12 - Objective Refraction
12 - Objective Refraction
AUTHOR (S)
Pirindhavellie Govender : University of KwaZulu Natal (UKZN) Durban, South Africa
Retinoscopes
Retinoscopy
INTRODUCTION
Objective refraction comprises a host of various techniques that may be employed to determine the patient’s refractive
correction. It is usually performed for the determination of the starting point of a subjective refraction. More importantly,
it is a technique which the practitioner has to rely on entirely when a subjective refraction cannot be ascertained. This
would be in cases of the patient malingering (i.e. feigning poor/better acuity than they actually have), uncooperative
patients like children, patients who are unable to communicate subjective responses to you and in patients who are
unreliable in terms of subjective responses. It is a procedure that becomes more accurate with a greater amount of
practice. Both retinoscopy and autorefraction are methods of objective refraction. In many cases, autorefraction is
used to replace retinoscopy, however, issues of instrument reliability, problems inherent due to the nearness of the
target in the instrument, and poor cooperation on the part of patients makes retinoscopy a far more superior and more
adaptable technique to perform than autorefraction.
Retinoscopy is a technique that is used to objectively determine the refractive error of a patient. It does not require the
patient’s responses and therefore can be performed on children and non-verbal individuals.
RETINOSCOPES
The first retinoscopes used a light source placed just behind the patient’s shoulder while modern retinoscopes have
the light source built into them (i.e. they are self-luminous retinoscopes). Retinoscopes allow us to shine a light into a
person’s eye and look at the light reflected back from their retina. This reflected light as seen in the person’s pupil is
called the retinoscopic reflex or “ret reflex”.
Objective Refraction
ADVANTAGES OF RETINOSCOPES
SPOT RETINOSCOPES
1. Astigmatism can be detected with spot retinoscopy by noting the shape of the ’ret reflex’. In an uncooperative
patient, it may not be possible to achieve prolonged viewing if the patient’s fixation is not steady, and the
examiner has to depend on occasional glimpses for assessment of the reflex. In streak retinoscopy the reflex
is always a slit, regardless of the presence of astigmatism, whereas in spot retinoscopy significant astigmatism
will make the reflex appear elliptical. The correct axis of the astigmatic correction is more rapidly determined
than with spot retinoscopy.
2. The time it takes for the examiner to rotate the streak from one position to another, an uncooperative patient
may change their fixation position to one requiring more or less accommodation. This might be interpreted as
astigmatism. With the spot retinoscope, the eye with no astigmatism (or corrected astigmatism) will always
return a circular reflex, even though the speed and/or direction may change.
STREAK RETINOSCOPES
1. The end point of retinoscopy is sometimes easier to observe with streaks than with spot retinoscopes.
2. All modern streak retinoscopes have an adjustment for changing the retinoscope beam vergence from being
divergent to convergent. This is not always available on spot retinoscopes.
Apart from the problems with uncooperative patients noted above, the choice of “spot” or “streak” is a matter
of individual preference.
PROJECTION SYSTEM
This part of the sytem illuminates the retina and comprises the following parts:
Light source, i.e. a small bulb that projects light onto the retina (RPE and choroid)
Condensing lens which lies in the path of the light projected from the bulb to focus the light onto the mirror
Mirror which is placed within the head of the instrument. It bends the light at right angles to the axis of the handle
so that the light is projected from the head of the instrument
Focusing sleeve which varies the distance between the bulb and the lens so that the light projected from the
retinoscope either diverges (plane mirror effect) or converges (concave mirror effect)
Current source or rechargeable/replaceable battery in the retinoscope handle
OBSERVATION SYSTEM
This part of the retinoscopic optical system allows the practitioner to view the retinal reflex. The light reflected off the
retina passes through an aperture in the mirror and out through the sight hole at the rear of the head. The light that is
reflected from the retina are acted upon by the optical components of the eye and therefore the observation of this
reflected light provides the practitioner information about the optics of the patient’s eye
In most cases, the practitioner will have the sleeve in the down position (divergent beam/plane mirror); however,
convergent beam or concave mirror retinoscopy may be preferred when:
The practitioner is unsure if the movement being observed is either “with” or “against”. By changing from divergent
to the convergent beam, the movement should change direction of the movement. This would make it easier to
decide if there really is movement present (“with” will become “against” and “against” will become “with”).
