ROTANODE
E7252X
E7252FX 0197
E7252GX
General Data
Electrical:
Circuit (Center-grounded) ................................................................... Three-phase full-wave rectified
Nominal Tube Voltage:
Radiographic ..................................................................................................... 40 ~ 150 kV Max.
Fluoroscopic ...................................................................................................... 40 ~ 125 kV Max.
Nominal Focal Spot Value:
Large Focus ...................................................................................................................... 1.2 mm
Small Focus ...................................................................................................................... 0.6 mm
Nominal Anode Input Energy (at 0.1s):
Large Focus ........................................................................................................................ 75 kW
Small Focus ........................................................................................................................ 27 kW
★The information contained herein is presented only as a guide for the applications of our products.
No Responsibility is assumed by TOSHIBA ELECTRON TUBES & DEVICES CO.,LTD.(TETD) for any infringements of patents or other rights of the
third parties which may result from its use. No license is granted by implication or otherwise under any patent or patent rights of TETD or others.
★The information contained herein may be changed without prior notice. It is therefore advisable to contact TETD before proceeding with the design of
equipment incorporating this product.
★For further particulars apply to TETD.
Motor Ratings:
Stator: XS-AL
Duty Starting Running
2) 2)
Power source (Hz) 180 60 180 60
Input power (W) 1100 910 83 83
4)
Voltage (V) 220 130 60 40
Current (A) 5.7 7.8 1.6 2.3
1) 5)
Min. Speed up (s) 1.2 0.8 - -
Capacitor (µF) 6 44 6 44
2)
Min. Braking (s) 3 / 90 V (DC)
Stator impedance:
Common-Main Winding ......................................................................................................... 9.4 Ω
Common-Auxiliary Winding.................................................................................................. 28.3 Ω
Stator: XS-R
Duty Starting Running
3) 3)
Power source (Hz) 180 50/60 180 50/60
Input power (W) 2300 1450 300 80
4)
Voltage (V) 460 240 130 58
Current (A) 5.4 6.5 2.0 1.5
1) 5)
Min. Speed up (s) 1.0 0.6 - -
Capacitor (µF) 3 24 3 24
2) 3)
Min. Braking (s) 1.5 / 90 V (DC)
Note 1) The speed up time from normal speed of 3000 min-1 (3600 min-1) to high speed of
10800 min-1 is 2/3 times of the specified speed up time from 0 to high speed, which is
described on motor rating table.
2) To be applied for high speed rotation.
3) To be obtained with AID starter Model 60/180.
4) The every applied voltage must be never exceeded 110% of the above specification.
5) The speed-up time is allowed up to 110% of the above specification.
Stator impedance:
Common-Main Winding ....................................................................................................... 27.5 Ω
Common-Auxiliary Winding.................................................................................................. 58.0 Ω
Anode Speed:
50 Hz ................................................................................................................... 2700 min-1 Min.
60 Hz ................................................................................................................... 3200 min-1 Min.
180 Hz ................................................................................................................... 9700 min-1 Min.
Resistance between Housing and Low Voltage Terminals ................................................ 2 MΩ Min.
