453564282181-G60g70g80-Lp-Sg-Eng 2.0
453564282181-G60g70g80-Lp-Sg-Eng 2.0
453564282181-G60g70g80-Lp-Sg-Eng 2.0
Service Guide
English
G60/G70/G80
Modular Patient Monitors
Service Guide
English
Product Information
Product model: G60/G70/G80
Product name: Modular Patient Monitor
Manufacturer
Philips Goldway (Shenzhen) Industrial Inc.
Registered/Production address:
Edition
Second Edition: May 2014
Standard
The product is made under the ISO9001 and ISO13485 quality system certified by TUVPS.
Declaration
The G60/G70/G80 modular patient monitor is a Class IIb device and complies with the requirements of the
Council Directive 93/42/EEC concerning medical devices and carries CE-marking accordingly.
Authorized EU Representative
Shanghai International Holding Corp.GmbH (Europe)
Eiffestrasse 80, 20537
Hamburg, Germany
Tel: 0049-40-2513175
Fax: 0049-40-255726
Rx Only
Caution: United States Federal Law restricts this device to sale by or on the order of a physician.
Parameter Specification
Protection Class Class I, anti-shock, externally and internally powered equipment, per IEC
60601-1
Philips Goldway is responsible for the safety, reliability and performance of the monitor when the:
The international call center can be reached during the following time:
E-mail: goldway.service@philips.com
Caution A Caution calls attention to a condition or possible situation that could damage or destroy the product
or the user’s work.
Warning A Warning calls attention to a condition or possible situation that could cause injury to the user and/or
patient.
Safety Requirements
Note — The safety indications in this chapter apply to general monitor use. Safety indications in other
chapters apply to specific monitor measurements.
Follow the instructions of this service guide when using the monitor. However, conventional medical practices
always supersede this document. Significance of safety requirements set forth here in this manual is not in
order of reading sequence.
The monitor is not an apnea monitor. The respiration measurement does not recognize obstructive and
mixed apneas it only indicates when a user-defined time has elapsed since the last detected breath.
Anyone who connects additional equipment to the signal input port or signal output port configures a
medical system and is therefore responsible to ensure that the system complies with the requirements of
standard IEC 60601-1. Never permanently install equipment connected to the signal input or output
ports. If in doubt, contact Philips Goldway.
To avoid mixture, clear all historical data of the last patient before monitoring a new patient.
Ensure that the monitor is in good working condition and is placed in proper position before clinical use.
Never rely exclusively on the alarm system for patient monitoring. You must periodically check that
monitor alarms are working properly. The most reliable method of patient monitoring combines close,
personal surveillance along with the correct operation of the monitor.
For pacemaker patients, rate meters may continue to count the pacemaker rate during occurrences of
cardiac arrest or come arrhythmia. Do not rely entirely upon rate meter alarms. Keep pacemaker
patients under close surveillance. See this manual for disclosure of the pacemaker pulse rejection
capability of this instrument.
Never simultaneously touch the patient and the input or output terminal of the monitor.
The monitor can be used with electrosurgical devices. To avoid burns or death, ensure that the
electrosurgical circuitry is properly connected.
Ensure that the patient is not connected to any metal, the surface of any conductors, or grounded parts of
any devices when a defibrillator is in use. Never touch the patient, table, or instruments during
defibrillation.
Disconnect the AC power supply and remove the battery before disassembling the monitor.
Caution To avoid personal injury, only use accessories and parts produced or recommended by Philips Goldway.
Otherwise, damage to the monitor can occur.
Position power cords and other cables to prevent tangling, patient contact, or electrical interference.
Always use a three-wire receptacle in order to properly ground the monitor. Use the monitor on battery
power only if you suspect any problem with external grounding or wiring.
Only connect the monitor to a power outlet with correct protective grounding. If no grounded power is
available, only use the monitor with internal batteries.
The battery will discharge over time when stored inside a monitor with no AC power. Keep the monitor
plugged in when storing it and perform a monthly check of battery power.
