Proline Promass 80E, 83E
Table of contents
2 Endress+Hauser
Proline Promass 80E, 83E
The amplitude of the Coriolis force depends on the moving mass m, its velocity v in the system, and thus on
the mass flow. Instead of a constant angular velocity , the Promass sensor uses oscillation.
In the sensor, two parallel measuring tubes containing flowing fluid oscillate in antiphase, acting like a tuning
fork. The Coriolis forces produced at the measuring tubes cause a phase shift in the tube oscillations (see
illustration):
• At zero flow, in other words when the fluid is at a standstill, the two tubes oscillate in phase (1).
• Mass flow causes deceleration of the oscillation at the inlet of the tubes (2) and acceleration at the outlet (3).
A A A
B B B
1 2 3
a0003385
The phase difference (A-B) increases with increasing mass flow. Electrodynamic sensors register the tube
oscillations at the inlet and outlet.
System balance is ensured by the antiphase oscillation of the two measuring tubes. The measuring principle
operates independently of temperature, pressure, viscosity, conductivity and flow profile.
Volume measurement
The measuring tubes are continuously excited at their resonance frequency. A change in the mass and thus the
density of the oscillating system (comprising measuring tubes and fluid) results in a corresponding, automatic
adjustment in the oscillation frequency. Resonance frequency is thus a function of fluid density. The density
value obtained in this way can be used in conjunction with the measured mass flow to calculate the volume
flow.
The temperature of the measuring tubes is also determined in order to calculate the compensation factor due
to temperature effects.
Endress+Hauser 3
Proline Promass 80E, 83E
Measuring system The measuring system consists of a transmitter and a sensor. Two versions are available:
• Compact version: transmitter and sensor form a mechanical unit.
• Remote version: transmitter and sensor are mounted physically separate from one another.
Transmitter
Esc
- + E
a0003671
a0003672
Sensor
a0002271
a0003673
a0003675
4 Endress+Hauser
Proline Promass 80E, 83E
H • Single bent tube. Low pressure loss and chemically Documentation No.
resistant material TI074D
• Nominal diameters DN 8 to 50 (3/8" to 2")
• Material: Zirconium 702/R 60702, Tantalum 2.5W
a0003677
a0003678
Endress+Hauser 5
Proline Promass 80E, 83E
Input
Measured variable • Mass flow (proportional to the phase difference between two sensors mounted on the measuring tube to
register a phase shift in the oscillation)
• Fluid density (proportional to resonance frequency of the measuring tube)
• Fluid temperature (measured with temperature sensors)
DN
x
[mm] [inch]
8 3/8" 85
15 ½" 110
25 1" 125
40 1 ½" 125
50 2" 125
80 3" 155
6 Endress+Hauser
Proline Promass 80E, 83E
Operable flow range Greater than 1000: 1. Flow rates above the preset full scale value do not overload the amplifier, i.e. the totalizer
values are registered correctly.
Output
Output signal Promass 80
Current output
Active/passive selectable, galvanically isolated, time constant selectable (0.05 to 100 s), full scale value
selectable, temperature coefficient: typically 0.005% o.f.s./°C, resolution: 0.5 μA
• Active: 0/4 to 20 mA, RL < 700 (for HART: RL 250 )
• Passive: 4 to 20 mA; supply voltage US 18 to 30 V DC; Ri 150
Pulse/frequency output
Passive, open collector, 30 V DC, 250 mA, galvanically isolated.
