanings of the
symbols that may appear alongside warnings in the documentation
or on the instrument itself are as follows:
I Mains power on
Fuse
Direct current
Plasma on
Plasma off
Color Coding
The various indicator lights appearing on Agilent instruments and
associated accessories are color coded to represent the status of the
instrument or accessory.
A green light indicates the instrument is in normal/standby
mode.
An orange light indicates that a potential hazard is present.
A blue light indicates that operator intervention is required.
A red light warns of danger or an emergency.
CE Compliance
Your Agilent 700 Series ICP-OES instrument has been designed to
comply with the requirements of the Electromagnetic Compatibility
(EMC) Directive and the Low Voltage (electrical safety) Directive
(commonly referred to as the LVD) of the European Union. Agilent
has confirmed that each product complies with the relevant
Directives by testing a prototype against the prescribed EN
(European Norm) standards.
Proof that a product complies with these directives is indicated by:
the CE Marking appearing on the rear of the product, and
the documentation package that accompanies the product
containing a copy of the Declaration of Conformity. The
Declaration of Conformity is the legal declaration by Agilent that
the product complies with the directives listed above, and shows
the EN standards to which the product was tested to
demonstrate compliance.
2. Introduction
Pre-installation Requirements 17
User Documentation 18
Conventions 18
Notes and Tips 18
Specifications 19
Environmental Conditions 19
Electrical Power Supplies 19
Gas Supplies 22
Water Cooling System 24
Drain Vessel 27
Laboratory Conditions 27
Temperature Control 27
Cleanliness 28
Instrument Cooling Air Supply 28
Weights and Dimensions 29
Pre-installation Requirements
Prior to receiving your instrument you will have been provided with
an Agilent 700 Series ICP Optical Emission Spectrometer pre-
installation Manual (publication number 8510233700), which
describes the environmental and operating requirements of the
ICP-OES system. You must prepare your laboratory according to
these instructions before the ICP-OES can be installed. You should
keep the pre-installation manual for future reference. If you have
misplaced your copy, you can obtain a free replacement from your
local Agilent office.
User Documentation
You have been provided with the following documentation to help
you set up and operate your Agilent ICP-OES system:
This operation manual, with Safety practices and hazards
information, instructions for finding information about installing
and maintaining the components of the ICP-OES and a brief
operation overview.
An extensive online Help system containing context-sensitive
Help, step-by-step instructions for frequently performed
operations and instructions for using any accessories you
ordered.
Conventions
The following conventions have been used throughout the
documentation:
Menu items, menu options and field names (e.g., select Copy
from the Edit menu) have been typed in bold. Bold is also used to
signify the pushbuttons appearing throughout the software (e.g.,
click OK).
ALL CAPITALS indicate keyboard and mouse commands (e.g.,
press the F2 key) and text you must type in from the keyboard
(e.g., type SETUP at the prompt).
References to the ICP-OES instrument are applicable to all
models (Agilent 710/715/720/725/730/735 ICP-OES) unless
otherwise stated.
Specifications
Environmental Conditions
The instrument is suitable for indoor use only and is classified
suitable under the following categories (EN 61010-1):
Installation category II
Pollution degree 2
Safety class 1
Condition Altitude Temp (C) Humidity (%RH)
non-condensing
NOTE Basic insulation is provided for single fault protection on the IEEE connector.
Fuses
NOTE For safety reasons, any other internal fuse or circuit breaker is not operator
accessible, and should only be replaced by Agilent-authorized personnel.
Electrical Requirements
The installation of electrical power supplies must comply with the
rules and/or regulations imposed by the local authorities responsible
for the use of electrical energy in the workplace.
All power supplies for the spectrometer, accessories and water cooler
should be single phase, AC, 3 wire systems (active, neutral, ground or
two active and ground). Each connection should be terminated at an
appropriate receptacle within reach of each assemblys power cable.
Use of power boards or extension cables is not recommended.
It is recommended that separate mains circuits, (individually
protected by fuses or circuit breakers) are used for each component
in the system such as the water cooler, autosampler and printer.
Avoid using power supplies from a source that may be subject to
electrical interference from other services (such as large electric
motors, elevators, welders and air conditioning units).
