X Ray Tube
X Ray Tube
X Ray Tube
General Data
★The information contained herein is presented only as a guide for the applications of our products. No responsibility is assumed by TOSHIBA for any
infringements of patents or other rights of the third parties which may result from its use. No license is granted by implication or otherwise under any patent
or patent rights of TOSHIBA or others.
★The information contained herein may be changed without prior notice. It is therefore advisable to contact TOSHIBA before proceeding with the design of
equipment incorporating this product.
TOKYO JAPAN
Motor Ratings:
Stator: XS-AL
Duty Starting Running
Power source (Hz) 180 ( 2 ) 60 180 ( 2 ) 60
Input power (W) 1100 910 83 83
Voltage (V) 220 130 60 40
Current (A) 5.7 7.8 1.6 2.3
Min. Speed up ( 1 ) (s) 1.2 0.8 - -
Capacitor (µF) 6 44 6 44
Min. Braking ( 2 ) (s) 3 / 90 V (DC)
Stator: XS-R
Duty Starting Running
Power source (Hz) 180 ( 3 ) 50/60 180 ( 3 ) 50/60
Input power (W) 2300 1450 300 80
Voltage (V) 460 240 130 58
Current (A) 5.4 6.5 2.0 1.5
Min. Speed up ( 1 ) (s) 1.0 0.6 - -
Capacitor (µF) 3 24 3 24
Min. Braking ( 2 ) (s) 1.5 / 90 V (DC) 3
( )
Note ( 1 ) The speed up time from normal speed of 3000 min-1 (3600 min-1) to high speed of
10800 min-1 is 2/3 times of the specified speed up time from 0 to high speed, which
is described on motor rating table.
2
( ) To be applied for high speed rotation.
( 3 ) To be obtained with AID starter Model 60/180.
Anode Speed:
50 Hz ....................................................................................................................... 2700 min-1 Min.
60 Hz ....................................................................................................................... 3200 min-1 Min.
180 Hz ...................................................................................................................... 9700 min-1 Min.
Resistance between Housing and Low Voltage Terminals ................................................ 2 MΩ Min.
-2-
E7252X / E7252FX / E7252GX
Mechanical:
Dimensions: ................................................................................................... See dimensional outline
Overall Length ................................................................................................................... 476 mm
Maximum Diameter ........................................................................................................ 152.4 mm
Target:
Angle ............................................................................................................................. 12 degrees
Construction ..................................................................... Rhenium-Tungsten-faced Molybdenum
Inherent Filtration .................................................................... At least 0.7 mm Al equivalent at 75 kV
Radiation Protection (To meet the requirements of IEC 60601-1-3):
Leakage Technique Factor .................................................................................. 150 kV 3.4 mA
X-ray Coverage .................................................................................... 354 × 354 mm at SID 970 mm
Weight (Approx.) ......................................................................................................................... 18 kg
High Tension Terminals ...................................................... To meet the requirements of IEC 60526
Cooling Method ................................................................................................... Natural or forced air
Housing (including both end plates):
Model Number ................................................................................................ XH-106V (X-Type)
............................................................................................... XH-180 (GX-Type)
............................................................................................... XH-181 (FX-Type)
-3-
E7252X / E7252FX / E7252GX
-4-
E7252X / E7252FX / E7252GX
Environmental Limits
Operating Limits:
Temperature .................................................................................................................. 10 ~ 40 ℃
Humidity .......................................................................................................................... 30 ~ 85 %
(No condensation)
Atmospheric Pressure ............................................................................................... 70 ~ 106 kPa
Shipping and Storage Limits:
Temperature ................................................................................................................. -20 ~ 70 ℃
Humidity .......................................................................................................................... 20 ~ 90 %
(No condensation)
Atmospheric Pressure ............................................................................................... 50 ~ 106 kPa
-5-
E7252X / E7252FX / E7252GX
Warning
1. Housing Rupture
Never input over-rated power to x-ray tube assembly.
If the input power is extremely higher than specification, it may cause the over temperature of
anode, insert tube glass shatter and ultimately the following serious problems due to generating
over-pressure by oil vaporization inside housing assembly.
In such a critical condition, the safety thermal switch can not protect x-ray tube even if it works.
We strongly request that the x-ray generator should have a protective function which manages
input power to x-ray tube assembly.
-6-
E7252X / E7252FX / E7252GX
Cautions
1. Over Rating
X-ray tube assembly can be broken with applying just one over rated shot.
Please read the technical data sheets carefully and follow the instructions.
2. Inherent Filtration
The total filtration and the distance between x-ray focal spot and human body are regulated legally.
They should be complied with the regulation.
4. Unexpected Malfunction
X-ray tube assembly may have the risk to be unexpectedly malfunctioning due to life termination or
failure. If the serious problems caused by the above risk is expected, we recommend to have a
contingency plan to avoid such a case.
5. New Application
If you use the product with new application not to be mentioned in this specification or with
different type of x-ray generator, please contact to us for confirming its availability.
-7-
E7252X / E7252FX / E7252GX
1. Qualified Persons
Only qualified persons who have technical training and professional knowledge can handle x-ray
tube assembly.
2. Fragile Glass
X-ray tube is assembled with glass, therefore, it can be broken with the mechanical vibration or
pulsed shock over 19.6m/s2 (2G).
Careful handling is required to treat or transport.
3. Ground Terminal
X-ray tube assembly has ground terminal. Ground cable should be connected.
4. High Voltage
All x-ray tubes operate at voltages high enough to kill through electrical shock. Never touch the
high voltage delivered plugs or terminals.
