AdultBrainPerfusionCT PDF
AdultBrainPerfusionCT PDF
AdultBrainPerfusionCT PDF
0 3/1/2016
Key Elements
• Time-resolved scans are used to track the flow of iodinated contrast media through the brain;
• Multiple images (20-40) are acquired over the same section of anatomy;
• Patients must be able to remain still during the exam in order to avoid motion misregistration;
• The table may remain stationary during the entire exam, or move back and forth between a few table
positions;
• Whole-brain perfusion CT can be accomplished using CT systems with wide detector arrays (8-16 cm);
alternatively, scan modes that move the patient back and forth over the desired scan volumes can be
used;
• Acquisitions are repeated at predetermined time intervals (e.g. every second to every 2-3 seconds) for
a predetermined duration (e.g. 40-90 seconds);
• Relatively thick image widths are used to minimize image noise (5-10 mm is common);
• Image quality is inferior to a routine head CT. That is, images are noisier and thicker.
• Data are used to generate color maps of hemodynamic significance:
--Blood volume (BV) and flow (BF), mean transit time (MTT), time to peak perfusion (TPP);
• A non-contrast-enhanced head CT and/or a CT angiogram may be combined with a perfusion CT scan.
Dose Management
• 80 kV should be used to increase iodine signal brightness;
• Low dose per single scan (i.e. one tube rotation) is critical, since repeated scanning will result in a
relatively high cumulative dose;
• Time interval between scans, and hence the total number of scans over the exam duration, should be
set carefully, taking into account the analysis algorithm (some approaches require relatively dense data
points);
• Dose (tube current) modulation should not be used, as it may interfere with the calculation of the BV
and BF parameters;
Additional Resources
• ACR Practice Guideline for the Performance of Computed Tomography (CT) Perfusion in
Neuroradiologic Imaging. (www.acr.org/Quality-Safety/Standards-Guidelines/Practice-Guidelines-by-
Modality/CT);
• AJNR Special Collection. Radiation Dose in Neuroradiology CT Protocols. Collection Editors: Max
Wintermark and Michael H. Lev (available at www.ajnr.org/specCol/specCollPCTToc.dtl).
2
The disclaimer on page 1 is an integral part of this document.
Copyright © March 1, 2016 by AAPM. All rights reserved.
Adult Brain Perfusion CT Protocols Version 2.0 3/1/2016
BRAIN PERFUSION CT: Sample Images
Cerebral Blood Flow (CBF, in mL/100 g/min) Cerebral Blood Volume (CBV, in mL/100 g)
3
The disclaimer on page 1 is an integral part of this document.
Copyright © March 1, 2016 by AAPM. All rights reserved.
Adult Brain Perfusion CT Protocols Version 2.0 3/1/2016
GE
Hitachi
Neurologica
Neusoft
Philips
Siemens
Toshiba
4
The disclaimer on page 1 is an integral part of this document.
Copyright © March 1, 2016 by AAPM. All rights reserved.
Adult Brain Perfusion CT Protocols Version 2.0 3/1/2016
BRAIN PERFUSION CT (Selected GE Scanners)
GENERAL: This protocol may include an optional, non-contrast-enhanced head scan and/or an optional head CT angiogram.
Center the table height, such that the external auditory meatus is located at the center of the gantry and the landmarked
at the level of the canthomeatal line (S0). The patient’s chin should be tilted toward the chest (i.e. in a “tucked”
position) to minimize the amount of tilt needed to better avoid the eyes especially for modes that do not support tilt.
Perfusion protocols are for adults; modifications must be done for pediatrics.
Perfusion computations are performed on an image-processing workstation or scan console after scan completion.
Continued
5
The disclaimer on page 1 is an integral part of this document.
Copyright © March 1, 2016 by AAPM. All rights reserved.
Adult Brain Perfusion CT Protocols Version 2.0 3/1/2016
BRAIN PERFUSION CT (Selected GE Scanners) (continued)
Continued
6
The disclaimer on page 1 is an integral part of this document.
Copyright © March 1, 2016 by AAPM. All rights reserved.
Adult Brain Perfusion CT Protocols Version 2.0 3/1/2016
BRAIN PERFUSION CT (Selected GE Scanners) (continued)
Option 3: Volume mode (table moves in and out of the gantry to increase coverage).
