E&H Radar Manual
E&H Radar Manual
E&H Radar Manual
Technical Information
Micropilot FMR62
Free space radar
Application
• Continuous, non-contact level measurement of liquids,
pastes and sludges
• Integrated PEEK antenna or PTFE-plated, flush-mounted antenna
• Maximum measuring range: 80 m (262 ft)
• Temperature: –40 to +200 °C (–40 to +392 °F)
• Pressure: –1 to +25 bar (–14.5 to +362.6 psi)
• Accuracy: ± 1 mm (0.04 in)
• Linearity protocol (3-point, 5-point)
Your benefits
• Reliable measurement thanks to improved focusing and small beam angle,
particularly in vessels with many internal fittings
• Compact design enables installation in the case of small vessels and small process
connections
• Safety by design - ensures highest safety
• Easy, guided commissioning with intuitive user interface
• Maximum reliability thanks to multi-echo tracking
• HistoROM configuration memory makes for easier commissioning,
maintenance and diagnostics
• SIL2 as per IEC 61508, SIL3 for homogeneous or diverse redundancy
• Easy proof testing for SIL and WHG
• RFID TAG - easy identification of measuring points for improved data access
• Heartbeat Technology
Micropilot FMR62
Table of contents
Mechanical construction . . . . . . . . . . . . . . . . . . . . 43
Function and system design . . . . . . . . . . . . . . . . . . . 8
Dimensions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
Measuring principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
Weight . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 50
Materials: GT18 housing (stainless steel, corrosion-
Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 resistant) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 51
Measured variable . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Materials: GT19 housing (plastic) . . . . . . . . . . . . . . . . . . 52
Measuring range . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 9 Materials: GT20 housing (die-cast aluminum, powder-
Operating frequency . . . . . . . . . . . . . . . . . . . . . . . . . . 13 coated) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53
Transmission power . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 Materials: antenna and process connection . . . . . . . . . . . 54
Materials: Weather protection cover . . . . . . . . . . . . . . . . 56
Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Output signal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 Operability . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Signal on alarm . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 Operating concept . . . . . . . . . . . . . . . . . . . . . . . . . . . . 57
Linearization . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 Local operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Galvanic isolation . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15 Operation with remote display and operating module
Protocol-specific data . . . . . . . . . . . . . . . . . . . . . . . . . . 16 FHX50 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 58
Remote operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . 59
SupplyCare inventory management software . . . . . . . . . . 60
Power supply . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Terminal assignment . . . . . . . . . . . . . . . . . . . . . . . . . . 17
Device plug connectors . . . . . . . . . . . . . . . . . . . . . . . . . 21 Certificates and approvals . . . . . . . . . . . . . . . . . . . 63
Supply voltage . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22 CE mark . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Power consumption . . . . . . . . . . . . . . . . . . . . . . . . . . . 23 RoHS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Current consumption . . . . . . . . . . . . . . . . . . . . . . . . . . 23 RCM-Tick marking . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Power supply failure . . . . . . . . . . . . . . . . . . . . . . . . . . 23 Ex approval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Potential equalization . . . . . . . . . . . . . . . . . . . . . . . . . 23 Dual seal according to ANSI/ISA 12.27.01 . . . . . . . . . . . . 63
Cable entries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24 Functional safety . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Cable specification . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24 WHG . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63
Overvoltage protection . . . . . . . . . . . . . . . . . . . . . . . . . 24 Hygienic compatibility . . . . . . . . . . . . . . . . . . . . . . . . . 63
NACE MR 0175 / ISO 15156 . . . . . . . . . . . . . . . . . . . . . 64
NACE MR 0103 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Performance characteristics . . . . . . . . . . . . . . . . . . 26
Pressure equipment with allowable pressure
Reference operating conditions . . . . . . . . . . . . . . . . . . . 26
≤ 200 bar (2 900 psi) . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Reference accuracy . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Marine approval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 64
Measured value resolution . . . . . . . . . . . . . . . . . . . . . . 26
Radio standard EN 302729-1/2 . . . . . . . . . . . . . . . . . . . 64
Response time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Radio standard EN 302372-1/2 . . . . . . . . . . . . . . . . . . . 65
Influence of ambient temperature . . . . . . . . . . . . . . . . . 27
FCC . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 65
Influence of gas phase . . . . . . . . . . . . . . . . . . . . . . . . . 27
Industry Canada . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
CRN approval . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 66
Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 28 Test, certificate . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 67
Installation conditions . . . . . . . . . . . . . . . . . . . . . . . . . 28 Hard-copy product documentation . . . . . . . . . . . . . . . . . 67
Installation: integrated antenna . . . . . . . . . . . . . . . . . . . 32 Other standards and guidelines . . . . . . . . . . . . . . . . . . . 68
Installation: flush mount antenna . . . . . . . . . . . . . . . . . . 33
Container with heat insulation . . . . . . . . . . . . . . . . . . . . 35
Ordering information . . . . . . . . . . . . . . . . . . . . . . . 69
Ordering information . . . . . . . . . . . . . . . . . . . . . . . . . . 69
Environment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36 3-point linearity protocol . . . . . . . . . . . . . . . . . . . . . . . 70
Ambient temperature range . . . . . . . . . . . . . . . . . . . . . 36 5-point linearity protocol . . . . . . . . . . . . . . . . . . . . . . . 71
Ambient temperature limits . . . . . . . . . . . . . . . . . . . . . 36 Customer-specific configuration . . . . . . . . . . . . . . . . . . . 72
2 Endress+Hauser
Micropilot FMR62
Tagging (TAG) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
Services . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 72
Application Packages . . . . . . . . . . . . . . . . . . . . . . . 73
Heartbeat Diagnostics . . . . . . . . . . . . . . . . . . . . . . . . . 73
Heartbeat Verification . . . . . . . . . . . . . . . . . . . . . . . . . 74
Heartbeat Monitoring . . . . . . . . . . . . . . . . . . . . . . . . . 75
Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 76
Device-specific accessories . . . . . . . . . . . . . . . . . . . . . . 76
Communication-specific accessories . . . . . . . . . . . . . . . . 79
Service-specific accessories . . . . . . . . . . . . . . . . . . . . . . 80
System components . . . . . . . . . . . . . . . . . . . . . . . . . . . 80
Supplementary documentation . . . . . . . . . . . . . . . 81
Standard documentation . . . . . . . . . . . . . . . . . . . . . . . . 81
Safety Instructions (XA) . . . . . . . . . . . . . . . . . . . . . . . . 81
Endress+Hauser 3
Micropilot FMR62
DANGER!
DANGER This symbol alerts you to a dangerous situation. Failure to avoid this situation will
result in serious or fatal injury.
WARNING!
WARNING This symbol alerts you to a dangerous situation. Failure to avoid this situation can
result in serious or fatal injury.
CAUTION!
CAUTION This symbol alerts you to a dangerous situation. Failure to avoid this situation can
result in minor or medium injury.
NOTE!
NOTICE This symbol contains information on procedures and other facts which do not result in
personal injury.
Direct current
Alternating current
Ground connection
A grounded terminal which, as far as the operator is concerned, is grounded via a
grounding system.
Preferred
Procedures, processes or actions that are preferred.
Forbidden
Procedures, processes or actions that are forbidden.
Tip
Indicates additional information.
Reference to documentation.
A Reference to page.
Reference to graphic.
Visual inspection.
1. , 2. , 3. … Series of steps
A, B, C, ... Views
4 Endress+Hauser
Micropilot FMR62
Symbol Meaning
Hazardous area
- Indicates a hazardous area.
Safety instructions
Observe the safety instructions contained in the associated Operating Instructions.
Endress+Hauser 5
Micropilot FMR62
PN Nominal pressure
FieldCare Scalable software tool for device configuration and integrated plant asset management
solutions
Operating tool The term "operating tool" is used in place of the following operating software:
FieldCare / DeviceCare, for operation via HART communication and PC
6 Endress+Hauser
Micropilot FMR62
Registered trademarks
HART®
Registered trademark of the HART Communication Foundation, Austin, USA
KALREZ®, VITON®
Registered trademark of DuPont Performance Elastomers L.L.C., Wilmington, USA
TEFLON®
Registered trademark of E.I. DuPont de Nemours & Co., Wilmington, USA
TRI CLAMP®
Registered trademark of Ladish Co. Inc., Kenosha, USA
Endress+Hauser 7
Micropilot FMR62
D
A0032017
The frequency of this wave is modulated in the form of a sawtooth signal between two limit
frequencies f1 and f2:
f 1
2
f2
Δf
Δt
f1
t
A0023771
This results in the following difference frequency at any time between the transmitted signal and the
received signal:
Δf = k Δt
where Δt is the run time and k is the specified increase in frequency modulation.
