Bodyflow - P2Ch/-P1Ch: Service Manual

Download as pdf or txt
Download as pdf or txt
You are on page 1of 24

SERVICE MANUAL

BodyFlowTM-P2CH/-P1CH

Revision: 11 / 2007
Bodyflow International PTY LTD

Legal Notice
Distribution, reproduction and translation of the software and its documentation (or excerpts
thereof) are prohibited without the prior written consent of Bodyflow™ International PTY LTD.
Bodyflow™ International PTY LTD reserves the right to change the software and associated data as
well as the documentation without notice. All other rights reserved.

Bodyflow™ International PTY LTD


Suite 12 Level 1
134 Cambridge Street
Collingwood
Victoria
Australia 3066
1300 BODYFLOW
1300 26393569
www.bodyflow.com.au
DISCLAIMER
The Bodyflow™ portable stimulation unit ("the Product") is manufactured in Germany by
PHYSIOMED ELEKTROMEDIZIN AG and is distributed in Australia by its distributor,
Bodyflow™ International PTY LTD.
Bodyflow™ International PTY LTD has endeavoured to ensure that the data analysis and assessment
of such data and other information as provided by the manufacturer is accurate.
Any projected information contained in the operating instructions is based on Bodyflow's or
PHYSIOMED ELEKTROMEDIZIN AG's analysis and subjective estimates and assumptions and may be
about circumstances, scenarios and events which may take place. As such, no representations are
made by Bodyflow™ International PTY LTD or PHYSIOMED ELEKTROMEDIZIN AG as to the accuracy
of such information.
The product must be used strictly in accordance with the operating instructions. Neither
Bodyflow™ International PTY LTD or PHYSIOMED ELEKTROMEDIZIN AG will not be liable for any
liability arising from any loss, damage or injury caused through any use of the product outside the
scope of the operating instructions.
To the extent permitted by law, neither PHYSIOMED ELEKTROMEDIZIN AG or
Bodyflow™ International PTY LTD assume any liability for any loss or damages incurred directly or
indirectly from any use of the Product or as a result of any person acting or refraining to act in
reliance of any information contained in any assessment or operating instructions relating to the
Product or in respect of any negligent act.
Where any Act implies any condition or warranty in respect to the use or supply of the Product,
Bodyflow™ International PTY LTD's liability will be limited to the re-supply of the relevant Product by
Bodyflow™ International PTY LTD.
Bodyflow™ is made in Germany in compliance with the quality requirements of ISO 9001 and the
applicable safety standards and regulations of the Council Directive 93/42/EEC of 14 June 1993
concerning medical devices.
A conformity check acc. to Annex II, approved by the notified body 1275, was carried out.

2 BodyFlowTM-P2CH/-P1CH
Bodyflow International PTY LTD

Table of Contents

1 Technical Data ...........................................................................................................4


2 Instrument Overview ..................................................................................................6
2.1 Setmenue..........................................................................................................7
2.2 Introduction .......................................................................................................7
2.3 General Notes ....................................................................................................7
2.4 Instrument Description ........................................................................................7
3 Calibration.................................................................................................................9
3.1 Additional Equipment Required .............................................................................9
4 Test Menu ............................................................................................................... 10
5 Selftest ................................................................................................................... 12
6 Controls and Indicators ............................................................................................. 13
6.1 Display <1>..................................................................................................... 13
Symbols in the Upper Status Bar ................................................................................ 13
6.2 Function Keys <2> ........................................................................................... 13
6.3 Intensity Control Circuit I <3> and Circuit II <4> ................................................. 14
6.4 Output Indicator <5> ........................................................................................ 14
6.5 Power Connector <6> ....................................................................................... 14
6.6 Power Switch <7> ............................................................................................ 14
6.7 Patient Lead Connector <9>............................................................................... 15
7 Operation of the Device............................................................................................. 16
7.1 Mains and Battery Operation .............................................................................. 16
7.2 Notes on Handling the Batteries .......................................................................... 16
7.3 Battery Charger................................................................................................ 17
7.4 Economy Mode ................................................................................................. 17
8 Monitoring Notes ...................................................................................................... 18
8.1 Selftest ........................................................................................................... 18
8.2 Therapy........................................................................................................... 18
8.3 Battery Charger................................................................................................ 18
9 Appendix................................................................................................................. 19

