Hooke's Law (4
Hooke's Law (4
Hooke's Law (4
Name
Institution Affiliation
Professor
Date
LAB REPORT 2
Abstract
This lab report gives an explanation to the experimental measures and the results
obtained from the test carried out to study the Hooke’s law on a spring balance subjected to loads
of different masses. It was found that the amount of mass added on the spring is directly
To examine the behavior of a spring when stretched out by an outside force and to
To prove that a lengthened spring can act as an excellent example to explain behavior of
To investigate the relationship between force and squeeze or stretch of a spring (Hooke’s
Law).
Introduction
Robert Hooke was extensively involved in science field and is principally remembered in
today’s learning for the Hooke’s Law. Currently, most of science textbooks credit Hooke
according to his contribution as the discoverer of the short and snappy spring’s mathematical
relationship F= ─kx.
Hooke’s law experiment is conducted to validate the property of all solid objects to
“stretch” or “squeeze”. All materials that are elastic and obey Hooke’s Law are generally
applicable today in many set ups such as the oscillating systems: waves and harmonic oscillators.
The elasticity properties of materials are experienced in a lot of physical observable facts. When
LAB REPORT 3
an object is compressed (deformed, stretched, twisted) while the compressing forces are
adequately little, the object will resume to its initial shape immediately when the compressing
forces are detached. In such instances, the compression is said to happen within the boundary of
the object’s elastic limit, thus, there will be no permanent compression. The slender stretching of
a plastic thread or rubber band is a good illustration of an elastic compression. Rods, metal
beams, steel wires, concrete columns and other solid materials can go through elastic
linear direction, the restoring (resisting) force will respond proportionately to the
compression/deformation.
If the string is not stretched until it reaches a point when it becomes permanently
compressed or deformed, the spring’s behavior when subjected to a squeezing force will obey
Hooke’s law. If the spring is stretched beyond its elastic limit, the objects spring is said to be
under a state called plastic range. Therefore, exerting extra force will cause the spring to undergo
permanent displacement.
Hooke’s Law
Hooke’s law therefore states that when you squeeze a string or an elastic material, it will
exert a restoring force that a proportional (equal) magnitude to the deformation or displacement
(distance stretched).
F= ─kx (Equation 1)
Fα x; it is read as; force is directly proportional to squeeze or stretch. This expression can
Fs= ─kx
Here, F stands for the restoring force provided by the material being squeezed or
stretched and x represents the displacement exerted by the material being squeezed. It is
important to put negative sign (-) since it means that the restoring force acts in opposite direction
against the displacement; the acceleration of the material in Simple Harmonic Motion (SMH) is
in the opposite direction. For instance, when a rubber band is stretched upwards by an object, the
string will automatically put forth a negative restoring force but in the downward direction
Figure 1
The equation above also means that for every object that obeys Hooke’s law for example
a spring and rubber band, the more it is squeezed or stretched, the larger will be the amount of
restoring force provided by the material doing the squeezing or the stretching. The force applied
(F) acting on the Hookean material make it to be moved (squeezed or stretched) by some
quantity (x). However, the ratio resulting from the change in the applied force, denoted by ∆F
LAB REPORT 5
k=∆F/∆x (Equation 2)
This experiment was done to test the relationship between these forces by using a spring.
