Sitransp Transmitters dsIII Fi01en
Sitransp Transmitters dsIII Fi01en
Sitransp Transmitters dsIII Fi01en
Transmitters for pressure, absolute pressure, differential pressure, flow and level
DS III series
■ Overview
2
EEPROM
6
Electronics
1 6
EEPROM
Sensor
Pe
Measuring cell
3
1 1 Measuring cell
2 Process connection
3 Seal diaphragm
4 Filling liquid Measuring cell for absolute pressure from differential pressure series,
5 Silicon pressure sensor functional diagram
pe Pressure as input variable The input pressure pe is transmitted through the seal diaphragm
2
(6, Figure "Measuring cell for absolute pressure from differential
pressure series, functional diagram") and the filling liquid (8) to
p the silicon pressure sensor (3).
e
The difference in pressure between the input pressure pe and
the reference vacuum (1) on the low-pressure side of the measu-
Measuring cell for pressure, functional diagram ring cell flexes the measuring diaphragm. The resistance of the
four piezo-resistors fitted in the diaphragm in a bridge circuit
The pressure pe is applied through the process connection (2, thus changes. This change in resistance results in a bridge out-
Figure "Measuring cell for pressure, functional diagram) to the put voltage proportional to the absolute pressure.
measuring cell (1). This pressure is subsequently transmitted
further through the seal diaphragm (3) and the filling liquid (4) to An overload diaphragm is installed to provide protection from
the silicon pressure sensor (5) whose measuring diaphragm is overloads. If the measuring limits are exceeded, the overload di-
then flexed. The resistance of the four piezo-resistors fitted in the aphragm (2) is flexed until the seal diaphragm rests on the body
diaphragm in a bridge circuit thus changes. This change in re- of the measuring cell (7), thus protecting the silicon pressure
sistance results in a bridge output voltage proportional to the in- sensor from overloads.
put pressure.
Measuring cell for differential pressure and flow
Measuring cell for absolute pressure from pressure series
2 3 4
1 Seal diaphragm
2 O-ring
3 Overload diaphragm
5 4 Silicon pressure sensor
4 1 - + 1 5 Process flange
5 6 Body of measuring cell
3
7 Filling liquid
1
1 Measuring cell
7 6
2 Process connection
3 Seal diaphragm
4 Filling liquid
5 Silicon absolute pressure sensor
pe Absolute pressure as input variable Measuring cell for differential pressure and flow, functional diagram
Parameterization PC or
HART laptop
Depending on the version, there are different possibilities for pa- modem
rameterizing the pressure transmitter and for setting or scanning
the parameters. RS 232
Adjustable parameters
HART communication between a PC communicator and a pressure
Parameters Input keys HART com- transmitter
munication
Start of scale x x When parameterizing with a PC, the connection is made through
a HART modem.
Full-scale value x x
The signals needed for communication in conformity with the
Electrical damping x x HART 5.x or 6.x protocols are superimposed on the output cur-
Start-of-scale value without application of a x x rent using the Frequency Shift Keying (FSK) method.
pressure ("Blind setting")
Physical dimensions available for the display
End-of-scale value without application of a x x
pressure ("Blind setting") Physical variable Physical dimensions
Zero adjustment x x Pressure (setting can also Pa, MPa, kPa, hPa, bar, mbar, torr, atm,
be made in the factory) psi, g/cm2, kg/cm2, inH20, inH20 (4 °C),
Current transmitter x x mmH20, ftH20, inHg, mmHg
Fault current x x Level (height data) m, cm, mm, ft, in
Disabling of keys, write protection x x 1) Volume m3, dm3, hl, yd3, ft3, in3, US gallon, lmp,
Type of dimension and actual dimension x x gallon, bushel, barrel, barrel liquid
Characteristic (linear / square-rooted) x 2) x 2) Volume flow m3/d, m3/h, m3/s, l/min, l/s, ft3/d, ft3/min,
ft3/s, US gallon/min, US gallon/s
Input of characteristic x
Mass flow t/d, t/h, t/min, kg/d, kg/h, kg/min, kg/s, g/d,
Freely-programmable LCD x g/h, g/min, g/s, lb/d, lb/h, lb/min, lb/s,
LTon/d, LTon/h, STon/d, STon/h, STon/min
Diagnostics functions x
1)
Total mass flow t, kg, g, lb, oz, LTon, STon
Cancel apart from write protection
2)
Only differential pressure Temperature K, °C, °F, °R
Miscellaneous %, mA
■ Technical specifications
2
Process connection Connection shank G½A to DIN
EN 837-1, female thread ½-14
SITRANS P pressure transmitters, DS III series, for pressure NPT or oval flange (PN 160
(MWP 2320)) with mounting
Input thread M10 or 7/16-20 UNF to EN
Measured variable Pressure 61518
• 0.01 ... 1 bar (0.145 ... 14.5 psi) 6 bar (87 psi) Terminal voltage on transmitter 10.5 ... 45 V DC
10.5 ... 30 V DC in intrinsically-
• 0.04 ... 4 bar (0.58 ... 58 psi) 10 bar (145 psi) safe mode
• 0.16 ... 16 bar (2.32 ... 232 psi) 32 bar (464 psi) Certificates and approvals
• 0.6 ... 63 bar (9.14 ... 914 psi) 100 bar (1450 psi) Classification according to pressure For gases of fluid group 1 and
equipment directive liquids of fluid group 1; complies
• 1.6 ... 160 bar (23.2 ... 2320 psi) 250 bar (3626 psi) (DRGL 97/23/EC) with requirements of article 3,
• 4.0 ... 400 bar (58 ... 5802 psi) 500 bar (7252 psi) paragraph 3 (sound engineering
practice)
Lower measuring limit
Explosion protection
• Measuring cell with silicone oil fil- 30 mark (0.435 psi) absolute
ling • Intrinsic safety "i" PTB 99 ATEX 2122
Upper measuring limit 100% of max. span (max. 160 bar - Identification Ex II 1/2 G EEx ia/ib IIB/IIC T6
(2320 psi) with oxygen measure- - Permissible ambient temperatu- -40 ... +85 °C (-40 ... +185 °F)
ment and inert liquid)
re temperature class T4;
Output -40 ... +70 °C (-40 ... +158 °F)
temperature class T5;
Output signal 4 ... 20 mA -40 ... +60 °C (-40 ... +140 °F)
Measuring accuracy temperature class T6
Reference conditions Increasing characteristic - Connection To certified intrinsically-safe cir-
Start-of-scale value 0 bar cuits with maximum values:
Stainless steel seal diaphragm Ui = 30 V, Ii = 100 mA,
Silicone oil filling Pi = 750 mW; Ri = 300 Ω
Room temperature (25 °C (77 °C)) - Effective internal inductance/ca- Li = 0.4 mH, Ci = 6 nF
r: Span ratio (r = max. span / set pacitance
span)
• Explosion-proof "d" PTB 99 ATEX 1160
Error in measurement and fixed-
point setting (including hysteresis - Identification Ex II 1/2 G EEx d IIC T4/T6
and repeatability) - Permissible ambient temperatu- -40 ... +85 °C (-40 ... +185 °F)
• Linear characteristic re temperature class T4;
-40 ... +60 °C (-40 ... +140 °F)
- r ≤ 10 ≤ (0.0029 ⋅ r +0.071)% temperature class T6
- 10 < r ≤ 30 ≤ (0.0045 ⋅ r +0.071)% - Connection To circuits with values:
- 30 < r ≤ 100 ≤ (0.005 ⋅ r +0.05)% UH = 10.5 ... 45 V DC
Influence of ambient temperature • Dust explosion protection for zone PTB 01 ATEX 2055
20
• With -10 ... +60 °C (14 ... 140 °F) ≤ (0.08 ⋅ r +0.1)%
- Identification Ex II 1 D IP65 T 120°°C
• With -40 ... -10 °C and ≤ (0.1 ⋅ r +0.15)% / 10 K Ex II 1/2 D IP65 T 120 °C
+60 °C ... +85 °C (≤ (0.1 ⋅ r +0.15)% / 18 °F)
(-40 ... +14 and 140 ... 185 °F) - Permissible ambient temperatu- -40 ... +85 °C (-40 ... +185 °F)
re
Rated conditions
- Max.surface temperature 120 °C (248 °F)
Degree of protection (to EN 60529) IP65
- Connection To certified intrinsically-safe cir-
Process temperature cuits with maximum values:
• Measuring cell with silicone oil fil- -40 ... +100 °C (-40 ... +212 °F) Ui = 30 V, Ii = 100 mA,
ling Pi = 750 mW, Ri = 300 Ω
• Measuring cell with inert filling li- -20 ... +100 °C (-4 ... +212 °F) - Effective internal inductance/ca- Li = 0.4 mH, Ci = 6 nF
quid pacitance
• In conjunction with dust explosion -20 ... +60 °C (-4 ... +140 °F) • Dust explosion protection for zone PTB 01 ATEX 2055
protection 21/22
Weight (without options) ≈ 1.5 kg (≈ 3.3 lb) - Connection To circuits with values:
UH = 10.5 ... 45 V DC;
Wetted parts materials Pmax = 1.2 W
• Connection shank Stainless steel, mat. No. • Type of protection "n" (zone 2) TÜV 01 ATEX 1696 X
1.4404/316L or Hastelloy C4, mat.