• Frequency output: full scale frequency 2 to 1000 Hz (fmax = 1250 Hz), on/off ratio 1:1, pulse width max. 2 s
• Pulse output: pulse value and pulse polarity selectable, pulse width configurable (0.5 to 2000 ms)
PROFIBUS PA interface
• PROFIBUS PA in accordance with EN 50170 Volume 2, IEC 61158-2 (MBP), galvanically isolated
• Profile Version 3.0
• Current consumption: 11 mA
• Permitted supply voltage: 9 to 32 V
• Bus connection with integrated reverse polarity protection
• Error current FDE (Fault Disconnection Electronic) = 0 mA
• Data transmission rate: 31.25 kBit/s
• Signal encoding: Manchester II
• Function blocks: 4 × Analog Input, 2 × Totalizer
• Output data: Mass flow, Volume flow, Density, Temperature, Totalizer
• Input data: Positive zero return (ON/OFF), Zero point adjustment, Measuring mode, Totalizer control
• Bus address can be configured via miniature switches or via the local display (optional)
Endress+Hauser 7
Proline Promass 80E, 83E
Promass 83
Current output
Active/passive selectable, galvanically isolated, time constant selectable (0.05 to 100 s), full scale value
selectable, temperature coefficient: typically 0.005% o.f.s./°C, resolution: 0.5 μA
• Active: 0/4 to 20 mA, RL < 700 (for HART: RL 250 )
• Passive: 4 to 20 mA; supply voltage US 18 to 30 V DC; Ri 150
Pulse/frequency output
active/passive selectable, galvanically isolated
• Active: 24 V DC, 25 mA (max. 250 mA during 20 ms), RL > 100
• Passive: open collector, 30 V DC, 250 mA
• Frequency output: full scale frequency 2 to 10000 Hz (fmax = 12500 Hz), on/off ratio 1:1,
pulse width max. 2 s
• Pulse output: pulse value and pulse polarity selectable, pulse width configurable (0.05 to 2000 ms)
PROFIBUS DP interface
• PROFIBUS DP in accordance with EN 50170 Volume 2
• Profile Version 3.0
• Data transmission rate: 9.6 kBaud to 12 MBaud
• Automatic data transmission rate recognition
• Signal encoding: NRZ Code
• Function blocks: 6 × Analog Input, 3 × Totalizer
• Output data: Mass flow, Volume flow, Corrected volume flow, Density, Reference density, Temperature,
Totalizers 1 to 3
• Input data: Positive zero return (ON/OFF), Zero point adjustment, Measuring mode, Totalizer control
• Bus address can be configured via miniature switches or via the local display (optional)
• Available output combination ä 11
PROFIBUS PA interface
• PROFIBUS PA in accordance with EN 50170 Volume 2, IEC 61158-2 (MBP), galvanically isolated
• Data transmission rate: 31.25 kBit/s
• Current consumption: 11 mA
• Permitted supply voltage: 9 to 32 V
• Bus connection with
integrated reverse polarity protection
• Error current FDE (Fault Disconnection Electronic): 0 mA
• Signal encoding: Manchester II
• Function blocks: 6 × Analog Input, 3 × Totalizer
• Output data: Mass flow, Volume flow, Corrected volume flow, Density, Reference density, Temperature,
Totalizers 1 to 3
• Input data: Positive zero return (ON/OFF), Zero point adjustment, Measuring mode, Totalizer control
• Bus address can be configured via miniature switches or via the local display (optional)
• Available output combination ä 11
MODBUS interface
• MODBUS device type: slave
• Address range: 1 to 247
• Supported function codes: 03, 04, 06, 08, 16, 23
• Broadcast: supported with the function codes 06, 16, 23
• Physical interface: RS485 in accordance with EIA/TIA-485 standard
• Supported baud rate: 1200, 2400, 4800, 9600, 19200, 38400, 57600, 115200 Baud
• Transmission mode: RTU or ASCII
• Response times:
Direct data access = typically 25 to 50 ms
Auto-scan buffer (data range) = typically 3 to 5 ms
• Possible output combination ä 11
8 Endress+Hauser
Proline Promass 80E, 83E
Pulse/frequency output
Failsafe mode selectable
Low flow cutoff Switch points for low flow are selectable.
Galvanic isolation All circuits for inputs, outputs, and power supply are galvanically isolated from each other.
Endress+Hauser 9
Proline Promass 80E, 83E
Power supply
Electrical connection
Measuring unit A B C
d d
g g
b b
a
a
a
b d/(g) (d)
– 27 PA(–)/FF(–) 27
+ 26 PA(+)/FF(+) 26
– 25 d – 25 d
+ 24 + 24
f – 23 e f – 23 e
+ 22 + 22
– 21 – 21
+ 20 + 20
N (L-) 2 c N (L-) 2 c
L1 (L+)1 L1 (L+) 1
b b
A (RxD/TxD-N) 27 A (RxD/TxD-N) 27
B (RxD/TxD-P) 26 B (RxD/TxD-P) 26
– 25 d – 25 d
+ 24 + 24
f – 23 e f – 23 e
+ 22 + 22
– 21 – 21
+ 20 g + 20 g
N (L-) 2 c N (L-) 2
c
L1 (L+) 1 L1 (L+) 1
b b
a0002441
10 Endress+Hauser
Proline Promass 80E, 83E
Order version 20 (+) / 21 (–) 22 (+) / 23 (–) 24 (+) / 25 (–) 26 (+) / 27 (–)
80***-***********D Status input Status output Frequency output Current output, HART
80***-***********H - - - PROFIBUS PA
80***-***********8 Status input Frequency output Current output 2 Current output 1, HART
Promass 83
The inputs and outputs on the communication board can be either permanently assigned (fixed) or variable
(flexible), depending on the version ordered (see table). Replacements for modules which are defective or
which have to be replaced can be ordered as accessories.
