Marquette Dash 3000-4000 - Service Manual

Download as pdf or txt
Download as pdf or txt
You are on page 1of 201

SERVICE AND INSTALLATION MANUAL

M189_EN: pag 1 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

M189_EN: pag 2 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

Index:
1. Safety.................................................................................................... 10
1.1. Convensions ..................................................................................................... 10
1.2. Safety systems.................................................................................................. 15
1.3. Resuming operation after an emergency stop .................................................. 17
1.4. General Safety precautions............................................................................... 19
1.5. Protection and safety regulation........................................................................ 19
1.6. Warnings and precaustions............................................................................... 19
1.7. General safety warning ..................................................................................... 21
1.8. Radiation protection warnings ........................................................................... 21
1.9. Electrical safety device ..................................................................................... 21
1.10. Exsplosion hazard ............................................................................................. 22
1.11. Risk associated witj cleaning the device ........................................................... 22
1.12. X-ray emission safety ........................................................................................ 22
1.13. Regulations, laws and normative reference ...................................................... 22
2. General Information............................................................................. 23
2.1. Labels ............................................................................................................... 23
2.2. Technical data................................................................................................... 31
3. Panoramic view of the system ............................................................ 60
3.1. Description of the system .................................................................................. 60
4. Transport and moving ......................................................................... 65
4.1. Contents of the delivery .................................................................................... 65
4.2. Packing characteristic ....................................................................................... 65
4.3. Components inside the gantry packing ............................................................. 66
4.4. Components inside the AWS packing ............................................................... 67
4.5. Lifting and moving of packing............................................................................ 68
4.6. Storage of packed equipment ........................................................................... 69
4.7. Removing the gantry from the pack .................................................................. 71
4.8. Removing the AWS from the pack .................................................................... 72
4.9. Disposal of the packing ..................................................................................... 72

M189_EN: pag 3 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

4.10. Instructions for the handling, storing and packing of digital detector ................. 73
5. Installation ........................................................................................... 77
5.1. Ambient requirements conditions ...................................................................... 77
5.2. Preliminary radiology room inspection .............................................................. 78
5.3. Preliminary mechanical clamping inspection..................................................... 78
5.4. Installation procedure ........................................................................................ 79
6. Preliminary operating check ............................................................... 87
6.1. Checks before using the unit............................................................................. 87
6.2. Initial start-up .................................................................................................... 88
6.3. Troubleshooting before initial start-up ............................................................... 90
6.4. Quality checks before clinical application .......................................................... 92
6.5. Connection to the DICOM ntework ................................................................... 92
6.6. Operator instructions ......................................................................................... 93
7. Method of use ...................................................................................... 94
7.1. Equipment ......................................................................................................... 94
7.2. Acquisizione Work Station .............................................................................. 109
7.3. System Messages........................................................................................... 114
7.4. Patient Registration ......................................................................................... 115
7.5. Exposure data setting ..................................................................................... 117
7.6. Performing the exposure ................................................................................. 118
7.7. Printing the exam images ............................................................................... 122
7.8. Concluding the exam ...................................................................................... 122
7.9. Manually sending images to other DICOM workstation .................................. 122
8. Image quality assessment ................................................................ 124
8.1. AEC Test for Installation ................................................................................. 125
8.2. Dose Calibration check ................................................................................... 133
8.3. Image Honogeneity and defective element evaluation .................................... 134
9. Components diagnosis ..................................................................... 140
9.1. Geometry of acquisition setting ....................................................................... 140
9.2. Detector .......................................................................................................... 141

M189_EN: pag 4 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

9.1. X-ray tube ....................................................................................................... 151


10. Electronic boards.......................................................................... 161
10.1. Electronic boards topographic diagram ........................................................... 161
10.2. I008 – Collimator board ................................................................................... 162
10.3. I009 Compressor board .................................................................................. 163
10.4. I009PB Compressor Piggy Back board ........................................................... 164
10.5. I010 Detector .................................................................................................. 165
10.6. I011 Biopsy board ........................................................................................... 166
10.7. I011AC Biopsy cabling switching .................................................................... 167
10.8. I012 Gantry boards ......................................................................................... 167
10.9. I013 SELEMA actuation .................................................................................. 168
10.10. I014 Gantry keyboard ..................................................................................... 168
10.11. I015 Tube keyboard ........................................................................................ 168
10.12. I017 Wireless foot pedals power board ........................................................... 169
10.13. I020 Power switching board ............................................................................ 169
10.14. I021 Operator control table command board .................................................. 170
10.15. I022 Operator control table Cabling switching board ...................................... 170
11. Maintenance .................................................................................. 171
11.1. Warnings on routine maintenance .................................................................. 171
11.2. Cleaning .......................................................................................................... 172
11.3. Preventive maintenance program ................................................................... 173
11.4. Disposal and scrapping of the obsolete equipment ......................................... 175
12. Breakdowns and diagnosis .......................................................... 176
12.1. Introduction ..................................................................................................... 176
12.2. Alarms management ....................................................................................... 177
12.3. List of alarms................................................................................................... 180
13. Interventions on the equipment ................................................... 181
13.1. Introduction ..................................................................................................... 181

M189_EN: pag 5 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

IMS S.r.l. All rights reserved.

No part of this manual can be reproduced, transmitted, transcribed or filed in a retrieval system or
translated into other languages in any form with any means without written consent from IMS S.r.l.
The buyer may reproduce copies for personal use.

This manual is considered an integral part of the equipment. It must be immediately replaced with
another copy if any part of it is ruined or illegible.

Before performing any type of operation on the equipment, IMS S.r.l. requires that anyone of any
title involved in using the equipment carefully read the entire contents of this manual, paying
special attention to the important warnings.

IMS S.r.l. may not be held responsible for the improper use of the equipment, and for damages
caused by unreasonable operations.
The equipment must only be used to meet the needs for which it was expressly designed. Any
other improper use is considered dangerous.

IMS S.r.l. is considered responsible for the equipment only if it is in its original configuration
determined in the design phase. Any modifications to the structure and the equipment’s operating
cycle must by expressly authorized by the IMS S.r.l. Technical Department.

IMS S.r.l. recommends that only original replacement parts be used and therefore cannot be held
responsible for any damages caused after the use of non-original replacement parts.

IMS S.r.l. reserves the right to modify the design and to make marketable improvements without
forewarning customers who already possess similar models.

IMS S.r.l. is considered responsible for the descriptions written in Italian. The translations cannot
be fully verified, therefore if the buyer finds a discrepancy in the text, this person should refer to the
Italian version and possibly contact our Technical Documentation Office, which will make any
necessary changes.

IMS S.r.l. would like to thank you for choosing one of our machines and we are sure that it will fully
satisfy your needs for a long period of time.

IMSS.r.l.
Internazionale Medico Scientifica
Via Sagittario 5 - 40037 Pontecchio Marconi - Bologna - Italy
Tel ++39 051 846851 - Fax ++39 051 846856
e-mail: Export Dept.: imscomm@imsitaly.com
Tecnical Dept.: imstech@imsitaly.com
http://www.imsitaly.com

M189_EN: pag 6 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

Intended use

The Giotto Class and its accessories are designed and built mammography examination for the
screening in modality of Full-Field Digital mammography (FFDM) and for diagnosis in modality of
Digital-Breast Tomosynthesis (DBT) for breast cancer detection and cannot be used for other
examinations or for purposes other than those specified by the Manufacturer. Its practical use is
therefore exclusively limited to personnel with medical training.
Using the accessory devices Smart Finder and Flexi table the system is able to perform
stereotactic biopsy examination with the optional possibility to perform the examination with patient
in prone position.

CE: 1936

CLASSIFICATION AND COMPLIANCE


Class I
IEC 60601-1
Type B
Medical Device Directive 93/42/CE Class II B

DOCUMENT N°: M189_EN


REVISON: 0.1
DATA: 01/07/15

REVISIONS OF THE MANUAL


N° REV. DATE Purpose of the revisions Modified PAGES Modified SECTION
01 01/07/15 Initial release

Please comply with that reported in Section 1 - “Safety”.

Carefully read Section 1 - “Safety” before operating the equipment.

MANUFACTURER’S WARNINGS

This product has the CE marking in conformance with the provisions outlined in Annex II of
93/42/EEC of 14 June 1993 concerning medical devices.

The CE marking is only valid for technical medical products/medical devices put on the market
during the validity of the EC directive indicated above.

If modifications are made to the product without our authorization, the declaration will no longer be
valid.

The original version of this manual was drafted in Italian.

M189_EN: pag 7 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

DO YOU HAVE ANYTHING TO REPORT WITH REGARD TO THIS MANUAL?

Your opinion means a great deal to us!

We strive to constantly improve the documentation of our equipment. To help do this, we give you
the opportunity to directly inform us of any need, suggestion or comment relating to this instruction
manual.

Use the following number to send notices via FAX:


• +39 051 846851.

If you prefer to communicate via e-mail, please use this address:


• imstech@imsitaly.com

In this case, please indicate the complete print reference code reported on the third page.
Thank you for your collaboration.

What I would like to report:


______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
______________________________________________________________________________
Non-mandatory data
Name _________________________________________________________________________
Hospetal_______________________________________________________________________
City/Country ____________________________________________________________________
E-mail _________________________________________________________________________
Tel./Fax _______________________________________________________________________
Number of Fax pages ____________________________________________________________

M189_EN: pag 8 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

M189_EN:Safety pag 9 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

1. Safety
1.1. Conventions
In order for the manual to be read quickly and rationally, symbols have been employed for
highlighting practical advice, simple information or situations in which great care must be taken.
Said symbols can be found alongside a section of text (and therefore refer only to that text),
alongside a figure (and refer to the subject illustrated in the figure and to the relative text) or at the
top of the page (in which case they refer to all the subjects treated on that page).

Pay maximum attention to the meaning of the symbols: their aim is not to have to repeat technical
concepts or safety warnings and therefore should be considered as proper “reminders”.
Thus, refer to the list of symbols whenever doubts arise as to their meaning. The symbols shown
in the following pages are not found on the device or its accessory parts; they are only present
in this manual. The series of the manuals supplied with the device usually contains some
documents not made up by I.M.S. S.r.l., which could use edition symbols that are graphically
different from the ones shown in this publication. Therefore, it is advisable to consult all the
documents supplied in order to “store” all the subjects for which the symbols themselves have
been used.
It is important not to confuse the edition symbols on the device with the “safety” plates, which are
applied in predetermined points on the device, auxiliary units, etc...

1.1.1. Structure of the manual

Sections This manual is made up of various sections, the titles of which are reported in the
heading.
Paragraphs Every section can have one or more paragraphs, the title of which is reported in
the text below the heading.
Page The page footer shows the progressive page numbers.
numbers

M189_EN:Safety pag 10 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

1.1.2. Text formatting


This manual contains certain text formats that are helpful in understanding the function of the text
more quickly.
The following characters were used:
Instruction  Fornisce istruzioni sul corretto utilizzo dell’apparecchiatura.
Questo testo è preceduto da un rombo.
LIST OF  This text is preceded by a dot.
EXPLANATORY Divides an instruction or a list into various sub-parts.
TEXT – This text is preceded by a dash.
REFERENCE  Refers to more detailed instructions reported in another page
 of the manual or another document.
 This text is marked by an arrow.
NOTE This is used in two ways:
 highlights important safety information, without which there is an
immediate risk
 contains a summary of the main information regarding a subject matter
This text is highlighted inside a grey box.
Menu The names of the menus are in bold type.

1.1.3. Warning, caustions and information


 The icons in a box give information .
 The symbols in a triangle are DANGER/WARNING symbols.
 The symbols in a circle represent an OBLIGATION/PROHIBITION..

1.1.3.1. Warning
Warnings indicate the possible presence of risks for the health and safety of patients, operators
and third parties.
WARNING
Firstly, the source of the danger is indicated.
The possible consequences are then reported.
Finally, you are informed on how to avoid the danger.
LIFT ONLY FROM THE TOP
Operations that require the use of qualified personnel and specific equipment, and the
respect of the conditions stated by the Manufacturer and current regulations.

M189_EN:Safety pag 11 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

1.1.1.1. Caution
This indication reports that an incorrect command could cause minor injuries or damage to the
equipment.
CAUTION
Firstly, the source of the danger is indicated.
The possible consequences are then reported.
Finally, you are informed on how to eliminate the danger.
NO!
Operations to be absolutely avoided.

1.1.1.1. Information
This indication provides further explanations about an issue.
INFORMATION
Il relativo testo è scritto in corsivo dentro il box grigio
EQUIPMENT TURNED OFF
With electrical power supplies isolated

1.1.1.1. Safety symbols


OBLIGATION symbols
Use PROTECTIVE GLOVES
This symbol indicates that the maintenance engineer must wear protection
gloves to avoid injuries.
Use PROTECTIVE GLASSES
This symbol indicates that the maintenance engineer must wear protective
glasses to avoid injuries
PROHIBITION symbols
No MAINTENANCE/REPAIRS ON MOVING MEMBERS
It is forbidden to repair, adjust, clean or lubricate moving members of the
machine
HAZARD/CAUTION symbols
CAUTION: ELECTRIC SHOCK HAZARD
It indicates the presence of electric devices and powered components and
warns the operator against electric shock hazards.
The electric maintenance engineer is the only operator that is qualified to carry
out adjustments or maintenance of the electric cabinets and junction boxes
CAUTION!
It indicates the presence of ionizing radiation.

M189_EN:Safety pag 12 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

1.1.1.1. Manuals symbols


SYMBOLS FOR THE MANUALS
THIS OPERATOR MANUAL
Descriptions inherent to this manual.

SUBCONTRACTING MANUALS
Consult the attached documents relating to subcontracting.

1.1.1.1. Qualification of the personnel in charge with operation of the


equipment
The operators that have access to the equipment must be specifically qualified, trained and
responsible for the tasks they are in charge with.
A description of the professional qualifications of the operators and maintenance personnel is
given below. Each profile is graphically described by an icon.

OPERATORS
Radiological technician: the person in charge of preparing the equipment,
positioning the patient and performing the mammography examination,
according to the indications given by the doctor.
Doctor: the person who visits the patient in advance in order to decide the
examination method. This person observes the plates taken by the radiological
technician. He/she carries out the invasive operations (biopsy) on the patient.
Mechanical maintenance personnel: qualified technician able to intervene
on the mechanical parts of the equipment to perform any necessary
maintenance interventions and repairs. This person is not qualified to work
on live-electrical circuits.
Electrician: qualified technician able to carry out any necessary interventions
of an electrical or electronic nature. This person is qualified to work on live-
electrical circuits.
Manufacturer’s technician: qualified personnel provided by the Manufacturer
to perform complex operations under particular situations or when agreed
upon with the Buyer

M189_EN:Safety pag 13 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

The names and data of all the patients and devices used in this manual as examples are fictitious.
Any similarity or correspondence to the actual names of people or institutions are purely
coincidental.
All the parameters and images reported in this manual are examples. The parameters seen in your
system are determining factors.

INFORMATION
In some figures the equipment may be highly equipped or outfitted with optional
accessories.

1.1.1. Regulations and laws


If there are regulations with legal implications for the installation and/or use of the device, the
installer and user must respect them. The national standards must be respected in every country.
With the exception of that given in this manual, the values can be set based on the national
standards.
This product has the CE marking in conformance with the provisions contained in directive
93/42/EEC of 14 June 1993 concerning medical devices,

Personal data are protected.


Please respect the regulations pertaining to this matter. During the manufacturing phase, the
equipment is checked according to applicable product standards.

INFORMATION
A test report shall be filed in I.M.S. S.r.l. archive.

The tests to be performed at the equipment installation site are the responsibility of the customer.

M189_EN:Safety pag 14 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

1.2. Safety systems


1.2.1. Emergency stop buttons
Three emergency stop buttons are installed on the equipment described in this manual. If
pressed, they immediately stop all equipment functions.

2
1) EMERGENCY STOP button located on both sides of the vertical X-ray unit (GANTRY) near two
columns push-button panels.
2) EMERGENCY STOP button located on control table of the acquisition workstation (AWS).
Never use the emergency stop button as a normal stop device to immediately stop all functions of
the equipment, but only in cases of an actual emergency and if malfunctions occur.

After the EMERGENCY STOP button has been activated, keep in mind the following:

INFORMATION
After pressing one of these three emergency stop buttons, the following message
appears on the touchscreen panel

and an acoustic signal is simultaneously emitted.

M189_EN:Safety pag 15 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

INFORMATION
The emergency stop button does not cut power to the workstation 1. The workstation
is also protected by an uninterruptible power supply UPS 2. The UPS takes power
from the equipment’s circuit; thus turning on “0” position the main switch, the UPS
switches to the batteries and the operator is informed of this operation by an acoustic
signal.
In the case of a power interruption, only the AWS acquisition workstation is powered
by the UPS for a few minutes (depending on the state of the battery). When the
battery of the uninterruptible power supply reaches 20%, the workstation
automatically turns off, following the operating system’s safety sequence.

M189_EN:Safety pag 16 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

1.3. Resuming operation after an emergency stop


INFORMATION
After pressing one of these three emergency stop buttons, the following message
appears on the touchscreen panel

and an acoustic signal is simultaneously emitted.


 Release the pressed emergency stop button by turning it (slightly). The acoustic signal will be
silenced and the alarm message on the touch screen panel will disappear ( see Section 12 -
“Breakdowns and diagnosis”).
 To resume operations, press the engage circuits push-button 1 sited on the acquisition
workstation control table switching it to the “OFF” position (LED off). What a few seconds and
then press the engage circuits pushbutton 1 again, switching it to the “ON” position (LED on).
 If the cause of the stop has been eliminated, the device will then be operational again.

M189_EN:Safety pag 17 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

1.3.1. Control and signals


The acquisition workstation has a main switch 1, an engage circuits push button 2 and a touch
screen panel 3 that is used to control and monitor the appliance. The touch screen panel displays
set parameters and alarm messages.

M189_EN:Safety pag 18 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

1.4. General Safety precautions


• The device may be dangerous to the patient and operator if the X-ray exposure values and
operating instructions are not respected. Observe all operating and safety instructions before
performing an X-ray exam.
• Check the following before installation: observance of minimal safety regulations, location and
operating efficiency of the equipment, measuring ambient conditions (temperature, humidity and
lighting) and checking the suitability of the work space.
• Always work with suitable clothing and the required personal protective equipment.
• While X-rays are being emitted, the operator must remain shielded behind the X-ray lead glass
panel supplied with the vertical X-ray unit (GANTRY).

1.5. Protection and safety regulation


• The person who is about to activate the radiological device must check that nobody else is in the
room, that the doors are closed and that all those present are protected by walls or by shielded
doors.
• During X-ray emissions, always stay behind the X-ray lead glass panel.
• Whenever necessary, always wear the personal dose-meter.
• Reduce the size of the beam and the value of the radiological parameters to a minimum (voltage,
current and radiography time) in so far as they are compatible with the diagnostic needs.

1.6. Warnings and precaustions


• The following is strictly forbidden: tampering with the equipment, controls and safety devices.
• Any modifications or tampering, however slight. The use of non-original spare parts, or spare
parts that are not compatible with the quality standards and electro-mechanical characteristics
specified by the Manufacturer, shall relieve the Manufacturer of all liability and shall void the
warranty rights.
• It is strictly forbidden to place and/or leave objects or anything else not foreseen by the
Manufacturer on the equipment that is potentially harmful to the safety of persons or the integrity of
the equipment.
• Carefully read the steps in this manual that are marked: “INFORMATION” and “CAUTION”.
• If you have any requests, always indicate the model and serial number of the equipment.
• Access to the workstation is password protected.
• It is forbidden to modify the configuration of the workstation.

M189_EN:Safety pag 19 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

• It is forbidden to install software different from the original software that was installed by I.M.S.
S.r.l. service personnel.
• The workstation can be connected to a LAN/WAN network, but the user is responsible for
protecting against viruses, intrusions or data leaks.
• The AWS acquisition workstation TFT monitor can only be used for QA before diagnosis. No
clinical evaluation is allowed.
• Diagnosis can be done both through analysis of laser printed images and the images on the
workstation of reporting workstation (RWS) (optional). The monitor type and performance must
comply with the I.M.S. S.r.l. specifications.
• The “DICOM CD” function of the “Raffaello” appliction software is not built for the permanent
filing of data.
This function is only used for data transfer.
• The “Raffaello” application software installed on the AWS acquisition workstation is able to
store a certain number of examinations (the number depends on the capacity of the hard disks).
This archive is not a legal archive, so any lost data is not the responsibility of I.M.S. S.r.l..
CAUTION!
It is strictly forbidden to operate the equipment if it is not permanently fixed to
the floor. Refer to the relative Section.

1.6.1. Mechanical safety


The assembly procedure of some part of the device can be critical for the functionality of the device
itself and to assure the safety performance of fine movement.
Following the list of screws and bolt must be used for each critical mechanical group.
Mechanical group List of screws Reference draw

M189_EN:Safety pag 20 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

1.7. General safety warning


The device was designed and built in compliance with the most severe safety Regulations and left
our production plant in perfect working order. In order to ensure that these conditions are
maintained and that the device is used correctly, the technician must follow the instructions
contained in this manual.
CAUTION!
X-ray machines can be extremely dangerous to the patient and the operator if
the specified protective measures are not scrupulously observed

RESIDUAL RISKS
• Although the equipment described in this manual was built in full compliance with the most
severe safety Regulations, the emission of X-rays is a potential hazard and the equipment
therefore cannot be used or handled by unappointed, unqualified, or unauthorized personnel.
• Over-exposure to X-rays can cause serious damage to the human body.
• Although X-ray radiation is in itself dangerous, the GIOTTO TOMO equipment does not cause
any dangerous conditions if used correctly.
• It is therefore essential that all technical and healthcare personnel be suitably informed and
instructed with regard to the hazards of X-ray exposure.

