Location via proxy:   [ UP ]  
[Report a bug]   [Manage cookies]                

DAT400 Manual EN

Download as pdf or txt
Download as pdf or txt
You are on page 1of 60

DAT 400

Digital Transmitter

TECHNICAL MANUAL
For Analog/Profibus/Devicenet versions

Software version PW13081


Page II
TABLE OF CONTENTS

PRECAUTIONS............................................................................................... Page 2

INTRODUCTION............................................................................................ Page 3

TECHNICAL FEATURES.................................................................................... Page 4

INSTALLATION............................................................................................... Page 5

FRONT PANEL OF THE INSTRUMENT............................................................... Page 10

USING THE KEYBOARD.................................................................................. Page 11

INFO DISPLAY................................................................................................ Page 13

VIEWING, ZEROING THE WEIGHT AND SELF-CALIBRATION.............................. Page 14

SETTING........................................................................................................ Page 18

DIAGRAM OF THE MENU............................................................................... Page 20

CONFIGURATION PARAMETERS...................................................................... Page 22

CALIBRATION................................................................................................ Page 25

WEIGHING PARAMETERS............................................................................... Page 27

INPUT/OUTPUT PARAMETERS.......................................................................... Page 29

SERIAL OUTPUT PARAMETERS.......................................................................... Page 32

ANALOG PARAMETERS.................................................................................. Page 36

SERIAL COMMUNICATION PROTOCOLS......................................................... Page 38

USB CONNECTION BETWEEN DAT 400 AND PC............................................. Pag. 47

FIELDBUS PROTOCOLS................................................................................... Page 48

TROUBLESHOOTING...................................................................................... Page 53

GROUNDING CONNECTIONS DAT400.......................................................... Page 54

INTRINSIC SAFETY BARRIERS........................................................................... Page 55

OPTION 24 VOLT POWER SUPPLY................................................................... Page 56


0316

Page 1
PRECAUTIONS
READ this manual BEFORE operating or servicing the instrument.
FOLLOW these instructions carefully.
SAVE this manual for future use.

CAUTION
The installation and maintenance of this instrument must be allowed
to qualified personnel only.
Be careful when you perform inspections, testing and adjustment
with the instrument on.
Perform the electrical connections in the absence of the power supply
Failure to observe these precautions may be dangerous.
DO NOT allow untrained personnel to work, clean, inspect, repair
or tamper with this instrument.

Page 2
INTRODUCTION
The DAT 400 is a transmitter of weight to be combined with the load cells to detect the weight in every
situation.
The module is easy to install and can be mounted on 35 mm DIN rail.
The display allows easy reading of the weight, the status of the instrument, the setting parameters and
errors.
The 4 keys located below the display allow the operator to perform the functions of ZERO, TARE,
GROSS/NET switching, setting of the setpoints weight, setting and tare both theoretical than real.
The DAT 400 uses the serial port RS232 with ASCII and Modbus RTU protocols for connecting to a PC,
PLC and remote units. In parallel with the RS232, a USB port available.
They are always available 2 programmable weight setpoints and the control of the maximum weight
value reached (peak).
The RS422/RS485 serial output allows you to connect up to 32 addressable devices.
The availability of the most common fieldbuses, as an alternative to the RS422/RS485 port, also allows
the transmitter to interface with any supervision device currently offered by the market.
Available versions:
DAT 400: weight transmitter with serial output RS232 (USB), RS485 and Peak function. Supported
protocols are Modbus RTU, continuous, slave and the ones upon request. Two programmable setpoints,
2 inputs and Peak function.
• DAT 400/A: version with the analog output.
• DAT 400/PROFIBUS: weight transmitter with serial output RS232 and PROFIBUS DP.
• DAT 400/DEVICENET: weight transmitter with serial output RS232 and DEVICENET.

IDENTIFICATION PLATE OF THE INSTRUMENT

It’s important to communicate this data in the event of a request for information or information concer-
ning the instrument together with the program number and version that are shown on the cover of the
manual and are displayed when the instrument is switched on.

WARNINGS
The following procedures must be performed by qualified personnel.
All connections must be performed when the instrument is switched off.

Page 3
TECHNICAL FEATURES
Power supply 24 Vdc ± 15 %
Max. absorption 5W
Insulation Class II
Installation category Cat. II
Operating temperature -10°C ÷ +40°C (max. humidity 85% non-condensing)
Storage temperature -20°C ÷ +50°C
Weight display Numerical 6 red led digits and 7 segments (h 14 mm)
Led 4 LEDs of 3 mm
Keyboard 4 mechanical keys
Overall dimensions 106 mm x 58 mm x 90 mm (l x h x w)
Installation Brackets for DIN profile section or OMEGA bar
Case material self-extinguishing Noryl (UL 94 V1)
Connections Screw terminal boards, pitch 5.08 mm
Load cells power supply 5 Vdc/120mA (max 8 cells of 350W in parallel), short-
circuit protected
Input sensitivity 0.02 mV min.
Linearity 0.01% of the full scale
Temperature drift 0.001% of the full scale / °C
Internal resolution 24 bits
Resolution of the weight displayed Up to 60,000 divisions on the net capacity
Measurement range –0.5 mV/V to +3.5 mV/V
Frequency of weight capture 5 Hz - 50 Hz
Digital filter To be selected from 0.2 Hz to 25 Hz
Number of weight decimals 0 ÷ 3 decimal places
Zero calibration and full scale Automatic (theoretical) or executable from the keyboard.
Logic outputs 2 opto-isolated (dry contact), max 24Vdc / 60 mA each
Logic inputs 2 opto-isolated at 24 Vdc PNP (external power supply)
Serial port (# 2) RS232C, RS422/RS485 and USB (in parallel)
Maximum cable length 15m (RS232) and 1000m (RS422 and RS485)
Serial protocols ASCII, Modbus RTU
Baud rate 2400, 9600, 19200, 38400, 115200 to be selected
Program code memory 64 Kbytes FLASH on-board reprogrammable from RS232
Data memory 2 Kbytes

Analog output (optional) Voltage or current


Resolution 16 bits
Calibration Digital from the keyboard
Impedance Voltage: min. 10KΩ; Current: max 300Ω
Linearity 0.03 % of the full scale
Temperature drift 0.001% of the full scale / °C

Fieldbus (optional PROFIBUS DP, DEVICENET


Buffer dimension 128 byte IN - 128 byte OUT

Page 4
INSTALLATION

GENERAL DATA
The DAT 400 is composed of a motherboard, on which you can add the options available; the mother-
board is housed in a plastic enclosure by a 35mm DIN rail.
The DAT 400 should not be immersed in water, subjected to jets of water and cleaned or
washed with solvents.
Do not expose to heat or direct sunlight.

OVERALL DIMENSIONS

90
PRECISE
DAT 400 8.8.8.8.8.8.
1 2 NET
46
SET FUN 0 PRG

106 58

ELECTRIC INSTALLATION
The transmitter DAT 400 uses screw terminal boards, pitch 5.08 mm, for the electrical con-
nection. The load cell cable must be shielded and channeled away from tension cables to
prevent electromagnetic interference.

INSTRUMENT POWER SUPPLY


1

PRECISE
DAT 400

The instrument is powered through the terminals 8 and 9. The power


cord must be channeled separately from other cables.
8.8.8.8.8.8.
SET

The supply voltage is electrically isolated.


+24 Vdc
1

Power supply voltage: 24 Vdc/ ± 15% max. 5W


2

0 Vdc
+
9

FUN

-
10

NET
0

Page 5
PRG
18
CONNECTIONS OF THE LOAD CELL/S
The cell/s cable must not be channeled with other cables, but must
follow its own path.
The instrument can be connected up to a maximum of 8 load cells
of 350 ohm in parallel. The supply voltage of the cells is 5 Vdc and
is protected by temporary short circuit.
The measuring range of the instrument involves the use of load cells
with a sensitivity of up to 3.5 mV / V.
The cable of the load cells must be connected to terminals 13-18.
SENSE+
SENSE-

SHIELD

In the case of 4-wire load cell cable, jumper the terminals 13 to 16


EXC+
SIG+

EXC-
SIG-

and 14 to 15.
18 10 9
Connect the cell cable shield to the terminal 9.
1

In the case of the usage of two or more load cells, use special junction
boxes (CEM4/C or CSG4/C). Below please find their connection.
PRG 0 FUN SET
NET 2 1

8.8.8.8.8.8. DAT 400


PRECISE

19 27J-BOX CGS4-C DAT400


+EXC 1 +EXC -SGN 1 17 SIGN- OUT1 1

-EXC 2 -EXC +SGN 2 18 SIGN+ OUT2 2

+SGN 3 +SGN -EXC 3 13 EXC- C OUT 3

-SGN 4 -SGN -SNS 4 16 SENSE- IN1 4


SHD 5 SHD +SNS 5 15 SENSE+ IN2 5

+EXC 6 14 EXC+ C IN 6
SHD 7
+EXC 1 +EXC

-EXC 2 -EXC

OPTIONAL
FIELDBUS
10 mA +
OPTIONAL
+SGN 3 +SGN

-SGN 4 -SGN 11 V +

SHD 5 SHD 12 A. COM


25 pin Connector

+24V
+EXC 1 +EXC S.GND 25
-EXC 2 -EXC 8 + RXD 26
24
+SGN 3 +SGN 9 - TXD 27
-SGN 4 -SGN

SHD 5 SHD

+EXC 1 +EXC
-EXC 2 -EXC
+SGN 3 +SGN
-SGN 4 -SGN
SHD 5 SHD

LOGIC INPUTS
The two logic inputs are opto-isolated.
The cable connecting the logic input should not be channeled with
the power cables.
1

PRECISE
DAT 400

+
24 VDC
INPUTS

INPUT 1 The function of the two inputs is as follows:


- INPUT 2
8.8.8.8.8.8.

INPUT1 Resetting the displayed value (gross, net or peak)


SET

COM. INPUT

INPUT 2 PRINT
1
2
9

FUN

The activation of the two functions is accomplished by bringing the


external power supply 24 Vdc to the corresponding terminals as
10

27
NET

shown in the figure.


0

Page 6
PRG
18

19
LOGIC OUTPUTS
OUTPUT 1
The two opto-isolated relay outputs are the normally open contact. OUTPUTS

1
24 Vdc OUTPUT 2
The capacity of each contact is 24 Vdc, 100 mA max. 100 mA Max COM. OUTPUT

The cable connecting the outputs should not be channeled with the
power cables. The connection should be as short as possible.

SET
1

PRECISE
DAT 400

1
2
9

FUN
SERIAL COMMUNICATION

8.8.8.8.8.8. 8.8.8.8.8.8.
SET

10

NET
1
RS232:

0
9

FUN
The RS232 serial port is always present and handles various pro-
tocols. TXD

(20m max)
10

27

RS232
RXD

NET

PRG
To achieve the serial connection, use a shielded cable, making sure S.GND

18
to connect the shield to one of the two ends: to pin 25 if connected SHIELD

on the side of the instrument, to the ground if it is connected on the


other side.

