4# Pour Point - NewLab 300

Download as pdf or txt
Download as pdf or txt
You are on page 1of 2

Automatic Analysers: NewLab Range

NewLab 300
Pour Point
Linetronic Technologies SA

CH-6864 Arzo,Mendrisio, Switzerland


tel. +41 91 6300703, fax +41 91 6300719
www.lin-tech.ch – info@lin-tech.ch
Via Onorio Longhi 2

ASTM D97 Measuring Pour Point Principle The software includes:


ASTM D5853 According to the methods, the sample
Analysis Menu
ASTM D5950 is cooled down at a specified rate and,
• Standard method as per ASTM / IP / ISO /
ASTM D6074 at the prescribed temperature intervals,
EN / DIN… norms of reference:
ASTM D6158 the mechanical arm of the analyser lifts
· (internal) with sample pre-heating
IP 15 the test jar from the cooling jacket and tilts
· (external) without sample pre-heating
IP 441 it in order to bring it in horizontal position
• Optional methods:
ISO 3016 to test the flow of the product.
· T-sample – T-bath (Delta T constant)
EN ISO 22995 The sample movement is detected by
· cooling rate °C / h
the thermal probes (PT100 detection) placed
· selectable bath steps
Subject above the sample surface which react if touched
· fast bath
Pour Point of petroleum products, crude oils, by the cooled sample.
· selectable tilt out test temperature
motor and engine oils, additives, lubricating
Measuring Pour Point Devices • Audible alarm and displayed messages
oils, …
• Two PT100 detection probes placed (at the end of the analysis and in case
on the surface of the product of errors and/or malfunctions)
• Mechanical moving arm bringing the test jar Diagnostic Menu
in horizontal position • Direct access to all analog, digital, inputs
and outputs
Measuring Temperature Probe
• Selectable value displaying: °C / Volt
• Platinum resistance PT100 class A
Calibration Menu
Measuring Parameters • Automatic calibration
• Temperatures: in °C of each temperature probe
• Measuring range: -110°C … +100°C • Last calibration date referred
• Range of analysis: -90°C … +60°C (300/2-SA) to each single probe displayed
• Resolution: 0.06 °C and relative data printable
• Accuracy: ± 0.1 °C • Display of calibration diagram
• Repeatability / Reproducibility: • Insertion of offset values
as per standards methods or better • Standard and advanced calibration modes
Data Utilities
Software Features
• Fields for introduction of operator
• New LabLink software able to manage
and product name
up to 6 analytical heads simultaneously
• Archive viewer for files recall
(stand alone)
• All analysis stored in Excel®
• User friendly interface
compatible format
• All analytical parameters recorded
• Storage capacity for more than
• Customizable analysis parameters
60’000 analysis
and methods
• LIMS compatible
• Customizable results report
• Printable graphs and results
• Self-identification of the typology
of the analysers connected

Automatic Analysers: NewLab Range

NewLab 300
Pour Point

Linetronic Technologies SA

CH-6864 Arzo,Mendrisio, Switzerland


tel. +41 91 6300703, fax +41 91 6300719
www.lin-tech.ch – info@lin-tech.ch
Via Onorio Longhi 2
NewLab 300 ST

Integrated Touch Screen Panel PC Cord Cable NewLab 300 ST


• TFT/LCD 12” • 3 conductors flexible cable 2 m (7 feet) • Measuring range: -110°C … +100°C
• Resolution 1024 × 768, 16.2 M colours length with PVC sheath oil and heat resistant • Range of analysis: -110°C … +55°C
• 2 USB ports for connection to an external printer as per CENELEC directives • Resolution: 0.01 °C
and/or external PC • Width: 34 cm
• Storage capacity for more than 60’000 analysis Ambient Temperature • Depth: 60 cm
• Max 32 °C • Height: 80 cm
Test Jar • H.R. 80% • Weight: 34 kg
• Same dimensions and volume as described
by the standard test methods Dimensions and weight
• Product level mark • 1 test pos.: w 66 × d 60 × h 80 cm, 60 kg
• Small edge on the top in order to fix the glass • 2 test pos.: w 66 × d 60 × h 80 cm, 90 kg / 100 kg
cell to the analytical head • 3 test pos.: w 100 × d 60 × h 80 cm, 130 kg
• 4 test pos.: w 134 × d 60 × h 80 cm, 160 kg
Cooling System • 6 test pos.: w 130 × d 75 × h 170 cm, 280 kg
• Integrated gas CFC free motor compressors:
· Single stage Spare Parts
(for temperatures up to -40°C / 1) • LAB-xxx/005-03: heater + auto adhesive + insulation
· Double stage • LAB-xxx/005-04: thermo switch
(for temperatures up to -80°C / 2) • LAB-xxx/005-06: PT100 bath
• Equipped with an automatic energy power • LAB-xxx/007-02: static relay
save system. After 15 minutes from • LAB-xxx/007-04: PCB fuse 1.6 A, box of 10 pcs.
the end of the analysis the cooling system • L AB-xxx/006-01: cooling fluid valve + fitting
goes in stand-by mode. (only for motor-compressor units)
• LAB-300/007-01: main electronic board Pour Point
Safety Devices • LAB-300/002-16: precision potentiometer
• Pressure controller for 1st stage motor • LAB-300/008-12: PT100 product w/connector
compressor • LAB-300/008-13: PT100 detection
• Pressure controller for 2nd stage motor • LAB-300/008-04: calibrated test jar
compressor • LAB-300/008-041: o-ring for test jar
• Thermostat for 2nd stage activation
• Thermo-switch for each cooling / heating jacket Calibration Tools
• Motor compressors equipped with internal • OilLab 80: calibration decade box – PT100 simulator
overload devices • OilLab 81: set of connectors and cables for cold range

Electrical Supply
• 220V ± 15% / 50 to 60 Hz
• 115V ± 15% / 60 Hz