1.8. Radiation protection warnings


• Exposure to X-rays is a health hazard, and therefore great care should be taken in using
adequate means of protection. Radiation can accumulate over time and its effects can be
manifested after many years.
• Primary radiation is the most dangerous kind, therefore, avoid exposure; any object stuck by
primary radiation produces secondary radiation, which is also highly dangerous.
• The unit is equipped with an protective X-ray temperate glass screen coupled to the console,
which protects people against diffused radiation.

1.9. Electrical safety device


• The device must have an electrical supply line equipped with an earth circuit complying to current
Standards.
For any type of simple maintenance, such as cleaning the equipment, always disconnect it from
the mains to avoid damage to persons and/or the electrical - electronic part of the mammography
equipment.
• To ensure the isolation integrity for the system, use only I.M.S. S.r.l. approved accessories on the
equipment. Any changes to the interconnections must be performed by I.M.S. S.r.l. authorized
personnel.
• To ensure proper isolation maintain a 1.5 meter distance between the patient and any
devices not approved for use in the Patient Area. Devices not approved for use in the patient
area (such as the data management computer, the medical recording workstation, laser printers)
must not be installed in this area (see IEC 60601-1-1).

M189_EN:Safety pag 21 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

CAUTION!
In accordance with EN 60601-1, the unit must be permanently connected to
the electrical mains.
(The protective earth conductor must be connected directly to the protective
earth terminal of the electrical panel).

1.10. Exsplosion hazard


• Gas or flammable vapors cannot be used with this equipment.
• Some types of disinfectants could vaporize, thus forming an explosive mixture. If used, they must
be allowed to disperse before powering up the equipment.

1.11. Risk associated witj cleaning the device


• Gas or flammable vapors cannot be used with this equipment.
• Some types of disinfectants could vaporize, thus forming an explosive mixture. If used, they must
be allowed to disperse before powering up the equipment.

1.12. X-ray emission safety


The device uses two separate methods to assure that the generator doesn’t exceed the maximum
exposure limit.
a) A software counter that includes the anode current over time and stops the exposure when the
maximum time has been reached; the counter is protected by a watch dog circuit.
b) A hardware counter that cuts off the electrical power supply to the generator if the maximum
allowed exposure time is exceeded.

1.13. Regulations, laws and normative reference


If there are regulations with legal implications for the installation and/or use of the device, the
installer and user must respect them. The national standards must be respected in every country.
With the exception of that given in this manual, the values can be set based on the national
standards.
This product has the CE marking in conformance with the provisions contained in directive
93/42/EEC of 14 June 1993 concerning medical devices,
Personal data are protected.
Please respect the regulations pertaining to this matter. During the manufacturing phase, the
equipment is checked according to applicable product standards.
INFORMATION
A test report shall be filed in I.M.S. S.r.l. archive.

The tests to be performed at the equipment installation site are the responsibility of the customer.

M189_EN:Safety pag 22 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

2. General Information
2.1. Labels

Keep the hands away

Radiation High Tension

Main Identification label Parts under voltage

Moving part IPX0 Protection class label

Earth point Focal Spot Electrostatic discharge

WEEE Label Component Identification

M189_EN:General Information pag 23 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

2.1.1. Labels on the X-ray unit

M189_EN:General Information pag 24 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

M189_EN:General Information pag 25 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

2.1.2. Labels on the operator console (AWS)

M189_EN:General Information pag 26 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

2.1.3. Labels on the internal component


Label
Label Component Description
ID

1 IMS accessory Identification data

High-Voltage
2 Identification data
Generator

3 Identification data

X-ray tube:

4 Focal spot position

M189_EN:General Information pag 27 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

Label
Label Component Description
ID

4 Earth point

Vertical
5 Identification data
movement motor

Identification and
6 Anti-scatter grid
specifications data

INSERIRE IMMAGINE CON Identification and


7 Collimator
IDENTIFICAZIONE E HVL specifications data

8 Monitor Identification data

M189_EN:General Information pag 28 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

M189_EN:General Information pag 29 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

M189_EN:General Information pag 30 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

2.2. Technical data


The equipment can only be used in rooms designed for medical use.
The current national standards apply. In order to ensure the safety of users, patients and third
parties, it is recommended that you fully comply with the standards mentioned in this document
(unless they are inconsistent with the national standards).

THIS IS A DRAFT VERSION; ALL THE DATA WITH THE * MARK ARE UNDER REVISION.

2.2.1.1. X-ray unit and operator console storage conditions


Min Max
Humidity 10% 80%
Temperature -10 °C = 14 °F 80 °C = 176 °F
Atmospheric pressure 700 hPa 1060 hPa

WARNING
If the equipment is stored while packed, do not stack the packing crates on top of one
another.

2.2.1.1. Digital image receptor storage conditions


Min Max
Humidity 10% 95%
Temperature 5 °C = 41 °F 45 °C = 113 °F
Atmospheric pressure 700 hPa 1060 hPa

WARNING
If the equipment is stored while packed, do not stack the packing crates on top of one
another.

CAUTION
Do not stack other crates on top of the digital detector crate!!!

CAUTION
If the temperature is below 5 °C or is over 45 °C, there is a risk of permanent damage
to the digital detector.

M189_EN:General Information pag 31 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

2.2.1.2. Operating conditions


Temperature range From +12 °C to +27 °C.
If the detector internal temperature drops below 5 °C or exceeds
45 °C, there is a risk that the detector may be permanently
damaged.
Maximum ratio 0 °C in 20 minutes
Relative humidity Up to 95% non condensing
range
Maximum warm-up 10 minutes
time for operation

2.2.1.3. Enclosures protection class


The Giotto Class is a IPXO protection class device. Device is not sealed, not suitable for
use in the presence of flammable anesthetic mixture from the air, oxygen, or nitrogen
oxides.

2.2.1.4. Electrical power supply


230 VAC ± 10%, 50/60 Hz ± 5%,
Tension single phase, permanent connection
Maximum Power 25 A – 230 VAC duty cycle 3s ON 60s OFF*
Fuses 40 A*
Permanent connection with Required network filter with minimum attenuation of 10
protective conductor dB*

M189_EN:General Information pag 32 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

2.2.1.5. X-ray unit specifications


Total 630 mm (850 mm with consequence of inclination
Vertical movement
movement)
Potter/Bucky - Floor: Min: 630 mm
(unit in vertical position) Max: 1260 mm
Potter/Bucky - Floor: Min: 1360 mm
(unit in horizontal position) Max: 1990 mm
Gantry rotation range - 177.0° / + 177.0° (Motorized)
Gantry inclination range - 15.0° / + 90.0 ° (Motorized)
Angular range of emission - 14.5° / + 14.5 ° (Motorized)
Source-to-Image Distance (SID) 685 mm
Source-to-Breast Support Distance 668 mm
Magnification factor 1.8 X (optional)
Dimension of x-ray unit
147.6 x 80.0 x 149.8 cm (in vertical position)
(H x W x D)
Dimension of operator console unit
195.0 x 88.6 x 66.0 cm *
(H x W x D)
Weight of x-ray unit 270 kg
Weight of operator console unit 120 kg
Weight of Flexi Table 250 kg

M189_EN:General Information pag 33 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

SOURCE-to-IMAGE DISTANCE (SID), DIRECT FOCAL DISTANCE, and SOURCE-to AXIS


ROTATION DISTANCE (SAD)

M189_EN:General Information pag 34 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

2.2.1.6. Breast compression system


Actuator type Stepper Motor - Pedal or Manually with control knob
Maximum compressor force 200 N (motorized) 300 N (manual)
Minimum Force for AEC activation 5N
Compression Force accuracy ± 1 kg
Compression Thickness ± 2 mm
Release after exposure Automatic or manually in case of inhibition by operator
Compression velocity 2 mm/s during approaching 1 mm/s during compression *
Compressor plates material PET-G
Compressor plate maximum flexion < 5 mm for asymmetric compression
Compressor plate absorption 16.5% ÷ 0.003 mm Al for 28 kV W/Ag

M189_EN:General Information pag 35 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

2.2.1.7. High Voltage Generator specifications


Manufacturer IMD Generators
Model HF1 Mammo 8kW
Input power supply
(Inverter + Filament + Single phase 230VAC 50/60Hz
Stator power supply)
11 kVApk (max 32Apk) Depending on the exposure
Rated input max power
parameters
Max input current
32A 400 Vac aM 20 kA
protection breakers
Un = 230 Vac 50/60 Hz
Recommended thermo-
magnetic circuit breaker In = 32 A;
Number of poles = 2
Protection class C as per IEC-EN 60898
Power output Single phase 20 kHz (Max 210 VACrms)
Filament power board
Single phase 16 kHz (Max 5.5 Arms)
output:
Stator driver power board
Single phase + shifted phase 150 Hz (Max 230 VACrms / 7Apk)
output:
Inverter
Pulsed load output power 8 kW (200mA @ 40 kV 0,1 sec)
kV Range 22 - 49 kVDC
Max Ripple < 1%
Max RAD mode load
200 mA @ 40 kV (0,1 sec)
output
Max frame per second rate 30 fps
Filament
Max load output power 65VA (5A @ 13V)
Stator driver
Steady state nominal
280 VA
power
Maximum surge power
1600 VA
(launch max 1msec)

M189_EN:General Information pag 36 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

High Voltage Generator Electrical connections

M189_EN:General Information pag 37 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

2.2.1.8. High Voltage Generator Electric characteristics and I/O (power and
signals)

All the control unit interface connectors are from Molex (except power GND).
In order to connect the unit, following headers are needed:
CP1 = 2ways Minifit SR 42816 series connector
CP2 = 3ways Minifit SR 42816 series connector
CP3 = 8ways (4x2) Minifit Jr 5557 series connector
CP4 = 3ways Microfit 43645 series connector
CP5 = 4ways (2x2) Microfit 43645 series connector
CP6 = 6ways (3x2) Microfit 43020 series connector
GND = M5 brass nut

Terminals Input / Output characteristic


Input power supply
CP1-1,2 Single phase
(Inverter + Filament + Stator power
230Vac 50/60Hz
supply)
11kVApk
(max 32Apk)
Rated input max power
Depending on the exposure
parameters
Max input current protection breakers 32A 400Vac aM 20kA
Un = 230Vac 50/60Hz
In = 32A
Recommended thermo-magnetic
Number of poles = 2
circuit breaker
Protection class = C
as per IEC-EN 60898
Connected to the
GND Screw GND
Control unit chassis
Single phase
CP2-1,2 Power output 20kHz
(Max 210Vacrms )
CP3,1: (LF) Filament power board output:
Single phase
CP3,2: (SF) (LF) = Large Focus
16kHz
CP3,3: (CM) (SF) = Small focus
(Max 5.5Arms)
CP3,4: GND (CM) = Common
CP3,5: (P) Stator driver power board output: Single phase +
CP3,6: (S) (P) = Main phase shifted phase
CP3,7: (C) (S) = Shifted 150Hz
CP3,8: GND (C) = Common (Max 230Vacrms / 7Apk)
M189_EN:General Information pag 38 of 201
Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

Terminals Signal Block diagram


1,3 INPUT
(CP4-2: GND) CAN BUS

CP5-1,2 INPUT
Digital*
Command
X-Ray
CP5-3,4 OUTPUT
Digital*
HV > 85%

CP6-4,5 INPUT This input manages the normally closed contact which comes
Digital from the safety switch in the HV transformer.
Thermal safety
switch
CP6-1 mA+ INPUT These inputs have to be connected to the load as indicated in
CP6-2 mA- Analog the chapter 8.3
Feedback Anodic
mA

2.2.1.9. High Voltage Generator: conditions of storage and usage

Transport and storage conditions


Range of temperature 0° – 50° C
Relative humidity (non condensing) 20 – 90 %
Atmospheric pressure 700 – 1060 hPa
Working conditions
Range of temperature 10° – 40° C
Relative humidity (non condensing) 30 – 75 %
Atmospheric pressure 700 – 1060 hPa

M189_EN:General Information pag 39 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

2.2.1.10. X-ray tube assembly

M189_EN:General Information pag 40 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

Manufacturer I.A.E.
Model XK116T
Focal Spot dimension 0.1-0.3 mm
Anode Angles 10°-16°
Anode rotation speed 3000-10000 rpm
Anodic Material RT-TZM
Inherent Filtration 0.5 mm Be
Termal Anodic capacity 225 kJ - 300 kHU
Continuous anode heat dissipation 715 W
Maximum anode heat dissipation 750 W
Equivalent anode input power 100 W - 38% of max
Dimensions 110 x 155 x 305 mm
Weight 5.5 kg
Temperature condition
-10 / + 80 °C
for transportation and storage
Humidity condition
max 80 %
for transportation and storage
Maximum High Tension 49 kV
Maximum heat dissipation with fan 300 W
Maximum leakage radiation
63 uGy/h - 7 mR/h
at 1 m from focal spots
93/42/CEE Classification IIb
IEC 60601-1 equipment class I
IEC 60601-1 equipment type B

M189_EN:General Information pag 41 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

M189_EN:General Information pag 42 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

M189_EN:General Information pag 43 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

M189_EN:General Information pag 44 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

M189_EN:General Information pag 45 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

Range kV/mA
kV Large Focus (0.3mm2) - mA Small Focus (0.1mm2) - mA

22 110 41

23 115 42

24 120 43

25 125 44

26 130 45

27 135 46

28 140 47

29 145 46

30 150 45

31 155 44

32 150 43

33 145 42

34 140 41

35 135 40
* Data under revision

Range current-time product (mAs):

- LARGE FUOCUS (0.3 mm2) 2-600 mAs with steps of 1 mAs


- SMALL FOCUS (0.1 mm2) 2-185 mAs with steps of 1 mAs

The Giotto Class unit does not apply the loading factors from TABLES R10 – R20.

M189_EN:General Information pag 46 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

2.2.1.11. X-ray beam filtration


Material Silver (Ag)
Thickness 0.0050 ± 5 mm
Purity ≥ 99.9 %
W/Ag (50 m)
25 kV = 0.48 ± 0.02 mm Al *
Half-Value Layer (HVL)
28 kV = 0.53 ± 0.02 mm Al *
31 kV = 0.57 ± 0.02 mm Al *

2.2.1.12. Radiation Output

Reproducibility < 5%

Linearity R2> 0.999

kV accuracy ± 1 kV of nominal value

mAs accuracy ± 5% o ± 1mAs of nominal value

~ 1 mGy for 26 kV W/Ag 20 mAs at SID distance


Dose output
(~ 7.0 mGy/s at 685 mm from source); *

2.2.1.13. Beam Limiting device


Automatic collimation depending on the compressor format and angle of emission.
The image dimension is depending on the collimation format (collimation diaphragm never
visible)
Light field source LED + mylar mirror

Light field intensity > 160 lux

Single pressure of keyboard button or when compressor is moving


Light field activation
with pedal

Light field duration 20 s

X-ray field / Light field


< 2 % of SID according to IEC 60601-2-45
misalignment

M189_EN:General Information pag 47 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

2.2.1.14. Radiation Protection

Operator protection  Anti-X ray Protective barrier

Anti-X ray protective barrier


Temperate glass 760 x 1960 x 5 mm
material and dimension
The protective barrier installed on the control cabinet guaranties an
Equivalent filtration equivalent absorption of 0.1 mm Pb as required by IEC 60601-1-3
and IEC 60601-2-45 standards.
 Patient face protection
Patient protection  Automatic collimator
 AEC system

2.2.1.15. Acquisition Work Station (AWS)

Operating system WINDOWS 7 Embedded SP1

Processor INTEL DUAL- CPU

Memory RAM 24 GB

Capacity hard drive 1TB or superior, DVD READER/BURNER

Graphic board Integrated display port

1600x1200 – 20” da 180 cd/m2


LCD Display
Optional Diagnostic 3 MP

Network interface 3GB Ethernet

Remote diagnostic Internet


 Machine state
 Work-list modality
 Patient data management
 Brightness and contrast control
G.U.I.
 Magnification Lens
 Tomosynthesis scroll tools with kinetic modality
 Q.A. test tools
 Biopsy examination modalities
Digital Images Formats DICOM 3.0 – See Informance Statement

Verification SCU/SCP Store SCU/SCP Query /retrieve SCU Print


DICOM Functions
SCU WORKLIST CU STORAGE COMMITMENT SCU MPPS SCU *
Image processing and
RAFFAELLO software
reconstructions

M189_EN:General Information pag 48 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

2.2.1.16. Digital image receptor


Manufacturer Analogic
Model AXS-2430
Sensor Physical
239.36 x 304.64 mm
Dimensions
Active Area 2816 x 3584 pixels
Image matrix 2812 x 3580 pixels
N.B. The first two lines and columns at the edge of the active area are declared dead
Pixel size 0.085 x 0.085 mm
Geometric Fill Factor 88%
Detection Method Direct Conversion (DR) using Amorphous Selenium (a-Se)
Read out technology TFT array
~ 3.0 mGy (mammographic context)
Saturation Dose
~ 0.7 mGy (tomosynthesis context)
Response Linear (R2 > 0.99)
DQE @ 1lp/mm > 50 %
DQE @ 5.8 lp/mm > 20 %
MTF @ 1lp/mm > 90 %
MTF @ 5.8 lp/mm > 40 %
Lag < 0.005 after 30 s
Ghost 0.05 (Euref Method)
≥ 4 Masses
≥ 3 Fibres
ACR Mammography
≥ 3 Speck goups
Accreditation Phantom
For every modality of acquisition
(MAMMO, TOMO, COMBO) using the AEC system
ADC Bit depth 16 bits min
13 Bit for mammographic and projections images
Image bit depth
14 Bit for reconstruction
19 MB for mammographic and projections images
Image dimension
Depending on the volumetric reconstruction dimension
Maximum defective pixels < 5000
Maximum correctable
< 10
defective lines/columns
Maximum correctable
150 cluster (max 8 pixel)
defective cluster

M189_EN:General Information pag 49 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

2.2.1.17. Automatic Exposure Ccontrol (AEC) system


The compressed breast thickness determinate the optimum values
of the kV and mAs for the pre-exposure (for tomosynthesis scan the
Principle of function first exposure id used as pre-exposure) Analyzing the image
received the AWS determinates the maximum density inside the
breast and select the mAs needed.
The analysis area is rectangular, aligned with chest wall and laterally
centered.
Sensible areas 1. 2D Mammo : 15 cm x 19 cm
2. 3D Tomo : 10 cm x 19 cm
3. Biopsy : 4 cm x 1 cm
- Manual: manual selection of kV, filter, mAs
Techniques
- Automatic: automatic selection of Focus type, kV, filter, mAs
Short term reproducibility < ± 5%
Long term reproducibility < ± 10%

2.2.1.18. Anti-scatter grid


Manufacturer JPI Healthcare solution
Model 2100
Type linear focused grid
Ratio 4:1
Line density 31 lp / cm
Focal Distance 65 cm
Interspace material Pure carbon ( C )
Absorbed material Lead alloy ( Pb )
Cover material CFRP
Application limit (f1 ~ f2) 43 ~ 180 cm
Transmission of primary
73%
radiation (Tp)
Transmission of scatter
26%
radiation (Ts)
Transmission of total
58%
radiation (Tt)
Grid selectivity (Σ) 2.81 ± 5%
Contrast improvement
1.26 ± 5%
fator (K)
Grid exposure factor (B): 1.72 ± 10%

M189_EN:General Information pag 50 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

2.2.1.19. Dimension of the X-ray unit

Max Height 2320 mm


Isocentric height 30 mm over the detector support
Breast support height 70 - 145 (± 2) cm
Source-to-image distance (SID) 690 mm
Gantry rotation range +180° (in anticlockwise) -120° (in clockwise)

Weight 450 kg

M189_EN:General Information pag 51 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

2.2.1.20. Dimension of the opreator console unit

Dimension (H x W x D) 1982 x 886 x 660 mm


Weight 180 kg

M189_EN:General Information pag 52 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

2.2.2. Electrical connections

Temporary images (is not the final version)

M189_EN:General Information pag 53 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

2.2.3. Warning on electromagnetic compatibility (EMC)


The unit is certified as a fixed installation electro medical device and was built in conformity with
the applicable provisions of directive 93/42/EEC concerning medical devices modified by directive
2007/47/EC.

The unit is built in compliance with the IEC 60601-1-2 standard.

CAUTION
In accordance with EN 60601-1, the unit must be permanently connected to the
electrical mains.
(The protective earth conductor must be connected directly to the protective earth
terminal of the electrical panel).

CAUTION
THE UNIT AND IT’S ACCESSORIES MUST ONLY BE INSTALLED AND USED IN
SHIELDED ROOMS (MINIMUM SHIELDING RATIO 20 dB).

INFORMATION
The fixed cables for systems and devices that cannot be disconnected by the user
are not listed. These cables are part of the system and were always taken into
consideration when measuring the EMC. The device or system would not function
without these cables.
CAUTION
Use of accessories, transformers or cables different than those indicated.
Possible increase in the emissions or reduction in the resistance to
disturbances for the device or system!
Use only transformers and cables sold by the manufacturer of the device or system
as replacement parts for the internal components.

M189_EN:General Information pag 54 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

2.2.4. Manufacturer’s declaration and guidelines - Electromagnetic emissions


The system is intended for operation in an electromagnetic environment indicated below. The
customer or user of the system must ensure that it is used within this environment.