No. 2.4610 - Identification Ex II 3 G EEx nA L IIC T4/T5/T6
• Seal diaphragm Stainless steel, mat. No. • Explosion protection to FM Certificate of Compliance
1.4404/316L or Hastelloy C276, 3008490
mat. No. 2.4819 - Identification (XP/DIP) or (IS); CL I, DIV 1, GP ABCD T4...T6;
Measuring cell filling Silicone oil or inert filling liquid (NI) CL II, DIV 1, GP EFG; CL III; CL I,
(max. 160 bar (2320 psi) with oxy- ZN 0/1 AEx ia IIC T4...T6; CL I,
gen measurement) DIV 2, GP ABCD T4...T6; CL II,
DIV 2, GP FG; CL III
2
• Explosion protection to CSA Certificate of Compliance
1153651
- Identification (XP/DIP) or (IS) CL I, DIV 1, GP ABCD T4...T6;
CL II, DIV 1, GP EFG; CL III; Ex ia
IIC T4...T6; CL I, DIV 2, GP ABCD
T4...T6; CL II, DIV 2, GP FG; CL III
■ Dimensional drawings
15 55 4)
138 (5.4) (0.6) ca. 30 (1.2) 100 (3.94)
(0.6)
69 50 28 ca. 96 (3.78) 6 2
(2.7) (1.97) (1.1)
(0.8)
20
(1.97)
50
(3.15)
ø 80
3 3
117 (4.6)
171 (6.7)
1) 8
237 (9.3)
1) 4
5
68 (2.7)
120 (4.7)
1
72 (2.8)
(1,44)
36,5
123 (4.8)
7
min. 77 (3.0) 5)
105 (4.1)
1 Process connection:
- ½-14 NPT,
- connection shank G½A or
- oval flange
2 Blanking plug
3 Electrical connection:
- screwed gland Pg 13,5 (adapter)2) 3),
- screwed gland M20x1,5 3),
- screwed gland ½-14 NPT or
- Han 7D/ Han 8U plug 2) 3)
4 Terminal side
5 Electronics side, digital display (longer overall
length for cover with window) 1) Allow approx. 20 mm (0.79 inch) thread length to permit unscrewing
6 Protective cover over keys 2) Not with type of protection "explosion-proof enclosure"
7 Mounting bracket (option) 3) Not with type of protection "FM + CSA [is + xp]"
8 Screw cover - safety bracket (only for type of protection 4) 45 mm (1.8 inch) for Pg 13,5 with adapter
"Explosion-proof enclosure", not shown in the drawing) 5) Minimum distance for rotation
SITRANS P pressure transmitters, DS III series for pressure, dimensional drawing, dimensions in mm (inch)
2
Selection and Ordering data Order No. 1) For oxygen application, add Order code E10.
SITRANS P pressure transmitter, DS III series, } 7 M F 4 0 3 3 - 2) Without cable gland, with blanking plug
for pressure 3) With enclosed cable gland EEx ia and blanking plug
77777 - 7777 4) Not together with type of protection "Explosion-proof"
2-wire system
Measuring cell filling Measuring cell
cleaning
Silicone oil Standard } 1
Inert liquid1) Grease-free 3
Measured span
0.01 ... 1 bar (0.15 ... 14.5 psi) } B
0.04 ... 4 bar (0.58 ... 58 psi) } C
0.16 ... 16 bar (2.32 ... 232 psi) } D
0.63 ... 63 bar (9.14 ... 914 psi) } E
1.6 ... 160 bar (23.2 ... 2320 psi) } F
4.0 ... 400 bar (58.0 ... 5802 psi) } G
Wetted parts materials
Seal diaphragm Process connection
Stainless steel Stainless steel } A
Hastelloy Stainless steel B
Hastelloy Hastelloy C
Version as diaphragm seal Y0
Process connection
• Connection shank G½B to EN 837-1 } 0
• Female thread ½-14 NPT 1
• Oval flange made of stainless steel,
max. span 160 bar (2320 psi)
- Mounting thread 7/16-20 UNF to 2
EN 61518
- Mounting thread M10 to DIN 19213 3
Non-wetted parts materials
• Housing made of die-cast aluminium } 0
• Housing stainless steel precision casting 3
Design
• Standard design 1
• International version, English label inscriptions, } 2
documentation in 5 languages on CD
Explosion protection
• without A
• with ATEX, Type of protection:
- "Intrinsic safety (EEx ia)" B
- "Explosion-proof (EEx d)" 2) D
- "Intrinsic safety and explosion-proof enclosure } P
(EEx ia + EEx d)" 3)
- "n (zone 2)" E
- "Intrinsic safety, explosion-proof enclosure and R
dust explosion protection (EEx ia + EEx d +
zone 1D/2D)" 3)
• with FM + CSA, Type of protection:
- "Intrinsic safety and explosion-proof (is + xp)" 2) NC
Electrical connection / cable inlet
• Screwed gland Pg 13.5 (adapter) 4) A
• Screwed gland M20x1.5 } B
• Screwed gland ½-14 NPT C
• Han 7D plug (plastic housing) incl. mating D
connector 4)
Display
• without (digital indicator hidden, setting: mA) } 1
• with visible digital indicator, setting: mA 6
2
Further designs Order code Additional data Order code
Please add "-Z" to Order No. and specify Order code. Please add "-Z" to Order No. and specify Order code.
Pressure transmitter with mounting bracket made of: Measuring range to be set Y01
• Steel A01 specify in plain text:
• Stainless steel A02 Y01: ... to ... mbar, bar, kPa, MPa, psi
Plug Measuring point number/identification Y15
• Han 7D (metal, gray) A30 max. 16 characters, specify in plain text:
• Han 8U (instead of Han 7D) A31 Y15: ...........................................
English rating plate B21 Setting of pressure indicator in pressure units Y21
Pressure units in inH2O or psi specify in plain text (standard setting: mA):
Y21: mbar, bar, kPa, MPa, psi, ...
Manufacturer's test certificate M C11 Note:
(calibration certificate) The following pressure units can be selected:
to DIN 55350, Part 18 and to ISO 8402 bar, mbar, mm H2O*), inH2O*), ftH2O*), mmHG, inHG,
Acceptance test certificate B C12 psi, Pa, kPa, MPa, g/cm2, kg/cm2, mA, Torr, ATM or %
to EN 10204-3.1.B *) Reference temperature 20 °C
Factory certificate C14 Setting of pressure indicator in non-pressure units Y22 + Y01
to EN 10204-2.2 specify in plain text:
Y22: ..... up to ..... l/min, m3/h, m, USgpm, ...