Normal operating range of the housing temperature .......................................................... 16 ~ 75 °C
-2-
E7252X / E7252FX / E7252GX
Mechanical:
Dimensions: ................................................................................................. See dimensional outline
Overall Length .................................................................................................................. 476 mm
Maximum Diameter ....................................................................................................... 152.4 mm
Target:
Angle ........................................................................................................................... 12 degrees
Diameter ............................................................................................................................ 74 mm
Construction .................................................................... Rhenium-Tungsten faced Molybdenum
Permanent Filtration ............................................................... 0.9 mm Al / 75 kV IEC60522 / 2003
Radiation Protection (To meet the requirements of IEC 60601-1-3):
Leakage Technique Factor ................................................................................. 150 kV 3.4 mA
X-ray Coverage .................................................................................. 354 × 354 mm at SID 835 mm
Weight (Approx.) ........................................................................................................................ 18 kg
High Tension Terminals ...................................................... To meet the requirements of IEC 60526
Cooling Method .................................................................................................. Natural or forced air
Tube Housing Model Number:
E7252X .......................................................................................................................... XH-106V
E7252FX ........................................................................................................................ XH-180
E7252GX ......................................................................................................................... XH-181
-3-
E7252X / E7252FX / E7252GX
-4-
E7252X / E7252FX / E7252GX
Environmental Limits
Operating Limits:
Temperature ................................................................................................................ 10 ~ 40 ℃
Humidity ........................................................................................................................ 30 ~ 85 %
(No condensation)
Atmospheric Pressure .............................................................................................. 70 ~ 106 kPa
Shipping and Storage Limits:
Temperature .............................................................................................................. -20 ~ 70 ℃
Humidity ........................................................................................................................ 20 ~ 90 %
(No condensation)
Atmospheric Pressure .............................................................................................. 50 ~ 106 kPa
-5-
E7252X / E7252FX / E7252GX
As the result of reviewing the test requirements in the following Tables, an X-ray tube assembly
as a component of X-ray system or apparatus does not fall within the scope of the EMC directive of
IEC60601-1-2: 2001. Therefore, X-ray system manufacturer shall conduct the required test and
evaluate EMC.
CAUTION: The diagnostic X-ray system can only allow to use the designated cables and
components. If the different cables or components are used, they may deteriorate
the performance of electromagnetic emissions and immunity.
-6-
E7252X / E7252FX / E7252GX
<5 % UT
(>95 % dip in UT )
for 5 sec
Power frequency 3 A/m Power frequency magnetic fields
(50/60 Hz) should be at levels characteristic of a
magnetic field typical location in a typical
IEC 61000-4-8 commercial or hospital environment.
NOTE UT is the a.c. mains voltage prior to application of the test level.
-7-
E7252X / E7252FX / E7252GX
10 Vrms [V2] V d = [
12
] P
150 kHz to 80 MHz V 2
in ISM bandsa
23
d = [ ] P 800 MHz to 2.5 GHz
E1
Where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in meters (m).b
Field strengths from fixed RF transmitters,
as determined by an electromagnetic site
survey,c should be less than the
compliance level in each frequency ranged.
Interference may occur in the vicinity of
equipment marked with the following
symbol:
-8-
E7252X / E7252FX / E7252GX
Note: 1 At 80 MHz and 800 MHz, the higher frequency range applies.
2 These guidelines may not apply in all situations. Electromagnetic propagation is affected
by absorption and reflection from structures, object and people.
a
The ISM (industrial, scientific and medical) bands between 150kHz and 80MHz are 6,765MHz
to 6,795MHz; 13,553MHz to 13,567MHz; 26,957MHz to 27,283MHz; and 40.66MHz to
40.70MHz.
b
The compliance levels in the ISM frequency bands between 150kHz and 80MHz and in the
frequency range 80MHz to 2.5GHz are intended to decrease the likelihood that
mobile/portable communications equipment could cause interference if it is inadvertently
brought into patient areas. For this reason, an additional factor of 10/3 is used in calculating
the recommended separation distance for transmitters in these frequency ranges.
c
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the [EQUIPMENT or
SYSTEM] is used exceeds the applicable RF compliance level above, the [EQUIPMENT or
SYSTEM] should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as reorienting or relocating the
[EQUIPMENT or SYSTEM].
d
Over the frequency range 150kHz to 80MHz, field strengths should be less [V1] v/m.
-9-
E7252X / E7252FX / E7252GX
- 10 -
E7252X / E7252FX / E7252GX
Warning
1. Housing Rupture
Never input over-rated power to x-ray tube assembly.
If the input power is extremely higher than specification, it may cause the over temperature of
anode, insert tube glass shatter and ultimately the following serious problems due to generating
over-pressure by oil vaporization inside housing assembly.
In such a critical condition, the safety thermal switch can not protect x-ray tube even if it works.
We strongly request that the x-ray generator should have a protective function which manages
input power to x-ray tube assembly.
- 11 -
E7252X / E7252FX / E7252GX
Cautions
1. Over Rating
X-ray tube assembly can be broken with applying just one over rated shot.