Measurement of the Compliance Guidelines - Electromagnetic Environment


emission
disturbances
Conducted RF Group 1 The system uses HF energy only for its internal
Radiated RF functioning. Therefore its HF emissions are very low and
CISPR 11 are not likely to cause any interference in nearby
electronic equipment.
Conducted RF Class A The unit/system is suitable for use in all environments
Radiated RF other than domestic and those directly connected to
CISPR 11 the public low voltage power supply network that
supply buildings used for domestic purposes.
The system has a nominal input current greater than 16 A
per phase.
Harmonic emissions Not applicable
IEC 61000-3-2
Voltage Not applicable
fluctuations/flicker
IEC 61000-3-3

M189_EN:General Information pag 55 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

2.2.5. Manufacturer’s declaration and guidelines - Immunity to


electromagnetic interferences
The system is intended for operation in an electromagnetic environment indicated below. The
customer or user of the system must ensure that it is used within this environment.
Electromagnetic Test level - Compliance Electromagnetic environment -
interference tests IEC 60601 level guidelines
Electrostatic discharge ±6 kV ±6 kV Floors should be wood, concrete or ceramic
(ESD) Contact Contact tile. If floors are covered with synthetic
discharge discharge material, the relative humidity should be at
IEC 61000-4-2
least 30%.
±8 kV ±8 kV
Air discharge Air discharge
Electrical fast transient/ ±2 kV ±2 kV The mains power quality must be that of a
burst for power for power typical commercial or hospital environment.
IEC 61000-4-4 supply cables supply cables
±1 kV ±1 kV
for input and for input and
output cables output cables
Surge ±1 kV ±1 kV The mains power quality must be that of a
differential differential typical commercial or hospital environment.
IEC 61000-4-5
mode mode
±2 kV ±2 kV
Common Common
mode mode
Voltage drops, short <5% UT Not applicable The mains power quality must be that of a
term interruptions and (> 95% drop <5 % UT (>95 typical commercial or hospital environment. If
fluctuations in the in UT) for 0.5 % dip in UT) the user of the system requires continuous
supply voltage cycle for half cycle operation during power outages, it is
IEC 61000-4-11 40% UT 40 % UT (>60 recommended that it be powered with an
(> 60% drop % dip in UT) uninterruptible power supply.
in UT) for 5 for 5 cycles The system has a nominal input current greater
cycles 70 % UT (>30 than 16A per phase.
70% UT % dip in UT)
(> 30% drop in for 25 cycles
UT) for 25 <5 % UT (>95
cycles % dip in UT)
<5% UT for 5 s
(> 95% drop
in UT) for 5 s

Magnetic field at the 3 A/m 3 A/m The magnetic fields at mains frequency should
supply frequency (50- correspond to the typical values seen in a
60 Hz) typical business or hospital environment.
IEC 61000-4-8
Note: UT is the alternative current voltage before the application of the test level.

M189_EN:General Information pag 56 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

CAUTION
Use of the device or system immediately next to other devices or stacked
on top of other appliances.
Proper operation cannot be guaranteed!
 If the device or system must operate alongside other appliances or be
stacked on top of them, it should be execute a functional test to check
that it functions properly in this layout.
Portable and mobile HF communication devices should be used no closer to any part of the
equipment, including cables, than the recommended separation distance. The separation distance
is calculated based on the applicable transmission frequency equation.
P is the maximum nominal power rating of the transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation distance in meters (m).
The field strength from the stationary RF transmitter, as determined by a local electromagnetic site
survey a should be less than the compliance level b at each frequency range. Interference may

occur near equipment marked with the following symbol


Electromagnetic Test level - IEC Compliance Recommended separation
interference tests 60601 level distance:
Conducted HF 3 Vrms 3 Vrms d = 1, 2√P
interference from 150 kHz to
IEC 61000-4-6 80 MHz
Radiated HF 3 V/m 3 V/m d = 1, 2√P per 80 MHz - 800
interference from 80 MHz to MHz
IEC 61000-3-2 2,5 GHz d = 2, 3√P per 800 MHz - 2,5
GHz
Note 1: The higher frequency range applies at 80 MHz and 800 MHz.
Note 2: These guidelines may not be applicable in all cases. The spread of electromagnetic
emissions is affected by the absorption and reflection by buildings, objects and people.
A) The field strengths of stationary transmitters, e.g. base stations for radio telephones
(mobile/cordless) and mobile radio equipment, amateur radio stations, AM and FM radio and television
transmitters cannot be predicted theoretically with accuracy. In order to determine the electromagnetic
environment with regard to stationary transmitters, a study of the location should be carried out. If the
field strengths measured at the location where the system will be used exceed the compliance level
given above, the system should be monitored to see that it operates properly. If unwanted performance
characteristics are observed, additional measures may be necessary, such as reorienting or
repositioning the system.
B) At the frequency range from 150 kHz to 80 MHz, the field strengths should be lower than 3 V/m.
CAUTION
PORTABLE AND MOBILE RADIO COMMUNICATION DEVICES CAN
INFLUENCE THE OPERATION OF THE UNIT AND ITS
ACCESSORIES.

M189_EN:General Information pag 57 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

2.2.6. Recommended separation distances between the system and portable


and mobile HF communication devices munication devices

The system is intended for use in an electromagnetic environment in which HF disturbances are
controlled. The customer or the user of the system can help prevent electromagnetic interference.
This can be achieved by maintaining a minimum distance between portable and mobile HF
communication devices (transmitters) and the system, based on the rated maximum output power
of the communication devices.

Separation distance according to frequency of the


Maximum rated output power of transmitter [m]
transmitter (W) da 150 kHz a 80 da 80 MHz a 800 da 800 MHz a
MHz MHz 2,5 GHz
d = 1, 2√P d = 1, 2√P d = 2, 3√P
0,01 0,12 0,12 0,23
0,1 0,38 0,38 0,73
1 1,2 1,2 2,3
10 3,8 3,8 7,3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended separation
distance d in meters (m) can be determined using the equation from the applicable column, where
P is the maximum output power rating of the transmitter in watts (W) according to the transmitter
manufacturer.
Note 1: The higher frequency range applies at 80 MHz and 800 MHz.
Note 2: These guidelines may not be applicable in all cases. The spread of electromagnetic
emissions is affected by the absorption and reflection by buildings, objects and people.

M189_EN:General Information pag 58 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

2.2.7. List of compatible components


The person who connects additional devices to the medical product is considered to be the system
configure and therefore must guarantee under his/her own responsibility that the configuration of
the current versions conform to the main standards (system standard IEC 60601-1-2 and/or other
applicable standards). Please contact the local reference person if further clarifications are
required.

2.2.7.1. Standard accessories


Description Product Code
Compressor Paddle 24x30 cm for Tomosynthesis scan XXXXXX
Face protection for mammo XXXXXX
Face protection for tomo XXXXXX
Foot switches for vertical movement XXXXXX
Foot switches for compressor XXXXXX
Visualization Monitor XXXXXX
Touch-Screen monitor XXXXXX
2MP Monitor XXXXXX
3MP Monitor XXXXXX
Operator Protection (Glass barrier) XXXXXX

2.2.7.2. Optional accessories


Description Product Code
Biopsy unit: Smart Finder 28000-00
Patient support: Flexi Table 29000-00
Review Work Station (RWS) with two 5 MPixel monitors 31010-00
Review Work Station (RWS) with two 5 MPixel monitors + 1 additional
31020-00
LCD 21.3” 2 MPixel monitor
Review Work Station (RWS) with one 10 MPixel monitor 31040-00
Review Work Station (RWS) with one 10 MPixel monitor + 1 additional
31060-00
LCD 21.3” 2 MPixel monitor
Mammographic Compressor Paddle 24x30 cm 25100-00
Compressor Paddle 15x30 cm 25200-00
Compressor Paddle 05x05 cm 25980-00
Compressor Paddle 10x10 cm 25900-00
Magnification Platform with Compressor Paddle 24 x 30 cm 26100-00
Software Raffaello Viewer 31003-00
Software CAD Galileo 31070-00
Software G-View (synthetic 2D image from 3D reconstruction) 31002-00
Supplement foot pedal 25030-00
Wireless foot pedal for vertical movement 25010-10
Wireless foot pedal for compressor movement 25020-00

M189_EN:General Information pag 59 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

3. Panoramic view of the system


3.1. Description of the system
3.1.1. Introduction
1.1.1.1 Field of application
This device is a Full Field Digital Mammography (FFDM) and Digital Breast Tomosynthesis (DBT)
system with an integrated digital image receptor based on amorphous selenium (a-Se) material
and TFT readout method.
Using the accessory device Smart Finder the device is able to perform biopsy examination in
stereotactic or tomosynthesis modalities.
Using the accessory device Flexi Table the Giotto Class permits to perform the biopsy examination
also with patient in prone position.
The equipment is used under the guidance of qualified medical personnel to perform digital
mammograms, that is to say for diagnostic imaging using the contact technique.
1.1.1.2 Targeted mammographic view using the magnification device (optional)
The targeted mammographic view is used to better represent a particular area of the breast.
The result is a better view of a small portion of the breast.
1.1.1.3 Automatic mode for X-ray optimization
The equipment has special automatic functions to optimize the X-ray parameters and therefore
obtain the best possible image quality with the lowest dose level possible.
The Automatic Exposure Control (AEC) system is integrated in the digital image receptor and is
available for all the modalities of acquisition present: Contact mammography, Geometrical
Magnification, Targeted mammographic views, Tomosynthesis scan, Combined modality and
biopsy.
1.1.1.4 Automatic Exposure control (AEC) system for 2D images
Principle of operation: using a short pre-exposure, based on the compressed breast thickness
read by the compression system, an initial image is taken and analyzed for the evaluation of the
densest area inside the breast, for the optimization of the final dose level for the best compromise
between the image quality obtained and the dose absorbed by the patient.
1.1.1.1 Automatic Exposure control (AEC) system for 3D images
Principle of operation: the first projection is based on the compressed breast thickness read by
the compression system, and the first image is taken and analyzed for the evaluation of the
densest area inside the breast for the optimization of the final dose level for the best compromise
between the image quality obtained and the dose absorbed by the patient. To maintain the
symmetry of the acquisition the final projection is made with the same dose level of the first one.
1.1.1.2 Manual mode
If you cannot or do not want to use the automatic exposure control functions, the equipment’s
exposure parameters can also be adjusted manually.

M189_EN:Panoramic view of the system pag 60 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

3.1.2. Over all view of the system


The equipment includes the following components:

1. Floor mount base 4 19


2. Columns for vertical
movement 14
3. Gantry 7 18
17
4. X-ray tube housing 12 16
5. High voltage 15 6
generator
6. Compressor unit
3
7. Collimator system
8. Digital detector
9. Retractable
potter-bucky system 13 8
10. Foot pedal control
for compressor 9
21
11. Foot pedal control
5
for columns
12. Compressor knob 2 20
13. Grab bars
14. Gantry movements
10
keyboards
15. LCD display
1 11
16. Emergency
push buttons
17. Touch screen 24
monitor 23
18. 5Mpixel Monitor
19. Anti X-ray barrier
20. PC (AWS) with
integrated CD reader/writer
21. Keyboard and mouse
22. UPS
23. Patient support
Flexi Table
24. Biopsy device
Smart Finder

M189_EN:Panoramic view of the system pag 61 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

3.1.3. Mammographic X-ray unit

3.1.3.1. Keyboards

1. Sequences of isocentric positioning


2. Manual selection of collimator format
3. Light indicator (automatic switch-off)
4. Gantry vertical movement to the top
5. Gantry vertical movement to the bottom
6. Rotation gantry clockwise
7. Rotation gantry anticlockwise
8. Preset isocentric position
9. Automatic gantry inclination to horizontal position
10. Automatic gantry inclination to vertical position
M189_EN:Panoramic view of the system pag 62 of 201
Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

1. Sequences of isocentric positioning


2. Manual selection of collimator format
3. Light indicator (automatic switch-off)
4. Gantry vertical movement to the top
5. Gantry vertical movement to the bottom
6. Rotation gantry clockwise
7. Rotation gantry anticlockwise

3.1.3.2. Display

Horizontal view (for tilt angle < 70°) Vertical view (for tilt angle > 70°)

1. Height of compressor paddle (in cm)


2. Compression force (in Kg)
3. Selected projection (laterality and point of view)
4. Collimation format selected
5. Gantry angle
6. Tube angle
7. Status
8. Alarms
9. Movement inhibition

M189_EN:Panoramic view of the system pag 63 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

3.1.4. Operator control console

3.1.4.1. Protective barrier


The protective wall prevents the operator performing the exam from being exposed to X-rays.

3.1.4.2. Touch screen panel


see the documentation supplied with the touch screen panel.

3.1.4.3. Visualization Monitor TFT


see the Documentation supplied with the monitor.

3.1.4.4. Workstation con unità CD-R/DVD-R


see Documentation supplied with the workstation.

3.1.4.5. Control table


The following devices are found on the control table:

- X-ray emission button


- Emergency stop button
- Keyboard
- Mouse

3.1.4.1. X-ray emission button


Pressed by the operator to begin the exposure to X-rays.

3.1.4.2. Emergency stop button


 see the “Emergency stop button” paragraph in Section 1 - “Safety”.

M189_EN:Panoramic view of the system pag 64 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

4. Transport and moving


4.1. Contents of the delivery

4.2. Packing characteristic

kg A B C
GANTRY 550 2070 870 1960
AWS 185 930 580 1270
DIGITAL DETECTOR 70 113 107 85
RWS (optional) 120 120 70 100

M189_EN:Transport and moving pag 65 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

4.3. Components inside the gantry packing

INFORMATION
Irrespective of the type of protective packing used for the GANTRY, the packing case
for individual parts or accessories contains the following components:

- Equipment secured to a wooden pallet, properly oriented on supporting brackets dedicated.


- X-ray emission tube, packed separately in a dedicated box.
- Accessories supplied with equipment (face protector), packed in cardboard box.
- Touchscreen panel and TFT monitor, protected by plastic material.
- Pedal controls protected by plastic material.
- Keyboard and mouse, protected by plastic material.

M189_EN:Transport and moving pag 66 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

4.4. Components inside the AWS packing


INFORMATION
Irrespective of the type of protective packing used for the AWS, the packing case for
individual parts or accessories contains the following components:

- Acquisition workstation, protected by plastic material.


- X-ray protective wall, properly protected by plastic material.

M189_EN:Transport and moving pag 67 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

4.5. Lifting and moving of packing

• Fully consider the packing characteristics (dimensions and weight).


• All the information below concerning lifting and handling of the packing, should be considered as
a guide;
these operations must be strictly entrusted to qualified operators who, according to the
characteristics of the
load, lifting equipment, means of transport, and available space, must organize all the operator

INFORMATION
The GANTRY or AWS packing crate can be lifted from above by a crane, hoist,
bridge crane, etc.., or from below by a fork lift truck; in both cases, take great care
(since the contents are “delicate”) and avoid sudden manoeuvres, jerks, impacts,
etc…

CAUTION
The Manufacturer declines all liability for damage of any kind caused by incorrect
lifting or handling.

M189_EN:Transport and moving pag 68 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

4.5.1. Forbidden lifting system

• The lifting systems having the characteristics indicated below, cannot be used:
- Lifting capacity less than three times the equipment’s weight.
- Inadequate mechanical characteristics (e.g. forks too short).
- Load-bearing structures, lifting ropes, cables and belts that are either worn or not up to standard.

CAUTION
Keep the pack in a vertical position. If the pack is inclined or resting on one of its
sides, this could seriously compromise the condition of the equipment.

4.6. Storage of packed equipment


The GANTRY or • AWS crates (with the exception of the DIGITAL DETECTOR) can be stored for
no more than 15 weeks, at the following environmental conditions:
• The digital detector crate can be warehoused or transported for no more than 2 weeks after the
thermos bottles are correctly preconditioned. A label on the Digital Detector crate reports the expiry
date.

4.6.1. Gantry and AWS storage conditions


• The table below gives the conditions that must be met during storage of the GANTRY and the
AWS.
MIN MAX
Humidity 10% 80%
Temperature 10°C = 14°F - 80°C = 176°F
Atmospheric pressure 700hPa 1060hPa

WARNING
If the equipment is stored while packed, do not stack the packing crates on top of one
another

M189_EN:Transport and moving pag 69 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

4.6.2. Digital image receptor storage conditions


• The table below gives the conditions that must be met during storage of the DIGITAL
DETECTOR.
MIN MAX
Humidity 10% 95%
Temperature 5 °C = 41°F 45°C = 113°F
Atmospheric pressure 700hPa 1060hPa

WARNING
If the equipment is stored while packed, do not stack the packing crates on top of one
another.

CAUTION
Do not stack other crates on top of the digital detector crate!!!

M189_EN:Transport and moving pag 70 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

4.7. Removing the gantry from the pack


• Be careful of any nails, especially on the top of the pack.
• Remove protective plastic and check that the contents correspond to the attached packing list;
promptly notify I.M.S. of any missing items or damage to the equipment.
• Remove the bags containing the dehumidifying salts and store them in a dry place.

Image under revision


WARNING
The equipment should be unpacked according the order shown in the figure. Make
sure not to bang any parts of the equipment as it may cause damage.
• Remove the other packs from the wooden pallet and open them with caution.
• Remove the packing material from the equipment.
INFORMATION
We recommend that at least two people be used to remove the GANTRY from the
wooden pallet.

M189_EN:Transport and moving pag 71 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

4.8. Removing the AWS from the pack


• Be careful of any nails, especially on the top of the pack.
• Remove protective plastic and check that the contents correspond to the attached packing list;
promptly notify I.M.S. of any missing items or damage to the equipment.
• Remove the bags containing the dehumidifying salts and store them in a dry place.

• Remove the other packs from the wooden pallet and open them with caution.
• Remove the packing material from the equipment.
WARNING
The equipment should be unpacked according the order shown in the figure. Make
sure not to bang any parts of the equipment as it may cause damage.
INFORMATION
We recommend that at least two people be used to remove the GANTRY from the
wooden pallet.

4.9. Disposal of the packing


• Separate the packing materials and dispose of them in compliance with current regulations on the
disposal of solid waste.
• The packing materials (wood, cardboard and pluribool) are 100% recyclable.

M189_EN:Transport and moving pag 72 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

4.10. Instructions for the handling, storing and packing of digital


detector
CAUTION
• If the temperature of the detector drops below 10°C, it may be irreparably
damaged (the nature of the damage is equal to about 80% of the total cost of the
detector).
• It has been statistically calculated that this pack has a duration of about 15
days.
• Once the detector has been received from I.M.S. S.r.l., it must be stored in a
room with a temperature between 5°C and 45°C.
• If the detector must be shipped by air or by truck in temperature conditions that
are outside the limits, only the original pack can be used and it must be prepared
as described below.
• The warranty will be void if there has been tampering with the anti-intrusion
labels.
• The warranty will be void if the temperature sensor shows that the limits have
been exceeded.
INFORMATION
If the detector is not returned in its original packaging, immediately take a
picture of the packaging and the detector and report this to the UT for
verifications of the case

M189_EN:Transport and moving pag 73 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

4.10.1. Instruction for packing the digital detector before shipment


1. Insert the mammography Detector inside a transparent “ESD” plastic bag.

2. Hermetically seal the bag and wrap it with adhesive packing tape.
3. Open the box and insert insulating material.
4. Place the prepared Detector inside the cardboard box, with the carbon fiber side
facing upwards.

5. Close the cardboard box and seal the flaps with adhesive packing tape.
6. Prepare the “EXOGELS” containers by heating them until the material contained in
the bottles become completely liquid (the oven thermostat must be positioned at the
temperature reported on the oven, or rather max. 45°C) for about 8-10 hours.
CAUTION
Put the containers in the oven at about 17:30 and leave the oven on overnight.
Once the EXOGELS containers are completely heated, remove them from the
oven and let then “rest” for about 15-20 minutes. After this period, check the
temperature of each container using the I.M.S. S.r.l. 118 infra-red thermometer;
their temperature must be approximately 35°C (not higher than 37°C).
7. The containers can now be placed in their casings. Once the EXOGELS containers
are inserted, make sure that the temperature inside the pack does not exceed 45°C.
M189_EN:Transport and moving pag 74 of 201
Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

8. Insert the box containing the Detector, inside the cardboard packing box.
9. Install the insulation foreseen for the EXOGELS containers, inside the pack. See
recommendations in part 7.

10. Insert a copy of these instructions (in English) in the pack along with a photograph
of the bottom part of the detector showing the status of the thermometers and the
detector serial number.Put the pack’s cover on. Apply a label containing all the
identification data of its content on the outside of the packing.

M189_EN:Transport and moving pag 75 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

11. Apply the label below (in A4 format) to all sides of the pack (besides the bottom).
12. Write the pack preparation date and the initials of the operator; indicate the expiry
date of the thermal package (15 days from preparation date).

ATTENZIONE / CAUTION!
Imballo termico / Thermic package
Preparato il / Prepared on _____ da / by _____

Scade il / Expires on _________

Oltre la data di scadenza immagazzinare tra 5°C e 45°C


After the expiring date store in a temperature within 5°C and 45°C

13. The duration of this thermal package depends on the external temperature. After
several shipments between Italy and Canada performed during the winter, we have
determined that the package lasts about 15 days. Therefore, the shipment must be
organized in such a way that the time between its departure from I.M.S. S.r.l. and its
receipt by the end customer does not exceed 15 days.

