Equotip 540 - Operating - Instruction PDF
Equotip 540 - Operating - Instruction PDF
Equotip 540 - Operating - Instruction PDF
60 Year s of IIn
nnovation Swiss Precision since 1954
6.2 Upload PDF-Files from an USB-stick ........................................ 35
Table of Contents
7. System ..................................................................... 36
1. Safety and Liability .................................................. 4 7.1 Features ............................................................................................ 36
1.1 General Information........................................................................... 4
7.2 Probes ............................................................................................... 37
1.2 Liability .............................................................................................. 4
7.3 Hardware .......................................................................................... 37
1.3 Safety Instructions ............................................................................ 4
7.4 Date & Time .................................................................................... 37
1.4 Correct Usage.................................................................................... 4
7.5 Language.......................................................................................... 37
1.5 Optimizing Performance of the Battery System .......................... 4
7.6 Device information ........................................................................ 37
2. Getting Started ......................................................... 5 8. Maintenance and Support ...................................... 38
2.1 Installation ......................................................................................... 5
8.1 Maintenance..................................................................................... 38
2.2 Main Menu ........................................................................................ 6
8.2 Support Concept............................................................................. 39
3. Measurement ............................................................ 7 8.3 Standard Warranty and Extended Warranty ............................... 39
3.1 Performing Measurements .............................................................. 7 8.4 Disposal............................................................................................ 39
3.2 Measurements Screen ...................................................................... 10
3.3 Measuring Methods .......................................................................12 9. Troubleshooting ........................................................ 40
3.4 Instrument Verification / Daily Performance Check .................24 9.1 Incorrect Measurements / Failed Performance Check ..................40
9.2 No Reading Displayed ......................................................................41
4. Settings .................................................................... 25
4.1 Measurements................................................................................... 25 9.3 Battery .................................................................................................42
4.2 Verification (Performance & Uncertainty Check) .....................27 9.4 Touchscreen Calibration....................................................................42
4.3 Conversions (Hardness Conversions)............................................. 28 10. Equotip Link Software .............................................. 42
4.4 Reporting .........................................................................................28 10.1 Starting Equotip Link ........................................................................42
5. Data (Explorer) ...................................................... 29 10.2 Application Settings ..........................................................................42
5.1 Measurements................................................................................... 29 10.3 Connecting to an Equotip 550 Touchscreen Unit ..........................43
5.2 Verifications ....................................................................................31 10.4 Connection to Portable Rockwell Probe .........................................43
10.5 Adjusting the Settings .......................................................................43
6. Information ............................................................ 35 10.6 Exporting Data ...................................................................................44
6.1 Documents........................................................................................ 35
© 2016 Proceq SA 2
10.7 Exporting and Importing of Setting Profiles.................................. 45
10.8 Exporting and Importing of Conversion Curves........................... 45
© 2016 Proceq SA 3
Table of Contents
4 © 2016 Proceq SA
Table of Contents
Connections
12 USB Host
Figure 1: Insert Battery Snap-in connectors USB Device
Ethernet
There are three status LEDs on the right side of the display. The middle light is Power Supply
the power indicator which is red when charging and turns to green when battery Figure 2: Connections
is fully charged. The lower LED is used for application spe- cific notification.
For Leeb Impact Devices USB Host:
NOTE! Only use the battery charger provided for charging.
use the Snap-in connector 1. Additionally connect a mouse, keyboard
For UCI Probe or USB stick.
USB Device:
use the Snap-in connectors 1 or 2. Connect to PC.
• A complete charge requires < 9 h (Instrument not operating)
For the Portable Rockwell Probe Ethernet:
• Charging time is much longer if the instrument is in use. use the USB Host connector. Connection to network.
• An optional Quick Charger (Part No. 327 01 053) can be used to charge a Power Supply:
spare battery or to charge the battery outside of the instru- ment. In this Connect the power supply
case it takes < 5.5 h for a complete charge. through this connection.
© 2016 Proceq SA 5
Table of Contents
6 © 2016 Proceq SA
Table of Contents
3. Measurement After the last of the impacts is performed, the hardness average and fur- ther
statistics of the measurement series are displayed.
3.1 Performing Measurements NOTE! Make sure the loading tube is allowed to slowly return
back to the starting position. Do take care so the loading tube does
3.1.1 Leeb Testing Procedure (except Leeb U) not spring back uncontrolled, which may result in
permanent device damage.
Select automatic compensation for impact direction “Automatic”, see
chapter “3.2.1 Controls”. If “Automatic” is not allowed, set the impact
Equotip Leeb Impact Devices DL doesn’t sup-port NOTE! If possible, follow the standard practice of Leeb re- bound
hardness testing as described in the standards DIN 50156-1
automatic mode. The impact direction must be selected manually. Select the
(metallic materials), ASTM A956 (steel, cast steel and cast iron
appropriate material group, hardness scales and number of
only), or other applicable standards. If these are not available, the
impacts per measurement series. For more information see chapter “4.
user is recommended to average a minimum of n = 3 impacts at an
Settings”. Conduct impacts by cycling through “load, position and trig- ger”
indentation distance of 3 to 5 mm (0.12 to 0.20”) for each location
mechanism:
of the sample that shall be tested.
