Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                

User Manual For Medincno

Download as pdf or txt
Download as pdf or txt
You are on page 1of 68

User manual for

medinCNO®
The medinCNO® contains batteries and electrical components. Consequently it cannot be
disposed of in domestic waste but must be collected separately and recycled in
accordance with local regulations. (WEEE)

0483
Classification (according to COUNCIL DIRECTIVE 93/42/EEC Annex IX): IIb

medin Medical Innovations GmbH

Adam-Geisler-Str. 1 – 82140 Olching – Germany

+49 8142-44846-0 – www.medingmbh.com – info@medingmbh.com

Made in Germany

OP_cno Rev 11 Correct as of: 26.05.2017

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 2 von 68
Table of Contents
Table of Contents .............................................................................................................................................. 3
1 Introduction .......................................................................................................................................... 5
1.1 About the instructions for use ......................................................................................................... 5
1.2 General liability conditions .............................................................................................................. 5
2 Description of the Device and Conditions for Use ................................................................................. 6
2.1 Intended purpose ............................................................................................................................ 6
2.2 Operating staff ................................................................................................................................ 6
2.3 Alternative respiration systems ....................................................................................................... 6
2.4 Combination with other devices and ancillary equipment .............................................................. 6
2.5 Technical description of function .................................................................................................... 7
2.5.1 Principles of function .......................................................................................................... 7
2.5.2 Technical data ..................................................................................................................... 7
2.5.3 Overview of the general technical description .................................................................... 8
3 Installation of the medinCNO® – Conditions ........................................................................................ 10
3.1 Mounting ...................................................................................................................................... 10
3.2 Power supply................................................................................................................................. 11
3.2.1 Mains operation ................................................................................................................ 11
3.2.2 Battery operation .............................................................................................................. 11
3.2.3 Power supply unit ............................................................................................................. 12
3.3 Gas feed ........................................................................................................................................ 13
3.4 External alarm interface and data interfaces................................................................................. 14
3.4.1 External alarm ................................................................................................................... 14
3.4.2 USB data interface ............................................................................................................ 15
3.4.3 RS232 interface ................................................................................................................. 16
3.4.4 Output data (live and trend data) ..................................................................................... 17
3.5 Environmental conditions (operation / transport / storage) ......................................................... 20
4 First Use .............................................................................................................................................. 23
4.1 CPAP system components ............................................................................................................. 23
4.2 Connecting the medinCNO® and setting up the CPAP system ........................................................ 24
4.3 Starting up the system .................................................................................................................. 26
5 Operation ............................................................................................................................................ 28
5.1 Operator console .......................................................................................................................... 29
5.2 Display screen ............................................................................................................................... 30
5.3 Basic CPAP settings ....................................................................................................................... 31
5.4 Push .............................................................................................................................................. 32
5.5 O2 flush ......................................................................................................................................... 34
5.6 Backup function ............................................................................................................................ 34
5.7 CPAP mode.................................................................................................................................... 35
5.8 Apnoea CPAP mode ...................................................................................................................... 36
5.9 SNIPPV mode ................................................................................................................................ 38
5.10 Oscillation mode ..................................................................................................................... 41
5.11 Switching between modes ...................................................................................................... 43
5.12 Alarm settings ......................................................................................................................... 44
5.13 Trend ...................................................................................................................................... 45
5.14 Standby ................................................................................................................................... 46
6 Switching Off ....................................................................................................................................... 47

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 3 von 68
6.1 Software switch-off ....................................................................................................................... 47
6.2 Hardware switch-off...................................................................................................................... 47
6.3 Disconnection ............................................................................................................................... 47
6.4 Disposal ......................................................................................................................................... 47
7 Accessories .......................................................................................................................................... 48
7.1 Accessories.................................................................................................................................... 48
7.2 Components of the rest of the CPAP system ................................................................................. 48
8 Alarms and Error Messages ................................................................................................................. 50
8.1 Error messages during system start-up ......................................................................................... 50
8.2 Error messages during operation .................................................................................................. 51
8.3 Emergency programme ................................................................................................................. 55
8.4 Indicator lights .............................................................................................................................. 56
9 Cleaning and Maintenance .................................................................................................................. 57
9.1 Cleaning ........................................................................................................................................ 57
9.2 Maintenance and exchange of the internal batteries .................................................................... 57
9.3 Repairs .......................................................................................................................................... 58
9.4 Device modifications ..................................................................................................................... 58
9.5 Power supply unit (cleaning, maintenance, repairs and modifications) ........................................ 58
9.6 External filter ................................................................................................................................ 59
10 Electromagnetic Compatibility ............................................................................................................ 60
10.1 Electromagnetic transmission ................................................................................................. 60
10.2 Electromagnetic immunity ...................................................................................................... 61
11 Symbols ............................................................................................................................................... 64
12 Abbreviations ...................................................................................................................................... 65
13 History of the Software Versions and Hardware Changes ................................................................... 67

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 4 von 68
1 Introduction

1.1 About the instructions for use

These instructions for use contain information about starting up, using and maintaining the medinCNO ®
(medin-cno). They also contain safety information, describe the device's functions and give an overview of
the ancillary equipment needed.

Users must be very familiar with the information and warnings given in these instructions for use in order to
operate the medinCNO® safely. However, they are no substitute for training.

The instructions and warnings are categorised as follows:

Warning: Warnings must be followed in order to prevent possible serious consequences for the patient or
the operator.

Attention: This indicates hazards that could damage the device or impair its functionality.

Please note: This indicates how the medinCNO® can be used more efficiently.

Keep these instructions for use close to the medinCNO® and easily accessible.

Please contact the dealer or go directly to the manufacturer if you have any questions or comments about
these instructions for use.

Warning:
- Read these instructions for use carefully before using the products, and follow the instructions,
warnings and tips closely.
- The medinCNO® may only be used by individuals with a thorough knowledge of these
instructions for use, in accordance with these instructions for use for the purpose described
u de the se tio headed I te ded pu pose .
- The use of the equipment and consumables used in combination with the medinCNO® (e.g.
nCPAP generators and humidifiers) is described in the relevant instructions for use. Users must
also comply with these instructions for use.

1.2 General liability conditions

The general terms of trade of medin Medical Innovations GmbH (hereinafter referred to as Medin) are
binding.

Medin accepts no responsibility or liability for the safe use of the medinCNO ® if the device is serviced or
repaired by individuals who have not received appropriate specialist training or if an operation that does not
comply with correct use is performed. Medin also accepts no responsibility or liability for the safe use of the
medinCNO® if the function tests and maintenance that are specified are not carried out. The user is
responsible for respecting maintenance and function test intervals.

The service life of the device is defined as 8 years.

Medin accepts no liability for damage caused by failure to comply with these instructions.

Warning:
The medinCNO® may not be modified without permission from the manufacturer and subsequent
appropriate examinations and testing to ensure continued safe use.

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 5 von 68
2 Description of the Device and Conditions for Use

2.1 Intended purpose

The medinCNO® CPAP driver is used in combination with the Medijet® nCPAP generator to administer CPAP
therapy to premature infants and newborns. The medinCNO® must be used under the supervision of expert,
specially trained staff i a li i al setti g, a d the patie t s o ge satu atio ust e o ito ed at the sa e
time.

Warning:
- The medinCNO® is intended for clinical use only.
- The medinCNO® a o l e used hile the patie t s o ge satu atio is ei g o ito ed at
the same time.
- The medinCNO® may only be used by and under the supervision of expert staff who have been
trained in the device.
- For use only in combination with the Medijet® nCPAP generator.
- The medinCNO® is a CPAP device. It is suitable only for patients who can breathe spontaneously
and not for patients who are dependent on artificial respiration.
- The medinCNO® may not come into contact with flammable anaesthesia gases or other
flammable materials while in use.
- The medinCNO® may not be operated in an oxygen-enriched environment.
- US legislation stipulates that this device may only be sold to or on the instructions of a doctor
(for USA: Rx only).

2.2 Operating staff

The medinCNO® must be operated by a doctor or by an individual with the necessary technical skills acting on
a do to s i st u tio s. Co se ue tl , all ope ati g staff ust ha e a detailed k o ledge of these
instructions for use and have undergone training in the device.

2.3 Alternative respiration systems

An alternative CPAP system or respiration system (e.g. bag valve unit) must always be available while the
medinCNO® is in use.

2.4 Combination with other devices and ancillary equipment

The medinCNO® may be combined with devices or consumables (e.g. CPAP ancillary equipment) not
mentioned in these instructions for use only after consulting the manufacturer.

Use of ancillary equipment that is not approved puts the patient at risk and may impair the proper
functioning of the device.

Warning:
If the medinCNO® is connected to another device or system, the requirements of IEC 60601-1:2005 chapter
16 and the requirements with regard to EMC according to IEC 60601-1-2 must be met, otherwise the device
function, electrical safety and EMC of the medinCNO® are not guaranteed.

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 6 von 68
2.5 Technical description of function

2.5.1 Principles of function

The medinCNO® is a CPAP driver that can be used in combination with the Medijet ® nCPAP generator to
administer CPAP therapy. The role of the medinCNO® in this CPAP system is to provide the necessary, possibly
oxygen-enriched drive flow, which is fed to the nCPAP generator Medijet ® via the tubes connected to the
patient and is converted into CPAP pressure within the generator.

An internal electronic blender combines the air and oxygen coming from an external source and administers
the total volume so that the drive flow reaching the patient can be enriched with oxygen as necessary to any
oxygen level between 21% and 100%.

The flow volume of the gas mix within a range of 4 L/min to 15 L/min can be set depending on the mode
selected. According to the flow volume set, the medinCNO ® provides a constant drive flow. This generates a
constant CPAP pressure in the Medijet® CPAP generator, or the medinCNO® assesses the CPAP pressure signal
measured in the Medijet® in order to react to the patient's breathing and trigger a higher flow impulse in the
event of apnoea or synchronously with the patient's inhalation. This triggers a pressure impulse with
elevated CPAP pressure in the Medijet® CPAP generator. The medinCNO® can also trigger oscillation between
5 Hz and 20 Hz in the CPAP pressure applied by Medijet® by means of an oscillating drive flow.

The monitor which is part of the medinCNO® displays the parameters selected by the user, the se so s
readings, the length of time that CPAP pressure has been applied by Medijet ® in the form of a graph, and any
alarms.

The medinCNO® can operate for up to three hours without mains power thanks to its rechargeable battery.

Warning:
The medinCNO® may not be modified without permission from the manufacturer and subsequent
appropriate examinations and testing to ensure continued safe use.

2.5.2 Technical data

Dimensions (L x W x H) 29 x 23.5 x 18.5 cm


Weight: 4.75 kg
Display . – colour, 800 x 480 pixel
Power supply: 100 to 240 V AC / 50 to 60 Hz
Internal battery 14.4 V DC, 3 hours run time, rechargeable
Gas connections: Air supply: 300 to 600 kPa (=3.0 to 6.0 bar)
Oxygen: 300 to 600 kPa (=3.0 to 6.0 bar)
Connector standard: DISS or NIST (as preferred)
Patient flow outlet Dimensions M22 (ED) or F15 (ID)
CPAP pressure meter connection
Luer-type – 4.3 mm ID
Flow: Setting range: 0 L/min to 15 L/min
Working range: 4 L/min to 10 L/min

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 7 von 68
Accuracy: ±1 L/min (in the working range) ±2 L/min (outside of the
working range)
Basis of measurement: DIN 1343 at 0°C, 1013.25 mbar and 0% air
humidity (0/1013)
Oxygen concentration: Setting range: 21% to 100% (in flow working range)
Measurement range: 21% to 100%
Accuracy ± 3% (Vol.)
MLF – 16 or OOM 102 measurement sensor
Pressure measurement: Measurement range 0 mbar to 15 mbar - pressure graph
Accuracy ±1.3 mbar
Verification: Redundant measurement by two sensors
Push (inspiration support by the Setting range
medinCNO®) additional flow during the inspiration push:
Min: 0 L/min
Max: 17.5 L/min minus set basic flow
Duration 200 ms to 2 seconds
Accuracy
± 1 L/min (if the total flow = basic flow + additional
inspiration flow - is within the flow working range)
± 2 L/min (if the total flow = basic flow + additional
inspiration flow - is outside of the flow working range)
Basis of measurement: DIN 1343 at 0°C, 1013.25 mbar and 0% air
humidity (0/1013)
O2 flush Duration: 1 minute
Level: +10, +20, +30 % (vol.) above set target oxygen concentration or
100% oxygen.
Electronic shut-off valve In the event of an error, interrupts the flow supply to the patient and
opens the tubing system to the atmosphere
Mechanical overpressure valve Opening pressure 4 kPa (= 40 mbar)
Operating Time The medinCNO® can be used for continuous, long-term operation up
to 4 weeks without a restart in the interim.
IP class IP20

2.5.3 Overview of the general technical description

Information on these points can be found in the chapters indicated:

 Conditions for use, transport and storage: see chapter 3.5


 Characteristics and accuracies of the device: see chapter 2.5.2
 Information on the installation of the device: see chapter 3
 Description of the supply gases required: see chapter 3.3
 Description of the power supply: see chapter 3.2
 The device is disconnected from the mains by pulling the power supply unit cable of the
medinCNO® out of the electrical outlet. Thus when installing the medinCNO ®, it should be
ensured that this plug is always freely accessible and can be disconnected from the mains

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 8 von 68
without difficulty. The medinCNO® does not contain any switch which disconnects it from
the mains.
 Changes and modifications to the medinCNO® are not permitted without the permission of
the manufacturer: see chapter 9.4
 Repairs and an exchange of parts may only be made by trained, professional service
personnel and only in accordance with the instructions in the Maintenance Manual,
observing warnings in the manual and in these instructions for use.
 The medinCNO® and its power supply unit are not suitable for use in the direct vicinity of
the patient. Only the patient tubes, the Medijet® and the masks and prongs used are used
in the direct vicinity of the patient.

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 9 von 68
3 Installation of the medinCNO® – Conditions

Prior to start-up, the medinCNO® must be securely mounted. In doing so, the following conditions must be
observed:

3.1 Mounting

Figure 3-1: a) Brackets holding the medinCNO® in position; b) Ventilation slits; c) Overpressure valve; d)
Knurled screws

The medinCNO® must always be securely mounted before the device is started up. To do this, the two fixing
brackets on the back of the device (Figure 3-1-a) are inserted into a standard rail (cross-section: 10 x 25 mm)
and the brackets are then fixed in place manually using the knurled screws (Figure 3-1– d). It is important to
ensure that the rear and lateral ventilation slits (Figure 3-1 – b), including prefilters, are not covered and that
the overpressure valve (Figure 3-1– c) on the base of the device is not blocked.

The medinCNO® can be transported when it is not in use. When it is in use, it must be securely mounted.
However, if the medinCNO® becomes damaged during transport or due to rough handling, its function must
be checked by a service technician. It is not allowed to be used until then.