M189_EN:Transport and moving pag 76 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

5. Installation
5.1. Ambient requirements conditions
 see the “Technical Data” paragraph in Section 2 - “General Information”.
CAUTION!
If the temperature drops below the temperature value indicated in the table, the
detector is at risk of being permanently damaged. In this case the warranty will
no longer be valid.
CAUTION!
The equipment must be installed in environmental conditions that respect the
conditions given in the table. It must not be installed in places where appliances
that alter these environmental characteristics are present.
The operating temperature must be maintained within the range of +12°C to
+27°C. If the detector temperature drops below +5°C or exceeds +45°C, there is
a risk that the detector may be permanently damaged.
CAUTION!
The detector needs a special pack for transportation. This pack ensures a
suitable temperature for a certain period of time. This time depends on the
transport conditions (cannot be more than two weeks). Consult the I.M.S. S.r.l.
Technical Service Dept. for more information about transportation and storage
conditions. If the detector temperature drops below +5°C or exceeds +45°C,
there is a risk that the detector may be permanently damaged.
The detector must be transported under controlled temperature conditions.
Using the pack provided by I.M.S. S.r.l., the internal temperature is guaranteed to
remain above +10°C with an outside temperature of -40°C for 48 hours (if the
pack wascorrectly prepared). See I.M.S. S.r.l. packing instructions.

M189_EN:Installation pag 77 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

5.2. Preliminary radiology room inspection


Before beginning installation, inspect the X-ray room to ensure that it complies with the following
specifications:
• Door minimum width: 75 cm
• Electric line with distribution box ( see the “Environmental requirements conditions” paragraph
at this Section).
• Floor plan with the layout of the equipment.
• ETHERNET data network.
• X-ray protection devices.

5.3. Preliminary mechanical clamping inspection

Check the integrity of the mechanical clamping after transport, in the following critical mechanical
gropu before to procede with the installation.

M189_EN:Installation pag 78 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

5.4. Installation procedure


5.4.1. Necessary tools

• HAMMER DRILL
• CONCRETE BITS O 10 - O 18
• VARIOUS WRENCHES (socket and Allen wrenches)
• STANDARD AND PHILIPS HEAD SCREWDRIVERS (in different sizes)
• HAMMER
• PINCERS (in different sizes)
• LEVEL

WARNING
The selecting of the position to install the equipment must not depend only on the it’s
actual dimensions, but one should also make sure that there is sufficient perimeter
space to ensure that all preparation, inspection, routine and non-routine maintenance
operations can be carried out in a highly rational manner, and that the work area is
ergonomic.

M189_EN:Installation pag 79 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

5.4.2. Positioning and fixing of the gantry


• Move all the equipment into the room and position them as close to their final locations as
possible. Leave working area around each equipment until final assembly is complete.
• Before beginning the final positioning, check that the floor of the room is able to withstand the
declared weight (with the safety margins specified by the current regulations) and that the surface
is as uniform as possible.
• When you have determined the final position where the equipment will be permanently installed
(fixed), position the base of the equipment and drill the holes 1 in the floor for anchoring the base-
plate.

CAUTION!
The supplied anchoring kit must be used to fasten the equipment to the floor.

Image under revision


CAUTION!
The levelling screws can protrude about 1 cm, insert a spacer if this is not
sufficient.

M189_EN:Installation pag 80 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

• Floor loading: 1000 kg/m2 (200 lb/ft2)


• Tension load 4KN
• Anchor point shear load 3KN

INFORMATION
Bolt hole locations are approximate, and represent the recommended
locations for equipment placement.
It is the responsibility of the customer, or their agents to verify that the floor is
capable of supporting the equipment floor loading, and anchorage
requirements.

CAUTION!
Before starting, it is recommended that at least two people carry out the listed
operations.

• Position the unit near the foreseen installation area.


• Remove the two lateral guards 3 and 4.

5.4.3. Installation of base cover

M189_EN:Installation pag 81 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

5.4.4. Touch Screen Panel and TFT Monitor Installation


Refer to the monitor replacement procedure.

Images under revision

5.4.5. Installation of the operator protective barrier


CAUTION!
The screen protecting the radiology technician during the mammography
examination must be handled very carefully.
INFORMATION
The protective screen conforms to the following standards: EN 60601-2-45 /
EN 60601-1-3 and is equivalent to 0.1 mm Pb
Proceed as follows to install it:
- Remove the front panel 1 from the electric cabinet.
- Fit (with extreme care) the X-ray lead glass panel 2 inside the special housing 3 on the control
table 4.
- “Accompany” the glass panel with both hands until it reaches the end of the housing.

M189_EN:Installation pag 82 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

INFORMATION
Install the X-ray lead glass panel, with the word “Giotto” at the top and legible
from the side opposite the operator.

Image under revision


INFORMATION
The weight of the glass and the “mandatory” housing should be sufficient to
ensure the stability of the glass. However, the glass panel must be anchored
to the structure of the electric cabinet by four nylon screws.
After installing the panel, reposition and secure the electric cabinet panel.
CAUTION!
It is strictly forbidden to use the vertical X-ray unit (GANTRY) without the electric
cabinet panel.

M189_EN:Installation pag 83 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

5.4.6. Removing the U.P.S. locking material


The upper and side part of the U.P.S. 1 may be locked (depending on the type of packaging) with a
plastic shim (or other material). This shim must be removed and stored for future moving of the
unit.

5.4.7. Connections between X-ray unit (Gantry) and Acquisition Work Station
(AWS)
The various cables coming from the vertical X-ray unit (GANTRY) enter through the lower part of
the acquisition workstation (AWS) and are connected as shown in the figure.

M189_EN:Installation pag 84 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

5.4.8. Connection of High-Voltage cable

5.4.9. Installation and connection of the X-ray tube

5.4.10. Digital detector installation

5.4.11. Connection of the equipment to the power supply line

5.4.12. Electrical safety checks

5.4.13. Auxiliary signaling and safety connections

5.4.13.1. Connection of door contact and the warning lights

M189_EN:Installation pag 85 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

M189_EN:Preliminary operating check pag 86 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

6. Preliminary operating check


6.1. Checks before using the unit
Each equipment is tested at the factory before shipment. List of the tests performed:
• Mechanical alignment of moving parts.
• Mechanical calibration of transducers (potentiometers, encoder pressure sensors).
• Calibration of mechanical actuators software through PARAM TOMO software.
• X-ray tube software calibration.
• Calibrate touch screen software.
• Calibrate acquisition geometry.
• High voltage check with voltage divider and oscilloscope.
• Check of the anode current with oscilloscope.
• Check of the mAs with mAs meter.
• High voltage check with non-invasive instrument.
• (HVL) filtration check.
• Check of dose and dose rate.
• Check of X-ray field and light field collimation.
• Check of the electrical interface with digital detector.
• Check of the digital detector’s electrical noise level.
• Digital detector high voltage and lamp voltage check.
• Digital detector calibration check.
• Check of defective pixels and image artefacts.
• Check of the Potter-Bucky diaphragm and grid erasing.
• Check of the automatic exposure meter.
• Check of digital image acquisitions with “Raffaello” application software.
• Check image quality.
• Check of DVD mastering function.
• Network connection check (service and image transfer).
• Check of USB interface.
• Check that all the accessories are included in the shipment.
• Check electrical safety.

M189_EN:Preliminary operating check pag 87 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

6.2. Initial start-up


The following must be checked the first time the system is started up:
- Emergency stop buttons.
- Surface of the compressors.
- Head rest surface.
- All parts in contact with patients.
- Breast support table.
- LCD display in the base of the equipment.
- Visual check of external parts to the equipment.
- Cable ducts.
- Mounting of the covers.
- Stability of the equipment.
- Tightness of fastening screws.
- Correct collimation of the X-ray field.
• Since the equipment was completely tested before shipment, the operations to be performed are
limited and the risk of damaging the equipment is very low.
• The connection cables between the control console and the stand are all different from one
another in order to avoid connection errors. All the boards have circuit breakers if the polarity of the
power supply is inverted.
Therefore, after the power supply check, the equipment can be switched ON.
Instruments: digital multimeter.
• With all cables connected and the equipment connected to the mains network and to the data
network:
• Main switch in OFF position. Apply power through the external electrical panel and check that the
voltage on
the input terminals of the main switch is correct. Also check the protective earth connection.
• Turn the main switch to the ON position. Verify that all the boards are powered, that LEDs are
illuminated and
fans are working, the message OFF STATE must appear after approximately 30 sec.
• Turn on the UPS switch. Wait a few seconds and verify that the UPS switches from battery to
mains power. (The UPS always uses the battery when turned ON and after a few seconds, if the
mains power is present, it switches to mains power. When operating on the battery, an intermittent

M189_EN:Preliminary operating check pag 88 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

buzzer sounds; the switch to the mains network is recognized by the sound of a relay and by the
fact that the buzzer no longer sounds.
 Check of digital image acquisitions with “Raffaello” application software.
 Check of DVD mastering function.
 Network connection check (service and image transfer).
 Check of USB interface.
• Turn ON the workstation using the start button.
• Wait for the operating system to start then enter the password (Username: administrator).

INFORMATION
For security reasons, the password occasionally changes.
(Consult the I.M.S. S.r.l. website http://www.imsitaly.com).
• Turn ON the power circuits, green button located under the main switch. The main contactor of
the power circuitry must close and the equipment must perform the following self-diagnostic
operations within a few seconds:
 Test collimator, change filter and lamp.
 Filter wheel and lamp test.
 Test the anti-scatter grid and potter movement.
 Test movement of the X-ray tube.
 Test the compressor.
• If all the tests are successful, the message AUTO will appear on the touch screen display and the
machine will be ready for use.
• Manually check the up/down movement of the column, the rotation both manual and programmed
and the tomo movement by performing the relative test (Consult the PARAM TOMO programme
guide for further details).
• If present, check the operation of the MAMMOBED and BIOPSY.
• Checking the correct collimation of the X-ray field. ( see the “Collimator “ paragraph in Section
13 -“Interventions on the unit”).

M189_EN:Preliminary operating check pag 89 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

6.3. Troubleshooting before initial start-up


• After turning the main switch ON, the message “OFF STATE” does not appear on the control
console:
• Based on the type of alarm message, consult the alarm message section for further information.
The typical problems are:

No communication between the boards:


Check that the cables for the Ethernet network and for the equipment are inserted correctly.

No power to the electronic boards:


Check that all the boards have +24V logic, check the fuses, power packs and connection cables.

Emergency chain open:


Check the emergency stop buttons and verify the connections of the equipment’s real time bus
cables. The control console S856 board has two red LEDs that must be ON (one is for the
emergency stop button and one for the equipment’s fault chain).
If there is no power to the 24V logic, check the fuses in the control console and in the boards,
check the insulation tester, etc.

Data network connection error:


If the hospital data network is connected to the SERVICE port (connector JC56) instead of the
external port (connector JC26), communication between the electronic boards may not be
activated and the unit does not start. The error could also have consequences on the hospital data
network. Check the connections in the AWS workstation.

The UPS does not turn on:


The UPS battery must be recharged at least every 3 months if inactive. If the battery is completely
dead the UPS may not start correctly.
Check that the power switch on the front of the UPS is in the ON position.

The workstation doesn’t turn on:


Check for the presence of electrical power.
Verify that the power switch on the workstation is in the ON position.
Check that the control connector from the workstation to the control console keyboard is correctly
connected.

M189_EN:Preliminary operating check pag 90 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

The LCD monitor doesn’t turn on:


Check the power supply voltage.
Check that the DVI cable is connected to the workstation.

WINDOWS doesn’t start:


Check that the mouse and keyboard are connected.
Use the USB pen drive attached to the equipment. Put it in the USB port located on the metal
frame of the control console and start the workstation; if Windows starts normally, this means that
the hard disk is corrupt.
The motors do not move:
Verify the compression force reading if the display on the base shows a compression force > 3 kg,
all the movements are inhibited and the compressor is free (without compression active). The
compression force reading mechanism needs to be checked (without compression force it must
always be 0 kg).
Verify the presence of the supply voltages: +24V power and 30V motors.

Vertical motor doesn’t move:


Verify that the 320V voltage is present in board S865, this voltage comes from the high voltage
inverter and there is a delay circuit to activate this voltage. This is due to the charging time of the
filter capacitors inside the inverter.

Errors with X-ray sequence in “M2” service mode:


Check the connection cables between the X-ray tube and the S866 board. All the X-ray functions
are managed by this board.

In normal operating mode with “Raffaello” application software the system is not ready for
X-ray emission (red or orange):
Wait at least 5 minutes.
Check the communication with the detector using the GMD SERVICE program.
Check with the ParamTomo program, under the menu Diagnostic->Detector->Digital IOs, that
the PLATE_
SYNC signal pulses correctly 2.8 sec. (MAMMO) / 1 sec. (TOMO). For versions of the “Raffaello”
application prior to 2.2.x.x. check that the PLATE_SYNC signal is pulsing correctly 3.5 sec.
(MAMMO) / 1 sec. (TOMO).

M189_EN:Preliminary operating check pag 91 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

Verify the power supply voltage for the detector, board S801.
Check the status LEDs on board S878.

6.4. Quality checks before clinical application


After checking that all the equipment’s functions are ON and operational, the radiological
parameters, image quality and compression values, collimation, etc. must be checked using
suitable instruments that are calibrated and in good working order before the clinical examination
can be performed on the patient.

CAUTION!
To perform mammograms on patients the quality checks must be carried out by
qualified personnel and must be within the acceptance limits. I.M.S. S.r.l. cannot
guarantee the clinical result if all these checks are not performed.

These are standard quality checks for mammography equipment. Refer to the “Quality assessment
of the Flat-Field image” Section in this manual for a brief guide or refer to the digital mammography
quality protocols (vary from country to country).
These checks must be done in accordance with local laws.
To evaluate Flat-Field images, or for problems relating to the correct operation of the X-ray tube or
the digital detector, refer to the specific chapter.

6.5. Connection to the DICOM ntework


The equipment can be connected to the DICOM network through the “Raffaello” application
software in order to exchange images and clinical data.
Refer to the DICOM CONFORMANCE STATEMENT for more information on the available
functions.
To configure the equipment for a DICOM connection, some parameters in “Raffaello” application
software and the equipment’s AWS acquisition workstation normally need to be set.
Read the of “Raffaello” application software guide for further details.

M189_EN:Preliminary operating check pag 92 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

6.6. Operator instructions


To complete the commissioning of the equipment the operators, radiologists and technicians need
to be trained on its use.

For this, the following must be done:


• Explain the operator’s manual.
• Power ON and power OFF procedure.
• Procedure to identify an alarm on the equipment and how to resolve it.
• Both manual and automatic movements for the equipment.
• Installation of accessories and use.
• Use of the “Raffaello” application software.
• Daily quality checks.
• Breast phantom and mammography simulation and technical explanations of radiographic work
techniques.
• Use of the BIOPSY.
• Use of the MAMMOBED.
• Transmission of the images to PACS and RWS.
• How to manipulate images for diagnostic use.
• Printing of images and image transfer to CD DICOM.

M189_EN:Preliminary operating check pag 93 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

7. Method of use

7.1. Equipment
7.1.1. Preliminary operations before switching on

7.1.1.1. Cleaning
see Secton 1 - “Safety”.

7.1.2. Switching on/Operating


The X-ray unit, the operator control console and the UPS are switched on separately by means of
their own switches.
The monitors can be switched off separately or using the main switch.

Image under revision


1) On/Off switch for the touchscreen panel
2) On/Off switch for the TFT monitor
3) On/Off switch for the workstation
4) Main On/Off switch
5) On/Off switch On/Off switch for insertion circuits
6) On/Off switch for the UPS

M189_EN:Method of use pag 94 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

Starting the system when off


If the equipment was switched off using the master switch, the entire system needs to be restarted.
CAUTION
Limited image quality caused by incorrect use.
The equipment’s detector needs to warm up for about 15 minutes before
images can be generated.

 Switch on the master switch located in the room.


 Press the On switch on the control unit to start the equipment.
– The internal control system automatically performs a functional check of the equipment.
Switch on the UPS.
Switch on the workstation, the touchscreen panel and the monitor of the acquisition
workstation.
– The workstation switches on.
 Log into the workstation.
 Let the equipment warm up for about 30 minutes before performing an optimal image quality
test or a calibration.
 Login with a username and password is required if access protection was set in the user
management section.
For further information refer to Section 9 - “User management and access protection” of the
GIOTTO CLASS Operator Manual.

7.1.1. Calibration
The detector settings may be modified by outside factors (for example: changes in temperature,
humidity and vibrations). This is why the detector must be regularly calibration.

The calibration procedures must be performed by IMS S.r.l. technical personnel at the frequency
indicated by the “Raffaello” application software”.
CAUTION
Limited image quality caused by incorrect use.
If the software indicates that a calibration has gone beyond its foreseen
interval, optimal image quality cannot be guaranteed.
Contact the Support Service as soon as possible.

7.1.1.1. Daily check


Prima di iniziare la sessione di esami è richiesta una verifica della qualità di immagine
mediante una procedura automatizzata.
see Section 7.11 - “Quality Control Procedures”.

M189_EN:Method of use pag 95 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

7.1.2. Method for basic use

7.1.2.1. Safety during the examination

WARNING
Risk of injury due to moving parts on the equipment.
The equipment’s motorized movements can crush or trap parts of the body.
Make sure that the patient is positioned correctly. Be aware of possible spots on
the equipment that can cause injuries or crushing.
see Section 1 - “Safety”.
In case of emergency, immediately press one of the red emergency stop buttons
located on the vertical X-ray unit and on the acquisition unit
see Section 1 - “Safety”.
Before activating the isocentric functions, make sure that the patient, staff and
third parties are not exposed to risks during the movement of the mobile arm.

WARNING
Incorrectly recorded patient data can lead to incorrect results.
Incorrect diagnosis or unnecessary exposure to radiation if the X-ray needs to
be repeated!
Before every exam, check that the patient corresponds to the data and to the
procedure loaded into the system.

M189_EN:Method of use pag 96 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

7.1.2.2. Performing the examination


How to perform the exam:
On the acquisition workstation

1) Open a new study using the appropriate procedure.


see in the present section at the paragraph “Patient Registration”.

2) Select the exposure mode (Auto/Man) and, if necessary, manually set the exposure parameters.
esposizione
see in the present section at the paragraph “Exposure data setting”.

3) Select the side of the first projection (if an exam protocol has been configured, the first
projection will be automatically proposed).
see in the present section at the paragraph “Manual setting the projections”.

On the x-ray unit


4) Mount a suitable compression paddle for the exam.
see in the present section at the paragraph “Compressor Paddle”.

5) Eventually apply the face protector.


see in the present section at the paragraph “Face protection”.

6) Set the gantry to the desired projection angle.


see in the present section at the paragraph “Gantry position setting”.

7) Set the gantry to the appropriate height and tilt angle.


see in the present section at the paragraphs “Gantry height adjustment” and “Gantry tilt angle
adjustment”.

8) Place the breast to be examined on the breast support surface based on the view selected from
the acquisition workstation.

9) Lower the compression paddle using the foot switch or the handle.
see in the present section at the paragraph “Compression using the foot-switch”.

10) Make sure that the breast is correctly compressed. Use the lighting to ensure that the
compression does not cause folds in the skin. If possible, the profile of the nipple should be visible.

11) Make sure that only the breast to be examined is in the irradiation field.

M189_EN:Method of use pag 97 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

On the acquisition workstation

12) Check the exposure parameters.

13) Wait for the X-ray button to light (green) before beginning the X-ray.

14) Press and hold the X-ray button on the control unit for the entire time you hear the acoustic
signal.

– During the X-ray, the radiation symbol will appear on all the displays (black clover with
yellow background).

– The irradiation begins and the first exam image is created.

15) The irradiation indicator must remain on only when the start X-ray button is pressed. Comply
with the measures regarding radiation protection.

16) After the image has been created, it will be displayed in the appropriate area.

17) Make sure that the breast is automatically or manually decompressed after the X-ray.

18) Check the image’s quality.

If necessary, cancel the image and repeat the operations from 10 to 20.
refer to the “Delete images” paragraph of this section.

19) Repeat the operations from 2 to 18 until all the projections have been performed.

20) Finish the exam.

refer to the “Concluding the exam” paragraph of this section.


CAUTION
Unnecessary exposure to radiation if the X-ray needs to be repeated.
If the button is released before the acoustic signal finishes, the exposure will
be interrupted and the image will be underexposed.

The digital X-ray is automatically displayed on the TFT monitor about 3 - 5 s after the end of the
entire acquisition. The AWS P.C. needs this time to acquire the image, apply the standard
corrections and to adapt the image to the mammography standard. During this stage, the image is
adapted to improve the aspects when viewed by the technician for the first time. In the case of
tomosynthesis or combo acquisition, the reconstructed volumetric image will instead be available
about 1-2 minutes later depending on the volume dimension.

M189_EN:Method of use pag 98 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

7.1.3. Compressor Paddle

Image under review

 Insert the compression paddle with the guide into the compression housing unit by making a
left to right movement.
– The compression paddle automatically locks.
 Keep lever 1 pressed.
 Slide the compression paddle out laterally by removing it from the housing guide 2.

Images under review

M189_EN:Method of use pag 99 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

7.1.4. Geometrical Magnification device

Images under review


The magnification device provides a 1.8 magnification factor.

Using the side handles, insert the magnification device into the support.
Grasp the handles on the side of the magnification device and pull towards yourself to
remove it from the support.

Images under review


For selective spot projection, use a compression paddle for zooming or for spot projections
and zooming.

M189_EN:Method of use pag 100 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

7.1.1. Compressor Paddle for Magnification device

Images under review

Insert the magnification device compression paddle with the guide in the compression housing unit
by making a left to right movement.
– The compression paddle automatically locks.

