Abstract
Impedance measurement is a promising technique for detecting pre-malignant
changes in epithelial tissue. This paper considers how the design of the
impedance probe affects the ability to discriminate between tissue types. To
do this, finite element models of the electrical properties of squamous and
glandular columnar epithelia have been used. The glandular tissue model
is described here for the first time. Glandular mucosa is found in many
regions of the gastrointestinal tract, such as the stomach and intestine, and
has a large effective surface area. Firstly, the electrical properties of a small
section of gland, with epithelial cells and supportive tissue, are determined.
These properties are then used to build up a three-dimensional model of a
whole section of mucosa containing many thousands of glands. Measurements
using different types of impedance probe were simulated by applying different
boundary conditions to the models. Transepithelial impedance, and tetrapolar
measurement with a probe placed on the tissue surface have been modelled.
In the latter case, the impedance can be affected by conductive fluid, such
as mucus, on the tissue surface. This effect has been investigated, and a new
design of probe, which uses a guard electrode to counteract this potential source
of variability, is proposed.
1. Introduction
1.5mm V
2.24mm
and permittivity is seen in breast, liver and muscle tissue (Pethig 1984). We are interested in
measurements of epithelial tissue with a view to screening applications—electrical impedance
probes offer the potential of rapid, non-invasive diagnosis at minimal cost. Brown et al
(2000), using the probe shown in figure 1, demonstrated a clear difference between normal
and dysplastic cervical epithelium.
The sensitivity and specificity was at least as good as for the Papanicolaou smear test.
Gonzalez-Correa et al (1999), using a similar probe, were able to differentiate between
normal squamous oesophageal tissue and Barrett’s oesophagus, a condition in which intestinal
columnar epithelium develops at the junction of the stomach and oesophagus. This is
associated with a substantially increased risk for adenocarcinoma (Falk 2002).
In the preceding in vivo measurements, the current is injected on the surface of the
epithelium. This is in contrast to transepithelial measurements, where current is injected
across the epithelium. The epithelium is mounted in an Ussing chamber, which creates an
electrical seal and allows close control of the experimental conditions. The impedance is
normalized to the area of exposed epithelium, and so has units of cm2. This type of
measurement is used by electrophysiologists to investigate the ion transportation and barrier
properties of epithelia (Barry 1989). Gitter et al (1997) describe an Ussing chamber which
can perform measurements in the frequency range 1 Hz–65 kHz. The fact that tissue must be
clamped in a measurement chamber means that this technique is limited to ex vivo samples. It
is possible for the tissue adjacent to the clamp to become damaged by crushing, which reduces
its impedance. This is a particular problem for small sample chambers (where the fraction of
damaged tissue will be high) and for high impedance epithelia.
Computer modelling allows us to investigate how changes in tissue structure affect the
impedance spectrum measured with different types of probe. In this way, the probe design can
be optimized to detect a particular feature of interest in the tissue. When making measurements
using surface probes, it is possible for the tissue to become ‘short-circuited’ by conductive fluid,
such as mucus, on the tissue surface. It is difficult to control the contribution of this component
to the measurement and it could lead to variability in the experimental data. For example—
fluid may be squeezed out if increased pressure is applied to the probe, thereby reducing the
conductance of the short-circuit path. The effect of surface fluid for measurements on an
isotropic, homogeneous medium has been investigated by Jones et al (2001). For epithelial
tissue, the fact that the epithelium acts as a barrier to current flow means that the effects of
surface fluid will be accentuated.
Because of this potential source of variability, a new design of probe has been considered.
This probe uses a concentrically arranged guard electrode in order to limit the lateral flow of
measurement current in the surface fluid.
Modelling of epithelial tissue impedance measured using three different designs of probe 607
surface fluid
columnar
cell
lamina
mucosa propria
gland
lumen
submucosa
In order to test probe designs, finite element models (FEM) of epithelial tissue have
been developed. An important characteristic of epithelial tissue is its ‘tightness,’ which
indicates the relative importance of the paracellular and transcellular conduction pathways.