1. Load the impact device – while not in
contact with the test piece – by hold-
ing it firmly with one hand and sliding NOTE! Do not carry out an impact in an area that has already been
the loading tube with the other hand until deformed by another impact. Also, do not load the device when it is
contact is grabbed by the clutch. already positioned in the new test location, since the material under the
device may be affected through prior stress, and the catch chuck of the
2. Position the support ring on the test device may get damaged.
piece. Take particular care to fully
position the support ring on the test piece,
but not coinciding with a previ- ous test
indentation.
3. To trigger an impact, press the trigger
button to release the impact body. To
perform another impact, repeat this
cycle.
Figure 4: Leeb Testing Procedure
© 2016 Proceq SA 7
Table of Contents
8 © 2016 Proceq SA
Table of Contents
© 2016 Proceq SA 9
Table of Contents
Impact Direction: To
manually set the impact
direction if required (Leeb
only, by default this is
automatic). Settings: direct shortcut to settings
menu.
(only applicable to the actual
measurement series)
10 © 2016 Proceq SA
Table of Contents
3.2.2 Measurement Views Equotip 550 is fully customizable as a device and can display three dif- ferent
measurement views simultaneously. Each view can be switched to meet the
user’s requirements by simply clicking on the icon related to the particular
display at the top right corner of each screen.
Signal View: Display the probe signal from the last active
measurement. This may be useful for advanced evaluations.
Statistic View: View statistics for the active measurement se- ries.
Number of impacts (n), Average (x ), Standard deviation (σ),
theprimary
scale.
Table View: Display the measurements for the active series in a table
format.
Info: Display the measurement settings e.g. series length, probe type,
material group etc.
User’s View: The user can choose between probe angle, mini- mum,
maximum, range and probe type for the field contents. To change, tap
on each individual box.
HL Single Record View: Display the last or selected measurement result
in both the primary and secondary hardness scales.
Sample ID’s: Defines the custom field.
ID +
NOTE! Screen views cannot be duplicated.
© 2016 Proceq SA 11
Table of Contents
3.3 Measuring Methods During measurement with Equotip 550 impact devices (D, DL, DC, C, G, S,
and E); an impact body with a ball indenter is launched by spring energy against
Equotip 550 family of instruments is capable of accepting three diffrent test
the sample to be measured, and then rebounds. Before and after the impact, a
methods using a single indicating unit.
permanent magnet inside the impact body passes through a coil in which a
voltage signal is induced by the forwards and backwards movement. This
3.3.1 Equotip Leeb
induction signal behaves proportionally to the velocities. The ratio of the
3.3.1.1 Test Principle rebound velocity vr to the impact velocity vi multiplied by 1000 yields the
hardness value HL (Leeb hardness). HL is a direct measure of the hardness. The
Release button third resp. fourth letter of the HL
impact device.
v
HL = r ·1000
vi
Loading tube Equotip Leeb U
Although the Equotip Leeb Impact Device U is constructed differently to simplify
the measurement process, the underlying principle is the same.
Impact Spring
Catch chuck
Guide tube
Connection cable – 4 pole
Figure 11: Equotip Leeb U Impact Device
Existing Parotester impact devices type U are fully supported by the Equotip
Sensor of impact device with ID
ROM Impact Body
550. Typ P and PG impact devices can be still used, but the unit is shown as
HLU although it would be actually LP resp. LPG.
Support ring
NOTE! HLU values can be directly compared to LU on exist-
Figure 10: Schematic View of a Leeb Impact Device ing Parotester instruments.
12 © 2016 Proceq SA
Table of Contents
© 2016 Proceq SA 13
Table of Contents
14 © 2016 Proceq SA
Table of Contents
Type G: Increased impact energy. For solid and inhomogenous components, e.g.
heavy castings and forgings.
Impact energy: 90 Nmm
Type DC: Short device. For use in very
confined spaces, e.g. in holes, cylinders or for
internal measurements on assembled machines. Type D: Universal unit. For the majority of your
Impact energy: 11 Nmm hardness testing requirements.
Impact energy: 11 Nmm
Type S: Si3N4 ball indenter. For testing Type C: Reduced impact energy. Surface
especially in the very high hardness range (in hardened components, coatings, thin walled or
excess of 50 HRC / 650 HV): Tool steels with impact sensitive components (small measuring
high carbide content inclusions. indentation).
Impact energy: 11 Nmm Impact energy: 3 Nmm
© 2016 Proceq SA 15
Table of Contents
NOTE! Clamping may strain the sample, which can affect the
hardness readings.