1) Area for grasping the magnification device compressor when mounting and dismounting.

Keep lever 1 pressed.
Slide the magnification device compression paddle out laterally by removing it from the
housing guide 2.

M189_EN:Method of use pag 101 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

7.1.1. Face protection


The face protector prevents the patient’s head from being in the trajectory of the X-rays.

Images under review


 Using the handles, insert the face protector in the support.
1) Face protector
2) Area for grasping the face protector when mounting and dismounting.
The face protector must be removed when using the magnification device.
Grasp the handles on the side of the face protector and pull towards yourself to remove it
from the support.

Images under review

M189_EN:Method of use pag 102 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

7.1.2. Gantry position setting


The setting of the projection angle and the gantry height is motor assisted.
The special keys on both sides of the mobile arm and the compression unit are used to control the
movements (see Section 3 - “Panoramic view of the system”).
The height can also be adjusted using the foot switch (see Section 3 - “Panoramic view of the
system”).

7.1.2.1. Gantry movements warnings


WARNING
Risk caused by the motorized movements of the equipment.
Injuries caused by collision or entrapment of the patient.

When adjusting the gantry height, make sure that no parts of the patient’s body
are in the risk zone ( see Section 1 - “Safety”).
When the gantry is rotating, make sure that the patient is at a safe distance from
the vertical X-ray unit. Be especially careful with patients in wheelchairs.
When the gantry is tilting in horizontal position, make sure that the patient is at a
safe distance from the vertical X-ray unit. Be especially careful with patients in
wheelchairs.

The rotation and height adjustment are always blocked when there is a compression force of 30 N
(3 kg) or higher.

M189_EN:Method of use pag 103 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

7.1.2.2. Gantry rotation angle adjustment


The gantry can isocentrically rotate between +180° (in a counter-clockwise direction) and -120° (in
a clockwise direction).

Positioning on the pre-set projection angle


The angle for every projection is predefined in the configuration of the workstation.
see Section 10 - “Configuration”.
Briefly press the Store isocentric position button.
– The gantry automatically moves to the predefined angular position for the next projection.

– The angular position appears on the vertical X-ray unit’s display.


Press any key to stop the movement before it is done.
– The rotation movement stops in its current angular position.
Press the Store isocentric position button again to restart the movement.
For safety purposes, the rotational movement will automatically stop if there is the risk of collision
with the floor.
In alternative to the protocol, 5 preset isocentric positions can be individually selected.

Fine adjustment of the projection angle


If necessary, the rotation angle can be adjusted in a more precise manner by pressing the rotation
movement set keys.
Press and hold the gantry rotation key and turn it in the counter-clockwise or clockwise
direction.
Release it when the desired angular position is reached.
– The rotation movement stops in its current angular position.
– The angular position appears on the vertical X-ray unit’s display.

For safety purposes, the rotational movement will automatically stop if there is the risk of
collision with the floor. To continue the movement, move the equipment upwards and then
continue the rotation.

M189_EN:Method of use pag 104 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

7.1.2.3. Gantry height adjustment


With reference to the 0° angular position, the gantry can be moved to a height between 70 cm and
145 cm above the ground.

Adjusting the height with the keys


Press and hold the gantry movement key up or down.
Release it when the desired height is reached.
– The gantry stops at the height reached at the moment it is released.

Adjusting the height with the foot switch


The height adjustment foot switch is marked with the words “UP” and “DOWN”.

Press and hold the foot switch up or down.

Release the foot switch when the desired height is reached.
The gantry stops at the height reached at the moment it is released.

7.1.2.1. Gantry tilt angle adjustment


With reference to the 0° angular position, the gantry can be tilted between -15° (with respect to the
patient direction) and +90° to permit use of patient support device Flexi Table for the biopsy exam
with patient in prone position.

Adjusting the tilt angle with the keys


Press and hold the gantry movement key in direction of the patient side or in the opposite
direction.
Release it when the desired height is reached.
– The gantry stops at the angle reached at the moment it is released.

M189_EN:Method of use pag 105 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

7.1.1. Compression/Decompression
Compression and decompression are motorised and are done using either the foot switch or the
handles located on the compression unit. The compression speed automatically decreases when
the compression paddle touches the breast and the compression force exceeds 10 N (1 kg).

The compression force and the compressed thickness are always shown on the display of the X-
ray unit and on the touchscreen panel.

7.1.1.1. Warnings regarding compression/decompression


WARNING
Risk caused by the involuntary activation of the compression foot switch.
Possible crushing injury to the patient.
As soon as the patient’s breast is compressed, move the foot switch out of
range of the patient and personnel.

CAUTION
An inappropriate compression setting can cause poor image quality.
Incorrect diagnosis or unnecessary exposure to radiation if the X-ray needs to
be repeated.
Make sure that the breast remains compressed during the entire X-ray.

In the case of an electrical black-out, the motorized decompression cannot be performed. In this
situation, manually raise the compression paddle using the handles,  see
“Compression/decompression adjustment handle” paragraph in Section 3 - “Panoramic view of the
system”).
INFORMATION
The maximum compression force that can be applied by the motor is 200 N (20 kg).
A force of 300 N (30 kg) can only be reached using the handles.

7.1.1.2. Compression using the foot-switch


The compression/decompression foot switches are marked with the compression/decompression
symbol ( ).

Image under revision


1) Compression paddle downwards (compression)
2) Compression paddle upwards (decompression)

M189_EN:Method of use pag 106 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

To move it, press and hold the foot switch up (compression) or down (decompression).
– When performing compression, the light will automatically switch on.
Release the foot switch when the desired compression force is reached.
– The compression paddle stops at the height reached at that moment.

7.1.1.1. Compression using the manual knobs


The compression/decompression handles are located on both sides of the compression unit.
Rotate the handle downwards (compression) or upwards (decompression).
Release the handle when the desired compression force is reached.
– The compression paddle stops at the height reached at that moment.

7.1.1.2. Automatic decompression


When automatic decompression is active, the compression paddle lifts approximately 8 cm after
every X-ray.
The use of automatic decompression is preset when the exam procedure is selected on the
workstation. If necessary, it can be manually activated or deactivated, see the
“Decompression in case of electrical black-out” paragraph in this section.
The automatic decompression is also active when an X-ray is interrupted before it is completed.

7.1.1.3. Decompression in case of black-out


If motorized decompression cannot be performed, you must activate the emergency release
located on top of the compression unit.
Firmly press the compression emergency release.
– The compression is mechanically released.
Manually raise the compression paddle.

INFORMATION
After the emergency release has been activated, an automatic system reset is
performed. Confirm the error message.

M189_EN:Method of use pag 107 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

7.1.2. Lighting
The lighting functions as an X-ray indicator for the correct positioning of the breast and normally
switches on automatically when the compression foot switch is pressed.
WARNING
If the breast is positioned incorrectly, other parts of the body may enter into the
radiation field
Possible unnecessary radiation exposure to the patient!
Using the lighting, position the patient so that only the breast to be examined is
in the radiation field.
Make sure that no part of the breast is in the shaded area.

The lighting can be manually switched on at any time using the control keys located on the sides of
the vertical X-ray unit.

Press the lighting key.

The lighting will automatically switch off after a certain amount of time (defined during installation of
the equipment, normally 20 seconds).

The lighting also switches off as soon as an X-ray is started (if the X-ray is performed before the
predefined time has expired)..

7.1.3. Shut down


Shut down the system at the end of every work day. To avoid any data loss, finish the exam and
correctly shut down the workstation before switching the equipment off.
WARNING
Interrupted exam.
Unnecessary exposure to radiation if the X-ray needs to be repeated!
Make sure that the Off key on the control unit is not pressed during the exam.

7.1.3.1. Shutting down the workstation


Shut down the system using the standard Windows procedure: Start > Close Session >
Shut down the system.
– The workstation switches off.

7.1.3.2. Shutting down the system


Press the Off switch on the control unit to shut down the equipment.
Shut down the workstation.
Switch off the UPS.
Turn off the master switch.
M189_EN:Method of use pag 108 of 201
Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

7.2. Acquisizione Work Station


7.2.1. Software Raffaello
After login, the interface of the “Raffaello” software will appear on the touchscreen panel, which is
divided into the following areas:

1) Menu bar
2) Functions bar
3) Study List (search option)

The workstation is only to be used for performing the X-rays and is not intended to function as a
report station or long term archive.

7.2.2. Monitor
In its standard configuration, the TFT monitor is a 2 MP colour monitor.
The X-rays are viewed on this second monitor.
Alternatively, a 3 MP TFT monitor can be installed on the workstation, which allows you to view the
X-rays images in high resolution with diagnostic capability.
INFORMATION
Even with the high resolution monitor, the workstation is only to be used for performing
X-rays and is not intended to function as a report station or long term archive.
After login, there are no active screens on the TFT monitor.

M189_EN:Method of use pag 109 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

During the acquisition phase, the TFT monitor shows a screen that is divided into six functional
zones:

1) Acquisition status

2) Patient and acquisition data

3) Preview of the acquired images

4) Hanging protocol information

5) Image display area

6) Task bar

 see Section 6 - “Operator Interface”.

M189_EN:Method of use pag 110 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

7.2.3. Mouse and keyboard


Use the mouse and/or keyboard to give commands or enter data into the computer.

7.2.3.1. Mouse
The mouse is equipped with three keys.

1) Use the left mouse key to select objects or launch commands and applications.
2) Use the central key to modify the contrast and brightness of images (“window/level”).
3) Use the right mouse key to open the context menus.

7.2.3.2. Keyboard
The keyboard is used to enter text and numbers. It is also used to access certain functions and to
launch programs using key combination and the symbols keypad.
A virtual keyboard can also be activated from the touchscreen panel.

1) Functions keys
2) Alphanumeric keyboard
3) Cursor movements button
4) Symbol keypad

M189_EN:Method of use pag 111 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

Typing text and numbers


The keys on the alphanumeric keyboard are used to enter text and numbers or commands.
Deleting characters
The Backspace key is used to cancel the character preceding the cursor and the Canc key is
used to cancel the character that follows it. When a text is highlighted, either of these two keys can
be used to delete it.
Moving the cursor
The arrow keys are used to move the cursor within an input field. The Pos 1 and Fine keys are
used to move the cursor to the beginning or end of an input field.
Function keys
see Section 6 - “Operator Interface”.
F1 Preset 1
F2 Preset 2
F3 Preset 3
F4 Preset 4

7.2.3.1. Using the keyboard


Almost all the commands can be given either with the mouse, the keyboard, or the touchscreen
panel.
Using combinations of keys
The combination of the Ctrl or Alt keys and another key let you quickly give certain commands to
the computer.
INFORMATION
All functions can be performed by simultaneously pressing the Alt key and the underlined
letter for the key or in the menu item. This allows you to use the program without the
mouse.

The table below shows the main key combinations:


Alt + F4 Closes the selected window
Ctrl + C Copy (only text)
Ctrl + X Cut (only text)
Ctrl + V Paste (only text)
Ctrl + I Shows information about the selected element
Ctrl + P DICOM printing of the selected element
Ctrl + Alt + W Allows the window/level to be manually configured

M189_EN:Method of use pag 112 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

7.2.4. Menu
The menu bar occupies the upper part of the screen and consists of a series of keys (icons) that
open menus in separate windows.

1 File
With this button you can import or export a DICOM file or close the program.
2 View
With this button you can choose whether the toolbar and the status bar are displayed or hidden.
3 Tools
With this key you can choose to configure or modify the default configuration data for the
“Raffaello” software.
4 Help
This indicates the version of the “Raffaello” software installed and gives access to the in-
line Help.

7.2.4.1. Reusorce control


The system keeps track of the occupied memory space, the capacity of the main database, the
daily calendar and the virtual memory.
Main database

INFORMATION
The disk space is insufficient when the “Raffaello” software reports “Free Space < 20%”.

M189_EN:Method of use pag 113 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

 To have a sufficient amount of free memory space, promptly file the patient and exam data
and cancel them from the patient list.
CAUTION
An insufficient working memory or space on the hard disk can cause system
instability or blocking.
Risk of data loss.
Pay attention to the warning symbols relating to memory space and make sure
that the available free space is sufficient.

7.2.5. Exit the program


Before shutting down the workstation or the equipment, you must exit the “Raffaello” software
using the Exit key. In this manner, all the programs being used are closed correctly.
The Windows 7 interface will appear when finished.

CAUTION
When exiting the program, unsaved data may be lost.
Possibility of data loss!
Save the data before exiting the program.
CAUTION
Danger if you switch off the computer from stand-by mode or before the system
has fully shut down.
Possibility of data errors, data losses, system blockage and system
damage!
Shut down the system before switching the computer off.

7.3. System Messages


7.3.1. Messages Classification
Various types of system messages can appear on the touchscreen panel.

ALARMS Error messages report an irregular situation.


Informational messages help the user perform the exam or search in the database.
WARNINGS The questions ask the user to confirm an operation started by the system or the
user before continuing with the procedure.

INFORMATION
Confirm the error message.
Reset the displayed error messages using the appropriate key.
The concerned components are automatically reset and the exam can continue.

 see Section 11 - “Breakdowns and diagnosis”, for more information on system messages.

M189_EN:Method of use pag 114 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

7.4. Patient Registration


WARNING
The system time is used for the internal identification of patient data such as exams,
series and images. If the system time is reset, two identification codes may be
created or the data may be assigned to the wrong patient.
Risk of interchanging patient data and errors in assigning the data!
Do not register patients, perform exams or process the images if the system
time is set to a value different from the original value.

A patient must be registered before undergoing an exam.


All the patient related information that are required during the exam are communicated to the
system during the patient registration.
The operator can select from the various registration procedures depending on how the exams are
performed in the hospital and the available time for the registration.
CAUTION
If the patient and exam data are only saved locally, it will no longer be available after
cancellation.
Risk of data loss!
Save the data in an external archive, for example in a HIS/RIS system or on
CD-ROM. The data can be transferred back to the local database at any time.

7.4.1. Open a new study


A new study can be opened in three different ways.
If it is decided that a patient will be the next to be visited during normal activities, then firstly enter
the patient’s data (New Study) or recall it from the database and then perform the exam (Other
Images).
New Study
This key, located on the task bar, is used to acquire images for a new patient. By pressing the New
Study key, the “Patient Info” dialog window appears. You must complete the fields, select the exam
mode and start the acquisition.

Other Images
This key allows other images of a patient already present in the database to be acquired.

Worklist
If there is a worklist system that assigns all the patients foreseen for that day to every equipment,
as soon as the individual exams start this data just needs to be recalled and possibly completed.
This saves time during the actual exams.
Since the equipment can interface with the worklist system, this requirement needs to be
communicated during the installation phase.
When the Worklist function key is pressed the dialog box will appear.
- Select the day or days to search for the scheduled exams.
- Select the exam mode to be searched.
- Select the workstations on which to perform the search.
M189_EN:Method of use pag 115 of 201
Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

- Press the Search Worklist key.


- Position the mouse over the patient data and left click to select it.
- Press New Study if you would like to open a study for this patient.
- The Patient Info Dialog box will open. Fill in the necessary information to open the new study.

Once the “Patient Info” dialog box opens, the necessary fields need to be set.

7.4.2. Patient Info Dialog


Patient data
The patient’s folder appears in the upper part of the control area and includes the patient’s
recorded data.

Select the acquisition protocol


If more than one acquisition protocol is configured, you must select the one to be used for the
exam. If only one protocol is configured, this will be automatically proposed by the interface. If no
protocol is configured or selected, the image acquisition sequence must be autonomously chosen.

“Quality Control” key


To acquire and process phantom images for quality control checks, mark the study as “Quality
Control”. This will disable the automatic image processing and the sending of the images to the
PACS and the medical recording workstation.

CAUTION
If the nature of the acquisition is not correctly indicated, the image will not be
appropriate for the quantitative analysis of the concerned physical sizes.

Breast implants
For the correct acquisition and processing of images, indicate the presence of any breast implants
when opening the study.
Even if automatic mode is selected, the AEC is not used and a preset exposure table is instead
used that is based on the thickness.
WARNING
If the implants are not taken into account, there can be treatment errors and incorrect
contrast settings during the exam.
Risk of injury during the compression. Unnecessary exposure to radiation if the
X-ray needs to be repeated!
Before performing the exam, ask the patient if she has breast implants. Inform
the patient of the risks of false indications.
Consequently, adjust the preparation and exam for the patient (i.e. compression,
system parameter settings).
Once the “Patient Info” dialog has been completed and the acquisition modality
(Mammo/Tomo/Combo) selected, press the “Start Acquisition” key to open the acquisition
interface.

M189_EN:Method of use pag 116 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

7.5. Exposure data setting


The acquisition can be performed in automatic or manual mode.

7.5.1. Automatic exposure mode


The equipment has special automatic functions to optimize the X-ray parameters and therefore
obtain the best possible image quality.
If the compression force is greater than 50 N (5 kg), the exposure mode is enabled by means of
the AEC (Automatic Exposure Control).
see the “Automatic modes for X-ray optimization” paragraph in Section 3 - “Panoramic view of
the system”.
If the presence of an implant is detected in the breast, the automatic mode will be enabled
but the preset system parameters will be used.

WARNING
An inappropriate compression setting can cause poor image quality.
Diagnosis error or unnecessary exposure to radiation if the X-ray needs to be
repeated!
If possible, us a compression greater than 10 kg (100 N).

7.5.2. Manual exposure mode


The individual exposure parameters (kV, mAs and target/filter combination) can be adjusted in
manual mode using the special keys on the acquisition interface.

Table of manual exposures (glandularity 50/50)


Compressed Breast Thickness MAMMO TOMO COMBO
Filter kV
(mm) (mAs) (mAs) (mAs)
10-20 Rh 23 40-45 60-70 80-90
20-30 Rh 24 50-60 75-90 100-120
30-40 Ag 25 60-70 90-100 120-140
40-50 Ag 26 80-90 110-130 160-180
50-60 Ag 28 90-100 130-150 180-200
60-70 Ag 30 100-110 150-170 200-220
70-80 Ag 31 110-120 170-190 220-240
80-90 Ag 33 120-140 190-210 240-280
>90 Ag 35 140-160 210-240 280-320

M189_EN:Method of use pag 117 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

INFORMATION
The parameters for the last exposure are displayed in the exposure data box.
In manual exposure mode, these will also be used for the next exposure, unless
changes are made before the emission of the X-rays.
WARNING
Changing the exposure parameters can cause inadequate anode/filter, mAs and kV
combinations.
Diagnosis error or unnecessary exposure to radiation if the X-ray needs to be
repeated!
When changing the settings manually, follow the recommended exposure
parameters (see table on the next page).
If possible, use the automatic AEC mode to optimize the exposure parameters.

The maximum mAs value that can be set depends on the set kV value.
kV 23 24 25 26 27 28 29 30 31 32 33 34 35
mAs max. 560 560 500 500 500 500 450 450 400 400 400 400 360

7.6. Performing the exposure


7.6.1. Safety during the irradiation
Follow the manufacturer’s instructions for use and the safety warnings.
see Section 1 - “Safety” and Section 7 - “Method of use”.

WARNING
Incorrectly recorded patient data can lead to incorrect results.
Diagnosis error or unnecessary exposure to radiation if the X-ray needs to be
repeated!
Before every exam, check that the patient corresponds to the data and to the
procedure loaded into the system.

7.6.2. Manual setting the projections


Set the equipment’s mobile arm based on the first desired projection.
Select the lateral projection on the touchscreen panel and check that the expected projection
appears.
Place the breast to be examined on the Potter/Bucky surface based on the projection
selected from the acquisition workstation.
WARNING
If the projection is set manually without the use of a protocol, only the projections
permitted at that position of the gantry will be selectable.
Before selecting the laterality, move the gantry to the desired position.

M189_EN:Method of use pag 118 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

7.6.3. Using the pre-set protocols for the sequence

7.6.3.1. Selezione del protocollo di acquisizione

Selecting the acquisition protocol

The acquisition protocol can be selected from the “Patient Info Dialog” page.
If only one protocol is configured, this will be always proposed for the acquisition.
If more than one protocol is configured, the first on the list that respects the operator filters, study
description and presence of implants will be automatically proposed. In any case, the selection can
be manually changed using the drop down menu.

Using the acquisition protocol

Press the isocentric position sequence key.


Adjust the column height (only for the first projection).
Place the breast to be examined on the Potter/Bucky surface based on the projection
indicated on the acquisition workstation.

When the acquisition session begins, the Giotto TOMO will automatically set the lateral projection,
the position for the projection and the projection modifier according to the first point in the
sequence for the protocol.

These indications and the progressive projection number in the sequence are also reported in the
upper left of both monitors (see Section 6 - “Operator Interface”).

INFORMATION
If the gantry position is not compatible with the selected view, the following warning
message will appear: “206: INCORRECT ISOCENTRIC POS”.

INFORMATION
If the operator manually modifies the lateral projection or position, the acquisition
protocol will be disabled. By returning to the programmed data for that projection, the
protocol will be re-enabled.

INFORMATION
If the exposure for the previous projection is done again, the protocol will be re-enabled
after the acquisition. If a different projection is acquired, the operator must then proceed
to the acquisition of the other images by manually setting the lateral projection and the
view position and manually move the gantry into the correct position (the Isocentric
position sequence key can no longer be used).

