Langer 2016.08.15 A100-200-300 Manual English

Download as pdf or txt
Download as pdf or txt
You are on page 1of 12

User manual

A100 / A200 / A300


Optical fibre probe

Contents: Page

1. Usage 2
2. Function 5
3. Operation 6
4. Safety instructions 8
5. Warranty 9
6. Technical data 10
7. Scope of delivery 11

Copyright (C) Dipl.- Ing. Gunter Langer


Nöthnitzer Hang 31
01728 Bannewitz
17.01.2000

Voltage measurement and oscillographic signal tracing in devices and


systems under extreme electromagnetic conditions.
2

1. Usage
Measurement under interference conditions
Measurement of analogue electrical signals under extreme electromagnetic stress such as:
 Coupling of radiated or conducted RF emissions
 Fast transients burst / ESD
 High potential (high voltage)

EMC tests as the main field of application:


 Radiated RF emissions EN 61000-4-3:
80 MHz-1 GHz, 80 % AM (1 kHz), anechoic chambers TEM cells stripline,
 Conducted RF emissions EN 61000-4-6:
40 Veff, 150 kHz-230 MHz, 80 % AM (1 kHz)
 Burst EN 61000-4-4
 ESD EN 61000-4-2

Measured signals:
Supply voltages (switching controllers, linear controllers),
Reference voltages,
Digital signals (optical couplers, optical receivers),
Analogue signals (operational amplifiers, ADC, DAC).

Specific measurement technology:


To measure analogue signals under extreme interference conditions, measurement technology is
needed that
a) itself is not affected by disturbance fields and thus does not simulate any disturbance in the
equipment under test (EUT).
b) can be connected to the EUT in a decoupled way, i.e. connecting the probe head does not
result in the development of additional disturbance current paths through which disturbances
can penetrate or be discharged.

The A100 / A200 / A300 optical fibre measurement systems meet these demands.

Specific measuring task


Analogue electronic modules are generally influenced in EMC tests when RF, modulated by 1 kHz,
is applied to the EUT. This influence is due to the fact that the infiltrated RF disturbance is de-
modulated at PN junctions of the electronic circuit. This generates signal level fluctuations or 1 kHz
disturbance signals. The 1 kHz disturbance signal is produced through modulation of the RF
disturbance by 1 kHz.
Relatively slow disturbance signals with a fundamental wave of 1 kHz, which mostly interfere with
analogue circuits, are characteristic for RF disturbance coupling.
Figures 1 to 6 on the following pages show examples of useful signals that were subjected to
disturbances. The deviation of the signal form from the sine wave varies, i.e. the disturbance signal
also contains a harmonic component as well as the fundamental one. The task is to correctly
measure these relatively slow disturbance signals under extreme RF interference conditions.

The A100 / A200 / A300 measurement systems are ideal for these conditions because of their high
disturbance immunity.
3

Examples of disturbed useful signals

 Signals were measured with the AS 100 optical fibre probe


 Radiated RF emissions: 250 MHz, 80 % amplitude-modulated by 1 kHz
 EUT: operational amplifier circuit; RF coupling via an operational amplifier input; the disturbance
signal was measured on the output.

Figure 1: The oscillogram shows a constant useful signal with a demodulated 1 kHz component.

Figure 2: The demodulated 1 kHz disturbance signal, superimposed on the useful signal, shows a large
harmonic component.

Figure 3: The demodulated 1 kHz disturbance signal is limited by the lower rail.
4

Figure 4: Useful signal without interference

Figure 5: Useful signal with 1 kHz disturbance signal

Figure 6: Useful signal with 1 kHz disturbance signal limited by the upper rail
5

2. Function
2.1. Method of measurement
An RF-immune probe (AS XXX optical fibre probe), of which the A/D (analogue-to-digital) converter
is an integral part, is autonomously operated in the EUT without interfering with the EUT's EMC
conditions.
The only connection between the AS XXX probe and the EUT is an optical fibre that sends the
measured values, which are continuously generated by the ADC as a serial data flow, to an optical
receiver.

A D/A (digital-to-analog) converter (DAC) converts the digital sampling values back into an
analogue signal that can be evaluated by an oscilloscope.

