Flow Computer Model 405A: 405A-M-V3 Sep 2013
Flow Computer Model 405A: 405A-M-V3 Sep 2013
Flow Computer Model 405A: 405A-M-V3 Sep 2013
MODEL 405A
2. Specification 5
3. Operation 7
4. Options 16
5. Calibration 26
405A-M-V3
6. Input Circuits 34
7. Installation 37
7.1 General 37
7.2 Wiring Designations for the Model 405A 39
7.3 Ex 410 Enclosure Dimensions 40
8. Trouble Shooting 41
Index 42
405A-M-V3
Introduction 3
1. INTRODUCTION
The Flow Computer Model 405A is a microprocessor based instrument
designed to measure 4-20mA, 0-20mA, 1-5 Volt or 0-10 Volt signals from
flowmeters and pressure transducers. The instrument can be programmed
to display directly in engineering units and includes such features as linear
or square law calculation, integration and digital filtering. For open channel
flowmetering, the power of the input relationship is fully programmable.
The instrument is fully programmable, with all calculation constants set via
the front panel switches and stored in a non-volatile memory which will
retain data indefinitely. The user can program span, zero, filtering levels,
display resolution and cutoff points.
The 4-20mA, 0-20mA, 1-5 Volt and 0-10 volt input signals are isolated from
the supply rails and outputs, and may therefore float independently. This
ensures that the input will be compatible with all transmitters and can be
used in current loops which have more than one receiver.
405A-M-V3
4 Introduction
Model 405 A. 1 0 E B
B for Backlight
C for Conformal Coating
Options
0 for no option
1 for 4-20mA output
2 for RS232/422/485
3 for Relay option
4 for 4-20mA and relays
5 for RS232/422/485 and
relays
Mounting
1 for panel
mounting
2 for field mounting
3 for explosionproof
405A-M-V3
Specification 5
2. SPECIFICATION
General
Analog Input
405A-M-V3
6 Specification
Relay Outputs
4-20mA Output
Resolution: 10 bits.
Accuracy: Better than 0.05%.
Maximum Load: 500 ohms internally powered.
950 ohms from 24VDC.
Isolation: Output is isolated.
Pulse Output
405A-M-V3
Operation 7
3. OPERATION
The Model 405A uses a low power CMOS microprocessor to perform all
control functions and calculations.
4-20 mA 4-20 mA
4-20mA Option
( 0 - 2 0 m A) Out put
Signal
RS232/422 RS232/422
Common
Option Out put
Pulse
Out put
Model 405A
Flow Alarm
DC Input Power
1 10 / 22 0 V DC Power Ground
AC Mains
DC Power to sensors
405A-M-V3
8 Operation
The display will normally show the Rate or resettable Total, as selected by
the RATE or TOTAL keys on the front facia. An LED in the key panel will
light to indicate which function is currently displayed.
The DISPLAY key can be used to display the Accumulated Total. On the first
press of the DISPLAY key, the display shows ACCTOT for one second followed
by the actual total. The Accumulated Total continuously totalises the flow
and is not resettable from the front panel.
On reaching the maximum displayed total, all totals will roll over to zero
and continue totalising. If, at any time, power is lost or the instrument is
switched off, the totals will be stored in the non-volatile memory. When
power is switched back on to the instrument, the stored totals will be
recalled from memory and the totals will be incremented from the last
values.
405A-M-V3
Operation 9
At the minimum input (ie 4mA, 0mA, 1 Volt or 0 Volts), A = 0, and at the
maximum input (ie 20mA, 5 Volts or 10 Volts), A = 1.
The Span, S, can be set during calibration anywhere in the range of 0.1000
to 50000.0000 and the Zero value C, set in the range 0.0000 to
50000.0000.
The Span, S, can be selected to display rate in any units desired, such
a litres/minute or kilograms/hour. This also means that the Total will
be displayed with the same unit of volume, ie litres or kilograms.
Note that when the input signal drops below 3.5mA (4-20mA input) or
0.875V (1-5V input), a signal error will occur. The instrument will beep and
the display will alternate between the current total and the word "SIGNAL".
405A-M-V3
10 Operation
Because many transducers do not always exactly transmit 4mA (0mA, 1 Volt
or 0V) when they are at zero rate, it is often necessary to define a rate
below which no integration takes place. This is termed the cutoff point and
is programmed as a percentage of the Span, S.
