Biacore™ Assay Handbook: GE Healthcare Life Sciences
Biacore™ Assay Handbook: GE Healthcare Life Sciences
Biacore™ Assay Handbook: GE Healthcare Life Sciences
Life Sciences
2 Application overview
2.1 Screening and detecting binding partners ............................... 9
2.2 Kinetics and affinity measurements ........................................ 10
2.2.1 Kinetic analysis...........................................................................................10
2.2.2 Affinity analysis ..........................................................................................11
2.3 Concentration measurements .................................................. 13
2.3.1 Calibrated assays......................................................................................13
2.3.2 Calibration-free concentration analysis (CFCA)............................14
2.4 Mapping binding sites ................................................................. 14
3 General considerations
3.1 Sensor surfaces ............................................................................ 17
3.1.1 General properties ....................................................................................17
3.1.2 Sensor surface types................................................................................17
3.2 Attaching the ligand .................................................................... 19
3.2.1 Covalent immobilization ........................................................................19
3.2.2 High affinity capture ................................................................................21
3.2.3 Response levels..........................................................................................21
3.3 General buffer considerations ................................................... 23
3.3.1 Buffer substances......................................................................................23
3.3.2 Ionic strength ..............................................................................................23
3.3.3 Additives........................................................................................................24
3.3.4 Buffer preparation ....................................................................................25
3.4 Matching sample and running buffer ...................................... 26
3.4.1 Matching refractive index......................................................................26
3.4.2 Matching buffer composition...............................................................27
3.4.3 Matching buffer environment in practice .......................................28
3.5 Dilution series and replicates .................................................... 28
3.5.1 Dilution series..............................................................................................28
3.5.2 Replicates .....................................................................................................29
3.6 Preparing vials and microplates ............................................... 30
6 Epitope mapping
6.1 Pair-wise binding ..........................................................................45
6.1.1 Principle.........................................................................................................45
6.1.2 Evaluation.....................................................................................................46
6.2 Peptide inhibition .........................................................................48
7 Troubleshooting assays
7.1 Troubleshooting ligand attachment .........................................49
7.1.1 Covalently attached ligand...................................................................49
7.1.2 Captured ligand .........................................................................................50
7.2 Troubleshooting analyte binding ..............................................51
7.2.1 Analyte capacity too low .......................................................................51
7.2.2 Analyte binding too high........................................................................51
7.3 Dealing with non-specific and unwanted binding .................51
7.3.1 Non-specific binding ................................................................................51
7.3.2 Unwanted binding ....................................................................................52
7.4 Unexpected sensorgram shapes ...............................................52
7.4.1 Unstable baseline......................................................................................53
7.4.2 Sample response below baseline.......................................................54
7.4.3 “Humpbacked” sensorgrams...............................................................55
7.4.4 Anomalous response during buffer flow ........................................55
7.4.5 Response subtraction spikes................................................................56
7.4.6 Regular disturbances ..............................................................................56
1 Introduction
Biacore systems can be used to study interactions involving (in principle) any
kind of molecule, from organic drug candidates to proteins, nucleic acids,
glycoproteins and even viruses and whole cells. Since the response is a measure
of the change in mass concentration, the response per molar unit of interactant
is proportional to the molecular weight (smaller molecules give lower molar
responses). The practical lower limit for detection of small molecules with
today’s instrumentation is about 100 Da.
The detection principle does not require any of the interactants to be labeled,
and measurements can be performed on complex mixtures such as cell culture
supernatants or cell extracts as well as purified interactants. The identity of the
interactant monitored in a complex sample matrix is determined by the
interaction specificity of the partner attached to the surface. The SPR detection
principle is non-invasive and works equally well on clear and colored or opaque
samples.
• screening for binding partners and ranking of binding ability, for instance
in drug discovery and biopharmaceutical development
• Entry-level systems like Biacore X100 have one or two pairs of flow cells
(where one cell in a pair is used for detecting the interaction and the other
as a reference cell where required) and can handle only relatively few
samples without user intervention.
• Advanced systems may have multiple detection spots (in Biacore 4000, for
example, 20 detection spots are divided between 4 flow cells for parallel
analysis of multiple samples) and unattended sample handling capacity
for 1500 samples or more.
While the available options for assay development and execution may vary
between the different systems, the same general principles apply to all systems.
This book will focus as far as possible on common principles of Biacore assays,
keeping system-specific information to a minimum.
• The interaction partner attached to the surface is called the ligand (Figure
1-1). In drug discovery and development work, the ligand is sometimes
referred to as the target molecule.
Note: The term “ligand” is applied here in analogy with terminology used in
affinity chromatography contexts, and does not imply that the
surface-attached molecule is a ligand for a cellular receptor.
analyte
analyte
ligand
ligand
capturing
molecule
sensor
surface
Figure 1-1. The ligand is the interaction partner that is attached to the sensor
surface. The ligand may be immobilized directly on the surface (left) or attached
through binding to an immobilized capturing molecule (right). The analyte is free in
solution and binds to the immobilized ligand.
Absolute
response (RU)
Report point
(response)
Relative
Report point response (RU)
(baseline)
Time
Figure 1-2. Schematic illustration of a sensorgram. The bars below the sensorgram
curve indicate the solutions that pass over the sensor surface.