PRG
1

PRECISE
DAT 400
The cable must not be channeled with power cables; the maximum
18

19
length is 15 meters (EIA RS-232-C), beyond which you should take
the optional RS485 interface.
SET
1
2

RS485:
9

FUN

The serial port RS485 (2-wire) is present in the model DAT 400/
10

27
RS485.
NET

To achieve the serial connection, use a suitable shielded cable,


0

making sure to connect the shield to one of the two ends: to pin S.GND
TXD+
23 if connected on the side of the instrument, to the ground if it is

(1000m max)
N°unità max
RS422/485
TXD-
connected on the opposite side.
PRG

RXD+
RXD-
The cable should not be channeled with the power cables.
18

19

PRECISE
DAT 400
8.8.8.8.8.8.
SET
ANALOG OUTPUT (OPTIONAL) 1

The transmitter provides an analog output in current and voltage.


2

SHIELD
9

FUN

Analog voltage output: range from -10 to 10 V or -5 to 5 V, 10 KΩ CURRENT (300 Ω max)


minimum load.
10

VOLTAGE (10 kΩ min)


NET

ANALOG COM.
Analog current output: range from 0 to 20 mA or 4 to 20 mA. The
0

maximum load is 300 Ω.


To achieve the serial connection, use a suitable shielded cable,
PRG

making sure to connect the shield to one of the two ends: to pin
18

9 if connected on the side of the instrument, to the ground if it is


connected on the opposite side.
Attention: do not connect the analog output to devices that are
switched on.

Page 7
PROFIBUS DP CONNECTION (OPTIONAL)

Pin Signal Description


1 - -
2 - -
3 B line +RxD/+TxD, level RS485
27

4 RTS Request to send


5 GND Ground (isolated)
1
6

3 B_LINE 6 + 5V Bus Output +5V termination (isolated)


7 - -
8 A_LINE
8 A line -RxD/-TxD, level RS485
9
5
19

9 - -
Housing Cable shield Internally connected to protective
earth according to Profibus
specification
For connection to the Profibus Master, use a standard Profibus cable.
The typical impedance of the cable should be between 100 and
130 Ohms (f> 100 kHz). The cable capacity (measured between
conductor and conductor) should be less than 60 pF / m and the
minimum cable cross section should not be less than 0.22 mm2
In a Profibus-DP network, you can use either cable type A to type B
cable, depending on the required performance. The following table
summarizes the features of the cable to be used:
Specification Type A Cable Type B Cable
Impedance from 135 to 165 ohm from 100 to 300 ohm
(f = 3 – 20 MHz) (f > 100 kHz)
Capacity < 30 pF/m < 60 pF/m
Resistance < 110 ohm/km -
Conductor cross > 0,34 mm2 > 0,22 mm2
section
The following table shows the maximum length of the wires line with
cable type A and type B, function of the different communication
speed required:
Baud rate
9.6 19.2 187.5 500 1500 3000 6000 12000
(kbit/s)
Cable A lenght
1200 1200 1000 400 200 100 100 100
(m)
Cable B lenght
1200 1200 600 200 - - - -
(m)
For a reliable operation of the Fieldbus, should be used a line ter-
mination at both ends.
In the case of multiple DAT 400 instruments, use the line termination
at only one instrument.
For configuring the instrument, the GSD file is available (hms_1810.
GSD) that must be installed in the master.

Page 8
DEVICENET CONNECTION (OPTIONAL)
Pin Signal Description
1 V- Negative power bus
2 CAN_L CAN low bus line
27

3 SHIELD Shield
25

2 CAN_L 4 CAN_H CAN high bus line


CAN_H 5 V+ Positive power bus
2 3 4 5

1 V- To connect to the DeviceNet master, use a standard DeviceNet cable


5 V+ or shielded twisted-pair cable as shown on the diagram.
1

3 SHIELD The cable must not be channeled with power cables. For reliable
operation of the Fieldbus, should be used as a line termination of
121 Ω value between the terminal CAN_L and CAN_H.
For the configuration of the card is available ESD file that must be
installed in the master.

Page 9
FRONT PANEL OF THE INSTRUMENT
The DAT 400 has a bright 6-digit display, 4 status LEDs and four keys.
In this operating mode the display shows the weight and the LEDs indicate the status of weight and the
setpoints.
The set-up parameters are easily accessed and modified through the use of the three front buttons used
to select, edit, confirm and save the new settings.

PRECISE
DAT 400 8.8.8.8.8.8.
1 2 NET
SET FUN 0 PRG

DISPLAY
On the 6-digit display, it’s usually shown the scale weight. According to the various programming
procedures, the display is used for programming of the parameters to be stored in the memory, or the
messages that indicate the type of operation being carried out and help therefore the Operator in the
management and programming of the instrument.

LED INDICATORS
Below the display there are 4 LED indicators:
1 State of the logic output 1 (ON = closed contact OFF = open contact)
2 State of the logic output 2 (ON = closed contact OFF = open contact)
NET The displayed value is the net weight
0 IT indicates the condition of stable weight.

Page 10
USING THE KEYBOARD
The instrument is programmed and controlled through the keyboard which has 4 keys, with double
functions. The selection of one of the key functions is established automatically by the instrument accor-
ding to the operation in progress. In general, the management of the programming menus is done by
using the SET and FUN keys to scroll through the items; the PRG key is used to enter its sub-menu or
programmable parameter, while the 0 button allows you exiting the menu or returning to the top level.

KEY FUNCTIONS DURING THE WEIGHT DISPLAY

SET
Access to the menu for the programming of the setpoints

FUN Select the display view (gross weight, net weight).


(Long press) Selection of the weight/peak display
Resetting the displayed value (gross weight, net weight or peak).
0
(Press and hold for 5 sec.) Calibration of zero, to be executed only if its function
is enabled in the PARAM menu (see item “0 ALL”).

PRG Sending the weight string on the serial line.


(Long press) Access to the quick set-up menu.

PRG SET
(Press for 3 sec) Access to the setup menu.
+

PRG 0 (Press for3 sec) It accesses the keypad lock/unlock menu and auto-off function of
+ the display.

KEY FUNCTION DURING THE MAIN MENU DISPLAY


SET
It selects the next parameter.

FUN
It selects the previous parameter.

0
It exits the programming menu or returns to the upper level.

PRG It accesses the corresponding sub-menu or programming or confirms the selected


parameter.

KEY FUNCTION WHEN SETTING THE NUMERICAL VALUES


SET
It increases the value of the flashing digit.

FUN
It decreases the value of the flashing digit.

0
It goes to the next digit.

PRG
It confirms the displayed value.

Page 11
KEY FUNCTION WHEN SETTING THE NUMERICAL VALUES
SET
It selects the next value.

FUN
It selects the previous value.

PRG
It confirms and store the displayed value.

EXIT FROM THE SETTING MENU

0 0
Press the key to return to the main menu. Press the . key again. It’s displayed “StORE?”.
PRG
Press the key to return to the main menu.

PRG
To exit without saving any changes, switch off the instrument instead of pressing the key.

Page 12
INFO DISPLAY
When the instrument is switched ON, you can test the display, then in sequence you can display the
identification code of the software and its version. Communication codes in the event of a request for
assistance.

ERRORS NOTIFICATION
In the operation mode, the display can report the following error codes.
ϩϩϩϩϩϩ the weight applied to the load cell exceeds by more than 9 divisions the maximum capacity
of the weighing system.
O-L Signal of the load cells absent or outside of the measurement range mV/V.
No CoN Fieldbus network disconnected
E-ProF PROFIBUS interface absent or not operating.
E-dnEt DEVICENET interface absent or not operating.
ϩϩϩϩϩϩ Dash that runs along the perimeter of the display: BLIND function enabled.
ErrNEN Memory error. Press the PRG key to reset the memory and return the parameters to their
default values. NOTE: it is also deleted the calibration performed.

Page 13
VIEWING, ZEROING THE WEIGHT AND SELF-CALIBRATION
After being calibrated, at the subsequent switches on, the display shows the current weight.

VIEWING THE NET WEIGHT/GROSS WEIGHT

FUN
Press the key to toggle between the net weight and the gross weight and vice versa. The value
displayed is signaled by the LED NET (lit: net weight). If you have not entered the tare, the net weight
is equal to the gross weight.
In case of negative weight, it is displayed the minus sign before the most significant digit.

ZEROING, WEIGHT AND SELF-CALIBRATION


These two functions are performed by pressing 0.
When the instrument is in the operation mode “Net” (“NET” LED on), the 0 key performs the self-
calibration.
When the instrument is in operation mode “Gross” (“NET” LED off), the 0 key clears the gross weight.

SELF-CALIBRATION
The execution of self-calibration is possible under the following conditions:
• Instrument under conditions of “Net” (NET” LED on).
• Positive gross weight.
• Gross weight not greater than the maximum capacity.
• Stable weight.
• Unstable weight. In this condition, we must distinguish two cases:
1. The weight stability control is enabled (the parameter “MOTION” (*) must be other than zero): the
command executed while the weight is unstable only has an effect if the weight stabilizes within 3
seconds from the moment the command was given.
2. The weight stability control is disabled (the parameter “MOTION” (*) is equal to zero): the executed
command takes effect immediately, even with unstable weight.
(*) The operating modes of the parameter “MOTION” are described on page 28.
The self-calibration is retained in memory even after the power is switched off.

ZEROING
The reset command of the gross weight is used to correct for small zero shifts of the weighing system
during normal operation.
Normally these zero shifts are due to thermal drifts or to residues of material that accumulate on the
weighing system over the time.
To run the command, it is necessary that the instrument is under conditions of “Gross” (“NET” LED off)
and that the deviation of the weight with respect to the zero of the scale (the one performed with the
calibration of zero) does not exceed (in positive or negative) the number of divisions set in parameter
“0 BAND” (inside the PARAM menu).
The reset command of the gross weight does not run if there is even one of the following conditions:

Page 14
• Unstable weight (with control of the stability of the weight enabled). In this case, the reset command
takes effect only if the weight stabilizes within 3 seconds or if the control of the weight stability is
disabled (parameter “MOTION “ equal to zero).
• Gross Weight greater (in positive or negative) than the number of divisions set in parameter “0
BAND” , when the setpoint of auto-calibration is not programmed.
The zero obtained with the resetting of the gross weight is retained in memory even after the power is
switched off.
The reset operation of the gross weight can be repeated several times, but the number of reset divisions
zero is added from time to time, so when the total exceeds the limit value set in parameter “0 BAND”,
the zero cannot be executed. In this case, it is necessary to calibrate the Zero.
Any auto-zero parameter setting when switching on (AUTO 0) reduces (or cancels, in the case of “AUTO
0”> “0 BAND”) the range of action of the reset command.

PEAK FUNCTION
The instrument continuously memorizes the peak value of the gross weight. The peak value is detected
at the same frequency of acquisition of the weight (see table on filters). In addition to visualization, the
peak value can be used in the following functions:

FUNCTION DESCRIPTION
The setpoints can be set to have the peak value as a reference. (See the
LOGIC OUTPUT
procedure for setting the logic outputs operations).
Acquisition of the peak value (peak hold) through the CONTIN, AUTO, DE-
SERIAL PORT
MAND, SLAVE and MODBUS protocols.
The analog output value can assume the value of the peak (peak old).
ANALOG OUTPUT
(See the procedure for setting the analog output).
Press the “FUN” key and hold it for 3 seconds until the left of the display shows the letter “P”.