Δt is given by the distance D between the reference point R and the product surface:
D = (c Δt) / 2
8 Endress+Hauser
Micropilot FMR62
Input
Measured variable The measured variable is the distance between the reference point and the product surface. The level
is calculated based on "E", the empty distance entered. Optionally, the level can be converted to other
variables (volume, mass) by linearization (32 value pairs).
Media groups
Media group DC (εr) Example
D > 10 conducting liquids, e.g. aqueous solutions, dilute acids and alkalis
For dielectric constants (DC values) of many media commonly used in various industries refer
to:
• the Endress+Hauser DC manual (CP01076F)
• the Endress+Hauser "DC Values App" (available for Android and iOS)
Endress+Hauser 9
Micropilot FMR62
FMR62
A0034565
A0034566 A0034567
A0 A B C D A0 A B C D A B C D
1 1
1.5 (3.3) 1.5 (3.3) 1,5
(5) 2.5 (5) (5)
(8) 3 3
(10) (10)
5 5
(16) 6 (16)
A0034585
(20)
8 8
(26) (26)
A0034580
10
(33)
A0034575
10 Endress+Hauser
Micropilot FMR62
FMR62
A0034565
A0034566 A0034567
A0 A B C D A0 A B C D A0 A B C D
1
1.5 (3.3) 1,5
(5) (5)
3 3 3
(10) (10) (10)
6 6
(20) (20) 7
(23)
11 11
(36) 13 (36)
15 (43) A0034586
(49)
20
22
(66)
(72) A0034581
A0034576
Endress+Hauser 11
Micropilot FMR62
FMR62
A0034565
A0034566 A0034567
A0 A B C D A0 A B C D A0 A B C D
2
(7)
4 4
(13) (13)
7 7
(23) (23)
7
(23)
12
13
(39)
(43) 15
(49)
23
25
(75) 28 (82)
(92) A0034587
40
44
(131) 50
(144)
(164) A0034582
A0034577
12 Endress+Hauser
Micropilot FMR62
FMR62
A0034565
A0034566 A0034567
A0 A B C D A0 A B C D A0 A B C D
7
(23)
12
13
(39)
(43)
22
(72) 23 25
(75) (82)
40
(131) 50 45
(148) 50
(164) 60 (164) 60
65 (197) 70 (197)
(213) (230)
A0034588
80 A0034583
(262)
A0034578
Endress+Hauser 13
Micropilot FMR62
14 Endress+Hauser
Micropilot FMR62
Output
Output signal HART
Switch output
For HART devices, the switch output is available as an option. See product structure, feature 20:
"Power Supply, Output", option B: "2-wire; 4-20mA HART, switch output"
Switch output
Switching behavior Binary (conductive or non-conductive), switches when the programmable switch
point is reached
Insulation voltage floating, Insulation voltage 1 350 VDC to power supply aund 500 VAC to ground
Switch point freely programmable, separately for switch-on and switch-off point
Switching delay freely programmable from 0 to 100 s, separately for switch-on and switch-off
point
Linearization The linearization function of the device allows the conversion of the measured value into any unit of
length or volume. Linearization tables for calculating the volume in cylindrical tanks are pre-
programmed. Other linearization tables of up to 32 value pairs can be entered manually or semi-
automatically.
Galvanic isolation All circuits for the outputs are galvanically isolated from each other.
Endress+Hauser 15
Micropilot FMR62
Manufacturer ID 17 (0x11)
HART device variables The measured values can be freely assigned to the device variables.
Measured values for PV (primary variable)
• Level linearized
• Distance
• Electronic temperature
• Relative echo amplitude
• Area of incoupling
• Analog output adv. diagnostics 1
• Analog output adv. diagnostics 2
Measured values for SV, TV, FV (second, third and fourth variable)
• Level linearized
• Distance
• Electronic temperature
• Terminal voltage
• Relative echo amplitude
• Absolute echo amplitude
• Area of incoupling
• Analog output adv. diagnostics 1
• Analog output adv. diagnostics 2
Start-up time 65 s
Set-up time 15 s
16 Endress+Hauser
Micropilot FMR62
Power supply
Terminal assignment 2-wire: 4-20mA HART
! 2- w
4-2 ire le
0 vel
H A mA
[21 RT
4-2
0m
A ]
ope
A
n Sp
a
711 re pa
HA 20mA
08x rt
xx
RT
4...
+
1 -
1 2 4 5 6
mm
2
10
+ +
–
–
# 4...20 mA
3 B
7 +
1-c 1 -
1 2 4 5 han
nel 4-2 2
ove 0m
[16
] rvo A
ltag
e pro
tec
tion
+ +
–
–
# 4...20 mA
3
A0011294
Endress+Hauser 17
Micropilot FMR62
10
9
8
% 2- w
4-2 ire
0
H A mA
[02 RT
PF
S
A /03
] ope
n Sp
a
711 re pa
HA 20 mA
08x rt
3+ ++ xx
31 --
RT
4...
42 +
7
1 -
mm
2 6
4-
10
+ +
–
-
≥250 Ω 4...20 mA
$
1 2 3 4 5 B
3+ 11 +
3 -
4-2 4
0m +
2-c
A/ 1 -
han 4-2 2
4- [17
nel
ove FIE 0mA
] rvo LD /
ltage BU
pro S
tec
tion
+ +
–
-
≥250 Ω 4...20 mA
$
1 2 3 4 5
A0013759
18 Endress+Hauser
Micropilot FMR62
14
13
12
# 2- w
4-2 ire le
0 v el
H A mA
[04 RT
4-2
0m
A /05
]
A
ope
A
n Sp
20m
a
711 re pa
HA 20mA
08x rt
xx
+
4 ...
3 -
RT
11
4 ...
4 +
1 -
mm
2
+ +
10
1 " –
-
4...20 mA
9 8 4
+ +
2 " –
-
4...20 mA
3 5 6 7 4
B
+ +
1 " –
- +
3 -
4...20 mA 4-2 4
10 0m +
9 8 4 2-c
A/ 1 -
han
nel 4- 2
[17
ove FIE 20mA
] rvo LD /
ltag
ep
B U
ro S
tec
tion
+ +
2 " –
-
4...20 mA
3 5 6 7 4
A0013923
This version is also suited for single-channel operation. In this case, current output 1 (terminals
1 and 2) must be used.
Endress+Hauser 19
Micropilot FMR62
+
3+ 1
2
Ri + 3+
-
Ri
4-
A0015909
4-
6 Connection of a relay A0015910
20 Endress+Hauser
Micropilot FMR62
Device plug connectors For the versions with fieldbus plug connector (M12 or 7/8"), the signal line can be connected
without opening the housing.
2 not connected
1 2 3 Signal -
A0011175
4 Ground
2 Signal +
2 4 3 Not connected
A0011176
4 Screen
Endress+Hauser 21
Micropilot FMR62
A: • Non- 14 to 35 V R [W]
2-wire; hazardous
4-20mA • Ex nA 500
HART • Ex ic
• CSA GP
Ex ia / IS 14 to 30 V
• Ex d(ia) / XP 14 to
• Ex ic(ia) 35 V 3)
• Ex nA(ia) 0
• Ex ta / DIP 10 20 30 35 U0 [V]
14 25
A0031745
Ex ia + Ex d(ia) / 14 to 30 V
IS + XP
B: • Non- 16 to 35 V R [W]
2-wire; hazardous
4-20 mA • Ex nA 500
HART, • Ex nA(ia)
switch • Ex ic
outpu • Ex ic(ia)
• Ex d(ia) / XP
• Ex ta / DIP
• CSA GP
0
• Ex ia / IS 16 to 30 V
U0 [V]
• Ex ia + Ex 10 20 30 35
16 27
d(ia) / IS + XP A0031746
22 Endress+Hauser
Micropilot FMR62
C: All 16 to 30 V R [W]
2-wire;
4-20mA 500
HART,
4-20mA
0
10 20 30 35 U0 [V]
16 27
A0031746
Current consumption
HART
Nominal current 3.6 to 22 mA the start-up current for multidrop mode can be parametrized
(is set to 3.6 mA on delivery)
Endress+Hauser 23
Micropilot FMR62
HART
• A normal device cable suffices if only the analog signal is used.
• A shielded cable is recommended if using the HART protocol. Observe grounding concept of the
plant.
Overvoltage protection If the measuring device is used for level measurement in flammable liquids which requires the use of
overvoltage protection according to DIN EN 60079-14, standard for
test procedures 60060-1 (10 kA, pulse 8/20 μs), overvoltage protection has to be ensured by an
integrated or external overvoltage protection module.
Technical data
1) The material of the gland is dependent on the housing type; GT18 (stainless steel housing): 316L (1.4404); GT19 (plastic housing) and GT20
(aluminum housing): nickel-coated brass (CuZn).