Objects of different masses were hanged on the spring and then the force acting on the object
was controlled. The extension (unit) was measured for every weight applied on the spring, the
extent to which the string was able to “stretch”. The relationship between equation 1 and
equation 2 is illustrated using a graph of weight (F) versus the stretch (x) whereby it must
The constant k will determine the spring stretchiness. A harder spring will produce a
The force exerted and displacement obtained is linear as shown the following graph:
Y-axis
Weight
(N)
Graph Figure 2
The graph above representing the relationship between the forces applied and the amount of
extension of the elongated elastic spring in meters has a straight line starting from zero mark
(point of origin) because the spring has not reached its elastic limit. If the elastic limit is
exceeded, the graph will begin to curve and stop apparently as shown in the graph below:
Y-axis
Stretching Force
The behavior of the above graph curves towards the end because the proportionality element is
not fulfilled by the force and extension factors; the sprig has exceeded its elastic limit due to the
Strength of the material used: This factor introduces Hooke’s Law through the
mechanical properties exhibited by materials. It is one of the key factors for choosing any
material to use in a fussy application. Strength focuses on the stiffness of materials when
subjected to comprehensive forces or linear tensile (forces that act in one direction in a specific
line). This factor does not include bulky forces (forces that act in three dimensions or throughout
a given volume) or shearing forces (forces that act in two dimensions or along a plane). The
LAB REPORT 7
strengths of all materials are measured using a logarithmic scale. The logarithmic scale does not
have a zero reading; it starts with a small value such as 0.01 or 0.001 but it will never get to zero
mark. From the logarithmic scale, some of few materials have been tested and their values have
been recorded for example diamond has logarithmic value from 10,000 to 100,000; glass
material 1000 to 10,000; kevlar, mild steel 100 to 1,000; wood 1 to 100 and formed polymers 0.1
to 1. Tension is direct proportion to the force applied on the material just as stretch is directly
That is; F = k Dx
The above formula represents the mathematical statement as used in Hooke’s Law,
whereby k stands for the stiffness of the material subjected to the force. When calculating the
strength of any material, the above statement is therefore equated to material’s tension per
F/Dx Nm-1(units). The formula represents the opposite of Hooke’s Law stiffness. The
stiffness of the material (k) will be calculated by plotting a graph of F-Dx, where the slope from
the linear part of the graph will give the stiffness (you need to take the averages for the slope
values on individual data points). An experiment between spring and fibres can be conducted to
Extension and Tension: The Hooke’s Law heavily relies on this factor to show the
behavior of each material. The elasticity of materials measures the extension and tension of the
materials. From Hooke’s Law, a conclusion has been made that Elastic materials can extent
(extension) until a given limit but they will never get damaged (tension). This point has been
referred to as elastic limit of materials. The extension part ends at the elastic limit of the material.
LAB REPORT 8
Extension and tension factors do not coincide because the behavior can be explained in terms of
Theoretical Explanation
Loading of several kinds of solid objects such as metals basically will stretch out the
bonds acting between the atoms of every solid material from which the object is made of. There
exist a linear relationship between the amount of load applied and the extension caused; this is
therefore a reflection to verify the fact that the magnitude of the bond’s force is linearly varying
with distance between one atom and the other (separation between atoms). The permanent
displacement results due to plastic behavior of the particular atoms with respect to their original
states. Many polymers for example rubber band display very great extensions. Their extensions
are far too big to be clarified using the theory of bond stretching. Initially, it operates under
straightening and uncoiling out of very long molecules residing within the structure of materials.
As the bonds existing between the material stretches over, the long molecules are pulled out
straightly.
Theory of Elasticity and Stiffness of materials: (Matt, 2015) stated that Hooke’s law is
an extension to explain stiffness and elasticity of materials. This theory is applied in Hooke’s
Law when calculating the elastic energy exerted by a given force on a spring.
Elasticity is a unique property found in materials or objects such as rubber band and
springs that causes the material to regain its previous shape after deformation. A material is said
to be ore elastic than another if it is able to restore itself to its previous shape configuration more
precisely. From Hooke’s law experiment, springs are considered to be the best materials to
demonstrate the effect of elasticity because, when squeezed or stretched, they exert a strong
F = C∆L.
The elastic energy of every spring only relies on distance (change in length). U = ½
C∆L2.
Stiffness is another characteristic which illustrates the extent to which materials can resist
the deformation effect when a corresponding force is applied. Therefore, the stiffness of an
elastic material depends on the measure of its resistance to deformation forces. For elastic
materials that have a single measure of freedom for example when compressing or stretching a
K = F │𝜕
From the formula above, stiffness is represented by k, the force applied on the material is
represented by F while 𝜕 stands for the displacement (for example the change in spring’s length
when stretched).
