Astell Duaclave Douboe Autoclave Operation Manua
Astell Duaclave Douboe Autoclave Operation Manua
Astell Duaclave Douboe Autoclave Operation Manua
SWIFTLOCK
DUACLAVE
SECURE-TOUCH+
TOUCHSCREEN
CONTROL
AUTOFILL
STERILIZER
FRONT-LOADING
DOUBLE
AUTOCLAVES
120 & 153 Litre x2
Instruction Manual
INSTALLATION
OPERATION
MAINTENANCE
FAQ
Optional Extras
Integral Data Printer
IQ/OQ Documentation
Direct Steam (heating option)
Load Sensed Process Timing RS232 Interface
Ethernet Interface
Technical details
Model No. Volume Internal Chamber Overall Dims Power Supply Heaters in Chamber
(Litres) dims (dia x depth) wxhxd
ASB260BTDV 120 450 x 635 (x2) 685 x 1855 x 1310 3 Phase 380-440 v15Amps x2 off (each
3 phase 415V (x2) chamber has a separate supply )
ASB260BTDV 120 450 x 635 (x2) 685 x 1855 x 1310 1 Phase 27Amps x2 off (each chamber
Single phase 230V (x2) has a separate supply )
USB260BTDV 120 450 x 635 (x2) 685 x 1855 x 1310 3 Phase 208v 27Amps x2 off (each
3 Phase 208V (x2) chamber has a separate supply )
ASB270BTDV 153 450 x 865 (x2) 685 x 1855 x 1310 3 Phase 380-440 v15Amps x2 off (each
Single phase 230V (x2) chamber has a separate supply )
ASB270BTDV 153 450 x 865 (x2) 685 x 1855 x 1310 1 Phase 27Amps x2 off (each chamber
Single phase 230V (x2) has a separate supply )
USB270BTDV 153 450 x 865 (x2) 685 x 1855 x 1310 3 Phase 208v 27Amps x2 off (each
3 Phase 208V (x2) chamber has a separate supply )
mxn746-d01ed a Swiftlock Duaclave #6&7 autofill & Dir Steam.doc SECURE-TOUCH + SWIFTLOCK 5
WARNING IEC61010
Although the safety record of laboratory Sterilizers is good, remember at all times that they store
considerable potential energy, and should be treated with respect and care . If correctly used and
cared for, your Sterilizer will give you long and safe service .
Pay proper attention to regular maintenance . Never force the locking mechanism, or operate the
machine with any leaks, or incorrectly operating parts.
Report any defects to your Supervisor . If deterioration or defects are noticed, record them in a log
book and contact our Service Department .
Record also the results of annual and periodic inspections . Every 4 to 6 weeks is recommended.
Check the logbook before you start using the machine, as someone else may have recorded a fault of
which you are not aware. ]
The only way to cancel this warning is to run a Safety TEST CYCLE IS NOW RUNNING
valve Test Cycle.
When steam is seen coming from the
safety-valve outlet, touch [ ] to
IMPORTANT * terminate the test and advance
THE MACHINE MUST BE machine into cooling.
ATTENDED AT ALL TIMES LEAVE UNTIL COMPLETE
UNTIL TEST IS COMPLETED. & THEN OPEN DOOR
CLIMATIC SPECIFICATIONS
OPERATING
Permitted Humidity not exceeding 75%
mean non-condensing.
Permitted Temperature range 0-40 C
Do not site the equipment in an area where
there are significant quantities of dust. this
will gather on the electronics and cause
malfunction.
Do not site the equipment in an area where
Spray or fumes from other equipment may
touch the cabinet or cooling vents.
STORAGE
Permitted Temperature range -10 / +70 C
Humidity not specified but an equalising time
is required- see above.
mxn746-d01ed a Swiftlock Duaclave #6&7 autofill & Dir Steam.doc 9
This equipment is heavy, not portable and is not Do Not position either side up against
provided with lifting handles. a wall.