16 © 2016 Proceq SA
Table of Contents
© 2016 Proceq SA 17
Table of Contents
D/DC DL S E G C
Steel and cast steel Vickers HV 81-955 80-950 101-964 84-1211 81-1012
Brinell HB 81-654 81-646 101-640 83-686 90-646 81-694
Rockwell HRB 38-100 37-100 48-100
HRC 20-68 21-68 22-70 20-72 20-70
HRA 61-88 61-88
Shore HS 30-99 31-97 28-104 29-103 30-102
Rm N/mm² σ1 275-2194 275-2297 340-2194 283-2195 305-2194 275-2194
σ2 616-1480 614-1485 615-1480 616-1479 618-1478 615-1479
σ3 449-847 449-849 450-846 448-849 450-847 450-846
Cold work tool steel Vickers HV 80-900 80-905 104-924 82-1009 * 98-942
Rockwell HRC 21-67 21-67 22-68 23-70 20-67
Stainless steel Vickers HV 85-802 * 119-934 88-668 * *
Brinell HB 85-655 105-656 87-661
Rockwell HRB 46-102 70-104 49-102
HRC 20-62 21-64 20-64
Cast iron lamellar graphite GG Brinell HB 90-664 * * * 92-326 *
Vickers HV 90-698
Rockwell HRC 21-59
Cast iron, nodular graphite GGG Brinell HB 95-686 * * * 127-364 *
Vickers HV 96-724
Rockwell HRC 21-60 19-37
Cast aluminium alloys Brinell HB 19-164 20-187 20-184 23-176 19-168 21-167
Vickers HV 22-193 21-191 22-196 22-198
Rockwell HRB 24-85 24-86 23-85
Copper/zinc alloys (brass) Brinell HB 40-173 * * * * *
Rockwell HRB 14-95
CuAI/CuSn-alloys (bronze) Brinell HB 60-290 * * * * *
Wrought copper alloys, low alloyed Brinell HB 45-315 * * * * *
*Custom conversion curve / correlation
Table 2: Overview of Available Conversions
18 © 2016 Proceq SA
Table of Contents
3.3.2 Equotip Portable Rockwell NOTE! By calculating the penetration depth between the
preload and total load, surface roughness discrepancies are
3.3.2.1 Test Principle significantly disregarded.
During measurement with Equotip 550 Portable Rockwell probe, a diamond indenter
is forced into the test piece, and then released back out of the ma- terial. The NOTE! The hardness testing principle in Portable Rockwell
indentation depth is measured continuously during this process. Indentation depth is follows the Rockwell stationary test. As for the Rockwell
calculated after decreasing the total load to preload. test, no adjustment for the test direction is required. How-
10N 50N 10N
ever, there are three main differences to traditional stationary
Rockwell tests:
• The test loads are lower.
• The Portable Rockwell indenter is sharper.
• The dwell times during the test are shorter.
© 2016 Proceq SA 19
Table of Contents
Maximum test 70 HRC • Use the 3 mm Allen key setup tool to release the cantilever. Turn it by 90°.
piece hardness
• Take the probe and remove the foot. The diamond indenter remains mounted.
Minimum spacing three times the diameter of a test indentation
• Screw the probe into the probe holder of the clamp clockwise (hand-tight).
Table 3: Portable Rockwell Test Piece Requirements • Turn the cantilever so its tip is centred over the probe; tighten the can- tilever
screw securely using the 3 mm Allen key setup tool.
• The recommended clearance between the bottom of the probe holder and the
sample surface should be between 2 and 5 mm. Adjust the height with the
two knurled screws.
20 © 2016 Proceq SA
Table of Contents
NOTE! In case the probe connector is in an inconvenient 3.3.2.7 Installation of Standard Foot or Tripod
position, release the set screw. Ensure that the springs in The round standard foot permits meas-
the mechanism do not get lost. Turn the mechanism into a urements on test objects that are only
convenient position, aligning the set screw with the guide accessible from one side, such as large metal
channel. Lock the set screw so that the probe holder will still sheets. The tripod is used when the flat foot
slide up and down without rubbing on the set screw. cannot be placed on the test piece without
wiggling.
3.3.2.6 Considerations 1. The diamond indenter remains
• When measuring cylindrical samples with adapters Z4 or Z4+28, make sure, mounted. Figure 17: Portable Rockwell
the sample is not twisted on the clamp support. This is best ensured when the with Tripod
2. Install the foot on the probe.
back part of the clamp rests on a table and only the sample support of the
clamp sticks out over the table’s edge. 3.3.2.8 Installing the Special Foot
• When applying the load, slowly squeeze the leavers and allow the sample to Two special feet extend the Portable Rockwell application range to cy- lindrical
adjust to the support. During the measurement, do not touch the sample, if test pieces.
possible. When releasing, grab the sample again.
1. The diamond indenter remains
• Whenever the sample geometry (i.e. the wall thickness) allows it, free- hand mounted.
measurements usually offer better measuring performance. This applies
particularly to measurements on cylinders. 2. Install the foot on the probe.
• For small diameter rods (or stiff enough pipes), the V-notch clamp adapter 3. Place the foot on the test piece and
Z2 has been designed. When installing the Z2 support ensure that the centre release the set screw on the foot. Then
of the V-notch is centred underneath the probe holder. press down the probe onto the test piece
and lock the set screw.
