269-251401 Rev A. Evolution 200 Series User Guide 2013
269-251401 Rev A. Evolution 200 Series User Guide 2013
269-251401 Rev A. Evolution 200 Series User Guide 2013
For U.S. Technical Support, please contact: For International Support, please contact:
Thermo Fisher Scientific Thermo Fisher Scientific
5225 Verona Road Telephone: +1 608 273 5017
Madison WI 53711-4495 U.S.A. E-mail: support.madison@thermofisher.com
Telephone: 1 800 532 4752
E-mail: us.techsupport.analyze@thermofisher.com
Thermo Fisher Scientific Inc. provides this document to its customers with a product purchase to use in the
product operation. This document is copyright protected and any reproduction of the whole or any part of this
document is strictly prohibited, except with the written authorization of Thermo Fisher Scientific Inc.
The contents of this document are subject to change without notice. All technical information in this
document is for reference purposes only. System configurations and specifications in this document supersede
all previous information received by the purchaser.
Thermo Fisher Scientific Inc. makes no representations that this document is complete, accurate or error-
free and assumes no responsibility and will not be liable for any errors, omissions, damage or loss that might
result from any use of this document, even if the information in the document is followed properly.
This document is not part of any sales contract between Thermo Fisher Scientific Inc. and a purchaser. This
document shall in no way govern or modify any Terms and Conditions of Sale, which Terms and Conditions of
Sale shall govern all conflicting information between the two documents.
For Research Use Only. This instrument or accessory is not a medical device and is not intended to be used
for the prevention, diagnosis, treatment or cure of disease.
Contents
Chapter 1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .1
About the Software and Hardware. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Setting Up the Instrument. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Before Using the Spectrophotometer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
About This Document . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2
Organization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Conventions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
Contacting Us . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
E-mailing Data Files . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Product Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Warranty Period . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Items Not Covered by Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Warranty Repair . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Shipping Cost for Items Covered by Warranty . . . . . . . . . . . . . . . . . . . . . . . . 6
Update Policy . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
Trademarks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Site Preparation and Safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Introduction
This document covers the Thermo Scientific™ INSIGHT™ software and the Evolution™ 200
series UV-Visible spectrophotometers and accessories.
Contents
• About the Software and Hardware
• Contacting Us
• Product Warranty
• Trademarks
• Site Preparation and Safety
Overview
The Evolution 200 series UV-visible spectrophotometers integrate advanced hardware
features with the power and flexibility of a wide range of Smart Accessories™. The 200 series
spectrophotometers include:
• Evolution 201
• Evolution 220
• Evolution 260 Bio
For a list of sampling and other accessories available for these instruments, see Evolution 200
Series Accessories.
All instruments include our INSIGHT software for data collection and analysis. The
INSIGHT software provides five modes of operation:
• Fixed, to measure the light passing through the sample at one or more wavelengths.
• Scan, to measure the light that passes through the sample over a range of wavelengths.
• Quant, to set up and perform quantitative analyses of sample data.
• Rate, to make time- and temperature-based kinetics measurements.
• Live Display, for quick measurements and simplified data collections in Fixed or Scan
mode.
The Evolution 260 Bio systems also have operating modes for the common biological
applications, including Nucleic Acid and Protein A280. See INSIGHT Bio Applications for
more information.
Organization
This document has the following main sections:
Section Description
Introduction Overview of the instrument and this document plus
information about registering your instrument, contacting us,
and the system warranty.
Evolution 200 Series Complete instructions for using and maintaining the
Spectrophotometers instrument.
INSIGHT Software Complete instructions for using software features other than
biological applications and Performance Verification (see below).
Bio Applications Instructions for using biological applications to analyze samples.
Accessories Complete instructions for using accessories with the instrument.
Performance Verification Instructions for setting up and running tests to check the
performance of the instrument.
INSIGHT Security Optional software that adds features for digitally signing files
and verifying digital signatures.
Conventions
This document uses the following conventions:
WARNING Indicates a hazardous situation which, if not avoided, could result in death or
serious injury.
CAUTION Indicates a hazardous situation which, if not avoided, could result in minor or
moderate injury.
NOTICE Follow instructions with this label to avoid damaging the system hardware or
losing data.
Contacting Us
For U.S. Technical Support, please contact:
Note Please have the instrument serial number available when you contact us.
Note Please do not send exported report files or user-defined workbooks—these do not
contain the information we need.
Product Warranty
Warranties herein are for products manufactured by Thermo Fisher Scientific or its authorized
dealers.
Thermo Fisher Scientific warrants that this product is free from defects in labor and materials
and shall conform to its product specifications as defined in the product user documentation.
This warranty covers parts (except those specified below) and labor, and applies only to
equipment which has been installed and operated in accordance with the documentation
supplied by Thermo Fisher Scientific, and which has been serviced only by authorized
Thermo Fisher Scientific dealers or service personnel. This warranty does not apply to
equipment and accessories that have been modified or tampered with in any way, misused, or
damaged by accident, neglect, or conditions beyond Thermo Fisher Scientific’s control.
Related Topics
Warranty Period
Warranty Repair
Update Policy
Warranty Period
With the exception of SPECTRONIC™ 20+, SPECTRONIC 20D+, GENESYS™ 20 and the
Helios™ Epsilon spectrophotometers, all UV Visible spectrophotometers come with a
standard warranty that is fourteen months from the date of shipment or twelve months from
the date of installation (whichever occurs first). The replacement parts and spare components
are warranted for 90 days from the date of shipment. If such parts are intended for long term
storage, it is recommended that their functionality is tested immediately upon receipt and any
problems reported to Thermo Fisher Scientific within the above specified warranty period.
Thermo Fisher Scientific reserves the right to request the return of failed components that are
being replaced under warranty.
User should complete and return the Product Registration card, and retain proof of delivery
date.
Warranty Repair
In the event of failure within the warranty period, Thermo Fisher Scientific will at Thermo
Fisher Scientific’s option, repair or replace the product not conforming to this warranty. There
may be additional charges, including freight, for warranty service performed in some
countries. For service, call Thermo Fisher Scientific (or its authorized dealer outside the
United States and Canada). Thermo Fisher Scientific reserves the right to ask for proof of
delivery date.
Warranty repairs for UV Visible instruments may vary per product line and region. UV
Visible products manufactured by Thermo Fisher Scientific are covered by a
“return-to-factory” parts and labor warranty. In some regions, on-site warranties, extended
warranties and service contracts may be purchased for all products.
In the event that on-site warranty service is required, and no on-site service contract is active
on this product, the customer is responsible for any travel expenses, labor costs, and additional
expenses which may be incurred in the performance of on-site warranty service.
Please contact your dealer or Thermo Fisher Scientific Technical Service department for
further information. A Return Authorization Number must be obtained from Thermo Fisher
Scientific Technical Service before returning any product for in-warranty repair or
replacement.
This warranty covers return shipment (standard ground) charges for returning the repaired or
replaced product to the customer.
Update Policy
Thermo Fisher Scientific may, from time to time, revise the performance of its products, and
in doing so incur no obligation to furnish any such revisions to any Thermo Fisher Scientific
customer.
Trademarks
Microsoft, Windows and Excel are either trademarks or registered trademarks of Microsoft
Corporation in the United States and/or other countries.
Teflon, Viton and Freon are either trademarks or registered trademarks of E. I. du Pont de
Nemours and Company (DuPont) or its affiliates in the United States and/or other countries.
Hellma is either a trademark or a registered trademark of Hellma GmbH in the United States
and/or other countries.
AccuVac and Hach are either trademarks or registered trademarks of Hach Company in the
United States and/or other countries.
All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries.
Contents
• Safety Considerations
• Operating Precautions
• Spectrophotometer Basics
• Installing and Removing Accessories
• Using the Spectrophotometer
• Maintenance
• Ordering Parts
Safety Considerations
Each person using the spectrophotometer should read the general safety information in the
Site and Safety Information guide (see the documentation CD) and the instrument-specific
safety information provided here.
Contents
• Safety Label Locations
• Lifting or Moving the Instrument
WARNING Do not operate this system without following the safety precautions described
in this manual and the documentation that came with your system.
Back view
Compliance label
Related Topics
Removing and Installing the Detector Module
Hand hold
Hand hold
Operating Precautions
The spectrophotometer contains precise optical components. Handle it carefully and follow
these precautions:
• Do not allow moisture to leak into the instrument interior.
• Wipe off spilled chemicals immediately.
• Do not drop the instrument.
• Protect the instrument from mechanical shock.
• Protect the instrument from dust.
Spectrophotometer Basics
Contents
• Spectrophotometer Components
• Connectors
• Removable Panels
• Corrosion Protection
Spectrophotometer Components
The following illustration identifies major components visible on the outside of a typical
spectrophotometer. (Some components may not be present on all instruments.)
The z-height (distance from the bottom of the cell to the center of the light beam) for the
spectrophotometer is 8.5 mm.
Reference compartment
Sample compartment
Optional touchscreen
Keypad
Related Topics
Opening and Closing the Sample and Reference Compartment Doors
INSIGHT Software
Keypad
Touchscreen
Related Topics
Sample Compartment Front Panel
Related Topics
Local Control Instruments
INSIGHT Software
All instruments include our INSIGHT™ software for data collection and analysis. See
INSIGHT Software.
Keypad
All instruments include the keypad shown below. For a list of keypad functions, see Using the
Spectrophotometer.
Related Topics
Keypad Operation
Touchscreen
The touchscreen is included only on Local Control versions of the spectrophotometer.
Related Topics
Touchscreen keyboard
Connectors
This section shows the locations of the connectors inside and outside the spectrophotometer.
Contents
• Outside Cover
• Inside the Sample Compartment
Outside Cover
This illustration shows the connectors on the instrument outside cover.
Back view
USB
USB
Label Function
Computer Dedicated USB port for connecting an external computer. See
Computer Control Instruments.
Hg lamp Connects to the optional Mercury Lamp accessory.
Power Connects to main power cord.
Power switch System power On/Off switch.
Purge Connects to optional purge source. See Purging the Sample
Compartment.
Label Function
Trigger Connects to optional accessories that accommodate external trigger
input or output. For more information, find the Integrated Fiber Optic
Module User Guide in the documentation media.
INSIGHT can be used to send a 3.3V TTL output signal that triggers
an installed accessory to begin an operation. The signal can be sent at the
start or end of a measurement. See Options > Preferences > Triggering.
WARNING Avoid shock hazard. Power off the spectrophotometer and disconnect the
power cord from the wall outlet or power strip before disconnecting the power cord from
the spectrophotometer.
To ensure proper shielding of USB devices, install a Ferrite ring (samples provided) on the end
of any USB cable connected to the instrument. Loop the cable once and clamp the ring over
the loop as shown below.
Ferrite ring
Note The spectrophotometer does not support high power USB devices.
Related Topics
Triggering
Smart Accessory
connector
Related Topics
Installing a Smart Accessory
Removable Panels
This section covers panels that can be removed to allow access to features on the instrument.
Contents
• Sample Compartment Front Panel
• Tubing Access Panel
• Mercury Lamp Access Cover
• Detector Access Cover
• Detector Access Panel
To replace the panel, align the tab on the bottom edge with the slot in the instrument cover
and press the panel in place.
Removable panel
Tab
Related Topics
Installing a Manual Accessory
Internal fluid ports for routing fluid from the instrument to an installed accessory are located
inside the sample compartment.
Foam
Internal fluid ports Screw
The connectors fit 1/4-inch internal diameter tubing. An adapter may be required to connect
the tubing to the accessory or recirculator. Purchase adapters separately.
Cords and cables for installed accessories can also be routed behind this panel. Push the cord
through the foam and the opening below it. The panel itself can be removed temporarily to
accommodate a large connector. To remove the panel, loosen the screw.
Related Topics
Installing a Liquid Thermostatted Accessory
WARNING Avoid UV radiation hazard in the mercury lamp compartment. Protect eyes
and skin from exposure.
Lift the cover to install the optional Mercury Lamp. For more information, refer to the
Mercury Lamp User Guide in the documentation media.
CAUTION Safety glasses with side shields or goggles with solid side pieces are the only
equipment that provides adequate eye protection against direct and reflected UV light.
Related Topics
Removing and Installing the Detector Module
Thumbscrew
Window (optional)
Removable panel
To replace the panel, hook the tabs on the bottom edge over the slots in the sample
compartment side wall and tighten the thumbscrew.
Related Topics
Removing and Installing the Detector Module
Corrosion Protection
The spectrophotometer contains precise optical components that may be damaged by a
corrosive environment. Before using the instrument to analyze volatile, corrosive or caustic
samples, install the optional sample compartment windows and, if necessary, purge the sample
compartment.
Related Topics
Installing Sample Compartment Windows
NOTICE The warranty does not cover damage to internal optics or electronics caused by
failure to use sample compartment windows.
Related Topics
Purging the Sample Compartment
Related Topics
Installing Sample Compartment Windows
Installing Purge
Related Topics
Installing Sample Compartment Windows
Installing Purge
Installing Purge
Use 1/4-inch internal diameter tubing to connect a purge gas source to the purge fitting on
the back of the spectrophotometer.
Related Topics
Installing Sample Compartment Windows
Related Topics
10 mm Rectangular Sample Cell Holder
Use the cell lifter to raise the cell for easy removal. This cell holder allows fine adjustment of
the vertical and horizontal position of the cell. This feature is useful for aligning reduced
volume and microcuvettes with the light beam. See z-height.
Cell lifter and vertical alignment
Horizontal alignment
Related Topics
Rectangular Reference Cell Holder
Spectrophotometer Z-height
The standard reference holder accommodates rectangular reference cells with these
pathlengths: 10 mm, 20 mm, 40 mm, 50 mm and 100 mm.
Related Topics
10 mm Rectangular Sample Cell Holder
It is not necessary to power off the instrument while installing or removing a manual
accessory.
Alignment pins
A B
Toggle
1. Open the sample compartment door. See Opening and Closing the Sample and
Reference Compartment Doors.
2. Remove any accessory from the compartment (see the instructions below).
3. Position the accessory so the two alignment holes on its baseplate fit over pins A and B in
the sample compartment.
4. Gently push down on the front edge of the baseplate until it snaps in place under the
toggle.
5. Close the sample compartment door. See Opening and Closing the Sample and Reference
Compartment Doors.
For operating instructions and installation details, see the user guide for the accessory.
1. Use the handle or finger hole on the accessory baseplate to release the plate from the
toggle.
2. Lift the plate off the alignment pins and remove the accessory from the sample
compartment.
Related Topics
Installing a Liquid Thermostatted Accessory
These accessories install in the sample compartment. A connector under the accessory
baseplate provides data communication and power.
It is not necessary to power off the instrument while installing or removing a Smart Accessory.
Alignment pins
A B
Alignment pins
1. Open the sample compartment door. See Opening and Closing the Sample and
Reference Compartment Doors.
2. Remove the sample compartment front panel. See Sample Compartment Front Panel.
3. Remove any cell holder or accessory. See To remove a Smart Accessory.
4. Insert the accessory.
a. Grasp the accessory by the handles and lower it into the sample compartment
aligning the connector under the baseplate with the connector in the sample
compartment floor.
Smart Accessory
b. Line up the two holes at the back of the accessory with pins A and B in the sample
compartment.
c. Press down on the front of the accessory to secure the connection.
The software displays a prompt to initialize the accessory.
CAUTION Avoid pinch hazard. Keep hands and objects clear of the accessory during
initialization.
5. Replace the sample compartment front panel. See Sample Compartment Front Panel.
6. Close the sample compartment door. See Opening and Closing the Sample and Reference
Compartment Doors.
7. Click OK to initialize the accessory.
Initialization reads information about the accessory and, for cell changers, moves it to
position 1.
For operating instructions and installation details, see the user guide for the accessory.
1. Open the sample compartment door. See Opening and Closing the Sample and
Reference Compartment Doors.
2. Remove the sample compartment front panel. See Sample Compartment Front Panel.
3. Grasp the handles and pull the accessory up and out of the sample compartment.
Related Topics
Installing a Liquid Thermostatted Accessory
The Recirculator Tubing Kit provides parts for connecting an accessory to a temperature
controller or fluid recirculator.
4. Use additional tubing and adapters (not included), if necessary, to connect the fluid ports
inside the sample compartment to the ports on the accessory.
For a cell changer, set the changer to position 1 before connecting the tubing to the ports
on the accessory. Use just enough tubing to allow the cell changer to move through all
possible positions. This requires approximately 10 inches (25 cm) of tubing for each port.
Note Ensure the tubing does not obstruct the light beam or prevent the sample
compartment door from closing. If necessary, secure the tubing using the supplied
clips.
Related Topics
Installing a Smart Accessory
1. Open the reference compartment cover. See Opening and Closing the Sample and
Reference Compartment Doors.
2. Grasp the cell holder by the handle and lower it into the reference compartment.
Handle
3. Insert the pin in the alignment hole and press the holder in place.
4. Close the reference compartment cover.
For descriptions of the available reference holders, see Reference Cell Holders or search for that
name in the INSIGHT Help system.
Related Topics
Installing a Smart Accessory
Contents
• Computer Control Instruments
• Local Control Instruments
All instrument features can be run from the computer using our INSIGHT software. Basic
operations such as powering the spectrophotometer on and off, measuring a sample or
baseline and running predefined macros can also be performed from the instrument keypad
(see Keypad Operation).
Related Topics
Setting up a Computer Control Instrument
1. Connect the supplied USB cable to the Computer port on the back of the
spectrophotometer.
2. Connect the other end of the cable to the USB port on the back of the computer.
3. Turn on the spectrophotometer main power switch (on the back panel).
4. If the main power switch is already on, press the Power button on the keypad.
Wait for the system to initialize. This may take several minutes. The power indicator on
the keypad stops blinking when initialization is complete.
Note If the instrument fails to initialize, the power indicator blinks rapidly. Contact us.
Related Topics
Turning off a Computer Control Instrument
Related Topics
Setting up a Computer Control Instrument
All instrument features can be run from the touchscreen or remote computer. Basic operations
such as powering the spectrophotometer on and off, measuring a sample or baseline and
running predefined macros can also be performed from the instrument keypad (see Keypad
Operation).
Related Topics
Setting up a Local Control instrument for touchscreen operation
1. Turn on the spectrophotometer main power switch (on the back panel).
2. Press the Power button on the keypad.
Wait for the system to initialize. The power indicator on the keypad stops blinking when
initialization is complete. INSIGHT software starts automatically and appears on the
touchscreen.
Tip For routine use, leave the power switch on the instrument back panel in the On
position and use the Power button on the keypad to turn the instrument on and off.
Related Topics
Setting up a Local Control instrument for remote operation
Touchscreen keyboard
1. Connect the supplied USB cable to the Computer port on the back of the
spectrophotometer.
2. Connect the other end of the cable to the USB port on the back of the computer.
3. Turn on the spectrophotometer main power switch (on the back panel).
4. Press the Power button on the keypad.
Wait for the system to initialize. The power indicator on the keypad stops blinking when
initialization is complete. INSIGHT software starts automatically and appears on the
touchscreen.
5. Make sure the remote computer is on.
6. Do one of the following:
• Hold down the Function 4 and Function 2 keys on the instrument keypad for
4 seconds.
• From the INSIGHT software displayed on the touchscreen, choose System Settings
> System. In the Instrument Control group, select Computer.
7. Double-click the INSIGHT desktop icon on the remote computer to start the software.
Note To return control to the touchscreen, from the INSIGHT software displayed on the
remote computer:
1. Do one of the following:
• Choose System Settings > System. In the Instrument Control group, select
Instrument Local.
• Hold down the Function 4 and Function 1 keys on the instrument keypad for
4 seconds.
2. Double-click the INSIGHT desktop icon on the touchscreen to start the software.
Related Topics
Setting up a Local Control instrument for touchscreen operation
Tip For routine use, leave the power switch on the instrument back panel in the On
position and use the Power button on the keypad to turn the instrument off.
Related Topics
Setting up a Local Control instrument for touchscreen operation
Keypad Operation
Use the keypad on the Evolution 200 instrument front panel to perform the following
operations.
Power Turn the instrument on or off (press and hold 4 seconds to power
off ).
Keypad Shortcuts
Each function key can be assigned to any executable program such as the Windows Notepad
application or a script created using our CUE application.
Note Function key assignments are shared by all users of the system computer. However,
a specific user’s ability to run a program from a function key is controlled by their
Windows user account privileges, which are assigned by the Windows administrator.
3. To associate an executable program with a function key, enter the program’s full path
name in one of the Button boxes or use the browse (...) button to the right of an entry box
to locate and select the program and then click Open.
This example shows how to select a script from the CUE Example Scripts folder provided
with INSIGHT software. CUE scripts have the file name extension *.TCUE.
5. When you are finished assigning programs to function keys, click Apply and then click
Close.
The button assignments are persistent and will be active the next time a user logs on.
Maintenance
Contents
• Routine Maintenance
• Cleaning the Instrument
• Cleaning the Touchscreen
• Removing and Installing the Detector Module
Routine Maintenance
Clean the instrument regularly using approved methods (wipe off spilled chemicals
immediately).
Related Topics
Maintenance
To clean the instrument exterior and the sample or reference compartment interior
1. Use a lint-free cloth dampened with a weak solution of detergent and water to wipe the
surface as necessary.
2. Repeat using a cloth dampened with plain water.
3. Dry the surface with another cloth.
Related Topics
Maintenance
1. Press the Power button on the keypad to turn off the spectrophotometer.
2. Wipe the screen gently with a soft, lint-free cloth dampened (not wet) with distilled
water.
To remove stubborn materials or oil, use a standard screen cleaner kit that includes
antistatic wipes. Spray a little solution on the wipe and rub it gently across the screen.
3. Dry the screen with a clean, lint-free cloth.
Related Topics
Maintenance
Avoid UV radiation hazard in the detector compartment. Protect eyes and skin from
exposure.
CAUTION Safety glasses with side shields or goggles with solid side pieces are the only
equipment that provides adequate eye protection against direct and reflected UV light.
1. Turn off the spectrophotometer main power switch (on the back panel).
2. Remove the detector access cover. See Detector Access Cover.
3. Loosen the thumbscrew that attaches the detector module to the instrument, and
carefully lift the module straight up.
NOTICE Avoid touching the connector pins on the bottom of the detector module.
Thumbscrew
Detector module
5. Line up the connector on the bottom of the detector module with the connector in the
detector compartment, and then press down on the module to secure the connection.
6. Tighten the thumbscrew by hand.
7. Replace the detector access cover.
Related Topics
Maintenance
Ordering Parts
To order replacement parts, contact us using the information provided at the beginning of this
document.
INSIGHT Software
Contents
• Applications
• The INSIGHT Window
• Thermo Software IQ
• Operation
• Options
• System Settings
• Keyboard Shortcuts
Applications
Contents
• Fixed
• Scan
• Quant
• Rate
• Live Display
Fixed
The Fixed application measures the light passing through
the sample at one or more wavelengths. To work with this
application, click Fixed in the right pane in Home.
Related Topics
The INSIGHT Window
Scan
The Scan application measures the light that passes
through the sample over a range of wavelengths. To work
with this application, click Scan in the right pane in
Home.
Related Topics
The INSIGHT Window
Quant
Use the Quant application to set up and perform
quantitative analyses of sample data for Fixed or Scan
applications. To work with this application, click Quant
in the right pane in Home.
Related Topics
The INSIGHT Window
Rate
Use the Rate application to make time- and
temperature-based kinetics measurements. To work with
this application, click Rate in the right pane in Home.
Related Topics
The INSIGHT Window
Live Display
Use the Live Display application for quick measurements
and simplified data collections in Fixed or Scan mode. To
work with this application, click Live Display in the right
pane in Home.
Related Topics
The INSIGHT Window
Above the panes is a menu bar that can include features for working with workbooks,
performing math, analysis and other operations, and getting help. The available features
depend on the selected application, whether spectral data is displayed, etc.
Alternatively, click this icon on the Windows taskbar. If the taskbar is not visible, click
the bottom (or other) edge of the Windows desktop to display it.
Related Topics
Navigation Pane
Right Pane
Menus
Navigation Pane
The navigation pane at the left side of the window contains features for the current
application:
Task Buttons
Action Buttons
Other Navigation Pane Features
Related Topics
Right Pane
Menus
Task Buttons
The main task buttons (present when the software starts) provide access to basic operations:
Button Description
Home Gives access to user groups and their applications. (The default is Classic.)
My Data Provides tools for storing and retrieving data and other information.
Options Includes features for controlling how users interact with the software.
System Settings Provides features for aligning accessories, calibrating lamps, updating
firmware and performing other tasks.
Button Description
Measure (current Displays data acquisition features for the current application.
application)
Reports Configures reports containing sample data and specifies how to
print reports.
Oligo Calculator Calculates molecular weight, extinction coefficients, concentration
factors and melting points for nucleic acid sequences.
Dye/Chrom. Editor For entering and editing new dyes rather than using one of the
precoded default dyes.
Related Topics
Action Buttons
Action Buttons
These action buttons appear at the top of the navigation pane when appropriate for the
opened application:
Button Description
Starts sample data acquisition, displaying prompts for confirming samples
information, loading samples, saving data or performing other actions. Available
only after a blank measurement for some applications. May measure single or
multiple samples, depending on the application settings.
Measures the baseline for the application.
Prints a copy of the spectrum and associated sample data to the default printer.
Note To start data acquisition automatically after a specified delay instead of responding
to a prompt, use Enable auto input trigger on the Preferences tab in Options.
Related Topics
Task Buttons
Feature Description
Overlay data Displays multiple spectra or multiple fixed data points as they are
acquired. The overlaid sample IDs appear in the upper-left corner of the
data display. To display the spectrum or data point of interest in red, click
its sample ID. The last sample measured appears in red at the top of the
legend. Available for Fixed and Scan applications only.
Advanced For selecting predefined formulas for additional data processing, with the
Calculations results appearing in a calculations table at the bottom of the right pane.
Options include basic math and statistics. The formulas can be applied
to specific samples and columns of data in the sample measurements
table or to selected rows and columns in the calculations table for the
current workbook or template. Calculations are applied to all
subsequently acquired data. Custom calculations are saved with the
workbook or template. Available for all applications except Rate.
Feature Description
To define calculations based on the samples table
1. In the calculations table, click the cell where you want the
calculation result to appear.
2. In the Advanced Calculations Cell Properties box, choose
Calculation.
3. Select an equation in the Equation list box.
The list of available equations is fixed.
4. Set Source to Samples table.
5. Select a sample measurement in the Column list box.
6. Specify the sample rows to include by entering their ID numbers in
the Range/Reference box.
Use a colon to specify a range (e.g., 2:5 or 1:N) or commas for
individual samples (e.g., 1,3,5).
7. To select the background color for the cell that displays the
calculated result, click Background, choose a color and click OK.
8. To display the cell contents in bold text, click Bold Text.
9. Choose OK.
The calculated result appears in the selected cell.
Feature Description
To define a calculation based on the calculations table
1. In the calculations table, click the cell where you want the
calculation result to appear.
2. In the Advanced Calculations Cell Properties box, choose
Calculation.
3. Select an equation in the Equation list box.
4. Set Source to Calculations Table.
5. Specify the table cells to include by entering their ID numbers in the
Range/Reference box.
If Equation is set to Std. Deviation, Mean, or %RSD, use a colon to
specify a range (e.g., A2:A5 or A1:AN) or commas for individual
cells (e.g., A1,A3,A5).