2 Application overview
This chapter gives an overview of the major application areas for Biacore
systems. These are
• Screening for binding partners, including immunogenicity testing
• Measurement of interaction kinetics and affinity
• Measuring concentration
• Mapping binding site topography
Note: The term “kinetics” is used in this handbook to refer to interaction kinetics.
The kinetics of other processes such as enzyme-catalyzed reactions are
not normally amenable to study in Biacore systems.
It is important to be aware that the results of kinetic analysis are relevant only
in the context of the interaction model chosen for the evaluation. It is not strictly
possible to derive a model from the observed binding behavior, although the
shape of the sensorgrams can sometimes give clues for the choice of
appropriate models. Fitting the data to a mathematical model does not provide
evidence of the interaction mechanism (Figure 2-1).
Figure 2-1. Fitting experimental data to a mathematical model does not prove that the
model is appropriate, and the same data may fit acceptably to several models. In this
example, the closeness of fit cannot distinguish confidently between a bivalent analyte
model (left) and a heterogeneous ligand model (right).
Assay set-up
There are currently two ways of setting up a kinetic analysis experiment (Figure
2-2):
Figure 2-2. In multi-cycle kinetics and affinity determinations (left), each sample is injected
in a separate cycle. The concentration series is presented as an overlay plot aligned at the
start of the injection in the evaluation software.
In single-cycle determinations (right), the samples are injected sequentially in the same
cycle.
Arrows in the illustrations mark the start of sample injections.
Note that these values, like the values for rate constants, are only valid in the
context of the model used to analyze the binding data. More complex binding
mechanisms do not always give straightforward relationships between affinity
and kinetics, and it may only be possible to derive affinity constants from kinetic
constants for values obtained with a simple interaction model.
Figure 2-3. Sensorgrams (left) and corresponding plot of steady state response against
concentration (right) for determination of binding affinity. The vertical line in the right-
hand plot indicates the value of the calculated equilibrium dissociation constant KD.
Analysis of steady-state binding data by fitting the curve of binding level against
concentration can only be applied to 1:1 binding. A model for 1:1 binding at two
independent ligand sites can be useful in some circumstances, but more
complex models are not sufficiently robust for analysis of the data. It is however
possible to apply classical Scatchard analysis to data obtained from Biacore
systems, to give some indication of the causes of deviation from 1:1 binding
behavior.
It may be thought that the same experiment can in principle provide data for
both kinetics and steady state binding, offering two independent affinity values
from one set of data. In practice, however, restrictions on the injected volume of
sample (and therefore the contact time between sample and the sensor surface)
mean that interactions that give measurable kinetics seldom reach steady state
and vice versa. The affinity for a given interaction can therefore usually be
derived from kinetics or steady state measurements but seldom both.
Affinity in solution
The principle of measuring affinity in solution with Biacore systems is to
determine the free concentration (and therefore the concentration of complex)
in mixtures of known interactant concentrations, using the Biacore system for
the concentration assay. Commonly, the assay is set up with one interactant A
(or an analogue thereof) on the surface. A calibration curve is established for
determining the concentration of the second interactant B. Mixtures containing
variable concentrations of A and a fixed concentration of B are allowed to reach
equilibrium and then analyzed for the free concentration of B. The results are
analyzed using standard tools for affinity measurements. A requirement of this
approach is that only free B and not complex AB can bind to the surface-
attached A.
• Direct assays measure analyte bound directly to the ligand on the sensor
surface. This approach is suitable for macromolecular analytes (molecular
weight > 5000 daltons): direct detection of smaller molecules is possible
but the useful range of the assay is generally limited in such cases.
Response enhancement or sandwich approaches can be used to amplify
the response obtained and/or to increase the selectivity of the assay.
• Indirect or competition assays provide an indirect measure of analyte
concentration, and are most useful for low molecular weight analytes. The
commonest format for indirect assays is the solution competition or
inhibition approach, where a known amount of an interacting partner
called the detecting molecule is mixed with the sample, and the amount of
free detecting molecule remaining in the mixture is measured. In the
surface competition method, analyte and a high molecular weight
analogue (often a protein conjugate) compete for binding to a common
partner on the sensor chip surface. In both competition assay formats, the
response obtained is inversely related to the concentration of analyte in
the sample.
Figure 2-4. Different antibodies can bind simultaneously to distinct and spatially
separated epitopes (left), but not to common or overlapping epitopes (right).
3 General considerations
This chapter discusses aspects of working with Biacore systems that are
common to all application areas.
Figure 3-1. Schematic illustration of the structure of the sensor chip surface on CM-series
chips.
Hydrophobic surfaces
Amine coupling
Immobilized: 10,000 RU
Activity: undetectable
Thiol coupling
Immobilized: 12,000 RU
Activity: 5%
Figure 3-2. Amine coupling is frequently successful, but in this example (immobilization of
a hormone receptor) amine coupling destroys the ligand activity completely. The best
results are obtained with high affinity capture.
• Capture does not require any chemical modification of the ligand and can
be performed in physiological buffer conditions. It can therefore be used
with ligands that are not amenable to covalent immobilization.
