Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                

Newlab 200 CFPP - Cold Filter Plugging Point: Automatic Analysers: Newlab Range

Download as pdf or txt
Download as pdf or txt
You are on page 1of 2

Automatic Analysers: NewLab Range

NewLab 200
CFPP – Cold Filter Plugging Point
CH-6864 Arzo,Mendrisio, Switzerland
tel. +41 91 6300703, fax +41 91 6300719
www.lin-tech.ch – info@lin-tech.ch
Linetronic Technologies SA
Via Onorio Longhi 2

ASTM D6371 Measuring Parameters • T hanks to an istogram (graph) that shows


IP 309 - IP 419 • Temperatures: in °C the aspiration and release times it is possible
EN 116 • M easuring range: +80°C … -80°C to observe the behaviour of the sample
• R esolution: 0.06 °C during its cooling phase
Subject
• A ccuracy: ± 0.1 °C • T his feature is an excellent tools
Cold Filter Plugging Point of diesel,
• R epeatability / Reproducibility: for the observation and evaluation of the
biodiesel and heating fuels.
as per standards methods or better additivations actions and behaviour

Measuring CFPP Principle Diagnostic Menu


Software Features
The sample is cooled down according • D irectaccess to all analog, digital, inputs
• A ble to manage up to 6 analytical heads
to the methods and when the preselected and outputs
simultaneously (stand alone)
temperature is reached a vacuum of 20 • S electable value displaying: °C / Volt
• U ser friendly interface
mBar is automatically applied to the sample. • Vacuum data displayed in mBars
• A ll analytical parameters recorded
The product is sucked through the filter into
• C ustomizable analysis parameters Calibration Menu
the calibrated aspiration pipette.
and methods • Automatic calibration
If the sample takes more than 60 seconds
• C ustomizable results report of each temperature probe
to reach the upper barrier detector
• P rintable graphs and results • A utomatic calibration of vacuum sensor
(during the aspiration phase), or it fails to
• S elf-identification of the typology • L ast calibration date referred to each single
return completely into the test jar before
of the analysers connected probe displayed and relative data printable
that the product has cooled by a further 1°C,
• D isplay of calibration diagram
the Cold Filter Plugging Point is reached. The software includes:
• I nsertion of offset values
Measuring CFPP Devices Analysis Menu • S tandard and advanced calibration modes
• Aspiration pipette • S tandard method as per ASTM / IP / ISO /
Data Utilities
• Filter assembly EN / DIN… norms of reference
• F ields for introduction of operator
• Light barrier • Optional methods:
and product name
· T-sample – T-bath (Delta T constant)
Measuring Temperature Probe • A rchive viewer for files recall
· cooling rate °C / h
• P latinum resistance PT100 class A • A ll analysis stored in Excel® compatible
· selectable bath steps
format
· fast bath with selectable temperature
Accessories • S torage capacity for more than 60’000
• A udible alarm and displayed messages
• O ilLab 250 – external vacuum generator: analysis
at the end of the analysis and in case
· Vacuum pump • LIMS compatible
of errors and/or malfunctions
· Two glass bottles
• T he parameters displayed and updated
· A glass cork with: u-tube, funnel, manual Integrated Touch Screen Panel PC
in real time are:
flow regulating valve • TFT/LCD 12”
· sample temperature
• O ilLab 255 – internal vacuum generator: • R esolution 1024 × 768, 16.2 M colours
· bath temperature
· 1 x micro-pump of 350 mBar • 2 USB ports for connection to an external
· vacuum pressure
· 1 x electronic pressure / vacuum regulator printer and/or external PC
· low level light value
composed by: proportional valve, pressure • S torage capacity for more than 60’000
· up level light value
/ vacuum control sensor, regulator for analysis
· aspiration time
reference vacuum generation at 20 mBar,
· release time
vacuum stabilizer
· intertime test
Automatic Analysers: NewLab Range

NewLab 200
CFPP – Cold Filter Plugging Point

CH-6864 Arzo,Mendrisio, Switzerland


tel. +41 91 6300703, fax +41 91 6300719
www.lin-tech.ch – info@lin-tech.ch
Linetronic Technologies SA
Via Onorio Longhi 2
Cleaning pipette procedure Cord Cable:
• B y using a suitable cleaning liquid • 3
conductors flexible cable 2 m (7 feet)
and pressing the relevant function button length with PVC sheath oil and heat resistant
the analyser performs a cleaning sequence as per CENELEC directives
of 10 aspirations cycles
• E asy removing of aspiration pipette Ambient Temperature
and filter assembly allows cleaning • M ax 32 °C
according to the methods • H.R. 80%

Test Jar Dimensions and weight


• S ame dimensions and volume as described • 1 test pos.: w 66  d 60  h 80 cm, 60 kg
by the standard test methods • 2 test pos.: w 66  d 60  h 80 cm, 90 kg / 100 kg
• P roduct level mark • 3 test pos.: w 100  d 60  h 80 cm, 130 kg
• S mall edge on the top in order to fix the • 4 test pos.: w 134  d 60  h 80 cm, 160 kg
glass cell to the analytical head • 6 test pos.: w 130  d 75  h 170 cm, 280 kg

Cooling System Spare Parts


• I ntegrated gas CFC free motor compressors: • L AB-xxx/005-03: heather + auto adhesive + insulation
· S ingle stage • L AB-xxx/005-04: thermo switch
(for temperatures up to -40°C / 1) • L AB-xxx/005-06: PT100 bath
· D ouble stage • L AB-xxx/007-02: static relay
(for temperatures up to -80°C / 2) • L AB-xxx/007-04: PCB fuse 1 AT (box of 10 pcs)
• E quipped with an automatic energy power • L AB-xxx/006-01: cooling fluid valve + fitting
save system. After 15 minutes from • L AB-200/002-02: vacuum valve + fitting
the end of the analysis the cooling system • L AB-200/007-01: main electronic board CFPP
goes in stand-by mode. • L AB-200/008-06: sensor up (orange)
• L AB-200/008-07: sensor down (yellow)
Safety Devices • L AB-200/008-08: emitter up (red)
• P ressure controller for 1st stage motor • L AB-200/008-09: emitter down (blue)
compressor • L AB-200/008-12: PT100 product w/connector
• P ressure controller for 2nd stage motor • L AB-200/008-04: CFPP calibrated glass cell
compressor • L AB-200/008-041: o-ring for CFPP test jar
• T hermostat for 2nd stage activation • L AB-200/008-13: calibrated aspiration pipette CFPP
• T hermo-switch for each cooling / heating • L AB-200/008-18: clamp + kness for vacuum tube
jacket • L AB-200/013-01: filter assembly
• M otor compressors equipped with internal • LAB-200/013-02: filter
overload devices • L AB-200/1288: o-ring (big) for CFPP filter
• L AB-200/1232: o-ring (small) for CFPP filter
Electrical Supply
Calibration Tools
• 2 20V ± 15% / 50 to 60 Hz
• O ilLab 80: calibration decade box – PT100 simulator
• 1 15V ± 15% / 60 Hz
• O ilLab 81: set of connectors and cables for cold range

You might also like