M189_EN:Method of use pag 119 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

7.6.4. Acquisition procedure


Wait until the status indicator on the TFT monitor indicates that the system is ready for the
next acquisition.
Compress the breast and return to the acquisition workstation.
Place the breast to be examined on the Potter/Bucky surface based on the projection
indicated on the acquisition workstation.
Wait for the X-ray button to light (green) before beginning the X-ray.
Press and hold the X-ray button on the control unit for the entire time the acoustic signal is
heard.
– During the X-ray, the radiation symbol will appear on all the displays (black clover with yellow
background).
– The irradiation begins and the first exam image is created.
After the image has been created, it will be displayed in the appropriate area.
WARNING
During the Combo examination a second type of acoustic signal indicates the
movement of the X-ray tube between the end of the tomo portion and the start of the
mammography.
Keep the button pressed until the entire examination is completed.

7.6.5. Correcting the projection


The exposures displayed in the image area must correspond to the expected projections. If there
are differences, a successive correction needs to be performed.
WARNING
If the exposure was done on the wrong side or if an incorrect exposure was
performed, the image will be saved with the wrong projection.
Possibility of diagnosis or treatment error!
Check that the correct projection is associated with the image and correct it if it’s
wrong.

Correction
Using the right mouse key click on the concerned image.
– A context menu will open.
Select the menu entry Other Toolbar Functions > Show Image Info.
– The dialog box showing the image information will open.
Using the drop down menu modify the incorrect data and press Close.
– A confirmation window will open.
Press Yes to make the change.
– Enter the password (if requested).

7.6.6. Repeat the image processing


If the quality of an image is unsatisfactory, it can be reprocessed by modifying the parameters and
without repeating the X-ray.
see Section 8 - “Image transfer and presentation” for more information.

M189_EN:Method of use pag 120 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

7.6.7. Discard images


If you were not able to sufficiently improve the quality of the image through reprocessing, is should
be discarded (i.e. if there are artefacts or noise) before closing the study.
Select the concerned image with the mouse.
Select the Review key and indicate the reason for refusal.

The selected image will not be sent to the archiving and recording workstations.

The discarded images will not be deleted and can be viewed together with the others in the
acquisition workstation.

Restore a discarded image:


Select the Review key and re-accept the image.
7.6.1. Delete images
Delete images using the Delete key.
CAUTION
Possible data loss.
Deleted images cannot be restored.
Check that the image that you really want to delete has been selected before
continuing with the operation.

7.6.2. Request image exposure data


The projection and the exposure parameters used are saved along with the image in the form or
attributes. This data can be viewed individually for all images.

Using the right mouse key click on the desired image.
– A context menu will open.

Select the menu entry Other Toolbar Functions > Show Image Info.
– The “Image Info” dialog box will open.
– The dialog box is composed of three tabs with the attributes about the image, the series and the
study to which the image belongs.

Select the image page and view the interested data.

Press Close.
– The dialog box will close.
see “Show Info” paragraph in Section 6 - “Operator Interface”.

M189_EN:Method of use pag 121 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

7.7. Printing the exam images


 see Section 8 - “Image transfer and presentation”.

7.8. Concluding the exam


After the exposures, the last image created appears on the monitor. At this point the exam on the
current patient can be concluded. If necessary, the relative documentation can be drawn up later.

Select Close on the TFT monitor’s task bar or press Close key on the touchscreen panel.
– The exam on the current patient is finished.

– If the system is configured for automatic forwarding to other DICOM archiving/reporting or


printing workstations the images marked QA will be sent when the study is closed.
The next patient can now be registered.

7.9. Manually sending images to other DICOM workstation


If you want to send part or all of the exam to DICOM workstation not configured for automatic
sending upon closing of the study, this can be done manually.
see Section 8 - “Image transfer and presentation”..

M189_EN:Method of use pag 122 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

M189_EN:Image quality assessment pag 123 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

8. Image quality assessment


INFORMATION
For further details on the checks described in this Section or the quality controls
recommended for the equipment, consult manuals M176 and M172 “Quality Control
Procedures for Medical Physicists”.

Image quality can be assessed visually or numerically through different tests.


This section reports three tests that cover both approaches, which are:
 Automatic Exposure Control (AEC) Test
 Dose Calibration Test
 Artefacts evaluation
CAUTION
When any quality control test is performed on the image, only the raw images are
visually and numerically analysed, since the processing is optimized for the clinical
application and artefacts can be created when applied to phantom images.
It is also recommended that the flat-field images be analysed that are obtained with
PMMA blocks certified for homogeneity, material purity and thickness precision, or
that are supplied directly by I.M.S. S.r.l..
The general acceptance criteria for the visual analysis of the Flat-Field images are the following:
There must be no:
• Areas of clear inhomogeneity  wrong Flat-Field correction, visible MCMs;
• White or black bands in the peripheral zones  incorrect collimation;
• Clear horizontal structures  grid visibility or noise;
• Clear vertical structures  visible MCMs or noise;
• Artefacts in terms of:
o rows/columns or clusters of uncorrectable defective pixels;
o central line;
o objects or dirt of any kind;
o shadows from previously radiographed objects (latent or ghost images).

M189_EN:Image quality assessment pag 124 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

8.1. AEC Test for Installation


This test is performed during installation by the service technician to perform a self-diagnosis of the
system to check that the equipment’s performance (in terms of image quality) is sufficient to carry
out clinical exams. The test includes the following checks:
• Communication of the detector with the AWS acquisition workstation;
• Correct collimation of the X-ray field;
• Flat Field correction;
• Homogeneity of the signal and uniformity of the SNR parameter;
• Presence of uncorrectable defects from the detector;
• Presence of any artefacts;
• Visibility of the anti-scatter grid;
• Efficiency of the digital detector;
• Clear and structured noise;
The results are automatically assessed by the software, which decides the outcome of the test
through testing parameters. It is however recommended that the image be visually evaluated in
terms of homogeneity, visible noise, visibility of the anti-scatter grid, the presence of uncorrectable
defects by the detector and the possible artefacts that cannot be detected through the numeric
analysis relative to this test ( see the “Assessment of artefacts and image homogeneity”
paragraph at this Section).

Required instrumentation:
• 2 PMMA blocks, thickness 20 mm each covering the entire detection area (supplied with the
equipment).

M189_EN:Image quality assessment pag 125 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

Click on QC key

Select the level “SERVICE” from the menu on section “Level Selection” and click on “OK”

M189_EN:Image quality assessment pag 126 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

Enter the password “master” and click on “OK”

Select from menu of the section “Quality Control Test Selection” the test AEC and clicck
on “Start Acquisition”.

M189_EN:Image quality assessment pag 127 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

Follow the instruction shown in the monitor and click on “OK”

Put 40 mm di PMMA (supplied with the Giotto Class) in the breast support surface and
apply a compression force of at least 5 kg (max suggested compression force 10 kg) , as
shown in picture

Image under review

M189_EN:Image quality assessment pag 128 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

Wait the ready state and select the view R-CC

M189_EN:Image quality assessment pag 129 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

Make an exposure to acquire an image.


When the image appears click on “QC Test”.

Confirm the analysis of the image.

Click on “OK” to visualize the details of the calculation performed.

M189_EN:Image quality assessment pag 130 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

Click on “OK” to visualize the Region Of Interest (ROI) used for the average pixel value
levels and for the signal-to-noise ratio (SNR) calculations.

M189_EN:Image quality assessment pag 131 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

Acceptance criteria:
The test is passed if:
1. The compressed thickness read is ± 3 mm with respect to the nominal value;
2. The maximum percentage deviation between the average pixel value of the single
ROI with respect to the total average value of the 5 ROIs used is minor than 5%;
3. The maximum percentage deviation between the SNR of the single ROI with
respect to the SNR average value of the 5 ROIs used is minor than 5%;

M189_EN:Image quality assessment pag 132 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

8.2. Dose Calibration check

M189_EN:Image quality assessment pag 133 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

8.3. Image Honogeneity and defective element evaluation


The purpose of this test is to check that the final image obtained is void of large artefacts caused
by the presence of dust or small objects on the surface of the detector or in the anti-scatter grid.
This test will also assess significant inhomogeneity caused by sensitivity differences in certain
parts of the detector or by an incorrect calculation of the detector’s gain map during the flat-field
correction, obtained with the long calibration.

Required instrumentation:
• 2 PMMA blocks, thickness 20 mm each covering the entire detection area (supplied with the
equipment).

Frequency:
Weekly, after replacement of the X-ray tube or the detector and after performing calibrations.

Procedure:
1. Open “Raffaello” application software (the image acquisition software) and open a new
study called “Artefact Assessment”, assign a patient ID number, enter a date of birth, click
on Quality Control button, select the MAMMO mode and then click on Start Acquisition;

M189_EN:Image quality assessment pag 134 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

2. Position the two 20 mm thick PMMA plates (total 40 mm PMMA) supplied with the
equipment on the resting surface for the breast (on the anti-scatter grid) so that the entire
active detection area is covered, as shown in the figure;

3. Remove the compressor;


4. Select the automatic exposure mode, select R-CC view (Right Cranio-Caudal) and perform an
exposure with the following parameters Filter Ag, 26kV, 100mAs, acquiring the image;

M189_EN:Image quality assessment pag 135 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

5. Repeat this procedure 6 times, changing the position of the PMMA plates (rotate or slide
slightly, always keeping the detector completely covered) for every image acquired.

ANALYSIS OF THE DATA AND INTERPRETATION OF THE RESULTS


1. Open the study just performed in “Raffaello” application software and select the raw images that
were just acquired;
2. Right click and the series of interest and select “Export DICOM file” and export the raw images
into a reference folder;
3. Open the “IMS Quality Control” program.
4. Click on the “Select Output Dir” button and select a folder to where the files containing the test
results are to be saved;
5. Click on the “Calculate Flat Field” button and select the exported images through the dialog
box;

M189_EN:Image quality assessment pag 136 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

6. Select the parameters shown in the following figure;

7. If the analysis is successful, the “Processing Successful” message should appear in the dialog
box;
8. A “Flat-Field.txt” file must have been created in the test results folder.
9. Open the file created and read the last line in the file, where the test results of all six images are
reported.

ACCEPTANCE CRITERIA

No pixel can be above the selected threshold for the result of the merging of the 6 images.
If there are pixels that are above the set threshold for the merging of the 6 acquired images,
identify the pixels from the coordinates reported on the Flat-Field.txt report and search for the
cause of this deviation through a visual analysis of the images. If these pixels are all in peripheral
areas, the collimation of the X-ray beam may be incorrect. If they are all on the same row or
column, there may be an uncorrectable row or column.

No ROI can deviate beyond the set threshold based on the mean count.
If the maximum percent deviation of the mean from the merging of the 6 images is not within the
threshold limit, the system fails the Artefact Assessment test.

After visual analysis of the raw images, the following must not be present at any of the thicknesses:
a. Areas of clear inhomogeneity (wrong Flat-Field correction);
b. White peripheral bands (incorrect collimation);
c. Clear horizontal structures (grid visibility or noise) or vertical structures (visible MCMs or noise);
d. Artefacts (uncorrectable defective rows or columns, objects of any kind).

M189_EN:Image quality assessment pag 137 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

If this test fails, the source of the problem must be identified and corrective action must be made by
a specialized technician. When the problem is resolved, the test must be performed again and it
must be successful before proceeding with clinical exams.
TROUBLESHOOTING
If the system fails this test, check these options for a possible cause:
 Clean the surface of the detector.

IMPORTANT WARING
The use of liquids of any kind is strictly forbidden.

 Presence of an object (dirt or dust) on the filter used: large, square blurred object that is not
present if an image is acquired with another filter (if available).
 Presence of an object (dirt or dust) on the mirror: large, square very blurred object, present
on images obtained with both filters.
 Presence of an object (dirt or dust) on the anti-scatter grid: small, focused object.
 Presence of an object (dirt or dust) on the surface of the detector: small, focused object.
 Presence of an object (dirt or dust) during the calibration without grid: black object that is
always present.
IMPORTANT WARING
If the problem turns out to be an object (dirt or dust) on a filter, clean the filter
very delicately using a clean, non abrasive cloth and avoid damaging it. In this
case, the fast calibration relating to the dirty filter must be repeated and you
must check that the resulting image is good. If the fast calibration is not enough
to give an image that can meet the acceptance criteria, all the calibrations must
be repeated! If the problem turns out to be an object (dirt or dust) that was
present during the long calibration, all the calibrations must be repeated! If the
problem turns out to be an object (dirt or dust) on a filter, clean the filter very
delicately using a clean, non abrasive cloth and avoid damaging it. In this
case, the fast calibration relating to the dirty filter must be repeated and you
must check that the resulting image is good. If the fast calibration is not enough
to give an image that can meet the acceptance criteria, all the calibrations must
be repeated! If the problem turns out to be an object (dirt or dust) that was
present during the long calibration, all the calibrations must be repeated!

M189_EN:Image quality assessment pag 138 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

M189_EN:Image quality assessment pag 139 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

9. Components diagnosis
9.1. Geometry of acquisition setting

M189_EN:Components diagnosis pag 140 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

9.2. Detector
To perform a complete diagnosis of the digital detector, a set of images must be acquired that is
composed of:
• CgrabbDiff.dat
• Dark Farme.dat
• Dark Sub.dat
• Flat-Field image with the anti-scatter grid
• Flat-Field image without the anti-scatter grid

Through a visual and numeric analysis, this set of images can be used to check for the presence of
structural artefacts from the detector, such as the presence of groups of defective lines or columns,
or the presence of structural noise, which degrade the quality of the clinical image.
The acquisition conditions are specified in the following section.

9.2.1. Dark Frame and Dark Sub analysis

DARK FRAME
Follow the procedure below to acquire a Dark Frame image:
1. Open the GMDService program (if closed) in the Mammo mode by following the path:
Start/Program/ Raffaello/GMDService;
2. Make sure that the Image Acquisition Services are ON; perform the following procedure to
check whether the services have started correctly:
a. Click on Test Acquisition and look at the dialog box. If the message “Acquisition Service Error:
The service is probably not active” appears, proceed as follows:

b. Click on Actions/Open Control Dialog;


c. Click on the Stop button in the Acquisition Services section;

M189_EN:Components diagnosis pag 141 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

d. A message confirming that the acquisition services have been disabled should appear in the
dialog box;
e. Click on the Start Console button in the Acquisition Services section;
f. Observe the sequence of events until (after about one minute) the detector status changes to 12
= IDLEMode.
3. Click on Test Acquisition and send high voltage to the detector by clicking on Start
Acquisition.
4. Observe the detector status from the control console (if activated in the special window). The
high voltage is active when the detector’s status is “ready”.
5. When the detector has high voltage, click on Dark Frame and wait a few seconds.

M189_EN:Components diagnosis pag 142 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

Acceptance Criteria
When the Dark Frame image is generated, visually analyse it using the following acceptance
criteria:
• There must be no clear defects in terms of large vertical or horizontal bands;

Correct Dark Frame image

M189_EN:Components diagnosis pag 143 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

Example pf Faulty MCM Example of Faulty gate drive

DARK SUB
Follow the procedure below to acquire a Dark sub image:
1. Open the GMDService program (if closed) in the Mammo mode by following the path:
Start/Program/ Raffaello/GMDService;
2. Make sure that the Image Acquisition Services are ON; perform the following procedure to
check whether the services have started correctly:
a. Click on Test Acquisition and look at the dialog box. If the message “Acquisition Service Error:
The service is probably not active” appears, proceed as follows:

b. Click on Actions/Open Control Dialog;

M189_EN:Components diagnosis pag 144 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

c. Click on the Stop button in the Acquisition Services section;


d. A message confirming that the acquisition services have been disabled should appear in the
dialog box;
e. Click on the Start Console button in the Acquisition Services section;
f. Observe the sequence of events until (after about one minute) the detector status changes to 12
= IDLEMode.
3. Click on Test Acquisition and send high voltage to the detector by clicking on Start
Acquisition.
4. Observe the detector status from the control console (if activated in the special window). The
high voltage is active when the detector’s status is “ready”.
5. When the detector has high voltage, click on Continue, then select the following conditions.

• W/Ag
• Large Focus
• 23kV
• 30mAs
• No Gain and Offset correction
• No defect correction
• Only Dark Frame correction
• No anti-scatter grid
• No Compressor
• No PMMA plates

IMAGE UNDER REVISION

M189_EN:Components diagnosis pag 145 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

Acceptance Criteria
When the Dark Sub image is generated, visually analyze it using the following acceptance criteria:
• There must be no completely black or white MCMs (large vertical bands formed by 128
channels).
• None of the edges of the image must be cut off  incorrect collimation.
• There must be no artefacts of any kind ( see the “Assessment of artefacts and image
homogeneity” paragraph in Section 8 - “Quality assessment of the flat-field image”).

Artefacts created by the presence of object on filter, mylar mirror or Be window.

M189_EN:Components diagnosis pag 146 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

9.2.2. Communication with ASW troubles


The first check on the correct communication between the AWS acquisition workstation and the
digital detector, can be done simply by opening the “Raffaello” application software and observing
the last icon in the bottom right of the applications bar.
If the icon is crossed out, this means that the detector is not communicating correctly with the
AWS.

INFORMATION
After the equipment is switched on, you must wait about one minute before the
detector begins to send data packets to the AWS. This time is needed before you
can proceed with other communication checking actions.

Follow the procedure below if the communication is not correct.


1. Make sure that the Image Acquisition Services are ON; perform the following procedure to
check whether the services have started correctly:
a. Click on Test Acquisition and look at the dialog box. If the message “Acquisition Service Error:
The service is probably not active” appears, proceed as follows:

b. Click on Actions/Open Control Dialog;


c. Click on the Stop button in the Acquisition Services section;
d. A message confirming that the acquisition services have been disabled should appear in the
dialog box.
e. Click on the Start Console button in the Acquisition Services section;
f. Observe the sequence of events until (after about one minute) the detector status changes to 12
= IDLEMode.
g. If the services are active, but the communication is still not correct, stop the services by closing
the console and click on Actions/Open Control Dialog; then click on the Check button in the
Detector section.

M189_EN:Components diagnosis pag 147 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

h. If communication is correct, the temperature value read by the detector should appear. If
communication
is not correct, an error message will appear relating to the failed attempt to open the corresponding
COM port.

9.2.3. Weak lines or columns


It’s possible that the lines or columns near defective lines/columns are affected by a different offset
than the others. To correct this visual effect as a greyscale value different from the background for
the lines and columns adjacent to the defective lines/columns, you need to identify these lines or
columns in the appropriate section of GMDService and also identify their references in order to
make the correction.
The image below shows an example of two weak lines and the relative references to enter into the
GMDService section.

M189_EN:Components diagnosis pag 148 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

9.2.4. Acquisition and analysis of defects map


Follow the procedure below to obtain the defect map after the defect calibration:
1. Open the “Raffaello” application software.
2. Open a new study by entering a patient name, surname, ID and date of birth;
3. When the detector’s status is “ready” (green light in upper left) click on the following combination
of buttons: Ctrl+Alt+D.
4. The image of the defect map from the detector obtained after the defect calibration will appear
after a few seconds. This image can be exported by Raffa “Raffaello” application software ello
either in DICOM format or in .dat binary format.

INFORMATION
The defect map does not include any weak lines/columns that were added by the
user.

Acceptance Criteria
For analysis of the defect map, the following acceptance criteria must be taking into consideration:
Single correctable pixel < 2500 pixels (3D); < 5000 pixels (3DL)
Correctable dead lines and columns No more than a total of 10 lines and columns (2 half-
(at least 3 good lines/columns in between) columns make 1 column).
Correctable dead lines and columns Adjoining dead columns/lines are not accepted (there
must be at least 3 good lines/columns in between).
No more than a total of 10 lines and
columns (2 half-columns make 1 column)
The score excludes all the pixels from the 2 lines
around the panel and the extreme corner of the high
voltage contact (150 x 150 pixels) from the chest
wall.
Correctable pixel groups (clusters) < 100 clusters (3D); <150 clusters (3DL) (max. 8
pixels) The score excludes all the pixels from the 2
lines around the panel and the extreme corner of the
high voltage contact (150 x 150 pixels) from the
chest wall.

M189_EN:Components diagnosis pag 149 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

M189_EN:Components diagnosis pag 150 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

9.1. X-ray tube


The X-ray tube is an instrument whose performance slowly decays over time. The emission
efficiency (intended as the emitted dose for the selected mAs) decays rapidly for approximately the
first 100 exposures, and then decays very slowly over time.

The X-ray tube can be operational until the tungsten filament, which emits electrons for thermoionic
emission, becomes so thin (due to the sublimation of the metals at high temperature) that it breaks
and creates electrostatic discharges. The formation of small craters at the anode (generated by the
constant impact of the electrons) also modifies the spectral quality; in particular, the half value
layer (HVL) parameter, or rather the thickness of aluminium needed to reduce the emitted dose by
half, which indicates its penetrating ability, or the energy carried by the photons, tends to increase
over time.

For mammograms, there is a serious operating limit in the need to have reduced exposure times
(or high output rates or rather, dose emitted per unit of time) in order to reduce the patient pain as
much as possible and to prevent organ movement, which reduce the quality of the diagnostic
image.

There are three tests for the diagnosis of the operating status of the X-ray tube:

1. Measurement of kV accuracy and reproducibility;


2. Measurement of the X-ray tube emission values - measurement of Efficiency;
3. Measurement of the exposure time at clinical conditions.

The operating status of the tube can be assessed through these simple tests and whether it needs
to be replaced.

M189_EN:Components diagnosis pag 151 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

INFORMATION
The curve of the graph was obtained using the following fixed exposure meter
parameters: 30kV 100mAs and is reported for 1 m focal distance.
At normal clinical use, the decay curve cannot be used as a preview curve because
the operating conditions are not comparable.

Before making efficiency measurements, check the number of exposures by the tube through the
appropriate section in the Param Class program.