Agilent ICP-OES spectrometers are supplied with a 2 meter long
(6 ft 6 in), mains supply cable. The power cable is hardwired to the
rear of the instrument. The cable is manufactured by Titanex and
rated to 750 V AC at 58 amps.
Local regulations in several European countries do not permit the
use of a breakable wall-mounted power supply connection for high
current single phase mains supplies. In these areas, the mains power
supply to the ICP-OES should be hardwired to the wall.
For Canada only, the power cable shipped from the factory must be
replaced with a locally sourced UL/CSA cable such as Carol 02728.
In areas where 208/220/240 volt single phase power supplies are not
commonly available, two active phases from a two or three phase
system may be used.
The instrument supply draws a maximum of 28 amps RMS at
187 volts with a power factor of approximately 0.70. In areas where
208 to 240 V AC supplies are not available at the required rating, a
380/415 volt, phase-to-phase, to 220 volt single phase step down
transformer may be installed.
220 V AC 10%
provides 230 V AC +6%, -14%
Personal computer (Typical) 115, 120, 220, 240 V AC, 50/60 Hz 300 VA
Gas Supplies
The installation of compressed or liquid gas supplies must comply
with the rules and/or regulations imposed by the local authorities
responsible for such use in the workplace.
Argon Nitrogen
Required flow rates 0.732 L/min 3.5 L/min (if nitrogen is used to
purge the polychromator)
Cryogenic Liquids
Cryogenic liquid gases are stored under pressure at very low
temperatures in Portable Liquid Cylinders (PLCs).
For high gas flow rates and/or low ambient temperatures, it may be
necessary to obtain the gas by passing the liquid through an external
evaporator rather than use the internal pressure building facility of
the PLCs.
Liquid argon and liquid nitrogen may not be stored for extended
periods and often have special storage requirements. Contact your
local authorities and cryogenic gas supplier for more detailed
information on storage requirements and boil-off rates for local types
of PLCs.
NOTE The operation manual, mounting and assembly hardware for the cooling system
are included in the water cooler packaging. Care should be exercised to locate
all of these articles before the pack is discarded.
NOTE Pressure regulation is recommended for supplies where the cooling water
pressure may be subject to fluctuations. Pressure regulation is necessary for
supplies that may exceed the maximum permissible pressure of 310 kPa
(45 psi).
The instrument is equipped with a water flow sensor, which will stop
operation of the plasma and camera Peltier assembly if the cooling
water flow through the instrument drops below 1.1 L/min (0.3 gpm).
CAUTION Always ensure the water cooling system is on before igniting the plasma.
Provided that the flow rate can be maintained above 1.1 L/min and
maximum supply pressure is below 310 kPa (45 psi), the cooling
water supply may be taken from an in-house domestic water system,
if local regulations permit.
An alternative is to install a recirculating water cooler system. The
system should provide cover for the reservoir to prevent evaporation
and stop contamination by dust or other impurities. Algicide should
be used.
A recirculating water cooling system has the advantage of reducing
the volume of water that will be required over the life of the
instruments operation. Because of the limited size of the water
cooler's reservoir, in the case of accidental damage, the amount of
water damage that can occur will also be limited.
Drain Vessel
The Agilent ICP-OES system needs a drain vessel for disposal of
excess fluids and vapors from the spraychamber. Suitable tubing is
supplied with the spectrometer for use with inorganic solvents. When
using organic solvents, different drain tubing that is suitable for the
solvent in use will be required.
A chemically inert container, not glass or of a narrow-necked style, to
hold a minimum of 2 liters (4 pints) of waste must be provided by the
user. It should be located underneath the sample compartment (or on
the right side of the instrument), where it is protected by the bench
and in full view of the operator.
Laboratory Conditions
The area selected for the operation of an AgilentICP-OES system
must be free from drafts, corrosive atmospheres and vibration.
Sample preparation areas and materials storage facilities should be
located in a separate room.
The area should be a dust-free, low humidity environment. Air
conditioning is strongly recommended for control of the
environment. For optimum analytical performance it is
recommended that the ambient temperature of the laboratory is
between 20 and 25 C (68 and 77 F).