INPUT / OUTPUT FUNCTIONS

INPUT
Resetting the displayed value (gross weight, net weight or peak). Closed for 5 sec. ->
1 Calibration of zero, to be executed only if its function is enabled in the PARAM menu (see
item “0 ALL”).
2 Sending the weight string on the serial line or print.
OUTPUT
1 Setpoint 1
2 Setpoint 2

Page 15
PROGRAMMING THE WEIGHT SETPOINTS
The set setpoint values are compared with the weight to drive its logic output. The comparison criterion
is established in the process of set-up of the logic I / O (see relevant paragraph).
To access the Setpoint setting, press the SET key and follow the instructions on the table below.


SET

sEt 1 PRG
123456 PRG

SET SET FUN 0

Increment Change
digit selected digit

FUN

SEt 2 PRG
123456 PRG

0 SET FUN 0

Increment Change
EXIT digit selected digit

During the step of setting the setpoints, both outputs are disabled. If the setpoint value in memory is 0,
the corresponding output is never enabled, regardless of the set-up of the selected setpoints. When the
weight is not detectable or out of range, all outputs are disabled (contact open or closed depending
on the MODE; see the relevant chapter).

Page 16
KEYBOARD LOCK/UNLOCK FUNCTION
KEYBOARD LOCK/UNLOCK A function that allows you to enable or disable the keys individually.
When the keys are locked, the only way to access these settings is to press and hold pressed the PRG
+ 0 keys for 3 seconds. For more information on the function, refer to the block diagram above.
SWITCHING THE DISPLAY OFF This function allows turning off the display after a programmable time.
KEYS CORRESPONDING
3 sec.
PRG + 0
 SET FUN 0 PRG

LOck PRG
0000 PRG

SET SET FUN 0

Change Select
selected digit digit

FUN

blind PRG
off PRG

SET FUN

0
on PRG

EXIT

00 PRG

SET FUN 0

Change Select
selected digit digit

You can select ON / OFF of the parameter BLIND and the setting of a time; the time count starts from
the moment when, after exiting the setup menu, the display shows the weight value. After the set time,
the display turns off and only a dash appears. This dash cycles through the perimeter of the display
counterclockwise. When the display is off, also the 4 keys are disabled, regardless of how you set the
keypad lock (LOCK). The only way to access the settings will be PRG + 0.

Page 17
SETTING

GENERAL DATA
All functions of the DAT 400 are activated and modified by accessing a simple setup menu, shown on
page 21. All settings selected or activated remain stored even after switching off the transmitter.
The DAT 400 is preconfigured with a default setting. The following pages show the values of “Default”
for each parameter.
With the first on-site installation, it’s necessary to change some parameters in order to obtain a correct
indication of the displayed weight (Theoretical adjustment).
The settings of the setup menu can be changed using the front keys or via the utility “INOVATION 2”
software supplied.

CHANGING AND ENTERING THE PARAMETERS:


The setup parameters are grouped into a number of main menus.
To access the setup menu press the PRG key and then the SET key and hold down simultaneously for
3 seconds.
The display shows the message Conf1G that is the first of the main menus.
Use the arrow keys to select the menu you want to change.
Press the PRG key to access the selected menu.

KEY FUNCTION DURING THE MAIN MENU PROGRAMMING


SET
It selects the next menu.

FUN
It selects the previous menu.

0
It exits the programming menu or returns to the upper level.

PRG It accesses the corresponding sub-menu or programming or confirms the selected


parameter.

KEY FUNCTION WHEN SETTING THE NUMERICAL VALUES


SET
It increases the value of the flashing digit.

FUN
It decreases the value of the flashing digit.

0
It goes to the next digit.

PRG
It confirms and store the displayed value.

Page 18
KEY FUNCTIONDIAGRAMA
WHEN SETTINGDEL MENÚVALUES
THE PROSED
SET
It selects the next value.

FUN
It selects the previous value.

PRG
It confirms and store the displayed value.

The menu parameters can assume values that


​​ can be set or selected.

0
NOTE To exit and save the modified data, press multiple times the key until the display shown
PRG
StorE, then press key to go back to the operating mode.

Page 19
ENTER MENU 0 0 0 0 0 0
FUN FUN FUN FUN FUN
3 sec.
F
PRG + SET Conf1G SET CaL1br SET ParaN SET 1n-oUt SET SEr1aL SET AnaLoG StorE?
0 PRG
PRG PRG PRG PRG PRG PRG
0
CapaC PRG
CaL 0
fILtEr 0
NodE 1 0
baUd r 0
f-sCaL EXIT MENU
SET FUN 0 SET FUN SET FUN SET FUN SET FUN
SEns1t 0 Notion Hyst-1 baud2r NodE
SET FUN SET FUN SET FUN SET FUN SET FUN
nEt CaL auto 0 t1NEr1 Prot-1 AnzEro
SET FUN SET SET FUN SET FUN SET FUN SET FUN
dEad L 123456 0 trac dELay1 prot-2 tEst
SET FUN SET FUN 0 SET FUN SET FUN SET FUN SET FUN
DIAGRAM OF THE MENU

dspd1J 60000 0 band NodE 2 AddrEs ranGE


SET FUN PRG SET FUN SET FUN SET FUN SET FUN
S1GnaL 0 ALL Hyst-2 Pr-Add OffsEt
SET FUN SET FUN SET FUN SET FUN SET FUN
CoUnts t1NEr2 dELay
SET FUN SET FUN SET FUN
DEVICENET
opNodE dELay2 rEN-Co dN-Add
SET FUN SET FUN SET FUN
SET FUN
UpLoad tEst1n data F
SET FUN SET FUN SET FUN
ETHERNET IP
PROFINET
dnLoad tstoUt t-Nap
SET FUN SET FUN SET FUN 1P-Add
SET FUN
r-NaP
SET FUN SubnEt
SET FUN
dELNap
SET FUN

Page 20
QUICK SETUP MENU

ENTER MENU

5 sec.


PRG

CapaC
SET FUN PRG
Enter Value PRG

SEns1t
SET FUN PRG
Enter Value PRG

nEt
SET FUN PRG
Enter Value PRG

dEad L
SET FUN PRG
Enter Value PRG

dspd1J
SET FUN PRG
Select Value PRG

S1GnaL
SET FUN PRG
See Value 0

CaL1br PRG
SET FUN PRG
CaL
0
AnaLoG
SET FUN PRG Enter Value PRG
0

CaL
SET
0
StorE?
PRG
123456
SET FUN 0

EXIT MENU
60000
PRG

Page 21
CONFIGURATION PARAMETERS
Through the setting of the parameters listed below, the theoretical Full Scale DAT 400 calibration is
performed. These steps with the zero calibration described on the next page must be completed. The
procedure ensures a good accuracy of the system (maximum error <1% FS) if there are no mechanical
problems.
Program the known values of total capacity and sensitivity of the load cells and the approximate values
of net capacity and calibration. If the parameter SENSIT is not programmed, it is taken the 2.0000
mV/V value.
If the parameter CAPAC is programmed other than 0, according to the data CAPAC, SENSIT, NET and
DEAD L , the instrument automatically runs the following functions:
Resetting the linearization points.
Selection of the value of the division, however, to be modified, to the best of 10,000 divisions.
Calibration of the theoretical approximate calibration of the weight (zero and full scale).
Automatic programming of the overload setpoint (= NET).
These functions are performed each time you change one of the 4 parameters shown.
When you change the DSPDIV selection., it is automatically recalculated to full-scale calibration. The
selections are incompatible with the calibration parameters or calibration in memory are not accepted.
The selection programmed in Opmode is read from the instrument when it is switched on and it makes
that the instrument operates in that way.

CAPAC CAPACITY OF THE WEIGHING SYSTEM FUN FUN

AnaLog
It defines the value corresponding to the sum of the rated capacity
SET Conf1G SET CaL1br
PRG

of the load cells. In the case of systems with only one load cell and
“N” fixed supports, enter the capacity value of the cell for the total 0
CapaC PRG
Enter Value PRG

number of supports. This figure represents the full scale value of the
SET FUN

weighing system. 0
SEns1t
SET FUN
PRG
Enter Value PRG

Following the change of the parameter value, the theoretical tare of


the weight is recalculated. 0
nEt
SET FUN
PRG
Enter Value PRG

Values: from 1 to 500000


Unit: the same of that displayed
0
dEad L
SET FUN
PRG
Enter Value PRG

Default: 10000
0
dspd1J
SET FUN
PRG
Select Value PRG

sEnsit LOAD CELLS SENSITIVITY


0
S1GnaL
Set the value corresponding to the average sensitivity of the load SET FUN
PRG
Read Value 0

cells, in mV / V. The instrument accepts values between 0.5 and 4


mV / V. If no value if programmed, it’s assumed it is 2mV/V.
0
CoUnts
SET FUN
PRG
Read Value 0

Following the change of the sensitivity value, the theoretical tare of


the weight is recalculated.
0
opNodE
SET FUN
PRG
Select Value PRG

Values: from 0.5000 to 4.0000 mV/V 0


UpLoad Receive setup data
Default: 2.0000 SET FUN
PRG

from serial port


0
dnLoad Transmit setup data
PRG
SET FUN

to serial port

Page 22
NEt CAPACITY OF THE WEIGHING SYSTEM
FUN FUN
Programming the net capacity of the weighing system. Values lower
AnaLog Conf1G CaL1br
than 1/10 of CAPAC are not accepted.
SET SET

PRG

0
CapaC Values: from 1 to 500000
SET FUN
PRG
Enter Value PRG

Unit: the same of that displayed


0
SEns1t Default: 10000
SET FUN
PRG
Enter Value PRG

0
nEt dEad L FIXED CALIBRATION OF THE WEIGHING SYSTEM
PRG
Enter Value PRG

Programming the fixed calibration value of the weighing system.


SET FUN

0
dEad L Values: from 1 to 500000
SET FUN
PRG
Enter Value PRG

Unit: the same of that displayed


0
dspd1J Default: 00000
SET FUN
PRG
Select Value PRG

0
S1GnaL dsPdIJ DIVISION VALUE
SET FUN
PRG
Read Value 0

The ratio between the capacity of the system and the division value
0
CoUnts represents the resolution of the system (number of divisions).
SET FUN
PRG
Read Value 0

Following the change of the capacity of the system, it is automatically


0
opNodE selected the division value to the best of 10000 divisions.
SET FUN
PRG
Select Value PRG

0
UpLoad Receive setup data Following the change of the division value, if the maximum capa-
SET FUN
PRG
city does not change, the calibration of the weight is automatically
from serial port
corrected.
0
dnLoad Transmit setup data
Value to be selected:
PRG
SET FUN

to serial port
0.0001 - 0.0002 - 0.0005
0,001 - 0,002 - 0,005
0.01 - 0.02 - 0.05
0.1 - 0.2 - 0.5
1 -2 - 5
10 - 20 - 50
Default: 1

S1GnaL TESTING THE LOAD CELLS SIGNAL


It’s displayed the signal acquired from the load cells expressed in
mV/V.