Figure 18: Portable Rockwell
Special Feet
© 2016 Proceq SA 21
Table of Contents
22 © 2016 Proceq SA
Table of Contents
© 2016 Proceq SA 23
Table of Contents
3.3.3.5 Installation of the Special Foot 3.4 Instrument Verification / Daily Performance Check
5º
5º The default foot allows the measurement to be performed on every See chapter “6.2 Device Verification” and follow the on-screen proce- dure.
surface. The probe must be perpendicular to the sur- face (± 5°). The After the verification process your instrument is fully operational and you can
special foot can be used to increase the re- peatability and avoid the now continue with your measurements.
distortion of the results, see chapter “14. Ordering Information”.
NOTE! The performance check should be done regularly be- fore
1. Unscrew the standard foot and remove it using the instrument to verify the mechanical and elec- tronic
2. Screw the special foot to the probe tightly functions of the probe and the indicating device. This requirement
is also included in the relevant hardness stand- ards, see chapter
NOTE! To measure in places with limited accessibility, the “13. Standards & Guidelines”.
probe can be used without any foot. If doing so, the side of
the rod of the probe must not touch any surface or be han-
dled, as this leads to biased readings.
24 © 2016 Proceq SA
Table of Contents
© 2016 Proceq SA 25
Table of Contents
Edit Entries NOTE! Storing signal data will cause measurement files to
The entries of the different sample ID fields can be deleted or edited here. For take up more memory.
easy increasing or decreasing, use the up- and down- arrows. For add- ing or
removing entry fields, please see chapter “8.1.2 General Features”. Enable Warnings
Choose to enable warning display signals and sounds to indicate false
4.1.5 Workflow measurements.
Activate User Guidance
Select to display on screen instructions and messages when taking a
measurement.
26 © 2016 Proceq SA
Table of Contents
© 2016 Proceq SA 27
Table of Contents
DIN 50156 Leeb hardness testing of metallic materials. Will be re- placed by 4.3.1 Standard Conversions
ISO 16859.
Proceq has developed correlations to convert the Leeb hardness meas- urements
DIN 50157 Hardness testing of metallic materials with portable measuring to other commonly used hardness scales based on groups of alloys that have a
instruments operating with mechanical pen- etration depth. close relationship. The conversions for HLD and Material Group 1 (Carbon
DIN 50159 Hardness testing with the UCI method. Steels) are standardized according to ASTM E140-12b.
ASTM A956 Standard test method for Leeb hardness testing of steel products.
ASTM A1038 Standard test method for portable hardness testing by the 4.3.2 Custom Conversion Curves
Ultrasonic Contact Impedance method
See chapter “6.4 Conversion Curve Creation”.
ISO 16859 Hardness testing of metalic materials by Leeb.
4.3.2.1 Custom Compensation
Extended Measurement uncertainty analysis is applied to under- stand
In some cases a user must measure hardness on many samples with identi- cal size
Uncertainty the differences in test results and to determine sources of
and shape that is below the ideal limits for accuracy. Studies have been published by
(Combined error. The uncertainty of an Equotip Leeb, Equotip Portable
ASME and Nordtest that have identified and confirmed the validity of the strategy to
Uncertainty) Rockwell or Equotip UCI hardness testing system consists of
apply a compensation factor to correct for the inaccuracies induced by the non-ideal
a statistical component, a component inherent to the
geometry. The methods outlined in chapters “6.4 Conversion Curve Creation” can be
measurement device and a component arising from the
applied to create this compensation factor to be applied automatically to the Equotip
metrological chain between national standard and the user
test result.
device (traceability).
Although uncertainty could be a complicated topic, Equotip
550 automatically calculates the combined un- certainty of the
4.4 Reporting
system. All the required information is al- ready available in The content of the measurement reports can be adjusted here.
the calibration certificates provided by Proceq. Therefore the
device only requires adding these values in the specified fields 4.4.1 Images Explorer
and following the simple steps on the display in order to Images, i.e. company logos can be loaded from an USB stick on the de- vice, for
complete the process. use in reports. Pictures must be in the *.png or *.jpg format, ideally come with
72dpi and a maximum resolution of 496x652 pixel.
4.3 Conversions (Hardness Conversions)
There is no direct correlation between any two hardness scales. There- fore Upload Images from an USB-stick
conversions must be determined by comparison testing for any given alloy. To do so follow the steps below.
• Create the folder “PQ-Import” in the main directory of the USB-stick (not
as a subfolder in another folder) and fill it with all the picture-files to be
uploaded to the Equotip Touchscreen
28 © 2016 Proceq SA
Table of Contents
• Connect the USB-stick to the USB Device plug on the left side of the 5. Data (Explorer)
Equotip Touchscreen
• Click on and confirm with click on
• The uploaded images appear in the Images Explorer
© 2016 Proceq SA 29
Table of Contents
Tap on a saved file to open it and return to the Data Explorer list by press- ing the
NOTE! If the file name already exists, the name will be ex-
back button.
tended by a number and increased with every additional file.
managesaved
measurements data. 5.1.3 Review of Data
Each folder and measurement series is shown as one line in the explorer view. In the detailed view of a measurement series, all information can be seen and the
settings are editable.
For each series the probe used, mean value of the series, series name, date &
time of the measurement can be seen. All the different views can be switched according to the users needs.