If in doubt, Remove sidepanels before lifting Lift Do Not site other equipment close to
at sides or rear of case, taking the load of the unit either side or the rear.
on the internal structure. Do not lift or handle
using the Plastic Door cover or Control Panel or Do not cover or obstruct Cooling
any other projection. Do not drag across surfaces vents.
or the Feet may be damaged.. Do not adjust Feet
on frontload models- these are fixed to tilt the Do not place objects on top of the
machine backwards equipment.
.
POSITIONING & LOCATION
The equipment should be placed on a suitable DISCLAIMER
surface and levelled before use. This must
allow the door or cover to open unobstructed,
and will permit the chamber to be operated The manufacturers or their agents cannot
and loaded by the operator whilst complying accept responsibility for damage that may be
with local regulations regarding lifting and caused by exhaust gases water or
manual handling - Please see ”HANDLING” condensate should the machine be operated
instructions. other than in accordance with the
instructions in the provided operating
The equipment should be positioned so that manuals, or if the Pipework system is
the Power switch is unobstructed and easily modified, misused, blocked or obstructed, or
accessible from the operating position ( in by use of non-approved alternative spare
front of the equipment ). parts .
Do not cover or obstruct Cooling vents. Do
not place objects on top of the equipment.
INSTALLATION
POWER, WATER & DRAINAGE A range of steam quench and cooling heat
exchange systems is available factory-fitted to
the the machine, which automatically cool the
WATER , STEAM ,DRAINAGE
SERVICES & CONNECTIONS. PULSE FREESTEAMING OPTION
!!! WARNING !!!
NOTE- This machine comprises TWO separate
sterilizers however for convenience the water If this option is fitted or configured
inlet, exhaust/drain and driptray drain are Installation to Plastic Drainage systems,
shared. or the Astell range of Quench /Cooling
boxes is not possible without danger to
1] The Chamber Exhaust must be connected to a premises or personnel. Operation in such
suitable Building Drain to carry away condensate. an installation constitutes misuse and the
manufacturers accept no liability for the
2] A Steam Vent is required to exhaust the steam effects of such action, and reserve the
produced during the Heatup and Freesteaming right to withhold warranty provisions for
parts of the cycle. parts concerned with the Chamber or
steam system .
3] The Safety Valve exhausts into the room.
hot exhaust to below 60C using cold mains
water . if one of these has been supplied please
4] These Electrically Heated models require a
install according to the separate instructions
supply of cold Mains water. and where fitted
supplied. At the rear of this manual
provision for the Autofill tank Overflow / Warning
Pipe.
Fitting to Plastic Drainage Systems; This should be arranged to allow the chamber to
be drained regularly. It may be plumbed into a
mxn746-d01ed a Swiftlock Duaclave #6&7 autofill & Dir Steam.doc 11
drain or tundish. As long as the temperature
requirements are met. Either-
A the small diameter COPPER pipe at the
KEEP DRAIN VALVE CLOSED DURING USE & bottom rear of the machine. It should be
DO NOT OPEN WHEN HOT OR connected to a suitable building drain capable of
PRESSURIZED. carrying steam condensate ( at up to 134C )
Connection into the boiler-house condensate
F] STEAM INLET.. return is not advisable due to the possible
Steam Heated Models Only bacteriological contamination of returned
This is a 3/8" BSP Screwed Gun-Metal Union. condensate if any of the load contaminates the
The steam supply should be connected to the inside of the machine. This outlet should be
gunmetal union. connected via an air-break to prevent back-
suction.
Pressure Reducing Valve
The autoclave is NOT fitted internally with a PRV, Or B Ilnked into the exhaust ouitlet- no
and steam supply pressure must be reduced to special arrangements are required.
