PHY370 Chapter 3 PDF
PHY370 Chapter 3 PDF
PHY370 Chapter 3 PDF
CHAPTER 3
Sound & Ultrasound
• Sound is a vibration that is transmitted in a medium, that
can be heard by a human ear.
The human ear cannot hear below 20 Hz and the
frequencies below this are called “infra sounds”.
We cannot also hear frequencies that is above
20,000 Hz and this frequencies is known as
“ultrasound”.
Ultrasound is sound having a frequency greater
than 20,000 cycles per second, that is, sound above
the audible range.
Frequencies in the range of millions of cycles per
second are used for medical diagnostic purposes.
There are at least four advantages of ultrasound energy in
medical diagnostic imaging:
a) ultrasound can be directed in a beam
b) ultrasound obeys the laws of reflection and refraction
c) ultrasound energy is reflected off of small objects
d) ultrasound brings no injuries to health.
There are two principal disadvantages of ultrasound.
a) The energy propagates poorly through a gaseous
medium (ultrasound transducers must have airless
contact with the body during examinations).
b) Ultrasound images are relatively noisy, and have
poorer contrast than X-ray and MR modalities.
• Sound waves are also known as ‘Acoustic’ waves.
• Sound waves are an organized vibrations of the molecules of
a medium that is able to support the propagation of these
waves.
• Usually the vibrations are organized in a sinusoidal fashion
(figure 1)
Figure 1
• Sinusoidal sound wave is actually a series of areas of
compression and rarefactions.
• Sine wave with the peak (or the “hill”) representing the
pressure maximum, and the nadir (or the “valley”)
representing the pressure minimum.
• The combination of one compression and one
rarefaction represents one cycle.
• The distance between the peak of one compression to
the peak of the next is a wavelength.
• The wavelength λ (m) of the sound wave shown in
Figure 1 :
2
k
• The velocity c (m/s), or the speed at which sound
waves travel through a particular medium:
2f
c
k k
1
c
o K
It is convenient way to express relevant tissue
properties by using the characteristic (or acoustic)
impedance Z (kg/m2s), which is defined by
Z oc
Generation of Ultrasound Waves (Pressure
Waves)
• The use of ultrasound as a medical imaging modality
became practical with the development of small
piezoelectric crystal transducers.
• “Piezo” means pressure, so piezoelectric means that
pressure is generated when electrical energy is applied to
the crystal.
Z oc
• Figure above shows what happens when an ultrasound wave (shown as
a red arrow in image above) tries to pass from one substance to
another substance with a different acoustic impedance.
1) Part of the ultrasound waves continues into the second substance, but
becomes slightly bent away from their original direction (pink arrow). The
bending away when ultrasound passes from one substance to another
substance with a different acoustic impedance is called refraction.
2) Part of the wave (shown as blue arrow in image below) is reflected back
to the probe .The amount reflected back depends on the difference of the
acoustic impedance between the two substances, more the difference,
more the reflection. Reflected waves are extremely important, since it is
only these waves that return back to the probe and provide information
for the machine to show an image.
• As the ultrasound wave crosses from one tissue to the next,
each with a different acoustic impedance, some of the wave is
reflected back at each crossing (two blue arrows in image
above).
• Therefore, multiple reflected waves return to the probe and the
machine uses this information to display an image showing the
different tissues.
• Irregular surfaced objects such as nerves scatter the
ultrasound waves in all directions.
• A small portion of the waves are reflected back to the probe
(shown as blue arrow in image above).
• This is called "scattered reflection".
If an object is large and smooth like a nerve blocking needle,
all the ultrasound wave is reflected back.
This is very useful since it helps us to clearly see needles when
performing ultrasound guided nerve blocks.
This mirror like reflection, where the waves are reflected back
mostly in one direction is officially called "specular reflection".
Ultrasound Scanning Modes
Probes are generally described by the size and shape of their face
(“footprint”).
Selecting the right probe for the situation is essential to get good
images, although there may be times where more than one probe
may be appropriate for a given exam.
There are three basic types of probe used in emergency and critical
care point-of-care ultrasound:
a) linear,
b) curvilinear, and
c) phased array.
Linear (also sometimes called vascular) probes are generally high
frequency, better for imaging superficial structures and vessels, and
are also often called a vascular probe.
Probe Selection – Introduction (2)
The curved face creates a wide field of view of almost 180° and its high
frequencies provide superior resolution (figure below).
Phased array probes have crystals that are grouped closely together.
The timing of the electrical pulses that are applied to the crystals
varies and they are fired in an oscillating manner.
The sound waves that are generated originate from a single point and
fan outward, creating a sector-type image (figure below)
Phase Array Probe
This probe has a smaller and flatter footprint than the curvilinear one,
which allows the user to maneuver more easily between the ribs and
small spaces.
These probes have frequencies between 2 and 8 MHz, but they usually
operate at the higher end making them ideal for echocardiography.
The phased array probe can also function at lower frequencies, which
is useful for viewing the abdomen, pelvis, and for procedural guidance.