Warning:
- The medinCNO® must always be securely mounted in an upright position throughout the entire
time that it is in use.
- The medinCNO® must not be covered while in use, and none of the openings or ventilation slits
must be blocked.
- The overpressure valve must be kept clear whenever the medinCNO ® is in use, and it must
always hang free in the air.
- A damaged medinCNO® may not be used.

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 10 von 68
3.2 Power supply

3.2.1 Mains operation

The power supply unit (REF 39-113) of the medinCNO® may only be connected to a power supply with the
following properties:

 Voltage: 100 – 240 V~ (alternating current)

 Frequency: 50 – 60 Hz

 Strength of current: at least 1.1 A

The power supply unit of the medinCNO® has a C8 connector for connection to the mains grid. The power line
between the power supply unit of the medinCNO® and the mains grid must therefore meet these conditions:

 Power supply side: plug appropriate for the country

 Device side: C7 plug in accordance with EN60320

 Minimum rated voltage: the minimum rated voltage of the power line must be at least as high as
the voltage of the local mains supply. (Recommended: 240 V)

 Minimum rated current: 1.1 A

 Maximum length 2 metres (in order to achieve EMC in accordance with DIN EN 60601-1-2:2016)

Spare power lines with EU plugs can be ordered under REF 39-115 (supplied as standard as part of the power
supply unit). Alternative power lines for other countries are available under these numbers:

 with UK plug REF 39-116

 with plug for USA/Japan REF 39-117

 with Australian plug REF 39-118

 With Chinese plug REF 39-127

Attention:
The medinCNO® may only be run on the local mains grid using the power supply unit provided by the
manufacturer (REF 39-113) and suitable power lines, otherwise the functionality and EMC of the device
cannot be guaranteed.

3.2.2 Battery operation

The medinCNO® contains an internal battery and can work on a fully charged battery for up to three hours.

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 11 von 68
Figure 3-2: Battery charge status displayed in the top left-hand area of the screen

The charge status of the internal battery is displayed in the top left-hand area of the medinCNO® screen
(Figure 3-2).

A full battery is shown with a symbol with five bars. As soon as only two bars remain, the battery symbol is
shown in red and the medinCNO® should be connected to the mains soon. As soon as there is only one bar
left, the battery alarm is triggered.

Attention:
When the medinCNO® is running on battery power, an alarm is sounded when the internal battery is low. If
this happens it is essential that the device is connected to an external power source without delay.

Please note:

If the medinCNO® battery is completely depleted, it can only be recharged very slowly which greatly increases
the recharging time (up to 5 hours). In addition, the battery will be damaged by this exhaustion and will need
to be replaced sooner. Since the battery of the medinCNO® also discharges while switched off, it is
recommended that the medinCNO® be regularly charged during longer periods of storage.

3.2.3 Power supply unit

The power supply unit of the medinCNO ® may only be used according to its imprinted input and output
power.

Input power: 100 to 240 V AC / 50 to 60 Hz / 1.1 A

Output power: 24 V DC / 2.2A (SELV - Safety Extra Low Voltage).

Disconnect the power supply unit from the mains during thunderstorms.

Warning:
If the medinCNO® is not connected to the mains, the internal battery will also discharge while switched off.
That is, if the medinCNO® is to later be used on rechargeable battery power, it will need to be charged
beforehand.
If the alarm Battery low – Connect power supply unit does not disappear despite connecting the power supply
unit and deleting the alarm text (e.g. by triggering a push notification or a higher-priority alarm (red)), there
may be a defect of the medinCNO®. Contact a service technician and do not use this medinCNO® for the time
being.

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 12 von 68
3.3 Gas feed

In order to operate the medinCNO®, it must be supplied with air and oxygen from an external source that
meets these conditions:

Medicinal oxygen:

- 300 to 600 kPa (=3.0 to 6.0 bar)

- O ge o te t: . %

- H2O pp V/V free from condensate)

Medicinal air (Aer medicalis):

- 300 to 600 kPa (=3.0 to 6.0 bar)

- Oil . g/ 3, ideally oil-free

- H2O pp V/V a d f ee f o o de sate! If as a result of the operating temperature it is


possible that condensate may form despite a relatively low concentration of H2O in the air, then
the concentration of H2O must be reduced accordingly.

Figure 3-3: Water trap in the medinCNO® air inlet; a) Bolt for removing the condensate from the water trap

The medinCNO® can be fitted with a water trap in the air inlet as an option (Figure 3-3) (REF 39-800). This
allows small amounts of condensate to be removed from the air supply provided to the medinCNO ® if the
water trap is checked for condensate and emptied on a regular basis. To empty the water trap, undo the
black bolt (Figure 3-3 – a) on the underside of the water trap and keep it open until the condensate has run
off (this can also be done under pressure).

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 13 von 68
Attention:
The functionality of the medinCNO® can only be guaranteed if both the oxygen and medicinal air supplied to
it are free from condensate water, oil and contaminants. If this is not guaranteed, the medinCNO ® could be
damaged. Even the use of a water trap can only remove small amounts of condensate from the air.

In order to avoid fires, oxygen must not come into contact with oil or grease. It is therefore essential to
ensure that all parts of the medinCNO® and the CPAP system which come into contact with oxygen (e.g. gas
feed tubes, O2 inlet to the medinCNO®, patient outlet, patient tubes and Medijet®) remain free from oil and
grease.

Gas feed connections:

The medinCNO® can be fitted as standard with NIST or DISS gas supply connections as the customer prefers.
The tube kits to connect the device to the external gas source must also meet the selected standard.

3.4 External alarm interface and data interfaces

Attention:
Only devices and systems which are approved for medical use and which meet the relevant electrical safety
and EMC standards may be connected to the medinCNO® data interface and alarm interface.

3.4.1 External alarm

The medinCNO® has an interface for connection to an external alarm system. The interface can open or close
an external current circuit in the event of an alarm. The interface thus acts only as a switch, that is, the power
supply must be ensured by the external alarm system.

Time delay:

The interface to the external alarm system is updated every 100 msec. That is, the alarm in the external
alarm system is activated with a maximum delay of 100 msec in comparison to the alarm in the device.
The alarm is likewise deactivated with a maximum delay of 100 msec in comparison to the alarm in the
device. The silencing of the alarm sound does not count here, but rather the actual deactivation of the alarm
(that is, correction of the alarm condition and acknowledgement of the alarm by the user).

The cable supplied by Medin must be used to connect the medinCNO® to an external alarm system in order to
achieve EMC according to DIN EN 60601-1-2:2016 (REF 39-114) and the connection must be set up initially by
an appropriately trained technician.

Attention:
The medinCNO® may only be connected to an external alarm system using the connection cable provided by
the manufacturer (REF 39-114), otherwise the functionality and EMC of the device cannot be guaranteed.

Warning:
The function of the external alarm interface must be initially activated by a service technician in the service
menu prior to its first use. The medinCNO®s are delivered with a deactivated interface as a standard feature.

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 14 von 68
3.4.2 USB data interface

The USB data interface of the medinCNO® can be used to retrieve internal device data and to install software
updates. In addition, there is the option to prompt an export of the trend data in the trend by pressing the
corresponding buttons. The USB interface thus delivers either live data or trend data which must be received
and saved by a computer or data management system. The data are serial data which are transformed
internally into a USB signal. Their receipt corresponds to that of normal serial signals (such as RS232) with
these settings:

- Data Type: serial

- Baud rate: 115200

- Data bits: 8

- Parity: none

- Stop bits: 1

- Handshaking: none

To connect the medinCNO® to a system for archiving and processing patient data, a standard USB cable (USB
2.0 type A to USB 2.0 type B) with double winding through a ferrite of the type: 742-711-32 from the
manufacturer Würth Elektronik can be used (see Figure 3-4: USB and RS232 cable with ferrite). Nonetheless,
the function of the medinCNO® must be observed to avoid disruptions by the entire system with regard to
electromagnetic influences.

Attention:
The data interface of the medinCNO® may only be connected with the cable specified above, including ferrite
core of the type: 742-711-32 from the manufacturer Würth Elektronik with double winding and a computer
system that meets the medically relevant requirements in terms of electrical safety and EMC, otherwise the
functionality and EMC of the medinCNO® cannot be guaranteed.
In doing so, only the following ferrite core may be used: Würth Elektronik - Type: 742-711-32.

The cable used in each case (USB cable or optional RS232 cable) must thus contain a ferrite (REF 39-125;
Würth Elektronik - Type: 742-711-32, gray or black) with a double wound cable in accordance with Figure 3-4.

a) b) c) d)

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 15 von 68
Figure 3-4: USB and RS232 cable with ferrite
a) RS232 cable without ferrite b) RS232 cable with a ferrite that is still open whereby the cable
must be carried through the core twice, that is, wrap the ferrite core once. c) RS232 cable with
ferrite d) USB cable with ferrite (the cable must also be carried through the ferrite core twice as
in the case of RS232 cable)

3.4.3 RS232 interface

The RS232 interface can be installed optionally. The function, data and settings are identical to the USB
interface. A ferrite (Würth Elektronik - Type: 742-711-32, gray or black; REF 39-125) with a double-wound
cable must likewise be present on the RS232 cable identical to the USB cable (see chapter 3.4.2).

Attention:
The medinCNO® data interface must be connected using the cable, including ferrite core (REF 39-125)
specified above (see Figure 3-4) and a computer system that meets the medically relevant requirements in
terms of electrical safety and EMC, otherwise the functionality and EMC of the medinCNO ® cannot be
guaranteed.

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 16 von 68
3.4.4 Output data (live and trend data)

3.4.4.1 Live data


The live data of the medinCNO® contain the patient pressure, the set flow, the measured oxygen
concentration, whether a push is currently being emitted, all alarms, the set additional push flow and the
current mode. In addition, date, time, device name and software version are regularly indicated. The live data
are automatically and constantly indicated in normal operation. However, they are not saved in the
medinCNO®. If the live data are of interest, they must be recorded and saved by a computer or data system or
the data will be lost. The exact structure of the live data is described below:

Every 10 min., a line with these data is indicated:

- ***;
- Date;
- Time;
- Device name (can be set in the service menu);
- VXXX (CPU software version);

In the interim, the following data are provided every 0.1 sec (separated by ; ) via the USB interface:

- The currently measured CPAP pressure (in the CPAP, Apnoea CPAP and SNIPPV mode) or the CPAP
average value (in oscillation mode) in XX.X mbar

- Set target flow in XX.X L/min for the basic CPAP level; in the case of additional activation of the
leak assist, the target flow + correction flow is emitted.

- Measured oxygen concentration FiO2 XXX%;

- Current push activity X: 2 = Backup actively running (independent of whether there is currently a
push or not), 1 = push active (without backup), 0 = no push active;

- High-priority alarms (active alarms are indicated by a bit set at the corresponding site). The sum of
all active alarms XXXX is given (calculation in hexadecimal):

 0x0001: Disconnection
 0x0002: CPAP pressure high (fixed)
 0x0004: Feed pressure high
 0x0008: Gas feed oxygen
 0x0010: Gas feed air
 0x0020: Device error patient pressure measurement
 0x0040: Flow stopped, reactivation via alarm reset
 0x0100: Device error restart necessary

- Medium-priority alarms XXXX:

 0x0001: Battery almost empty – Connect power supply unit


 0x0002: CPAP pressure high (adjustable)
 0x0004: CPAP pressure low (adjustable)
 0x0008: Oxygen concentration high
 0x0010: Oxygen concentration low
 0x0020: Internal temperature too high

- Low-priority alarms XXXX:

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 17 von 68
 0x0001: Fan blocked
 0x0004: Leak-Assist

- Additional flow set during the push XX.X L/min

- Mode X:

 C = CPAP
 A = Apnoea CPAP
 S = SNIPPV
 O = Oscillation

Example of live data:

***;2014-04-04;14:09:11;medin-cno1004;V1.3;
04.1;06.0;039;0;0000;0008;0000;03.7;O
04.1;06.0;039;0;0000;0008;0000;03.7;O

04.2;06.0;039;0;0000;0008;0000;03.7;O
04.2;06.0;039;0;0000;0008;0000;03.7;O

07.0;08.7;021;0;0000;0000;0000;03.8;C
07.9;08.7;021;0;0000;0000;0000;03.8;C
07.9;08.7;021;0;0000;0000;0000;03.8;C
07.7;08.7;021;0;0000;0000;0000;03.8;C
07.5;08.7;021;0;0000;0000;0000;03.8;C
06.7;08.7;021;0;0000;0000;0000;03.8;C

3.4.4.2 Trend data


The trend data are divided into two parts: The first part contains logfiles of saved alarms and device starts.
This recording contains the data of the last device uses and will continue to be saved even if the medinCNO ®
is switched off. The second part of the trend data is deleted each time the medinCNO ® is switched off. It
saves the data since the last device start for a maximum of 28 days. It contains the CPAP pressure, the set
flow, the measured oxygen concentration, the set additional push flow, the mode, respiratory rate, whether
the backup was active, the pushrate R insp, the minimum and maximum possible push rate, the maximum
pressure during the pushes, the oscillation frequency, the set level of the oscillation amplitude, the resultant
pressure of the oscillation amplitude, the set inspiration time and the set apnoea time.

The structure of the trend data is shown below as an example:

###;2014-04-07;14:20:14;medin-cno1004;V1.3;alarm record;
start;2014-04-04;10:59;
alarm;04;11:07;0000;0008;0000;for 14:48
alarm;04;13:50;0000;0004;0000;for 00:30
alarm;04;14:14;0002;0002;0000;for 00:36
###;2014-04-07;14:20:15;medin-cno1004;V1.3;trend record;
CPAP;Flow;FiO2;Finsp;Mode;RR;BU;Rinsp;min;max;Pinsp;Freq;AmpS;AmpP;Tinsp;Tapn;
05.6;07.0;021;00.0;C;018; ; ; ; ; ; ; ; ; ; ;
05.6;07.0;021;00.0;C;018; ; ; ; ; ; ; ; ; ; ;

5.5;07.0;024;01.0;A;024;-;000; ; ;00.0; ; ; ;0.3;10.0;
5.9;07.5;024;01.0;A;018;-;000; ; ;00.0; ; ; ;0.3;10.0;

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 18 von 68

05.8;08.0;021;02.6;S; ;B;070;25;120;10.6; ; ; ;0.2;02.2;
07.3;08.0;021;02.6;S; ;B;070;25;120;11.2; ; ; ;0.2;02.2;

02.5;06.0;026;02.5;O; ; ; ; ; ; ;07;10;05.4; ; ;
04.1;06.0;036;03.7;O; ; ; ; ; ; ;14;06;04.1; ; ;
04.1;06.0;039;03.7;O; ; ; ; ; ; ;19;06;04.9; ; ;
***;2014-04-07;14:22:51;medin-cno1004;record complete;

Start of alarm recording:

- ###;
- Date: YYYY-MM-DD; (current date, set in the service menu)
- Time: HH:MM:SS; (current time, set in the service menu,
the switch to local time or summer
time must
be performed manually)
- Device name: medin-cnoZZZZZ (set in the service menu)
- SW version: VXXX (installed main software version)
- Alarm record;

Start and alarm recording:


- For each device start, date and time are saved as such:
start; YYYY-MM-DD;HH:MM;

- For each new alarm, day, time, all active alarms of high, medium and low priority and the
duration (min:sec) of the alarms are saved:
alarm; DD; HH:MM;HHHH;MMMM;LLLL;for MM:SS

The alarm coding takes place analogously to the coding in the live data (for detailed information, see chapter
3.4.4.1). The date is only shown as a day. The month can be determined using the last prior start entry. The
alarm duration is indicated in minutes and seconds from the start of the alarm until its end.

