EPM Manual DensityPRO
EPM Manual DensityPRO
EPM Manual DensityPRO
“Microsoft” and “Windows” are either trademarks or registered trademarks of Microsoft Corporation in the United
States and/or other countries. “HART” is a registered trademark of the HART Communication Foundation.
“FOUNDATION Fieldbus” and “Fieldbus Foundation” are registered trademarks of Fieldbus Foundation.
“National Instruments” is a registered trademark of National Instruments Corporation.
All other trademarks are the property of Thermo Fisher Scientific Inc. and its subsidiaries.
Thermo Fisher Scientific Inc. (Thermo Scientific) makes every effort to ensure the accuracy and completeness of this
manual. However, we cannot be responsible for errors, omissions, or any loss of data as the result of errors or omissions.
Thermo Scientific reserves the right to make changes to the manual or improvements to the product at any time without
notice.
The material in this manual is proprietary and cannot be reproduced in any form without expressed written consent
from Thermo Scientific.
This page intentionally left blank
Revision History
Revision Level Date Comments
A 08-2015 Initial release per ERO 8698
B 02-2017 ECO 9079 - Added handling of calibration and
standardization in Primary Measurement Type for
Main CPU version >= 3.600
The following admonitions are used throughout this manual to alert users to potential
Warnings, hazards or important information. Failure to heed the warnings and cautions in this
Cautions & manual can lead to injury or equipment damage.
Warning: The triangular icon displayed with a warning advises the user about
Notes the type of hazard covered by the warning. See the table below for the types of
warning symbols used in this manual.
Table 1. Types of Warnings
NOTE: Please read the notes below before upgrading from CPU firmware
≤ 3.500 to CPU firmware version 3.600 and above.
Allowed STD/CAL Operation In Notes 1-4 below See Note 1 See Note 2 See Note 3 See Note 4
Note 1: g/cc
Note 2: g/cc
The user will be able to perform STD and CAL using items 3 -8 listed in Case 4.
However, the user shall enter the STD and CAL values in Density Primary
Measurement type and NOT in "g/cc" as shown on the User Interfaces (UI).
The user must update items 3-8 as necessary to reflect correct STD/CAL unit
on the Standardization and Calibration page of the UI.
Start
This
Upgrading CPU firmware from procedure
No
Rev <= 3.500 to version => 3.6 does not
apply
Yes
End
Quick Setup
The procedures described in this section assume you will access the menu items
directly.
There are four methods of standardization: on water, on process, deferred, and other.
Standardization
There are two types of calibration: a one point calibration and a two point calibration.
Calibration
The two calibration points allow you to have two calibration points in the region of
interest, and standardization can then be used to compensate for pipe wear and process
buildup on the pipe walls. The first point calibration can be thought of as an offset. It
will move the response curve up or down.
The second point calibration is a slope correction. The slope correction pivots around
the first cal point. The slope can be set using the second point calibration or entered
directly.
If the gauge is standardized on water or on process, the standardization serves as the
first calibration point.
Function
A Cesium (Cs-137) radioisotope source is used for most applications, and a Cobalt
Source (Co-60) source is available for applications requiring a higher energy source. The
radioisotope is bound in ceramic or glass pellets and doubly encapsulated in a pair of
sealed stainless steel containers. The resulting source capsule is highly resistant to
vibration and mechanical shock.
The source capsule is further enclosed in the source head, a lead-filled, welded steel
housing. A shaped opening in the lead shielding directs the gamma radiation beam
through the process material towards the detector. Outside of the beam path, the
energy escaping the source head is very low and well within prescribed limits. Closing
the source shutter allows the beam to be turned off (the shutter blocks the radiation)
during installation or servicing of the gauge. All source housings meet or exceed the
safety requirements of the U.S. Nuclear Regulatory Commission (NRC) and
Agreement State regulations. Refer to the gamma radiation safety guide (P/N 717904).
The gauge uses a scintillator-type detector to measure the radiation reaching the
Detector detector from the source. The detector consists of either a special NaI (sodium iodide)
Transmitter scintillator material or a PVT (polyvinyl toluene) scintillator material coupled to a
photomultiplier tube with the associated electronics. When radiation strikes the
scintillator material, small flashes of light are emitted. As the density of the process
material increases, more gamma radiation is absorbed by the process material and fewer
light pulses are generated by the scintillator material. The photomultiplier tube and
associated detector electronics convert the light pulses into electrical pulses that are
processed to determine the process material density and related measurement values.
Communication with the gauge is via the RS485 and RS232 serial ports, USB, or
Communications & Ethernet from a PC running EZ Cal II software. Once the gauge is set up, the primary
Measurement measurement (density) can be viewed on the display and on the EZ Cal II software.
Display
Display Background The display on the DensityPRO units provides measurement data to the user.
Display & Keypad The purpose of the display and keypad on the DensityPRO remote transmitter is to
provide the user with the ability to access the configuration menus and system data
(Remote Unit) information.
2. When the user first starts up the system, the display informs the user that the
system is booting.
THERMO SCIENTIFIC
MS2011
SYSTEM IS BOOTING
Inputs & Outputs The DensityPRO gauges provide the user with numerous inputs and outputs, which
can be found in the table below.
Type Description
Inputs – Three 4–20 mA inputs, full scale ± 0.3% over operating
temperature range; fault high/low detection
– Two 0 to 10 VDC voltage inputs, full scale ± 0.3% over
operating temperature range
– Two digital inputs (DI) provide contact input with internal +5
VDC wetting voltage
– Temperature compensation circuitry with 100-ohm Platinum
RTD, 3- or 4-wire; full scale ± 0.4ºC over operating
temperature
Current outputs – 4–20 mA output, full scale ± 0.3% over operating temperature
range
Isolated, loop-powered (default)
Isolated, self-powered output
– Optional Intrinsically Safe Input/Output 4–20 mA output, full
scale ± 0.3% over operating temperature range temperature
range
Isolated, loop-powered (default)
Isolated, self-powered output
Contact closure (relay) outputs Two relays, DPDT-fully sealed 8 A at 250 VAC
Serial outputs – RS485 half duplex
– RS232 full duplex
– Fieldbus: A DensityPRO gauge is available
from the Fieldbus Foundation™ website.
The DD is a DD4 or DD5, interpreted by a
host implementing DD Services 4.x or 5.x.
– (Pending)
– (Pending)
Features
The setup wizard enables you to quickly configure the gauge by requiring you to enter
Setup Wizard all of the basic parameters. Additional menu groups contain fields in which you can
enter specialized parameters and commands, allowing you to customize the gauge for a
wide variety of applications.
Note: The setup wizard is only available when using the PC-based EZ Cal II
software.
Thermo Fisher’s Dynamic Process Tracking (DPT) ensures there is no lag time in the
Instantaneous system’s response to significant changes in process density. When changes reach a user
Response set threshold, the DPT feature reduces the normal averaging time constant to a faster,
user-set time constant, ensuring a rapid, smooth output response. When the process
stabilizes, the time constant automatically goes back to the original setting, to reduce
the fluctuations inherent in radiation-based measurements. In this way, process density
changes are immediately reflected in the transmitter output, while the effects of
statistical variations in the radiation measurement are greatly reduced.
Select up to four measurement values for display: density, bulk density, solid and
Multiple Readouts carrier concentrations, the ratio of solids to carrier, bulk mass flow, bulk volume flow,
and the rate of change of any of these measurements.
Up to sixteen process alarms may be assigned in the system, in addition to system fault
Extensive Alarms alarms and warning alarms
In addition to this guide, the following documents must be read and understood by all
Additional persons installing, using, or maintaining this equipment:
Documents – DensityPRO NAI installation guide (P/N 1-0702-015)
– DensityPRO installation guide (P/N 1-0702-144)
– Gamma Radiation Safety (P/N 717904)
The EZ Cal II PC user interface software provides a way for users to interact with the
Features DensityPRO gauges.
The DensityPRO integrated units provide the option of having a local display screen,
but do not have a keypad. This prohibits users from controlling and configuring the
gauge. Connecting the integrated unit to a PC running the EZ Cal II software provides
the user with these abilities.
The DensityPRO remote units include a keypad and display screen. All of the
operations available to the user through use of the gauge’s keypad are also available in
the EZ Cal II software. Using the software allows the user to be away from the gauge
and provides a more user-friendly interface for interaction.
The EZ Cal II PC user interface software provides the user with:
– The ability to connect to the gauge via USB, RS232, RS485 (2/4wire), and
Ethernet.
– The ability to configure communication ports A and B.
– The ability to configure the Ethernet port.
– The option to configure the system control setup.
– The option to display the system status.
– The option to configure the alarms and display the status of alarms.
– The option to configure the execution of system commands.
– The option to configure physical Input and Output setup and display the current
status of each.
– The option to configure detector setup.
– The option to configure density application setup.
– The option to flash application firmware to the gauge’s main board.
– Access to a gauge setup wizard for quick gauge configuration.
– Access to a calibration wizard to calibrate all physical inputs and outputs.
– An option to upload configuration from a gauge to the PC and save it to a file.
– An option to download a file configuration from a PC to the gauge.
– The option to enter a special password for cold and warm start operation.
– The option to display user interface menu text in the following languages:
English
Chinese (available in future releases)
Portuguese (available in future releases)
Spanish (available in future releases)
– The option to change the password mode entry.
Connect the serial port on a PC (Com A or Com B) to the RS232 serial port of the
Startup DensityPRO gauge. This connection enables you to communicate with the gauge from
a PC running EZ Cal II software.
The setup wizard provides you with a step-by-step procedure for entering the data
The Setup required for gauge operation. To start the wizard, open the EZ Cal II software and
Wizard click on the small blue wizard’s cap on the task bar, underneath the View menu
option.
If the user adjusts the contrast during the booting process, the gauge will save the new
Adjusting Contrast contrast setting once booting is complete. If no change in contrast is made, the
While Booting previously saved contrast setting will be applied to the display screen.
The keypad is a five by five membrane keypad comprised of input keys and operational
Keypad keys. The numeric keys 0 through 9 and the character keys representing a decimal
Overview point (.) and a dash (-) allow the user to input data into the gauge using the keypad.
The arrow keys allow the user to scroll through information on the screens.
Additionally, the up and down arrow keys are used to adjust the display contrast. The
F1, F2 and F3 keys provide different functionalities based on the information on the
display. Some of the functionalities include moving the user back to the previous
screen, moving the user forward to the next screen, selecting a field for editing, and
submitting newly-input information.
The menu screens organize data collected by the DensityPRO gauge into categories
Using the Keypad and subjects designed to help direct the user to the correct data. The organization of
with the Menu the display menu screens is very similar to that of the EZ Cal II software. Using the up
and down arrows allows the user to scroll through the menu items on the display
Screens screen until reaching the desired category of information. Pressing the F1 key will
bring the user back to the previous screen, while the F3 key will move the user to the
next screen associated with the selected menu.
Note: The up and down arrows can only scroll through menu items when
viewing the menu screens. The contrast control function is not available.
The edit screens allow the user to analyze data, interact with the gauge, and take
Using the Keypad action. Edit screens run in a read mode and a write mode.
with the Edit Read mode displays data to the user. The screen continues to display up-to-date
Screens information. Pressing the F2 button will open the screen to write mode.
Note: Once a screen has been opened for editing, the up and down arrows
regain contrast control functionality.
If the screen is opened in write mode and the user is not logged in, the screen will
change to the password screen so the user may log in with a password. Only users with
engineering access may submit changes to the information stored in the gauge’s
database.
In writing mode, the function keys perform different operations.
– F1 will exit write mode without saving any changes.
– F2 will submit and save the new data.
– F3 will move the user through the different fields available for editing on the
screen.
An arrow in front of a displayed field indicates a dropdown menu associated with that
Editing Fields with field. Once the field is selected, the up and down arrow keys enable the user to explore
Dropdown Menus the dropdown selections. If a field contains more characters than the screen can
display, using the right and left arrows will shift the data provide visibility.
Once changes are complete, press the F2 button to submit and save the data to the
gauge.
Configuration
Uploading a configuration file provides the user with a way to save all configured
Upload parameters from the gauge to the computer. In the event of a cold start, the file can
Configuration then be downloaded back to the gauge to reinstall the set parameters.
1. To save a file configuration from the gauge to the computer, click the Upload
CFG from Gauge button.
2. Enter a name for the configuration file and click Save.
3. To abort the upload, click the Abort button.
4. In order to save the configuration to the gauge permanently, click the Write to
Flash Again button. This will ensure that the specified parameters are loaded each
time the system boots.
Sometimes it may be preferable to load parameters into the gauge for only a
short time, such as while running a specific test. In this instance, once the
parameters have been downloaded to the gauge, the process is complete. Once
the gauge is shut down, the configuration parameters will revert to the last
configuration written to flash.
The following buttons are seen consistently throughout the EZ Cal II software:
EZ Cal II Control
– Abort – Stops the current process.
