Pure Water
Pure Water
Pure Water
**from equivalent values from USP Table for Conductivity Limits as a Function of Temperature.
*The measurement of pH in Type I, II, and III reagent water has been eliminated because these grades of water do not
contain constituents in sufficient quantity to alter pH.
CLSI (CLINICAL LABORATORY STANDARDS INSTITUTE)*
Clinical Laboratory Reagent Water (CLRW)
CLSI
Resistivity @ 25°C . . . . . . . . . . . . . . . . . . . . . . . . . 10
(megohms-cm)
Conductivity @ 25°C . . . . . . . . . . . . . . . . . . . . . . . 0.1
(micromhos/cm)
Bacterial Growth (cfu/mL) . . . . . . . . . . . . . . . . . . . . 10
Organics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . <500 ppba
Particulate Matter . . . . . . . . . . . . . . . . . . . . . . . . . . 0.22 µm Filtera
a
This specification is a process specification, and not measured by the end-user.
*Formerly NCCLS (National Committee for Clinical Laboratory Standards).