Recording the set and measured values:

The trend data contain all values saved in the trend and additional values since the last time the device was
switched on. Every 20 seconds, a line is saved. If a value is not available in the current mode, it is replaced
with blank spaces.

The structure is as follows:

- Average value of the CPAP pressure in XX.X mbar


- Set flow in XX.X L/min + if applicable, correction flow with activated leak assist
- Measured FiO2 in XXX%
- Set, additional push flow in XX.X L/min
- Current mode: C = CPAP, A = Apnoea CPAP, S = SNIPPV, O = Oscillation, - = Standby
- Measured respiration rate in XXX / min
- Status of backup (B = actively running; -= currently not actively running or generally off )
- Push frequency Rinsp in XXX / min

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 19 von 68
- Minimum reachable push frequency R insp in XX / min caused by apnoea
- Maximum reachable push frequency R insp in XX / min caused by triggering synchronised with
respiration
- Maximum CPAP value during the push in XX.X mbar
- Set oscillation frequency in XX Hz
- Set amplitude level of oscillation in level XX
- Measured amplitude during the oscillation in XX.X mbar
- Set push length Tinsp in X.X sec
- Set apnoea time in XX.X sec

Please note:

When reading the trend data, there may be time delays upon receipt at the computer or data system. For
this reason, when reading the data, wait until the last line (***;2014-04-07;14:22:51;medin-
cno1004;record complete;) has been issued. Only then have the trend data been fully received.

3.5 Environmental conditions (operation / transport / storage)

The medinCNO® must not be used in rooms at risk of explosion, near to flammable substances or close to
locations exposed to splash water.

Warning:
The medinCNO® is not suitable for use near magnetic resonance imaging equipment (MR-unsafe) or
electrosurgical equipment.

During operation, the medinCNO® should not be used in the direct vicinity of other electronic devices or
stacked. If use as described above is nonetheless necessary, this device and other devices should be observed
to ensure that they are working properly.

Environmental conditions for the medinCNO® during operation:

Temperature: 15°C to 35°C

Relative air humidity: Air humidity 20% to 80% (not condensing)

Moisture: The device is to be kept dry during operation.

Ambient pressure: 700 hPa to 1100 hPa

Oxygen environment: O2 content <25%

Altitude: a o e sea le el

Cleanliness: The hygiene regulations valid for the hospital are to be observed. In
addition, the medinCNO® is to be operated in standard hospital ambient
conditions.

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 20 von 68
Environmental conditions of the power supply unit during operation:

Temperature: - between 0°C and 40°C


- do not use power supply unit in locations exposed to significant
temperature fluctuations.

Moisture: - use only in dry locations


- relative humidity %, o -condensing
- do not use in locations exposed to significant amounts of moisture or
condensate
- do not use in locations exposed to significant environmental stress
- do not use outdoors

Vibrations: do not use in locations exposed to constant vibration

Environmental conditions for the medinCNO® during transport/short-term storage:

Temperature: -20°C to 50°C (longer-term storage only at room temperature)

Relative air humidity: 20% to 80% (not condensing)

Moisture: During transport and storage, the medinCNO® must be protected from
wet conditions.

Cleanliness: The medinCNO® must be protected from contamination during


transport and storage.

Attention:
If the medinCNO® is e posed to te pe atu es of °C du i g t a spo t, it ust e a climatised at room
temperature for at least 24 hours at the location where it is to be used before being switched on and used for
the first time. If this is not done, the condensate which has formed could damage the medinCNO ®.

The medinCNO® may only be transported in the original packaging, since it contains internal lithium batteries
which may only be transported under special conditions. If the medinCNO ® contains defective batteries,
these may not be transported and must be removed in advance and disposed of in accordance with local
regulations. (See separate instructions on transporting the medinCNO® and the batteries it contains).

Long-term storage of the medinCNO®:

Temperature: Room temperature (approximately 20°C)

Relative air humidity: 20% to 80% (not condensing)

Moisture: During storage, the medinCNO® must be protected from wet conditions.

Cleanliness: The medinCNO® must be protected from contamination during storage.

Please note:

Long-term storage must take place under room temperature conditions, since the batteries may become
damaged otherwise and need to be replaced sooner.

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 21 von 68
Environmental conditions of the power supply unit during transport and storage:

Temperature: not in locations exposed to significant temperature fluctuations.

Moisture: - only in dry locations


- not in locations exposed to significant amounts of moisture or
condensate
- not in locations exposed to significant environmental stress
- not outdoors

Vibrations: not in locations exposed to constant vibration

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 22 von 68
4 First Use

The CPAP system must be set up and the device checked before starting any CPAP therapy using the
medinCNO®. This check starts automatically whenever the device is started up.

The device should not be connected to a patient during set-up and system start. In addition, during set-up,
system start and operation, the user should not simultaneously touch the patient and the device. This applies
above all to the power supply, the alarm, RS232 or USB interface and the accessible metal parts of the
pneumatic unit (gas connections, overpressure valve and mounting screws).

Warning:
If errors are detected when the medinCNO® is checked before first use, if the medinCNO® is found to be
damaged or does not behave as expected or as intended in the user manual, it must not be connected to a
patient under any circumstances.

The power supply unit should not be used if it has visible damage to the housing or cord.

The power supply unit should not be used in the direct vicinity of the patient since there may be
temperatures of >71°C.

4.1 CPAP system components

A CPAP system based on a medinCNO® comprises the following components:

- A CPAP driver: the medinCNO®

- A Medijet® nCPAP generator together with a suitable prong or mask

- An active respiratory gas humidifier to humidify the respiratory gas (recommended)

- A patient tube circuit set that fits the patient gas outlet, CPAP pressure meter inlet of the
medinCNO®, the Medijet® and the humidifier selected

- A bonnet of the correct size

- A s ste fo o ito i g the patie t s o ge satu atio a d he ki g the effi a of CPAP


therapy (this is an essential condition for the use of the medinCNO ®)

Please note:

The medinCNO® should always be used in combination with an active respiratory gas humidifier in order to
a oid the patie t s ai a s d i g out a d ooli g.

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 23 von 68
4.2 Connecting the medinCNO® and setting up the CPAP system

Figure 4-1: Connections of the medinCNO®: a) Gas feed oxygen; b) Gas feed air;
c) Mains power supply (power supply unit); d) Patient gas outlet; e) CPAP pressure meter line inlet,
f) Overpressure valve, g) USB interface, h) RS232 interface (optional), i) External alarm

Gas feed connection:

The oxygen inlet "O2-IN" (Figure 4-1 -a) and air inlet "Air-IN" (Figure 4-1 -b) of the medinCNO® must be
connected to an external air and oxygen source using gas feed tubes before the device is switched on. The air
and oxygen must comply with the quality requirements laid down in chapter 3.3.

Attention:
The medicinal air and oxygen supplied to the medinCNO® must meet the quality requirements laid down in
chapter 3.3 as otherwise the medinCNO® could be damaged.

Power supply connection:

The power connection of the medinCNO® "Power" (Figure 4-1-c) must be connected to the mains by means of
the power supply unit for the device. Users are recommended to keep the medinCNO ® connected to the
mains at all times while it is in use. The internal rechargeable battery in the medinCNO® allows the device to
keep working in the event of a brief interruption in mains power supply. The medinCNO ® can run on a fully
charged battery for up to approximately three hours. The green LED of the power supply unit shows the
operating status of the power adapter.

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 24 von 68
Figure 4-2: Battery charge status displayed in the top left-hand area of the screen

The charge status of the internal battery is displayed in the top left-hand area of the medinCNO®screen.

Attention:
When the medinCNO® is running on battery power, an alarm is sounded when the internal battery is low. It
must then be immediately connected to an external power source, as otherwise the medinCNO® will stop
working.

To disconnect the medinCNO® from the mains, the plug of the power supply unit of the medinCNO ® must be
pulled out of the electrical outlet. Otherwise the medinCNO ® is still connected to the mains, even when it is
turned off.

Therefore, when setting up the medinCNO ®, it should be ensured that the plug of the power supply unit is still
accessible even when set up and the medinCNO® can thus be easily disconnected from the mains.

Disconnect the power supply unit from the mains during thunderstorms.

Warning:
Since the surface temperature of the power supply unit can become too hot for the patient, it should always
be ensured that the power supply unit is not within the patient's reach.

The plug of the power supply unit of the medinCNO ® must also always be freely accessible even when set up,
in order to allow the medinCNO® to be quickly disconnected from the mains in the event of a hazardous
situation.

Do not insert the power supply plug of the medinCNO® into ceiling outlets since, due to its weight, it can be
pulled out of the outlet in this case.

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 25 von 68
Connecting the patient tube circuit set:

Figure 4-3: Set-up of a CPAP system based on the medinCNO®

The patient tube circuit set is used to connect the patient to the medinCNO ®. This involves inserting the
humidifier chamber into the humidifier and then connecting it to the "Flow out" patient gas outlet via the
shorter tube (Figure 4-1- d). The longer tube connects the humidifier chamber to the Medijet ®. The humidifier
temperature probes and heating cable adapters must be inserted into this tube if required. The thin pressure
meter tube connects the pressure meter connection of the Medijet ® to the CPAP pressure meter inlet
"Patient Pressure" (Figure 4-1 – e) of the medinCNO®.

Warning:
The use of the Medijet®, the humidifier and the patient tube kits is described in the user manuals for these
devices, along with important additional information and warnings regarding their use. Users must also know
and comply with these other user manuals even though they may be familiar with the medinCNO ® user
manual.

Only suitable tube systems with appropriate adapters should be used to connect the patient and the Medijet®
to the medinCNO®, since otherwise there will be an increased likelihood of disconnections.

4.3 Starting up the system

The medinCNO® is started and its system test is automatically triggered by pressing the "On/Off" button
(Figure 5-1 – a). This test examines the functions of the medinCNO ®, checks the oxygen calibration and
determines the properties of the tube system to which it is connected.

Make sure that no patient is connected to the medinCNO® throughout the entire duration of the system test,
a d p ess OK to o fi this.

If an error is detected during the system test, this is indicated by an error message. See chapter 8.1 for the
meaning of the individual error messages.

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 26 von 68
The function of the mechanical overpressure valve must also be checked during the system test. This is done
by briefly blocking the patient gas outlet (Figure 4-1 - d) and observing during this time whether the
overpressure valve (Figure 4-1 - f) opens and closes.

In order to determine the characteristic flow CPAP pressure course of the patient tube system used and to
check the CPAP pressure sensors, the entire tube system and the Medijet® must now be connected to the
medinCNO® and the Medijet® prong adapter must be closed. Once this has been done, the process of
easu i g the tu e s ste a e sta ted p essi g the OK utto .

If oxygen calibration is necessary, this is carried out automatically.

If the system test is completed without errors, the medinCNO ® starts up in CPAP mode and can be used on
the patient.

Warning:
- The system start test must always be conducted and completed before connecting the
medinCNO® to a patient.
- No patient may be connected to the medinCNO® during the system test.
- If an error is detected during the system test, the device must not be used or connected to a
patient under any circumstances.

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 27 von 68
5 Operation

The medinCNO® can be used to administer pure CPAP therapy, push-supported CPAP therapy or oscillating
CPAP therapy. All versions are based on pure CPAP therapy and the push or modulated oscillation options are
added if required. The medinCNO® offers four CPAP modes:

- CPAP: pure CPAP mode in which a constant CPAP pressure in the Medijet ® is generated by a
constant flow. This mode operates without automatic pushes. However, operators can trigger
manual pushes if necessary.
In addition, the leak assist can be switched on (see chapter 5.7).

- Apnoea CPAP: In this mode, the basic CPAP mode described under CPAP is combined with apnoea
monitoring. The medinCNO® assesses the CPAP pressure signal and recognises from this
information whether or not respiratory activity is taking place. If no respiratory activity can be
detected for a length of time that can be set, the medinCNO ® automatically triggers a push to
stimulate the patient.
In addition, the backup function can be used (see chapter 5.6) and the special case NIPPV can be
activated (see chapter 5.8).

- SNIPPV: (=synchronised non-invasive positive pressure ventilation). In the SNIPPV mode, the
medinCNO® has the capability not only of automatically triggering pushes during the apnoea
phases but also of synchronising them with the patient's inspiration. This means that the patient
can be stimulated by automatic pushes both during apnoea and when he or she is breathing
normally. In addition, it is possible to use the backup function.
Nevertheless, the SNIPPV mode is only a CPAP mode and is not a substitute for ventilator
respiratory support by a mechanical ventilator that the patient might need.

- Oscillation: The oscillation mode offers a Medijet® pressure of varying frequency within a range of
5 to 20 Hz instead of constant CPAP pressure.

The fifth mode, Standby, can be used for the time between preparation of the medinCNO® and its actual use
as well as for pauses in the CPAP therapy for a patient.

Some operating steps and settings are common to all medinCNO ® modes and are transferred when switching
from one mode to another, but some settings are specific to particular modes.

Warning:
- The medinCNO® is intended for clinical use only.
- The medinCNO® a o l e used hile the patie t s o ge satu atio is ei g o ito ed at
the same time.
- The medinCNO® may only be used by and under the supervision of expert staff who have been
trained in the device.
- For use only in combination with the Medijet® nCPAP generator.
- The medinCNO® is a CPAP device. It is suitable only for patients who can breathe spontaneously
and not for patients who are dependent on artificial respiration.
- Ensure that resuscitation equipment is always available while the medinCNO ® is in use.
- A reliable, adequate power supply (mains or battery) must be available throughout the entire
duration of medinCNO® operation.
- The medinCNO® must not be switched off while it is connected to a patient.
- If unusual or inexplicable events occur while the medinCNO ® is in use, or if you hear unusual
noises or observe alarm messages that cannot be switched off, shut down the medinCNO ® and
contact a service technician.

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 28 von 68
5.1 Operator console

Figure 5-1: medinCNO® controls: a) ON/OFF; b) Buttons for entering basic CPAP parameters; c) Dial for use in
changing and confirming settings; d) Trigger button for manual pushes; e) Alarm reset button; f) Button for
mode-specific and other settings

The medinCNO® is operated by means of several buttons and a dial. The usual procedure is to press a button
to select the parameter indicated next to that button on the display, then use the dial to select the required
value. This is confirmed by pressing the dial.