Temperature Control
The heat generated by the ICP-OES system into the laboratory is
about 3650 watts, (joules per second) or 13140 kilojoules per hour
(12455 BTU per hour).
NOTE Most of the 3650 watts generated by the system is vented outside the
laboratory by the exhaust system.
Total load with both the water cooler and ICP-OES in the same room
would be up to 4650 watts or 16740 kilojoules per hour (15867 BTU
per hour).
Cleanliness
Limit dust levels to less than 36000000 particles (0.5 microns or
larger) per cubic meter of air.
NOTE If an external cooling air supply system is required, an air inlet duct attachment
(part number 0110595300) must be ordered with the instrument.
Personal computer
Printer
Van der Heijden Typ 001VD for ICP 350 mm 405 mm 505 mm 42 kg
and Van der Heijden Typ Minore IIVD 13.8 in 15.9 in 19.9 in 93 lb
AGM 1
3. Installation
Personal Computer Requirements 31
Installing the ICP Expert II Software 32
PCI-GPIB Driver Installation 35
PCI-GPIB Communications Card Installation 36
Hardware Installation 40
Minimum Recommended
Pentium 4 processor, 2.8 GHz or higher Pentium 4 processor, 3.0 GHz or higher
1 GB of RAM 2 GB of RAM
Super VGA monitor with high color (16 bit) Super VGA monitor with high color (16 bit)
display, 800 x 600 resolution display, 1024 x 768 resolution
16 bit sound card and speakers 16 bit sound card and speakers
One full length PCI-compatible slot for I/O card One full length PCI-compatible slot for I/O card
Other PC Ports at least one serial (RS232) port Other PC ports at least one USB port, two serial
(RS232) ports
NOTE Do not plug the PCI-GPIB card into the PCI slot of the controlling computer
before installing the ICP Expert II Software. When installing the ICP Expert II
Software for the first time, it is important to make sure that the PCI-GPIB card
driver is installed before plugging in the PCI-GPIB card. PCI-GPIB driver
installation automatically occurs during installation of the ICP Expert II
Software.
1 If you have turned off the auto-start capability of your CD-ROM drive, you will need to run the setup.exe file on the
CD-ROM.
TIP To familiarize yourself with the ICP Expert II software, after installation, click the
Windows Start button and choose Programs > ICP Expert II > ICP Expert II
Help. When the ICP Expert II Help appears, click the How to link to view step-
by-step instructions on how to use the Agilent 700 Series ICP-OES.
CAUTION The components on the communications card and in the PC are highly static-
sensitive. To avoid damaging these components you must drain any static
charges from your body before installing the board, and prevent the generation
of any new static charges during the installation. This can be done by wearing
an ESD (electrostatic discharge) wrist strap attached to a grounding point. You
can obtain a disposable ESD strap from Agilent (part number 7910031300); or
you can obtain one from your local electronics supplier.
Hardware Installation
Your ICP-OES should be ready to operate after the Agilent field
service engineer has installed it. However, you may need to set up
items such as the peristaltic pump, nebulizer, torch or spraychamber.
For instructions on how to do this, please refer to the online Help.
Click the Windows Start button and choose Programs > ICP Expert
II > ICP Expert II Help.
When the ICP Expert II Help appears, click How to > Install
instrument components to view step-by-step instructions on how to
remove and install the various instrument components.
4. Operation
Analysis Checklist 42
Turning on the Instrument and Software 42
Turning On the Instrument for the First Time (or From
Shutdown Mode) 42
Turning On the Instrument From Standby Mode 43
Creating/Opening a Worksheet 44
Developing a Method 45
Preparing for Analysis 46
Calibrating the Detector 47
Aligning the Torch 49
Setting Operating Conditions 50
Setting up Standards 52
Running Samples 53
Printing a Report 54
Turning the Instrument Off 55
Using Standby Mode 56
Using Shutdown Mode 58
Analysis Checklist
You need to complete the following steps in turn to measure a
sample/s. You will find details on each step in this chapter.