June 2012, Version 1 Clinical Optometric Procedures 1, 12-3
Objective Refraction
Convergent beam allows more light to enter the eye, making it easier to perform retinoscopy on patients with small
pupils or opacities in the media.
Many practitioners favor the “with” movement as opposed to the “against” movement since it is easier to perceive.
This is especially the case in high ametropia where the reflex is very dull and the “against” movement is difficult to
discern. Practitioners therefore slide the sleeve up to the concave or convergent position so that the dull against
movement becomes a distinct slow with movement which is now easier to neutralize.
While the practitioner should take cognizance of these changes in the direction of the movement of the “ret reflex”,
he must keep in mind that the refractive correction is the same, the refractive state of the patient is the same, it is
only the direction of the ret reflex that is different.
RETINOSCOPY
TYPES OF RETINOSCOPY
STATIC RETINOSCOPY
A technique which allows the determination of the refractive error for distance while maintaining accommodation in a
fixed state. Patient fixation is maintained at a distance of 6m.
DYNAMIC RETINOSCOPY
A technique that determinesthe accommodative response at near, while not having control over accommodation. This
method of retinoscopy investigates the patient’s accommodative response. Patient fixation is less than 6m.
MOHINDRA RETINOSCOPY
A technique that determines the distance refractive error when patient fixation is more difficult to maintain. This
method encourages the patient to fixate the light source.
RADICAL RETINOSCOPY
A variation of static retinoscopy that is employed especially in cases where media opacities are present and in cases
of very high refractive error.
For the purposes of this module, only static retinoscopy will be discussed, however, this and other methods of
retinoscopy will be discussed in other specialised modules such as low vision.
Rays of light from a normal (emmetropic) eye are reflected off the retina (from the retinoscope) and exit the pupil
as parallel rays. If the retinoscope is held at a distance of one meter from the patient's eye, then a +1.00 D
spherical ('working lens") is required to focus the parallel light rays at a one-meter distance. This is known as
making the patient’s eye conjugate with the optometrist's eye. Similarly, working at any distance
can be compensated for with an appropriate working distance lens. For example, if refracting at a distance of
50cm, a +2.00D spherical working lens is needed.The aim of retinoscopy is to find the lens that would produce a
neutral reflex. This neutral reflex is one that appears stationery and fills the pupil. However, it must be noted that
this reflex is at an infinite speed making it appear to be stationery. Neutrality is achieved by adding lenses to
make the reflex brighter, move faster, and ultimately neutralize the movement.
While autorefractors have become quite accurate, an experienced retinoscopist can achieve the same degree of
accuracy. Performing retinoscopy also allows the examiner to view the quality of the optical medium of the eye,
picking up cases of keratoconus and providing information that the autorefractor cannot.
The practitioner will observe the streak reflected from the rim of a trial lens held in front of a patient’s eye. Through the
trial lens and pupil a narrower band of light being reflected from the person’s retina will be observed. This is the ’ret
reflex’ (Fig. 12.3).
When you move the retinoscope (spot or streak), the ‘ret reflex’ moves. The movements of the ret reflex may be
‘‘with’’, ‘’against’’ and neutral. This reflex is not always positioned in the principal meridians but can appear in oblique
orientations as well (Fig. 12.4).
WITH MOTION
‘’With’’ motion, the ‘ret reflex’ moves in the same direction as the sweeping of the retinoscope streak. Figure 12.5
shows the streak moving across the rim of the trial lens from your left to right. If the ret reflex moves in the same
direction as the streak projected from the trial lens rim, this is termed “with” motion.