Theoretical explanation
Elasticity is better understood by considering the stress and strain exhibited by solid
materials. Stress takes the assumption that the object experiencing elasticity behaves like any
object that has been placed on a Cartesian plane system where there are many forces acting on it.
The stress condition is obtained by summing up all the vector forces acting on the object. The
resulting sum of the forces is zero; this is the stress effect. Therefore, for a spring or any elastic
material subject to stretch or squeezing force, the total force acting on the object behaves like a
stress thus leaving the object with zero forces; this makes it restore its shape after deformation
LAB REPORT 10
after the vectors forces are released. During the condition of vector forces, the elastic object
Strain refers to the internal regaining (restoring) force that is experienced by any elastic
material that undergoes deformation. The quantity of the resulting force caused by deformation is
Robert Hooke was the first scientist to realize the relationship between strain and stress
forces. Hooke, through the law of elasticity, determined the approximation that states that the
extent by which a material is distorted (the strain effect) relates linearly to force that causes the
distortion (the stress effect) under the normal conditions. Generally the relationship between
Tij = Cijkl∈ij
From the formula above, elastic stiffness is represented by Cijkl which is the fourth order tensor.
The strain mechanisms are obtained by overturning the Hooke’s law to utilize the conformity
coefficients: Sijkl. .
Hooke’s law (F α x) can be calculated using a combined formula (using stress and strain effect)
F/A = Y (∆ L / L).
LAB REPORT 11
Equipment used
Meter Rule
C-clamps
Rubber band
Stopwatch
Ruler support
A Mass holder
A supporting base
Safety
While conducting the experiment, it was important to observe safety measures and
considerations. The set up must be designed such that the apparatus are assembled within the
central dimension of the table and on the base of the stand to prevent from toppling over. The
stand used was strong and the masses were hooked to reduce the chances for them to
unintentionally fall off. Also, all the springs did not fall off since they were tightly fixed on the
stand.
Considerations were made to keep my feet and body out of the region where the masses
and the meter stick will fall whenever the rubber band and the spring break. Finally, enough
LAB REPORT 12
masses were hung to the end of the hook so as to obtain a measurable squeeze or stretch. This
was done cautiously as per the instructions to avoid too much weight on the spring which could
cause great damage to the spring. A phenomenon referred by Hooke’s Law as the “elastic limit”
of a spring. In most physics laboratories, it is a rule for students to buy another spring if they
damage whilst the instructions specified the masses to be used. However, it is important to
Besides the safety measure used for the experiment; there were other general measures
that apply to the laboratory environment, most of which were observed during the experiment
such as:
Everyone in the lab was required to exercise extensive care near radioactive materials,
lasers cryogenic liquids and lasers. Warning signs were placed near the places where such
It was the responsibility of everyone inside the lab to secure his/her owns safety and also
It was a rule that no one starts an experiment or start working in a particular area without
We were instructed not to touch any equipment apart from the ones indicted in the
For any query or uncertainty about the safety rules and procedures in the lab, we were
instructed to consult the lab demonstrator. Uncertainty issues include the health condition where
someone might think he/she may be affected by the techniques used to conduct the Hooke’s Law
experiment.
LAB REPORT 13
Lastly, as an important rule, it was advisable for everyone working in the lab to report
accidents arising in the working area to the lab demonstrator immediately, who will then take the
necessary action.