24 Endress+Hauser
Micropilot FMR62
Endress+Hauser 25
Micropilot FMR62
Performance characteristics
Reference operating • Temperature = +24 °C (+75 °F) ±5 °C (±9 °F)
conditions • Pressure = 960 mbar abs. (14 psia) ±100 mbar (±1.45 psi)
• Humidity = 60 % ±15 %
• Reflector: metal plate with diameter ≥ 1 m (40 in)
• No major interference reflections inside the signal beam
Reference accuracy Typical data under reference operating conditions: DIN EN IEC 61298-2 / DIN EN IEC 60770-1;
percentage values in relation to the span.
Accuracy (Sum of non- Measuring distance up to 0.8 m (2.62 ft): max. ±0.03 %
linearity, nonrepeatability ±4 mm (±0.16 in)
and hysteresis) 2)
Measuring distance > 0.8 m (2.62 ft): ±1 mm (±0.04 in) ±0.02 %
3)
Non-repeatability ≤ 1 mm (0.04 in)
4 (0.16)
∆ [mm (in)]
1 (0.04)
0
-1 (-0.04)
-4 (-0.16)
Measured value resolution Dead band according to DIN EN IEC 61298-2 / DIN EN IEC 60770-1:
• Digital: 1 mm
• Analog: 1 µA
Response time The response time can be configured. The following step response times (in accordance with
DIN EN IEC 61298-2 / DIN EN IEC 60770-1) 2) are when damping is switched off:
2) According to DIN EN IEC 61298-2 / DIN EN IEC 60770-1, the step response time is the time that elapses after an abrupt change in the input
signal until the change in the output signal has adopted 90% of the steady-state value for the first time.
26 Endress+Hauser
Micropilot FMR62
Influence of ambient The measurements are performed according to DIN EN IEC 61298-3 / DIN EN IEC 60770-1
temperature • Digital (HART): average TK = 3 mm/10 K
• Analog (current output):
– Zero point (4 mA): average TK = 0.02 %/10 K
– Span (20 mA): average TK = 0.05 %/10 K
Influence of gas phase High pressure reduces the speed of propagation of the measuring signals in the gas/vapor above the
medium. This effect depends on the type of gas phase and its temperature. This results in a
systematic measured error that increases with increasing distance between the reference point of
the measurement (flange) and the surface of the product. The following table shows this measured
error for a few typical gases/vapors (with regard to the distance, a positive value means that a too
large distance is measured):
310 590 - - -
364 687 - - -
With a known, constant pressure, it is possible to compensate for this measured error with a
linearization, for example.
Endress+Hauser 27
Micropilot FMR62
Installation
Installation conditions Mounting position
1 2 3
A0016882
• Recommended distance A wall - nozzle outer edge: ~1/6 of the container diameter. However, the
device must not under any circumstances be mounted closer than 15 cm (5.91 in) to the tank wall.
• Not in the center (2) as interference can cause signal loss.
• Not above the filling curtain (3).
• The use of a weather protection cover (1) is recommended to protect the transmitter from direct
sunlight or rain.
28 Endress+Hauser
Micropilot FMR62
A0031777
Avoid the location of internal fittings (limit switches, temperature sensors, struts, vacuum rings,
heating coils, baffles etc.) inside the signal beam. Take into account the beam angle → 30.
Endress+Hauser 29
Micropilot FMR62
A0031813
Metal deflection plates installed at an angle to scatter the radar signals help prevent interference
echoes.
Optimization options
• Antenna size
The larger the antenna the smaller the beam angle α,resulting in fewer interference echoes
→ 30.
• Mapping
Measurement can be optimized by electronically suppressing interference echoes.
Beam angle
α
W= 2 . D . tan_
2
D
W
A0031824
The beam angle is defined as the angle α where the energy density of the radar waves reaches half
the value of the maximum energy density (3dB width). Microwaves are also emitted outside the
signal beam and can be reflected off interfering installations.
30 Endress+Hauser
Micropilot FMR62
FMR62
A0032081 A0032082
1)
Antenna integrated integrated, PTFE cladded PTFE cladded
, PEEK, 40 mm / 1-1/2" flush mount flush mount 80 mm / 3"
PEEK, 20 50 mm / 2"
mm / 3/4"
Beam angle α 14 ° 8° 7° 3°
5 m (16 ft) 1.32 m 0.70 m (2.29 ft) 0.61 m 0.26 m (0.85 ft)
(4.33 ft) (2.00 ft)
10 m (33 ft) 2.63 m 1.40 m (4.58 ft) 1.22 m 0.52 m (1.71 ft)
(8.63 ft) (4.00 ft)
Endress+Hauser 31
Micropilot FMR62
A0034564
• Measurements can be performed through an open full bore ball valve without any problems.
• At the transitions, no gap may be left exceeding 1 mm (0.04 in).
• Diameter of opening of ball valve must always be equivalent to pipe diameter; avoid edges and
constrictions.
1) or an integer multiple of this value; however, take into account that the microwave transparency is
significantly reduced when increasing the thickness of the window.
øD
A0032208
32 Endress+Hauser
Micropilot FMR62
Note the following if the antenna does not project out of the nozzle:
• The end of the nozzle must be smooth and free from burrs. The edge of the nozzle should be
rounded if possible.
• Mapping must be performed.
• Please contact Endress+Hauser for applications with nozzles that are higher than indicated in
the table.
øD
A0032206
Endress+Hauser 33
Micropilot FMR62
Note the following if the antenna does not project out of the nozzle:
• The end of the nozzle must be smooth and free from burrs. The edge of the nozzle should be
rounded if possible.
• Mapping must be performed.
• Please contact Endress+Hauser for applications with nozzles that are higher than indicated in
the table.
Minimum Maximum
EN
DN50/PN16 4 45 65
DN80/PN16 8 40 55
DN100/PN16 8 40 60
DN150/PN16 8 75 115
ASME
2"/150lbs 4 40 55
3"/150lbs 4 65 95
4"/150lbs 8 45 70
6"/150lbs 8 85 125
JIS
10K 50A 4 40 60
10K 80A 8 25 35
10K 100A 8 35 55
34 Endress+Hauser
Micropilot FMR62
A0032207
If process temperatures are high, the device should be included in the usual container insulation
system (2) to prevent the electronics from heating as a result of thermal radiation or convection. The
insulation should not be higher than the neck of the device (1).
Endress+Hauser 35
Micropilot FMR62
Environment
Ambient temperature range Measuring device –40 to +80 °C (–40 to +176 °F)
Local display –20 to +70 °C (–4 to +158 °F), the readability of the display may be impaired at
temperatures outside the temperature range.
Ambient temperature limits The following diagrams only consider functional aspects. Additional restrictions may apply for
certified device versions. See the separate Safety Instructions for more information → 81.
In the event of temperature (Tp) at the process connection, the permitted ambient temperature (Ta)
is reduced as indicated in the following diagram (temperature derating) in the table header.
FMR62
Antenna 1)
• GE:
Integrated, PEEK, 3/4"
• GF:
Integrated, PEEK, 1-1/2" Ta
• GM: P1 P2
PTFE cladded flush mount DN50
• GN:
PTFE cladded flush mount DN80 P3
Seal 2) Tp
• A6: P5 P4
FKM Viton GLT, –40 to 200 °C (–40 to 392 °F)
A0032024
• C2:
FFKM Kalrez, –20 to 200 °C (–4 to 392 °F)
• F6:
PTFE cladded, –40 to 200 °C (–40 to 392 °F)
Temperature specifications: °C (°F)
Housing type 3) P1 P2 P3 P4 P5
Tp Ta Tp Ta Tp Ta Tp Ta Tp Ta
36 Endress+Hauser
Micropilot FMR62
FMR62
Antenna 1)
• GE:
Integrated, PEEK, 3/4"
• GF:
Integrated, PEEK, 1-1/2"
• GM:
Ta
PTFE cladded flush mount DN50
• GN: P1 P2 P2+
PTFE cladded flush mount DN80
P3
Seal 2)
• A5:
FKM Viton GLT,
Tp
P5 P4
–40 to 150 °C (–40 to 302 °F)
• C1: A0032025
FFKM Kalrez,
–40 to 150 °C (–40 to 302 °F)
• F5:
PTFE cladded,
–40 to 150 °C (–40 to 302 °F)
Temperature specifications: °C (°F)
Tp Ta Tp Ta Tp Ta Tp Ta Tp Ta Tp Ta
3) This restriction is valid if the following options of the product structure have been selected at the same time: 030 ("Display, Operation") = C
("SD02") or E ("SD03"); 040 ("Housing") = A ("GT19").
