Operators Manual: For In-Vitro Diagnostic Use Only!
Operators Manual: For In-Vitro Diagnostic Use Only!
Operators Manual: For In-Vitro Diagnostic Use Only!
CoaDATA 501
Revision History
Date Changes
Version Manual (mm/dd/yy) Print
Software-Version
Copyright of Software
All software by LABiTec LAbor BioMedical Technologies GmbH (in the following LABiTec-Software) is the
intellectual property of the LABiTec LAbor BioMedical Technologies GmbH. Intellectual property rights shall
remain with LABiTec LAbor BioMedical Technologies GmbH. You are entitled to use LABiTec-Software and
the printed accompanying material at a place of work that cannot be transferred. Any violation of property
rights or copyright or trademark or using conditions may be subject to legal action. LABiTec reserves the
rights to modify the software, documentation as well as this operator manual without prior written notice.
Your Distributor:
(Please do not hesitate to contact your local distributor if you have any questions or problems.)
Contents
1 INTRODUCTION ................................................................. 3
1.1 Application.................................................................................... 3
1.2 Instrument Description................................................................ 3
1.3 Installation .................................................................................... 6
1.3.1 Connect an external printer............................................... 6
1.4 Measuring Principle..................................................................... 7
1.5 Reagents ....................................................................................... 8
2 OPERATION........................................................................ 9
2.1 Steps for Instrument Operation.................................................. 9
2.1.1 Turn on analyzer ............................................................... 9
2.1.2 STANDBY ......................................................................... 10
2.1.3 How to measure................................................................ 11
2.1.4 How to change methods ................................................... 13
2.1.5 How to change methods with a ChipCARD ...................... 15
2.2 Method Parameterization ............................................................ 16
2.2.1 PT-parameterization ......................................................... 16
2.2.2 aPTT - parameterization ................................................... 22
2.2.3 Fibrinogen 1 [g/l] - parameterization ................................. 25
2.2.4 Fibrinogen 2 [mg/dl] - parameterization ............................ 28
2.2.5 Thrombin time parameterization ....................................... 28
2.2.6 Intrinsic Factor parameterization....................................... 28
2.2.7 Extrinsic Factor – parameterization .................................. 28
2.2.8 Utilities............................................................................... 29
2.2.8.1 Menu printer.............................................................. 29
2.2.8.2 Menu computer............................................................ 30
2.2.8.3 Menu beeper................................................................ 30
2.2.8.4 Menu clock .................................................................. 30
2.2.8.5 Menu calibrate temp .............................................. 31
2.2.8.6 Menu secret number................................................. 32
2.2.8.7 Menu cuvette test................................................... 32
2.3 Printer............................................................................................ 33
2.3.1 Sample print-outs PT and calibration................................ 34
2.4 Errors ............................................................................................ 37
2.4.1 Application errors .............................................................. 37
2.4.2 Error Messages................................................................. 38
2.4.3 Errors during operation ..................................................... 39
2.4.4 Warnings ........................................................................... 39
2.4.5 How to change fuses ........................................................ 39
3 SOFTWARE ........................................................................ 40
3.1 Software overview ....................................................................... 41
3.2 Flow Chart of different application methods ............................ 42
3.3 Method Parameters...................................................................... 43
4 SAFETY ISSUES................................................................. 44
4.1 Hazard and Precautions .............................................................. 44
4.2 Maintenance and Hygiene........................................................... 46
4.2.1 Disposal of analyzer.......................................................... 46
5 APPENDIX .......................................................................... 47
5.1 Disposables .................................................................................. 47
5.2 Materials Supplied ....................................................................... 47
5.3 Technical Data.............................................................................. 48
5.4 Safety Specifications................................................................... 49
5.5 Mathematics ................................................................................. 50
5.6 Terminology.................................................................................. 52
1 Introduction
1.1 Application
The instrument type CoaDATA 501 (in the following titled as analyzer) as
described in this manual is an opto-mechanical coagulation analyzer which
applies the turbodensitometric measuring principle.
The measuring channel is integrated into the 37.4°C incubation block with 1
position for reagent bottle and 4 positions for cuvettes.