Conditions for ligand capture are dictated by the requirements of the capturing
interaction, and capturing is normally performed in the same buffer conditions
as the interaction being studied. It is however important that the capturing
interaction is reasonably stable under the chosen conditions, so that ligand
does not dissociate significantly during sample analysis. Demands on capturing
stability vary with different applications, and are highest for kinetic and affinity
analyses, since loss of ligand will cause a fall in response as well as reducing the
analyte binding capacity of the surface. Simpler applications such as detection
and screening may tolerate more loss of ligand during sample injection.
Figure 3-3. Sensorgram from a typical amine coupling, illustrating the distinction between
the amount of ligand bound and the amount immobilized.
Note: Even activation of the surface with EDC/NHS can be detected as a small
increase in response. When the response from immobilized ligand is low,
it may be more appropriate to estimate the ligand level from the response
after activation rather than the baseline at the start of the immobilization
procedure.
analyte MW
R max RU = ------------------------------ immobilized ligand level (RU)
ligand MW
The shape of the sensorgram from the attachment procedure can provide
guidance in choosing appropriate conditions. Bear in mind that the amount of
ligand immobilized will generally be less than the amount bound during
preconcentration. Most Biacore systems support a mode of ligand
immobilization using amine coupling that exploits the information from a brief
• The sensorgram slope and shape during protein injection indicates the
efficiency of electrostatic preconcentration. Try to use a combination of
ligand concentration and injection time so that the amount bound reaches
a plateau. This will contribute to the robustness of the immobilization
procedure.
• The relative response levels before and after deactivation of the surface
indicate how well the electrostatically bound protein has become
covalently attached. Deactivation washes out non-covalently bound
material: a reduction in response may be observed after deactivation but
this does not affect the assay as long as sufficient ligand is attached to the
surface.
Phosphate-based buffers are often more suitable for work with low molecular
weight analytes since organic buffers such as HEPES can bind to the ligand and
interfere with accurate detection of low molecular weight compounds.
3.3.3 Additives
Detergent
Inclusion of non-ionic detergent (0.05% Surfactant P20, Tween™ or equivalent)
in all buffers is recommended to minimize deposition of protein and other
biomolecules in the flow system. The detergent should be included at a
concentration above the critical micelle concentration (CMC). At lower
concentrations, the detergent itself can accumulate in the flow system and lead
to unwanted response effects as it is released (see FIgure 3-4).
Figure 3-4. Schematic illustration of typical “detergent effects” caused by using detergent
at concentrations below CMC. The broken line shows an undisturbed sensorgram for
comparison.
EDTA
HBS-EP and HBS-EP+ general purpose buffers from GE Healthcare include 3 mM
EDTA to remove traces of free divalent metal ions that may be present in the
buffer. This is a general precautionary measure and is not specifically required
for work with Biacore systems. Buffers with no added EDTA are also available.
NSB Reducer
Addition of NSB reducer (a preparation of soluble carboxymethyl dextran) to
samples can in some cases help to reduce non-specific binding of sample
components to the dextran matrix on the sensor surface. The recommended
concentration of NSB Reducer is 1 mg/ml. NSB Reducer should only be used in
situations where non-specific binding is a problem (typically with complex
samples such as serum or whole cell extracts).
Organic solvents
Many low molecular weight organic compounds, relevant particularly in
pharmaceutical development work, are sparingly soluble in aqueous buffers
and require addition of organic solvents. Dimethyl sulfoxide (DMSO) is commonly
used. Biacore systems can be used with buffers containing up to 10% DMSO.
However, organic solvents contribute significantly to the bulk refractive index of
samples and buffers: 1% DMSO contributes approximately 1200 RU to the
response level. Expected responses from low molecular weight analytes, which
often need organic solvents to maintain solubility, may be as low as 10-20 RU or
less. It is therefore crucial that the measured responses are accurately
compensated for any variations in organic solvent concentration. Procedures
for adjusting measured sample responses for solvent effects, called solvent
correction, are described in principle in Appendix B and in more detail in the
documentation for Biacore systems where the correction is supported.
Other additives
Avoid additives that can compete with the ligand for immobilization chemistry,
for example sodium azide with amine coupling chemistry. Even at low
concentrations, sodium azide can effectively prevent amine coupling of
proteins.
Figure 3-6. Airspikes and transient baseline displacements are often attributable to
inadequately degassed running buffer.
Filtering
Buffers should be filtered through a 0.22 µm filter to remove particles.
Degassing
Degas running buffer before use unless you are working with a Biacore system
fitted with an in-line degasser. Alternatively use degassed ready-to-use buffers
from GE Healthcare.
response
time
Figure 3-7. Schematic illustration of bulk shifts: negative (left), zero (center) and positive
(right).
response
time
Figure 3-8. If dissociation is slow enough, binding levels can be assessed from a report
point placed shortly after the end of the sample injection (left). However, if dissociation is
rapid (right) a report point at this position will not give a reliable measure of the binding
level.
It is not always possible to match the refractive index of sample and running
buffer, for example in analysis of unpurified samples such as serum. In such
cases it is important that measurements are made after the end of the sample
injection to avoid bulk shift effects.
Reliable pipettes are essential for creating accurate dilution series by volume.
Where possible (particularly for small volumes), use repeated aliquots of the
same volume for dilution series, to avoid errors that may be introduced by
repeatedly changing the volume on variable volume pipettes.
• Serial dilution where each dilution is used as the sample solution for the
next step.
• Parallel dilution where sample is always taken from the same stock
solution and mixed with increasing volumes of buffer.
