Bodyflow - P2Ch/-P1Ch: Service Manual
Bodyflow - P2Ch/-P1Ch: Service Manual
Bodyflow - P2Ch/-P1Ch: Service Manual
BodyFlowTM-P2CH/-P1CH
Revision: 11 / 2007
Bodyflow International PTY LTD
Legal Notice
Distribution, reproduction and translation of the software and its documentation (or excerpts
thereof) are prohibited without the prior written consent of Bodyflow™ International PTY LTD.
Bodyflow™ International PTY LTD reserves the right to change the software and associated data as
well as the documentation without notice. All other rights reserved.
2 BodyFlowTM-P2CH/-P1CH
Bodyflow International PTY LTD
Table of Contents
3 BodyFlowTM-P2CH/-P1CH
Bodyflow International PTY LTD
1 Technical Data
Table 1: Treatment
Table 2: Charging
Weight 0.485 kg
Standard Light
Current 40 mA 40 mA
4 BodyFlowTM-P2CH/-P1CH
Bodyflow International PTY LTD
Manufacturer Address
PHYSIOMED ELEKTROMEDIZIN AG
Hutweide 10
91220 Schnaittach/Laipersdorf
Germany
5 BodyFlowTM-P2CH/-P1CH
Bodyflow International PTY LTD
2 Instrument Overview
Table 6: Legend
Table 7: Symbols
6 BodyFlowTM-P2CH/-P1CH
Bodyflow International PTY LTD
2.1 Setmenue
In the Setmenue, you can adjust the following device parameters:
Symbol Meaning
2.2 Introduction
With your Bodyflow™ you have acquired a high-quality and extremely versatile unit for
stimulation current therapy. The instrument will only show its true potential, however, if you are
well informed about its functions. For this reason, carefully read the Operating Instructions and
familiarize yourself with the use of the instrument.
7 BodyFlowTM-P2CH/-P1CH
Bodyflow International PTY LTD
The instrument complies with all current safety standards. It meets the requirements of the EC
directive concerning medical devices (93/42/EEC) and is therefore CE-labelled.
Bodyflow™ has two modes of operation:
• Treatment: In this mode, the instrument is disconnected from the mains. When the battery
charger is plugged in, the instrument cannot be switched on and treatment is not possible.
Plugging in the battery charger into the instrument during treatment has the consequence
that treatment is being interrupted and the intensity will be automatically turned down to zero
and the instrument switches off.
• Charging: Charging is only possible when the device is switched off (refer to Mains and
Battery Operation on page 16).
8 BodyFlowTM-P2CH/-P1CH
Bodyflow International PTY LTD
3 Calibration
The following service instructions allow you to do a basic calibration.
Usually for later balancing (e.g. repairs) the entire balancing procedure is performed. Only
occasionally, if severe errors concerning the balancing procedure are not apparent or have
been cleared, it may be limited to individual components.
This service manual covers one channel devices (BodyFlowTM-P1CH) as well as two channel
devices (BodyFlowTM-P2CH).
When a second channel is indicated, the respective instructions are only valid for
BodyFlowTM-P2CH, but not for BodyFlowTM-P1CH!
Parameters only valid for the second channel are written in italics.
State of calibration*
Oscillograph with probe B
Digital multimeter B
Frequency meter B
For information necessary for the calibrating procedure such as positions of the
trimmers or test points, refer to the wiring scheme and the parts list from page 19.
9 BodyFlowTM-P2CH/-P1CH
Bodyflow International PTY LTD
4 Test Menu
To access the test menu, a jumper has to be set at the pins 1 + 2 of the socket X1 PM0615-1
before starting up the instrument. When switching on the instrument, you can access the test
menu then.
Ti 1 – 10ms Ri 0 – 950ms
Select the required parameter using the NAV buttons and press ENTER to confirm (e.g. the timer
value will be highlighted and flash).
Use the NAV buttons to modify the parameter value. To check the output signals of CH1 (and
CH2), use an oscillograph and a resistive load of 500 Ohm.
