Vaisala HUMICAP® Humidity and Temperature Transmitters HMT120 and HMT130
Vaisala HUMICAP® Humidity and Temperature Transmitters HMT120 and HMT130
Vaisala HUMICAP® Humidity and Temperature Transmitters HMT120 and HMT130
com
Performance
The HMT120/130 incorporates Vaisala
HUMICAP technology that measures
relative humidity accurately and
reliably. The Vaisala HUMICAP is
resistant to dust and most chemicals.
The transmitter enclosure is
optimized for use in cleanrooms.
The smooth surface of the enclosure
makes it easy to clean and the
enclosure material is chosen
to tolerate purifying agents.
Furthermore, the cabling can be
done through the back wall of the
transmitter.
Interchangeable Probe
The HMT120/130 transmitters use
a fully interchangeable relative
humidity probe. The probe can
be easily removed and replaced
with a new one without having to
adjust the transmitter, which allows
for easy and quick recalibration
of the transmitter. The probe can
be adjusted using one of Vaisalas
portable meters as a reference.
Also available is a constant output
probe with fixed RH and T output
for convenient inspection of the
monitoring system and signal
transfer line.
Available Options
The HMT120 and HMT130
transmitters are available as wall
mounted or with remote probe. For
high temperature applications or
where space is limited, the remote
Interchangeable probe
(easy field calibration)
Wall-mounted or with a
remote probe
Enclosure IP65
Technical Data
Performance
Operating Environment
RELATIVE HUMIDITY
Measurement range
0 ... 100 %RH
Accuracy including non-linearity, hysteresis, and repeatability
temperature range
0 C ... +40 C (32 F ... 104 F)
0 ... 90 %RH
1.5 %RH
90 ... 100 %RH
2.5 %RH
temperature range
-40 ... 0 C, +40 ... +80 C
(-40 F ... 32 F, 104 F ... 176 F)
0 ... 90 %RH
3.0 %RH
90 ... 100 %RH
4.0 %RH
Factory calibration uncertainty at +20 C (+68 F)
0 ... 90 %RH
1.1 %RH
90 ... 100 %RH
1.8 %RH
Humidity sensor
Vaisala HUMICAP 180R
Stability
2 %RH over 2 years
In typical HVAC applications
0.5 %RH per year
TEMPERATURE
Measurement range
4 ... 20 mA
10 ... 30 VDC (RL= 0 ohms)
20 ... 30 VDC (RL< 500 ohms)
Mechanics
Material
transmitter housing
display window
probe body
probe grid filter
Housing classification
Connections
inputs and outputs
probe interface
Probe cable lengths
Display (optional)
Weight (including probe)
PBT plastic
PC plastic
Stainless steel (AISI 316)
Chrome coated ABS plastic
IP65
Screw terminals 0.5 ... 1.5 mm2
4-pin M8 female panel connector
3 m, 5 m, 10 m - up to 50 m
128 x 64 resolution full graphics
B&W display without backlight
270 g
Accessories
Humidity and temperature probe
Humidity and temperature replacement probe
Constant output probe
Standard humidity sensor
Catalytic humidity sensor for H2O2
Probe mounting flange
Probe mounting clamps, 10 pcs
HMP110 sensor protection
plastic grid filter
plastic grid with membrane filter
stainless steel sintered filter
Probe cable 3 m
Probe cable 5 m
Probe cable 10 m
Probe cable 20 m
Radiation shield
Rain shield with installation kit
Duct installation kit
HM70 connection cable
USB serial interface cable
* See separate order form
HMP110*
HMP110R*
HMP110REF*
HUMICAP 180R
HUMICAP 180V
226061
226067
DRW010522SP
DRW010525SP
HM46670SP
HMT120Z300
HMT120Z500
HMT120Z1000
HMT120Z2000
DTR504A
215109
215619
211339
219685
Dimensions
Dimension in mm
12 [0,47]
72
145
71 [2,8]
120
Please contact us at
www.vaisala.com/requestinfo