If multiple devices are connected to a patient, the sum of the leakage currents may cause a dangerous
situation. Therefore, qualified service representatives should always perform current leakage tests
before setting up the connections. This ensures that leakage is within limits and prevents personal injury
and environmental damage. If any question remains, consult the manufacturer for correct usage.
The monitor must perform with the international standard IEC 60601-1-2 and other applicable EMC
standards. Interference takes place when electromagnetic energy is extremely high. Ensure that any
nearby instruments are also in compliance with EMC standards. Never turn on or use portable
communication devices like mobile phones or portable dual-channel radios near a monitor.
Any devices connected to the monitor must comply with the IEC standard for that device. For example,
data processing devices must comply with IEC 60950 and medical devices must conform to IEC 60601-1.
The whole system must comply with the most recent IEC 60601-1-1 standards.
To ensure monitor safety, all service parts and accessories must comply with IEC 60601 standards. The
system configuration of the monitor must comply with IEC 60601-1-1 medical electrical standards.
If patient data is lost, closely monitor the patient or replace the monitor immediately until the monitor
restores normal function again.
Monitoring will be stopped while user configurations will be saved automatically when the monitor is
powered off.
To reduce the hazard of burns in high frequency surgical neutral connections, the leads and connectors
must be located away from the surgical site. Keep cables away from other devices.
Ensure that qualified service representatives calibrate and maintain the monitor on an annual basis.
Periodically check all reusable accessories for damage. Replace and dispose of damaged accessories
when necessary according to your local hospital waste disposal regulations.
Clean and sterilize the monitor and accessories according to local requirements. Turn off the monitor
and disconnect the power cords before cleaning or sterilization.
Keep all monitor packing materials away from children, or dispose of them in accordance with your
local environmental regulations.
Always properly dispose of the monitor and all accessories at the end of their service life. Dispose of
batteries according to your local regulations. Never incinerate batteries or expose them to high
temperatures.
Ensure that no water condenses into or on the monitor. Condensation can occur from changes in
temperature or exposure to humidity.
Environmental Specifications
Temperature Range Operating 10 oC to 40 oC (50 oF to 104 oF)
Typical operating time 2 hours (with ECG, SpO2 measurement, and NIBP
measured every 15 minutes)
ECG (12-lead) ○ ○ ○
ECG ST/ARR ● ● ●
RESP ● ● ●
Multi-parameter
NIBP ● ● ● module
SpO2 ● ● ●
TEMP1/TEMP2 ● ● ●
IBP1/IBP2 ○ ○ ○
Recorder ○ ○ ○ N/A
Note — If a Multi-parameter module with IBP1/IBP2 and a dual IBP module are placed into the unit, there
will be four IBP channels numbered IBP1, IBP2, IBP3, and IBP4 displayed on the screen. The channel
positions are changed according to the order in which the measurement modules are recognized and/or placed
into the unit.
Warning Never attempt to open the monitor case. Only qualified personnel should service the monitors.
Never block the monitor’s air vent during the operation.
Note — Only connect the monitor to a power outlet with correct protective grounding. If no grounded power
is available, only use the monitor with the internal battery.
2. Connect the power cord to the monitor and to a single-phase AC power source.
3. Check that the power light on the front panel is yellow. This means that the AC power is correctly
connected.
Warning Personnel connecting additional devices to input and output connectors are responsible for system
compliance with the IEC 60601-1-1 standard. If you cannot determine the safety of the monitor when
combining it with another medical device, contact Philips Goldway to ensure that the combined devices
are safe and will not cause electric shock or other hazards.
When safety (for example an electrical shock caused by the sum of current leakage) cannot be determined
from the specific combination of the monitor and other medical devices, please contact your service
representative.
Caution To maintain normal operation of the monitor, develop a good maintenance plan for periodic cleaning,
maintenance and service. For more information, see Chapter 2 "Testing and Maintenance".
Opening the monitor for any reason Power supply and battery test
Display test
Touchscreen test
Speaker test
All safety tests
Caution The frequency recommendations in the following table do not supersede local requirements. Always
perform locally required testing in addition to the testing outlined in the table.