3 BodyFlowTM-P2CH/-P1CH
Bodyflow International PTY LTD

1 Technical Data
Table 1: Treatment

Protection class acc. to VDE 0750 / IEC


Battery Mode Only, Type BF
601

Table 2: Charging

Protection class acc. to VDE 0750 / IEC


II
601

Input voltage 17 VDC

Input current 0.8 ADC

Power Supply 7,2 V 1350mAh Ni-MH Accumulator

Table 3: General Technical Data

acc. to Council Directive concerning medical devices


CE characterization
(93/42 EEC)

Class acc. to Council Directive concerning


IIa
medical devices

Ambient temperature (operation) + 10 °C ... + 40 °C

Storage temperature + 10 °C ... + 40 °C

Dimensions (W x H x D) 17.5 cm x 4.5 cm x 10 cm

Weight 0.485 kg

Table 4: Battery Charger

Type (to be used exclusively) Switchmode Charger FW 7219 / NI 4-10 NTC

Mains supply 100 ... 240 VAC

Input current 0.1 ... 0.3 A

Mains frequency 50 ... 60 Hz

Output voltage 17 VDC

Output current 0.8 ADC

Table 5: Stimulation Current Output Parameters

Standard Light

Current 40 mA 40 mA

Resistance 500 Ohm 500 Ohm

4 BodyFlowTM-P2CH/-P1CH
Bodyflow International PTY LTD

Manufacturer Address

PHYSIOMED ELEKTROMEDIZIN AG

Hutweide 10

91220 Schnaittach/Laipersdorf

Germany

5 BodyFlowTM-P2CH/-P1CH
Bodyflow International PTY LTD

2 Instrument Overview

Table 6: Legend

1 Display 2 Function Keys

3 Intensity Control Circuit I 4 Intensity Control Circuit II (only BodyFlowTM-P2CH)

5 Output Indicator 6 Power Connector

7 Power Switch 8 Patient Lead Connector

Table 7: Symbols

Type BF component, not connected to protective ground wire!

6 BodyFlowTM-P2CH/-P1CH
Bodyflow International PTY LTD

2.1 Setmenue
In the Setmenue, you can adjust the following device parameters:

Symbol Meaning

Contrast of the Display <1>

Brightness of the Display <1>

Back to start screen

2.2 Introduction
With your Bodyflow™ you have acquired a high-quality and extremely versatile unit for
stimulation current therapy. The instrument will only show its true potential, however, if you are
well informed about its functions. For this reason, carefully read the Operating Instructions and
familiarize yourself with the use of the instrument.

2.3 General Notes


The instrument complies with the technical specifications of IEC 601, VDE 0750 and is assigned
to class IIa according to the Council Directive concerning Medical Devices.
The instrument may only be operated by qualified personnel who have undergone special
training. You must operate the instrument properly, i.e. in accordance with the Operating
Instructions.
It is not intended for operation in explosion hazard zones or hydrotherapy rooms. Drastic
temperature changes should be avoided, since condensation could be caused within the
instrument. Do not start up the instrument until it is in temperature equilibrium with its
environment!
Operating the instrument in the proximity (e.g. 1 m) of a short-wave or micro-wave therapy unit
may cause output irregularities and should be avoided for this reason. Simultaneous connection
of the patient to high-frequency surgical instrument should also be avoided.
Using the electrodes near the chest can increase the risk of heart beat irregularities.

2.4 Instrument Description


Bodyflow™ is a portable stimulation current therapy unit. The device is equipped with a
rechargeable battery and is intended to be used as a mobile unit, e.g. in situations where no
connection to the mains is available. This unit can only be used on battery power and not whilst
plugged into mains power.
The function of Bodyflow™ is controlled by a microprocessor. Essential components are
permanently controlled by the processor and thus malfunctions are prevented. After switching on,
all instrument functions are checked during an automatic self-test routine.

7 BodyFlowTM-P2CH/-P1CH
Bodyflow International PTY LTD

The instrument complies with all current safety standards. It meets the requirements of the EC
directive concerning medical devices (93/42/EEC) and is therefore CE-labelled.
Bodyflow™ has two modes of operation:
• Treatment: In this mode, the instrument is disconnected from the mains. When the battery
charger is plugged in, the instrument cannot be switched on and treatment is not possible.
Plugging in the battery charger into the instrument during treatment has the consequence
that treatment is being interrupted and the intensity will be automatically turned down to zero
and the instrument switches off.
• Charging: Charging is only possible when the device is switched off (refer to Mains and
Battery Operation on page 16).