Endress+Hauser 37
Micropilot FMR62
Vibration resistance DIN EN 60068-2-64 IEC 60068-2-64 at 5 to 2 000 Hz: 1.5 (m/s2)2/Hz
A limited vibration resistance applies according
DIN EN 60068-2-64/IEC 60068-2-64 at 5 to 2 000 Hz of 0.39 (m/s2)2/Hz,
if the following order features are selected in the product structure at the same time:
• 040 ("Housing"): B ("GT18 two-chamber, 316L)
and
• 090 ("Seal"):
– A6 ("FKM Viton GLT, –40 to 200 °C (–40 to 392 °F), incl. gas-tight feedthrough")
or
– C2 ("FKM Viton GLT, –40 to 200 °C (–40 to 392 °F), incl. gas-tight feedthrough")
or
– F6 ("FKM Viton GLT, –40 to 200 °C (–40 to 392 °F), incl. gas-tight feedthrough")
Electromagnetic Electromagnetic compatibility in accordance with all of the relevant requirements outlined in the
compatibility (EMC) EN 61326 series and NAMUR Recommendation EMC (NE 21). For details, please refer to the
Declaration of Conformity 4).
A normal device cable suffices if only the analog signal should be used. Use a shielded cable for
digital communication (HART/ PA/ FF).
Maximum measured error during EMC testing: <0.5 % of the span. By way of derogation, for devices
with a plastic housing and see-through lid (integrated display SD02 or SD03) the measured error can
be up to 2 % of the span in the event of strong electromagnetic radiation in the 1 to 2 GHz frequency
range.
38 Endress+Hauser
Micropilot FMR62
Process
Process temperature, process
The pressure ranges indicated can be reduced by the choice of process connection. The nominal
pressure
pressure (PN) indicated on the nameplate refers to a reference temperature of 20 °C, and of
100 °F for ASME flanges. Observe pressure-temperature dependency.
Please refer to the following standards for the pressure values permitted at higher
temperatures:
• EN 1092-1: 2001 Tab. 18
In terms of their stability-temperature property, the materials 1.4435 and 1.4404 are
grouped in EN 1092-1 table 18 under 13E0. The chemical composition of the two materials
can be identical.
• ASME B 16.5a – 1998 Tab. 2-2.2 F316
• ASME B 16.5a – 1998 Tab. 2.3.8 N10276
• JIS B 2220
[bar] ([psi]) p
1 2
20 (290)
0 (0)
-1 (-14.5) Tp
-40 -20 0 +150 [°C]
(-40) (-4) (+32) (+302) ([°F])
A0032182
10 FMR62: Permitted range for process temperature and process pressure, integrated antenna, PEEK (feature
070: GE, GF), 150 °C (302 °F)
1 Feature 90, seal: A5, FKM Viton GLT
2 Feature 90, seal: C1, FFKM Kalrez
1) The pressure range may be further restricted in the event of a CRN approval
Endress+Hauser 39
Micropilot FMR62
[bar] ([psi]) p
1 2
20 (290)
0 (0)
-1 (-14.5) Tp
-40 -20 0 +200 [°C]
(-40) (-4) (+32) (+392) ([°F])
A0032183
11 FMR62: Permitted range for process temperature and process pressure, integrated antenna, PEEK (feature
070: GE, GF), 200 °C (392 °F)
1 Feature 90, seal: A6, FKM Viton GLT
2 Feature 90, seal: C2, FFKM Kalrez
1) The pressure range may be further restricted in the event of a CRN approval
[bar] ([psi]) p
1 2
16 (232)
0 (0)
-1 (-14.5) Tp
-40 0 +150 +200 [°C]
(-40) (+32) (+302) (+392) ([°F])
A0032618
12 FMR62: Permitted range for process temperature and process pressure, Tri-Clamp 2" process connection
1 Feature 90, seal: F5
2 Feature 90, seal: F6
40 Endress+Hauser
Micropilot FMR62
1) The pressure range may be further restricted in the event of a CRN approval
[bar] ([psi]) p
1 2
14 (203)
0 (0)
-1 (-14.5) Tp
-40 0 +150 +200 [°C]
(-40) (+32) (+302) (+392) ([°F])
A0032619
13 FMR62: Permitted range for process temperature and process pressure, Tri-Clamp 3" or Tri-Clamp 4"
process connection
1 Feature 90, seal: F5
2 Feature 90, seal: F6
1) The pressure range may be further restricted in the event of a CRN approval
Endress+Hauser 41
Micropilot FMR62
[bar] ([psi]) p
1 2
25 (362.6)
0 (0)
-1 (-14.5) Tp
-40 0 +150 +200 [°C]
(-40) (+32) (+302) (+392) ([°F])
A0032184
14 FMR62: Permitted range for process temperature and process pressure, flange or DIN11851 process
connection
1 Feature 90, seal: F5
2 Feature 90, seal: F6
• AFK: F5:
–40 to +150 °C
NPS 2" Cl.150 RF, PTFE > 316/316L PTFE
(–40 to +302 °F)
• AGK: cladded
NPS 3" Cl.150 RF, PTFE > 316/316L
• AHK:
NPS 4" Cl.150 RF, PTFE > 316/316L
• AJK:
NPS 6" Cl.150 RF, PTFE > 316/316L
• CFK:
DN50 PN10/16 B1, PTFE > 316L
• CGK:
DN80 PN10/16 B1, PTFE > 316L
• CHK: prel =
DN100 PN10/16 B1, PTFE > 316L –1 to 25 bar
• CJK: F6: (–14.5 to
DN150 PN10/16 B1, PTFE > 316L –40 to +200 °C 362.6 psi) 1)
PTFE
• KFK: (–40 to +392 °F)
cladded
10K 50A RF, PTFE > 316L
• KGK:
10K 80A RF, PTFE > 316L
• KHK:
10K 100A RF, PTFE > 316L
• KJK:
10K 150A RF, PTFE > 316L
• MRK:
DIN11851 DN50 PN25 slotted nut, PTFE > 316L
• MTK:
DIN11851 DN80 PN25 slotted nut, PTFE > 316L
1) The pressure range may be further restricted in the event of a CRN approval
42 Endress+Hauser
Micropilot FMR62
Mechanical construction
Dimensions Dimensions of the electronics housing
R110 98 (3.86)*
78 (3.07) 90 (3.54)
07)
(4.
3.5
146 (5.75)
ø108.5 (4.27)
142 (5.59)
115.25 (4.54)
ø10
A0011666
R110
99 (3.9)*
78 (3.07) 90 (3.54)
)
.17
6 (4
144 (5.67)
ø106 (4.17)
ø10
113.5 (4.47)
A0011346
R110 98 (3.86)*
78 (3.07) 90 (3.54)
07)
(4.