Certificate "Functional Safety (SIL)" C20 (specification of measuring range in pressure units
"Y01" is essential, unit with max. 5 characters)
Setting of upper limit of output signal to 22.0 mA D05
Only the settings for "Y01", "Y21", "Y22" and "D05" can be made in the
Acid gas version to NACE D07 factory
(only together with seal diaphragm made of Hastelloy)
Ordering example
Type of protection IP68 D12 Item line: 7MF4033-1EA00-1AA7-Z
(not together with Han 7D / Han 8U plug, B line: A01 + Y01 + Y21
Pg 13.5 screwed gland)
C line: Y01: 10 ... 20 bar (145 ... 290 psi)
Digital indicator along side the input keys D27 C line: Y21: bar (psi)
(only together with the devices 7MF4033-....0-.A.6
or -.A.7-Z, Y21 or Y22 + Y01)
Use in or at zone 1D/2D E01
(only together with type of protection "Intrinsic safety
(EEx ia)"
Use at zone 0 E02
(only together with type of protection "Intrinsic safety
(EEx ia)"
Oxygen application E10
(max. 160 bar (2320 psi) with oxygen measurement
and inert liquid)
■ Technical specifications
2
Process connection Connection shank G½A to DIN
EN 837-1, female thread ½-14
SITRANS P pressure transmitters, DS III series for absolute pres- NPT or oval flange (PN 160
sure, from the pressure series (MWP 2320)) with mounting
thread M10 or 7/16-20 UNF to
Input EN 61518
Measured variable Absolute pressure Power supply UH
Span Maximum working pressure Terminal voltage on pressure trans- 10.5 ... 45 V DC
• 8.3 ... 250 mbar (0.12 ... 3.6 psi) 6 bar (87 psi) mitter 10.5 ... 30 V DC in intrinsically-
safe mode
• 43 ... 1300 mbar (0.62 ... 18.9 psi) 10 bar (145 psi)
Certificates and approvals
• 160 ... 5000 mbar 30 bar (435 psi)
(2.32 ... 72.5 psi) Classification according to pressure For gases of fluid group 1 and
equipment directive liquids of fluid group 1; complies
• 1 ... 30 bar (14.5 ... 435 psi) 100 bar (1450 psi) (DRGL 97/23/EC) with requirements of article 3,
paragraph 3 (sound engineering
Lower measuring limit practice)
• Measuring cell with silicone oil fil- 0 mark (0 psi) absolute Explosion protection
ling
• Intrinsic safety "i" PTB 99 ATEX 2122
Upper measuring limit 100% of max. span (max. 160 bar
(2320 psi) with oxygen measure- - Identification Ex II 1/2 G EEx ia/ib IIB/IIC T6
ment and inert liquid)
- Permissible ambient temperatu- -40 ... +85 °C (-40 ... +185 °F)
Output re temperature class T4;
-40 ... +70 °C (-40 ... +158 °F)
Output signal 4 ... 20 mA temperature class T5;
Measuring accuracy -40 ... +60 °C (-40 ... +140 °F)
temperature class T6
Reference conditions Increasing characteristic
Start-of-scale value 0 bar - Connection To certified intrinsically-safe cir-
Stainless steel seal diaphragm cuits with maximum values:
Silicone oil filling Ui = 30 V, Ii = 100 mA,
Room temperature (25 °C (77 °F)) Pi = 750 mW; Ri = 300 Ω
r: Span ratio
- Effective internal inductance/ca- Li = 0.4 mH, Ci = 6 nF
(r = max. span / set span) pacitance
Error in measurement and fixed- • Explosion-proof "d" PTB 99 ATEX 1160
point setting (including hysteresis
and repeatability) - Identification Ex II 1/2 G EEx d IIC T4/T6
• Linear characteristic - Permissible ambient temperatu- -40 ... +85 °C (-40 ... +185 °F)
re temperature class T4;
- r ≤ 10 ≤ (0.0029 ⋅ r +0.071)% -40 ... +60 °C (-40 ... +140 °F)
- 10 < r ≤ 30 ≤ (0.0045 ⋅ r +0.071)% temperature class T6
Influence of ambient temperature - Connection To circuits with values:
UH = 10.5 ... 45 V DC
• With -10 ... +60 °C (14 ... 140 °F) ≤ (0.1 ⋅ r +0.2)%
• Dust explosion protection for PTB 01 ATEX 2055
• With -40 ... -10 °C and ≤ (0.1 ⋅ r +0.15)% / 10 K zone 20
+60 °C ... +85 °C (≤ (0.1 ⋅ r +0.15)% / 18 °F)
(-40 ... +14 and 140 ... 185 °F) - Identification Ex II 1 D IP65 T 120 °C
Ex II 1/2 D IP65 T 120 °C
Rated conditions
- Permissible ambient temperatu- -40 ... +85 °C (-40 ... +185 °F)
Degree of protection (to EN 60529) IP65 re
Process temperature - Max.surface temperature 120 °C (248 °F)
• Measuring cell with silicone oil fil- -40 ... +100 °C (-40 ... +212 °F) - Connection To certified intrinsically-safe cir-
ling cuits with maximum values:
• Measuring cell with inert filling li- -20 ... +100 °C (-4 ... +212 °F) Ui = 30 V, Ii = 100 mA,
quid Pi = 750 mW, Ri = 300 Ω
• In conjunction with dust explosion -20 ... +60 °C (-4 ... +140 °F) - Effective internal inductance/ca- Li = 0.4 mH, Ci = 6 nF
protection pacitance
Design • Dust explosion protection for zone PTB 01 ATEX 2055
21/22
Weight (without options) ≈ 1.5 kg (≈ 3.3 lb)
- Identification Ex II 2 D IP65 T 120 °C
Wetted parts materials
- Connection To circuits with values:
• Connection shank Stainless steel, mat. No. UH = 10.5 ... 45 V DC;
1.4404/316L or Hastelloy C4, mat. Pmax = 1.2 W
No. 2.4610
• Type of protection "n" (zone 2) TÜV 01 ATEX 1696 X
• Seal diaphragm Stainless steel, mat. No.
1.4404/316L or Hastelloy C276, - Identification Ex II 3 G EEx nA L IIC T4/T5/T6
mat. No. 2.4819 • Explosion protection to FM Certificate of Compliance
Measuring cell filling Silicone oil or inert filling liquid 3008490
(max. 160 bar (2320 psi) with oxy- - Identification (XP/DIP) or (IS); CL I, DIV 1, GP ABCD T4...T6;
gen measurement) (NI) CL II, DIV 1, GP EFG; CL III; CL I,
ZN 0/1 AEx ia IIC T4...T6; CL I,
DIV 2, GP ABCD T4...T6; CL II,
DIV 2, GP FG; CL III
2
• Explosion protection to CSA Certificate of Compliance
1153651
- Identification (XP/DIP) or (IS) CL I, DIV 1, GP ABCD T4...T6;
CL II, DIV 1, GP EFG; CL III; Ex ia
IIC T4...T6; CL I, DIV 2, GP ABCD
T4...T6; CL II, DIV 2, GP FG; CL III
■ Dimensional drawings
15 55 4)
138 (5.4) (0.6) ca. 30 (1.2) 100 (3.94)
(0.6)
69 50 28 ca. 96 (3.78) 6 2
(2.7) (1.97) (1.1)
(0.8)
20
(1.97)
50
(3.15)
ø 80
3 3
117 (4.6)
171 (6.7)
1) 8
237 (9.3)
1) 4
5
68 (2.7)
120 (4.7)
1
72 (2.8)
(1,44)
36,5
123 (4.8)
7
min. 77 (3.0) 5)
105 (4.1)
1 Process connection:
- ½-14 NPT,
- connection shank G½A or
- oval flange
2 Blanking plug
3 Electrical connection:
- screwed gland Pg 13,5 (adapter)2) 3),
- screwed gland M20x1,5 3),
- screwed gland ½-14 NPT or
- Han 7D/ Han 8U plug 2) 3)
4 Terminal side
5 Electronics side, digital display (longer overall
length for cover with window) 1) Allow approx. 20 mm (0.79 inch) thread length to permit unscrewing
6 Protective cover over keys 2) Not with type of protection "explosion-proof enclosure"
7 Mounting bracket (option) 3) Not with type of protection "FM + CSA [is + xp]"
8 Screw cover - safety bracket (only for type of protection 4) 45 mm (1.8 inch) for Pg 13,5 with adapter
"Explosion-proof enclosure", not shown in the drawing) 5) Minimum distance for rotation
SITRANS P pressure transmitters, DS III series for absolute pressure, from the pressure series, dimensional drawing, dimensions in mm (inch)
2
Selection and Ordering data Order No. E) Combinations of the versions marked with E) are subject to the export regu-
SITRANS P pressure transmitters, DS III series 7MF 4 2 3 3 - lations AL: 2B230, ECCN: N.
for absolute pressure from the pressure series
2-wire system 77777 - 7777
Measuring cell filling Measuring cell
cleaning
Silicone oil Standard 1
Inert liquid Grease-free 3
Measured span
8.3 ... 250 mbar (0.12 ... 3.63 psi) E) D
43 ... 1300 mbar (0.62 ... 18.9 psi) E) F
0.16 ... 5 bar (2.32 ... 72.5 psi) E) G
1 ... 30 bar (14.5 ... 435 psi) H
Wetted parts materials
Seal diaphragm Process connection
Stainless steel Stainless steel A
Hastelloy Stainless steel E) B
Hastelloy Hastelloy E) C
Version for diaphragm seal 1) Y0
Process connection
• Connection shank G½B to EN 837-1 0
• Female thread ½-14 NPT 1
• Oval flange made of stainless steel,
max. span 160 bar (2320 psi)
- Mounting thread 7/16-20 UNF to 2
EN 61518
- Mounting thread M10 to DIN 19213 3
Non-wetted parts materials
• Housing made of die-cast aluminium 0
• Housing stainless steel precision casting 3
Design
• Standard design 1
• International version, English label inscriptions, 2
documentation in 5 languages on CD
Explosion protection
• without A
• with ATEX, Type of protection:
- "Intrinsic safety (EEx ia)" B
- "Explosion-proof (EEx d)" 2) D
- "Intrinsic safety and explosion-proof enclosure P
(EEx ia + EEx d)" 3)
- "n (zone 2)" E
- "Intrinsic safety, explosion-proof enclosure and R
dust explosion protection (EEx ia + EEx d +
zone 1D/2D)" 3)
• with FM + CSA, Type of protection:
- "Intrinsic safety and explosion-proof (is + xp)" 2) NC
Electrical connection / cable inlet
• Screwed gland Pg 13.5 4) A
• Screwed gland M20x1.5 B
• Screwed gland ½-14 NPT C
• Han 7D plug (plastic housing) incl. mating D
connector 4)
Display
• without (digital indicator hidden, setting: mA) 1
• with visible digital indicator 6
• with customer-specific digital indicator (setting 7
as specified, Order code "Y21" required)
Power supply units see "SITRANS I power supply units and input isola-
tors".
The device is delivered together with brief instructions (Leporello) and
a CD-ROM containing detailed documentation.
2
Further designs Order code Further designs Order code
Please add "-Z" to Order No. and specify Order code. Please add "-Z" to Order No. and specify Order code.
Pressure transmitter with mounting bracket made of: Additional data
• Steel A01 Measuring range to be set Y01
• Stainless steel A02 specify in plain text:
Plug Y01: ... to ... mbar, bar, kPa, MPa, psi
• Han 7D (metal, gray) A30 Measuring point number/identification Y15
• Han 8U (instead of Han 7D) A31 max. 16 characters, specify in plain text:
Rating plate inscription Y15: ...........................................
(instead of German) Measuring point text Y16
• English B11 max. 27 characters, specify in plain text:
• French B12 Y16: ...........................................
• Spanish B13 Entry of HART address (TAG) Y17
• Italian B14 max. 8 characters, specify in plain text:
English rating plate B21 Y17: ...........................................