The list can be sorted by tapping on the corresponding header. The small arrow For more details about the different views, please refer to chapter “3.2.2
indicates which list is sorted. Measurement Views”.
30 © 2016 Proceq SA
Table of Contents
5.2 Verifications
From the main menu select “Data”, and then “Verifications” to review and manage saved verification data “6.2 Device Verification”.
The verification data is stored and managed in the same manner as the measurement data. Except no deletion is allowed.
Each folder and measurement series is shown as one line in the explorer view.
Additionally the result for each verification data series, either a “passed” or a “failed” is displayed.
© 2016 Proceq SA 35
Table of Contents
© 2016 Proceq SA 35
Table of Contents
Folder Manager
Here the desired path can be edited. A maximum of four subfolders can be
created with selectable information. As soon as one of this information changes,
a new folder will be created automatically.
Figure 26: System Menu
Use File Manager
7.1 Features Activate this option to use automatic file naming as configured in the File
Manager.
7.1.1 Device Lock Settings
File Manager
Lock/Unlock: Select this to lock the instrument and protect it from un-
An automatic name can be configured here consisting of four different
intentional changes.
information fields.
Password: A password can be set for the lock/unlock function. If this field is
left empty, no password is required to unlock the user settings. Long Filename Viewing
Choose here between a full view of the filename, or if only a selected range
7.1.2 General Features should be displayed in the measurement screen. This setting influ- ences only
Measurement Wizard: There are three options available on how the the name on the measurement screen, but not in the explorer or reports.
measurements wizards are enforced.
36 © 2016 Proceq SA
Table of Contents
8.3 Hardware
General settings about the user interface and power options can be ed- ited here.
Sound: The volume of the audio notifications of the instrument can be adjusted
respectively switched off.
Display: The user can adjust the brightness of the display backlight.
Power: The time after when the instrument dims the display, or shuts down can
be adjusted, for both battery and AC powered operation.
© 2016 Proceq SA 37
Table of Contents
8. Maintenance and Support NOTE! Never immerse the device in water. Do not use com-
pressed air, abrasives, solvents or lubricants to clean the device.
8.1 Maintenance
The instrument should be calibrated annually to ensure consistent, reli-
able and accurate measurements. However, the service interval may be based on 9.1.3 Storage
actual experience and usage. Consult the applicable standards for more
Only store the Equotip 550 in the original packaging and in a dry room free of
guidance.
dust.
8.1.1 Regular Device Check
Performance checks, see chapter “4.2 Verification (Performance & Uncer- tainty
9.1.4 Re-calibration for Impact Direction (Leeb only)
Check)” of the instrument should be carried out at least once a day or at the latest For Leeb impact devices the compensation function relies on parameters
after 1000 impacts. In the case of infrequent use, carry out the check before the specific to each impact device, which are stored in the device. Validity of the
beginning and at the end of a test series. In addition, have the device calibrated by Angle
an authorized Proceq Service Center once a year. Calibration and then pressing the “Test” button. For each impact
direction, the deviation from the curve shall be less than ±0.2 Leeb (HL).
NOTE! The unit is working properly when the average is within the The parameters may change with time or due to external influences. Re-
target range. Otherwise, please see chapter “10. Troubleshooting”. calibration of the automatic compensation function in Equotip Leeb im- pact
devices (except type DL) is recommended particularly when:
9.1.2 Cleaning
• Impact device has been cleaned, or
Leeb Impact device: Unscrew the support ring. Remove the impact body from
• Impact device has not been used for a long time, or
the guide tube. Clean the guide tube with the cleaning brush.
Reassemble. • Impact body has been replaced.
A re-calibration is done by consecutive selection of “0° (vertical down)”, “90°
Leeb Indenters: Clean the ball of the Leeb impact bodies and Portable
(horizontal)” and “180° (vertical up)”.
Rockwell diamond indenter with acetone or similar solvent. (Do not use water
or water based detergents!)
9.1.5 Updating the Equotip 550 Operating System and
Portable Rockwell and UCI probes: Clean the probes and the dia- mond Application
indenters with a clean, dry cloth.
Connect the device to the computer. Updates can be done using Equotip Link as
Housing: Clean the display and housing with a clean, dry cloth after use. Also follows:
clean the connector sockets with a clean, dry brush.
• Select update symbol in Equotip Link
• Select “Express” and confirm with “Next”.
38 © 2016 Proceq SA
Table of Contents
• Select the device type and confirm with “Next”. 9.4 Disposal
• In the “Choose Communication Type” dialog box, select the type of Disposal of electric appliances together with household waste is not
communication used between the Equotip and PC, and then click “Next”. permissible. In observance of European Directives 2002/96/ EC,
• In the “Device search result and selection” dialog box, make sure the serial 2006/66/EC and 2012/19/EC on waste, electrical and elec-
number of the device in the drop-down field is the device to be updated, tronic equipment and its implementation, and in accordance with nation- al and
and then click “Next”. local law, electric tools and batteries that have reached the end of their life must
be collected separately and returned to an environmentally compatible recycling
• PqUpgrade will now search the Proceq servers for any available up- dates. To
facility.
do so, a working internet connection is required.