M189_EN:Components diagnosis pag 152 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

9.1.1. Tube Tension, accuracy and reproducibility

This test is designed to check that the kV supplied by the X-ray system are accurate and
reproducible. This test and its relative tolerance limits make reference to the Fourth Edition of the
European protocol on Quality Control “European protocol for the quality control of the physical and
technical aspects of mammography screening 4th edition (2b.2.1.3.3 and 2b.2.1.3.4)”.

Required instrumentation:

• Digital multimeter calibrated for measuring kV in the mammographic range (~20-40 kV) with
accuracy of ± 1 kV and precision not less than 0.5 kV;

IMPORTANT WARING
• The system uses different Anode/Filter combinations. Before making the
measurement, make sure that the multimeter can correctly read the kV for the
required spectrum (W/Rh (50um) Paddle or No Paddle).

• Plate of steel or lead large enough to cover the entire detector area, able to completely shield it
from the X-ray field.

Frequency:

Every six months and when there are inconsistencies between the displayed mean glandular dose
and that measured through the special test.

Procedure:

1. Turn off the generator;


2. Perform the M2 service sequence in order to use the system without acquiring images;

IMPORTANT WARING
In this mode, the X-ray tube can emit radiation without waiting the time needed
for image creation. It is however recommended that no more than one
emission be performed per minute in order to avoid damaging the anode of
the X-ray tube.
In this manner the detector does not send signals to the AWS acquisition
M189_EN:Components diagnosis pag 153 of 201
Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

workstation, thus preventing the generation of latent or “ghost” images. It is


therefore absolution necessary that
all the detection surfaces of the detector be covered with a shielding
material suitable for the energy range used in mammography (2 mm of steel,
iron or lead) that can absorb 99% of the incident radiation.
Do not use this mode if a shield of this type is not available, but instead use the
GMDService/Test Acquisition program.
If a solid state multimeter is used, be careful not to irradiate the dosimeter
sensor numerous times, placed in the same position in order to avoid creating
latent or ghost images.

3. Turn on the generator;


4. Cover all the detection surfaces of the detector with a plate of X-ray shielding material;
5. Position the multimeter sensor in the standard dosimetric measuring position (6 cm from the
chest wall side and laterally centred, as shown in the figure below); set it for the correct
measuring of the configuration W/Rh 50 μm with or without the compressor, depending on the
selected measuring conditions.
6. Select the manual exposure mode, or use the GMDService program and perform an exposure
for every possible kV value (from 22 to 35 kV) using W/Rh and 20 mAs.
7. After performing this series of measurements, make 10 repeated exposures, using W/Rh 28kV
20mAs.

ANALYSIS OF THE DATA AND INTERPRETATION OF THE RESULTS


1. Compare the set nominal kV values with the measured values.
2. To determine the repeatability of the measurement, calculate the mean and the standard
deviation.
Calculate the percent variation as a ratio between the standard deviation and the mean, and check
that every value measured is within ± 5%.

ACCEPTANCE CRITERIA
The tolerance limit for the accuracy test is ± 1kV and the reproducibility of the measurement must
be within ± 0.5 kV of the mean value.
If the nominal kV value differs by more than 1 kV from the value measured with the multimeter, the
system fails the test.
M189_EN:Components diagnosis pag 154 of 201
Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

If the variation of the mean value is > ± 0.5 kV, the system fails the reproducibility test.
If this test fails, the source of the problem must be identified and corrective action must be made by
a specialized technician. When the problem is resolved, the test must be performed again and it
must be successful before proceeding with clinical exams.

TROUBLESHOOTING
If the system fails this test, check these options for a possible cause:
• Check that the measurement configuration used and the digital multimeter setting correspond
exactly.
• Check the position the digital multimeter sensor: standard dosimetric measuring position (6 cm
from the chest wall side and laterally centred).
• Check the X-ray emission graphs and the nominal emission values read by the system in the
appropriate section of the Param Class.

M189_EN:Components diagnosis pag 155 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

9.1.2. Radiation output: X-ray tube emission efficiency


The purpose of this test is to assess the correct emission of the X-ray tube in terms of efficiency
(uGy/mAs), output rate, exposure time, and quality of the X-ray beam (half value layer - HVL).

Required instrumentation:
• Solid state digital dosimeter (or digital multimeter) or ionization chamber calibrated for dose
measurements in the energy range, and for the spectra used, in mammography applications (~20-
40 keV);
• Plate of steel or lead large enough to cover the entire detector area, able to completely shield it
from the X-ray field.

IMPORTANT WARING
The system uses different Anode/Filter combinations. Before making the
measurement, make sure that the multimeter is able to correctly measure the
dose for the required spectrum (W/Ag (50um) Paddle or No Paddle).

Frequency:
Upon acceptance, every six months, and when problems occur relating to the mean glandular dose
and exposure time.

Procedure:
1. Turn off the generator;
2. Perform the M2 service sequence in order to use the system without acquiring images;

IMPORTANT WARING
In this mode, the X-ray tube can emit radiation without waiting the time needed
for image creation. It is however recommended that no more than one
emission be performed per minute in order to avoid damaging the anode of
the X-ray tube.
In this manner the detector does not send signals to the AWS acquisition
workstation, thus preventing the generation of latent or “ghost” images. It is
therefore absolution necessary that all the detection surfaces of the
detector be covered with a shielding material suitable for the energy range
used in mammography (2 mm of steel, iron or lead) that can absorb 99% of

M189_EN:Components diagnosis pag 156 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

the incident radiation.


Do not use this mode if a shield of this type is not available, but instead use the
GMDService program-Test Acquisition button.
If a solid state multimeter is used, be careful not to irradiate the dosimeter
sensor numerous times, placed in the same position in order to avoid creating
latent or ghost images.

3. Turn on the generator;


4. Cover all the detection surfaces of the detector with a plate of X-ray shielding material;
5. Position the multimeter sensor in the standard position (6 cm from the chest wall side and
laterally centred, as shown in the figure below); set it for the correct measuring of the
configuration W/Rh 50 μm with or without the compressor, depending on the selected measuring
conditions.
6. Select manual exposure mode and perform an exposure using W/Rh 28kV LF 20 mAs.

ACCEPTANCE CRITERIA
Reference values for installation:

Reference values for the periodic check:

If this test fails, perform the test below (exposure time at clinical conditions) and the source of the
problem must be identified and corrective action must be made by a specialized technician. When
the problem is resolved, the test must be performed again and it must be successful before
proceeding with clinical exams.

M189_EN:Components diagnosis pag 157 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

9.3.2 POSSIBLE SOLUTIONS


If the system fails this test, check these options for a possible cause:
• Check that the measurement configuration used and the digital multimeter setting correspond
exactly.
• Check the position the digital multimeter sensor: standard dosimetric measuring position (6 cm
from the
chest wall side and laterally centred).
• Perform the test below (Exposure time at clinical conditions) or the exposure test in automatic
mode.

M189_EN:Components diagnosis pag 158 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

9.1.3. Exposure time


The purpose of this test is to evaluate the exposure time at clinical conditions.

Required instrumentation:
• Dedicated instrument or solid state digital multimeter for measuring exposure time
• PMMA plates with a total thickness of 45 mm

Frequency:
Annually or when incorrect exposure problems occur

Procedure:
1. Position 45 mm of PMMA on the surface of the anti-scatter grid;
2. Open “Raffaello” application software and open a new study called “EXPOSURE TIME TEST”,
assign a patient ID number and enter a date of birth;
3. Select the automatic exposure mode and perform an exposure;
4. Put the sensor of the digital multimeter or the dedicated instrument in the standard dosimetric
measuring position (6 cm from the chest wall side and laterally centred);
5. Perform an exposure using the same parameters obtained with the automatic exposure;
6. Record the exposure time obtained and compare it with that shown on the monitor.

ACCEPTANCE CRITERIA
The exposure time must be within the limits reported in the reference table below.

Acceptable: < 1.5 s Optimal: < 1.0 s

If this test fails, the source of the problem must be identified and corrective action must be made by
a specialized technician (X-ray tube must be replaced). When the problem is resolved, the test
must be performed again and it must be successful before proceeding with clinical exams.

TROUBLESHOOTING
If the system fails this test, check these options for a possible cause before asking for the X-ray
tube to be replaced:
• Check that the measurement configuration used and the digital multimeter setting correspond
exactly.

M189_EN:Components diagnosis pag 159 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

• Check the position the digital multimeter sensor: standard dosimetric measuring position (6 cm
from the chest wall side and laterally centred).
Perform the previous test again (X-ray tube emission values - Efficiency) or the exposure test in
automatic mode.

M189_EN:Components diagnosis pag 160 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

10. Electronic boards


10.1. Electronic boards topographic diagram
INSERIRE IMMAGINE POSIZIONE SCHEDE
Electronic boards present on the equipment are the following:
Code Description
I008 Collimator
I009 Compressor
I009 PB Compressor Piggy Back
I009 RF RFID board
I010 Detector
I011 Biopsy
I012 Gantry
I013 Actuation of SELEMA
I014 Gantry keyboard
I015 Tube keyboard
I017 Wireless footpedal power
I020 Powers switching
I021 Operator control table commands
I022 Cabling switching of operator control table
Master Can Open IXXAT Central managment

M189_EN:Electronic boards pag 161 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

10.2. I008 – Collimator board


Connettore Funzione
X400 Connessione di terra piste di shield segnali
X401 Alimentazioni In
X402 Connettore motore paletta 1
X403 Connettore motore paletta 2
X404 Connettore motore paletta 3
X405 Connettore motore paletta 4
X406 Connettore motore carrello/ruota filtri
X407 Connettore sensori fine corsa palette 1
X408 Connettore sensori fine corsa palette 2
X409 Connettore sensori fine corsa palette 3
X410 Connettore sensori fine corsa palette 4
X411 Connettore sensori posizione CW-CX-CCW filtro
X412 Connettore lampada
X413 Sensori termici e pilotaggio ventola tubo
X414 Connessione tastiere tubo
X415 Ingresso segnale sensore raggi
X416 Alimentazioni Out
X417 CAN bus in-out
X950 Connettore ICSP PIC

M189_EN:Electronic boards pag 162 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

10.3. I009 Compressor board


Connettore Funzione
JC76
JC77
X491 Cella di carico 1 misurazione forza di compressione
X492 Cella di carico 2 misurazione forza di compressione
X493 Antenna RFID
X950 Connettore ICSP PIC

M189_EN:Electronic boards pag 163 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

10.4. I009PB Compressor Piggy Back board


Connettore Funzione
X484 Alimentazioni Out modulo display 1
X485 Seriale RS232 modulo display 1
X486 Alimentazioni Out modulo display 2
X487 Seriale RS232 modulo display 2
X488 Interfacciamento transceiver CAN a periferica FlexCAN Modulo Display 1
Interfacciamento transceiver CAN a periferica
X489 FlexCAN Modulo Display 2
JC75LT Segnali interconnessione I021-I009PB via flat lato sinistra paziente
JC75RT Segnali interconnessione I021-I009PB via flat lato destra paziente
JC76 Connessione piggyback alla I009
JC77 Connessione piggyback alla I009
JC81 CAN bus via cavo singolo

M189_EN:Electronic boards pag 164 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

10.5. I010 Detector


JC72 Segnali interconnessione I012-I010
JC75 Segnali interconnessione I010-I021
JC78 Segnali interconnessione I010-I021
JC79 Bus Segnali Real Time
JC80 Segnali abilitazione movimenti supporto paziente
X440 Connessione di terra piste di shield segnali
X441 Alimentazioni In-Out
X442 Uscita alimentazioni azionamento Compressore
X443 Segnali verso azionamento compressore
X444 Segnali verso azionamento asse Z biopsia
X445 Iniezione tensione in circuiti Fault/Emergenza
X446RT Segnali fungo emergenza destra
X446LT Segnali fungo emergenza sinistra
X447 Segnali seriale terminale biopsia
X448LT Alimentazioni e segnali terminale biopsia, attacco di sinistra
X448RT Alimentazioni e segnali terminale biopsia, attacco di destra
X449 Segnali verso azionamento carrello Potter
X450 Ingresso sensore ingrandimento x1.5
X451 Ingresso sensore ingrandimento x1.8
X452 Segnali interfacciamento Detector LMAM II
X453 Sensore fine corsa 1 asse Z Biopsia
X454 Sensore fine corsa 2 asse Z Biopsia
X455 Segnali Encoder braccio Tomo, segnale pilotaggio Relè statico freno braccio Tomo
X456 Segnali verso azionamento braccio Tomo
X457 Segnali azionamento griglia anti-diffusione
X458 Sensore fine corsa carrello Potter IN (griglia anti-diffusione in campo raggi)
X459 Sensore fine corsa carrello Potter OUT (griglia anti-diffusione fuori campo raggi)

M189_EN:Electronic boards pag 165 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

10.6. I011 Biopsy board


Connettore Funzione
X530 Connessione di terra piste di shield segnali
JC82 Alimentazioni, segnali, CAN azionamenti
X532 Segnali verso azionamento asse X biopsia
X533 Segnali verso azionamento asse Y biopsia
X534 Segnali CAN
X535 Pulsante enable movement
X536 Sensore fine corsa asse X Biopsia
X537 Sensore fine corsa asse Y Biopsia
X538 alimentazioni e segnali encoder angolo biopsia
X539 alimentazioni e segnali encoder carrello pistola

M189_EN:Electronic boards pag 166 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

10.7. I011AC Biopsy cabling switching


Connettore Funzione
X530 Connessione di terra piste di shield segnali
JC78RT Connessione segnali biopsia/accessori tramite cavo flat lato dx paziente
JC78MD Connessione segnali biopsia/accessori tramite cavo flat lato TBD paziente
JC78LT Connessione segnali biopsia/accessori tramite cavo flat lato sx paziente
JC82 Alimentazioni, segnali, CAN degli azionamenti Assi X-Y biopsia

10.8. I012 Gantry boards


Connettore Funzione
JC70 Segnali fungo emergenza tastierino operatore AWS
JC71 Segnali tastierino operatore AWS
JC72 Segnali da/verso scheda I010
X460 Connessione di terra piste di shield segnali
X461 Alimentazioni In
X462 CAN bus in-out
X463 Segnali controllo e monitoraggio globale alimentatore XP XM10
X464 Segnali controllo e monitoraggio moduli alimentatore XP XM10
X465 Segnali controllo e monitoraggio modulo spare alimentatore XP XM10
X466 Connessione tastiere stativo
X467 Ingresso sensore 1 antichiacciamento
X468 Ingresso sensore 2 antichiacciamento
X469A Pedaliere movimento verticale Gantry e Compressore
X469B Pedaliere movimento verticale Gantry e Compressore (clone di X469A)
X469C Pedaliere movimento verticale Gantry e Compressore (clone di X469A)
X470 Segnali inverter alta tensione raggi-X
X471 Uscita Comando raggi automatici
X472 Ingresso comando ausiliario raggi automatici
X473 Contatto porta stanza raggi e segnalazione apparecchio attivo
X474 Comando segnalazione luce esposizione in corso, per stanza raggi
X475 Comando Teleruttore Potenza Inverter
X950 Connettore ICSP PIC
X951 Segnali JTAG CPLD

M189_EN:Electronic boards pag 167 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

10.9. I013 SELEMA actuation


Connettore Funzione
X500 Connessioni CAN/RS422 di servizio
X501 Alimentazioni In-Out
X502 I/O Digitali
X503 I/O Digitali APPS
X504 Segnali Encoder
X505 Motore e freno
X506 I/O TEN
X507 Sicurezza termica
X508 Segnali fasatura motore brushless
X509 CAN bus In-Out
X510 Indirizzo cablato nodo CAN
X511 Ingresso potenziometro o sensore analogico
Connessione SPI tra azionamento
X512 Master e Slave in modalità accoppiata.
Master e Slave in modalità accoppiata.

10.10. I014 Gantry keyboard


Connettore Funzione
X466RT Connessione tastiera di destra paziente
X466LT Connessione tastiera di sinistra paziente

10.11. I015 Tube keyboard


Connettore Funzione
X414RT Connessione tastiera di destra paziente
X414LT Connessione tastiera di sinistra paziente

M189_EN:Electronic boards pag 168 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

10.12. I017 Wireless foot pedals power board


Connettore Funzione
X380 Ingresso alimentazione 18-36Vdc
X381 Uscita alimentazione +12Vdc 500mA

10.13. I020 Power switching board


Connettore Funzione
PL1_OUT Uscita alimentazione +24V logica
PL1_IN Ingresso alimentazione +24V logica
PL1_THRU1 Presa intermedia distribuzione alimentazione +24V logica
PL1_THRU2 Presa intermedia distribuzione alimentazione +24V logica
PL1_THRU3 Presa intermedia distribuzione alimentazione +24V logica
PL1_THRU4 Presa intermedia distribuzione alimentazione +24V logica
PL2_OUT Uscita alimentazione +24V potenza
PL2_IN Ingresso alimentazione +24V potenza
PL2_THRU1 Presa intermedia distribuzione alimentazione +24V potenza
PL2_THRU2 Presa intermedia distribuzione alimentazione +24V potenza
PL2_THRU3 Presa intermedia distribuzione alimentazione +24V potenza
PL2_THRU4 Presa intermedia distribuzione alimentazione +24V potenza
PL3_OUT Uscita alimentazione +36V potenza interrompibile
PL3_IN Ingresso alimentazione +36V potenza interrompibile
PL3_THRU1 Presa intermedia distribuzione alimentazione +36V potenza interrompibile
PL3_THRU2 Presa intermedia distribuzione alimentazione +36V potenza interrompibile
PL3_THRU3 Presa intermedia distribuzione alimentazione +36V potenza interrompibile
PL4_OUT Uscita alimentazione +48V potenza non interrotta
PL4_IN Ingresso alimentazione +48V potenza non interrotta
PL4_THRU1 Presa intermedia distribuzione alimentazione +48V potenza non interrotta
PL4_THRU2 Presa intermedia distribuzione alimentazione +48V potenza non interrotta
Presa intermedia distribuzione alimentazione +48V potenza non interrotta,
PL4_THRU3 con diodo antirimonto

M189_EN:Electronic boards pag 169 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

10.14. I021 Operator control table command board


Connettore Funzione
X490 Connessione di terra piste di shield segnali
X481 Alimentazioni In
X482RT CAN bus in
X482LT CAN bus out
X483 Iniezione tensione in circuiti Fault/Emergenza
X495 Alimentazioni Out per azionamento asse Z biopsia
X496 Sensore di collisione carrello Compressore contro carrello asse Z Biopsia
X497 Biopsy CAN and signals to Z axis
X444A Intercettazione segnali I010 verso azionamento asse Z biopsia
JC75 Segnali interconnessione I010-compressore
JC75LT Segnali interconnessione compressore via flat lato sinistra paziente
JC75RT Segnali interconnessione compressore via flat lato destra paziente
JC78 Segnali interconnessione I010-biopsia/accessori
JC78RT Connessione segnali biopsia/accessori tramite cavo flat lato dx paziente
JC78MD Connessione segnali biopsia/accessori tramite cavo flat lato TBD paziente
JC78LT Connessione segnali biopsia/accessori tramite cavo flat lato sx paziente
JC81RT CAN bus via cavo singolo (alternativo CAN andata)
JC81LT CAN bus via cavo singolo (alternativo CAN ritorno)
X950 Connettore ICSP PIC

10.15. I022 Operator control table Cabling switching board


Connettore Funzione
JC71 Segnali tastierino operatore AWS
JC71LO Diramazione segnali 1-12 connettore JC71
JC71HI Diramazione segnali 13-26 connettore JC71
X521 Connessione pulsante ON/OFF e Buzzer esterno
X522A Connessione pulsante XRAY
X522B Connessione pulsante XRAY ausiliario
X522C Connessione pedaliera XRAY ausiliaria

M189_EN:Electronic boards pag 170 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

11. Maintenance
11.1. Warnings on routine maintenance
INFORMATION
The equipment requires only minimum maintenance. The normal precautions in
using the equipment will ensure optimal, continuous and reliable operation.
Nevertheless, we advise you to run periodic checks and controls on the operation
of the equipment and its parts, in order to ensure that its performance and
reliability are maintained.

WARNING
Whenever the equipment is started, make sure that the safety devices are functioning
correctly.

CAUTION
If you note any faults or malfunctions, stop using the equipment immediately and
contact the Authorized Technical Service.

CAUTION
For any type of preventive maintenance, always disconnect the equipment from the
mains to avoid damage to persons and/or the electrical - electronic parts of the
mammography equipment.
Maintenance jobs must only be carried out by authorized personnel.

M189_EN:Maintenance pag 171 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

11.2. Cleaning
CAUTION
When cleaning, take care not to allow solvents to penetrate inside the equipment, as
this could cause serious damage to the various parts.
INFORMATION
For the normal cleaning of the various guards, do not use solvents, thinners or
products that contain “aggressive” substances towards paint and plastic. Use
exclusively neutral liquid soap and a soft sponge.
Do not use chemical products, thinners or solvents to clean the equipment.
All parts of the equipment in contact with the patient must be disinfected with specific
products like, for example, CHLORHEXIDINE.

Do this before every radiological examination.


Never use alcohol to clean the acrylic components (compressor, surface for magnification
technique, headrest).

The anti-Xray lead glass panel should be cleaned with a glass cleaning product, or denatured
alcohol. Use specific products for cleaning the monitor (water based and neutral soap).

CAUTION
The incorrect use of products can cause damage to the monitor.