Pressure units in inH2O or psi Setting of pressure indicator in pressure units Y21
Manufacturer's test certificate M C11 specify in plain text (standard setting: mA):
(calibration certificate) Y21: mbar, bar, kPa, MPa, psi, ...
to DIN 55350, Part 18 and to ISO 8402 Note:
The following pressure units can be selected:
Acceptance test certificate B C12 bar, mbar, mm H2O*), inH2O*), ftH2O*), mmHG, inHG,
to EN 10204-3.1.B psi, Pa, kPa, MPa, g/cm2, kg/cm2, mA, Torr, ATM or %
Factory certificate C14 *) Reference temperature 20 °C
to EN 10204-2.2 Setting of pressure indicator in non-pressure units Y22 + Y01
Certificate "Functional Safety (SIL)" C20 specify in plain text:
Y22: ..... up to ..... l/min, m3/h, m, USgpm, ...
Setting of upper limit of output signal to 22.0 mA D05 (specification of measuring range in pressure units
"Y01" is essential, unit with max. 5 characters)
Acid gas version to NACE D07
(only together with seal diaphragm made of Hastelloy) Only the settings for "Y01", "Y21", "Y22" and "D05" can be made in the
factory
Type of protection IP68 D12
(not together with Han 7D / Han 8U plug,
Pg 13.5 screwed gland)
Digital indicator along side the input keys D27
(only together with the devices 7MF4233-....0-.A.6
or -.A.7-Z, Y21 or Y22 + Y01).
Use in or at zone 1D/2D E01
(only together with type of protection "Intrinsic safety
(EEx ia)"
Use at zone 0 E02
(only together with type of protection "Intrinsic safety
(EEx ia)"
Oxygen application E10
(max. 160 bar (2320 psi) with oxygen measurement
and inert liquid)
■ Technical specifications
2
Measuring cell filling Silicone oil or inert filling liquid
(max. 160 bar (2320 psi) with oxy-
SITRANS P pressure transmitters, DS III series for absolute pres- gen measurement)
sure, from the differential pressure series Process connection Female thread ¼-18 NPT and
Input flange connection to DIN 19 213
with mounting thread M10 or
Measured variable Absolute pressure 7/ -20 UNF
16
Span Maximum working pressure Power supply UH
• 8.3 ... 250 mbar (0.12 ... 3.6 psi) 32 bar (464 psi) Terminal voltage on pressure trans- 10.5 ... 45 V DC
• 43 ... 1300 mbar (0.62 ... 18.9 psi) 32 bar (464 psi) mitter 10.5 ... 30 V DC in intrinsically-
safe mode
• 160 ... 5000 mbar 32 bar (464 psi)
(2.32 ... 72.5 psi) Certificates and approvals
• 1 ... 30 bar (14.5 ... 435 psi) 160 bar (2320 psi) Classification according to pressure For gases of fluid group 1 and
equipment directive liquids of fluid group 1; complies
• 5.3 ... 100 bar (76.9 ... 1450 psi) 160 bar (2320 psi) (DRGL 97/23/EC) with requirements of article 3,
With connection thread M10 and paragraph 3 (sound engineering
7/ -20 UNF in the process flan- practice)
16
ges Explosion protection
Lower measuring limit • Intrinsic safety "i" PTB 99 ATEX 2122
• Measuring cell with silicone oil fil- 0 mark (0 psi) absolute - Identification Ex II 1/2 G EEx ia/ib IIB/IIC T6
ling
- Permissible ambient temperatu- -40 ... +85 °C (-40 ... +185 °F)
Upper measuring limit 100% of max. span (max. 160 bar re temperature class T4;
(2320 psi) with oxygen measure- -40 ... +70 °C (-40 ... +158 °F)
ment and inert liquid) temperature class T5;
Start of scale Continuously adjustable between -40 ... +60 °C (-40 ... +140 °F)
the measuring limits temperature class T6
- r ≤ 10 ≤ (0.0029 ⋅ r + 0.071)% • Dust explosion protection for zone PTB 01 ATEX 2055
20
- 10 < r ≤ 30 ≤ (0.0045 ⋅ r + 0.071)%
- Identification Ex II 1 D IP65 T 120 °C
Influence of ambient temperature Ex II 1/2 D IP65 T 120 °C
• With -10 ... +60 °C (14 ... 140 °F) ≤ (0.1 ⋅ r +0.2)% - Permissible ambient temperatu- -40 ... +85 °C (-40 ... +185 °F)
Twice the value with 20-mbar re
(0.29 psi) measuring cell
- Max.surface temperature 120 °C (248 °F)
• With -40 ... -10 °C and ≤ (0.1 ⋅ r +0.15)% / 10 K
+60 °C ... +85 °C (≤ (0.1 ⋅ r +0.15)% / 18 °F) - Connection To certified intrinsically-safe cir-
(-40 ... +14 and 140 ... 185 °F) Twice the value with 20-mbar cuits with maximum values:
(0.29 psi) measuring cell Ui = 30 V, Ii = 100 mA,
Pi = 750 mW, Ri = 300 Ω
Rated conditions
- Effective internal inductance/ca- Li = 0.4 mH, Ci = 6 nF
Degree of protection (to EN 60529) IP65 pacitance
Process temperature • Dust explosion protection for PTB 01 ATEX 2055
zone 21/22
• Measuring cell with silicone oil fil- -40 ... +100 °C (-40 ... +212 °F)
ling - Identification Ex II 2 D IP65 T 120 °C
• Measuring cell with inert filling li- -20 ... +100 °C (-4 ... +212 °F) - Connection To circuits with values:
quid UH = 10.5 ... 45 V DC;
Pmax = 1.2 W
• In conjunction with dust explosion -20 ... +60 °C (-4 ... +140 °F)
protection • Type of protection "n" (zone 2) TÜV 01 ATEX 1696 X
Design - Identification Ex II 3 G EEx nA L IIC T4/T5/T6
Weight (without options) ≈ 4.5 kg (≈ 9.9 lb) • Explosion protection to FM Certificate of Compliance
3008490
Wetted parts materials
- Identification (XP/DIP) or (IS); CL I, DIV 1, GP ABCD T4...T6;
• Seal diaphragm Stainless steel, mat. No.
(NI) CL II, DIV 1, GP EFG; CL III; CL I,
1.4404/316L, Hastelloy C276,
ZN 0/1 AEx ia IIC T4...T6; CL I,
mat. No. 2.4819, Monel, mat. No.
DIV 2, GP ABCD T4...T6; CL II,
2.4360, tantalum or gold
DIV 2, GP FG; CL III
2
• Explosion protection to CSA Certificate of Compliance
1153651
- Identification (XP/DIP) or (IS) CL I, DIV 1, GP ABCD T4...T6;
CL II, DIV 1, GP EFG; CL III; Ex ia
IIC T4...T6; CL I, DIV 2, GP ABCD
T4...T6; CL II, DIV 2, GP FG; CL III
■ Dimensional drawings
15
(0.6) 138 (5.4) approx. 30 (1.2) 5) 100 (3.94)
50 6 2
(1.97)
(0.8)
20
(1.97)
50
(3.15)
Ø 80
117 (4.6)
198 (7.8)
8
1
of housing
134 (5.27)
96 (3.8)
11
ca. 96 (3.78) 52 (2.05)
72 (2.83)
105 (4.1) 68 (2.7)
120 (4.7)
1 Process connection: ¼-18 NPT (EN 61518)
2 Blanking plug
3 Electrical connection:
- screwed gland Pg 13,5 (adapter) 2) 3),
- screwed gland M20x1,5 3),
- screwed gland ½-14 NPT or
- Han 7D/ Han 8U plug 2) 3)
4 Terminal side
5 Electronics side, digital display (longer overall length for cover with window)
6 Protective cover over keys
7 Mounting bracket (option)
8 Sealing screw with valve (option) 1) Allow approx. 20 mm (0.79 inch) thread length to permit unscrewing
9 Screw cover - safety bracket (only for type of protection 2) Not with type of protection "explosion-proof enclosure"
"Explosion-proof enclosure", not shown in the drawing) 3) Not with type of protection "FM + CSA [is + xp]"
10 Lateral venting for liquid measurement 4) 92 mm (3.62 inch) for minimum distance to permit rotation with indicator
11 Lateral venting for gas measurement (suffix H02) 5) 45 mm (1.8 inch) for Pg 13,5 with adapter
SITRANS P pressure transmitters, DS III series for absolute pressure, from the differential pressure series, dimensional drawing, dimensions in mm
(inch)
2
Selection and Ordering data Order No. Selection and Ordering data Order No.
SITRANS P pressure transmitters, DS III series 7MF 4 3 3 3 - SITRANS P pressure transmitters, DS III series 7MF 4 3 3 3 -
for absolute pressure from the series for absolute pressure from the series
Differential pressure Differential pressure
2-wire system 77777 - 7777 2-wire system 77777 - 7777
Measuring cell filling Measuring cell Display
cleaning • without (digital indicator hidden, setting: mA) 1
Silicone oil Standard 1 • with visible digital indicator 6
Inert liquid1) Grease-free 3 • with customer-specific digital indicator (setting 7
Measured span as specified, Order code "Y21" required)
8.3 ... 250 mbar (0.12 ... 3.63 psi) E) D Power supply units see "SITRANS I power supply units and input isola-
43 ... 1300 mbar (0.62 ... 18.9 psi) E) F tors".