Buttons – Auto Refresh – Continuously refreshes the screen with data from the gauge.
– Back – Moves the user to the previous screen.
– Cancel – Closes the current screen.
– Home – Returns the user to the Detector and Application Type Selection screen,
Figure 3-3.
– Next – Moves the user to the next available screen.
– Refresh – Updates the screen with the latest information from the gauge.
– Submit – Saves any information entered on the current screen to the gauge.
If a screen has multiple Submit buttons, each button will apply to a separate,
portioned-off area of the screen.
Password protection limits access to the gauge’s setup parameters based on the access
User Modes & level rights granted through the four user modes below.
Passwords – Display List –This is normal mode of operation. The local display will scroll
through a list of measurement data with no ability to view or modify the database
via Remote Backplane (RPB) display/keypad. Remote read access to the database is
available however no writes to are allowed.
– Operator Mode – The local display on the RBP will be running the user menu
interface, allowing a user to view the configuration of the gauge. Remote read
access to the database is available, however no writes to are allowed.
– Technician Mode – The local user can configure a small sub-set of the database
within the gauge via the front panel menu system on the RBP. Remote write access
is also available to a small subset of the gauge database so that calibration and
standardization can be performed.
– Engineer Mode – Local and remote users have full read/write access to the gauge’s
database.
The password configuration is accessible from the following interface ports:
– Com A
– Com B
– Ethernet
– USB
– Local Display/Keypad
– Fieldbus Interface
Only one port may have access to Engineer mode at any time. The exception to this is
the Fieldbus port, which always has Engineering access, regardless of the operational
modes of the other ports.
The default operation mode is Display List. If no activity is detected on the port for
five minutes, the unit will default back to Display List. See System Timeout for
additional information.
The following values are available for diagnostics purposes, to be used in other
processes and for display purposes.
– Com A Password Mode
– Com B Password Mode
– Ethernet Password Mode
– USB Password Mode
– Remote Display Password Mode
– Current Password Mode
– Engineering Mode Count
– Operator Mode Count
– Technician Mode Count
– Invalid Password Count
The following values shall be available for editing purpose when the gauge is in
engineering mode.
– Engineering Password
– Technician Password
– Operator Password
– Engineering Mode Count
The Password Entry/Validation screen can be accessed either by selecting the screen
Password from the Functions dropdown menu at the top of the screen, or by clicking the last
Entry/Validation icon button, which is circled in Figure 3-2.
The fields representing the inputs will display the current operational mode of
each port. If the user connects to the gauge through Com A and enters the
Engineer mode password, Com A Password Mode will display Engineer, while the
other port fields say Display List.
– The Mode Count fields indicate how many times the gauge has been accessed in
each respective mode.
– The passwords for the different access levels may be changed by manually
changing them in the Password Setup section of this screen.
Note: Passwords must be numeric only so as to be accessible when working at
the keypad display.
If a user does not interact with the gauge for five minutes, the display will time out and
System Timeout begin showing the scrolling measurement screens. Each time a button on the keypad is
pressed or the user interacts with the EZ Cal II software, the timeout period resets to
five minutes.
During standardization and calibration the timeout function is inactive and will not
occur.
Begin the density setup using the setup wizard. To start the wizard, click on the small
The Setup blue wizard’s cap on the task bar, underneath the View menu option.
Wizard
4. After selecting the type of application, click the Next button to proceed to the
Wizard Type Selection screen.
5. Like the Upload / Download Configuration screen, the Setup Wizard allows the
user to upload and download configuration files. However, while the
Upload/Download Configuration screen manages system-wide gauge parameters,
the buttons on this screen supply the user with a smaller set of parameters.
a. To save setup parameters to the computer for future reference, click Upload
CFG from Gauge.
i. Only the parameters related to the selected wizard type will be uploaded.
b. Name the file and click Save.
c. To download density setup parameters back to the gauge, click the Open file
button and select the desired file.
i. Only the parameters related to the selected wizard type will be uploaded.
d. Click Download File to Gauge.
6. Click the Next button to proceed to the next screen.
If slurry was selected, enter the carrier density (specific density of the
carrier liquid). The default value is 1.0. Enter the solids density (the dry
solid density of suspended solids) in g/cc. The default value is 2.65. Enter
the attenuation coefficients of the carrier and solid in cm2/g. The default
values for these options are .086 and .077, respectively.
If solution was selected, enter the solvent density (the specific density of
the solvent liquid) in g/cc. The default value is 1.0. Enter the solute
density (the dry solid density of the suspended solute) in g/cc. The default
value is 2.65. Enter the attenuation coefficients of the solvent and solute
in cm2/g. The default values for these options are .086 and .077,
respectively.
Select single phase when it is unnecessary or impossible to describe the
process material as slurry, emulsion, or solution. For example, foam
plastic, a mixture of plastic and gas, might be measured as a single-phase
material if the gas in the mixture only varies the material’s density and has
little effect on the measurement.
For an emulsion, the Fluid 1 Density is the specific density of the carrier
liquid in g/cc. The default value is 1.0. The Fluid 2 Density is the specific
density of the suspended liquid in g/cc. The default value is 2.65 g/cc.
Enter the attenuation coefficients of the carrier liquid and the suspended
liquid in cm2/g. The default values for these options are .086 and .077,
respectively.
Note: If you want to measure the overall density of the process material only,
you can select single-phase, regardless of the material’s makeup.
9. After making the appropriate selections on this screen, click the Next button to
proceed to the Gravity & Attenuation screen.
The DensityPRO family of gauges will accept a process temperature input from a 3- or
4-wire RTD, as a 4–20 mA signal from an existing temperature transmitter, or as a
manual value entered into the transmitter by the operator.
1. Select Temperature Compensation and click the Next button to bring up the
following screen. If your application requires temperature compensation, ensure
that the parameters on this screen are entered correctly.
Standardization 1. Select Standardization and click the Next button to advance to the following
screen.
STD Sample Time (sec) – The standardization sample time is the amount of
time, in seconds, that the detector will average the incoming signal. Choosing
an appropriate value for this parameter will depend on your process
conditions. In situations where the process density is quite steady, such as
when running water through a pipe, there should not be any significant swings
in the density value, and, therefore, the detector count should be stable. Under
these conditions, a relatively short standardization time, in the order of 60 to
300 seconds may be appropriate. When standardizing on a live process, it is
more likely that the density value will change over time. In these conditions it
is desirable to use a much longer sampling time, more in the order of 600 to
900 seconds.
STD Density – Enter the standardized density of the material in g/cc.
2. Click the Next button to view the Remaining Time Screen.
1. Select Gauge Calibration and click the Next button to advance to the following
Gauge Calibration screen.
3. Enter the calibrated density in the CAL Density textbox, and choose which
calibration point to use from the CAL Point dropdown. After point 1 is calibrated,
this dropdown list will give the user the option of selecting 1 or 2 points, and so
on. Up to 10 points can be configured.
4. Click Start CAL to begin the gauge calibration.
5. The Time Remaining (sec) indicator will count down the remaining sample time.
Detector 1 has now been successfully set up for a density application. Once the data
has been saved, click the Home button to return to the first screen of the Setup
Wizard.
Once Detector 1 has been configured for a density application and saved, there are
additional parameters that can be programmed from the EZ Cal II software.
See the section regarding Application for further information.
Communication with the DensityPRO gauges is via an RS232 single-drop serial port,
Communication an RS485 multi-drop serial port, a USB cable, or an Ethernet cable from a PC running
Setup EZ Cal II software. Once the gauge is set up, the primary measurement (density) can
be viewed on the display and on the EZ Cal II software.
To communicate with the gauge from a PC, the PC must be running the Thermo
Scientific EZ Cal II software. To access the Communication Setup screens using only
the gauge display, see Figure A-1.
The serial port on a PC can connect directly to the gauge’s RS232 serial port (Com A
Com A / Com B or Com B).
(RS232) To configure the Com A and Com B ports, select Communication Setup from the EZ
Cal II menu tree.
The default communication settings for the RS232 (Com A and Com B) ports of the
gauge and for the Thermo Scientific EZ Cal II are:
– Unit ID: 1
– Baud Rate: 9600
– Parity: None
– Data Bits: 8
– Stop Bits: 1
1. Enable or disable the selected port.
2. When configuring an RS485 port, assign a unit identification number between 1
and 32.
3. Select the appropriate baud rate from the dropdown list. The dropdown contains
options for 9600, 19200, 38400, 57600 and 115200. The higher the baud rate,
the faster the data transfer.
4. Select Even, Odd or None from the Parity dropdown.
5. Choose the appropriate number of data bits, either 7 or 8.
6. Select 1, 1.5 or 2 stop bits. For most communications, 1 stop bit should be
appropriate.
7. Port Mode:
a. For Com A, select from RS232, RS232 with RTS/CTS and UART Logic
Level.
b. For Com B, select fromRS232, RS232 with RTS/CTS, 2-Wire RS485 and 4-
Wire RS485.
8. Indicate ASCII or RTU as the proper Modbus Protocol.
9. Select the Floating Format, either Normal or Reversed. This selection determines
the order in which bytes and words will be sent.
10. The Rx Count, Tx Count and Error Count fields will auto-populate.
11. After selecting the appropriate parameters, click Submit to save the data, Refresh
to update the data throughout, or Cancel to exit Communication Setup without
saving.
12. Repeat steps 1 – 11 to configure the Com B port, if desired.
RS485 To communicate with multiple gauges via RS485 party line, each unit must be
assigned a unique unit identification number so it can be addressed individually. By
default, all gauges are assigned unit number one (1).
Note: Connecting a PC serial port (COM) to the RS485 port on the gauge
requires an RS485/RS232 converter.
To assign a unique unit number to each gauge, you must be able to communicate with
each one individually. Disconnect each gauge from the party line in turn and
communicate with the disconnected gauge directly. Alternatively, remove power from
all gauges except one and assign a unit number to the powered gauge. Repeat this
procedure for the remaining gauges.
If trouble arises when using another device on the RS485 chain, verify that the device
is properly terminated for its position on the chain. To terminate a device, connect a
120-ohm resistor between its RS485 +/- data terminals. Never terminate more than the
first and last device in the chain.
Each DensityPRO unit includes a 10 Base-T minimum Ethernet port on the Main
Ethernet CPU PCA. Operators should ensure the area is non-hazardous before connecting or
disconnecting the Ethernet cable.
The Communication Setup screens also allow for configuration of the Ethernet port.
16. Select the Floating Format, either Normal or Reversed. This selection determines
the order in which bytes and words will be sent.
17. The MAC address should only be configured or changed by Thermo Scientific
technical support personnel.
Note: For help in establishing the correct information to input onto this screen,
please see your system administrator.
18. Once all information has been entered, click Submit to save the data, Refresh to
update the data throughout, or Cancel to exit Communication Setup without
saving.
The Main CPU PCA includes a USB port, which allows the user to connect to the
USB Port system using a type A Male to Mini 5-pin Male USB cable. Operators should ensure
the area is non-hazardous before connecting or disconnecting the USB cable.
The options under the System Control menu tree in EZ Cal II are not identical to the
System Control options available in the transmitter display screens of the DensityPRO gauges. This
section covers the EZ Cal II System Control screen. For the complete menu tree of the
DensityPRO integrated unit, see Figure A-2.
From the EZ Cal II menu tree, select System Control.
The System Control Configuration screen provides the user with the ability to set up
Configuration system parameters related to items such as the system clock and the LCD screen. The
gauge should be able to read and write any of these parameters as requested by the user.
To access the information on the System Control Configuration screen using only the
remote transmitter keypad, see Figure A-2.
1. Set the system date and time.
a. Choose the desired date format. The available options are:
MM/DD/YY
DD/MM/YY
YY/MM/DD
b. In the System Date textbox, enter the current date.
c. Enter the current time in the System Time textbox, using the format
HH:MM:SS.
To synchronize the date and time of the DensityPRO system with that of the PC
in use, check the Use PC Date/Time textbox.
2. The LCD scroll time dictates the length of time information takes to scroll across
the LCD screen. If the system in use is an integrated system, enter the desired
scroll time in the Integrate Scroll Time (sec) textbox. If the system in use is a
remote system, enter the desired scroll time in the Remote Scroll Time (sec)
textbox.
3. The LCD contrast can be adjusted on the Remote unit by adjusting the LCD
Contrast value. The value can range from 0 – 99.
Note: At this time, it is strongly recommended that configuration information
be saved to your computer. If the work has been completed off-line, connect
to the density gauge via one of the serial ports and upload the configuration
file to the gauge.
The System Status screen auto-populates with information about the system status,
System Status and information setup of the detector.
To access the information on the System Status screen using only the remote
transmitter keypad, see Figure A-4.
The purpose of the mode/fault alarms is to detect change in the alarm status and drive
Mode/Fault the selected current or relay outputs based on the alarm action setup.
Alarm Setup There are six Mode/Fault Alarm tabs.