Please read the technical data sheets carefully and follow the instructions.
2. Permanent Filtration
The total filtration and the distance between x-ray focal spot and human body are regulated legally.
They should be complied with the regulation.
4. Unexpected Malfunction
X-ray tube assembly may have the risk to be unexpectedly malfunctioning due to life termination or
failure. If the serious problems caused by the above risk is expected, we recommend to have a
contingency plan to avoid such a case.
5. New Application
If you use the product with new application not to be mentioned in this specification or with different
type of x-ray generator, please contact to us for confirming its availability.
- 12 -
E7252X / E7252FX / E7252GX
1. Qualified Persons
Only qualified persons who have technical training and professional knowledge can handle x-ray
tube assembly.
2. Fragile Glass
X-ray tube is assembled with glass, therefore, it can be broken with the mechanical vibration or
pulsed shock over 19.6m/s2 (2G).
Careful handling is required to treat or transport.
3. Ground Terminal
X-ray tube assembly has ground terminal. Ground cable should be connected.
4. High Voltage
All x-ray tubes operate at voltages high enough to kill through electrical shock. Never touch the
high voltage delivered plugs or terminals.
When direct access to such parts is required, the primary circuit should be disabled and high
voltage capacitors/cables discharged.
6. Operation Atmosphere
X-ray tube assembly is not allowed to use in the atmosphere of flammable or corrosive gas.
7. Protective Cover
X-ray tube assembly is not allowed to use without the protective cover attached.
8. Handling
Appropriate jig or tools are required for tube installation to avoid physical damages.
9. Returning Tube
X-ray tube assembly should be repackaged with the original material when it is returned back for
quality examination in our factory.
Be careful to put the tube upside cathode. If the packaging is not proper, the tube may not be
correctly examined.
- 13 -
E7252X / E7252FX / E7252GX
Caution in Operation
1. X-Ray Radiation
X-ray tube assembly should have the beam limiting equipment mounted on the x-ray port to protect
unnecessary radiation.
2. Dielectric Oil
X-ray tube assembly has dielectric oil contained for high voltage stability. As it is poisonous for
human health, if it is exposed to the non-restricted area, it should be disposed as following to the
local regulation.
3. Operation Atmosphere
X-ray tube assembly is not allowed to use in the atmosphere of flammable or corrosive gas.
4. Lead Disposition
X-ray tube housing is lined with lead to protect unnecessary radiation. As the lead powder or vapor
is harmful for human health, it should be disposed as following to the local regulation or returned
back to us with your cost of transportation. We dispose it in our facility with free of charge.
6. Any Malfunction
Please contact to your system service person immediately, if any malfunction is noticed.
- 14 -
E7252X / E7252FX / E7252GX
Caution Label
(a) This label is a caution label to notify the user of the following point.
"Housing end cap is used to protect the electric shock and x-ray leakage."
- 15 -
E7252X / E7252FX / E7252GX
400
150
300
100
200
50
100
0 0
0.01 0.03 0.1 0.2 0.3 0.5 1 2 3 5 10 0.01 0.03 0.1 0.2 0.3 0.5 1 2 3 5 10
0.02 0.05 0.02 0.05
EXPOSURE TIME [s] EXPOSURE TIME [s]
80kV
TUBE CURRENT [mA]
90kV 200
500 100kV 90kV
110kV 100kV
110kV
400 125kV 150
125kV
150kV
300 150kV
100
200
50
100
0 0
0.01 0.03 0.1 0.2 0.3 0.5 1 2 3 5 10 0.01 0.03 0.1 0.2 0.3 0.5 1 2 3 5 10
0.02 0.05 0.02 0.05
EXPOSURE TIME [s] EXPOSURE TIME [s]
- 16 -
E7252X / E7252FX / E7252GX
1000 16
100kV
900 125kV
800 150kV 14
700
600 12
Ef
500
80kV
400 10
60kV
300
40kV
200 8
100
0 6
3.9 4.1 4.3 4.5 4.7 4.9 5.1 5.3 5.5
400 10
100kV
125kV
FILAMENT VOLTAGE [V]
300 8
TUBE CURRENT [mA]
150kV
Ef
200 6
80kV
100 4
60kV
40kV
0 2
3.6 3.8 4.0 4.2 4.4 4.6 4.8 5.0 5.2
- 17 -
E7252X / E7252FX / E7252GX
Thermal Characteristics
Housing Thermal Characteristics
900
200W
800
700
HEAT STORAGE [kJ]
600
500
400
COOLING
300
200
100
HEATING
0
0 60 120 180 240 300 360
TIME [min]
180 COOLING
160
HEAT STORAGE [kJ]
140
120
100
80
60
40
HEATING
20
0
0 2 4 6 8 10 12
TIME [min]
The heating curves are showing examples of
average input power to the anode in operation.