Figure 7: Optical fibre measuring system

2.2. AS XXX ADC probe (sensor)


The voltage that is applied to input E of the AS XXX probe (sensor) is converted to a serial data
flow by an A/D converter.
The variable voltage divider generates the control ranges:

AS 100 50 / 10 VDC
AS 110 10 / 1 VDC
AS 120 1 / 0.1 VAC
AS 200 50 / 10 VDC
AS 300 ±10 VDC

In the AS 120 optical fibre probe a capacitive coupled amplifier is integrated with a low cut-off
frequency of about 300 Hz.
The sensor is protected against polarity reversal.

2.3. AE XXX DAC receiver


An optical receiver converts the optical fibre signal to a serial data signal. A sequential logic system
synchronizes the receiver and controls the serial-parallel conversion of the measured values and
the D/A conversion.

Operational amplifiers are connected to the DAC output to filter and adjust the control ranges.
6

3. Operation
3.1. Connection of the probe (example AS 100) to the EUT

Figure 8: AS 100 pin assignment

Attention: Only connect and disconnect an AS XXX sensor when it is not subjected to any
interference.

The connector between the optical fibre probe and the EUT has to be extremely short to ensure
satisfactory measurements of burst and ESD under radiated RF interference conditions. Common
prods and clips are too big.

The optical fibre probe has to be connected to the circuit board or IC via a socket to ensure the
requisite small-scale set-up. Because the socket, rather than the probe itself, is soldered to the
pins or lines, the measuring point can be quickly changed by plugging the probe into other sockets.

 To ensure the socket doesn’t move, glue it to the circuit board or IC of the EUT. Connect the
socket to the EUT via a CuL wire according to the sensor's pin assignment.
 Power is supplied via the EUT or a battery.

To avoid measuring errors pay attention:

 To position the AS XXX close to the EUT's GND areas.


 To solder the socket directly to the line to be measured with short connecting wires
(10...20 mm).
 That the power is supplied by immediate signal environment via the pins or block capacitor of
the relevant IC.
 To connect the GND of the socket and the GND of the module via a short connector
(10...20 mm) when power is supplied by a battery. Position a block capacitor (approx. 100 nF) at
the connecting point.

Basic rule: The closer the sensor (with its housing and GND connection) is positioned to the GND
system of the EUT, the higher its disturbance immunity will be and the fewer disturbances will
occur.
7

Figure 9: AS 100 is connected to the EUT; position the AS 100 on the IC package if there isn't enough free space
above GND

3.2. Starting the AE XXX optical receiver


 Connect the AE XXX to the oscilloscope (BNC plug).
 The receiver output supplies a zero-delay voltage of 10 V independent of the measuring range
that is selected on the AS XXX. Thus, the vertical deflection on the oscilloscope should be 0.5-
2 V/div.
 Connect the power supply to AE XXX (plug-in power supply unit)  lower LED "ON“ must be
on. Connect Sensor and receiver with the optical fibre  upper LED "Sync“ must be on. Loose
the clamped joint on the receiver, insert the optical fibre completely and slightly fasten the loose
joint.

Figure 10: AE 200 connectors and displays

 The oscilloscope must show an output voltage and/or signal form that corresponds to that
applied to the sensor input.
8

4. Safety instructions
When using a product from Langer EMV-Technik GmbH, please observe the following safety
instructions to protect yourself from electric shock or the risk of injuries.

Read and follow the instructions in the user manual and keep it in a safe place for later reference.
The device may only be used by personnel who are qualified in the field of EMC and who are fit to
work with and possibly be influenced by disturbance voltages and (electric and magnetic) burst
fields.

 Read the operating and safety instructions for all devices used in the set-up.
 Never use any damaged or defective devices.
 Carry out a visual check before using a measurement set-up with a Langer EMV-Technik GmbH
product. Replace any damaged connecting cables before using the product.
 Never leave a product from Langer EMV-Technik GmbH unattended whilst it is in operation.
 The Langer EMV-Technik GmbH product may only be used for its intended purpose. Any other
use is forbidden.
 People with a pacemaker are not allowed to work with this device.
 The test set-up should always be operated via a filtered power supply.