Note: The Mode/Fault Alarm Setup screens are organized slightly differently
within the gauge. System and application alarm information are often together
on the same display screen. Reviewing the complete Mode/Fault Alarm Setup
map will aid in navigation of the keypad display.
– The System tab, shown in Figure 4-6, allows the user to configure system-wide
alarms.
– The System Status tab, shown in Figure 4-7, displays the alarms from the System
screen and shows whether or not the alarms are currently in an active state.
– The Application tabs (#1 - #4), shown in Figure 4-8, allow the user to configure
application-specific alarms.
Alarms have three operational modes.
– Disable – The alarm is disabled, and no alarm action is executed.
– Enable – The alarm is enabled. The selected alarm action is executed based on the
alarm set condition.
– Inhibit – The alarm is temporary disabled. Another operation may change the
mode of this alarm from Inhibit to Enable.
After selecting an operational mode, the user may select an output source for the alarm
action. The actions available are:
– Do Nothing (Do not take any actions)
– Relay Output A
– Relay Output B
– Current Output A
– Current Output B
– Current Output C
Once an output source has been selected, an alarm action can be selected on either the
Current Output or Relay Outputs screens. For further information on these screens,
see these respective sections.
To access the information on the Mode/Fault Alarm Setup System tab using only the
System Tab remote transmitter keypad, see Figure A-6.
As previously stated, the System Status tab displays the alarms from the System tab and
System Status Tab shows whether or not the alarms are currently in an active state.
To access the information on the Mode/Fault Alarm Setup System Status tab using
only the remote transmitter keypad, see Figure A-7.
The Application tabs are organized so that the alarm’s operational mode is in the left
Application Tabs column, with the options for action on the right. The bottom of the tab includes
information on the status of each alarm, showing whether or not they are currently in
an active state.
To access the information on the Mode/Fault Alarm Setup Application tabs using only
the remote transmitter keypad, see Figure A-6 for the top half of the screen, and Figure
A-7 for the bottom half of the screen.
The system commands provide the user with the ability to directly control certain
Commands functions of the gauge.
Once the desired command or commands have been selected, click the Submit button
to execute. Click the Refresh button to reset the screen or the Cancel button to exit.
To access the information on the Commands tab using only the remote transmitter
keypad, see Figure A-8.
The Common Actions dropdown provides the user a way to change a large group of
Common Action parameters with one command.
Below the Common Action dropdown list on the Commands screen are the Hold
Hold Current Current Output A, B and C dropdown lists. Each of these lists contains the same
Output options for holding the preferred current.
– Do Nothing – No action is taken when the submit button is pressed.
– Hold at minimum – The current output value is held at the minimum value
entered.
– Hold at maximum – The current output value is held at the maximum value
entered.
– Hold at set value – The current output value is held at the set hold value.
– Hold at Fault Low – The current output value is held at the Fault Low value.
– Hold at Fault High – The current output value is held at the Fault High value.
– Clear Hold value – If the Hold/Live value is set to Hold, then it is reset to Live.
The Hold Relay Output A and B sections are for display purposes only and are not
Hold Relay Output configurable by the user.
The Alarm Action dropdown list allows the user to make changes to alarms as a group,
Alarm Action either system-wide or by application.
– Do Nothing – No action is taken when sent to the gauge.
– Clear all alarms for all applications – Currently this option is non-functional and
no action is taken upon its selection. This option will be utilized in future versions
of the EZ Cal II software.
– Disable all alarms – Any process alarm or mode/fault alarm that is in an enabled
state is changed to the inhibit state. This option is available for all applications or
individual applications.
– Enable all alarms – Any process alarm or mode/fault alarm that is in the inhibit
state is changed to an enabled state. This option is available for all applications or
individual applications.
– Erase all process alarm assignments – All configurations associated with setting up
a process alarm are set to the default values. These values include Measurement
ID, Action, Action Delay, Set Point, Clear Point, and the Enable/Disable setting.
This option is available for all applications or individual applications.
– Erase all mode/fault alarm assignments – All configurations associated with setting
up a mode/fault alarm are reset to the default value. Enable/Disable setting,
Action. This option is available for all applications or individual applications.
– Erase all alarm assignments, all mode/fault/process alarms for all applications –
This option allows the user to perform the Erase all process alarm assignments
action and the Erase all mode/fault alarm assignments action simultaneously. This
option is not available for individual applications.
Physical Inputs
& Outputs
The function of an analog input is to connect external signals, such as temperature,
Analog (Current & pressure, flow or density, to the gauge for conversion into a form to be used by the
Voltage) Inputs internal measurement process. Further, the analog input signal in the gauge is mapped
to an input type such as temperature, pressure, flow, or density, and appropriate unit
and measurement ranges are assigned.
The analog inputs offer a way for the user to provide additional process information to
the gauge. This information may be used by the gauge to provided additional types of
measurements.
For a density application, the analog inputs will be used to provide temperature and
flow information. The temperature information may be used for temperature-
compensated density calculation. The flow information may be used for mass and/or
volume flow calculation.
Two types of analog inputs are available on the gauge: current inputs, labeled 4–20 In
1 and 4–20 In 2 on the main board, and voltage inputs, labeled Vdc In 1 and Vdc In 2
on the main board.
Note: All four of the Current/Vdc Input tabs are essentially the same, except
the current tabs reference milliamps and the voltage tabs reference volts.
Current Tabs Within the EZ Cal II software, select Current/Vdc Input from the menu tree on the
left to bring up the tabs containing the analog input tabs. The Current Input #1 (mA
#1) tab will display by default.
Display Description
ºC Degree Celsius
ºF Degree Fahrenheit
Display Description
PSI_G Pounds per square inch gauge
KPA_G Kilopascal
BAR_G Bar
PASCAL_G Pascal
Display Description
g/cc grams per cubic centimeter
lb/US gal pounds per US gallon
lb/UK gal pounds per UK gallon
lb/cu ft pounds per cubic foot
ston/cu yd short tons per cubic yard
lton/cu yd long tons per cubic yard
g/l grams per liter
oz/cu m ounces per cubic meter
lb/cu in pounds per cubic inch
g/cu in grams per cubic inch
lb/cu yd pounds per cubic yd
kg/cu m kilograms per cubic meter
deg API degrees API
degBaum lt degrees Baume light
degBaum hv degrees Baume heavy
degree Twaddle degrees Twaddle
2. Define the input operating range by entering the Minimum mA and Maximum
mA into the corresponding textboxes.
3. In the Min Value @ Min mA textbox, enter the minimum value, in engineering
units, that corresponds to the value entered in the Minimum mA textbox.
4. In the Max Value @ Max mA textbox, enter the maximum value, in engineering
units, that corresponds to the value entered in the Maximum mA textbox.
5. Additional fields on this screen, which are listed below, are for display purposes
only and are not configurable by the user.
Live Value (EU) – This field displays the live calibrated value in engineering
units in both Live and Manual modes.
Raw Value (mA) – This field displays the live calibrated value in milliamps in
both Live and Manual modes.
#1 mA Min Value – This field displays the expected minimum analog input
value provided by the user during calibration.
#2 mA Mid Value – This field displays the expected midpoint analog input
value in a three point calibration provided by the user during calibration.
#3 mA Max Value – This field displays the expected maximum analog input
value provided by the user during calibration.
#1 Raw Min Value – This field displays the measured minimum value read
from the device.
#2 Raw Mid Value – This field displays the measured midpoint value read
from the device.
#3 Raw Max Value – This field displays the measured maximum value read
from the device.
Calibration Points – This field displays the number of points the current
information used to calibrate the current input.
Status – This field displays Live or Manual, depending on the operating mode
selected, as well as low and high alarm conditions.
Calibration All four of the current inputs can be calibrated using two or three calibration points.
The calibration process for the current inputs should be completed by selecting
Calibration from the Functions menu or by clicking the fourth icon button, the
wrench. Selecting either of these options will guide the user through the calibration
process step-by-step. For further information, see Calibration.
Voltage Tabs To access the information on the Voltage Input tabs using only the remote transmitter
keypad, see Figure A-10.
3. In the Min Value @ Min Volt textbox, enter the minimum value, in engineering
units, that corresponds to the value entered in the Minimum Volt textbox.
4. In the Max Value @ Max Volt textbox, enter the maximum value, in engineering
units, that corresponds to the value entered in the Maximum Volt textbox.
5. Additional fields on this screen, which are listed below, are for display purposes
only and are not configurable by the user.
Live Value (EU) – This field displays the live calibrated value in engineering
units in both Live and Manual modes.
Raw Value (Volt) – This field displays the live calibrated value in volts in both
Live and Manual modes.
#1 Volt Min Value – This field displays the expected minimum voltage input
value provided by the user during calibration.
#2 Volt Mid Value – This field displays the expected midpoint voltage input
value in a three point calibration provided by the user during calibration.
#3 Volt Max Value – This field displays the expected maximum voltage input
value provided by the user during calibration.
#1 Raw Min Value – This field displays the measured minimum value read
from the device.
#2 Raw Mid Value – This field displays the measured midpoint value read
from the device.
#3 Raw Max Value – This field displays the measured maximum value read
from the device.
Calibration Points – This field displays the number of points the current
information used to calibrate the voltage input.
Status – This field displays Live or Manual, depending on the operating mode
selected, as well as low and high alarm conditions.
Digital inputs trigger the execution of functions associated with a selected detector’s
Digital Inputs open and close contact actions configured by the user. The two digital inputs available
to the user can be accessed by selecting Digital Input from the EZ Cal II menu tree.
To access the information on the Digital Input tabs using only the remote transmitter
keypad, see Figure A-14.
Action Description
Do Nothing No Action is required. (Default)
1. The standardization has to have been executed at least once
from the user interface before using the discrete input.
2. Start the standardization using the last values entered for
Start STD and use results standardize on (bypass, pipe empty, pipe full), standardization
time, and density.
3. After the standardization has finished accept the average
counts during standardization. The action is ignored if the
standardization has not been executed from the user interface.
Hold Measurement #1@ hold value The value for measurement #1 Hold/Live register is set to hold
Hold Measurement #2@ hold value The value for measurement #2 Hold/Live register is set to hold
Hold Measurement #3@ hold value The value for measurement #3 Hold/Live register is set to hold
Hold Measurement #4@ hold value The value for measurement #4 Hold/Live register is set to hold
Any process alarm and mode/fault alarm in the inhibit state is set to
Enable all alarms
the enable state
Any process alarm and mode/fault alarm in the enable state is set to
Disable all alarms
the inhibit state
Clear relay and totalizers Set all the totalizers value to zero and zero any pending relay outputs
Fast catch-up The filtered data counts are initialized to the raw data counts
Action Description
Note: When a sudden change in raw data counts occurs, the response time of the filtered data
counts is determined by the time constant and may take a significant amount of time to reflect the
average raw data count value. The Fast Catch-up function initializes the filtered data counts to the
value of the raw data counts. This can currently be accomplished by changing the time constant
and setting it back to the previous value.
Dynamic tracking performs a similar function. When the dynamic tracking threshold is exceeded, a
faster time constant is used until the filtered data counts catch up to the raw data counts. The
Fast Catch-up function can be used when dynamic tracking is enabled.
3. Define the output operating range by entering the Min Current and Max Current.
4. In the Min Value @ Min Current textbox, enter the minimum value, in
engineering units, that corresponds to the value entered in the Min Current
textbox.
5. In the Max Value @ Max Current textbox, enter the maximum value, in
engineering units, that corresponds to the value entered in the Max Current
textbox.
6. Choose an operational mode from the Hold/Live dropdown.
Live – The selected Measurement ID is currently driving the current output.
Hold Minimum – The current output is set to the Min Current value.
Hold Maximum – The current output is set to the Max Current value.
Fault High – The current output is set to the high value determined by the
Namur Standard (20.5 mA).
Fault Low – The current output is set to the low value determined by the
Namur Standard (3.8 mA).
Hold Value – The current output is set to the value entered by the user in the
Hold Value (EU) textbox.
Once a value in engineering units have been entered into the Hold Value
(EU) field, the EZ Cal II software will convert that value into milliamps
and display that value in the Present Value field.
7. The Alarm Action dropdown allows the user to designate what happens when an
alarm associated with the selected current output is triggered.
None – Do not take any alarm action.
Set Output to minimum – Sets the current output to the Min Current value.
Set Output to maximum – Sets the current output to the Max Current value.
Set Output to fault low – Sets the current output to the low value established
by the Namur standard (3.8 mA).
Set Output to fault high – Sets the current output to the high value
established by the Namur standard (20.5 mA).
Hold Output at set value – Sets the current output to the value designated in
the Hold Value (EU) field.
Clear Hold value – Sets the operational mode of the current output to Live.
8. Enter the HART ID in the appropriate textbox if a Hart Communication Protocol
is connected to the gauge. This option is only available on the Current Output C
tab.
9. Additional fields on these screens, which are listed below, are for display purposes
only and are not configurable by the user.