Order version 20 (+) / 21 (–) 22 (+) / 23 (–) 24 (+) / 25 (–) 26 (+) / 27 (–)
83***-***********B Relay output Relay output Frequency output Current output, HART
FOUNDATION
83***-***********G - - -
Fieldbus Ex i
83***-***********H - - - PROFIBUS PA
83***-***********C Relay output 2 Relay output 1 Frequency output Current output, HART
83***-***********D Status input Relay output Frequency output Current output, HART
83***-***********E Status input Relay output Current output 2 Current output 1, HART
83***-***********L Status input Relay output 2 Relay output 1 Current output, HART
83***-***********W Relay output Current output 3 Current output 2 Current output 1, HART
Endress+Hauser 11
Proline Promass 80E, 83E
Order version 20 (+) / 21 (–) 22 (+) / 23 (–) 24 (+) / 25 (–) 26 (+) / 27 (–)
83***-***********0 Status input Current output 3 Current output 2 Current output 1, HART
83***-***********2 Relay output Current output 2 Frequency output Current output 1, HART
83***-***********3 Current input Relay output Current output 2 Current output 1, HART
83***-***********4 Current input Relay output Frequency output Current output, HART
83***-***********5 Status input Current input Frequency output Current output, HART
83***-***********6 Status input Current input Current output 2 Current output 1, HART
4 5 6 7 8 9 10 11 12 41 42
d
d
d
e
c
4 5 6 7 8 9 10 11 12 41 42
+ + + +
S1 S1 S2 S2 GND TM TM TT TT
a0003681
Cable specification remote • 6 × 0.38 mm PVC cable with common shield and individually shielded cores
version • Conductor resistance: 50 /km (0.015 /ft)
• Capacitance: core/shield: 420 pF/m (128 pF/ft)
• Cable length: max. 20 m (65 ft)
• Permanent operating temperature: max. +105 °C (+221 °F)
Operation in zones of severe electrical interference:
The measuring device complies with the general safety requirements in accordance with EN 61010-1,
the EMC requirements of IEC/EN 61326, and NAMUR recommendation NE 21/43.
12 Endress+Hauser
Proline Promass 80E, 83E
Promass 83
Lasting min. 1 power cycle:
• EEPROM and T-DAT save the measuring system data if the power supply fails.
• HistoROM/S-DAT: exchangeable data storage chip with sensor specific data (nominal diameter, serial
number, calibration factor, zero point, etc.)
Potential equalization No special measures for potential equalization are required. For instruments for use in hazardous areas, observe
the corresponding guidelines in the specific Ex documentation.
Endress+Hauser 13
Proline Promass 80E, 83E
Performance characteristics
Reference operating • Error limits following ISO/DIN 11631
conditions • Water, typically +15…+45 °C (+59…+113 °F); 2…6 bar (29…87 psi)
• Data according to calibration protocol ±5 °C (±9 °F) und ±2 bar (±29 psi)
• Accuracy based on accredited calibration rigs according to ISO 17025
Maximum measured error The following values refer to the pulse/frequency output. Measured error at the current output is typically
±5 μA. Design fundamentals ä 16.
o.r. = of reading; 1 g/cm = 1 kg/l; T = medium temperature
Density (liquids)
Temperature
±0.5 °C ± 0.005 · T °C (±1 °F ± 0.003 · (T - 32) °F)
14 Endress+Hauser
Proline Promass 80E, 83E
[%]
±1.0
±0.5
±0.2
0
0 1 2 4 6 8 10 12 14 16 18 t/h
a0006891
o.r. = of reading
Density (liquids)
±0.00025 g/cm³
Temperature
±0.25 °C ± 0.0025 · T °C (±1 °F ± 0.003 · (T – 32) °F)
Influence of fluid temperature When there is a difference between the temperature for zero point adjustment and the process temperature,
the typical measured error of the Promass sensor is ±0.0002% of the full scale value / °C (±0.0001% of the
full scale value / °F).
Influence of fluid pressure The table below shows the effect on accuracy of mass flow due to a difference between calibration pressure
and process pressure.
Endress+Hauser 15
Proline Promass 80E, 83E
DN Promass E
8 3/8" no influence
15 ½" no influence
25 1" no influence
40 1½" no influence
50 2" –0.009
80 3" –0.020
o.r. = of reading
16 Endress+Hauser
Proline Promass 80E, 83E
Mounting location
Entrained air or gas bubbles in the measuring tube can result in an increase in measuring errors.
Therefore, avoid the following mounting locations in the pipe installation:
• Highest point of a pipeline. Risk of air accumulating.
• Directly upstream of a free pipe outlet in a vertical pipeline.
a0003605
Mounting location
Notwithstanding the above, the installation proposal below permits installation in an open vertical pipeline.
Pipe restrictions or the use of an orifice with a smaller cross-section than the nominal diameter prevent the
sensor running empty while measurement is in progress.
5
a0003597
Endress+Hauser 17
Proline Promass 80E, 83E
8 3/8" 6 0.24
15 ½" 10 0.40
25 1" 14 0.55
40 1½" 22 0.87
50 2" 28 1.10
80 3" 50 2.00
Orientation
Make sure that the direction of the arrow on the nameplate of the sensor matches the direction of flow
(direction of fluid flow through the pipe).