In a leaky epithelium, the paracellular current pathway predominates. In a tight epithelium,
the paracellular pathway is sealed off by high resistance tight junctions, and the transcellular
current pathway is more important (Frömter and Diamond 1972, Powell 1981). The low
frequency transepithelial impedance for leaky epithelia is typically much lower than that for
tight epithelia.
In this paper, we describe, for the first time, a model of a glandular mucosa which has
straight glands. It is possible to change the parameters of the model to give it characteristics of
a ‘leaky’ epithelium, such as that found in the small intestine, or a ‘tight’ epithelium, such as
that found in the body of the stomach. This model has been used to investigate the distinction
between abnormal intestinal-type Barrett’s oesophageal tissue and the normal gastric-type
epithelium of the stomach.
We have also included results for a model of cervical squamous epithelium developed
previously by Walker et al (2002). This model allows us to investigate the changes occurring
with dysplasia, where the epithelium becomes much more leaky than normal.
2. Tissue models
Finite element analysis was used to determine the voltage distribution produced in the tissue
when current is applied through external electrodes. This is a numerical technique in which the
volume to be modelled is split up into discrete regions of space (elements) whose interactions
are mediated through common points (nodes) on the boundaries of the elements. Given
the conductivities of the individual elements, a matrix equation can be constructed which
relates the nodal voltages (Φ) to the currents injected at each node (I) through a conductivity
matrix (k):
kΦ = I.
The solution process involves inverting this equation to determine Φ. A review of finite
element analysis applied to bioelectric phenomena is given by Miller and Henriquez (1990).
100µm
columnar
cell
gland
lumen
lamina
propria
mucosa, which has straight glands. The lumen of the glands is surrounded by an epithelium
of a single layer of columnar cells. Connective tissue, known as the lamina propria, supports
this epithelium and provides a blood supply (Young and Heath 2000, pp 249–73). To cross
the mucosa, current flows down the gland lumen (assumed to be filled with conductive fluid),
crosses the epithelium and then flows into the lamina propria.
The glandular tissue model is similar in conception to that described by Clausen et al
(1983). They model a single gland and calculate transepithelial impedance using analytic
techniques. The gland is modelled as a transmission line, where the lumen of the gland
is connected to the lamina propria at regular intervals by leaky capacitors which represent
the epithelial cells. This leads to a distributed impedance and a frequency spectrum with a
characteristically broad dispersion.
We have created a FEM with many glands in it, allowing us to extend their idea to model
the transverse conduction which occurs when surface tissue probes are used. The analysis
is split into two stages. In the first stage, the electrical properties of a section of gland are
calculated. In the second stage, the whole mucosa is built up from these gland sections and
the boundary conditions for the probe design of interest are applied.
2.1.1. Modelling—stage 1. Figure 4 shows the FEM used for the first part of the analysis.
The central section is the lumen of the gland, and this is surrounded by four cells. The lateral
intercellular spaces (LIS) are filled with a thin layer of conductive fluid. Around all of the cells,
at the edge of the model, is the lamina propria. In figure 4(a), parts of the model have been cut
away to allow details of the interior to be seen—one cell has been removed completely,
and two of the cells have been opened up to show the cytoplasm and cell membranes.
Figure 4(b) shows a sectional view of the model. The connectivity between sections in
the FEM is shown in table 1. The cell cytoplasm is joined to other sections of the model via
elements modelling a cell membrane with a given conductance and capacitance per unit area.
The lateral intercellular space is connected to the gland lumen via a tight junction. The tight
junction determines how much current can flow along the paracellular current path. Two types
of epithelium have been modelled—a ‘perfectly tight’ epithelium, in which the tight junction
has infinite resistance, and a ‘perfectly leaky’ epithelium, in which the tight junction has zero
resistance. These two extreme cases are used as approximations to gastric- and intestinal-type
epithelia, respectively.