0.16 ... 5 bar (2.32 ... 72.5 psi) E) G Included in delivery of the device:
1 ... 30 bar (14.5 ... 435 psi) H • Brief instructions (Leporello)
5.3 ... 100 bar (76.9 ... 1450 psi) KE • CD-ROM with detailed documentation
• Sealing plug(s) or sealing screw(s) for the process flanges(s)
Wetted parts materials
Seal diaphragm Parts of measuring cell 1) For oxygen applications, add Order code E10.
2) Version 7MF4433-1DY... only up to max. span 200 mbar (2.9 psi)
Stainless steel Stainless steel A 3) Not for span "5.3 ... 100 bar (76.9 ... 1450 psi)"
Hastelloy Stainless steel E) B 4) Without cable gland, with blanking plug
Hastelloy Hastelloy E) C 5) With enclosed cable gland EEx ia and blanking plug
Tantalum Tantalum E 6) Not together with type of protection "Explosion-proof"
Monel Monel E) H E) Combinations of the versions marked with E) are subject to the export
Gold Gold L regulations AL: 2B230, ECCN: N.
Version for diaphragm seal 2) Y
Process connection
Female thread ¼-18 NPT with flange connection
• Sealing screw opposite process connection
- Mounting thread M10 to DIN 19213 0
- Mounting thread 7/16-20 UNF to EN 61518 2
• Vent on side of process flange 3)
- Mounting thread M10 to DIN 19213 4
- Mounting thread 7/16-20 UNF to EN 61518 6
Non-wetted parts materials
Process flange screws Electronics housing
Stainless steel Die-cast aluminium 2
Stainless steel Stainless steel precision 3
casting
Design
• Standard design 1
• International version, English label inscriptions, 2
documentation in 5 languages on CD
Explosion protection
• without A
• with ATEX, Type of protection:
- "Intrinsic safety (EEx ia)" B
- "Explosion-proof (EEx d)" 4) D
- "Intrinsic safety and explosion-proof enclosure P
(EEx ia + EEx d)" 5)
- "n (zone 2)" E
- "Intrinsic safety, explosion-proof enclosure and R
dust explosion protection (EEx ia + EEx d +
zone 1D/2D)" 5)
• with FM + CSA, Type of protection:
- "Intrinsic safety and explosion-proof (is + xp)" 4) NC
Electrical connection / cable inlet
• Screwed gland Pg 13.5 6) A
• Screwed gland M20x1.5 B
• Screwed gland ½-14 NPT C
• Han 7D plug (plastic housing) incl. mating D
connector 6)
2
Further designs Order code Further designs Order code
Please add "-Z" to Order No. and specify Order code. Please add "-Z" to Order No. and specify Order code.
Pressure transmitter with mounting bracket made of: Additional data
• Steel A01 Measuring range to be set Y01
• Stainless steel A02 specify in plain text:
O-rings for process flanges Y01: ... up to ... mbar, bar, kPa, MPa, psi
(instead of FPM (Viton)) Measuring point number/identification Y15
• PTFE (Teflon) A20 max. 16 characters, specify in plain text:
• FEP (with silicone core, approved for food) A21 Y15: ...........................................
• FFPM (Kalrez, compound 4079) A22 Measuring point text Y16
• NBR (Buna N) A23 max. 27 characters, specify in plain text:
Plug Y16: ...........................................
• Han 7D (metal, gray) A30 Entry of HART address (TAG) Y17
• Han 8U (instead of Han 7D) A31 max. 27 characters, specify in plain text:
Sealing screws A40 Y17: ...........................................
¼-18 NPT, with valve in material of process flanges Setting of pressure indicator in pressure units Y21
Rating plate inscription specify in plain text (standard setting: mA):
Y21: mbar, bar, kPa, MPa, psi, ...
(instead of German)
Note:
• English B11 The following pressure units can be selected:
• French B12
bar, mbar, mm H2O*), inH2O*), ftH2O*), mmHG, inHG,
• Spanish B13 psi, Pa, kPa, MPa, g/cm2, kg/cm2, mA, Torr, ATM or %
• Italian B14 *) Reference temperature 20 °C
English rating plate B21 Setting of pressure indicator in non-pressure units Y22 + Y01
Pressure units in inH2O or psi specify in plain text:
Y22: ..... up to ..... l/min, m3/h, m, USgpm, ...
Manufacturer's test certificate M C11
(specification of measuring range in pressure units
(calibration certificate)
"Y01" is essential, unit with max. 5 characters)
to DIN 55350, Part 18 and to ISO 8402
Only the settings for "Y01", "Y21", "Y22" and "D05" can be made in the
Acceptance test certificate B C12 factory
to EN 10204-3.1.B
Factory certificate C14
to EN 10204-2.2
Certificate "Functional Safety (SIL)" C20
Setting of upper limit of output signal to 22.0 mA D05
Acid gas version to NACE D07
(only together with seal diaphragm made of Hastelloy)
Type of protection IP68 D12
(not together with Han 7D / Han 8U plug,
Pg 13.5 screwed gland)
Digital indicator along side the input keys D27
(only together with the devices 7MF4333-....0-.A6 or
-.A.7-Z, Y21 or Y22 + Y01)
Use in or at zone 1D/2D E01
(only together with type of protection "Intrinsic safety
(EEx ia)"
Use at zone 0 E02
(only together with type of protection "Intrinsic safety
(EEx ia)"
Oxygen application E10
(max. 160 bar (2320 psi) with oxygen measurement
and inert liquid)
Interchanging of process connection side H01
Vent on side for gas measurements H02
Process flange
• Hastelloy K01
• Monel K02
• Stainless steel with PVDF insert K04
max. PN 10 (MWP 145 psi),
Max. temperature of medium 90 °C (194 °F)
■ Technical specifications
2
Influence of ambient temperature
• With -10 ... +60 °C (14 ... 140 °F) ≤ (0.08 ⋅ r +0.1)%
SITRANS P pressure transmitters, DS III series, for differential pres-
sure and flow • With -40 ... -10 °C and ≤ (0.1 ⋅ r +0.15)% / 10 K
+60 °C ... +85 °C (≤ (0.1 ⋅ r +0.15)% / 18 °F)
Input (-40 ... +14 and 140 ... 185 °F) Twice the value with 20-mbar
Measured variable Differential pressure and flow (0.29 psi) measuring cell.
Span Maximum working pressure Rated conditions
• PN 32 (MWP 464 psi) Degree of protection (to EN 60529) IP65
- 1 ... 20 mbar (0.0145 ... 0.29 psi) 32 bar (464 psi) Process temperature
• PN 160 (MWP 2320 psi) • Measuring cell with silicone oil fil- -40 ... +100 °C (-40 ... +212 °F)
- 1 ... 60 mbar (0.0145 ... 0.87 psi) 160 bar (2320 psi) ling
- 2.5 ... 250 mbar 160 bar (2320 psi) • Measuring cell with inert filling li- -20 ... +100 °C (-4 ... +212 °F)
(0.036 ... 3.63 psi) quid
- 6 ... 600 mbar (0.087 ... 8.7 psi) 160 bar (2320 psi) • In conjunction with dust explosion -20 ... +60 °C (-4 ... +140 °F)
protection
- 16 ... 1600 mbar 160 bar (2320 psi)
(0.232 ... 23.2 psi) Design
- 50 ... 5000 mbar 160 bar (2320 psi) Weight (without options) ≈ 4.5 kg (≈ 9.9 lb)
(0.725 ... 72.5 psi) Wetted parts materials
- 0.3 ... 30 bar (4.35 ... 435 psi) 160 bar (2320 psi) • Seal diaphragm Stainless steel, mat. No.
• PN 420 (MWP 6092) 1.4404/316L, Hastelloy C276,
mat. No. 2.4819, Monel, mat. No.
