The instrument also provides constant column flow rate operation to improve analysis results, whilst preserving the
analytical column.
The Digital Pressure Control (DPC) of carrier gas allows to operate in three different ways: Pressure Programming,
Constant Flow and Constant Pressure.
DANI GC1000 offers a complete range of injectors, all equipped with electronic pressure control, which allow an accurate
injection of all samples in packed or capillary columns.
DANI not only provides traditional split/splitless and packed column injectors, but also the programmable temperature
injector PTV.
Thanks to cold injection of the sample, the PTV solves all problems of discrimination of complex mixture as well as the
treatment of thermally labile compounds.
DANI PTV also combines the best repeatability and accuracy injection with easy-to-use together with all advantages of the
traditional split/splitless injector.
Solvent split, large volume, and sample focusing injections and the compatibility with SPME technique, make the DANI PTV
a universal injector.
DANI GC1000 features a full range of highly sensitive detectors. Each detector incorporates a Digital Pressure Control
(DPC) that permits digital setting of the detector gas flowrate.
DANI selective detectors specifically eliminate matrix interferences while providing maximum sensitivity.
Page 41 GC1000
DANI GC1000 – Digital Gas Chromatograph
Technical Specifications
DIMENSIONS
Size 420x670x580 mm (W x H x D)
ENVIRONMENTAL CONDITIONS
Operating temperature 10–40°C
Operating humidity 0-90%
Line voltage requirements 230V ~ 50 Hz (±10% max) 2300W
HEATED ZONES
Independent heated zones, not including 4 injectors
oven 4 detectors
2 auxiliary (max temperature 450°C)
COLUMN OVEN
Dimensions 290x300x190 mm (W x H x D)
Operating temperature 5°C above ambient to 500°C
-50°C to 500°C with liquid CO2 cryogenic cooling option
-100°C to 500°C with liquid N2 cryogenic cooling option
Temperature resolution 0.1°C
Thermal accuracy ± 0.1°C
Thermal stability ± 0.1°C
Temperature programming 5 ramps, 6 isotherms
Max Heating Ramp Rate 50°C/min, 0.1 °C/min resolution
Typical cool-down rate 300°C to 50°C in 5 min
Isothermal time 0.00 to 200.00 min, 0.01 min resolution
PNEUMATIC CONTROLS
Injector Pressure range 0-60 psi
Injector Pressure/flow programming Constant flow
Constant or programmed pressure
Pressure programming: 5 ramps/6 isobars
Page 42 GC1000
DANI GC1000 – Digital Gas Chromatograph
Technical Specifications
PNEUMATIC CONTROLS
Injectors Carrier Gas Control Digital Pressure Control (DPC)
Injectors Split Gas Control Digital Readout Module (DRM)
Detectors Gas Control Digital Pressure Control (DPC)
INJECTORS
Injectors available Packed (PK)
Split/splitless (SL/IN)
Programmable Temperature Vaporizer (PTV)
Gas sampling valves (GSV)
Carrier gas He, N2, H2, Ar, Ar+CH4
Maximum number installed 4
simultaneously
Maximum number of PTV 2
DETECTORS
Detectors available FID flame ionization detector
NPD nitrogen and phosphorous detector
ECD electron capture detector
PID photo ionization detector
TCD thermal conductivity detector
mTCD micro thermal conductivity detector
FPD flame photometric detector
Maximum number installed 4
simultaneously
Maximum number of TCD 2
Maximum number of µTCD 2
Output signals Analog outputs (0-1V, 0-10V)
DATA COMMUNICATION
Remote control Start in; start and ready out (contact relays normally open)
Communication RS232
Time events n.5 voltage free contacts (contact relays normally open)
USER INTERFACE
Display 2 lines / 20 characters/line
Keyboard Soft Touch keyboard
VARIOUS
Analytical Methods Capacity 40
Page 43 GC1000