Figure 12.5 View of with motion through streak retinoscope sight hole
AGAINST MOTION
In an ‘’against’’ motion, if the streak moves from left to right, the ret reflex will move from right to left (opposite
direction). The ret reflex is moving in the opposite direction as the streak projects from the trial lens rim (Fig. 12.6).
This is an “against” movement.
Figure 12.6 View of against motion through streak retinoscope sight hole
ASTIGMATIC REFLEX
As previously said, astigmatic reflex is characterised by different movements in different meridians. These meridians
can be in any direction (horizontal, vertical, oblique) but are usually perpendicular to one another. An astigmatic reflex
present with several types of movements: (if neutralizing with minus cylinders and using the divergent beam)
Both meridians presenting with ‘’against’’ motion. The practitioner will neutralize the meridian with faster motion
first.
Both meridians presenting with ‘’with’’ motion. The practitioner will neutralize the meridian with slower motion first.
If you find it difficult identifying the difference in the speed of the movements, check both meridians with each
addition of lens, until one gets neutralised. The remaining meridian should have an ‘’against’’ movement which
you will now neutralize with minus cylinders with the axis parallel to the streak. If instead you observe a “with”
movement, neutralize this meridian with plus spherical lenses. Revert to the first meridian which will now show an
“against” movement and neutralize it with minus cylinders.
Example 1; You observe the reflexes along each of the meridians when conducting retinoscopy and notice that in
the vertical meridian the reflex moves with the movement of the retinoscope and in the horizontal meridian, the
reflex moves against the movement of the retinoscope. The following scenario presents itself during
neutralisation:
Example 2: You observe the reflexes along each of the meridians when conducting retinoscopy and notice that in
the vertical meridian the reflex moves with the movement of the retinoscope and in the horizontal meridian, the
reflex moves with the movement of the retinoscope. You start to neutralise the slower with movement. The
scenario presents itself during neutralisation;
Figure 12.8 The pathway of the retinoscopes light and the examiner’s visual axis coincide
Set the trial frame to the correct PD and make sure it fits the patient comfortably. This ensures that the lenses
fitted in the trial frame have their optical center coincident with the patient’s visual axis. The vertex distance may
also be set on the trial frame to the standard employed in your country. In many cases it can range from 12 to
14mm. This measurement is especially important when taking into consideration high prescriptions where even
slight adjustments of the vertex distance can produce a resultant difference in the prescription being offered
to the patient. This can result in problems when patients are finally prescribed their spectacles and they are
unable to view through them unlike through the trial frame.
PROCEDURE
1. Static retinoscopy is a binocular technique in which both eyes may be fogged with the working distance lenses
or the technique can be performed without the working distance lenses. However, adjustments (subtraction of
plus power appropriate to working distance used) must be made to the final Rx. Each method has its inherent
advantages.
The advantage of using this method is that “with” movements indicates hyperopia and “against” movements
indicates myopia (while using divergent beam). In addition, the fogging lenses are able to fully relax the
patient’s accommodation, especially in cases of low hyperopia.
The advantage of this technique is that it frees up trial lens spaces in the trial frame. It also prevents additional
reflection surfaces from the working distance lenses and the correcting lenses.
The working distance lens is determined by the dioptric equivalent of the working distance (FW). For example, if the practitioner used a
working distance of 50cm (fW), then the appropriate working distance lens is given by:
One must make note that there are set lenses when working with phoropters. For example, most phoropters have a working distance lens
of +1.50DS, indicating that the practitioner needs to perform their retinoscopy at a working distance of 67cm.
2. The room illumination should be dim, otherwise it will be difficult to observe the “ret reflex”. If the room is
too dim, it will be difficult for the practitioner to see what he/she is doing. The patient’s pupil will be larger
and therefore the examiner would be faced with more aberrations as a result of the larger pupil diameter.
3. Check that the retinoscopic beam is divergent by moving the sleeve to its lowest position. That is, closest
to the battery handle.