If Equation is set to Addition, Subtraction, Multiplication or
Division, specify one cell location in each Reference box. (e.g., A1).
If Equation is set to a Factor, use the first (Reference) box to specify
a cell location and the second (Factor) box to enter the factor.
6. To select the background color for the cell that displays the
calculated result, click Background, choose a color and click OK.
7. To display the cell contents in bold text, click Bold Text.
8. Choose OK.
The calculated result appears in the selected cell.
Related Topics
Adding Custom Calculations
Task Buttons
Action Buttons
Right Pane
The right pane in Home includes:
Feature Description
Select a user group to access its applications (see
“Application buttons” below). See Configuring the Home
Page for information on associating applications with
user groups.
Click a button (for example, Fixed or Scan) to open an
application. The available applications depend on the
selected user group (see “Group” above).
When the Home screen is not displayed, the right pane is used for application settings, sample
measurement results, available workbooks, and other features.
Related Topics
Data Display
Navigation Pane
Menus
1. Click Options.
Enter the desired group name in the Groups box and click Add.
After you click Add, the new group name appears in the Group box and the available
applications appear in the Applications box.
Drag the desired applications from the Applications box to the buttons at the right. To
remove an application from the group, drag it from its button to the Applications box.
Click Clear App Buttons and then click Yes when prompted.
Drag the List of Templates item from the Applications box to one of the top nine menu
buttons.
Related Topics
Applications Tab
Right Pane
Data Display
All data acquired with the current workbook are listed in the sample measurements table. To
view data, select them in the list. To select multiple spectra (or sample points), hold down
Shift or Ctrl. To display data from multiple samples as they are measured, select Overlay data
in the navigation pane.
The sample IDs appear at the top of the data display, with the selected spectrum (or sample
points) and sample ID in red. To turn off the sample IDs, right-click the data display, point to
Sample Legend and choose OFF.
For scan data, the X and Y values of the pointer location appear below the data display. To
enlarge a region, draw a box around it and click inside the box. Alternatively, click (if
available) to reveal the view finder which has additional display tools.
To change the appearance of the axes or data, right-click an axis and choose Display
Attributes. View the effects on the display while adjusting the attributes. To change the color
of the axes or their labels, click the appropriate color box. Your selections will affect the
appearance of data on the display and in printed reports.
The Sample ID readout at the right of the display shows the name of the sample being
measured (or to be measured). The results table below the Sample ID shows the results of
analysis operations (Peak Pick and Value Level Crossings for Scan data) and whether sample
points fall within specified control limits (for Fixed data).
Right-click the display to access these features (if available for the current data type):
Feature Description
Copy To Clipboard Copies the plotted data as it is currently displayed to the
Windows® Clipboard.
Autoscale Adjusts the vertical scale of each spectrum for optimal viewing.
Double-clicking the display has the same effect.
Set Scale For setting minimum and maximum values for the axes. To set
them automatically, select Autoscale.
Auto Arrange Labels Positions labels for easier reading.
Sample Legend Specifies whether to display sample IDs near the upper-left corner
of the pane.
Sample Labels Specifies whether to display spectra with annotations (see
“Annotate” below) and labels, such as those showing absorbance
values for specific wavelengths.
Line Type Specifies whether to display scan spectra using solid lines or
individual data points.
Feature Description
Annotate Provides features for adding annotation to displayed data. (Added
annotation is displayed only for the selected data.) See Working
With Labels for information about manipulating added labels.
Right-click an edge of an added ellipse, rectangle or image to access
ways to manipulate or delete the object. The availability of features
depends on the current application.
More:
Ellipse draws an ellipse. To move it, drag its top edge. To resize it,
drag a side or bottom edge. Its position, size and shape are affected
by zooming in and out.
Floating Image works the same as Image (see above), except the
image position and size are not affected by zooming.
Feature Description
3D Display (Scan Rate Shows the selected scan data in a three-dimensional surface plot
only) with these axes: absorbance, wavelength and time since the start of
the experiment (vertical axis).
You can rotate the image, zoom in or out, or drag the image within
the pane.
More:
Zoom. To enlarge or shrink the image, hold down the right mouse
button and drag vertically across the image. You can also enlarge
or shrink the image by clicking it and turning the mouse wheel (if
present).
Pan. Hold down the Shift key and drag the image in any direction
to move it within the pane.
When you click a point in the 3-D image, the spectral display pane
changes to show the spectrum collected at the clicked time value.
To display multiple spectra, hold down the Shift key while clicking
locations in the 3-D image.
To adjust the display limits of the 3-D image, draw a box around
an area of interest in the spectral pane and click inside the box.
To redisplay the full data range and adjust the vertical scale to fill
the vertical axis, right-click the spectral display pane and choose
Autoscale.
To remove the 3-D image from the window, right-click the spectral
display pane and choose 3D Display to clear the check mark or
choose View > 3D Display.
Feature Description
More Display Options Specifies how to display fixed sample data and control limits:
(Fixed only)
Connect points (in Spectrum) draws connecting lines between
sample points.
Related Topics
Working With Labels
Right Pane
Measuring Samples
Measurement Results
To move a label
To edit a label
Double-click it and use the dialog box that appears to change the text, its font or its color, or
right-click the label and choose Edit, Color or Font.
To rotate a label
Right-click it and choose Send to Front or Send to Back. Click here for information about
adding annotation.
To delete a label
Related Topics
Using the Palette Tools and View Finder
Data Display
Measuring Samples
If the current application provides a Measure button in the navigation pane, use it to access
data acquisition features for the application. The name of the application appears on the
button; for example:
Clicking the button displays action buttons for starting sample measurement or other data
acquisition. The provided buttons and their availability depend on the current application
and whether necessary operations have been performed. See Action Buttons.
See Setting Application Parameters for information about setting parameters that affect data
acquisition, including those for controlling sampling accessories. For information about
sampling accessories for the Evolution 200 Series instruments, refer to the user guide for the
accessory or search for “Evolution 200 Series Accessories” in the INSIGHT Help system.
Related Topics
Performing Fixed Measurements
Measurement Results
Below the data display in some applications is a table of sample measurement results. To
display a spectrum, click its row in the table.
To print a report containing results information for a sample, select it in the table
and click Print or choose File (menu) > Print Report. To preview the report,
choose File > Print Preview. To specify the information to include, use the Print
tab and Reportable Data tab in Reports.
To view, copy or print information about a spectrum, including its history, right-click it in the
table and choose Properties.
To measure a sample again, right-click its row in the sample measurements table and choose
Remeasure (if available). (If Remeasure is not available, you can choose Options >
Preferences and clear Prevent removal of data to enable it.) After the remeasurement, the
previous information for the sample is either crossed out or removed from the table,
depending on the status of Prevent removal of data on the Preferences tab in Options.
To delete a spectrum from the workbook, right-click it in the table and choose Remove (if
available). To delete all the spectra from the workbook, right-click the table and choose
Remove All (if available).
Any Peak Pick or Value Level results appear in the results table to the right of the data display.
Menus
The File menu, when available, includes:
Command Description
New Workbook Opens a new workbook and closes and saves the current workbook.
Merge Workbooks If available, allows you to copy data between workbooks created
with the same application.
Instructions:
3. Open a workbook that was created using the desired application
such as Fixed or Scan. See My Data.
4. Choose File > Merge Workbooks.
5. In Merge Workbooks, use the navigation (left) pane to locate
the workbook that contains the data you want to copy.
A list of workbooks created using the selected application
appears in the Workbooks pane. If the workbook contains
sample data, the sample names are listed in the Samples pane.
The data associated with the selected sample are displayed in the
Data pane.
6. Select the sample measurements to import by clicking their
names in the Samples pane. To select multiple samples, hold
down Shift or Ctrl.
7. Click OK.
8. The sample names and other information are appended to the
sample measurements table in the open workbook.
9. Choose Save Workbook or Save As Workbook to save the new
data to the current or another workbook.
The merged workbook is unchanged.
Save Workbook If available, saves the current workbook using its current file name
and location.
Save As Workbook If available, saves the current workbook with the specified file name
and location.
Close Workbook and Closes and saves the current workbook and returns to Home.
Go Home
Close All Workbooks Closes all open workbooks (all applications) and returns to Home.
and Go Home
Command Description
Import If available, imports an externally collected spectrum (.csv file or .tsv
file) into the current Scan workbook.
Print Preview Shows the current report before printing, with the header and footer
specified on the Report Master Page tab in Options.
Print Report Prints the current report to the default printer.
Use Current Settings Makes the current workbook settings and report configuration
as Default selections the default choices for a newly opened workbook for the
application.
Save Workbook If available, saves the current software settings as a template that can
Settings as Template be used later by the specified user group.
More
Command Description
3D Display (Scan Shows scanned rate data in a three-dimensional surface plot.
Rate only)
Command Description
Smooth Reduces noise in wavelength scan data from Scan and Rate experiments.
The smoothing algorithm uses the Savitzky-Golay method.
Derivative Converts scan data from Scan and Rate experiments to a derivative.
Convert Spectra Converts wavelength scan data to another Y-axis format.
Add Adds two scan spectra.
Normalize Adjusts the Y scale of scan data so that a selected data point has the
desired Y value.
Command Description
Subtract Subtracts one scan spectrum from another.
Ratio Divides one scan spectrum by another.
Average Averages two or more scan spectra.
Factor Adds a constant to scan data, subtracts a constant, multiplies by a
constant or divides by a constant.
Command Description
Peak Pick Finds peaks, valleys, or maximum and minimum values in a spectrum or
region.
Value Level Finds the wavelengths where a spectrum crosses a specified ordinate value.
Modify Rate Perform rate calculations or modify existing rate calculations on the data in a
Curve plot.
Command Description
Help Displays the INSIGHT Help system, which can also be accessed from many
software screens by pressing F1.
About Provides information about the software version and instrument type.
Related Topics
Navigation Pane
Right Pane
If the system is operating normally, a green check mark appears on the Instrument
Status icon at the bottom of the INSIGHT window.
A yellow exclamation mark appears briefly on the icon when the system is busy
performing an operation.
If a system problem occurs, a red X may appear on the icon. Click it to see
information about the problem.
Related Topics
Evolution 200 Series Accessories
Thermo Software IQ
Thermo Software IQ performs Installation Qualification (IQ) for the software. IQ verifies
that the correct software files were installed and can also be used to verify that these files have
not been changed, deleted, or overwritten since they were installed. (For more information,
see the Thermo Software IQ user guide available through the Help menu of the software.)
Contact Information
For help with questions or problems with the software, use the information at the beginning
of this document to contact us.
Operation
Contents
• Acquiring Data using Live Display
• Performing Fixed Measurements
• Performing Scan Measurements
• Performing a Quantitative Analysis
• Performing a Rate Measurement
• Adding Custom Calculations
• Using the Palette Tools and View Finder
• Setting Application Parameters
• Configuring a Report
• Exporting Data
• Managing Data
• Math and Analysis Operations
Feature Description
Selects the absorbance display mode.
Feature Description
Selects the single wavelength data acquisition mode for the next
measurement.
Use the right and left arrows in the left pane to select the wavelength. The
measurement results appear in the right pane.
Selects the full scan data acquisition mode for the next measurement.
Use the top set of right and left arrows in the left pane to select the
starting wavelength for the scan; use the bottom set of arrows to select the
ending wavelength.
Records the currently displayed measurement in the computer’s internal
memory. Data from subsequent measurements are appended. To save the
data, choose Print. The memory location is cleared automatically when
you exit Live Display (or choose Clear).
Pressing Store while in Fixed mode stores the measurement result along
with the wavelength used and the display mode settings (Abs or %T).
Starts sample data acquisition with Live Display. Available only after a
blank measurement.
Pressing Measure while in Fixed mode collects the data at the selected
wavelength using a bandwidth of 1 nm and an integration time of
0.5 sec. After you press Measure, the Measure button is disabled and the
software automatically takes a measurement every 2 seconds until you
press Stop or Scan or exit Live Display (or after 5 minutes of inactivity).
Feature Description
Stops data acquisition.
The selection tool is available in Scan mode only. Use it to zoom in the
spectrum. To zoom in, draw a box and click inside it. (Double-click
anywhere in the data display to zoom out.)
The spectral cursor tool is available in Scan mode only. Use it to view
the X and Y values of a point in a spectrum.
To view the X and Y values of a point, select the spectral cursor tool and
click the data display. Cross hairs appear. The X and Y coordinates of
their intersection with the spectrum appear below the data pane. To move
the cross hairs, drag across the pane or use the left and right arrow keys on
the keyboard.
Note Change Fixed workbook settings as desired before making sample measurements.
All Fixed measurements are saved in a workbook that includes the data acquisition
settings. See Setting Application Parameters for information about unlocking settings.
Related Topics
Fixed
Note If you opened an existing workbook in step 1 above, the settings will be
unavailable unless you unlock them. See Setting Application Parameters for
information about unlocking settings.
6. Click Measure.
More:
If Sample averaging on the Samples tab in Settings was set to Duplicate, “D” at the end of
a sample name indicates the second measurement to be made of the sample. If Sample
averaging was set to Triplicate, “D” and “T” at the end of sample names indicate the
second and third measurements to be made, respectively.
To enter previously saved samples information, use Load Samples. To save the samples
information, use Save Samples.
9. If only one sample will be measured, install it.
10. Click Continue.
11. Follow any instructions that appear, such as to install a specified sample.
The Data Display shows the acquired data (a fixed data point) for the sample selected in
the table. (Right-click the data to access commands for customizing the display, including
adding annotation.)
The table contains the columns of information specified on the Reportable Data tab in
Reports. Examples include sample identification, user name, and the results of replicate,
duplicate or triplicate measurements and their average values.
To measure a sample again, right-click its row in the sample measurements table and
choose Remeasure (if available). (If Remeasure is not available, you can choose Options
> Preferences and clear Prevent removal of data to enable it.) After the remeasurement,
the previous information for the sample is either crossed out or removed from the table,
depending on the status of Prevent removal of data on the Preferences tab in Options.
Related Topics
Fixed
Data Display
Related Topics
Scan
Note If you need to select a new workbook or template, click My Data, double-click
a workbook or template, and then click Settings.
5. Click Baseline.
6. Click Measure.
More:
To enter previously saved samples information, use Load Samples. To save the samples
information, use Save Samples.
9. If only one sample will be measured, install it.
10. Click Continue.
11. Follow any instructions that appear, such as to install a specified sample.
The Data Display shows the acquired data (a scan spectrum) for the sample selected in
the table. (Right-click the data to access commands for customizing the display, including
adding annotation.)
The table contains the columns of information specified on the Reportable Data tab in
Reports. Examples include sample identification, user name, date and time, and
integration time.
To measure a sample again, right-click its row in the sample measurements table and
choose Remeasure (if available). (If Remeasure is not available, you can choose Options
> Preferences and clear Prevent removal of data to enable it.) After the remeasurement,
the previous information for the sample is either crossed out or removed from the table,
depending on the status of Prevent removal of data on the Preferences tab in Options.
Related Topics
Scan
Data Display
Related Topics
Settings for Quant Applications
Contents
• Quantifying Samples Without Using Standards
• Quantifying Samples Using Standards
Related Topics
Performing a Quantitative Analysis
Data Display
Note Select Manually entered factor on the Type tab before using this procedure.
Note To start an analysis immediately after viewing or changing the template settings,
click the Measure action button instead and skip to step 3.
More:
If Sample averaging on the Samples tab in Settings was set to Duplicate, “D” at the end of
a sample name indicates the second measurement to be made of the sample. If Sample
averaging was set to Triplicate, “D” and “T” at the end of sample names indicate the
second and third measurements to be made, respectively.
To enter previously saved samples information, use Load Samples. To save the samples
information, use Save Samples.
7. If only one sample will be measured, install it.
8. Click Continue.
9. Follow any instructions that appear, such as to install a specified sample.
The Run Chart tab plots the concentration of the measured component versus sample
number. If Use control limits was selected on the Samples tab, horizontal limit lines
show whether the concentrations are within the specified limits:
The Data tab displays the acquired data (a fixed data point or scan spectrum) for the
sample selected in the table. (Right-click the data to access commands for customizing the
display, including adding annotation.) See Data Display for more information.
The table contains the columns of information specified on the Reportable Data tab in
Reports. Examples include sample identification, user name, and the results of replicate,
duplicate or triplicate measurements and their standard deviation.
To measure a sample again, right-click its row in the sample measurements table and
choose Remeasure (if available). (If Remeasure is not available, you can choose Options
> Preferences and clear Prevent removal of data to enable it.) After the remeasurement,
the previous information for the sample is either crossed out or removed from the table,
depending on the status of Prevent removal of data on the Preferences tab in Options.
Related Topics
Quantifying Samples Using Standards
Data Display
Note Select Measure single standard, Standard curve, Standard curve with two
wavelengths or Advanced standard curve on the Type tab before using this procedure.
Note To start an analysis immediately after viewing or changing the template settings,
start with the next step.
2. If standards need to be acquired, click Measure. If standards have already been acquired,
skip to step 7 to acquire sample data.
3. Follow the instructions that appear. The instructions that appear are dependant on the
current settings for the features on the Standards tab in Settings.
More:
If Standard averaging on the Standards tab in Settings was set to Duplicate, “D” at the
end of a standard name in the Confirm Standards List box indicates the second
measurement to be made of the standard. If Sample averaging was set to Triplicate, “D”
and “T” at the end of standard names indicate the second and third measurements to be
made, respectively.
To enter previously saved standards information, use Import Standards.
If Calculate from weight/volume was selected on the Standards tab, enter the weight and
volume for each standard in the table.
To remove a standard from the list, select it, right-click the table and choose Clear
Standard. To remove all the standards, choose Clear Table.
To save standards information for later use, use Export Standards.
5. Click Continue.
6. Follow the instructions that appear, installing the specified standards.
The Standard Curve(s) tab displays the resulting standard curve (or curves). Specify the
standards to use for the curve (or curves) by selecting Yes or No in the Use column in the
table. These selections can always be changed. After each change, the standard curve is
updated automatically.
More:
The table contains the columns of information specified on the Reportable Data tab in
Reports. Examples include sample identification, user name, and the results of replicate,
duplicate or triplicate measurements and their standard deviation.
To permanently remove an acquired standard from the analysis, right-click it and choose
Remove. Its information is crossed out but not removed from the table.
The Data tab displays the acquired data (a fixed data point or scan spectrum) for the
standard selected in the table. To access commands for customizing the display, including
adding annotation, right-click the data. See Data Display for more information.
Note After a calibration curve is used to measure a sample, the curve can no longer be
edited. However, you can open a new workbook based on the existing calibration data
and then edit the curve before you begin acquiring new sample data.
7. Click Measure.
8. Follow the instructions that appear. The instructions that appear are dependant on the
current settings for the features on the Samples tab in Settings.
9. When samples information appears, modify it if desired.
More:
If Sample averaging on the Samples tab in Settings was set to Duplicate, “D” at the end
of a sample name in the Confirm Sample List box indicates the second measurement to
be made of the sample. If Sample averaging was set to Triplicate, “D” and “T” at the end
of sample names indicate the second and third measurements to be made, respectively.
To enter previously saved samples information, use Load Samples. To save the samples
information, use Save Samples.
10. If only one sample will be measured, install it.
11. Click Continue.
12. Follow any instructions that appear, such as to install a specified sample.
The sample measurements table contains the columns of information specified on the
Reportable Data tab in Reports. If Sample averaging was set to Duplicate, “D” at the end
of a sample name indicates the second measurement. If Sample averaging was set to
Triplicate, “D” and “T” indicate the second and third measurements, respectively.
The Standard Curve(s) tab shows graphically the relationship between the standard curve,
measured spectral intensity and calculated concentration for the selected sample: A
horizontal line connects the sample spectral intensity value on the Y-axis to the standard
curve. A vertical line connects that point to the sample concentration value on the X-axis.
The information at the top shows the measured wavelength, the equation for the standard
curve and the calculated r2 value, which indicates how well the standard curve fits the
standard data points (1.0 is a perfect fit).
Note If there are two analysis wavelengths, the lines appear for each of the two
standard curves.
The Run Chart tab plots the concentration of the measured component versus sample
number. (If there are two analysis wavelengths, the concentration value is the average of
the values for those wavelengths.) If Use control limits was selected on the Samples tab,
horizontal limit lines show whether the concentrations are within the specified limits:
To measure a sample again, right-click its row in the sample measurements table and
choose Remeasure (if available). (If Remeasure is not available, you can choose Options
> Preferences and clear Prevent removal of data to enable it.) After the remeasurement,
the previous information for the sample is either crossed out or removed from the table,
depending on the status of Prevent removal of data on the Preferences tab in Options.
To copy the data on the Standard Curve(s) tab or Run Chart tab, right-click the plot and
choose Copy to Clipboard.
Related Topics
Quantifying Samples Without Using Standards
Data Display
Contents
• Performing a Fixed Rate Measurement
• Performing a Scanning Rate Measurement
Related Topics
Performing a Fixed Rate Measurement
Data Display
The Time Status icon automatically appears above the data display during rate measurements.
1. Make sure Single wavelength or Multiple wavelength is selected on the Type tab in
Settings in Rate.
2. Click Measure Rate.
5. Click Measure.
If necessary, the Zero action button can be used to remeasure the zero sample.
6. When samples information appears, modify it if desired.
To enter previously saved samples information, use Load Samples. To save the samples
information, use Save Samples.
Enter values in any special columns that were specified on the Samples tab in Settings.
These values may appear in the rate data report.
7. Click Continue.
To extend the collection time for the current sample measurement (without interrupting
the data collection), click Extend Time and enter the amount of time to extend the
measurement. If you are collecting data in multiple stages, only the current stage is
affected.
Scroll bars are provided if plots are out of view.
Right-click the Data tab to access features for changing the scale or adding annotation.
See Data Display for more information.
The sample measurements table contains the columns of information specified in the
Reported Columns pane of the Reportable Data tab in Reports.
9. To perform rate calculations or modify existing rate calculations on the data in a plot,
double-click it or select Analyze (menu) < Modify Rate Curve.
Note The features to set up rate calculations are available for time-based single and
multiple wavelength rate measurements only. To calculate rate curves for
temperature-based data (temperature and time are recorded), display the data with
time on the X-axis and then double-click the data display or select Analyze (menu)
< Modifying a Rate Curve.
If Calculate rate constants automatically at end of data acquisition is selected on the Rate
Calculations tab, after data collection completes, the rate calculations will appear in the
sample measurements table along with the other sample data. To perform rate calculations or
modify existing rate calculations on the data in a plot, double-click it or select Analyze (menu)
> Modify Rate Curve. A window displays the plot and a table containing the default rate
calculation values based on the Rate Calculations tab.
Note To calculate rate curves for temperature-based data (temperature and time are
recorded), display the data with time on the X-axis and then double-click the data
display or select Analyze (menu) < Modifying a Rate Curve.
These operations are available from the Modify Rate Curve window.
• If desired, add or change sets of rate calculation values for the displayed data. Each set
represents a curve that describes the rate behavior over the specified time period.
To delete a row from the table, right-click it and choose Delete Selected Row. To delete
all the information from the table, right-click it and choose Clear Table.
• If desired, use the vertical markers on the rate curve to adjust the data range used for rate
calculations.
• Click Update to generate kinetic rate models, each of which appears as a line across the
data.
The window closes, the lines appear across the specified plots, and the sample
measurements table is updated with information about the model lines.
Related Topics
Settings for Rate Applications
Data Display
1. Make sure Scan data acquisition is selected on the Type tab in Settings in Rate.
2. Click Measure Rate.
5. Click Measure.
If necessary, the Zero action button can be used to remeasure the zero sample.
6. When samples information appears, modify it if desired.
To enter previously saved samples information, use Load Samples. To save the samples
information, use Save Samples.
Enter values in any special columns that were specified on the Samples tab in Settings.
These values may appear in the rate data report.
7. Click Continue.
8. Follow the instructions that appear.
Each grid on the Data tab displays all the spectra acquired for a particular sample.
Related Topics
Settings for Rate Applications
Data Display
Related Topics
Use sample correction factor
Advanced Calculations
Arithmetic Operations
Comparison Operations
Logical Operations
The following equation functions produce a value based on one or more arguments you
supply.
Argument Description
<number>, <angle>, <length>, An expression that yields a number.
<value>, <power>, <start>, <len>
<list> Any expression that yields a list of numbers.
<text>, <sub> An expression that yields a text.
Arithmetic Functions
Ceiling(2.2) is 3.0
floor(<number>) Largest integer less than or equal to <number>.
Example:
Floor(2.7) is 2.0
random() Random fractional number between 0 and 1.
round(<number>) Rounds a double value to the nearest integer value.
sqrt(<number>) Square root of a number.
truncate(<number>) Truncated number. Everything to the right of the decimal point
is removed, leaving only the integer portion of the number. The
returned value is not rounded.
Logarithmic Functions
log(<number>)
pow(<value>, <power>) Value raised to the power. Example:
pow(2,3) is 8.0
Trigonometric Functions
Note For functions acos and asin, <number> is an expression that yields a number in the
range -1 to 1.
Note The formula for converting degrees to radians is radians = degrees * (pi/180).
Statistical Functions
Text Functions
Measurement Functions
Related Topics
Measurement Tab for Fixed
To zoom in, draw a box and click inside it. To move the spectrum, drag it up or down.
Related Topics
Using the Spectral Cursor Tool
Cross hairs appear. The X and Y coordinates of their intersection with the spectrum
appear below the palette.
To move the cross hairs, drag across the pane or use the left and right arrow keys on the
keyboard.
To remove the cross hairs from the display, select another tool such as the selection tool.
Related Topics
Using the Selection Tool
Related Topics
Using the Selection Tool
To measure a peak
The X values of the endpoints appear in the Baseline readout. The defined area is shaded,
and its measurement appears in the Corrected area readout. Here is an example showing
the region limits and baseline endpoints adjusted to measure the corrected area of a peak:
To annotate the spectrum with the measured area, right-click the spectrum, point to
Annotate and choose Tool Value. See Working With Labels.
To remove the vertical lines and baseline from the display, select another tool such as the
selection tool.
Related Topics
Using the Selection Tool
Drag the vertical lines in the data display or drag horizontally between them to move both at
once.
Related Topics
Using the Selection Tool
The currently displayed region is indicated by the region markers, the bold vertical lines.
To expand horizontally
To expand vertically
Click the left half of the Vertical Expand/Contract button. To contract vertically,
click the right half.
To expand or contract vertically while keeping the top or bottom of spectra in place
To expand vertically while keeping the bottom of the spectra in place, click the
symbol at the top of this button.
To expand vertically while keeping the top of the spectra in place, click the symbol
at the right.
To contract vertically while keeping the bottom of the spectra in place, click the
symbol at the bottom.
To contract vertically while keeping the top of the spectra in place, click the symbol
at the left.
Related Topics
Using the Selection Tool
The tabs in the right pane (and their features) depend on the application.
If appears on a tab, point to it and read the information to correct the problem.
Button Description
Displays data acquisition features for the application and starts an acquisition.
Resets the parameters on all the tabs to their default values. For Quant and Rate,
also steps through each tab in sequence.
Makes the parameters on all the tabs available, closes the current workbook, and
creates a new workbook.
Change Fixed workbook settings as desired before making sample measurements. All Fixed
measurements are saved in a workbook that includes the data acquisition settings.