 Attention! Functional near fields and interference emissions may occur when the probe is
operated. The user is responsible for taking appropriate precautions to prevent any inter-
ference with the correct function of products outside the operational EMC environment (in
particular through interference emissions. This can be achieved by:

- observing an appropriate safety distance


- using of shielded or shielding rooms

 The disturbances that are injected into the modules can destroy the device under test (latch-up) if
their intensity is too high. Protect the device under test by:

- increasing the disturbance gradually and stopping when a functional fault occurs
- interrupting the power supply to the device under test in the event of a latch-up.

Attention! Make sure that internal functional faults are visible from outside. The device
under test may be destroyed due to an increase in the injection intensity if the faults are not
visible from outside. Take the following precautions if necessary:

 monitor the representative signals in the device under test


 use a special test software
 monitor visible reaction of the device under test to inputs (reaction test of the device under test).

We cannot assume any liability for the destruction of devices under test!
9

5. Warranty
Langer EMV-Technik GmbH will remedy any fault due to defective material or defective manufac-
ture, either by repair or by delivery of spare parts, during the statutory warranty period.

This warranty is only granted on condition that:

- the information and instructions in the user manual have been followed.

The warranty will be forfeited if:

- an unauthorized repair is performed on the product,


- the product is modified,
- the product is not used for its intended purpose.
10

6. Technical data
6.1. System

System A 100 A 200 A 300


Resolution 12 Bit 12 Bit 10 Bit
Conversion rate 125 ksps 3 Msps 12.5 Msps
Bandwidth 25 kHz 500 kHz 5 MHz
Transfer rate (optical fibre) 4 Mbps 48 Mbps 150 Mbps
Operating range (optical fibre) 1 – 20 m *
* The device is delivered with an optical fibre of 1.5 m – 1 channel or 6 m – 2 channels
Please order other dimensions separately.

6.2. Sensors

Sensor type AS 100 AS 110 AS 120 AS 200 AS 300


Measuring range, 50 V / 10 V / 1V/ 50 V / ±10 VDC
switchable 10 VDC 1 VDC 0,1 VAC 10 VDC
300Hz – 25kHz

Input resistance 100 k 1 M 1 M 100 k 100 k


Radiated immunity* >200 V/m >100 V/m >100 V/m >100 V/m 200 V/m
Current consumption ca. 3 mA ca. 30 mA 70 – 30 mA
Operating voltage 3 - 16 V 4.5 - 16 V
Dimensions (36x12.5x7.5) mm
Operating
0 – 70 °C
temperature range
* Connectors  20 mm, on GND area, disturbances at output < 50 mV

6.3. Receivers

Receiver type AE 100 AE 200 AE 300


Current consumption ca. 30 mA ca. 100 mA 90 mA
Operating voltage 12 - 16 V 12 V
Output voltage 0 - 10 V ±1 V
Dimensions (71x47x14) mm
Operating
0 – 70°C
temperature range
11

7. Scope of delivery
7.1. A100-1 / A200-1 / A300-1 (1-channel) set
Pos. Designation Type Quantity
---------------------------------------------------------------------------------------------------------------------------------
01 Optical receiver (DAC) AE 100 / 200 / 300 1
02 Optical fibre probe (ADC), of choice* AS XXX 1
03 Power supply unit 12 V / 300 mA 1
04 Optical fibre LWL Ø 2.2 mm, 1.5 m 1
05 Adapter socket 3-poles 3
06 wire CuL 34 m reel 1
07 System case (240x185x50) mm 1
08 Quick guide 1
09 User manual 1
* Further sensors are available on request and can be added to the set.
---------------------------------------------------------------------------------------------------------------------------------
12

7.2. A100-2 / A200-2 / A300-2 (2-channels) set


Pos. Designation Type Quantity
---------------------------------------------------------------------------------------------------------------------------------
01 Optical receiver (DAC) AE 100 / 200 / 300 2
02 Optical fibre probe (ADC), of choice* AS XXX 2
03 Power supply unit 12 V / 300 mA 1
04 Power split cable 20 cm 1
05 Optical fibre LWL Ø 2.2 mm 6 m 2
06 Adapter socket 3-poles 6
07 wire CuL 34 m reel 1
08 Transport case with foam insert (340x260x55) mm 1
09 Quick guide 1
10 User manual 1