The various buttons are divided into several groups:

- ON/OFF: (Figure 5-1–a) to switch the medinCNO® on and off

- Basic CPAP settings: (Figure 5-1 –b) to select and adjust the two basic CPAP parameters of oxygen
concentration (FiO2) and Flow. In addition, to set the additional flow for pushes (Finsp) and to
trigger an O2 Flush.

- Push button (PUSH): (Figure 5-1 –d) to trigger a push manually and immediately

- Alarm reset: (Figure 5-1 –e) to confirm and briefly acknowledge alarms. By pressing this button,
the current alarm sounds are suppressed for 120 seconds and then reactivated. However, if a new
alarm becomes active during this time, this will also be emitted acoustically.

- Other settings: (Figure 5-1 –f) to select and alter mode-specific and other settings

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 29 von 68
5.2 Display screen

Figure 5-2: Display and button assignment of the medinCNO® a) Flow; b) FiO2 value or start and
discontinuation of the O2 flush; c) Additional push flow; d) Manual push trigger button; e) Readings; f) Mode;
g) Status displays;
h) Pressure course graph; i) Alarm messages; k) Alarm reset button; m) to r) Buttons for other settings

The medinCNO® display screen shows the values set, the readings, status messages and the names of the
buttons surrounding the screen:

- Mode: displays the selected mode - position f in Figure 5-2

- Basic CPAP settings: the target values set for flow, FiO2 and the additional push flow selected are
shown on the right of the display screen - Figure 5-2 position a (Flow), b (FiO2) and c (additional
push flow - Finsp) After pressing button b twice, the setting of the O2 flush and its start button are
indicated instead of the set target value for FiO2.

- Manual push: the manual push trigger button, position d in Figure 5-2.

- Readings: the values measured by the medinCNO® for the average CPAP pressure (CPAP) and the
measured oxygen concentration (FiO2). In addition, mode-dependent readings such as the
respiratory (RR) or push frequencies (Rinsp - corresponds to the number of inspiration pushes per
minute automatically emitted by the medinCNO®) are shown above in the coloured bar of the
medinCNO® (Figure 5-2 position e)

- Status displays: the medinCNO® displays battery status and interrupted alarm sounds at position g
in Figure 5-2 .

- Pressure course curve: the CPAP pressure measured over time is shown as a graph in position h in
Figure 5-2 . In addition, the maximum measured pressure for each push is marked numerically.
The representation of the pressure course can be helpful in selecting CPAP parameter settings.
The pressure curve can be indicated on a scale of 0-10, 0-15 or 0-20 mbar/cmH2O.

- Active alarms: information about active alarms is given in position i in Figure 5-2. If several alarms
are active at the same time, they are displayed in turn. See chapter 8 for a detailed description of
alarms and tips on how to avoid them occurring.

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 30 von 68
- Button assignment: the meaning of adjacent buttons is defined in fields k to r in Figure 5-2 . Field
k and its button are always used to acknowledge alarms. In some menus, two different
parameters are assigned to a single button in fields n to r. In this case, the operator can press the
relevant button several times to switch between the two parameters.

5.3 Basic CPAP settings

The CPAP therapy generated in the Medijet® is largely determined by two factors: the flow volume (Flow)
emitted by the CPAP driver and the oxygen concentration (FiO2) contained in this. These two fundamental
factors are set in the same way in all modes and are transferred when the operator switches from one mode
to another.

Flow:

Press button a as shown in Figure 5-2 to change the flow value that has been selected. This action activates
the flow setting (a blue background appears). The target value can then be changed by turning the dial. As
this also changes the flow volume produced by the medinCNO®, the operator can see the effect of this
change on the CPAP pressure produced as shown in graph h in Figure 5-2. The operator can terminate
activation, and thereby the ability to change the flow value, by pressing the dial. If the dial is not pressed to
conclude the flow change, the medinCNO® still accepts the flow volume change.

If the flow value is altered in situations when no patient is connected to the Medijet ®, the Medijet® prong
adapter must be closed by the operator during this process (e.g. by holding it shut). This simulates the CPAP
pressure which is produced. If a patient is connected to the Medijet ® during flow adjustment, the prong or
mask must be checked for leaks before any adjustment is made as the presence of leaks could affect the
CPAP pressure which is produced.

The medinCNO® automatically produces a flow of 7 L/min on start-up. The user must adjust this flow while
keeping the Medijet® prong adapter closed until the flow produces the desired CPAP pressure.

Warning:
Since the set basic flow (Flow) and additionally the set push flow (Finsp) are emitted during a push, an increase
in the basic flow also increases the CPAP pressure during a push. Therefore, the setting for Finsp must always
be monitored when the Flow setting is changed. (Adjustment according to chapter 5.4)

Please note:

The mask or prong should always be checked for leaks before adjusting the flow that has been selected, as
the presence of leaks affects the CPAP pressure which is produced.

If the leak assist function is activated in CPAP mode, it is automatically deactivated by adjusting the flow. If it
is once again desired, it must then be manually restarted (see chapter 5.7).

FiO2 oxygen concentration:

Alter the oxygen concentration by using button b in Figure 5-1 to highlight the current oxygen concentration
and then turning the dial to select the desired value. Press the dial to confirm the new figure. This completes
the process of adjusting the oxygen concentration. It is only now that the medinCNO® alters the oxygen
concentration in the patient flow. If the oxygen concentration adjustment is not confirmed by pressing the
dial, the medinCNO® continues to blend the oxygen concentration that was set before the target value was
adjusted.

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 31 von 68
Depending on the FiO2 alarm preset in the service menu, the change in the oxygen concentration affects the
limits of the oxygen alarm:

- FiO2 alarm automatic: At the same time as the change in the oxygen concentration, the limits for
the oxygen alarm are also adapted to the new oxygen concentration.

- FiO2 alarm manual: The change in the target oxygen concentration does not affect the limits of
the oxygen alarm. After changing the oxygen concentration, these limits have the same values as
before the change. They can only be adjusted manually to the new oxygen values. For this
purpose, the user must open the Alarm settings menu and manually update the alarm
thresholds (see chapter 5.12).

Please note:

The oxygen concentration in the patient flow does not change until the new oxygen concentration has been
confirmed. Only then can the effect of the changed oxygen concentration on the patient be observed.

As long as an O2 flush is active, the target oxygen concentration FiO2 cannot be changed. Only after a manual
premature end or after an automatic end of the O 2 flush can the FiO2 setting be changed once again (see
chapter 5.5).

5.4 Push

In addition to the basic CPAP mode, the medinCNO® can increase the CPAP pressure if necessary for a short
ti e. These sho t pe iods a e alled pushes a d a e used to sti ulate the patie t. A push is dete i ed
by the additional flow used for it and its duration.

Additional flow during a push - Finsp:

In field c in Figure 5-1, set an additional flow Finsp by pressing the button to activate the setting, turning the
dial to select a value which immediately becomes active for the next push and then pressing the dial to
confirm. If the dial is not pushed, the change is made nonetheless. As flow during a push, the medinCNO ®
emits the sum from the basic flow Flow set in Figure 5-1 field a and the additional push flow Finsp.

As an additional push flow Finsp, a value between 0 L/min (that is, essentially no push is triggered) and a
maximum value dependent on the basic flow set can be set. The maximum value is 17.5 L/min minus the set
basic flow Flow.

Warning:
A push does not generate a breath; instead it can only stimulate but not ventilate!

The setting Finsp may only be changed in small increments. The effect of each small increment (= CPAP
pressure that develops during a push) must be monitored by triggering a manual push or by waiting for an
automatic push. It should be noted that any change becomes immediately active for the next push, even
without pressing the dial.
The CPAP pressure that develops during a push depends on the leakage of the CPAP system (particularly the
leakage from prongs or the mask). An increase in the push flow by +1 L/min can cause an increase of up to
3.5 mbar in the CPAP pressure.

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 32 von 68
Since leakage from the mask and prongs causes a decrease in the pressure during a push, this leakage must
be checked before increasing the push flow. An increase in the push flow is not a substitute for leak-proof
positioning of the mask or prong!
During a push, the set basic flow (Flow) and additionally the set push flow (Finsp) are emitted such that an
increase in the basic flow also causes an increase in the CPAP pressure during a push.
When setting the additional flow during a push, it should be ensured that:
 The CPAP pressure high alarm sounds and briefly stops the flow emission to the patient starting
from a CPAP pressure of 18 mbar. That is, if the push is set such that its pressure is very high, a
slight change in the leakage or expiration against push by the patient will cause the medinCNO®
to briefly stop the flow and trigger an alarm. To prevent this, the push flow must be reduced.
 If a very high overall flow is set in the event of a push, high dynamic pressure may develop in the
tube system. If this exceeds 40 mbar, the Feed pressure high alarm is triggered and the flow
emission to the patient is stopped. In this case, the push flow set must be reduced and the
leakage at the patient reduced. Depending on the duration of the high feed pressure, the
medinCNO® automatically restarts the flow after a short time or not until after confirmation
from the user (if Press alarm reset to restart flow is displayed).
 During a push, if a flow is set that is so high that it triggers the Feed pressure high alarm and
simultaneously a CPAP pressure high alarm, it is possible that the text Press alarm reset to start
flow may also be displayed. In this case, the medinCNO® stops the flow until the user restarts it
by pressing the Alarm Reset button.

Please note:

In order to be able to use the manual push function at all times, the level of the additional push flow Finsp
should be set in advance (e.g. at the start of the CPAP therapy) and adjusted as needed in the event of flow
changes.

Factors that can exert an influence on the adjusting CPAP pressure during a push:

- Patie t s espi atio : I a si ila a to the p essu e flu tuatio s that a e aused the
patie t s eathi g du i g CPAP the ap a d a e displa ed o the CPAP p essu e urve along
ith othe pa a ete s, the patie t s espi ato a ti it has a i flue e o the a tual le el of
push which is achieved.
- Tube system volume and push level: The volume of air in the tube system must be compressed
before the increased flow triggered by the medinCNO® causes the CPAP pressure in the Medijet®
to rise. If large tube systems are used, or if the difference between the basic flow and the total
flow during the push is considerable, it will take some time to reach the desired CPAP pressure. If
the duration of the push is shorter than this time lag, it will not be possible to achieve the
desired push pressure.
- Changes to the CPAP system and changes in leakage conditions: Since the user determines the
necessary amount of additional flow during a push (Finsp), taking into account any leakage
situation present at this point in time, the level of the push changes when the leakage changes. If
the pressure values during a push are too low, check for leaks and resolve them. Decreases in
the leakage, changes to the Medijet® and tube system or temperature effects may cause
increased pressure values during a push.

Triggering a manual push

A push can be triggered manually in all medinCNO® modes. This is simply done by pressing button d in Figure
5-1. The medinCNO® responds to this by immediately triggering a push. The duration of the push triggered
depends on the current mode: The push duration set in the SNIPPV and Apnoea CPAP mode is triggered. The
set push duration corresponds to the set duration of the inspiration. In CPAP and oscillation mode, pressing

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 33 von 68
the button triggers a push that lasts for 500 milliseconds. If the button is held down, no additional push is
triggered after the first one has been completed. Another push can only be triggered by releasing the push
button and pressing it again. A condition for triggering a manual push is that an Finsp push flow must be set in
advance.

5.5 O2 flush

The O2 flush offers the possibility of increasing the oxygen concentration emitted for one minute. Then the
set target oxygen concentration is automatically emitted once again.

Setting the level:

The automatic duration of the O2 flush is 1 min. However, it can be discontinued early by pressing the button.
The level of the O2 flush can be changed by the user in the respective Settings mode. It is opened by pressing
the p button in Figure 5-2. Here the setting of the level of the O2 flush can now be activated by pressing the r
button in Figure 5-2. The following are available for selection: +10, +20, +30 and 100, and this can be selected
by turning the dial. A final pressing of the dial confirms the new setting which takes effect as of the next
newly started O2 flush. O2 flushes already in progress will continue to be performed with the old setting. The
selected level of the O2 flush determines the emitted oxygen concentration during an O 2 flush: Thus, for
example, during an O2 flush, an oxygen concentration should be emitted which is 10, 20 or 30 vol. % greater
than the set target value of the oxygen concentration (affects setting +10, +20, +30). Given a target oxygen
concentration of 25% and an O2 flush with the setting +20, an oxygen content of 45% would be emitted
during the O2 flush. However, a maximum of 100% oxygen can be emitted. The setting 100 of the O 2 flush
causes pure oxygen (100%) to always be emitted during the O2 flush.

Start of an O2 flush:

The start button for an O2 flush is located behind the FiO2 button and can be selected by pressing this button
twice (button b in Figure 5-2). If the start button has a blue background, pressing the dial can start the O2
flush. Now the medinCNO® automatically emits an increased oxygen concentration for one minute.

Discontinuation of an O2 flush:

As long as an O2 flush is active, the O2 flush discontinuation button can be found at position b in Figure 5-2
instead of the FiO2 setting button. If this is selected by pressing button b in Figure 5-2 and then confirmed by
pressing the dial, the O2 flush is immediately discontinued. If, on the other hand, the O 2 flush is not
discontinued early by the user, it ends automatically after one minute and the button FiO2 target value is
displayed once again.

Please note:

As long as an O2 flush is active, the target oxygen concentration FiO 2 cannot be changed. Only after a manual
premature end or after an automatic end of the O2 flush can the FiO2 setting be changed once again.

5.6 Backup function

The Backup function can only be activated in the Apnoea CPAP and SNIPPV modes. Under the respective
mode settings, the Backup function can be switched on or off and the desired Backup Rate can be set.

Setting the backup rate is possible in intervals of 5 between 5/min and 120/min. The rate setting depends on
the inspiration time. The greater the inspiration time, the smaller the maximum backup rate.
The push alarm cannot be deactivated during the active backup function. There are only the following setting
options available: optical, optical & acoustic and only apnoea.

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 34 von 68
If the backup function is activated, the backup pushes with the set backup rate will be issued in the event of
apnoea after the apnoea time has passed and support or stimulate the patient. If patient respiration is
detected, the backup will automatically end.

In addition, the button Push becomes the Abort backup button. This can be used if the medinCNO® does not
independently detect respiration and the backup function therefore does not automatically end. Then the
backup can be manually discontinued by the user with this button.

In the trend, backup pushes administered are indicated with a blue line.

5.7 CPAP mode

CPAP mode provides a pure basic CPAP therapy. In this mode the medinCNO® produces a constant flow all
the time which is converted into a constant CPAP pressure in the Medijet ®.

