Ultrasonic Guide
Ultrasonic Guide
Ultrasonic Guide
FAKOPP
ULTRASONIC TIMER
www.fakopp.com
1
Introduction
Operation principle
received
1.5 excitation pulse/100
200 signal
Amplitude [v]
0.5
µ s]
Time [µ
-0.5 0 20 40 60 80
threshold
Figure 1. Timer starts, when excitation pulse rise and stops when
received signal reaches the threshold level. (The received
signal polarity is opposite in this unit!)
2
Time correction:
The transit time appears on the screen, includes the transit time inside
the transducer. The wave guide (triangle or spike) is part of the
transducer. If we measure the distance between the screw locations
(center of triangle wave guide) we have to subtract 6.1µs from the
readout, to get the correct transit time in the sample. Using this
corrected time and distance data, correct ultrasonic velocity
determination is possible. Distance defined as the distance between the
transducer center, see figure 2.
Distance
3
120
y = 2.254x + 6.1
100 2
R = 0.9999
80
time [µ s]
60
40
20
0
0 10 20 30 40 50
distance [cm]
Velocity determination:
V[m/s]=1000*distance[mm] / ( read out[µs]-time correction)
distance
4
100
y = 2,5x + 9,4
80
R2 = 0,9998
time [µs]
60
40
20
0
0 5 10 15 20 25 30 35 40
distance [cm]
Installation
Sensitivity considerations
y = 20,242e-0,0033x
R2 = 0,9636
Amplitude (mV)
10
1
0 200 400 600 800 1000
y = 53,245e-0,0167x
R2 = 0,9918
0,1
distance (mm)
After a point the amplitude of the first peak becomes lower than the
threshold level and the read out becomes 11 µs higher because the low
signal.
140
120
100
tim e ( s)
80
60
40
20
0
0 10 20 30 40 50
distance (cm )
7
To avoid the situation in figure 8, we recommend to use short
transducer distance. The recommended transducer distance is 25-30
cm only.
Transducers
for veneer
The 45 kHz piezo electric transducers are designed specially for wood
and veneer evaluation. A special triangle shape wave guide makes the
transducer suitable for longitudinal test when the side (no end surface)
of the sample is accessible. The required force for proper coupling is
100-200N.
Maintenance
Battery change
The ultrasonic timer has two independent circuits, the use of two
batteries is a mist. The timer unit has “LOW BAT” indicator. When
voltage goes below 8,5 V the “LOW BAT” sign appears at the top right
corner of the display. This case please change battery.
When the excitation circuit battery is low, no signal is generates and the
start led stops flashing. This case please change battery.
IMPORTANT: Always switch of unit before battery change!!!
Open battery compartment (figure 11.) and change the battery.
You can recharge the battery. Unfortunately we are not able to provide
battery charger for 110V.
9
Figure 9: The open battery compartment.
Guarantee:
For the Ultrasonic Timer device the guarantee is one year. The
guarantee starts at the date displayed on the bill and valid for the
following one year. The repair can be asked from the manufacturer,
fax: +36 99 33 00 99. The guarantee does not concern to the break or
deliberate damage of the machine.
10
Technical data
11