Mode – On the Current Output A tab, this field always displays Normal at
this time. On the Current Output C tab, this field displays whether the gauge
is connected to a Hart Communication Protocol.
Live Value (EU) – This field displays the live calibrated value in engineering
units in both Live and Manual modes.
Calibration Points – This field displays the number of points the current
information used to calibrate the current input.
#1 mA Min Value – This field displays the expected minimum analog input
value provided by the user during calibration.
#2 mA Mid Value – This field displays the expected midpoint analog input
value in a three point calibration provided by the user during calibration.
#3 mA Max Value – This field displays the expected maximum analog input
value provided by the user during calibration.
#1 Raw Min Value – This field displays the measured minimum value read
from the device.
#2 Raw Mid Value – This field displays the measured midpoint value read
from the device.
#3 Raw Max Value – This field displays the measured maximum value read
from the device.
Status – This field displays one of the following:
Live – Current output is being driven by the selected measurement ID.
Hold – Current output is being driven based on the Hold Value (EU).
Low Alarm – The current output value is below the Min Current value.
High Alarm – The current output value is above the Max Current value.
Fault Low Alarm – The current output is below the value defined by the
Namur standard.
Fault High Alarm – The current output is above the value defined by the
Namur standard.
Calibration Current outputs can be calibrated using two or three points. The calibration process
for the current outputs should be completed by selecting Calibration from the
Functions menu or by clicking the fourth icon button, the wrench. Selecting either of
these options will guide the user through the calibration process step-by-step. For
further information, see Calibration.
There are two relay outputs on the system that allow the user to provide additional
Relay Outputs process information from the gauge to the outside world. The function of a relay
output is to provide alarm signal and/or totalizer pulses to an external system for
monitoring or processing purposes.
To access the information on the Digital Input tabs using only the remote transmitter
keypad, see Figure A-16.
Hold Relay ON – When the alarm assigned to the relay is active, the relay
will be energized. When the alarm assigned to the relay is inactive, the
relay will return to a de-energized state.
Hold Relay OFF – When the alarm assigned to the relay is active, the
relay will be de-energized. When the alarm assigned to the relay is
inactive, the relay will return to an energized state.
Clear Hold – The Hold/Live field will be set to Live mode when the
alarm assigned to the relay is active, and it will remain in Live mode once
the alarm becomes inactive.
b. If Totalizer is selected from the Process Function dropdown, the Total
Selection dropdown allows the user to select a totalizer.
i. If no totalizer is selected from the Total Selection dropdown, the relay
remains at its default state.
ii. When a totalizer is selected from the Total Selection dropdown, the
indicated totalizer associated with the detector designated in the Detector
ID field will drive the relay.
4. Enter a pulse width value in the Pulse Width (30-200 ms) field.
a. Entering 0 will generate no relay pulse output.
b. A value of 30 – 200 bases the relay pulse output on pulse width at 10 ms
intervals.
Note: If the process function is set to Totalizer, the relays will operate at a
maximum frequency with a period of twice the pulse width.
The gauge outputs totalizer pulses on a relay as long as there are pulses
available from the selected totalizer – even when the relay pulses accumulate at
a faster rate than they can be output.
5. Selecting No from the Invert Output dropdown maintains the output of the relay
in its normal state. Selecting Yes from the dropdown will invert the output of the
relay.
6. Additional fields on this screen, which are listed below, are for display purposes
only and are not configurable by the user.
Present Value – This field displays the current live status (On or Off) of the
relay.
Status – This field displays Live or Hold, depending on the selected
operational mode.
From the EZ Cal II menu tree, click Detector to view all of the detector counts in a
Detector graphic representation.
1. Select the number of detectors to view on the graph by using the Detector
indicator. Each detector will be represented by a different color on the graph.
2. The Update Period (ms) field allows the user to determine the update rate of the
counts. The entry in this field can range from 1 – 2000 ms.
3. Click Run Char to begin a real-time graphic view of the detector counts.
4. Click Freeze Char to pause the detector count chart.
From the EZ Cal II menu tree, click Detector #1. By default, the Setup tab is
Detector displayed.
Screens
To access the information on the Detector Setup tab using only the remote transmitter
Setup Tab keypad, see Figure A-3, Figure A-5, Figure A-17.
Detector Setup Many of the fields on the Detector Setup tab provide the user with the ability to set up
the detector parameters based on application-specific requirements.
1. Enter the density time constant, in seconds, in the Density Time Constant (sec)
textbox.
2. Choose whether to enable or disable dynamic tracking from the Dynamic
Tracking Enable dropdown.
a. Enabling dynamic tracking ensures rapid, smooth filtered counts in response
to significant changes in the process level. The dynamic tracking system
contains a slow filter and a fast filter. Initially, filter data counts use the slow
filter. The fast filter uses a normal averaging time constant. The difference
between the data counts of the slow and fast filters are monitored constantly.
When the difference exceeds the dynamic tracking threshold percentage,
which has a default value of 5.3% of filtered counts, dynamic tracking is
activated. The output then switches to the faster filter, with the fast signal
using the value of slow signal as a starting point. After one fast filter time
constant, the difference of the slow and fast filters is calculated. If the
difference falls below the threshold value, dynamic tracking is de-activated and
the output switches back to the slow filter, with the slow signal using the value
of fast signal as a starting point.
Detector The detector configuration parameters on the Detector Setup tab provide the user with
the ability to view and configure high voltage power supply parameters for the
Configuration detector. The gauge has the ability to read and write any of these parameters as
requested by the user. When the system is shut down, the last stable high voltage
control information will be used at the next startup.
Each gauge can utilize up to two high voltage power supplies.
1. The Force Hi Voltage #0 Hold Value dropdown provides the user with the option
to place the detected high voltage power supplies in hold mode.
a. If Yes is selected, the system will hold the high voltage at the value designated
in the Hi Voltage #0 Value textbox.
b. If No is selected, the system will use live system data as the Hi Voltage #0
Value.
2. The following fields are intended for display only, however, the user has the ability
to modify the values in these fields.
Last HV Control Time Constant
Last CPLD Window Count
Last Board Temp ºC
Last Hi Voltage #0 Control
Last Hi Voltage #0 Setup
Min CPLD Channel 1 Window Count
Warning: Making changes to these parameters may cause the system to
become unstable.
The Count tab displays the specified detector counts in a graphic representation.
Count Tab
There is one available current input per detector. Information regarding the current
Current Input Tab input can be found on the Current Input tab.
The RTD Input tab is used for temperature compensated density measurement or to
RTD Input Tab monitor the process temperature.
To access the information on the Detector RTD Input tab using only the remote
transmitter keypad, see Figure A-13.
Calibration RTD inputs can be calibrated using two or three points. The calibration process for
the RTD inputs should be completed by selecting Calibration from the menu or by
clicking the fourth icon button, the wrench. Selecting either of these options will guide
the user through the calibration process step-by-step. For further information, see
Calibration.
The Diagnosis Tab displays values to the user for diagnostic purposes.
Diagnostics Tab
To access the information on the Detector Diagnosis tab using only the remote
transmitter keypad, see Figure A-5.
Each set of Application screens relates to the Detector screens of the same number. For
Application instance, information entered in the screens accessed by selecting Application #1 from
the menu tree will be applied to Detector #1, whereas information entered on the
Application #2 screens will be applied to Detector #2.
Density Setup The Density Setup tab provides the user with a way to configure the primary
measurement. This can also be accomplished by using the Setup Wizard. The Setup
Wizard can be accessed through the Functions dropdown menu or by clicking the blue
wizard’s cap icon near the top right corner of the screen.
To access the information on the Application Density Setup tabs using only the
remote transmitter keypad, see Figure A-17.
Carrier content/vol
Solids/carrier
% by weight solids
% by weight carrier
% by volume solids
% by volume carrier
b. If Single phase is selected as the material, the primary measurements available
are:
Density
Bulk Density
c. If Emulsion is selected as the material, the primary measurements available are:
Density
Fluid 2 Content/vol
Fluid 1 Content/vol
Fluid 2/Fluid 1
% by weight Fluid 2
% by weight Fluid 1
% by volume Fluid 2
% by volume Fluid 1
d. If Solution is selected as the material, the primary measurements available are:
Solute content/vol
Solvent content/vol
Solute/Solvent
% by weight Solute
% by weight Solvent
% by volume Solute
% by volume Solvent
Bulk Density
5. Select a primary unit of measure from the Density Unit (Primary) dropdown. The
selection made on this tab will provide the unit information for the Action and
Measurement Data tabs.
g/cc – gram/cubic centimeter
lb/US gal – pound/US gallon (U.S. liquid)
lb/UK gal – pound/UK gallon (UK or imperial gal.)
lb/cu ft – pound/cubic foot
ston/cu yd – short ton (2000 pounds)/cubic yard
lton/cu yd – long ton (2240 pounds)/cubic yard
g/l – gram/liter
oz/cu in – ounce/cubic inch
lb/cu in – pound/cubic inch
g/cu in – gram/cubic inch
lb/cu yd – pound/cubic yard
kg/cu m – kilogram/cubic meter
deg API – degree American Petroleum Institute
degBaum lt – degree Baume, light scale
degBaum hv – degree Baume, heavy scale
6. Next, specify the unit of measure for the inside diameter of the pipe. The available
selections are:
cm – centimeter
mm – millimeter
in – inch
ft – foot
yd – yard
m – meter
7. Finally, enter the inside diameter of the pipe in the Pipe Inside Diameter textbox
and click the Submit button to save the data.
Input Configuration The Input Configuration Setup tab let the user assign the type of input connected to
the DensityPRO gauge and to set that input’s operating range for use with additional
Setup parameters in calculations. The user can assign analog inputs to the density,
temperature, pressure and/or flow parameters, as well as the units in which those
parameters should be displayed.
The information entered on this tab affects the information displayed on the
Additional Measurement tab. If analog input Main Board 4-20mA #1 is assigned to
Temperature Input and the Temperature Unit is set to ºF on this tab, choosing
Temperature from one of the Measurement # dropdowns on the Additional
Measurement tab will associate the input with that measurement number on the
Action and Measurement Data tabs.
While the Input Configuration tab allows the user to link four analog and
measurement inputs, there are only three measurements available on the Additional
Measurement tab.
To access the information on Application Input Configuration Setup tabs using only
the remote transmitter keypad, see Figure A-17 and Figure A-18.
Input
Available Selection
Density Temperature Pressure Flow
Not Used/Calculated
Main Board 4-20mA #1
Main Board 4-20mA #2
Main Board 0-10V #1
Main Board 0-10V #2
Detector Board 4-20mA input
Detector Board RTD input
2. Choose engineering units from the dropdown that corresponds with the selected
input type. The following tables provide the units available for selection based on
the various types of input.
Table 4-7. Density Input, Density Units
Display Description
g/cc grams per cubic centimeter
lb/US gal pounds per US gallon
lb/UK gal pounds per UK gallon
lb/cu ft pounds per cubic foot
ston/cu yd short tons per cubic yard
lton/cu yd long tons per cubic yard
g/l grams per liter
oz/cu m ounces per cubic meter
lb/cu in pounds per cubic inch
g/cu in grams per cubic inch
lb/cu yd pounds per cubic yd
kg/cu m kilograms per cubic meter
deg API degrees API
degBaum lt degrees Baume light
degBaum hv degrees Baume heavy
degree Twaddle degrees Twaddle
Display Description
ºC Degree Celsius
ºF Degree Fahrenheit
Display Description
g/cc grams per cubic centimeter
lb/US gal pounds per US gallon
lb/UK gal pounds per UK gallon
lb/cu ft pounds per cubic foot
Temperature The Temperature Compensation Setup tab allows the user to establish how the process
temperature is provided to the system.
Compensation Setup
To access the information on the Application Temperature Compensation Setup tabs
using only the remote transmitter keypad, see Figure A-18.
Additional There are a total of four measurements available on a DensityPRO gauge. The gauge
can convert the basic density measurement into a variety of output measurements
Measurement Setup appropriate for specific applications based on the type of material selected. Given
temperature input, the gauge can compensate the density measurement for changes in
the process temperature. If flow input is provided via additional inputs, the gauge is
capable of calculating a mass flow or a volume flow.
The purpose of providing additional measurements is to allow the user to configure a
second, third and forth measurement as appropriate for a specific application.
To access the information on Application Additional Measurement Setup tabs using
only the remote transmitter keypad, see Figure A-18.
Note: Additional measurements are closely linked with an application’s input
configuration setup. See the section Input Configuration Setup for additional
details.
Note: The Analog Input Type must be configured correctly for each selection
made on the Input Configuration Setup tab. See the section Analog (Current
& Voltage) Inputs for additional details.
The Additional Measurement tab allows the user to assign measurement inputs to
measurements two, three and four.
1. The Material Type will auto-populate based on information configured on the
Density Setup screen.