Tests Frequency
Safety (In accordance with IEC 60601-1)
• System enclosure leakage current Once every year and after repairs where the unit has been
opened (front and back separated) or the monitor has been
• Ground integrity
damaged by impact.
• Patient leakage current
Test Equipment
"Visual Test" None
Test Equipment
"Touchscreen Test" None
10. Check the item “RELATIVE_CHARGE” and record the number displayed.
11. Connect the monitor to an AC power source and ensure the charging LED flashes.
12. Wait 2-5 minutes and check the RELATIVE_CHARGE value again.
Each time you connect AC power and observe a flashing charging LED, the RELATIVE_CHARGE value
should increase. The battery capacity test is complete.
4. Rotate the navigation wheel to Alarm Volume and press the wheel.
5. Rotate the navigation wheel to adjust the alarm volume.
You should hear a tone from the speaker.The speaker test is complete.
Method 2:
1. Turn on the G60/G70/G80 Modular Patient Monitor.
2. Insert a parameter module (see G Series Modular Monitor Parameter Modules Service Guide) into
module slot.
You should hear a tone from the speaker. The speaker test is complete.
Figure 2-2 S(1) Part 2: System Enclosure Leakage current - Single Fault (open earth)
5. Touch the green circle with a touch pen or your finger. It will appear at 5 different locations.
6. Restart the monitor.
The touchscreen test is complete.
Note — The items displayed in red rectangle in Figure 3-1 indicates the external connection functional
modules, and those in the blue rectangle are optional configuration.
3.2.6 Speaker
The speaker is driven by an amplifier. The speaker provides the monitor’s audio during alarm conditions. The
microprocessor outputs different audio tones specified by different alarm priorities and conditions.
3.2.7 Recorder
The recorder contains the following functions:
• Motor (paper drive) and driver circuitry
• Selectable print speed
The fuse is blown. Replace the fuse. See "Removing the AC Power
Connector Fuse" on page 5-2, "Removing the AC
Power Connector Fuse" on page 6-3 or "Removing
the AC Power Connector Fuse" on page 7-3.
The AC socket board has Replace the AC socket board. See "Removing the
malfunctioned. AC/DC Power Module" on page 5-4, "Removing the
AC/DC Power Module" on page 6-3 or "Removing
the AC/DC Power Module" on page 7-3.
The DC power cable has Replace the AC connector.See "Removing the AC/
malfunctioned. DC Power Module" on page 5-4, "Removing the
AC/DC Power Module" on page 6-3 or "Removing
the AC/DC Power Module" on page 7-3.
The keypad board cable is Ensure that the AC power cord is plugged into.
unplugged or broken. Replace the keypad board cable. See "Removing the
(G60) Keypad Board" on page 5-7.
The keypad board is Replace the keypad board. See "Removing the
broken.(G60) Keypad Board" on page 5-7.
The main board cable is Ensure that the AC power cord is plugged into.
unplugged or Replace the main board cable. See "Removing the
broken.(G70/G80) Main Board" on page 6-16 or "Removing the Main
Board" on page 7-13.
The main board has Replace main board. See "Removing the Main
malfunctioned. Board" on page 5-13, "Removing the Main Board"
on page 6-16 or "Removing the Main Board" on
page 7-13.
The keypad board cable is Ensure that the AC power cord is plugged into.
unplugged or broken. Replace the keypad board cable. See "Removing the
(G60) Keypad Board" on page 5-7.
The keypad board is Replace the keypad board. See "Removing the
broken.(G60) Keypad Board" on page 5-7.
The main board cable is Ensure that the AC power cord is plugged into.
unplugged or broken. Replace the main board cable. See "Removing the
(G70/G80) Main Board" on page 6-16 or "Removing the Main
Board" on page 7-13.
The main board has Replace the main board. See "Removing the Main
malfunctioned. Board" on page 5-13, "Removing the Main Board"
on page 6-16 or "Removing the Main Board" on
page 7-13.