Results can be printed via on optional, external printer or can be read from
the display.
At the left side of the analyzer is the connector for the external power
adapter located. The external power adapter can be connected to the mains
with a voltage range from 100V - 240V, 47 - 63Hz.
For data output an RS 232 C 6-pin interface is also located on the left side
of the analyzer.
Analyzer
Made in Germany
Type: CoaDATA 501
SN A 00 0 0000 IVD
P-ID CD500-96.0000 - +
: XXXXXX
Mode 4 5 6 0
Figure 1 Analyzer
Description of keys
1 2 3
Mode 4 5 6 0
Esc 7 8 9 Enter
Reset Start
Esc-key
Esc
Switch from measuring to STANDBY
Exit a submenu
Enter Enter-key
Confirm selection, advance printer paper
1 2 3 Number-keys
Enter parameters
0-key
A print-out of the respective parameters for the selected method is
generated by pressing 0 during measuring.
Mode-key
Mode
1. Calibration
2. Menu selection, analyzer settings, and method parameterization
3. Exit a menu and save entered or modified data.
1.3 Installation
Remove the analyzer from its packaging and verify that the accessories kit
is complete. Please notify your distributor immediately in the event that the
shipment was incomplete. Refer also to Materials Supplied in chapter 5.2
• Prior to installation of the analyzer read the instructions under Hazard and
Precautions in chapter 4.1.
• Connect the power adapter between the analyzer and a power supply
(100V - 240V) free from interferences by large power users such as
elevators and centrifuges.
1 2 3
Mode 4 5 6 0
Esc 7 8 9 Enter
Danger!
Printer
Reset Start
• Connect the data-cable between the analyzer and the printer. Ask your
dealer for recommended printer types.
• Connect the power adapter to the printer (see figure 3). The printer will be
set ON by connecting to the mains.
NOTE Never operate the printer without paper! Read the instructions
manual from the manufacturer of the printer for further details.
A light beam passes through the cuvette containing the test plasma onto a
photo detector. Any change in the intensity of the transmitted light, that is
light increase or decrease, is converted into an electric signal. Hence, even
the most unstable clot can be detected.
The period from adding the start reagent until clot formation is measured. It
then can be converted into the appropriate units (%, ratio, INR, mg/dl, g/l).
Once the start reagent has been added, the measuring channel is adjusted,
that is the lamp intensity automatically adjusts up or down depending on the
turbidity of the test sample. In this process the turbidity of the sample
plasma and the reagent are adjusted.
A mixer is located in the cuvette. During the measuring process the mixer
provides homogeneity of the reagent-plasma medium. At the same time a
small whirl emerges through the mixer movement which assures that even
the smallest fibrin clot is formed in front of the photo detector.
This stirring action combined with the optical measurement constitutes the
basic features of the patented "turbodensitometric measuring principle".
Test cuvette
Lamp Detector
perm. magnet
Stirrer - Motor
ELECTRONIC
DISPLAY
1.5 Reagents
For proper coagulation analysis we recommend to use reagents, controls
and buffers from well known reagent manufacturers.
Contamination
With the application of different reagents, and here especially reagents
containing thrombin, there is a danger of reagent carry-over.
Useful hints:
Use the pipettor supplied with the respective pipette tips (Thrombi-Tips).
pipettes or tips.
20-200µl
Make sure no air bubbles are generated during the pipetting process.
Read the reagent packaging insert prior to use and follow the instructions.
Only use original cuvettes and stir bars from the manufacturer which
NOTE
are subject to strict quality control measures. Please understand
that the use of non-original cuvettes or instruments problems
caused by the use of non-original cuvettes may lead to the warranty
obligation become null and void.
2 Operation
2.1 Steps for Instrument Operation
Communication with the analyzer is performed via the liquid crystal display.
We assume that you are familiar with the function of the individual keys as
described in chapter 1.2
LAbor GmbH
Use the warm-up phase to load the analyzer with cuvettes and reagents for
testing.
Each cuvette must be equipped with a stir bar.
• Remove the existing cuvette from the measuring channel and close the
light protection caps.