2. Measurements #2, #3 and #4 allow the user to choose the measurement type.
a. The Density measurement is derived from the primary density measurement
or the measurement from density information on the Input Configuration tab.
b. The Solids content/volume measurement is derived from the primary density
measurement value if no additional density information is configured on the
Input Configuration tab.
c. The Carrier content/volume measurement is derived from the primary density
measurement value if no additional density information is configured on the
Input Configuration tab.
d. The Solids/carrier measurement is derived from the primary density value and
is primarily used for oilfield applications.
e. The % by weight solids measurement is derived from the primary density
measurement.
f. The % by weight carrier measurement is derived from the primary density
measurement.
g. The % by volume solids measurement is derived from the primary density
measurement.
h. The % by volume carrier measurement is derived from the primary density
measurement.
i. The Bulk Mass Flow measurement is derived from the primary density
measurement and the flow measurement selected from the Input
Configuration tab.
j. The Solids Mass Flow measurement is derived from the primary density
measurement and the flow measurement selected from the Input
Configuration tab.
k. The Carrier Mass Flow measurement is derived from the primary density
measurement and the flow measurement selected from the Input
Configuration tab.
l. The Bulk Volume Flow measurement is derived from the flow measurement
selected from the Input Configuration tab.
m. The Solids Volume Flow measurement is derived from the primary value,
Carrier Density constant, Solid Density constant and in conjunction with the
flow measurement selected from the Input Configuration tab.
n. The Carrier Volume Flow measurement is derived from the primary density
value, Carrier Density constant, Solid Density constant and in conjunction
with the flow measurement selected from the Input Configuration tab.
o. The Bulk Solids Flow measurement is derived from the primary density value,
Carrier Density value, Solid Density value, Bulk Solid Density constant and in
conjunction with the flow measurement selected from the Input
Configuration tab.
p. The Temperature measurement is derived from the analog input selection
made in Input Configuration.
q. The Velocity measurement value is derived based on the flow selection made
under Input Configuration and based on the pipe inside diameter. The
measurement will be automatically converted to the user-selected unit, either
ft/sec or m/sec.
r. The Proppant measurement is derived from the primary density value, the
primary density constant carrier density value and the solid density value.
3. Each measurement unit will auto-populate based on the measurement type
selected.
4. Use the fields under Number of Decimal Places to determine the number of
decimal places to display. These fields are configurable from 0 to 4.
5. In each instance where the Density, Temperature, Pressure or Flow Inputs are set
to Not Used, then Measurement #1 will be used on the Measurement Data tab.
General Setup The General setup screen allows the user to read and modify parameters associated
with the system. The DensityPRO gauge will be read and will write any of these
parameters as requested by the user.
To access the information on Application General Setup tabs using only the remote
transmitter keypad, see Figure A-3.
Using as a Default By default, the gauge uses carrier density as the calibration (CAL) point. For some
applications, this default CAL point may provide adequate measurement accuracy
Calibration Value without performing any additional calibration measurements. For example, if the
standardization is performed on a pipe full of clean carrier (for a slurry material type)
and solids concentration is selected as the primary measurement, the measurement
readout should be reasonably accurate.
To access the information on Application Standardization tabs using only the remote
transmitter keypad, see Figure A-18.
Accept STD – Accept the standardization value before or after the cycle
completes. When the user accepts the standardization value before the cycle
completes, the date, time and value of the standardization counts at the time
of acceptance are stored in the gauge.
Reject STD – Reject the standardization value after completion of the
standardization cycle. If the user rejects the standardization, the value of the
standardization counts and the standardization date and time stored in the
gauge are unchanged.
Note: The high voltage must be stable during the standardization cycle. If the
high voltage is unstable, the cycle will abort.
CAL Data Tab 1. Select a method of density calibration from the CAL Method dropdown.
Polynomial
Breakpoint Table
2. Choose the point to be calibrated using the CAL Point field. Initially, the only
selection available is 1. Once Point 1 has been calibrated and accepted, Point 2 will
become available for selection from the dropdown. The CAL Point field will allow
the user to calibrate up to 10 points.
Note: When a standardization cycle is run and accepted on a full pipe, the
system uses this data as Point 1 for calibration purposes. The user will then
have the option of recalibrating Point 1 or calibrating Point 2.
3. Calibration counts are averaged over a user-defined period of time. Enter a time
period, in seconds, in the CAL Sample Time (sec) textbox. The time can range
from 1 second to 65,535 seconds.
4. Enter a calibration density value. The units displayed are those associated with the
primary density measurement.
5. The following commands are available for the gauge calibration process:
Start CAL – Begin the calibration cycle.
Abort CAL – Abort or reject the calibration cycle before cycle completion.
When the user terminates the calibration cycle, the value of the calibration
counts and the density stored in the gauge remain unchanged.
Accept CAL – Accept the calibration value before or after the cycle completes.
When the user accepts the calibration value before the cycle completes, the
values of the calibration counts and the density at the time of acceptance are
stored in the gauge.
Reject CAL – Reject the calibration value after completion of the calibration
cycle. If the user rejects the calibration, the value of the calibration counts and
the density stored in the gauge are unchanged.
6. The fields listed below are for display purposes only and are not configurable by
the user.
Time Remaining (sec)
Detector Average Count
Calibration/Reference Latest
Calibration Temperature
Figure 4-31. Application Screen, Gauge Calibration Point Data Tab (Polynomial)
1. Selecting Polynomial as the density calibration method on the CAL Data tab
introduces a distinct set of fields on the Point Data tab of the Gauge Calibration
screen.
The accepted calibration values calculated on the CAL Data tab will auto-populate
the fields listed below on the Point Data tab. The user may also manually change
these values by entering new data into the textboxes.
Cal Density Point #1
Cal/Ref for Point #1
Count Rate for Point #1
Cal Density Point #2 (If Point 2 is calibrated)
Cal/Ref for Point #2 (If Point 2 is calibrated)
Count Rate for Point #2 (If Point 2 is calibrated)
Slope Correction factor
2. The slope correction factor is automatically calculated by the gauge using the
calibration points accepted by the user. However, changes to data fields, both on
the Point Data tab and Density Setup tab, will affect the slope correction factor. If
any of the fields listed below are changed, clicking the Calculate Slope Correction
Factor button will recalculate the slope correction factor based on the current data.
The result of this calculation will then be displayed in the Slope Correction factor
textbox.
Point Data Tab Selecting Breakpoint Table as the density calibration method on the CAL Data tab,
introduces a different set of fields on the Point Data tab of the Gauge Calibration
(Breakpoint Table) screen. The breakpoint table displays the density of each calibrated point, as well as the
point’s count rate and calibration/standardization ratio.
Figure 4-32. Application Screen, Gauge Calibration Point Data Tab (Breakpoint Table)
Additionally, this tab allows the user to view displayed point data, to change the point
data by manually entering new values, and to submit or delete point data.
The purpose of the totalizers is to calculate the amount of mass or volume flow
Totals measured by a gauge over a period of time. There are four totalizers available for each
detector, and each totalizer can be configured to totalize mass or volume flow.
To access the information on Application Totals tabs using only the remote
transmitter keypad, see Figure A-19.
The following outline provides basic steps for configuration the totalizer. Each totalizer
can be setup to totalize on the same or different measurements.
Totalizers #1 – #4 1. Choose an action state for Totalizer #1 from the Selection dropdown.
Disable – If the totalizer is disabled, the totalizer value stops accumulating,
even when the value of the input is greater than or equal to the threshold
limit. If the totalizer has not been previously enabled, the Totalizer Value field
will be set to zero.
Enable – This allows the totalizer to start accumulating when the value of the
input is greater than or equal to the threshold limit. If the totalizer has been
enabled previously, the Totalizer Value resumes accumulating at the previous
value.
Clear – The Totalizer Value and any pending relay counts are set to zero. The
totalizer returns to its previous state of disabled, enable or inhibit.
Inhibit – The Inhibit state is for use with the Enable All action. When the
Enable All action is requested, all Totalizers in the Inhibit state will be
changed to the Enable state. All Totalizers that are in the Disable state will
stay disabled.
2. Make a selection from the Input dropdown to designate which input to totalize.
None – No input is selected to totalize. The Totalizer Value will not increase.
Measurement #1: The value of Measurement #1 is totalized. This will be
either the Live or Hold value, depending on what the Hold/Live field is set to
for Measurement #1 on the Application Action tab.
Measurement #2: The value of Measurement #2 is totalized. This will be
either the Live or Hold value, depending on what the Hold/Live field is set to
for Measurement #2 on the Application Action tab.
Measurement #3: The value of Measurement #3 is totalized. This will be
either the Live or Hold value, depending on what the Hold/Live field is set to
for Measurement #3 on the Application Action tab.
Measurement #4: The value of Measurement #4 is totalized. This will be
either the Live or Hold value, depending on what the Hold/Live field is set to
for Measurement #4 on the Application Action tab.
3. Select a measurement unit from the Totalizer Value dropdown. The available units
are:
US gallon
Imperial gallon
Cubic centimeter
Cubic meter
Cubic inch
Cubic feet
Cubic yard
Ounces
Pounds
Short tons
Long tons
Kilograms
Metric tons
Grams
4. The totalizers give the user the option to link measurement data to a totalizer by
selecting Measurement # from the Input dropdown. The measurement value of
the specified Measurement # will then be input into the totalizer. The value of the
input must be greater than or equal to the value entered in the Threshold Limit
field before the totalizer value will accumulate. When this condition is met, the
total of the input value is accumulated.
5. Enter a value that represents the magnitude of each pulse output in the
Units/Pulse textbox.
6. The following fields are for display purposes on this screen, though their data may
be used in other processes. These fields are not configurable by the user.
Threshold Unit – The unit of measure associated with the Input. The
Threshold Limit should use the same units.
Totalizer Value – Displays the current value accumulated by the totalizer since
it was last cleared. When a totalizer is active, the value updates every second.
Any flow that occurred while the totalizer was disabled or inhibited is not
included in the value, nor is any flow that occurred when the input value was
below the Threshold Limit.
Note: The Totalizer Value will roll over at 1,000,000.
All Totalizer Choosing an action state from the Selection dropdown under All Totalizer at the
bottom of the Totals tab will apply the selected state to any of the configured totalizers.
– Disable – The Disable state in the All Totalizer Selection dropdown sets all four
totalizers to a disabled state. In a disabled state, the totalizer values stop
accumulating, even when the values of the inputs are greater than or equal to the
threshold limits. If a totalizer has not been previously enabled, the Totalizer Value
field in that totalizer’s area will be set to zero.
– Enable – When the Enable state is chosen from the All Totalizer Selection
dropdown, all Totalizers in the Inhibit state will be changed to the Enable state.
This allows the totalizers to begin accumulating when the values of the inputs are
greater than or equal to the threshold limits. If a totalizer has been enabled
previously, the Totalizer Value resumes accumulating at the previous value.
– Clear – For all of the totalizers, the Totalizer Value and any associated relay counts
are set to zero when the Clear state is selected. The totalizers then return to their
previous state of disabled, enable or inhibit.
– Inhibit – When Inhibit is selected from the All Totalizer Selection dropdown, all
enabled totalizers will change to an inhibit state. All totalizers that are in the
Disable state will stay disabled.
The Application Action tab provides the user the ability to hold the measurements
Action configured on the Additional Measurement Setup tab at the desired value. The
following options apply to Measurements #1 – #4.
To access the information on Application Action tabs using only the remote
transmitter keypad, see Figure A-20.
The Measurement Data tab displays the live measurement values and configured
Measurement Data measurement units for Measurements #1 – #4 to allow for further analysis. These fields
are not configurable by the user.
To access the information on Application Measurement Data tabs using only the
remote transmitter keypad, see Figure A-19.
The process alarms provide a signal to the external device when the process value goes
Process Alarm above or below a set point value. The gauge provides sixteen process alarms. The
system can be configured to setup a process alarm based on one of the four
measurements. An alarm is activated when the measurement value reaches the specified
set point. The relative values assigned to the set point and clear point determines
whether the alarm is a high alarm or low alarm. The alarm condition must exist for at
least the duration of action delay time for the alarm to trigger and take the selected
action.
To access the information on Application Process Alarm tabs using only the remote
transmitter keypad, see Figure A-20.
High and Low Alarms If the set point value is greater than the clear point, the alarm is a High alarm. In this
case, the alarm is activated as the measurement value increases above the set point
value. The alarm stays active until the measurement value again goes below the clear
point. If the clear point value is greater than the set point, then the alarm is a Low
alarm. In this case, the alarm is activated as the measurement value goes below the set
point. The alarm stays active until the measurement value again goes above the clear
point.
In the high alarm depiction below, the measurement rises above the set point and
remains there longer than 10 seconds, the time entered in the Action Delay (sec)
textbox. Therefore, after 10 seconds, the alarm is triggered and continues until the
measurement drops below the clear point. In the second instance, the measurement
does not rise above the set point for 10 seconds, so the alarm is not triggered.
In the low alarm depiction, the measurement drops below the set point and remains
there longer than 10 seconds, the time entered in the Action Delay (sec) textbox.