• Follow the on-screen instructions to finish the update.
© 2016 Proceq SA 39
Table of Contents
40 © 2016 Proceq SA
Table of Contents
© 2016 Proceq SA 41
Table of Contents
Double click on the Equotip Link Icon on your desktop or start the program via
the start menu.
42 © 2016 Proceq SA
Table of Contents
10.3 Connecting to an Equotip 550 Touchscreen Unit 10.4 Connection to Portable Rockwell Probe
Connect the Equotip 550 Touchscreen Unit to an USB port or connect • Connect the Portable Rockwell probe to a PC using the provided probe cable.
it to the Ethernet (DHCP server requried)., then select to download data • Start the Equotip Link software, and click on the Portable Rockwell icon
from the Equotip 550 Touchscreen Unit. to detect the Portable Rockwell probe. Click the “New” button at the bottom
of the screen.
The following window will be displayed: Select the appropriate communi- • Select the hardness scale to be displayed (Hardness scales).
cation type. If connecting through Ethernet, please enter the instruments IP • Select the number of readings “n” per measurement series.
address in the appropriate field. Click on “Next >”.
10.4.1 Viewing Data
The transferred measurements from your Equotip 550 will be displayed on the
screen:
When an Equotip 550 has been found its details will be displayed on screen.
Click on the “Finish” button to establish the connection
Click on the measurement file to be transferred. If Click on the double arrow icon in the first column to see more details.
multiple files have to be transferred, select them while
holding the “shift”or “ctrl” button, or click on “Select
10.5 Adjusting the Settings
All”. “All settings, such as material group, scale, impact direction and both limits can
be changed afterwards in Equotip Link.
If the settings of multiple measurement series have to changed, select each of
them while holding the “shift” or “ctrl” button.
© 2016 Proceq SA 43
Table of Contents
10.5.1 Adjusting Date and Time Click on the “Export as CSV file(s)” icon. The data is then
exported as a Microsoft Office Excel comma separated file. The
Right click in the “Date & Time” column.
export options may be chosen in the following window:
44 © 2016 Proceq SA
Table of Contents
© 2016 Proceq SA 45
Table of Contents
Environment Indoor use only Impact energy • 11.5 Nmm for D, DC, E, S probes
• 11.1 Nmm for DL probe
Pollution degree 2 • 3.0 Nmm for C probe
Installation category 2 • 90.0 Nmm for G probe
• 200.0 Nmm for U probe
Mass of impact body • 5.45 g (0.2 ounces) for D, DC, E, S probes
• 7.25 g (0.26 ounces) for DL probe
• 3.10 g (0.11 ounces) for C probe
• 20.0 g (0.71 ounces) for G probe
• 26.0 g (0.92 ounces) for U probe
Ball indenter • Tungsten carbide, 3.0 mm (0.12“) diameter
for C, D, DC probes
• Tungsten carbide, 2.78 mm (0.11“) diameter for
DL probe
• Tungsten carbide, 5.0 mm (0.2“) diameter for
G probe
• Ceramics, 3.0 mm (0.12”) diameter for S
probe
• Polycrystalline diamond, 3.0 mm (0.12”)
diameter for E probe
• Hardened Steel, 50.0 mm (1.97”) diameter for U
probe
Operating temperature -10 to 50˚C (14 to 122°F)
46 © 2016 Proceq SA
Table of Contents
© 2016 Proceq SA 47
Table of Contents
13. Ordering Information 356 10 005 Equotip 550 UCI consisting of Equotip Touchscreen
incl. Battery, Equotip UCI Probe HV1-HV5, UCI Probe
13.1 Units Cable, UCI Test Block ~850 HV, Power Supply, USB-
Cable, Surface Roughness Comparator Plate, DVD with
Part No. Description
Software, Documentation, Carrying Strap and Carrying
356 10 001 Equotip 550 consisting of Equotip Touchscreen incl. Case
Battery, Power Supply, USB-Cable, Surface Roughness
356 10 006 Equotip 550 Leeb U (for paper, film and foils) consisting
Comparator Plate, DVD with Software, Documentation,
of Equotip Touchscreen incl. Battery, Equotip Leeb
Carrying Strap and Carrying Case
Impact Device U, Cleaning Brush, Probe Cable,
356 10 002 Equotip 550 Leeb D consisting of Equotip Touchscreen Power Supply, USB-Cable, DVD with Software,
incl. Battery, Equotip Basic Leeb Impact Device D, Im- Documentation, Carrying Strap and Carrying Case
pact Body D, Support Rings (D6,D6a), Cleaning Brush,
356 10 020 Equotip 550 Portable Rockwell & UCI Kit consisting
Impact Device Cable, Test Block ~775 HLD / ~56 HRC,
of Equotip 550 UCI (356 10 005) and Equotip Portable
Coupling Paste, Power Supply, USB-Cable, Surface
Rockwell Probe 50 N (356 00 600)
Roughness Comparator Plate, DVD with Software,
Documentation, Carrying Strap and Carrying Case 356 10 021 Equotip 550 Portable Rockwell & Leeb D Kit consist-
ing of Equotip 550 Leeb D (356 10 002) and Equotip
356 10 003 Equotip 550 Leeb G consisting of Equotip Touchscreen
Portable Rockwell Probe 50 N (356 00 600)
incl. Battery, Equotip Basic Leeb Impact Device G, Im-
pact Body G, Support Rings (G6,G6a), Cleaning Brush, 356 10 022 Equotip 550 Leeb D & UCI Kit consisting of Equotip
Impact Device Cable, Test Block ~570 HLG /~340 HB, 550 Leeb D (356 10 002), Equotip UCI Probe HV1-HV5
Coupling Paste, Power Supply, USB-Cable, Surface (356 00 700) and Equotip UCI Test Block ~850 HV, ISO
Roughness Comparator Plate, DVD with Software, 6507-3 HV5 Calibration (357 54 100)
Documentation, Carrying Strap and Carrying Case
356 10 004 Equotip 550 Portable Rockwell consisting of Equotip 13.2 Impact Devices & Probes
Touchscreen incl. Battery, Equotip Portable Rockwell
Probe 50 N, Protective Rubber Sleeve, Probe Cable, Description
Part No.