8 BodyFlowTM-P2CH/-P1CH
Bodyflow International PTY LTD

3 Calibration
The following service instructions allow you to do a basic calibration.
Usually for later balancing (e.g. repairs) the entire balancing procedure is performed. Only
occasionally, if severe errors concerning the balancing procedure are not apparent or have
been cleared, it may be limited to individual components.
This service manual covers one channel devices (BodyFlowTM-P1CH) as well as two channel
devices (BodyFlowTM-P2CH).
When a second channel is indicated, the respective instructions are only valid for
BodyFlowTM-P2CH, but not for BodyFlowTM-P1CH!
Parameters only valid for the second channel are written in italics.

3.1 Additional Equipment Required

State of calibration*
Oscillograph with probe B
Digital multimeter B
Frequency meter B

*State of calibration B = precision 2 %

For information necessary for the calibrating procedure such as positions of the
trimmers or test points, refer to the wiring scheme and the parts list from page 19.

9 BodyFlowTM-P2CH/-P1CH
Bodyflow International PTY LTD

4 Test Menu
To access the test menu, a jumper has to be set at the pins 1 + 2 of the socket X1 PM0615-1
before starting up the instrument. When switching on the instrument, you can access the test
menu then.

Screen 1: Output Current

Selectable parameters for functional testing


0 – 60 min

CH1 0 – 40mA CH2 0-40mA

Ti 1 – 10ms Ri 0 – 950ms

pole plus pole auto pole minus

-> Proceed to screen 2 (maximum intensity)

Select the required parameter using the NAV buttons and press ENTER to confirm (e.g. the timer
value will be highlighted and flash).
Use the NAV buttons to modify the parameter value. To check the output signals of CH1 (and
CH2), use an oscillograph and a resistive load of 500 Ohm.

Screen 2: Checking the displayed value at maximum intensity without output load

CH1 Required value Actual value Output voltage


DCmA DCmA DCV
CH2 Required value Actual value Output voltage
DCmA DCmA DCV
CH1 / CH2 digital value

Imax Test function “Trigger safety shutoff”


Pressing ENTER key: Imax = 45 -55 DCmA
< - Return to screen 1 -> Proceed to screen 3

10 BodyFlowTM-P2CH/-P1CH
Bodyflow International PTY LTD

Screen 3: Checking the displayed value at maximum dose with resistive output load
CH1 500 Ohm / CH2 500 Ohm
Calibrating actual value vs. required value
CH1 Required value Actual value Output voltage
DCmA DCmA DCV
CH2 Required value Actual value Output voltage
DCmA DCmA DCV
CH1 / CH2 digital value

Imax Test function “Trigger safety shutoff”


Pressing ENTER key: Imax = 45 -55 DCmA
< - Return to screen 2 -> Proceed to screen 4

Screen 4: Calibration of Ubatt (DCV)


Attention! Battery charger is not connected
Enable the selection field Ucal (cursor flashes).
Measure Ubatt at the battery using the digital multimeter.
Set the measured value in the display using the NAV buttons.

Charging current

<- Return to screen 2 Contrast of the Display <1>

Screen 5: Charging current


Attention! Battery charger is connected

Enable the selection field Ucal (Cursor flashes).


Measure Ubatt at the battery using the digital multimeter.
Set the measured value in the display using the NAV buttons.

Charging current
The measured value DCA is calculated by the microprozessor!

<- Return to screen 2 Contrast of the Display <1>

Finally :
1. Unplug the battery charger when charging is finished.
2. Remove the jumper at X1 (pins 1 and 2) and switch of the device!

11 BodyFlowTM-P2CH/-P1CH
Bodyflow International PTY LTD

5 Selftest
To execute the instrument selftest, press the following key sequence when in therapy mode:
Pres ENTER <2> and POWER <7> simultaneously.
You have to press both keys until you have checked the following display values:

Screen 6: Instrument Selftest

CH1 Optocoupler Value Value


check 10mA +/- 40mA +/- 10%
value ≤ 5 10%
CH2 Optocoupler Value
(only for check 25mA +/-
P2CH) value ≤ 5 10%
Imax Test function “Trigger safety shutoff”
Pressing ENTER key: Imax = 45 -55 DCmA

When the instrument selftest has detected no problems, the following message is displayed:
“NF . . OK “ + “Imax . . OK “

In the case of a problem, a monitoring note is displayed instead of OK (refer to page 18)!