Method
The apparatus for the experiment were assembled on the table and the set up was placed in the
correct position. The ruler (half meter) was supported using a clamp stand. A helical spring was
attached to the clamp stand at the point having a jutting screw; the pointer was then suspended
below the screw. The mass (load) hanger was clasped under the spring’s pointer. A paper
containing blank table was prepared to note down the results in an organized. Ray diagrams used
F= ─kx
Where;
F refers to the quantity of Force applied on the spring; it is measured in Newton’s (N)
Fx = ─k (x ─ xo) Force
𝐹x
k= the spring constant or force constant
x ─ xo)
𝐹𝑥
𝑥 = 𝑥𝑜 ─ the distance (Length) from equilibrium
𝑘
LAB REPORT 14
𝐹𝑥
𝑥𝑜 = +𝑥 the equilibrium position of the spring
𝑘
The potential energy of the spring can be obtained using the following formulas:
1𝑘𝑥 2
𝑈= Total potential Energy of the spring
2
2𝑈
𝑥= Force constant of the spring
𝑥2
√2U
𝑥= Stretch Length of the spring
𝑘
The total energy for an elastic object is the sum of potential energy of the spring and the total
E = ½ mv2 + ½ kr2
Where
Fx =Force
Procedure
The appropriate apparatus were provided for the experiment. The materials were arranged to
Only elastic materials such as helical spring and rubber band are used. The equation for
The force exerted on the elastic material should not be too large nor too small to facilitate
logical readings of the material and to stay within the elastic limit.
Experiment steps:
2. Attach the hook and the pointer with its pan to the edge of the spring.
3. Place the meter rule to stand vertically (along the vertical stand provided) in a way that
the arrow of the pointer keeps up a correspondence to the readable markings on the meter
ruler.
4. Note down on paper the preliminary reading of the arrow pointer on the ruler (call it x).
make sure that the spring is not bouncing whenever you are taking the readings against
the ruler.
6. Measure the time taken for the oscillations (T). Repeat this step to obtain two trials for
the oscillations (x oscillations) and the time taken for a single (1) oscillation; call it T.
7. Note down on your paper the displacement of the pointer after adding the weight (call it
x-x0).
8. Use masses of different weights (40g, 60g, 80g and 100g) and repeat steps 6 and 8.
9. Draw a graph of displacement (x─xo) cm against mass (g). For each scenario, calculate
the spring or force constant (k) by using the formula k =mg÷x; where m stands for the
LAB REPORT 16
mass of material used, g refers to the gravity (9.8 m/s2). Calculate the average of the
10. Draw a graph to show the values for T2 against mass (g).
Recording of the results: in order to record both the oscillation’s results and the mass
against extension results and also to capture all the calculations, a table with seven columns
was prepared. The experiment used masses which were hung on the spring; therefore, the
table required a column to record the masses of different magnitudes. The extension
column was needed to record the readings from the meter ruler. To increase the likelihood
of obtaining accurate results for the oscillations, the table provided three columns whereby,
results for two trials were recorded and the average (mean) for the two trials were recorded.
oscillations
20g 6.4 cm 2.71 sec 2.85 sec 2.78 sec 0.554 sec 0.307 sec
40g 12.9 cm 3.87 sec 3.81 sec 3.84 sec 0.768 sec 0.590 sec
60g 19.0 cm 4.47 sec 4.63 sec 4.55 sec 0.910 sec 0.828 sec
80g 25.0 cm 5.31 sec 5.19 sec 5.25 sec 1.050 sec 1.103 sec
100g 31.6 cm 5.88 sec 5.75 sec 5.82 sec 1.163 sec 1.355 sec
Table 1
Observations
LAB REPORT 17
When the masses were loaded on the spring, the spring pulls or pushes the masses
downwards and upwards until an equilibrium state is achieved when the spring is brought
to halt. The time taken by the oscillations are counted and recorded by setting a timer. It
was noted that when the mass is increased (using a load of bigger mass). The oscillations
were proportional to the amount of masses added for example the mass of 40 g results in
40g
Table 2
LAB REPORT 18
∆ in x
35
30
25
20
15
10
Figure 3
2
80 g = 4 𝜋 × (3.15 ÷ 14)
1.3
1.2
1.1
1.0
0.9
0.8
0.7
0.6
0.5
0.4
0.3
0.3
0.2
0.1
1. It was found that the spring or force constant is truly calculated from the slope of the
graph (Force against stretch). This was done by calculating the gradient of the slope.