ø112 (4.41)
145.5 (5.73)
3.5
141.5 (5.57)
ø108.5 (4.27)
ø10
61 (2.4)
A0020751
Endress+Hauser 43
Micropilot FMR62
1
A B
186 (7.32)
143 (5.63)
22 (0.87)
R
ø32.2 (1.27)
ø23.8 (0.94)
A0031557
18 Dimensions: mm (in)
A Seal: FKM Viton GLT, -40 to 150°C/-40 to 302°F or FFKM Kalrez, -20...150oC/-4...302°F
B Seal: FKM Viton GLT, -40 to 200°C/-40 to 392°F or FFKM Kalrez, -20...200oC/-4...392°F
R Reference point of the measurement
1 Bottom edge of housing
44 Endress+Hauser
Micropilot FMR62
1
A B
227 (8.94)
184 (7.24)
30 (1.18)
ø54.8 (2.18)
ø42.5 (1.67)
A0032787
19 Dimensions: mm (in)
A Seal: FKM Viton GLT, -40 to 150°C/-40 to 302°F or FFKM Kalrez, -20...150oC/-4...302°F
B Seal: FKM Viton GLT, -40 to 200°C/-40 to 392°F or FFKM Kalrez, -20...200oC/-4...392°F
R Reference point of the measurement
1 Bottom edge of housing
Endress+Hauser 45
Micropilot FMR62
FMR62: flush mount antenna with sanitary adapter DIN11851, PTFE cladded
100: Process Connection • MRK: DIN11851 DN50 PN25 slotted nut, PTFE>316L
• MTK: DIN11851 DN80 PN25 slotted nut, PTFE>316L
A B 1
L2
L1
L3
L4
R
øD1
øD2
A0032175
20 Dimensions: mm (in)
A Seal: PTFE cladded, -40 to 150°C/-40 to 302°F
B Seal: PTFE cladded, -40 to 200°C/-40 to 392°F
R Reference point of the measurement
1 Bottom edge of housing
MRK:
48 mm 65.5 mm 148 mm 191 mm 5 mm 19 mm
DN50 PN25 slotted nut,
(1.89 in) (2.58 in) (5.83 in) (7.52 in) (0.20 in) (0.75 in)
PTFE>316L
MTK:
75 mm 97 mm 161 mm 204 mm 6 mm 32 mm
DN80 PN25 slotted nut,
(2.95 in) (3.82 in) (6.33 in) (8.03 in) (0.24 in) (1.26 in)
PTFE>316L
46 Endress+Hauser
Micropilot FMR62
A B 1
L2
L1
2.6 (0.10)
1.6 (0.06) R
L4
øD1
øD2
øD3
A0032176
21 Dimensions: mm (in)
A Seal: PTFE cladded, -40 to 150°C/-40 to 302°F
B Seal: PTFE cladded, -40 to 200°C/-40 to 392°F
R Reference point of the measurement
1 Bottom edge of housing
GM:
48 mm (1.89 in) 145 mm (5.71 in) 188 mm (7.40 in) 19 mm (0.75 in)
PTFE cladded flush mount DN50
GN:
75 mm (2.95 in) 158 mm (6.22 in) 201 mm (7.91 in) 32 mm (1.26 in)
PTFE cladded flush mount DN80
TDK:
56.5 mm (2.22 in) 64 mm (2.52 in)
DN40-51 (2")
TFK:
83.5 mm (3.29 in) 91 mm (3.58 in)
DN70-76.1 (3")
THK:
110 mm (4.33 in) 119 mm (4.69 in)
DN101.6 (4")
Endress+Hauser 47
Micropilot FMR62
FMR62: flush mount antenna with flange EN1092-1, ASME B16.5 or JIS B2220; PTFE cladded
100: Process Connection • AFK: NPS 2" Cl.150, PTFE>316/316L flange ASME B16.5
• AGK: NPS 3" Cl.150, PTFE>316/316L flange ASME B16.5
• AHK: NPS 4" Cl.150, PTFE>316/316L flange ASME B16.5
• AJK: NPS 6" Cl.150, PTFE>316/316L flange ASME B16.5
• CFK: DN50 PN10/16, PTFE>316L flange EN1092-1
• CGK: DN80 PN10/16, PTFE>316L flange EN1092-1
• CHK: DN100 PN10/16, PTFE>316L flange EN1092-1
• CJK: DN150 PN10/16, PTFE>316L flange EN1092-1
• KFK: 10K 50A, PTFE>316L flange JIS B2220
• KGK: 10K 80A, PTFE>316L flange JIS B2220
• KHK: 10K 100A, PTFE>316L Flansch JIS B2220
• KJK: 10K 150A, PTFE>316L flange JIS B2220
1
A B
L2
L1
4 (0.16)
L3
R
L4
øD1
øD2
øD3
A0032177
22 Dimensions: mm (in)
A Seal: PTFE cladded, -40 to 150°C/-40 to 302°F
B Seal: PTFE cladded, -40 to 200°C/-40 to 392°F
R Reference point of the measurement
1 Bottom edge of housing
GM:
48 mm (1.89 in) 147 mm (5.79 in) 186 mm (7.32 in) 19 mm (0.75 in)
PTFE cladded flush mount DN50
GN:
75 mm (2.95 in) 159 mm (6.26 in) 202 mm (7.95 in) 32 mm (1.26 in)
PTFE cladded flush mount DN80
48 Endress+Hauser
Micropilot FMR62
AFK:
92 mm (3.62 in) 150 mm (5.91 in) 17.5 mm (0.69 in)
NPS 2" Cl.150, PTFE>316/316L
AGK:
127 mm (5.00 in) 190 mm (7.48 in) 22.3 mm (0.88 in)
NPS 3" Cl.150, PTFE>316/316L
AHK:
158 mm (6.22 in) 230 mm (9.06 in) 22.3 mm (0.88 in)
NPS 4" Cl.150, PTFE>316/316L
AJK:
212 mm (8.35 in) 280 mm (11.02 in) 23.9 mm (0.94 in)
NPS 6" Cl.150, PTFE>316/316L
CFK:
102 mm (4.02 in) 165 mm (6.50 in) 18 mm (0.71 in)
DN50 PN10/16, PTFE>316L
CGK:
138 mm (5.43 in) 200 mm (7.87 in) 20 mm (0.79 in)
DN80 PN10/16, PTFE>316L
CHK:
158 mm (6.22 in) 220 mm (8.66 in) 20 mm (0.79 in)
DN100 PN10/16, PTFE>316L
CJK:
212 mm (8.35 in) 285 mm (11.22 in) 22 mm (0.87 in)
DN150 PN10/16, PTFE>316L
KFK:
96 mm (3.78 in) 155 mm (6.10 in) 16 mm (0.63 in)
10K 50A, PTFE>316L
KGK:
127 mm (5.00 in) 185 mm (7.28 in) 18 mm (0.71 in)
10K 80A, PTFE>316L
KHK:
151 mm (5.94 in) 210 mm (8.27 in) 18 mm (0.71 in)
10K 100A, PTFE>316L
KJK:
212 mm (8.35 in) 280 mm (11.02 in) 22 mm (0.87 in)
10K 150A, PTFE>316L
Endress+Hauser 49
Micropilot FMR62
Weight Housing
Part Weight
50 Endress+Hauser
Micropilot FMR62
5.2
6
5.1
7
4 3 9 8
A0013788
5.1 Dummy plug, cable gland, adapter or plug • Dummy plug, depending on the device version:
(depending on the device version) – PE
– PBT-GF
• Cable gland: 316L (1.4404) or nickel-plated brass
• Adapter: 316L (1.4404/1.4435)
• Seal: EPDM
• M12 plug: Nickel-plated brass 1)
• 7/8" plug: 316 (1.4401) 2)
5.2 Dummy plug, cable gland or adapter (depending on • Dummy plug: 316L (1.4404)
the device version) • Cable gland: 316L (1.4404) or nickel-plated brass
• Adapter: 316L (1.4404/1.4435)
• Seal: EPDM
6 Dummy plug or M12 socket (depending on the • Dummy plug: 316L (1.4404)
device version) • M12 socket: 316L (1.4404)
1) For the version with M12 plug the sealing material is Viton.
2) For the version with 7/8" plug, the sealing material is NBR.
Endress+Hauser 51
Micropilot FMR62
5.2
6
5.1
7
4 9 8
A0013788
1 Housing PBT
5.1 Dummy plug, cable gland, adapter or plug • Dummy plug, depending on the device version:
(depending on the device version) – PE
– PBT-GF
• Cable gland, depending on the device version:
– Nickel-plated brass (CuZn)
– PA
• Adapter: 316L (1.4404/1.4435)
• Seal: EPDM
• M12 plug: Nickel-plated brass 1)
• 7/8" plug: 316 (1.4401) 2)
5.2 Dummy plug, cable gland or adapter (depending on • Dummy plug, depending on the device version:
the device version) – PE
– PBT-GF
– Nickel-plated steel
• Cable gland, depending on the device version:
– Nickel-plated brass (CuZn)
– PA
• Adapter: 316L (1.4404/1.4435)
• Seal: EPDM
6 Dummy plug or M12 socket (depending on the • Dummy plug: Nickel-plated brass (CuZn)
device version) • M12 socket: Nickel-plated GD-Zn
1) For the version with M12 plug the sealing material is Viton.
2) For the version with 7/8" plug, the sealing material is NBR.
52 Endress+Hauser
Micropilot FMR62
5.2
6
5.1
7
4 3 9 8
A0013788
2.1 Cover of the electronics compartment; RAL 7035 • Cover: AlSi10Mg(<0,1% Cu)
(gray) • Window: Glass
• Cover seal: NBR
• Seal of the window: NBR
• Thread-coating: Graphite-based lubricant varnish
2.2 Cover of the terminal compartment; RAL 7035 • Cover: AlSi10Mg(<0,1% Cu)
(gray) • Cover seal: NBR
• Thread-coating: Graphite-based lubricant varnish
5.1 Dummy plug, cable gland, adapter or plug • Dummy plug, depending on the device version:
(depending on the device version) – PE
– PBT-GF
• Cable gland, depending on the device version:
– Nickel-plated brass (CuZn)
– PA
• Adapter: 316L (1.4404/1.4435)
• Seal: EPDM
• M12 plug: Nickel-plated brass 1)
• 7/8" plug: 316 (1.4401) 2)
5.2 Dummy plug, cable gland or adapter (depending on • Dummy plug, depending on the device version:
the device version) – PE
– PBT-GF
– Nickel-plated steel
• Cable gland, depending on the device version:
– Nickel-plated brass (CuZn)
– PA
• Adapter: 316L (1.4404/1.4435)
• Seal: EPDM
6 Dummy plug or M12 socket (depending on the • Dummy plug : Nickel-plated brass (CuZn)
device version) • M12 socket: Nickel-plated GD-Zn
Endress+Hauser 53
Micropilot FMR62
1) For the version with M12 plug the sealing material is Viton.
2) For the version with 7/8" plug, the sealing material is NBR.
