Micropilot FMR51, FMR52

Examples for the construction of stilling wells

A B A-A

A A 1
1 C
2 3

100%
100%

4
5
(5.91...19.7)
150…500

≤ 1 (0.04)

(5.91...19.7)
150…500

≤ 1 (0.04)
D 6

0%

0%
mm (in)
A0019009

A Micropilot FMR50/FMR51: Horn 40mm(1½")


B Micropilot FMR50/FMR51/FMR52/FMR54: Horn 80mm(3")
C Stilling well with slots
D Full bore ball valve
1 Marking for axial alignment
2 Threaded connection
3 e.g. welding neck flange DIN2633
4  hole max. 1/10  stilling well
5  hole max. 1/10  stilling well; single sided or drilled through
6 Inside of holes deburred
7 Diameter of opening of ball valve must always be equivalent to pipe diameter; avoid edges and constrictions.

46 Endress+Hauser