3.04 Bar maximum.[ 44 P.S.I.G /304 kPa) with a
suitable Pressure Reducing Valve before
reaching the autoclave. J] DRIP-TRAY DRAIN
It is important to fit a pressure gauge into this This is a separate plastic pipe 13mm internal
supply line, after the pressure reduction valve, bore, emerging from the rear of the machine.
which will indicate actual steam pressure This carries away water that trickles into the
available to the machine . driptray beneath the door. It should be led to a
suitable drain (Below the level of the Driptray.)
A condensate draining steam trap must be Do not connect to the Autoclave drain, without a
fitted between the PRV & Autoclave, trap or suitable U-Bend or similar, as this may
immediately upstream of the Autoclave, to cause hazardous blow-backs if pressure
remove all condensate as Free Water will stop develops in the drain . As this water is generally
Sterilizing action. fairly clean, a simple loose connection to a
separate drain will suffice or a bucket or
Steam Supply Problems;- container may beadequate if drainage cannot be
A common fault with installations is that long pipe arranged..
runs, poor layout or poor steam quality fill the
autoclave Chamber up with water. This will be K] MAINS WATER INLET
avoided if attention is paid to Condensate AUTOFILL option
separation. Ensure that the steam supply flow is WATER-COOLING option
adequate to feed the machine when operating - EXHAUST QUENCH SYSTEM option
The steam pressure in the line will fall when the
machine is HEATING UP or FREESTEAMING. This is a Plastic washing-machine type Hose
with a ¾ BSP ( ¾ RP) Female Threaded tap
F] SUPPLY CONDENSATE RETURN connector as for a Washing Machine Inlet.
Steam Heated Models Only
The steam Supply line must be properly installed. This should be connected to a Cold Water
If provision is not made to automatically drain off Mains Supply ,
the condensate that forms in the supply line the Minimum 0.3 Bar , Maximum 6 Bar
autoclave will draw in large quantities of water. & pressure. Max Flow rate required 20
this will stop the Sterilising action, and may litres/minute.
cause damage. This supply Must be provided with a Shutoff
Valve ( eg a Washing-Machine-Isolator Valve ).
Warning ; On new steam systems or those that
have cooled beware condensate build-up in lines L] AUTOFILL TANK OVERFLOW (warning pipe)
which may cause water hammer & damage AUTOFILL option
valves and instruments. Correct installation and This is a White Polypropylene 20mm pipe or on
Steam Trap arrangements will avoid this. some models a 22 mm copper pipe which
projects from the rear of the machine .
H] STEAM DRAIN OUTLET This will carry overflow water if the Level in the
Steam Heated Models Only Autofill tank is too high. It should be fed to an
( Chamber Condensate ) visible open Tundish or drain.;- U.K. Water
On Steam Heated Models the Chamber is fitted Research Council and the U.K. water byelaws
with an Automatic Chamber Drain. The outlet require Warning pipe outlets to be visible to the
from this is arranged in to ways operator.
mxn746-d01ed a Swiftlock Duaclave #6&7 autofill & Dir Steam.doc 12
The autoclave shown is an example only & does not represent any particular model.
All pipework positions are typical only and may differ from model to model.
All connections are labelled for identification.
EXHAUST VENT
This MUST be routed to the outside of the DRIPTRAY DRAIN
building or to a purpose-designed extraction This carries the driptray waste which may
system to remove the steam outflow. It Must collect from runoff from the door inside surface
be free-venting & installed with a slope when the door is opened. It should be routed
minimum 1:50 so that it is self-draining along to a building drain. Where this is the same
the whole length to prevent blockage by water. drain as the exhaust a u-bend or trap must be
fitted in line with this pipe to prevent steam
Where an outside wall is not available , It is pressure in the drain causing backflow up to
acceptable to install an extraction hood and the driptray.