5
1
A0032019
4 4 4
3 3 3
2 2 2
5 5 5
6
1 1 1
A0031815
1 Antenna PTFE
Antenna seal: Viton (FKM)
54 Endress+Hauser
Micropilot FMR62
Endress+Hauser 55
Micropilot FMR62
Materials: Weather
protection cover 1
2
3
7 4
5
6
A0015473
No Part: Material
6 Ground terminal
• Screw: A4
• Spring washer: A4
• Clamp: 316L (1.4404)
• Holder: 316L (1.4404)
7 • Washer: A4
• Cheese head screw: A4-70
56 Endress+Hauser
Micropilot FMR62
Operability
Operating concept Operator-oriented menu structure for user-specific tasks
• Commissioning
• Operation
• Diagnostics
• Expert level
Operating languages
• English
• Deutsch
• Français
• Español
• Italiano
• Nederlands
• Portuguesa
• Polski
• русский язык (Russian)
• Svenska
• Türkçe
• 中文 (Chinese)
• 日本語 (Japanese)
• 한국어 (Korean)
• Bahasa Indonesia
• tiếng Việt (Vietnamese)
• čeština (Czech)
Feature 500 of the product structure determines which of these languages is preset on delivery.
Endress+Hauser 57
Micropilot FMR62
A0032219 A0032221
Format for displaying measured variables and status variables can be individually configured
Permitted ambient temperature for the display: –20 to +70 °C (–4 to +158 °F)
The readability of the display may be impaired at temperatures outside the temperature
range.
Operating local operation with 3 push buttons (, , external operation via touch control; 3 optical
elements ) keys: , ,
A0032215
58 Endress+Hauser
Micropilot FMR62
2 3 6 8
1 4 5 7
A0028746
1
A0032466
1 Service interface (CDI) of the measuring device (= Endress+Hauser Common Data Interface)
2 Commubox FXA291
3 Computer with DeviceCare/FieldCare operating tool
Endress+Hauser 59
Micropilot FMR62
SupplyCare inventory SupplyCare is a web-based operating program for coordinating the flow of material and information
management software along the supply chain. SupplyCare provides a comprehensive overview of the levels of
geographically distributed tanks and silos, for instance, providing complete transparency over the
current inventory situation, regardless of time and location.
Based on the measuring and transmission technology installed onsite, the current inventory data are
collected and sent to SupplyCare. Critical levels are clearly indicated and calculated forecasts provide
additional security for material requirements planning.
The main functions of SupplyCare:
Inventory visualization
SupplyCare determines the inventory levels in tanks and silos at regular intervals. It displays current
and historical inventory data and calculated forecasts of future demand. The overview page can be
configured to suit the user's preferences.
Master data management
With SupplyCare you can create and manage the master data for locations, companies, tanks,
products and users, as well as user authorization.
Report Configurator
The Report Configurator can be used to create personalized reports quickly and easily. The reports
can be saved in a variety of formats, such as Excel, PDF, CSV and XML. The reports can be
transmitted in many ways, such as by http, ftp or e-mail.
Event management
Events, such as when levels drop below the safety stock level or plan points, are indicated by the
software. In addition, SupplyCare can also notify pre-defined users by e-mail.
Alarms
If technical problems occur, e.g. connection issues, alarms are triggered and alarm e-mails are sent to
the System Administrator and the Local System Administrator.
Delivery planning
The integrated delivery planning function automatically generates an order proposal if a pre-set
minimum inventory level is undershot. Scheduled deliveries and disposals are monitored
continuously by SupplyCare. SupplyCare notifies the user if scheduled deliveries and disposals are not
going to be met as planned.
Analysis
In the Analysis module, the most important indicators for the inflow and outflow of the individual
tanks are calculated and displayed as data and charts. Key indicators of material management are
automatically calculated and form the basis for optimizing the delivery and storage process.
Geographical visualization
All the tanks and the tank inventories are represented graphically on a map (based on Google Maps).
The tanks and inventory situations can be filtered by tank group, product, supplier or location.
Multi-language support
The multi-language user interface supports 9 languages, thereby enabling global collaboration on a
single platform. The language and settings are recognized automatically using the browser settings.
SupplyCare Enterprise
SupplyCare Enterprise runs by default as a service under Microsoft Windows on an application server
in an Apache Tomcat environment. The operators and administrators operate the application via a
Web browser from their workstations.
60 Endress+Hauser
Micropilot FMR62
2
3
10
5 7 8
9
.
A0034288
Endress+Hauser 61
Micropilot FMR62
1
3
WWW
4
10
5 7 8
9
.
A0034289
With SupplyCare Hosting, users do not need to make the initial software purchase or install and run
the IT infrastructure needed. Endress+Hauser constantly update SupplyCare Hosting and enhance
the capability of the software in conjunction with the customer. The hosted version of SupplyCare is
thus always up-to-date and can be customized to meet different customer requirements. Other
services are also offered in addition to the IT infrastructure and the software that is installed in a
secure, redundant Endress+Hauser data center. These services include defined availability of the
global Endress+Hauser Service and Support Organization and defined response times in a service
event.
62 Endress+Hauser
Micropilot FMR62
RoHS The measuring system complies with the substance restrictions of the Restriction on Hazardous
Substances Directive 2011/65/EU (RoHS 2).
RCM-Tick marking The supplied product or measuring system meets the ACMA (Australian Communications and Media
Authority) requirements for network integrity, interoperability, performance characteristics as well
as health and safety regulations. Here, especially the regulatory arrangements for electromagnetic
compatibility are met. The products are labelled with the RCM- Tick marking on the name plate.
A0029561
Ex approval • ATEX
• IECEx
• CSA
• FM
• NEPSI
• KC
• INMETRO
• TIIS 5)
• EAC
Additional safety instructions must be followed for use in hazardous areas. Please refer to the
separate "Safety Instructions" (XA) document included in the delivery. Reference to the applicable XA
can be found on the nameplate.
For details on the available certificates and associated XAs: → 81
Dual seal according to The devices have been designed according to ANSI/ISA 12.27.01 as dual seal devices, allowing the
ANSI/ISA 12.27.01 user to waive the use and save the cost of installing external secondary process seals in the conduit
as required by the process sealing sections of ANSI/NFPA 70 (NEC) and CSA 22.1 (CEC) These
instruments comply with the North-American installation practice and provide a very safe and cost-
saving installation for pressurized applications with hazardous fluids.
Further information can be found in the Safety Instructions (XA) of the relevant devices.
Functional safety Use for level monitoring (MIN, MAX, range) up to SIL 3 (homogeneous or diverse redundancy),
independently evaluated by TÜV Rheinland in accordance with IEC 61508, refer to the "Functional
Safety Manual" for information.
Hygienic compatibility FMR62 with flange cladding made of PTFE complies with FDA 21 CFR 177.1550 and
USP <88> Class VI.
5) in preparation
Endress+Hauser 63
Micropilot FMR62
3A and EHEDG approval for versions with Tri-Clamp and DIN11851 process connections. 6)
To avoid any risk of contamination, install the device according to the “Hygienic Equipment
Design Criteria (HDC)” which was published by the Subgroup "Design Principles" of the EHEDG
on April 8 2004.
Suitable fittings and seals must be used to ensure hygiene-compliant design according to the
specifications of 3-A SSI and EHEDG.
The gap-free connections can be cleaned of all residue using the usual cleaning methods within
this industry.
The 316L housing (GT18), option B in order feature 40, is not suitable for use in applications
which require a 3A approval.
NACE MR 0175 / ISO 15156 • The wetted, metal materials comply with the requirements of NACE MR 0175 / ISO 15156.
• Declaration of Conformity: see the product structure, feature 580, version JB
NACE MR 0103 • The wetted, metal materials comply with the requirements of NACE MR 0103.
• The Declaration of Conformity is based on NACE MR 0175.
The hardness and intergranular corrosion have been tested, and heat treatment (solution
annealed) has been performed. The materials used therefore meet the requirements of
NACE MR 0103.
• Declaration of Conformity: see the product structure, feature 580, version JE.
Pressure equipment with Pressure instruments with a flange and threaded boss that do not have a pressurized housing do not
allowable pressure fall within the scope of the Pressure Equipment Directive, irrespective of the maximum allowable
≤ 200 bar (2 900 psi) pressure.
Reasons:
According to Article 2, point 5 of EU Directive 2014/68/EU, pressure accessories are defined as
"devices with an operational function and having pressure-bearing housings".
If a pressure instrument does not have a pressure-bearing housing (no identifiable pressure chamber
of its own), there is no pressure accessory present within the meaning of the Directive.
FMR62
Radio standard The devices FMR62 with antenna 40 mm / 2" integrated PEEK, comply with the Level Probing Radar
EN 302729-1/2 (LPR) radio standard EN 302729-1/2. The devices FMR62 with antenna 40 mm / 2" integrated
PEEK, are approved for unrestricted use inside and outside closed containers in countries of the EU
and the EFTA . that have already implemented this standard.