CoUnts A/D CONVERTER INTERIOR POINTS TEST


View of the interior points of the instrument (1,000,000 at the ma-
ximum input signal).

oPNodE SELECTION OF THE OPERATING MODE


Selection of the operating mode of the device (display) when it is
switched on:
Value to be selected:
GROSS, NET, PEAK
Default: GROSS

Page 23
UpLoad RECEIVING DATA FUNCTION
Receiving function from a serial of a file containing the setup data FUN FUN

AnaLog Conf1G CaL1br


that will be automatically set in the instrument.
SET SET

PRG

0
CapaC
dnLoad SENDING DATA FUNCTION SET FUN
PRG
Enter Value PRG

Sending function from a serial of a file with the content of the setup 0
SEns1t
memory of the instrument. SET FUN
PRG
Enter Value PRG

0
nEt
EXAMPLE OF SETTING/CALIBRATION SET FUN
PRG
Enter Value PRG

To weigh a tank, with empty weight of 750 kg and with a capacity 0


dEad L
of 1000 liters, containing a product with a specific gravity of 1.33 SET FUN
PRG
Enter Value PRG

to read the weight with a display resolution of 0.2 Kg. 0


dspd1J
Before proceeding with the configuration, make sure that the load SET FUN
PRG
Select Value PRG

cells are connected properly to the unit and the tank is empty and 0
S1GnaL
set the parameters. SET FUN
PRG
Read Value 0

They are used: 0


CoUnts
SET FUN
PRG
Read Value 0

Nr 3 load cells with capacity of 1000 kg


Sensitivity of respectively 2.0015, 2.0008 and 1.9998 mV/V (ave-
0
opNodE
SET FUN
PRG
Select Value PRG

rage value = 2.0007 mV/V)


0
UpLoad Receive setup data
Set the following values in the configuration parameters: SET FUN
PRG

from serial port


CapaC = 3000 0
dnLoad Transmit setup data
PRG

sEnsit 2.0007
SET FUN

to serial port

NEt 1500
dEad L 0
dsPd1J 0.2
Make sure that the value read in the SIGNL parameter corresponds
with the calibration weight of the system according to the following
proportion:
3000:2.0007=750:X
Where X is the value of the signal expressed in mV/V corresponding
to the theoretical value of the weight of the empty tank. The value
should be about 0.5 mV/V.
Proceed to the calibration described in the following paragraph,
or you can exit the configuration menu by saving the data entered.
The instrument should indicate the value corresponding to the weight
of the empty tank (for example 756.8).
Re-access the configuration menu and enter the weight value read
in the dEad L parameter and enter the value 756.8.
Quit the configuration menu by saving the data.
For greater accuracy, prepare some sample weights or the pre-
weighed material on a certified system and calibrate as described
in the next paragraph.

Page 24
CALIBRATION
The calibration described herein should be performed with the use of sample weights and/or product
pre-weighed on a weighing system.
Before proceeding with the calibration of the full scale, always perform the zero calibration.
During the calibration phase, the display shows the weight intermittently with the inscription CaL.
ATTENTION: At switching off the instrument without exiting the set-up menu, the programming executed
is not stored.
NOTE In the event that after calibration, the system shows linearity errors, you should verify that the
weighed structure is completely free of mechanical constraints.

ZERO CALIBRATION FUN FUN

Conf1G CaL1br ParaN


Perform the operation when the scale has no items (including the
SET SET

PRG

fixed tare), and when the weight is stable. The zero of the system is
done by pressing the key 0.
PRG
CaL With unload
35 scale
The weight displayed resets and the display shows Cal alternated 0

by 0. You can repeat this operation more times. 0


PRG
CaL Load Weight
CALIBRATION OF FULL SCALE SET

Prior calibration load the sample weight on the system and wait for
123450
SET FUN 0

the stabilization; the display shows a weight value. Enter Load


120000 Value
Press the SET key to adjust the weight. The display shows the theo- PRG

retical weight value with the first digit to the left flashing. With the PRG
CAL
arrow keys, enter the actual weight loaded on the system starting
from the first flashing digit. Switch to the next digit by pressing 0.
The confirmation of the last digit (far right) with the PRG key corrects
the weight. The display shows CAL, by altering the actual weight
value entered.
If the set value is higher than the resolution offered by the instrument,
the weight is not accepted and the display shows an error message
for a few seconds.
This procedure can be repeated.
Press the PRG key again to return to the CaLIb. menu.

EXIT FROM CALIBRATION MENU

0
The exit from the menu CaLIb is performed by pressing the
key until the appearance of the message store?.
To store the new calibration and exit the setup menu, press the PRG
key.
To cancel the calibration of zero and full scale:
FUN +0 it cancels the zero calibration.
FUN + SET it cancels the calibration of the full scale.

Page 25
FUN FUN
LINEARIZATION PROCESS
Config CaL1br Param
When planning the sample weight, values greater than the full scale,
SET SET

PRG

or lower than the previous point, or when the weight is not stable,
PRG
CaL With unload
are not accepted. If the entered value is accepted, it is proposed the
35 scale
0
next step, otherwise still the same.
0 The linearization points are automatically reset by any change of
PRG
CaL Load Weight P1 the data of the theoretical calibration or it is performed a full-scale
SET 4 sec.
calibration.

0
LIn P1 PRG

000000
SET FUN 0

12000 Enter Load


PRG
Value P1
0
LIn P2 Load Weight P2PRG

000000
SET FUN 0

24000 Enter Load


Value P2PRG

0
LIn P9 Load Weight P9PRG

000000
SET FUN 0

120000 Enter Load


Value P9 PRG

0
CaL

Page 26
WEIGHING PARAMETERS
The parameters in this menu allow the adjusment of the acquisition and updating timing of the display
and the manual or automatic zeroing performed by transmitter

F1LtEr WEIGHT FILTER


FUN FUN

CaL1br
This parameter adjusts the refresh speed of the display and the serial SET ParaN SET In-oUt
and analog output.
PRG

Low values of the filter speed up the display refresh.


0
f1LtEr
SET FUN
PRG
Select Value PRG

High values of the filter slow down the display refresh. 0


Not1on
Value Update Response SET FUN
PRG
Select Value PRG

0 50 Hz 25 Hz 0
auto 0
1 50 Hz 16 Hz SET FUN
PRG
Enter Value PRG

2 25 Hz 8 Hz 0
0 trac
3 25 Hz 5 Hz SET FUN
PRG
Select Value PRG

4 25 Hz 2.5 Hz 0
0 band
5 10 Hz 1.5 Hz Enter Value
PRG PRG
SET FUN

6 10 Hz 1 Hz 0
0 ALL
7 10 Hz 0.7 Hz SET FUN
PRG
Select Value PRG

8 5 Hz 0.4 Hz
9 5 Hz 0.2 Hz
Default: 5

Not1on WEIGHT STABILITY


This parameter defines the divisions number needed to deem the
weight stable.
A large number of divisions allows the transmitter to detect quickly
the weight stability, which is needed when executing tare and print
commands.
Value Change
0 Always stable weight
1 Stability determined quickly
2 Stability determined with medium parameters
3 Stability determined accurately
4 Stability determined with the highest accuracy
Default: 2

Auto 0 AUTOZERO UPON SWITCHING ON


This parameter defines the value of the maximum resettable weight
when the instrument is switched on.
This operation corresponds to a zero calibration of the system and
is executed only if the weight is stable and below the set value.
Value from 0 to the value of the CAPAC parameter.
Default: 0

Page 27
0 trAc TRACKING THE ZERO
FUN FUN

CaL1br SET ParaN SET In-oUt This function allows you to perform a momentary zero calibration
compensating for the temperature drift of the weight.
PRG

0
f1LtEr At swithching off, it automatically returns to the previous zero cali-
PRG
Select Value PRG

bration.
SET FUN

0
Not1on The maximum weight resettable by this parameter is 2% of the range
SET FUN
PRG
Select Value PRG

of the system.
0
auto 0 To disable this feature, use the value 0.
SET FUN
PRG
Enter Value PRG

0
0 trac Value Change
SET FUN
PRG
Select Value PRG

0 Control OFF
0
0 band 1 0.5 div/sec
SET FUN
PRG
Enter Value PRG

2 1 div/sec
0
0 ALL
SET FUN
PRG
Select Value PRG
3 2 div/sec
4 3 div/sec
Default: 0

0 band ZERO BAND


This parameter defines the number of divisions resettable by the
pressure of the front button of zero or by Input 1.
Values: from 0 to 200
Default: 100

0 ALL ZERO CAL


Enables the function that allows to perform zero calibration by pres-
sing and holding the 0 key or input 1 for 5 sec.
Value:
Off Function disabled
On Function Enabled
Default: Off

Page 28
INPUT/OUTPUT PARAMETERS

NodE 1 SETPOINT 1 OPERATION MODE


FUN FUN

ParaN SET In-oUt SET SEr1aL Select 4 operation criteria of the setpoint 1 in sequence:
PRG

NET The relay output is active in Net Weight mode


0
NodE 1 GROSS The relay output is active in Gross Weight mode
SET FUN
PRG
Select Value
PRG
PEAK The relay output is active in Peak mode
Select Value
PRG
Default: GROSS
Select Value
PRG Comparison with the net weight, gross weight or the peak. In this
Select Value PRG
last case, the comparison is made with the last peak value acquired,
0
Hyst-1 even when the peak function is not active.
PRG
Enter Value PRG

N.O. The relay 1 is normally open


SET FUN

0
t1NEr1 N.C. The relay 1 is normally closed
SET FUN
PRG
Enter Value PRG

Default N.O.
0
dELay1 POS. The output is operating with positive weight
SET FUN
PRG
Enter Value PRG

NEG. The output is operating with negative weight


0
NodE 2 Default: POS
SET FUN
PRG
Select Value
NORML Output 1 is active with unstable weight
PRG

Select Value
PRG
STABL The output is active with stable weight
Select Value
PRG Default: Norml
Select Value PRG

0
Hyst-2 Hyst-1 HYSTERESIS OF THE SETPOINT 1
SET FUN
PRG
Enter Value PRG

Hysteresis value than the setpoint value set.