fan over the autoclave and to lead the Vent
pipe up into this hood.. This extraction system
must be rated to accept steam and extract at a
rate which prevents steam penetration back
into the room
mxn746-d01ed a Swiftlock Duaclave #6&7 autofill & Dir Steam.doc 13
The recorder options include an optional The type of heater used in this machine means
that Operation from a RCCD residual or earth-
automatically controlled 240 Volts 1 Amp power
outlet via a CEE22 Outlet connector. This is leakage trip may cause tripping problems ., This
intended to power ONLY an Astell approved may occur especially if the machine is not operated
Recorder System. Instructions for use are for some time and is due to moisture ingress to the
provided with the Recorder option. Some Inconel-Sheathed heaters. If this happens connect
to a non- protected supply and run for 5-10 cycles.
versions may use a different connector and the
Then revert to Protected supply. If the tripping then
output may also include remote control low-
continues please consult the manufacturer.
voltage signals.
NOTE
The rating plate on the rear of the
cabinet gives the electrical
specifications for this machine.
Please refer to the data on this plate
before connecting your sterilizer.
mxn746-d01ed a Swiftlock Duaclave #6&7 autofill & Dir Steam.doc 16
200/240 V (208V) 3-PHASE SUPPLY However- please note that this requires
the fitting of a special transformer unit and
significant detailed wiring changes, and
Power Cable is connected as shown in drawing requires the services of a factory-trained
367- 500 at the rear of this manual qualified engineer.
C Starting
When you press START the system will heat
immediately - there is no “ Charging With
Water” stage
E Chamber DRAIN
On these models this is NOT a Drain for
cleaning the chamber, but it is a Condensate
Drain. This discharges the ccondensed
steam to the drainage system.
F Fuses
FS10,11,12 are not fitted in the power input
box. FS05 is the mains input protection fuse.
G Overtemperature
As standard There is no Overtemperature
cutout on Directly Heated machines.
Generally these cannot heat above the
temperature of the incoming steam so that this
and Boil-dry protection, is not required.
mxn746-d01ed a Swiftlock Duaclave #6&7 autofill & Dir Steam.doc 19
INITIAL POWER UP
1. Check that display Power is On 12 Open door fully & unload chamber.
2. If display shows “Tank water level
low”, add water to Tank 13 If [DOOR] button is pressed and
3. Press “Door” Button. Door is not opened it will cancel
after 10 seconds, and will re-lock.
The bolt will unlock . After this , to OPEN , just press
4. Turn Handle, & open door. [DOOR] button again .
5. Check Gasket is correctly seated
and inspect for damage. Notes-
6 LOAD STERILIZER. If Door is closed by accident, press
[Door] key To Open .
7. CLOSE Cover and rotate handle to If handle is rotated to LOCKED position
lock. with door open , press [DOOR]
button and return handle to
8 Select Cycle “unlocked” position
press [DOOR]
11 Door unlocks
rotate handle & Open door.
1. Check that display Power is On 12 Open door fully & unload chamber.
2. If display shows “Tank water level
low”, add water to Tank 13 If [DOOR] button is pressed and Door
3 Press “Door” Button. is not opened it will cancel after 10
seconds, and will re-lock. After this ,
The bolt will unlock . to OPEN , just press [DOOR] button
4. Turn Handle, & open door. again .
5. Check Gasket is correctly seated and
inspect for damage.
6. LOAD STERILIZER. Notes-
If the Cycle requires use of LOAD SENSING
Place Load Sensing Probe in position in If Door is closed by accident, press
centre of load [Door] key To Open .
If cycle does NOT require use of the Load If handle is rotated to LOCKED
Sensing probe, place the probe safely to one position with door open , press
side of the load . [DOOR] button and return handle to
“unlocked” position.
7. CLOSE Cover & rotate handle to lock.
8 Select Cycle
press [DOOR]
11 Door unlocks
rotate handle & Open door.
OPERATING INSTRUCTIONS
DOOR CONTROLS
OPEN
Touch “Closed Door”
button on the display screen to
Unlock the door.