- 2.5 ... 250 mbar 420 bar (6092 psi) 2.4360, tantalum or gold
(0.036 ... 3.63 psi)
Measuring cell filling Silicone oil or inert filling liquid
- 6 ... 600 mbar (0.087 ... 8.7 psi) 420 bar (6092 psi) (max. 160 bar (2320 psi) with oxy-
gen measurement)
- 16 ... 1600 mbar 420 bar (6092 psi)
(0.232 ... 23.3 psi) Process connection Female thread ¼-18 NPT and
flange connection to DIN 19 213
- 50 ... 5000 mbar 420 bar (6092 psi) with mounting thread M10 or
(0.725 ... 72.5 psi) 7/ -20 UNF to EN 61518
16
- 0.3 ... 30 bar (4.35 ... 435 psi) 420 bar (6092 psi) Power supply UH
Lower measuring limit Terminal voltage on pressure trans- 10.5 ... 45 V DC
• Measuring cell with silicone oil fil- -100% of max. span mitter 10.5 ... 30 V DC in intrinsically-
ling (-33% with 30 bar (435 psi) safe mode
measuring cell) or Certificates and approvals
30 mbar (0.44 psi) absolute
Classification according to pressure
Upper measuring limit 100% of max. span (max. 160 bar equipment directive
(2320 psi) with oxygen measure- (DRGL 97/23/EC)
ment and inert liquid)
• PN 32/160 (MWP 464/2320) For gases of fluid group 1 and
Output liquids of fluid group 1; complies
Output signal 4 ... 20 mA with requirements of article 3,
paragraph 3 (sound engineering
Measuring accuracy practice)
Reference conditions Increasing characteristic • PN 420 (MWP 6092) For gases of fluid group 1 and
Start-of-scale value 0 bar liquids of fluid group 1; complies
Stainless steel seal diaphragm with basic safety requirements of
Silicone oil filling article 3, paragraph 1 (appendix
Room temperature (25 °C (77 °F)) 1); assigned to category III, con-
r: Span ratio formity evaluation module H by
(r = max. span / set span) the TÜV Nord
Error in measurement and fixed- Explosion protection
point setting (including hysteresis
and repeatability) • Intrinsic safety "i" PTB 99 ATEX 2122
• Linear characteristic - Identification Ex II 1/2 G EEx ia/ib IIB/IIC T6
- r ≤ 10 ≤ (0.0029 ⋅ r + 0.071)% - Permissible ambient temperatu- -40 ... +85 °C (-40 ... +185 °F)
re temperature class T4;
- 10 < r ≤ 30 ≤ (0.0045 ⋅ r + 0.071)% -40 ... +70 °C (-40 ... +158 °F)
- 30 < r ≤ 100 ≤ (0.005 ⋅ r +0.05)% temperature class T5;
-40 ... +60 °C (-40 ... +140 °F)
• Square-root characteristic (flow temperature class T6
> 50%)
- Connection To certified intrinsically-safe cir-
- r ≤ 10 ≤ 0.1% cuits with maximum values:
Ui = 30 V, Ii = 100 mA,
- 10 < r ≤ 30 ≤ 0.2%
Pi = 750 mW; Ri = 300 Ω
• Square-root characteristic (flow
- Effective internal inductance/ca- L i = 0.4 mH, Ci = 6 nF
25 ... 50%)
pacitance
- r ≤ 10 ≤ 0.2%
- 10 < r ≤ 30 ≤ 0.4%
2
• Explosion-proof "d" PTB 99 ATEX 1160
- Identification Ex II 1/2 G EEx d IIC T4/T6
- Permissible ambient temperatu- -40 ... +85 °C (-40 ... +185 °F)
re temperature class T4;
-40 ... +60 °C (-40 ... +140 °F)
temperature class T6
- Connection To circuits with values:
UH = 10.5 ... 45 V DC
• Dust explosion protection for PTB 01 ATEX 2055
zone 20
- Identification Ex II 1 D IP65 T 120 °C
Ex II 1/2 D IP65 T 120 °C
- Permissible ambient temperatu- -40 ... +85 °C (-40 ... +185 °F)
re
- Max.surface temperature 120 °C (248 °F)
- Connection To certified intrinsically-safe cir-
cuits with maximum values:
Ui = 30 V, Ii = 100 mA,
Pi = 750 mW, Ri = 300 Ω
- Effective internal inductance/ca- Li = 0.4 mH, Ci = 6 nF
pacitance
• Dust explosion protection for zone PTB 01 ATEX 2055
21/22
- Identification Ex II 2 D IP65 T 120 °C
- Connection To circuits with values:
UH = 10.5 ... 45 V DC;
Pmax = 1.2 W
• Type of protection "n" (zone 2) TÜV 01 ATEX 1696 X
- Identification Ex II 3 G EEx nA L IIC T4/T5/T6
• Explosion protection to FM Certificate of Compliance
3008490
- Identification (XP/DIP) or (IS); CL I, DIV 1, GP ABCD T4...T6;
(NI) CL II, DIV 1, GP EFG; CL III; CL I,
ZN 0/1 AEx ia IIC T4...T6; CL I,
DIV 2, GP ABCD T4...T6; CL II,
DIV 2, GP FG; CL III
• Explosion protection to CSA Certificate of Compliance
1153651
- Identification (XP/DIP) or (IS) CL I, DIV 1, GP ABCD T4...T6;
CL II, DIV 1, GP EFG; CL III; Ex ia
IIC T4...T6; CL I, DIV 2, GP ABCD
T4...T6; CL II, DIV 2, GP FG; CL III
■ Dimensional drawings
2
15
(0.6) 138 (5.4) approx. 30 (1.2) 5) 100 (3.94)
50 6 2
(1.97)
(0.8)
20
(1.97)
50
(3.15)
Ø 80
117 (4.6)
3
198 (7.8)
8
1
of housing
134 (5.27)
96 (3.8)
11
ca. 96 (3.78) 52 (2.05)
72 (2.83)
105 (4.1) 68 (2.7)
120 (4.7)
1 Process connection: ¼-18 NPT (EN 61518)
2 Blanking plug
3 Electrical connection:
- screwed gland Pg 13,5 (adapter) 2) 3),
- screwed gland M20x1,5 3),
- screwed gland ½-14 NPT or
- Han 7D/ Han 8U plug 2) 3)
4 Terminal side
5 Electronics side, digital display (longer overall length for cover with window)
6 Protective cover over keys
7 Mounting bracket (option)
8 Sealing screw with valve (option) 1) Allow approx. 20 mm (0.79 inch) thread length to permit unscrewing
9 Screw cover - safety bracket (only for type of protection 2) Not with type of protection "explosion-proof enclosure"
"Explosion-proof enclosure", not shown in the drawing) 3) Not with type of protection "FM + CSA [is + xp]"
10 Lateral venting for liquid measurement 4) 92 mm (3.62 inch) for minimum distance to permit rotation with indicator
11 Lateral venting for gas measurement (suffix H02) 5) 45 mm (1.8 inch) for Pg 13,5 with adapter
SITRANS P pressure transmitters, DS III series for differential pressure and flow, dimensional drawing, dimensions in mm (inch)
2 15
(0.6) 138 (5.4)
approx. 30 (1.18)8)
100 (3.94)
50 ca. 96 (3.78) 6 2
69 (2.7) (1.97)
(0.8)
20
(1.97)
50
(3.15)
Ø 80
3 3
4 1)
ca. 85 (3.35)6)
+ 8
8
SITRANS P pressure transmitters, DS III series for differential pressure and flow, with process covers for vertical differential pressure lines, dimensional
drawing, dimensions in mm (inch)
7
4 6 2
183 (7.2)
115 (4.5) 1
77
(3.0)
2
193 (7.6)
TERMINALS
FIELD 1 Process connection, ¼-18 NPT (EN 61 518)
2 Terminal side, analog indicator (optional)
3 External digital indicator (optional)
4 Sealing plug
5 Fixing screw
6 Blanking plug
7 Protective cover for keys
3 8 Electrical connection
SITRANS P pressure transmitters, DS III series for differential pressure and flow, with digital indicator beside control keys, dimensional drawing, dimen-
sions in mm (inch)
2
Selection and Ordering data Order No. Selection and Ordering data Order No.
SITRANS P pressure transmitters, } 7MF 4 4 3 3 - SITRANS P pressure transmitters, } 7MF 4 4 3 3 -
DS III series for differential pressure and flow DS III series for differential pressure and flow
77777 - 7777 77777 - 7777
PN 32/160 (MWP 464/2320 psi) PN 32/160 (MWP 464/2320 psi)
2-wire system 2-wire system
Measuring cell filling Measuring cell Display
cleaning • without (digital indicator hidden, setting: mA) } 1
Silicone oil standard } 1 • with visible digital indicator 6
Inert liquid1) grease-free 3 • with customer-specific digital indicator (setting 7
Measured span as specified, Order code "Y21" required)
PN 32 (MWP 464 psi) } Available ex stock
1 ... 20 mbar 2) (0.0145 ... 0.29 psi) } B Power supply units see "SITRANS I power supply units and input isola-
PN 160 (MWP 2320 psi) tors".
1 ... 60 mbar (0.0145 ... 0.87 psi) } C Included in delivery of the device:
2.5 ... 250 mbar (0.036 ... 3.63 psi) } D • Brief instructions (Leporello)
6 ... 600 mbar (0.087 ... 8.70 psi) } E • CD-ROM with detailed documentation
16 ... 1600 mbar (0.232 ... 23.2 psi) } F • Sealing plug(s) or sealing screw(s) for the process flanges(s)
50 ... 5000 mbar (0.725 ... 72.5 psi) } G 1) For oxygen application, add Order code E10.
0.3 ... 30 bar (4.35 ... 435 psi) } H 2) Not suitable for connection of remote seal
Wetted parts materials 3) Only together with max. spans 250, 1600, 5000 and 30000 mbar
(3.63, 23.2, 72.5 and 435 psi).