Using the same fixed volume pipettes for both sample and buffer provides the
best accuracy for dilution factor (although not necessarily for absolute
volumes). If different pipettes are used with different errors of delivery, the
dilution error will propagate through the series more significantly with serial
dilution than with parallel. However, serial dilution is a simpler procedure,
requiring fewer operations with a fixed-volume pipette, so that parallel dilution
is more susceptible to errors of reproducibility.
For the best accuracy in preparing dilution series, volumes of buffer and sample
used may be determined by weighing and the dilution factor calculated
individually for each sample. This avoids issues of both calibration accuracy and
reproducibility for pipetted volumes.
3.5.2 Replicates
Measuring replicate samples to establish the reproduciibility of an assay should
be a standard procedure in all laboratory work, but it is not always clear how the
replicates should be prepared and measured in order to provide the most useful
information. Replication can cover the whole assay procedure from raw sample
material to evaluation of the final results, or it can focus on one or more specific
• Place replicate samples in separate vials or microplate wells for the assay
wherever possible, so that only one sample is taken from each position.
Injecting multiple samples from the same autosampler position is not
recommended since evaporation from vials or wells that have been
penetrated can affect the concentration.
Beware of using samples that might precipitate or aggregate with time (for
example, low molecular weight organic compounds close to the limit of
solubility).
Cover samples as soon as they are prepared, and keep them covered for the
duration of the assay to prevent evaporation. Always use vial caps and
microplate foil for Biacore systems from GE Healthcare: the autosampler needle
may not penetrate caps or foil from other sources satisfactorily, resulting in
failed assays or even damage to needle. When sealing microplates with foil,
make sure the foil is correctly placed so that the wells are covered by areas free
from adhesive. If the foil is wrongly applied, adhesive may accumulate on the
needle and cause malfunction.
Screening applications using Biacore systems fall into two main categories that
differ significantly in challenges and experimental design:
• Small molecule (LMW) screening, where response levels are low and
organic solvents are often needed to maintain analyte solubility. These
interactions are usually rapid and regeneration is not needed.
Table 4-1 summarizes the main experimental conditions for LMW and antibody
screening.
Table 4-1. Main experimental conditions for LMW and antibody screening.
Additives:
- DMSO Yes (1 to 10%)1 No
- detergent Yes Yes
4.1.1 Goals
The primary goal of small molecule screening in pharmaceutical development
is to identify candidate molecular structures for further development, on the
basis of their binding to selected target molecules. The first stages of the
process is often screening of candidate libraries or fragment libraries to identify
promising binders for further development and refinement. Information-rich
screening with Biacore systems can provide a wealth of valuable data in
addition to yes/no binding results, including comparative binding to multiple
targets to identify promiscuous (and therefore less interesting) binding
candidates and assessment of kinetics and affinity properties that become
increasingly important as the candidate molecules are refined.
Many small organic compounds are sparingly soluble in aqueous buffers and
require inclusion of organic solvents (commonly 1% to 3% DMSO) to maintain
solubility. DMSO has a strong refractive index contribution (about 1200 RU for
1% DMSO), and both careful matching of sample and running buffer and
correction for small differences in DMSO concentration are required (see below).
4.1.4 Buffers
Phosphate buffers are generally recommended for work with small molecules.
Using organic buffers such as HEPES can bind to the ligand and interfere with
accurate detection of small organic compounds. Physiological ionic strength
(150 mM monovalent cations) should be used to reduce non-specific binding of
Flow rate
Use low flow rates (10 µl/min) for LMW screening applications to conserve
sample. Issues such as mass transfer limitations (Section A.1) and resolution of
fast binding events are not relevant.
Contact time
Most interactions involving small molecules are rapid, and contact times can be
kept short (30 to 60 seconds) to increase screening throughput.
Regeneration
Regeneration is generally not necessary in LMW screening applications, since
dissociation is rapid and is usually complete within 1 minute or thereabouts.
Report points
Because of the rapid dissociation of many LMW compounds from the target
molecule, report points for screening should be placed before the end of the
sample injection. LMW compounds sometimes show rapid binding to specific
sites followed by slower and more indiscriminate binding (Figure 4-1). Placing
report points early in the sample injection can reduce the impact of this
behavior and give a more accurate indication of specific binding ability.
Figure 4-1. Simulated sensorgrams illustrating 1:1 binding (dark green) and binding to one
specific site followed by slower and more indiscriminate binding (red).
Other considerations
Some LMW compounds are “sticky” and can be difficult to wash off the sensor
surface and out of the flow system, causing carry-over of material to the next
analysis cycle. This can be detected by routinely including a “carry-over
injection” of buffer after the sample injection: the response from a “sticky”
compound will be carried over into this buffer injection. Include a flow system
wash with 50% DMSO after every cycle to minimize problems with sticky
compounds. If a screening experiment involves compounds that are known to
be sticky, place these at the end of the experiment to avoid interference with
other compounds or omit them from the experiment altogether.
Figure 4-2. Plotting association rate constant ka against dissociation rate constant kd
(here on logarithmic scales) creates a plot where the affinity is represented by diagonal
lines. Compounds on the same diagonal have the same affinity but differ in kinetics.
These approaches are described in more detail in the handbooks for Biacore
systems that support fragment screening.