The setting, action and information options are:

- Flow (chapter 5.3)

- Oxygen concentration FiO2 (chapter 5.3) and start and discontinuation of the O2 flush (chapter
5.5)

- Finsp (additional push flow) and manual PUSH (chapter 5.4)

- Alarm settings (chapter 5.12)

- Trend graphs (chapter 5.13)

In addition, the CPAP settings also include the Leak-Assist: This compensates for the difference between the
desired target average pressure and average pressure measured at the patient, with the compensation flow
being limited to ±2 L/min. In addition, an alarm (leak assist (yellow)) is triggered if the target pressure is not
reached after 30 s.
If there is a higher-priority alarm (red), this function is discontinued. If the alarm ends, the leak assist then
starts automatically on its own.
The function is ended either by direct switching off or changing into the Standby mode.

In addition, the CPAP settings contain the setting option for the level of the O2 flush (see chapter 5.5).

Please note:

The patie t s espi atio auses the CPAP p essu e p odu ed i the Medijet ® to change, and brief fluctuations
in the CPAP pressure course occur. These are visible on the CPAP graph. This can occur to a greater degree
with small prongs which is why the CPAP alarm thresholds should be set as narrowly as possible.

Use of the leak assist does not replace regular monitoring of the patient interface. This is still the task of the
medical staff and should be performed regularly.

Warning:
The CPAP mode does not contain any apnoea monitoring and no apnoea alarm.

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 35 von 68
5.8 Apnoea CPAP mode

In Apnoea CPAP, the basic CPAP of the CPAP mode is complemented by apnoea monitoring. The apnoea
o ito i g s ste assesses flu tuatio s i the CPAP p essu e ou se hi h a e aused the patie t s
respiration and responds to phases in which the patient does not breathe by automatically triggered pushes.

In addition to the setting, action and information options which are non-mode-dependent, such as:

- Flow (chapter 5.3)

- Oxygen concentration FiO2 (chapter 5.3) and start and discontinuation of the O2 flush (chapter
5.5)

- Finsp (additional push flow) and manual PUSH (chapter 5.4)

- Alarm settings (chapter 5.12)

- Trend graphs (chapter 5.13)

there are special settings for the apnoea mode. Open these apnoea settings by pressing button p in Figure
5-2. You can then change six parameters. To change a parameter, press the corresponding button to select it,
turn the dial to the desired setting, and press the dial to confirm your selection.

Figure 5-3: Submenu Apnoea settings to change the a) Inspiration time, b) Apnoea time c) Sensitivity d)
Backup on/off e) Backup rate f) the level of the O2 flush. The only purpose of the orange lines is to clarify the
significance of the sensitivity setting as a demarcation between respiration and apnoea. They are not
displayed by the medinCNO®. Instead, the sensitivity is displayed on the left-hand edge of the CPAP curve as a
pair of arrows (blue).

Warning:
Significant amounts of condensate formation in the tube system must be avoided since, as a result, the
respiratory signal in the CPAP pressure curve will be masked, detection of apnoea phases can be prevented,
and the triggering of pushes can be influenced.

Inspiration: Figure 5-3 a

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 36 von 68
The setting in this field determines the duration of pushes automatically triggered in the event of apnoea, the
backup pushes, and the duration of pushes triggered manually. It can be set to a time between 200
milliseconds and 2 seconds.

Apnoea time: Figure 5-3 b

The apnoea time determines how long the medinCNO® waits before triggering an automatic push if no
respiratory activity can be detected in the CPAP pressure course. If no respiration activity can be detected
even after a push, either the backup pushes are started or the device waits for the length of time set as
apnoea time before the next push is triggered, depending on the backup setting. The length of time that can
be set as the apnoea time is any time between 2 seconds and 20 seconds.

Please note:

The medinCNO® has an internal block preventing another apnoea push being triggered immediately after a
push for twice the length of time set as inspiration time. This means that for a selected inspiration time of
two seconds, the medinCNO® does not emit any apnoea push for four seconds after the push, even if an
apnoea time of less than four seconds is set.

Pushes based on the backup rate and manual pushes are not affected by this.

Sensitivity: Figure 5-3 c

The sensitivity setting determines the minimum pressure fluctuations in the CPAP signal which are
considered as indicative of respiration. For clarification, the sensitivity is shown as a pair of blue arrows on
the left-hand side of the CPAP pressure curve around the current CPAP average value. When setting the
se siti it , the alue sele ted he e ust e lo e tha the p essu e flu tuatio aused the patie t s
normal respiration. However, the value must also be greater than the pressure fluctuations caused by the
CPAP system. For the pressure signal shown in Figure 5-3 (without respiration on the left, with respiration on
the right), the sensitivity shown in orange would be correct as it is lower than the pressure fluctuations
caused by respiration and greater than the noise of the pressure signal without respiration. The purpose of
this setting is to allow the medinCNO® to deal with periods in which the CPAP pressure signal remains within
the limits as apnoea and with situations in which these limits are exceeded as respiratory activity.

In addition, a switch can be made to the NIPPV special case. In the case of an activated backup function
(chapter 5.6), the trigger sensitivity can be switched off. In this case, the NIPPV display appears in the upper
corner of the display. A backup rate between 5/min and 120/min can be set, depending on the inspiration
time.

Please note:

The sensitivity must be set to less than the patient's respiration and greater than the pressure fluctuations
caused by other sources (e.g. condensate in the flow tube).

Backup on/off: Figure 5-3 d and backup rate Figure 5-3 e - see chapter 5.6

Level of the O2 flush Figure 5-3 f - see chapter 5.5

Finsp – Additional push flow: – see chapter 5.4

Push pressure: The level of the automatic push triggered in the event of apnoea is identical to the pressure of
a manual push. The developing pressure results from the additional flow Finsp used for it, which is adjusted
using button c in Figure 5-2 as described in chapter 5.4.

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 37 von 68
Please note:

When selecting the push flow and sensitivity, it is important to ensure that the pressure during the push will
become greater than the upper threshold of the sensitivity selected. Otherwise the apnoea phase will not be
interrupted by the push.

Display:

When operating in Apnoea CPAP mode, the current respiration rate (RR - breaths >1 mbar per minute) and
number of pushes automatically generated by the medinCNO® per minute (Rinsp) are shown in the orange bar
on the top of the screen in addition to the information referred to in chapter 5.2. In the particular case of
NIPPV, the current respiration rate RR is not displayed.

If the backup function is activated, this is also displayed in the top left-hand area.

5.9 SNIPPV mode

The difference between the SNIPPV mode and the Apnoea CPAP mode is that in this case, an automatic push
is triggered not only in the event of apnoea but also synchronously with inspiration.

Also, in addition to the general settings and displays, such as:

- Flow (chapter 5.3)

- Oxygen concentration FiO2 (chapter 5.3) and start and discontinuation of the O2 flush (chapter
5.5)

- Finsp (additional push flow) and manual PUSH (chapter 5.4)

- Alarm settings (chapter 5.12)

- Trend graphs (chapter 5.13)

there are special settings for this mode. Its overview opens by pressing button p in Figure 5-2. They can be
changed by selecting, setting the desired value using the dial and confirming this selection by pressing the
dial.

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 38 von 68
Figure 5-4: SNIPPV mode settings a) Inspiration time; b) Expiration pause; c) Apnoea time; d) Sensitivity e)
Backup on/off and f) Backup rate g) the level of the O2 flush. The sensitivity is displayed on the left-hand
edge of the CPAP curve as a pair of blue arrows. The blue lines serve only to clarify the meaning of sensitivity,
the green line corresponds to the expiration pause and the orange arrow indicates the apnoea time. None of
these lines is displayed in actual operation.

Warning:
Significant amounts of condensate formation in the tube system must be avoided since, as a result, the
respiratory signal in the CPAP pressure curve will be obscured, detection of inspiration and apnoea phases
may be prevented, and the triggering of pushes may be influenced.

Inspiration: Figure 5-4 a

The setting in this field determines both the duration of the push which is automatically triggered by the
medinCNO® and the duration of the push triggered manually. It can be set to a time between 200
milliseconds and 2 seconds.

Expiration pause (Exp – Pause): Figure 5-4 b

The expiration pause time determines the length of time after a push during which another push cannot be
automatically triggered. The medinCNO® stops a push being triggered after each automatically triggered
push, even if the pressure again falls below the trigger threshold for a synchronous push. The purpose of this
is to restrict the number of pushes and allow the patient to exhale undisturbed. In the example shown in
Figure 5-4, the selected expiration pause after the first push was shown in green. The effect of this is to stop
a push being triggered even though the CPAP pressure falls below the trigger threshold again a short time
after the push. The next push is not produced until after the end of the expiration pause.

The expiration pause time determines the maximum push frequency (Tr. rate max), as the longer the time
selected, the fewer pushes can be triggered automatically by the medinCNO ®. In addition, the minimum
adjustable expiration pause is dependent on the duration of the inspiration time.

The expiration pause does not affect backup pushes or manually triggered pushes. Manual pushes can still
also be triggered in this period by pressing button d in Figure 5-2 .

The sensitivity cannot be switched off. This is only possible in the Apnoea CPAP mode in the particular case of
NIPPV.

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 39 von 68
Tip

If a value is chosen as expiration pause (Exp – Pause) which is larger than the set apnoea time, the expiration
pause prevails and the push is not triggered until it has ended. That is, in the case of an expiration pause of,
e.g., five seconds and an apnoea time of three seconds, the push is triggered after five seconds. Since this
setting can also negatively influence the trigger, it is recommended to always set the time of the expiration
pause to be less than the apnoea time.

The medinCNO® only emits manual or backup pushes for the duration of the expiration pause (Exp – Pause),
even after a push triggered by apnoea time.

If respiration that exceeds the trigger threshold takes place after a push within the expiration pause (Exp –
Pause) during which pushes cannot take place, no synchronous push is triggered but this respiration is taken
into account when determining apnoea, and consequently the apnoea time waiting time only starts
subsequently.

Apnoea time: Figure 5-4 c

Apnoea time determines the maximum time between two pushes. If no respiratory activity can be detected
in the CPAP pressure course, then the medinCNO® cannot trigger a push synchronised with respiration. In this
case, the medinCNO® waits for the length of the apnoea time and then triggers an automatic push. If no
respiration activity can be detected even after a push, either the backup pushes are started or the device
waits for the length of time set as apnoea time before the next push is triggered, depending on the backup
setting. This means that, as shown in Figure 5-4 in orange, the apnoea time corresponds to the maximum
waiting time if the sensitivity trigger threshold is not exceeded. Consequently, the length of apnoea time
determines the minimum number of pushes if no respiration can be detected, and this is converted into a
minimum push frequency (Tr. Rate min) and displayed in the trend.

Sensitivity: Figure 5-4 d

Sensitivity in the SNIPPV mode determines the threshold pressure fluctuations in the CPAP signal at which
the medinCNO® should respond. The lower threshold of this sensitivity is used as a trigger for a push
synchronised with respiration. If the CPAP pressure reading is below this threshold, the medinCNO ® triggers a
push. The sensitivity setting is also used to identify apnoea. Consequently, it is important to select a value
that is greater than the pressure fluctuations (e.g. due to condensate in the tube system) caused by the CPAP
system and smaller than the pressure changes caused by respiration. For the pressure signal displayed in
Figure 5-4 (without respiration on the right, with respiration on the left), the sensitivity shown in blue could
be selected. In this case, the synchronous pushes would be triggered when pressure falls below the blue line,
and the pressure in the Medijet® would rise after a short delay. At the same time, the pressure signal remains
within the blue limits in the event of no respiratory activity and a push is triggered after the length of time
defined as apnoea time (on the right). The sensitivity set is displayed on the left-hand edge of the CPAP
course curve as a pair of blue arrows.

Please note:

When selecting sensitivity, it is important to bear in mind that there may be a slight delay of up to 0.1
seconds between the time at which the pressure falls below the trigger threshold and the time when CPAP
pressure is increased.

Backup on/off: Figure 5-4 d and backup rate Figure 5-4 e - see chapter 5.6

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 40 von 68
Please note:

During a backup that is actively running, no synchronised pushes are triggered. Only after the backup is
discontinued (automatically or manually) can synchronised pushes be issued once more.

Finsp – Additional push flow - see chapter 5.4

Push pressure: Both the pushes synchronised with respiration and the automatic pushes triggered in the
event of apnoea have a level identical to the pressure of a manual push. This pressure is determined from the
Finsp flow used for it, which is adjusted using button c in Figure 5-2 as described in chapter 5.4.

Level of the O2 flush Figure 5-4 g - see chapter 5.5

Possible way of setting the SNIPPV mode:

1. Set the desired inspiration time


2. Set the desired apnoea time
3. Set the expiration pause – and/or decide on the maximum number of breaths to be supported
by pushes (Tr. rate max):
a. If the patient should be supported by pushes to the maximum possible point, this
time is set to the minimum value. If each breath is to be supported, the tr. rate max
must be larger than the patient's respiration rate.
b. On the other hand, if the patient requires less support, a greater value is selected.
The less support needed, the greater the expiration pause that can be selected.
But attention: The expiration pause time should be less than the set apnoea time.
The maximum resultant push frequency is likewise displayed in the orange header.

Displays:

In SNIPPV mode, the operator sees the current push frequency (Rinsp - pushes per minute) in the orange bar
at the top of the screen in addition to the information referred to in chapter 5.2. Additionally, the maximum
push frequency (Tr. rate max), which is determined by the length of time of the expiration pause (Exp –
Pause), is displayed. If the backup function is additionally activated, this is likewise displayed in the top left-
hand area.

5.10 Oscillation mode

In oscillation mode, an oscillating fluctuation is available in addition to the basic CPAP. This is modulated on
the basis of the background basic CPAP. In this case the medinCNO ® produces a flow which is periodically
blocked and thus reduced in accordance with the oscillation frequency that has been set. This flow results in
an oscillating CPAP pressure in the Medijet®, which is measured by the pressure meter line and displayed on
the medinCNO® screen.

As the medinCNO® itself only produces a reduced flow periodically, and the CPAP pressure is only formed at
the end of the tube system in the Medijet®, the oscillation produced in the CPAP pressure is heavily
dependent on the physical effects in the tube circuit set and Medijet® that are being used. Consequently, the
only oscillation setting options in the medinCNO® are levels and not absolute values. In order to assess the
effect of the system as a whole, the user must assess oscillation on the basis of the CPAP pressure measured
in the Medijet® and the effect on the patient.

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 41 von 68
Figure 5-5: Possible settings in oscillation mode: a) Flow volume Flow emitted b) Frequency; c) Amplitude
level Amplitude set; and d) Level of the O2 Flush

The oscillation produced is determined by the following factors. Where these factors can be adjusted, this is
done by pressing a button to select the factor, turning the dial to reach the desired value and confirming
selection by pressing the dial.

Warning:
As the oscillation produced is heavily dependent on the physical effects in the CPAP system (tube circuit set
and Medijet®), oscillation must always be set carefully and gradually, while at the same time checking the
results on the basis of the pressure course curve displayed in the medinCNO ® and the effect on the patient.