2. When the force is increased, the stretch or the squeeze of the spring increases in a
proportionate manner.
3. Based on the readings from the pointer scale, it was found that the force/spring constant
In most physics laboratories, you must compare the results you obtain with experimental
or other known results. These other calculations will guide you in evaluating the relevance of the
results, thus, you will be able to make correct inferences and conclusions. It is important to know
the extent by which your experimental results differ from other values obtained in laboratories
concerning the same phenomenon. Most scientific experiments accept values that are within the
of less than 10 % error. If the comparison of your results shows a percentage difference that is
greater than 10%, there is a high likelihood that you have many mistakes. This comparison will
help you to look back to your experiment and identify the potential sources to the errors. The
calculations used will constitute the integral part of your discussion and the analysis. A greater
percentage error should be taken into consideration during error analysis. In most cases, great
error percentage will indicate that the objective of the lab experiment has not been achieved.
Percentage error: this is used when comparing the results obtained to an accepted value or a
known result. Percentage error is the supreme value of the deviation from the two results divided
│𝐷𝑖𝑓𝑓𝑒𝑟𝑒𝑛𝑐𝑒│
% error = (100 %)
𝐾𝑛𝑜𝑤𝑛
Percentage Difference:
Percentage difference is obtained by comparing the results with other experimental results. It
shows the fixed value of the deviation from the results divided by the average of the two; it is
│𝐷𝑖𝑓𝑓𝑒𝑟𝑒𝑛𝑐𝑒│
% difference = (100 %)
𝐴𝑣𝑒𝑟𝑎𝑔𝑒
relies on approximation. This introduces uncertainties during its measurement that arises due to
many types of errors. During the experiment, there were three kinds of errors which prone to
deter the objectives of the experiment: Random errors, Reliability errors (systemic and parallax)
Random errors
These errors are difficult to trace because their causes are indeterminate or
unknown. Random errors are normally small errors that follow the rules of chance. These
errors are reduced and avoided by taking the averages of large amount of observations.
While taking the experiment on Hooke’s law, the following types of random errors were
considered:
1. Instrument resolution: All instruments used in physics labs have restricted precision
that confines the capability to resolve little measurement differences. In this case, the
meter stick used in the experiment cannot differentiate distances (length) to exactness
much better than when using a meter rule which uses smaller units such as (0.5 mm). one
LAB REPORT 22
of the best ways to avoid this error is to use the null difference method which yields more
accurate measurements. The balance or null method is quite different from measuring
two or more similar quantities. The degree of unknown quantity will be found by
comparing the sample using the difference between the two samples. However, this
method helps in eliminating the problems associated with source instability. With this
method, the instruments will be very susceptible and will not require a scale.
were experienced by when using a meter rule with larger units thus ignoring smaller units
measurements during lab experiments is that there are no clear measurement definitions.
For example, when two students are given the same thread to measure, they would
probably produce different results since each student might stretch the thread with
specify and consider the state of affairs that will affect the accuracy of the measurement.
While taking the Hooke’s law experiment, incomplete definitions resulted from the
springs used. The degree of stretching the spring was not uniform for all the masses, that
is, there were other factors that contributed to the stretching of the spring other than the
force exerted by the masses; for example holding the spring by hand caused small stretch
to the spring.
3. Physical variations: this error usually occurs when limited measurement trials are
phenomena being investigated when doing any scientific practical. Doing so will help to
reveal hidden variations that might perhaps go undetected. The variations will therefore
call for detailed examination; these variations can be pooled to get an average figure. The
experiment on Hooke’s Law suffered physical variation error since it only involved two
trials. The trials were limited to discover other variations in measuring the length and the
4. Instrument drift error: most readings on electronic instruments drift over a period of
time. This type of error is usually not a big concern but it is significant if the amount of
deviation is larger. Therefore, it is advisable for students and scientists to regularly check
the readings of any electronic device used for experiments to adjust these drift errors.