12 BodyFlowTM-P2CH/-P1CH
Bodyflow International PTY LTD

6 Controls and Indicators


The design of Bodyflow™ allows for easy operation. Because of its small size, the instrument is
very easy to transport. It has been designed for operation both inside and outside of therapy
rooms, and is fed by rechargeable batteries for that reason (refer to Mains and Battery Operation
on page 16).
All controls and indicators are integrated into the housing, thus allowing for easy cleaning of the
instrument's surface and protecting it from dust.
The instrument's microprocessor monitors the safety-related components, prevents from
malfunctions and checks the instrument after switching it on.

6.1 Display <1>

The Display <1> shows all menu items


including the therapy parameters of the
instrument. You can select the parameters
using the Function Keys <2>.

Symbols in the Upper Status Bar


The upper status bar shows the following symbols:

Button to open the popup menu

Currently selected menu

State of charge of the battery (refer to Mains


and Battery Operation on page 16)

6.2 Function Keys <2>

The function keys <2> are used to select the therapy parameters
and to operate the instrument. After switching on the instrument, the
Display <1> shows the start screen. You can now access the desired
therapy program by pressing the right or left key. To select an item,
simply press the ENTER button in the middle.

13 BodyFlowTM-P2CH/-P1CH
Bodyflow International PTY LTD

6.3 Intensity Control Circuit I <3> and Circuit II <4>

Intensity Control Circuit I <3> and Intensity Control Circuit II


<4> (only BodyFlowTM-P2CH) serve to set the intensity in the circuits I
and II in steps of 0.5 mA. When turning up the intensity of the two
intensity controls, the associated therapy timer in the Display <1> will
be started as well.

Automatic Output Current Switch-off


Bodyflow™ features an automatic output current switch-off, activated in case the current
flow of the electrodes is interrupted (electrode falls off or is disconnected from the
instrument). The symbol will appear in alternation with the intensity on the Display
<1> and the current will be automatically turned down to a minimum basic current in the
respective circuit. The timer stops the therapy time.
To eliminate the error, you have to press the button with the minus symbol of Intensity
Control Circuit I <3> or Intensity Control Circuit II <4> one time to be able to
continue. The message will disappear and you can increase the intensity again.

6.4 Output Indicator <5>

The Output Indicator <5> tells you to be cautious when handling the electrodes.

Caution

When the Output Indicator <5> flashes, the Patient Lead Connector <8> is
under voltage!
Make sure you do not touch the electrodes when the current is turned up! Touching
the electrodes with your fingers may startle as the electrodes need to disperse the
current evenly!

6.5 Power Connector <6>

The Power Connector <6> is located at the front side of the instrument. Here,
you plug in the supplied battery charger if you want to charge the batteries.

6.6 Power Switch <7>

The Power Switch <7> is on the bottom of the instrument's upper side. By
means of this switch, you can switch the instrument on and off. After switching
on, a selftest is automatically carried out by the instrument.

14 BodyFlowTM-P2CH/-P1CH
Bodyflow International PTY LTD

6.7 Patient Lead Connector <9>

The Patient Lead Connector <8> on the front side of the instrument serves to
plug in the electrodes.
The two connectors on the left (seen with the instrument facing up) are assigned
to circuit I, the two on the right to circuit II.
The polarity is of no importance, since the instrument operates in biphase mode.

15 BodyFlowTM-P2CH/-P1CH
7 Operation of the Device
The operating steps not directly relating to the therapy are described in the following paragraphs.

7.1 Mains and Battery Operation


At battery operation, the battery has to be fully charged before operating it for the first time. The
typical life expectancy of this battery and it's recharge life is 500 cycles or recharges.
The charging status of the batteries is displayed on the Display <1>:

Battery Charge Symbol

0%

25%

50%

75%

100%

How to Charge the Battery


If you want to charge the battery, proceed as follows:
1. Plug the supplied battery charger into the Power Connector <6> on the front side of the
device.
2. The batteries are being charged. When the batteries are completely discharged, the charging
procedure will take approx. 3 hours.

Important

In order to ensure a long battery life, the batteries must be charged


completely when first charged. The first charging procedure should not be
interrupted!

7.2 Notes on Handling the Batteries


If the battery capacity is very low during operation, the 3-step warning system is activated:
a. The charging status symbol flashes.
b. An acoustic signal sounds every second and the charging status symbol flashes. The intensity
is reduced prematurely.
c. The device shuts down to avoid complete discharging of the batteries.
In this case, recharge the battery, as described in section How to Charge the Battery on page 16.