The following countries are those that have currently implemented the directive:
Belgium, Bulgaria, Germany, Denmark, Estonia, France, Greece, UK, Ireland, Iceland, Italy,
Liechtenstein, Lithuania, Latvia, Malta, The Netherlands, Norway, Austria, Poland, Portugal,
Romania, Sweden, Switzerland, Slovakia, Spain, Czech Republic and Cyprus.
Implementation is still underway in all of the countries not listed.
Please note the following for operation of the devices outside of closed vessels:
1. Installation must be carried out by properly trained, expert staff.
2. The device antenna must be installed in a fixed location pointing vertically downwards.
6) EHEDG pending
64 Endress+Hauser
Micropilot FMR62
3. The installation site must be located at a distance of 4 km from the astronomy stations listed
below or otherwise approval must be provided by the relevant authority. If the device is
installed at a distance of 4 to 40 km from one of the listed stations, it must not be installed at a
height of more than 15 m (49 ft) above the ground.
Astronomy stations
Country Name of the station Latitude Longitude
Radio standard The devices comply with the Tanks Level Probing Radar (TLPR) radio standard EN 302372-1/2 and
EN 302372-1/2 are approved for use in closed containers. For installation, points a to f in Annex B of EN 302372-1
must be taken into consideration.
FCC This device complies with Part 15 of the FCC rules. Operation is subject to the following two
conditions: (1) This device may not cause harmful interference, and (2) this device must accept any
interference received, including interference that may cause undesired operation.
[Any] changes or modifications not expressly approved by the party responsible for compliance could
void the user's authority to operate the equipment.
The devices are compliant with the FCC Code of Federal Regulations, CFR 47, Part 15, Sections
15.205, 15.207, 15.209.
In addition, the devices are compliant with Section 15.256 . For these LPR (Level Probe Radar)
applications the devices must be professionally installed in a downward operating position. In
addition, the devices are not allowed to be mounted in a zone of 4 km around RAS stations and
within a radius of 40 km around RAS stations the maxium operation height of devices is 15 m (49 ft)
above ground.
Endress+Hauser 65
Micropilot FMR62
CRN approval Some device versions have CRN approval. Devices are CRN approved if the following two conditions
are met:
• The device has a CSA approval (product structure: feature 010 "Approval")
• The device has a CRN-approved process connection as per the following table:
1) The CRN certificate applies in a limited pressure range; see the following table.
• Process connections that do not have CRN approval are not listed in this table.
• Please refer to the product structure to discover which process connections are available for a
specific device type.
• CRN-approved devices bear the registration number (tbd.) on the nameplate.
An additional pressure restriction applies for the device versions listed in the table below if they
are CRN-approved. The pressure range indicated in the "Process" section → 39 continues to
apply for device versions not listed in the table below (in preparation).
66 Endress+Hauser
Micropilot FMR62
Test reports, declarations and inspection certificates are available in electronic format in the
W@M Device Viewer:
Enter the serial number from nameplate (www.endress.com/deviceviewer)
This concerns the options for the following order codes:
• 550 "Calibration"
• 580 "Test, certificate"
• 590 "Additional approval", option LW: "CoC-ASME BPE"
Hard-copy product Hard-copy versions of the test reports, declarations and inspection certificates can also be order via
documentation order code 570 "Service", option I7 "Hard-copy product documentation". The documents are then
supplied with the product.
Endress+Hauser 67
Micropilot FMR62
68 Endress+Hauser
Micropilot FMR62
Ordering information
Ordering information Detailed ordering information is available from the following sources:
• In the Product Configurator on the Endress+Hauser website: www.endress.com -> Click "Corporate"
-> Select your country -> Click "Products" -> Select the product using the filters and search field ->
Open product page -> The "Configure" button to the right of the product image opens the Product
Configurator.
• From your Endress+Hauser Sales Center: www.addresses.endress.com
Product Configurator - the tool for individual product configuration
• Up-to-the-minute configuration data
• Depending on the device: Direct input of measuring point-specific information such as
measuring range or operating language
• Automatic verification of exclusion criteria
• Automatic creation of the order code and its breakdown in PDF or Excel output format
• Ability to order directly in the Endress+Hauser Online Shop
Endress+Hauser 69
Micropilot FMR62
3-point linearity protocol The following points must be considered if option F3 (3-point linearity protocol) was selected in
feature 550 ("Calibration").
The 3 points of the linearity protocol are defined as follows:
A
1
6000 (236)
500 (19.7)
2
A0032285
2nd measuring point In the middle between the 1st and 3rd measuring point
3rd measuring point 6 000 mm (236 in) below the reference point R
70 Endress+Hauser
Micropilot FMR62
5-point linearity protocol The following points must be considered if option F4 (5-point linearity protocol) was selected in
feature 550 ("Calibration").
The 5 points of the linearity protocol are evenly distributed over the measuring range (0% - 100%).
Empty calibration (E) and Full calibration (F) must be specified in order to define the measuring
range 7).
The following restrictions must be considered when selecting E and F:
A
E
F
A0032286
Minimum distance between reference point R and Minimum span Maximum value for "Empty
100% mark calibration"
A ≥ length of antenna + 200 mm (8 in) F ≥400 mm (16 in) E ≤24 m (79 ft)
Minimum value: 400 mm (16 in)
The selected values for Empty calibration and Full calibration are only used to create the
linearity protocol. Afterwards, the values are reset to the default values specific for the
antenna. If values other than the default values are required, they must be ordered as a
customized parameterization → 72.
7) If (E) and (F) are not specified, antenna-dependent default values will be used instead.
Endress+Hauser 71
Micropilot FMR62
Customer-specific If the option IJ: "Customized parameterization HART", IK "Customized parameterization PA" or IL
configuration "Customized parameterization FF" has been selected in feature 570 "Service", presettings that differ
from the default settings can be selected for the following parameters:
Designation in the Electronic The first 32 characters of the measuring point designation
Name Plate (ENP)
Designation on the display The first 12 characters of the measuring point designation
module
Services The following services can be selected via the product structure in the Product Configurator 8):
• PWIS-free (PWIS = paint-wetting impairment substances)
• Customized parameterization HART → 72
• Customized parameterization PA → 72
• Customized parameterization FF → 72
• W/o tooling DVD (FieldCare)
• Hard-copy product documentation
72 Endress+Hauser
Micropilot FMR62
Application Packages
Heartbeat Diagnostics Availability
Available in all device versions.
Function
• Continuous self-monitoring of the device.
• Diagnostic messages output to
– the local display.
– an asset management system (e.g. FieldCare/DeviceCare).
– an automation system (e.g. PLC).
Advantages
• Device condition information is available immediately and processed in time.
• The status signals are classified in accordance with VDI/VDE 2650 and NAMUR recommendation
NE 107 and contain information about the cause of the error and remedial action.
Detailed description
See Operating Instructions of the device (→ 81); chapter "Diagnostics and trouble shooting"
Endress+Hauser 73
Micropilot FMR62
Advantages
• No onsite presence is required to use the function.
• The DTM 9) triggers verification in the device and interprets the results. No specific knowledge is
required on the part of the user.
• The verification report can be used to prove quality measures to a third party.
• Heartbeat Verification can replace other maintenance tasks (e.g. periodic check) or extend the
test intervals.
• The Heartbeat Verification module contains a wizard for the proof test which must be performed
at appropriate intervals for the following applications:
– SIL (IEC61508/IEC61511)
– WHG (German Water Resources Act)
• To perform a proof test, the device must be locked (SIL/WHG locking).
• The wizard can be used via FieldCare, DeviceCare or a DTM-based process control system.
In the case of SIL-locked and WHG-locked devices, it is not possible to perform verification
without additional measures (e.g. by-passing of the output current) because the output current
must be simulated (Increased safety mode) or the level must be approached manually (Expert
mode) during subsequent re-locking (SIL/WHG locking).
Detailed description
SD01870F
9) DTM: Device Type Manager; controls device operation via DeviceCare, FieldCare or a DTM-based process control system.
10) Only relevant for devices with SIL or WHG approval: order code 590 ("Additional approval"), option LA ("SIL") or LC ("WHG").
74 Endress+Hauser
Micropilot FMR62
Function
• In addition to the verification parameters, the corresponding parameter values are also logged.
• Existing measured variables, such as the echo amplitude, are used in the Foam detection and
Build-up detection wizards.
The Foam detection and Build-up detection wizards cannot be used together.
Advantages
• Early detection of changes (trends) to ensure plant availability and product quality.
• Use of information for the proactive planning of measures (e.g. cleaning/maintenance).
• Identification of undesirable process conditions as the basis to optimizing the facility and the
processes.
• Automated control of measures to remove foam or buildup.