Screen 2: Checking the displayed value at maximum intensity without output load
10 BodyFlowTM-P2CH/-P1CH
Bodyflow International PTY LTD
Screen 3: Checking the displayed value at maximum dose with resistive output load
CH1 500 Ohm / CH2 500 Ohm
Calibrating actual value vs. required value
CH1 Required value Actual value Output voltage
DCmA DCmA DCV
CH2 Required value Actual value Output voltage
DCmA DCmA DCV
CH1 / CH2 digital value
Charging current
Charging current
The measured value DCA is calculated by the microprozessor!
Finally :
1. Unplug the battery charger when charging is finished.
2. Remove the jumper at X1 (pins 1 and 2) and switch of the device!
11 BodyFlowTM-P2CH/-P1CH
Bodyflow International PTY LTD
5 Selftest
To execute the instrument selftest, press the following key sequence when in therapy mode:
Pres ENTER <2> and POWER <7> simultaneously.
You have to press both keys until you have checked the following display values:
When the instrument selftest has detected no problems, the following message is displayed:
“NF . . OK “ + “Imax . . OK “
In the case of a problem, a monitoring note is displayed instead of OK (refer to page 18)!
12 BodyFlowTM-P2CH/-P1CH
Bodyflow International PTY LTD
The function keys <2> are used to select the therapy parameters
and to operate the instrument. After switching on the instrument, the
Display <1> shows the start screen. You can now access the desired
therapy program by pressing the right or left key. To select an item,
simply press the ENTER button in the middle.
13 BodyFlowTM-P2CH/-P1CH
Bodyflow International PTY LTD
The Output Indicator <5> tells you to be cautious when handling the electrodes.
Caution
When the Output Indicator <5> flashes, the Patient Lead Connector <8> is
under voltage!
Make sure you do not touch the electrodes when the current is turned up! Touching
the electrodes with your fingers may startle as the electrodes need to disperse the
current evenly!
The Power Connector <6> is located at the front side of the instrument. Here,
you plug in the supplied battery charger if you want to charge the batteries.
The Power Switch <7> is on the bottom of the instrument's upper side. By
means of this switch, you can switch the instrument on and off. After switching
on, a selftest is automatically carried out by the instrument.
14 BodyFlowTM-P2CH/-P1CH
Bodyflow International PTY LTD
The Patient Lead Connector <8> on the front side of the instrument serves to
plug in the electrodes.
The two connectors on the left (seen with the instrument facing up) are assigned
to circuit I, the two on the right to circuit II.
The polarity is of no importance, since the instrument operates in biphase mode.
15 BodyFlowTM-P2CH/-P1CH
7 Operation of the Device
The operating steps not directly relating to the therapy are described in the following paragraphs.
0%
25%
50%
75%
100%
Important
16 BodyFlowTM-P2CH/-P1CH
Important
If the unit is not used for a longer period of time, please fully charge the
battery once a month. This will help to avoid exhaustive discharge.
Battery Charger
Depending on the current state of the batteries, the LED is illuminated or flashing in green or
yellow. This has the following significance:
The battery charger can be equipped with different primary adaptors to match the
line voltage of the destination country. One primary adaptor for the respective
country is in the scope of delivery.
17 BodyFlowTM-P2CH/-P1CH
8 Monitoring Notes
8.1 Selftest
MUX…ERR 101
DAC…ERR 102
NF…ERR 105
Imax…ERR 106
8.2 Therapy
19 BodyFlowTM-P2CH/-P1CH
Sparpart List - Client: 1 / Physiomed Elektromedizin AG Date: 28.11.2007 12 Page: 1
p.ID: 01151 var./sw: 1/0-0-0 desc 1/2:Bodyflow portable mit Zubehör Zweikanalversion
O#: 01151 desc 3/4:Drainagegerät
ND: BODYFLOW desc 5/6:
Ersteller:Legl desc 7/8:
Typ: Bodyflow BF-P 1CH/2CH Bem: Sparepartlist