CLOSE
CLOSE AND lock the door
This is sited inside the cabinet & cuts out the FAULTS
heater in the event of excessive temperature rise Should there be an instrumentation problem a
( typically >150C). It is only resettable by manual monitor system detects out-of-range values and
operation when the covers have been removed. reports them via the FAULT system
mxn746-d01ed a Swiftlock Duaclave #6&7 autofill & Dir Steam.doc 26
IF an ALARM is ACTIVE
the ALARM button FLASHES RED
Login with your Username and Password
Demineralised Distilled water This means that the internal water tank is low on
Tap-water Softened water water . The water inlet supply is turned off, or not
connected. Please see installation pages for details
WATER QUALITY of water supply inlet.
Untreated tap water of drinking quality is
satisfactory in this Sterilizer although the In normal operation the Tank is maintained fuill of
use of treated or partially demineralised water automatically and this message is not shown.
water is acceptable, particularly in
extreme hard water areas. This will help WATER SYSTEM – DIRECT STEAM
reduce furring on the elements the water HEATED MODELS
must be conductive , requiring ionic
content. The machine takes a supply of steam from the
external supply. No water is required for the
DEIONISED OR DISTILLED- steam generation.
water is not acceptable
******HOWEVER models with WATER COOLING
require a cald water supply
The water used must be conductive, ie have an
ionic content. DEIONISED OR DISTILLED- water
will not work in this sterilizer unless a suitable ionic
salt is added. Please consult your dealer for advice
if you wish to use treated water or full fault see later section
HANDLE LOCKED WITH DOOR OPEN
SOFT WATER
Water that has been chemically softened is usually
satisfactory but rain-water or water that is from a If the door is open and you move the handle
naturally soft source may not work. accidentally to the LOCKED position it will lock in
that position.
Make sure that the load is correctly made ( note temperatures and times may vary from those
examples shown here )
up and in the correct containers to suit the
program cycle to be used.
B If cycle does NOT require use of the Load This will take you to a screen where you can
Sensing probe, place the probe safely to one select any one of the available programs.
side of the load .
WARNING!
The NEXT press
on the Button
STARTING A CYCLE
The Chamber Closure Door /Cover must be
closed and locked
CYCLE COMPLETE -- PASSED Clear the Alarm and then return from the alarm
page. ( see ALARMS section )
ENTERING PASSWORDS
Press the Button
Press Here
Notes-
Select
Use buttons
Then press the
to select stage;-
eg COOLING button
SAFETY INTERLOCKS
PRESSURE INTERLOCK
Preset to [ less than +0.15 Bar ] this is an Typical
independent electrical interlock operated by a View of
precision pressure switch preventing the electrical Cooling
release of the Closure lock if the chamber is Lock
pressurised. Thermostat
OVERPRESSURE CUTOUT
This is preset to approx 3.00 Bar. Above the
safety valve point, & operated by the
electronics control software. If the pressure
sensed by the controller pressure transducer NOTE;-The temperatures will not be quite the
exceeds this level , which suggests safety same, due to differing thermal inertia.between the
valve problems, then the heating is shut down, a Chamber Wall and the Load. Normally the
warning shown and a fault condition generated. Chamber wall & Cooling Lock sensor attatched to it
will cool down faster than the load.
Should the Fault system indicate a Pressure IF there is a water level problem or a valve sticks
switch fault This usually indicates a Pressure open, blowing out the water, & the system goes to
bad adjustment or Drift , and is seldom a Boil Dry then the heater will be exposed and the
Pressure switch Failure. Overtemperature Thermostat will start to operate.
This will take over control and shut down the heater
The Pressure switch should release at approx + until the temp. falls when it will come back on.. This
0.15 Bar Gauge. Due to age or stress this setting will cause the temperature to vary very widely up
may drift. and down as the heater cuts in & out and the
Pressure will not match the Temperature (- eg the
The system monitors this switch as a safety feature Chamber might be above 100C but the pressure
to prevent opening the chamber if the pressure might be zero or much lower than normal )
exceeds 0.15Bar. Service attention will be required.