Start with the streak of the retinoscope orientated vertically. Observe the position of the reflex within
the eye. In cases where the principle meridians lie at 90 and 180 degree, this vertical streak will coincide
with one of the principal meridians. The streak is then orientated horizontally to observe the reflex in this
meridian (Fig. 12.9a). However, in cases where they don’t coincide (Fig. 12.9b), then the practitioner
must rotate the streak on the retinoscope so it is in line with the “ret reflex” (Fig. 12.9c).
Figure 12.9 Aligning the beam of the retinoscopic light source with that of the ret reflex
If you are doing retinoscopy on the person’s right eye, you need to place the retinoscope over your right
eye. Your right eye should be in line with the person’s right eye. Hold the retinoscope in your right hand.
Ask the patient to look at a fixation target like a spot of light or a 6/60 target at least 3 metres away. This
helps to relax accommodation and keep the eyes steady.
To get an accurate measurement of refractive error you need to keep the retinoscope very close to the
patient’s line of sight (visual axis).
Ask the patient to inform you if your head gets in the way of the target. Encourage him to look at the
distance target and keep reminding him to keep looking at the distance target and not at the ret light. Also
ensure that the patient keeps both eyes open.
For retinoscopy on the person’s left eye, move to the left side of the person and use the retinoscope in
your left hand with your left eye (examiner’s right eye for patient’s right eye, examiner’s left eye for
patient’s left eye). Remember to keep close to the patient’s visual axis.
If you add too much minus the movement will change to “with”. This means you have passed neutrality
and you need to remove some of the minus to return to the neutral point
COMMON ERRORS
1. Performing retinoscopy off-axis.
2. Performing retinoscopy at an incorrect working distance affects the working distance lens used.
3. Blocking the patient’s view of the distance chart which results in accommodation being stimulated.
4. Confusing the retinoscope sleeve positions. Using concave mirror retinoscopy instead of plane mirror
retinoscopy and not making the necessary adjustment.
5. Holding lenses away from the spectacle plane.
6. Not concentrating on the movement in the centre of the pupil in patients with large pupils.
When you move forward, the ret reflex will become “with”
When you move backward, the ret reflex will become “against”.
RECORDING FINDINGS
If using a working distance lens:
The refractive correction is recorded in spherocylindrical format after removing the working distance lenses.
e.g. OD: -1.00 / - 2.00 x 90 OS: -1.00 / -1.75 x 95
Do not use the DS for dioptres of spherical power, DC for dioptres of cylindrical power or the degree sign
when recording the prescription in this form. Also use the symbol “x” for the word axis. If the axis is 0º, then it
is recorded preferably as an axis of 180. These are one and the same.
The refractive correction is recorded in spherocylindrical format after adjusting the retinoscopy result in the
trial frame by the appropriate working distance correction.
e.g. If a working distance of 50cm was used and the retinoscopy result in the trial frame was as follows:
One would need to subtract a correction of 2.00DS from this spherical aspect of the final result. The final
prescription would thus be recorded as:
Similarly, if a retinoscopy result of OD: +5.00 / - 2.00 x 90 OS: +3.00 / -1.75 x 95 was found, the final Rx,
with assuming 50cm working distance, would be recorded as: OD: +3.00 / - 2.00 x 90 OS: +1.00 / -1.75 x 95.
OTHER SCENARIOS
Latent hyperopes have a tendency to display more sphere on retinoscopy than on dry subjective
refraction.
Patients with media opacities may display dim/dull reflexes which can make retinoscopy difficult
to perform. In these cases one may need to perform radical retinoscopy, with a reduced working
distance or off-axis position.
BIBLIOGRAPHY
Eskridge JB, Amos JF, Bartlett JD, Clinical Procedures in Optometry, Philadelphia, PA: J.B.Lippincott Company, 1991.
Benjamin W, Borish's Clinical Refraction, Butterworth-Heinemann, 2007