Modelling of epithelial tissue impedance measured using three different designs of probe 609
B
tight junction
gland lumen
cytoplasm
C A B
lamina
propria
A cell D
intercellular membrane
space tight lateral cell membranes
D junction and intercellular space
(a) (b)
Figure 4. Gland section. Dimensions A to D are given in table 2. (a) 3D cutaway diagram of
FEM. (b) Section perpendicular to the gland axis.
The properties used in the model are shown in table 2. The cytoplasm conductivity
and the cell membrane conductance and capacitance are consistent with the range of values
given in a review by Geddes (1972). In addition, the value of cell membrane capacitance
divided by conductance (0.01 F for apical, lateral and basal membranes) is of the same
order of magnitude as values derived by Clausen et al (1983) specifically for gastric mucosa
(0.025 F for apical membrane and 0.0035 F for basolateral membrane). For simplicity,
the conductivities of all of the volume conductor sections in the model have been given the
same value, 1 S m−1. For comparison, Geddes and Baker (1967) give values in the range 0.72
to 1.60 S m−1 for animal serum, and values in the range 1.43 to 1.59 S m−1 for human plasma
(all values at body temperature). The thickness of lateral intercellular space is consistent with
the values given by Spring and Hope (1978). Other linear dimensions used in the model
are representative of typical values taken from our observations of micrographs of glandular
tissue.
The lamina propria provides a low resistance path for current flow. Its dimensions can
be changed by inflammation or oedema. The model has been solved with two thicknesses of
lamina propria, which lead to either 8% or 15% of the total mucosa volume being occupied
by lamina propria.
The electrical properties of the stage 1 FEM of the gland section are characterized by
applying four different boundary condition sets, as shown in table 3.
610 D M Jones et al
2.1.2. Modelling—stage 2. The gland section model in stage 1 consists of 13 344 elements
and occupies a volume of 50 µm × 50 µm × 20 µm = 5 × 104 µm3. To simulate a
measurement made using a probe of the size of the order of 1 mm, a tissue volume of at
least 109 µm3 must be modelled—a factor 20 000 greater. If the gland section model was
simply replicated to cover this volume, then this would produce a model with 267 million
elements. It is impractical to solve a model of this size using the currently available computing
power. The model of the gland section must be reduced to a single element encapsulating
its essential electrical properties. The essential properties are exactly those four described
in table 3, obtained by solving the stage 1 model with four different boundary condition
sets. The single ‘gland section’ element which is given these properties has ten nodes
(figure 5(c)) and is constructed as follows. It is based on the standard eight node solid
hexahedral element used to model conduction in a continuum (figure 5(a)). In general, the
conductivity can be anisotropic, with different principal conductivities, ‘KXX’, ‘KYY’ and
‘KZZ’ being specified along the coordinate axes. ‘Z’ is taken to be along the direction of the
gland. The symmetry in the model indicates that the ‘x’ and ‘y’ directions are interchangeable,
and so KXX = KYY. In order to model conduction along glands, two nodes are added to
this element, in the centre of the top and bottom faces. The eight original nodes can be
regarded as being in the lamina propria, and the two new nodes are in the gland lumen. This
additional compartment to the element is required because it is possible for the gland lumen
and adjacent lamina propria to be decoupled electrically from each other so that the electric
Modelling of epithelial tissue impedance measured using three different designs of probe 611
(a)
Bulk conduction in
uniform medium
K ZZ
KYY (=KXX )
KXX
(b)
+
Conduction along gland Z Z
lumen and between lumen
and lamina propria Z
Z
R
Z
Z
Z Z
(c)
10 node’gland section’
=
hybrid element
R
Z
KXX
KZZ
Figure 5. The hybrid ‘gland section’ element (c) is made up from a section representing conduction
in a uniform medium (a) and a section representing conduction between the lamina propria and the
gland lumen and conduction along the gland lumen (b).
potential cannot be represented in a single continuum. Figure 5(b) shows the additional
conduction pathways introduced into the element. Resistance ‘R’ models conduction along
the gland lumen, and impedances ‘Z’ model conduction between the lumen and the lamina
propria.