Detailed description
SD01870F
Endress+Hauser 75
Micropilot FMR62
Accessories
Device-specific accessories Weather protection cover
Accessory Description
222 (8.74)
A0015466
4)
38
35°
65°
(6.1
156
A0015472
Enclosed",
The weather protection cover can be ordered together with the device (product structure, feature 620 "Accessory
option PB "Weather Protection Cover").
Alternatively, it can be separately ordered as an accessory; order code 71162242.
76 Endress+Hauser
Micropilot FMR62
Accessories Description
2.
1.
4.
3.
A0019128
• Material:
– Plastic PBT
– 316L/1.4404
– Aluminum
• Degree of protection: IP68 / NEMA 6P and IP66 / NEMA 4x
• Suitable for display modules:
– SD02 (push buttons)
– SD03 (touch control)
• Connecting cable:
– Cable supplied with device up to 30 m (98 ft)
– Standard cable supplied by customer up to 60 m (196 ft)
• Ambient temperature range: –40 to 80 °C (–40 to 176 °F)
• IfL ortheM).remote display should be used, order the device version "Prepared for display FHX50" (feature 030, version
For the FHX50, you must select option A: "Prepared for display FHX50" under feature 050 "Measuring
device version".
• If the device version "Prepared for display FHX50" was not originally ordered and a FHX50 display is to be
retrofitted, you must select version B "Not prepared for display FHX50" under feature 050: "Measuring device
version" when ordering the FHX50. In this case, a retrofit kit for the device is supplied with the FHX50. The kit
can be used to prepare the device so that the FHX50 can be used.
FHX50
Use of the FHX50 may be restricted for transmitters with an approval. A device can only be retrofitted with the
if the option L or M ("Prepared for FHX50") is listed under Basic specifications, item 4 "Display, operation" in
the Safety Instructions (XA) for the device.
Also pay attention to the Safety Instructions (XA) of the FHX50.
Endress+Hauser 77
Micropilot FMR62
Overvoltage protection
Accessory Description
A0021734
Technical data
• Resistance per channel: 2 * 0.5 Ωmax
• Threshold DC voltage: 400 to 700 V
• Threshold impulse voltage: < 800 V
• Capacitance at 1 MHz: < 1.5 pF
• Nominal arrest impulse voltage (8/20 μs): 10 kA
• Suited for wire cross-sections: 0.2 to 2.5 mm2 (24 to 14 AWG)
Gas-tight feedthrough
Accessories Description
Gas-tight feedthrough Chemically inert glass feedthrough; prevents gases from entering the electronics housing
To order with the device: product structure, feature 610 "Accessory mounted", option NC "Gas-tight feedthrough"
78 Endress+Hauser
Micropilot FMR62
Accessory Description
Commubox FXA291 Connects Endress+Hauser field devices with CDI interface (= Endress+Hauser
Common Data Interface) to the USB interface of a computer.
Order code: 51516983
Accessory Description
HART Loop Converter Evaluates the dynamic HART variables and converts them to analog current signals
HMX50 or limit values.
Order code: 71063562
Accessory Description
Accessories Description
Connect Sensor Fully integrated, battery-powered gateway for simple applications with SupplyCare Hosting.
FXA30/FXA30B Up to 4 field devices with 4 to 20 mA communication (FXA30/FXA30B), serial Modbus
(FXA30B) or HART (FXA30B) can be connected. With its robust design and ability to run for
years on the battery, it is ideal for remote monitoring in isolated locations. Version with LTE
(USA, Canada and Mexico only) or 3G mobile transmission for worldwide communication.
BA01710S.
For details, see "Technical Information" TI01356S and Operating Instructions
Accessories Description
Fieldgate Fieldgates enable communication between connected 4 to 20 mA, Modbus RS485 and
FXA42 Modbus TCP devices and SupplyCare Hosting or SupplyCare Enterprise. The signals are
transmitted either via Ethernet TCP/IP, WLAN or mobile communications (UMTS). Advanced
automation capabilities are available, such as an integrated Web-PLC, OpenVPN and other
functions.
BA01778S.
For details, see "Technical Information" TI01297S and Operating Instructions
Accessories Description
SupplyCare Inventory management software that visualizes levels, volumes, masses, temperatures,
Enterprise pressures, densities or other tank parameters. The parameters are recorded and transmitted
SCE30B by means of gateways of the type Fieldgate FXA42.
This Web-based software is installed on a local server and can also be visualized and
operated with mobile terminals such as a smartphone or tablet.
BA00055S
For details, see "Technical Information" TI01228S and Operating Instructions
Endress+Hauser 79
Micropilot FMR62
Accessories Description
SupplyCare Inventory management software that visualizes levels, volumes, masses, temperatures,
Hosting SCH30 pressures, densities or other tank parameters. The parameters are recorded and transmitted
by means of gateways of the type Fieldgate FXA42, FXA30 and FXA30B.
SupplyCare Hosting is offered as a hosting service (Software as a Service, SaaS). In the
Endress+Hauser portal, the user is provided with the data over the Internet.
BA00050S.
For details, see "Technical Information" TI01229S and Operating Instructions
Accessory Description
Field Xpert SFX350 Field Xpert SFX350 is a mobile computer for commissioning and maintenance. It
enables efficient device configuration and diagnostics for HART and FOUNDATION
fieldbus devices in the non-Ex area.
Accessory Description
Field Xpert SFX370 Field Xpert SFX370 is a mobile computer for commissioning and maintenance. It
enables efficient device configuration and diagnostics for HART and FOUNDATION
fieldbus devices in the non-Ex area and the Ex area.
DeviceCare SFE100 Configuration tool for HART, PROFIBUS and FOUNDATION Fieldbus devices
Graphic Data Manager The graphic data manager Memograph M provides information on all the relevant
Memograph M process variables. Measured values are recorded correctly, limit values are
monitored and measuring points analyzed. The data are stored in the 256 MB
internal memory and also on an SD card or USB stick.
RN221N Active barrier with power supply for safe separation of 4 to 20 mA current circuits.
Provides bi-directional HART transmission.
RNS221 Transmitter supply for 2-wire sensors or transmitters exclusively for non-Ex areas.
Provides bi-directional communication using the HART communication sockets.
80 Endress+Hauser
Micropilot FMR62
Supplementary documentation
For an overview of the scope of the associated Technical Documentation, refer to the following:
• The W@M Device Viewer: enter the serial number from the nameplate
(www.endress.com/deviceviewer)
• The Endress+Hauser Operations App: Enter the serial number from the nameplate or scan the
2-D matrix code (QR code) on the nameplate.
The following document types are available:
In the Download Area of the Endress+Hauser Internet site: www.endress.com → Downloads
Safety Instructions (XA) Depending on the approval, the following Safety Instructions (XA) are supplied with the device. They
are an integral part of the Operating Instructions.
The nameplate indicates the Safety Instructions (XA) that are relevant to the device.
BC ATEX II 1/2G Ex ia/db [ia Ga] IIC T6 Ga/Gb XA01552F XA01552F XA01552F
BD ATEX II 1/2/3G Ex ia/ic [ia Ga] IIC T6 Ga/Gb/Gc XA01550F XA01550F XA01550F
BL ATEX II 1/2/3G Ex ia/ec [ia Ga] IIC T6 Ga/Gb/Gc XA01550F XA01550F XA01550F
B2 ATEX II 1/2G Ex ia IIC T6 Ga/Gb, 1/2D Ex ia IIIC T85°C Da/Db XA01555F XA01555F XA01555F
B3 ATEX II 1/2G Ex ia/db [ia Ga] IIC T6, Ga/Gb 1/2D Ex ta/tb IIIC T85°C Da/Db XA01556F XA01556F XA01556F
B4 ATEX II 1/2G Ex ia IIC T6 Ga/Gb, Ex ia/db [ia Ga] IIC T6 Ga/Gb XA01553F XA01553F XA01553F
Endress+Hauser 81
Micropilot FMR62
I3 IEC Ex ia/db [ia Ga] IIC T6 Ga/Gb, Ex ta/tb IIIC T85°C Da/Db XA01556F XA01556F XA01556F
I4 IEC Ex ia IIC T6 Ga/Gb, Ex ia/db [ia Ga] IIC T6 Ga/Gb XA01553F XA01553F XA01553F
KA KC Ex ia IIC T6 Ga * * *
KB KC Ex ia IIC T6 Ga/Gb * * *
MA INMETRO Ex ia IIC T6 Ga * * *
MH INMETRO Ex ic IIC T6 Gc * * *
NA NEPSI Ex ia IIC T6 Ga * * *
NG NEPSI Ex ec IIC T6 Gc * * *
NH NEPSI Ex ic IIC T6 Gc * * *
82 Endress+Hauser
*71375858*
71375858
www.addresses.endress.com