Vertical (Fig. V)
Recommended orientation with upward direction of flow. When fluid is not flowing, entrained solids will sink
down and gases will rise away from the measuring tube. Thus the measuring tubes can be completely drained
and protected against solids buildup.
Fig. V:
ÃÃ ÃÃ
Vertical
a0004572
Fig. H1:
Horizontal ÃÃ ÃÃ
Transmitter head up
a0004576
Fig. H2:
Horizontal ÃÃ ÃÃ
Transmitter head down
a0004580
Fig. H3:
Horizontal ✘ ✘
Transmitter head to the side
a0007558
ÃÃ = Recommended orientation
à = Orientation recommended in certain situations
✘ = Impermissible orientation
In order to ensure that the permissible ambient temperature range for the transmitter is not exceeded, we
recommend the following orientations:
• For fluids with very high temperatures, we recommend the horizontal orientation with the transmitter head
pointing downwards (Fig. H2) or the vertical orientation (Fig. V).
18 Endress+Hauser
Proline Promass 80E, 83E
• For fluids with very low temperatures, we recommend the horizontal orientation with the transmitter head
pointing upwards (Fig. H1) or the vertical orientation (Fig. V).
" Caution!
When using a bent measuring tube and horizontal installation, the position of the sensor has to be matched to
the fluid properties!
1 2
a0004581
Heating
Some fluids require suitable measures to avoid heat transfer at the sensor. Heating can be electric, e.g. with
heated elements, or by means of hot water or steam pipes made of copper or heating jackets.
" Caution!
• Risk of electronics overheating! Make sure that the maximum permissible ambient temperature for the
transmitter is not exceeded. Consequently, make sure that the adapter between the sensor and transmitter
and the connection housing of the remote version always remain free of insulating material. Note that a
certain orientation might be required, depending on the fluid temperature ä 21.
• If using an electric trace heating system whose heating is regulated via phase angle control or pulse packages,
influence on the measured values cannot be ruled out due to magnetic fields (i.e. for values that are greater
than the values approved by the EN standard (sine 30 A/m)). In such cases, the sensor must be magnetically
shielded.
The housing can be shielded with tin plates or electric sheets without preferential direction (e.g. V330-35A)
with the following properties:
– Relative magnetic permeability μr 300
– Plate thickness d 0.35 mm (d 0.014")
• Information on permitted temperature ranges ä 21
Special heating jackets, which can be ordered separately from Endress+Hauser as an accessory, are available
for the sensors.
Thermal insulation
Some fluids require suitable measures to avoid loss of heat at the sensor. A wide range of materials can be used
to provide the required thermal insulation.
Endress+Hauser 19
Proline Promass 80E, 83E
Inlet and outlet runs There are no installation requirements regarding inlet and outlet runs.
System pressure It is important to ensure that cavitation does not occur, because it would influence the oscillation of the
measuring tube. No special measures need to be taken for fluids which have properties similar to water under
normal conditions.
In the case of liquids with a low boiling point (hydrocarbons, solvents, liquefied gases) or in suction lines, it is
important to ensure that pressure does not drop below the vapor pressure and that the liquid does not start to
boil. It is also important to ensure that the gases that occur naturally in many liquids do not outgas. Such effects
can be prevented when system pressure is sufficiently high.
Therefore, the following locations should be preferred for installation:
• Downstream from pumps (no danger of vacuum)
• At the lowest point in a vertical pipe
! Note!
• Install the device at a shady location. Avoid direct sunlight, particularly in warm climatic regions.
• At ambient temperatures below –20 °C (–4 °F) the readability of the display may be impaired.
Storage temperature –40 to +80 °C (–40 to +175 °F), preferably +20 °C (+68 °F)
20 Endress+Hauser
Proline Promass 80E, 83E
Secondary containment
The sensor Promass E has no secondary containment.
Rupture disk in the sensor The sensor housing protects the inner electronics and mechanics and is filled with dry nitrogen. The housing
housing (optional) of this sensor does not fulfill any additional secondary containment function. However, 15 bar (217.5 psi) can
be specified as a reference value for the pressure loading capacity.
For increased safety, a version with rupture disk (triggering pressure 10 to 15 bar / 145 to 217.5 psi) can be
used, which is available for order as a separate option.
Further information ä 40.