When conducting the lab experiment about Hooke’s Law, the sources of drift error
resulted from the stopwatch used. Small drift error occurred the time between readings
5. Hysteresis and Lag time errors: most measurement instruments require more time to
attain its equilibrium state. Measurements that are taken when the device is still in motion
will produce results that are generally low and vague, hence resulting in lag time error.
From the experiment on Hooke’s Law, lag time errors were involved when waiting for
the oscillations to stop. Some readings both from stopwatch and the meter scale were
taken whilst the masses were still oscillating. Additional errors were encountered when
the readings were done long time after the oscillating masses had reached a balance since
it was difficult to tell when the masses had come to a stop. Hysteresis effect happens
when the devices undergo memory effect such that it moves up or down or stops
subsequently when taking the records of values. When doing the Hooke’s Law
LAB REPORT 24
experiment, hysteresis effect was experienced with the stopwatch device when taking the
values for the oscillations. This effect was commonly associated to the material
incorporated with the stopwatch components. The components get magnetized, thus there
Reliability Errors
A reliability error affects the quality of measurement. In the experiment, reliability errors
1. Systematic error
This error usually occur when there faulty equipment in the laboratory or when some of
the apparatus have not been accurately calibrated in order to robust with the experiment.
Systematic errors resulted from environmental conditions inside the laboratory. For
example, when the doors and windows were open, air from outside interrupted with spring’s
oscillations thus manipulating with the swinging of the spring or it cause unnecessary
oscillations when the student is taking the scale readings. Thus, the incorrect measurements were
In Hooke’s law experiment, systemic errors often happen when the lab technicians
provide meter rulers containing “zero errors”. Zero error will either make the readings on the
scale to become lower or higher than the genuine result. Therefore, it is the responsibility of the
lab technicians to ensure that the rulers do not contain any zero error before the supplying to the
students. Furthermore, it is mandatory for the students to check all the apparatus as listed in the
instructions.
LAB REPORT 25
The helical spring should be in good condition since an inelastic or damaged spring will
change the force or spring constant of the experiment. Inaccurate spring will lead to deviations in
2.Parallax Error: This error occurs due to the difference or displacement of actual
reading of the apparatus used for the experiment. Since our eyes are not positioned
From the Hooke’s law experiment, parallax errors resulted when students who managed
to observe and read the results did not put total concentration on the amount of decimal places
during the calculation. When taking the readings from the spring pointer, the student who is
handling the experiment should position his/her eyes directly to the level of the measurement
reading.
Personal errors
These errors occur due the poor technique, bias and carelessness on the side of the
experimenter. This error can be minimized by avoiding and correcting blunders or human
mistakes. This experiment was taken while considering the correct procedures; therefore there
were no human errors or biasness such as inadvertently forcing the graph values to fit the line.
The readings from the ruler were carefully observed from all the angles. This was done so
that the readings could appear vertical unlike when the readings are viewed from right
angle.
LAB REPORT 26
The ruler and the stopwatch were adjust at the end of every trial readings and
immediately before the start of every experiment. This was done to avoid reading repeat
The experiment readings were taken when the spring arrangement had attained a
stationary position.
The values were obtained from the observations on the experiment and there were no
mental arithmetic fixed to the results. Readings from the stopwatch and the ruler were
noted down and calculations were done later after the end of the experiment.
The eye level had to be matched with the pointer scale and also the stopwatch’s scale to
avoid parallax error. The pointer assisted in reading the tiny values on the scale.
The figure 2 (graph of extension against mass) provided us with scattered points. This
was a clear indication that the readings were precise and no external data were used to
manipulate the graph. It also showed that the trend values of the experiment were changing due
to the factors related to the theory. A line of best fit was drawn and it touched majority of the
circled points on the graph. These majority points along the line designated a trend line of
proportionality, therefore, an even straight line extending from the zero origin was drawn. The
general pattern of the graph and the results was a major consideration while drawing the graph.