B the temperature is near to room temperature Please note that this testing is mandatory in the
And E.U.. and many other countries.
the display still shows the fault as above Selecting this will force system to use
a preset set of parameters specifically
, then ;-
the Pressure switch may be out of adjustment. for testing the safety-valve. This runs at
excess temperature and pressure.
It may resolve itself if left for 1-2 hrs to cool down .
As standard, all models have an AUTOMATIC
If this continues, the Pressure switch should be SAFETY VALVE TEST WARNING SYSTEM .
checked and adjusted - please contact your service Performing a safety valve Test Cycle and
agent Or Astell Scientific for advice. getting a PASSED result resets the “ SAFETY
VALVE TEST OVERDUE” warning message, for
the next test period.
Typical view of Pressure Switch
Do Not Use This Feature for
Sterilization- Safety-valve test ONLY.
Please see the SAFETY VALVE TEST
instructions at the beginning of this manual.
mxn746-d01ed a Swiftlock Duaclave #6&7 autofill & Dir Steam.doc 43
Temperature
Hold Time
Maximum
Minimum
Minimum
Sterilize
Sterilize
Sterilize
** AIRPURGE TIMES
Airpurge times longer than 15 minutes on an
unloaded or lightly loaded chamber may cause
problems with boiling Dry . Long times are usually
only necessary with full loads. 134 138 3 min
126 129 10 min
“Open” load containers without drain holes which
121 124 15 min
can collect and trap condensate from draining back
down to the chamber base – can retain water 115 118 30 min
causing boiling dry .
Lower temperatures or shorter times may have to be
Long Airpurge times can mean excessive heating in used to prevent degradation of bacterial growth
the condensate bottle and potential hazards from media. This may be adequate for culture purposes ,
steam and condensation. but does may not constitute full sterilization. The
manufacturer of the media will usually specify the
Please ensure that the load and containers sterilizing temperature.
match the selected cycle.
mxn746-d01ed a Swiftlock Duaclave #6&7 autofill & Dir Steam.doc 46
Stabilise Time
Set to between 1 and 5 minutes
COOLING
The COOL parameter Must be set to COOL1
COOLING LOCK(S)
Eg
To change
the DAY
OPTIONAL PRINTERS
The printer is mounted on the Front face of the Grasp at both ends and pull firmly away from
machine. The standard printer supplied is the the printer assy.
Astell “Clamshell-Load” Type Shown above .
This is a 40 character easy-load dot-matrix To fit the new ribbon
impact printer which prints on 58mm wide plain
roll paper. It uses a replaceable ink ribbon Hold the new Ribbon Cassette by the ends and
cartridge and prints in one single colour ( Black) rotate the small wheel using the tip of a Biro Pen
in the direction shown by the arrow on the
cassette. This tensions the ribbon.
CLEANING
This Sterilizer is designed to Heat clean Water in Do NOT use Abrasives or Chlorine-based cleaners
the Chamber. The filling water may become or bleaches on the cabinet or chamber
contaminated by unsterile condensate or drainage
from the Chamber.If the filling water becomes Cleaning of LABELS & WORKING PARTS Clean
contaminated with AGAR, BLOOD, FAECES , or Labelling, aluminium parts or working parts with a
solid rubbish it will cause fumes, smells, potential soft cloth moistened with a little methylated spirits.
health problems and accelerated machine failure. Do not use Solvents IPA, or Aqueous Cleaners
which may remove the printing or melt the
The Chamber and all wetted parts should be plastics.