4.2.1 Goals
The goal of antibody screening is to identify cell clones that produce antibodies
appropriate for the purpose (usually biopharmaceuticals or biochemical tools).
Screening needs to be rapid so that clones can be selected before they grow to
unsuitable cell densities. Obtaining kinetic information early in the screening
process is often a significant advantage. Information that is generally sought
includes:
• Which antibodies have kinetic and/or affinity properties suitable for the
purpose
4.2.4 Buffers
HBS-EP+ (Hepes-buffered saline with 0.3 mM EDTA and 0.05% Surfactant P20,
available from GE Healthcare) or similar is recommended as running buffer.
Samples may be diluted if required in the same buffer. Precise matching of
sample and running buffer is not practicable for screening work.
Flow rate
Antibody screening by capture on a generic capturing molecule can be run at
low flow rates (typically 10 µl/min) if sample consumption is an issue.
Characterization of the antibody-antigen interaction, whether performed
directly on immobilized antigen or by a secondary injection over captured
antibody, should be run at moderate flow rates (typically 30 µl/min) to reduce
mass transfer limitations (see Section A.1).
Contact time
Contact times should be sufficient to give confidently measurable response
levels without compromising screening throughput. Contact times of 1 to 2
minutes are usually sufficient.
Regeneration
Regeneration conditions for antibody screening kits from GE Healthcare are
specified in the respective Instructions for Use. For assay development using
custom antibodies, regeneration at low pH (glycine-HCl, pH 1.5 to 3) is usually
effective.
Report points
For screening, place a report point shortly after the end of the sample injection.
This will avoid any bulk refractive index differences between sample and
running buffer.
Figure 4-3. Sensorgram shape can in some cases distinguish weak and strong binders
that reach the same response level as a result of different concentrations (left). In other
cases, depending on the relative values of the rate constants and concentrations, the
sensorgrams may be identical. These sensorgrams are simulated: those in the right-hand
panel are displaced slightly from each other on the time axis.
ADAs are released from drug complexes under acidic conditions (pH 1 to 2), and
can be detected if the sample is neutralized and the assay is performed
immediately, before the complexes have had time to re-form fully. This
approach is implemented in some Biacore systems, using a specially designed
IFC that mixes acidified samples with neutralizing solution immediately before
injection over the sensor surface. Details of the approach are described in the
manuals for systems that support immunogenicity testing (e.g. Biacore T200
Immunogenicity Handbook).
5.1 Approaches
Determination of both kinetics and affinity rely on measurements over a range
of analyte concentrations. Kinetic (rate) constants are obtained by analyzing the
sensorgram shape in relation to a mathematical model of the interaction
mechanism: affinity constants are obtained from the steady state binding
levels. Affinity constants can also be calculated from the ratio of the rate
constants. Details of the data analysis principles are given in Appendix A.
Both approaches have been shown to give equivalent results in test systems,
and both can be used for steady-state affinity measurements if the sample
injection time is long enough for steady state to be reached.
The single-cycle approach requires less time for a complete analysis, and also
benefits from the lack of requirement for optimized regeneration. This can
significantly reduce the demands on assay development, and permits kinetic
determinations for interactions where regeneration conditions are not
available. However, the cycle time is longer in single-cycle kinetics, so the
approach is more sensitive to drift such as dissociation of ligand in a capture
assay.
ligand densities in each flow cell. Combining the data from two ligand densities
compensates for the lack of analyte concentrations, and the approach can
provide reliable kinetic data from experiments that maintain a moderately high
sample throughput.
For kinetic analysis, the amount of immobilized ligand should be kept low, so
that the maximum response from analyte binding is typically in the region of 30
to 50 RU or lower. There are two main reasons for this:
• Low ligand levels help to reduce limiting effects of mass transfer. Mass
transfer refers to the diffusion-controlled supply of analyte molecules to
the surface from bulk solution: if transfer is slow in relation to the
association rate, the observed binding will be a measure of the diffusion
process and not of the interaction rates. Mass transfer is discussed in
more detail in Appendix A.
• Low ligand levels also help to minimize artifacts that can arise from
crowding of molecules on the surface. Crowding effects are difficult to
quantitate but can give rise to unnecessarily complex binding behavior.
Reference surface
Kinetic and affinity measurements should always use reference-subtracted
data. The reference surface may be untreated, activated and deactivated, or
prepared with an inactive protein to mimic the physical properties of the active
surface. These alternatives are discussed in more detail in the Biacore Sensor
Surface Handbook.
5.3 Buffers
The same buffer must be used for sample and running buffer, so that
association (during sample injection) and dissociation (after the end of the
injection) occur in the same environment. Moreover, evaluation of kinetics is
more robust if the sample and running buffer are matched in refractive index, so
that bulk shifts at the beginning and end of the sample injection are kept as
small as possible.
Figure 5-1. Plot of response against analyte concentration evaluated with offset to allow
for bulk effects.
Samples may be diluted into running buffer from stock solution if the dilution
factor is sufficiently high so that residual stock solution components can be
ignored. Alternatively, samples may be prepared using buffer exchange
techniques.
Figure 5-2. Analyte concentrations for kinetic analysis should cover a wide range (center
panel). If the concentrations are too low (left panel) or too high (right panel) the
sensorgrams may be tightly grouped and evaluation will be less robust. The vertical scale
for these simulated sensorgrams is the same in all three panels.