The values of R, Z, KXX and KZZ are specified so that the element behaves in the same
way as the FEM in stage 1 when subjected to the same four boundary condition sets.
The model for the second stage of analysis is constructed by stacking gland section
elements vertically to form complete glands, and then repeating glands horizontally to form
a three-dimensional section of mucosa. A conductive mucous layer is added to the surface
of the mucosa. Because this layer is thin, two-dimensional ‘shell’ elements are used. These
model transverse conduction across the surface of the mucosa, but have zero impedance in the
vertical direction. A thick, uniform layer of submucosa is added to the bottom of the model.
612 D M Jones et al
The connectivity of the FEM is such that current enters the top of the mucosa only through the
gland lumen nodes in the centre of the gland section elements. Conversely, current leaves the
bottom of the mucosa only through the lamina propria nodes on the edges of the elements.
This ensures that at some point, the current must cross the epithelial cell barrier, represented
by impedance ‘Z’ in figure 5(b). The properties of the components of the stage 2 model are
shown in table 4. The complete glandular tissue models were solved at 36 frequencies in the
range 1 Hz–10 MHz.
SF M M
SF
C E C E
(a) M (b) I
G M G SF I G M G I
SF
E
(c) I (d)
Because this probe makes in vivo measurements, submucosal tissue has been included in the
models in this case.
Conceptually, the guarded probe design was developed from the transepithelial measurement
configuration. In a transepithelial measurement, the current flow is perpendicular to the surface
fluid layer. We wanted to reproduce this situation with a probe which could be placed on the
lumenal surface of the epithelium.
Figure 6(a) shows a schematic diagram of a transepithelial measurement, with an electrical
seal being made by fixing the epithelium in a clamp (‘C’ in the figure). In order to make
in vivo measurements, the clamp must be removed from the tissue (figure 6(b)). Since it is the
impedance of the epithelium itself which is of interest, and not the underlying submucosa, the
serosal electrode is made into a very large indifferent electrode (I). Any conductive fluid on
the tissue surface (SF) will cause the impedance to decrease. This is because the current can
spread out laterally, increasing the effective area of the measurement electrode (M).
614 D M Jones et al
I
G
M
1
V
Figure 7. Guarded probe with concentric electrodes. M is the measurement electrode (radius 0–
400 µm), G is the guard electrode (radius 400–840 µm) and I is the indifferent electrode (radius
2435–5985 µm).
To prevent this from happening, an annular guard electrode (G) can be placed just outside
the measurement electrode (figure 6(c)). The guard electrode is maintained at the same
voltage as the measurement electrode. This ensures a well-defined area for current injection
by preventing lateral current flow.
The final step is to consider the position of the indifferent electrode (I). It is possible to
move this from the serosal surface to the lumenal surface of the epithelium (figure 6(d)). The
indifferent electrode is now in the shape of a large annulus surrounding both the measurement
and the guard electrodes. There will now be some contribution to the overall impedance from
the epithelium underlying the indifferent electrode. However, if the area of the indifferent
electrode is much greater than the area of the measurement electrode, most of the contribution
to the impedance will arise from the epithelium underneath the measurement electrode.
The electrical connections to the electrodes are shown in figure 7. A feedback loop with
a unity gain amplifier maintains the guard electrode at the same voltage as the measurement
electrode. The current injected through the measurement electrode is controlled, and this
is termed the measurement current. The impedance is calculated as the voltage at the
measurement electrode divided by this measurement current. Clearly, current is also injected
at the guard electrode, but this does not form part of the measurement current used in the
calculation. The purpose of the guard electrode is simply to prevent lateral flow of the
measurement current. Figure 7 also shows the dimensions of the electrodes used in the model,
with the gap between the guard and measurement electrodes being infinitesimally small.
Boundary conditions were applied to the models to simulate this drive scheme.
Submucosal tissue has been included in the models for this measurement.
4. Results
Three pairs of comparisons between epithelial models were made (table 5). The first
comparison considers the effect of changing the tight junction resistance in the glandular
tissue model. The second comparison considers the effect of lamina propria volume fraction
in the glandular tissue model. Finally, normal and CIN3 squamous models are compared.