Turn on the instrument and software
Create/open a worksheet
Develop a method
Prepare for analysis
Calibrate the detector
Align the torch
Set operating conditions
Set up standards
Run samples
Print a report
Turning On the Instrument for the First Time (or From Shutdown
Mode)
To turn on the instrument for the first time or from Shutdown
mode:
1 Check that all tubing on the torch, spraychamber, nebulizer and
peristaltic pump is correctly connected.
2 Close the torch compartment door, ensuring that the locking
lever is fully latched.
3 Switch on the computer, monitor and printer.
NOTE Both the RF supply circuit breaker and instrument power switch must be
switched on for the instrument to work.
NOTE You can check if the Peltier is on by looking at the Instrument Setup window
Status page. It may take some time to reach operating temperature. It is
recommended that you wait three to five minutes before starting an analysis.
Creating/Opening a Worksheet
Any new worksheets created are saved to a database. You will not be
able to create or open worksheets unless you have set up the VSDA
software. Instructions for setting up VSDA can be found in the online
Help (the online Help is accessed from within the VSDA software by
pressing the F1 key) and the software installation instructions that
came with your ICP Expert II software.
To create a new worksheet:
1 From the Main Index window, click the Worksheet button. The
Load Worksheet dialog box will be displayed.
2 Click New in the Load Worksheet dialog box. The Create a New
Worksheet dialog box will be displayed.
3 Highlight the name of the server where your data is to be stored
(this may be the local PC or a client-based server). A list of
available databases is displayed on the right side of the dialog
box, as well as below the server name on the left.
4 From the list below the server, click the database you are going to
use. Any worksheets already in existence will be displayed in the
panel on the right.
5 In the Name field, type a name for the worksheet that you are
about to create.
6 From the Menu bar, select the type of worksheet that you are
creating. Choose from quantitative or time resolved scan
worksheet files. You can also create a worksheet from a template.
Some templates are supplied in the ICP Expert II software.
7 Click Save. The Worksheet window will appear with the new
worksheet loaded.
Developing a Method
To develop a method:
1 Open the Method Editor by clicking the Method Editor button
(or the Method Editor icon on the Toolbar). The Element page
will be displayed.
2 On the Element page, ensure that the Periodic Table is displayed.
If it is not visible, choose Periodic Table from the View menu.
3 Select the required elements by clicking on each element in the
Periodic Table. This will display the Add Analysis Line dialog
box, which allows the addition of wavelengths, internal
standards and potential interferences.
NOTE You can also add multiple elements by holding down the CONTROL key and
clicking the required elements in the Periodic Table. The most prominent line
will be automatically added to the method as an analyte. Selected analytes will
be highlighted in blue.
NOTE See the ICP Expert II Help for a more detailed description of setting up a
method.
NOTE The plasma will take between 30 to 40 seconds to ignite. If it fails to ignite, refer
to the Troubleshooting section in the online Help for further information.
NOTE See the online Help for instructions on how to adjust the pressure bars.
NOTE This You can save the dark current measurement data by clicking Store Dark
Current.
4 Remove the sample tubing from the wash solution. Wipe the
tubing and transfer it into the wavelength calibration solution.
5 Once the wavelength calibration solution is aspirating, click
Calibrate to perform the wavelength calibration. After a short
while, a message will appear stating that the wavelength
calibration is complete, as well as indicating the wavelength
calibration range.
NOTE Preferably, this step should be performed by an Agilent field service engineer.
3 Remove the sample tubing from the wash solution. Wipe the
tubing and transfer it into the solution to be analyzed. Once the
solution is aspirating, click the Read Spectrum icon (or press the
F8 key on the keyboard) to perform a scan.
4 When the scan is complete, you may alter the settings for the
selected line, and perform another wavelength scan to compare
the effects of different operating conditions on the analytical
signal. This will allow you to determine the optimum settings for
the selected line. Ensure that Overlay is selected in the Graphics
menu so that you can compare the signals. Power, nebulizer flow
and viewing height (radial instruments) will have the greatest
impact on sensitivity.
NOTE For Agilent 720/730 Series ICP-OES instruments, the software can
automatically determine and set the optimal operating conditions for each line.
Refer to the online Help for further details.