Pressure loss Pressure loss depends on the fluid properties and on the flow rate. The following formulae can be used to
approximately calculate the pressure loss:
2·g
Reynolds number Re =
p·d·n·r
a0004623
Dp = K · n ·r
0.25 1.85 –0.86
Re 23001) ·g
a0004626
K2 · n · g
0.25 2
Re < 2300 Dp = K1 · n · g +
r
a0004628
Endress+Hauser 21
Proline Promass 80E, 83E
DN d[m] K K1 K2
[mbar]
10000
DN 40
DN 25
DN 15 DN 50
DN 8
DN 80
1000
100
10
0.1
0.001 0.01 0.1 1 10 100 1000 [t/h]
a0004606
22 Endress+Hauser
Proline Promass 80E, 83E
Mechanical construction
Design, dimensions Dimensions:
Transmitter remote version, wall-mount housing (non hazardous area and II3G / zone 2) ä 27
VCO connections ä 32
Tri-Clamp ä 33
VCO connections ä 38
Tri-Clamp ä 39
Rupture disc ä 40
Endress+Hauser 23
Proline Promass 80E, 83E
A B
A* C
E
E
G
F
di L
A0007638
Dimensions SI units
DN A A* B C D E F G L di
1) 1)
8 227 207 187 168 160 224 93 317
1) 1)
40 227 207 187 168 160 237 121 358
1)
dependent on respective process connection
* Blind version (without local display)
All dimensions in [mm]
Dimensions in US units
DN A A* B C D E F G L di
2) 2)
½" 9.08 8.28 7.48 6.72 6.40 8.90 4.13 13.03
2) 2)
2" 9.08 8.28 7.48 6.72 6.40 9.96 6.69 16.65
1)
dependent on respective process connection
* Blind version (without local display)
All dimensions in [inch]
24 Endress+Hauser
Proline Promass 80E, 83E
A B
A* C
e r Sp annu
ht unt ng
Nic öffnen
Nicht-eigensichere
ove r tight wh
e
IP40-Abdeckung geschützt
Non-intrinsically safe
D
circuits Ip40 protected
ep c
Boucles de courant
c
sans sécurité intrinsèque
protégées par Ip40
E
sou s nsion l Ne
pas o t e pparei
uvrir l’a
G
F
di L
A0015135
Dimensions SI units
DN A A* B C D E F G L di
1) 1)
8 240 217 206 186 178 242 93 335
1) 1)
40 240 217 206 186 178 255 121 376
1)
dependent on respective process connection
* Blind version (without local display)
All dimensions in [mm]
Dimensions in US units
DN A A* B C D E F G L di
1) 1)
½" 9,45 8,54 8,11 7,32 7,01 9,60 4,13 13,73
1) 1)
2" 9,45 8,54 8,11 7,32 7,01 10.66 6,69 17,35
1)
dependent on respective process connection
* Blind version (without local display)
All dimensions in [inch]
Endress+Hauser 25
Proline Promass 80E, 83E
A
A*
B C
B* D
e r Sp annu
ht unt ng
Nic öffnen
Nicht-eigensichere
ove r tight wh
e
c
IP40-Abdeckung geschützt
p cover tight w
Non-intrinsically safe
circuits Ip40 protected
E
ep c
Esc
Boucles de courant
c
– +
ve
E
sans sécurité intrinsèque
protégées par Ip40
sou s nsion l Ne
pas o t e pparei
uvrir l’a
M
L
J
K
F
G
a0002128
Dimensions in SI units
A A* B B* C D E F G H J K L M
8.6
265 242 240 217 206 186 178 100 130 100 144 170 348
(M8)
Dimensions in US units
A A* B B* C D E F G H J K L M
10.4 9.53 9.45 8.54 8.11 7.32 7.01 3.94 5.12 3.94 5.67 6.69 13.7
(M8)
26 Endress+Hauser
Proline Promass 80E, 83E
Transmitter remote version, wall-mount housing (non hazardous area and II3G / zone 2)
D
Esc
- + E
B
C
F G
A E
H J K J
N
S S
M
R
L
O
P Q P
a0001150
Dimensions in SI units
A B C D E F G H J
K L M N O P Q R S
Dimensions in US units
A B C D E F G H J
K L M N O P Q R S
Endress+Hauser 27
Proline Promass 80E, 83E
C
a0007637
Dimensions in SI units
DN A B C
Dimensions in US units
DN A B C
28 Endress+Hauser
Proline Promass 80E, 83E
LK
G
U
E
+1.5 (+0.06)
di L –2.0 (–0.08) mm (inch)
a0007640-ae
Surface roughness (flange): EN 1092-1 Form B1 (DIN 2526 Form C), Ra 3.2 to 12.5 μm
DN G L N S LK U di
Surface roughness (flange): EN 1092-1 Form B1 (DIN 2526 Form C), Ra 3.2 to 12.5 μm
DN G L N S LK U di
Flange according to EN 1092-1 (DIN 2501 / DIN 2512N 1)) / PN 63: 1.4404/316L
Surface roughness (flange): EN 1092-1 Form B2 (DIN 2526 Form E), Ra 0.8 to 3.2 μm
DN G L N S LK U di
Endress+Hauser 29
Proline Promass 80E, 83E