The battery has fully Charge the battery. See "Removing/Replacing the
exhausted. Battery" on page 5-1, "Removing/Replacing the
Battery" on page 6-1 or "Removing/Replacing the
Battery" on page 7-1.
The battery has fully Charge the battery. See "Removing the Main Board"
exhausted. on page 5-13, "Removing the Main Board" on
page 6-16 or "Removing the Main Board" on
page 7-13.
The battery can not charge Change the battery. See "Removing/Replacing the
Battery" on page 5-1, "Removing/Replacing the
Battery" on page 6-1, or "Removing/Replacing the
Battery" on page 7-1.
The main board has Replace the main board. See "Removing the Main
malfunctioned. Board" on page 5-13, "Removing the Main Board"
on page 6-16 or "Removing the Main Board" on
page 7-13.
The inverter has Replace the inverter. See "Removing the Keypad
malfunctioned.(G60) Board" on page 5-7.
The display has Check the display, replace the display. See
malfunctioned. "Removing the LCD Display Assembly" on
page 5-9, "Removing the LCD Display Assembly"
on page 6-10 or "Removing the LED Backlight
Board" on page 7-8.
There is a bad inverter. Replace the inverter. See "Removing the Keypad
(G60) Board" on page 5-7.
The keypad board has Replace the keypad board. See "Removing the Keypad
malfunctioned. (G60) Board" on page 5-7.
The main board has Replace the main board. See "Removing the Main
malfunctioned. (G70/G80) Board" on page 6-16 or "Removing the Main Board"
on page 7-13.
The monitor The LCD cable is not Open the monitor and reseat the LCD signal cable.
displays random/ attached correctly
distorted graphics
with a white The LCD has Replace the Display assembly. See "Removing the
background. malfunctioned. LCD Display Assembly" on page 5-9, "Removing the
LCD Display Assembly" on page 6-10 or "Removing
the LED Backlight Board" on page 7-8.
The keypad board has Replace the keypad board. See "Removing the Keypad
malfunctioned. (G60) Board" on page 5-7.
The main board has Replace the main board. See "Removing the Main
malfunctioned. (G70/G80) Board" on page 6-16 or "Removing the Main Board"
on page 7-13.
I turn on the The main board has Replace the main board. See "Removing the Main
monitor and the malfunctioned. Board" on page 5-13, "Removing the Main Board" on
opening screen page 6-16 or "Removing the Main Board" on
and/or color bar page 7-13.
starts, but then
just hangs.
The monitor does There is a bad connection Ensure that all connectors to the main board are seated
not respond to the to the main board. correctly.
navigation wheel
or front panel The monitor keypad or Replace the front case. See "Removing the Front Case
buttons navigation wheel is Assembly" on page 5-5, "Removing the Front Case
broken. Assembly" on page 6-6 or "Removing the Front Case
Assembly" on page 7-6.
The main board has Replace the main board. See "Removing the Main
malfunctioned. Board" on page 5-13, "Removing the Main Board" on
page 6-16 or "Removing the Main Board" on
page 7-13.
Can not display the The battery is broken. See "Removing/Replacing the Battery" on
battery information page 5-1, "Removing/Replacing the Battery" on
screen. page 6-1, or "Removing/Replacing the Battery"
on page 7-1.
The keypad board is Replace the keypad board. See "Removing the
broken.(G60) Keypad Board" on page 5-7.
Continuously charging The battery has Replace the battery. See "Removing/Replacing
the battery more than 4 malfunctioned. the Battery" on page 5-1, "Removing/Replacing
hours, the charging the Battery" on page 6-1, or "Removing/
LED is not green. Replacing the Battery" on page 7-1.
The speaker is bad. Replace the fan and speaker assembly. See
"Removing the Fan and Speaker Assembly" on
page 5-16, "Removing Fan and Speaker Assembly"
on page 6-19 or "Removing Fan and Speaker
Assembly" on page 7-16.
There is a bad connection Ensure that all connectors to the keypad board are
to the keypad board. (G60) seated correctly.