Rc = 2200 A + 100
At 20% cutoff,
Note that integration will not occur if A=0 (ie 4mA, 0mA, 1 Volt or 0 Volts),
even with an offset programmed.
405A-M-V3
Operation 11
3.2.3 Filtering
The Flow Computer has a digital filter which will average out these
fluctuations and enable the Rate to be read to four digit accuracy. The
degree of filtering is fully programmable which means that highly accurate
and stable readings can be obtained without excessive lag.
When the Rate is retransmitted via the 4-20mA output, the filtering will
also average out any fluctuations on the output.
The diagram below shows a pulsating signal input together with the effect
of filtering.
Rate
Filtered Response
Unfiltered Response
Time
405A-M-V3
12 Operation
A 90% 99%
1 0 0
2 1 2
4 2 4
6 3 6
10 5 11
15 8 17
20 11 22
25 14 28
35 20 40
45 25 51
60 34 69
75 43 86
90 52 103
99 57 113
405A-M-V3
Operation 13
The Scaling Factor is always programmed in the unit relating to Rate and
The Total Conversion constant is a division factor which can be used to
convert the totals to the different unit. The Total Conversion factor affects
the net, accumulated and gross totals, and is limited between 0.01 and
2000.
For Example.
Some common units are given below together with the Total Conversion
constant (TOTCON) which should be programmed.
405A-M-V3
14 Operation
Note that due to the uneven pulse output spacing on this output, the pulse
output cannot be used to drive rate indicators.
405A-M-V3
Operation 15
Relay or
Impulse Counter
5.6 ohms 10
33V
Zener
12
DC Supply
DC Supply 11
Out (8-24V)
External Load
Resistor 10K
Logic Input
5.6 ohms 10
33V
Zener
12
405A-M-V3
16 Options
DC to DC
Convertor T erminal
+15V 24
23
Digital to
22
Analog
Convertor
Opto-Isolation Amplifiers 0V 20
-15V 21
405A-M-V3
Options 17
4.1.2 Calculation
For rates or displayed values above and below the maximum and
minimum values the output will remain at its 20mA or 4mA level
respectively.
It should be noted that the output will be updated every 0.25 seconds in
unison with the display and, between updates, the output value is constant.
405A-M-V3
18 Options
+15V 24
Link
23
I(+)
I(-) Load
22
500 ohm
maximum
Amplifiers 0V 20
-15V 21 *
405A-M-V3
Options 19
T erminal
+15V 24
23
I(+)
I(-)
22
Load
Amplifi ers 20
0V
-15V 21
External Supply
of 23 to 30 VDC
405A-M-V3
20 Options
With this option installed, the circuits for both the RS232 and RS422/485
are provided as standard. They can be used to interface to both printers
and computers, and a number of standard protocols are built into the
instrument.
4.2.1 Hardware
24 (+)
RS422 Out
23 (-)
26 (+)
RS422 In
25 (-)
21 Data In
RS232
22 Data Out
27 CTS
20 Ground
405A-M-V3
Options 21
To convert the RS422 interface to an RS485 interface, the RS422 (-) Data In
Terminal must be connected to the RS422 (-) Data Out Terminal and the
RS422 (+) Data In Terminal must be connected to the RS422 (+) Data Out
Terminal.
These connections will convert the RS422 4 wire interface to the RS485 2
wire interface, as shown in figure 2.
Load
Host
120 ohms
Computer
405A-M-V3
22 Options
T wisted Pair
+ Load
120 ohms
Host
Computer -
Gnd
- + Gnd - + - + G n d -- +
In Out In Out
400 Series 400 Series
Instrument Instrument
405A-M-V3
Options 23
The RS232/422/485 option has a real time clock and enables the time and
date to be set and printed on tickets. The date format can be European
(days/months/years) or USA (months/days/years), while the time is on a 24
hour clock.
A battery backup is provided. The battery will typically need replacing every
two years or more frequently if extended power downs are a feature of the
installation. Battery type is a CR2032 coin cell.
All new instruments are supplied with a ‘pullout battery life protection tab’
Please do not remove the tab until you are ready to install and apply
power to the instrument.