Therefore, after 10 seconds, the alarm is triggered and continues until the
measurement rises above the clear point. In the second instance, the measurement does
not drop below the set point for 10 seconds, so no alarm occurs.
Alarm Functions The function of the process alarms is to provide a critical signal to the gauge based on
the user’s configuration. There are a total of sixteen alarms available per detector. Each
alarm can be set to trigger on any one of the four measurements. The process alarms
report high-limit alarms and low-limit alarms based on the set and clear point values
and on the action delay time set by the user.
1. Choose an operational mode from the Setup dropdown.
Disable – The alarm is disabled; no alarm action is executed.
Enable – The alarm is enabled. The alarm action will be executed based on Set
Point and Clear Point conditions.
Inhibit – The alarm is temporary disabled. An alarm action may change the
setup of this alarm from Inhibit to Enable.
2. Assign the alarm to a measurement by making a selection from the Measurement
ID dropdown. The alarm will then monitor the selected measurement and trigger
the alarm based on the set point and clear point. The following selections are
available in the Measurement ID dropdown:
None
Measurement #1
Measurement #2
Measurement #3
Measurement #4
3. The alarm is only triggered if the alarm condition exists after a user-defined
amount of time. If the alarm condition does not exist for the designated time, the
alarm will not trigger. Enter the desired wait time before the alarm triggers in the
Action Delay (Sec) textbox.
4. Select an alarm action from the Action dropdown. The available options are listed
below.
a. Do Nothing – No Action is taken.
b. Relay Output A
i. At the Set Point, action must be taken based on the Alarm Action setting
located on the Relay Output A tab.
ii. At the Clear Point, relay operation is reset to Normal operation.
c. Relay Output B
i. At the Set Point, action must be taken based on the Alarm Action setting
located on the Relay Output B tab.
ii. At the Clear Point, relay operation is reset to Normal operation.
d. Current Output A
i. At the Set Point, action must be taken based on the Alarm Action setting
located on the Current Output A tab.
ii. At the Clear Point, Current Output A is set back to normal operation.
e. Current Output B
i. At the Set Point, action must be taken based on the Alarm Action setting
located on the Current Output B tab.
ii. At the Clear Point, Current Output B is set back to normal operation.
f. Current Output C.
i. At the Set Point, action must be taken based on the Alarm Action setting
located on the Current Output C tab.
ii. At the Clear Point, Current Output C is set back to normal operation.
g. When Clear all Holds is selected, all Hold/Live values are set to Live for the
following items:
Measurements #1 through #4
Current Outputs A through C
Relays A and B
h. Clear all alarms (Not Applicable)
i. No action is taken when sent to the gauge. This will be used in future
applications to clear latching alarms.
i. Clear all Totalizers (Do not enable):
i. The value of totalizers 1 through 4 are set to zero and the totalizers are
disabled.
j. Inhibit all Totalizers
i. Any totalizer (totalizers 1 through 4) in the enable state will be placed in
the inhibit state.
k. Enable all Totalizers
i. Any totalizer in the inhibit state will be placed in the enable state.
l. Inhibit totalizer 1
i. If totalizer 1 is in the enable state, it will be placed in the inhibit state.
m. Enable totalizer 1
i. If totalizer 1 is in the inhibit state, it will be placed in the enable state.
n. Zero totalizer 1
i. Set the totalizer count to zero while the alarm is active. The totalizer will
start accumulating when the alarm condition clears.
o. Inhibit totalizer 2
i. If totalizer 2 is in the enable state, it will be placed in the inhibit state.
p. Enable totalizer 2
i. If totalizer 2 is in the inhibit state, it will be placed in the enable state.
q. Zero totalizer 2
i. Set the totalizer count to zero while the alarm is active. The totalizer will
start accumulating when the alarm condition clears.
r. Inhibit totalizer 3
i. If totalizer 3 is in the enable state, it will be placed in the inhibit state.
s. Enable totalizer 3
i. If totalizer 3 is in the inhibit state, it will be placed in the enable state.
t. Zero totalizer 3
i. Set the totalizer count to zero while the alarm is active. The totalizer will
start accumulating when the alarm condition clears.
u. Inhibit totalizer 4
i. If totalizer 4 is in the enable state, it will be placed in the inhibit state.
v. Enable totalizer 4
i. If totalizer 4 is in the inhibit state, it will be placed in the enable state.
w. Zero totalizer 4
i. Set the totalizer count to zero while the alarm is active. The totalizer will
start accumulating when the alarm condition clears.
5. Enter a set point value, in the same units as the selected measurement, in the Set
Point textbox.
6. Enter a clear point value, in the same units as the selected measurement, in the
Clear Point textbox.
7. The following fields display values that are available to be used in other processes.
These fields are for display purposes and are not configurable by the user on this
screen.
Unit – Displays the units of the selected measurement.
Alarm Status – Displays Set or Clear, based on whether the alarm is active or
not.
The DensityPRO gauges have the ability to calibrate all physical inputs and outputs.
Calibration To begin input/output calibration, select Calibration from the Functions dropdown
menu at the top of the screen or click the fourth icon button, the wrench.
To access the information on Input/Output Calibration using only the remote
transmitter keypad, see Figure A-9 through Figure A-15. Beginning with Figure A-9,
each Physical I/O subsection has screens for calibration.
1. Select an output from the list. The available outputs for calibration are:
Output Calibration
Current Output A
Current Output B
Current Output C
2. Choose a 2-point or 3-point calibration.
3. Click the Next button to capture the calibration minimum point.
Note: A digital multi-meter must be connected to the output for accurate
calibration.
b. To recalculate the point, click Start Calibrate before clicking the Submit
button.
c. If the point data is satisfactory, click Submit to save and Next to calibrate the
next point.
In a 2-point calibration, the next screen will calibrate the maximum
values.
In a 3-point calibration, the next screen will calibrate the mid-point
values. Submit that information and click the Next button to proceed to
the maximum point value screen.
7. Follow steps 4 – 6 to calibrate the remaining point(s).
PASSWORD VALIDATION
USB PASSWORD MODE
B DISPLAY
SMART PASSWORD MODE
B DISPLAY
CURRENT PASSWORD MODE
B DISPLAY
BACK NEXT
PASSWORD ENTRY PASSWORD SETUP ENGINEER PASSWORD
PASSWORD
ENGINEER B 0
PASSWORD VALIDATION TECHNICIAN CHANGE TIMES
PASSWORD SETUP OPERATOR B 0
MODE COUNT
B 0
BACK NEXT BACK NEXT BACK EDIT
PASSWORD SETUP TECHNICIAN PASSWORD
PASSWORD
ENGINEER B 0
TECHNICIAN CHANGE TIMES
OPERATOR B 0
MODE COUNT
B 0
BACK NEXT BACK EDIT
PASSWORD SETUP OPERATOR PASSWORD
PASSWORD
ENGINEER B 0
TECHNICIAN CHANGE TIMES
OPERATOR B 0
MODE COUNT
B 0
BACK NEXT BACK EDIT
Figure A-2. System Control (Sheet 1 of 2)
HIGH VOLTAGE (HV)
LAST HV
CONTROL TC
B 2
MIN CPLD 1
WINDOW COUNT
B 200
BACK NEXT
DIAGNOSIS MAIN BOARD
NaI MAIN Err STATUS
TEMPERATURE B Sys None
HIGH VOLTAGE (HV) B SysInit None
MAIN BOARD B RUNTIME None
MAIN MENU MAIN MENU
COMMUNICATION SETUP
SYSTEM CONTROL COMMANDS
SYSTEM STATUS PHYSICAL I/O
Mode/Fault Alarm Setup DETECTOR
MORE… PREV…
BACK NEXT BACK NEXT BACK EDIT NEXT BACK EDIT NEXT
Figure A-17. Detector (Sheet 1 of 4)
1) A-
2) A
1. Use a three-wire cable to connect PC-to- J2A of the Main CPU PCA. The cable
Programming the connections are shown in the table below.
Main CPU Table B-1. RS232 Local Port Connection
The basic density calculation converts the measured intensity of the radiation to a
Basic Density density value. This calculation is the basis for all of the various density calculations
Calculation used in the gauge. The terms used to calculate density consist of constants, user entered
values, measured values, and calculated values.
– Constants – These terms are fixed values that the user may be able to read.
G: Geometry factor (0.85) – The geometry factor does not have any units. It is
used to correct for the location of the source relative to the detector with
different source heads and mounting configurations. History has determined
that this value can be a constant.
– User Inputs – These are values entered by the user. The user shall be able to read
and write the value for the items in this section.
L: Pipe ID – The inside diameter of pipe in centimeters (cm). The default
value is 0 cm.
Dc: Density of the carrier – The density of the carrier fluid in grams per cubic
centimeter (g/cm3). The default value is 1 g/cm3.
µc: Attenuation Coefficient of the carrier – The attenuation coefficient of the
carrier fluid in square centimeters per gram (cm2/g). The default value is
0.086 cm2/g.
Ds: Density of the solid – The density of the solid material in grams per cubic
centimeter (g/cm3). The default value is 2.65 g/cm3.
µs: Attenuation Coefficient of the solid – The attenuation coefficient of the
solid material in square centimeters per gram (cm2/g). The default value is
0.077 cm2/g.
Ibkg: Background radiation – The radiation measured at the detector with the
shutter on the source closed in counts per second (cps). The default value is 3
cps.
The background radiation is determined by measuring the radiation intensity
at the detector with the shutter closed and using the following equation.
Ibkg (cps) = Measured Radiation Intensity (mr/hr) x 300 (cps/(mr/hr))
Dcal1: Density for calibration point 1 – The density of the process material in
grams per cubic centimeters (g/cm3) during calibration. The default value is 0
g/cm3.
Dcal2: Density for calibration point 2 – The density of the process material in
grams per cubic centimeters (g/cm3) during calibration. The default value is 0
g/cm3. Refer to Slope Correction Factor for the use of this value.
– Measured Values – These are the process values read by the gauge.
I: Radiation Intensity – The intensity of the radiation from the detector in
counts per second (cps).
Ical1: Radiation Intensity for calibration point 1 – The intensity of the
radiation from the detector in counts per second (cps) when measuring the
density of calibration point 1 (Dcal1).
Ical2: Radiation Intensity for calibration point 2 – The intensity of the
radiation from the detector in counts per second (cps) when measuring the
density of calibration point 2 (Dcal1). Refer to Slope Correction Factor for
the use of this value.
Istd: Radiation Intensity during standardization – The intensity of the
radiation from the detector in counts per second (cps) during the
standardization. Refer to Standardization Term for additional information.
– Calculated Values – These are values the gauge calculates using the constants, user
inputs and measured values.
D: Density – The bulk density of the process material.
S: Slope Correction Factor – A unit less correction used to improve the
accuracy of the density calculation after a second calibration point is entered.
The user shall also be able to read and write this value.
To calculate the basic density, follow Equation 1.
In
1
1
Equation 1
The standardization term is the average radiation intensity in counts per second over a
Standardization period of time. This value should be measured with a repeatable point in the process.
Term An example of a repeatable point in the process is an empty pipe or a pipe full of a
material with a known density. The standardization term can be used to correct for
buildup in the pipe or pipe wear.
To include the standardization term in the basic density calculation, the radiation
intensity (I), background intensity (Ibkg), and calibration intensity (Ical) are divided
by the standardization intensity (Istd).
This results in the following equation.
In
1
1
Equation 2
To improve the accuracy of the density calculation, the slope correction factor (S) is
Slope Correction included in the density calculation. The slope correction factor is calculated when the
Factor second calibration point is complete. To obtain the information for the second
calibration point, the user must enter the second calibration density and the gauge
must measure the radiation intensity for the second calibration density.
The user has the option to edit the slope correction factor. Additionally, if any of the
terms used to calculate the slope correction factor are modified after calibrating the
second calibration point, the slop correction factor can be recalculated.
The following equation is used to calculate the slope correction.
∗ 1 2
2
In
1
Equation 3
The following is the equation to calculate density including slope correction factor.
General Density
Calculation
Equation 4
The density for a single-phase material should be calculated using Equation 4. Initially
Single-Phase the slope correction factor will be 1. When a second calibration point is used, the slope
Density correction factor should be calculated using Equation 3. The user shall have the option
to modify and/or recalculate the slope correction factor at any time.
Calculation
The parameters below are used to calculate the density of a single-phase material.
– Constants
G: Geometry factor (Default = 0.85) – The geometry factor does not have any
units. It is used to correct for the location of the source relative to the detector
with different source heads and mounting configurations. History has
determined that this value can be a constant.
– User Inputs
Material Type – The user shall be able to select one of the following material
types. The section on Single-Phase Material only applies to the selection of
“Single Phase” for the material type.
Single Phase
Slurry
Emulsion
Solution
Primary Measurement – When the user selects a material type of single phase,
the primary measurement will be density.
