Ca MW: Table 1 CAS: Table 1 RTECS: Table 1: CALCIUM and Compounds, As Ca 7020
Ca MW: Table 1 CAS: Table 1 RTECS: Table 1: CALCIUM and Compounds, As Ca 7020
Ca MW: Table 1 CAS: Table 1 RTECS: Table 1: CALCIUM and Compounds, As Ca 7020
SYNONYMS: Quicklime (CaO); limestone (CaCO 3 ); marble (CaCO 3 ); hydrated lime (Ca(OH) 2 )
SAMPLING MEASUREMENT
BACKGROUND
CORRECTION: none used
RANGE STUDIED: 2.6 to 10.2 mg/m 3 [1] RANGE: 0.08 to 1.7 mg per sample [2]
(85-L samples)
ESTIMATED LOD: 0.001 mg per sample [3]
BIAS: - 0.39%
OVERALL PRECISION (ŜrT): 0.063 [1] PRECISION (Sr): 0.02 [1,3]
ACCURACY: ± 11.5%
APPLICABILITY: The working range is 1 to 20 mg/m 3 for an 85-L air sample. This is an elemental analysis, not compound
specific. Verify that the compounds in the samples are soluble with the ashing procedure. Aliquots of the samples can be
analyzed separately for many additional metals.
INTERFERENCES: The use of 1000 µg/mL Cs controls ionization in the flame caused by metals such as Na, K, Li, and Mg. The
presence of Si, Al, or H 3 PO 4 require the use of 1% (w/w) La as a releasing agent.
OTHER METHODS: This method combines and replaces P&CAM 173 [4] and S205 [2]. Method 7300 (ICP-AES) is an alternate
analytical method.
REAGENTS: EQUIPMENT:
SPECIAL PRECAUTIONS: Perform all perchloric acid digestions in a perchloric acid fume hood.
SAMPLING:
SAMPLE PREPARATION:
NOTE: The following sample preparation gave quantitative recovery (see EVALUATION OF
METHOD). Steps 4 through 9 of Method 7300 or other quantitative ashing techniques
may be substituted, especially if several metals are to be determined on a single filter.
3. Open the cassette filter holders and transfer the samples and blanks to separate clean beakers.
4. Add 5 mL conc. HNO 3 and cover with a watchglass. Start reagent blanks at this point.
5. Heat on hotplate (140 °C) until most of the acid has evaporated.
6. Add 2 mL conc. HNO 3 and 1 mL 60% HClO 4.
7. Heat on 400 °C hotplate until dense fumes of perchloric acid appear.
8. Remove watchglass and rinse into the beaker with distilled water.
9. Place the beakers on the 400 °C hotplate and allow to go to dryness.
10. Cool each beaker and dissolve the residues in 5 mL 5% HCl.
11. Transfer the solution quantitatively to a 100-mL volumetric flask containing 2 mL of 50 mg/mL
Cs solution and 2 mL of 50 mg/mL La solution.
NOTE: Dilute to a smaller volume (e.g., 10 mL) with proportionally less Cs and La if required for
sensitivity of analysis for other metals in the sample.
12. Dilute to volume with 5% HCl.
13. Add known amounts, covering the range 0 to 500 µg Ca per sample, of calibration stock
solution to 100-mL volumetric flasks containing 2 mL each of the 50 mg/mL Cs and La solutions
and dilute to volume with 5% HCl.
14. Analyze the working standards along with the samples and blanks (steps 19 and 20).
15. Prepare a calibration graph of absorbance vs. solution concentration (µg/mL).
16. Aspirate a standard for every 10 samples to check instrument drift.
17. Check recoveries with at least one spiked media blank per 10 samples.
18. Use method of additions occasionally to check for interferences.
MEASUREMENT:
CALCULATIONS:
21. Using the measured absorbances, calculate the corresponding concentrations (µg/mL) of
calcium in the sample, C s, and average media blank, C b, from the calibration graph.
22. Using the solution volumes (mL) of the sample, V s, and media blanks, V b, calculate the
concentration, C (mg/m 3), of calcium in the volume of air sampled, V (L):
EVALUATION OF METHOD:
Method S205 [2] was issued on September 26, 1975, and validated over the range 2.6 to 10.2 mg/m 3
using an 85-L air sample by lab testing with spiked filters and generated atmospheres of CaO, verified
by EDTA titration [1]. Collection efficiency of 1.00 was determined for the sampling device. Precision
and accuracy data are given on page 7020-1. An additional check showed an estimated LOD of 1 µg
Ca per sample [3].
REFERENCES:
[1] Documentation of the NIOSH Validation Tests, S205, U.S. Department of Health, Education, and
Welfare, Publ. (NIOSH) 77-185 (1977).
[2] V. 3, S205, U.S. Department of Health, Education, and Welfare, Publ. (NIOSH) 77-157-C (1979).
[3] User check, UBTL, NIOSH Seq. #3990-Q (unpublished, November 29, 1983).
[4] Ibid, NIOSH Manual of Analytical Methods, 2nd ed., Vol. 5, P&CAM 173, U.S. Department of Health,
Education, and Welfare, Publ. (NIOSH) 79-141 (1979).
Mark Millson, NIOSH/DPSE and R. DeLon Hull, Ph.D., NIOSH/DBBS; S205 originally validated under
NIOSH Contract CDC-99-74-45.