0
t1NEr2
SET FUN
PRG
Enter Value PRG
Value: from 0 to 999
Default: 2
0
dELay2
SET FUN
PRG
Enter Value PRG

t1NEr1 SETPOINT 1 TEMPORIZATION


0
tEst1n Read and
Value of time, in tenths of a second, during which, when the weight
PRG 0
SET FUN

change status
value set is overcome, the output relative to setpoint 1 remains
0
tstoUt
SET FUN
PRG Read and 0
enabled.
change status
After this time, even if the weight value is still above the setpoint, the
output is automatically disabled.
The function is not activated if the programmed time is equal to zero.
Value: from 0 to 999
Default: 0

dELay1 SETPOINT 1 DELAY


Value of time, in tenths of a second, after which, when the weight va-
lue set is overcome, the output relative to setpoint 1 remains enabled.
The function is not activated if the programmed time is equal to zero.
Value: from 0 to 999
Default: 0

Page 29
NodE 2 SETPOINT 2 OPERATION MODE FUN FUN

ParaN In-oUt SEr1aL


Select 4 operation criteria of the setpoint 2 in sequence:
SET SET

PRG

NET The relay output is active in Net Weight mode 0


NodE 1
GROSS The relay output is active in Gross Weight mode SET FUN
PRG
Select Value
PRG

PEAK The relay output is active in Peak mode Select Value


PRG

Default: GROSS Select Value


Comparison with the net weight, gross weight or the peak. In this
PRG

Select Value PRG

last case, the comparison is made with the last peak value acquired,
even when the peak function is not active.
0
Hyst-1
SET FUN
PRG
Enter Value PRG

N.O. The relay 2 is normally open 0


t1NEr1
N.C. The relay 2 is normally closed SET FUN
PRG
Enter Value PRG

Default N.O. 0
dELay1
POS. The output is operating with positive weight Enter Value
PRG PRG
SET FUN

NEG. The output is operating with negative weight 0


NodE 2
Default: POS SET FUN
PRG
Select Value
PRG

NORML Output 2 is active with unstable weight Select Value


PRG

STABL Output 2 is active with stable weight Select Value


Default: Norml
PRG

Select Value PRG

0
Hyst-2
Hyst-2 HYSTERESIS OF THE SETPOINT 2 SET FUN
PRG
Enter Value PRG

Hysteresis value than the setpoint value set 0


t1NEr2 PRG
Enter Value PRG

Value: from 0 to 999


SET FUN

Default: 2 0
dELay2
SET FUN
PRG
Enter Value PRG

t1NEr2 SETPOINT 2 TEMPORIZATION 0


tEst1n PRG Read and 0
SET FUN

Value of time, in tenths of a second, during which, when the weight change status
value set is overcome, the output relative to setpoint 2 remains 0
tstoUt PRG Read and 0

enabled.
SET FUN

change status
After this time, even if the weight value is still above the setpoint, the
output is automatically disabled.
The function is not activated if the programmed time is equal to zero.
Value: from 0 to 999
Default: 0

dELay2 SETPOINT 2 DELAY


Value of time, in tenths of a second, after which, when the weight va-
lue set is overcome, the output relative to setpoint 2 remains enabled.
The function is not activated if the programmed time is equal to zero.
Value: from 0 to 999
Default: 0

Page 30
tEst1n LOGIC INPUTS TEST PROCEDURE
FUN FUN

ParaN SET In-oUt SET SEr1aL The display shows the inputs status.
PRG

0 = input disabled
0
NodE 1
SET FUN
PRG
Select Value 1= input activated.
PRG

Select Value The input 1 corresponds to the 1st value on the left.
PRG

Select Value Enable and disable the inputs to check the corresponding state on
the display. During this procedure, the normal function of the inputs
PRG

Select Value PRG

is not active. Use this procedure only to check the hardware.


0
Hyst-1
SET FUN
PRG
Enter Value PRG

tstoUt LOGIC OUTPUTS TEST PROCEDURE.


0
t1NEr1
Enter Value The display shows the outputs status.
PRG PRG
SET FUN

0
dELay1 0 = output disabled, 1= output activated.
SET FUN
PRG
Enter Value PRG

The input 1 corresponds to the 1st value on the left.


0
NodE 2 During this procedure, the LEDs reflect the state of the outputs. To set
SET FUN
PRG
Select Value
PRG
the digits, use the keys as for the numeric settings.
Select Value
PRG
During this procedure, the normal function of the outputs is not active.
Select Value
PRG Use this procedure only to check the hardware.
Select Value PRG

0
Hyst-2
SET FUN
PRG
Enter Value PRG

0
t1NEr2
SET FUN
PRG
Enter Value PRG

0
dELay2
SET FUN
PRG
Enter Value PRG

0
tEst1n PRG Read and 0
SET FUN

change status
0
tstoUt PRG Read and 0
SET FUN

change status

Page 31
SERIAL OUTPUT PARAMETERS
This menu allows you to set the serial ports COM1 and COM2 and the communication parameters.
The instrument has two independent serial ports:
COM1 with RS232 or RS422/RS485 interface
COM2 with
ENTER MENU 0
optional FIELDBUS
0
interface. 0 0 0 0
FUN FUN FUN FUN FUN

3 sec.

F
PRG + SET
r BAUD RATE
Conf1G
baUd SET
COM1
CaL1br SET ParaN SET 1n-oUt SET SEr1aL SET AnaLoG S
It defines the baudrate of the RS232 serial port.
PRG PRG PRG PRG PRG PRG

The
0
CapaC
value must be set CaL
PRG to the same value
fILtEr
0 of the PC / PLC 0or remote
NodE 1 0
baUd r 0
f-sCaL
display.
SET FUN 0 SET FUN SET FUN SET FUN SET FUN

Value to be selected:
SEns1t Notion Hyst-1 baud2r NodE
0
2400 SET FUN SET FUN SET FUN SET FUN SET FUN

9600
nEt CaL auto 0 t1NEr1 Prot-1 AnzEro
19200
SET FUN SET SET FUN SET FUN SET FUN SET FUN

38400
dEad L
115200 123456 0 trac dELay1 prot-2 tEst
SET FUN SET FUN 0 SET FUN SET FUN SET FUN SET FUN

Default: 9600
dspd1J 60000 0 band NodE 2 AddrEs ranGE
baud2r
SET FUNBAUD RATE COM2 (DEVICENET
PRG
VERSION)
SET FUN SET FUN SET FUN SET FUN

It defines the baudrate of the DEVICENET interface.


S1GnaL 0 ALL Hyst-2 Pr-Add OffsEt
The value
SET FUN must be set to the same value of
SET the
FUNPC/PLC. SET FUN SET FUN SET FUN

Value to be selected:
CoUnts t1NEr2 dELay
125 SET
250FUN500 DeviceNet SET FUN SET FUN

Default: DEVICENET
opNodE
125 SET
DeviceNet
dELay2 rEN-Co dN-Add
FUN SET FUN SET FUN
SET FUN

UpLoad
Prot-1 COM1 PROTOCOL tEst1n data F
SET FUN
It defines how to use the RS232 serial port
SET FUN SET FUN
ETHERNET IP
PROFINET
dnLoad
None: Serial communication OFF tstoUt t-Nap
SET FUN
Contin: Continuous transmission of the weight string. It can be used,
SET FUN SET FUN 1P-Add
SET FUN
for example, to drive a weight repeater. See details in the relevant r-NaP
paragraph. SET FUN SubnEt
SET FUN
Demand: When the Operator presses the front button or through
Input 2, a string of weight is transmitted. The command is accepted
dELNap
SET FUN
if the weight is stable. Between two consecutive transmissions the
weight must have a variation of at least 20 divisions.
Autom-: It’s automatically transferred to a string of weight when the
weight stabilizes at a value higher than the minimum weight (20
divisions). Between two consecutive transmissions, the weight must
have a variation of at least 20 divisions.
Slave: ASCII protocol. See details in the relevant paragraph.

Page 32
0 0 0
UN FUN FUN
Modbus: MODBUS RTU (slave) protocol. used only if PROT-2 is
1n-oUt SET SEr1aL SET AnaLoG StorE?
configured equal NONE. See details in the relevant paragraph.
0 PRG
PRG PRG PRG
Selectable communication patrameters:
NodE 1 0
baUd r 0
f-sCaL
n-8-1 EXIT MENU
SET FUN SET FUN SET FUN
n-8-2
Hyst-1 baud2r NodE E-8-1
SET FUN SET FUN SET FUN
o-8-1
Default: n-8-1
t1NEr1 Prot-1 AnzEro
SET FUN SET FUN SET FUN Print: Data transfer to the printer.

dELay1 prot-2 tEst Value


None
to be selected:
SET FUN SET FUN SET FUN

Contin
NodE 2 AddrEs ranGE Demand
SET FUN SET FUN SET FUN
Autom-
Hyst-2 Pr-Add OffsEt Slave
SET FUN SET FUN SET FUN
Modbus
t1NEr2 dELay Print
SET FUN SET FUN Default: None
DEVICENET
dELay2 rEN-Co dN-Add
Prot-2 COM2 PROTOCOL:
SET FUN SET FUN
SET FUN
It defines the Fieldbus use mode
tEst1n data F None: Serial communication OFF
SET FUN SET FUN
ETHERNET IP
PROFINET PROFIB: PROFIBUS fieldbus (if there is an optional board)
tstoUt t-Nap
SET FUN SET FUN 1P-Add
DEVNET: DEVICENET fieldbus (if there is an optional board)
SET FUN
Value to be selected:
r-NaP
SET FUN SubnEt None
SET FUN Profib
dELNap Devnet
SET FUN
Default:
None

Page 33
0
prot-2
SET FUN
PRG
Select Value PRG

NU 0 0 0 0 0
0
AddrEs 0
FUN FUN FUN FUN SET FUN
FUN
PRG
Enter Value PRG

NTER MENU
Conf1G CaL1br
AddrEs
SET
FUN
0
SET
FUN
ParaN 1n-oUt SEr1aL
0 COM1 SERIAL COMMUNICATION ADDRESS
0
SET
IP-AddAnaLoG StorE? FUN
0
SET
FUN
0
0
SET
FUN
0

0 PRG
Conf1Gof theCaL1br ParaN protocols 1n-oUt SEr1aL AnaLoG Sto
PRG
PRG PRG PRG PRG PRGFUN PRG
Configuration address used in the transmission and
SET
ec.

F in
PRG + SET

the MODBUS protocol.


