ASTM® Reagent Water Specifications

CLSI® (formerly NCCLS) Reagent Water Specifications


USP® 30 Pharmaceutical Grade Water Standard

UNITED STATES PHARMACOPOEIA 30 STANDARDS


USP 30 Pharmaceutical Grade Water

Organics . . . . . . . . . . . . . . . . . . . . . . . . . . <0.5 ppm TOC


Conductivity . . . . . . . . . . . . . . . . . . . . . . . <1.3 µS/cm at 25°C in-line measurement**
Endotoxin by LAL
Purified Water . . . . . . . . . . . . . . . . . . . . No specification
WFI . . . . . . . . . . . . . . . . . . . . . . . . . . . . <0.25 EU/mL
Bacteria (guideline only)
Purified Water . . . . . . . . . . . . . . . . . . . . <100 cfu/mL
WFI . . . . . . . . . . . . . . . . . . . . . . . . . . . . <10 cfu/100 mL

**from equivalent values from USP Table for Conductivity Limits as a Function of Temperature.

AMERICAN SOCIETY FOR TESTING AND MATERIALS


Standard Specifications for Reagent Water
ASTM Type I II III IV

Conductivity @ 25°C (micromhos/cm) . . . . . . . . . . 0.056 . . . . . . 1.0 . . . . . . . 0.25 . . . . . . . . 5.0


Resistivity @ 25°C (Megohms-cm). . . . . . . . . . . . . 18 . . . . . . . . 1.0 . . . . . . . 4.0 . . . . . . . . . 0.2
Total Silica (µg/L) . . . . . . . . . . . . . . . . . . . . . . . . . . 3 . . . . . . . . 3. . . . . . . .. 500 . . . . . . . . . –
Total Organic Carbon (µg/L) . . . . . . . . . . . . . . . . . . 100 . . . . . . . 50. . . . . . . . 200 . . . . . . . . . –
Chlorides (µg/L) . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 . . . . . . . . . 5. . . . . . . . . 10 . . . . . . .. . . 50
Sodium (µg/L) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 . . . . . . . . . 5. . . . . . . . . 10 . . . . . . . . . . 50
pH* . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . – . . . . . . . . . –. . . . . . .. . – . . . . . . . . 5.0-8.0

Type A Type B Type C


Max Heterotropic Bacteria Count . . . . . . . . . . . . . . 10/1000 mL. . . 10/100 mL. . . 100/10 mL
Endotoxin, EU/mL . . . . . . . . . . . . . . . . . . . . . . . . . . < 0.03 . . . . . . . 0.25 . . . . . . . . N/A

*The measurement of pH in Type I, II, and III reagent water has been eliminated because these grades of water do not
contain constituents in sufficient quantity to alter pH.
CLSI (CLINICAL LABORATORY STANDARDS INSTITUTE)*
Clinical Laboratory Reagent Water (CLRW)
CLSI
Resistivity @ 25°C . . . . . . . . . . . . . . . . . . . . . . . . . 10
(megohms-cm)
Conductivity @ 25°C . . . . . . . . . . . . . . . . . . . . . . . 0.1
(micromhos/cm)
Bacterial Growth (cfu/mL) . . . . . . . . . . . . . . . . . . . . 10
Organics . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . <500 ppba​
Particulate Matter . . . . . . . . . . . . . . . . . . . . . . . . . . 0.22 µm Filtera​
a​
This specification is a process specification, and not measured by the end-user.
*​Formerly NCCLS (National Committee for Clinical Laboratory Standards).

9016 Oak Street., Vancouver, BC V6P 4B9 604-677-6777 Fax 264-9777 ​www.Filterco.net