Figure 5-3. The analyte concentration range for steady state affinity determination
should cover more than twice the KD value (left). Do not trust reported affinity constants
that are higher than half the highest analyte concentration (right).
In addition to accurate sample preparation, correct values for the stock solution
concentration are important. The ideal value is the concentration of analyte
that is capable of binding to the surface-attached ligand: this may not
necessarily be the same as the total analyte concentration. If half of the analyte
molecules are in a conformation that cannot bind, the total concentration will
be twice the value that should be used for evaluation of binding kinetics. If the
sample is amenable to concentration determination by CFCA, this is the value
that should ideally be used: it represents the concentration capable of
interaction and is independent of calibration considerations. Figure 5-4
illustrates an example of the importance of using the correct analyte
concentration.
Figure 5-4. Importance of using the correct analyte concentrations. The experiment
compared the kinetics of wild type and four mutants of a papain inhibitor binding to
papain. Using concentrations obtained from A280 measurements (green bars), mutant
number 2 had an association rate constant about one order of magnitude lower than the
wild type and the other mutants. However, using concentrations obtained by CFCA
(orange bars), association rate constants obtained were similar for all variants of the
inhibitor.
6 Epitope mapping
6.1.1 Principle
The basic principle of pair-wise mapping studies in Biacore systems is
straightforward: one antibody is attached to the surface, antigen is bound to the
antibody, and a second antibody is tested for simultaneous binding ability. In
practice, the assay set-up involves additional steps (see Figure 6-1):
1 2 3 4 5
• Pairs where the first and second antibody are identical provide a check on
the assay performance. If simultaneous binding is observed this may
indicate that the antigen contains multiple copies of the epitope, which
may invalidate the mapping with respect to that epitope.
• Pairs where the roles of first and second antibody are reversed provide a
control for unexpected effects such as conformational changes induced
by antibody binding. Ideally, the results should be independent of the order
in which antibodies are tested, but this is not always the case in practice.
6.1.2 Evaluation
The label-free detection in Biacore systems offers advantages over label-based
methods for epitope mapping, in that all steps in the binding sequence are
monitored with the same technique. It is easy to relate the observed level of
antigen binding to the level of first antibody, and the level of second antibody to
that of antigen. This addresses a potential issue in interpreting the results, in
that the absence of second antibody binding may result either from interference
between the epitopes or from poor binding of the antigen to the first antibody.
This cannot be resolved in methods where antigen binding is not monitored
directly.
Report points for each evaluation step are set just after the end of the respective
injections.
Antigen ranking
Even with sufficient first antibody captured, binding of antigen may vary, and
results for binding of the second antibody will be inconclusive if the level of
antigen bound is too low. As for first antibody binding, a cut-off boundary can
be set to exclude points with response values below an acceptable limit. The
binding data may be shown as relative response values or as percent of
expected binding (based on the amount of first antibody captured and the
relative sizes of antibody and antigen).
Dissociation fit
An additional component in the evaluation is estimation of the dissociation rate
of antibodies classed as positive binders in the second antibody ranking step.
This information is not essential for pair-wise mapping itself, although it can be
useful in judging the status of borderline results. However, it is often useful in the
context of antibody development for immunoassay reagents, and is obtained
without additional experimental design or set-up from determinations with
Biacore systems.
Result matrix
The results of pair-wise epitope mapping are collected in a matrix of binding
values, with first and second antibodies in columns and rows respectively. Cells
in the matrix are classed as positive (meaning independent epitopes) or
negative (interfering epitopes) on the basis of the cut-off boundary set for the
second antibody ranking step. If the cut-off boundaries have been judiciously
set in the first antibody and antigen ranking steps, the classification in the
matrix reflects only those pairs where second antibody binding is potentially
reliable.
Figure 6-2. Principle of epitope mapping by peptide inhibition principle. In the absence of
peptide (left), antibody can bind to the epitope. Added peptide that mimics the epitope
binds to the antibody (right) and blocks binding to the antigen.
An assay of this kind can be set up with either the antigen or antibody attached
to the surface. Antigen may be attached permanently and regenerated:
antibodies are captured to allow investigation of multiple antibodies using the
same sensor surface.
7 Troubleshooting assays
• Ligand attachment
• Analyte binding
Poor response increase Too high ionic strength in Use low ionic strength (10 mM
during ligand injection immobilization buffer monovalent cations).
Ligands kept in stock solution
with physiological ionic
strength should be diluted at
least 10-fold or otherwise
desalted.
Poor response increase Ligand concentration too low Test higher concentrations
during ligand injection
Ligand binds during Ligand lacks accessible target Try a different chemistry or
injection but is not groups for immobilization chemistry use capturing
retained
Ligand binds during Buffer components such as Tris or Make sure the buffer is
injection but is not additives such as sodium azide appropriate for the
retained compete with the ligand for immobilization chemistry.
immobilization
Immobilization level low Old EDC/NHS solutions Use fresh EDC and NHS
solutions.
One or more injections do Solutions wrongly placed in Check the Rack Positioning
not show the autosampler information in the software
characteristic response for the immobilization
procedure.
the reference surface and the non-specific component of binding to the active
surface are equivalent (high ligand levels can mask dextran binding sites and
reduce non-specific binding to the surface matrix). Non-specific binding of this
kind can often be reduced by addition of soluble dextran to the samples, to
compete for binding to the dextran on the sensor surface. Soluble dextran is
available for this purpose from GE Healthcare as NSB Reducer.