< 1 PT >
Printer
If the printer is set to AUTO in the menu UTILITIES the parameterization of
the selected method as well as the result of the first measurement is printed
as soon as the first measurement is completed.
2.1.2 STANDBY
< 1 PT >
cuv in
37.4°C
Single/double determinations
The user can switch to single determination prior to or after a
measurement in the method menu <replication> (refer to chapter 2.2.1
PT-parameterization).
Sample incubation
Sample incubation is always performed in the measuring channel!
• Switch to measuring mode.
cuv in 1
The analyzer automatically recognizes the cuvette and starts the timer for
sample incubation (timer count down). An acoustic signal indicates 5 sec
remaining incubation time.
After sample incubation the measuring channel will be adjusted for sample.
(adjS = adjust Sample).
Once the sample has been adjusted the following display 100 ul alternately
GO – S1 appears:
• Press Reset-key.
cuv in 2
The analyzer automatically recognizes the cuvette and starts the timer for
sample incubation (timer count down). An acoustic signal indicates 5 sec
remaining incubation time.
After sample incubation the measuring channel will be adjusted for sample.
(adjS = adjust Sample).
Once the sample has been adjusted the following display 100 ul alternately
GO – S2 appears:
Once the second measured value has been obtained, the mean from the
measured values will be determined and converted into %, ratio, and INR
via the entered calibration curve.
The results will be displayed consecutively for duration of 5 sec. The printer
will automatically print the results. The last message cuv out then press
"Reset" will request the removal of the cuvettes from the measuring
channels.
cuv in 1
• Press the right arrow; the next method aPTT is displayed as ready-to-
measure.
• Press the arrow-key --> (arrow-key <-- back); the following methods and
the UTILITIES Menu will be displayed:
< 1 PT >
< 2 aPTT >
< 3 Fib. 1 > g/l
< 4 Fib. 2 > mg/dl
< 5 Thrmb >
< 6 Intr. >
< 7 Extr. > Can be overwritten by ChipCARD!
< UTILIT >
The new method has been initialized. Incubation of the first samples can
begin.
cuv in 1
The ChipCARD will be inserted into the ChipCARD reader with the contact
ahead and method description readable.
To load a method:
• Insert the ChipCARD with the selected method into the ChipCARD
Reader the method name and the lot-no. will be displayed automatically.
<-reading card
aPTT
Lot XXXXX
remove card
cuv in 1
UTILITIES
The menu Utilities is a group of menus in which instrument settings can be
performed after a "Secret no." (code number) has been entered.
2.2.1 PT-parameterization
Method parameters in the analyzer have been preset by the manufacturer.
Prior to performing clotting analysis you must update the method parameter
for the reagent used.
• Press Mode. The analyzer will request you to enter an up to 5 digit long
secret number (factory setting: 11111). For additional information refer to
chapter 2.2.8.6 Menu secret number.
<1.conv>
Press the arrow key -> to display the following menus: <1. conversion>,
<2. conversion>, <replication>, <measurement> and <cuv remove
detec.>.
Overview PT parameterization:
<1.conversion>
<curve> input of a 9-point calibration curve under or
<quick> to enter the 100%-value and slope or
<2.conversion>
<INR> input of the ISI-value for INR or
1st conversion
curve
• Select 1. conversion and press Enter to confirm selection.
• Press Enter to type in the first calibration point. This point is defined as
point of greatest activity and shortest clotting time.
1.point
100.0 %
12.0 s
The entry field for the next calibration curve point is displayed.
2.point
The following display appears once the last calibration curve point has been
confirmed:
Quick
The curve for the PT calibration curve can be entered by typing in the 100
%-value and the factor for the slope of the calibration curve.
factor = 54
Please use the value for factor as provided with the reagent package insert.
If you need to calculate the factor by yourself please refer to chapter 5.5.
If complete data has been entered for <curve> and <quick > the
NOTE calibration curve type selected last is active for con-versions in the
ready-to-measure mode. This is also valid for INR and ratio under
2nd conversion.
none
If < none > was selected no conversion will be performed.