Feature Description
Name (if present) User group name followed by the name of the template of settings
(entered when the template was saved using Save Workbook Settings
as Template in the File menu).
Feature Description
Description Optional description of the template.
(optional)
Calculate additional For selecting or defining formulas for additional data processing, with
results the results appearing in the analysis report. The formulas are written
in a form similar to that used in a command script language, with
constants, mathematical functions, etc. (See “Operations and
Functions” for definitions.) All functions are not case-sensitive.
Spaces are not allowed between a function name and “(”. After adding
formulas, edit their information in the formula table as desired. To
delete a table row, right-click it and choose Delete Row. Some
information in the table cannot be modified or deleted.
Instructions:
1. Click Select.
2. Select the desired formulas.
Hold down the Ctrl or Shift keys on the keyboard while making
multiple selections. The available formulas depend on the
Formulas & Units tab settings in Options.
3. Click OK.
To define a formula
1. Click Build.
2. To enter a formula in its entirety, set Equation type to User
defined and type the equation. To use a provided formula, select
it and enter values for the variables.
The available formulas depend on the Formulas & Units tab
settings in Options.
3. Click OK.
4. For selected provided formulas, enter a name and concentration
unit in the table.
The available formulas depend on the Formulas & Units tab
settings.
Related Topics
Setting Application Parameters
Feature Description
Data mode Y-axis format for acquired data. Available options include
Absorbance and % Transmittance.
Integration time How long the system acquires and averages data at each measured
wavelength. Increasing the integration time improves the
signal-to-noise ratio.
Bandwidth Available options include:
• Bandwidth. A larger spectral bandwidth increases the amount
of energy that passes through the sample. Increasing the
bandwidth can improve the signal-to-noise ratio; however, a
larger spectral bandwidth decreases the ability to resolve closely
spaced spectral features.
• Materials, Fiber and Micro. These AFBG (Application
Focused Beam Geometry) options match the optical
configuration to the application for optimized use of
associated accessories.
• Blocked. Blocks the beam completely.
Feature Description
Result mode Determines how results are calculated and displayed:
Normal displays the results in the units specified in the Data mode
setting (see above).
Feature Description
Use control limits If available and selected, specifies control limits for each
wavelength in the table. Enter the desired upper and lower limits
for each wavelength in the table, using the appropriate unit for the
selected data mode.
Related Topics
Setting Application Parameters
Related Topics
Setting Application Parameters
Feature Description
Number of samples Number of samples for the experiment.
Base name Beginning of the name for sample spectra, to which a sequential
number is appended. For example, the names of three sample
spectra using the base name “dye” would be dye1, dye2 and
dye3.
Sample averaging Whether and how to average concentration values from multiple
measurements of the same sample or from multiple samples. To
average multiple measurements of the same sample, select
Replicate and set Number of replicates to the total number of
measurements. To average values from two or three different
samples, select Duplicate or Triplicate, respectively.
Use sample correction If available and selected, specifies a multiplication factor for each
factor sample result. Can be used to correct for sample properties and
preparation steps such as a sample dilution that affects the
measured result. Enter the desired factor for each sample in the
table.
Load Samples For locating and selecting a .tsv (tab separated values) file or .csv
(comma separated values) file containing sample names and
descriptions, which are entered in the samples table on the
Samples tab.
Save Samples Saves the contents of the samples table in a .tsv (tab separated
values) file or .csv (comma separated values) file in a specified
location.
Samples table Lists samples by their names and descriptions, up to a maximum
of 1000. To change a sample name, click its cell and edit the text.
To enter or change text in a table cell, click it and type. To order
the list by sample name or description, click the appropriate
column heading.
If the table contains just one sample and data is acquired with an
accessory that automatically changes samples, sample names are
incremented.
Related Topics
Setting Application Parameters
Feature Description
Name (if present) User group name followed by the name of the template of settings
(entered when the template was saved using Save Workbook Settings
as Template in the File menu).
Description Optional description of the template.
(optional)
Feature Description
Calculate additional For selecting or defining formulas for additional data processing, with
results the results appearing in the analysis report. The formulas are written
in a form similar to that used in a command script language, with
constants, mathematical functions, etc. (See “Operations and
Functions” for definitions.) All functions are not case-sensitive.
Spaces are not allowed between a function name and “(”. After adding
formulas, edit their information in the formula table as desired. To
delete a table row, right-click it and choose Delete Row. Some
information in the table cannot be modified or deleted.
Instructions:
1. Click Select.
2. Select the desired formulas.
Hold down the Ctrl or Shift keys on the keyboard while making
multiple selections. The available formulas depend on the
Formulas & Units tab settings in Options.
3. Click OK.
To define a formula
1. Click Build.
2. To enter a formula in its entirety, set Equation type to User
defined and type the equation. To use a provided formula, select
it and enter values for the variables.
The available formulas depend on the Formulas & Units tab
settings in Options.
3. Click OK.
4. For selected provided formulas, enter a name and concentration
unit in the table.
The available formulas depend on the Formulas & Units tab
settings.
Feature Description
Baseline Correction For defining a baseline correction for the raw data. These options are
Type available:
• 100%T baseline. Standard baseline correction. The data may be
acquired using an empty holder or using solvent.
• 0%T or 0%R baseline. Corrects for any inherent variations in
the electronic zero line of the instrument.
More:
More:
The software multiplies the scan result at each data point by the
corresponding value in the table (explained below). To delete the
information from the table, click Clear Table.
Select Standard reference correction with 0%R or 0%T baseline to display additional
features:
Feature Description
Clear Table Deletes all entries from the table.
Load Table Fills the table with wavelength and intensity values from a .tsv (tab
separated values) file or .csv (comma separated values) file containing X
delimiter Y pairs.
Save Table Saves in a .tsv file, .csv file, or .xml (extensible markup language) file the
wavelength and intensity values in the table.
Related Topics
Setting Application Parameters
Feature Description
Data mode Y-axis format for acquired data. Available options include Absorbance
and % Transmittance.
Feature Description
Bandwidth A larger spectral bandwidth increases the amount of energy that
passes through the sample. Increasing the bandwidth can improve the
signal-to-noise ratio; however, a larger spectral bandwidth decreases
the ability to resolve closely spaced spectral features, especially in
scanning measurements.
Integration time How long the system acquires and averages data at each data interval.
Increasing the integration time improves the signal-to-noise ratio but
reduces the scan speed.
Data interval Difference in wavelength between two consecutive data points.
Scan speed Wavelength range covered per unit time. Varies inversely with
integration time. Increases as the data interval increases.
Estimated time Estimated duration of data acquisition. Increases as integration time
increases and as the data interval decreases. Varies inversely with scan
speed.
Related Topics
Setting Application Parameters
Related Topics
Setting Application Parameters
Feature Description
Number of samples Number of samples for the experiment.
Base name Beginning of the name for sample spectra, to which a sequential
number is appended. For example, the names of three sample
spectra using the base name “dye” would be dye1, dye2 and
dye3.
Load Samples For locating and selecting a .tsv (tab separated values) file or .csv
(comma separated values) file containing sample names and
descriptions, which are entered in the samples table on the
Samples tab.
Save Samples Saves the contents of the samples table in a .tsv (tab separated
values) file or .csv (comma separated values) file in a specified
location.
Samples table Lists samples by their names and descriptions, up to a maximum
of 1000. To change a sample name, click its cell and edit the text.
To enter or change text in a table cell, click it and type. To order
the list by sample name or description, click the appropriate
column heading.
If the table contains just one sample and data is acquired with an
accessory that automatically changes samples, sample names are
incremented.
Related Topics
Setting Application Parameters
Feature Description
Result Data analysis operation to perform automatically after sample measurement.
The parameters for a selected operation appear to the right. See Finding Peaks in
Scan Data or Finding Value Level Crossings in Scan Data for information on
setting them and how results are displayed.
Related Topics
Setting Application Parameters
Note Change Quant settings only before sample measurements. All Quant measurements
saved in a workbook must be made with the same settings. See Setting Application
Parameters for information about unlocking settings.
Related Topics
Formulas and Units Tab
Feature Description
Name (if present) User group name followed by the name of the template of settings
(entered when the template was saved using Save Workbook Settings as
Template in the File menu).
Description Optional description of the template.
(optional)
Quant Type Type of quantitative analysis (see the table below). An image
representing the selected type appears to the right. The required tabs also
appear (click the Reset action button to step through them).
Pathlength Distance the light travels through the sample.
Related Topics
Settings for Quant Applications
Feature Description
Analysis wavelength(s) Wavelength(s) to use for the quantitative analysis.
Correction Baseline correction to apply to raw data. Available options:
• Single point. Enter a wavelength value to define the
endpoint for the single point baseline.
This option generates a baseline correction for each sample
measurement by drawing a straight line through the
specified baseline point and calculating the absorbance
value on the drawn baseline at the specified analytical
wavelength. The software subtracts the calculated baseline
absorbance value from the measured absorbance value and
reports the corrected absorbance.
• Sloping baseline. Enter two wavelength values to define
the endpoints for the sloping baseline.
This option generates a baseline correction for each sample
measurement by drawing a line between the two specified
baseline points and calculating the absorbance value on the
drawn baseline at the specified analytical wavelength. The
software subtracts the calculated baseline absorbance value
from the measured absorbance value and reports the
corrected absorbance.
• None Uses uncorrected data.
Component name Component to quantify.
Unit Component concentration unit. The available units depend on
the Formulas & Units tab settings in Options.
Use user-defined factor Calculates component concentration using the entered factor
(K in the equation C = K *A), with the selected unit.
Use extinction coefficient Calculates component concentration with absorbance (A) equal
to
c
where is the entered extinction coefficient (a constant
dependent on the component and wavelength), c is
concentration in the selected unit, and is 1 cm (the
pathlength).
Feature Description
Molecular weight Molecular weight of the component. Available only if the units
specified on this tab are different and require this value for a
conversion.
Calculate additional For selecting or defining formulas for additional data
results processing, with the results appearing in the analysis report. The
formulas are written in a form similar to that used in a
command script language, with constants, mathematical
functions, etc. (See “Operations and Functions” for definitions.)
All functions are not case-sensitive. Spaces are not allowed
between a function name and “(”. After adding formulas, edit
their information in the formula table as desired. To delete a
table row, right-click it and choose Delete Row. Some
information in the table cannot be modified or deleted.
Instructions:
1. Click Select.
2. Select the desired formulas.
Hold down the Ctrl or Shift keys on the keyboard while
making multiple selections. The available formulas depend
on the Formulas & Units tab settings in Options.
3. Click OK.
To define a formula
1. Click Build.
2. To enter a formula in its entirety, set Equation type to User
defined and type the equation. To use a provided formula,
select it and enter values for the variables.
The available formulas depend on the Formulas & Units
tab settings in Options.
3. Click OK.
4. For selected provided formulas, enter a name and
concentration unit in the table.
The available formulas depend on the Formulas & Units
tab settings.
Feature Description
Analysis wavelength(s) Wavelength(s) to use for the quantitative analysis.
Correction See Correction above.
Component name Component to quantify.
Unit Component concentration unit. The available units depend on
the Formulas & Units tab settings in Options.
Standard concentration Concentration of the single standard.
Std averaging Whether and how to average concentration values from multiple
measurements of the same standard or from multiple standards
prepared under the same conditions. To average multiple
measurements of the same standard, select Replicate and set
Number of replicates to the total number of measurements. To
average values from two or three similar standards, select
Duplicate or Triplicate, respectively.
Calculate additional See Calculate additional results above.
results
Standard curve
Feature Description
Analysis wavelength(s) Wavelength(s) to use for the quantitative analysis.
Correction See Correction above.
Component name Component to quantify.
Unit Component concentration unit. The available units depend on the
Formulas & Units tab settings in Options.
Calculate additional See Calculate additional results above.
results
Feature Description
Analysis wavelengths Wavelengths to use for the quantitative analysis.
Correction See Correction above.
Component name Component to quantify.
Unit Component concentration unit. The available units depend on the
Formulas & Units tab settings.
Calculate additional See Calculate additional results above.
results
Feature Description
Standard Curve Equation Equation used to quantify samples, written in a form similar to
that used in a command script language, with constants,
mathematical functions, etc. (See Operations and Functions for
definitions of operations and functions.) All functions are not
case-sensitive. Spaces are not allowed between a function name
and “(”.
For this Quant type the equation defines the value for the Y-axis
of the calibration curve.
Instructions:
1. Click Select.
2. Select the desired equation and click OK.
The available equations depend on the Formulas & Units
tab settings in Options.
To define an equation
1. Click Build.
2. To enter an equation in its entirety, set Equation type to
User defined and type the equation. To use a provided
equation, select it and enter values for the variables.
The available equations depend on the Formulas & Units
tab settings.
3. Click OK.
Component name Component to quantify.
Unit Component concentration unit. The available units depend on
the Formulas & Units tab settings in Options.
Calculate additional See Calculate additional results above.
results
Related Topics
Operations and Functions
Feature Description
Mode Specifies Scan or Fixed data acquisition.
Data mode Y-axis format for acquired data. Available options include
Absorbance and % Transmittance.
Related Topics
Settings for Quant Applications
Related Topics
Settings for Quant Applications
Feature Description
Curve fit type Type of equation used to create the standard curve from standard
concentration values.
Standard averaging Whether and how to average concentration values from multiple
measurements of the same standard or from multiple standards
prepared under the same conditions. To average multiple
measurements of the same standard, select Replicate and set Number
of replicates to the total number of measurements. To average values
from two or three similar standards, select Duplicate or Triplicate,
respectively.
Feature Description
2
Minimum r The r2 value indicates how well the standard curve fits the standard
data points, with 1.0 a perfect fit. If Minimum r2 is selected, samples
can be quantified only after that value is achieved for the standard
curve.
Calculate from If available for the current unit, calculates concentration when the
weight/volume weight and volume of material used to prepare each standard is
known. Enter the weight and volume for each standard in the
appropriate table cells, or enter it later as part of running an analysis.
Use correction If available and selected, specifies a multiplication factor for each
factor standard. Can be used to correct for standard properties (percent
purity, water content, etc.) or preparation steps that affect the
measured result such as a dilution.
When this option is selected, two columns are added to the standards
table: Correction Factor and Corrected Concentration. In the
Correction Factor column, enter the desired factor for each standard
in the table. The values in the Corrected Concentration column are
used to create the calibration curve. If a report is generated that
includes information about the standards, the report will also include
the correction factors and corrected concentrations.
Standards Table Contains concentration and other information about the standards.
To add information from a .csv (comma separated values) file or .tsv
(tab separated values) file, click Import Standards. To save the
information in a .csv file or .tsv file, click Export Standards.
Related Topics
Settings for Quant Applications
Feature Description
Number of samples Number of samples in the analysis.
Base name Beginning of the name for sample spectra, to which a sequential
number is appended. For example, the names of three sample spectra
using the base name “dye” would be dye1, dye2 and dye3.
Sample averaging Whether and how to average concentration values from multiple
measurements of the same sample or from multiple samples. To
average multiple measurements of the same sample, select Replicate
and set Number of replicates to the total number of measurements.
To average values from two or three samples, select Duplicate or
Triplicate, respectively.
Use sample If available and selected, specifies a multiplication factor for each
correction factor sample result. Can be used to correct for sample properties and
preparation steps such as a sample dilution that affects the measured
result. Enter the desired factor for each sample in the table.
Use weight/volume If available for the current unit, corrects sample concentrations using
correction the entered target weight and volume:
If the table contains just one sample and data is acquired with an
accessory that automatically changes samples, sample names are
incremented.
Related Topics
Settings for Quant Applications
Related Topics
Rate
Feature Description
Name (if User group name followed by the name of the template of settings (entered
present) when the template was saved using Save Workbook Settings as Template in
the File menu).
Description Description of the template.
Rate Type Type of rate measurement to perform (see the table below). An image
representing the selected type appears to the right. The required tabs also
appear (click the Reset action button to step through them).
Mode These options are available:
Related Topics
Settings for Rate Applications
Feature Description
Integration time How long the system acquires and averages data to obtain each data
point at each measured wavelength. Available for single and multicell
experiments only. (To set integration time for scan rate experiments,
see Instrument Tab for Rate.)
Dwell time How long the sample remains in the light beam. Available for single
and multicell experiments only.
Use this setting to vary the number of data points collected per cell
per cycle. Dwell time is always a multiple of integration time (see
above). Dwell time can be very useful when performing multicell
rate measurements. This is because the cell changer accessory will
position a cell in the light beam for a specified period of time (dwell
time), rather than moving back and forth between cells and
increasing the time between measurements (when no sample is in
the light beam).
For example, if the integration time is 0.1 second and dwell time is
0.5 second, five data points will be recorded for each cell before the
cell changer moves to the next cell. For fairly fast reactions, multiple
samples can be measured, with over ten times as many data points
acquired within a short time. This can improve the accuracy of rate
calculations for faster multicell rate measurements.
Number of stages Defines the number of measurement stages over which the frequency
of data acquisition can be determined. For example, if a reaction has
two components, the first component could be fast and require a
high data density. The second component could be much slower and
longer lived, therefore requiring a lower data density. Defining
multiple stages enables the use of the proper data density over a
given period of time.
Feature Description
Stages table Available for single and multiple wavelength experiments only.
Defines for each stage the start time, end time and interval (how
frequently a measurement is made). For example, using an interval
of 10 seconds would measure the sample every 10 seconds from the
start time to the end time. In a multicell experiment the interval
setting is the measurement interval for each sample.
Defines for each stage the end temperature, ramp rate, hold type,
hold time, interval and whether data collection will occur.
Hold Time. When Hold Type is set to Time, enter the length of
time to hold at the target temperature before continuing to the next
stage.
Related Topics
Settings for Rate Applications
Feature Description
Data mode Y-axis format for acquired data. Available options include Absorbance
and % Transmittance.
Feature Description
Estimated time Estimated duration of data acquisition. Increases as integration time
(Scan only) increases and as the data interval decreases. Varies inversely with scan
speed. Available for scan experiments only.
Reference When selected, automatically acquires a reference measurement at a
wavelength specified wavelength each time a sample measurement is taken. The
correction reference measurement is then subtracted from the corresponding
sample measurement to produce a corrected measurement result.
Only the corrected results are reported.
Related Topics
Settings for Rate Applications
Related Topics
Instrument Status Monitors
Feature Description
Number of samples Number of samples for the experiment.
Base name Beginning of the name for sample spectra, to which a sequential
number is appended. For example, the names of three sample spectra
using the base name “dye” would be dye1, dye2 and dye3.
Set up sample table Specifies the columns of sample information to be entered during the
columns experiment. To add a column, type the desired name in an empty cell
and press the Tab key. To delete a column (except the Sample Title
column), right-click it and choose Delete Column. To delete all the
columns (except the Sample Title column), right-click the table and
choose Delete All Columns.
Load Samples For locating and selecting a .tsv (tab separated values) file or .csv
(comma separated values) file containing sample names and
descriptions, which are entered in the samples table on the Samples
tab.
Save Samples Saves the contents of the samples table in a .tsv (tab separated values)
file or .csv (comma separated values) file in a specified location.
Samples table Lists samples by their names and descriptions, up to a maximum of
1000. Additional columns are defined by the Set up sample table
columns option (see above).
To change a sample name, click its cell and edit the text. To enter or
change text in a table cell, click it and type. To order the list by
sample name or description, click the appropriate column heading.
If the table contains just one sample and data is acquired with an
accessory that automatically changes samples, sample names are
incremented.
Related Topics
Settings for Rate Applications
Feature Description
Calculate rate constants Calculates rate constants automatically at the end of the
automatically at end of experiment. When this option is selected, the rate curve appears in
data acquisition the data display and the rate settings and calculations appear in
the sample measurements table after the measurement is
completed. To modify the rate curve, double-click the data
display or choose Analyze (menu) > Modify Rate Curve.
Related Topics
Settings for Rate Applications
Configuring a Report
A report contains a table of sample data and other specified information. Reports can be saved
and printed or the sample data can be exported.
Note See the Report Master Page tab in Options to define global settings for all reports
including headers and footers and a logo or other image.
To configure a report
The sample data that were measured and archived in the opened workbook are displayed
on the Samples tab.
3. Select the sample results to include by clicking a row in the table.
Use “Shift + click” or “Ctrl + Click” to select additional rows.
4. To specify the columns to include and their order, click the Reportable Data tab.
5. To specify the items to include in printed reports for this workbook, click the Layout tab.
Note To specify that all items in printed reports span the full width of each page,
select Fit To Width in the navigation pane. See Layout tab for more information
about formatting options for printed report.
6. To preview all the pages in the report, click the Preview action button.
Related Topics
Report Master Page Tab
Reports
Exporting Data
You can export acquired data and associated measurement results from any opened report or
configure INSIGHT to export all data and results automatically.
Contents
• Exporting Data from a Report
• Exporting Data Automatically
Related Topics
Exporting Data
Reports
Related Topics
Exporting Data
Reports
Managing Data
Contents
• My Data
• Reports
My Data
Tab Description
Workbooks Select a folder in the tree to list its workbooks. Select a listed workbook to see
its spectra. To zoom in on the displayed spectra, draw a box and click inside it
or use the view finder. To view the full data range, double-click the spectral
display.
Templates Select a group folder in the tree to list its templates.
PV Reports Select the listed result reports of interest. To select multiple reports, hold
down Shift or Ctrl. (One window will display all the opened results.)
Use the links below the buttons to display archived workbooks. Click Active workbook
library to view the directory where a workbook was last saved. To change the path to the
default workbook, right-click My favorite workbook library and choose Edit Link.
Related Topics
Managing Data
Reports
A report contains a table of sample data and other specified information.
Button Description
Shows the current report before printing, with the header and footer specified
on the Report Master Page tab in Options.
Saves the specified spectral data or sample measurement results in the specified
format (see the table below).
Format Description
Report, Excel XML Can be opened in Excel®. Only the columns selected on the
Spreadsheet (*.xml) Reportable Data tab are saved and only for the rows selected on
the Samples tab. Configure those tabs to include the desired
information before exporting.
For Rate applications, the data is displayed with one column per
spectrum, and one sample per sheet.
Report, Tab Separated Can be opened in Notepad or Excel. Only the columns selected
Values (*.tsv) on the Reportable Data tab are saved and only for the rows
selected on the Samples tab. Configure those tabs to include the
desired information before exporting.
Report, Comma Can be opened in Notepad or Excel. Only the columns selected
Separated Values (*.csv) on the Reportable Data tab are saved and only for the rows
selected on the Samples tab. Configure those tabs to include the
desired information before exporting.
Spectrum, Excel XML Can be opened in Excel. Saves absorbance values along with the
Spreadsheet (*.xml) corresponding wavelengths only for the rows selected on the
Samples tab. If multiple sample rows are selected, the
corresponding absorbance values for each wavelength are saved
onto separate worksheets within one Excel file.
Spectrum, Tab Separated Saves absorbance values along with the corresponding
Values (*.tsv) wavelengths for the rows selected on the Samples tab in a format
that can be opened in Notepad or Excel. If multiple sample rows
are selected, the corresponding absorbance values for each
wavelength are saved sequentially in one column.
Format Description
Spectrum, Comma Saves absorbance values along with the corresponding
Separated Values (*.csv) wavelengths for the rows selected on the Samples tab in a format
that can be opened in Notepad or Excel. If multiple sample rows
are selected, the corresponding absorbance values for each
wavelength are saved sequentially in one column.
Spectra, New Workbook Saves the settings and information about any samples, standards
(*iwbk) and advanced calculations for the selected application to a new
INSIGHT workbook.
Spectrum 3D, Excel Saves absorbance values along with the corresponding time and
XML Spreadsheet (*.xml) wavelengths for the rows selected on the samples tab in a format
that can be opened in Excel. Only available in Scan Rate
applications.
If the computer is not mapped to recognize that .xml files should be opened with Excel, open
the exported file from within Excel.
Tab Description
Samples Displays the sample data archived in the opened workbook. To sort the
information in the table, click the desired column heading.
To view particular data, select it in the table. For most applications this
displays a sample spectrum. For fixed Rate data, this displays spectral data
points for the indicated wavelength and the line resulting from the rate
calculations, if specified. For scan Rate data, all of the sample’s spectra
appear.
To view, print or copy data acquisition and other information about data,
right-click it in the list and choose Properties.
To change a sample ID, click its cell and type a new ID. (If sample IDs are
not editable, choose Options > Preferences and clear Prevent removal of
data.)
Reportable Specifies the report columns to include and their order. The available
Data columns depend on the current application and workbook settings.
Feature Description
Header/Footer Includes a header and footer in the printout.
Header Items Title and subtitle to appear in the header. Use the Font buttons to specify
their fonts.
Feature Description
Settings info Includes a table of workbook settings in printed reports including the
name, serial number and settings for any accessories that were used. Use the
Font button to specify the font for the information.
Samples table Includes a table of sample measurement information for the rows selected
on the Reports > Samples tab in printed reports. Use the Font button to
specify the font for the information.
Results/ This option adds the following if available to printed reports:
Calculations
• Results table, which appears to the right of the data display below the
table
Sample ID readout. The results table shows the results of analysis
operations and whether sample points fall within specified control
limits.
• Calculations table, which appears below the sample measurements
table when Advanced Calculations is selected in the navigation pane
(available for all applications except Rate and DNA Melting).
Use the Font button to specify the font for the information.
Standards For Quant applications that include standards, includes a table and graph
of information for the standards in printed reports. Use the Font button to
specify the font for the information.
X-Y pairs For Rate applications, includes a table of x-y data for each sample row
selected on the Samples tab in printed reports. For each sample, one X-Y
pair is logged for each wavelength, sampling interval and stage. Use the
Font button to specify the font for the information.
Graphs Includes sample spectra and other data plots in printed reports. Data can
be placed in individual graphs or overlaid on one graph.
• Select Overlay to print data plots overlaid. (Note that this is the only
setting that allows tables and graphs to be printed on the same page.)
• Select Separate to print each data plot in a separate graph, with the
specified number of graphs on a page.
Report notes Includes a Report Notes box in printed reports. To add text to the Report
Notes box, click Edit. Use the Font button to specify the font for the
information.
Page Setup Specifies the paper size, orientation and other printing attributes.
Fit to page Changes the layout of all included items so they fill the page width.
width (Column headings may be truncated.)
Note We recommend first defining a master page on the Report Master Page tab in
Options.
Related Topics
Configuring a Report
Managing Data
The smoothing algorithm uses the Savitzky-Golay method. For more information, see About
Smoothing.
To smooth data
c. Click OK. Each added spectrum is listed with asterisks at the beginning of its name
to indicate the data is not in the current workbook.
2. Select any other listed spectra to smooth and click Add.
To remove a spectrum from the Operands pane, click it and click Remove. To remove all
the spectra, click Clear All.
3. Select the polynomial order.
4. Specify the degree of smoothing (number of points in the smoothing formula) by
dragging the slider:
5. To perform a math operation on the smoothed result, click Operate on Result and
choose an operation. Otherwise, go to the next step.
Note To use the result in another Smooth operation, use Workbook or Temporary
to add it to the list (see step 6), click Add to make it an operand, and continue with
step 2.
Related Topics
About Smoothing
About Smoothing
The smoothing feature of INSIGHT software uses a Savitzky-Golay algorithm modified to
reduce high frequency breakthrough; that is, to reduce noise in the spectrum. The algorithm
is based on performing a least-squares linear regression fit of a polynomial of order k over at
least k+1 data points around each point in the spectrum.