SET
IP-Ad1
SET SET SET SET

0
CapaC CaL fILtEr NodE 1 baUd r f-sCaL
PRG PRG 0 PRG
EXIT ME
0 PRG 0 PRG 0 PRG 0 PRG PRG

Value
SET
CapaCfrom 000 to 99.
FUN
0
CaL fILtEr NodE 1 baUd r
PRG
0 SET FUN
0 000f-sCaL
SET FUN
0
SET FUN
0
SET FUN
0
Enter Value
Default:01
SEns1t SET
0 Notion Hyst-1 baud2r
FUN 0
NodE123 SET FUN SET FUN SET FUN SET FUN

PRG
SET FUN
SET
SEns1t
Pr-Add
FUN

PROGRAMMING OF THE Notion


0 PROFIBUS ADDRESS Hyst-1 baud2r … NodE
SET FUN SET FUN SET FUN

nEt CaL SET FUN


auto 0 t1NEr1 Prot-1 AnzEro
Configuration of the address used in the PROFIBUS protocol.
SET FUN SET FUN SET FUN
SET FUN
SET FUN

SET
nEt
FUN
CaL
SET
auto 0 t1NEr1 Prot-1
SET
IP-Ad4
FUN
AnzEro
SET FUN SET FUN SET FUN

Values: from 0 to 126 0

dEad L 0 trac dELay1 prot-2 …


tEst
Default: 01 123456
SET FUN SET SET FUN SET FUN SET FUN SET FUN

SET FUN SET FUN SET FUN


dEad L 123456 0 trac dELay1 prot-2
SET FUN SET FUN
tEst
PRG
SET FUN 0

subnEt 0

dspd1J 60000 OF THE


dC-Add PROGRAMMING SET FUN
0 DEVICENET
band ADDRESS NodE 2 AddrEs ranGE
SET FUN 0 SET FUN SET FUN SET
SET
FUN
FUN
PRG SET FUN

SET FUN SET SnEt 1


FUN SET FUN SET FUN SET FUN
dspd1Jof the address
Programming 60000 0 bandprotocol.
used in DEVICENET NodE 2 AddrEs
PRG
ranGE 0 PRG

S1GnaL
Value: from 0 to 63
SET FUN
0 ALL Hyst-2 Pr-Add OffsEt PRG
000 SET FUN SET FUN SET FUN SET FUN

SET FUN SET Enter Value


FUN SET FUN SET FUN SET FUN
S1GnaL
Default:01 0 ALL Hyst-2 Pr-Add 123 OffsEt
CoUnts SET FUN
t1NEr2 dELay …
SET FUN SET FUN SET FUN
PRG
SET FUN

dELay DELAYED RESPONSE OF THE SLAVE AND MODBUS RTU COM1


SET FUN SET FUN SET FUN
CoUnts
PROTOCOLS t1NEr2 dELayDEVICENET SET FUN

opNodE SET FUN


dELay2 rEN-Co SnEt
Indicative delay of the response string used in the SLAVE protocol. dN-Add 4
DEVICENET
SET FUN SET FUN

0
SET FUN SET FUN SET FUN
opNodE
(expressed in 1/100 sec., max 1 sec). dELay2 rEN-Co … dN-Add SET FUN

UpLoad SET FUN


tEst1n
This value is expressed in milliseconds and represents the delay with data F
SET FUN SET FUN PRG
SET FUN
SET
UpLoad
which
FUN
tEst1n
the instrument sends the response to the request received from dELay ETHERNET
data FEnter ValueIP
SET FUN SET FUN 0

PRG

PROFINET
PRG

the master.
SET FUN

dnLoad SET FUN


tstoUt t-Nap ETHERNET IP SET FUN SET FUN

rEN-Co1P-Add
Value: from 0 to 999 msec
tstoUt t-Nap RemotePROFINET
0
SET FUN SET FUN SET FUN
dnLoad SET FUN
PRG

SET FUN
PRG

Default: 000 SET FUN


r-NaP
data F 1P-Add
connection SET FUN
0
SET FUN

r-NaPSubnEt
SET FUN
Enter Value SET FUN SET FUN
PRG PRG

SET FUN
t-NAP
dELNap SubnEt 0 SET FUN

trasN- SET FUN


SET FUN
PRG
SET FUN

dELNapEnd-oh 0

SET FUN
0
r-Nap
SET FUN
PRG
rECEIV
End-oh 0

0
dELNap
SET FUN
PRG
ErasE
End-oh

Page 34

PRG

0
dELay
SET FUN
PRG
Enter Value PRG

0
rEN-Co rEN-Co REMOTE COMMUNICATION
PRG
Remote PRG
SET FUN
It enables communication with a PC for the setting via the PC program
connection
0
data F PRG
Enter Value PRG

data F REMOTE COMMUNICATION


SET FUN

0
t-NAP Parameters of the serial COM1 protocols (parity, bits n., stop bits)
SET FUN
PRG
trasN-
End-oh 0
except MODBUS.

0
r-Nap t-Nap SENDING DATA TO THE PC
SET FUN
PRG
rECEIV
End-oh 0
This function allows to transfer the mapping of the registers from the
DAT 400 instrument to the PC program. Before starting the transmis-
0
dELNap sion of the mapping, you should enable the reception of the map
SET FUN
PRG
ErasE
End-oh (Receive button) on the PC program. During transmission, the display
of the instrument displays TRASM, at the end of the transmission it
shows END-OK.
To end the mapping transferring process, press the key 0.

r-Nap RECEIVING DATA FROM THE PC


This function allows to receive the registers mapping from the PC
program.
Before starting the mapping transmission in the PC (Send key), the
receiving function must be enabled on the DAT 400 by pressing the
PRG key. When receiving, the display shows RECEIV, at the end
of the receiving it shows END-OK. To end the process of mapping
transferring, press the 0 key.

dELNap DATA RESET


This function allows you to restore the default mapping of the instru-
ment, while restoring the default mapping of the instrument, ERASE
is displayed.

Page 35
ANALOG - ANALOG OUTPUT PARAMETERS (DAT 400/A ONLY)

F-SCaL FULL SCALE FUN FUN

SEr1aL AnaLoG ConF1G


It’s the weight corresponding to the full scale of the analog output
SET SET

PRG

that can be different from the capacity of the weighting system.


0
f-sCaL
Value to be set from 000 to 99999. SET FUN
PRG
Enter Value PRG

Default: the same value of the CAPACITY parameter 0


NodE
SET FUN
PRG
Select Value PRG

NodE ANALOG OUTPUT OPERATION MODE 0


AnzEro
Selection of the value to be associated to the analog output, corre- SET FUN
PRG
Enter Value PRG

sponding to the net weight, gross weight or the peak value. 0


tEst PRG
Select Value PRG

Value to be selected: SET FUN

NET 0
ranGE PRG
Select Value PRG

GROSS
SET FUN

PEAK 0
offsEt PRG
SET FUN

Default: GROSS
-0- PRG

0 SET FUN
Set Offset
AnZEro ZERO VALUE OF THE ANALOG OUTPUT -Fs- PRG

Analog value related to the full scale of the analog output to be


subtracted.

tEst ANALOG OUTPUT TEST PROCEDURE


With this procedure it is possible to check the operation of the analog
output, causing the output value through the use of the keyboard.
The display shows the percentage of the output value than the full
scale set.
FUN
Use the SET and keys to increase/decrease the output value.

ranGE ANALOG OUTPUT RANGE


Select the analog output range.
Value to be selected:
0÷20mA
4÷20mA
0÷10Vdc
0÷5Vdc
Default: 4÷20mA

Page 36
0
tEst
SET FUN
PRG
Select Value PRG

0
ranGE
SET FUN
PRG
Select Value PRG

offsEt ADJUSTING THE OFFSET (CALIBRATION)


0
offsEt
Measure the analog output value with a multimeter to perform the
PRG
SET FUN

-0- PRG
calibration of zero (0) and full scale (FS).
0 SET FUN
Set Offset
SET FUN
-Fs- PRG

Use the and keys to adjust the analog output. Press and
hold down the key for a rapid change.
0
Press the key to toggle between offset of zero and that of full
scale.
PRG 0
Press the key to return to the OFFSET sub-menu. Press the
key to return to the ANALOG menu.

Page 37
SERIAL COMMUNICATION PROTOCOLS

CONTINUOUS, AUTOMATIC AND MANUAL TRANSMISSION PROTOCOL


These protocols have been programmed into their programming menu.
The string is transmitted as follows:

STX <status> <net weight> <gross weight> <peak> ETX <chksum> EOT
Where
STX (start of text) = 0x02h
ETX (end of text) = 0x03h
EOT (end of transmission) = 0x04.
<status> = an ASCII character that can take the following values:
“S” = stable weight.
“M” = weight that is not stable (moving).
“O” = weight greater than the maximum capacity.
“E” = weight that cannot be detected.
<net weight> = field consisting of 6 ASCII characters of net weight.
<gross weight> = field consisting of 6 ASCII characters of gross weight.
<peak> = field consisting of 6 ASCII characters of peak.
<chksum> = 2 ASCII control characters calculated considering the characters between STX and ETX
excluded. The control value is obtained by executing the operation of XOR (or exclusive) of the 8-bit
ASCII codes of the characters considered. The result is a character that is expressed in hexadecimal
with 2 digits that can take values from “0” to “9” and “A” to “F”.
<chksum> is the ASCII encoding of the two hexadecimal digits.
In the case of continuous communication protocol, the given string is transmitted at a frequency of 10
Hz, regardless of the weight filter selected.
In the case of automatic and manual communication protocols, between 2 weight transmissions, the
weight must have a variation corresponding at least 20 divisions.

SLAVE TRANSMISSION PROTOCOL


LIST OF THE CONTROLS AVAILABLE:
• Request for the net and gross weight and current peak.
• Change in gross weight.
• Change in net weight.
• Command of reset or automatic calibration or peak reset.
• Programming the two setpoints of weight
• Requesting the programmed setpoints.
• Control of setpoints storage in permanent memory.
The unit connected to the instrument (typically a personal computer) acts as a MASTER and is the only
unit that can start a process of communication.
The process of communication must be made by the transmission of a string by the MASTER, followed
by a reply from the SLAVE concerned.

Page 38
CONTROLS FORMAT DESCRIPTION:
The double quotes enclose constant characters (observe upper and lower case); the <and> symbols
contain variable numeric fields.
REQUEST FOR THE NET AND GROSS WEIGHT AND CURRENT PEAK
Master: <Addr> “N” EOT
DAT 400: “N” <Addr> <status> <net> <gross> <peak> ETX <chksum> EOT
CHANGE IN GROSS WEIGHT
Master: <Addr> “C” “L” EOT
DAT 400: <Addr> “C” “L” ACK EOT
CHANGE IN NET WEIGHT
Master: <Addr> “C” “N” EOT
DAT 400: <Addr> “C” “N” ACK EOT
COMMAND OF RESET OR AUTOMATIC CALIBRATION OR PEAK RESET
Master: <Addr> “A” “A” EOT
DAT 400: <Addr> “A” “A” ACK EOT
PROGRAMMING TWO WEIGHT SETPOINS
Master: <Addr> “S” <s1> <s2> ETX <csum> EOT
DAT 400: <Addr> “S” ACK EOT
REQUESTING FOR THE PROGRAMMED SETPOINT
Master: <Addr> “R” EOT
DAT 400: <Addr> “R” <s1> <s2> ETX <csum> EOT
STORING THE WEIGHT Setpoint IN A PERMANENT MANNER
Master: <Addr> “M” EOT
DAT 400: <Addr> “M” ACK EOT
In the case of communication error or otherwise unrecognized command from DAT 400, it will respond
with the following string:
DAT 400: <Addr> NAK EOT
FIELDS DECRIPTION
The double quotes enclose constant characters (observe upper and lower case); the <and> symbols
contain variable numeric fields.
<addr> = Serial communication address of the instrument; it is the ASCII character obtained by adding
80h to the number of address (i.e. address 1: <Addr> = 80h + 01h = 81h).
<csum> = checksum of the string data. It is calculated by performing the exclusive OR (XOR) of all cha-
racters from <Addr> to ETX excluded the latter; the result of the XOR is decomposed into 2 characters
by considering separately the upper 4 bits (first character) and lower 4 bits (second character); the
2 characters obtained are then ASCII encoded (example: XOR = 5Dh; <csum> = “5Dh” namely 35h
and 44h).
ETX (end of text) = 0x03h,
EOT (end of transmission) = 0x04h,
ACK (acknowledgment) = 0x06h,
NAK (No acknowledgment) = 0x15h.