Sometimes the analyte binds promiscuously to the ligand (for example, many
low molecular weight compounds bind with low affinity to multiple sites on
serum albumin, giving behavior that resembles non-specific binding). In such
cases, it may be possible to obtain useful information about high-affinity
binding sites from sensorgrams obtained at relatively low sample
concentrations, where low affinity binding does not dominate the observed
response. Adjustment of buffer conditions for the assay can also help to reduce
the effect of promiscuous binding.
• The baseline should be stable within the analysis cycle. Minor baseline
shifts between cycles may be acceptable.
• The response should not go below the baseline level during dissociation.
response
2,3
4
1 5
time
Figure 7-1. Generalized features of the “expected” shape of reference-subtracted
sensorgrams.
1: stable baseline.
2: positive response during sample injection.
3: increasing or stable response during injection.
4: decreasing response during dissociation
5: small reference subtraction spikes at the start and end of injection
6: response does not go below baseline
binding capacity is not affected (as shown by reproducible binding levels for
control samples).
response
time
Figure 7-2. Sample response below baseline is usually be caused by a negative bulk
refractive index shift.
Responses may fall below baseline during sample injection if the refractive
index of the sample is lower than that of running buffer, so that a negative bulk
refractive index shift is associated with the sample injection. In such cases, there
should be a corresponding positive shift at end of the injection: response levels
measured after the end of the injection will still be reliable.
Higher relative response in the reference flow cell than the active flow cell,
resulting from sample binding to the reference cell can also cause a shift below
baseline in the reference-subtracted sensorgram.
Both of these situations can usually be identified by examining the active and
reference sensorgrams separately.
Figure 7-3. Reference subtraction spikes at the beginning and end of a blank (buffer)
injection.
Some overshoot at the start and end of the sample injection is normal in
reference-subtracted sensorgrams from systems with serially arranged flow
cells, particularly at low flow rates. This results from slight misalignment of the
reference and active sensorgrams together with a significant bulk refractive
index effect (Figure 7-3). This should not affect more than about 1 second of the
sensorgram. More serious overshoot may indicate malfunction of the sample
injection system, and requires attention from your GE Healthcare service
representative.
Figure 7-4. Disturbances that occur at the same time in each cycle are attributable to
automatic aspects of instrument operation such as pump flow-fill.
d AB -
-------------- = k a A B – k d AB
dt
d AB -
-------------- = – k d AB
dt
dR
------ = k a CR max – k a C + k d R
dt
Because interaction occurs at the surface of the sensor chip, analyte must be
transferred laterally in the flow cell from the bulk solution to the surface before
interaction with ligand can take place. In a controlled flow system such as
Biacore, this transfer is a diffusion-limited process, referred to as mass transfer,
that can be described by a well-defined mathematical model. The analyte
The mass transfer coefficient km is a function of the flow rate, flow cell
dimensions and diffusion properties of the analyte molecule:
D 2 3 f 13
k m = 0.98 --- ----------------------
h 0.3 w l
f is the volume flow rate of solution through the flow cell (m3s-1)
The mass transfer coefficient can be adjusted approximately for the molecular
weight of the analyte and for the conversion of surface concentration to RU to
give a term called the mass transfer constant kt:
9
k t = k m MW 10
kt
t c = ------
3 f
For globular proteins with molecular weight of the order of 50,000 daltons,
typical values for tc are of the order of 108 RU·M-1s-1. Reported values that differ
greatly in order of magnitude (e.g. 1012 or 1014) may indicate that the parameter
is not significant for the fitting (i.e. that the observed binding is not limited by
mass transfer).
dR
------ = k a CR max – k a C + k d R = 0
dt
so
k a CR max = k a C + k d R
ka ka
R eq -----C + 1 = ----- CR max
kd kd
K A CR max
R eq = -----------------------
KA C + 1
Kinetics
The rate constants reported by kinetic evaluation are determined in terms of a
1:1 interaction model, and it is important to realize that they are only valid in that
context. If the interaction mechanism is not a simple 1:1 binding, the fitted
curves will deviate to some extent from the experimental data and the reported
constants will not be a true representation of the interaction kinetics. The
apparent 1:1 binding constants can still be used for comparative studies of
observed binding rates, but it is important in reporting the values to emphasize
that they are empirical and not mechanistic constants.
There are two major tools for assessing the significance of the reported
constants: the closeness of fit between the fitted and experimental curves and
the statistical significance of the parameters.
Closeness of fit
Visual inspection of the residual plot or of the fitted curves overlaid on the
experimental data gives an indication of the closeness of fit. Ideally, the
residuals will scatter randomly around zero over a range that corresponds to
the short-term noise in the detection system (approximately 1 to 2 RU).
Deviations from ideal fitting appear as systematic variations in the residuals,
imparting a non-linear shape to the residual plot. Judge the residual range and
shape in proportion to the response ranges in the experimental sensorgrams
and in relation the goal of the investigation.
Figure A-1. Residual plots from a good fit (top) and a poor fit (bottom). Note the different
response scales in the two plots (top ±2 RU, bottom ±60 RU).