WARNING
To keep the equipment in perfect working order and to avoid malfunctions and/or
faults, organize systematic, rational cleaning of the various parts of the equipment
and of the room where it is installed. When cleaning, take care not to allow solvents
to penetrate inside the equipment, as they could cause serious damage to the parts.
CAUTION
The scrapping of waste materials and parts of the equipment, including the batteries
of the UPS, must only be done by SERVICE PERSONNEL.

M189_EN:Maintenance pag 172 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

11.3. Preventive maintenance program

MECHANICAL

ELECTRICAL

CONFORMITY OF SETTING PARAMETERS

M189_EN:Maintenance pag 173 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

DESCRIPTION FEQUENCY PROCEDURE


Check the safety bolt;
Clean and lubricate the columns;
Vertical movement 12 months
Check elevator movement speed;
Check the cables positions inside the chain;
Check compression force setting;
Check maximum compression force (30 kg);
Compression system 12 months Check programmed compression force limit (20 kg);
Check thickness value during compression;
Check the cables positions inside the chain;
Check integrity of the brake system
Tube Arm 12 months Check the integrity of the belt;
Lubricate the guide;
Gantry rotation 2 years Check the integrity of the moto-redactor
Gantry inclination 2 years Check the integrity of the moto-redactor
Check the x-ray field collimation;
Collimator 12 months
Check and lubricate the four screws of the motors;
12 months Defect calibration
12 months Fast calibration
Digital Detector
12 months Check the efficiency of the detector fans
12-24 months Long Calibration
Lubricate the rail;
Potter/Bucky system 12 months Check the anti-scatter grid motor;
Check the cables positions inside the chain;
Lubricate the screw;
Biopsy Z axis 12 months
Check the cables positions inside the chain;
Enclosure 2 years Check the integrity of the enclosures;
Check the integrity of the breaking system of the control table;
AWS Workstation 12 months
Check the proper functionality of the UPS system;
Image quality 12 months See dedicated chapter
X-ray protection Refer to local laws

M189_EN:Maintenance pag 174 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

11.4. Disposal and scrapping of the obsolete equipment


According to Directive 2012/19/UE, on Waste Electrical and Electronic Equipment (WEEE), and in
reference to Italian Legislative Decree no. 151 of 25/07/05, we inform users of this device of the
following aspects:
1) When the equipment is decommissioned, it must not be disposed of as unsorted municipal
waste and must be collected and disposed of separately;
2) WEEE must be delivered to appropriate collection facilities, or returned to the distributor if a new
equipment is purchased;
3) If disposed of improperly, WEEE may have negative effects on human health and on the
environment;
4) Every WEEE has the symbol reported below, indicating that the equipment must not be dispose
of as unsorted municipal waste;
5) In case of unlawful disposal of WEEE, the penalties indicated in art. 16 of Italian Legislative
Decree no. 151 of 25/07/05 will be applied.

COMPOSITION

The Giotto Class medical device and its components have been built according to Directive
2011/65/UE, on Restriction of Hazardous Substances (RoHS).

• Steel: 50%
• Aluminum: 10%
• Plastic: 10%
• Copper/Brass/Bronze: 27%
• Glass: 2%
• Oil: 1%

There are no radioactive materials present in the equipment.

M189_EN:Maintenance pag 175 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

12. Breakdowns and diagnosis


12.1. Introduction
During operation the equipment’s touchscreen panel shows a series of messages indicating the
status and any emergency situations.
This section gives a list of the signals and, for each one, a brief explanation of the possible causes
and the operations to be performed to restore the equipment to normal working order. These
explanations are only valid for the cases in which the problem may be easily and safely resolved.

ATTENTION
All operations must be performed by specialized personnel.

INFORMATION
If the problem may not be resolved by means of the instructions given in this
section, the Manufacturer’s Technical Support Service should be immediately
contacted.
The progressive number in front of the message is also the message ID number
and appears on the panel during signalling.
The numbering of the messages in the following list might not be consecutive
since the equipment in question may not have certain optional parts.
All interventions to resolve the problems must be done respecting the same
prevention and safety standards that are valid for any other type of intervention on
the equipment.
I.M.S. S.r.l. declines all liability for damage of any kind caused by insufficient
maintenance, and the use of unoriginal spare parts or spare parts that are
incompatible with the quality standard and electro-mechanical characteristics of
the equipment.

M189_EN:Breakdowns and diagnosis pag 176 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

12.2. Alarms management


Giotto Class signals the presence of an alarm acoustically by means of three repeated sounds and
visually by showing the alarm code and an error message on the touchscreen. The presence of an
alarm inhibits exposure to X/rays. There are 2 types of alarm: those that can be reset by pressing a
button and those requiring the machine to be switched off and on again.
Some alarms can derive from different events. A good way of understanding what problem
generates an alarm is the Param Class “Error Log”. The Error Log is a register of alarms, errors
and other significant events taking place while Giotto Class is operating

12.2.1. Opening and consulting the Error Log


1. Execute the “Param Class” application. Once the software is running open the “File” menu and
select “Connection”.

IMAGE UNDER REVISION

M189_EN:Breakdowns and diagnosis pag 177 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

3. Open the “Diagnostic” menu and select “Error Log”

IMAGEs UNDER REVISION

M189_EN:Breakdowns and diagnosis pag 178 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

Each row corresponds to an event. The following information is available for each event:
• Electronic board that recorded the event.
• Event code: the codes are univocal for the alarms. The other event types are identified by their
board-code combination.
• Event level: this can be of 4 types:
o ALARM: alarm thrown by GiottoTomo. Inhibits exposure to X-rays (e.g.: Compressor fault
alarm).
o ERROR : operating error. This can generate an alarm. It refers to a fault that must be
analysed in order to understand what action to take on the machine. (e.g.: Error in rotation
movement potentiometer)
o WARNING: Reporting of situations that are on the verge of an error, but that do not
compromise correct machine operation. (e.g.: Signalling elevated temperature in the
vertical movement motor)
o INFO: Information about significant events that do not generate operating problems. (e.g.:
Information concerning switching on the machine).
• Description of event.
• Date and time the event was recorded.
• Number of pieces of information associated with the event. An event can have one or more
associated data, such as potentiometer values, current values, machine status, and so on.
• Code of the first piece of associated information. Double click on the event row to turn the codes
into descriptive text.
• Value of the first piece of associated information.
•…
•…
• Code of the last piece of associated information.
• Value of the last piece of associated information.

To understand what generated an alarm, check if an associated error is present in the adjacent
rows. If the date and time of an error coincide with that of an alarm, it was probably that error which
triggered the alarm.

M189_EN:Breakdowns and diagnosis pag 179 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

12.3. List of alarms


List of machine status messages and instructions to restore operation:

INSERT THE LIST OF THE ALARMS UPDATED

M189_EN:Breakdowns and diagnosis pag 180 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

13. Interventions on the equipment


13.1. Introduction
This chapter covers all operations to repair or replace groups or components that are worn,
defective or damaged.

WARNING
After every intervention, check that the equipment’s various cable harnesses are like
they were originally.

CAUTION
Only specialized personnel can perform these interventions.

INFORMATION
I.M.S. is not responsible for damage of any kind due to improper maintenance, the
using of non original replacement parts, or that do not meet the equipment’s quality
standards and electromechanical characteristics.

13.2. X-RAY TUBE


• Replacement:
• Calibration and auxiliary operations:
• Possible Faults:
o X-ray tube thermal safety switch:
o Rotating anode jammed:
o Reduction in X-ray tube efficiency:
o Filament fault:

13.3. COLLIMATOR
• Replacement:
• Calibration and auxiliary operations:
• Possible Faults:
o Collimator filters change disk:
o Collimator size change:

M189_EN:Interventions on the equipment pag 181 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

o Centring led lamp:

13.4. FILTER WHEEL


• Replacement:
• Calibration and auxiliary operations:
• Possible Faults:
o Faulty coupling between motor and disc. Excessive noise. Disc blocked.
o Faulty motor
o Faulty position sensors
o Damaged filters

13.5. COMPRESSOR MOTOR


• Replacement:
• Calibration and auxiliary operations:
• Possible Faults:
o The compressor does not move.
o Compression impossible.
o The compressor is not automatically released.

13.6. LOAD CELL


• Replacement:
• Calibration and auxiliary operations:
• Possible Faults:
o The grid does not move.
o The grid doesn’t move correctly, increase in the noise in the acquired images.

M189_EN:Interventions on the equipment pag 182 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

13.7. DETECTOR
• Replacement:
• Calibration and auxiliary operations:
 Necessary tools

• Supporto dedicato;
• 2 blocchi di PMMA dello spessore di 20 mm ciascuno e di dimensioni di 150x145 mm;
• 2 blocchi di PMMA dello spessore di 20 mm ciascuno coprenti l’intera area di rivelazione (in
dotazione con l’apparecchiatura);
• Dosimetro (opzionale) oppure Multimetro digitale (consigliato UNFORS Xi).

M189_EN:Interventions on the equipment pag 183 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

In questa sezione sono descritte le procedure complete per la calibratura del detector che
comprendono la calibrazione del guadagno (che dura all’incirca 5 minuti) che ha lo scopo di
correggere l’immagine acquisita per la sottrazione del the dark frame, del guadagno degli
amplificatori di segnale e dell’offset;
la procedura per la calibrazione dei difetti che ha lo scopo di correggere l’immagine acquisita per
la presenza di pixel/righe/colonne difettose (che dura all’incirca 2 minuti); e le calibrazioni
disuniformità filtri che comprendono la correzione delle disomogeneità dei filtri e della griglia anti-
scatter, della piattaforma d’ingrandimento e della biopsia (che durano all’incirca 3 minuti ciascuna);
infine vi è descritta la procedura per effettuare la calibrazione della dose emessa dal tubo
radiogeno, che ha lo scopo di compensare la visualizzazione della dose ghiandolare media, per il
decadimento naturale dell’efficienza dello stesso.
Le calibrazioni vanno eseguite per l’installazione dell’apparecchiatura e vanno ripetute
periodicamente, o quando insorgono problemi relativi ad artefatti o relativi alla qualità
dell’immagine.

M189_EN:Interventions on the equipment pag 184 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

Prima di effettuare una qualsiasi calibrazione del detector, e nel caso in cui una di queste non
dovesse avviarsi correttamente, seguire attentamente le seguenti avvertenze:
IMPORTANT WARING
• Assicurarsi di avere eseguito il Login come
Amministratore.
• Assicurarsi che l’apparecchiatura non sia in modalità M2;
• Assicurarsi che il cavo Raggi Automatici (connettore
X34) sia connesso;
• Chiudere l’applicativo “Raffaello”;
• Assicurarsi che i Servizi di Acquisizione siano attivi; Per
controllare che i servizi siano partiti correttamente seguire la
seguente procedura:
1) Aprire il programma GMDService da Start->Programs-
>Raffaello->GMD Service;
2) Premere “Service Procedure”, e inserire la password di
servizio: “giottomdsrv”.
3) Premere “Test Acquisition” e osservare la finestra di
dialogo. Entro un minuto deve apparire il messaggio
“Detector ready”. Se è presente il messaggio “Acquisition
Service Error: Probably the service is not active“ seguire la
procedura:
a. Premere “Actions/Open Control Dialog”;
b. Nella sezione “Servizi di Acquisizione” premere il tasto
“Stop”;
c. Nella finestra di Dialogo deve apparire un messaggio di
conferma dello stato
disattivo dei servizi di acquisizione;
d. Nella sezione “Servizi di Acquisizione” premere il tasto
“Start Consolle”;
e. Osservare la sequenza di eventi sulla consolle.
• Assicurarsi che la collimazione del campo raggi sia
corretta.

M189_EN:Interventions on the equipment pag 185 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

CALIBRAZIONE DEL GUADAGNO


La calibrazione del guadagno dura circa 5 minuti per essere completata ed è composta
dall’acquisizione di diversi dark frame, per il calcolo della mappa di guadagno del detector, al fine
di ottenere la correzione di flat-field sull’immagine finale.

Strumentazione richiesta:

M189_EN:Interventions on the equipment pag 186 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

• Nessuno

Frequenza:
All’accettazione e ogni anno.
Questa calibrazione è da ripetersi nel caso in cui appaiano sull’immagine finale delle evidenti
disomogeneità.
AVVERTENZE SPECIFICHE PER LA CALIBRAZIONE DEL GUADAGNO IN
MODALITÀ
MAMMOGRAFICA
Se dal momento in cui viene effettuata la calibrazione del guadagno, al momento in
cui si acquisisce l’immagine finale, vi è una variazione di temperatura superiore a 10°C,
è possibile che nell’immagine finale siano presenti evidenti disomogeneità, con la
possibilità di vedere la struttura di amplificazione (MCM) del detector.
Se si dovesse verificare questa condizione, occorre necessariamente ripetere
l’intera procedura di calibrazione.

M189_EN:Interventions on the equipment pag 187 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

CALIBRAZIONE DEI DIFETTI


La calibrazione dei difetti dura 2 minuti per essere completata ed è composta dall’acquisizione di 2
immagini (una al minuto), per il calcolo della mappa dei pixel, righe o colonne difettose, al fine di
ottenere la correzione dei difetti sull’immagine finale. Terminata la calibrazione dei difetti viene
generato un file di testo con il report dei risultati ottenuti, e un’immagine (file dati .dat) caricabile
attraverso Raffaello (vedi apposita sezione), per la visualizzazione della mappa dei difetti.

Strumentazione richiesta:
• Supporto dedicato;
• 2 blocchi di PMMA dello spessore di 20 mm ciascuno e di dimensioni di 150x145 mm.

Frequenza:
All’accettazione e ogni anno.
Questa calibrazione è da ripetersi anche nel caso in cui qualche nuovo pixel difettoso viene
scoperto nel detector, o sia presente su uno dei filtri (o sullo specchio) un qualche oggetto che
altera l’immagine creando un artefatto visibile. Va inoltre eseguita nuovamente nel caso in cui si
sostituisca il detector.

Procedura:
1. Accendere il generatore;
2. Assicurarsi che l’apparecchiatura non sia in modalità M2;
3. Assicurarsi che il cavo Raggi Automatici (connettore X34) sia connesso;
4. Chiudere l’applicativo “Raffaello”;
5. Aprire il programma GMDService da Start->Programs->Raffaello->GMD Service;
6. Premere “Service Procedure”, e inserire la password di servizio: “giottomdsrv”;
7. Assicurarsi che i “Servizi di Acquisizione” siano attivi;
8. Premere il tasto “Complete Calibration”;
9. Selezionare “Mammo Gain” e premere il tasto “Start”;
10. Assicurarsi che la procedura sia partita correttamente. Se la calibrazione non parte
correttamente ricontrollare le avvertenze riportate a inizio Sezione!

M189_EN:Interventions on the equipment pag 188 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

CALIBRAZIONI DISUNIFORMITÀ FILTRI IN MODALITÀ MAMMOGRAFICA


Le calibrazioni disuniformità filtri durano 5 minuti ciascuna e sono composte dall’acquisizione di 5
immagini (una ogni minuto), per la correzione sull’immagine finale delle disomogeneità dei filtri,
della griglia anti-scatter, della piattaforma d’ingrandimento e per la correzione delle immagini di
Biopsia.

Strumentazione richiesta:
• 2 blocchi di PMMA dello spessore di 20 mm ciascuno coprenti l’intera area di rivelazione (in
dotazione con l’apparecchiatura).

Frequenza:
All’accettazione e ogni anno.
Questa calibrazione è da ripetersi anche nel caso in cui venga sostituita la griglia anti-scatter, o la
piattaforma d’ingrandimento e nel caso in cui vengano ripetute la calibrazione del guadagno o dei
difetti.

M189_EN:Interventions on the equipment pag 189 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

M189_EN:Interventions on the equipment pag 190 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

M189_EN:Interventions on the equipment pag 191 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

M189_EN:Interventions on the equipment pag 192 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

M189_EN:Interventions on the equipment pag 193 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

Procedura:
1. Se le calibrazioni disuniformità filtri sono state appena effettuate, saltare i seguenti punti e
proseguire
direttamente dal punto 9;
2. Accendere il generatore;
3. Assicurarsi che l’apparecchiatura non sia in modalità M2;
4. Assicurarsi che il cavo Raggi Automatici (connettore X34) sia connesso;
5. Chiudere l’applicativo Raffaello;
6. Aprire il programma GMDService da Start->Programs->Raffaello->GMD Service;
7. Premere “Service Procedure”, e inserire la password di servizio: “giottomdsrv”;
8. Assicurarsi che i “Servizi di Acquisizione” siano attivi;
9. Rimuovere il compressore, se presente;
10. Premere il tasto “Complete Calibration” e assicurarsi della corretta configurazione seguendo
le avvertenze;

11. Selezionare “Mammo Dose”;


12. Assicurarsi che la procedura sia partita correttamente;
13. Dopo queste 3 immagini il programma chiede di acquisire la quarta immagine con la misura
della dose emessa tramite dosimetro tarato per spettro W/Rh. Se si è in possesso del dosimetro,
leggere attentamente nella finestra di dialogo il messaggio per settare correttamente il dosimetro e
poi premere “yes”, oppure selezionare “no”; Se si sceglie l’opzione “no”, il programma chiede se si
vuole cancellare il precedente valore di riferimento.

M189_EN:Interventions on the equipment pag 194 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

M189_EN:Interventions on the equipment pag 195 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

M189_EN:Interventions on the equipment pag 196 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

• Possible Faults:
o The detector does not acquire images.
o Communication error between the AWS and the detector.

13.8. DIGITAL DISPLAYS FOR THE VERTICAL X-RAY UNIT


• Replacement:
• Calibration and auxiliary operations:
• Possible Faults:
o Display not switched on
o Display switches off unexpectedly
o Displayed information not updated
o Unintelligible data displayed

13.9. HIGH VOLTAGE CABLE


• Replacement:
• Calibration and auxiliary operations:
• Possible Faults:
o Cable discharge:
In this case, there is an anode power spike that is intercepted by both the high voltage
inverter and the filament control circuit, which causes the immediate interruption of the
exposure. If the inverter alarm is tripped, one of the following LEDs will light on the inverter
board: LD8 KV>110%, LD9 KV min, LD7 mA max; the power must be cut to reset the
alarm.

M189_EN:Interventions on the equipment pag 197 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

CAUTION
When a cable discharges H.V. it can’t be detected by a normal ohm meter because
the testing voltage is too low.
o Interrupted cable:
This type of fault cannot be safely detected with a normal ohm meter because the testing
voltage is too low. The cable may be good with the ohm meter but applying H.V. it may be
open.
CAUTION
Don’t bend or twist the H.V. cable, as it could become damaged.
When locking the connectors together, be careful not to break the plastic body.

13.10. UNINTERRUPTABLE POWER SUPPLY (UPS)


• Replacement:
• Calibration and auxiliary operations:
• Possible Faults:
o The UPS does not turn on.
o The UPS does not keep the system running during a power failure.

13.11. FILAMENT DC-DC CONVERTER


• Replacement:
• Calibration and auxiliary operations:
• Possible Faults:
oo Communication error of auxiliary processor on S866
oo Filament current error
oo Anodic current error
oo Filament power supply voltage error

13.12. AWS WORKSTATION


• Replacement:
• Calibration and auxiliary operations:
• Possible Faults:
M189_EN:Interventions on the equipment pag 198 of 201
Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

oo The workstation does not turn on.


oo There are problems in the workstation’s internal boards.

13.13. DVD – CD BURNER


• Replacement:
• Calibration and auxiliary operations:
• Possible Faults:
oo The DVD – CD burner does not turn on.
oo The DVD – CD burner does not write data.
oo The data recorded on the disc are illegible.

13.14. MONITOR
• Replacement:
• Calibration and auxiliary operations:
• Possible Faults:
oo The monitor doesn’t turn on.
oo There are parts of the monitor that do not work.

13.15. IXXAT
• Replacement:
• Calibration and auxiliary operations:
• Possible Faults:

13.16. Board I008: Collimator


• Replacement:
• Calibration and auxiliary operations:
• Possible Faults:

13.17. Board I009: Compressor


• Replacement:
M189_EN:Interventions on the equipment pag 199 of 201
Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

• Calibration and auxiliary operations:


• Possible Faults:

13.18. Board I010: Detector group


• Replacement:
• Calibration and auxiliary operations:
• Possible Faults:

13.19. Board I011: Biopsy


• Replacement:
• Calibration and auxiliary operations:
• Possible Faults:

13.20. Board I012: Gantry


• Replacement:
• Calibration and auxiliary operations:
• Possible Faults:

13.21. Board I013: Actuator


• Replacement:
• Calibration and auxiliary operations:
• Possible Faults:

13.22. Board I014: Gantry keyboards


• Replacement:
• Calibration and auxiliary operations:
• Possible Faults:

13.23. Board I015: Tube keyboards


• Replacement:
• Calibration and auxiliary operations:
• Possible Faults:

M189_EN:Interventions on the equipment pag 200 of 201


Draft version_0.1 del 01/07/15
SERVICE AND
INSTALLATION Via Sagittario 5 - 40037
Pontecchio Marconi (Bo)
MANUAL Tel. ++39 051 846851
Fax. ++39 051 846856
www.imsitaly.com

13.24. Board I017: Wireless pedals control power supply


• Replacement:
• Calibration and auxiliary operations:
• Possible Faults:

13.25. Board I020: Power supply switching


• Replacement:
• Calibration and auxiliary operations:
• Possible Faults:

13.26. Board I021: Operator commands


• Replacement:
• Calibration and auxiliary operations:
• Possible Faults:

13.27. Board I022: Operator console connections


• Replacement:
• Calibration and auxiliary operations:
• Possible Faults:

M189_EN:Interventions on the equipment pag 201 of 201