Surface roughness (flange): EN 1092-1 Form B2 (DIN 2526 Form E), Ra 0.8 to 3.2 μm
DN G L N S LK U di
DN G L N S LK U di
DN G L N S LK U di
DN G L N S LK U di
30 Endress+Hauser
Proline Promass 80E, 83E
DN G L N S LK U di
DN G L N S LK U di
DN G L N S LK U di
DN G L N S LK U di
Endress+Hauser 31
Proline Promass 80E, 83E
VCO connections
U
G
E
+1.5 (+0.06)
di L –2.0 (–0.08) mm (inch)
a0007641-ae
DN G L U di
32 Endress+Hauser
Proline Promass 80E, 83E
Tri-Clamp
G
U
E
+1.5 (+0.06)
di L –2.0 (–0.08) mm (inch)
a0007643-ae
DN Clamp G L U di
½"-Tri-Clamp: 1.4404/316L
DN Clamp G L U di
Endress+Hauser 33
Proline Promass 80E, 83E
G
U
+1.5 (+0.06)
di L –2.0 (–0.08) mm (inch)
a0007644-ae
DN G L U di
3A version also available (Ra 0.8 μm/150 grit.); All dimensions in [mm]
E
G
U
+1.5 (+0.06)
di L –2.0 (–0.08) mm (inch)
a0007649-ae
DN G L U di
3A version also available (Ra 0.8 μm/150 grit.); All dimensions in [mm]
34 Endress+Hauser
Proline Promass 80E, 83E
LK
G
U
+1.5 (+0.06)
di L –2.0 (–0.08)
S
mm (inch)
a0007649-ae
DN G L N S LK U di
8 54 249 4 × Ø9 10 37 10 5.35
15 59 293 4 × Ø9 10 42 16 8.30
25 70 344 4 × Ø9 10 53 26 12.0
40 82 456 4 × Ø9 10 65 38 17.6
50 94 562 4 × Ø9 10 77 50 26.0
Endress+Hauser 35
Proline Promass 80E, 83E
G
U
+1.5 (+0.06)
di L –2.0 (–0.08) mm (inch)
a0007651-ae
DN G L U di
E
G
U
+1.5 (+0.06)
di L –2.0 (–0.08) mm (inch)
a0007653-ae
DN G L U di
3A version also available (Ra 0.8 μm/150 grit.); All dimensions in [mm]
36 Endress+Hauser
Proline Promass 80E, 83E
LK
G
U
E
+1.5 (+0.06)
di L –2.0 (–0.08) mm (inch)
a0007640-ae
DN G L N S LK U di
DN G L N S LK U di
DN G L N S LK U di
Endress+Hauser 37
Proline Promass 80E, 83E
VCO connections
U
G
E
+1.5 (+0.06)
di L –2.0 (–0.08) mm (inch)
a0007641-ae
DN G L U di
38 Endress+Hauser
Proline Promass 80E, 83E
Tri-Clamp
G
U
E
+1.5 (+0.06)
di L –2.0 (–0.08) mm (inch)
a0007643-ae
DN Clamp G L U di
½"-Tri-Clamp: 1.4404/316L
DN Clamp G L U di
Endress+Hauser 39
Proline Promass 80E, 83E
G
U
+1.5 (+0.06)
di L –2.0 (–0.08) mm (inch)
a0007653-ae
DN G L U di
Rupture disc Sensor housings with integrated rupture disks are optionally available.
# Warning!
• Make sure that the function and operation of the rupture disk is not impeded through the installation.
Triggering overpressure in the housing as stated on the indication label. Take adequate precautions to ensure
that no damage occurs, and risk to human life is ruled out, if the rupture disk is triggered.
Rupture disk: Burst pressure 10 to 15 bar (145 to 217.5 psi).
• It is not permitted to open the connections or remove the rupture disk.
" Caution!
Rupture disks can not be combined with separately available heating jacket.
! Note!
• Please note the indication labels.
RUPTURE DISK
12,5 BAR +/-20%@20°C
A0007823
The position of the rupture disk is indicated by an adhesive label on top of the disk. If the rupture disk is
triggered, the adhesive label is damaged and can thus be visually monitored.
40 Endress+Hauser
Proline Promass 80E, 83E
Weight in SI units
DN [mm] 8 15 25 40 50 80
Compact version 8 8 10 15 22 31
Weight in US units
Compact version 18 18 22 33 49 69
Process connections
• Stainless steel 1.4404/316L
– Flanges according to EN 1092-1 (DIN 2501) and according to ASME B16.5
– DIN 11864-2 Form A (flat flange with groove)
– Threaded hygienic connection: DIN 11851, SMS 1145, ISO 2853, DIN 11864-1 Form A
– VCO connections
• Stainless steel SUS 316L
– Flanges to JIS B2220
Measuring tubes
• Stainless steel EN 1.4539 / ASTM 904L
• Finish quality: Ramax 0.8 μm/150 grit (30 μin/150 grit)
Seals
Welded process connections without internal seals
Endress+Hauser 41
Proline Promass 80E, 83E
# Warning!
The following material load curves refer to the entire sensor and not just the process connection.