2nd conversion
< 2.conv >
INR
• Press Enter and the following display appears:
ISI= 1.05
ATTENTION: If <none> was selected under 1st conversion the 100 % sec
value, e.g. 100 % = 12,6 sec. will be requested.
ratio
< ratio >
none
If < none > was selected no conversion will be performed.
replication
In the menu "replication" you can select between "double" for double
determinations or "single" for single determinations by pressing the arrow
key.
• Press Enter.
Only if double determination has been selected the following display for
entry of the coefficient of variation of the individual values will be displayed.
If this value is exceeded the message "mean error" will appear in the display
and print-out.
measurement
< measur >
• Press Enter
Incubation time
incubat.
= 60 s
A display which requests the entry of start reagent volume and lot number
appears.
= 100 ul
lot no.=
12345678 Example!
By selecting:
<2.conv>
<2.conversion>
<ratio> input of normal value for ratio calculation or
<none> for no conversion
2nd conversion
<2.conv> Selection Conversion
ratio
< ratio >
100% = Example!
replication
< replic >
Only if double determination has been selected the following display for
entry of the coefficient of variation of the individual values will be displayed.
If this value is exceeded the message "mean error" will appear in the display
and print-out.
n = 5% 1% - 20%
measurement
< measur >
<- incubat.
=120 s
<-Start Reagent
= 50 ul
lot. no. =
12345678
By selecting:
<1.conversion>
<curve> input of a 9-point calibration curve under or
<none> no conversion
1st conversion
A 9-point calibration curve can be entered with this menu.
Calibration curve points that are not to be used must have the entry 0.0 s.
For information on interpolation refer to chapter 5.5 Mathematics. The
system requires two points to be defined but we would recommend a
minimum of 3 points. You can exit the calibration curve menu with Esc if no
entry has been made. Once an entry has been made, all additional
calibration curve points must be retrieved and verified.
• Enter the concentration and clotting time and confirm the entry with
Enter.
2.point:
2.53 g/l
11 s
• Enter the concentration and clotting time and confirm the entry with
Enter.
The following display appears once the last calibration curve point has been
confirmed:
replication
< replic >
• Press Enter.
Only if double determination has been selected the following display for
entry of the coefficient of variation of the individual values will be displayed.
If this value is exceeded the message "mean error" will appear in the display
and print-out.
n = 10% 1% - 20%
measurement
< measur >
<- incubat.
= 60 s
A dialog will request the entry of start reagent volume and reagent lot no.:
<-Start Reagent
= 100 ul
lot. no. =
12345678
By selecting:
<1.conversion>
<curve> input of a 9-point calibration curve under or
<none> no conversion
<1.conversion>
<curve> input of a 9-point calibration curve under or
<none> no conversion
2.2.8 Utilities
To enter the Utilities Menu:
• finish an actual measurement
• change to the Standby-Mode
• press the left arrow-key
The following menus can be accessed under Utilities provided the secret
number (11111) has been entered: <printer>, <computer>, <beeper>,
<clock>, <calibrate-temp>, <secret no>, <cuvette test>.
Each alteration in one of the menus is to be confirmed with the Enter- key.
The following display appears:
A selection between <AUTO>, <MANUAL>, <Par Pro>, and OFF for the
optional printer EPSON P40 can be made from this menu.
A print-out includes:
Test results, method parameters, complete parameter list, and error
messages.
If you select
In this menu you can select between <ON>, <OFF>, and <CLICK>.
If you select:
You can enter the current date and time with this menu. Once a selection
has been made the following dialog will appear:
• Enter the current hour, minute, and second and press Enter.
Measuring tools:
Digital multimeter with 4 1/2-digit display. For example Voltkraft M-4650B
and temperature measurement adapter for multimeter, e.g. Fluke 80T-
150U, output in °C.
To verify the temperature place the measuring sensor into position "T"
which is filled with bi-distilled water.
• Check the temperature after approx. 10 min.
ext 37.4
• Wait until the temperature in the display has stabilized internally to 37.4
°C.
• Next enter the temperature displayed by the digital thermometer via the
numerical keypad. The entry will be displayed under extern.
