temperature measurement
Applications
Special features
■■ Application ranges from -200 ... +500 °C (-328 ... +932 °F)
■■ Compact design
■■ Universal application
■■ Direct installation into the process
■■ Explosion-protected versions
Description
3157966.01
Legend:
Connection head
Process connection
Thermowell
Terminal block
Connection head
JS
JS BS BSZ, BSZ-K BSZ-H, BSZ-HK BSS BS
Model Material Cable outlet Ingress protection Cap Surface
JS Aluminium M16 x 1.5 1) IP65 Cap with 2 screws Blue, lacquered 2)
1) Standard
2) RAL 5022
1) Not with 2-wire connection method Ferrules from stainless steel are only adjustable once; once
the fitting has been loosened, sliding along the thermowell is
For detailed specifications for Pt100 sensors, see Technical no longer possible.
information IN 00.17 at www.wika.com.
Ferrules from PTFE can be adjusted numerous times; once
■■ At the head: -40 ... +80 °C the fitting has been loosened it can again be tightened onto
■■ Storage: -40 ... +80 °C the thermowell.
■■ Max. temperature at process connection: 150 °C
Process connections
Double nipple
All process connections are manufactured from stainless
steel. Other materials are available on request. Using a double-sided threaded nipple, the thermometer
can be screwed directly into the process. In this case the
The insertion length A (U1 or U2) can be customised. permissible temperature ranges must be observed.
The neck length, N (MH), depends on the type of the process The neck length, N (MH), for parallel threads depends on the
connection selected. height of the hexagon. This is 10 mm.
The neck length N (MH) of NPT threads not only includes the
Extended process connection hexagon height but also half of the thread height. This gives
us a neck length, N (MH), of approx. 19 mm.
Connection heads, connecting leads/wires and the optional
transmitter must only be used within the above-mentioned
temperature ranges. Without process connection
If the thermometer will operate at temperatures outside of the This version is designed mainly for mounting in one of the
temperature limits, the clearance between the connection available compression fittings.
head and the hot or cold surfaces must be increased.
The neck length, N (MH), in this case only specifies the
This neck length is dependent upon the intended application height of the hexagon on the head of the thermowell.
and generally serves as isolation or as a cooling element N (MH) is always 7 mm.
between the process and the connection head.
The resistance thermometers of the TR10-D series are designed for direct installation into the process. Using it in an additional
thermowell only makes sense in exceptional cases.
11346418.01
Without process Threaded connection Compression fitting
connection Double nipple (welded) (adjustable)
Thread
Legend:
A (U1) Insertion length N (MH) Neck length
(with parallel threads) (U) Insertion length per DIN 43772
A (U2) Insertion length (U1) Insertion length in accordance with international standards
(with tapered threads) Ød Measuring insert diameter
Electrical connection
black yellow
For the electrical connections of built-in temperature transmitters see the corresponding data sheets or operating instructions.
Resistance thermometers of the TR10-D series are available Certification type Measuring Material
with an EC-type examination certificate for “intrinsically safe”, accuracy certificate
Ex i, ignition protection. 2.2 test report x x
These instruments comply with the requirements of the ATEX
3.1 inspection certificate x -
directive for gases and dusts.
DKD/DAkkS calibration x -
certificate
The permissible power Pmax as well as the permissible
ambient temperature for the respective category can be seen The different certifications can be combined with each other.
on the EC-type examination certificate and in the operating
instructions.
Approvals
Ordering information
Model / Version / Measuring insert / Explosion protection / Process connection / Version and material of threaded connection /
Thread size / Measuring element / Connection method / Temperature range / Design of the sensor tip / Sensor diameter /
Insertion length A / Neck length N (MH) / Certificates / Options
© 2008 WIKA Alexander Wiegand SE & Co. KG, all rights reserved.
The specifications given in this document represent the state of engineering at the time of publishing.