16 BodyFlowTM-P2CH/-P1CH
Important

If the unit is not used for a longer period of time, please fully charge the
battery once a month. This will help to avoid exhaustive discharge.

7.3 Battery Charger


The supplied battery charger (Ref. No 00584) has an LED to indicate the current state of the
batteries.

Battery Charger

Depending on the current state of the batteries, the LED is illuminated or flashing in green or
yellow. This has the following significance:

State of Battery Charger LED Light Code

Standby LED is permanently yellow

Precharge LED flashes slowly in yellow

Rapid Charge LED flashes quickly in green

Maintain LED flashes slowly in green

Error LED flashes quickly in yellow

Ready LED is permanently green

Wait LED flashes slowly in green and yellow (alternating)

The battery charger can be equipped with different primary adaptors to match the
line voltage of the destination country. One primary adaptor for the respective
country is in the scope of delivery.

7.4 Economy Mode


The unit automatically switches over to the economy mode to save power. This will occur after
approx. 20 seconds. The Display <1> is no longer illuminated. Pressing any key will re-activate
the illumination.

17 BodyFlowTM-P2CH/-P1CH
8 Monitoring Notes

8.1 Selftest

ERROR CODE Monitoring note

MUX…ERR 101

DAC…ERR 102

NF…ERR 105

Imax…ERR 106

NF…ERR CH2 125

8.2 Therapy

ERROR CODE Monitoring note

I20% ERROR CH1 / CH2 201

Imax H_ERROR CH1 / CH2 202

Imax S_ERROR CH1 / CH2 203

T/R ERROR CH1 / CH2 204

0mA/0V ERROR CH1 / CH2 205

Imax S_INT_ERROR CH1 / CH2 206

8.3 Battery Charger

Test Monitoring note


Battery charger connected,
***
TEST OKAY
Battery charger connected,
901
TEMPERATURE SENSOR FAULT
Battery charger connected,
902
CHARGING CURRENT MEASUREMENT FAULT
9 Appendix

19 BodyFlowTM-P2CH/-P1CH
Sparpart List - Client: 1 / Physiomed Elektromedizin AG Date: 28.11.2007 12 Page: 1

part number / O-No description / ND step/cost perpos.no. item notice quantity

p.ID: 01151 var./sw: 1/0-0-0 desc 1/2:Bodyflow portable mit Zubehör Zweikanalversion
O#: 01151 desc 3/4:Drainagegerät
ND: BODYFLOW desc 5/6:
Ersteller:Legl desc 7/8:
Typ: Bodyflow BF-P 1CH/2CH Bem: Sparepartlist

51696 [¶] housing lower part 1 1 1 1,0000 Stck


bodyflow BF-P
51695 [¶] housing upper part 1 1 2 1,0000 Stck
bodyflow BF-P
51697 [¶] vision panel 1 channel 1 100 3 1,0000 Stck
bodyflow BF-P
51720 [¶] vision panel 2 channel 1 100 4 1,0000 Stck
bodyflow BF-P
51698 [¶] keyboard pad 1-channel 1 100 5 1,0000 Stck
bodyflow BF-P
51721 [¶] keyboard pad 2-channel 1 100 6 1,0000 Stck
bodyflow BF-P
51723 [¶] 0615-1/2-CH mainboard 1 1 7 1,0000 Stck
bodyflow BF-P
51726 [¶] 0503-1/PVB processor board 1 1 8 1,0000 Stck
bodyflow BF-P
51724 [»] 0616-1 keypad board 1 1 9 1,0000 Stck
bodyflow BF-P
51725 [»] 0617-1 display board 1 1 10 1,0000 Stck
bodyflow BF-P
51734 [¶] accumulator 1 1 11 1,0000 Stck
FT 7,2-02
51735 [¶] insulating plate/ Accu 1 1 12 1,0000 Stck
bodyflow BF-P
51699 [¶] rubber foot 1 1 13 4,0000 Stck
bodyflow BF-P
51739 [¶] TEXAS socket 3-pole 1 1 14 1,0000 Stk
bodyflow BF-P
51700 [¶] plug socket holder 1 100 15 1,0000 Stck
bodyflow BF-P
51740 [¶] plug socket 2mm (red) MLA2-G 1 100 16 2,0000 Stck
bodyflow BF-P
51741 [¶] plug socket (black) MLA2-G 1 100 17 2,0000 Stck
bodyflow BF-P