CAUTION High plasma operating powers can result in torch meltdown. When running the
torch at a power setting above 1.4 kW, always monitor the torch and decrease
the power if it starts to overheat.
Plasma gas flows that are too low may also cause the torch to overheat and
melt.
NOTE Recommended values for the line settings are given in the Conditions used by,
Sample introduction and General settings online Help pages. Use these values
as a guide when entering your own values until you become more familiar with
the system.
5 Repeat the above steps until you are satisfied with your results.
Generally, this involves finding the optimum intensity.
Setting up Standards
To set up standards:
1 Click the Standards tab. Ensure the Calibration mode is
Quantitative, not Standard Additions.
2 Enter the number of standards required. As you add standards,
columns will be added to the Standard Concentrations table.
3 Enter the concentration of each standard in the table.
4 Select the units required by clicking on the required cell and
choosing the units from the drop-down menu.
Leave the correlation coefficient at its default value of 0.995000. The
MultiCal Parameters table can remain unchanged.
NOTE Clicking on the element name in the Standard Concentrations table will
highlight the row. If your standards have the same values, you can quickly copy
and paste the row values by right-clicking on the table and using the Copy and
Paste options.
Running Samples
To run samples:
1 From the Worksheet window, click the Sequence tab.
2 Ensure the Sample source is set to Manual.
3 Click the Sequence Editor button. Enter the required number of
samples in the Sample count field.
4 Select Begin with calibration and Include a blank in
calibration. Select the other boxes as required.
5 Click OK.
6 From the Sequence page, click the Sequence Parameters button.
7 Set On calibration failure to Stop.
8 Click OK. When you are asked to confirm your selections, click
Yes.
9 Click the Analysis tab and do the following:
a Ensure your samples are selected. This will be indicated by a
yellow highlight. To select all solutions, click the Tube cell in
the top left corner of the table.
b Click the Start Analysis icon (or press the SHIFT +F8 keys)
in the Toolbar to begin the analysis and follow the
subsequent prompts.
NOTE If the Start icon is grayed out, hold the cursor over the icon until the hint text
(enabled in the View menu) is revealed. This will indicate what action is
required to enable the icon.
NOTE For further information about running an analysis, refer to the ICP Expert II Help.
Printing a Report
To print a report:
1 From the File menu, choose Report Settings.
NOTE When printing a report, you can simply select Print from the File menu.
However, if you do so, a report will be generated based upon the last used
Report Settings. You will not be able to select the data to be included in the
printout.
2 In the Report Settings dialog box on the General page, specify the
following:
whether to generate a native or HTML report
the report style
whether to report all of the results or only those highlighted
whether to print the report automatically at the end of the
analysis run
the data content to be included
what calibration data is to be included
the match sample label string, if required
whether to print all or only the visible columns
the units
3 On the Customize page, specify the appearance of the report
including whether you want to print the report in color, the size
of the graphs, and add a header, footer and logo.
NOTE When selecting the report Style, please note the following:
Tabular produces a report in the same form as the worksheet. Calibration data
cannot be selected for printing.
Mean Only reports are designed for analysts who are only interested in final
results. As a result, many of the Content options and the option to print
Calibration data are disabled.
An All Data report allows you to select all of the Report Settings options to
include in your report.
NOTE You can preview the report if required by clicking the Preview button.
Previewing the report allows you to ensure that you have included all of the
data you require.
NOTE The yellow Plasma emergency off button, located next to the mains power
switch on the front of the instrument, is designed to be used only in an
emergencyfor example if the torch is melting. It is not intended to be used
every time you want to extinguish the plasma. If used to extinguish the plasma,
you will need to reset the Plasma emergency off button to the on position
before the plasma can be re-ignited.
Figure 8 Peristaltic pump with pressure bars pushed up and resting bar
freely down, allowing tubing to be loosened
6 It is recommended that you switch off the water cooler.
NOTE If the water cooler is turned off, the Peltier will automatically turn off.
NOTE The yellow Plasma emergency off button, located next to the mains power
switch on the front of the instrument, is designed to be used only in an
emergencyfor example if the torch is melting. It is not intended to be used
every time you want to extinguish the plasma.