Page 39
<status> = an ASCII character that can take the following values:
“S” = stable weight
“M” = weight that is not stable (moving)
“O” = weight greater than the maximum capacity
“E” = weight that cannot be detected.
<s1>...<s2> = 6 ASCII characters of setpoint.
<net weight> = 6 ASCII characters of net weight.
<gross weight> = 6 ASCII characters of gross weight.
<peak> = 6 ASCII characters of peak.
If the request is made cyclically, the weight is acquired with a maximum frequency of:

Frequency Baud Rate


200Hz 115200
50Hz 38400
35Hz 19200
25Hz 9600
8Hz 2400

Page 40
MODBUS RTU PROTOCOL
The addresses listed in the tables below follow the standard address specified in the guidelines of
the Modicon PI-MBUS-300. Below please find an excerpt that helps the user to communicate with the
instrument.
“All data addresses in Modbus messages are referenced to zero. The first occurrence of a data item
is addressed as item number zero. For example:
The coil known as ‘coil 1’ in a programmable controller is addressed as coil 0000 in the data
address field of a Modbus message.
Coil 127 decimal is addressed as coil 007E hex (126 decimal).
Holding register 40001 is addressed as register 0000 in the data address field of the message. The
function code field already specifies a ‘holding register’ operation. Therefore the ‘4XXXX’ reference
is implicit.”

To confirm a new value in E2prom, run the function of MAKE – BACKUP. If this function is not performed
by switching off, the DAT will return to the value before the change.
If not specified otherwise, the numerical values (such as addresses, codes and data) are expressed as
decimal values​​.
For any hardware configuration of the instrument (FIELDBUS or Analog), the MODBUS RTU protocol
is always available on COM1 RS232; in the event of Fieldbus absence, the MODBUS RTU protocol is
also available on COM2 RS485.

INSTRUMENT RESPONSE TIMES


In order to respond to most requests, the instrument takes a maximum time of 20 msec.
Exceptions are:
• the e2prom Backup command (max time = 350mSec.)
• writing of the registers of the cells capacity, cells sensitivity, net weight, system calibration, filter
(max time = 550mSec).

COMMUNICATION ERRORS HANDLING


The communication strings are controlled by the CRC (Cyclic Redundancy Check). In the case of a
communication error, the slave does not respond with a string. The master must consider a timeout for
the receipt of the response. In case of no answer, a communication error has occourred.

RECEIVED DATA ERROR HANDLING


In the case of string received correctly, but that cannot be executed, the slave responds with an EXCEP-
TION RESPONSE according to the following table.

Code Description
1 ILLEGAL FUNCTION (The function is not valid or not supported)
2 ILLEGAL DATA ADDRESS (The address of the specified data is not available)
3 ILLEGAL DATA VALUE (The received data have invalid value)

Page 41
SUPPORTED FUNCTIONS

Function Description
01 READ COIL STATUS (Reading the state of the logic outputs)
02 READ INPUT STATUS (Reading the state of the logic inputs)
03 READ HOLDING REGISTERS (Reading the programmable registers)
04 READ INPUT REGISTERS (Reading the “read only” registers”)
05 FORCE SINGLE COIL (Writing the status of each output)
06 PRESET SINGLE REGISTER (Writing a programmable register)
15 FORCE MULTIPLE COILS (Multiple writing of outputs)
16 PRESET MULTIPLE REGISTERS (Multiple writing of registers)
Funct + 80h EXCEPTION RESPONSE

Page 42
LIST OF THE MODBUS PROTOCOL HOLDING REGISTERS
Addres Holding Register R/W Format Note
40001 Status Register R INT See table A
40002 Gross weight (MSB) R
DINT
40003 Gross weight (LSB) R
40004 Net weight (MSB) R
DINT
40005 Net weight (LSB) R
40006 Peak value (MSB) R
DINT
40007 Peak value (LSB) R
40008 Load cell signal in mV/V R INT
40009 Logic inputs R INT LSB = Input 1
40010 Output R/W INT LSB = Output 1 (it writes only if the setpoint = 0)
40011 Keys status R INT See table B, even if the key lock is enabled
40012 Firmware code and version R INT See table C
40201 Setpoint 1 (MSB) R/W
DINT
40202 Setpoint 1 (LSB) R/W
40203 Setpoint 2 (MSB) R/W
DINT
40204 Setpoint 2 (LSB) R/W
40501 Data Register (MSB) W Data related to the Command Register
DINT
40502 Data Register (LSB) W
40503 Command Register W INT See table D
41001 Cells capacity (MSB) R/W
DINT
41002 Cells capacity (LSB) R/W
41003 Cells sensitivity R/W INT
41004 Weight division value R/W INT See table E
41005 Tare of the system (MSB) R/W
DINT
41006 Tare of the system (LSB) R/W
41007 System capacity (MSB) R/W
DINT
41008 System capacity (LSB) R/W
41101 Weight filter R/W INT 0-9
41102 Weight stability R/W INT 0-4
41103 Auto-zero setpoint in % (MSB) R/W 0-100%.
DINT
41104 Auto-zero setpoint in % (LSB) R/W
41105 Zero tracking R/W INT 0-4
41106 Zero band in divisions R/W INT
41201 Operation mode Set 1 R/W INT See table F
41202 Hysteresis Set 1 (MSB) R/W
DINT
41203 Hysteresis Set 1 (LSB) R/W
41204 Timer Set 1 R/W INT
41205 Delay Set 1 R/W INT
41206 Operation mode Set 2 R/W INT See table F
41207 Hysteresis Set 2 (MSB R/W
DINT
41208 Hysteresis Set 2 (LSB) R/W
41209 Timer Set 2 R/W INT
41210 Delay Set 2 R/W INT
41401 Analog full scale (MSB) R/W
DINT
41402 Analog full scale (LSB) R/W
41403 Analog mode R/W INT See table G
41404 Analog range R/W INT See table H
Points of analog output (da 0 a 65535).Analog output
41405 Analog output value R/W INT
uses this value only if FS (41402) = 0
The programmed value is automatically copied to
42000 Monitor Register W INT
Monitor Register R (42100)
Copy of the value entered in Monitor Register W
42100 Monitor Register R INT
(42000)

Page 43
TABLE A - REGISTER STATUS CODING

BIT 13 12 11 10 9 8 6 5 4 3 2 1 0
Output Output Input Input Memory Lock Off Over- Under- Tare Zero Stable Zero
Description
2 1 2 1 Flag keyboard Range load load entered band weignt center
WARNING: Bits 15, 14 and 7 are not managed and are always equal to 0.
OPERATION Flag memory (bit 9): When modifying a register that requires saving in the E2prom (see
table “Data stored in memory with the command 0x020” on the next page), this bit is set to 1 to remind
the user to run the command 0x0020 (saving data in the permanent memory). After performing this
operation, the bit is automatically reset to zero.
OPERATION Bit related to a weight error (bit 6 of the STATUS REGISTER) When the cell is disconnected
or the measured values are out of range, this bit is set to 1.
OPERATION Bit related to the Band of zero (bit 2 of the STATUS REGISTER) When the gross weight is
less than or equal to the parameter “Band of zero in divisions”, bit 2 of the status register is set to 1.

TABLE B - KEYS CODING (40011)

bit Keys Status


0 SET key
1 FUN key
2 0 key
3 PRG key
WARNING: Bits 4 to 15 are not managed and are always equal to 0.

TABLE C - FIRMWARE CODING

bit Firmware Code


0…7 Version code
8…15 Firmware code

TABLE D- COMMAND REGISTER CODING TABLE FOR MODBUS PROTOCOL

Register value Command Register Function


0x0000 No command
0x0001 Semiautomatic zero
0x0002 Autotare
0x0003 Peak reset
0x0010 Calibration of the weight zero
0x0011 Calibration of the full weight scale
0x0020 Saving the data in the permanent memory
0x7FFF Direct access to memory

TABLE E- CODING DIVISION VALUE

Register value 0 1 2 3 4 5 6 7 8
Division value 0.001 0.002 0.005 0.00 0.02 0.05 0.1 0.2 0.5

Register value 9 10 11 12 13 14
Division value 1 2 5 10 20 50

Page 44
Addr. of Modbus register Data stored in memory with the 0x0020 command
41001-41002 Cells capacity
41003 Cells sensitivity
41004 Weight division value
41005-41006 Tare of the system
41007-41008 System capacity
41101 Weight filter
41102 Weight stability
41103-41104 Auto-zero setpoint
41105 Zero tracking
41106 Zero band in divisions
41201 Operation mode Set 1
41202-41203 Hysteresis Set 1
41204 Timer Set 1
41205 Delay Set 1
41206 Operation mode Set 2
41207-41208 Hysteresis Set 2
41209 Timer Set 2
41210 Delay Set 2
41401-41402 Analog full scale
41403 Analog mode
41404 Analog range

TABLE F - WEIGHT SETPOINT OPERATION CODING

bit Setpoint operation mode


0…1 0 = Net weight, 1 = Gross weight, 2 = Peak
2 0 = N.O. 1 = N:C.
3 0 = Positive values 1 = Negative values
4 0 = Always controlled 1 = Only with stable weight

EXAMPLE: CALIBRATION FUNCTION VIA MODBUS


To perform the calibration of full scale (that require the weight value in the sample set in the data register),
the value in the data register must be present when the command register is programmed. For example:
Calibrate the full scale using a sample weight of 2000 kg.
Write 2000 in the data register.
Write 0x0011 in the command register.
Use the multiple registers write function and write the registers of data register and command register
in a single command.

Page 45
PRINT PROTOCOL
This protocol enables the communication with a printer. The data on the printed report are the following:
NET, GROSS and TARE (+ PEAK, if enabled).
PRG
The print command takes place by pressing the key or by activating the remote input #2. (*)
The printed report is issued only if the following conditions are met:
• Gross weight positive
• Net weight positive
• Stable weight (**)
• BLIND function disabled (***)
PRG
(*) In case the key is locked (see page 18) the print command can take place through the remote
input #2 only.
(**) The weight stability condition is controlled only if the “MOTION” parameter (see page 28) is set
to a value different than 0.
If MOTION = 0 the printed report is issued even if the weight is unstable.
(***) In case the BLIND function is enabled (see page 18) the print command can take place through
PRG
the remote input #2 only, not via the key.
Between one printout and the next one the weight must change for at least 20 counts.

CONNECTION TO THE PRINTER

PRINTED REPORT FORMATS


Standard Peack enabled
Net 9.488 kg Net 9.488 kg
Gross 19.874 kg Gross 19.874 kg
Tare 10.386 kg Tare 10.386 kg
Peak 35.294 kg

Page 46
THE USB CONNECTION BETWEEN THE DAT-400 AND THE PC

The hardware module installed inside the DAT-400 and allowing the instrument to be interfaced with a
PC through the USB port, is named CP210x, manufactured by Silicon Laboratories.
The CP210x module needs the drivers to be installed in the PC:
By installing on PCs the INNOVATION 2 utility software drivers will be loaded automatically. For in-
stallation refer to the manual.