Unit for Measurement #1 – The user has the ability to select the units for
Measurement #1. The following unit selections are available.
Grams per cubic centimeter (g/cm3 or g/cm3)
Pounds per US gallon (lb/gal)
Pounds per UK gallon (lb/ImpGal)
Pounds per cubic foot (lb/ft3 or lb/ft3)
Short tons per cubic yard (STon/yd3 or STon/yd3)
Long tons per cubic yard (LTon/yd3 or LTon/yd3)
– Calculated Values
L: Pipe ID (cm) – Inside diameter of the pipe converted from user selected
units to centimeters.
(Dc): Density of the carrier material – For a single-phase material Dc = 0
(g/cm3).
(µc): Attenuation of carrier material – For a single-phase material µc = 0
(cm2/g).
(Ds): Density of the solid material – For a single-phase material Ds = 1
(g/cm3).
(Dcal1): Calibration point 1 Density – Density for calibration point 1 (Dc1u)
converted to grams per cubic centimeter (g/cm3).
(Dcal2): Calibration point 2 Density – Density for calibration point 2 (Dc2u)
converted to grams per cubic centimeter (g/cm3).
D: Density of process material – The density of the process material in g/cm3
calculated using Equation 4.
Slurry Material
When the process material is slurry, the gauge will be able to calculate the density of
Slurry Density the material, and will also be able to provide information on the carrier and solids.
Calculation The density for a slurry material shall be calculated using Equation 4. Initially the slope
correction factor will be 1. If a second calibration point is used, calculate the slope
correction factor using Equation 3. The user has the option to modify and/or
recalculate the slope correction factor at any time.
The parameters below are used to calculate the density of a slurry material.
– Constants
G: Geometry factor (Default = 0.85) – The geometry factor does not have any
units. It is used to correct for the location of the source relative to the detector
with different source heads and mounting configurations. History has
determined that this value can be a constant.
– User Inputs
Material Type – The user has the option of selecting any of the following
material types.
Single Phase
Slurry
Emulsion
Solution
This section only applies to the selection of Slurry for the material type.
Primary Measurement – When the user selects a material type of slurry, the
following options are available for the Primary Measurement.
Density – The density of the slurry is calculated using Equation 4. The
value will be displayed in the unit’s selected from the list in Units for
Measurement #1.
Solids content/volume
D = density of the slurry from Equation 4 (g/cc)
Dc = density of the carrier (g/cc)
Ds = density of the solids (g/cc)
Ms = Solid content per volume (g/cc)
Equation 5
The value will be displayed in the unit’s selected from the list in Units for
Measurement #1.
Carrier content/volume
D = density of the slurry from Equation 4 (g/cc)
Dc = density of the carrier (g/cc)
Ds = density of the solids (g/cc)
Mc = Carrier content per volume (g/cc)
Equation 6
The value will be displayed in the unit’s selected from the list in Units for
Measurement #1.
Solids/carrier – The solids/carrier is the same calculation as the proppant
calculation. (Refer to Proppant.)
Percent by weight solids
D = density of the slurry from Equation 4 (g/cc)
Dc = density of the carrier (g/cc)
Ds = density of the solids (g/cc)
Spw = percent by weight solids (%)
∗ 100
Equation 7
The value for Measurement #1 shall be displayed in percent (%).
Percent by weight carrier: The percent by weight carrier is calculated using
Equation 8.
D = density of the slurry from Equation 4 (g/cc)
Dc = density of the carrier (g/cc)
Ds = density of the solids (g/cc)
Cpw = percent by weight carrier (%)
100
Equation 8
The value for measurement #1 shall be displayed in percent (%).
Percent by volume solids
D = density of the slurry from Equation 4 (g/cc)
∗ 100
Equation 9
The value for measurement #1 shall be displayed in percent (%).
Percent by volume carrier
D = density of the slurry from Equation 4 (g/cc)
Dc = density of the carrier (g/cc)
Ds = density of the solids (g/cc)
Cpv = percent by volume carrier (%)
100
Equation 10
The value for measurement #1 shall be displayed in percent (%).
Percent Consistency (Available after implementation of the 3680
CutPRO application)
SGU 15 / SGU 60 (Available after implementation of the 3680 CutPRO
application)
Units for Measurement #1 – When the user designates the Primary
Measurement as density, carrier content per volume, or solids content per
volume, the units displayed for Measurement #1 are user selectable. The
following unit selections are available.
Grams per cubic centimeter (g/cm3 or g/cm3)
Pounds per US gallon (lb/gal)
Pounds per UK gallon (lb/ImpGal)
Pounds per cubic foot (lb/ft3)
Short tons per cubic yard (STon/yd3)
Long tons per cubic yard (LTon/yd3)
Grams per liter (g/L)
Ounces per cubic inch (oz/in3)
Pounds per cubic inch (lb/in3)
Grams per cubic inch (g/in3)
1 ∗
Equation 11
Carrier content/volume: Units shall be from the list in Units for
Measurement #1.
Dc1u = density for calibration point 1 (g/cc) calculated using
Equation 12
Dc = density of the carrier (g/cc) Ds = density of the solids (g/cc)
Mc = Carrier content per volume (g/cc) entered by the user and
converted to grams per cubic centimeter.
1 ∗
Equation 12
Solids/Carrier:
Note: Gauge will not be calibrated in these units. This is the
proppant calculation.
Percent by weight solids (%):
Dc1u = density for calibration point 1 (g/cc) calculated using
Equation 13
Dc = density of the carrier (g/cc) Ds = density of the solids (g/cc)
Spw = Percent by weight solids in percent (%) entered by the user.
∗
1
∗
100
Equation 13
Percent by weight carrier (%):
Dc1u = density for calibration point 1 (g/cc) calculated using
Equation 14
Dc = density of the carrier (g/cc) Ds = density of the solids (g/cc)
Cpw = Percent by weight carrier in percent (%) entered by the user
∗
1
100
∗
100
Equation 14
Percent by volume solids (%):
Dc1u = density for calibration point 1 (g/cc) calculated using
Equation 15
Dc = density of the carrier (g/cc) Ds = density of the solids (g/cc)
Spv = Percent by volume solids in percent (%) entered by the user
1 ∗
100
Equation 15
Percent by volume carrier (%):
Dc1u = density for calibration point 1 (g/cc) calculated using
Equation 16
Dc = density of the carrier (g/cc) Ds = density of the solids (g/cc)
Cpv = Percent by volume carrier in percent (%) entered by the user
100
1 ∗
100
Equation 16
(Dc1u): Density for calibration point 1 – The user enters the density for the
first calibration point in the selected units, reference. The density shall be
entered when standardizing on a full pipe or when the first point of the
calibration is performed.
(Dc2u): Density for calibration point 2 – Calibration point 2 shall be an
optional calibration point. The user enters the density for the second
calibration point in the same units as selected for the first calibration point.
The density shall be entered when the second point of the calibration is
performed.
– Measured Inputs
I: Radiation Intensity – Counts per second from the detector. The radiation
intensity shall be the filtered counts.
Istd: Standardization Counts – The average count rate per second from the
detector during standardization. The detector will calculate the average count
rate during the standardization cycle.
Ical1: Calibration point 1 counts – The average count rate per second from
the detector during calibration of the first point. The detector will calculate
the average count rate during the calibration cycle or during the
standardization cycle when standardized on a full pipe.
Ical2: Calibration point 2 counts – The average count rate per second from
the detector during calibration of the second point. The detector will calculate
the average count rate during the calibration cycle.
– Calculated values
L: Pipe ID (cm) – Inside diameter of the pipe converted from user selected
units to centimeters.
Dcal1: Calibration point 1 Density – Density for calibration point 1 (Dc1u)
converted to grams per cubic centimeter.
Dcal2: Calibration point 2 Density – Density for calibration point 2 (Dc2u)
converted to grams per cubic centimeter.
D: Density of process material – Calculated using Equation 4.
∗ ∗
100
Equation 18
– Carrier Mass Flow rate – The flow rate of the carrier in the slurry calculated using
the density of the carrier.
D = Density of slurry in grams per cubic centimeter calculated using Equation
4.
Qv = Volumetric flow rate of slurry in cubic centimeters per second; measured
by the gauge using an analog input.
Cpw = Percent by weight carrier: The percent by weight carrier is calculated
using Equation 8.
Qmc = Carrier mass flow rate calculated using Equation 19.
∗ ∗
100
Equation 19
– Bulk Volumetric Flow rate – The volumetric flow rate of the slurry measured by
the flow input.
Solids Volumetric Flow rate – The volumetric flow rate of solids flowing in
the slurry.
Qv = Volumetric flow rate of slurry in cubic centimeters per second; measured
by the gauge using an analog input.
∗
100
Equation 20
– Carrier Volumetric Flow rate – The volumetric flow rate of the carrier flowing in
the slurry.
Qv = Volumetric flow rate of slurry in cubic centimeters per second; measured
by the gauge using an analog input.
Cpv = percent by volume carrier (%) calculated using Equation 10.
Qvc = Carrier volumetric flow rate calculated using Equation 21.
∗
100
Equation 21
– Bulk Solids Flow rate (Reference In Situ Calculation)
– Temperature – The temperature measured by the temperature input.
– Velocity – The velocity of the slurry
Qv = Volumetric flow rate of slurry in cubic centimeters per second; measured
by the gauge using an analog input.
L = Inside diameter of the pipe in centimeters.
V = Velocity of the slurry calculated using volumetric flow rate.
∗
4
Equation 22
– Proppant – Reference Proppant.
The proppant calculation is used when oil wells are fractured. It calculates the amount
Proppant of solids per gallon of fluid. Two units have been used in the past for the amount of
solids added per gallon of fluid: Pounds Proppant Added (PPA) and Pounds Sand
Added (PSA). To accommodate the English and metric systems the following units
will be available.
– lb/gal (PPA) – pounds per gallon (Pound Proppant Added) for English units.
– g/l (PPA) – grams per liter (Pounds Proppant Added) for metric units.
The proppant calculation will be available for Measurements #2 – #4.
The proppant calculation is also the calculation for solids/carrier.
To calculate Pounds Proppant Added, the density of the slurry is calculated using
Equation 4.
– User Inputs
Dc: Density of the carrier material – Density of the carrier in grams per cubic
centimeter.
Ds: Density of the solid material – Density of the solids in grams per cubic
centimeter.
– Calculated Values
D: Density of the slurry – Calculated using Equation 4 in grams per cubic
centimeter.
PPA: Pounds Proppant Added – Calculated using Equation 23 in grams per
cubic centimeter.
1
Equation 23
The value displayed for PPA shall have user selectable units.
lb/gal (PPA) – pounds per gallon (Pound Proppant Added) for English units.
g/l (PPA) – grams per liter (Pounds Proppant Added) for metric units.
The in situ volume is the volume of loose material. The in situ calculation uses the
In Situ Calculation density of the slurry to calculate the volume of dry solids. Dredging would be an
example of an application that would use in situ volume.
– User Inputs
Dc: Density of the carrier material – Density of the carrier in grams per cubic
centimeter.
Ds: Density of the solid material – Density of the solids in grams per cubic
centimeter.
DB: Bulk Density of the solid material – Density of the dry solids in grams
per cubic centimeter.
– Measured Values
Qv: Volume Flow: Volume flow of slurry in cubic centimeters per second.
– Calculated Values
D: The density of the slurry calculated in grams per cubic centimeter using
Equation 4.
DDS: Grams of dry solid per cubic centimeter of slurry (g/cm3) calculated
using Equation 24.
QmDS: Mass flow of dry solids in the slurry calculated in grams per second
using Equation 25.
QBS: Volume flow of dry solids. This is the Bulk Solids Flow in cubic
centimeters per second, and is calculated using Equation 26.
Equation 24
∗
Equation 25
Equation 26
An emulsion material is one fluid suspended in another fluid. Emulsions will be treated
Emulsion similarly to slurries by the gauge. Fluid 1 will be treated as the carrier and fluid 2 will
Material be treated as the solid.
When the process material is an emulsion the gauge will be able to calculate the density
Emulsion Density of the material, as well as be able to provide information on the fluid 1 and fluid 2.
Calculation Calculate the density for an emulsion material using Equation 4. Initially, the slope
correction factor will be 1. When a second calibration point is used, calculate the slope
correction factor using Equation 3. The user has the option to modify or recalculate
the slope correction factor at any time.
The parameters below are used to calculate the density of an emulsion material.
– Constants
G: Geometry factor (Default = 0.85) – The geometry factor does not have any
units. It is used to correct for the location of the source relative to the detector
with different source heads and mounting configurations. History has
determined that this value can be a constant.
– User Inputs
Material Type – The user has the option of selecting any of the following
material types.
Single-Phase
Slurry
Emulsion
Solution
This section only applies to the selection of Emulsion for the material type.
Primary Measurement – When the user selects emulsion as the material type,
the following options are available for the Primary Measurement.
Density – The density of the emulsion is calculated using Equation 4.