If used to extinguish the plasma, you will need to reset the Plasma emergency
off button to the on position before the plasma can be re-ignited.
3 Turn off the Peltier by choosing Peltier Off from the Analyze
menu.
CAUTION The ICP-OES gas control unit supplies uninterrupted gas purge to the camera
(Agilent 720/730 Series ICP-OES only) and polychromator assemblies to
minimize the ingress of moisture. It is strongly recommended that gas supplies
be left on at all times except during long periods of non-use. It is recommended
to turn off the camera Peltier before turning off the instrument gas supply.
4 Switch off the mains power switch (the green light will go off).
This will turn off the complete instrument, including the Peltier
(if not done so manually earlier) as well as the polychromator
thermostating system.
5 Loosen the peristaltic pump tubes by releasing the pressure bars
and lift the tubes out of the grooves. To do this:
a Push up the pressure bar screws. This releases them from the
resting bar (refer to Figure 8).
b Allow the resting bar to swing downwards.
c Lift the tubing out of the grooves.
6 Switch off the water cooler, laboratory exhaust system and any
accessories (where applicable), and shut down the argon gas
supply at the cylinder.
7 Exit the ICP Expert II software, if it is no longer required, by
choosing Exit from the File menu. Switch off the printer and
monitor.
8 Restarting the instrument will take several hours due to gas
purge and polychromator thermal stabilization time.
NOTE If the instrument is not going to be in use for a period of time the torch, cone,
snout and torch compartment should be cleaned of any deposits, dirt or residue.
5. Maintenance
Routine Maintenance 62
Maintenance of Consumables 63
Cleaning 64
Spare Parts 64
This chapter includes the Agilent 700 Series ICP optical emission
spectrometer maintenance requirements that may be carried out by
an operator. Any maintenance procedures not specifically mentioned
in this chapter should be carried out only by Agilent-trained, Agilent-
qualified or Agilent-authorized field service engineers.
NOTE This section refers to maintenance procedures for the ICP-OES instrument. You
should refer to your PC and printer manuals for their maintenance procedures,
and to the ICP Expert II Help for the maintenance procedures for any
accessories you ordered.
Routine Maintenance
The following parts of the ICP-OES require routine maintenance.
Maintenance instructions are included in the ICP Expert II Help. To
access these instructions, click the Windows Start button, then
Programs > ICP Expert II > ICP Expert II Help. Click the
Maintenance link.
Hourly
Check and, if necessary, empty the drain vessel.
Daily
Check the water level in the Argon Saturator Accessory (ASA)
before every use (if applicable).
Clean the surface of your ICP-OES (spills should be cleaned up
immediately).
Inspect the pump tubing and replace if it has lost its elasticity.
Unclip the pump tubes when the pump is not in use.
Weekly
Clean the torch.
Clean the cone (710/720/730 ICP-OES instruments).
Clean the snout (715/725/735 ICP-OES instruments).
Clean the bonnet (715/725/735 ICP-OES instruments).
Clean the spraychamber.
Clean the nebulizer.
Monthly
Clean the cooling air intake filter on top of your instrument.
Inspect the state of the induction coil. Contact your local Agilent
office or representative if maintenance is required.
Check the water level in the water cooler (refer to manual
supplied with the water cooler for details).
Perform a wavelength calibration.
Maintenance of Consumables
For information on maintaining consumables (such as replacing the
torch, cleaning the cone, replacing the peristaltic pump tubes and
replacing fuses), refer to the online Help. Specifically:
1 Click the Windows Start button and choose Programs > ICP
Expert II > ICP Expert II Help.
2 When the ICP Expert II Help appears, click Maintenance to view
step-by-step instructions on how to maintain instrument
consumables.
Cleaning
Any spills in the sample compartment should be wiped up
immediately.
The exterior surfaces of the ICP-OES should be kept clean. All
cleaning should be done with a soft cloth. If necessary, this cloth can
be dampened with water or a mild detergent. Do not use organic
solvents or abrasive cleaning agents.
Spare Parts
For spare parts and consumables ordering information, refer to the
Agilent Technologies. Web site:
www.agilent.com