Erbe T-175 ESU - Service Guideline
Erbe T-175 ESU - Service Guideline
Erbe T-175 ESU - Service Guideline
1. General
Applies to model T175. Comment from a website:
“exclusively designed for being used in the field of veterinary
medicine or for export purposes”.
2. Reference Documents
Erbotom T175 Operating Instructions.
Erbotom T175 Service Manual”
T175 ESU
4.1. Inspection
Check unit for any damage; i.e. connectors, knobs, the
power cord and the footswitch with its cable.
Note: This generator’s output power maximizes for 300 ohm load
resistance. The analyzer suggested has load settings of only 500 and 125 PATIENT PLATE CABLE
ohms. Refer to load curves in Appendix for corrections.
Note: Bipolar can be activated with either the footswitch or by the load sensing of the bipolar forceps.
Activate the bipolar mode by touching the bipolar output (via forceps for example) to a load.
After a couple of seconds (internally adjustable) the unit will activate in bipolar. The activation
can also be used to measure power as above.
4.11. RF Leakage -- ESU analyzers read RMS current, and when appropriately connected, can
display RF leakage. Limits are below. RF leakage is an infrequent field check but should not
be overlooked for older equipment.
The standard IEC 60601-2-2, particular requirements for the safety of high frequency surgical
equipment, limits this parameter to a maximum of 4.5 W measured on a 200 Ω non-inductive
resistor for monopolar applications (150mA in this case).
APPENDIX
160
140
CUT
120 COAG
WATTS
100
80
60
40
20
DIAL SETTING
0
0 1 2 3 4 5 6 7 8 9 10
55 60
50 Bipolar Output Power vs Dial @ 75 Ohm Load
45
Bipolar Wattage vs Load 50
40
35 40
Watts
30
30
25
20
20
15
10
10
5
0 BIPOLAR DIAL SETTING
0
10 100 OHMS 1000 10000 1 2 3 4 5 6 7 8 9 10