• Wait until the temperature has stabilized at 37.4 °C on the display intern.
In the menu <sec.no.> you are able to select a code number to restrict
access to the parameter menus.
Once a sec.no. has been selected the following display will appear:
>11111< (0=none)
• Press Enter to save the new number. A print-out of secret number as well
as serial number follows (printer mode: AUTO).
-----------------------------
Analyzer name
ser. no. C1770711 Sample print-out!
secret no=xxxxx x= number
-----------------------------
If zero is entered and saved instead of a number the secret number will not
be queried when a parameter menu has been selected.
The entry must be completed with Enter if a secret number shorter 5 digit is
stored.
You can control the function of the automatic cuvette detection with this
menu.
CUV-TEST
cuv
If a cuvette was placed in the measuring channel the display cuv appears.
The display will show the respective status once a cuvette is placed into or
removed from the analyzer.
2.3 Printer
It is possible to link an external printer (optional) to the analyzer, refer to
chapter 1.3.1 Connect an external printer. The connection to the printer
shall be done via the RS 232C interface of the analyzer. Ask your local
dealer for recommended printer models.
General print-outs
Once a method has been selected the programmed calibration curve
parameters will be printed followed by the results. The print-out is automatic
as soon as a result has been obtained by the measuring channel.
=======================
PARAMETER-PROTOCOL
( 980 Bytes)
1-channel
V X.xx, mm/dd/yy
math-vers V XX.xx
-- device-specific: --
ser.no. j582074 Serial number of analyzer
secret no. = 11111
ntc_soll = 509
mode SINGLE
incubat = 60 s
start-reagent:
Lot 1
reagent = 100 ul
100.0 % 11.6 s
50.0 % 17.7 s
25.0 % 29.9 s
10.0 % 66.6 s
---- method store 2 ----
aPTT
print-outs PT
PT Documentation
Example: Conversions via a 4-point calibration curve in % and INR.
start-reagent:
Lot 101xxx Reagent lot no.
reagent = 100 ul Reagent start volume
PT Method
patient _____________ Patient name
02.12.99, 10:52:55 Date, Time
time 1 = 12.0 s 1st measured time
time 2 = 12.8 s 2nd measured time
Mean = 12.4 s Mean of measured times
INR = 1.7 Conversion to INR
% = 88.4% Conversion to PT %
PT documentation
Example: Conversion via Factor calibration curve in % and INR.
start-reagent:
Lot 101xxx Reagent lot no.
reagent = 100 ul Reagent start volume
-----------------------
results:
PT Method
patient _____________ Patient name
02.12.99, 12:58:38 Date, Time
time 1 = 11.7 s 1st measured time
time 2 = 12.1 s 2nd measured time
Mean = 11.9 s Mean of measured times
INR = 1.01 Conversion to INR
quick = 98.1% Conversion to PT %
2.4 Errors
Errors can be generated by the user and/or the system itself. The analyzer
displays error messages and warnings in the display. If the printer has been
activated these messages will be printed.
Should any of these errors occur and they are recognized in time, they must
be remedied immediately.
BREAK dark preparation for measurement is too turbid dilute plasma or mix reagent.
BREAK top lim exceeded measuring range (too high) repeat test
possibly caused by air bubbles
BREAK bot lim exceeded measuring range (too low) repeat test
BREAK noise loud noise after sample adjustment check for air bubbles or other
particles.
BREAK drift measured curve drifted after reagent has check sample for air bubbles
been added
break jump Measuring break because of a measuring repeat test and if necessary
value jump (no clotting) contact technical Service.
break readjust message is displayed if the light value is too Repeat test and dilute plasma if
dark during the adjustment phase necessary.
mg/dl <2.0 converted value is lower than the check analytical steps and the
parameterized minimum value. conversion parameter.
mg/dl <200.0 converted value is bigger than the see above (mg/dl <2.0)
parameterized maximum value.
SYSTEM FAILURE: external RAM on address Sxxxx in error contact technical Service
ext. RAM: Sxxxx
parameter-error! check sum error for parameters in EEPROM contact technical Service
press any key ...