The baudrate, parity and word length can be programmed during calibration
and the user must ensure that these correspond to the setting on the printer
or computer with which the 405 is communicating.
Printer
A ticket is printed each time the RESET key is pressed. The instrument
prints the ticket before resetting the resettable total. Protocols are
provided to drive the following printers:
The tickets can also be printed with a number of different units, including
litres and gallons. The units are selectable from a pre-programmed list.
405A-M-V3
24 Options
A CTS input is provided, and will prevent the instrument from transmitting
any further characters to a printer if the printer buffer is full. The CTS input
is usually connected to the "Data Buffer Full" output from the printer.
If the printer buffer is large enough to handle the message output from
the instrument, then this input need not be used and should be left
unconnected.
Computer
For more information on the computer interface please consult the manual
"The RS232/422/485 Communications Option for the 400 Series, Version 2".
405A-M-V3
Options 25
The Relay output option consists of two Form C relays which can be
preset during calibration to energise when the rate or displayed value
exceeds or drops below the preset values.
The "low" relay is energised whenever the rate is below the preset value, and
the "high" relay is energised whenever the rate exceeds the preset value.
The preset values are programmed during calibration as described in section
5.
34 Normally Open
Relay 1
35 N o r m a l l y Cl o s e d
Low Alarm
36 Common
31 Normally Open
Relay 2 32 N o r m a l l y Cl o s e d
High Alarm
33 Common
405A-M-V3
26 Calibration
5. CALIBRATION
The Calibration routine enables the Setup Parameters to be programmed, as
well as enabling the input signals to be checked.
2 By pressing the TOTAL key and, while still holding, pressing the
RESET key. Both keys must then be held for approximately 6
seconds. This second method of access can be disabled during
the calibration so that it is only possible to enter the calibration
routine via the link across terminals 1 and 2.
Note that the arrows in the Rate and Total key switches indicate that
these switches can be used to change and increment digits respectively.
405A-M-V3
Calibration 27
On first entering the Calibration routine, the display will show the Model
number followed by:
The user can toggle between these modes using the TOTAL switch and by
using the DISPLAY switch, select the appropriate mode.
To exit Calibration, step through the Setup program or Test program until
the end, and press the DISPLAY switch when End is displayed, (ensure the
calibration link is removed).
405A-M-V3
28 Calibration
405A-M-V3
Calibration 29
405A-M-V3
30 Calibration
3 OP 4 Flowrate at 4mA
xxxx Enter flowrate.
405A-M-V3
Calibration 31
10 PARITY Parity.
NP No Parity.
OP Odd Parity.
EP Even Parity.
11 SIGNAL Signal Type.
rs232 RS232.
rs422 RS422/485.
12 ID NO Unit Identification Number.
0 None.
1-99 Id Number.
13 PTYPE xx Printer/Computer Type.
Computer.
405A-M-V3
32 Calibration
00 None.
01 Litres (Ltrs).
02 Gallons (Gals).
03 Barrels (bbls).
04 Pounds (lbs).
05 Grams (gms).
06 Kilograms (kgs).
07 Tons (tons).
405A-M-V3
Calibration 33
405A-M-V3
34 Input Circuits
6. INPUT CIRCUITS
6.1 THE SIGNAL INPUT
The basic circuit of the input is shown below. Both the current and
voltage signals are fed to a data selector but only one signal is processed,
depending upon whether a current (4-20mA or 0-20mA) or a voltage (1-5
V or 0-10 V) input configuration is selected. The signal is fed to a voltage
to frequency convertor and transmitted to the microprocessor via an
opto-coupler.
V oltage to
Current
9 Frequency
V ol tage
14
Microprocessor
10K
250
Ohm Reference
8
DC to DC
Regulator
Convertor
Isolation
405A-M-V3
Input Circuits 35
11
+ 8-24 VDC
Transmitter
4-20mA 9
-
250 ohms
8
Ground
2
Flow Computer
+
Transmitter
4-20mA 9
-
250 ohms
8
Chart Recorder
-
INPUT CONNECTIONS
405A-M-V3
36 Input Circuits
RATE
3
TOTAL
4
RESET
DISPLAY
6
405A-M-V3
Installation 37
7. INSTALLATION
7.1 GENERAL
The terminal designations for the Model 405A Flow Computer are given on
the following pages. The cutout hole in the panel should be 5.5" (139mm)
wide x 2.6" (67mm) high. Two side clips are supplied to secure the
instruments into the panel.