Modelling of epithelial tissue impedance measured using three different designs of probe 615
Table 5. Comparison between three pairs of tissue models. Models can be compared using the
impedance at a single frequency, indicated by ‘f ’ in the table.
The modelled impedance spectra are shown in figure 8. The spectrum for the tight glandular
epithelium shown in figure 8(a) has two frequency dispersions. The step size of the low
frequency dispersion depends upon the ‘tightness’ of the epithelium—high resistance tight
junctions are needed in order to see a dispersion. For the leaky epithelium model, the low
frequency dispersion disappears. The step size of the high frequency dispersion depends upon
on the lamina propria volume fraction, as shown by figure 8(b).
For the normal squamous model (figure 8(c)), there is just one dispersion. This dispersion
is significantly reduced in size for the leaky CIN3 model.
For both squamous and glandular tissue, the epithelial barrier integrity determines the
magnitude of the low frequency impedance. Pre-cancerous (CIN3) tissue presents a much
poorer barrier than normal tissue and hence has a lower impedance. Similarly, for glandular
tissue, leaky intestinal type epithelium is less of a barrier than tight gastric type epithelium. In
squamous tissue, however, the impedance attains its maximum value at a higher frequency—
approximately 100 Hz—as compared to glandular tissue, where the impedance reaches its
maximum at approximately 1 Hz.
The modelled impedance spectra are shown in figure 9. Comparing these spectra to the
corresponding spectra in figure 8, it can be seen that the general form is the same. There
are slight differences in shape, the most noticeable being the low frequency dispersion in
tight glandular tissue. For the transepithelial measurement (figure 8(a)), there is quite a long
tail to this dispersion in the region 102–104 Hz, but this tail is absent in the four-electrode
measurement (figure 9(a)).
Another significant difference is the fact that the four-electrode probe measurement can
be affected by conductive surface fluid. The spectra shown are for a surface layer of thickness
10 µm and conductivity 1 S m−1. In order to further investigate the effect of surface fluid,
its conductivity was fixed at 1 S m−1 and its thickness varied in 13 steps from 0.01 to
100 µm. Note that because the conductive layer is thin, doubling the conductivity will have
the same effect as doubling the thickness. In experiments, the conductivity can be decreased
by washing the tissue surface with tap water, and this has the same effect as reducing the fluid
thickness.
The impedance measured at a particular frequency is plotted against the thickness of
surface fluid (figure 10). The frequencies used for the three comparisons are shown in
table 5, and the corresponding points in figure 9 have been circled. Comparisons (a) and (c)
are between tight and leaky epithelia. In these cases, the lowest modelled frequency (1 Hz for
the glandular model, 100 Hz for the squamous model) is used to give the greatest separation.
616 D M Jones et al
400
300
200
100
0 0 1 2 3 4 5 6 7
10 10 10 10 10 10 10 10
frequency, Hz
(a)
70
60
magnitide of impedance, Ωcm2
50
40
8% lamina propria, leaky
30 15% lamina propria, leaky
20
10
0 0 1 2 3 4 5 6 7
10 10 10 10 10 10 10 10
frequency, Hz
(b)
350
300
250
200
150
100
50
0 2 3 4 5 6
10 10 10 10 10
frequency, Hz
(c)
Figure 8. Modelled transepithelial impedance spectra. (a) Glandular tissue, effect of epithelial
tightness, (b) glandular tissue, effect of lamina propria volume fraction, (c) squamous tissue,
comparison of normal and pre-cancerous (CIN3) states.