[psi] [bar]
1500
100
1400 PN 100
1300 90
1200
80
1100
1000 70
900 60 PN 63
800
50
700
600 40
PN 40
500
30
400
300 20
200
10
100
0 0
-60 -40 -20 0 20 40 60 80 100 120 140 160 180 200 [°C]
-80 -40 0 40 80 120 160 200 240 280 320 360 400 [°F]
a0006904-ae
[psi] [bar]
1500
100
1400
1300 90
1200
80 Class 600
1100
1000 70
900 60
800
50
700
600 40
Class 300
500
30
400
300 20
200
10 Class 150
100
0 0
-60 -40 -20 0 20 40 60 80 100 120 140 160 180 200 [°C]
-80 -40 0 40 80 120 160 200 240 280 320 360 400 [°F]
a0006905-ae
42 Endress+Hauser
Proline Promass 80E, 83E
[psi] [bar]
1000 70
900 60 63K
800
50
700
600 40
40K
500
30
400
300 20
20K
200
10
100 10K
0 0
-60 -40 -20 0 20 40 60 80 100 120 140 160 180 200 [°C]
-80 -40 0 40 80 120 160 200 240 280 320 360 400 [°F]
A0006906-ae
[psi] [bar]
1500
100
1400 PN 100
1300 90
1200
80
1100
1000 70
-60 -40 -20 0 20 40 60 80 100 120 140 160 180 200 [°C]
[°F]
-80 -40 0 40 80 120 160 200 240 280 320 360 400
a0006908-ae
[psi] [bar]
50
700
600 40
DN 8…40
500
30
400
300 20 DN 50…80
200
10
100
0 0
-60 -40 -20 0 20 40 60 80 100 120 140 160 180 200 [°C]
-80 -40 0 40 80 120 160 200 240 280 320 360 400 [°F]
A0006909
DIN 11851 allows for applications up to +140 °C (+284 °F) if suitable sealing materials are used. Please take this into
account when selecting seals and counterparts as these components can limit the pressure and temperature range.
Endress+Hauser 43
Proline Promass 80E, 83E
[psi] [bar]
500
30
400
300 20
200 PN 16
10
100
0 0
-60 -40 -20 0 20 40 60 80 100 120 140 160 180 200 [°C]
-80 -40 0 40 80 120 160 200 240 280 320 360 400 [°F]
A0012947
SMS 1145 allows for applications up to 6 bar (87 psi) if suitable sealing materials are used. Please take this into account
when selecting seals and counterparts as these components can limit the pressure and temperature range.
[psi] [bar]
50
700
600 40
DN 8...40
500
30
400
300 20 DN 50…80
200
10
100
0 0
-60 -40 -20 0 20 40 60 80 100 120 140 160 180 200 [°C]
-80 -40 0 40 80 120 160 200 240 280 320 360 400 [°F]
A0006910-ae
[psi] [bar]
600 40
500
30 DN 8...40
400
300 20
200 DN 50…80
10
100
0 0
-60 -40 -20 0 20 40 60 80 100 120 140 160 180 200 [°C]
-80 -40 0 40 80 120 160 200 240 280 320 360 400 [°F]
A0006911-ae
44 Endress+Hauser
Proline Promass 80E, 83E
[psi] [bar]
30
400
300 20
200
10
100
0 0
-60 -40 -20 0 20 40 60 80 100 120 140 160 180 200 [°C]
-80 -40 0 40 80 120 160 200 240 280 320 360 400 [°F]
A0006912-ae
Human interface
Display elements • Liquid-crystal display: backlit, two lines (Promass 80) or four lines (Promass 83) with 16 characters per line
• Selectable display of different measured values and status variables
• At ambient temperatures below –20 °C (–4 °F) the readability of the display may be impaired.
Promass 83
• Local operation with three optical keys (S/O/F)
• Application-specific Quick Setup menus for straightforward commissioning
Promass 83
Remote operation via HART, PROFIBUS DP/PA, FOUNDATION fieldbus, MODBUS RS485
Endress+Hauser 45
Proline Promass 80E, 83E
C-Tick symbol The measuring system complies with the EMC requirements of the "Australian Communications and Media
Authority (ACMA)"
Ex approval Information about currently available Ex versions (ATEX, FM, CSA, IECEx, NEPSI etc.) can be supplied by your
Endress+Hauser Sales Center on request. All information relevant to explosion protection is available in
separate Ex documents that you can order as necessary.
FOUNDATION Fieldbus The flow device has successfully passed all the test procedures carried out and is certified and registered by the
certification Fieldbus Foundation. The device thus meets all the requirements of the following specifications:
• Certified to FOUNDATION Fieldbus Specification
• The device meets all the specifications of the FOUNDATION Fieldbus H1.