Using the Newton’s second Law: F = ma; the Hooke’s Law can be futher explained by
relating the two laws since they apply the same concept. Therefore;
(F = ma) = (F = k∆x)
F = ma = k∆x;
LAB REPORT 27
Where; a stands for the acceleration exerted by the mass (load) on the spring. When the mass is
suspended on the spring, the acceleration occurs due to the forces of gravity that acts on the
mass, thus pulling it down. Because the spring is elastic (ideal spring), when the mass is
displaced by the gravity, it will start oscillating with more than two oscillations depending on the
The figure 1 shows the behavior of a spring as subjected to forces of different magnitudes
(F and 2F). The spring is shorter when not stretched and maintains its structure unlike the spring
under F and 2F forces. The displacement for the spring under F force is x while under 2F force it
reads 2x. x and 2x represents the elasticity of the spring. The force 2F is twice force F thus their
displacement 2x and x respectively indicates that there exist a linear relationship between the two
forces applied. This implies that when a force is applied, the spring will be displaced
F is an intricate function to cause the displacement 2x and x in the figure. The first spring
is at equilibrium position where there are no forces acting on it. The net force in this case is said
to be zero; x = 0. So,
W0 = mg/k ; when we subject this equilibrium to Hooke’s Law, we can define the equation for
F = 0 kg/s2
We can calculate the potential energy for the equilibrium state as shown below:
For minimal displacements away from the equilibrium, any F can be placed with a
F = A + Bx + Cx2 +Dx3 . For minimal displacements away from the equilibrium, any F can be
F = A + Bx + Cx2 +Dx3 . For minimal displacements away from the equilibrium, any F can be
F = A + Bx + Cx2 +Dx3 . For minimal displacements away from the equilibrium, any F can be
F = A + Bx + Cx2 +Dx3 + …
F = Bx + Cx2 + Dx2 …
From our figure 1, when the displacement is x (small displacement), using the formula
above for this case, if we add higher powers of x we will still get a smaller value. Therefore, to
obtain a good approximation, we have to neglect them all. We will then get a better
F = Bx
When a positive value of B is supplied, the resultant force will thrust downwards with
respect to the direction depicted by the displacement. This thrust movement will speed up the
mass attached to the spring to move away from its point of equilibrium as indicated by the arrow
in the diagram. This scenario is referred to as unstable equilibrium of an object; a good example
is when a ball is placed on top of a sharp hill with it having balanced equilibrium. However,
when the value of B supplied is negative, the resultant force will try to maintain the direction and
LAB REPORT 29
restore the entire system of the spring to a stable an equilibrium state. This scenario is referred to
as stable equilibrium.
When calculating the effect of Hooke’s Law, we have to make the value of B to be negative and
define our formula to be B = -k (because this is the area we have much interest). Therefore we
B = -k
F = -k
Conclusion
The results from the above experiment can be used to confirm the truth of the Hooke’s
Law. Data from table one and the summary provided in the graph showed that the spring has
conformed to the Hooke’s law whereby the line drawn was straight; meaning that the extent to
which the spring stretch is directly proportional to the quantity of weight added to the spring
hanger. In order to observe and record accurate behavior of the spring during the experiment, it is
important to consider other factors that contribute to correct conclusion such as errors during the
set up.
LAB REPORT 30
References
Matt,W. (2015). What Is Hooke’s Law?: The Universe Today. London: Nova Science
OpenStax College. (2012). “Hooke’s Law: Stress and Strain Revisited”: creative Commons
Attribution.
Bird, R. (2012). Mechanical Engineering Principles, 2nd edition. London and New York:
Croft, A., and Davison, R. (2008). Mathematics for Engineers,3rd edition. London : Pearson