When direct access to such parts is required, the primary circuit should be disabled and high
voltage capacitors/cables discharged.
6. Operation Atmosphere
X-ray tube assembly is not allowed to use in the atmosphere of flammable or corrosive gas.
7. Protective Cover
X-ray tube assembly is not allowed to use without the protective cover attached.
8. Handling
Appropriate jig or tools are required for tube installation to avoid physical damages.
9. Returning Tube
X-ray tube assembly should be repackaged with the original material when it is returned back for
quality examination in our factory.
Be careful to put the tube upside cathode. If the packaging is not proper, the tube may not be
correctly examined.
-8-
E7252X / E7252FX / E7252GX
Caution in Operation
1. X-Ray Radiation
X-ray tube assembly should have the beam limiting equipment mounted on the x-ray port to protect
unnecessary radiation.
2. Dielectric Oil
X-ray tube assembly has dielectric oil contained for high voltage stability. As it is poisonous for
human health, if it is exposed to the non-restricted area, it should be disposed as following to the
local regulation.
3. Operation Atmosphere
X-ray tube assembly is not allowed to use in the atmosphere of flammable or corrosive gas.
4. Lead Disposition
X-ray tube housing is lined with lead to protect unnecessary radiation. As the lead powder or vapor
is harmful for human health, it should be disposed as following to the local regulation or returned
back to us with your cost of transportation. We dispose it in our facility with free of charge.
6. Any Malfunction
Please contact to your system service person immediately, if any malfunction is noticed.
-9-
E7252X / E7252FX / E7252GX
Caution Label
(a) This label is a caution label to notify the user of the following point.
"Housing end cap is used to protect the electric shock and x-ray leakage."
- 10 -
E7252X / E7252FX / E7252GX
400
150
300
100
200
50
100
0 0
0.01 0.03 0.1 0.2 0.3 0.5 1 2 3 5 10 0.01 0.03 0.1 0.2 0.3 0.5 1 2 3 5 10
0.02 0.05 0.02 0.05
EXPOSURE TIME [s] EXPOSURE TIME [s]
125kV
600 125kV 200
60kV
50kV 150kV
500 150kV
150
400
300 100
200
50
100
0 0
0.01 0.03 0.1 0.2 0.3 0.5 1 2 3 5 10 0.01 0.03 0.1 0.2 0.5 1 2 3 5 10
0.02 0.05 0.02 0.05 0.3
EXPOSURE TIME [s] EXPOSURE TIME [s]
- 11 -
E7252X / E7252FX / E7252GX
80kV
90kV 200
500 100kV 90kV
100kV
110kV
110kV
400 125kV 150
125kV
150kV
300 150kV
100
200
50
100
0 0
0.01 0.03 0.1 0.2 0.3 0.5 1 2 3 5 10 0.01 0.03 0.1 0.2 0.3 0.5 1 2 3 5 10
0.02 0.05 0.02 0.05
EXPOSURE TIME [s] EXPOSURE TIME [s]
100kV
90kV
400 110kV
100kV
125kV 150
110kV
300 150kV 125kV
150kV
100
200
50
100
0 0
0.01 0.03 0.1 0.2 0.3 0.5 1 2 3 5 10 0.01 0.03 0.1 0.2 0.3 0.5 1 2 3 5 10
0.02 0.05 0.02 0.05
EXPOSURE TIME [s] EXPOSURE TIME [s]
- 12 -
E7252X / E7252FX / E7252GX
Three-Phase
600 12 300 8
Ef
500
Ef
400 10 200 6
300
80KV 80KV
200 8 100 4
60KV
60KV
100 40KV
40KV
0 6 0 2
3.9 4.1 4.3 4.5 4.7 4.9 5.1 5.3 5.5 3.6 3.8 4.0 4.2 4.4 4.6 4.8 5.0 5.2
FILAMENT CURRENT [A] FILAMENT CURRENT [A]
Single-Phase
900 100KV
100KV
800 14 400 125KV 10
125KV
FILAMENT VOLTAGE [V]
FILAMENT VOLTAGE [V]
TUBE CURRENT [mA]
TUBE CURRENT [mA]
700 150KV
150KV
600 12 300 8
Ef
Ef
500
400 10 200 6
300
- 13 -
E7252X / E7252FX / E7252GX
Thermal Characteristics
800
1000 700
HEAT STORAGE [kJ]
600
800 COOLING
500
600
400
300
400
200
200
100
0 0
0 10 20 30 40 50 60 70 80 90 100 ・・・(1)
0 20 40 60 80 100 120 140 160 180 200 ・・・(2)
TIME [min]
(1) Cooling with Air Circulator.
(2) Cooling Without Air Circulator.
180 COOLING
250
160
HEAT STORAGE [kJ]
200 140
120
150
100
80
100
60
40
50 HEATING
20
0 0
0 2 4 6 8 10 12
TIME [min]
The heating curves are showing examples of
average input power to the anode in operation.
- 14 -
E7252X / E7252FX / E7252GX
- 15 -
E7252X / E7252FX / E7252GX
- 16 -
E7252X / E7252FX / E7252GX
- 17 -
E7252X / E7252FX / E7252GX
Toshiba Nasu Operations meets the Environmental Toshiba Corporation, Display Devices & Components
Management System Standard, ISO14001 Company, Electron Tubes and Devices Division, Nasu
Operations-Electron Tubes meets internationally
recognized standards for Quality Management System
ISO 9001, ISO13485