Since oscillation therapy affects the CO2 concentration in the patient's blood, it is recommended that this be
regularly checked during oscillation.

Frequency: The frequency of oscillation in CPAP pressure depends on the frequency of flow impulses
produced by the medinCNO®. This frequency can be changed in the sub-menu of oscillation settings by means
of button b in Figure 5-5. However, the frequency-dependent damping effect of the volume of the tube
system means that the frequency choice indirectly affects the CPAP pressure amplitude produced in the
Medijet®: higher amplitudes are reached at low frequencies than at high frequencies.

Flow: The basic CPAP flow setting (set using button a in Figure 5-5) has a significant influence on the
oscillation produced. As the medinCNO® generates oscillation by interrupting the Medijet® drive flow, the
flow volume produced affects the maximum pressure during oscillation. Increasing the drive flow results in
higher maximum pressure.

Amplitude: The button c in Figure 5-5 can be used to select between various amplitude levels. Depending on
the tube system used and the frequency setting, these amplitude levels have an effect on the CPAP system
and generate pressure oscillation in the Medijet®. The level of the pressure oscillations produced can be
taken from the pressure course displayed on the medinCNO ® or read in the orange readings line. The main
effects of increasing the amplitude level are to reduce the minimum oscillation pressure and slightly increase
the maximum pressure.

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 42 von 68
Tube system volume: as the air in the tube system has a damping effect on the oscillating flow impulses, the
oscillation amplitude produced in the case of large tube circuit sets is smaller than in the case of smaller tube
circuit sets.

Display: In oscillation mode, the medinCNO® provides the user with information about the amplitude that has
been measured and the selected frequency. This information is displayed in the orange bar at the top of the
screen. Since the measured amplitude is determined over the past 5 seconds, changes only have an effect
after a time delay.

Graph: In the graph of the CPAP pressure course, a dotted line is marked in addition to the CPAP pressure
curve and the two dashed alarm lines. This presents the CPAP average value of the past 5 seconds and is used
as a basis for the alarms. That is, if the dotted line crosses one of the two dashed lines, the alarm is triggered
after the alarm-dependent time delay. Since the dotted line always evaluates the past 5 seconds, it does not
react to changes until after a 5-second delay.

Push settings – see chapter 5.4

Level of the O2 flush - see chapter 5.5

Please note:

If, due to an alarm, there is a high CPAP pressure (red), feed pressure high (or Press Alarm Reset to restart
flow) or a disconnection alarm for an interruption in the oscillation, the values originally set for amplitude
and flow must not be reset but instead the values will be gradually reverted to.

In the case of all other higher-priority alarms (red), the values for amplitude and flow must be reset!

5.11 Switching between modes

Pressing button m in Figure 5-2 opens mode selection. In this case the available modes are shown at the
bottom of the screen. The user can now select the desired mode by pressing the corresponding button. It is
important to note that CPAP mode and Apnoea CPAP mode are operated by the same button. Pressing this
button repeatedly allows the user to switch between the two modes. The selected mode appears on a blue
background. The mode selected must then be confirmed by pressing the Confirm button. Only then does the
medinCNO® switch to this mode.

The user can check whether mode switch has taken place correctly by checking the mode display in the top
left-hand area of the screen after making the switch.

If the NIPPV special case is active in Apnoea CPAP mode, this is displayed instead of the mode display.

When switching between two modes, the medinCNO® transfers all the settings that are available in both
modes.

The following situations represent exceptions:

- Os illatio ode: If this e ds, the set flo alue is o l t a sfe ed if it is 8 L/min. Otherwise it is
reduced to 8 L/min.
- Apnoea CPAP: If sensitivity is deactivated, this is adjusted in the event of a change in the SNIPPV
to a value of ±0.5.

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 43 von 68
Please note:

If a highest-priority alarm (e.g. high pressure or lack of gas) and the safety mode triggered as a result are
active, the mode may not be changed. The mode can only be changed once again after the error has been
corrected. A mode change during the safety mode is only possible if the safety mode was activated by the
Disconnection alarm.

5.12 Alarm settings

The alarm sub-menu is available in each medinCNO® mode. This sub-menu can be opened by pressing button
n in Figure 5-2.

The points described below can be changed after opening this sub-menu. If two functions are allocated to a
button, the operator can press the relevant button repeatedly to switch between the two functions.
Individual values are changed by selecting the corresponding value (visible thanks to its blue background),
then using the dial to change the value and finally confirming the new value by pressing the dial.

CPAP high: this corresponds to the freely selectable upper CPAP pressure alarm. If this value is exceeded by
the CPAP average value for a certain amount of time, a CPAP pressure high alarm is triggered after a delay
(see chapter 8.2).

CPAP low: this alarm monitors the lower threshold of the CPAP pressure. If this value is undershot by the
CPAP average value for a certain amount of time, a CPAP pressure low alarm is triggered after a delay (see
chapter 8.2).

Please note:

Because of the scale, it is possible that the CPAP high and CPAP low alarm thresholds (generally shown as
dashed lines) are not visible on the display (e.g. if the scale is set from 0 to 10 and the value for CPAP high is
13). In this case, the value of the limit is visible as a number to the upper right next to the graph.

FiO2: the FiO2 setting determines the alarm interval of the oxygen alarm. The thresholds of this alarm depend
on the target oxygen concentration selected and are displayed next to the target oxygen levels. If this value
(see chapter 8.2) is exceeded/undershot, an alarm will be triggered after a delay. There are two different
setting options for this alarm, which must be preset in the service menu by a service technician.

- Automatic FiO2 alarm: Here the FiO2 limit values are automatically adapted to the new target
value for any change in the target oxygen concentration.
- Manual FiO2 alarm: In this case, if there are changes to the target oxygen concentration, the
FiO2 alarm threshold values are not changed. This adaptation can only be performed manually
by the user by opening the Alarm setting menu, selecting the button FiO2 alarm refresh and
confirming this by pressing the dial. If adapting the alarm thresholds is not possible because
they already refer to the currently set oxygen concentration, the button FiO2 alarm refresh is
dimmed and instead, the button to change the alarm interval will be visible.

Push alarm: The push alarm setting can be used to determine how the medinCNO® informs the user in the
event of a triggered push. The following settings are available: information via an acoustic signal and an
optical signal, information via an optical signal alone or no information. In addition, there is an option to use
a sound and light signal to indicate pushes caused only by apnoea. When backup is activated, the push alarm
cannot be turned off; rather it is at least indicated by a visual signal.

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 44 von 68
Alarm Silent: The Alarm Silent function can deactivate the acoustic alarm for five minutes. After this time, the
acoustic alarm reactivates itself without any intervention by the user. If the medinCNO ® sound was
deactivated for five minutes, the sound can be reactivated with this button before the five minutes are up.

Warning:
This setting may only be used while the medinCNO® is not connected to a patient (e.g. when replacing
individual components of the CPAP system) and the operator must remain by the device all the time that the
alarm is off.

Brightness: The brightness setting changes the brightness of the backlighting of the display.

Scale: Using this setting, the scale of the pressure graphs can be changed. The scales 0-10 mbar/cmH2O, 0-15
mbar/cmH2O or 0-20 mbar/cmH2O are available.

Please note:

The value for the CPAP pressure high alarm (red) is 18 mbar. Even if the scale up to 20 mbar is used, the value
is still 18 mbar.

Sound: The user can determine the repetition frequency of the alarm sounds by altering the sound setting. If
+5 is set here, the alarm sounds of the medinCNO® will be repeated very frequently. This setting is suitable
for environments with many devices and a significant amount of background noise, since it facilitates the
identification of the source of the alarm. If the value is set to +1 here, the alarm sound of high and medium
alarms will be repeated as infrequently as possible and the alarm sound of low alarms will only be emitted
once and not repeated. This setting is only suitable for quiet, manageable environments.

Warning:
If the user alters the brightness or sound settings, he/she must take account of the specific surroundings and
may only change the settings to such an extent that the screen content remains visible and the alarm sounds
can still be heard.

5.13 Trend

The trend display shows the course taken by the values set for each mode. Eight hours are shown per page.
Overall the values since the last device start, however a maximum of values from the past 28 days, can be
displayed by pressing the arrow buttons. If, at the time indicated, the medinCNO® was in a mode in which the
value shown is not available, this time period will be shown as empty in the trend. All trend values, except for
the alarms, are deleted when the device is switched off.

CPAP / FiO2 trend: Both of these trend courses are available in all modes. They show the course of the
oxygen concentration set and the average CPAP pressure.

RR / Rinsp: In Apnoea CPAP mode, the course of the respiration rate (breaths >1 mbar per minute) and push
frequency (number of pushes per minute) can also be shown. In the NIPPV special case, no respiration rate is
shown and if a backup was actively running, the line is shown in blue.

Rinsp / Tr. rate min/max: In SNIPPV mode, a trend is available which provides information about the actual
push frequency (Rinsp - number of pushes per minute) and the maximum and minimum push frequencies (per
minute) theoretically possible at this point in time due to the settings. If backup is actively running, the line is
shown in blue.

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 45 von 68
Frequency/amplitude: In oscillation, the set frequency and the measured amplitude are recorded in addition
in the trend.

USB export: The USB export of the trends contains the readings and settings and additionally a logfile of the
alarms described in chapter 3.4.4.

5.14 Standby

The standby mode can be used if the medinCNO® is not being used to treat an acute case but should be kept
ready for use. If the device is switched from one of the other modes to this mode, the medinCNO ® stores all
the settings that have been selected, stops producing a flow and deactivates the alarms. Pressing the End
button takes the device out of standby mode and causes the medinCNO ® to revert to the previous mode.

If the standby mode is used during oscillation, the oscillation is only generated once again with an amplitude
of +1 and the flow is reduced to 8 L/min upon ending the standby. Thereafter, the amplitude and the flow
must be reset manually by the user.

If the medinCNO® is switched during CPAP mode with activated leak assist into standby mode, the leak assist
needs to be manually restarted after ending the standby.

Ending the standby and reactivating normal operation of the medinCNO ® lasts approximately ten seconds.

Warning:
The standby mode can only be used when no patient is attached to the medinCNO ®.

Please note:

In standby mode, no gas flow can be emitted from the device. When combining with an active respiratory gas
humidifier, the following points must be borne in mind:

- If no gas flow is issued and if the respiratory gas humidifier is continuing to heat, there can be
so-called "hot shots". This describes the brief delivery of overly hot respiratory gas to the
patient which collected in the respiratory gas humidifier chamber during time spent in standby
mode.

Recommendation:
In the case of a brief period in standby mode (manipulation such as bonnets and/or mask/prong
changes, suctioning, etc.) and activated humidifier, the medinCNO ® should be restarted before
the patient is connected to the device. This will avoid the emission of hot shots to the patient.
In the case of a longer period in standby mode (pause in CPAP therapy, skin-to-skin care, etc.),
the respiratory gas humidifier should be switched off.

- In the case of a longer period in standby mode and deactivated respiratory gas humidifier, the
tubing system should be hung as vertically as possible so that any condensate can flow back
into the respiratory gas humidifier. The Medijet® generator should be neatly covered. In
addition, the 7-day rule regarding replacement of the tubing system applies.

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 46 von 68
6 Switching Off

There are two ways of switching off the medinCNO®.

6.1 Software switch-off

Holding the ON/OFF button down for at least three seconds switches off the medinCNO ®. The medinCNO®
stops supplying air and oxygen and switches off the electronic components. If the medinCNO® is connected to
the power supply, the internal rechargeable battery is charged, if necessary. However, this is not shown on
the display.

6.2 Hardware switch-off

Holding button m in Figure 5-2 down while at the same time pressing the ON/OFF button for four seconds
immediately switches off the medinCNO® and the device stops supplying gas. If the combination of buttons is
not held down long enough, the medinCNO® switches off and the gas feed is interrupted, but an alarm sound
lasting for two minutes is additionally triggered.
To suppress this two-minute alarm sound early, the medinCNO® must be started until the step: Attention: Do
not connect a patient to the medin-cno during system start-up and then the ON/OFF button should be held
down briefly once again.

6.3 Disconnection

Once the medinCNO® has been switched off it can be unhooked from the air and oxygen supply and its power
supply unit can be disconnected. The patient tube circuit sets must be removed and disposed of, and the
surface of the medinCNO® cleaned as described in chapter 9.1.

Disconnect the power supply unit from the mains during thunderstorms. Do not pull on the cable to
disconnect the power supply unit from the mains.

Please note:

Since the medinCNO® battery even discharges when the device is switched off, it is recommended to charge
the medinCNO® if it is stored for longer period of time or to charge it completely before using it on
rechargeable battery power.

6.4 Disposal

Please dispose of the device and the associated power supply unit properly (WEEE - Waste Electrical and
Electronic Equipment).

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 47 von 68
7 Accessories

7.1 Accessories

Various accessories and spare parts are available from Medin for operation of the medinCNO ®:

Gas supply tubes Medin provides gas supply tubes compliant with various standards.
Water trap A water trap for the air supply of the medinCNO® can be ordered
together with the medinCNO®. This is then mounted in the air
supply line of the medinCNO®.
Adapter for patient flow outlet: Spare adapter, aluminium M22/F15, REF 51091
Spare adapter, white REF 39-902
Spare power supply unit: REF 39-113
Power plug, EU (without power REF 39-115
supply unit)
Power plug, UK (without power REF 39-116
supply unit)
Power plug, USA/Japan (without REF 39-117
power supply unit)
Power plug, Australia (without REF 39-118
power supply unit)
Power plug, China (without power REF 39-127
supply unit)
Spare rechargeable battery REF 39-101
Spare 3V battery REF 39-102
Spare oxygen cell REF MLF–16 or OOM-102
Spare filter for water trap REF 39-819
Ferrite for RS232 or USB cable REF 39-125
Filter REF 39-317
Filter grid REF 39-318
Retrofit kit REF 39-320

7.2 Components of the rest of the CPAP system

The medinCNO® must be used in combination with the Medijet® nCPAP generator and the appropriate
accessories (bonnets, masks, prongs and tube circuit sets).

For more specific information on this, visit the homepage www.medingmbh.com.

Warning:
The medinCNO® can only be operated in combination with the Medijet® nCPAP generator and the appropriate
tube circuit sets, bonnets, masks and prongs. If used with any other nCPAP generator, the functionality of the

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 48 von 68
device cannot be ensured, the correct alarm function cannot be guaranteed, sufficient CPAP pressure will not
be generated and neither apnoea detection nor synchronisation in SNIPPV mode will work.