6. To perform a math operation on the converted result, click Operate on Result and
choose an operation. Otherwise, go to the next step.
Related Topics
Math and Analysis Operations
Note To use the result in another Convert Spectra operation, use Workbook or
Temporary to add it to the list (see step 5), click Add to make it an operand, and
continue with step 2.
Related Topics
Math and Analysis Operations
Type a value in the Factor text box or use its up and down arrow buttons. Use Increment
to set the amount of change these buttons make to the factor when clicked. For example,
with Factor set to 1.000 and Increment set to 0.500, clicking the Factor up arrow button
changes the factor to 1.500.
4. To perform a math operation on the addition result, click Operate on Result and choose
an operation. Otherwise, go to the next step.
Note To use the result in another Add operation, use Workbook or Temporary to
add it to the list (see step 5), click Add in the Operand 1 box, and continue with step
2.
Related Topics
Math and Analysis Operations
c. Click OK. Each added spectrum is listed with asterisks at the beginning of its name
to indicate the data is not in the current workbook.
2. Select any other listed spectra to normalize and click Add.
To remove a spectrum from the Operands pane, click it and click Remove. To remove all
the spectra, click Clear All.
3. Select the desired Y value for the data point specified in the next step.
4. Select the wavelength of the data point.
5. To perform a math operation on the normalized result, click Operate on Result and
choose an operation. Otherwise, go to the next step.
Related Topics
Math and Analysis Operations
a. Click Browse.
b. Select a folder in the left pane (use the features above it to navigate), a workbook in
the middle pane, and spectra in the upper-right pane. To select multiple spectra, hold
down Shift or Ctrl. Selected spectra appear in the lower-right pane.
c. Click OK. Each added spectrum is listed with asterisks at the beginning of its name
to indicate the data is not in the current workbook.
2. Select the spectrum to subtract and click Add in the Operand 2 box.
3. To multiply the second spectrum by a factor other than 1, drag the slider or use Factor.
Type a value in the Factor text box or use its up and down arrow buttons. Use Increment
to set the amount of change these buttons make to the factor when clicked. For example,
with Factor set to 1.000 and Increment set to 0.500, clicking the Factor up arrow button
changes the factor to 1.500.
4. To perform a math operation on the subtraction result, click Operate on Result and
choose an operation. Otherwise, go to the next step.
Note To use the result in another Subtract operation, use Workbook or Temporary
to add it to the list (see step 5), click Add in the Operand 1 box, and continue with
step 2.
Related Topics
Math and Analysis Operations
Note To use the result in another Ratio operation, use Workbook or Temporary to
add it to the list (see step 5), click Add in the Operand 1 box, and continue with step
2.
Related Topics
Math and Analysis Operations
To use Average
Note A spectrum appears in the Result pane only if two or more spectra are in the
Operands pane.
3. To perform a math operation on the result, click Operate on Result and choose an
operation. Otherwise, go to the next step.
Note To use the result in another Average operation, use Workbook or Temporary
to add it to the list (see step 4), click Add to make it an operand, and continue with
step 2.
Related Topics
Math and Analysis Operations
To use Factor
4. Specify the factor: type a value, click the up and down arrow buttons, or drag the slider.
5. To perform a math operation on the result, click Operate on Result and choose an
operation. Otherwise, go to the next step.
Note To use the result in another Factor operation, use Workbook or Temporary to
add it to the list (see step 6), click Add to make it an operand, and continue with step
2.
Related Topics
Math and Analysis Operations
Note Peaks can be found automatically after data acquisition. See Measurement Tab for
Scan.
To locate peaks
Parameter Description
Find Specifies whether to find peaks, valleys, both peaks and
valleys, or maximum and minimum Y values.
Max number of peaks (or Maximum number of items to find.
valleys, or peaks and valleys)
Parameter Description
Sort peaks (or valleys, or Order for labeling and listing found items in the Peak
peaks and valleys) by Pick results table.
Sensitivity How readily small peaks or valleys are found. To
optimize the sensitivity, select Auto. To set it manually,
deselect Auto and drag the slider.
Wavelength Range Wavelength range in which to find items. To use the
entire range, select Full. To limit the range, deselect Full
and type the desired limits or use the region-threshold
tool:
Use threshold Specifies whether to find only those peaks whose Y values
are greater than or equal to a specified value, the
threshold. (For % transmittance, the software finds
valleys less than or equal to the threshold.) To optimize
the threshold value, select Auto. To set it manually,
deselect Auto and set Threshold by typing a value or
dragging the horizontal line in the data display.
Smooth spectrum Specifies whether or not to smooth the data before
finding peaks or valleys. This tends to reduce the number
of small peaks, valleys and shoulders. (For more
information, see About Smoothing.) The (blue)
smoothed spectrum is overlaid on the original. To
optimize the degree of smoothing, select Auto. To adjust
it manually, deselect Auto and drag the slider.
To adjust the data display, use the view finder. To zoom in or move the spectrum, use the
selection tool:
The results appear in the data display pane and the table below it:
Related Topics
Math and Analysis Operations
Note Crossings can be found automatically after data acquisition. See Measurement Tab
for Scan.
Parameter Description
Level for crossing Specifies the ordinate (Y) value. One can also drag the horizontal
line in the data display.
Sort crossings by Order for listing crossings in the value level crossings results table.
Parameter Description
Wavelength Range Wavelength range in which to find crossings. To use the entire
range, select Full. To limit the range, deselect Full and type the
desired limits or use the region-threshold tool:
To adjust the data display, use the view finder. To zoom in or move the spectrum, use the
selection tool:
The results appear in the data display pane and the table below it:
3. Click OK.
The found crossings have labels, and the crossing results are listed at the right.
The settings are saved in the workbook and will be used for measuring samples with
Result on the Peak Pick tab set to Value Level.
Related Topics
Math and Analysis Operations
Options
Applications Tab
Report Master Page Tab
Preferences Tab
Data Store Tab
Formulas and Units Tab
Applications Tab
Click Options to display the Applications tab in the
right pane. Use it to configure the home page by adding
new user groups and giving them access to particular
applications. The Group drop-down list box shows the
name of the current group.
Enter the desired group name in the Groups box and click Add.
Drag the desired applications from the Applications list to the buttons at the right. To
remove an application from the group, drag it from its button to the Applications list.
Click Clear App Buttons and then click Yes when prompted.
Drag the List of Templates application to one of the top nine menu buttons.
Feature Description
Company name Enter a company name and press Enter. Select an orientation for the
name, and use Font to specify the font.
Date Date of the report creation. Use Font to specify the font for Date and
Time.
Time Time of the report creation. Use Font to specify the font for Time and
Date.
Logo Select an orientation for the logo, and use Browse to locate and select a
logo image file.
Footer text Text at the bottom of the report. Use Font to specify the font for the
footer.
Page number Format for page numbers. Select None to not include page numbers.
Related Topics
Configuring a Report
Reports
Preferences Tab
Click Options to display the Preferences tab in the right
pane.
Feature Description
Data Value Digits Specify the number of digits to the right of the decimal point to use
for X and Y values for spectral data.
Triggering Starting data acquisition normally requires a response to a prompt
(displayed by Measure or another action button). To start acquisition
automatically a specified number of minutes after the prompt
appears, select Enable auto input trigger and enter the number of
minutes for the delay.
Feature Description
Prevent removal of Prevents the removal of acquired data. When this option is selected,
data the following conditions are true:
• Remeasure right-click option in the sample measurements table
is enabled (if available). After a remeasurement, the previous
information for the sample is crossed out.
• Remove and Remove all (available for Scan applications only)
are disabled.
• Sample IDs cannot be edited in the sample measurements table
below the data display and on the Samples tab in Reports.
Directories The paths used for saving template files, Performance Verification
reports, and workbooks. Type a path for each or click the button to
specify a path.
Feature Description
Workbook Save Options To be prompted to save data before acquiring it, select Data
saved on each measure. Save Workbook and Save As Workbook
will not be available in the File menu.
In the appropriate table, select in the Show column the items to be available.
To add an item, enter it (or the appropriate information) in the row that starts with a plus (+)
sign.
To delete an item, right-click it and choose Delete Row. Items with a lock icon cannot be
edited or deleted.
Feature Description
Default formulas These formulas appear when you click Select on the
Measurement tab in Quant Settings.
Default units These units appear in the Unit list box on the Measurement tab
in Quant Settings.
Advanced user-defined These formulas appear when you click Build on the
calibration equations Measurement tab in Quant Settings.
Related Topics
Settings for Quant Applications
System Settings
These are available in System Settings:
Alignment Tab
Calibrations Tab
Lamp Tab
System Tab
Diagnostics tab (for our use only)
Alignment Tab
Click System Settings to display the Alignment tab in
the right pane. Use it to view sample and reference
detector intensities while aligning an accessory.
To perform an alignment
1. To measure intensities using a green light (510 nm), select Green. To use no filter, select
White.
The correct setting depends on the installed accessory. For more information, refer to the
user guide for the accessory or search for “Evolution 200 Series Accessories” in the
INSIGHT Help system.
2. Set Bandwidth as desired.
The available settings depend on the instrument.
3. Click Start.
Every half second the sample and reference detector intensities are plotted and displayed
by the Sample beam and Reference readouts.
Note If the instrument has variable slit wheels, the start of intensity measurement is
delayed while they move to their 2 nm position.
The Time readout shows the elapsed time. Intensities can be measured and plotted for up
to 15 minutes without restarting the process.
If the plot of sample detector intensity is not at the desired position for viewing, adjust
the detector gain by dragging the Gain slider up or down.
To adjust the display, use the view finder. To zoom in, draw a box and click inside it. To
return to full display, double-click the pane.
4. Align the accessory to achieve the desired sample detector intensity. For more
information, refer to the user guide for the accessory or search for “Evolution 200 Series
Accessories” in the INSIGHT Help system.
5. When finished, click Stop.
Related Topics
System Settings
Calibrations Tab
Click System Settings to display the Calibrations tab in
the right pane. Use it to calibrate the instrument
wavelength or accessories.
To calibrate using a particular lamp, click its Calibrate button and follow the instructions that
appear. See Mercury Lamp for instructions for using a mercury lamp safely.
Related Topics
System Settings
Lamp Tab
Click System Settings to display the Lamp tab in the
right pane. Use it to display status information about
the xenon lamp.
Related Topics
System Settings
System Tab
Click System Settings to display the System tab in the
right pane. Use it to switch the instrument between
local and computer control and to update firmware.
Feature Description
Instrument Control Switches instrument control between the Local Control option and
INSIGHT software running on an external computer. Select the
desired setting and respond appropriately to the message that appears.
Firmware Update Click Load Update to install the selected instrument firmware.
Follow the instructions that appear.
Instrument Boot Click Display to view the instrument’s power-up boot record, which
Record can be saved or copied.
Related Topics
System Settings
Keyboard Shortcuts
These are available in INSIGHT software:
Contents
• About the Bio Applications
• Nucleic Acid
• Nucleic Acid Labels
• DNA Melting
• Protein A280
• Proteins and Labels
• Pierce BCA
• Protein Bradford
• Pierce Modified Lowry
• Pierce 660 nm Protein Assay
• Protein Biuret
Below are the default values for the nucleic acid methods. Wavelengths are in nanometers.
Protein Assays
There are two types of protein assay methods included, direct UV and colorimetric assays.
The colorimetric assays include several assays that are similar. Likewise, the direct protein
methods are also very similar. Use the table below to find the appropriate assay procedure.
Click the links to access complete information about the assays.
Below are the default values for the protein methods. Wavelengths are in nanometers.
Nucleic Acid
Contents
• Overview
• Nucleic Acid Concentration Calculations
• Application Settings
• Unique Screen Features
• Making Nucleic Acid Measurements
• Oligo Calculator
Overview
Use the Nucleic Acid application to measure nucleic acid
samples for concentration and purity.
See Unique Screen Features for information about the results the application calculates. See
Nucleic Acid Concentration Calculations for information about the calculations.
Related Topics
Nucleic Acid Concentration Calculations
Application Settings
Data Display
Oligo Calculator
c = (A * )/
where c is the nucleic acid concentration in ng/microliter, A is the absorbance in AU, is the
wavelength-dependent extinction coefficient in ng-cm/microliter, and is the pathlength in
centimeters.
Coefficient Value
Double-stranded DNA 50 ng-cm/?l
Single-stranded DNA 33 ng-cm/?l
RNA 40 ng-cm/?l
Related Topics
Overview
Application Settings
Data Display
Oligo Calculator
The 260/230 ratio assay determines the purity of nucleic acids in the presence of phenol,
which absorbs strongly at 230 nm.
A ratio of approximately 1.8 is generally accepted as “pure” for DNA, approximately 2.0 for
RNA. If either ratio is appreciably lower, protein, phenol or other contaminants that absorb
strongly near 280 nm may be present.
The 260/230 values for a “pure” nucleic acid are often higher than the respective 260/280
values and are commonly in the range 1.8 to 2.2. If the ratio is appreciably lower, co-purified
contaminants may be present.
Application Settings
To set data acquisition parameters for a Nucleic Acid
workbook, click Settings.
Related Topics
Overview
Data Display
Oligo Calculator
Feature Description
Name Application name.
Description (optional) Description of the template.
Nucleic Acid Type Type of analysis to perform.
Related Topics
Measurement Tab for Nucleic Acid
Feature Description
Analysis wavelength Wavelength(s) to use for the analysis.
Correction Specifies bichromatic normalization of the absorbance data.
• Single point. Enter a wavelength value to define the endpoint for
the single point baseline.
This option generates a baseline correction for each sample
measurement by drawing a straight line through the specified
baseline point and calculating the absorbance value on the drawn
baseline at the specified analytical wavelength. The software
subtracts the calculated baseline absorbance value from the
measured absorbance value and reports the corrected absorbance
• Sloping baseline. Enter two wavelength values to define the
endpoints for the sloping baseline.
This option generates a baseline correction for each sample
measurement by drawing a line between the two specified
baseline points and calculating the absorbance value on the
drawn baseline at the specified analytical wavelength. The
software subtracts the calculated baseline absorbance value from
the measured absorbance value and reports the corrected
absorbance.
• None. Uses uncorrected data. Without baseline correction,
spectra may be offset from the baseline. If this offset is significant,
the calculated protein concentration may be higher than the true
value.
Component name Component to quantify.
Feature Description
Unit Component concentration unit. The available units depend on the
Formulas & Units tab settings in Options.
Calculate additional For selecting or defining formulas for additional data processing, with
results the results appearing in the analysis report. The formulas are written
in a form similar to that used in a command script language, with
constants, mathematical functions, etc. All functions are not
case-sensitive. Spaces are not allowed between a function name and
“(”. After adding formulas, edit their information in the formula table
as desired. To delete a table row, right-click it and choose Delete Row.
Some information in the table cannot be modified or deleted.
Instructions:
1. Click Select.
2. For each desired formula, select it in the list and click Add.
The available formulas depend on the Formulas & Units tab
settings in Options.
3. Click Close.
To define a formula
1. Click Build.
2. To enter a formula in its entirety, set Equation type to User
defined and type the equation. To use a provided formula, select
it and enter values for the variables.
The available formulas depend on the Formulas & Units tab
settings.
3. Click OK.
4. For selected provided formulas, enter a name and concentration
unit in the table.
The available units depend on the Formulas & Units tab settings.
Related Topics
Type Tab for Nucleic Acid
Feature Description
Mode Specifies Scan or Fixed data acquisition.
Data mode Y-axis format for acquired data.
Start wavelength and End Starting and ending values of the wavelength range for acquiring
wavelength (Scan only) data.
Bandwidth A larger spectral bandwidth increases the amount of energy that
passes through the sample. Increasing the bandwidth can
improve the signal-to-noise ratio; however, a larger spectral
bandwidth decreases the ability to resolve closely spaced spectral
features.
Integration time How long the system acquires and averages data at each data
interval (for scanning measurements) or at each measured
wavelength (for fixed-wavelength measurements). Increasing the
integration time improves the signal-to-noise ratio but reduces
the scan speed.
Data interval (Scan only) Difference in wavelength between two consecutive data points.
Scan speed (Scan only) Wavelength range covered per unit time. Varies inversely with
integration time. Increases as the data interval increases.
Estimated time (Scan Estimated duration of data acquisition. Increases as integration
only) time increases and as the data interval decreases. Varies inversely
with scan speed.
Table of wavelengths Shows the analysis wavelength(s) entered on the Measurement
(Fixed only) tab.
Related Topics
Type Tab for Nucleic Acid
The status of accessories can be monitored during measurements. See Instrument Status
Monitors.
Related Topics
Type Tab for Nucleic Acid
Feature Description
Number of samples Number of samples in the analysis.
Base name Beginning of the name for sample spectra, to which a sequential
number is appended. For example, the names of three sample spectra
using the base name “dye” would be dye1, dye2 and dye3.
Sample averaging Whether and how to average concentration values from multiple
measurements of the same sample or from multiple samples. To
average multiple measurements of the same sample, select Replicate
and set Number of replicates to the total number of measurements.
To average values from two or three samples, select Duplicate or
Triplicate, respectively.
Use sample If available and selected, specifies a multiplication factor for each
correction factor sample result. Can be used to correct for sample properties and
preparation steps such as a sample dilution that affects the measured
result. Enter the desired factor for each sample in the table.
Feature Description
Use control limits Displays high and low limit lines on the Run Chart tab to show
whether sample concentrations are within the specified limits.
Load Samples For locating and selecting a .tsv (tab separated values) file or .csv
(comma separated values) file containing sample names and
descriptions, which are entered in the samples table.
Save Samples Saves the contents of the samples table in a .tsv (tab separated values)
file or .csv (comma separated values) file in a specified location.
Samples table Lists samples by their names and descriptions, up to a maximum of
1000. To change a sample name, click its cell and edit the text. To
enter or change text in a table cell, click it and type. To order the list
by sample name or description, click the appropriate column heading.
If the table contains just one sample and data is acquired with an
accessory that automatically changes samples, sample names are
incremented.
Related Topics
Type Tab for Nucleic Acid
Feature Description
Type Type of analysis to perform. To the right is the factor value from the
Type tab.
Conc. Concentration determined by multiplying the absorbance at 260 nm
by a factor (default, 50) after correction is applied. Beer’s law is used to
calculate the nucleic acid concentration (see Nucleic Acid
Concentration Calculations).
A260 Absorbance at 260 nm.
A280 Absorbance at 280 nm.
Feature Description
260/280 Ratio of sample absorbance at 260 nm and 280 nm, used to assess the
purity of DNA and RNA. A ratio of about 1.8 is generally accepted as
“pure” for DNA, about 2.0 for RNA. If either ratio is appreciably
lower, protein, phenol or other contaminants that absorb strongly
near 280 nm may be present. See Nucleic Acid Purity Measurements
for more information.
260/230 Ratio of absorbance at 260 nm and 230 nm, a secondary measure of
nucleic acid purity.
Advanced calculations
For selecting predefined formulas for additional data processing, with the results appearing in
a calculations table at the bottom of the right pane. Options include basic math and statistics.
The formulas can be applied to specific samples and columns of data in the sample
measurements table or to selected rows and columns in the calculations table for the current
workbook or template. Calculations are applied to all subsequently acquired data. Custom
calculations are saved with the workbook or template. Available for all applications except
Rate.
To delete the contents of a cell in the calculations table, right-click the cell and choose Delete.
1. In the calculations table, double-click the cell where you want the calculation result to
appear.
2. In the Cell Properties box, choose Calculation.
3. Select an Equation in the list.
The available equations are fixed.
1. In the calculations table, double-click the cell where you want the calculation result to
appear.
2. In the Cell Properties box, choose Calculation.
3. Select an Equation in the list.
4. Set Source to Calculations Table.
5. Specify the table cells to include by entering their IDs in the Range/Reference box.
If Equation is set to Std. Deviation, Mean, or %RSD, use a colon to specify a range (e.g.,
A2:A5 or A1:AN) or commas for individual cells (e.g., A1,A3,A5).
If Equation is set to Addition, Subtraction, Multiplication, or Division, specify one cell
location in each Reference box. (e.g., A1).
If Equation is set to a Factor, use the first (Reference) box to specify a cell location and
the second (Factor) box to enter the factor.
6. To select the background color for the cell that displays the calculated result, click
Background, choose a color and click OK.
7. To display the cell contents in bold text, click Bold Text.
8. Choose OK.
The calculated result appears in the selected cell.
Related Topics
Overview
Application Settings
Oligo Calculator
Data Display
Note To start an analysis immediately after viewing or changing the template settings,
click the Measure action button instead and skip to step 3.
2. Click Measure.
More:
If Sample averaging on the Samples tab in Settings was set to Duplicate, “D” at the end of
a sample name indicates the second measurement to be made of the sample. If Sample
averaging was set to Triplicate, “D” and “T” at the end of sample names indicate the
second and third measurements to be made, respectively.
To enter previously saved samples information, use Load Samples. To save the samples
information, use Save Samples.
5. If only one sample will be measured, insert it.
6. Click Continue.
7. Follow any instructions that appear, such as to install a specified sample.
The Run Chart tab plots the concentration of the measured component versus sample
number. If Use concentration limits was selected on the Samples tab, horizontal limit
lines show whether the concentrations are within the specified limits:
Related Topics
Overview
Application Settings
Data Display
Oligo Calculator
Oligo Calculator
Use Oligo Calculator to calculate molecular weight,
extinction coefficients, concentration factors and
melting points for specific nucleic acid sequences.
Clicking this task button displays:
Tab Description
Oligo Calculator For entering sequences of interest and selecting appropriate sample type
variables.
Melting Displays the calculated melting points of a DNA strand. Available only
for DNA sequences.
More:
Use:
• The buttons below the Base Sequence box.
• The A, C, G, T and U keys on the keyboard.
• Copy and paste a base sequence (A, C, G, T and U only) from another application.
To clear the sequence, click Clear. Individual bases can be deleted only manually.
2. Select the nucleic acid type to analyze.
3. Select the degree of phosphorylation if applicable: mono-phosphate for DNA; mono- or
tri-phosphate for RNA.
4. Select Double-stranded if applicable. The complementary base sequence will be
included in the analysis.
5. Enter the molecular weights of any additions to the base sequence in the Modification
box.
The calculated results appear:
Result Description
Molecular weight Calculated base sequence molecular weight.
Molar extinction Calculated 260 nm extinction coefficient in ng-cm/microliter.
coefficient
Concentration factor Factor, based on the extinction coefficient, used to calculate the
concentration of the base sequence.
Result Description
Number of bases Number of bases in the entered sequence.
% GC Percentage of the total number of bases made up by guanine
and cytosine.
Parameter Description
Oligo molarity Concentration of the Oligo in molar units.
Cation molarity Concentration of cations in the sample.
% Formamide Percentage concentration of formamide in the sample.
Result Description
Salt-adjusted Calculated melting point of the base sequence, corrected for the
concentration of cation in the sample. This method does not
account for the effect of interaction between neighboring bases.
Nearest-neighbor Melting point of the base sequence when the effect of interaction
between neighboring bases is taken into account.
Related Topics
Overview
Application Settings
Data Display
• Overview
• Dye/Chromophore Editor
• Application Settings
• Unique Screen Features
• Making Nucleic Acid Labels Measurements
• Oligo Calculator
Overview
Use the Nucleic Acid Labels application to measure
nucleic acid samples for concentration and purity.
Related Topics
Dye/Chromophore Editor
Application Settings
Data Display
Oligo Calculator
Dye/Chromophore Editor
Use the Dye Chromophore Editor to select predefined
dyes, modify existing dyes for protocols, or enter new
dyes.
To enter a new dye, type its information in the appropriate cells of the bottom row of the table
and select a unit. Refer to the dye manufacturer for appropriate correction factors. The
260 nm corrections will be used for nucleic acid sample concentration calculations. Entered
information is saved automatically.
To delete a user-defined dye, right-click its row and choose Delete Row. Predefined dyes,
identified by the lock icon, cannot be edited or deleted.
Related Topics
Overview
Application Settings
Data Display
Oligo Calculator
Application Settings
To set data acquisition parameters for a Nucleic Acid
Labels workbook, click Settings.
Related Topics
Overview
Dye/Chromophore Editor
Data Display
Oligo Calculator
Feature Description
Name Application name.
Description (optional) Description of the template.
Nucleic Acid Type Type of analysis to perform.
Set Dye 1 and Dye 2 to the fluorescent dyes whose absorbance will
be used in concentration correction calculations.
Pathlength Distance the light travels through the sample.
Related Topics
Measurement Tab for Nucleic Acid Labels
Feature Description
Analysis wavelength Wavelength(s) to use for the analysis.
Correction Specifies bichromatic normalization of the absorbance data.
• Single point. Enter a wavelength value to define the endpoint for
the single point baseline.
This option generates a baseline correction for each sample
measurement by drawing a straight line through the specified
baseline point and calculating the absorbance value on the drawn
baseline at the specified analytical wavelength. The software
subtracts the calculated baseline absorbance value from the
measured absorbance value and reports the corrected absorbance
• Sloping baseline. Enter two wavelength values to define the
endpoints for the sloping baseline.
This option generates a baseline correction for each sample
measurement by drawing a line between the two specified
baseline points and calculating the absorbance value on the
drawn baseline at the specified analytical wavelength. The
software subtracts the calculated baseline absorbance value from
the measured absorbance value and reports the corrected
absorbance.
• None. Uses uncorrected data. Without baseline correction,
spectra may be offset from the baseline. If this offset is significant,
the calculated protein concentration may be higher than the true
value.
Component name Component to quantify.
Feature Description
Unit Component concentration unit. The available units depend on the
Formulas & Units tab settings in Options.
Calculate additional For selecting or defining formulas for additional data processing, with
results the results appearing in the analysis report. The formulas are written
in a form similar to that used in a command script language, with
constants, mathematical functions, etc. All functions are not
case-sensitive. Spaces are not allowed between a function name and
“(”. After adding formulas, edit their information in the formula table
as desired. To delete a table row, right-click it and choose Delete Row.
Some information in the table cannot be modified or deleted.
Instructions:
1. Click Select.
2. For each desired formula, select it in the list and click Add.
The available formulas depend on the Formulas & Units tab
settings in Options.
3. Click Close.
To define a formula
1. Click Build.
2. To enter a formula in its entirety, set Equation type to User
defined and type the equation. To use a provided formula, select
it and enter values for the variables.
The available formulas depend on the Formulas & Units tab
settings.
3. Click OK.
4. For selected provided formulas, enter a name and concentration
unit in the table.
The available units depend on the Formulas & Units tab settings.
Related Topics
Feature Description
Mode Specifies Scan or Fixed data acquisition.
Data mode Y-axis format for acquired data.
Start wavelength and End Starting and ending values of the wavelength range for acquiring
wavelength (Scan only) data.
Bandwidth A larger spectral bandwidth increases the amount of energy that
passes through the sample. Increasing the bandwidth can
improve the signal-to-noise ratio; however, a larger spectral
bandwidth decreases the ability to resolve closely spaced spectral
features.
Integration time How long the system acquires and averages data at each data
interval (for scanning measurements) or at each measured
wavelength (for fixed-wavelength measurements). Increasing the
integration time improves the signal-to-noise ratio but reduces
the scan speed.