Modelling of epithelial tissue impedance measured using three different designs of probe 617
magnitude of impedance, kΩ
1
0.8
0.6
0.4
0.2
0 0 1 2 3 4 5 6 7
10 10 10 10 10 10 10 10
frequency, Hz
(a)
0.16
0.14
magnitude of impedance, kΩ
0.12
0.1
0.08
0.02
0 0 1 2 3 4 5 6 7
10 10 10 10 10 10 10 10
frequency, Hz
(b)
0.8
0.6
0.4
0.2
0 2 3 4 5 6
10 10 10 10 10
frequency, Hz
(c)
Figure 9. Modelled spectra for four-electrode probe measurement. (a) Glandular tissue, effect of
epithelial tightness, (b) glandular tissue, effect of lamina propria volume fraction, (c) squamous
tissue, comparison of normal and pre-cancerous (CIN3) states. Circled points are used to
distinguish between tissue types, and correspond to the circled points in figure 10.
618 D M Jones et al
0.8
0.6
0.4
0.2
0
0.01 0.1 1 10 100
thickness of 1S/m surface fluid, µm
(a)
0.18
magnitude of impedance at 10kHz, kΩ
0.16
0.14
0.12
0.1
0.08
0.02
0
0.01 0.1 1 10 100
thickness of 1S/m surface fluid, µm
(b)
normal
magnitude of impedance at 100Hz, kΩ
0.8
0.6
0.4
0.2
0
0.01 0.1 1 10 100
thickness of 1S/m surface fluid, µm
(c)
Figure 10. Modelled effect of conductive surface fluid on the four-electrode probe measurement
at a specific frequency. (a) Glandular tissue, effect of epithelial tightness, frequency = 1 Hz,
(b) glandular tissue, effect of lamina propria volume fraction, frequency = 10 kHz, (c) squamous
tissue, comparison of normal and pre-cancerous (CIN3) states, frequency = 100 Hz. Common
points in figures 9 and 10 have been circled.
Modelling of epithelial tissue impedance measured using three different designs of probe 619
A frequency of 10 kHz is used to compare glandular tissue models with different lamina propria
volume fractions. The impedance at this frequency is unaffected by epithelial tightness.
It can be seen that the surface fluid has a large effect on the impedance of normal squamous
tissue at 100 Hz and tight glandular tissue at 1 Hz. It also has an effect, albeit smaller, on the
impedances measured for the other models.
5. Discussion
The general shape of the impedance spectra is similar for the three different types of
measurement configuration. Slight differences are seen with the four-electrode probe, where
the dispersions appear to be narrower. In addition, the four-electrode measurement can be
significantly affected by surface fluid, as shown in figure 10.
In both squamous and glandular tissue, the epithelial tightness affects the low frequency
impedance measured with all types of probe. However, the structure of squamous and glandular
tissue is very different. The squamous model is composed of several layers of cells, whereas
the glandular model has only a single cell layer, which is folded to create glands. These
differences mean that maximum impedance occurs at a lower frequency for glandular tissue
than for squamous tissue. The folded nature of the glandular epithelium means that the
supporting lamina propria is present throughout the depth of the mucosa. The lamina propria
provides a low resistance current pathway, and the effect of this is seen in a second dispersion
at higher frequencies in the impedance spectrum.
For glandular tissue, it is possible that other model parameters will affect the impedance
spectrum, e.g., the epithelial depth, gland size and the spacing between glands. In particular,
changes to parameters which have the effect of increasing the effective epithelial surface area
will make the epithelium appear to be more conductive. In certain glandular epithelia, such as
the small intestine, villi are present, but these have not been included in our model. Their effect
would be to increase the epithelial surface area, and hence increase the overall conductance of
the tissue.
The transepithelial measurement is independent of surface fluid. At low frequency, it
distinguishes well between tight and leaky epithelia. For glandular tissue, the impedance at
higher frequencies (∼10 kHz) can be used to provide an indication of the lamina propria
volume fraction. However, it is not practical to measure transepithelial impedance for
in vivo clinical diagnosis because access to the serosal surface is required, and the tissue
sample must be fixed within a measurement chamber. Non-invasive measurements must be
considered.
Tetrapolar surface probes can and have been used to make in vivo measurements, but
they are affected by surface fluid, and this complicates interpretation of experimental results.