Significance of parameters
Fitting experimental data to a mathematical interaction model will return values
for all parameters in the model, regardless of whether they are significant or
not. As an example, sensorgrams that are completely limited by mass transfer
will give values for the kinetic rate constants even though the experimental data
does not contain any kinetic information.
to changes in the parameter value: in other words, the value returned for the
parameter has little significance.
For the 1:1 binding model, an additional indicator of the parameter significance
is the U-value. This is a parameter that represents the uniqueness of the
calculated rate constants and Rmax, determined by testing the dependence of
fitting on correlated variations between selected variables. Lower values
indicate greater confidence in the results. A high value (above about 25)
indicates that the reported kinetic constants contain no useful information.
Affinity
To obtain a robust fit to a plot of Req against C for affinity determination, it is
important that the range of analyte concentration is wide enough to reveal the
curvature of the plot in full. In particular, if the highest concentration is too low
in relation to the calculated KD value, the reported values will be uncertain even
if the fit is apparently good and the chi-square value is low.
As a general guideline, the fit will be reliable if the highest analyte concentration
is at least twice the calculated KD value. If the highest analyte concentration is
less than 2KD, you should treat the reported KD with caution.
This model does not include a mass transfer coefficient. If mass transfer
limitations are significant, dissociated analyte will be washed away from the
surface with reduced efficiency and the dissociation rate constant will be
underestimated.
where ka1 and kd1 are the association and dissociation rate constants for the
first site
and ka2 and kd2 are the association and dissociation rate constants for the
second site
Once binding has occurred at the first analyte site, binding at the second site is
facilitated by the proximity of analyte and ligand. Similarly, analyte molecules
that are attached at both sites are not released from the surface until
dissociation occurs at both sites, so the observed dissociation rate is much
slower than that seen for a single site. The resulting enhanced binding is often
called avidity to distinguish it from single site binding affinity.
Note: The association rate constant for interaction at the second site is reported
in units of RU-1s-1. This is because both the interacting components are
present on the surface and are measured in RU, not in true
concentrations.
where ka1 and kd1 are the association and dissociation rate constants for one
ligand
and ka2 and kd2 are the association and dissociation rate constants for the
other ligand
K A CR max
R eq = ----------------------- + offset
KA C + 1
where offset is a constant term that gives the intercept of the fitted curve on
the y-axis. Without this term the fitted curve is forced to go through the
origin of the plot.
The parameters KA , Rmax and offset are fitted as variables. The affinity is
reported as a KD value, which is the inverse of KA . This function gives reliable
fitting if the highest analyte concentration used is at least twice the KD value.
y = slope * x + intercept
R hi – R lo
y = R hi – -------------------------
-
x A2
1 + ------
A 1
Rhi and Rlo are fitting parameters that correspond to the maximum and
minimum response levels respectively
B.1 Introduction
Solvent correction refers to the procedures used to adjust measured responses
for the effects of varying concentrations of organic solvents. These procedures
are relevant primarily for work with small organic analytes that require inclusion
of organic solvents (commonly dimethyl sulfoxide, DMSO) in the buffer to
maintain analyte solubility.
Figure B-1. Bulk solution is excluded from the volume occupied by ligand molecules on the
active surface, so the contribution of bulk solution to the relative response is smaller than
on the reference surface.
In drug discovery and development work, the analytes are often small
molecules which give correspondingly low response values (typically of the
order of 10-50 RU or less). High levels of immobilized ligands (several thousand
RU) are used to maximize the analyte response, enhancing the excluded volume
effect described above. Samples generally include DMSO to maintain solubility,
resulting in high bulk responses. A difference of 1% (percentage points) in DMSO
concentration corresponds to a difference of about 1200 RU in response, so that
small variations in DMSO concentration, unavoidable in the preparation of
diverse samples, easily lead to variations in bulk response of the same order of
magnitude as the expected sample responses.
Figure B-2. The principle of solvent correction. 1. The sensorgram from the reference
surface shows a bulk displacement (-150 RU in the illustration) during sample injection
because the sample and running buffer are not exactly matched. 2. From the solvent
correction curve, a displacement of –150 RU in the reference sensorgram corresponds to
a solvent effect of +5 RU in the reference-subtracted sensorgram. 3. The reference-
subtracted sensorgram is corrected for the solvent error. This procedure is applied to
every point during sample injection.
It is however important that the solvent correction curves fit the experimental
points closely. Poor fitting of the curves to the experimental points (Figure B-3,
center panel) can introduce uncertainties in the solvent correction factors that
are the same order of magnitude as the factors themselves, potentially
introducing additional errors in the sample measurements.
Sample responses that lie outside the range of the solvent correction curves
(Figure B-3, right panel) cannot be corrected and will be omitted from evaluation
of solvent corrected data. Solvent correction curves may be extrapolated in the
software, but this option should only be used for small extensions of the solvent
correction range (less than 10% of the measured range) since the curve shape
is not predictable.
T
target molecule 7, 32
tc 58
terminology 7
thiol coupling 20
Tris buffer 23, 50
GE Healthcare Bio-Sciences KK
Sanken Bldg. 3-25-1, Hyakunincho, Shinjuku-ku, Tokyo 169-0073, Japan
imagination at work
29-0194-00 AA 05/2012