There is a bad connection Ensure that all connectors to the main board are
to the main board. (G70/ seated correctly.
G80)
The main board has Replace the main board. See "Removing the Main
malfunctioned. Board" on page 5-13, "Removing the Main Board"
on page 6-16 or "Removing the Main Board" on
page 7-13.
The NIBP module has not Reset the NIBP module case.
set in correctly.
There are no NIBP Wrong cuff size or Use proper cuff size, ensure proper cuff
readings. incorrect cuff placement. placement.
The tube is kinked or there Ensure that the tube is straight and not kinked.
is air leakage in the air Replace the cuff and ensure that there is no air
tube or cuff. leakage in tube.
There has been some Ensure that all external blood pressure reading
external problem. requirements are met and that the patient is not
moving excessively.
The NIBP module has not Reset the NIBP module case.
set in correctly.
The main board has Replace the main board. See "Removing the
malfunctioned. Main Board" on page 5-13, "Removing the Main
Board" on page 6-16 or "Removing the Main
Board" on page 7-13.
I’m getting unreliable Wrong cuff size or Use proper cuff size, ensure proper cuff
NIBP readings. incorrect cuff placement. placement.
There have been some Ensure that all external blood pressure reading
external problems. requirements are met and that the patient is not
moving excessively.
I’m getting unreliable The probe has Replace the temperature probe.
temperature readings malfunctioned.
The main board has Replace the main board. See "Removing the
malfunctioned. Main Board" on page 5-13, "Removing the Main
Board" on page 6-16 or "Removing the Main
Board" on page 7-13.
The SpO2 readings are There has been some Ensure that all external SpO2 reading
unreliable external problem. requirements are met and that the patient is not
moving excessively.
The CO2 tube is kinked. Check the tube to ensure that it is straight.
The main board has Replace the main board. See "Removing the
malfunctioned. Main Board" on page 5-13, "Removing the Main
Board" on page 6-16 or "Removing the Main
Board" on page 7-13.
The CO2 readings area The CO2 tube is kinked. Check the tube and make sure that it is straight.
unreliable
There have been some Ensure that all external CO2 reading
external problems. requirements are met and that the patient is not
moving excessively.
The ECG cable is broken. Ensure that the cable is in good working order.
The main board has Replace the main board. See "Removing the
malfunctioned. Main Board" on page 5-13, "Removing the Main
Board" on page 6-16 or "Removing the Main
Board" on page 7-13.
I’m getting unreliable The ECG leads are not Ensure that the ECG leads are connected.
ECG readings connected.
There has been some Ensure that all external ECG reading
external problem. requirements are met and that the patient is not
moving excessively. Ensure proper skin
preparation has been performed.
The respiration There has been some Ensure that all respiration reading requirements
readings are unreliable external interference. are met and that the patient is not moving
excessively. Ensure that the correct ECG lead set
is being used.
The recorder does not The paper is out and/or the Install a new paper supply. Open the recorder
print. recorder door is open. door, and then close it again.
The main board has Replace the main board. See "Removing the
malfunctioned. Main Board" on page 5-13, "Removing the Main
Board" on page 6-16 or "Removing the Main
Board" on page 7-13.
Caution Before removing the battery, perform a monitor shutdown and disconnect the monitor from AC
power.
Before putting a new battery into use, perform at least two complete reconditioning cycles. This improves the
battery life and increases performance specifications. A reconditioning cycle is complete (charging and
discharging the battery) when the monitor powers off from lack of battery power. Perform a recondition when
battery performance is shortened or when the battery is stored for 2 months or more.
To recondition the battery:
1. Disconnect the monitor from the patient and turn it off.
2. Place the battery you want to recondition into the monitor battery compartment (see "Removing/
Replacing the Battery" on page 5-1).
3. Connect the monitor to AC power and continuously charge the battery for more than 8 hours.
4. Disconnect the monitor from AC power and power the monitor with the battery until the monitor turns
off.
5. Connect the monitor to AC power and continuously charge the battery for more than 8 hours.
Reconditioning is complete.