Data interval (Scan only) Difference in wavelength between two consecutive data points.
Scan speed (Scan only) Wavelength range covered per unit time. Varies inversely with
integration time. Increases as the data interval increases.
Estimated time (Scan Estimated duration of data acquisition. Increases as integration
only) time increases and as the data interval decreases. Varies inversely
with scan speed.
Table of wavelengths Shows the analysis wavelength(s) entered on the Measurement
(Fixed only) tab.
Related Topics
Type Tab for Nucleic Acid Labels
The status of accessories can be monitored during measurements. See Instrument Status
Monitors.
Related Topics
Type Tab for Nucleic Acid Labels
Feature Description
Number of samples Number of samples in the analysis.
Base name Beginning of the name for sample spectra, to which a sequential
number is appended. For example, the names of three sample spectra
using the base name “dye” would be dye1, dye2 and dye3.
Sample averaging Whether and how to average concentration values from multiple
measurements of the same sample or from multiple samples. To
average multiple measurements of the same sample, select Replicate
and set Number of replicates to the total number of measurements.
To average values from two or three samples, select Duplicate or
Triplicate, respectively.
Use sample If available and selected, specifies a multiplication factor for each
correction factor sample result. Can be used to correct for sample properties and
preparation steps such as a sample dilution that affects the measured
result. Enter the desired factor for each sample in the table.
Use control limits Displays high and low limit lines on the Run Chart tab to show
whether sample concentrations are within the specified limits.
Feature Description
Load Samples For locating and selecting a .tsv (tab separated values) file or .csv
(comma separated values) file containing sample names and
descriptions, which are entered in the samples table.
Save Samples Saves the contents of the samples table in a .tsv (tab separated values)
file or .csv (comma separated values) file in a specified location.
Samples table Lists samples by their names and descriptions, up to a maximum of
1000. To change a sample name, click its cell and edit the text. To
enter or change text in a table cell, click it and type. To order the list
by sample name or description, click the appropriate column heading.
If the table contains just one sample and data is acquired with an
accessory that automatically changes samples, sample names are
incremented.
Related Topics
Type Tab for Nucleic Acid Labels
Feature Description
Type Type of analysis to perform. To the right is the factor value from the
Type tab.
Conc. Concentration based on absorbance at 260 nm, the default or
user-defined extinction coefficient, and the specified pathlength. (The
extinction coefficient is reported for a 10 mm pathlength.) Beer’s law
is used to calculate the nucleic acid concentration (see Nucleic Acid
Concentration Calculations).
Feature Description
A260 Absorbance at 260 nm.
260/280 Ratio of sample absorbance at 260 nm and 280 nm, used to assess the
purity of DNA and RNA. A ratio of ~1.8 is generally accepted as
“pure” for DNA, ~2.0 for RNA. If either ratio is appreciably lower,
protein, phenol or other contaminants that absorb strongly near
280 nm may be present. See Nucleic Acid Purity Measurements for
more information.
Dye(s) Conc. Concentration of the user-selected dye(s).
Advanced calculations
For selecting predefined formulas for additional data processing, with the results appearing in
a calculations table at the bottom of the right pane. Options include basic math and statistics.
The formulas can be applied to specific samples and columns of data in the sample
measurements table or to selected rows and columns in the calculations table for the current
workbook or template. Calculations are applied to all subsequently acquired data. Custom
calculations are saved with the workbook or template. Available for all applications except
Rate.
To delete the contents of a cell in the calculations table, right-click the cell and choose Delete.
1. In the calculations table, double-click the cell where you want the calculation result to
appear.
1. In the calculations table, double-click the cell where you want the calculation result to
appear.
2. In the Cell Properties box, choose Calculation.
3. Select an Equation in the list.
4. Set Source to Calculations Table.
5. Specify the table cells to include by entering their IDs in the Range/Reference box.
If Equation is set to Std. Deviation, Mean, or %RSD, use a colon to specify a range (e.g.,
A2:A5 or A1:AN) or commas for individual cells (e.g., A1,A3,A5).
If Equation is set to Addition, Subtraction, Multiplication, or Division, specify one cell
location in each Reference box. (e.g., A1).
If Equation is set to a Factor, use the first (Reference) box to specify a cell location and
the second (Factor) box to enter the factor.
6. To select the background color for the cell that displays the calculated result, click
Background, choose a color and click OK.
7. To display the cell contents in bold text, click Bold Text.
8. Choose OK.
The calculated result appears in the selected cell.
Related Topics
Overview
Dye/Chromophore Editor
Application Settings
Data Display
Oligo Calculator
Note To start an analysis immediately after viewing or changing the template settings,
click the Measure action button instead and skip to step 3.
2. Click Measure.
More:
If Sample averaging on the Samples tab in Settings was set to Duplicate, “D” at the end of
a sample name indicates the second measurement to be made of the sample. If Sample
averaging was set to Triplicate, “D” and “T” at the end of sample names indicate the
second and third measurements to be made, respectively.
To enter previously saved samples information, use Load Samples. To save the samples
information, use Save Samples.
The Run Chart tab plots the concentration of the measured component versus sample
number. If Use concentration limits was selected on the Samples tab, horizontal limit
lines show whether the concentrations are within the specified limits:
Related Topics
Overview
Dye/Chromophore Editor
Application Settings
Data Display
Oligo Calculator
Oligo Calculator
Use Oligo Calculator to calculate molecular weight,
extinction coefficients, concentration factors and
melting points for specific nucleic acid sequences.
Clicking this task button displays:
Tab Description
Oligo Calculator For entering sequences of interest and selecting appropriate sample type
variables.
Melting Displays the calculated melting points of a DNA strand. Available only
for DNA sequences.
More:
Use:
• The buttons below the Base Sequence box.
• The A, C, G, T and U keys on the keyboard.
• Copy and paste a base sequence (A, C, G, T and U only) from another application.
To clear the sequence, click Clear. Individual bases can be deleted only manually.
2. Select the nucleic acid type to analyze.
3. Select the degree of phosphorylation if applicable: mono-phosphate for DNA; mono- or
tri-phosphate for RNA.
4. Select Double-stranded if applicable. The complementary base sequence will be
included in the analysis.
5. Enter the molecular weights of any additions to the base sequence in the Modification
box.
Result Description
Molecular weight Calculated base sequence molecular weight.
Molar extinction Calculated 260 nm extinction coefficient in ng-cm/microliter.
coefficient
Concentration factor Factor, based on the extinction coefficient, used to calculate the
concentration of the base sequence.
Number of bases Number of bases in the entered sequence.
% GC Percentage of the total number of bases made up by guanine
and cytosine.
Parameter Description
Oligo molarity Concentration of the Oligo in molar units.
Cation molarity Concentration of cations in the sample.
% Formamide Percentage concentration of formamide in the sample.
Result Description
Salt-adjusted Calculated melting point of the base sequence, corrected for the
concentration of cation in the sample. This method does not
account for the effect of interaction between neighboring bases.
Nearest-neighbor Melting point of the base sequence when the effect of interaction
between neighboring bases is taken into account.
Related Topics
Overview
Dye/Chromophore Editor
Application Settings
Data Display
DNA Melting
Contents
• Overview
• DNA Melting Point Calculations
• Application Settings
• Unique Screen Features
• Making DNA Melting Measurements
Overview
Use the DNA Melting application to determine the
temperature at which a DNA sample denatures based on
the change in absorbance as the temperature is increased.
As the temperature is increased, the electronic interactions between the DNA bases are
modified and the entire absorption spectrum increases in intensity. This phenomenon is
called the hyperchromic effect and the increase in absorption due to the abrupt change in
structure is usually around 30–40%.
See Unique Screen Features for information about the DNA Melting data acquisition results.
For information about the calculations, see DNA Melting Point Calculations.
• Sloping Intercept. Used when the absorbance varies with temperature inside and outside
the melting range. Requires three temperature values (t1, t2 and t3).
T1 is a temperature value in °C that occurs below the temperature range in which the
DNA melting point occurs.
T3 is a temperature value in °C that occurs above the DNA melting range.
T2 is an intermediate temperature value in °C.
The software constructs a best fit line between the point at T2 and the point at the other
temperature that is closest to it. A second line is drawn through the third point and
parallel with the first line. Am (the absorbance at Tm) is midway between the points where
the two parallel lines intercept the Y-axis.
A third line parallel to the others is drawn from Am to the Absorbance vs Temperature
curve, and a perpendicular line is dropped from that intersection to the X-axis. Tm is the
temperature at which the perpendicular line intercepts the X-axis.
Related Topics
Overview
Application Settings
Application Settings
To set data acquisition parameters for a DNA Melting
workbook, click Settings.
Related Topics
Overview
Feature Description
Name (if User group name followed by the name of the template of settings (entered
present) when the template was saved using Save Workbook Settings as Template in
the File menu).
Description Description of the template.
Feature Description
Rate Type Type of rate measurement to perform (see the table below). An image
representing the selected type appears to the right. The required tabs also
appear (click the Reset action button to step through them). These options
are available:
• Single wavelength. For each sample, measures a specified wavelength at
specified intervals of time.
• Multiple wavelengths. For each sample, measures multiple specified
wavelengths at specified intervals of time. Up to 40 wavelengths can be
specified.
Mode Temperature is the only available option for the DNA Melting application. It
performs rate measurements based on elapsed time and also records
temperature settings.
Related Topics
Overview
Feature Description
Integration time How long the system acquires and averages data to obtain each data
point at each measured wavelength for single and multicell
experiments.
Dwell time How long the sample remains in the light beam. Available for single
and multicell experiments.
Use this setting to vary the number of data points collected per cell
per cycle. Dwell time is always a multiple of integration time (see
above). Dwell time can be very useful when performing multicell
rate measurements. This is because the cell changer accessory will
position a cell in the light beam for a specified period of time (dwell
time), rather than moving back and forth between cells and
increasing the time between measurements (when no sample is in
the light beam).
For example, if the integration time is 0.1 second and dwell time is
0.5 second, five data points will be recorded for each cell before the
cell changer moves to the next cell. For fairly fast reactions, multiple
samples can be measured, with over ten times as many data points
acquired within a short time. This can improve the accuracy of rate
calculations for faster multicell rate measurements.
Number of stages Defines the number of measurement cycles over which the frequency
of data acquisition can be determined. For example, if a reaction has
two components, the first component could be fast and require a
high data density. The second component could be much slower and
longer lived, therefore requiring a lower data density. Defining
multiple stages enables the use of the proper data density over a
given period of time.
Temp unit Determines the temperature unit used in the stages table (see below).
Feature Description
Time unit Determines the time unit used in the stages table (see below).
Stages table Defines data acquisition for each stage of an experiment including
the target temperature, ramp rate, hold type, hold time, interval and
whether data collection will occur.
Note that the temperature in the final stage can be set to a value
around 10 °C to simulate storage in a refrigerator. This feature
allows proper storage conditions for the sample if DNA melting
experiments will be completed when the instrument is unattended.
Hold Time. When Hold Type is set to Time, enter the length of
time to hold at the target temperature before continuing to the next
stage.
Related Topics
Overview
Feature Description
Data mode Y-axis format for acquired data. Absorbance is the only available
option for the DNA Melting application.
Bandwidth A larger spectral bandwidth increases the amount of energy that
passes through the sample. Increasing the bandwidth can improve the
signal-to-noise ratio; however, a larger spectral bandwidth decreases
the ability to resolve closely spaced spectral features.
Reference When selected, automatically acquires a reference measurement at a
wavelength specified wavelength each time a sample measurement is taken. The
correction reference measurement is then subtracted from the corresponding
sample measurement to produce a corrected measurement result.
Only the corrected result is reported.
Related Topics
Overview
The status of accessories can be monitored during measurements. See Instrument Status
Monitors.
Related Topics
Evolution 200 Series Accessories
Overview
Feature Description
Number of samples Number of samples for the experiment.
Base name Beginning of the name for sample spectra, to which a sequential
number is appended. For example, the names of three sample spectra
using the base name “dye” would be dye1, dye2 and dye3.
Set up sample table Specifies the columns of sample information to be entered during the
columns experiment. To add a column, type the desired name in an empty cell
and press the Tab key. To delete a column (except the Sample ID
column), right-click it and choose Delete Column. To delete all the
columns (except the Sample ID column), right-click the table and
choose Delete All Columns.
Load Samples For locating and selecting a .tsv (tab separated values) file or .csv
(comma separated values) file containing sample names and
descriptions, which are entered in the samples table on the Samples
tab.
Feature Description
Save Samples Saves the contents of the samples table in a .tsv (tab separated values)
file or .csv (comma separated values) file in a specified location.
Samples table Lists samples by their names and descriptions, up to a maximum of
1000. Additional columns are defined by the Set up sample table
columns option (see above).
To change a sample name, click its cell and edit the text. To enter or
change text in a table cell, click it and type. To order the list by
sample name or description, click the appropriate column heading.
If the table contains just one sample and data is acquired with an
accessory that automatically changes samples, sample names are
incremented.
Related Topics
Overview
Feature Description
Calculate Tm values Calculates DNA melting point temperature (Tm) values
automatically at end of automatically at the end of the experiment. When this option is
data acquisition selected, the entered temperature values are marked in the data
display and the calculated Tm values appear in the sample
measurements table after the measurement is completed. To
modify the Tm settings, double-click the data display or choose
Analyze (menu) > Tm Calculations.
Feature Description
Tm Calculation Select a method for calculating DNA melting point temperature
values. These options are available:
• Horizontal Intercept. Uses the absorbance at two specified
temperature values and their horizontal intercepts to calculate
Tm. Useful when absorbance does not vary with temperature
outside the melting range.
• Sloping Intercept. Uses the absorbance at three specified
temperature values and their sloped intercepts to calculate Tm.
Select this option when the absorbance varies with
temperature inside and outside the melting range.
• Inflection. Calculates the first derivative of the Absorbance vs
Temperature curve. Tm is the temperature at which the first
derivative curve has its maximum value (i.e., the slope).
Related Topics
Overview
Feature Description
Extend Time To extend the collection time for the current sample measurement
(without interrupting the data collection), click Extend Time and
enter the amount of time to extend the measurement. If you are
collecting data in multiple stages, only the current stage is affected.
Related Topics
Overview
Application Settings
Data Display
Measurements are taken at specified intervals of time for each stage defined on the
Measurements tab in Settings. Temperature settings and ramp rate are controlled and
recorded for the duration of the experiment.
It may be helpful to display the following status monitors during your DNA Melting
experiments:
• Cell Changer status
• Temperature Controller status
• Time status
• Temperature probe status
To display status monitors, click the Configure Instrument Status icon above the data
display.
In the Configure Instrument Status box, select the status monitors to display and click OK.
Note To configure the software to display status monitors automatically, display the status
monitors and then save the workbook as a template.
Each row in the sample measurements table below the plots shows the measurement
results for one sample at a specified wavelength. The table contains information specified
on the Reportable Data tab in Reports, plus the Tm value if Calculate Tm values
automatically at end of data acquisition is selected on the Settings > Tm Calculations
tab.
To extend the collection time for the current sample measurement (without interrupting
the data collection), click Extend Time in the navigation pane and enter the amount of
time to extend the measurement. If you are collecting data in multiple stages, only the
current stage is affected.
Right-click the Data tab to access features for changing the scale or adding annotation.
See Data Display for more information.
To measure a sample again, right-click its row in the sample measurements table and
choose Remeasure (if available). (If Remeasure is not available, you can choose Options
> Preferences and clear Prevent removal of data to enable it.) After the remeasurement,
the previous information for the sample is either crossed out or removed from the table,
depending on the status of Prevent removal of data on the Preferences tab in Options.
To copy the data, right-click the plot and choose Copy to Clipboard.
7. To calculate or modify the DNA melting temperature (Tm) calculations on the data in a
plot, double-click the plot or select Analyze (menu) > Tm Calculations.
Related Topics
Overview
Application Settings
Modifying Tm Calculations
Data Display
Modifying Tm Calculations
DNA melting measurements must define how the DNA melting point will be measured. This
is done on the Tm Calculations tab in Settings.
• If desired, change the Tm Calculation setting to best fit the shape of the DNA melting
point curve. If you select a different calculation method, add or change the required
temperature values as needed. For more information, see DNA Melting Point
Calculations.
• Click Update to recalculate Tm values based on the selected calculation method.
• Specify how to apply the calculations and then click Accept.
After you click Accept, the window closes and the sample measurements table is updated with
the new Tm values.
Related Topics
Overview
Application Settings
Protein A280
Contents
• Overview
• Application Settings
• Unique Screen Features
• Making Protein A280 Measurements
Overview
The Protein A280 method is used to quantify proteins on
the basis of their absorbance at 280 nm.
Tyrosine, tryptophan, phenylalanine and to a small extent lysine and Cys-Cys disulphide
bonds have an absorbance peak around 280 nm. This method does not require generation of
a standard curve and is ready for protein sample quantitation at software startup. Colorimetric
assays such as Pierce BCA, Pierce 660 nm, Bradford and Lowry are more commonly used for
uncharacterized protein solutions and cell lysates.
The Protein A280 application displays the UV spectrum, measures the protein’s absorbance at
280 nm (A280), and calculates the concentration (mg/ml).
Related Topics
Application Settings
Data Display
Application Settings
To set data acquisition parameters for a Protein A280
workbook, click Settings.
Related Topics
Overview
Data Display
Feature Description
Name Application name.
Description (optional) Description of the template.
Feature Description
Protein A280 Type Type of analysis to perform.
Related Topics
Measurement Tab for Protein A280
Feature Description
Analysis wavelength Wavelength(s) to use for the analysis.
Correction Specifies bichromatic normalization of the absorbance data.
• Single point. Enter a wavelength value to define the endpoint for
the single point baseline.
This option generates a baseline correction for each sample
measurement by drawing a straight line through the specified
baseline point and calculating the absorbance value on the drawn
baseline at the specified analytical wavelength. The software
subtracts the calculated baseline absorbance value from the
measured absorbance value and reports the corrected absorbance
• Sloping baseline. Enter two wavelength values to define the
endpoints for the sloping baseline.
This option generates a baseline correction for each sample
measurement by drawing a line between the two specified
baseline points and calculating the absorbance value on the
drawn baseline at the specified analytical wavelength. The
software subtracts the calculated baseline absorbance value from
the measured absorbance value and reports the corrected
absorbance.
• None. Uses uncorrected data. Without baseline correction,
spectra may be offset from the baseline. If this offset is significant,
the calculated protein concentration may be higher than the true
value.
Component name Component to quantify.
Feature Description
Unit Component concentration unit. The available units depend on the
Formulas & Units tab settings in Options.
Calculate additional For selecting or defining formulas for additional data processing, with
results the results appearing in the analysis report. The formulas are written
in a form similar to that used in a command script language, with
constants, mathematical functions, etc. All functions are not
case-sensitive. Spaces are not allowed between a function name and
“(”. After adding formulas, edit their information in the formula table
as desired. To delete a table row, right-click it and choose Delete Row.
Some information in the table cannot be modified or deleted.
Instructions:
1. Click Select.
2. For each desired formula, select it in the list and click Add.
The available formulas depend on the Formulas & Units tab
settings in Options.
3. Click Close.
To define a formula
1. Click Build.
2. To enter a formula in its entirety, set Equation type to User
defined and type the equation. To use a provided formula, select
it and enter values for the variables.
The available formulas depend on the Formulas & Units tab
settings.
3. Click OK.
4. For selected provided formulas, enter a name and concentration
unit in the table.
The available units depend on the Formulas & Units tab settings.
Related Topics
Type Tab for Protein A280
Feature Description
Mode Specifies Scan or Fixed data acquisition.
Data mode Y-axis format for acquired data.
Start wavelength and End Starting and ending values of the wavelength range for acquiring
wavelength (Scan only) data.
Bandwidth A larger spectral bandwidth increases the amount of energy that
passes through the sample. Increasing the bandwidth can
improve the signal-to-noise ratio; however, a larger spectral
bandwidth decreases the ability to resolve closely spaced spectral
features.
Integration time How long the system acquires and averages data at each data
interval (for scanning measurements) or at each measured
wavelength (for fixed-wavelength measurements). Increasing the
integration time improves the signal-to-noise ratio but reduces
the scan speed.
Data interval (Scan only) Difference in wavelength between two consecutive data points.
Scan speed (Scan only) Wavelength range covered per unit time. Varies inversely with
integration time. Increases as the data interval increases.
Estimated time (Scan Estimated duration of data acquisition. Increases as integration
only) time increases and as the data interval decreases. Varies inversely
with scan speed.
Table of wavelengths Shows the analysis wavelength(s) entered on the Measurement
(Fixed only) tab.
Related Topics
Type Tab for Protein A280
The status of accessories can be monitored during measurements. See Instrument Status
Monitors.
Related Topics
Type Tab for Protein A280
Feature Description
Number of samples Number of samples in the analysis.
Base name Beginning of the name for sample spectra, to which a sequential
number is appended. For example, the names of three sample spectra
using the base name “dye” would be dye1, dye2 and dye3.
Sample averaging Whether and how to average concentration values from multiple
measurements of the same sample or from multiple samples. To
average multiple measurements of the same sample, select Replicate
and set Number of replicates to the total number of measurements.
To average values from two or three samples, select Duplicate or
Triplicate, respectively.
Use sample If available and selected, specifies a multiplication factor for each
correction factor sample result. Can be used to correct for sample properties and
preparation steps such as a sample dilution that affects the measured
result. Enter the desired factor for each sample in the table.
Use control limits Displays high and low limit lines on the Run Chart tab to show
whether sample concentrations are within the specified limits.
Feature Description
Load Samples For locating and selecting a .tsv (tab separated values) file or .csv
(comma separated values) file containing sample names and
descriptions, which are entered in the samples table.
Save Samples Saves the contents of the samples table in a .tsv (tab separated values)
file or .csv (comma separated values) file in a specified location.
Samples table Lists samples by their names and descriptions, up to a maximum of
1000. To change a sample name, click its cell and edit the text. To
enter or change text in a table cell, click it and type. To order the list
by sample name or description, click the appropriate column heading.
If the table contains just one sample and data is acquired with an
accessory that automatically changes samples, sample names are
incremented.
Related Topics
Type Tab for Protein A280
Feature Description
Type Type of analysis to perform.
/1000 and Molecular /1000 and molecular weight for the Other protein ( &
Weight (kDa) MW) sample type.
Ext. Coeff, 1% (L/g-cm) Extinction coefficient and 1% for the Other protein (
1%) sample type.
Conc. Concentration based on absorbance at 280 nm, after
correction is applied, and the selected extinction coefficient.
A280 Absorbance at 280 nm for the protein sample being
measured.
260/280 Ratio of absorbance at 260 nm and 280 nm.
Advanced calculations
For selecting predefined formulas for additional data processing, with the results appearing in
a calculations table at the bottom of the right pane. Options include basic math and statistics.
The formulas can be applied to specific samples and columns of data in the sample
measurements table or to selected rows and columns in the calculations table for the current
workbook or template. Calculations are applied to all subsequently acquired data. Custom
calculations are saved with the workbook or template. Available for all applications except
Rate.
To delete the contents of a cell in the calculations table, right-click the cell and choose Delete.
1. In the calculations table, double-click the cell where you want the calculation result to
appear.
2. In the Cell Properties box, choose Calculation.
3. Select an Equation in the list.
The available equations are fixed.
4. Set Source to Samples table.
5. Select a sample measurement in the Measurements list.
6. Specify the sample rows to include by entering their ID numbers in the Range/Reference
box.
Use a colon to specify a range (e.g., 2:5 or 1:N) or commas for individual samples (e.g.,
1,3,5).
7. To select the background color for the cell that displays the calculated result, click
Background, choose a color and click OK.
8. To display the cell contents in bold text, click Bold Text.
9. Choose OK.
The calculated result appears in the selected cell.
1. In the calculations table, double-click the cell where you want the calculation result to
appear.
2. In the Cell Properties box, choose Calculation.
3. Select an Equation in the list.
4. Set Source to Calculations Table.
5. Specify the table cells to include by entering their IDs in the Range/Reference box.
If Equation is set to Std. Deviation, Mean, or %RSD, use a colon to specify a range (e.g.,
A2:A5 or A1:AN) or commas for individual cells (e.g., A1,A3,A5).
If Equation is set to Addition, Subtraction, Multiplication, or Division, specify one cell
location in each Reference box. (e.g., A1).
If Equation is set to a Factor, use the first (Reference) box to specify a cell location and
the second (Factor) box to enter the factor.
6. To select the background color for the cell that displays the calculated result, click
Background, choose a color and click OK.
7. To display the cell contents in bold text, click Bold Text.
8. Choose OK.
The calculated result appears in the selected cell.
Related Topics
Overview
Application Settings
Data Display
Note To start an analysis immediately after viewing or changing the template settings,
click the Measure action button instead and skip to step 3.
2. Click Measure.
More:
If Sample averaging on the Samples tab in Settings was set to Duplicate, “D” at the end of
a sample name indicates the second measurement to be made of the sample. If Sample
averaging was set to Triplicate, “D” and “T” at the end of sample names indicate the
second and third measurements to be made, respectively.
To enter previously saved samples information, use Load Samples. To save the samples
information, use Save Samples.
5. If only one sample will be measured, insert it.
6. Click Continue.
7. Follow any instructions that appear, such as to install a specified sample.
The Run Chart tab plots the concentration of the measured component versus sample
number. If Use concentration limits was selected on the Samples tab, horizontal limit
lines show whether the concentrations are within the specified limits:
Related Topics
Overview
Application Settings
Data Display
Overview
The Proteins & Labels application determines protein
concentration using the protein’s intrinsic absorption at
280 nm as well as the concentration of a fluorescent dye
modification (protein array conjugates). It can also
measure the purity of metalloproteins (such as
hemoglobin) using wavelength ratios.
Related Topics
Dye/Chromophore Editor
Application Settings
Data Display
Dye/Chromophore Editor
Use the Dye Chromophore Editor to select predefined
dyes, modify existing dyes for protocols, or enter new
dyes.
To enter a new dye, type its information in the appropriate cells of the bottom row of the table
and select a unit. Refer to the dye manufacturer for appropriate correction factors. The
260 nm corrections will be used for nucleic acid sample concentration calculations. Entered
information is saved automatically.
To delete a user-defined dye, right-click its row and choose Delete Row. Predefined dyes,
identified by the lock icon, cannot be edited or deleted.
Related Topics
Overview
Application Settings
Data Display
Application Settings
To set data acquisition parameters for a Proteins and
Labels workbook, click Settings.
Related Topics
Overview
Dye/Chromophore Editor
Data Display
Feature Description
Name Application name.
Description (optional) Description of the template.
Feature Description
Type Type of analysis to perform.
Related Topics
Measurement Tab for Proteins and Labels
Feature Description
Analysis wavelength Wavelength(s) to use for the analysis.
Correction Specifies bichromatic normalization of the absorbance data.
• Single point. Enter a wavelength value to define the endpoint for
the single point baseline.
This option generates a baseline correction for each sample
measurement by drawing a straight line through the specified
baseline point and calculating the absorbance value on the drawn
baseline at the specified analytical wavelength. The software
subtracts the calculated baseline absorbance value from the
measured absorbance value and reports the corrected absorbance
• Sloping baseline. Enter two wavelength values to define the
endpoints for the sloping baseline.