Test Block ~62HRC, Power Supply, USB-Cable, Sur- Impact Device incl. Support Ring, Impact Body, Cable
face Roughness Comparator Plate, DVD with Software,
Documentation, Carrying Strap and Carrying Case 356 00 500 Equotip Leeb Impact Device C
356 00 100 Equotip Leeb Impact Device D
356 00 110 Equotip Leeb Impact Device DC
48 © 2016 Proceq SA
Table of Contents
356 00 120 Equotip Leeb Impact Device DL 351 90 018 USB-Cable 1.8 m (6 ft)
356 00 400 Equotip Leeb Impact Device E 327 01 061 Power Supply
356 00 300 Equotip Leeb Impact Device G 711 10 013 Power Supply Cable USA 0.5 m (1.7 ft)
356 00 200 Equotip Leeb Impact Device S 711 10 014 Power Supply Cable UK 0.5 m (1.7 ft)
360 04 600 Equotip Leeb Impact Device U 711 10 015 Power Supply Cable EU 0.5 m (1.7 ft)
327 01 053 Quick Charger
Impact Device only
356 00 081 Equotip Surface Roughness Plate
353 00 501 Equotip Basic Leeb Impact Device C
350 01 015 Equotip Coupling Paste
353 00 101 Equotip Basic Leeb Impact Device D
356 00 082 Display Antiglare Protection Film for Touchscreen Unit
353 00 111 Equotip Basic Leeb Impact Device DC
353 00 121 Equotip Basic Leeb Impact Device DL 356 00 080 Equotip Leeb Impact Device Cable 1.5 m (5 ft)
353 00 401 Equotip Basic Leeb Impact Device E 353 00 086 Equotip Leeb Impact Device Extension Cable 5 m (15 ft)
353 00 301 Equotip Basic Leeb Impact Device G 356 00 083 Equotip Leeb Impact Device U Cable 1.5 m (5 ft)
353 00 201 Equotip Basic Leeb Impact Device S 350 01 004 Equotip Impact Body D/DC
360 04 032 Equotip Basic Leeb Impact Device U 350 71 311 Equotip Impact Body DL
350 71 413 Equotip Impact Body S
356 00 600 Equotip Portable Rockwell Probe 50N (for 350 08 002 Equotip Impact Body G
Equotip 550 or PC)
350 07 002 Equotip Impact Body E
356 00 700 Equotip UCI Probe HV1-HV5
350 05 003 Equotip Impact Body C
13.3 Parts and Accessories 360 04 504 Equotip Impact Body U
327 01 043 Carrying Strap complete 350 01 010 Equotip Support Ring D6a
© 2016 Proceq SA 49
Table of Contents
350 08 005 Equotip Support Ring G6a Equotip Portable Rockwell support Z4 for measuring clamp
354 01 228
(for tubes and pipes up to Ø 28 mm)
350 71 314 Equotip Support Ring DL
356 00 720 Equotip UCI Special Foot
360 04 531 Equotip Support Ring U
Set of Support Rings (12pcs) for Equotip Leeb Impact 13.4 Test Blocks
353 03 000
Device D/DC/C/E/S
Part No. Description
Equotip Leeb Impact Device Cleaning Brush
350 01 008
D/DC/C/E/S 357 11 500 Equotip Test Block C, ~565 HLC / <220 HB,
350 08 006 Equotip Leeb Impact Device Cleaning Brush G Proceq Factory Calibration
360 04 502 Equotip Leeb Impact Device Cleaning Brush U 357 12 500 Equotip Test Block C, ~665 HLC / ~325 HB,
Proceq Factory Calibration
350 01 007 Equotip Leeb Impact Device DC Loading Stick
357 13 500 Equotip Test Block C, ~835 HLC / ~56 HRC,
350 71 316 Equotip Leeb Impact Device DL Plexiglas Sleeve Proceq Factory Calibration
360 04 530 Assembly Gauge for Equotip Leeb Impact Device U 357 11 100 Equotip Test Block D/DC, <500 HLD / <220 HB,
Proceq Factory Calibration
357 12 100 Equotip Test Block D/DC, ~600 HLD / ~325 HB,
354 01 139 Equotip Portable Rockwell Probe Cable 2 m (6 ft)
Proceq Factory Calibration
354 01 200 Equotip Portable Rockwell Measuring Clamp
357 13 100 Equotip Test Block D/DC, ~775 HLD / ~56 HRC,
354 01 130 Equotip Portable Rockwell Tripod Proceq Factory Calibration