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 49 von 68
8 Alarms and Error Messages

8.1 Error messages during system start-up

Error message text Meaning


Connect power supply unit - battery low The internal battery is almost discharged. Please connect
the power supply unit to the mains.
Fan blocked The fan is blocked, please contact a service technician.
Gas feed pressure low, check O2 The oxygen feed pressure is too low (<3 bar). Please check
oxygen feed.
Gas feed pressure high, check O2 The oxygen feed pressure is too high (>6 bar). Please check
oxygen feed.
Gas feed pressure low, check air The air feed pressure is too low (<3 bar). Please check air
feed.
Gas feed pressure high, check air The air feed pressure is too high (>6 bar). Please check air
feed.
Oxygen calibration failed, check O2 The oxygen calibration could not be performed correctly.
Please check whether the oxygen cell is working properly
and whether the medinCNO® is correctly connected to the
oxygen feed. The medinCNO® cannot be operated without
oxygen.
Device error - device shut down necessary The e is a defe t i the de i e s p eu ati s ste , at the
overpressure valve or rechargeable battery. The
medinCNO® can no longer be used and must be shut down.
Please contact a service technician.
Error during tube test - restart An error occurred during the tube test (e.g. leakage).
Please repeat the tube test.
Device error - patient pressure. Device must The patient pressure sensor is causing an error. The device
be replaced cannot be used and must be replaced. Please contact a
service technician.
3V battery empty - battery has to be The internal 3V battery is empty and must be replaced.
exchanged. Please contact a service technician.
Battery needs replacing - device function in The internal rechargeable battery is too old or was used
battery mode cannot be guaranteed. Device too frequently and must be replaced. Do not use the
must be connected to mains for use! medinCNO® on rechargeable battery power and contact a
service technician.
Maintenance necessary! The last maintenance of the medinCNO® took place at least
Next maintenance necessary: 11 months ago and must be repeated no later than the
YY.MM.DD date indicated. Contact a service technician.

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 50 von 68
8.2 Error messages during operation

Error message text Colour/ ∆t (1) Meaning and device response


Priority
Disconnection (CPAP, Apnoea red/ <6s(2) The CPAP pressure is constantly <2.0 mbar
CPAP, SNIPPV mode) high for >5 seconds - the flow tube or pressure
meter tube are no longer connected to the
medinCNO®. Please reconnect the tube.
This alarm starts the emergency
programme.
Disconnection (oscillation mode) red/ <12s The CPAP pressure is constantly <2.0 mbar
(2) (3) for >5 seconds - the flow tube or pressure
high
meter tube are no longer connected to the
medinCNO®. Please reconnect the tube.
The average value interprets the CPAP
readings from the past five seconds which
delays the start and end of the alarm up to
12 seconds.
This alarm starts the emergency
programme.
CPAP pressure high red/ <1s CPAP pressure >18 mbar - the medinCNO®
high briefly interrupts the flow.
This alarm starts the emergency
programme.
Feed pressure high red/ <1s Feed pressure >40 mbar, the flow supply
high tube is blocked or is generating resistance
that is too high. Please check. - The
medinCNO® briefly interrupts the flow.
This alarm starts the emergency
programme.
Device error - patient pressure. red/ <1s The patient pressure sensor is causing an
Device must be replaced! high error. The two CPAP sensors are measuring
a difference of >2 mbar for 500
milliseconds. The device can no longer be
used and must be replaced. Please contact
a service technician.
This alarm starts the emergency
programme.
Device error - device shut down red/ <1s A serious error occurred within the
necessary high medinCNO®. The medinCNO® can no longer
be used and must be replaced. Contact a
service technician.
This alarm starts the emergency
programme.
Check gas feed pressure O2 red/ 3s The O2 feed pressure is too low (<3 bar) or
high too high (>6.0 bar) - the medinCNO® is only
providing air.
This alarm starts the emergency
programme.

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 51 von 68
Check gas feed pressure air red/ 3s The air feed pressure is too low (<3 bar) or
high too high (>6.0 bar) - the medinCNO® is only
providing oxygen.
This alarm starts the emergency
programme.
CPAP pressure high yellow/ <87s The CPAP average pressure was
(3) continuously greater for >1 second than
medium
the set upper CPAP pressure alarm
threshold.
- In a normal case, the average value
interprets the CPAP readings from the past
five seconds which delays the start of the
alarm up to <8 seconds.
- If the backup is activated, there may be a
maximum delay of 85 seconds until the
average value has been fully updated,
which can result in a total alarm delay of
<87 seconds.
- In SNIPPV mode without active backup,
the average value is updated every 15
seconds, and for this reason, there is a
total alarm delay of <17 seconds.
In this case, the medinCNO® continues to
work normally.
CPAP pressure low yellow/ <101s For 15 seconds, the CPAP average pressure
(2)(3) was constantly less than the set lower
medium
CPAP pressure alarm threshold.
- In a normal case, the average value
interprets the CPAP readings from the past
five seconds which delays the start of the
alarm up to <22 seconds.
- If the backup is activated, there may be a
maximum delay of 85 seconds until the
average value has been fully updated,
which can result in a total alarm delay of
<101 seconds.
- In SNIPPV mode without active backup,
the average value is updated every 15
seconds, and for this reason, there is a
total alarm delay of <31 seconds.
In this case, the medinCNO® continues to
work normally.
Oxygen concentration high yellow/ 100s The oxygen concentration measured is
medium higher than the alarm threshold set for
~1.5 min - the medinCNO® continues to
operate normally.
Oxygen concentration low yellow/ 100s The oxygen concentration measured is
medium higher than the alarm threshold set for
~1.5 min - the medinCNO® continues to
operate normally.

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 52 von 68
Temperature yellow/ 90s The temperature in the medinCNO® is too
medium high. Check the environmental conditions
and/or the prefilter and contact a service
technician.
Connect power supply unit - yellow/ 6s The internal battery is almost discharged.
battery low medium The power supply unit must be connected
to the mains immediately.
Fan blocked yellow/ 15s The fan is blocked, please contact a service
low technician.

Leak-Assist yellow/ <36s This alarm is triggered if the average CPAP


low is different from the target CPAP for 30
seconds and if the leak flow of ±2 L/min is
not sufficient to compensate.
In this case, the medinCNO® continues to
work normally.
Pure alarm sound, medinCNO® - <1s This alarm is triggered if the medinCNO®
switched off power supply fails completely, for example
because the power supply unit was not
connected to the mains although the
battery was discharged.
The medinCNO® was switched off.
Press Alarm Reset to restart flow red/ - No alarm – informational message only
high Triggered by a prolonged high feed
pressure (e.g. if a flow tube is blocked), by
several high CPAP pressures successively or
by a combination of high feed pressures
and high CPAP pressures, the flow supply is
interrupted for the time being and is only
reactivated after the Alarm Reset button is
pressed. Reactivation of the flow takes
approximately 10 seconds starting from
this moment.
Push activated turquoise/ - Not actually an alarm, just notification that
info a push was triggered. Users can decide
whether or not to receive this information
visually and aurally.
Apnoea – push activated turquoise - Not actually an alarm, just a notification
/ that a push was triggered due to apnoea.
info Users can decide whether or not to receive
this information visually and aurally.
Backup Push gray/ - Not actually an alarm, just a notification
info that a backup push was triggered due to
apnoea. Users can decide whether or not
to receive this information aurally.
However, this information is always
indicated visually.

Dela ∆t: This o espo ds to the maximum time which passes from the time the alarm condition occurs until the alarm is
triggered.

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 53 von 68
(2) After start-up, the disconnection and CPAP pressure low alarm is suppressed for the first 5 minutes of operation. However,
the alarms are activated no later than after 5 min or if a CPAP pressure >1 mbar for at least 30 sec has already been measured.

(3) Maximum time delay, including maximum required time to completely re-calculate the average value.

The alarm volume is:

In the case of A-weighted


Type of alarm Volume in (dB)
background noise volume (dB)
High priority (red) 69 53
Medium priority (blinking yellow) 67 52
Low priority (steady yellow) 67 52
Information signal (push) 68 53
Alterable alarms:

The user can set the CPAP pressure low, CPAP pressure high (yellow) alarms and the permitted deviation
interval for the oxygen concentration high / low alarms. This can be done using the setting in the Alarm
Settings menu (see chapter 5.12). After the numerical values are adjusted, the modified settings become
active and are saved when this change is confirmed by pressing the dial. No other alarms can be modified.

Alarm settings upon start-up:

Upon start-up, the CPAP pressure high (yellow) alarm is set to 12 mbar and the CPAP pressure low alarm is set
to 3 mbar. The alarm interval for monitoring the oxygen levels set is always set to ±5% at start-up.

If the medinCNO® is restarted, the alarms are reset to these levels. This cannot be changed by the user.

The alarms - CPAP pressure low (yellow) and CPAP pressure high (yellow) analyse whether the average value
of the CPAP readings is outside of the alarm thresholds (can be set under Alarm Settings – see chapter 5.12).
Thus the alarm CPAP pressure high is triggered if the CPAP average value was above the alarm threshold for
more than one second. The alarm CPAP pressure low is triggered if the CPAP average value was below the
alarm threshold for more than 15 seconds.

In oscillation mode, the Disconnection alarm is triggered if the average CPAP value was less than 2 mbar for
five seconds.

Since the Disconnection (only in oscillation mode), CPAP pressure low and CPAP pressure high alarms refer to
the average CPAP pressure, the start and end of the alarm is thus further delayed (see table at the start of
the chapter). This likewise causes the CPAP pressure low or Disconnection alarm during oscillation to not be
reset until the CPAP average value is again above the alarm threshold and not directly after adjusting the
alarm threshold or correcting the problem.

The alarm oxygen concentration high/low is triggered when the measured oxygen concentration is outside of
the set limits. These thresholds are always automatically set by the medinCNO ® in automatic FiO2 alarm
thresholds mode at the level of the set interval (can be set under alarm settings; see chapter 5.12) around
the target oxygen concentration. In FiO2 alarm thresholds manual mode, they are set via manual updating at
the level of the set intervals around the oxygen target value active at this time. The thresholds that are active
due to the current setting are displayed in the target oxygen concentration field as superscript and subscript
numerals. Thus, for example, at a set target oxygen concentration of 30% and an alarm interval setting of
±3%, the oxygen concentration low alarm is triggered if the measured oxygen concentration is less than 27%,

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 54 von 68
and the oxygen concentration high alarm is triggered if the measured oxygen concentration is greater than
33%.

Warning:
When setting the alarm thresholds for CPAP pressure high (yellow), CPAP pressure low and oxygen
concentration, it must be ensured that the alarm thresholds are set so as to be as limited as necessary for the
patient since otherwise these alarms are unusable.

If a very high flow is used due to significant leakage or if a very small prong is used, the Disconnection and
CPAP pressure low alarms can be negatively affected by the pressure that develops in the generator as a
result. In these cases, the threshold of the CPAP pressure low alarm must be set very carefully and as high as
possible.

If the alarm Battery low – Connect power supply unit does not disappear despite connecting the power supply
unit and deleting the alarm text (by triggering a push notification or another alarm (red)), there may be a
defect of the medinCNO®. Contact a service technician and do not use this medinCNO® for the time being.

Please note:

When using pushes which generate a CPAP pressure of 18 mbar or more and which are supplied by means of
high push flows or a long patient tube containing a constriction, the feed pressure high and CPAP pressure
high (red) alarms may be generated during pushes. This combination generally also activates the text
message Press alarm reset to restart flow, whereby the medinCNO® stops the flow emission until the Alarm
Reset button is pushed. After pressing the button, the reactivation of the flow takes approximately 10
seconds. To avoid this, the push flow must be decreased and constrictions in the tube system must be
eliminated.

Alarm resetting:

All high-priority alarms (all red alarms) are self-sustaining. That is, these alarms are not deactivated until after
confirmation by the user and elimination of the alarm condition. Medium- and low-priority alarms (yellow)
are not self-sustaining. That is, these alarms are automatically cancelled when the alarm condition is no
longer present. The text message is displayed here for approximately one minute more and is then also
deleted, if no other text message is displayed.

8.3 Emergency programme

If high-priority alarms occur during operation of the medinCNO® due to errors or supply breakdowns, the
medinCNO® tries to maintain operation as long as possible. In doing so, the normal function of the
medinCNO® is limited as follows:

- Instead of oscillation, leak-assist or pushes, the medinCNO® limits its operation to normal basic
CPAP. That is, the oscillation is stopped and the medinCNO® does not emit any more manual
pushes, apnoea pushes, backup pushes and SNIPPV pushes. Likewise, the leak-assist function
pauses. After the emergency programme has ended, the oscillation is not automatically restarted
but instead the user must reset this unless only the following alarms were present: CPAP pressure
high, Feed pressure high, Press Alarm Reset to restart flow or Disconnection. In the case of these
alarms, the amplitude and flow values originally set do not need to be reset, but instead the
values are gradually reset automatically.

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 55 von 68
- The mode change of the medinCNO® is limited: The mode can be changed only if the emergency
programme has been triggered due to a Disconnection alarm. Otherwise the error must be
corrected prior to changing the mode.
- The flow emitted by the medinCNO® is li ited to a a i u of L/ i . If a flo of L/min was
set previously, this setting will be maintained. Otherwise, only a flow of 8 L/min will be emitted
instead of the flow set. After the emergency programme, the set flow will again be automatically
emitted in CPAP, Apnoea CPAP and SNIPPV mode. In the oscillation mode, the user must manually
perform this together with reactivation of the oscillation. (Exception: Only these alarms were
active: Disconnection, CPAP pressure high, Feed pressure high or Press Alarm Reset to restart flow)
- If the emergency programme is triggered by a CPAP pressure high or Feed pressure high alarm,
the medinCNO® stops its flow. If the triggering high pressure decreases immediately (<200 msec),
the medinCNO® restarts its flow after approximately 4 seconds. If it takes longer than this amount
of time for the pressure to decrease, the medinCNO® restarts its flow gradually, whereby it takes
approximately 10 seconds for the medinCNO® to once again emit its set flow. If the alarm was
triggered by a blocked flow supply tube or a combination of several alarms of the CPAP pressure
high or feed pressure high type, the flow is stopped for the time being and in addition to the feed
pressure high/CPAP pressure high alarm, a text message appears, informing the user that the flow
will be restarted by pressing the Alarm Reset button.
- If the emergency programme is caused by the breakdown of one of the supply gases, the
medinCNO® continues its basic CPAP operation with the remaining gas so that an oxygen
concentration of 21% or 100% develops.
- If there is a serious error in the device, the flow is deactivated and the device must be restarted.

8.4 Indicator lights

The medinCNO® contains two LEDs with the following meanings:

Alarm LED:

Colour Meaning
Red A highest-priority alarm is present – Immediate action by the operator is necessary.
Yellow A medium- or low-priority alarm is present – Prompt action by the operator is required.
White Information signal: two short blinks indicate that a push was triggered. This signal can be
deactivated in many situations by the operator (see chapter 5.12) as needed.

LED in the dial:

This LED is activated if the operator can change or confirm settings by turning or pressing the dial. If this LED
is deactivated, the dial does not have any function at the present moment.