LightSpeed VCT and LightSpeed VCT and
GE
Discovery CT750 HD Discovery CT750 HD
Volume Shuttle Volume Shuttle
Scan Type
Axial Helical
Rotation Time (s) 0.4 0.4
Detector Rows 64 64
Total Exposure Time 13.6 sec 47.38 sec
Pitch N/A 0.984:1
Passes 17 28
kVp 80 80
Manual mA 400 200
AutomA/SmartmA OFF OFF
SFOV Head Head
Prep Group 5 5
ISD N/A N/A
DFOV 25 25
Image Thickness 5mm x 8i 5 mm
Interval (mm) 40 10
Reconstruction
Standard Standard
Algorithm
SS50
ASiR SS50 (ON)
(optional, if available)
Coverage (mm) 80 110-120
Temporal Sampling (s) 3 sec 3 sec
CTDI-vol (mGy) 178 @ 400 mA 261 @ 200 mA
7
The disclaimer on page 1 is an integral part of this document.
Copyright © March 1, 2016 by AAPM. All rights reserved.
Adult Brain Perfusion CT Protocols Version 2.0 3/1/2016
BRAIN PERFUSION CT (Selected HITACHI Scanners)
Images obtained using this Dynamic Scan protocol can be processed using TeraRecon iNtuition™ TDA clinical application to obtain
quantitative perfusion parameters.
Use of other manufacturer’s perfusion software applications may be appropriate, but the accuracy of the results has not been validated
by either Hitachi or the AAPM. Users should work with the software vendor to determine whether the data obtained with this scan
protocol are appropriate for the specific algorithm used to calculate perfusion parameters.
8
The disclaimer on page 1 is an integral part of this document.
Copyright © March 1, 2016 by AAPM. All rights reserved.
Adult Brain Perfusion CT Protocols Version 2.0 3/1/2016
BRAIN PERFUSION CT (Selected NEUROLOGICA Scanners)
Scout: PA and Lateral, 200 mm coverage, 120kV at 3mA for both views.
Notes:
1. Scan voltage is limited to 100kV and cannot be changed by the CT technologist.
2. Maximum scan current is set to 6 mA.
3. Default scan time is 30 seconds. Lower mA setting is recommended for longer scans.
4. Low dose reference scan is available for scan localization prior to CTP scan.
9
The disclaimer on page 1 is an integral part of this document.
Copyright © March 1, 2016 by AAPM. All rights reserved.
Adult Brain Perfusion CT Protocols Version 2.0 3/1/2016
BRAIN PERFUSION CT (Selected NEUSOFT Scanners)
GENERAL: Protocols are designed and intended for adult patients only. The gantry should be tilted to position the scan field
parallel to the Frankfort horizontal plane. Functional Imaging is used measure physiological processes in the brain.
Using a continuous or sequentially acquired dataset, at one anatomical position, the absorption of contrast material into
the brain tissue can be measured and quantified.
S COUT (NV 16): Lateral and PA, Base of skull through vertex. 250 mm in length, 40 mA, 90 kVp
S COUT (NV 64): Lateral and PA, base of skull through vertex. 250 mm in length, 40 mA, 80 kVp
After perfusion scan is complete, load perfusion images to the Brain Perfusion Application.
kVp 80 90
Reference
150 96
mAs
Pitch n/a n/a
RECON 1
Type Axial Axial
Filter F10 SA
Thickness
5 6
(mm)
Increment 5 6
(mm)
10
The disclaimer on page 1 is an integral part of this document.
Copyright © March 1, 2016 by AAPM. All rights reserved.
Adult Brain Perfusion CT Protocols Version 2.0 3/1/2016
BRAIN PERFUSION CT (Selected PHILIPS Scanners)
GENERAL: These protocol parameters should not be used for pediatric patients.
Option 2: Jog Mode (Table moves back and forth between two positions).
Brilliance Brilliance Brilliance Brilliance
PHILIPS
16 slice 40/64 slice iCT SP iCT
Rotation Time (s) 0.5 0.5 0.4 0.4
Collimation 16 × 1.5 mm 32 × 1.25 mm 32 × 1.25 mm 64 × 1.25 mm
Coverage (mm) 48 80 80 160
kVp 90 80 80 80
mAs 125 125 100 100
ACS/DOM OFF OFF OFF OFF
Cycle Time (s)* 4 4 4 4
# of Jog Cycles 15 15 15 15
Thickness (mm) 6.0 5.0 5.0 5.0
Increment (mm) 0.0 0.0 0.0 0.0
Resolution Standard Standard Standard Standard
FOV (mm) 250 250 220 220
Filter UB UB UB UB
WC/WL 80/40 80/40 80/40 80/40
CTDI-vol (mGy) 120 66 80 72
* Cycle time represents the time from the start of one scan to the start of the next scan over the same piece of
anatomy (i.e., the sampling interval of the time attenuation curve). For the 4 s cycle time, the manufacturer’s
perfusion analysis software reports relative, rather than absolute, perfusion parameters. Absolute, quantitative
perfusion parameters are reported for cycle times less than or equal to 2.5 s.