Cleaned regularly in order to maintain
adequate operating conditions. Load Cleaning of LCD TOUCHSCREEN ;- Clean with a
Containers should be chosen to minimise loss soft cloth moistened with a little water. For best
of the contents into the Sterilizer, e.g.:- do not results you can use 2 Spectacle Cleaning Wipes”
use plastic bags inside Sterilizer Baskets but suitable for Plastic lenses.
use a purpose-made Sterilizer container
designed to catch leaking waste. Please ask If stubborn marks persist use a soft cloth
your supplier for details of containers for your moistened with a very small amount of methylated
machine. spirits. Do not use IPA or similar solvents or
Aqueous cleaners.as these will damage the
The Chamber should be cleaned internally to transparent window.
prevent build-up of contaminants , and we
would suggest that for a machine in daily use,
if used for Fluid Loads or WASTE DESTRUCT
the Chamber should be checked once a week
and cleaned if necessary, . For other
applications Cleaning should be at monthly
intervals or more frequently.
Open the DOOR as you would normally.
Remove all trays or shelves.
Clean the base of the chamber with Hot Water. Do
not use an abrasive cleaner or one which may
leave soluble residues that would remain in the
pipework and contaminate the water.
Locate solenoid-operated
Locking bolt.
PERIODIC MAINTENANCE
Safety Valve Test Quick Reference
There are very few moving or wearing parts in this
sterilizer , HOWEVER , It is essential that the unit IMPORTANT *
is kept clean , especially the water in the tank or THE MACHINE MUST BE ATTENDED AT
Chamber and the Gasket. ALL TIMES UNTIL TEST IS COMPLETED.
A GASKET
Keep the mating surface of the Chamber flange
clean. This bears on the gasket to seal the TOUCH
Chamber If necessary apply a little Silicone
Grease. SELECT “SETTINGS”
Check that gasket is not bulging out of the groove TOUCH
and, and that the edge is smooth without any cuts
or abrasions.
SELECT “SITE ENGINEER”
IMPORTANT- Check that the gasket is fitted ENTER “333333”
correctly - the small “Holes” should be on the
outside lip. SELECT “EDIT MACHINE SETTING”
FUSES
FUSE REPLACEMENT
Consult manufacturer before replacing fuses
This section describes how to derive the settings In most cases the behaviour of a load in this
that you may need to enter for your intended arrangement is unknown, so it is necessary to use a
Cycle(s) . The description applies to Media Loads or chart recorder or multi-thermocouple indicator to
Discard Loads but a similar approach can be applied monitor the load, and find the ideal AIRPURGE and
to Instrument/Glassware. Sterilize times.
9] PROGRAMMABLE COOLING LOCK. To ensure that sterilizing conditions are met, the
You should have Set the PRESET COOL LOCK temperatures should all be amended as follows to
THERMOSTAT dial to Minimum . allow for calibration tolerances before setting into the
profile :-
The load will cool down slower than the Chamber
Temperature Display; Monitor the load temperatures AIRPURGE TIME - Set to the time measured, from
on the Recorder/Indicator and note down the the start of airpurge Period to all points in the load
Displayed CHAMBER TEMPERATURE at the reaching 95C. ( at this moment the display was
moment when the HOTTEST part of the load has Recorded as “T1”).
cooled to a temperature safe to handle. Call this
"CYCLE COOL-LOCK TEMP" AIRPURGE TIME = ( (Original Time Set) - T1 )
minutes
10] PRESET COOL LOCK THERMOSTAT
It is important to Correctly Adjust the PRESET e.g. If “T!” was Display=2.00, then ;-
COOLING-LOCK THERMOSTAT. However, it can 8 min - 2 min = 6 minutes airpurge
only be set to One Temperature, which should be time to set in Cycle profile
that required for the worst-case Cycle from a point of
view of operator safety. STERILIZE TIME - Set to the total of the time taken
for the load to catch-up to sterilize temperature ,
Monitor the load temperature as it cools and when added to the desired Sterilize Time in minutes.
it falls below a safe temperature to handle (usually
80C) rotate the PRESET COOLING LOCK
mxn746-d01ed a Swiftlock Duaclave #6&7 autofill & Dir Steam.doc SECURE-TOUCH + SWIFTLOCK 60
The Catch-up time is the original set sterilize time , When the load reaches the Sterilizing Temperature the
less the reading taken as "T2” Sterilizing Timer will START , and the cycle will proceed. (
It is not necessary to note down any value or time .)