(stainless steel process flanges) 4) Without cable gland, with blanking plug
Seal diaphragm Parts of measuring cell 5) With enclosed cable gland EEx ia and blanking plug
Stainless steel Stainless steel } A 6) Not together with type of protection "Explosion-proof"
Hastelloy Stainless steel B
Hastelloy Hastelloy C
Tantalum 3) Tantalum E
Monel 3) Monel H
Gold 3) Gold L
Version for diaphragm seal Y
Process connection
Female thread ¼-18 NPT with flange connection
• Sealing screw opposite process connection
- Mounting thread M10 to DIN 19213 } 0
- Mounting thread 7/16-20 UNF to EN 61518 2
• Vent on side of process flange 2)
- Mounting thread M10 to DIN 19213 4
- Mounting thread 7/16-20 UNF to EN 61518 6
Non-wetted parts materials
Process flange screws Electronics housing
Stainless steel Die-cast aluminium } 2
Stainless steel Stainless steel precision 3
casting
Design
• Standard design 1
• International version, English label inscriptions, } 2
documentation in 5 languages on CD
Explosion protection
• without A
• with ATEX, Type of protection:
- "Intrinsic safety (EEx ia)" B
- "Explosion-proof (EEx d)" 4) D
- "Intrinsic safety and explosion-proof enclosure } P
(EEx ia + EEx d)" 5)
- "n (zone 2)" E
- "Intrinsic safety, explosion-proof enclosure and R
dust explosion protection (EEx ia + EEx d +
zone 1D/2D)" 5)
• with FM + CSA, Type of protection:
- "Intrinsic safety and explosion-proof (is + xp)" 4) NC
Electrical connection / cable inlet
• Screwed gland Pg 13.5 6) A
• Screwed gland M20x1.5 } B
• Screwed gland ½-14 NPT C
• Han 7D plug (plastic housing) incl. mating D
connector 6)
2
Further designs Order code Further designs Order code
Please add "-Z" to Order No. and specify Order code. Please add "-Z" to Order No. and specify Order code.
Pressure transmitter with mounting bracket made of: Oxygen application E10
• Steel A01 (max. 160 bar (2320 psi) with oxygen measurement
• Stainless steel A02 and inert liquid)
O-rings for process flanges Interchanging of process connection side H01
(instead of FPM (Viton)) Vent on side for gas measurements H02
• PTFE (Teflon) A20
Stainless steel process flanges for vertical H03
• FEP (with silicone core, approved for food) A21 differential pressure lines
• FFPM (Kalrez, compound 4079) A22 (not together with K01, K02 and K04) 1)
• NBR (Buna N) A23
Process flange
Plug • Hastelloy K01
• Han 7D (metal, gray) A30 • Monel K02
• Han 8U (instead of Han 7D) A31 • Stainless steel with PVDF insert K04
Sealing screws Max. PN 10 (MWP 145 psi)
¼-18 NPT, with valve in material of process flanges A40 Max. temperature of medium 90 °C (194 °F)
2
Selection and Ordering data Order No. Selection and Ordering data Order No.
SITRANS P pressure transmitters, DS III series 7MF 4 5 3 3 - SITRANS P pressure transmitters, DS III series 7MF 4 5 3 3 -
for differential pressure and flow for differential pressure and flow
77777 - 7777 77777 - 7777
PN 420 (MWP 6092 psi) PN 420 (MWP 6092 psi)
2-wire system 2-wire system
2
Further designs Order code Further designs Order code
Please add "-Z" to Order No. and specify Please add "-Z" to Order No. and specify
Order code. Order code.
Pressure transmitter with mounting bracket made of: Additional data
• Steel A01 Measuring range to be set
• Stainless steel A02 specify in plain text:
O-rings for process flanges • with linear characteristic: Y01
(instead of FPM (Viton)) Y01: ... up to ... mbar, bar, kPa, MPa, psi
• PTFE (Teflon) A20 • with square-root characteristic: Y02
• FEP (with silicone core, approved for food) A21 Y01: ... up to ... mbar, bar, kPa, MPa, psi
• FFPM (Kalrez, compound 4079) A22 Measuring point number/identification Y15
• NBR (Buna N) A23 max. 16 characters, specify in plain text:
Y15: ...........................................
Plug
• Han 7D (metal, gray) A30 Measuring point text Y16
• Han 8U (instead of Han 7D) A31 max. 27 characters, specify in plain text:
Y16: ...........................................
Sealing screws A40
¼-18 NPT, with valve in material of process flanges Entry of HART address (TAG) Y17
max. 27 characters, specify in plain text:
Rating plate inscription Y17: ...........................................
(instead of German)
Setting of pressure indicator in pressure units Y21
• English B11
specify in plain text (standard setting: mA):
• French B12 Y21: mbar, bar, kPa, MPa, psi, ...
• Spanish B13 Note:
• Italian B14 The following pressure units can be selected:
English rating plate B21 bar, mbar, mm H2O*), inH2O*), ftH2O*), mmHG, inHG,
Pressure units in inH2O or psi psi, Pa, kPa, MPa, g/cm2, kg/cm2, mA, Torr, ATM or %
*) Reference temperature 20 °C
Manufacturer's test certificate M C11
(calibration certificate) Setting of pressure indicator in non-pressure units Y22 +
specify in plain text: Y01 or Y02
to DIN 55350, Part 18 and to ISO 8402
Y22: ..... up to ..... l/min, m3/h, m, USgpm, ...
Acceptance test certificate B C12 (specification of measuring range in pressure units
to EN 10204-3.1.B "Y01" or "Y02" is essential, unit with max. 5 characters)
Factory certificate C14 Only the settings for "Y01", "Y21", "Y22" and "D05" can be made in the
to EN 10204-2.2 factory
Certificate "Functional Safety (SIL)" C20
Setting of upper limit of output signal to D05
22.0 mA
Acid gas version to NACE D07
(only together with seal diaphragm made of Hastelloy)
Type of protection IP68 D12
(not together with Han 7D / Han 8U plug,
Pg 13.5 screwed gland)
Digital indicator along side the input keys D27
(only together with the devices 7MF4533-....2-.A.6 or
-.A.7-Z, Y21 or Y22 + Y01)
Use in or at zone 1D/2D E01
(only together with type of protection "Intrinsic safety
(EEx ia)"
Use at zone 0 E02
(only together with type of protection "Intrinsic safety
(EEx ia)"
Interchanging of process connection side H01
Stainless steel process flanges for vertical H03
differential pressure lines
■ Technical specifications
2
Measuring cell with silicone oil filling
- High-pressure side pabs ≥ 1 bar: -40 ... +175 °C
SITRANS P pressure transmitters, DS III series, for level (-40 ... +347 °F)
Input pabs < 1 bar: -40 ... +80 °C
Measured variable Level (-40 ... +176 °F)
Span Maximum working pressure - Low-pressure side -40 ... +100 °C (-40 ... +212 °F)
(nominal pressure) -20 ... +60 °C (-4 ... +140 °F) in
conjunction with dust explosion
• 25 ... 250 mbar (0.36 ... 3.63 psi) See "Mounting flange" protection
• 25 ... 600 mbar (0.36 ... 8.7 psi) Design
• 53 ... 1600 mbar (0.77 ...23.2 psi) Weight (without options)
• 160 ... 5000 mbar • To DIN (pressure transmitter with ≈ 11 ... 13 kg (≈ 24.2 ... 28.7 lb)
(2.32 ... 72.5 psi) mounting flange, without tube)
Lower measuring limit (measuring -100% of max. span or 30 mbar • To ANSI (pressure transmitter with ≈ 11 ... 18 kg (≈ 24.2 ... 39.7 lb)
cell with silicone oil filling) (0.435 psi) abs., depending on mounting flange, without tube)
mounting flange
Wetted parts materials
Upper measuring limit 100% of max. span
• High-pressure side: Seal dia- Stainless steel, mat. No.
Output phragm of mounting flange 1.4404/316L,
Output signal 4 ... 20 mA Monel 400, mat. No. 2.4360,
Hastelloy B2, mat. No. 2.4617,
Measuring accuracy Hastelloy C276, mat. No. 2.4819,
Reference conditions Increasing characteristic Hastelloy C4, mat. No. 2.4610,
Start-of-scale value 0 bar tantalum, PTFE, ECTFE
Stainless steel seal diaphragm Measuring cell filling Silicone oil
Mounting flange without tube
Silicone oil filling Process connection
Room temperature (25 °C (77 °F)) • High-pressure side Flange to DIN and ANSI
r: Span ratio
(r = max. span / set span) • Low-pressure side Female thread ¼-18 NPT and
flange connection to DIN 19213
Error in measurement and fixed- with mounting thread M10 or
point setting (including hysteresis 7/ -20 UNF to EN 61518
16
and repeatability)
Power supply UH
• Linear characteristic
Terminal voltage on pressure trans- 10.5 ... 45 V DC
- r ≤ 10 ≤ 0.15% mitter 10.5 ... 30 V DC in intrinsically-
- 10 < r ≤ 30 ≤ 0.3% safe mode
- 30 < r ≤ 100 (0.0075 ⋅ r + 0.075)% Certificates and approvals
Influence of ambient temperature Classification according to pressure For gases of fluid group 1 and
equipment directive liquids of fluid group 1; complies
• With -10 ... +60 °C (14 ... 140 °F) (DRGL 97/23/EC) with requirements of article 3,
- 250-mbar (3.63 psi) measuring ≤ (0.5 ⋅ r + 0.2)% (0.4 instead of paragraph 3 (sound engineering
cell 0.2 with 10 < r ≤ 30) practice)
- 600-mbar (8.7 psi) measuring ≤ (0.3 ⋅ r + 0.2)% (0.4 instead of Explosion protection
cell 0.2 with 10 < r ≤ 30)
• Intrinsic safety "i" PTB 99 ATEX 2122
- 1,600 and 5,000 mbar (23.2 and ≤ (0.25 ⋅ r + 0.2)% (0.4 instead of
0.2 with 10 < r ≤ 30) - Identification Ex II 1/2 G EEx ia/ib IIB/IIC T6
72.5 psi) measuring cells
- Permissible ambient temperatu- -40 ... +85 °C (-40 ... +185 °F)
• With -40 ... -10 °C and re temperature class T4;
+60 °C ... +85 °C -40 ... +70 °C (-40 ... +158 °F)
(-40 ... +14 and 140 ... 185 °F) temperature class T5;
- 250-mbar (3.63 psi) measuring ≤ (0.25 ⋅ r +0.15)% / 10 K -40 ... +60 °C (-40 ... +140 °F)
cell (≤ (0.25 ⋅ r +0.15)% / 18 °F) temperature class T6
(Twice the value for 10 < r ≤ 30) - Connection To certified intrinsically-safe cir-
- 600-mbar (8.7 psi) measuring ≤ (0.15 ⋅ r +0.15)% / 10 K cuits with maximum values:
cell (≤ (0.15 ⋅ r +0.15)% / 18 °F) Ui = 30 V, Ii = 100 mA,
(Twice the value for 10 < r ≤ 30) Pi = 750 mW; Ri = 300 Ω
- 1,600 and 5,000 mbar (23.2 and ≤ (0.12 ⋅ r +0.15)% / 10 K - Effective internal inductance/ca- Li = 0.4 mH, Ci = 6 nF
72.5 psi) measuring cells (≤ (0.12 ⋅ r +0.15)% / 18 °F) pacitance
(Twice the value for 10 < r ≤ 30) • Explosion-proof "d" PTB 99 ATEX 1160
Rated conditions - Identification Ex II 1/2 G EEx d IIC T4/T6
Degree of protection (to EN 60529) IP65 - Permissible ambient temperatu- -40 ... +85 °C (-40 ... +185 °F)
Process temperature Note: Note the assignment of the re temperature class T4;
max. permissible operating tem- -40 ... +60 °C (-40 ... +140 °F)
perature to the max. permissible temperature class T6
operating pressure of the respec- - Connection To circuits with values:
tive flange connection! UH = 10.5 ... 45 V DC
2
• Dust explosion protection for PTB 01 ATEX 2055 Mounting flange
zone 20
Nom. diam. Nom. press.