This option generates a baseline correction for each sample
measurement by drawing a line between the two specified
baseline points and calculating the absorbance value on the
drawn baseline at the specified analytical wavelength. The
software subtracts the calculated baseline absorbance value from
the measured absorbance value and reports the corrected
absorbance.
• None. Uses uncorrected data. Without baseline correction,
spectra may be offset from the baseline. If this offset is significant,
the calculated protein concentration may be higher than the true
value.
Component name Component to quantify.
Feature Description
Unit Component concentration unit. The available units depend on the
Formulas & Units tab settings in Options.
Calculate additional For selecting or defining formulas for additional data processing, with
results the results appearing in the analysis report. The formulas are written
in a form similar to that used in a command script language, with
constants, mathematical functions, etc. All functions are not
case-sensitive. Spaces are not allowed between a function name and
“(”. After adding formulas, edit their information in the formula table
as desired. To delete a table row, right-click it and choose Delete Row.
Some information in the table cannot be modified or deleted.
Instructions:
1. Click Select.
2. For each desired formula, select it in the list and click Add.
The available formulas depend on the Formulas & Units tab
settings in Options.
3. Click Close.
To define a formula
1. Click Build.
2. To enter a formula in its entirety, set Equation type to User
defined and type the equation. To use a provided formula, select
it and enter values for the variables.
The available formulas depend on the Formulas & Units tab
settings.
3. Click OK.
4. For selected provided formulas, enter a name and concentration
unit in the table.
The available units depend on the Formulas & Units tab settings.
Related Topics
Type Tab for Proteins and Labels
Feature Description
Mode Specifies Scan or Fixed data acquisition.
Data mode Y-axis format for acquired data.
Start wavelength and End Starting and ending values of the wavelength range for acquiring
wavelength (Scan only) data.
Bandwidth A larger spectral bandwidth increases the amount of energy that
passes through the sample. Increasing the bandwidth can
improve the signal-to-noise ratio; however, a larger spectral
bandwidth decreases the ability to resolve closely spaced spectral
features.
Integration time How long the system acquires and averages data at each data
interval (for scanning measurements) or at each measured
wavelength (for fixed-wavelength measurements). Increasing the
integration time improves the signal-to-noise ratio but reduces
the scan speed.
Data interval (Scan only) Difference in wavelength between two consecutive data points.
Scan speed (Scan only) Wavelength range covered per unit time. Varies inversely with
integration time. Increases as the data interval increases.
Estimated time (Scan Estimated duration of data acquisition. Increases as integration
only) time increases and as the data interval decreases. Varies inversely
with scan speed.
Table of wavelengths Shows the analysis wavelength(s) entered on the Measurement
(Fixed only) tab.
Related Topics
Type Tab for Proteins and Labels
The status of accessories can be monitored during measurements. See Instrument Status
Monitors.
Related Topics
Type Tab for Proteins and Labels
Feature Description
Number of samples Number of samples in the analysis.
Base name Beginning of the name for sample spectra, to which a sequential
number is appended. For example, the names of three sample spectra
using the base name “dye” would be dye1, dye2 and dye3.
Sample averaging Whether and how to average concentration values from multiple
measurements of the same sample or from multiple samples. To
average multiple measurements of the same sample, select Replicate
and set Number of replicates to the total number of measurements.
To average values from two or three samples, select Duplicate or
Triplicate, respectively.
Use sample If available and selected, specifies a multiplication factor for each
correction factor sample result. Can be used to correct for sample properties and
preparation steps such as a sample dilution that affects the measured
result. Enter the desired factor for each sample in the table.
Use control limits Displays high and low limit lines on the Run Chart tab to show
whether sample concentrations are within the specified limits.
Feature Description
Load Samples For locating and selecting a .tsv (tab separated values) file or .csv
(comma separated values) file containing sample names and
descriptions, which are entered in the samples table.
Save Samples Saves the contents of the samples table in a .tsv (tab separated values)
file or .csv (comma separated values) file in a specified location.
Samples table Lists samples by their names and descriptions, up to a maximum of
1000. To change a sample name, click its cell and edit the text. To
enter or change text in a table cell, click it and type. To order the list
by sample name or description, click the appropriate column heading.
If the table contains just one sample and data is acquired with an
accessory that automatically changes samples, sample names are
incremented.
Related Topics
Type Tab for Proteins and Labels
Feature Description
Type Type of analysis to perform.
/1000 and Molecular Weight /1000 and molecular weight for the Other protein ( &
(kDa) MW) sample type.
Ext. Coeff, 1% (L/g-cm) Extinction coefficient and 1% for the Other protein (
1%) sample type.
Feature Description
Conc. Concentration based on absorbance at 280 nm, after
correction is applied, and the selected extinction
coefficient.
Dye(s) Conc. Concentration of the user-selected dye(s).
Advanced calculations
For selecting predefined formulas for additional data processing, with the results appearing in
a calculations table at the bottom of the right pane. Options include basic math and statistics.
The formulas can be applied to specific samples and columns of data in the sample
measurements table or to selected rows and columns in the calculations table for the current
workbook or template. Calculations are applied to all subsequently acquired data. Custom
calculations are saved with the workbook or template. Available for all applications except
Rate.
To delete the contents of a cell in the calculations table, right-click the cell and choose Delete.
1. In the calculations table, double-click the cell where you want the calculation result to
appear.
2. In the Cell Properties box, choose Calculation.
3. Select an Equation in the list.
The available equations are fixed.
1. In the calculations table, double-click the cell where you want the calculation result to
appear.
2. In the Cell Properties box, choose Calculation.
3. Select an Equation in the list.
4. Set Source to Calculations Table.
5. Specify the table cells to include by entering their IDs in the Range/Reference box.
If Equation is set to Std. Deviation, Mean, or %RSD, use a colon to specify a range (e.g.,
A2:A5 or A1:AN) or commas for individual cells (e.g., A1,A3,A5).
If Equation is set to Addition, Subtraction, Multiplication, or Division, specify one cell
location in each Reference box. (e.g., A1).
If Equation is set to a Factor, use the first (Reference) box to specify a cell location and
the second (Factor) box to enter the factor.
6. To select the background color for the cell that displays the calculated result, click
Background, choose a color and click OK.
7. To display the cell contents in bold text, click Bold Text.
8. Choose OK.
The calculated result appears in the selected cell.
Related Topics
Overview
Dye/Chromophore Editor
Application Settings
Data Display
Note To start an analysis immediately after viewing or changing the template settings,
click the Measure action button instead and skip to step 3.
2. Click Measure.
More:
If Sample averaging on the Samples tab in Settings was set to Duplicate, “D” at the end of
a sample name indicates the second measurement to be made of the sample. If Sample
averaging was set to Triplicate, “D” and “T” at the end of sample names indicate the
second and third measurements to be made, respectively.
To enter previously saved samples information, use Load Samples. To save the samples
information, use Save Samples.
5. If only one sample will be measured, insert it.
6. Click Continue.
7. Follow any instructions that appear, such as to install a specified sample.
The Run Chart tab plots the concentration of the measured component versus sample
number. If Use concentration limits was selected on the Samples tab, horizontal limit
lines show whether the concentrations are within the specified limits:
Related Topics
Overview
Dye/Chromophore Editor
Application Settings
Data Display
Data Display
Pierce BCA
Contents
• Overview
• Application Settings
• Unique Screen Features
• Making Pierce BCA Measurements
Overview
The BCA Protein Assay, which was patented by Pierce
Biotechnology, part of Thermo Fisher Scientific, remains a
cornerstone for accurately measuring protein
concentration in biological samples. The method uses
bicinchoninic acid (BCA) as the detection reagent for
Cu+, which is formed when Cu2+ is reduced by protein in
an alkaline environment. A purple-colored reaction
product is formed by the chelation of two molecules of
BCA with one cuprous ion (Cu+). The resulting Cu-BCA
chelate formed in the presence of protein is measured at
562 nm. While this assay is linear over a wide
concentration range, the calibration curves are best
represented as second order. Preformulated reagents of
BCA and CuSO4 are available in kit form from us.
Follow the assay kit manufacturer’s recommendations for all standards and samples
(unknowns). Ensure each is subjected to the same timing and temperature throughout the
assay.
Note If running the assay at 60 °C, doubling the volumes may afford greater insurance
against skewed results from evaporation/condensation within the sealed reaction tube.
Protein standards (BSA) for generating a standard curve may also be provided by the
manufacturer for the BCA assay.
Related Topics
Application Settings
Data Display
Application Settings
To set data acquisition parameters for a Pierce BCA
workbook, click Settings.
Related Topics
Overview
Data Display
Feature Description
Name Application name.
Description (optional) Description of the template.
Standard curve Type of quantitative analysis used.
Pathlength Distance the light travels through the sample.
Related Topics
Measurement Tab for Pierce BCA
Feature Description
Analysis wavelength Wavelength(s) to use for the analysis.
Correction Specifies bichromatic normalization of the absorbance data.
• Single point. Enter a wavelength value to define the endpoint for
the single point baseline.
This option generates a baseline correction for each sample
measurement by drawing a straight line through the specified
baseline point and calculating the absorbance value on the drawn
baseline at the specified analytical wavelength. The software
subtracts the calculated baseline absorbance value from the
measured absorbance value and reports the corrected absorbance
• Sloping baseline. Enter two wavelength values to define the
endpoints for the sloping baseline.
This option generates a baseline correction for each sample
measurement by drawing a line between the two specified
baseline points and calculating the absorbance value on the
drawn baseline at the specified analytical wavelength. The
software subtracts the calculated baseline absorbance value from
the measured absorbance value and reports the corrected
absorbance.
• None. Uses uncorrected data. Without baseline correction,
spectra may be offset from the baseline. If this offset is significant,
the calculated protein concentration may be higher than the true
value.
Component name Component to quantify.
Feature Description
Unit Component concentration unit. The available units depend on the
Formulas & Units tab settings in Options.
Calculate additional For selecting or defining formulas for additional data processing, with
results the results appearing in the analysis report. The formulas are written
in a form similar to that used in a command script language, with
constants, mathematical functions, etc. All functions are not
case-sensitive. Spaces are not allowed between a function name and
“(”. After adding formulas, edit their information in the formula table
as desired. To delete a table row, right-click it and choose Delete Row.
Some information in the table cannot be modified or deleted.
Instructions:
1. Click Select.
2. For each desired formula, select it in the list and click Add.
The available formulas depend on the Formulas & Units tab
settings in Options.
3. Click Close.
To define a formula
1. Click Build.
2. To enter a formula in its entirety, set Equation type to User
defined and type the equation. To use a provided formula, select
it and enter values for the variables.
The available formulas depend on the Formulas & Units tab
settings.
3. Click OK.
4. For selected provided formulas, enter a name and concentration
unit in the table.
The available units depend on the Formulas & Units tab settings.
Related Topics
Type Tab for Pierce BCA
Feature Description
Curve fit type Type of equation used to create the standard curve from standard
concentration values.
Standard averaging Whether and how to average concentration values from multiple
measurements of the same standard or from multiple standards
prepared under the same conditions. To average multiple
measurements of the same standard, select Replicate and set Number
of replicates to the total number of measurements. To average values
from two or three similar standards, select Duplicate or Triplicate,
respectively.
Minimum r2 The r2 value indicates how well the standard curve fits the standard
data points, with 1.0 a perfect fit. If Minimum r2 is selected, samples
can be quantified only after that value is achieved for the standard
curve.
Calculate from If available for the current unit, calculates concentration when the
weight/volume weight and volume of material used to prepare each standard is
known. Enter the weight and volume for each standard in the
appropriate table cells, or enter it later as part of running an analysis.
Use correction If available and selected, specifies a multiplication factor for each
factor standard. Can be used to correct for standard properties (percent
purity, water content, etc.) or preparation steps that affect the
measured result such as a dilution.
When this option is selected, two columns are added to the standards
table: Correction Factor and Corrected Concentration. In the
Correction Factor column, enter the desired factor for each standard
in the table. The values in the Corrected Concentration column are
used to create the calibration curve. If a report is generated that
includes information about the standards, the report will also include
the correction factors and corrected concentrations.
Standards Table Contains concentration and other information about the standards.
To add information from a .csv (comma separated values) file or .tsv
(tab separated values) file, click Import Standards. To save the
information in a .csv file or .tsv file, click Export Standards.
Related Topics
Working With Standard Curves
Feature Description
Mode Specifies Scan or Fixed data acquisition.
Data mode Y-axis format for acquired data.
Start wavelength and End Starting and ending values of the wavelength range for acquiring
wavelength (Scan only) data.
Bandwidth A larger spectral bandwidth increases the amount of energy that
passes through the sample. Increasing the bandwidth can
improve the signal-to-noise ratio; however, a larger spectral
bandwidth decreases the ability to resolve closely spaced spectral
features.
Integration time How long the system acquires and averages data at each data
interval (for scanning measurements) or at each measured
wavelength (for fixed-wavelength measurements). Increasing the
integration time improves the signal-to-noise ratio but reduces
the scan speed.
Data interval (Scan only) Difference in wavelength between two consecutive data points.
Scan speed (Scan only) Wavelength range covered per unit time. Varies inversely with
integration time. Increases as the data interval increases.
Estimated time (Scan Estimated duration of data acquisition. Increases as integration
only) time increases and as the data interval decreases. Varies inversely
with scan speed.
Table of wavelengths Shows the analysis wavelength(s) entered on the Measurement
(Fixed only) tab.
Related Topics
Type Tab for Pierce BCA
The status of accessories can be monitored during measurements. See Instrument Status
Monitors.
Related Topics
Type Tab for Pierce BCA
Feature Description
Number of samples Number of samples in the analysis.
Base name Beginning of the name for sample spectra, to which a sequential
number is appended. For example, the names of three sample spectra
using the base name “dye” would be dye1, dye2 and dye3.
Feature Description
Sample averaging Whether and how to average concentration values from multiple
measurements of the same sample or from multiple samples. To
average multiple measurements of the same sample, select Replicate
and set Number of replicates to the total number of measurements.
To average values from two or three samples, select Duplicate or
Triplicate, respectively.
Use sample If available and selected, specifies a multiplication factor for each
correction factor sample result. Can be used to correct for sample properties and
preparation steps such as a sample dilution that affects the measured
result. Enter the desired factor for each sample in the table.
Use weight/volume If available for the current unit, corrects sample concentrations using
correction the entered target weight and volume:
If the table contains just one sample and data is acquired with an
accessory that automatically changes samples, sample names are
incremented.
Related Topics
Type Tab for Pierce BCA
Advanced calculations
For selecting predefined formulas for additional data processing, with the results appearing in
a calculations table at the bottom of the right pane. Options include basic math and statistics.
The formulas can be applied to specific samples and columns of data in the sample
measurements table or to selected rows and columns in the calculations table for the current
workbook or template. Calculations are applied to all subsequently acquired data. Custom
calculations are saved with the workbook or template. Available for all applications except
Rate.
To delete the contents of a cell in the calculations table, right-click the cell and choose Delete.
1. In the calculations table, double-click the cell where you want the calculation result to
appear.
2. In the Cell Properties box, choose Calculation.
3. Select an Equation in the list.
The available equations are fixed.
4. Set Source to Samples table.
5. Select a sample measurement in the Measurements list.
6. Specify the sample rows to include by entering their ID numbers in the Range/Reference
box.
Use a colon to specify a range (e.g., 2:5 or 1:N) or commas for individual samples (e.g.,
1,3,5).
7. To select the background color for the cell that displays the calculated result, click
Background, choose a color and click OK.
8. To display the cell contents in bold text, click Bold Text.
9. Choose OK.
The calculated result appears in the selected cell.
1. In the calculations table, double-click the cell where you want the calculation result to
appear.
2. In the Cell Properties box, choose Calculation.
3. Select an Equation in the list.
4. Set Source to Calculations Table.
5. Specify the table cells to include by entering their IDs in the Range/Reference box.
If Equation is set to Std. Deviation, Mean, or %RSD, use a colon to specify a range (e.g.,
A2:A5 or A1:AN) or commas for individual cells (e.g., A1,A3,A5).
If Equation is set to Addition, Subtraction, Multiplication, or Division, specify one cell
location in each Reference box. (e.g., A1).
If Equation is set to a Factor, use the first (Reference) box to specify a cell location and
the second (Factor) box to enter the factor.
6. To select the background color for the cell that displays the calculated result, click
Background, choose a color and click OK.
7. To display the cell contents in bold text, click Bold Text.
8. Choose OK.
The calculated result appears in the selected cell.
Related Topics
Overview
Application Settings
Data Display
Note Standards diluted from a stock standard must cover the expected range of the
unknown samples. Protein concentrations are not extrapolated beyond measured
standard concentrations.
Procedure
Note To start an analysis immediately after viewing or changing the template settings,
click the Measure action button instead and skip to step 3.
2. If standards need to be acquired, click Measure. If standards have already been acquired,
skip to step 7 to acquire sample data.
More:
If Standard averaging on the Standards tab in Settings was set to Duplicate, “D” at the
end of a standard name indicates the second measurement to be made of the standard. If
Sample averaging was set to Triplicate, “D” and “T” at the end of standard names indicate
the second and third measurements to be made, respectively.
To enter previously saved standards information, use Import Standards.
If Calculate from weight/volume was selected on the Standards tab, enter the weight and
volume for each standard in the table.
To remove a standard, select it, right-click the table and choose Clear Standard. To
remove all the standards, choose Clear Table.
To save standards information for later use, use Export Standards.
5. Click Continue.
6. Follow the instructions that appear, installing the specified standards.
The Standard Curve(s) tab displays the resulting standard curve (or curves). Specify the
standards to use for the curve (or curves) by selecting Yes or No in the Use column in the
table.
More:
The table contains the columns of information specified on the Configuration tab in
Reports. Examples include sample identification, user name, and the results of replicate,
duplicate or triplicate measurements and their standard deviation.
If Standard averaging was set to Duplicate, “D” at the end of a standard name indicates
the second measurement. If Standard averaging was set to Triplicate, “D” and “T”
indicate the second and third measurements, respectively.
To permanently remove an acquired standard from the analysis, right-click it and choose
Remove. Its information is crossed out but not removed from the table.
The Data tab displays the acquired data (a fixed data point or scan spectrum) for the
standard selected in the table. To access commands for customizing the display, including
adding annotation, right-click the data. See Data Display for more information.
7. Click Measure.
More:
If Sample averaging on the Samples tab in Settings was set to Duplicate, “D” at the end of
a sample name indicates the second measurement to be made of the sample. If Sample
averaging was set to Triplicate, “D” and “T” at the end of sample names indicate the
second and third measurements to be made, respectively.
To enter previously saved samples information, use Load Samples. To save the samples
information, use Save Samples.
10. If only one sample will be measured, install it.
11. Click Continue.
12. Follow any instructions that appear, such as to install a specified sample.
The sample results table contains the columns of information specified on the
Configuration tab in Reports. If Sample averaging was set to Duplicate, “D” at the end of
a sample name indicates the second measurement. If Sample averaging was set to
Triplicate, “D” and “T” indicate the second and third measurements, respectively.
The Standard Curve(s) tab shows graphically the relationship between the standard curve,
measured spectral intensity and calculated concentration for the selected sample: A
horizontal line connects the sample spectral intensity value on the Y-axis to the standard
curve. A vertical line connects that point to the sample concentration value on the X-axis:
The Run Chart tab plots the concentration of the measured component versus sample
number. If Use concentration limits was selected on the Samples tab, horizontal limit
lines show whether the concentrations are within the specified limits:
To measure a sample again, right-click its row in the results table and choose Remeasure.
After the remeasurement, the previous information for the sample is crossed out (but not
removed from the table).
To copy the data on the Standard Curve(s) tab or Run Chart tab, right-click the plot and
choose Copy to Clipboard.
Related Topics
Overview
Application Settings
Data Display
Protein Bradford
Contents
• Overview
• Application Settings
• Unique Screen Features
• Making Protein Bradford Assay Measurements
Overview
The Protein Bradford assay measures absorbance at
595 nm; it determines concentration for either standard or
micro sample concentrations.
This assay is commonly used for determining protein concentration. It is often used for more
dilute protein solutions where lower detection sensitivity is needed and/or in the presence of
components that also have significant UV (280 nm) absorbance. Like the other colorimetric
assays, the Bradford assay requires generating a standard curve before measuring sample
proteins.
The Bradford procedure uses the protein-induced absorbance shift of Coomassie Blue dye to
595 nm to measure protein concentration. The bound protein-dye complex is measured at
595 nm and normalized at 750 nm. A single stabilized reagent mixture containing Coomassie
Blue dye, alcohol, and surfactant in kit form is available from numerous manufacturers.
Protein standards (BSA) for generating a standard curve may also be provided by the Bradford
assay manufacturer.
Related Topics
Application Settings
Data Display
Application Settings
To set data acquisition parameters for a Protein Bradford
workbook, click Settings.
Related Topics
Overview
Data Display
Feature Description
Name Application name.
Description (optional) Description of the template.
Standard curve Type of quantitative analysis used.
Pathlength Distance the light travels through the sample.
Related Topics
Measurement Tab for Protein Bradford
Feature Description
Analysis wavelength Wavelength(s) to use for the analysis.
Correction Specifies bichromatic normalization of the absorbance data.
• Single point. Enter a wavelength value to define the endpoint for
the single point baseline.
This option generates a baseline correction for each sample
measurement by drawing a straight line through the specified
baseline point and calculating the absorbance value on the drawn
baseline at the specified analytical wavelength. The software
subtracts the calculated baseline absorbance value from the
measured absorbance value and reports the corrected absorbance
• Sloping baseline. Enter two wavelength values to define the
endpoints for the sloping baseline.
This option generates a baseline correction for each sample
measurement by drawing a line between the two specified
baseline points and calculating the absorbance value on the
drawn baseline at the specified analytical wavelength. The
software subtracts the calculated baseline absorbance value from
the measured absorbance value and reports the corrected
absorbance.
• None. Uses uncorrected data. Without baseline correction,
spectra may be offset from the baseline. If this offset is significant,
the calculated protein concentration may be higher than the true
value.
Component name Component to quantify.
Feature Description
Unit Component concentration unit. The available units depend on the
Formulas & Units tab settings in Options.
Calculate additional For selecting or defining formulas for additional data processing, with
results the results appearing in the analysis report. The formulas are written
in a form similar to that used in a command script language, with
constants, mathematical functions, etc. All functions are not
case-sensitive. Spaces are not allowed between a function name and
“(”. After adding formulas, edit their information in the formula table
as desired. To delete a table row, right-click it and choose Delete Row.
Some information in the table cannot be modified or deleted.
Instructions:
1. Click Select.
2. For each desired formula, select it in the list and click Add.
The available formulas depend on the Formulas & Units tab
settings in Options.
3. Click Close.
To define a formula
1. Click Build.
2. To enter a formula in its entirety, set Equation type to User
defined and type the equation. To use a provided formula, select
it and enter values for the variables.
The available formulas depend on the Formulas & Units tab
settings.
3. Click OK.
4. For selected provided formulas, enter a name and concentration
unit in the table.
The available units depend on the Formulas & Units tab settings.
Related Topics
Type Tab for Protein Bradford
Feature Description
Curve fit type Type of equation used to create the standard curve from standard
concentration values.
Standard averaging Whether and how to average concentration values from multiple
measurements of the same standard or from multiple standards
prepared under the same conditions. To average multiple
measurements of the same standard, select Replicate and set Number
of replicates to the total number of measurements. To average values
from two or three similar standards, select Duplicate or Triplicate,
respectively.
Minimum r2 The r2 value indicates how well the standard curve fits the standard
data points, with 1.0 a perfect fit. If Minimum r2 is selected, samples
can be quantified only after that value is achieved for the standard
curve.
Calculate from If available for the current unit, calculates concentration when the
weight/volume weight and volume of material used to prepare each standard is
known. Enter the weight and volume for each standard in the
appropriate table cells, or enter it later as part of running an analysis.
Use correction If available and selected, specifies a multiplication factor for each
factor standard. Can be used to correct for standard properties (percent
purity, water content, etc.) or preparation steps that affect the
measured result such as a dilution.
When this option is selected, two columns are added to the standards
table: Correction Factor and Corrected Concentration. In the
Correction Factor column, enter the desired factor for each standard
in the table. The values in the Corrected Concentration column are
used to create the calibration curve. If a report is generated that
includes information about the standards, the report will also include
the correction factors and corrected concentrations.
Standards Table Contains concentration and other information about the standards.
To add information from a .csv (comma separated values) file or .tsv
(tab separated values) file, click Import Standards. To save the
information in a .csv file or .tsv file, click Export Standards.
Related Topics
Working With Standard Curves
Feature Description
Mode Specifies Scan or Fixed data acquisition.
Data mode Y-axis format for acquired data.
Start wavelength and End Starting and ending values of the wavelength range for acquiring
wavelength (Scan only) data.
Bandwidth A larger spectral bandwidth increases the amount of energy that
passes through the sample. Increasing the bandwidth can
improve the signal-to-noise ratio; however, a larger spectral
bandwidth decreases the ability to resolve closely spaced spectral
features.
Integration time How long the system acquires and averages data at each data
interval (for scanning measurements) or at each measured
wavelength (for fixed-wavelength measurements). Increasing the
integration time improves the signal-to-noise ratio but reduces
the scan speed.
Data interval (Scan only) Difference in wavelength between two consecutive data points.
Scan speed (Scan only) Wavelength range covered per unit time. Varies inversely with
integration time. Increases as the data interval increases.
Estimated time (Scan Estimated duration of data acquisition. Increases as integration
only) time increases and as the data interval decreases. Varies inversely
with scan speed.
Table of wavelengths Shows the analysis wavelength(s) entered on the Measurement
(Fixed only) tab.
Related Topics
Type Tab for Protein Bradford
The status of accessories can be monitored during measurements. See Instrument Status
Monitors.
Related Topics
Type Tab for Protein Bradford
Feature Description
Number of samples Number of samples in the analysis.
Base name Beginning of the name for sample spectra, to which a sequential
number is appended. For example, the names of three sample spectra
using the base name “dye” would be dye1, dye2 and dye3.
Feature Description
Sample averaging Whether and how to average concentration values from multiple
measurements of the same sample or from multiple samples. To
average multiple measurements of the same sample, select Replicate
and set Number of replicates to the total number of measurements.
To average values from two or three samples, select Duplicate or
Triplicate, respectively.
Use sample If available and selected, specifies a multiplication factor for each
correction factor sample result. Can be used to correct for sample properties and
preparation steps such as a sample dilution that affects the measured
result. Enter the desired factor for each sample in the table.
Use weight/volume If available for the current unit, corrects sample concentrations using
correction the entered target weight and volume:
If the table contains just one sample and data is acquired with an
accessory that automatically changes samples, sample names are
incremented.
Related Topics
Type Tab for Protein Bradford
Advanced calculations
For selecting predefined formulas for additional data processing, with the results appearing in
a calculations table at the bottom of the right pane. Options include basic math and statistics.
The formulas can be applied to specific samples and columns of data in the sample
measurements table or to selected rows and columns in the calculations table for the current
workbook or template. Calculations are applied to all subsequently acquired data. Custom
calculations are saved with the workbook or template. Available for all applications except
Rate.
To delete the contents of a cell in the calculations table, right-click the cell and choose Delete.
1. In the calculations table, double-click the cell where you want the calculation result to
appear.
2. In the Cell Properties box, choose Calculation.
3. Select an Equation in the list.
The available equations are fixed.