• Interoperability Test Kit (ITK), revision status 5.01 (device certification number: on request)
• The device can also be operated with certified devices of other manufacturers
• Physical Layer Conformance Test of the Fieldbus Foundation
PROFIBUS DP/PA The flow device has successfully passed all the test procedures carried out and is certified and registered by the
certification PNO (PROFIBUS User Organization). The device thus meets all the requirements of the following
specifications:
• Certified in accordance with PROFIBUS Profile Version 3.0 (device certification number: available on
request)
• The device can also be operated with certified devices of other manufacturers (interoperability)
MODBUS certification The measuring device meets all the requirements of the MODBUS/TCP conformity test and has the
"MODBUS/TCP Conformance Test Policy, Version 2.0". The measuring device has successfully passed all the
test procedures carried out and is certified by the "MODBUS/TCP Conformance Test Laboratory" of the
University of Michigan.
Pressure Equipment Directive The measuring devices can be ordered with or without PED (Pressure Equipment Directive). If a device with
PED is required, this must be ordered explicitly. For devices with nominal diameters less than or equal to DN
25 (1"), this is neither possible nor necessary.
• With the identification PED/G1/III on the sensor nameplate, Endress+Hauser confirms conformity with the
"Basic safety requirements" of Appendix I of the Pressure Equipment Directive 97/23/EC.
• Devices with this identification (with PED) are suitable for the following types of fluid:
- Fluids of Group 1 and 2 with a steam pressure of greater or less than 0.5 bar (7.3 psi)
- Unstable gases
46 Endress+Hauser
Proline Promass 80E, 83E
• Devices without this identification (without PED) are designed and manufactured according to good
engineering practice. They correspond to the requirements of Art. 3, Section 3 of the Pressure Equipment
Directive 97/23/EC. Their application is illustrated in Diagrams 6 to 9 in Appendix II of the Pressure
Equipment Directive 97/23/EC.
Promass 80
Promass80***-************A
Promass80***-************D
Promass80***-************S
Promass80***-************T
Promass80***-************8
Promass 83
Promass83***-************A Promass83***-************M Promass83***-************Ø
Promass83***-************B Promass83***-************R Promass83***-************2
Promass83***-************C Promass83***-************S Promass83***-************3
Promass83***-************D Promass83***-************T Promass83***-************4
Promass83***-************E Promass83***-************U Promass83***-************5
Promass83***-************L Promass83***-************W Promass83***-************6
Ordering Information
The Endress+Hauser service organization can provide detailed ordering information and information on the
order codes upon request.
Accessories
Various accessories, which can be ordered separately from Endress+Hauser, are available for the transmitter
and the sensor.
Documentation
• Flow measuring technology (FA005D)
• Technical Information
– Promass 80A, 83A (T054D)
– Promass 80F, 83F (TI101D)
– Promass 80H, 83H (TI074D
– Promass 80I, 83I (TI075D)
– Promass 80P, 83P (TI078D)
– Promass 80S, 83S (TI076D)
• Operating Instructions/Description of Device Functions
– Promass 80 HART (BA057D/BA058D)
– Promass 80 PROFIBUS PA (BA072D/BA073D)
– Promass 83 HART (BA059D/BA060D)
– Promass 83 FOUNDATION Fieldbus (BA065D/BA066D)
– Promass 83 PROFIBUS DP/PA(BA063D/BA064D)
– Promass 83 MODBUS (BA107D/BA108D)
• Supplementary documentation on Ex-ratings: ATEX, FM, CSA, IECEx NEPSI
• Functional safety manual Promass 80, 83 (SD077D)
Endress+Hauser 47
Proline Promass 80E, 83E
Registered trademarks
TRI-CLAMP ®
Registered trademark of Ladish & Co., Inc., Kenosha, WI, USA
HART®
Registered trademark of HART Communication Foundation, Austin, TX, USA
PROFIBUS®
Registered trademark of the PROFIBUS User Organization, Karlsruhe, Germany
FOUNDATION™ Fieldbus
Registered trademark of the Fieldbus FOUNDATION, Austin, USA
MODBUS®
Registered trademark of the MODBUS Organization
HistoROM™, S-DAT®, T-DAT™, F-CHIP®, Fieldcheck®, FieldCare®, Applicator®
Registered or registration-pending trademarks of Endress+Hauser Flowtec AG, Reinach, CH
48 Endress+Hauser
Proline Promass 80E, 83E
Endress+Hauser 49
Proline Promass 80E, 83E
50 Endress+Hauser
Proline Promass 80E, 83E
Endress+Hauser 51
Instruments International
Endress+Hauser
Instruments International AG
Kaegenstrasse 2
4153 Reinach
Switzerland
TI00061D/06/EN/13.11
71129459
FM+SGML6.0 ProMoDo