Interestingly, it appears that some surface fluid is actually required to see a difference between
620 D M Jones et al
magnitude of impedance, kΩ
100
80
60
40
20
0 0 1 2 3 4 5 6 7
10 10 10 10 10 10 10 10
frequency, Hz
(a)
14
magnitude of impedance, kΩ
12
10
6
8% lamina propria, leaky
4 15% lamina propria, leaky
0 0 1 2 3 4 5 6 7
10 10 10 10 10 10 10 10
frequency, Hz
(b)
normal
100
CIN3 (pre–cancer)
magnitude of impedance, kΩ
80
60
40
20
0 2 3 4 5 6
10 10 10 10 10
frequency, Hz
(c)
Figure 11. Modelled spectra for concentric guarded probe measurement. (a) Glandular tissue,
effect of epithelial tightness, (b) glandular tissue, effect of lamina propria volume fraction,
(c) squamous tissue, comparison of normal and pre-cancerous (CIN3) states. Circled points
are used to distinguish between tissue types, and correspond to the circled points in figure 12.
Modelling of epithelial tissue impedance measured using three different designs of probe 621
120
60
40
20
0
0.01 0.1 1 10 100
thickness of 1S/m surface fluid, µm
(a)
10
0
0.01 0.1 1 10 100
thickness of 1S/m surface fluid, µm
(b)
100
80
normal
60 CIN3 (pre–cancer)
40
20
0
0.01 0.1 1 10 100
thickness of 1S/m surface fluid, µm
(c)
Figure 12. Modelled effect of conductive surface fluid on the guarded concentric probe
measurement at a specific frequency. (a) Glandular tissue, effect of epithelial tightness, frequency =
1 Hz, (b) glandular tissue, effect of lamina propria volume fraction, frequency = 10 kHz,
(c) squamous tissue, comparison of normal and pre-cancerous (CIN3) states, frequency = 100 Hz.
Common points in figures 11 and 12 have been circled.
622 D M Jones et al
the tight and leaky epithelia—if the surface fluid thickness is reduced to zero, then the
distinction between the two types becomes very small (figures 10(a) and (c)). For low values
of surface fluid thickness, the slope of the graphs in this figure is positive. This corresponds to
a negative sensitivity for the transfer impedance, i.e. increasing the conductance of the surface
path increases the transfer impedance. For high values of surface fluid thickness, the slope of
the graphs becomes negative, corresponding to a positive sensitivity.
Although the tetrapolar impedance is generally higher in a tight epithelium than in a leaky
epithelium, the amount by which the impedance is higher is strongly dependent on the surface
fluid properties. This makes it impossible to make quantitative predictions about the tightness
of epithelia with intermediate properties from a four-electrode surface probe measurement.
An average of many readings may show differences between tissue types, but there will be
a significant amount of scatter on individual data points. Similarly, although an increase
in the lamina propria volume fraction will, on average, decrease the impedance at 10 kHz
(figure 10(b)), individual readings may be affected by variations in surface fluid properties.
Experimental impedance measurements suggest that for unwashed tissue samples, the surface
fluid thickness is somewhere in the region of 10 to 100 µm.
Ideally, we would like the plots of impedance versus surface fluid thickness to show
horizontal lines, with a consistent, large separation between different tissue classifications.
The results for the guarded concentric probe (figure 12) are very encouraging from this point
of view, with large separations between the different pairs of tissue types. This shows that in
principle, the guarding scheme is effective in preventing the surface fluid from affecting the
results.
However, the practicalities of probe measurement must also be considered. Because
the guarded probe makes a bipolar measurement, the electrode impedance contributes to the
measurement, and this must be minimized as far as possible. This is especially important
given that electrode impedance can become very large at the low frequencies required to
investigate epithelial barrier integrity. The electrode impedance could become much greater
than the tissue impedance, and it could also prevent the guarding mechanism from functioning
effectively. There are several techniques for reducing the impedance of an electrode by treating
it to increase its effective surface area. One example is the platinization process for platinum
electrodes (Geddes 1972) which can decrease the electrode impedance by a factor of 100
to 1000. More recently, Hubmann et al (1992) have investigated treatments which involve
sputtering the electrode surface with either titanium nitride or iridium. This results in a fractal
surface which has a very high effective surface area.