STERILIZE TIME = ( set Sterilize time - T2 ) +
( Desired Sterilize Time ) minutes Omit Stage 7 completely ( the cycle will proceed
under automatic control ).
e.g. If “T2” was Display=51.00, then ;- Carry out stages 8,9,10,11 as described above.
( 60 min - 51 min ) + 15 min = 24
minutes Sterilize time to set in Cycle profile Enter "Supervisor Mode" ,select "enter Cycle
parameters " and Enter the CYCLE VALUES
COOLING LOCK SETTINGS determined as follows;-
A The type of heater used in this machine means Q Difficult to close Door against gasket (2)
that Operation from a RCCD residual or earth-
leakage trip may cause tripping problems ., This A If gasket is more than 6 months old consider
may occur especially if the machine is not replacement.
operated for some time and is due to moisture
ingress to the Inconel-Sheathed heaters. If this Q Steam leaks from around Gasket (1)
happens connect to a non- protected supply
and run for 5-10 cycles. Then revert to A Gasket MUST be fitted correctly into Groove
Protected supply. If the tripping then continues and must have no damage to the edges of the
please consult the manufacturer. lips or cuts or abrasions. The opposite Sealing
face that it touches must be clean with no
Q Why cannot I operate a control and Fast- cuts or damage.
Exhaust the steam when I want to?
Q Steam leaks from around Gasket (2)
This action would however probably be dangerous
and could be in breach of health and safety A Possibly Gasket needs replacing -it must be
legislation. Firstly the amount of teanm replaced before the silicone rubber loses its
produced when a chamber pressurised to 120C sealing properties- any time after 6 months
is exhausted is very large and constitutes a of use.
serious hazard.. Unless special high
temperature exhaust pipework is installed in the Q Gasket has been changed but steam leaks
building to take this exhaust away top a safe .
discharge , then this would be unsafe. If the A Gasket MUST be fitted correct way round-
load being processed contained any fluids then there are small holes in the gasket lip
the action of exhausting the steam from 120C These must be on the OUTSIDE edge.
would cause all the fluid load to flash-boil. This
would cause the load to violently boil over & Q Cycles that are provided pre-set as
empty the load contents into the chamber, delivered are not suitable for my loads.
which quickly will "gum-up" the valves and
pipework. Also if such a load was removed A Consult the sections on selecting Times ,
aven after it had appeared to stop boiling over temperatures , etc, then in SUPERVISOR
then it would be very likely to flash-boil again level enter the new Cycle values you require.
into the face of the user carrying the load.
Q Delayed Start would be useful / is Not
Required
Q I have connected to Deionized water and
the unit does not work. A Turn ON or OFF Delayed Start - See
Programming manual
A Please connect to The correct water supply -
Not Deionized or Distilled. Q I have started the wrong Cycle
CERTIFICATION
It is a statutory requirement of the U.K. Health & Safety at Work Act and
in many other countries that Sterilizers be thoroughly inspected by a
competent person prior to use ( usually an Engineering Surveyor from
an Insurance Company ) , and at least every 14 months thereafter.
(Section 35(5) of the Factories Act 1961.)
Test certificates for Electrical safety and Calibration of the control system
are supplied as standard, Full test analysis results are available to
special request.
mxn746-d01ed a Swiftlock Duaclave #6&7 autofill & Dir Steam.doc SECURE-TOUCH + SWIFTLOCK 65
Pressure Temperature Correlation
Pressure/Temperature Calibration
This chart shows Pressure and Temperature correlation for Saturated Steam, and the
Maximum/Minimum acceptable Temperature display Values for a Correlation
Accuracy of +/- 2%. It applies to ASTELL Sterilizers only.
Notes Space
This space is for your own notes on the system