Measuring cell polluted Additional pipetting of plasma or reagent Remove liquid with pipettor, clean
with liquids into the measuring cell without cuvette with appropriate absorbent cloth,
refer to chapter 4.2
2.4.4 Warnings
Warning Meaning
Cool down This message will appear if the incubation block is too warm during a measuring
pause. No other measurements can be started during cool down.
TEM. WARN If during a measurement the temperature of the incubation block deviates
significantly from the set value, the measurement is not cancelled. Instead this
warning appears in the display and is also printed out via the active printer.
mean error Signals a wrong result after a double determination in relation to the coefficient.
max-time reached If a point of calibration curve meets the no more points maximum measured
- no more points time no additional points can be entered as they need to increase from point to
point. Input will be blocked when this message appears.
min-value reached During the input of the calibration curve, the range of values (%, g/l, mg/dl) is
- no more points limited for each method due to manufacturer’s settings. In addition the points
or must increase or decrease depending on the presetting. If the largest or smallest
max-value reached permitted value for a point has been entered, no additional points can be
- no more points entered. Input will be blocked when this message appears.
3 Software
The software for the analyzer is stored in a memory and will be activated as
soon as the analyzer is switched on. It controls the analyzer via start
functions for the analytic program.
The menu Utilities has been integrated into the method menu so that
system settings can be performed for the following menus <printer>,
<computer>, <beeper>, <clock>, <calibrate temp>, <secret number>, and
<cuvette test>.
At this point the optical blank value will be determined and stored for the
measuring channel. No cuvettes may be located in the measuring channels
at this time!
Storing of parameter
After a parameterization of an instrument or test-data, short information
"write parameter" will be shown on the display.
Analyzer name
Power ON
Initialising
WARM UP
Remove cuvettes
then press any key
auto blanking
keep channels clear
< 1 PT >
cuv in 2 Mode
cuv out, *
* then press “Reset”
Figure 7 Software Overview
cuv in 1 cuv in
incu 47 incu 47
adj – S1 adj – S
GO - S1 GO - S
100 ul 100 ul
1.2 s 1.2 s
t= 12.6 s time =…
cuv in 2 % = 91.0
GO - S2
100 ul
6.9 s
mean time=
% = 91.0
INR = 1.05
cuv out, *
Calibration curve P1 - P9
5 Thrmb Value 0 0 0 0 0 0 0 0 0 0 0 0 -
Time 0 0,0 0,0 0,0 0,0 0,0 0,0 0,0 0,0 sec
7 Extr. Value 200,0 0,5 0,0 0,0 0,0 0,0 0,0 0,0 0,0 % 0 0 0 -
Time 5,0 150,0 0,0 0,0 0,0 0,0 0,0 0,0 0,0 sec
method store
max. 5 characters
method name
start-reagent [µl]
(quick)
1st conversion reference curve
decimal place
4 Safety issues
1. Electrical safety
Check that the operating voltage is set correctly before you connect the
device to the main power supply.
To connect the device to the power supply, use only sockets which are
grounded in order to keep the risk of an electrical shock as low as possible.
All current carrying parts are sources of danger for an electrical shock.
Surfaces (floors, work table) must not be moist when you are working with
any electrical device.
Carry out only the maintenance work and/or the replacement of parts
described in these operating instructions.
All work which requires the analyzer to be opened may only be carried
out by a technician who is familiar with the risks related thereto.
Use only replacement fuses of the stated type and with the stated nominal
current.
Never use fuses which have been "repaired".
Never short-circuit the fuse holder.
3. Mechanical safety
(Analyzer is operating)
Never open screw-attached housing parts while the instrument is ON. There
is a risk of injury due to moving parts (fan, motor, drives).
Caution!
Risk of infections
4. Samples
Avoid any direct contact with samples which are potentially infectious or
which may generate other risks to the human body.
If sample material is spilled onto the analyzer, wipe it off immediately and
decontaminate the surface. Refer to chapter 4.2.
5. Reagents
Observe the suggestions in the package inserts for a correct use of the
reagents.
Caution!
8. Operator qualification
The analyzer should only be operated by trained personnel. Ask you local
dealer or distributor for further information on the availability of user
trainings.