Page 47
FIELDBUS PROTOCOL
FIELDBUS data exchange takes place on two separate memory areas, which are described in the
tables below.

WARNING:
The “Input Data Area” and the “Output Data Area” are of 128 bytes
To transfer the parameters of the Output Data Area to the instrument you need to enable direct access
to the memory, writing the hexadecimal value 0x7FFF in the Command Register (1 word in writing) in
order to avoid that the instrument resets all its variables in the case of uninitialized Output Data Area
This command must be sent before the connection to inform the instrument that the parameters have
been initialized by the master. Now, the instrument continually tests the changes made​to the parameters
and stores them only in case of real change.

INPUT DATA AREA

ADDRES
Byte Bit Description Format Note
0 0 Keyboard_Look BOOL See table A Page 45
0 1 Memory_Flag BOOL See table A Page 45
0 2 Input_1 BOOL See table A Page 45
0 3 Input_2 BOOL See table A Page 45
0 4 Output_1 BOOL See table A Page 45
0 5 Output_2 BOOL See table A Page 45
0 6 Spare BOOL See table A Page 45
0 7 Spare_1 BOOL See table A Page 45
1 0 Center_of_Zero BOOL See table A Page 45
1 1 Stable_Weight BOOL See table A Page 45
1 2 Zero_Band BOOL See table A Page 45
1 3 Tare_entered BOOL See table A Page 45
1 4 Under_Load BOOL See table A Page 45
1 5 Over_Load BOOL See table A Page 45
1 6 Off_Range BOOL See table A Page 45
2 0 Gross_Weight DINT
6 0 Net_Weight DINT
10 0 Peak_Weight DINT
14 0 Load_Cells_Signal_mV_V INT
16 0 Input_Status INT
18 0 Output_Status INT
20 0 Not used BYTE
21 0 SET_Key BOOL See table B Page 45
21 1 FUN_Key BOOL See table B Page 45
21 2 0_Key BOOL See table B Page 45
21 3 PRG_Key BOOL See table B Page 45

Page 48
22 0 Cod_Firmware BYTE See table C Page 45
23 0 Cod_Versione BYTE See table C Page 45
24 0 SP_1 DINT
28 0 SP_2 DINT
32 0 Load_Cell_Capacity DINT
36 0 Load_Cell_Sensitivity INT
38 0 Weight_Division_Value INT
40 0 Tare_of_the_System DINT
44 0 Capacity_of_the_System DINT
48 0 Filter_Value INT
50 0 Weight_Stability_value INT
52 0 Autozero_Set DINT
56 0 Tracking_Factor INT
58 0 Zero_Band INT
60 0 Not used BYTE
61 0 Set_1_Net_Weight BOOL See table F Page 46
61 1 Set_1_Gross_Weight BOOL See table F Page 46
61 2 Set_1_NO_NC BOOL See table F Page 46
61 3 Set_1_Pos_Neg_Values BOOL See table F Page 46
61 4 Set_1_Control BOOL See table F Page 46
62 0 Set_1_Hysteresys DINT
66 0 Set_1_Timer INT
68 0 Set_1_Delay INT
70 0 Not used BYTE
71 0 Set_2_Net_Weight BOOL See table F Page 46
71 1 Set_2_Gross_Weight BOOL See table F Page 46
71 2 Set_2_NO_NC BOOL See table F Page 46
71 3 Set_2_Pos_Neg_Values BOOL See table F Page 46
71 4 Set_2_Control BOOL See table F Page 46
72 0 Set_2_Hysteresys DINT
76 0 Set_2_Timer INT
78 0 Set_2_Delay INT
80 0 Monitor_Register INT

READING EXAMPLE
To read the gross weight on the DAT 400 it is needed to read the addresses from 2 to 5 of the Input Area.
To read the net weightit is needed to read the addresses from 6 to 9 of the Input Area.
Whwn the display shows the gross weight value of 12351 in the corresponding bytes there will be:

Byte 2 Byte 3 Byte 4 Byte 5


Hex 00 00 30 3F

Page 49
OUTPUT DATA AREA

Dirección
Byte Bit Descripción Formato Note
128 0 Command_Register INT See table D Page 45
130 0 Logic_Output INT
132 0 SP_1 DINT
136 0 SP_2 DINT
140 0 Data_Register DINT
144 0 Load_Cell_Capacity DINT
148 0 Load_Cell_Sensitivity INT
150 0 Weight_Division_Value INT See table E Page 45
152 0 Tare_of_the_System DINT
156 0 Capacity_of_the_System DINT
160 0 Filter_Value INT
162 0 Weight_Stability_value INT
164 0 Autozero_Set DINT
168 0 Tracking_Factor INT
170 0 Zero_Band INT
172 0 Not used BYTE
173 0 Set_1_Net_Weight BOOL See table F Page 46
173 1 Set_1_Gross_Weight BOOL See table F Page 46
173 2 Set_1_NO_NC BOOL See table F Page 46
173 3 Set_1_Pos_Neg_Values BOOL See table F Page 46
173 4 Set_1_Control BOOL See table F Page 46
174 0 Set_1_Hysteresys DINT
178 0 Set_1_Timer INT
180 0 Set_1_Delay INT
182 0 Not used BYTE
183 0 Set_2_Net_Weight BOOL See table F Page 46
183 1 Set_2_Gross_Weight BOOL See table F Page 46
183 2 Set_2_NO_NC BOOL See table F Page 46
183 3 Set_2_Pos_Neg_Values BOOL See table F Page 46
183 4 Set_2_Control BOOL See table F Page 46
184 0 Set_2_Hysteresys DINT
188 0 Set_2_Timer INT
190 0 Set_2_Delay INT
192 0 Monitor_Register INT

WRITING EXAMPLES
To write the set-up parameters following the example:
In the byte 128 (Command Register) write value Hex 7FFF. This value opens the writing area of the
DAT 400.

Page 50
Example: to change the default values of the DAT 400 like the Capacity of the load cells, the Sensitivity
and Division value to 15000, 2.9965 and 2:
Capacity Byte 144 Byte 145 Byte 146 Byte 147
Hex 00 00 3A 98
Dec 15000

Sensitivity Byte 148 Byte 149


Hex 75 0D
Dec 29965

Division Byte 150 Byte 151


Hex 00 0D
Dec 13
Save the data by writing the value Hex 20 in Command Register.
N.B. The DAT 400 does not accept writing of the same values already written.
To perform Zero and FS Calibration it is not needed to abilitate the internal Writing Area of the DAT 400.
Zero Calibration:
Whit empty system put Hex 10 in Command Register (byte 128). The new Zero value is stored.
Full Scale Calibration:
Put a know weight on the system and write its value in the Data Register (from byte 140 to 143). Put
value Hex 11 in Command Register. The weight value will be displayed.

Page 51
TROUBLESHOOTING
PROBLEM POSSIBLE CAUSE SOLUTION
The weight is not detectable because
The display shows the
the cell is absent or incorrectly Check the connections of the cells.
O-L message
connected
The weight cannot be shown
The display shows the
because it exceeds the available five
hyphen in the upper
digits or is greater than the capacity
display
of the cells.
The display shows
The weight cannot be shown
the underscore on the
because negative, more than -9999.
lower display.
The number of
Select the correct division value in
decimal places is Incorrect division value selected.
the main menu.
wrong.
The serial Installation wrong. Check the connections as described
communication does The selection of the operation of the in the installation manual.
not work properly. serial interface is incorrect. Select the settings as appropriate.
To re-establish the zero, calibrate the
The function of The gross weight exceeds the action weight.
semiautomatic zero limit of semi-automatic zero.
Wait for the stabilization of the
doesn’t work. The weight doesn’t stabilize. weight or adjust the weight filter
parameter.
Check the gross weight.
The semiautomatic The gross weight is negative or
tare function does not exceeds the maximum capacity. Wait for the stabilization of the
work. weight or adjust the weight filter
The weight doesn’t stabilize.
parameter.

Page 52
GROUNDING CONNECTIONS DAT-400

“0 Vdc”

3
2

1: Silo. A: Load receptor-base Earthing.


2: Load cell Accessory. B: Accessory Earthing.
3: Junction Box. C: Junction Box Earthing.
4: DAT indicator. D: DAT indicator Earthing.

Page 53

SAFE AREA HAZARDOUS AREA
Model 400 
Digital / Analog Transmitter Intrinsic Safety Barriers Summing Junction Box
 Signal  Signal
8 Z961H 1
 Exc. 13 + Signal + Signal 1  Sig + Exc 1
5 4 2 + Sig  Exc 2
+ Exc. 14
+ Sens 15  Exc.  Exc. 3  Exc + Sig 3
 Sens 16 6 3 4  Sens  Sig 4
Z966H
 Sig. 17 + Exc. + Exc. 5 + Sens Shield 5
+ Sig. 18 5 4 6 + Exc
 Sense  Sense
7 Shield
8 Z961H 1
+ Sense + Sense + Exc 1
5 4  Exc 2
+ Sig 3
48

 Sig 4
Shield 5
FUNCTION BARRIER TYPE QTY 1 + Exc + Exc 1
2  Exc  Exc 2
+/ Signal Z961H 1 3 + Sig + Sig 3
FUNCTION COLOR
4  Sig  Sig 4
+/ Exc.  Z966H 1 5 Shield Shield 5
+ Exc. Green
+/ Sense Z961H 1
+ Sense Blue
 Exc. Black
 Sense Yellow
+ Signal Red
 Signal White

Page 54
Shield GND
OPTION 24 VOLT POWER SUPPLY
INSTALLATION

• Make sure the installation complies with local regulations and electrical codes.
• Connect AC voltage to the terminals marked “L” and “N”.
• A red LED is illuminated when the power supply is “ON”.
• The DC voltage is available on the terminals marked “+” and “-”. The second set of
Terminals are used when powering two transmitters from a single power supply (See notes).

Notes: The power supply should be used to power two transmitters only when a single load
cell is connected to each transmitter. For multiple load cell applications it is recommended
that an individual power supply be used for each transmitter. This enables the power supply
to compensate for any fluctuations in the incoming line voltage. Recent field experiences
have shown that voltage drops of greater than 10% can cause the transmitters to drift and
become unstable.

Fuse

L N LED + -
230 VAC 24 VDC

FUSE REPLACEMENT

The following procedures require work inside the power supply enclosure and should be
performed by qualified service personnel.

• Before opening the unit, disconnect the AC voltage.


• Remove the front cover from the power supply.
• Press down gently on the cover of the fuse holder, and turn counter-clockwise.
• Pull out the cover and fuse as an assembly, replace fuse with a new one.
• Re-install fuse and cover as an assembly, press down gently and turn clockwise.
• Replace the front cover on the power supply.
• Re-apply AC voltage to the unit.

In the event of a malfunction, please contact the nearest distributor for assistance.
Any attempt to modify or repair the power supply will void the manufacturers warranty.

Page 55
NOTES

Page 56
NOTES

Page 57

You might also like