- 18 -
E7252X / E7252FX / E7252GX
140
90
2-M63.5
P=1.27
ANODE
4-M6 8 DEEP
146.5
12°
90
50.8
476
134.5
4-M5 10 DEEP
90 S
L
11.1
CATHODE
105
TERMINAL CONNECTIONS
18.28
9 8 7 6 5 1 2 3 4 E
FOCAL SPOT
NC NC NC NC
10°
53
M A
81
ET
TEMPERATURE RELAY
192
(NORMALLY CLOSED)
Note) Do not connect terminal No.1 and No.5 or 6
in series circuit.
70
71
B
191
A:±1.5mm,B:±1.5mm
C : COMMON NC : NON-CONNECTION
L : LARGE FOCUS ET : EARTH TERMAINAL
S : SMALL FOCUS : CENTERAL X-RAY
M : MAIN WINDING OF THE STATOR ANODE & CATHODE TERMINAL
A : AUX. WINDING OF THE STATOR : IEC 60526 TYPE
- 19 -
E7252X / E7252FX / E7252GX
φ140
90
ANODE
2-M63.5
P=1.27
φ152.4
146.5
4-M6 8DEEP
11.1
12°
50.8
476
90
φ
92
45°
134.5
4-M5 10DEEP
90
C
11.1
CATHODE
L
105
TERMINAL CONNECTIONS
18.28
FOCAL SPOT
10°
9 8 7 6 5 1 2 3 4 E
NC NC NC NC 53
M A 81
ET 192
TEMPERATURE RELAY
(NORMALLY CLOSED)
Note) Do not connect terminal No.1 and
φ70
71
B 191
A:±1.5mm,B:±1.5mm
C : COMMON NC : NON-CONNECTION
L : LARGE FOCUS ET : EARTH TERMINAL
S : SMALL FOCUS : CENTRAL X-RAY
M : MAIN WINDING OF THE STATOR ANODE & CATHODE TERMINAL
A : AUX. WINDING OF THE STATOR : IEC60526 TYPE
- 20 -
E7252X / E7252FX / E7252GX
φ70
TERMINAL CONNECTIONS
9 8 7 6 5 1 2 3 4 E
NC NCNC NC
M A FOCAL SPOT
ET
191
TEMPERATURE RELAY
(NORMALLY CLOSED)
Note) Do not connect terminal No.1 and
71
No.5 or 6 in series circuit.
81
10°
CENTRAL OF FOCAL SPOT
53
A
B 18.28
A:±1.5mm,B:±1.5mm
φ140
CENTRAL RAY AND REFERENCE AXIS
90
ANODE
φ152.4
4-M6 8DEEP
146.5
11.1
12°
50.8
476
90
φ
92
45
134.5
4-M5 10DEEP
90 C
11.1
CATHODE
L
105
- 21 -
E7252X / E7252FX / E7252GX
EU REPRESENTATIVE
・TOSHIBA ELECTRONICS EUROPE
RIVERSIDE WAY, CAMBERLEY, SURREY. GU15 3YA U.K.
PHONE (0) 1276 694600 FAX (0) 1276 694800
For Sales & Technical Services, please contact the following representative:
Toshiba Electron Tubes & Devices Co., Ltd. Toshiba Electron Tubes & Devices Co., Ltd. meets
meets the Environmental Management internationally recognized Standards for Quality
System Standard, ISO 14001 Management System ISO 9001, ISO 13485