4. Set Source to Samples table.
5. Select a sample measurement in the Measurements list.
6. Specify the sample rows to include by entering their ID numbers in the Range/Reference
box.
Use a colon to specify a range (e.g., 2:5 or 1:N) or commas for individual samples (e.g.,
1,3,5).
7. To select the background color for the cell that displays the calculated result, click
Background, choose a color and click OK.
8. To display the cell contents in bold text, click Bold Text.
9. Choose OK.
The calculated result appears in the selected cell.
1. In the calculations table, double-click the cell where you want the calculation result to
appear.
2. In the Cell Properties box, choose Calculation.
3. Select an Equation in the list.
4. Set Source to Calculations Table.
5. Specify the table cells to include by entering their IDs in the Range/Reference box.
If Equation is set to Std. Deviation, Mean, or %RSD, use a colon to specify a range (e.g.,
A2:A5 or A1:AN) or commas for individual cells (e.g., A1,A3,A5).
If Equation is set to Addition, Subtraction, Multiplication, or Division, specify one cell
location in each Reference box. (e.g., A1).
If Equation is set to a Factor, use the first (Reference) box to specify a cell location and
the second (Factor) box to enter the factor.
6. To select the background color for the cell that displays the calculated result, click
Background, choose a color and click OK.
7. To display the cell contents in bold text, click Bold Text.
8. Choose OK.
The calculated result appears in the selected cell.
Related Topics
Overview
Application Settings
Data Display
Note To start an analysis immediately after viewing or changing the template settings,
click the Measure action button instead and skip to step 3.
2. If standards need to be acquired, click Measure. If standards have already been acquired,
skip to step 7 to acquire sample data.
More:
If Standard averaging on the Standards tab in Settings was set to Duplicate, “D” at the
end of a standard name indicates the second measurement to be made of the standard. If
Sample averaging was set to Triplicate, “D” and “T” at the end of standard names indicate
the second and third measurements to be made, respectively.
To enter previously saved standards information, use Import Standards.
If Calculate from weight/volume was selected on the Standards tab, enter the weight and
volume for each standard in the table.
To remove a standard, select it, right-click the table and choose Clear Standard. To
remove all the standards, choose Clear Table.
To save standards information for later use, use Export Standards.
5. Click Continue.
6. Follow the instructions that appear, installing the specified standards.
The Standard Curve(s) tab displays the resulting standard curve (or curves). Specify the
standards to use for the curve (or curves) by selecting Yes or No in the Use column in the
table.
More:
The table contains the columns of information specified on the Configuration tab in
Reports. Examples include sample identification, user name, and the results of replicate,
duplicate or triplicate measurements and their standard deviation.
If Standard averaging was set to Duplicate, “D” at the end of a standard name indicates
the second measurement. If Standard averaging was set to Triplicate, “D” and “T”
indicate the second and third measurements, respectively.
To permanently remove an acquired standard from the analysis, right-click it and choose
Remove. Its information is crossed out but not removed from the table.
The Data tab displays the acquired data (a fixed data point or scan spectrum) for the
standard selected in the table. To access commands for customizing the display, including
adding annotation, right-click the data. See Data Display for more information.
7. Click Measure.
More:
If Sample averaging on the Samples tab in Settings was set to Duplicate, “D” at the end of
a sample name indicates the second measurement to be made of the sample. If Sample
averaging was set to Triplicate, “D” and “T” at the end of sample names indicate the
second and third measurements to be made, respectively.
To enter previously saved samples information, use Load Samples. To save the samples
information, use Save Samples.
10. If only one sample will be measured, install it.
11. Click Continue.
12. Follow any instructions that appear, such as to install a specified sample.
The sample results table contains the columns of information specified on the
Configuration tab in Reports. If Sample averaging was set to Duplicate, “D” at the end of
a sample name indicates the second measurement. If Sample averaging was set to
Triplicate, “D” and “T” indicate the second and third measurements, respectively.
The Standard Curve(s) tab shows graphically the relationship between the standard curve,
measured spectral intensity and calculated concentration for the selected sample: A
horizontal line connects the sample spectral intensity value on the Y-axis to the standard
curve. A vertical line connects that point to the sample concentration value on the X-axis:
The Run Chart tab plots the concentration of the measured component versus sample
number. If Use concentration limits was selected on the Samples tab, horizontal limit
lines show whether the concentrations are within the specified limits:
To measure a sample again, right-click its row in the results table and choose Remeasure.
After the remeasurement, the previous information for the sample is crossed out (but not
removed from the table).
To copy the data on the Standard Curve(s) tab or Run Chart tab, right-click the plot and
choose Copy to Clipboard.
Related Topics
Overview
Application Settings
Data Display
Overview
The Pierce Modified Lowry assay measures absorbance at
750 nm to determine concentration.
This assay is an alternative method for determining protein concentration based on the widely
used and cited Lowry procedure for protein quantitation. Like the other colorimetric assays,
the Lowry assay requires generating a standard curve before measuring sample proteins.
The Pierce Modified Lowry procedure involves reaction of protein with cupric sulfate in
alkaline solution, resulting in formation of tetradentate copper-protein complexes. The
Folin-Ciocalteu Reagent is effectively reduced in proportion to the chelated
copper-complexes, resulting in a water-soluble blue product that is measured at 650 nm and
normalized at 405 nm. Preformulated reagents, used in the assay, are available in kit form
from numerous manufacturers.
Protein standards (BSA) for generating a standard curve may also be provided by the
manufacturer for the Lowry assay.
Related Topics
Application Settings
Data Display
Application Settings
To set data acquisition parameters for a Pierce Modified
Lowry workbook, click Settings.
Related Topics
Overview
Data Display
Feature Description
Name Application name.
Description (optional) Description of the template.
Standard curve Type of quantitative analysis used.
Pathlength Distance the light travels through the sample.
Related Topics
Measurement Tab for Pierce Modified Lowry
Feature Description
Analysis wavelength Wavelength(s) to use for the analysis.
Correction Specifies bichromatic normalization of the absorbance data.
• Single point. Enter a wavelength value to define the endpoint for
the single point baseline.
This option generates a baseline correction for each sample
measurement by drawing a straight line through the specified
baseline point and calculating the absorbance value on the drawn
baseline at the specified analytical wavelength. The software
subtracts the calculated baseline absorbance value from the
measured absorbance value and reports the corrected absorbance
• Sloping baseline. Enter two wavelength values to define the
endpoints for the sloping baseline.
This option generates a baseline correction for each sample
measurement by drawing a line between the two specified
baseline points and calculating the absorbance value on the
drawn baseline at the specified analytical wavelength. The
software subtracts the calculated baseline absorbance value from
the measured absorbance value and reports the corrected
absorbance.
• None. Uses uncorrected data. Without baseline correction,
spectra may be offset from the baseline. If this offset is significant,
the calculated protein concentration may be higher than the true
value.
Component name Component to quantify.
Feature Description
Unit Component concentration unit. The available units depend on the
Formulas & Units tab settings in Options.
Calculate additional For selecting or defining formulas for additional data processing, with
results the results appearing in the analysis report. The formulas are written
in a form similar to that used in a command script language, with
constants, mathematical functions, etc. All functions are not
case-sensitive. Spaces are not allowed between a function name and
“(”. After adding formulas, edit their information in the formula table
as desired. To delete a table row, right-click it and choose Delete Row.
Some information in the table cannot be modified or deleted.
Instructions:
1. Click Select.
2. For each desired formula, select it in the list and click Add.
The available formulas depend on the Formulas & Units tab
settings in Options.
3. Click Close.
To define a formula
1. Click Build.
2. To enter a formula in its entirety, set Equation type to User
defined and type the equation. To use a provided formula, select
it and enter values for the variables.
The available formulas depend on the Formulas & Units tab
settings.
3. Click OK.
4. For selected provided formulas, enter a name and concentration
unit in the table.
The available units depend on the Formulas & Units tab settings.
Related Topics
Type Tab for Pierce Modified Lowry
Feature Description
Curve fit type Type of equation used to create the standard curve from standard
concentration values.
Standard averaging Whether and how to average concentration values from multiple
measurements of the same standard or from multiple standards
prepared under the same conditions. To average multiple
measurements of the same standard, select Replicate and set Number
of replicates to the total number of measurements. To average values
from two or three similar standards, select Duplicate or Triplicate,
respectively.
Minimum r2 The r2 value indicates how well the standard curve fits the standard
data points, with 1.0 a perfect fit. If Minimum r2 is selected, samples
can be quantified only after that value is achieved for the standard
curve.
Calculate from If available for the current unit, calculates concentration when the
weight/volume weight and volume of material used to prepare each standard is
known. Enter the weight and volume for each standard in the
appropriate table cells, or enter it later as part of running an analysis.
Use correction If available and selected, specifies a multiplication factor for each
factor standard. Can be used to correct for standard properties (percent
purity, water content, etc.) or preparation steps that affect the
measured result such as a dilution.
When this option is selected, two columns are added to the standards
table: Correction Factor and Corrected Concentration. In the
Correction Factor column, enter the desired factor for each standard
in the table. The values in the Corrected Concentration column are
used to create the calibration curve. If a report is generated that
includes information about the standards, the report will also include
the correction factors and corrected concentrations.
Standards Table Contains concentration and other information about the standards.
To add information from a .csv (comma separated values) file or .tsv
(tab separated values) file, click Import Standards. To save the
information in a .csv file or .tsv file, click Export Standards.
Related Topics
Working With Standard Curves
Feature Description
Mode Specifies Scan or Fixed data acquisition.
Data mode Y-axis format for acquired data.
Start wavelength and End Starting and ending values of the wavelength range for acquiring
wavelength (Scan only) data.
Bandwidth A larger spectral bandwidth increases the amount of energy that
passes through the sample. Increasing the bandwidth can
improve the signal-to-noise ratio; however, a larger spectral
bandwidth decreases the ability to resolve closely spaced spectral
features.
Integration time How long the system acquires and averages data at each data
interval (for scanning measurements) or at each measured
wavelength (for fixed-wavelength measurements). Increasing the
integration time improves the signal-to-noise ratio but reduces
the scan speed.
Data interval (Scan only) Difference in wavelength between two consecutive data points.
Scan speed (Scan only) Wavelength range covered per unit time. Varies inversely with
integration time. Increases as the data interval increases.
Estimated time (Scan Estimated duration of data acquisition. Increases as integration
only) time increases and as the data interval decreases. Varies inversely
with scan speed.
Table of wavelengths Shows the analysis wavelength(s) entered on the Measurement
(Fixed only) tab.
Related Topics
Type Tab for Pierce Modified Lowry
The status of accessories can be monitored during measurements. See Instrument Status
Monitors.
Related Topics
Type Tab for Pierce Modified Lowry
Feature Description
Number of samples Number of samples in the analysis.
Base name Beginning of the name for sample spectra, to which a sequential
number is appended. For example, the names of three sample spectra
using the base name “dye” would be dye1, dye2 and dye3.
Feature Description
Sample averaging Whether and how to average concentration values from multiple
measurements of the same sample or from multiple samples. To
average multiple measurements of the same sample, select Replicate
and set Number of replicates to the total number of measurements.
To average values from two or three samples, select Duplicate or
Triplicate, respectively.
Use sample If available and selected, specifies a multiplication factor for each
correction factor sample result. Can be used to correct for sample properties and
preparation steps such as a sample dilution that affects the measured
result. Enter the desired factor for each sample in the table.
Use weight/volume If available for the current unit, corrects sample concentrations using
correction the entered target weight and volume:
If the table contains just one sample and data is acquired with an
accessory that automatically changes samples, sample names are
incremented.
Related Topics
Type Tab for Pierce Modified Lowry
Advanced calculations
For selecting predefined formulas for additional data processing, with the results appearing in
a calculations table at the bottom of the right pane. Options include basic math and statistics.
The formulas can be applied to specific samples and columns of data in the sample
measurements table or to selected rows and columns in the calculations table for the current
workbook or template. Calculations are applied to all subsequently acquired data. Custom
calculations are saved with the workbook or template. Available for all applications except
Rate.
To delete the contents of a cell in the calculations table, right-click the cell and choose Delete.
1. In the calculations table, double-click the cell where you want the calculation result to
appear.
2. In the Cell Properties box, choose Calculation.
3. Select an Equation in the list.
The available equations are fixed.
4. Set Source to Samples table.
5. Select a sample measurement in the Measurements list.
6. Specify the sample rows to include by entering their ID numbers in the Range/Reference
box.
Use a colon to specify a range (e.g., 2:5 or 1:N) or commas for individual samples (e.g.,
1,3,5).
7. To select the background color for the cell that displays the calculated result, click
Background, choose a color and click OK.
8. To display the cell contents in bold text, click Bold Text.
9. Choose OK.
The calculated result appears in the selected cell.
1. In the calculations table, double-click the cell where you want the calculation result to
appear.
2. In the Cell Properties box, choose Calculation.
3. Select an Equation in the list.
4. Set Source to Calculations Table.
5. Specify the table cells to include by entering their IDs in the Range/Reference box.
If Equation is set to Std. Deviation, Mean, or %RSD, use a colon to specify a range (e.g.,
A2:A5 or A1:AN) or commas for individual cells (e.g., A1,A3,A5).
If Equation is set to Addition, Subtraction, Multiplication, or Division, specify one cell
location in each Reference box. (e.g., A1).
If Equation is set to a Factor, use the first (Reference) box to specify a cell location and
the second (Factor) box to enter the factor.
6. To select the background color for the cell that displays the calculated result, click
Background, choose a color and click OK.
7. To display the cell contents in bold text, click Bold Text.
8. Choose OK.
The calculated result appears in the selected cell.
Related Topics
Overview
Application Settings
Data Display
Note To start an analysis immediately after viewing or changing the template settings,
click the Measure action button instead and skip to step 3.
2. If standards need to be acquired, click Measure. If standards have already been acquired,
skip to step 7 to acquire sample data.
More:
If Standard averaging on the Standards tab in Settings was set to Duplicate, “D” at the
end of a standard name indicates the second measurement to be made of the standard. If
Sample averaging was set to Triplicate, “D” and “T” at the end of standard names indicate
the second and third measurements to be made, respectively.
To enter previously saved standards information, use Import Standards.
If Calculate from weight/volume was selected on the Standards tab, enter the weight and
volume for each standard in the table.
To remove a standard, select it, right-click the table and choose Clear Standard. To
remove all the standards, choose Clear Table.
To save standards information for later use, use Export Standards.
5. Click Continue.
6. Follow the instructions that appear, installing the specified standards.
The Standard Curve(s) tab displays the resulting standard curve (or curves). Specify the
standards to use for the curve (or curves) by selecting Yes or No in the Use column in the
table.
More:
The table contains the columns of information specified on the Configuration tab in
Reports. Examples include sample identification, user name, and the results of replicate,
duplicate or triplicate measurements and their standard deviation.
If Standard averaging was set to Duplicate, “D” at the end of a standard name indicates
the second measurement. If Standard averaging was set to Triplicate, “D” and “T”
indicate the second and third measurements, respectively.
To permanently remove an acquired standard from the analysis, right-click it and choose
Remove. Its information is crossed out but not removed from the table.
The Data tab displays the acquired data (a fixed data point or scan spectrum) for the
standard selected in the table. To access commands for customizing the display, including
adding annotation, right-click the data. See Data Display for more information.
7. Click Measure.
More:
If Sample averaging on the Samples tab in Settings was set to Duplicate, “D” at the end of
a sample name indicates the second measurement to be made of the sample. If Sample
averaging was set to Triplicate, “D” and “T” at the end of sample names indicate the
second and third measurements to be made, respectively.
To enter previously saved samples information, use Load Samples. To save the samples
information, use Save Samples.
10. If only one sample will be measured, install it.
11. Click Continue.
12. Follow any instructions that appear, such as to install a specified sample.
The sample results table contains the columns of information specified on the
Configuration tab in Reports. If Sample averaging was set to Duplicate, “D” at the end of
a sample name indicates the second measurement. If Sample averaging was set to
Triplicate, “D” and “T” indicate the second and third measurements, respectively.
The Standard Curve(s) tab shows graphically the relationship between the standard curve,
measured spectral intensity and calculated concentration for the selected sample: A
horizontal line connects the sample spectral intensity value on the Y-axis to the standard
curve. A vertical line connects that point to the sample concentration value on the X-axis:
The Run Chart tab plots the concentration of the measured component versus sample
number. If Use concentration limits was selected on the Samples tab, horizontal limit
lines show whether the concentrations are within the specified limits:
To measure a sample again, right-click its row in the results table and choose Remeasure.
After the remeasurement, the previous information for the sample is crossed out (but not
removed from the table).
To copy the data on the Standard Curve(s) tab or Run Chart tab, right-click the plot and
choose Copy to Clipboard.
Related Topics
Overview
Application Settings
Data Display
Overview
The Thermo Scientific Pierce 660 nm Protein Assay
reagent is a ready-to-use formulation that offers rapid,
accurate and reproducible colorimetric detection of
minute amounts of protein in solution. The reagent is
ideal for measuring total protein concentration in samples
containing reducing agents and/or detergents.
The proprietary dye-metal complex binds to protein in acidic conditions, causing a shift in
the dye’s absorption maximum, which is measured at 660 nm. The complex is reddish-brown
and turns green upon protein binding. This color change is produced by deprotonation of the
dye at low pH facilitated by interactions with positively charged amino acid groups in the
protein. Consequently, the complex interacts primarily with basic residues in the protein, such
as histidine, arginine and lysine, and to a lesser extent tyrosine, tryptophan and phenylalanine.
Protein standards (BSA) for generating a standard curve may also be provided by the Pierce
660 nm assay manufacturer.
Related Topics
Application Settings
Data Display
Application Settings
To set data acquisition parameters for a Pierce 660 nm
Protein Assay workbook, click Settings.
Related Topics
Overview
Data Display
Feature Description
Name Application name.
Description (optional) Description of the template.
Standard curve Type of quantitative analysis used.
Pathlength Distance the light travels through the sample.
Related Topics
Measurement Tab for Pierce 660 nm Protein Assay
Feature Description
Analysis wavelength Wavelength(s) to use for the analysis.
Correction Specifies bichromatic normalization of the absorbance data.
• Single point. Enter a wavelength value to define the endpoint for
the single point baseline.
This option generates a baseline correction for each sample
measurement by drawing a straight line through the specified
baseline point and calculating the absorbance value on the drawn
baseline at the specified analytical wavelength. The software
subtracts the calculated baseline absorbance value from the
measured absorbance value and reports the corrected absorbance
• Sloping baseline. Enter two wavelength values to define the
endpoints for the sloping baseline.
This option generates a baseline correction for each sample
measurement by drawing a line between the two specified
baseline points and calculating the absorbance value on the
drawn baseline at the specified analytical wavelength. The
software subtracts the calculated baseline absorbance value from
the measured absorbance value and reports the corrected
absorbance.
• None. Uses uncorrected data. Without baseline correction,
spectra may be offset from the baseline. If this offset is significant,
the calculated protein concentration may be higher than the true
value.
Component name Component to quantify.
Feature Description
Unit Component concentration unit. The available units depend on the
Formulas & Units tab settings in Options.
Calculate additional For selecting or defining formulas for additional data processing, with
results the results appearing in the analysis report. The formulas are written
in a form similar to that used in a command script language, with
constants, mathematical functions, etc. All functions are not
case-sensitive. Spaces are not allowed between a function name and
“(”. After adding formulas, edit their information in the formula table
as desired. To delete a table row, right-click it and choose Delete Row.
Some information in the table cannot be modified or deleted.
Instructions:
1. Click Select.
2. For each desired formula, select it in the list and click Add.
The available formulas depend on the Formulas & Units tab
settings in Options.
3. Click Close.
To define a formula
1. Click Build.
2. To enter a formula in its entirety, set Equation type to User
defined and type the equation. To use a provided formula, select
it and enter values for the variables.
The available formulas depend on the Formulas & Units tab
settings.
3. Click OK.
4. For selected provided formulas, enter a name and concentration
unit in the table.
The available units depend on the Formulas & Units tab settings.
Related Topics
Type Tab for Pierce 660 nm Protein Assay
Feature Description
Curve fit type Type of equation used to create the standard curve from standard
concentration values.
Standard averaging Whether and how to average concentration values from multiple
measurements of the same standard or from multiple standards
prepared under the same conditions. To average multiple
measurements of the same standard, select Replicate and set Number
of replicates to the total number of measurements. To average values
from two or three similar standards, select Duplicate or Triplicate,
respectively.
Minimum r2 The r2 value indicates how well the standard curve fits the standard
data points, with 1.0 a perfect fit. If Minimum r2 is selected, samples
can be quantified only after that value is achieved for the standard
curve.
Calculate from If available for the current unit, calculates concentration when the
weight/volume weight and volume of material used to prepare each standard is
known. Enter the weight and volume for each standard in the
appropriate table cells, or enter it later as part of running an analysis.
Use correction If available and selected, specifies a multiplication factor for each
factor standard. Can be used to correct for standard properties (percent
purity, water content, etc.) or preparation steps that affect the
measured result such as a dilution.
When this option is selected, two columns are added to the standards
table: Correction Factor and Corrected Concentration. In the
Correction Factor column, enter the desired factor for each standard
in the table. The values in the Corrected Concentration column are
used to create the calibration curve. If a report is generated that
includes information about the standards, the report will also include
the correction factors and corrected concentrations.
Standards Table Contains concentration and other information about the standards.
To add information from a .csv (comma separated values) file or .tsv
(tab separated values) file, click Import Standards. To save the
information in a .csv file or .tsv file, click Export Standards.
Related Topics
Working With Standard Curves
Feature Description
Mode Specifies Scan or Fixed data acquisition.
Data mode Y-axis format for acquired data.
Start wavelength and End Starting and ending values of the wavelength range for acquiring
wavelength (Scan only) data.
Bandwidth A larger spectral bandwidth increases the amount of energy that
passes through the sample. Increasing the bandwidth can
improve the signal-to-noise ratio; however, a larger spectral
bandwidth decreases the ability to resolve closely spaced spectral
features.
Integration time How long the system acquires and averages data at each data
interval (for scanning measurements) or at each measured
wavelength (for fixed-wavelength measurements). Increasing the
integration time improves the signal-to-noise ratio but reduces
the scan speed.
Data interval (Scan only) Difference in wavelength between two consecutive data points.
Scan speed (Scan only) Wavelength range covered per unit time. Varies inversely with
integration time. Increases as the data interval increases.
Estimated time (Scan Estimated duration of data acquisition. Increases as integration
only) time increases and as the data interval decreases. Varies inversely
with scan speed.
Table of wavelengths Shows the analysis wavelength(s) entered on the Measurement
(Fixed only) tab.
Related Topics
Type Tab for Pierce 660 nm Protein Assay
The status of accessories can be monitored during measurements. See Instrument Status
Monitors.
Related Topics
Type Tab for Pierce 660 nm Protein Assay
Feature Description
Number of samples Number of samples in the analysis.
Base name Beginning of the name for sample spectra, to which a sequential
number is appended. For example, the names of three sample spectra
using the base name “dye” would be dye1, dye2 and dye3.
Feature Description
Sample averaging Whether and how to average concentration values from multiple
measurements of the same sample or from multiple samples. To
average multiple measurements of the same sample, select Replicate
and set Number of replicates to the total number of measurements.
To average values from two or three samples, select Duplicate or
Triplicate, respectively.
Use sample If available and selected, specifies a multiplication factor for each
correction factor sample result. Can be used to correct for sample properties and
preparation steps such as a sample dilution that affects the measured
result. Enter the desired factor for each sample in the table.
Use weight/volume If available for the current unit, corrects sample concentrations using
correction the entered target weight and volume:
If the table contains just one sample and data is acquired with an
accessory that automatically changes samples, sample names are
incremented.
Related Topics
Type Tab for Pierce 660 nm Protein Assay
Advanced calculations
For selecting predefined formulas for additional data processing, with the results appearing in
a calculations table at the bottom of the right pane. Options include basic math and statistics.
The formulas can be applied to specific samples and columns of data in the sample
measurements table or to selected rows and columns in the calculations table for the current
workbook or template. Calculations are applied to all subsequently acquired data. Custom
calculations are saved with the workbook or template. Available for all applications except
Rate.
To delete the contents of a cell in the calculations table, right-click the cell and choose Delete.
1. In the calculations table, double-click the cell where you want the calculation result to
appear.
2. In the Cell Properties box, choose Calculation.
3. Select an Equation in the list.
The available equations are fixed.
4. Set Source to Samples table.
5. Select a sample measurement in the Measurements list.
6. Specify the sample rows to include by entering their ID numbers in the Range/Reference
box.
Use a colon to specify a range (e.g., 2:5 or 1:N) or commas for individual samples (e.g.,
1,3,5).
7. To select the background color for the cell that displays the calculated result, click
Background, choose a color and click OK.
8. To display the cell contents in bold text, click Bold Text.
9. Choose OK.
The calculated result appears in the selected cell.
1. In the calculations table, double-click the cell where you want the calculation result to
appear.
2. In the Cell Properties box, choose Calculation.
3. Select an Equation in the list.
4. Set Source to Calculations Table.
5. Specify the table cells to include by entering their IDs in the Range/Reference box.
If Equation is set to Std. Deviation, Mean, or %RSD, use a colon to specify a range (e.g.,
A2:A5 or A1:AN) or commas for individual cells (e.g., A1,A3,A5).
If Equation is set to Addition, Subtraction, Multiplication, or Division, specify one cell
location in each Reference box. (e.g., A1).
If Equation is set to a Factor, use the first (Reference) box to specify a cell location and
the second (Factor) box to enter the factor.
6. To select the background color for the cell that displays the calculated result, click
Background, choose a color and click OK.
7. To display the cell contents in bold text, click Bold Text.
8. Choose OK.
The calculated result appears in the selected cell.
Related Topics
Overview
Application Settings
Data Display
1. Click Measure Pierce 660 nm Protein Assay in Pierce 660 nm Protein Assay.
Note To start an analysis immediately after viewing or changing the template settings,
click the Measure action button instead and skip to step 3.
2. If standards need to be acquired, click Measure. If standards have already been acquired,
skip to step 7 to acquire sample data.
More:
If Standard averaging on the Standards tab in Settings was set to Duplicate, “D” at the
end of a standard name indicates the second measurement to be made of the standard. If
Sample averaging was set to Triplicate, “D” and “T” at the end of standard names indicate
the second and third measurements to be made, respectively.
To enter previously saved standards information, use Import Standards.
If Calculate from weight/volume was selected on the Standards tab, enter the weight and
volume for each standard in the table.
To remove a standard, select it, right-click the table and choose Clear Standard. To
remove all the standards, choose Clear Table.
To save standards information for later use, use Export Standards.
5. Click Continue.
6. Follow the instructions that appear, installing the specified standards.
The Standard Curve(s) tab displays the resulting standard curve (or curves). Specify the
standards to use for the curve (or curves) by selecting Yes or No in the Use column in the
table.
More:
The table contains the columns of information specified on the Configuration tab in
Reports. Examples include sample identification, user name, and the results of replicate,
duplicate or triplicate measurements and their standard deviation.
If Standard averaging was set to Duplicate, “D” at the end of a standard name indicates
the second measurement. If Standard averaging was set to Triplicate, “D” and “T”
indicate the second and third measurements, respectively.
To permanently remove an acquired standard from the analysis, right-click it and choose
Remove. Its information is crossed out but not removed from the table.