2. Pull out the fuse holder and remove and replace the fuse as shown.
3. To replace the fuse holder, push it back into the power connector until it snaps into the place.
2. Remove the paper roll, and remove the two M3X8 screws, round head with a 3mm Phillips head screw
driver. Note the screw gaskets.
3. Use your thumbs to grip the inside walls of the recorder case and pull it straight out of the assembly and
remove the recorder connector cable.
Power Cord
Lock Catch
VGA Protective
Cover
DC Power Cables
Navigation Wheel
Rubber Foot
3. To open the front case assembly, lift the bottom edge of the assembly about 1- 2 cm by holding the two
sides of the assembly. Gently push the front case assembly in the direction of the arrow.
4. Turn the monitor over and slightly pull the front case assembly away from the monitor case.
6. Remove the ground wire connecting the LED screen and the main frame.
Ground Wire
2. Take off the keypad board, and then the silicon keypad is released as well.
2. Separate the front panel ASSY and the LCD plate gently.
LCD
3. Remove the four M3X6 screws to separate the LCD from the LCD plate.
Infrared connection
oard cable
ainboard
Infrared able
2. Remove the five M3X6 screws marked in yellow and two M4X10 screws with gasket.
3. Remove the four M4X12 screws with gasket on the bottom of the case.
Battery Cable
2. Remove the seven M3X6 screws as shown.
3. Carefully remove the main board from the external communication board. Note the connection pins
between the main board and the external communication board.
5.7.5.1 Removing the External Connection Assembly from the Main Frame
To remove the data output board from the data connectors assembly:
1. Remove the two M3X6 screws.
2. Install the stud bolt with the spring into the installation hole on the modular frame.
Caution Before removing the battery, perform a monitor shutdown and disconnect the monitor from AC
power.
Before putting a new battery into use, perform at least two complete reconditioning cycles. This improves the
battery life and increases performance specifications. A reconditioning cycle is complete (charging and
discharging the battery) when the monitor powers off from lack of battery power. Perform a recondition when
battery performance is shortened or when the battery is stored for 2 months or more.
To recondition the battery:
2. Pull out the fuse holder, remove and replace the fuse as shown. Then push the fuse holder back into the
power connector until it snaps into the place.
Power Cord
Lock Catch
VGA Protective
Cover
2. Use the left hand to open the left side of the back cover, and use the right hand to push the right side of
the back cover towards the direction of the arrow.
Grounding Plug
AC Power Cable
7. Remove four M3x8, round head screws fixing one AC/DC power module as shown and repeat the same
step to remove the second.
2. Turn over the monitor and make the LED screen towards up. Open the front case assembly from the
bottom as shown.
3. Lift the bottom edge of the front case assembly about 1~2 cm and hold the two sides of the front case
assembly. Gently push the front case assembly towards the direction as shown to loosen the top edge of
the front case assembly.
4. Turn over the monitor and half-open the front case assembly as shown.
Grounding Wire
Keyboard Cable
Backlight
Power Cable
Backlight
Power Cable
2. Separate the front panel ASSY and the LCD plate gently.
LCD
LCD Plate
4. Separate the LCD plate and the LCD ASSY. Remove the LCD cable and backlight power cable.
LCD ASSY
LCD Plate
LCD Cable Backlight Power Cable
Note — No need to remove the cable on the right on the LCD ASSY. When replacing LCD ASSY, stick four
dust foam on the front edge of the LCD ASSY.
2. Pull up the alarm board and remove the alarm LED cable.
2. Remove the paper roll, and remove the two M3x8 screws, round head with a 3 mm Phillips head screw
driver. Note the screw gaskets.
3. Remove the cable tie and remove the recorder connector cable.
4. Use your thumbs to grip the inside walls of the recorder case and pull it straight out of the assembly.
2. Remove five M3x6 screws (black circle) and two M4x10 (blue circle) screws
3. Remove four M 4x12 screws on the bottom of the rear case, with washers.
Note — When installing a new clock battery, please note the battery’s polarity.