11
The disclaimer on page 1 is an integral part of this document.
Copyright © March 1, 2016 by AAPM. All rights reserved.
Adult Brain Perfusion CT Protocols Version 2.0 3/1/2016
BRAIN PERFUSION CT (Selected SIEMENS Scanners)
GENERAL: This protocol may include an optional, non-contrast-enhanced head scan and/or an optional head CT angiogram
Center the table height, such that the external auditory meatus is located at the center of the gantry.
The patient’s chin should be tilted toward the chest (i.e. in a “tucked” position).
TOPOGRAM: PA and Lateral, 512 mm coverage, 120 kV, 100 mA. Craniocaudal direction.
Perspective 64
Emotion 16 S cope Power S ensation 64
S IEMENS Perspective 128
Scan M ode M ultiscan M ultiscan M ultiscan M ultiscan
Rotation Time (s) 1.0 1.0 1.0 1.0
Table M otion None None None None
Detector Configuration (mm) 6x3 16 x 1.2 32 x 1.2 24 x 1.2
Coverage per Rotation (mm) 18 19.2 38.4 28.8
Scan Range (mm) 18 19.2 38.4 28.8
Cycle Time (s) 3.75 3 3 1.0 (continuous)
Feed (mm/rot) - - - -
kV 80 80 80 80
mAs 150 150 150 200
Scan Field (mm) 200 200 200 200
Scan time (s) 40 40 40 40
CARE Dose4D OFF OFF OFF OFF
-
CARE kV - - -
CTDIvol (mGy) 394 394 391 297
RECON 1
Kernel H31s H31s H31s H31s
Slice (mm) 9.6 9.6 9.6 9.6
Increment (mm) 9.6 9.6 9.6 9.6
12
The disclaimer on page 1 is an integral part of this document.
Copyright © March 1, 2016 by AAPM. All rights reserved.
Adult Brain Perfusion CT Protocols Version 2.0 3/1/2016
BRAIN PERFUSION CT (Selected SIEMENS Scanners) (Continued)
RECON 1
Kernel H20f H20f H20f H20f Hr36
Slice (mm) 5.0 5.0 5.0 5.0 5.0
Increment (mm) 3.0 3.0 3.0 3.0 3.0
13
The disclaimer on page 1 is an integral part of this document.
Copyright © March 1, 2016 by AAPM. All rights reserved.
Adult Brain Perfusion CT Protocols Version 2.0 3/1/2016
BRAIN PERFUSION CT (Selected TOSHIBA Scanners)
GENERAL: This protocol may include an optional, non-contrast-enhanced head scan and/or an optional head CT angiogram
Center the table height, such that the external auditory meatus is located at the center of the gantry.
S CANOGRAM: PA and Lateral, 240 mm coverage, 120 kV, 50 mA, caudal-cranial direction.
Scan Type Dynamic Volume Continuous Dynamic Volume Intermittent Dynamic Volume Intermittent
Rotation Time (s) 1.5 0.75 0.75
Table M otion None None None
Detector Configuration 10 x 4 mm 160 x 0.5 mm 320 x 0.5 mm
Coverage per Rotation (mm) 40 80 160
Scan Range (mm) 40 80 160
Acquisition Interval (s)* Continuous 2 s initially, then 5 s 2 s initially, then 5 s
kV 80 80 80
310 (mask), 300 (arterial phase), 310 (mask), 300 (arterial phase),
mA 40
150 (elsewhere) 150 (elsewhere)
SURE
Exposure No No No
SFOV (mm) 240mm (S) 240mm (S) 240mm (S)
Delay after injection (s) 5 7 7
Scan Time (s) 49.5 53 53
CTDI-vol (mGy) 128.8 196.8 214.6
RECONS TRUCTION
Kernel 43 41 41
Thickness (mm) 10 0.5 0.5
Interval (mm) 10 0.5 0.5
DFOV (cm) 240 240 240
14
The disclaimer on page 1 is an integral part of this document.
Copyright © March 1, 2016 by AAPM. All rights reserved.