Caution!
Keep the instrument clean and do not spill liquids onto the analyzer. To
protect the instrument from dust, cover it with the supplied dust cover or
store instrument in a cabinet when not in use.
In case liquids were spilled onto the instrument, immediately absorb liquid
with an appropriate cloth.
If liquid has accidentically run or was pipetted into the measuring channel,
remove liquid with a pipette and clean the measuring channel with a lint-free
cloth.
- Electronic parts must be disposed off in accordance with the guidelines for
the disposal of electronic parts
- Make sure that the analyzer has been decontaminated before disposal.
5 Appendix
5.1 Disposables
Cuvettes
1 x Dispo-System micro 40.613.0002
500 cuvettes / stir bars 1,0 x 4,0 mm
Thrombi-Tips
1 x Thrombi-Tips Racks, 16 x 144 Tips 40.673.0016
Manuals
Operators Manual CoaDATA 501 30.000.1613
(German Version)
Pipettes
Thrombi-Pette 20 - 200 µl 20.000.2712
Thrombi-Pette 20 µl 20.000.2713
Thrombi-Pette 50 µl 20.000.2714
200
1 x CoaDATA 501
Operation manual
Measuring channels 1
Cuvette positions 4
Operating voltage 12 V DC
Weight 0,6 kg
Warranty
Warranty is granted for a period of 12 months starting from the date of
delivery.
If the instrument's type plate bears an IVD symbol it complies with the
following directive:
- 98/79/EC in-vitro Diagnostics directive
If the instrument bears no IVD symbol on the type plate it complies with the
following directives:
- 73/23/EEC Low-voltage directive
- 89/336/EEC-EMC directive
5.5 Mathematics
Computation of Factor:
(25 % Value in sec) – (100 % Value in sec)
(1/25) – (1/100 %) = factor
Example:
Please only enter the first two digits into the menu, as for example 57!
Computation of PT %
1
(( meas. time-100 % = PT %
Computation of mean
If in a dual determination the individual results deviate to a larger extent
than permitted by the coefficient of variation, the error message mean error
will be displayed and printed-out.
Extrapolation
PT >100 % linear extrapolation over the last two higher points.
PT <10 % linear extrapolation over the last two lower points.
Conversion to ratio
and INR:
ratio = measured clotting time / normal value
5.6 Terminology
A
adju = adjustment
B
break = cancel refer to chapter 2.4.2
break noisy =refer to chapter 2.4.2
break drift = refer to chapter 2.4.2
break dark = refer to chapter 2.4.2
break top lim = refer to chapter 2.4.2
break bot lim = refer to chapter 2.4.2
break jump = refer to chapter 2.4.2
break readjust = refer to chapter 2.4.2
break motor = refer to chapter 2.4.2
C
ChipCARD = memory card (size of a credit card)
clock = menu, date and time
curve = calibration curve
cal = calibration / adjustment
cuvette test = automatic cuvette detection
cuv lrn = blank value, cuvette detection
cool down = refer to chapter 2.4.4
E
ERROR = refer to chapter 2.4.2
err div0 = refer to chapter 2.4.2
err log0 = refer to chapter 2.4.2
err over = refer to chapter 2.4.2
F
Fib = Fibrinogen
G
Go = continue, next step
I
incu = incubate
INR = International Normalized Ratio / constant for Thromboplastin
J
K
L
lot/incu = lot number / incubation
Lo/IN = menu lot number / incubation
Lot = lot number
lrn = learn
M
meas = measuring
motor = refer to chapter 2.4.2
mg/dl <2.0 = refer to chapter 2.4.2
mg/dl <200.0 = refer to chapter 2.4.2
P
PAR.TYPE = parameter type
Par Pro = parameter protocol
parameter error = refer to chapter 2.4.2
Q
R
S
Secret no. = security code number
T
THROMBIN T = Thrombin time
TIME = hour:minute:second
Temperat. warning = refer to chapter 2.4.4
SYSTEM FAILURE .... = refer to chapter 2.4.2
V, W
writing parameter to internal = parameter storage