If you can not remove the main frame after you finished steps 1-4, remove the battery compartment door and
try again.
Battery Cable
2. Unplug the main board from the external communication module very carefully.
Connector Film
3. Remove the four M3x4 flat head screws to loosen the speaker and four screws to loosen the fan.
2. Install the stud bolt with the spring into the installation hole on the modular frame.
7.1 G60
2
1
1 2 3
4 5
7 8
9
10 11 12
Figure 7-2 G60 Front Case Assembly Exploded Diagram
1
2
4
3 6 8
5
18
7
9
17
16
12
13
10 11 14 15
Figure 7-3 G60 Rear Case Assembly Exploded Diagram
not shown SLC-YD MACH PWR Cord Lock Catch, SUS304 453564297361
7.2 G70
b
1
5
4
1 2 3
10
7
11
6
5
4
3
1 2
7.3 G80
b
1
5
4
1 2 3
10
7
11
6
5
4
3
1 2
Warning Use of accessories, transducers, and cables other than those specified may result in increased emissions
and/or decreased immunity of the G60/G70/G80 Modular Patient Monitor.
G60/G70/G80 Modular Patient Monitor should not be used adjacent to or stacked with other
equipment. If adjacent or stacked use is necessary, the monitor should be observed to verify normal
operation in the configuration in which it is used.
Caution The following ports on G60/G70/G80 Modular Patient Monitor are sensitive to electrostatic discharge
during patient monitoring:
• SpO2
• ECG
Follow these guidelines to reduce electrostatic discharge:
• Do not touch the connector pins of the port.
• When connecting and disconnecting cables to the ports, use ESD precautionary measures.
Precautionary measures may include the following:
– Make sure you are grounded prior to connection or touch grounded metal.
– Wear an ESD grounding strap and an ESD protective garment.
– Provide ESD training to all users. Users should understand what electrostatic discharge is, how
damage can occur, and how to prevent ESD.
RF emission Class A
CISPR 11 The G60/G70/G80 Modular Patient Monitor is suitable for use in
all establishments other than domestic and those directly
Harmonic emissions Class A connected to a low-voltage power supply network which supplies
IEC 61000-3-2 buildings used for domestic purposes.
Electrical fast ±2 kV for power supply ±1 kV for power supply Mains power quality should be that of
transient/burst lines lines a typical commercial or hospital
IEC 61000-4-4 environment. If interference occurs, the
using of filter on power lines or enough
separation between signal lines and
power line are recommended.
Surge ±1 kV differential mode ±1 kV differential mode Mains power quality should be that of
a typical commercial or hospital
IEC 61000-4-5 ±2 kV common mode ±2 kV common mode
environment.
Table 8-2 Electromagnetic Immunity for all Equipment and Systems (Continued)
70 % UT 70 % UT
(30 % dip in UT)
(30 % dip in UT)
for 25 cycles for 25 cycles
<5 % UT <5 % UT
(>95 % dip in UT)
(>95 % dip in UT)
for 5 sec
for 5 sec
1. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile
radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be con-
sidered. If the measured field strength in the location in which the G60/G70/G80 Modular Patient Monitor is used ex-
ceeds the applicable RF compliance level above, the G60/G70/G80 Modular Patient Monitor should be observed to
verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reori-
enting or relocating the G60/G70/G80 Modular Patient Monitor.
2. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Table 8-4 Recommended Separation Distances Between Portable and Mobile RF Communications Equipment
and G60/G70/G80 Modular Patient Monitor
3.5 3.5 7
d = ------- P d = ------- P d = ----- P
V1 E1 E1
For transmitters rated at a maximum output power not listed above, the recommended separation dis-
tance (d) in metres (m) can be estimated using the equation applicable to the frequency of the transmit-
ter, where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer.
Note — At 80MHz and 800MHz, the higher frequency range applies. These guidelines may not apply in all
situations. Electromagnetic propagation is affected by absorption and reflection from structures, objects and
people.
453564282181 453564282181
2.0 2.0