- Identification Ex II 1 D IP65 T 120 °C
Ex II 1/2 D IP65 T 120 °C • To EN 1092-1
- Permissible ambient temperatu- -40 ... +85 °C (-40 ... +185 °F) - DN 80 PN 40
re - DN 100 PN 16
- Max.surface temperature 120 °C (248 °F) PN 40
- Connection To certified intrinsically-safe cir- • To ASME B16.5
cuits with maximum values: - 3 inch class 150
Ui = 30 V, Ii = 100 mA, class 300
Pi = 750 mW, Ri = 300 Ω
- 4 inch class 150
- Effective internal inductance/ca- Li = 0.4 mH, Ci = 6 nF class 300
pacitance
• Dust explosion protection for zone PTB 01 ATEX 2055
21/22
- Identification Ex II 2 D IP65 T 120 °C
- Connection To circuits with values:
UH = 10.5 ... 45 V DC;
Pmax = 1.2 W
• Type of protection "n" (zone 2) TÜV 01 ATEX 1696 X
- Identification Ex II 3 G EEx nA L IIC T4/T5/T6
• Explosion protection to FM Certificate of Compliance
3008490
- Identification (XP/DIP) or (IS); CL I, DIV 1, GP ABCD T4...T6;
(NI) CL II, DIV 1, GP EFG; CL III; CL I,
ZN 0/1 AEx ia IIC T4...T6; CL I,
DIV 2, GP ABCD T4...T6; CL II,
DIV 2, GP FG; CL III
• Explosion protection to CSA Certificate of Compliance
1153651
- Identification (XP/DIP) or (IS) CL I, DIV 1, GP ABCD T4...T6;
CL II, DIV 1, GP EFG; CL III; Ex ia
IIC T4...T6; CL I, DIV 2, GP ABCD
T4...T6; CL II, DIV 2, GP FG; CL III
■ Dimensional drawings
2
approx. 30 (1.18)5)
185
15
100 (3.94) 138 (5.4) (0.6)
6 50
(1.97) 69 (2.72)
2
3
51)
165 (6.5)
L 77 (3)4)
9
d4
d5
D
54 (2.1) 108 (4.25) k
3 1
2
Selection and Ordering data Order No. Further designs Order code
SITRANS P pressure transmitters, DS III series 7MF 4 6 3 3 - Please add "-Z" to Order No. and specify Order code.
for level
O-rings for process flanges
2-wire system 77 Y77 - 7777 (instead of FPM (Viton))
Measuring cell filling Measuring cell cleaning • PTFE (Teflon) A20
Silicone oil Standard 1 • FEP (with silicone core, approved for food) A21
• FFPM (Kalrez, compound 4079) A22
Measured span
• NBR (Buna N) A23
25 ... 250 mbar (0.363 ... 3.63 psi) D
25 ... 600 mbar (0.363 ... 8.70 psi) E Plug
53 ... 1600 mbar (0.77 ... 23.2 psi) F • Han 7D (metal, gray) A30
0.16 ... 5 bar (2.32 ... 72.5 psi) G • Han 8U (instead of Han 7D) A31
Process connection of low-pressure side Sealing screws A40
Female thread ¼-18 NPT with flange connection ¼-18 NPT, with valve in material of process flanges
• Mounting thread M10 to DIN 19213 0 Rating plate inscription
• Mounting thread 7/16-20 UNF to EN 61518 2 (instead of German)
Non-wetted parts materials • English B11
Process flange screws Electronics housing • French B12
• Spanish B13
Stainless steel Die-cast aluminium 2
• Italian B14
Stainless steel Stainless steel precision 3
casting English rating plate B21
Design Pressure units in inH2O or psi
• Standard design 1 Manufacturer's test certificate M C11
• International version, English label inscriptions, 2 (calibration certificate)
documentation in 5 languages on CD to DIN 55350, Part 18 and to ISO 8402
Explosion protection Acceptance test certificate B C12
• without A to EN 10204-3.1.B
• with ATEX, Type of protection: Factory certificate C14
- "Intrinsic safety (EEx ia)" B to EN 10204-2.2
- "Explosion-proof (EEx d)" 1) D
- "Intrinsic safety and explosion-proof enclosure P Certificate "Functional Safety (SIL)" C20
(EEx ia + EEx d)" 2) Setting of upper limit of output signal to 22.0 mA D05
- "n (zone 2)" E
Type of protection IP68 D12
- "Intrinsic safety, explosion-proof enclosure and R
dust explosion protection (EEx ia + EEx d + (not together with Han 7D / Han 8U plug,
Pg 13.5 screwed gland)
zone 1D/2D)" 2)
• with FM + CSA, Type of protection: Use in or at zone 1D/2D E01
- "Intrinsic safety and explosion-proof (is + xp)" 1) NC (only together with type of protection "Intrinsic safety
(EEx ia)"
Electrical connection / cable inlet
Use at zone 0 E02
• Screwed gland Pg 13.5 3) A
(only together with type of protection "Intrinsic safety
• Screwed gland M20x1.5 B (EEx ia)"
• Screwed gland ½-14 NPT C
• Han 7D plug (plastic housing) incl. mating D Overfilling safety device for flammable and E08
connector 3) non-flammable liquids
(max. PN 32, only together with type of protection
Display "Intrinsic safety (EEx ia)"), to WHG and VbF
• without (digital indicator hidden, setting: mA) 1
Interchanging of process connection side H01
• with visible digital indicator 6
• with customer-specific digital indicator (setting 7
as specified, Order code "Y21" required)
Ordering information:
1st order item: Pressure transmitter 7MF4633-...
2nd order item: Mounting flange 7MF4912-3...
Example of ordering:
Item line 1: 7MF4633-1EY20-1AA1-Z
B line: Y01
C line: Y01: 80 to 143 mbar (1.16 to 2.1 psi)
Item line 2: 7MF4912-3GE01
Power supply units see "SITRANS I power supply units and input isola-
tors".
Included in delivery of the device:
• Brief instructions (Leporello)
• CD-ROM with detailed documentation
• Sealing plug(s) or sealing screw(s) for the process flanges(s)
2
Further designs Order code Selection and Ordering data Order No. Ord. code
Please add "-Z" to Order No. and specify Order code. Mounting flange 7MF 4 9 1 2 -
Additional data directly fitted to pressure transmitter 3 7777 777
SITRANS P (converter part) for level,
Measuring range to be set Y01 for DS III series
specify in plain text:
Y01: ... up to ... mbar, bar, kPa, MPa, psi Wetted parts materials
• Stainless steel 316L A
Measuring point number/identification Y15
- Coated with PFA 1) D
max. 16 characters, specify in plain text:
Y15: ........................................... - Coated with PTFE 1) E0
- Coated with ECTFE 1) F
Measuring point text Y16
max. 27 characters, specify in plain text: • Monel 400, mat. No. 2.4360 G
Y16: ........................................... • Hastelloy B2, mat. No. 2.4617 H
• Hastelloy C276, mat. No. 2.4819 J
Entry of HART address (TAG) Y17
• Hastelloy C4, mat. No. 2.4610 U
max. 8 characters, specify in plain text:
Y17: ........................................... • Tantalum K