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 56 von 68
9 Cleaning and Maintenance

Warning:
In order to avoid fires, oxygen must not come into contact with oil or grease. It is therefore essential to
ensure that all parts of the medinCNO® and the CPAP system which come into contact with oxygen (e.g. gas
feed tubes, O2 inlet to the medinCNO®, patient outlet, patient tubes and Medijet®) remain free from oil and
grease even during cleaning, repairs and maintenance.

Medin offers two difference warranty terms:

- 3 years (REF 39-912) from the date of production


- 5 years (REF 39-913) from the date of production

These warranty expansions only relate to components with defects caused by production, wear parts, as well
as software problems. In addition, the customer must demonstrate that all required maintenance was
performed.

Please note:

The following are not included in the warranty: physical damage, environmental damage (e.g. water or
grease in the device connection, excessively high voltage), improper use as well as consumables (e.g. oxygen
cells, filters, batteries).

9.1 Cleaning

The surface of the medinCNO® should be disinfected before first use using a slightly damp cloth 1 and every
time after it has been used on a patient. 70% isopropyl alcohol solution can be used for this purpose.

Warning:
- The device must never be sterilised or immersed in a liquid solution.
- No abrasive substances may be applied to the surface.
- No liquid may penetrate the device.

9.2 Maintenance and exchange of the internal batteries

The medinCNO® must undergo maintenance and a function check in accordance with the manufacturer's
instructions every 12 months. This must be carried out by trained specialists using appropriate measuring and
testing equipment.

The following actions must be performed during this maintenance:

- The oxygen cells are regenerated or their service life is checked.


o Type OOM-102 (REF OOM-102) oxygen cells must be replaced annually.
o Type MLF-16 (REF MLF-16) oxygen cells must be replaced if they have less than one
year of service life left or no later than after three years.

- Test of the de i e s fu tio alit a d se sor calibration.

1Slightly damp = Take a wet cloth and wring it out. Then wrap a dry cloth around the wet cloth and wring out
both of these cloths together once again.

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 57 von 68
- The internal 3-volt battery must be replaced as required or every three years at the latest. The
replacement battery must be a CR123A powerone from VARTA Microbattery GmbH (REF 39-102).

- The internal rechargeable battery must be replaced as required or every three years at the latest.
The replacement battery must be a rechargeable battery of the same type (can be ordered from
Medin, REF 39-101).

- If the medinCNO® contains a water trap, the filter contained within it (REF 39-819) should be
changed as needed (if soiled) or no later than after three years.

If an error is detected during this function test and maintenance, the device must not be used or connected
to a patient under any circumstances.

Warning:
The two internal batteries (3V and rechargeable) may only be replaced by the battery types listed above and
only by trained, professional service personnel and in accordance with the instructions in the service manual.

After each maintenance or repair of the medinCNO®, a complete function test must be performed and passed
before this medinCNO® can be used on a patient once again.

9.3 Repairs

Warning:
Repairs may only be made to the medinCNO® by specially trained, professional service personnel in
accordance with the instructions and warnings in the service manual. After each maintenance or repair of the
medinCNO®, a complete function test must be performed and passed before this medinCNO ® can be used on
a patient once again.

9.4 Device modifications

Warning:
The medinCNO® may not be modified without permission from the manufacturer and subsequent
appropriate examinations and testing to ensure continued safe use.

9.5 Power supply unit (cleaning, maintenance, repairs and modifications)

Cleaning:

Prior to cleaning, the power supply unit of the medinCNO ® must be disconnected from the mains.

Do not clean with chemical cleaning agents.

Maintenance/repairs/modifications:

The power supply unit FW7405M/24 of the medinCNO® is maintenance-free. It should not be opened.

The power supply unit may only be repaired by authorized specialists.

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 58 von 68
Warning:
To avoid an electrical shock, the housing of the power supply unit of the medinCNO ® should not be opened.

Modification of the power supply unit is not permitted (termination of the warranty).

9.6 External filter

The filter must be cleaned every 2 months by tapping or knocking (dry). The filter needs to be replaced after
12 months.

To change the filter, the filter grid must be pulled out from behind. The filter can be removed and cleaned.
The filter grid together with the filter is then pushed back in as far as the stop. You will hear a click as it snaps
into place.

Figure 6: External filter

Please note:

Only a type PPI 30 filter should be used.

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 59 von 68
10 Electromagnetic Compatibility

The information in this section is provided in order to enable the operator of the medinCNO ® to decide
whether the medinCNO® is suitable for its electromagnetic environment.

Attention:
- The medinCNO® is a medical electrical device. Consequently, in order to guarantee the function
of the medinCNO®, precautions with regard to electromagnetic compatibility must be taken and
the medinCNO® must be set up and used in accordance with the conditions set out below.
- The characteristics of the medinCNO® determined by transmissions permit it to be used in
industrial settings and hospitals (CISPR 11, class A). When used in a residential setting (for which
class B is usually necessary, according to CISPR 11), this device may not offer appropriate
protection from radiocommunication services. If necessary, the user must take corrective
measures such as moving or reorienting the device.
- Portable or mobile RF communications equipment (e.g. mobile phones) should not be used less
than 30 cm near Medin parts and cables associated with the medinCNO ®. Failure to observe
these instructions can lead to a reduction in the performance of the device.

Please note:

Electromagnetic interference may affect the device and this can be manifested through alarms (for a detailed
description, see chapter 8) or limitations in normal function of the medinCNO®(see chapter 8.3).

10.1 Electromagnetic transmission

Guida ce a d a ufacturer’s declaratio - electromagnetic emissions

The medinCNO®is intended for use in the electromagnetic environment specified below. The customer or
user of the medinCNO®, or the organisation responsible, should assure that it is used in such an
environment (CISPR11, group 1, class A).
Emissions test Compliance Electromagnetic environment - guidance
The medinCNO® uses RF energy only for its
internal function. Therefore, its RF
RF emissions, CISPR11 Group 1 emissions are very low and not likely to
cause any interference in nearby electronic
equipment.
RF emissions, CISPR11 Class A
Harmonic emissions, The medinCNO® is intended for use in
Class A
IEC 61000-3-2 professional healthcare institutions
(CISPR11, group 1, class A).
Voltage fluctuations/flicker
Complies
emissions, IEC 61000-3-3

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 60 von 68
10.2 Electromagnetic immunity

Guida ce a d a ufacturer’s declaratio - electromagnetic immunity

The medinCNO® is intended for use in the electromagnetic environment specified below. The customer or
user of the medinCNO®, or the organisation responsible, should assure that it is used in such an
environment.

Electromagnetic
Immunity test IEC 60601 test level Compliance level
environment - guidance

Floors should be wood,


Electrostatic ± 8 kV contact ± 8 kV contact concrete or ceramic tile.
discharge (ESD) If floors are covered with
according to ± 2 kV, ± 4 kV, ± 8 kV, ± 2 kV, ± 4 kV, ± 8 kV, synthetic material, the
IEC 61000-4-2 ± 15 kV air ± 15 kV air relative humidity should
be at least 30%.

± 2 kV for power ± 2 kV for power supply


supply lines lines
100 kHz repetition 100 kHz repetition
frequency frequency Mains power quality
Electrical fast
should be that of a typical
transient/burst
commercial or hospital
IEC 61000-4-4 ± 1 kV for input/output ± 1 kV for input/output environment.
lines lines
100 kHz repetition 100 kHz
frequency repetition frequency
± 0.5 kV, ± 1 kV ± 0.5 kV, ± 1 kV
differential mode differential mode Mains power quality
Surges, should be that of a typical
IEC 61000-4-5 commercial or hospital
± 0.5 kV, ± 1 kV, ± 2 kV ± 0.5 kV, ± 1 kV, ± 2 kV environment.
common mode common mode
Mains power quality
should be that of a typical
commercial or hospital
environment.
Thanks to its built-in
0% UT; ½ cycle battery, the medinCNO®
Voltage dips 0% UT; ½ cycle
according to At 0, 45, 90, 135, 180, continues to operate
At 0, 45, 90, 135, 180,
IEC 61000-4-11 225, 270 and 315 during power mains
225, 270 and 315 degrees
degrees interruptions. It therefore
does not need to be
powered from an
uninterruptible power
supply or external
battery.

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 61 von 68
0% UT; 1 cycle 0% UT; 1 cycle
and and
70% UT, 25/30 cycles 70% UT, 25/30 cycles
Single-phase; at 0 Single-phase; at 0
degrees degrees

Voltage
interruptions 0% UT; 0% UT;
according to 250/300 cycle 250/300 cycle
IEC 61000-4-11

The strength of power-


Power frequency frequency magnetic fields
(50/60 Hz) magnetic 30 A/m 30 A/m should correspond to that
field IEC 61000-4-8 of a typical commercial or
hospital environment.
Note: UT is the a.c. supply voltage prior to application of the test level

Electromagnetic
Immunity test IEC 60601 test level Compliance level
environment - guidance

Portable and mobile RF


communications equipment
should be used no closer to
3V any part of the medinCNO®,
3V
0.15 MHz to 80 MHz including cables, than the
Conducted RF 0.15 MHz to 80 MHz recommended separation
disturbances 6 V in ISM frequency
6 V in ISM frequency distance of 30 cm.
according to IEC bands a between a
bands between
61000-4-6 0.15 MHz and
0.15 MHz and 80 MHz
80 MHz
80% AM at 1 kHz
80% AM at 1 kHz

Portable and mobile RF


communications equipment
Radiated RF 3 V/m 3 V/m should be used no closer to
disturbances
80 MHz to 2.7 GHz 80 MHz to 2.7 GHz any part of the medinCNO®,
according to
80% AM at 1 kHz 80% AM at 1 kHz including cables, than the
IEC 61000-4-3
recommended separation
distance of 30 cm.

a The ISM bands between 150 kHz and 80 MHz are 6.765 MHz to 6.795 MHz; 13.553
MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz and 40.66 MHz to 40.70 MHz.

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 62 von 68
Determination of test values for electromagnetic immunity of casings against high-frequency, wireless
communication equipment

Electromag
Test Frequency Radiocommu Maximum netic
Modu Distance
frequency band nication output immunity
lation m
MHz MHz service W test level
V/m
Pulse
modulati
385 380 to 390 TETRA 400 1.8 0.3 27
on
18 Hz
FM
GMRS 460, ± 5 kHz
450 430 to 470 2 0.3 28
FRS 460 deviation
1 kHz sine
710 Pulse
LTE band 13,
745 704 to 787 modulati 0.2 0.3 9
17
780 on 217 Hz

810 GSM
800/900, Pulse
TETRA 800, modulati
870 800 to 960 2 0.3 28
iDEN 820, on
930 CDMA 850, 18 Hz
LTE band 5

1720 GSM 1800;


CDMA 1900,
Pulse
1700 to GSM 1900;
1845 modulati 2 0.3 28
1990 DECT;
on 217 Hz
LTE band 1,
1970 3, 4, 25;
UMTS
Bluetooth,
Pulse
2400 to WLAN 802.11
2450 modulati 2 0.3 28
2570 b/g/n,
on 217 Hz
RFID 2450,
LTE band 7
5240
Pulse
5100 to WLAN 802.11
5500 modulati 0.2 0.3 9
5800 a/n
on 217 Hz
5785

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 63 von 68
11 Symbols

The symbols with the meanings indicated below are used as labels on the medinCNO ® or in these instructions
for use:

Symbol Meaning Symbol Meaning

CPAP mode symbol Apnoea CPAP mode symbol

SNIPPV mode symbol Oscillation mode symbol

Indicates devices that contain


Standby mode symbol RF transmitters or
deliberately emit RF energy

Charge status of the internal


Manual push trigger
rechargeable battery

Alarm sound paused for


the time indicated. Alarm
Conformity mark
LED and alarm text
continue to be displayed.
The device contains
batteries and electrical
components. Consequently
Direct current
it cannot be disposed of in
domestic waste but must
be collected separately.

Manufacturer Device serial number

Do not use device if


Keep dry
packaging is damaged

Date of manufacture - year REF number, order number

I st u tio sig Follo


Patient flow outlet
the i st u tio s fo use

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 64 von 68
Patient connection
USB interface
pressure meter line

Electrical connection to
connect the medinCNO® Air gas inlet
power supply unit

Interface for external


Oxygen gas inlet
alarm system

Power supply unit symbols:

Symbol Meaning Symbol Meaning

Declaration of conformity Approval mark from


according to UL valid in the USA and
93/42/EEC Canada.

WEEE guideline Class II device

Manufacturer's trademark IP code numbers

Measurement of
ambient Attention – Follow
temperature/converter instructions
Temperature class

Read user manual

12 Abbreviations

Abbreviation Meaning
nCPAP / CPAP (nasal) continuous positive airway pressure
SNIPPV / NIPPV (synchronized) non-invasive positive pressure ventilation
Rinsp Push frequency (of number of pushes per minute emitted by the medinCNO®)
Finsp Push flow
FiO2 Oxygen concentration

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 65 von 68
RR Respiratory rate
Tr rate min/max maximum/minimum possible push frequency limited by setting
EMC Electromagnetic Compatibility

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 66 von 68
13 History of the Software Versions and Hardware Changes

The software of the medinCNO® consists of various software packets: Part of them determines the operating
properties of the medinCNO®, the alarms and the functional range. The display macro version establishes the
possible display languages. In doing so, each SW version is dependent on a particular version of the display
macro.
Software version Hardware changes Description
SW version V 1.0 / First version
2.46.0 together with
display macro version
V100-1
SW version V1.1 / 2.46.0 - Addition of the push pressure display
together with display - Measurement of the oscillation amplitude
macro version V101-X - Addition of the O2 flush
(X: depends on language - Addition of the push alarm for "only
pack) apnoea pushes"
- Addition of the manual FiO2 alarm
adjustment
- Reference of the CPAP alarms in oscillation
mode to the average value and display of
the average value in the graph
- Expansion of the trend recording including
export option

SW version V1.2 / 2.47.0 - Internal software changes regarding timing


together with display and internal processes – no changes for
macro version V101-X the user
(X: depends on language
pack)
SW version V1.3 / 2.47.0 Battery (other model), RS232, - Software change to RS232 interface
together with display O2 cell new - Tables for immunity and separation
macro version V102-X distances for life-sustaining device revised
(X: depends on language - Leak assist, backup function
pack) - Optional RS232 interface, ferrites on USB
or RS232 cable
- Change in scale
- New housing design
- Alarm conditions
- Trend data, live data
- Trend
- - Priority in the case of alarm messages
during operation added.
- List of Abbreviations
- Explanation of alarm calculation
- Electromagnetic compatibility according to
DIN EN 60601-1-2:2016
- No ETL anymore
- Power supply China, device service life,
slightly damp

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 67 von 68
Notes:

medin Medical Innovations GmbH  +49 8142 448 460 OP_medinCNO EN


Adam-Geisler-Str. 1 info@medingmbh.com Rev 11 Stand 25.08.2017
D-82140 Olching www.medingmbh.com 68 von 68

You might also like