A case earthing point is provided via an earth lug on the side of the case.
Note that this earthing point is for the case only and there is complete
electrical isolation between this point and all electronic circuits. For EMC
purposes, or when the instrument is connected to mains, this point must be
connected to a good earth using a multi-stranded, braided wire or strap.
All relay outputs are totally isolated from the case and from the internal
circuitry.
A Supply Output voltage is provided to power sensors. This output will provide
a regulated voltage of 8 to 24 volts and the voltage is adjustable by means of
the potentiometer on the rear panel. Maximum current is 50mA and the
instrument comes with the voltage factory set at 24 Volts. When the
instrument is powered from a DC power source, the maximum output voltage
on the Supply Output is the DC Input Voltage less 3.5 volts.
The instrument will operate from either 12 - 28 volts DC or from the mains.
The mains voltage is factory set to either 95 - 135 VAC (110 VAC nominal) or
190 - 260 VAC (220 VAC nominal). An internal mains transformer provides full
isolation between the mains and the electronic circuits.
It is good practice to use shielded cables for all signal connections to the
Model 405. Care must be taken to separate signal cables from power cables
so as to minimise interference.
405A-M-V3
38 Installation
Overall shields should be connected to the case earth at the instrument end
only.
This connection should be as short as possible and connected to the earthing
lug on the side of the case.
405A-M-V3
Installation 39
1 Calibration Link
2 Signal Ground
3 Rate Switch
4 Total Switch
5 Reset Switch
6 Program Switch
7 Not To Be Used
8 Flow Common (-)
9 Flow 4-20mA In (+) or 0-20mA In (+)
10 Pulse Out
11 DC Power Out (8-24 VDC)
12 DC Ground
13 DC Power Input
14 Flow 1-5 V In (+) or 0-10 V In (+)
2B Signal Ground
405A-M-V3
40 Installation
Door
115.0
Hinge 288
302 221
Bottom View
Earth Point
70.0 70.0
71.5 71.5
Enclosure with 3 x M20 Gland holes Enclosure with 3 x ¾“ NPT Gland holes
405A-M-V3
Trouble Shooting 41
8. TROUBLE SHOOTING
8.1 ERROR CODES
The instrument has extensive self test facilities and will display an error
code if it detects an invalid condition. If the instrument displays an error
code other than those listed below, please contact the factory.
Error codes are displayed as "Err 12" and a list of commonly encountered
codes are given below:
Error Codes
Input Errors
Output Errors
Calibration Errors
405A-M-V3
42 Index
Power Requirements, 5
Index F
Filtering, 11
Printer, 23
Pulsating Signal, 11
Pulse Output, 14
4-20mA Output, 16 Front Panel, 8
R
A G
Regulated Voltage, 37
Access, 29 Ground, 37
Relay Output, 25
Accumulated Total, 8 Remote Push-buttons,
I 36
B Identification Response, 12
Number, 31 RS232/422/485
Battery, 23 Inductive Loads, 38 Interface, 20
Battery Life Input Circuits, 34
Protection Tab, 23 Installation, 37 S
Baudrate, 23, 31 Interference, 38 Self Test, 41
Isolation, 37 Setup Parameters, 26
Snubbers, 38
M Specification, 5
C Supply Output, 37
Mains Voltage, 37
Calibration, 26 Model Number, 4
Chart Recorders, 17 Multipoint T
Clock, 23 Communication, 21 Terminal, 39
Communication Terminal
Protocol, 23 N Designations, 37
Communications, 20 Test, 33
Non-volatile Memory,
Computer, 24 8 Tickets, 23
Cutout, 5 Time, 23
O Timebase, 29
D Total Conversion, 13
Operating Transducer Supply, 5
Date, 23 Temperature, 5
Decimal Points, 29 Trouble Shooting, 41
Operation, 7
Dimensions, 5 Options, 16
Display Key, 8 W
Output Pulse, 14
Wiring Designations, 39
E P
Parity, 23, 31
405A-M-V3