The Data tab displays the acquired data (a fixed data point or scan spectrum) for the
standard selected in the table. To access commands for customizing the display, including
adding annotation, right-click the data. See Data Display for more information.
7. Click Measure.
More:
If Sample averaging on the Samples tab in Settings was set to Duplicate, “D” at the end of
a sample name indicates the second measurement to be made of the sample. If Sample
averaging was set to Triplicate, “D” and “T” at the end of sample names indicate the
second and third measurements to be made, respectively.
To enter previously saved samples information, use Load Samples. To save the samples
information, use Save Samples.
10. If only one sample will be measured, install it.
11. Click Continue.
12. Follow any instructions that appear, such as to install a specified sample.
The sample results table contains the columns of information specified on the
Configuration tab in Reports. If Sample averaging was set to Duplicate, “D” at the end of
a sample name indicates the second measurement. If Sample averaging was set to
Triplicate, “D” and “T” indicate the second and third measurements, respectively.
The Standard Curve(s) tab shows graphically the relationship between the standard curve,
measured spectral intensity and calculated concentration for the selected sample: A
horizontal line connects the sample spectral intensity value on the Y-axis to the standard
curve. A vertical line connects that point to the sample concentration value on the X-axis:
The Run Chart tab plots the concentration of the measured component versus sample
number. If Use concentration limits was selected on the Samples tab, horizontal limit
lines show whether the concentrations are within the specified limits:
To measure a sample again, right-click its row in the results table and choose Remeasure.
After the remeasurement, the previous information for the sample is crossed out (but not
removed from the table).
To copy the data on the Standard Curve(s) tab or Run Chart tab, right-click the plot and
choose Copy to Clipboard.
Related Topics
Overview
Application Settings
Data Display
Protein Biuret
Contents
• Overview
• Application Settings
• Unique Screen Features
• Making Protein Biuret Assay Measurements
Overview
The Protein Biuret application is similar to the Pierce
Modified Lowry application; however, it involves a single
incubation and requires more protein for the analysis. The
protein-dye complex that is formed in the assay has a deep
purple color and is measured at 545 nm. The calibration
curve for this assay is linear with standards that range in
concentration from 0.5 mg/ml to 10 mg/ml.
• The multipoint curve capability allows multiple replicates for each standard. Standards
can be run in any order; however, best practice dictates that the standards be measured
from the lowest concentration to the highest.
• Determine whether or not to use a standard when analyzing samples against the standard
curve.
• Establishing a new standard curve requires creating a new workbook.
• If a previously saved workbook is used, all concentration calculations for newly measured
samples will be based on the standard absorbance values saved in the workbook.
Related Topics
Application Settings
Data Display
Application Settings
To set data acquisition parameters for a Protein Biuret
workbook, click Settings.
Related Topics
Overview
Data Display
Feature Description
Name Application name.
Description (optional) Description of the template.
Standard curve Type of quantitative analysis used.
Pathlength Distance the light travels through the sample.
Related Topics
Measurement Tab for Protein Biuret
Feature Description
Analysis wavelength Wavelength(s) to use for the analysis.
Correction Specifies bichromatic normalization of the absorbance data.
• Single point. Enter a wavelength value to define the endpoint for
the single point baseline.
This option generates a baseline correction for each sample
measurement by drawing a straight line through the specified
baseline point and calculating the absorbance value on the drawn
baseline at the specified analytical wavelength. The software
subtracts the calculated baseline absorbance value from the
measured absorbance value and reports the corrected absorbance
• Sloping baseline. Enter two wavelength values to define the
endpoints for the sloping baseline.
This option generates a baseline correction for each sample
measurement by drawing a line between the two specified
baseline points and calculating the absorbance value on the
drawn baseline at the specified analytical wavelength. The
software subtracts the calculated baseline absorbance value from
the measured absorbance value and reports the corrected
absorbance.
• None. Uses uncorrected data. Without baseline correction,
spectra may be offset from the baseline. If this offset is significant,
the calculated protein concentration may be higher than the true
value.
Component name Component to quantify.
Feature Description
Unit Component concentration unit. The available units depend on the
Formulas & Units tab settings in Options.
Calculate additional For selecting or defining formulas for additional data processing, with
results the results appearing in the analysis report. The formulas are written
in a form similar to that used in a command script language, with
constants, mathematical functions, etc. All functions are not
case-sensitive. Spaces are not allowed between a function name and
“(”. After adding formulas, edit their information in the formula table
as desired. To delete a table row, right-click it and choose Delete Row.
Some information in the table cannot be modified or deleted.
Instructions:
1. Click Select.
2. For each desired formula, select it in the list and click Add.
The available formulas depend on the Formulas & Units tab
settings in Options.
3. Click Close.
To define a formula
1. Click Build.
2. To enter a formula in its entirety, set Equation type to User
defined and type the equation. To use a provided formula, select
it and enter values for the variables.
The available formulas depend on the Formulas & Units tab
settings.
3. Click OK.
4. For selected provided formulas, enter a name and concentration
unit in the table.
The available units depend on the Formulas & Units tab settings.
Related Topics
Type Tab for Protein Biuret
Feature Description
Curve fit type Type of equation used to create the standard curve from standard
concentration values.
Standard averaging Whether and how to average concentration values from multiple
measurements of the same standard or from multiple standards
prepared under the same conditions. To average multiple
measurements of the same standard, select Replicate and set Number
of replicates to the total number of measurements. To average values
from two or three similar standards, select Duplicate or Triplicate,
respectively.
Minimum r2 The r2 value indicates how well the standard curve fits the standard
data points, with 1.0 a perfect fit. If Minimum r2 is selected, samples
can be quantified only after that value is achieved for the standard
curve.
Calculate from If available for the current unit, calculates concentration when the
weight/volume weight and volume of material used to prepare each standard is
known. Enter the weight and volume for each standard in the
appropriate table cells, or enter it later as part of running an analysis.
Use correction If available and selected, specifies a multiplication factor for each
factor standard. Can be used to correct for standard properties (percent
purity, water content, etc.) or preparation steps that affect the
measured result such as a dilution.
When this option is selected, two columns are added to the standards
table: Correction Factor and Corrected Concentration. In the
Correction Factor column, enter the desired factor for each standard
in the table. The values in the Corrected Concentration column are
used to create the calibration curve. If a report is generated that
includes information about the standards, the report will also include
the correction factors and corrected concentrations.
Standards Table Contains concentration and other information about the standards.
To add information from a .csv (comma separated values) file or .tsv
(tab separated values) file, click Import Standards. To save the
information in a .csv file or .tsv file, click Export Standards.
Related Topics
Working With Standard Curves
Feature Description
Mode Specifies Scan or Fixed data acquisition.
Data mode Y-axis format for acquired data.
Start wavelength and End Starting and ending values of the wavelength range for acquiring
wavelength (Scan only) data.
Bandwidth A larger spectral bandwidth increases the amount of energy that
passes through the sample. Increasing the bandwidth can
improve the signal-to-noise ratio; however, a larger spectral
bandwidth decreases the ability to resolve closely spaced spectral
features.
Integration time How long the system acquires and averages data at each data
interval (for scanning measurements) or at each measured
wavelength (for fixed-wavelength measurements). Increasing the
integration time improves the signal-to-noise ratio but reduces
the scan speed.
Data interval (Scan only) Difference in wavelength between two consecutive data points.
Scan speed (Scan only) Wavelength range covered per unit time. Varies inversely with
integration time. Increases as the data interval increases.
Estimated time (Scan Estimated duration of data acquisition. Increases as integration
only) time increases and as the data interval decreases. Varies inversely
with scan speed.
Table of wavelengths Shows the analysis wavelength(s) entered on the Measurement
(Fixed only) tab.
Related Topics
Type Tab for Protein Biuret
The status of accessories can be monitored during measurements. See Instrument Status
Monitors.
Related Topics
Type Tab for Protein Biuret
Feature Description
Number of samples Number of samples in the analysis.
Base name Beginning of the name for sample spectra, to which a sequential
number is appended. For example, the names of three sample spectra
using the base name “dye” would be dye1, dye2 and dye3.
Feature Description
Sample averaging Whether and how to average concentration values from multiple
measurements of the same sample or from multiple samples. To
average multiple measurements of the same sample, select Replicate
and set Number of replicates to the total number of measurements.
To average values from two or three samples, select Duplicate or
Triplicate, respectively.
Use sample If available and selected, specifies a multiplication factor for each
correction factor sample result. Can be used to correct for sample properties and
preparation steps such as a sample dilution that affects the measured
result. Enter the desired factor for each sample in the table.
Use weight/volume If available for the current unit, corrects sample concentrations using
correction the entered target weight and volume:
If the table contains just one sample and data is acquired with an
accessory that automatically changes samples, sample names are
incremented.
Related Topics
Type Tab for Protein Biuret
Advanced calculations
For selecting predefined formulas for additional data processing, with the results appearing in
a calculations table at the bottom of the right pane. Options include basic math and statistics.
The formulas can be applied to specific samples and columns of data in the sample
measurements table or to selected rows and columns in the calculations table for the current
workbook or template. Calculations are applied to all subsequently acquired data. Custom
calculations are saved with the workbook or template. Available for all applications except
Rate.
To delete the contents of a cell in the calculations table, right-click the cell and choose Delete.
1. In the calculations table, double-click the cell where you want the calculation result to
appear.
2. In the Cell Properties box, choose Calculation.
3. Select an Equation in the list.
The available equations are fixed.
4. Set Source to Samples table.
5. Select a sample measurement in the Measurements list.
6. Specify the sample rows to include by entering their ID numbers in the Range/Reference
box.
Use a colon to specify a range (e.g., 2:5 or 1:N) or commas for individual samples (e.g.,
1,3,5).
7. To select the background color for the cell that displays the calculated result, click
Background, choose a color and click OK.
8. To display the cell contents in bold text, click Bold Text.
9. Choose OK.
The calculated result appears in the selected cell.
1. In the calculations table, double-click the cell where you want the calculation result to
appear.
2. In the Cell Properties box, choose Calculation.
3. Select an Equation in the list.
4. Set Source to Calculations Table.
5. Specify the table cells to include by entering their IDs in the Range/Reference box.
If Equation is set to Std. Deviation, Mean, or %RSD, use a colon to specify a range (e.g.,
A2:A5 or A1:AN) or commas for individual cells (e.g., A1,A3,A5).
If Equation is set to Addition, Subtraction, Multiplication, or Division, specify one cell
location in each Reference box. (e.g., A1).
If Equation is set to a Factor, use the first (Reference) box to specify a cell location and
the second (Factor) box to enter the factor.
6. To select the background color for the cell that displays the calculated result, click
Background, choose a color and click OK.
7. To display the cell contents in bold text, click Bold Text.
8. Choose OK.
The calculated result appears in the selected cell.
Related Topics
Overview
Application Settings
Data Display
Note To start an analysis immediately after viewing or changing the template settings,
click the Measure action button instead and skip to step 3.
2. If standards need to be acquired, click Measure. If standards have already been acquired,
skip to step 7 to acquire sample data.
More:
If Standard averaging on the Standards tab in Settings was set to Duplicate, “D” at the
end of a standard name indicates the second measurement to be made of the standard. If
Sample averaging was set to Triplicate, “D” and “T” at the end of standard names indicate
the second and third measurements to be made, respectively.
To enter previously saved standards information, use Import Standards.
If Calculate from weight/volume was selected on the Standards tab, enter the weight and
volume for each standard in the table.
To remove a standard, select it, right-click the table and choose Clear Standard. To
remove all the standards, choose Clear Table.
To save standards information for later use, use Export Standards.
5. Click Continue.
6. Follow the instructions that appear, installing the specified standards.
The Standard Curve(s) tab displays the resulting standard curve (or curves). Specify the
standards to use for the curve (or curves) by selecting Yes or No in the Use column in the
table.
More:
The table contains the columns of information specified on the Configuration tab in
Reports. Examples include sample identification, user name, and the results of replicate,
duplicate or triplicate measurements and their standard deviation.
If Standard averaging was set to Duplicate, “D” at the end of a standard name indicates
the second measurement. If Standard averaging was set to Triplicate, “D” and “T”
indicate the second and third measurements, respectively.
To permanently remove an acquired standard from the analysis, right-click it and choose
Remove. Its information is crossed out but not removed from the table.
The Data tab displays the acquired data (a fixed data point or scan spectrum) for the
standard selected in the table. To access commands for customizing the display, including
adding annotation, right-click the data. See Data Display for more information.
7. Click Measure.
More:
If Sample averaging on the Samples tab in Settings was set to Duplicate, “D” at the end of
a sample name indicates the second measurement to be made of the sample. If Sample
averaging was set to Triplicate, “D” and “T” at the end of sample names indicate the
second and third measurements to be made, respectively.
To enter previously saved samples information, use Load Samples. To save the samples
information, use Save Samples.
10. If only one sample will be measured, install it.
11. Click Continue.
12. Follow any instructions that appear, such as to install a specified sample.
The sample results table contains the columns of information specified on the
Configuration tab in Reports. If Sample averaging was set to Duplicate, “D” at the end of
a sample name indicates the second measurement. If Sample averaging was set to
Triplicate, “D” and “T” indicate the second and third measurements, respectively.
The Standard Curve(s) tab shows graphically the relationship between the standard curve,
measured spectral intensity and calculated concentration for the selected sample: A
horizontal line connects the sample spectral intensity value on the Y-axis to the standard
curve. A vertical line connects that point to the sample concentration value on the X-axis:
The Run Chart tab plots the concentration of the measured component versus sample
number. If Use concentration limits was selected on the Samples tab, horizontal limit
lines show whether the concentrations are within the specified limits:
To measure a sample again, right-click its row in the results table and choose Remeasure.
After the remeasurement, the previous information for the sample is crossed out (but not
removed from the table).
To copy the data on the Standard Curve(s) tab or Run Chart tab, right-click the plot and
choose Copy to Clipboard.
Related Topics
Overview
Application Settings
Data Display
Performance Verification
Contents
• Overview of PV
• PV Test Descriptions
• Performing PV Tests
Overview of PV
Use Performance Verification to set up and run tests to
check the performance of the instrument. Individual
performance tests can be run manually or automatically
using a compatible CVC (Calibrated Validation Carousel)
accessory. For information about the available CVC
configurations, refer to the CVC User Guide in the
“Accessories” section or your documentation media.
PV Test Descriptions
This section describes the performance tests that can be run manually if you have the required
standards. Many of the tests described here may also be included in the CVC test suites. For
information about the test samples included with each CVC and the corresponding test suite,
refer to the to the CVC User Guide in the “Accessories” section or your documentation media.
Test Description
Wavelength Accuracy Locates the peaks from a mercury lamp accessory and
(Hg lamp) displays their measured and allowed range of
wavelengths. A mercury lamp has strong, fundamental
lines throughout the UV-visible range. These emission
lines are a property of the lamp and serve as a
fundamental wavelength standard that does not require
calibration. The wavelengths and tolerance values
cannot be changed.
To run this test, use an empty cell holder and install the
Mercury Lamp calibration accessory properly in the
instrument. For more information, refer to the Mercury
Lamp User Guide in the “Accessories” section of your
documentation set.
Wavelength Repeatability Measures the ability of the spectrophotometer to return
(Hg lamp) to an identical wavelength in a repeatable manner. The
test uses a mercury lamp accessory, which has a strong,
fundamental line near 546 nm. This emission line is an
intrinsic property of the lamp and serves as a
fundamental wavelength standard that does not require
calibration. The wavelength and tolerance values cannot
be changed.
To run this test, use an empty cell holder and install the
Mercury Lamp calibration accessory properly in the
instrument. For more information, refer to the Mercury
Lamp User Guide in the “Accessories” section of your
documentation set.
Wavelength Accuracy Locates the peak near 542 nm of the internal xenon
(Xe lamp) lamp and displays the measured and allowed range of
wavelengths. A xenon lamp has strong, fundamental
lines throughout the UV-visible range. These are an
intrinsic property of the lamp and serve as a
fundamental wavelength standard that does not require
calibration. The wavelength and tolerance values cannot
be changed.
Test Description
Wavelength Repeatability Measures the ability of the spectrophotometer to return
(Xe lamp) to an identical wavelength in a repeatable manner. The
test uses the internal xenon lamp, which has a strong,
fundamental emission line near 542 nm. This line is a
property of the lamp and serves as a fundamental
wavelength standard that does not require calibration.
The wavelength and tolerance values cannot be
changed.
Test Description
Photometric Accuracy Automates the comparison of measured absorbance (or
(customized) %T) of photometric accuracy standards with their
certified values.
Test Description
Baseline Flatness Measures the flatness of a baseline scan over the region
(Abs, 800-200 nm) from 800 nm to 200 nm with smoothing applied to the
data.
Related Topics
Overview of PV
Performing PV Tests
Performing PV Tests
To perform PV tests
Note Select the Classic group in Home if Performance Verification is not available
for the current group.
The available tests appear in the right pane. If no CVC is installed, all of the tests
described in the previous section will be available. If a compatible CVC is properly
installed and a valid calibration file has been loaded, the Performance Verification screen
will display the names of the performance tests that CVC is designed to run. Refer to
Calibrated Validation Carousels (CVCs) for more information.
Note Tests that require the Mercury Lamp calibration accessory can be selected only
after the accessory has been properly connected and installed in the instrument.
CAUTION Avoid pinch hazard. If using a CVC accessory, keep hands and objects clear
of the accessory during operation.
Related Topics
Overview of PV
PV Test Descriptions
INSIGHT Security software, together with Thermo Security Administration software, also
lets the system administrator set system policies and control user access to INSIGHT software
features. See Access Control, System Policies, and Signature Meanings for complete
information.
NOTICE You must install Security Administration software and associate INSIGHT
Security software with it before running INSIGHT Security software for the first time.
Contents
• Getting Started with INSIGHT Security Software
• Access Control, System Policies, and Signature Meanings
• Event Logging
• Sample History Information
• INSIGHT Security Software Icons
• Signing a File
• Verifying a Signature
2. Install Security Administration software and add INSIGHT Security software to the
security database—performed by the system administrator or lab manager. Security
Administration can be installed on the computer running INSIGHT Security software or
on a computer that will be the server for Security Administration (recommended if
multiple computers will run INSIGHT Security software).
Use the Access Control and System Policy features to set the rights of individual users or
groups of users to use the protected features of INSIGHT software (the features included in
the security database). A protected feature will be available only if the logged-in user has the
right to use it.
Use the Signature Meanings features to specify the meanings that will be available for
electronic signatures for each user or user group. For example, the Signature Meanings can be
set so that only a particular user—for instance, the lab manager—is allowed to sign a file with
the “Approval” meaning.
To set access control features, system policies and signature meanings for INSIGHT
Note If a client application was running while you used Security Administration to
change its security policy settings, the new settings will not take effect until the
application is exited and restarted.
For more information, refer to the “Setting Security Policies for Client Applications” chapter
of the Security Administration User Guide.
The table below describes INSIGHT access control features and policies.
Feature Description
Ability to run INSIGHT Specifies which users can run INSIGHT software. (See
“Controlling access to client application features” in the
“Using Security Administration” chapter of the Security
Administration User Guide for general instructions.) If a user
has not been granted the ability to run the software, an error
message appears when the user attempts to start it.
Ability to delete workbooks Specifies which users can delete workbook files with the
Delete button in My Data.
Ability to delete PV reports Specifies which users can delete Performance Verification
reports with the Delete button in My Data.
Ability to delete templates Specifies which users can delete template files with the Delete
button in My Data.
Ability to delete Scan Specifies which users can delete Scan sample data files from
samples the current workbook by right-clicking a sample in the results
table in Measure Scan and choosing Remove.
Ability to edit templates Specifies which users can set parameters in Settings.
Ability to configure reports Specifies which users can change settings in Reports.
Ability to perform PV tests Specifies which users can run Performance Verification tests.
Ability to perform Specifies which users can use the Wavelength Calibration
wavelength calibration buttons on the Calibrations tab in System Settings.
Ability to calibrate Sipper Specifies which users can use the Sipper Calibrate button on
the Calibrations tab in System Settings.
Ability to calibrate Linear Specifies which users can use the Linear Cell Changer
Cell Changer Calibrate button on the Calibrations tab in System Settings.
Ability to update CVC Specifies which users can use the Load CVC Calibration
calibration button in Performance Verification.
Ability to select CVC tests Specifies which users can select listed tests for the CVC in
to perform in PV Performance Verification.
Math menu Specifies which users can use the commands in the Math
menu.
Analyze menu Specifies which users can use the commands in the Analyze
menu.
Ability to set Rate Specifies which users can set parameters in Settings for the
parameters Rate application.
Feature Description
Options tabs Specifies which users can change settings on the tabs in
Options.
System Settings tabs Specifies which users can change settings on the tabs in
System Settings.
Authenticate on startup Requires users in the specified policy groups to enter a valid
name and password when starting INSIGHT software.
Prevent overwriting of files Ensures that existing files are not overwritten when users in
the specified policy groups save workbook or template files.
If a user saves a file while this policy is selected, the file will be
set to read-only in its properties. In the future, users will not
be able to overwrite this file, even when this policy is not
selected.
Prevent delete, rename and Prevents users in the specified policy groups from performing
right-click in file dialog these operations in file dialog boxes:
boxes
• Deleting a file by selecting it and pressing the Delete key
on the keyboard.
• Renaming a file by selecting its name (by slowly clicking
the name twice) and typing a new name.
• Displaying a shortcut menu by right-clicking a file.
Normally a menu displayed this way lets the user perform
operations such as copying, deleting or renaming the file.
Feature Description
Prevent signature reason Prevents users in the specified policy groups from entering a
entry when signing files custom signature meaning when signing a file.
Feature Description
Directory for PV reports Designates for each policy group a directory for automatically
saving Performance Verification reports.
Allow access only to listed Specifies the INSIGHT software user groups that are available
button groups (blank field to users in the specified policy group. Selecting a user group in
allows access to all groups) Home displays the application buttons for that group. See
Applications Tab for more information.
Type the desired user groups in the text box in the Description
box, with a comma and a space between them; for example:
Classic, Quality Control
Leave the text box blank to give users in the specified policy
group access to all applications.
The signature meanings in the default list for INSIGHT software are described below.
Note Users can be allowed to enter custom signature meanings. See “Prevent signature
reason entry when signing files” in Access Control, System Policies, and Signature
Meanings for details.
Event Logging
To provide a system audit trail, Security Administration uses the Event Log service in
Windows software to record Security Administration and client application operations, or
“events,” in a log that you can view with Event Viewer. Examples of logged events include the
date the event occurred, the type of event, and the user name of the person who was logged in
when the event occurred. The Event Log service allows all file operations to be logged, both
within and outside of applications that are run on the system, even if a client application is
not running. By recording this information, Security Administration allows your system to
meet the audit trail requirements of 21 CFR Part 11.
To view the log of events for Security Administration and your client applications
1. Start the Windows Event Viewer on the computer where Security Administration is
installed.
Refer to your Windows documentation for information about locating the Event Viewer
feature in your version of Windows software.
2. In the left pane of the Event View window, click the Thermo Electron icon.
A log of significant events that occurred while your client applications were being used
appears in the right pane. You can sort the events according to date, category, user and so
on by clicking the column headings.
3. To see detailed information about an event, double-click the event (use the up and down
arrows to see information about the previous or next event).
You can also export the list of events.
4. When finished, close the Event Viewer.
For more information on the Windows Event Viewer, see its associated Help system.
This icon indicates that current workbook has not been signed since it was last
changed.
This icon indicates that current workbook has not been saved (or signed).
To see information about the signature status of the current workbook, click any of these
icons.
Related Topics
Signing a File
Verifying a Signature
Signing a File
If INSIGHT Security Software is installed, use the following procedures to digitally sign
workbook and template files. The visible portion of a digital signature consists of a user name,
a date and a stated reason for signing. A digital signature also contains encrypted information
used to detect whether the file has changed since it was signed.
Related Topics
INSIGHT Security Software
Verifying a Signature
Verifying a Signature
If INSIGHT Security Software is installed, use the following procedures to verify digital
signatures for workbook and template files.
Related Topics
INSIGHT Security Software
Signing a File
Index
Symbols Scan 108
accessories, manual
%T button 66
installing 25
removing 26
Numerics accessories, Smart
3D image installing 27
displaying 63 removing 28
rotate, zoom, pan 57 accessories, thermostatted
installing 29
A acquiring data 60, 71, 73, 79, 172, 186, 201, 214, 227, 240, 254,
268, 282, 296
Abs button 66
acquiring fixed wavelength data 68
absorbance values action button
display for quant 111 Export 132
access control 307–308 Measure 98
access cover Preview 132
instrument detector 20 Print 132
mercury lamp 20 Reset 98
sample compartment 18 Unlock 98
tubing 19 action buttons 48
accessories adapter
aligning 155 tubing 19, 30
parameters 102, 108, 117, 125, 167, 182, 196, 211, 223, adapters
236, 250, 264, 278
for connecting tubing 19
standard 24
Add 140
status 64
tubing connectors 19 adding
constant to spectrum 146
Accessories tab
group 151
DNA Melting 196
spectra 140
Fixed 102
Advanced 294
Nucleic Acid 167
Nucleic Acid Labels 182 advanced calculations 49, 85, 170, 184, 213, 225, 238, 252, 266,
Pierce 660 nm Protein Assay 278 280, 294
Pierce BCA 236 advanced equations for Quant 154
Pierce Modified Lowry 264 air, for instrument purge 23
Protein A280 211 aligning
Protein Biuret 292 accessory 155
Protein Bradford 250 detector module 41
Proteins & Labels 223 single cell holder 24
Quant 117 Alignment tab 155
Rate 125 Analyze menu 64, 147, 149
N P
navigation pane 46 page
nitrogen, for instrument purge 23 layout 152
Normalize 141 number 152
normalizing spectrum 141 pan
notes 3D image 57
adding to reports 134 panel
Nucleic Acid for external tubing connectors 19
Accessories tab 167 instrument 36
application 161–162, 168, 172 sample compartment front 18
Instrument tab 166 sample compartment side 21
Measurement tab 164 panels
overview 161 removable 17
Samples tab 167 parameters
Type tab 163 resetting 98
nucleic acid 161 parts
concentration 161 ordering 42
purtiy 162 path 130, 152
Nucleic Acid Labels 183, 186 pathlength 163, 177, 206, 218, 231, 245, 259, 273, 287
Accessories tab 182 for quant 110
Instrument tab 181 peak
Measurement tab 178 area 94
overview 176 height 93
Samples tab 182 peak area tool 94
Type tab 177 Peak Pick 147
nucleic acids Peak Pick tab
measuring 172 Scan 109
number of cycles 121 peak/valley measurement tool 93
number of samples 102, 108, 118, 126, 167, 182, 197, 211, 223, peaks
236, 250, 264, 278, 292 finding 109, 147
number of stages 193 Performance Verification 301, 305
report 130
O test descriptions 302
off button Performance Verification test
for Computer Control instrument 33 baseline flatness 305
for Local Control instrument 35 noise 304
off switch 15 photometric accuracy 304
photometric drift 305
oil
photometric repeatability 304
removing from touchscreen 40
resolution 304
Oligo Calculator 174, 188
stray light 304
on switch 15 photometric accuracy PV test 302
opening
z-height
spectrophotometer 13
zoom
3D image 57
spectrum 66, 92