For practical probe designs, a finite space between the measurement and guard electrodes
is required for insulation purposes. Computer modelling has shown that a large gap can
reduce the efficacy of the guarding process, and so the separation should be kept as small as
possible.
6. Conclusions
A new FEM of the electrical properties of glandular mucosa has been constructed. A previously
described model of squamous tissue was also used. These models allow us to determine how
the impedance spectrum is affected by changes in the tissue structure. In particular, the
tightness of the epithelium affects the low frequency impedance. In the glandular model, the
properties of the lamina propria affect the impedance at intermediate frequencies (∼10 kHz).
Measurements using Ussing chambers and tetrapolar surface probes have been simulated. The
tetrapolar surface probe measurements were found to be highly influenced by the presence of
even a thin layer of conductive fluid on the tissue surface. To counteract this effect, a new
Modelling of epithelial tissue impedance measured using three different designs of probe 623
design of probe, which uses a concentric guard electrode, has been considered. Modelling
has shown this design to be very effective at making measurements which are independent of
the surface fluid properties. However, in order to make practical measurements, the effect of
electrode impedance would need to be taken into account.
Acknowledgments
References
Barry P H 1989 Ionic permeation mechanisms in epithelia: biionic potentials, dilution potentials, conductances, and
streaming potentials Methods Enzymol. 171 678–715
Brown B H, Tidy J A, Boston K, Blackett A D, Smallwood R H and Sharp F 2000 Relation between tissue structure
and imposed electrical current flow in cervical neoplasia Lancet 355 892–95
Clausen C, Machen T E and Diamond J M 1983 Use of AC impedance analysis techniques to study membrane
changes related to acid secretion in amphibian gastric mucosa Biophys. J. 41 167–78
Falk G W 2002 Barrett’s esophagus Gastroenterology 122 1569–91
Frömter E and Diamond J 1972 Route of passive ion permeation in epithelia Nature New Biol. 235 9–13
Geddes L A 1972 Electrodes and the Measurement of Bioelectric Events (New York: Wiley)
Geddes L A and Baker L E 1967 The specific resistance of biological material—a compendium of data for the
biomedical engineer and physiologist Med. Biol. Eng. 5 271–93
Gitter A H, Schulzke J D, Sorgenfrei D and Fromm M 1997 Using chamber for high-frequency transmural impedance
analysis of epithelial tissues J. Biochem. Biophys. Methods 35 81–8
Gonzalez-Correa C A, Brown B H, Smallwood R H, Kalia N, Stoddard C J, Stephenson T J, Haggie S J, Slater D N
and Bardhan K D 1999 Virtual biopsies in Barrett’s esophagus using an impedance probe Ann. NY Acad. Sci.
873 313–21
Hubmann M, Hardt R, Bolz A and Schaldach M 1992 Long-term performance of stimulation and sensing behaviour
of TiN and Ir coated pacemaker leads having a fractal surface structure Proc. 14th Ann. Int. Conf. IEEE (Paris)
pp 2379–80
Jones D M, Smallwood R H, Hose D R and Brown B H 2001 Constraints on tetrapolar tissue impedance measurements
Electron. Lett. 37 1515–7
Miller C E and Henriquez C S 1990 Finite element analysis of bioelectric phenomena Crit. Rev. Biomed. Eng. 18
207–33
Pethig R 1984 Dielectric properties of biological materials: biophysical and medical applications IEEE Trans. Electr.
Insulat. 19 453–74
Powell D W 1981 Barrier function of epithelia Am. J. Physiol. 241 G275–88
Spring K R and Hope A 1978 Size and shape of the lateral intercellular spaces in a living epithelium Science 200
54–8
Walker D C, Brown B H, Smallwood R H, Hose D R and Jones D M 2002 Modelled current distribution in cervical
squamous tissue Physiol. Meas. 23 159–68
Young B and Heath J W 2000 Wheater’s Functional Histology (London: Harcourt)