354 01 250 Equotip Portable Rockwell Special Foot RZ 18 - 70 357 13 105 Equotip Test Block D/DC, ~775 HLD, one side,
354 01 253 Equotip Portable Rockwell Special Foot RZ 70 - Proceq Factory Calibration
354 01 137 Equotip Portable Rockwell Protective Rubber Sleeve 357 11 120 Equotip Test Block DL, <710 HLDL / <220 HB,
Proceq Factory Calibration
Equotip Portable Rockwell support Z2 for measuring
354 01 243 357 12 120 Equotip Test Block DL, ~780 HLDL /~325 HB,
clamp
Proceq Factory Calibration
Equotip Portable Rockwell Support Z4+28 for
354 01 229 357 13 120 Equotip Test Block DL, ~890 HLDL / ~56 HRC,
measuring clamp (for tubes and pipes over Ø 28 mm)
Proceq Factory Calibration
50 © 2016 Proceq SA
Table of Contents
357 13 400 Equotip Test Block E, ~740 HLE / ~56 HRC, Test Blocks Calibrations
Proceq Factory Calibration
Part No. Description
357 14 400 Equotip Test Block E, ~810 HLE / ~63 HRC, Equotip Leeb Test Block Additional Calibration HLD/
Proceq Factory Calibration 357 10 109
HLDC
357 31 300 Equotip Test Block G, <450 HLG / <200 HB, 357 10 129 Equotip Leeb Test Block Additional Calibration HLDL
Proceq Factory Calibration 357 10 209 Equotip Leeb Test Block Additional Calibration HLS
357 32 300 Equotip Test Block G, ~570 HLG / ~340 HB, 357 10 409 Equotip Leeb Test Block Additional Calibration HLE
Proceq Factory Calibration
357 10 509 Equotip Leeb Test Block Additional Calibration HLC
357 13 200 Equotip Test Block S, ~815 HLS / ~56 HRC,
Proceq Factory Calibration 357 30 309 Equotip Leeb Test Block Additional Calibration HLG
357 14 200 Equotip Test Block S, ~875 HLS / ~63 HRC, Equotip Leeb Test Block Additional Calibration HL,
357 90 909
Proceq Factory Calibration DIN 50156-3
Equotip Leeb Test Block Additional Calibration HB,
360 04 503 Equotip Test Block U, ~560 HLU, 357 90 919
ISO 6506-3
Proceq Factory Calibration
Equotip Leeb Test Block Additional Calibration HV,
357 90 929
ISO 6507-3
357 41 100 Equotip Portable Rockwell Test Block ~20 HRC, Equotip Leeb Test Block Additional Calibration HR,
357 90 939
ISO 6508-3 HRC Calibration ISO 6508-3
© 2016 Proceq SA 51
Proceq Europe Proceq Middle East
Ringstrasse 2 P. O. Box 8365, SAIF Zone,
CH-8603 Schwerzenbach Sharjah, United Arab Emirates
Phone +41-43-355 38 00 Phone +971-6-557-8505 Fax
Fax +41-43-355 38 12 +971-6-557-8606
info-europe@proceq.com info-middleeast@proceq.com
Proceq UK Ltd. Proceq SAO Ltd.
Bedford i-lab, Priory Business Park South American Operations
Stannard Way Rua Paes Leme, 136, cj 610
Bedford MK44 3RZ Pinheiros, São Paulo
United Kingdom Brasil Cep. 05424-010
Phone +44-12-3483-4515 Phone +55 11 3083 38 89
info-uk@proceq.com info-southamerica@proceq.com
Proceq USA, Inc. Proceq China
117 Corporation Drive Unit B, 19th Floor
Aliquippa, PA 15001 Five Continent International Mansion, No. 807 Zhao Jia
Phone +1-724-512-0330 Bang Road
Fax +1-724-512-0331 Shanghai 200032
info-usa@proceq.com Phone +86 21-63177479
Fax +86 21 63175015
Proceq Asia Pte Ltd info-china@proceq.com
12 New Industrial Road
#02-02A Morningstar Centre
Singapore 536202
Phone +65-6382-3966
Fax +65-6382-3307
info-asia@proceq.com
Proceq Rus LLC
Ul. Optikov 4
korp. 2, lit. A, Office 410
197374 St. Petersburg
Russia
Phone/Fax + 7 812 448 35 00
info-russia@proceq.com
Subject to change. Copyright © 2017 by Proceq SA, Schwerzenbach. All rights reserved. 82035601E ver 05
2017