The value will be displayed in the unit’s selected from the list in Units for
Measurement #1. Some of the terms of Equation 4 are redefined for an
emulsion.
Dc: Density of the fluid 1
µc: Attenuation Coefficient of the fluid 1
Ds: Density of the fluid 2
µs: Attenuation Coefficient of the fluid 2
– Calculated values
L: Pipe ID (cm) – Inside diameter of the pipe converted from user selected
units to centimeters.
Dcal1 – Calibration point 1 Density – Density for calibration point 1 (Dc1u)
converted to grams per cubic centimeter.
Dcal2 – Calibration point 2 Density – Density for calibration point 2 (Dc2u)
converted to grams per cubic centimeter.
D: Density of process material – Calculated density of the process material in
grams per cubic centimeter using Equation 4.
When the process material is a solution the gauge will be able to calculate the density
Solution of the material and provide information on the solvent and solute.
Calculation The density for a solution material should be calculated using Equation 4. Initially, the
slope correction factor will be 1. When a second calibration point is used, use Equation
3 to calculate the slope correction factor. The user has the option to modify or
recalculate the slope correction factor at any time.
The parameters below are used to calculate the density of a solution material.
– Constants
G: Geometry factor (Default = 0.85) – The geometry factor does not have any
units. It is used to correct for the location of the source relative to the detector
with different source heads and mounting configurations. History has
determined that this value can be a constant.
– User Inputs
Material Type – The user has the option of selecting any of the following
material types.
Single-Phase
Slurry
Emulsion
Solution
This section only applies to the selection of Solution for the material type.
Primary Measurement – When the user selects a material type of solution, the
following options are available for the primary measurement.
Density – The density of the solution is calculated using equation (4).
The value shall be displayed in the unit’s selected from the list in Units for
Measurement #1. Some of the terms of Equation 4 are redefined for a
solution.
Dc – Density of the solvent
µc – Attenuation Coefficient of the solvent
Ds – Density of the solute
µs – Attenuation Coefficient of the solute
Solute content/volume – Solute content per volume is calculated using a
fourth order polynomial.
D = density of the solution from Equation 4 (g/cc)
Dc = density of solvent (g/cc)
Ddiff = difference in solution density and solvent density (g/cc)
Equation 27
A, B, C, D = Coefficients provided by the user or from a list of
materials defined in Table B-1, Solution Coefficients.
Mconc = Concentration of solute or solute content per volume (g/cc)
∗ ∗ ∗ ∗
Equation 28
The value of Mconc shall be displayed in the unit’s selected from the list
in Units for Measurement #1.
Table C-1. Solution Coefficients
Material A B C D
Sucrose (0-100%) 2.598 0.1775 0.3503 0.0
D-Fructose (0-60%) 2.559 0.4315 0.0 0.0
D-Glucose (0-10%) 2.639 -0.09384 0.0 0.0
Sodium Chloride (NaCl) (0-50%) 1.408 1.050 -1.346 0.0
Sodium Hydroxide (NaOH) (0-50%) 0.8871002 1.138 -1.151 1.981
Potassium Chloride (KCl) (0-24%) 1.571 1.082 -1.786 0.0
Potassium Hydroxide (KOH) (0-52%) 1.098 0.885498 -0.3265 0.0
Hydrogen Chloride (HCl) (0-40%) 2.035 2.411 -12.50 48.56
Phosphoric Acid (H3PO4) (0-40%) 1.866 1.288 -8.047 18.07
A-Lactose (0-18%) 2.518 1.053 -6.338 0.0
H-Lactose (0-18%) 2.656 0.8647003 -4.504 0.0
Solvent content/volume – The solvent content per volume is calculated.
D = density of the solution from equation (4) (g/cc)
Mconc = solute content per volume from Equation 28 (g/cc) Msolv =
Solution content per volume
Equation 29
The value of Msolv shall be displayed in the unit’s selected from the list in
Units for Measurement #1.
Solute/Solvent – The solute/solvent is similar to the solids/carrier ratio for
slurries.
D = density of the solution from Equation 4 (g/cc)
Dc = density of the solvent (g/cc)
Equation 30
The value of Solute per Solvent shall be displayed in the unit’s selected
from the list in Units for Measurement #1.
Percent by weight solute – The percent by weight of solute is calculated.
D = density of the solution from Equation 4 (g/cc)
Mconc = solute content per volume from Equation 28 (g/cc) Sspw =
percent by weight solute (%)
∗ 100%
Equation 31
The value for measurement #1 shall be displayed in percent (%).
Percent by weight solvent – The percent by weight of solvent is calculated.
D = density of the solution from Equation 4 (g/cc)
Mconc = solute content per volume from Equation 28 (g/cc) Cspw =
percent by weight solvent (%)
1 ∗ 100%
Equation 32
The value for measurement #1 shall be displayed in percent (%).
Percent by volume Solute – The percent by volume solute is calculated.
Mconc = solute content per volume from Equation 28 (g/cc) Ds =
density of the solute (g/cc)
Sspv = percent by volume solute (%)
∗ 100%
Equation 33
The value for measurement #1 shall be displayed in percent (%).
Percent by volume solvent – The percent by volume of solvent is
calculated.
D = density of the solution from Equation 4 (g/cc) Dc = density of
the solvent (g/cc)
∗ 100%
Equation 34
The value for measurement #1 shall be displayed in percent (%).
Bulk Density – Bulk density is the same as item 1 Density above unless
temperature compensation is used. When temperature compensation is
used, Bulk Density is the uncompensated density. For more information
on temperature compensation, see Temperature Compensated Density.
Units for Measurement #1 – When the Primary Measurement is selected to be
Density, solvent content per volume or solute content per volume, the user
may select the units displayed for Measurement #1 from the following
options.
Grams per cubic centimeter (g/cm3 or g/cm3)
Pounds per US gallon (lb/gal)
Pounds per UK gallon (lb/ImpGal)
Pounds per cubic foot (lb/ft3)
Short tons per cubic yard (STon/yd3)
Long tons per cubic yard (LTon/yd3)
Grams per liter (g/L)
Ounces per cubic inch (oz/in3)
Pounds per cubic inch (lb/in3)
Grams per cubic inch (g/in3)
Pounds per cubic yard (lb/yd3)
Kilogram per cubic meter (kg/m3)
ID: Pipe ID – Inside diameter of pipe in units selected by the user.
Pipe ID Units – Units of measure for the inside diameter of the pipe.
Centimeters (cm)
Millimeters (mm)
Inches (in)
Feet (ft)
Yards (yd)
Meters (m)
Dc: Density of solvent material – Density of the solvent in grams per cubic
centimeter.
µc: Attenuation of solvent material – The value of the attenuation coefficient
depends on the composition of the process material and the type of source
used, Cesium-137 (Cs-137) or Cobalt-60 (Co-60).
Ds: Density of solute material – Density of the solute in grams per cubic
centimeter.
µs: Attenuation coefficient of solute – The value of the attenuation coefficient
depends on the composition of the process material and the type of source
used, Cesium-137 (Cs-137) or Cobalt-60 (Co-60).
Ibkg: Background radiation – Background radiation at the detector in counts
per second (CPS).
Calibration density units – The user may select the density units used for
calibration from the following options.
Grams per cubic centimeter (g/cm3)
Pounds per US gallon (lb/gal)
Pounds per Imperial gallon (lb/ImpGal)
Pounds per cubic foot (lb/ft3)
Short tons per cubic yard (STon/yd3)
Long tons per cubic yard (LTon/yd3)
Grams per liter (g/L)
Ounces per cubic inch (oz/in3)
Pounds per cubic inch (lb/in3)
Grams per cubic inch (g/in3)
Pounds per cubic yard (lb/yd3)
Kilogram per cubic meter (kg/m3)
Solute content per volume – Units shall be from the list in Units for
Measurement #1 .
Dsc1u = density for solution calibration point 1 (g/cc) Need to complete
the rest of the calibration units.
(Dc1u): Density for calibration point 1 – The user enters the density for the
first calibration point in the selected units. The density should be entered
when standardizing on a full pipe or when the first point of the calibration is
performed.
Temperature
Compensated
Density
The temperature compensation formula will compensate single phase, the carrier for
Temperature slurry, solvent for solution, or fluid 1 for emulsion. Use the formula a second time for
Compensated the solid, solute or fluid 2.
Calculation – ∆ ρ – Change in density of carrier/solvent/fluid 1 or bulk density with
temperature.
– A, B, C – Coefficients of the single phase, carrier, solvent, or fluid one
characterization polynomial.
– ∆T – Difference in process temperature and reference temperature.
– ∆T = TProcess – TReference
– TProcess – The temperature of the material in the process.
– TReference – The temperature of the material the density is corrected to.
– ρTref = ρBulk + ∆ρ
– ρTref – The density of the material at the reference temperature.
– ρBulk – The density of the material measured in the process at the process
temperature.
∆ρ A∗ ∆T B ∗ ∆T2 C ∗ ∆T3
Equation 35
When Use Temp Comp during STD is set to Yes the density will be compensated for
Standardization temperature if the standardization is done with the pipe full.
with Temperature The density during standardization (Cal Density Point #1) will be the density of the
Compensation process at the reference temperature. Internally, the gauge will store the density at the
process temperature. This will be at a location other than the modbus register.
The gauge will use the temperature compensation input source selected by the user to
determine the process temperature during standardization. At the end of the
standardization, the temperature from the temperature compensation input source
chosen by the user will be saved. Before the temperature is saved, it is corrected with
the Temperature Offset Correction.
Equation 36
– Tstored – The temperature value stored during the standardization and/or
calibration cycle.
– Tinput – The temperature for the user-selected input, determined by selecting Not
Used, Manual Value, Detector RTD Input or Detector 4-20 mA Input as the
Temperature Input Source.
– Toffset – The value entered by the user for Temperature Offset Correction. This is
a quick way to correct temperature offset. Normally when the RTD or current
input is calibrated, the Temperature Offset Correction is not needed and can be
set to zero.
The parameters below are used to calculate the density of a solution material.
– Constants
G: Geometry factor (Default = 0.85) – The geometry factor does not have any
units. It is used to correct for the location of the source relative to the detector
with different source heads and mounting configurations. History has
determined that this value can be a constant.
– User Inputs
T: Pipe ID – The inside diameter of the pipe converted from user-selected
units to centimeters.
Dc@ref – The density of the carrier, solvent, or fluid 1 at the reference
temperature.
µc – The attenuation coefficient of the carrier, solvent or fluid 1.
Ds@ref – The density of the solid, solute, or fluid 2 at the reference
temperature.
µs – The attenuation coefficient of the solid, solute, or fluid 2.
Ibkg – Background radiation at the detector in counts per second (CPS)
Dstd@ref – The density of the process material at the reference temperature
when standardizing on process.
Dcal1@ref – The density of the process material at the reference temperature
during calibration of point 1.
Dcal2@ref – The density of the process material at the reference temperature
during calibration of point 2.
Tref – Reference temperature
A1, B1, C1 – Coefficients of the carrier, solvent or fluid 1characterization
polynomial.
Temperature Input Source – The user selects how the process temperature is
provided to the system.
Not Used – Temperature compensation is disabled. This is the default
selection.
Manual Value – The user will provide a fixed value for the process
temperature.
Detector RTD Input – The process temperature is measured using the RTD
connected to the detector.
Detector 4-20mA Input – The user will configure the 4-20mA input on the
detector to provide the process temperature.
Manual Temperature – A user input value for the process temperature when
Manual Value is selected as the Temperature Input Source.
– Measured Inputs
I – Counts per second during process
Istd@proc – Average counts per second at process temperature during
standardization
Ical1@proc – Average counts per second at process temperature during
calibration for point 1
Ical2@proc – Average counts per second at process temperature during
calibration for point 2
Tproc – Temperature of the process material
– Calculated Values
Tdelta – The difference between the Process temperature and the reference
temperature.
Equation 37
Dc@proc – Density of the carrier/solvent/fluid one at the process temperature.
@ @ 1∗ 1∗
Equation 38
Dcal1@proc – Density of the process at the process temperature during
calibration of point 1.
@ ∗ 1@ @
1@ ∗ @ @
Equation 39
@ @
Equation 40
@ ∗ @
1@
@ ∗ @ @
Equation 41
Dcal2@proc – Density of the process at the process temperature during
calibration of point 2.
@ ∗ 2@ @
2@ ∗ @ @
Equation 42
@ @
Equation 43
@ ∗ @
2@
@ ∗ @ @
Equation 44
Dproc – Density of the process material at the process temperature.
@
In
1@
@
1@
@ @
@ @
Equation 45
Ws – Percent weight by solid/100
@ ∗ @
∗ @ @
Equation 46
Dref – Density of the process material at the reference temperature.
@ ∗ @
@ ∗ @ @
Equation 47
S – Slope correction for density calculation
@ @
@ @
2@
In
1@
∗ 1@ 2@
Equation 48
www.thermofisher.com